Upload
nenadpesic
View
221
Download
0
Embed Size (px)
Citation preview
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
1/46
HADRON THERAPY FOR CANCER USING
HEAVY IONS!"
Khalid Mohammed S Yamani
A dissertation submitted to the Department of Physics,
University of Surrey, in partial fulfilment of the degree of
Master of Radiation & Environment Protection
$%&'()*%+) ,- ./"0120
3'245)" ,- 65%2)(,+120 7 ./"012'5 821%+2%0
9+1:%(01)" ,- 84((%"
September 2009
;
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
2/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @
"#$%& '( )'*"&*"+
DE8FGDHF I
D%=14* PSR E('CC .%'T UIR G'+C% D=V40)*%+) WOR G'=1')1,+ 31%5= WPR 31%5= X,*,C%+%1)" ,+ D F('+0:%(0% 8%2)1,+ WUR X,*,C%+%1)" (%C1,+ 1+ ' &5'+ ,()/,C,+'5 ), )/% !%'* =1(%2)1,+ BWR 31%5= /,*,C%+%1)" '5,+C )/% !%'* 'Y10
BBR Field symmetry QA10. Lateral penumbra QAQQRDistal dose fall-off QAH/'&)%( QZ [JFGK$9HF[KJ K3 GD$[KFX6GD.?
QRQ H'+2%( '+= ('=1')1,+ )/%('&" QQ
QR@ X10),(12'5 =%:%5,&*%+) ,- ('=1')1,+ )/%('&" QQ
H/'&)%( @Z PHHYSICAL AND RADIOBIOLOGICAL PROPERTIES OF
HADRONS QI
2.1 Physical characteristic of the beamQO
2.1.1 Depth-dose distribution and Bragg-peakQO
2.1.2 Defined range:QP
2.1.3 Lateral scattering:QP
2.1.4 Non-elastic nuclear reaction:QU
2.1.5 Positron emission tomography technique (PET)QU
2.2 Biological effectQW
2.2.1 The Relative Biological Effectiveness (REB) and Linear Energy
Transfer (LET):QB
2.2.2 Oxygen Enhancement Ratio (OER)@A
H/'&)%( SZHadrons Therapy equipments @@
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
3/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% S
3.1 Conventional Accelerator Radiotherapy@@
3.2 Accelerator for heavy ions and proton @S
3.2.1 Fixed Field Alternation Gradient accelerator (FFAG)@O
3.2.2. The dielectric-wall accelerator (DWA)@O
3.4 Beam delivery systems and techniques @O
3.4.1 Passive beam shaping@P
3.4.2 Active beam shaping:@U
3.5 Gantry design for proton and carbon hadrontherapy facilities @W
3.6 Cost-effectiveness of hadron therapy in cancerS@
H/'&)%( IZTREATMENT PLANNING SS
4.1 Imaging protocolSI
4.2Organs contouringSI
4.3Dose calculation algorithmsSI
4.4Choice of the portsSI
4.5Positioning and treatmentsSI
H/'&)%( IZCOCULUSION SP
G%-%(%+2%0 SU
Appendix A SB
Appendix B IA
Appendix C I@
Appendix D IO
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
4/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% I
#,-./01.
The application of hadron accelerators (proton and light ions) in cancer treatment is
relatively new. They are far from standardized, but the use of hadron therapy as an alternative
to conventional radiation has led to important developments and refinements in conventional
treatment technique. Hadron beams allow highly conformal treatment of deep-seated tumours
with great accuracy, while delivering minimal doses to normal surrounding tissue.
This overview of hadron therapy covers the following topics: an introduction of
radiotherapy represents the relationship between cancer and radiotherapy, and the historical
development of radiation therapy, physical and radiobiological properties of hadrons, hadrons
therapy equipments with the cost of hadron facilities.
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
5/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% O
#1234567897:73.
I would like to thank Pro. Paddy Regan for his supervision, and guidance throughout this
project, I did appreciate his help and advice.
I would like to thank my wife for support I was provided with during the completion of
dissertation masters.
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
6/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% P
Glossary*:
;
F/% *%=12'5 '+= '22%5%('),( &/"01210)0 ), *%'+ \41)% =1--%(%+) )/1+C0 40% )/%
)%(* !%'* 1+)%+01)"R F/% '224(')% =%-1+1)1,+0 ,- )%(*0 (%5')1+C )/% +4*!%( ,-
'22%5%(')%= &'()125%0 '(% '0 -,55,]0Z
J4*!%( ,- &'()125%0^ J
354%+2%^ !Z &'()125%0_ 4+1) '(%'^ !" #$%#&
354Y^ 3 Z &'()125%0_ 4+1) )1*%^ ' " #$%#(
[+)%+01)") *Z %+%(C"_ 4+1)% '(%'_ 4+1) )1*%
* " +#,%#(#& " +#!%#( " +#'%#& -./.0
L/%(% + 10 )/% %+%(C" ,- &'()125%0
3,( ,4( &4(&,0% ,- =%02(1!1+C *,+,`%+%(C%)12 &'()125%0^ 1) 10 041)'!5% ), =%-1+%Z
[+)%+01)"^ *Z &'()125%0a 4+1) '(%'_ 4+1)% )1*%
* " #,%#(#& " #!%#( " #'%#& -./,0
X,]%:%(^ '22%5%('),( &/"01210) 40% )/% )%(* !%'* 1+)%+01)" ), *%'+ )/% +4*!%(
,- &'()125%0 &%( 4+1)% )1*%^ 1R%R^ '0 )/% -54Y =%-1+%= '!,:%R
[+)%+01)"^ *Z &'()125%0a 4+1)% )1*%
* " #$%#( " ' -./10
'+= 23 /* " 23#$%#( " 43&5 678839( -./:0
F/% !%'* 1+)%+01)" ]1)/ )/10 =%-1+1)1,+ =,%0 +,) :'(" ]/%+ )/% !%'* 2(,00`
0%2)1,+ 2/'+C%0R D+= )/% !%'* 1+)%+01)" *,=45')1,+ *%'+0 )/% !%'* -54Y
*,=45')1,+R
?
