1 Instruction Manual HI 8614 • HI 8614L HI 8615 • HI 8615L pH and ORP Transmitters HI 8614 L pH INDICATOR & TRANSMITTER pH www.hannainst.com

HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

Embed Size (px)

Citation preview

Page 1: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

1

Instruction Manual

HI 8614 • HI 8614LHI8615 • HI8615L

pH and ORP Transmitters

HI 8614 LpH INDICATOR

& TRANSMITTER

pH

w w w . h a n n a i n s t . c o m

Page 2: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

32

PRELIMINARY EXAMINATION

Remove the instrument from the packing ma-terial and examine it carefully to make surethat no damage has occurred during ship-ping. If there is any noticeable damage, notifyyour Dealer.

Note: Save all packing material until you aresure that the instrument functions cor-rectly. All defective items must be re-turned in the original packing materialtogether with the supplied accessories.

GENERAL DESCRIPTION

HI 8614 / HI 8614L (pH) and HI 8615 / HI 8615L(ORP) are 2-wire water-resistant transmittersspecially designed for long distance mea-surement of pH or ORP in industrial applica-tions.

Two versions are available: the standardHI8614 and HI8615, and the HI8614L andHI8615L models with LCD.

The LCD allows easy verification and moni-toring of measured values, and provides easiercalibration and maintainance.

The pH or ORP signal is transmitted in a 2-wire current loop in the range of 4 to 20 mA.

HI8614 and HI8614L can be connected tothe Hanna process instruments HI8510T,HI 8710T or HI 8711T, or to any recorder,computer or data monitoring device that ac-cepts 4 to 20 mA input.

HI8615 and HI8615L can be connected tothe Hanna meters HI8512T, HI 8720T or toany recorder, computer or data monitor thataccepts 4 to 20 mA input.

Dear Customer,

Thank you for choosing a Hanna InstrumentsProduct.

Please read this instruction manual carefullybefore using the instrument.

This manual will provide you with all the neces-sary information for the correct use of theinstrument.If you need additional technical information,do not hesitate to e-mail us [email protected].

These instruments are in compliance with the directives.

TABLE OF CONTENTS

Preliminary Examination ............................. 3

General Description .................................... 3

Specifications of HI 8614 & HI 8614L .......... 5

Specifications of HI8615 & HI 8615L........... 6

Terminal Board Connections ....................... 7

pH Calibration with Automatic TemperatureCompensation (HI8614 & HI8614L) .......... 9

pH Calibration with Manual TemperatureCompensation (HI 8614 & HI 8614L) ....... 13

ORP Calibration (HI 8615 & HI 8615L) .... 14

Electrode Conditioning and Maintenance .. 19

Temperature-Resistance Correlation forHANNA pH Sensitive Glass .................... 23

Installation Procedure and Examples ...... 25

Accessories .............................................. 28

Warranty .................................................. 34

CE Declaration of Conformity ................. 35

Page 3: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

54

SPECIFICATIONS OF HI8614 & HI8614L

HI 8614pH TRANSMITTER HI 8614 L

pH INDICATOR

& TRANSMITTER

pH

The transmitters use a universal BNC socketfor quick and secure connection to any elec-trode with a BNC connector.

For HI8614 and HI8614L the temperaturecompensation is performed by the transmitter'sATC circuitry if the temperature probe(HI 76608, optional) is attached. If ATC is notrequired, it is possible to substitute the tem-perature probe with a fixed resistor.

The input is isolated from the current loop toeliminate problems related to ground loop,low insulation cables, multiple electrode con-nections, and a common mode voltage of upto 100 V ensures true differential readings.

The terminal board of the transmitter providesfor connection of power supply, pH or ORPelectrodes and temperature probe (for HI8614and HI 8614L models only).

The unit is enclosed in a protective casingconforming to IP 65 standards.

