The First
International Proficiency Testing Conference
Sinaia, Romnia 11th 13th October, 2007
386
HEAVY METAL DETERMINATIONS FOR THE EMISSIONS FROM THE
INCINERATION PLANTS BY ATOMIC ABSORBTION SPECTROMETRY WITH GRAPHITE FURNACE
Alice Pop, Margareta Toader, Tatiana Paraschiv S.C. CEPROCIM S.A., B-dul Preciziei, nr. 6, Sector 6, Cod: 062203 Bucuresti
[email protected] Abstact The European Union Directive 61/96 establishes the general conditions to prevent and control polution. As follows, all factories which incinerate or co-incinerate wastes, among them there is also the cement industry, must survey and limit the emissions of poluants from their incineration plants. European standard EN 14385:2004 is the basis used to develop the methods for the analysis of heavy metals from the emissions of the incineration plants. This paper presents the working conditions, the technical parameters of methods and the steps of developing the testing methods for the dust from the emissions of the incinerating plants, which were reteined on glass fibre filters. Key words Heavy metals, emissions, atomic absorbtion spectrometry 1. INTRODUCERE
Directiva Consiliului Uniunii Europene 61/96 stabileste cadrul general pentru prevenirea si controlul poluarii, preciznd masurile necesare implementarii integrate a prevenirii si controlului poluarii, n scopul obtinerii unui nivel ridicat de protectie pentru mediu ca ntreg, avand n vedere ca aplicarea principiului de dezvoltare durabila va fi promovata printr-o abordare integrata a controlului poluarii. Directiva 76/2000, transpusa in legislatia nationala prin Hotararea de Guvern nr. 128/2002, modificata si completata prin HG 268/2005, frecventa masuratorilor si valorile limit pentru emisii n aer prezentate n tabelul 1.
Alice Pop, Margareta Toader, Tatiana Paraschiv: Heavy metal determination for the emissions from the incineration plants by atomic absorbtion spectrometry with graphite furnace
387
Tabelul 1 Valorile limita admise pentru emisiile n aer conform Directivei 76/2000
Nr. crt. Elememtele analizate Conditia [mg/Nm] 1 Cd + Tl 0.05 2 Hg 0.05 3 As + Co + Cr + Cu + Mn + Ni + Pb + Sb + V 0.5
Decelarea elementelor de analizat la concentratiile maxime admise implica utilizarea unor metode de analiza instrumentala de inalta rezolutie. Lucrarea prezinta rezultatele obtinute la stabilirea parametrilor si a conditiilor de lucru la analiza pulberilor din emisiile de la instalatiile de ardere, retinute pe filtre din fibra de sticla. 2. MATERIALE SI APARATURA 2.1 Materiale S-au utilizat solutii etalon, monoelement, cu concentratia de 1000 mg/l. Reactivii utilizati pentru pregatirea probelor de analizat au fost acid fluorhidric, acid azotic, acid boric, acid clorhidric, borohidrura de sodiu, hidroxid de sodiu, apa ultrapura. Filtrele din fibra de sticla cu proba de analizat si filtre din fibra de sticla curate utilizate ca probe oarbe, pentru stabilirea aportului de elemente de analizat din filtre si reactivi. 2.2 Aparatura Pentru aducerea in solutie a probelor de analizat s-a utilizat un mineralizator cu microunde, care permite dezagregarea in conditii de presiune si temperatura ridicate, in timp scurt. Mineralizatorul cu microunde are capacitatea de 27 l, este prevazut cu un megatron de 2450 MHz, cu posibilitate de programare a duratei de ncalzire si a puterii microundelor. Vasele de mineralizare sunt confectionate din teflon si au un volum de 70 ml. Stabilirea parametrilor de lucru si analiza probelor s-a efectuat pentru elementele Cd, Co, Cr, Cu, Mn, Ni, Pb, Sb si V, cu un spectrometru de absorbtie atomica tip Nova 400, cu cuptor de grafit ncalzit transversal, cu monocromator cu 1600 linii/mm, fanta reglabila la 0.2, 0.5, 0.8, si 1.2 nm si corectie de fond spectral cu lampa de deuteriu. 3 MOD DE LUCRU 3.1 Stabilirea conditiilor de mineralizare Aducerea n solutie a probelor de analizat se face, n prima etapa, cu amestec de acid azotic si acid fluorhidric, urmata, n etapa a doua, de solubilizarea fluorurilor cu acid boric. Procesele au loc la temperatura si presiune ridicate, n mineralizatorul cu microunde. Programele de mineralizare cuprind cte 3 pasi pentru fiecare din cele doua etape de mineralizare.