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
7/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% U
(=9B/7 ;C $70: /039 =3 :78=B:
[- ' =%&)/ =,0% 24(:% 10 2,+01=%(%=^ )/% ('+C% (%5')%0 ), )/% =10)'+2% !%)]%%+ )/%
%+)('+2% 04(-'2%0 ,- )/% !%'* '+= )/% =10)'5 &,1+) ,- WAb =,0% c31CRQdR [+ )/% 251+12'5
'&&512')1,+^ )/% ('+C% 10 %Y&(%00%= 1+ C_2*@^ 1+ ,(=%( ), !% 1+=%&%+=%+) ,- 0&%21-12
2'0%0 ]/%(% 1+/,*,C%+%1)1%0 2'+ !% &(%0%+)R [- 1) 10 &(%0%+) ,+ )/% !%'* )('2T c-,(
%Y'*&5% 54+C ,( !,+%d^ )/% ('+C% ]155 !% 0,*%]/') =1--%(%+) -(,* )/% :'54% (%5')1:%
), ]')%(R
D
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
8/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% W
L
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
9/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% B
5%'0) O 2* C(%')%( )/'+ )/% =%&)/ ,- )/% E('CC &%'TR [+ '==1)1,+^ )/% &/'+),*
04(-'2% 0/,45= !% ') )/% +,(*'5 )(%')*%+) =10)'+2% -,( -1Y%= !%'*R
R /79=43 =3 0 I603 4/.H494306 .4 .H7 ,70: 8=/71.=43
F/% /,*,C%+%1)" (%C1,+ 10 =%-1+%= !" )/% 51+%0 2,++%2)1+C )/% &,1+)0 ,+ )/%
&(1+21&5% 'Y%0 '+= ,+ )/% =1'C,+'50 ,- ' 0\4'(% -1%5=R F/% =1*%+01,+0 =*'+= =%'(%
=%-1+1+C )/% /,*,C%+%1)" (%C1,+ '0 0/,]+ 1+ 31CRIR
(=9B/7 LC Homogeneity region in a plan orthogonal to the beam direction
S 06439 .H7 ,70: 0T=-
It is necessary to define the homogeneity along the beam direction in correspondence
to the SOBP region for example with proton beams. The region of homogeneityis defined
as the region between two distal fall-off widths (below) inside the 50% dose point distally
and one distal fall-off width inside 90% dose point proximately. It is assumed that 100%
corresponds to the point of maximum dose in the SOBP as shown in Fig. 2 .
F/% /,*,C%+%1)" 10 =%-1+%= '0 )/% (')1,Z
;( "-
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
10/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QA
U
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
11/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QQ
)H0I.7/ ;C W3./48B1.=43 4G@08=4.H7/0I>C
;
Cancer can be defined as the uncontrolled growth and proliferation of groups of cells.
It is reported that is affecting the industrial countries in Europe, 26% of deaths are in relation
with cancer. The average survival of a 5 years-cancer is 40%. Many factors are involved in
cancer development: food, medicines, genetics, infections, and professional. The
radiotherapy, chemotherapy and surgery are the main therapist against cancer.
Figure 1.1:Present situations of the different treatments against solid tumors by EC (Haberer,
2002)
Figure1.1 present situation in cancer treatment, as we can see that of 45% of all patient
are cured, among various cancer treatments; surgical removal of the tumor tissue,
radiotherapy, and chemotherapy. Surgery and radiotherapy are today of crucial important and
successful in 22% and 12% of cases respectively. Moreover, radiotherapy is involved in
almost half of the curative treatment of the localized cancer when combined the 12% for
another 6% of the cases. For 18% of all cancer patients the local control of the primary
tumour without metastasis fails. And these patients could be cured successfully if improved
treatment modalities as the application of proton and ion beams in radiation therapy (haberer,
2002)
;
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
12/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% Q@
ions, alpha particles, mesons, etc) to target cancerous tissue by; external beam therapy, sealed
source therapy (Brachytherapy) and unsealed source therapy. External beam therapy is the
most common form of radiation therapy. It is performed with electron or photon beams; these
are produced by a linear accelerator
Charged particles beams have been used for radiotherapy such as protons and heavier
ions. E. Rutherford who is discovered the protons atthe end of the second decade of the 19th
Century. In 1955, the Lawrence Berkeley Laboratory (LBL) was the first organisation in the
world that is performed the first clinical application for a proton beams in California. During
this period 1919-1955, there are two significant on proton therapy occurred; in 1930, E. O.
Lawrence was built the first cyclotron with high energy for application of cancer treatment,
and in 1946, Robert Wilson was used the proton beams for deep-seated tumour treatment.Moreover, Wilson referred to the physical aspects of the dose distribution of the proton
beams, such as an advantage for proton compared to X-ray therapies. (Subramania,
1996;Aferd.2009)
There is an on going level of interest in using charge particles as beams of hadrons in
radiation therapy such as protons, neutrons and heavy ions. Particles therapy act to damage
the DNA of cancerous tissues that they cannot multiply and grow, and minimizing damage to
the surrounding healthy cells. At the present, there are now about 25 developed facilities
around the world using protons and light ions (carbon) for cancer treatment; they have been
treated more than 55,000 patients with proton therapy and 5000 patients with carbon ions
(Aferd.2009).
According to Dale et al.project that the Royal College of Radiologists (2003) in the
European study was found 11% tumours that are treated by chemotherapy, 40% by
radiotherapy and 49% are cured by surgery. As a result, the 40% is very respectable and
provides a strong indication of why the research is continuing and investment in this modality
is justified (Dale et al, 2009).
In general proton therapy and light ions therapy is more expensive than photon therapy.
The ratio of the cost was found about 2.4 relating to Goitein et al. project in 2002, and they
expected that the cost will be reduced to 1.7 or less in the next few years. Also Alfred pointed
out on his project that the refund rate is currently somehow sufficient to develop hadrons
facilities with the progress of technologies and a reasonable incoming margin. In the future,
the cost of these modalities will decrease more and will be spread more than before among
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
13/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QS
more patient.
