Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
1
GUIDELINE FORELECTRICITY METERS:APPROVAL, TESTING AND INITIAL
VERIFICATION REQUIREMENTS
2
DOCUMENT CONTROL
Revision number Date Review by Justification for reviewDraft 01/2012 28 August 2012 Approved by Energy
CommissionRevision 01-2013 7 March 2013 Working Group Metering
Guideline & JabatanStandard Malaysia
Ratifying for existingprocess
Revision 02-2013 28 April 2013 ST & SIRIM QAS To include additionalrequirement forimported meter
3
CONTENTS
PART 1: PRELIMINARY 3
PART 2: APPROVAL REQUIREMENT 6
PART 3: TESTING & INITIAL VERIFICATION 12
PART 4: GENERAL 14
FIRST SCHEDULE: METROLOGICAL REQUIREMENT 15
SECOND SCHEDULE: INITIAL VERIFICATION PROCEDURE
Annex 1
4
PART 1
PRELIMINARY
Interpretation
1. In these Guidelines, unless the context otherwise requires:
“Act” means Electricity Supply Act 1990.
“Accredited Laboratory” means a testing facility accredited to MS ISO/IEC 17025 general
requirements for the competence of testing and calibration laboratories or other laboratory
recognised by national accreditation and standardisation authority.
“Accredited Certification Bodies” means accredited by Standards Malaysia
“Accreditation and Standardisation Authority” means Department of Standards Malaysia
“Applicant” means any company who is supplied electricity meter to licensee used for the
sale of electricity
“Authorised Manufacturer” means any company who manufactures electricity meters which
complies with these guidelines and operates a quality management system which conforms
to MS ISO 9001 or equivalent.
“Batch Certification” means conformity assessment requirement for imported meters
through consignment or batch testing
“Certificate of Approval” means approved document issued by Commission (In line with
regulation 97 of Electricity Regulation 1994)
“Certificate of Pattern Approval” means approved document issued by National
Measurement Standard Laboratory
“Commission” means Energy Commission, Malaysia.
“Consumer” means as prescribed in the Act
Revision01-2013
Revision01-2013
Revision02-2013
5
“Electricity Meter” means self contained measuring instrument intended to measure
electrical energy continuously by integrating power with respect to time and to store the
result
“International Recognised Laboratory” means an international testing and calibration
laboratory which is accredited by recognised international accreditation and standardization
body which signatory to Asia Pacific Laboratory Accreditation Cooperation (APLAC) or
International Laboratory Accreditation Cooperation (ILAC), Mutual Recognition Arrangement
(MRA)
“Licensee” means as prescribed in the Act
“National Measurement Standards Laboratory” means National Metrology Laboratory
(NML), SIRIM Berhad
“Pattern Approval” means a meter type which has been evaluated and approved by a
National Measurement Standards Laboratory through compliance with International
Organization of Legal Metrology (OIML) standards, and by then issued a Certificate of Pattern
Approval
“Product Certification” means a conformity assessment system operated by Accredited
Certification Bodies related to product to which the same product certification requirements,
specified standard and procedure including relevant regulatory provisions apply
“Product Certification License” means a document issued by Accredited Certification Bodies
recording the conformance of a product to the specified standard(s) and the Accredited
Certification Bodies certification requirements in accordance to the product certification
agreement
“Reference Standard” means electricity measurement standard which traceable to national
standards by periodic calibration interval National Measurement Standards Laboratory
“Testing facility” means a combination of meter test bench and working standards, used for
the testing and verification of electricity meters
“Transfer Standard” means a Standard that has been verified by comparison to a Reference
Standard and is used to compare Working Standards indirectly against Reference Standards
Revision01-2013
Revision01-2013
Revision01-2013
6
“Type Test Report” means a full type test report issued by an independent accredited test
laboratory recognised by accredited certification bodies
“Working Standard” means a standard that has been verified by comparison to a Reference
Standard or to a Transfer Standard and is used for testing and verification of electricity
meters. Test equipment used for onsite testing and dispute resolution is deemed ‘Working
Standards’
“Year” means calendar year, according to the Gregorian calendar
Revision01-2013
7
PART 2
APPROVAL REQUIREMENT
Obligation of Parties to Ensure Fairness
2. The sale of electricity through electricity meters shall be subjected to practices that
ensure fairness and equity for both the customer and the supplier. Both parties must
understand their rights and responsibilities. Meters and their installation must be practical
and where appropriate, consistent with international standards.