Range 4 to 20 mA / 0.00 to 14.00 pHResolution 0.01 mA / 0.01 pHAccuracy (@20°C/68°F)

±0.02 mA / ±0.02 pHTypical EMC Deviation

±0.25 mA / ±0.2 pHCalibration Offset: ±2.2 mA / ±2 pH

Slope: ±0.5 mA / 86 to 116%Temperature Compensation

Fixed or automatic from 0 to 100°C(32 to 212°F) with HI76608 probe

Input Impedance 1012 OhmOutput 4-20 mA (isolated)Installation Category I IProtection IP 65Environment 0 to 50°C (32 to 122°F);

RH max 95% non-condensingPower Supply

HI8614: 18 to 30 VdcHI8614L: 20 to 36 Vdc

Load Max 500 OhmDimensions 165 x 110 x 90 mm

(6.5 x 4.3 x 3.5'')Weight 1 kg (2.2 lb.)

Page 4: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

76

TERMINAL BOARD CONNECTIONS

Unscrew the 4 screws and remove the top cover.

There are three cable glands on the transmittercover. The large cable gland with the split in therubber is for the electrode.

ELECTRODECABLE GLAND

WIRE CABLE GLANDS

Connect the positive supply to the strip terminal"+VE LOOP" and the negative supply to theterminal "-VE LOOP" of the transmitter terminalblock.

SPECIFICATIONS OF HI8615 & HI8615L

HI 8615

ORP TRANSMITTER

mV

HI 8615

ORP TRANSMITTER

Range 4 to 20 mA / ±1000 mVResolution 0.01 mA / 1 mVAccuracy (@20°C/68°F)

±0.02 mA / ±5 mVTypical EMC Deviation

±0.25 mA / ±15 mVCalibration Offset: ±0.8 mA / ±100 mV

Slope: ±0.8 mA / 90 to 110%Input Impedance 1012 OhmOutput 4-20 mA (isolated)Installation Category I IProtection IP 65Environment 0 to 50°C (32 to 122°F);

RH max 95% non-condensingPower Supply

HI8615: 18 to 30 VdcHI8615L: 20 to 36 Vdc

Load Max 500 OhmDimensions 165 x 110 x 90 mm

(6.5 x 4.3 x 3.5'')Weight 1 kg (2.2 lb.)

Page 5: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

98

pH CALIBRATION WITH AUTOMATICTEMPERATURE COMPENSATION

(HI 8614 & HI 8614L)

Initial preparation

• Connect the pH electrodeto the BNC socket.

• Connect the temperatureprobe to the transmitter.

Pour small quantities of pH 7.01 and pH 4.01solution into two clean beakers.

For accurate calibration use two beakers foreach buffer solution, the first one for rinsingthe electrode, the second one for calibration.

To get accurate readings, use pH 7.01 andpH 4.01 if you are going to measure acidsamples or pH 7.01 and pH 10.01 for alkalinemeasurements.

Note: with HI8614L the instruments displaycan be used during calibration withoutthe need to connect the ammeter andthe reading is directly expressed in pHunits.

HI 7004 HI 7007

HI 7

004

The wire between the transmitter and therecorder/indicator/controller should be a PVCinsulated two wire with a wire diameter of atleast 0.7 mm. This wire is fed through one ofthe smaller cable glands. The maximum dis-tance between the power supply and theamplifier is 300 m (1000'). It is not necessaryto use shielded cable. The transmitter isprotected against inversion of supply voltage.

Electrode connection: connectthe BNC of the cable to the BNCsocket on the transmitter.

For HI8614 & HI8614L only: forautomatic temperature com-pensation, connect the 2 termi-nals of the temperature probe(HI 76608, optional) to "TEMP.PROBE" terminals.If automatic temperature compensation is notrequired, short the "TEMP.PROBE" terminals with a resis-tance according to the externaltemperature:

Temperature (°C) Resistance (Ohms)

0 1634

10 1774

20 1922

30 2078

40 2242

50 2412

60 2590

A 2 kohm resistor is factory mounted for 25°Ctemperature compensation.

HI 7007

CALIBRATION

HI 7007

RINSE

Page 6: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

1110

For other buffer temperatures, refer to page13 for the appropriate mA / pH reading.

• Rinse the electrode with tapwater or distilled water anda small amount of pH 4.01or 10.01 buffer solution orimmerse it in the pH 4 rinsesolution (2nd calibrationpoint). Dip the electrode andthe temperature probe intopH 4.01 (or 10.01) calibra-tion buffer solution, shakebriefly and wait for a fewminutes for reading to stabi-lize.