.Fig 1.2:Development of radiation therapy linear accelerator (a)-4MeV AEI linac with an in line
treatment head, mfd. in 1955;(b)- the SL75 Philips linac with a horizontal accelerating structure
and 90 bending magnet, suitable for full rotation therapy, mfd. in 1964;(c)- the Dynaray 4 linac
with 270 achromatic magnet, mfd. in 1971; (d)- a present-day system with an in- line treatment
head, without beam binding, e.g. 4MeV LMR-4 Toshiba Linac with 35 cm along accelerating
structure.
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
14/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QI
)H0I.7/ ?C Physical and Radiobiological Properties of Hadrons
Hadrons are compound subatomic particles made of quark and antiquarks, bound
either in doublets or in triplets by strong force. The hadrons that are employed for
radiotherapeutic purpose are protons, neutrons and light ions (such as carbon, helium, neon
and oxygen). Among the use hadrons, protons show approximately the same LET as
conventional photons and electron beams when they going into treated body, but have an
increasing LET at the end of their track in tissues. However, neutrons and light ions are high-
LET particles (table 2.1)
Table 2.1: LET (L ) of ionizing particles and radiations of radiations of interest inradiotherapy and radiobiology, after (Wioletta & Waldemer, 2001)
The ability to change the radiative properties is due to the fact that the biological
affects are correlated with the distribution of energy deposited. This distribution is influenced
by the atomic number Z (for carbon=8, for proton=1) and the energy of the impinging
hadrons. The basic physical and radiobiological properties of photons and ions are compared
in table 2.2
Hadron therapy is using of charge particle beams of protons or carbon ions for
treatment of cancer. Ion beam therapies have a major of important advantages over
conventional radiation therapy for deep-seated cancerous cells, also for tumors closed to a
critical tissue of organ. The development of the physical characteristic of particles beamsfrom proton to carbon and of ion accelerator, they hold promise due to its capability to deliver
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
15/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QO
high dose to target volume. The heavy charge particles have rather reduced depth-dose
outline, but enhanced responses biological effectiveness (Wioletta & Waldemer, 2001)
Table 2.2: Differences between photons and ions, after (Wioletta & Waldemer, 2001)
2.1 Physical characteristic of the beam
The physical properties of charge particles interaction with critical organs introduce an
eminent benefit of hadron therapy compared to conventional radiation therapy, it allows a
delivery adequate high dose to stop a tumors growth, destroy it, and prevent damage to the
normal cell that is surrounding tumors as much as possible. According to Weyrather and
Debus project that they have reviewed the physical properties of particle beams interaction
with tissue, and summarized it as the following (2003):
2.1.1 Depth-dose distribution and Bragg-peak
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
16/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QP
The dose decreases exponentially for penetrating depths beyond the build-up depth for
photon conventional radiation therapy. At the same time as a charge particle beam loses just a
small quantity of its energy at the access until the maximum penetration depth where the
remaining energy lost over a short distance consequences in a high dose presented as Bragg-
peak (Fig.1.3)
Figure 2.1: Comparison of Depth-dose curves for photons, proton, and carbon beams
(Weyyrather&Debus, 2003|)
As we know that the target volume of tumours is much larger than the unmodified
Bragg peak width. In clinical applications a relatively consistent dose distribution be able to
deliver to tumours by spreading out the Bragg peak (SOBP) through a Range Modulated
Wheel (RMW) or through modulating the energy from pulse to pulse as can achieved with a
synchrotron.
2.1.2 Defined range
The particles beam prevents damage to healthy tissue beyond the tumour by the
defined range, and after the Bragg peak drops almost to zero. It is determined the density of
the target tissue and the energy of particles. A rang straggling in Bragg peak due to multiple
scattering of the particles which depend on the atomic number of the particles, and this
straggling for protons is highest and it decreases quadratically with atomic number of the
projectile ion.
2.1.3 Lateral scattering:
Heavy ions such as carbon have the advantage of the small scattering angle as they
penetrate the target. The expansion of carbon beam is less than 1mm at 10 cm depth, while for
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
17/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QU
proton it goes more than 2mm. this advantage give more accuracy for irradiating of deep-
seated tumour close to a critical tissue (Fig.3.2)
Figure 2.2: Comparison of the lateral scattering of photon,proton and carbon beams asfunction of the penetration depth(Weyyrather&Debus, 2003|)
2.1.4 Non-elastic nuclear reaction:
The interactions depend on the type of the charge particles and the cross-sectional of
the target volumes, as we know the particle beams undergoes nuclear interaction with target
materials. In the proton case, the non-elastic nuclear interaction results in attenuation of the
primary beam and emission of secondary proton and other particle with relatively lower
energy. However, for carbon, a small amount of the primary beam transferred to lighter
nuclear fragments causing a long dose tail beyond the Bragg peak, but it continues within the
acceptable limits.
2.1.5 Positron emission tomography technique (PET)
The PET technique allows 3D retrospective analysis for the dose distribution inside
the body of patients. There is a small percentage of the passage particle beam that going to
body tissue can be transferred to nuclear fragments with the same atomic number of the
projectile due to the only one or two neutrons are lost. As unstable carbon isotopes (10C, 11C,
and15
O) are produced. The decay of these isotopes with 19 second, 20 minutes, and 2 minutes
respectively of its half-life are under emission of a positron and neutrino. In carbon therapy, a
10C and 11C ranges differ only slightly from the range of stable 12C and the stopping point
can be screened by measuring the coincident emission of tow annihilation quanta of the
positron decay, by using PET-camera outside the body without extra dose to the patient.
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
18/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QW
However, the range of positron emitters is few mm in proton therapy shorter than the range of
primary proton due to the threshold of energy of nuclear reaction and fragmentation. On the
other hand, the positron emitter stopping point is somehow correlated with the Bragg peak.
This is an advantage for proton therapy to using PET; also the activity of positron emitters is
higher then for carbon approximately three times at the same depth. This is help to increase
the sensitivity of PET imaging in proton therapy over carbon therapy.
2.2 Biological effect:
The biological effects of ionizing radiation result generally from ionization and
excitation of atoms and molecules in the matter via radiation. These primary processes initiate
a complex chain of events in a living cell, which lead to chemical changes in some
biomolecules (DNA that is the most important target) and to bio functional changes such as
cell death, transformation and mutation (.'C'+%))1 3( &E),??,0.
