1

www.hannains t . com

Instruction Manual

HI 9811-5Portable

pH/EC/TDS/°C Meters

2

Dear Customer,Thank you for choosing a Hanna product.Please read this instruction manual carefully before using the meter.This manual will provide you with the necessary information forcorrect use of the instrument, as well as a precise idea of its versatility.If you need additional technical information, do not hesitate to e-mailus at [email protected].

WARRANTY .................................................................................. 2PRELIMINARY EXAMINATION ........................................................ 3GENERAL DESCRIPTION ................................................................ 3FUNCTIONAL DESCRIPTION .......................................................... 4SPECIFICATIONS ........................................................................... 5OPERATIONAL GUIDE ................................................................... 6pH CALIBRATION ......................................................................... 8pH VALUES AT VARIOUS TEMPERATURES ................................... 10EC/TDS CALIBRATION ................................................................. 11EC/TDS CONVERSION FACTOR ..................................................... 11BATTERY REPLACEMENT .............................................................. 12PROBE MAINTENANCE ................................................................ 13ACCESSORIES ............................................................................. 14

HI 9811-5 is guaranteed for two years against defects in workmanshipand materials when used for their intended purpose and maintainedaccording to instructions. Electrodes and probes are guaranteed for sixmonths. This warranty is limited to repair or replacement free ofcharge.Damage due to accidents, misuse, tampering or lack of prescribedmaintenance is not covered.If service is required, contact the dealer from whom you purchasedthe instrument. If under warranty, report the model number, dateof purchase, serial number and the nature of the problem. If therepair is not covered by the warranty, you will be notified of thecharges incurred. If the instrument is to be returned to HannaInstruments, first obtain a Returned Goods Authorization numberfrom the Technical Service department and then send it with shippingcosts prepaid. When shipping any instrument, make sure it is properlypacked for complete protection.

WARRANTYWARRANTYWARRANTYWARRANTYWARRANTY

TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTS

3

All rights are reserved. Reproduction in whole or in part is prohibited without the writtenconsent of the copyright owner.

PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine itcarefully to make sure that no damage has occurred during shipment.If noticeable damage is found, notify your Dealer.

Each meter is supplied with:• HI 1285-5, pH/EC/TDS/°C combination probe with 8-pin DIN

connector and 1 m (3.3') cable• HI 70007, pH 7.01 sachet, 1 pc• HI 70031, 1413 µS/cm sachet, 1 pc• HI 70032, 1382 ppm (mg/L) sachet, 1 pc• HI 700661 cleaning solution sachet, 2 pcs• Instruction manual• 1 x 9V alkaline battery.

Note: Save all packing material until you are sure that theinstrument functions correctly. Any defective item must bereturned in the original packing together with the suppliedaccessories.

GENERAL DESCRIPTIONGENERAL DESCRIPTIONGENERAL DESCRIPTIONGENERAL DESCRIPTIONGENERAL DESCRIPTION

HI 9811-5 is the complete, versatile, with a large LCD and splash-proof portable combination meter designed with the utmost precisionand simplicity. The instrument provides measurements for pH, EC andTDS ranges, which are easily selectable through the keypad on thefront panel.Conductivity measurements are automatically compensated fortemperature changes with a built-in temperature sensor. Thetemperature coefficient is fixed at 2%/°C.HI 9811-5 is a pH/EC/TDS meter designed for simplicity of use intaking pH, µS/cm, mg/L and temperature measurements. Suited forhydroponics, greenhouses, farming and ground water applications.

