How to operate the HemoCue ® Hb 801 System

HemoCue® Hb 801 System

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: HemoCue® Hb 801 System

How to operate the HemoCue® Hb 801 System

Page 2: HemoCue® Hb 801 System

HemoCue® Hb 801 System

How to Run a Test – Capillary & Venous Samples 3Cleaning and Disinfection 9Quality Control 11Quality Control – External Control 12Settings – Date and Time 13Settings – Bluetooth 14Troubleshooting 15

2GPM512INT 190423

Page 3: HemoCue® Hb 801 System

How to Run a Test

In order to obtain reliable and accurate results, correct capillary sampling collection technique is key.

• Always wear protective gloves• HemoCue® Hb 801 Microcuvettes are for single

use only • Follow local safety procedures for disposal of used

microcuvettes

3GPM512INT 190423

Page 4: HemoCue® Hb 801 System

How to Run a Test – Capillary Samples

Make sure the patient’s hand is warm and relaxed. Use the middle or ring finger for sampling – Avoid fingers with rings on. Sample at the side of the fingertip for best blood flow and comfort.

1.

2.

4.

3.

5.

Clean the fingertip with disinfectant, and allow to dry.

Using your thumb, lightly press the finger from the top of the knuckle towards the fingertip to stimulate blood flow.

Press lightly towards the fingertip, and puncture using a high-flow lancet.

Wipe away the first 2 or 3 drops of blood. Press lightly towards the fingertip until another drop of blood appears.

4GPM512INT 190423

Page 5: HemoCue® Hb 801 System

How to Run a Test – Capillary SamplesFill the microcuvette. Make sure the blood drop is large enough to fill the microcuvette completely in one single step. Do not refill. Do not let more than 40 seconds pass between step 6 (filling) and step 9 (inserting the microcuvetteinto the microcuvette holder).

6.

7.

8.

Wipe off excess blood from the outside of the microcuvette. Make sure that no blood is drawn out from the microcuvette.

Make a visual inspection. Precaution: If the microcuvette is not completely filled with blood, or if there are air bubbles, discard and fill a new microcuvette.

When a result is displayed, remove and discard the microcuvette. The result is displayed for 10 seconds after measurement. To change this setting, see section Connectivity and External Applications.

10.

9. Make sure the analyzer is in Ready State. Insert the filled microcuvette into the microcuvette holder and press down. The result will be displayed within a second.

5GPM512INT 190423

Page 6: HemoCue® Hb 801 System

How to Run a Test – Venous Samples

• Use anticoagulant (K2-EDTA or Li-Heparin) • Venous samples can be stored and transported for up to 24 hours, at room

temperature or if in refrigerator before measuring• Make sure venous samples are properly mixed before use

6GPM512INT 190423

Page 7: HemoCue® Hb 801 System

How to Run a Test – Venous Samples

If refrigerated, allow the sample to reach operating conditions. Mix thoroughly on mixer for at least 2 minutes or invert manually 8-10 times.

When mixing controls, follow instructions for the control material used.

Place a drop of blood or control material onto a hydro-phobic surface, using a suitable transfer device.

Fill the microcuvette completely – in one single step. Do not refill.

Follow immediately steps 7-10 in section Measuring Capillary Blood, to perform the measurement. Do not let more than 40 seconds pass between filling and inserting the microcuvette into the microcuvette holder.

1.

3.

2.

7GPM512INT 190423

Page 8: HemoCue® Hb 801 System

How to Run a Test – Venous Samples

Wipe off excess blood from the outside of the microcuvette. Make sure that no blood is drawn out from the microcuvette.

Make a visual inspection. Precaution: If the microcuvette is not completely filled with blood, or if there are air bubbles, discard and fill a new microcuvette.

Make sure the analyzer is in Ready State. Insert the filled microcuvette into the microcuvette holder and press down. The result will be displayed within a second.

When a result is displayed, remove and discard the microcuvette. The result is displayed for 10 seconds after measurement. To change this setting, see section 14 in the Operating Manual.

To see the result again, see section 13 in the Operating Manual.

4.

5.

6.

7.

8GPM512INT 190423

Page 9: HemoCue® Hb 801 System

1. Turn off the analyzer, and remove the microcuvetteholder.

2. Lightly dampen a cotton swab with cleaning agent. Clean all surfaces in the cavity; make sure to clean all the way down.

3. Clean the microcuvette holder with cleaning agent. Let the microcuvette holder dry outside of the analyzer while moving on to step 4.

4. Lightly dampen a wipe with cleaning agent, and clean all outer surfaces. Now the analyzer is ready for disinfection.

Cleaning and DisinfectionCleaning

9GPM512INT 190423

Page 10: HemoCue® Hb 801 System

6

8

55. Wipe the microcuvette holder with disinfectant.

6. Wipe all outer surfaces with disinfectant.

7. Allow the disinfectant to remain on the microcuvetteholder and analyzer surfaces according to instructions for the disinfectant used (wet-time/contact time).