Regulatory Requirement on Electricity Meters
3. (1) In ensuring all meters when manufacture are in compliance with the regulatory
requirement, reference to Annex 1 stipulate the overall regulatory requirements that need to
be complied with by manufacturer for the electricity meter before it can be installed at any
premises.
(2) With a view to use for trade, the pattern or design of electricity meters shall be
approved by National Measurement Standard Laboratory.
(3) All electricity meters used for billing purposes by a licensee must be certified by
Accredited Certification Bodies to show that, when tested, they conformed to the original
pattern approval and operated within the prescribed levels of accuracy.
(4) The licensee shall be supplied with appropriate documentation to certify that all
meters are manufactured and individually tested in accordance with MS Standards, where the
standards states it is applicable to newly manufactured electricity meters and the
requirements of this guideline.
Appropriate Meter
4. For the purpose of subsection (32) (1) of the Act, a meter is appropriate if it complies
with the following requirements;
(a) MS 62053-11: Electromechanical meters for active energy (classes 0.5, 1 and
2);
(b) MS 62053-21: Static Meters for active energy (classes 1 and 2);
8
(c) MS 62053-22: Static Meters for active energy (classes 0.2S and 0.5S);
(d) MS 62053-23: Static Meters for reactive energy (classes 2 and 3);
(e) MS 62052-11: Electricity metering equipment (AC) – General requirements,
tests and test conditions, Part 11: Metering equipment;
(f) First Schedule of this Guideline ; and
(g) Metering Code as specified in the Malaysian Distribution Code.
Pattern Approval
5. (1) Before any electricity meters can be manufactured, the authorised manufacturer
shall apply to the National Measurement Standards Laboratory for Pattern Approval, where
such approval may include metrological performance and maintaining calibration in the
range of operating conditions that is stated in this guideline.
(2) The process for Pattern Approval includes an assessment of the prototype of theelectric meter to ensure that they meet the required standards. Type Test Report of theproduct shall be included.
(3) Once approved, a Certificate of Pattern Approval will be issued by NationalMeasurement Standards Laboratory. Evaluation report and recommendation for new PatternApproval will then issued by National Measurement Standards Laboratory.
(4) Any changes to the pattern found during Product Certification audit, shall bereported to National Measurement Standards Laboratory and Commission.
Certificate of Approval (COA)
6. (1) The authorised manufacturer shall apply to Commission for new meter to haveCertificate of Approval (COA).
(2) The application shall be done with online application through e-Permit. Type TestReport shall be included and Commission will verify the validity and authenticity of the report.
(3) Once approved, a Certificate of Approval will be issued by Commission and must berenew every year.
Revision01-2013
Revision01-2013
Revision01-2013
Revision01-2013
Revision01-2013
Revision01-2013
9
Product Certification
7. (1) Accredited Certification Bodies, with the participation of the Commission wherenecessary, will carry out preliminary examination and evaluation of manufacturing productionof electricity meters.
(2) Authorised manufacturer shall submit the Certificate of Pattern Approval,Certificate of Approval and Type Test Report during application to Accredited CertificationBodies.
(3) Accredited Certification Bodies will evaluate the Type Test Report in whichreferring to stipulated requirements and international standards. The validity of Type TestReport is not more than 10 years and for new application, the validity period must not lessthan one year.
(4) Pre-requisite for the audit, authorised manufacturer shall have AccreditedLaboratory.
(5) Accredited Certification Bodies will perform the audit on the production ofelectricity meters to ensure that the quality system and product conform to the stipulatedrequirements and standard. Verification test may be conducted using in depth surveillancemethod according to accredited test laboratory practices.
(6) Examination of the facts from the verification tests will be documented in the testreport that will determine the basis for the certification of the product.
(7) Only authorised manufacturer with electricity meters that complies with productcertification processes and received a valid Certificate of Approval by the Commission willhave the right to affix the Accredited Certification Bodies certification mark or label onelectricity meters.
(8) Surveillance audit by Accreditation Certified Bodies with the participation ofCommission where necessary will be performed every year.
(9) Where non-conformity or defects identified in the production process or during
the auditing process control of the electricity meters, Accredited Certification Bodies will
require the manufacturers to do corrective action within a reasonable period of time.
Accredited Certification Bodies will confirm satisfactory action has been implemented by the
Authorised Manufacturer before issuing the product certification license.