4

1) 8-pin DIN connector for probe2) Liquid Crystal Display3) pH range selection key4) µS/cm (EC) range selection key5) mg/L (TDS) selection key6) pH offset calibration knob7) EC/TDS calibration knob8) °C (Temperature) selection key9) On/Off key

FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION

5

egnaR

Hp0.41ot0.00006ot0 µ mc/S

L/gm0003ot0C°0.07ot0.0

noituloseR

Hp1.001 µ mc/S

L/gm01C°1.0

ycaruccA)F°86/C°02@(

Hp1.0±.s.f%2± µ mc/S

L/gm.s.f%2±C°5.0±

noitaiveDCMElacipyT

Hp1.0±.s.f%2± µ mc/S

L/gm.s.f%2±C°5.0±

rotcaFnoisrevnoC 5.0

noitarbilaCHptesffohguorhttniop-1,launaM

remmirt

noitarbilaCSDT/CEepolshguorhttniop-1,launaM

remmirt

erutarepmeTSDT/CEnoitasnepmoC

C°05ot0morfcitamotuAhtiw)F°221ot23( β C°/%2=

)dedulcni(eborP 5-5821IH

epyTyrettaB enilaklaV9x1

efiLyrettaB esuesuounitnocfosruoh051.xorppA

tnemnorivnE)F°221ot23(C°05ot0

HR%001

snoisnemiD )"4.1x1.3x7.5(mm63x08x541

thgieW ).zo1.8(g032

SPECIF ICATIONSSPECIF ICATIONSSPECIF ICATIONSSPECIF ICATIONSSPECIF ICATIONS

6

INITIAL PREPARATIONEach meter is supplied complete with a 9V battery. Remove thebattery compartment cover on the back of the meter and install thebattery while observing its polarity.Connect the probe to the DIN socket on the top of the meter byaligning the pins with the socket and pushing in the plug.Always remove the electrode protective cap before taking anymeasurements and make sure the meter has been calibrated beforetaking any measurements.Turn the meter on by pressing the On/Off key.

TAKING pH MEASUREMENTSIf the probe has been left dry, soak the tip in HI 70300 storagesolution for 30 minutes to reactivate it.

• To take a pH measurement simplysubmerge the tip (4cm/1½") of theprobe into the sample to be tested.

• Select the pH mode.

• Stir briefly and wait a couple ofminutes for the reading to adjustand stabilize. The display showsthe pH value.

• If measurements are taken in different samples successively, it isrecommended to rinse (clean) the probe thoroughly to eliminatecross-contamination. After cleaning, it is recommended to rinsethe probe with some of the next sample to be measured.

TAKING EC/TDS MEASUREMENTS

• Immerse the tip of the probe (4cm/1½")into the sample to be tested. If possible,use plastic beakers or containers to minimizeany EMC interference.

• Tap the probe lightly on the bottom ofthe beaker to remove any air bubbleswhich may be trapped inside the tip.

OPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDE

7

• Select the appropriate measurement range (EC or TDS).

• Wait a couple of minutes for the temperature sensor to reachthermal equilibrium. The display will then show the measurementautomatically temperature compensated for temperature with theappropriate indication among the following:“µS” symbol indicates the meteris in EC mode;

“mg/L” symbol indicates the meteris in TDS mode.

TAKING TEMPERATURE MEASUREMENTS

• Submerge the tip (4cm/1½") of theprobe into the sample to be tested.

• Select the °C mode.

• Stir briefly and wait a couple ofminutes for the reading to adjustand stabilize. The display shows thetemperature value.

Notes: • If the display shows only a “E”on the right side of the LCD,the reading is out of range.

• It is recommended to cleanoften the probe with HI 700661Cleaning Solution.

• After measurements have been completed, the instrumentshould be switched off, and the probe cleaned and coveredwith the protective cap.

8

For greatest accuracy, frequent calibration of the instrument isrecommended. The instrument should be recalibrated for pH:

a) Whenever the electrode is replaced.b) At least once a month.c) After testing aggressive chemicals.d) Where extreme accuracy is required.

PREPARATIONPour small quantities of pH 7.01 (HI 7007) or pH 4.01 (HI 7004)or pH 10.01 (HI 7010) solution into a clean beaker.To obtain accurate readings, use pH 7.01 (HI 7007) if you are goingto measure neutral or close to neutral samples, pH 4.01 (HI 7004)if you are going to measure acidic samples or pH 10.01 (HI 7010)for alkaline measurements.If you need to calibrate to NIST standards, use pH 6.86 (HI 7006)instead of pH 7.01 and pH 9.18 (HI 7009) instead of pH 10.01.