Figure 2.3: Comparison of tracks of sparsely and densely ionizing radiation with the relevant
biological targets (Amaldi & Siliari,1996)2.2.1 The Relative Biological Effectiveness (REB) and Linear Energy Transfer (LET):
The different biological effect depending on the energy atomic number of the ion and
can occur from the same dose in high-LET as a hadron beams, while the low-LET radiation
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
19/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% QB
such as "-rays, X-ray and electrons that is used in conventional therapy, exhibit the same
biological effect when applied the same dose.
The Relative Biological Effectiveness (REB) can be calculated from measured data as
we can show that in Fig 3.3. For example, the ratio between the dose of X-ray and the ion
particle dose necessary to produce the same biological effect.(Weyrather&Debus, 2003|)
Figure 2.4: Definition of the relative biological electiveness, RBE, illustrated for cell
survival curves with carbon ions of differentenergy(Weyrather&Debus, 2003|)
As we can see at Fig.2.4, the shoulder in heavy ions (carbon ions) dose-effect curve is
much smaller, while in the proton represents repair capacity of the cell at low doses. This
represents a high ionization density is reduced the repair capacity of cell at the same dose.
And the most interesting conclusion is that protons in plateau are as the lethal as "-rays, while
protons in the Bragg peak show a higher effectiveness.
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
20/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @A
Table 2.3: RBE for inactivation of cultured mammalian cells irradiated with proton
beams of high initial energies (Amaldi & Siliari, 1996)
The RBE in hadron therapy is dependent on several biological and physical factors.
As the Weyratherand Kraftmentioned to Experiments revealed a dramatic dependence of
RBE on the dose level, atomic number and the repair capacity of the biological system. It also
showed that carbon ion are the optimal choice for high-LET therapy; later scattering and its
range are small producing an optimum precision in dose delivery system(2004). The
efficiency of ions to induce biological damage changes along their path. Local ionisation
density increases at the end of the range, and causing difficulty repairing for high DNA
damage and RBE increases up to maximum, while the high energies and light ions at the
entrance channel are easily repaired for the induced damage, and RBE is close to those
induced by low-LET radiation (Weyrather&Debus, 2003). At Fig. 3.3 that clears the RBE is
inversely dependent on the dose, its large for low doses and small for high dose. And high
RBE is related with lower dependence on Oxygen to have an effect on the cancerous tissues.
Therefore, heavier ions and carbon can effective to slow or stop growth of radio-resistance
and poorly oxygenated tumours (Dale et al, 2009).
2.2.2 Oxygen Enhancement Ratio (OER)
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
21/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @Q
The Oxygen enhancement ratio is an additional indicator for realising the therapeutic
effects giving by a certain type of radiation. The OER is the ratio of doses with or without
oxygen to produce the same biological effect. Therefore, the OER has a strong dependence on
the LET; it decreases as LET exceeds approximately 10 and approaches unity at very high
LET, as a result, the better oxygen enhancement ratio when its lower value.
Figure 3.1 is shown the comparison of the relative biological effectiveness (RBE) and
oxygen enhancement ratio (OER) values for different radiation particles and ions. It is clear
that to obtain the optimum curative effect; the RBE value should be high while the OER
should be low.
(=9B/7 D
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
22/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @@
)H0I.7/ DCHadrons Therapy equipments
Nowadays, the main source of radiation therapy carried with types of accelerators for
electron, and X-ray radiotherapy such as Linear accelerator, betatrons and microtrons. All
these accelerate electrons to high velocities that have varying energy, also as we mention
before the electrons are accelerated, the tissue penetration depth, the tissue at that greatest
dose of radiation is absorbed. (Bertil J, Alan M, 2006).
Hadron therapy is the one of the therapeutic technique that used to treatment cancer, and
its varied application of treatment that uses the widespread range of particle accelerators from
the use of small electron linear accelerators for radiotherapy, to the use of proton cyclotron for
radionuclide production is 10-20 MeV, and to the application of 50-60 MeV proton
accelerator for proton therapy or neutron treatment of eye melanomas to the hadron
accelerators that it is installed in radiation treatment centres for cancer therapy with ion beam
or proton (Amaldi U, et all, 1996).
3.1 Conventional Accelerator Radiotherapy
(=9B/7 D
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
23/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @S
Hadron accelerators have two types (cyclotron and synchrotron) to produce a charge
particle beam with sufficient energy that passing trough the beam transport system to reach
the target volume with enough intensity, due to achieve a typical medical treatment.
The cyclotrons are made up of the particle beam and a large magnet to moves starting
from machine center with a long spiral path, and it is working at fixed field and fixed
frequency of energy due to continuous beam, also operating reliability and simplicity. In
synchrotrons the radio frequency and magnetic fields are varying that provide a sufficiently
high dose rate for treatment compared to cyclotrons. However, cyclotrons have a capacity to
produce sufficient beam with high intensity (flux). (Chu et al, 1993).
In recent years, the building accelerators have increased throughout the world for
medical treatment, and the development and improvement is providing a new generation of
accelerators, which are more compact, efficient and inexpensive.
3.2 Accelerator for heavy ions and proton
Theoretically, the hadron beams that used in cancer (protons, light ions such as carbon)
has developed considerably since the middle of the last century and the hadron therapy
presents the potential of both physical (protons and ions) and biological (neutrons and ions)
advantages in the therapy of localised tumour (Lodge M, et all, 2007). The use of protons in
the treatment cancer depend on the superior dose distribution which can be achieved with
respect to electrons and photons, as a result, it is due to the low lateral scattering undergone
by protons, their will-defined range and the increasing dose deposition (ionization) with
increasing penetration in cell tissue, that produces theBragg peak (Fig. 2.1)
The depth at which the Bragg peak occurs depends on the initial energy of the protons
and its width on the energy spread of the beam. Fig. 4.1 shows that, an even enhanced dose
distribution and the additional advantages of an increasing the relative biological
effectiveness (RBE) at the end of range in tissues for protons. The therapy beam which used
in treatment cancer must have enough range to reach the deepest point of volume of the body,
in order to irradiate tumors that the minimum energy required for carbon ions are 400 MeV/u
and for protons are 200 MeV. All the clinical procedures can be requests satisfied among light
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
24/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @I
ions range energies between 120-400 MeV/u and 109-1010 ions/s of beam intensities, which
depends on the ion type. For protons corresponding figure are 60-250 MeV and 5 #1010-1011
(Fig. 4.3). (Amaldi, et al.1996).