• Adjust the slope trimmer until the amme-ter reads 8.58 mA or the display shows"4.01" at 25°C. For other buffer tempera-tures, refer to page 13 for the appropriatemA / pH reading (HI 8614L).

If you are using pH 10.01buffer solution adjust theslope trimmer until the am-meter read 15.43 mA at25°C (77°F) or the value in-dicated at page 13.

FOR HI8614L ONLY

The Display Module is factory calibrated, sothat the LCD display results are referred tothe 4-20 mA loop current (e.g. LCD displays0.00 pH when loop current is 4.00 mA anddisplays 14.00 pH when current is 20.00 mA).

HI 7004

mA

mA

SLOPE ADJUSTMENT

OFFSET ADJUSTMENT

Procedure

• Disconnect the +ve supply cable from"+VE LOOP" terminal and connect a 20mAf.s. ammeter between the +ve cable and"+VE LOOP" terminal.

• Remove the protective capfrom the electrode, rinse itwith some pH 7.01 solutionor immerse it in the pH 7rinse solution, then immersethe pH electrode and tem-perature probe into pH 7.01calibration buffer solution;shake briefly and wait forthe reading to stabilize.

Note: the tip of the elec-trode should be sub-merged approximately4 cm (1½") into thesolution. The tempera-ture probe should belocated as close tothe pH electrode aspossible.

• Adjust the offset trimmer until the amme-ter reads 12mA or the display shows "7.01"(HI 8614L only) if the temperature of thebuffer is at 25°C.

HI 7007

4 cm1½"

mASLOPE ADJUSTMENT

OFFSET ADJUSTMENT

Page 7: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

1312

pH CALIBRATION WITH FIXEDTEMPERATURE COMPENSATION

(HI 8614 & HI 8614L only)

• Take the temperature of thebuffer solutions using aChecktemp or a thermom-eter with a resolution of atleast 1°.

• Connect the appropriate re-sistor to the "TEMP.PROBE" terminals (seepage 6) depending on thetemperature of the calibra-tion solution.

• Follow the procedure as outlined in thecalibration with automatic temperature com-pensation and use the thermometer read-ing to adjust the trimmer until the amme-ter reads the value according to the follow-ing table.

°C

Under normal application, adjustment on thismodule may not be necessary.

If routine check is required, the following pro-cedures shall be performed.

• Calibrate the transmitter module using aammeter, as explained in the previouspages.

• Simulate a 4.00 mA loop current for thetransmitter (i.e. pH 0.00 as Electrode in-put) and check for display reading.

• Simulate a 20.00 mA loop current for thetransmitter (i.e. pH 14.00 as Electrodeinput) and check for display reading.

Note: when the instrument is used in con-junction with the Hanna indicatorHI 8510T or controllers HI 8710T,HI 8711T, the calibration can also beperformed on the indicator/controller.In this case slight calibration adjust-ment can be made on the indicator/controller even if the whole system cali-bration is advised, always starting fromthe transmitter.

05

1015202530354045505560

3241505968778695

104113122131140

4.014.004.004.004.004.014.024.034.044.054.064.074.09

8.5768.5658.5658.5658.5658.5768.5878.5998.6108.6228.6338.6458.667

7.137.107.077.047.037.017.006.996.986.986.986.986.98

12.13712.10312.06912.03412.02312.00011.98911.97711.96611.96611.96611.96611.966

10.3210.2410.1810.1210.0610.01

9.969.929.889.859.829.799.77

15.78915.69715.62915.56015.49115.43415.37715.33115.28515.25115.21715.18215.159

°C °F pH m A pH m A pH m ATemperature pH 4.01 Buffer pH 7.01 Buffer pH 10.01 Buffer

Page 8: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

1514

SLOPE ADJUSTMENT

OFFSET ADJUSTMENT

• If the reading lies outsidethis range, adjust the slopeadjustment trimmer on thetransmitter for a reading justwithin this range.

The unit is now calibrated.

A complete calibration of the transmitter mod-ule is advised periodically.

This calibration procedure requires the HI 8427or the HI 931001 pH and ORP simulator tosimulate the ORP electrode.