Revision01-2013
Revision01-2013
Revision01-2013
Revision01-2013
Revision01-2013
10
(10) If the action taken is not satisfactory or if additional testing fails to meet specifiedcriteria, the Accreditation Certified Bodies in consultation with the Commission shall issue anotice of suspension of production to the manufacturer.
(11) If a Certificate of Approval has been issued, the Commission has the right to
revoke or suspend it under the following circumstances:
i) if the Authorised Manufacturer has fail to maintain its quality control such that
the electricity meters installed in the consumers’ premises are found to be of
non-compliance to the requirement of this guideline;
ii) if the Authorised Manufacturer is found to have made false declaration or
giving misleading information;
iii) if the Authorised Manufacturer ceases its operation or in the process of winding
up;
iv) if there is a Court order to the effect of preventing the use of such meter;
v) if the usage of the particular meter can pose danger to the consumer;
vi) any other circumstances that the Commission feels reasonable to suspend or
revoke in view of public interest.
(12) Once the Certificate of Approval is suspended or revoked, the manufactureris notpermitted tousethe Accredited Certification Bodies certification mark or labels on electricitymeters.
(13) The Authorised Manufacturer shall be required to comply with the AccreditedCertification Bodies procedures with regards to suspension, revocation, withdrawal andtermination of product certification license.
Batch Certification
8. (1) Every imported electricity meters shall be verified through the batch certificationprocedure, performed by Accredited Certification Bodies.
(2) Applicant shall submit application to Accredited Certification Bodies with completedocumentation. Document required as mentioned in sub-requirement 7 (2).
(3) The evaluation of document by applicant as mentioned in sub-requirement 7 (3).
Revision01-2013
Revision02-2013
Revision02-2013
Revision02-2013
11
(4) The consignment test shall be performed by sampling method. Approval is givenonly to consignments which have satisfactorily gone through the inspection and testingprocess.
(5) Only applicant with electricity meters that complies with batch certificationprocesses and received a valid Certificate of Approval by the Commission will have the rightto affix the Accredited Certification Bodies certification mark or label on electricity meters.
Markings
9. (1) The electricity meter shall be legibly & durably marked as a minimum with the
followings:
a) Manufacturer
b) Vnom
c) Imax
d) fnom
e) Ib / In
f) Serial number
g) Number of phases
h) Number of wires
i) Meter constant (s)
j) Year of manufacture
k) Accuracy class
l) Directionality of energy flow if the meter is bidirectional or unidirectional. No
marking is required if the meter is capable only of positive direction of positive
direction energy flow
m) Meter type
n) The connection mode (s) for which the meter is specified
o) Connection terminals uniquely identified to distinguish between terminals
p) “Product Certification” mark or labels approved by Commission
(2) The marking on the meters shall be indelible, distinct and legible from outside the
meter. The markings of electricity meter intended for outdoor locations shall withstand solar
radiation. If the serial number is affixed to dismountable parts, the serial number shall also be
provided in a position where it is not readily dissociated from parts determining the
metrological characteristics.
Revision02-2013
Revision02-2013
Revision01-2013
12
(3) Electricity meters shall be provided with suitable means of sealing metrological
sensitive parts. A seal is used to provide security for the electricity meter from tampering and
it can be in the form of a crimped security seal or any suitable form.
13
PART 3
TESTING & INITIAL VERIFICATION
Initial verification
10. (1) Every electricity meter shall be tested and calibrated with the applicable
standard(s) by Authorised Manufacturer with an Accredited Laboratory.
(2) Initial verification process shall be performed in the manner as stipulated in Second
Schedule of this guideline or in such manner as determine by the Commission in consultation
with National Measurement Standards Laboratory.
(3) Each meter shall display the year it was last tested and verifed.
(4) Once verified, a meter may remain in service for a period of 10 years of any suchperiod as approved by the Commission, except where the meter is suspected to bemalfunctioning.
(5) Each meter type shall have ascribed to it a certification life (time to allow oncircuit). If no supporting evidence is available, a period of 10 years or such time that themeter to operate within permitted margins of error, whichever is least. In any event no metershall remain on circuit for periods exceeding 20 years.
Records
11. (1) The licensee shall maintain auditable records of all meters on circuit and theirprojected date of replacement. The minimum retention period for the records is 10 years.