PROCEDURE

• Connect the probe and switch the meter on, then press the pHkey to display pH measurement.

• Remove the protective cap from the probe, rinse and immerse itin the buffer and stir gently. Wait a couple of minutes for thereading to stabilize.

• Take the temperature of the buffer solution,e.g. “10.0°C”, as follow:Select the °C mode and read the displayedvalue.

• Adjust the pH calibration knob until theLCD shows the pH value at the abovetemperature (see the pH versus temperaturechart).

• The pH calibration is now complete.

pppppH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATION

9

Notes: • The probe should be submerged approximately 4 cm(1½") into the solution. The thermometer has to be close tothe probe.

• If turning the knob the needed value can not be reached,clean the probe (see the “Probe Maintenance” section). Ifalso after the probe cleaning the value can not bereached, replace the probe.

10

For temperature compensation during calibration, please refer to thefollowing chart.

For instance, if the buffer temperature is 25°C, the display shouldshow pH 4.0 or 7.0 or 10.0.If the buffer temperature is 10°C, the display should show pH 4.0 or7.0 or 10.1.

PMET SEULAVHp

Cº Fº 10.4 68.6 10.7 81.9 10.01

0 23 10.4 89.6 31.7 64.9 23.01

5 14 00.4 59.6 01.7 93.9 42.01

01 05 00.4 29.6 70.7 33.9 81.01

51 95 00.4 09.6 50.7 72.9 21.01

02 86 00.4 88.6 30.7 22.9 60.01

52 77 10.4 68.6 10.7 81.9 10.01

03 68 20.4 58.6 00.7 41.9 69.9

53 59 30.4 48.6 99.6 11.9 29.9

04 401 40.4 48.6 89.6 70.9 88.9

54 311 50.4 38.6 89.6 40.9 58.9

05 221 60.4 38.6 89.6 10.9 28.9

55 131 80.4 48.6 89.6 99.8 97.9

06 041 90.4 48.6 89.6 79.8 77.9

56 941 11.4 48.6 99.6 59.8 67.9

07 851 21.4 58.6 99.6 39.8 57.9

pppppH VALUES AT VARIOUS TEMPERATURESH VALUES AT VARIOUS TEMPERATURESH VALUES AT VARIOUS TEMPERATURESH VALUES AT VARIOUS TEMPERATURESH VALUES AT VARIOUS TEMPERATURES

11

The TDS value in aqueous solutions is directly proportional to theconductivity. The ratio between the two parameters depends on thesolution.

HI 9811-5 model has a fixed conversion factor set to 0.5. Thismeans that 1 µS/cm is equal to 0.5 mg/L of TDS.

Accessories needed:• Use HI 70031 (1413 µS/cm) EC calibration solution or

HI 70032 (1382 ppm, or mg/L) TDS calibration solution.Note: The conversion between EC and TDS is made by a built-in

circuit, hence it is requested to calibrate the meter only inEC or TDS range. The other range is thus automaticallycalibrated.

PROCEDURE• Pour approximately 4 cm (1½") of a conductivity calibration

solution (e.g. HI 70031) into a beaker. If possible, use plasticbeaker to minimize any EMC interference.

• Immerse the probe in the solution.• Wait for a couple of minutes for thermal equilibrium to be

reached.• Tap the probe on the bottom, then shake it lightly while rotating

to make sure no air bubbles remain trapped inside the probe.• Press the µS/cm (or mg/L) key.• Turn the EC/TDS calibration knob until

the display shows the EC or TDS readingat 25°C.

EC/TDSEC/TDSEC/TDSEC/TDSEC/TDS CALIBRATION CALIBRATION CALIBRATION CALIBRATION CALIBRATION

EC/TDS CONVERSION FACTOREC/TDS CONVERSION FACTOREC/TDS CONVERSION FACTOREC/TDS CONVERSION FACTOREC/TDS CONVERSION FACTOR

12

The meters is powered by a 9V battery that is located on the rear ofthe instrument.When battery symbol is empty a low battery condition is indicated.