These requirements stand for hadron accelerator considered for cancer treatment that the
basic specification must be met by ion radiation or proton therapy facility to represent an
Fig. 3.2: Depth-dose curves for photons, electrons, neutrons and 200MeV
protons.
Fig. 3.3: Range-energy curves in tissues for light ions and photons
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
25/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @O
effective and reliable instrument against cancer. The performance standards of the accelerator
and of delivery systems and beam transport must satisfy a number of clinical requirements.
These performance standards are expressed in terms of beam intensity, energy variability and
range, distal dose fall-off, time structure of extracted beam, lateral penumbra, beam abort time
and raster scanning specifications (Goitein M, 2007).
According to Alfred project that is summarized the new accelerator technologies for
proton therapy designed such as:
3.2.1Fixed Field Alternation Gradient accelerator (FFAG):
In Japan, the Energy Accelerator Research Organization (KEK) was built the first
proton FFAG accelerator in 1999. The energy range of several FFAG accelerators which is
manufactured tell this point are 150-250 MeV, and it is join the fixed magnetic field of
cyclotrons and the energy variability of the synchrotrons. In addition, the easy operation and
compact size of the FFAG accelerators make them more attractive for hadron therapy. In the
UK, the British Accelerator Science and Radiation Oncology Consortium (BASROC) aim to
build non-scaling FFAG accelerators to be used for hadron therapy.
3.2.2
The dielectric-wall accelerator (DWA)
In USA, the Lawrence Livermore National laboratory (LLNL) has developed the
dielectric-wall accelerator. This technology uses to accelerate the proton approximately
100MV/m of high electric gradients. It will be employed the development of a high
accelerator gradient LINAC for proton therapy.
3.3 Beam transport system
The extracted beam is transported from the accelerator to treatment radiation therapy
room by beam transport system. A beam lines consist of quadruple and dipole magnets,
vacuum chamber and diagnostic instrumentation. An efficient and stable transport of the
beam from the accelerator to the treatment room is needed to achieve a reliable therapy and
reproducible dosimetry. The stability of the centroid of the beam position is adjusted by
focusing magnets to control the beam position and profile, which is called tuning beam line. It
means adjusting the beam optics to transport the given beam to the required location with the
desired physical parameters (Chu et al, 1993).
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
26/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @P
3.4 Beam delivery systems and techniques
This system is located at the treatment room; it lies between the beam end line at the
vacuum window and the patient. The beam delivery system has a technical components and a
number of devices to adapt the beam such as; modulating devices and range changing, also
laterally spreading and shaping devices. It also has a number of monitors to manage the
prescribed 3D dose distribution inside the target volume (Wioletta & Waldemer, 2001). In
practice, the role of the beam delivery system technique are divided into two principle
according to the method of the beam spreading as following:
3.4.1Passive beam shaping
Passive beam shaping is the most commonly used in proton and heavy ion therapy. It
has three steps; a modulator designed modifies the monoenergetic beam from the accelerator
to provide a predefined depth dose modulation. Numerous modulators are available to cover
different depth modulation in tissue. In order additional range shifts are needed to adjust the
modulated Brag peak to the desired radiological depth. Finally, a clinically useful field widthis produced using a double scattering system or a magnetic beam wobbler. It has three major
disadvantages (IAEA, 2007):
The depth dose can only tailored to the distal end of the target due to the factwhich the compensator shifts the SOBP towards the entrance region. A
considerable amount of the high dose region (and high LET region) is located
in the normal tissue in front of the target volume, especially at the lateral field
borders
The amount of material in the beam line is considerable, leading to an increasein nuclear fragments produced by nuclear interactions with the material of the
beam modifiers. These nuclear fragments have lower energies and lead to a
higher LET and thus an increased biological effective dose of the beam already
in the entrance region
The large number of patient specific beam modifiers, which have to bemanufactured (a compensator and a collimator for each treatment field) and the
necessity to produce a number of modulators that may have to be exchanged
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
27/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @U
for different patients.
At HIMAC using the passive depth dose system, the depth dose is fixed by the
modulator hardware, which is designed to achieve a prescribed homogenous biological dose
effective dose for single field. The design of the modulators reflects the fixed dependence of
the RBE with depth for dose level.
3.4.2Active beam shaping:
This system takes advantage of the electric charge of ions to produce a firmly focused
pencil beam that is deflected laterally by a magnetic field instated the foils in the passive
scattering technique to allow a scanning of the beam over the treatment field. In clinical
applications the target volume is dissected into layers of equal ion energy, each layer is
covered by a grid pixels and the beam is scanned in a row pattern over these pixels (Alferd,
2009). Moreover, the energy from a synchrotron can be switched from pulse to pulse to adapt
the range of the particles in tissue. This technique is able to scan the target volume in three-
dimensional and the dose distribution can be tailored to any irregular shape without any
patient specific devices (such as collimators or compensators) or passive absorbers. Thus, the
high dose region can be conformed to the proximal end of the target volume and the integral
dose as well as the volume receiving high LET radiation is summarized (IAEA, 2007).
Fig. 3.4: Principal of passive beam shaping
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
28/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @W
3.5 Gantry design for proton and carbon hadrontherapy facilities
The tumor is treated several times with beams coming from different direction, so the
device that is used to allow directing beam from several directions onto the patient who is
lying on the treatment table and to reduce the dose to the surrounding health tissues is called
gantry. However, the necessity of choosing a gantry or not is influenced by many factors,
mainly by the design of each treatment facility and the kind of tumor to be treated. Building a
gantry for heavy charged particles is a technological challenge. It is lies in the production of
large and heavy gantries still achieving the same position stability of the ion beam. The first
hospital-based gantries had a diameter of 12 meters. The innovation PSI project compact
spot- scanning gantry (Pedroniet al, 1995) is a very good optimization of the size (diameter of
4m) predominantly because of the alternative chosen to move patient during the treatment (Fig
4.5).