Alice Pop, Margareta Toader, Tatiana Paraschiv: Heavy metal determination for the emissions from the incineration plants by atomic absorbtion spectrometry with graphite furnace
388
Obtinerea programului de mineralizare s-a facut prin stabilirea conditiilor de temperatura, putere a microundelor si durata de mineralizare pentru fiecare dintre cei 3 pasi ai programului, pentru fiecare din etapele de dizolvare a probei. Dupa ncercari succesive s-au stabilit conditiile de mineralizare necesare pentru dizolvarea completa si n conditii de siguranta a probelor de emisii de la instalatiile de ardere, asa cum rezulta din tabelul 2.
Tabelul 2 Conditiile de lucru pentru mineralizarea probelor prelevate de la instalatiile de ardere
Etapa Mineralizare cu HNO3 si HF (2 : 1) Mineralizare cu H3BO3 5% Temperatura [C]
145 150 160 140 140 140
Durata [min] 5 5 10 5 2 4 Putere [W] 450 450 400 450 450 400
3.2 Stabilirea parametrilor de lucru ai spectrometrului de absorbtie atomica cu cuptor de grafit
Functionarea unui spectrometru de absorbtie atomica cu cuptor de grafit implica stabilirea conditiilor de lucru pentru partea optica, cuptorul de grafit si autosampler [1, 2]. 3.2.1 Partea optica a spectrometrului Stabilirea caracteristicilor optice ale spectrometrului de absorbtie atomica se face n functie de elementul de analizat, concentratia acestuia n proba, densitatea de linii spectrale corespunzatoare matricii probei si energia necesara lampilor pentru domeniul de lungimi de unda la care se lucreaza. Parametrii optici care se stabilesc sunt: lungimea de unda si marimea fantei optice, precum si curentul de lampa si nivelul de amplificare ale lampilor catodice si lampii cu deuteriu. In tabelul 3 se prezinta caracteristicile optice pentru elementele analizate.
Tabelul 3 Caracteristicile optice ale spectrometrului de absorbtie atomica utilizat pentru determinari
Elementul Lungimea de unda [nm] Fanta [nm] Cd 228.8 0.8 Co 240.7 0.2 Cr 357.9 0.8 Cu 324.8 0.8 Mn 279.5 0.2 Ni 232.0 0.2 Pb 217.0 0.2 Sb 217.6 0.2 V 318.4 0.8
Pentru curentul de lampa si nivelul de amplificare se utilizeaza valorile recomandate de soft-ul aparatului si acestea pot varia n timp cu cresterea gradului de epuizare al lampilor.
Alice Pop, Margareta Toader, Tatiana Paraschiv: Heavy metal determination for the emissions from the incineration plants by atomic absorbtion spectrometry with graphite furnace
389
3.2.2 Cuptorul de grafit Stabilirea parametrilor de lucru ai cuptorului de grafit implica stabilirea temperaturilor de procesare a probelor, corespunzator proceselor de uscare, piroliza si atomizare, a vitezelor de ncalzire si a temperaturii de curatare a cuptorului de grafit la finalul determinarii. Acesti parametri sunt caracteristici fiecarui element determinat si se stabilesc prin ncercari succesive, urmarind forma si naltimea peak-urilor obtinute.In figurile 1 si 2 se prezinta peak-urile pe baza carora s-au stabilit temperaturile de piroliza si atomizare pentru analiza probelor de emisii de la instalatiile de ardere.
