Upload
ankita-pujar
View
217
Download
0
Embed Size (px)
Citation preview
8/7/2019 Hermans1
http://slidepdf.com/reader/full/hermans1 1/8
Telemetric Monitoring of Foetal ECG
B.Hermans and R. Puers
KULeuven, Department ESAT-MICAS, Belgium
L. Lewi, J. Jani and J. DeprestKULeuven, Dept Obstetrics-Gynaecology, Belgium
ABSTRACTIn this paper we propose an implantable telemetric system that meets the clinical needs for
accurate foetal ECG monitoring during labour for all pregnancies where monitoring is
indicated, as well as during pregnancy for a selected number of foetuses undergoing foetal
surgery.
INTRODUCTIONAs a consequence of recent developments, one can state that the foetus has become a patient,
accessible for diagnosis and therapy. Foetuses may become critically ill during their in uterolife, and may benefit from timely delivery or foetal therapy. Until now, continuous in utero
foetal monitoring has not been possible, since it requires permanent and direct access to the
foetus. Therefore, information from the unborn patient can only be retrieved by indirect and
often inaccurate means (ultrasound, doppler and cardiotocography). If these findings are
suggestive of foetal distress, a foetal blood sample must be taken for confirmation. Foetal
blood sampling is an invasive procedure with its associated risks and offers only momentary
information on the foetal status. The availability of a device that allows for continuous and
accurate monitoring of foetal well-being will reduce the inappropriate preterm delivery or
therapy of foetus, with good foetal well-being and avoid foetal death or brain damage, where
foetal distress has gone unrecognised.
Figure 1: Foetal Endoscopy
(Foetoscopy)This device will be an indispensable tool for foetus undergoing foetal surgery using a recently
developed technique ofoetoscopy (foetal - endoscopy). The rationale is to take advantage of
the already established access to the foetus during the operation, and at that time to attach
an autonomous monitoring system to it, through an adapted trocart. See Figure 1 for
surgical foetoscopic technique.
The same system can also be used for monitoring during labour for all pregnancies
where foetal monitoring is required. Whereas foetal ECG monitoring is currentlyavailable only non-telemetricly, which implies that the woman is immobilized to the
trocart
8/7/2019 Hermans1
http://slidepdf.com/reader/full/hermans1 2/8
recording device during labour, the telemetric intrapartum sensor may be applied on the foetal
scalp like the foetal heart rate electrodes, which are currently used. However, mobilization
during labour will be unrestricted, which benefits the progress of labour and increases the
sense of well-being of the woman.
In the present paper two consecutive approaches are highlighted. The first consists of a simple
telemetric device that continuously transmits the foetal information to the outside world. The
analysis is performed at the receiver site.
The second approach is a novel telemetric unit that will also contain on board circuitry to
record a local analysis. This system will be miniaturised in the future to allow easy and safe
access to the foetus.
METHODS
FIRST EXPERIMENTSWe started this research by investigating the possibility of measuring a foetal ECG by
telemetric means. Therefore we used a commercial available sensor
EA-F20 (DSI, Minnesota, figure 2). This 2 lead miniature sensor measures ECG on a
continuous base when it is magnetically activated.
Figure 2: ECG sensor
The signal of the sensor is transmitted and subsequently captured via the maternal abdomen of
the ewe by an antenna. This antenna also provides the necessary electronics to convert the RF
signal into an analog ECG signal. This signal is then further digitalised using a Keithley
Analog-to-Digital PCI/MCIA card. The received digital information is extracted from this
card and stored into the PC using a program developed with Testpoint.
Medical research has shown that the use of ECG to obtain a beat-to-beat analysis (Nijhuis, J.,
et al) can, in combination with ST-analysis, provide the gynaecologist enough information todetect and prevent foetal stress (Sundström, A., et al). Beat-to-beat analysis consists of the
calculation of the R-R interval of the ECG, whereas ST-analysis consists of determining the
T/QRS ratio.
Figure 3 illustrates the above used terms.
8/7/2019 Hermans1
http://slidepdf.com/reader/full/hermans1 3/8
RR-interval
QRS-amplitude
Q
S
R
Q
S
R
PT
T-amplitude
QRS-complexST-interval
contraction
heart chamberschanges during
hypoxia
contraction of
the atria
eart requency
‘
Figure 3: ECG description
To perform the medical analysis, a program in Visual Basic able to extract automatically beat-to-beat and ST analysis, has been written. Figure 4 shows a typical screenshot generated
by this program. The upper trace shows the raw received ECG signal. The extraction of the
QRS-complex is calculated by an algorithm based on a level decision criterion. The result of
this algorithm can be found on the second trace. The third trace is a conversion of the second
one into a beat/minute scale. This trace can be used for a medical beat-to-beat evaluation.
Mean, maximum and minimum heart rate is also automatically calculated. The next trace
shows the variation in heart rate and is also useful for medical interpretation. The ST analysis
can be found on the last trace. With the display of these traces the gynaecologist has sufficient
information to decide on the necessary therapy.
‘
8/7/2019 Hermans1
http://slidepdf.com/reader/full/hermans1 4/8
Figure 4: Analysis program
The telemetric device was implanted subcutaneously in a foetal lamb at day 119 of
gestation (term = 147 days). (Figure 5)
Figure 5: Surgical procedure
After the implantation of the sensor on the foetal lamb, telemetric data was obtained until day
11. On the last day of recording, an elevation in the ST-segment and bradycardia were
present, suggesting foetal hypoxia. On day 12 no foetal ECG signal could be obtained and
indeed the lamb was stillborn 2 days later. Figure 6 shows traces of the first and last day of recording which illustrates clearly the bradycardia and ST changes as described.