(2) The following records shall be kept and made available for inspection by theCommission or its representative at any time;
(a) Test certificate of all in service electricity meters,(b) Meter test equipment / standards calibration history,(c) Meter re-certification or replacement schedules,(d) Result of all ad hoc on site or laboratory meter testing initiated by the
customer or licensee,(e) For every account, details of the meter (s) installed.
Revision01-2013
Revision01-2013
14
Meter testing equipment
10. (1) The following pieces of test equipment are deemed by this guideline to beStandards;
(a) Reference standards,(b) AC/DC Transfer standards,(c) AC transfer standard,(d) Working standards.
(2) All Standards shall be maintained and calibration at the following intervals by anapproved accredited laboratory and/or National Measurement Standards Laboratory;
Standard type Maximum period between calibrations
Reference Standards 2 year
AC/DC Transfer Standards 1 year
AC Transfer Standards 1 year
Working Standards 1 year
(3) Test reports provided by approved accredited laboratories, shall be retained for
inspection on request by the Commission
(4) Apparatus used for the regulating and testing of electricity meters shall cease to be
used if periodic verification shows the instrument to be inaccurate or unstable. Such
apparatus must not be brought back into service until it has been repaired and re-calibrated.
(5) For any test load, the load applied to a working standard integrating meter shall
not be less than 25% or more than 125% of its full load rating. For a working standard, the
applied load shall not be less than 40 per cent or more than 100% of its full scale or range
reading.
15
PART 4
GENERAL
Public information
11. (1) The licensee shall provide metering information as public handouts available from
their offices and other suitable facilities. These include, but not limited to the following
advice;
a) How to read a meter,
b) Customer obligations on meter care,
c) What to do if you think your bill is too high,
d) Advice who to contact with queries,
e) How to get meter checked,
f) Dispute resolution procedure,
(2) Once metered consumption has commenced the licensee is responsible for
maintenance and replacement of the meter. However both the licensee and consumer
shall have obligations in keeping electricity meter in good order, and these obligations
should be emphasized in published documentation.
Transition
12. (1) The licensee, in implementing this guideline shall produce a transition plan to
ensure that all metering procedures are in compliance by nominated future dates.
(2) The plan, which shall be provided to the Commission, shall include but not be
limited to;
a) Phasing out of non compliance meters,
b) Meter replacement programs which take account of failure rates,
c) Production of Public Information,
d) Procurement of meters only in compliance with this regulation,
e) Upgrading of Test & Calibration facilities,
f) Implementation and phasing of 100% customer metering for electricity;
g) Phasing out of meters older than 20 years.
16
FIRST SCHEDULE
METROLOGICAL REQUIREMENTS
1.0 Metrological characteristics
1.1 Current (I)
R.M.S value of the electrical current flowing through the electricity meter
1.2 Starting current (Ist)
Lowest value of current specified by the manufacturer at which the electricity meter should
register electrical energy at unity power factor and, for poly-phase electricity meters, with
balanced load
1.3 Minimum current (Imin)
Lowest value of current at which the electricity meter is specified by the manufacturer to
meet the accuracy requirement by this guideline
1.4 Basic current (Ib)
Value of current in accordance with the relevant performance of a direct connected electricity
meter are fixed
1.5 Rated current (In)
Value of current in accordance with the relevant performance of a transformer operated
electricity meter are fixed
1.6 Transitional current (Itr)
Value of current at and above which the electricity meter is specified by the manufacturer to
lie within smallest maximum permissible error corresponding to the accuracy class of the
electricity meter
1.7 Maximum current (Imax)
Highest value of current at which the electricity meter is specified by the manufacturer to
meet the accuracy requirements of this guideline
17
1.18 Active energy
Active power integrated over time. Usually expressed in kWh or MWh
T T
E(T ) = ∫p(t) ⋅dt = ∫u(t) ⋅ i(t) ⋅dt
0 0
where:
E is the active energy. Other symbols are as defined in Active power definition
1.19 Relative error (of indication)
Measured quantity value minus reference quantity value, divided by the reference quantity
value. The relative error is usually expressed as a percentage of the reference quantity
value.
1.20 Maximum permissible error (m.p.e)
Extreme value of a measurement error, with respect to a known reference quantity value,
permitted by specifications or regulations for a given electricity meter. (measurement,
measuring instrument, or measuring system)
1.21 Base maximum permissible error
Extreme values of the error (of indication) of a meter, when the current and power factor
are varied within the intervals given by the rated operating conditions, and when the meter
is otherwise operated at reference conditions.