Abs
orba
nta
250C 300C 350C 400C 450C 500C
Pasul
Abs
orba
nta
250C 300C 350C 400C 450C 500C
Pasul
Cadmiu Cobalt Crom
Cupru Mangan Nichel
Plumb Stibiu Vanadiu
Abs
orba
nta
500C 550C 600C 650C 700C
Pasul
Abs
orba
nta
500C 550C 600C 650C 700C
Pasul
Abso
rban
ta
700C 800C 900C 1000C 1100C
Pasul
Abso
rban
ta
700C 800C 900C 1000C 1100C
Pasul
Abs
orba
nta
900C 1000C 1100C 1200C 1300C
Pasul
Abs
orba
nta
900C 1000C 1100C 1200C 1300C
Pasul
Abs
orba
nta
500C 600C 700C 800C 900C 1000C
Pasul
Abs
orba
nta
500C 600C 700C 800C 900C 1000C
Pasul
Abs
orba
nta
900C 1000C 1100C 1200C 1300C
Pasul
Abs
orba
nta
900C 1000C 1100C 1200C 1300C
Pasul
Abso
rban
ta
1200C 1300C 1400C 1500C 1600C
Pasul
1700C
Abso
rban
ta
1200C 1300C 1400C 1500C 1600C
Pasul
1700C
Abs
orba
nta
1 1 0 0 C 1 2 0 0 C 1 3 0 0 C 1 4 0 0 C
P a s u l
Abs
orba
nta
1 1 0 0 C 1 2 0 0 C 1 3 0 0 C 1 4 0 0 C
P a s u l
Abso
rban
ta
600 C 700 C 800 C 900 C 1000 C
Pasul
1100 C
Abso
rban
ta
600 C 700 C 800 C 900 C 1000 C
Pasul
1100 C
Figura 1 Stabilirea temperaturilor de piroliza pentru analiza probelor de emisii de la
instalatiile de ardere Pentru vitezele de ncalzire, timpii de uscare, piroliza si atomizare precum si pentru temperaturile de curatare ale tubului s-au utilizat valorile recomandate de soft-ul aparatului. 3.2.3 Autosampler Stabilirea parametrilor de lucru ai autosampler-ului implica:
- stabilirea volumului de proba utilizata; - numarul de determinari efectuate pentru fiecare etalon si proba; - metoda de dilutie a solutiilor etalon de lucru si a probelor prea concentrate; - unitatea de masura utilizata pentru exprimarea concentratiei; - momentul spalarii tubului capilar; - solutia etalon utilizata pentru verificarea finala; - declararea concentratiei solutiei etalon de lucru si a concentratiilor etalon
utilizate pentru tarsarea curbei de etalonare;
Alice Pop, Margareta Toader, Tatiana Paraschiv: Heavy metal determination for the emissions from the incineration plants by atomic absorbtion spectrometry with graphite furnace
390
Cadmiu Cobalt Crom
Cupru Mangan Nichel
Plumb
Stibiu Vanadiu
Abso
rban
ta
800C 900C 1000C 1100C 1200C 1300C
Pasul
1400C
Abso
rban
ta
800C 900C 1000C 1100C 1200C 1300C
Pasul
1400C
Abso
rban
ta
2 250C 23 00C 2 350C 2400C 24 50C
P asu l
Abso
rban
ta
2 250C 23 00C 2 350C 2400C 24 50C
P asu l
Abs
orba
nta
2000C 2100C 2200C 2300C 2400C
P asu l
Abs
orba
nta
2000C 2100C 2200C 2300C 2400C
P asu l
Abs
orba
nta
1800C 1900C 2000C 2100C 2200C
P asul
Abs
orba
nta
1800C 1900C 2000C 2100C 2200C
P asul
Abs
orba
nta
1500C 1600C 1700C 1800C 1900C
Pasul
Abs
orba
nta
1500C 1600C 1700C 1800C 1900C
Pasul
Abs
orba
nta
2100C 2150C 2200C 2250C 2300C
Pasul
2350C
Abs
orba
nta
2100C 2150C 2200C 2250C 2300C
Pasul
2350C
Abs
orba
nta
1550C 1600C 1650C 1700C 1750C
Pasul
1800C 1850C 1900C 1950C 2000C
Abs
orba
nta
1550C 1600C 1650C 1700C 1750C
Pasul
1800C 1850C 1900C 1950C 2000C
Abso
rban
ta
1000 C 1100C 1200 C 1300 C 1400 C
Pasul
1500 C 1600 C
Abso
rban
ta
1000 C 1100C 1200 C 1300 C 1400 C
Pasul
1500 C 1600 C
Figura 2 - Stabilirea temperaturilor de atomizare pentru analiza probelor de emisii de la instalatiile de ardere
In tabelul 4 sunt prezentate conditiile impuse pentru modul de functionare al autosamplerului.