8. Remove any excess disinfectant, or allow to air dry. Make sure all parts are completely dry before reattaching the microcuvette holder, and turning on the analyzer.

Cleaning and Disinfection

7

Disinfection

10GPM512INT 190423

Page 11: HemoCue® Hb 801 System

• An internal quality control (self-test) automatically verifies analyzer performance:

• Every time the analyzer is turned on• When the cuvette holder is put back into

place after removal• Every hour when in use

• Quality control and cleaning reminders can be set via HemoCue® 801 Connect

Quality Control

11GPM512INT 190423

Page 12: HemoCue® Hb 801 System

If external quality control is required by local or other regulations, use liquid controls recommended by HemoCue

HemoTrol WB• Package: 2x1.0 mL• 3 levels

• Low: ~9.4 g/dL (94 g/L, 5.8 mmol/L)• Normal: ~13 g/dL (130 g/L, 8.0 mmol/L)• High:~16 g/dL (160 g/L, 9.9 mmol/L)

• Shelf-life: 3 months

Quality Control – External Control

12GPM512INT 190423

Page 13: HemoCue® Hb 801 System

Settings – Date and Time

Connectivity and External Applications

The analyzer can be connected to external software applications, using either the USB cable, or Bluetooth® Low Energy (wireless connection).*

This allows for change of certain default features or transfer of data.

* Software applications are optional items. For more information, contact the local distributor or HemoCue AB.

Follow these procedures to change date and time settings.To exit settings at any time, press both buttons once.

Date and Time Buttons Extra information

1. Make sure the analyzer is in Ready State.

2. Enter: Press and hold buttons until the date, time, and Bluetooth areas on the display starts to flash.

3. To change date and time, press left button.

4. Set the correct values.

Go to next value.

5. Exit: Press both buttons once.

Display showing:

To quickly scroll between digits, press and hold the left button.

13GPM512INT 190423

Page 14: HemoCue® Hb 801 System

Settings – BluetoothFollow these procedures to change Bluetooth settings.To exit settings at any time, press both buttons once.

Bluetooth Buttons Extra information

1. Make sure the analyzer is in Ready State.

2. Enter: Press and hold buttons until the date, time, and Bluetooth areas on the display starts to flash.

3. To turn Bluetooth on/off, press right button.

4. Choose on or off.

5. Exit: Press both buttons once.

Display showing:

On – Bluetooth icon visibleOff – no Bluetooth icon visible

14GPM512INT 190423

Page 15: HemoCue® Hb 801 System

Actions to solve most problems:

1. Remove the microcuvette from the microcuvetteholder

2. Turn off the analyzer and turn it on again3. Clean the analyzer (surfaces in the cavity) and let the

analyzer dry4. Make sure the analyzer has reached operating and

ambient conditions before start-up5. Avoid direct sunlight

Troubleshooting

15GPM512INT 190423

Page 16: HemoCue® Hb 801 System

Troubleshooting, cont.If the problem remains, the below Further Actions should be implemented. • If the problem still remains after Further Actions, contact the local distributor or

HemoCue AB

Perform Main Actions above.

Perform Main Actions above. Make sure to press down the microcuvette during measurement.

If the display shows --- instead of a result; fill a new microcuvette and perform a measurement, see section 8 or 9 in the Operating Manual.

Possible causes are e.g. incorrect sampling technique or incorrect mixing of sample. Fill a new microcuvette and perform a measurement, see section 8 or 9 in the Operating Manual. Check expiration date for the materials used. Make sure all result segments on the display are working properly, see section 5 in the Operating Manual.

The results is below (LLL) or above (HHH) measurement range.

Symptom Further Actions

analyzer not measuring

unexpected result*

16GPM512INT 190423

Page 17: HemoCue® Hb 801 System

The microcuvette is empty.

If the analyzer is continuously busy, turn the analyzer off and then on again.

Change or charge the battery/batteries, or make sure the adapter is properly connected and that the USB cable is not damaged.

Change or charge the battery/batteries, or make sure the adapter is properly connected and that the USB cable is not damaged. Force a restart: Press and hold both buttons for 15 seconds.

Set date and time. If the problem occurs again: Set date and time, and keep the analyzer powered for at least 24 hours (either by batteries, or USB cable).

Follow instructions for software application used.

The microcuvette holder is either not in position, missing or broken.

Troubleshooting, cont.

*If the result is still deviating from expected value after performing all actions in the list, it is recommended to confirm the result with a suitable laboratory method. Follow local recommendations regarding obtaining and shipping blood samples for confirmation testing.

sudden analyzer shut down, or battery indicator flashing

analyzer not responding or starting

date and time area flashing in Ready State

bad wireless or USB connection

Clean the analyzer according to steps 1-2 in section 10 in the Operating Manual. Let the analyzer dry.

17GPM512INT 190423

Page 18: HemoCue® Hb 801 System

Because when it comes to caring for people, we refuse

to compromise.

18GPM512INT 190423