HI 8427 or HI 931001 produce a known signalinto the system so that the faults of thesystem can be isolated.

• Connect the ORP transmitter to the simu-lator.

• Set the simulator to 0 mV and adjust theoffset trimmer to read 12 mA on the am-meter or 0 mV on the HI 8615L display (HI8615L only).

SLOPE ADJUSTMENT

OFFSET ADJUSTMENTmA

ORP CALIBRATION (HI8615 & HI8615L)

Initial preparationDisconnect the +ve supply cable from the"+VE LOOP" terminal and connect a 20 mAf.s. ammeter between the +ve cable and the"+VE LOOP" terminal. With HI 8615L theinstrument display can be used during cali-bration without the need to connect the am-meter. In this case the values are directlyexpressed in mV units.

Pour a small quantity of HI7020ORP solution into a beaker.

Procedure

• Connect the shorting BNCconnector to the ORP trans-mitter.

• Adjust the OFFSET AD-JUSTMENT trimmer on the module for adisplay of 12 mA on the ammeter or 0 mVon the instrument display (HI 8615L only).This sets the zero point for the transmit-ter.

• Connect the ORP electrode to the transmit-ter and immerse the tip of the electrodeinto a beaker of HI7020 ORP calibrationsolution, and check that the ammeter read-ing lies between 13.6 and 14.2 mA or theinstrument reading is between 200 and 275mV at 25°C (HI 8615L).

HI 7

020

mA

SLOPE ADJUSTMENT

OFFSET ADJUSTMENT

mA

HI 7020

Page 9: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

1716

FOR HI8615L ONLY

The HI 8615L is factory calibrated, and thedisplayed values are referenced to the 4-20mA loop current (e.g. LCD displays -1000mV when loop current is 4.00 mA and dis-plays +1000 mV when current is 20.00 mA).Under normal application, adjustment on thismodule may not be necessary. If routinecheck is required, the following proceduresshall be performed.

• Follow the above procedure for HI8615(see page 12).

• Simulate a 12.00 mA loop current for thetransmitter (i.e. 0 mV at Electrode input)and check display reading.

• Simulate a 20.00 mA loop current for thetransmitter (e.g. +1000 mV at Electrodeinput) and check display reading.

• Set the simulator to 350 mV and adjustthe slope trimmer to read 14.8 mA on theammeter or 350 mV on the HI 8615Ldisplay (HI 8615L only).

• Connect the ORP electrode to the moduleand immerse the tip of the electrode intothe beaker of HI 7020 ORP solution andcheck that the ammeter reading lies be-tween 13.6 and 14.2 mA or the instrumentreading is between 200 and 275mV at25°C (HI 8615L only).

• Only if the reading lies out-side this range, adjust theslope adjustment trimmer onthe transmitter to reflect areading within this range.

SLOPE ADJUSTMENT

OFFSET ADJUSTMENT

mA

mA

HI 7020

SLOPE ADJUSTMENT

OFFSET ADJUSTMENT

Page 10: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

1918

ELECTRODE CONDITIONINGAND MAINTENANCE

ReferenceFilling Hole

ReferenceFilling Hole

SensitiveWire

ReferenceWire

ReferenceJunction

GlassBulb

GlassBulb

ReferenceJunction

ReferenceJunction

ReferenceJunction

Platinum orGold tip

Platinum orGold tip

ReferenceWire

ReferenceWire

SensitiveWire

Glass BodypH Electrode

Plastic BodypH Electrode

Plastic BodyORP Electrode

Glass BodyORP Electrode

ReferenceWire

PREPARATIONRemove the protective cap.DO NOT BE ALARMED IF ANY SALT DE-POSITS ARE PRESENT.This is normal with electrodes and they willdisappear when rinsed with water.During transport tiny bubbles of air may haveformed inside the glass bulb. The electrodecannot function properly under these condi-tions. These bubbles can be removed by

Please note:

-1000 mV = 4 mA

0 mV = 12 mA

1000 mV = 20 mA

350 mV = 14.8 mA

200 mV = 13.6 mA

275 mV = 14.2 mA

1 mV = 12.008 mA

Note: when the meter is used in conjunctionwith the Hanna indicator HI 8512T, orthe controller HI 8720T, the calibrationcan also be performed on the indicator/controller. In this case slight adjust-ment can be made on the indicator/controller even if the whole system cali-bration is advised, always starting fromthe transmitter.