The hydrotherapy aims for gantries-despite the higher magnetic rigidity. Th!rigidity i"
d!fin!d #" th!$r%duct %f th! b!nding r#diu" #nd th! r!quir!d m#gn!tic fi!ld "tr!ngthR In
proton therapy several facilities are already running gantries in routine process with a
magnetic rigidity of about 2.5 Tm. However, this situation is not yet reached in carbon
therapy due to higher magnetic rigidity of about 6.5 Tm.
Fig. 3.5: The exocentric PSI Gantry (PSI Web pages)
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
29/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% @B
The area of the treatment for proton gantry (Fig. 4.6), the gantry typically directs the beam from on
In radiotherapy and are called isocentric gantries. Therefore, at PSI used the exocentric
gantry for proton Gantry 1, which the patient table has a different position for each gantry
angle. A set of the used gantry angles is exemplified in Figure 3.7
Fig. 3.6: Layout of proton gantry delivered by Mitsubishi; after (Weinrich, 2006)
Fig. 3.7: Calcification of the different types of gantries; after (Kraft, 2000).
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
30/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SA
Then the advantage is that it can be built in a more compact manner; and it is shown in figure
3.8
In tables 4.1 and 4.2 an overview the important parameters of some proton and carbon
gantries in Europe and Japan.
Table 4.1: Proton Gantries in Europe & Japan;after (Weinrich, 2006)
Town No. Status Type Energy Length Radius Dipoles Quads
Germany
Munich 4 Validation Isocentric 250 MeV 10.1 m 5 m 2 7
Switzerland
Villingen Gantry 1 Operation Excentric 230 MeV 10.2 m 1.4 m 3 7
Switzerland
Villingen Gantry 2 Assembly Isocentric 230 MeV 11.6 m 3.2 m 3 7
Japan
Fig. 3.8: the exocentric Gantry 1 at PSI; after (Weinrich, 2006)
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
31/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SQ
Hyogo 2 Operation Isocentric 230 MeV 9.5 m 4.8 m 2 7
Chiba 2 Operation Isocentric 235 MeV 10.7 m 5 m 2 9
Tsukuba 2 Operation Isocentric 250 MeV 9 m 5 m 3 6
Shizuoka 2 Operation Isocentric 235 MeV 9 m 4.8 m 3 4
In addition, the treatments with carbon ions are currently delivered in three facilities;
HIMAC, HIBM in Japan and GSI Germany, table 4.2 shown the properties of some carbon
gantries.
Table 4.2: Carbon Gantries in Germany & Japan;after (Weinrich, 2006)
Town No. Status Type Energy Length Radius Dipoles Quads
Germany
Heidelberg 1 Assembly Isocentric430
MeV/u19 m 5.6 m 3 8
Japan
Chiba 1 Design Isocentric400
MeV/u16.9 m 7.1 m 3 7
T%#chi!v!#r!#"%n#bl!b!nding r#diu", much high!r fi!ld "tr!ngth"#nd thu"l#rg!r #nd
h!#vi!r m#gn!t"#r!n!c!""#ry f%r i%n". Whil!th!w!ight %f #$r%t%n g#ntry i"#lr!#dy #r%und
100 t%n"(#t length %f 10m), #n i"%c!ntric g#ntry f%r c#rb%n i%n"i"!x$!ct!d t%h#v!w!ight %f
#b%ut 600 t%n"#t length about %f 18 m.
Th!!n%rm%u""iz!#nd w!ight %f "uch #g#ntry t%g!th!r with th!high "$#ti#l #ccur#cy
r!quir!d f%r th!b!#m $%"iti%n #t th!i"%c!nt!r i"$r%b#bly th!r!#"%n why n%"uch g#ntry h#"
b!!n built u$ t% n%w. In"t!#d %f fl!xibl! b!#m d!liv!ry "y"t!m", fix!d inclin!d b!#m lin!"
h#v!b!!n r!#liz!d #t th! tw%%$!r#ting clinic#l i%n f#ciliti!" in J#$#n, wh!r!v!rtic#l b!#m"
#nd b!#m" with 45inclin#ti%n #r! #v#il#bl! t%g!th!r with h%riz%nt#l b!#m". &n%th!r
$%""ibility i" t%m%v! th!$#ti!nt r#th!r th#n th!b!#m. &t "%m!$r%t%n #"w!ll #"h!#vy i%n
f#ciliti!", tr!#tm!nt ch#ir"#r!#v#il#bl!%r m%uld"th#t c#n b!r%t#t!d ("!!$#ti!nt $%"iti%ning
f%r d!t#il"). It "h%uld b!m!nti%n!d th#t "iz!#nd w!ight %f #g#ntry c%uld b!r!duc!d by th!
u"! %f "u$!rc%nducting m#gn!t". Th! cry%g!nic" inv%lv!d, h%w!v!r, r!quir!" c%n"id!r#bl!
"$#c!#nd i"m%r!d!lic#t!t%h#ndl!in t!rm"%f "t#bility #nd $#ti!nt "#f!ty.
Hadrontherapy is developing to provide the same geometrical flexibility, as photon
therapy gantries have to be used for also these facilities. At the same time as the proton
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
32/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% S@
gantries are in routine operation now this is not achieved for carbon treatment. However,
advanced design efforts have reached the implantation step at GSI, but are under
consideration Riesenrad gantry. Because of the high costs of a gantry, alternatives to gantry
system have been experienced and considered at Chiba, Hyogo and NIRS.
3.6 Cost-effectiveness of hadron therapy in cancer
In the study by Mark Lodge et al (2007), they identified a total of 7209 articles, after
they de-duplication it, they reduced to 5089 articles, then they found the following number of
articles to be relevant to his review: neutron therapy (563), proton therapy (137), ion therapy
(49); economic viability and cost-effectiveness (10). Table 4.3 is shown the results literature
review in comparison with conventional therapy classified by tumour site.