1.22 Maximum permissible error shift
Extreme values of the change in error (of a meter, permitted by this guideline, when a
single influence factor is taken from its value at reference conditions and varied within
the rated operating conditions. For each influence factor there is one corresponding
maximum permissible error shift.
18
1.23 Rated operating condition
Operating conditions that must be fulfilled during measurement in order that an electricity
meter performs as designed
1.24 Reference condition
Operating conditions prescribed for evaluating the performance of an electricity meter for
comparison of measurement results
1.25 Accuracy class
Class of measuring instruments or measuring systems that stated metrological requirements
that are intended to keep measurement errors or instrumental uncertainties within specified
limits under specified operating conditions
1.26 Durability
Ability of a measuring instrument to maintain its performance characteristics over a period of
use
1.27 Measurand
The particular quantity subject to measurement
2.0 Metrological requirements
2.1 Units of measurement
The units of measurement for active electrical energy shall be one of the following units:
Wh, kWh, MWh, GWh
2.2 Rated operating conditions
Rated operating conditions are specified in Table 1.
19
Table 1: Rated operating conditions
Conditions or
influence quantity
Values, Ranges
Frequency fnom ±2% where fnom is to be specified by the manufacturer.
If the manufacturer specifies more than one nominal frequency, the
rated operating conditions shall be the combination of all fnom ±2%
intervals
Voltage Unom ±10% where Unom is to be specified by the manufacturer.
Meters designed to operate across a range of voltages shall have
applicable Unom ‘s specified by the manufacturer. If the manufacturer
specifies more than one nominal voltage the rated operating conditions
shall be the combination of all Unom ±10% intervals
Current Ist to Imax
Imax, Itr, Imin and Ist are to be specified by the manufacturer in
accordance with the following:
Direct
Connected
Accuracy Class
2 1 0.5 0.2
Imax / Itr ≥ 50 ≥ 50 ≥ 50 ≥ 50
Imax / Imin ≥ 100 ≥ 125 ≥ 250 ≥ 250
Imax / Ist ≥ 1000 ≥ 1250 ≥ 1250 ≥ 1250
Transformer
operated
Accuracy Class
2 1 0.5 0.2
Imax / Itr ≥ 24 ≥ 24 ≥ 24 ≥ 24
Imax / Imin ≥ 60 ≥ 100 (1) ≥ 100 ≥ 100
Imax / Ist ≥ 480 ≥ 600 ≥ 1200 ≥ 1200
20
Power factor From 0.5 inductive to 1 to 0.8 capacitive, except for classes 0.5 and 0.2
where the operating range is from 0.5 inductive to 1 to 0.5 capacitive.
For bi-directional meters the power factor range limits are valid in both
directions.
Temperature As pre-scribed in MS 62052-11:2009
Humidity and
water
As pre-scribed in MS 62052-11:2009
Connections modes The manufacturer shall specify whether the meter is intended for
direct connection, connection through current transformers or
through current and voltage transformers.
The manufacturer shall specify the connection mode(s), the number
of measurement elements of the meter and the number of phases of
the electric system for which the meter is intended.
A meter in accordance with this recommendation can be (but is not
limited to) one or more of the following:
Description
Single-phase two wire, 1 element
Three-phase four wire 3 element
Three-phase three-wire 2-element (applicable only in cases where
leakage currents can be ruled out)
The manufacturer may specify alternative connection modes for
poly-phase meters. These alternative connection mode(s) shall also
be part(s) of the operating conditions.
21
Harmonics The voltage and current shall be allowed to deviate from the
sinusoidal form.
Load balance The load balance shall be allowed to vary from fully balanced
conditions to current in only one current circuit for poly-phase
meters.
2.3 Accuracy requirements
2.3.1 General
a) The manufacturer shall specify the accuracy class of the electricity meter to be one of
class: 2, 1, 0.5 or 0.2
b) The electricity shall be designed and manufactured such that its error does not exceed
the maximum permissible error for the specified class under rated operating
conditions
c) The meter shall be designed and manufactured such that, when exposed to
disturbances, significant faults do not occur
Note: Type and accuracy of electricity meter consumption as prescribed in Malaysia
Distribution Code; Clause 9.5.3: Type and Accuracy of Revenue Metering Equipment
2.3.2 Direction of energy flow
Where a manufacturer has specified that a meter shall be capable of bidirectional
energy flow, the electricity meter shall correctly handle both positive and negative mean
energy flow and the electricity meter shall fulfill the requirement of this guideline for energy
flow in both directions. The polarity of energy flow shall be defined by the manufacturer’s
connection instructions for the meter. Mean energy flow refers to the active power
integrated over at least one cycle of the nominal frequency.