Page 11: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

2120

PERIODIC MAINTENANCE

Inspect the electrode and the cable. The cableused for connection to the meter must beintact and there must be no points of brokeninsulation on the cable or cracks on the elec-trode stem or bulb. Connectors must be per-fectly clean and dry. If any scratches or cracksare present on the electrode body, replacethe electrode. Rinse off any salt deposits withwater.

For refillable electrodesRefill it with fresh electrolyte (HI7071 for singlejunction or HI7082 for double junction elec-trodes). Allow the electrode to stand uprightfor 1 hour.Follow the Storage Procedure above.

CLEANING PROCEDURE

General Soak in Hanna HI7061 generalcleaning solution for approxi-mately 1 hour.

Removal of films, dirt or deposits on the mem-brane/junction:

Protein Soak in Hanna HI7073 proteincleaning solution for 15 minutes.

Inorganic Soak in Hanna HI 7074 inorganiccleaning solution for 15 minutes.

Oil /grease Rinse with Hanna HI7077 Oil &Fat cleaning solution.

IMPORTANT: After performing any of thecleaning procedures rinse the electrode thor-oughly with distilled water, drain and refill thereference chamber with fresh electrolyte, (notnecessary for GEL filled electrodes) and soakthe electrode in HI70300 storage solution forat least 1 hour before taking measurements.

"shaking down" the electrode as you woulddo with a glass thermometer.If the bulb and/or junction are dry, soak theelectrode in HI70300 storage solution for atleast one hour.

For refillable electrodesIf the fill solution (electrolyte) is more than1 cm (½") below the fill hole, add HI70823.5M KCl electrolyte solution for double junc-tion or HI7071 3.5M KCl+AgCl electrolytesolution for single junction electrodes. For afaster response unscrew the fill hole screwduring measurements.

For AmpHel® electrodesIf the electrode does not respond to pHchanges, the battery is run down and theelectrode should be replaced.

MEASUREMENT

Rinse the electrode tip with distilled water.Immerse the tip (4 cm / 1½") in the sampleand stir gently for approx. 30 seconds.For a faster response and to avoid crosscontamination of the samples, rinse the elec-trode tip with a few drops of the solution to betested, before taking measurements.

STORAGETo minimize clogging and assuring a quickresponse time, the glass bulb and the junc-tion should be kept moist and not allowed todry out. Replace the solution in the protectivecap with a few drops of HI70300 storagesolution or, in its absence, filling solution(HI7071 for single junction or HI7082 fordouble junction electrodes). Follow the Prepa-ration Procedure above before taking mea-surements.

Note: NEVER STORE THE ELECTRODE INDISTILLED WATER OR DRY.

Page 12: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

2322

TEMPERATURE-RESISTANCECORRELATION FOR HANNA

pH SENSITIVE GLASS

The resistance of glass electrodes partiallydepends on the temperature. The lower thetemperature, the higher the resistance. It takeslonger time for the reading to stabilize if theresistance is higher. In addition, the responsetime will suffer to a greater degree at tem-peratures below 10°C.

Since the resistance of the pH electrode is inthe range of 200 Mohm, the current acrossthe membrane is in the pico Ampere range.Large currents can disturb the calibration ofthe electrode for many hours.

For these reasons high humidity environ-ments, short circuits and static dischargesare detrimental for a stable pH reading.

The pH electrode's life also depends on thetemperature. If constantly used at high tem-peratures, the electrode life is drastically re-duced.

TROUBLESHOOTING

Evaluate your electrode performance basedon the following.

• Noise (Readings fluctuate up and down)could be due to:

- Clogged/Dirty Junction: refer to theCleaning Procedure above.

- Loss of shielding due to low electrolytelevel (in refillable electrodes only): HI7071for single junction or HI7082 for doublejunction electrodes.

• Dry Membrane/Junction: soak inHI70300 storage solution for at least onehour.