Table 4.2: Results literature review in comparison with conventional therapy classified
by tumour site;after (Lodge et al, 2007)
The current literature shows that the existing data do not suggest the rapid expansion of
hadron therapy, as a major treatment modality would be appropriate. The formation of a
European hadron therapy would offer a simple way of accelerating the rate at which they
obtain high-quality evidence that should be used in assessing the role of hadron therapy in the
management of cancer.
The heavy charge particle therapy will be more expensive than the conventional
radiotherapy (magnetic fields required, concrete thickness, training and level of the
staff)(Goitein, 1997). Oliv
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
33/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SS
)H0I.7/ LCTreatment planning
In this section we well describe the procedure that used for intracranial treatments for
proton therapy (at CPO: Centre de Protonthrapie d'Orsay, French)(8'*4%5^ @AAQd.
4.1Imaging protocol
Physician counter the tumour and all surrounding critical organ by using the Computed
Tomography (CT scan) data with transverse slices of 1 mm to simulate the dose deposition
inside the patient. They also use Magnetic Resonance Image (MRI). Technical progress in the
software contributed the achievement of the fusion between CT and MRI images. Because CT
scans are based upon the photons interactions in the middle, then a calibration curve helps to
convert from X-rays absorption to protons stopping power (Fig. 4.1)
Fig. 4.1: Calibration curves for intracranial treatment (CPO); after (8'*4%5^ @AAQd
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
34/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SI
4.2Organs contouring
ICRU is fixed standardised volumes of contouring:
I. Gross Tumour Volume (GTV); tumour visible on the images.II. Clinical Target volumes (CTV); tumour with possible microscopic
extension.
III. Planning Target Volumes (PTV); positive margins of treatment includingthe treatment uncertainties.
4.3Dose calculation algorithms
Numerous models are used for the dose calculating as the following:
I. The ray-tracing model is a one-dimension model. The SOBP parameters arematched to the different densities and thickness for one direction of the
tissue. It is incorrect and limited for example to bone and air cavities,
because of scattering effect.
II. The classical Monte-Carlo model is very efficient.III. The pencil beam models are good compromise between accuracy and
calculation time.
4.4Choice of the ports
To deliver the prescribed dose to the tumour and do not exceed the maximal dose for
surrounding healthy critical organs; the dosimetrist must find the good compromise. For
example the maxima(empiric data known by physician) dose for the optic chiasma is 55
Gy.
4.5Positioning and treatments
The ballistic properties that are used for hadron therapy have a specific positioning
system. At CPO, 1mm and 1 used for basic requirement for intracranial proton therapy.
Therefore, a specific system (frames, mask, foams, byte blocks,.) is used with patient
immobilizer to adjust the real position to the reference position prescribed by the planning
system of treatment 4 or 5 of fiduciary markers are fixed in the patient skull (Fig. 4.2). The
DRR (Digital Reconstructed radiography) extracted from CT data are compared with the axial
and the perpendicular X-rays radiography of the patient seated. The robot moves to position
the patient in several iterations by specific software, while there are other techniques such as
anatomic references or CCD camera (NAC) avoids the implantation of marks (Fig. 4.2). Thenthe dose during the treatment is monitored with transmission ionisation chamber. And in
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
35/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SO
passive scattering technique, the classical irradiation last is about 1 minute (2 Gy EQ CO per
session, 2 Gy/min). Besides, there is no specific perception or pain of the patient during the
treatment.
On other hand, the comparison of treatment plans using photon and proton is shown in
Figure 4.3 with different number of fields for cancer of Glandular parotid that locates nearest
to the ear, as we can see that on the left image, 100% dose is delivered by overlapping two
beams of photon from two entry points. The attenuation after beams hits the tumour; the
remaining doses are received by other part of the head up to 30%. Moreover, the use of
photon beam of 5 fields in middle image; the doses receive by other parts beyond the tumour
is lessen since lower energy beams can be used to overlapping at the tumour location but
remain delivering 100% dose to tumour. As a result, a large amount of healthy tissues are
irradiated. However, the proton beams in 3 fields on the right image; as described with the
depth-dose profile of proton treatment that the doses received by other body parts after the
Bragg peak is very small as shown in this treatment due to doses are very conformed to thetumour site.
Fig. 4.2: Patient in position treatment at CPO, after (8'*4%5^@AAQd
Fig. 4.3: Treatment plan comparison of using photon and proton
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
36/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SP
)H0I.7/ NCConclusion
There is a worldwide interest in the development of heavy charged particles therapy as
a conformal tool in radiotherapy. Due to hadron that has its unique depth-dose track with low
entry dose and an ionising intensity peak near the of its range, i.e Bragg Peak.
Wh!r!#" th! fir"t f#ciliti!" w!r! b#"!d %n r!"!#rch c!nt!r" in nucl!#r $hy"ic", th!
curr!nt t!nd!ncy g%!" t%w#rd" #$r%gr!""iv! int!gr#ti%n in"id! th! h%"$it#l, with m%d!rn
!qui$m!nt d!v!l%$!d "$!cific#lly f%r thi" #$$lic#ti%n. Du! t% it" "!v!r#l inn%v#ting
t!chniqu!"#nd %rg#niz#ti%n", %n!%f th!m#in ch#ll!ng!%f thi" th!r#$y i" t%r!#ch #r%utin!
r#t!%f tr!#tm!nt.
F%r it"inh!r!nt b#lli"tic qu#liti!", h!#vy ch#rg!d $#rticl!"b!#m"c#n incr!#"!th!d%"!
t%th!t#rg!t v%lum!, l%w!ring th!int!gr#l d%"!d!liv!r!d t%th!$#ti!nt #nd $r%t!cting critic#l
%rg#n" . F%r # c!rt#in numb!r %f clinic#l indic#ti%n", th!$r%t%n"c%n"titut!#lr!#dy th!b!"t
ch%ic!$r%v!n by clinic#l tri#l". F%r %th!r indic#ti%n", th!$%t!nti#l b!n!fit %f $r%t%n" "till
r!quir!"#clinic#l v#lid#ti%n, #"w!ll #"c%"t-!ff!ctiv!n!"""tudi!"c%m$#ring with #lt!rn#tiv!
t!chniqu!""uch #"c%nf%rm#l #$$r%#ch!"with $h%t%n b!#m". H!#vy i%n"b!#m"will h#v!t%
c%nfirm in th!n!xt y!#r"th!ir #ttr#ctiv!bi%l%gic#l $r%$!rti!".