When the low battery indication appears, only a few hours of workingtime remains.It is recommended to replace the battery immediately.When the battery level is so low that it may cause unreliablemeasurements, the meter turns off.Battery replacement must only take place in a nonhazardous areausing a 9V alkaline battery.Unscrew the three screws on the rear of the meter, remove the batterycompartment cover and replace the 9V battery with a new one whileobserving its polarity.

Make sure the battery contacts are tight and secure before replacingthe cover.

BATTERY REPLACEMENTBATTERY REPLACEMENTBATTERY REPLACEMENTBATTERY REPLACEMENTBATTERY REPLACEMENT

13

PERIODIC MAINTENANCEInspect the probe and the cable. The cable used for the connection tothe meter must be intact and there must be no points of brokeninsulation on the cable or cracks on the probe stem or bulb.Connector must be perfectly clean and dry. If any scratches or cracksare present, replace the electrode. Rinse off any salt deposits withwater.

CLEANING PROCEDUREFor better accuracy in measurements and to ensure a goodperformance of the probe, a frequent cleaning is recommended.For this purpose, soak it in Hanna HI 700661 Cleaning Solution for5 minutes.Notes: • For particular dirty (as for example protein, oil or grease)

see the “Accessories” section for Hanna specific solutions.• After cleaning the probe, it is recommended to recalibrate

the meter. If it is not possible to calibrate, the probe hasto be replaced with a new one.

• For field applications, it is always recommended to keep aspare probe handy. When anomalies are not resolved withsimple maintenance, change the probe and recalibrate themeter.

PROBE MAINTENANCEPROBE MAINTENANCEPROBE MAINTENANCEPROBE MAINTENANCEPROBE MAINTENANCE

14

PROBESHI 1285-0 Combination, amplified pH/EC/TDS probe with built-in

temperature sensor, 8-pin DIN connector and 1 m(3.3') cable

HI 1285-5 Combination, amplified pH/EC/TDS/temperature probewith built-in temperature sensor, 8-pin DIN connectorand 1 m (3.3') cable

pH BUFFER SOLUTIONSHI 7004L pH 4.01 buffer solution, 500 mL bottleHI 7006L pH 6.86 buffer solution, 500 mL bottleHI 7007L pH 7.01 buffer solution, 500 mL bottleHI 7009L pH 9.18 buffer solution, 500 mL bottleHI 7010L pH 10.01 buffer solution, 500 mL bottle

CONDUCTIVITY & TDS CALIBRATION SOLUTIONSHI 7031L 1413 µS/cm solution, 500 mL bottleHI 7032L 1382 ppm (mg/L) solution, 500 mL bottle

OTHER SOLUTIONSHI 700661P Cleaning Solution, 20 mL sachet (25 pcs)HI 70300L Storage Solution, 500 mL bottleHI 7073L Protein Cleaning Solution, 500 mL bottleHI 7074L Inorganic Cleaning Solution, 500 mL bottleHI 7077L Oil & Fat Cleaning Solution, 500 mL bottle

OTHER ACCESSORIESHI 710007 Shockproof rubber boot, blueHI 710008 Shockproof rubber boot, orangeHI 710050 Blue protective case

ACCESSORIESACCESSORIESACCESSORIESACCESSORIESACCESSORIES

15

RECOMMENDATIONS FOR USERS

Before using these products, make sure they are entirely suitable forthe environment in which they are used.Operation of these instruments in residential areas could causeunacceptable interferences to radio and TV equipment requiring theoperator to follow all necessary steps to correct interferences.The glass bulb at the end of the electrode is sensitive to electrostaticdischarges. Avoid touching this glass bulb at all times.During operation, ESD wrist straps should be worn to avoid possibledamage to the electrode by electrostatic discharges.Any variation introduced by the user to the supplied equipment maydegrade the instrument's EMC performance.To avoid electrical shock, do not use these instruments when voltagesat the measurement surface exceed 24 VAC or 60 VDC.To avoid damages or burns, do not perform any measurement inmicrowave ovens.

Hanna Instruments reserves the right to modify the design, construction andappearance of its products without advance notice.