Figure 6: ECG recording on first and last day
First day
-0,4
-0,3
-0,2
-0,1
0
0,1
0,2
0,00 0,17 0,34 0,50 0,67 0,84 1,00
time (s)
Last day
-0,6
-0,4
-0,2
0
0,2
0,4
0,00 0,17 0,34 0,50 0,67 0,84 1,00
time (s)
8/7/2019 Hermans1
http://slidepdf.com/reader/full/hermans1 5/8
Table 1 is generated from the written program and demonstrates the daily extracted results
until day 11. The high ST ratio on the first day may be explained by the foetal stress of the
surgical procedure.
First day Day 2 Day 5 Day 6 Day 7 Day 8 Last day(day 11)
ST (T/QRS) 0,36 0,24 0,25 0,23 0,26 0,22 0,47
Baseline HeartRate
212 177 174 185 153 152 108
Table 1: Daily recordings
NOVEL MONITORING SYSTEM
The experiments described above have proven the feasibility of telemetric monitoring
of foetal ECG. Nevertheless this sensor cannot be used in a clinical environment for foetal monitoring. First of all the size of the used sensor is too big for inserting the
sensor through the trocar during fetoscopy. The maximum diameter of a devicecapable of entering the trocar is 0.5 cm; length is not a critical parameter. Another
shortcoming is the absence of any intelligence in the sensor. The used sensor has no
capacity of making any decisions. The monitoring system that we propose will consist
of following parts as shown in figure 7 :
Figure 7: Monitoring System
1 2 3
1 : Sensor 2 : Intelligent Core3 : Bi-directional telemetric unit
8/7/2019 Hermans1
http://slidepdf.com/reader/full/hermans1 6/8
1: Sensor
A 2 lead ECG amplifier has been developed. The biggest challenge is to eliminate the
common mode signal that disturbs the ECG signal, whereas a 3 electrode ECG amplifier
would be used the third electrode will overcome this common mode interference (Nitish V., et
al.). The 2 electrodes may be implanted subcutaneously on the foetus at a distance of 2 – 3
cm.
2: Intelligent core
Instead of transmitting the entire signal, the recorded parameter will be analysed to eliminate
superfluous data. The system will contain algorithms that can automatically detect foetal
health condition. These algorithms will be based on the already described beat-to-beat and ST
analysis. The ultimate goal is to have a system that contains the intelligence to detect any
abnormality and transmit the necessary information.
3: Bi-directional telemetric unit
The device must be able to send and receive data wirelessly with low power consumption.
The possibility to communicate with the device while it is implanted, offers the unique
opportunity to alter the registration parameters or algorithms at the end-users’ request.
Prototype
A first prototype has been assembled. The 2 lead ECG amplifier has been developed using
bootstrapped amplifiers to reduce 50 Hz interference, followed by low and high analog filters
to eliminate signals that are no part of the ECG bandwidth. This creates an operating
frequency range between 0.5 and 150 Hz. The ECG is amplified in such a way that it is
adapted to the full range of the A/D converter. The board also provides a bandpass filter
between 5 and 15 Hz that will extract the QRS complex of the ECG signal (Jiapu P., et al).
A Flash PIC microcontroller PIC12F629 is used. It contains a 10 bit A/D converter. This
microcontroller is programmed to record a sample at a sampling frequency of 300Hz. The
program also provides what is necessary to treat the samples in a certain way that they can be
sent to a RS-232 converter. By doing so, data can be directly recorded on a PC. The boardalso has a transceiver TR3000 (RFM) with the antenna integrated on the PCB. Through this,
ECG can also be transmitted wirelessly. The PCB is powered with +3V and –3V obtained
through silver oxide button cells.
Figure 8 gives the schematic overview of the board. Figure 9 shows a picture of the developed
circuit.
Figure 8: Schematic overview
Figure 9: Developed circuit (9 cm x 4 cm)
ECG amp fil ters & amp
bandpass
filter battery
A/D
RS-232
transceiver
433Mhz
µ c o n t r o l l e r
8/7/2019 Hermans1
http://slidepdf.com/reader/full/hermans1 8/8
ACKNOWLEDGEMENTSThe authors which to acknowledge the support for this program “Verkennende Internationale
Samenwerking” (stimulation of research projects within the fifth framework program),
sponsored by the Flemish Regional Government; VIS/00/009, telemetric and non-telemetric
microsensors for perinatalogy.
LITERATURE CITEDDSI. Data Sciences International. www.transomamedical.com. Minnesota
Nijhuis, J., et al. Foetale bewaking. Elsevier/Bunge. 1998
Sundström A., et al.’ Foetale bewaking met STAN’. Neoventa Medical . 2000
Nitish V. Thakor and John G. Webster. ‘Ground-free ECG Recording with Two Electrodes’.
Trans. Biomed. Eng . Vol 12:pp 699-704. IEEE 1980.
Jiapu P. and Willis J. Tompkins.. ‘A Real-Time QRS Detection Algorithm’. Biomed. E ng . 3:
Vol pp 230-236. IEEE 1985 Trans.