• Drifting: soak the electrode tip in warmHanna solution HI7082 for one hour andrinse the tip with distilled water. Refill withfresh HI7071 for single junction electrodesand HI7082 for double junction electrodes.

• Low Slope: refer to the Cleaning Proce-dure above.

• No Slope: check the electrode for cracksin glass stem or bulb and replace theelectrode.

• Slow Response/Excessive Drift: soak thetip in Hanna HI7061 cleaning solution for30 minutes, rinse thoroughly in distilledwater and then follow the Cleaning Proce-dure above.

-20 -20

-10

0 +10+20 +30+40 +50+60+70 +80+90 °C

-10

-10

Ω

-10

-10

-10

-10

2x109

1x109

2x108

1x108

2x107

1x107

Page 13: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

2524

INSTALLATION PROCEDUREAND EXAMPLES

The HI 8614, HI 8614L, HI 8615 and HI 8615Ltransmitters may be wall mounted in any con-venient location near the measurement point.To minimize thermal drift due to extreme tem-perature fluctuations during the measurementprocess, particularly if the measurement isconducted outdoors, it is best to protect thetransmitter in an enclosed casing.

General Installation Procedure

For most industrial application involving longterm monitoring and control, it is also recom-mended to use tank electrode holders (HI 6050or HI6051) to protect the pH electrode and thetemperature probe from contamination by thetest solution.

Controlling the pH / ORP with a recorder

Typical Electrode LifeAmbient Temperature 1- 3 years90 °C Less than 4 months120°C Less than 1 month

High concentrations of sodium ions interferewith readings in alkaline solutions; the pH atwhich the interference starts to be significantdepends upon the composition of the glass.This interference is the alkaline error andcauses the pH to be underestimated. Hanna'sglass formulations have the indicated charac-teristics.

Alkaline Error

1.0 Mol L-1 Na+

0.1 Mol L-1 Na+

Sodium Ion Correction for the Glassat 20-25°C

Concentration pH Error13.0013.5014.0012.5013.0013.5014.00

0.100.140.200.100.180.290.40

Page 14: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

2726

Monitoring and Controlling the pH with(HI 8711) Panel Mounting Indicator/

Regulator with Independent DosageControl for Acid and Base

Monitoring the pH/ORP with Panel MountingpH (HI8510)/ORP (HI8512) Indicator

Controlling the pH/ORP withan Industrial Regulator

Monitoring and Controlling the pH/ORPwith Panel Mounting Indicator/Regulator

and Dosage Control of either Acid or Base

Page 15: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

2928

pH ELECTRODES

HI 1090T Screwcap PG13.5 connector,double junction, glass body

HI 1110S Screw connector, single junction,glass body

HI 1130B/3 BNC connector, 3 m (9.9') cable,single junction, glass body

HI 1110S HI 1130B/3

HI 1110T Screwcap PG13.5 connector,double junction, glass body

HI 1114S Screw connector, double junctionplastic body

HI 1134B/3 BNC connector, 3 m (9.9') cable,double junction plastic body

HI 1114S HI 1134B/3

HI 1115S Screw connector, single junction,glass body

HI 1135B/3 BNC connector, 3 m (9.9') cable,single junction, glass body

HI 1115S HI 1135B/3

PG13.5 THREADφ 12mm φ 9.5mm

110mm30mm

DIA 9.5mm3/4 x 16 UNF

DIA 12mm

110mm38.5mm

DIA 20.5mm

M13 x 1.5

DIA16mm

7mm

25mm

PG13.5 THREADφ 12mm φ 9.5mm

110mm30mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

M13 x 1.5

DIA16mm

7mm

25mm

DIA 12mmDIA 7.6mm

150mm25mm

DIA 16.5mm5mm

M13 x 1.5

DIA16mm

7mm

25mm

ACCESSORIES

pH CALIBRATION SOLUTIONSHI7004M pH 4.01 buffer solution, 230 mLHI7004L pH 4.01 buffer solution, 500 mLHI7006M pH 6.86 buffer solution, 230 mLHI7006L pH 6.86 buffer solution, 500 mLHI7007M pH 7.01 buffer solution, 230 mLHI7007L pH 7.01 buffer solution, 500 mLHI7009M pH 9.18 buffer solution, 230 mLHI7009L pH 9.18 buffer solution, 500 mLHI7010M pH 10.01 buffer solution, 230mLHI7010L pH 10.01 buffer solution, 500 mL