Mutu#l t!chn%l%gic#l tr#n"f!r b!tw!!n c%nf%rm#l t!chniqu!" m#k!" it $%""ibl! t%
incr!#"! th! qu#lity #nd th! $r!ci"i%n %f tr!#tm!nt" #nd c%n"%lid#t! th! $l#c! %f h!#vy
ch#rg!d $#rticl!"th!r#$y within th!th!r#$!utic #r"!n#l %f tr!#tm!nt #g#in"t th!c#nc!r.
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
37/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SU
@7G7/7317-
QR Goitein M, Radiation Oncology; A Physicists-Eye view, (Biological and MedicalPhysics, Biomedical engineering),Springer-Verlag New York, LLC, 2007
@R Goitein M, the technology of hadrontherapy; the context within which technicalchoices are made, Advances in hadron therapy, Elsevier scince-1997,, pp. 141-160
SR Amaldi U., Silari M. Biomedical Applications of Hadron Accelerators, Nuclearinstruments & methods in physics research. Section B, Beam interactions with
materials and atoms, Elsevier Science, Amsterdam, 1996, vol. 113, pp. 513-521
:/ Lodge M, Pijls-Johannesma, et al, A systematic literature review of the clinical andcost-effectiveness of hadron therapy in cancer, Radiotherapy and oncology : journal
of the European Society for Therapeutic Radiology and Oncology, Elsevier,
Ireland,2007, vol. 83, pp. 110-122.
A/ Jones, D.T.L., Schreuder, A.N. and Symons, J.E. Particle therapy at NAC: physicalaspects. In Proc.14th Int. Conf. on Cyclotrons and their Applications, ed. Comell,
J.C., World Scientific, Singapore, 491-498, 1996.
D/ Chu W T, Ludewigt B A and Renner T R, Instrumentation for treatment of cancerusing proton and light-ion beams, [ Review of Scintific Instruments 64, 2055(1993)]
UR Alfred R S, Vision 20/20: proton therapy, Medical physics (Med Phys), vol 36 (issue2): pp 556-568(2009).
F/ Bertil M, Alan M,Medical Devices: Use and Safety, Elsevier Health Sicences,(2006)BR Pedroni E, Bacher R, Blattman H et al The 200-MeV proton therapy project at the
Paul Scherrer Institute: conceptual design and practical realization. Med.Phys.1995;
22:37-53
QARWeinrich, Udo(GSI, Darmstadt), Gantry Design for Proton and CarbonHadrontherapy Facilities, 10th European Particle Accelerator Conference,Edinburgh, UK, 26 - 30 Jun 2006, pp.964-968
QQRG. Kraft, Tumor Therapy with Heavy Charged Particles, Prog. Part. Nucl. Phys. 45:S473-S544, 2000.
Q@RTakashi Nakamura & Lawrence Heilbronn, Handbook on Secondary ParticleProduction & Transport by High-Energy Heavy Ions, World Scientific Publishing
Company (2006)
QSR8'*4%5 >^ 3( &E) !"#$" &'($) *+, ")(-. $"(,/)0 &(,1#$')2 #3 ,(0#+1"),(&.4.(,2%%=1+C0 -,( )/% 2,4(0%e F/%('&"^ )&@* #11767/0.4/ +1H446 ZXE0/.=167
#11767/0.4/- G4/ :78=1=37 038 W38B-./=7-Y QA)/ `QU)/ >'" @AAQ ` .(4/,+12%
cHfd
QIRS. Lacombe, C. Le Sech, Advances in radiation biology: Radiosensitization inDNA and living cells, Surface Science, Volume 603, Issues 10-12, 1 June 2009,
Pages 1953-1960
QORg'+ gR L15T%+0^ 9]% K%5-T%^ 5#,)$1 6+7&(,#2+3 +* 8#+'+/#$(''. 9&1#7#:)0;&,)(0)(?2 *+, >,+1+32 (30 6(,@+3 A+32) *9(389&(>G9&E HG789&E GI
;>&(>G9 J96GEGKLMN>GEGKLM6O^ PGE753 Q?) *OO73 .^ . H&97&8L ,??F^ G9&E HG789&E GI ;>&(>G9 J96GEGKLMN>GEGKLM6O^ PGE753 A1)
*OO73 ,^ . H793 ,??,^
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
38/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SW
.F/Dale R G, et al, A 2009 why more needs to be known about RBE effect in modernradiotherapy, Applied Radiation & Isotopes; including data, instrumentation, and
method for use in agriculture, industry and medicine, vol 67, p.p 387-92
.S/. IAEA-TECDOC-1560, Dose Reporting in Ion Beam Therapy, Proceedings of ameeting organized jointly by the International Atomic Energy Agency and the
International Commission on Radiation Units and Measurements, Inc. and held inOhio, United States of America, 1820 March 2006
,?/T. Haberer,Advances n Charged Particles Therapy, Nuclear Physics in 21st Century:Int. Nucl. Phys. Conf. INPC 2001. AIP Conf. Proc., Vol. 610, pp. 157-166, 2002
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
39/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% SB
Appendix A: Proton Beam Facilities at the world & Particle Therapy Facilities in
operation
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
40/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% IA
Appendix B: Hadron therapy patient statistics (data received per end of Feb 2009);
( from PTCOG website:
http://ptcog.web.psi.ch/Archive/Patientstatisticsupdate02Mar2009.pdf) log on 2/9/2009.
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
41/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% IQ
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
42/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% I@
Appendix C: Hadron therapy facilities in planning stages or under construction and in
operation
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
43/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% IS
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
44/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% II
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
45/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
.'C% IO
Appendix D: Hadron therapy design and layout of;
1. HIMAC FACILITIES
2. HIT FACILITIES
8/13/2019 HADRON THERAPY FOR CANCER USING HEAVY IONS
46/46
HADRON THERAPY FOR CANCER USING HEAVY IONS
3. PIMMS FACILITES
4. CANO FACILITIES