A meter shall fall into at least one of the following categories:
Single-register, bi-directional; where the meter is specified as capable of measuring
both positive and negative mean energy flow, and where the net result will be placed
in a single register.
22
Two-register, bi-directional: where the meter is specified as capable of measuring
both positive and negative mean energy flow, as defined by connection of the meter,
and where the positive result and negative result are placed in different registers.
Single-register, positive direction only; where the meter is specified as capable of
measuring and registering only positive mean energy flow. It may inherently, by its
design, register only positive mean energy flow or it may be equipped with reverse
running detent.
Single-register, uni-directional; where the meter is specified as capable of measuring
and registering the absolute value of the mean energy flow. Normally such a meter
will register all energy as consumed energy independent of the true direction of the
energy flow or of how the meter is connected.
For bi-directional meters, energy registration shall occur in the correct register when the
direction of flow changes.
Note: The terms “single-register” and “two register” in the list above refer to the basic
energy register(s) only. A meter may have other registers, e.g. for storage of tariff and/or
phase information.
2.3.3 Base maximum permissible errors
The intrinsic error (expressed in percent) shall be within the base maximum permissible
error stated in Table 2 when the current and power factor are varied within the limits
given by Table 2 (operating range), and when the meter is otherwise operated at
reference conditions.
Table 2: Base maximum permissible errors and no load requirements
Quantity Base maximum permissible errors (%) for meters of class
Current (I) Power factor 2 1 0.5 0.2
Itr ≤ I ≤ Imax
Unity ± 2.0 ± 1.0 ± 0.5 ± 0.2
0.5 inductive to 1 to 0.8
capacitive
± 2.5 ± 1.5 ± 0.6 ± 0.3
Imin ≤ I < Itr
Unity ± 2.5 ± 1.5 ± 1.0 ± 0.4
0.5 inductive to 1 to 0.8
capacitive
± 2.5 ± 1.8 ± 1.0 ± 0.5
Ist ≤ I < Imin Unity ± 2.5*Imin / I ± 1.5*Imin / I ± 1.0*Imin / I ± 0.4 *Imin /I
23
2.3.4 No load
No significant energy shall be registered under conditions of no load. The meter is always
allowed to stop for currents below Ist
2.3.5 Allowed effects of influence quantities
The temperature coefficient of the meter shall fulfill the requirement of Table 3 when the
meter is otherwise operated at reference conditions.
Table 3: Limits for temperature coefficient of error
Influence quantity Power
factor
Limits for temperature coefficient
(%K) for meters of class
2 1 0.5 0.2
Temperature Coefficient (%K), over any
interval, within the temperature range,
which is not less than 15K and not greater
than 23K, for current
Itr ≤ I ≤ Imax
1± 0.1 ± 0.05 ± 0.03 ± 0.01
0.5
inductive± 0.15 ± 0.07 ± 0.05 ± 0.02
2.4 Protection against corruption
a) The metrological characteristics of the electricity meter shall not be influenced in any
inadmissible way by the connection to it of another device, by any feature of the
connected device itself or by any remote device that communicates with the
electricity meter
b) A hardware component that is critical for metrological characteristics shall be
designed so that it can be secured. Security measures foreseen shall provide for
evidence of an intervention
c) Software that is critical for metrological characteristics shall be identified as such and
shall be secured
d) Software identification shall be easily provided by the electricity meter
e) Evidence of a software intervention shall be available for a reasonable period of time
24
f) Measurement data, software that is critical for measurement characteristics and
metrologically important parameters stored or transmitted shall be adequately
protected against accidental or intentional corruption
g) The display of the total quantity supplied or the displays from which the total quantity
supplied can be derived, whole or partial reference to which is the basis for payment,
shall not be able to be reset during use
2.5 Suitability for use
2.5.1 Readability of result
The meter shall have one (or more) indicating device (s) which is (are) capable of
presenting or displaying numerical value of each legal unit of measure for which the meter is
approved. The indicating device shall be easy to read and the characters of measurement
results shall as minimum be 4 mm high. Any decimal fractions shall be clearly indicated; for
mechanical registers, any decimal fraction drum shall be marked differently.