ORP SOLUTIONSHI 7020M 200-275 mV test solution, 230 mLHI 7020L 200-275 mV test solution, 500 mLHI 7091M Reducing pretreatment solution, 230 mLHI 7091L Reducing pretreatment solution, 500 mLHI 7092M Oxidizing pretreatment solution, 230 mLHI 7092L Oxidizing pretreatment solution, 500 mL

ELECTRODE MAINTENANCE SOLUTIONSHI70300M Storage solution, 230 mLHI70300L Storage solution, 500 mLHI7061M General cleaning solution, 230 mLHI7061L General cleaning solution, 500 mLHI7073M Protein cleaning solution, 230 mLHI7073L Protein cleaning solution, 500 mLHI7074M Inorganic cleaning solution, 230 mLHI7074L Inorganic cleaning solution,500 mLHI7077M Oil & Fat cleaning solution, 230 mLHI7077L Oil & Fat cleaning solution, 500 mLHI7071 3.5M KCl + AgCl electrolyte solu-

tion, 4 x 50 mL bottle, for single junc-tion electrodes

HI7072 1M KNO3 electrolyte solution, 4 x

50 mL bottleHI7082 3.5M KCl electrolyte solution, 4 x

50 mL bottle, for double junctionelectrodes

Page 16: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

3130

HI 2910B/5 BNC connector, 5 m (16.5') cable,double junction, plastic body, built-in amplifier

ORP ELECTRODESHI 2930B/5 BNC connector, 5 m (16.5') cable,

Pt, plastic body, built-in amplifier

HI 3110S Screw-type connector, Pt, glassbody

HI 3130B/3 BNC connector, 3 m (9.9') cable,Pt, glass body

HI 3110S HI 3130B/3

HI 3110T Screwcap PG13.5 connector, Pt,glass body

HI 3115S Screw-type connector, side-arm,Pt, glass body

HI 3135B/3 BNC connector, 3 m (9.9') cable,side-arm, Pt, glass body

HI 3115S HI 3135B/3

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

M13 x 1.5

DIA16mm

7mm

25mm

PG13.5 THREADφ 12mm φ 1mm

110mm30mm

DIA 12mmDIA 7.6mm

150mm25mm

DIA 16.5mm5mm

M13 x 1.5

DIA16mm

7mm

25mm

HI 1210T Screwcap PG13.5 connector,double junction, plastic body

HI 1910B BNC connector, 1 m (3.3') cable,double junction, plastic body, built-in amplifier

HI 1911B BNC connector, 1 m (3.3') cable,double junction, plastic body, built-in amplifier

HI 1912B BNC connector, 1 m (3.3') cable,double junction, plastic body, built-in amplifier

HI 1912B/5 BNC connector, 5 m (16.5') cable,double junction, plastic body, built-in amplifier

HI 2114B/5 BNC connector, 5 m (16.5') cable,double junction, plastic body

PG13.5 THREADφ 12mm

110mm30mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

DIA 20.5mm

DIA 9.5mm3/4 x 16 UNFDIA 12mm

110mm38.5mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

Page 17: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

3332

EXTENSION CABLES FOR SCREW-TYPEELECTRODES ONLY (SCREW TO BNC CON-NECTOR)

HI7855/1 Extension cable 1 m (3.3') long

HI7855/3 Extension cable 3 m (9.9') long

HI7855/5 Extension cable 5 m (16.5') long

HI7855/10 Extension cable 10 m (33') long

HI7855/15 Extension cable 15 m (49.5') long

OTHER ACCESSORIES

BL PUMPS Dosing pumps with flow rate from1.5 to 20 lph

HI 98501 Pocket-size, thermometer withpenetration probe and 0.1°Cresolution (range -50.0 to150.0°C)

HI6050 & HI 6051 Submersible electrodeholders

HI 6054 & HI 6057 Electrode holders for in-lineapplications

HI 7871 & HI 7873 Level controllers

HI8427 pH / ORP electrode simulator

HI931001 pH / ORP electrode simulatorwith LCD

CONNECT TOSCREW TYPEELECTRODES

CONNECT TO THEBNC SOCKET

OF THE METER

HI 7855 SERIES CABLE CONNECTORSCONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC

HI 3210T Screwcap PG13.5 connector, Pt,plastic body

HI 3410S Screw connector, Pt, plastic bodyHI 3430B/3 BNC connector, 3 m (9.9') cable,

Pt, plastic body

HI 3410S HI 3430B/3

HI 3932B/5 BNC connector, 5 m (16.5') cable,Pt, plastic body, built-in amplifier

HI 4110S Screw-type connector, Au, glassbody

HI 4130B/3 BNC connector, 3 m (9.9') cable,Au, glass body

HI 4110S HI 4130B/3

HI 4932B/5 BNC connector, 5 m (16.5') cable,Au, plastic body, built-in amplifier

PG13.5 THREADφ 12mm

110mm30mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

M13 x 1.5

DIA16mm

7mm

25mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

M13 x 1.5

DIA16mm

7mm

25mm

DIA 20.5mm

3/4 x 16 UNF DIA 12mm

110mm38.5mm

Page 18: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

3534

Recommendations for Users

Before using these products, make sure that they are entirely suitable for theenvironment in which they are used.

Operation of these instruments in residential area could cause unacceptableinterferences to radio and TV equipments, requiring the operator to take all necessarysteps to correct interferences.

Any variation introduced by the user to the supplied equipment may degrade theinstruments' EMC performance.

To avoid damages or burns, do not perform any measurement in microwave ovens.

CE DECLARATION OF CONFORMITYWARRANTY

All Hanna Instruments meters are warrantedfor two years against defects in workman-ship and materials when used for their in-tended purpose and maintained according toinstructions.

The probes and the electrodes are war-ranted for a period of six months.

Damages due to accident, misuse, tamper-ing or lack of prescribed maintenance are notcovered. This warranty is limited to repair orreplacement free of charge.

If service is required, contact the dealer fromwhom you purchased the instrument. If underwarranty, report the model number, date ofpurchase, serial number and the nature of thefailure. Obtain a Returned Goods Authoriza-tion from the Customer Service departmentfirst and then return the instrument with theAuthorization # included along with shipmentcosts prepaid. If the repair is not covered bythe warranty, you will be notified of the chargefor repair or replacement. When shipping anyinstrument, make sure it is properly pack-aged for complete protection.

All rights are reserved. Reproduction in wholeor in part is prohibited without the writtenconsent of the copyright owner.

Hanna Instruments reserves the right tomodify the design, construction and appear-ance of its products without advance notice.

Page 19: HI 8614 • HI 8614L HI8615 † HI8615L - Hanna Instruments · PDF fileHI 8614 • HI 8614L HI8615 † HI8615L ... pH Calibration with Manual Temperature Compensation ... the need

36

MA

NP

HT

RR

3 11/05

SALES AND TECHNICAL SERVICE CONTACTS

Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699

China:Tel. (10) 88570068 • Fax (10) 88570060

Egypt:Tel. & Fax (02) 2758.683

Germany:Tel. (07851) 9129-0 • Fax (07851) 9129-99

Greece:Tel. (210) 823.5192 • Fax (210) 884.0210

Indonesia:Tel. (21) 4584.2941 • Fax (21) 4584.2942

Japan:Tel. (03) 3258.9565 • Fax (03) 3258.9567

Korea:Tel. (02) 2278.5147 • Fax (02) 2264.1729

Malaysia:Tel. (603) 5638.9940 • Fax (603) 5638.9829

Singapore:Tel. 6296.7118 • Fax 6291.6906

South Africa:Tel. (011) 615.6076 • Fax (011) 615.8582

Taiwan:Tel. 886.2.2739.3014 • Fax 886.2.2739.2983

Thailand:Tel. 66.2619.0708 • Fax 66.2619.0061

United Kingdom:Tel. (01525) 850.855 • Fax (01525) 853.668

USA:Tel. (401) 765.7500 • Fax (401) 765.7575

For e-mail contacts and complete list of Sales andTechnical offices, please see www.hannainst.com