Tabelul 4 Conditii pentru stabilirea modului de lucru al autosamplerului
Caracteristica Conditia Volumul de solutie 20 l Nr. de determinari pentru solutiile etalon 2 Nr. de determinari pentru o proba 2 Dilutie Dilueaza automat la concentratii prea mariMetoda de dilutie n tub Exprimarea concentratiei g/l Prepararea solutiilor de calibrare Dilutie n tub Spalarea ntre injectii Nr ciclurilor de spalare 2
Conditiile de lucru prezentate n tabelul 4 sunt valabile pentru toate elementele analizate. 3.3 Trasarea curbelor de etalonare Trasarea curbelor de etalonare se efectuaeza naintea fiecarei serii de determinari pentru fiecare element analizat, cu ajutorul solutiilor de calibrare si a solutiei de dilutie.
Alice Pop, Margareta Toader, Tatiana Paraschiv: Heavy metal determination for the emissions from the incineration plants by atomic absorbtion spectrometry with graphite furnace
391
Pentru fiecare element de analizat, solutiile de calibrare se obtin prin dilutia automata a solutiilor etalon de lucru, obtinute prin dilutii succesive ale solutiilor etalon monoelement, prezentate la punctul 2.1. Toate dilutiile se efectueaza cu solutie de acid azotic 0.5 %. In tabelul 5 sunt prezentate concentratia solutiilor etalon de lucru si concentratiile solutiilor de calibrare utilizate pentru trasarea curbei de etalonare.
Tabelul 5 - Concentratiile solutiilor utilizate la trasarea curbelor de etalonare pentru cele 9 elemente analizate
Elementul Conc. sol. etalon de lucru [g/l]
Concentratiile solutiilor de calibrare [g/l]
Cd 10 0 2 4 6 8 Co 50 0 5 10 15 20 Cr 50 0 5 10 15 20 Cu 50 0 5 10 15 20 Mn 30 0 3 7.5 12 24 Ni 50 0 5 10 15 20 Pb 50 0 5 10 15 20 Sb 50 0 5 10 15 20 V 100 0 10 30 50 100 Stabilirea concentratiei solutiei etalon de lucru si domeniul de concetratii utilizat pentru trasarea curbei de etalonare s-a facut pe baza concentratiei caracteristice a fiecarui element analizat. 4. REZULTATE SI DISCUTII S-au efectuat determinari pentru verificarea valabilitatii parametrilor stabiliti si a modului de lucru abordat pentru trecerea n solutie si analiza elementelor Cd, Co, Cr, Cu, Mn. Ni, Pb, Sb si V din emisiile de la instalatii de ardere din industria cimentului. In conditiile de lucru, stabilite la punctul 3.1, s-au mineralizat 2 filtre curate, care au constituit probe oarbe pentru cele 2 loturi de cte 3 probe de filtre din fibra de cuart, cu probe de pulbere din emisiile de la instalatiile de ardere. Pentru aceste probe s-au efectuat determinari ale continutului de Cd, Co, Cr, Cu, Mn. Ni, Pb, Sb si V, cu respectarea parametrilor de lucru prezentati la punctul 3.2. Rezultatele obtinute sunt prezentate n tabelul 6. Corelnd cantitatile de pna la 20 mg proba retinute pe filtre si domeniile de concentratii prezentate n tabelul 4, rezulta ca tehnica de analiza, prin spectrometrie de absorbtie atomica cu cuptor de grafit, este cea corespunzatoare pentru dozarea elementelor de analizat n probele de pulbere colectate de la instalatiile de ardere din industria cimentului. S-au efectuat de asemenea teste de repetabilitate pentru fiecare din elementele analizate. Acestea au constat n serii de cte 10 determinari pentru probe si solutiile etalon de lucru.