The indicating device shall not be significantly affected by exposure to normal
operating conditions over the maximum duration of the meter lifetime.
The indicating device shall be able to display all data relevant for billing purposes. In
the case of multiple values presented by a single indicating device it shall be possible to
display the content of all relevant memories. For automatic sequencing displays, each display
of register for billing purposes shall be retained for a minimum of 5 s.
For multi tariff meters, the register which reflects the active tariff shall be indicated. It
shall be possible to read each tariff register locally and each register shall clearly identify.
Electronic registers shall be non-volatile so that they retain stored values upon of
power. Stored values shall not be overwritten and shall be capable of being retrieved upon
restoration of power. The register shall be capable of storing and display an amount of energy
that corresponds to the meter running at P = Vnom x Imax x n for at least 4000 h, where n is the
number of phases. This capability for storage and display applies to all registers relevant for
billing including positive and negative flow registers for bi-directional meters and tariff
registers for multi tariff meters.
In the case of electronic registers, the minimum retention time for results is one year
for a disconnected meter. Electronic indicating devices shall be provided with a display test
25
that switches all the display segments on then off for the purpose of determining whether all
display segments are working.
2.5.2 Testability
The meter shall be equipped with a test output for efficient testing, such as rotor with
a mark or a test pulse output. If the design of the test output is such that the pulse rate does
not correspond to the measured power in every given relevant time interval, the
manufacturer shall declare the necessary number of pulses to ensure a standard deviation of
measurement less than 0.1 base m.p.e, at Imax, Itr and Imin.
The relation between the measured energy given by the test output and the measured
energy given by the indicating device shall comply with the marking on the name plate.
The wavelength of the radiated signals for emitting systems shall be between 550 nm
and 1000 nm. The output device in the meter shall generate a signal with a radiation strength
ET over a defined reference surface (optically active area) at a distance of 10 mm ± 1 mm from
the surface of the meter, with the following limiting values:
ON condition : 50 µ W/cm2 ≤ ET ≤ 7500 µ W/cm2
OFF condition : ET ≤ 2 µ W/cm2
2.5.3 Durability
The meter shall be designed to maintain an adequate stability of its metrological
characteristics over a period of time specified by the manufacturer, provided that it is
properly installed, maintained and used according to the manufacturer’s instructions when in
the environmental conditions for which it is intended. The manufacturer shall provide
evidence to support the durability claim.
The meter shall be designed to reduce as far as possible the effect of a defect that
would lead to an inaccurate measurement result.
The meter shall be designed and manufactured such that either
a) Significant errors do not occur, or
b) Significant durability errors are detected and acted upon by means of a
durability protection.
26
SECOND SCHEDULE
INITIAL VERIFICATION PROCESS
1.0 General
This schedule applies to the initial verification of electricity meters in use for trade
which have been either manufactured within or imported into Malaysia. This schedule does
not apply to the re-verification or in service testing or electricity meters.
2.0 Testing
2.1 Calibration status
Check that the test system used has sufficient accuracy to verify the meters under test, and
that the calibration is valid.
2.2 Conformity check
Check that the electricity meter is manufactured in conformity with the pattern approval
documentation
2.3 Warming up
It may be necessary to warm the meter up before full operation. The length of the warming
up period depends on the actual type of instrument and shall be determined in advance.
During the test for initial intrinsic error the electricity meter shall be allowed to stabilize at
each current level before measurements for a period, no longer than 5 minutes, to be
nominated by the manufacturer. The order of the testpoints shall be from lowest current to
highest current and then from highest current to lowest current. For each testpoint, the
resulting error shall be the mean of these measurements. For Imax, the maximum
measurement time shall be 10 minutes including stabilizing time.
2.4 Minimum test program
The minimum test program consists of:
a) No load check
For this test, there shall be no current circuit. The test shall be performed at Vnom.
27
For electricity meters with test output of the meter shall no produce more than one
pulse. For an electromechanical meter, the rotor of the meter shall not make a
complete revolution.
An electricity meter with more than one connection mode shall be tested in all modes.
For transformer operated meters with primary rated registers where the value of k
(and possibly Vnom) are given as primary side values, the constant k (and Vnom) shall be
recalculated to correspond to secondary side values (of voltage and current)
b) Starting current check
The test is performed at Ist and unity power factor
For initial verification of meters produced from a continuously operating process
resulting in a large number of identical units, it is sufficient from the error curve Ist to
Imin to be recorded on a sample batch every 3 months for the particular meter type.