Alice Pop, Margareta Toader, Tatiana Paraschiv: Heavy metal determination for the emissions from the incineration plants by atomic absorbtion spectrometry with graphite furnace
392
Tabelul 6 Continutul de metale grele din pulberi de emisii de la instalatii de ardere, raportat la cantitatea de pulbere
Continutul de element [mg/kg] Elementul Proba 1 Proba 2 Proba 3 Proba 4 Proba 5 Proba 6
Cd 1.6 1.2 1.0 1.6 0.8 0.6 Co 1.1 1.3 1.5 1.1 1.1 1.3 Cr 1.1 25.3 16.4 0.8 23.3 14.3 Cu 215.4 75.1 150.0 22.6 130.3 10.2 Mn 52.7 22.1 89.6 63.4 37.2 24.4 Ni 14.8 2.0 18.9 25.6 20.4 5.4 Pb 66.2 15.1 35.1 54.4 60.9 36.3 Sb 0.6 0.8 0.1 0.6 0.1 0.5 V 28.3 18.4 24.8 12.7 12.4 7.0 Rezultatele obtinute au evidentiat o buna stabilitate a rezultatelor n cazul elementelor analizate si obtinerea unor coeficienti de variatie care se ncadreaza n limita de 10 % prevazuta de standardele de metoda [3, 4] utilizate pentru aceste determinari. 5. CONCLUZII Decelarea elementelor de analizat la concentratiile maxime admise de Directiva 76/2000, transpusa in legislatia nationala prin Hotararea de Guvern nr. 128/2002, modificata si completata prin HG 268/2005, implica utilizarea unor metode de analiza instrumentala de inalta rezolutie. S-au stabilit parametrii de lucru cu mineralizatorul cu microunde prin stabilirea conditiilor de temperatura, putere si durata de ncalzire pentru cei 3 pasi din fiecare etapa de mineralizare a filtrelor din fibra de cuart. S-au stabilit parametrii de lucru cu spectrometrul de absorbtie atomica cu cuptor de grafit si generator de hidruri, prin stabilirea conditiilor de lucru ale spectrometrului, ale cuptorului de grafit si ale autosamplerului si s-au efectuat determinari ale continutului de elemente din probe de emisii prelevate de la instalatiile de ardere din industria cimentului si determinari de repetabilitate care au confirmat ca parametri de lucru sunt corespunzatori cu elementele care trebuie determinate si domeniile de concentratie uzuale pentru. BIBLIOGRAFIE [1] R. D. Beaty, J. D. Kerber, Concepts, instruments and techniques in Atomic Absorbtion, 2004 [2] B. Welz, H. Becker-Ross, S. Florek, U. Hitemann, High-Resolution Continuum Source AAS, WILEY-VCH Verlag GmBH & Co., 2006 [3] SR EN 14385:2004 Emisii de la surse fixe. Determinarea emisiei totale de As, Cd, Co, Cr, Cu, Mn. Ni, Pb, Sb, Tl si V [4] SR EN ISO 15586:2004 Calitatea apei. Determinarea elementelor n urme prin spectrometrie de absorbtie atomica cu cuptor de grafit
52_Jitaru_RDE_P_P0072_1.doc Environmental electrochemical technology, as example the electrochemical way for the reduction of some nitro derivatives. Monitoring of the environment, as example copper, lead and nitro derivatives monitoring in the wastewaters after the electrochemical reduction of the nitro derivatives on copper and lead electrodes. 3.4. Education and training to the development of environmental electrochemistry REFERENCES
53_Modrogan_RDE_P_P0087.doc1 INTRODUCTION 2 EXPERIMENTAL 2.1 The adsorbent mass preparation 2.2 The study of the adsorption 3 RESULTS AND DISCUSSIONS 3.1 Effect of pH 3.2 Adsorption dynamics
4 CONCLUSIONS REFERENCES
54_Maria(2)_RM_P_P0129 .doc 1 INTRODUCTION REFERENCES
55_Pop_RDE_P_P0118_f.docAlice Pop, Margareta Toader, Tatiana Paraschiv Abstact 1. INTRODUCERE 2. MATERIALE SI APARATURA 3 MOD DE LUCRU EtapaPutere [W]Cd
BIBLIOGRAFIE
56_Seletchi_RDE_P_P0083_refacuta.doc1 INTRODUCTION 2 ADDITIVE DOSE METHOD 3 ISOTHERMAL ANNEALING STUDIES 4 EXPERIMENTAL 4.1 Sample preparation 4.2 EPR measurements 5 APPLICATIONS OF ELECTRON PARAMAGNETIC RESONANCE METHOD 6 CONCLUDING REMARKS 7 REFERENCES
57_Seletchi_RDE_P_P0084.doc1 INTRODUCTION 2 EXPERIMENTAL 2.1 CT Instrumentation 2.2 Computer Software 3 RESULTS AND DISCUSSIONS 3.1 Image processing 3.2 Data analyze and interpretations
4 CONCLUSIONS REFERENCES
58_Spataru_RDE_P_P0078.doc59_Toader_RDE_P_P0116_refacuta.docMargareta Toader, Simona Ungureanu, Alice Pop
60_Zelic_RDE_P_P0058.docJelica Zeli, Leticija Ugrina and Draan Jozi
61_Coman_RDE_P_P0069.docREFERENCES