For initial verification of meters produced by other means, it will be sufficient if the
meter is observed to run continuously when the starting current is applied.
c) Current dependence
Meters shall comply with the accuracy requirements of Table 2. As minimum these
shall be checked at the following currents:
Imin, PF = 1;
Itr, PF = 1;
Itr, PF = 0.5 inductive;
10 Itr PF = 1;
10 Itr PF = 1;
Imax PF = 1;
Imax PF = 0.5 inductive;
In the case of three phase meters with alternative single phase connection mode or
which are being used as two phase meters, this test shall be performed separately for
each phase at:
10 Itr, PF = 1 and
10 Itr PF = 0.5 inductive
28
For meters with alternate connection modes, such as one phase connections for poly
phase meters being used as two phase meters, this test shall be performed separately
for each connection mode.
d) Check of the register
If test (pulse) outputs are used for tests of accuracy requirements, a test must be
performed to ensure that the relation between the basic energy register and relevant
test output (s) complies with that specified by the manufacturer.
The energy put through the meter shall be calculated using the number of pulses from
the test output; the relative difference must not be greater than one tenth of base
maximum permissible error.
The test shall be performed at a single arbitrary current I ≥ Itr
3.0 Reference conditions for initial verifications in a laboratory
Reference conditions and load conditions for initial and subsequent verifications in a
laboratory are given in Table 4 and 5.
Table 4: Reference conditions and their tolerances for initial verification
Quantity Reference conditions Tolerance
Voltage (s) Vnom ± 2%
Ambient temperature 23ºC ± 5ºC
Frequency Fnom ± 0.5%
Wave form Sinusoidal d ≤ 2%
Electromagnetic RF fields 30 kHz – 6 Ghz 0 V / m < 2 V / m
Operating position for instruments sensitive to
position
Mounting as stated by
manufacturer
± 3.0 º
Phase sequence for poly phase meters L1, L2, L3 -
Load balance Equal current in all circuits ± 5% and ± 5º
Table 5: Load conditions and their tolerance in tests for initial verification
Current (s) Current range of device under test Class 2,1 : ± 10%
Class 0.5, 0.2 : ± 10%
Power factor Power factor range of device under test Current to voltage phase
difference ± 5 o
29
3.0 Additional Requirements For Statistical Verifications
This section contains additional requirements for verification on a statistical basis.
3.1 Lot Acceptance
A lot shall consist of meters with homogenous characteristics. All meters that comprise the lot
shall correspond to the same type approval, and shall have the same year of manufacture.
3.2 Samples
Samples shall be taken randomly taken from a lot
3.3 Statistical testing
The statistical control shall be based on attributes. The sampling system shall ensure:
An Acceptance Quality Level (AQL) of not more than 1%; and
A Limiting Quality (LQ) of not more than 7%
The AQL is the maximum percentage of non conforming items in a lot at which the lot has a
probability of 95% to be accepted.
The LQ is the percentage of non conforming items in a lot at which the lot has a maximum
probability of 5% to be accepted.
Examples are given below based on a lot of 1000 meters
Number of meters tested 40 70 100 1000
Maximum number of non conforming meters 0 1 2 10
30
ANNEX 1: NEW A PPROVAL REQUIREMENT FOR ELECTRICITY METER
Pattern ApprovalProcess by
National M easurementStandard Laboratory
(NM SL)
Certificate of Approvalfrom Commission
Application forPattern Approval to
NM SL
Product CertificationProcess by Accredited
Certification Body(ACB)
New meter
Pattern ApprovalProcess
Document submission:-Type Test Report-Pattern document-Other supporting
document
Issue Certificate ofPattern Approval
Yes
No
Register withDagangnet
Document submissionthrough online, e-permit:
-Type Test Report
Certificate ofApproval process
Issue Certificate ofPattern Approval
Application forProduct Certification
to ACB
Document submission:-Certificate of Pattern Approval
-Certificate of Approval (Commission)-Type Test Report
DocumentEvaluation
Factory Audit by ACB jointly withCommission.
Pre-requisite:-Authorised M anufacturer shall have
Accredited Laboratory-Product sample for conformity test
-Audit Evaluation-Conformity Test
-Award licence for the productto be affixing with Certified M arking-Yearly renew and the audit will be
performed once a year
No
Yes
Yes
Yes
No
No