Guide to Successful Operation of Your LC System.pdf

Embed Size (px)

Citation preview

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    1/214

    Guide to SuccessfulOperation of Your LC System

    34 Maple Street

    Milford, MA 01757WAT022378TP, Revision 0

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    2/214

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    3/214

    Table of Contents 3

    How to Use This Guide ...................................................................... 9

    Chapter 1Introduction to System Troubleshooting ........................................... 13

    1.1 Get the Facts.......................................................................... 141.2 Check Simple Things First ..................................................... 161.3 Compare System Performance to Established

    Benchmarks .......................................................................... 171.4 Identify Possible Causes....................................................... 19

    1.5 Use a Systematic Troubleshooting Approach ....................... 211.6 Getting Help .......................................................................... 221.7 Troubleshooting References.................................................. 22

    Chapter 2Troubleshooting an LC System ........................................................ 23

    2.1 Initial Survey of System Problems ........................................ 252.2 Isolating System Problems ................................................... 29

    2.2.1 System Pressure ....................................................... 292.2.2 Baseline Noise........................................................... 39

    2.2.3 Changes in Chromatographic Resolution orResults ...................................................................... 67

    Chapter 3Troubleshooting System Components ........................................... 109

    3.1 Pump Troubleshooting ........................................................ 1103.1.1 Pump Troubleshooting Table.................................... 111

    Table of Contents

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    4/214

    4 Table of Contents

    3.1.2 Isolating a Defective Pump Head(Modified Ramp Test) .............................................. 120

    3.2 Manual Injector Troubleshooting ......................................... 1223.3 Autoinjector Troubleshooting............................................... 1253.4 Column Troubleshooting ..................................................... 1373.5 Detector Troubleshooting .................................................... 1393.6 Data-Handling Device Troubleshooting............................... 151

    Chapter 4Good Chromatography/ Operating Practices ................................. 165

    4.1 Mobile Phase Preparation and Use .................................... 1654.1.1 Mobile Phase Preparation ....................................... 1664.1.2 Solvent Degassing................................................... 1694.1.3 Solvent Use ............................................................. 1704.1.4 Solvent Changeover Practices................................. 173

    4.2 System Plumbing ................................................................ 174

    4.2.1 Tubing Connection Practices ................................... 1754.2.2 Cutting Tubing.......................................................... 179

    4.3 Chromatographic Performance Tests.................................. 1824.3.1 Measuring Resolution.............................................. 1834.3.2 Measuring Capacity Factor (k') ................................ 1864.3.3 Measuring Selectivity............................................... 1884.3.4 Measuring Column Efficiency (N) ............................ 190

    4.4 Measuring System Bandspreading ..................................... 200

    Appendix AReference Information .................................................................... 202

    A.1 Solvent Properties .............................................................. 202

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    5/214

    Table of Contents 5

    A.1.1 Solvent Properties Table ......................................... 203A.1.2 Using Miscibility Numbers (M-Numbers)................. 205

    A.2 Refractive Index of Common Solvents................................ 206A.3 Solvent UV Cutoffs ............................................................. 207A.4 Wavelength Selection for Chromophore Detection............. 210A.5 Column Backpressure ........................................................ 212

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    6/214Table of Contents 6

    1-1 The System Troubleshooting Process ........................................... 151-2 System Map Example.................................................................... 17

    2-1 Pressure Release Points within a Typical LC System.................... 312-2 High System Pressure Troubleshooting......................................... 322-3 Low System Pressure Troubleshooting.......................................... 352-4 Erratic System Pressure Troubleshooting...................................... 382-5 Isolating the Source of Changing Chromatographic

    Resolution/Results......................................................................... 68

    4-1 Comparison of Water Stored in Glass and PlasticContainers ................................................................................... 1674-2 Typical Tubing/Compression Fitting Connection .......................... 1764-3 Adapter Assembly for Incompatible Fittings................................. 1774-4 Cutting Tubing with a Circular Cutter ........................................... 1804-5 Cutting Tubing with a Sharp Object ............................................. 180

    4-6 Breaking Stainless Steel Tubing .................................................. 1814-7 Using a Polymeric Tubing Cutter.................................................. 1814-8 Calculating Resolution................................................................. 1834-9 Chemical Factors that Affect System Resolution ......................... 1854-10 Calculating Capacity Factor......................................................... 1874-11 Calculating Selectivity.................................................................. 1894-12 Examples of Low and High Efficiency Peaks............................... 1914-13 Methods of Calculating Column Efficiency .................................. 1924-14 5 Sigma Method Calculation........................................................ 1934-15 4 Sigma Method Calculation....................................................... 1944-16 Tangent Method Calculation ........................................................ 195

    4-17 3 Sigma Method Calculation........................................................ 1964-18 1/2 Height Method Calculation..................................................... 196

    List of Figures

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    7/214Table of Contents 7

    4-19 2 Sigma (Inflection) Method Calculation ..................................... 1974-20 Asymmetry-Based Method Calculation ....................................... 198

    4-21 Example of Calculating Column Efficiency (N) ............................ 199

    A-1 Viscosity for Aqueous Solvent Mixtures....................................... 213

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    8/214

    Table of Contents 8

    2-1 System Problems .......................................................................... 252-2 Baseline Noise Summary ............................................................. 412-3 Fluid Path-Related Baseline Noise Troubleshooting ..................... 452-4 Detector Electronics-Related Baseline Noise

    Troubleshooting ............................................................................. 60

    2-5 Incorrect/Changing Retention Time Troubleshooting .................... 702-6 Abnormal Peak Shape Summary ................................................. 812-7 Abnormal Peak Shape Troubleshooting ........................................ 872-8 Incorrect Qualitative/Quantitative Results Troubleshooting ......... 101

    3-1 Pump Troubleshooting ................................................................ 1113-2 Pressure Recording Examples for Modified Ramp Test ............. 1213-3 Manual Injector Troubleshooting ................................................. 1233-4 Autoinjector Troubleshooting ....................................................... 1253-5 Column Troubleshooting ............................................................. 1383-6 Detector Troubleshooting ............................................................ 140

    3-7 Computer Hardware Troubleshooting ......................................... 1523-8 Integrator Hardware Troubleshooting .......................................... 1583-9 Chart Recorder Troubleshooting ................................................. 161

    A-1 Physical Properties of Solvents .................................................. 203A-2 Refractive Indices of Common Solvents ..................................... 206A-3 UV Cutoffs for Chromatographic Solvents ................................. 207A-4 Wavelength Cutoffs for Common Mobile Phases ....................... 209A-5 Electronic Absorption Bands for Representative

    Chromophores ............................................................................ 210

    List of Tables

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    9/214

    How To Use This Guide 9

    How to Use This GuidePurpose

    This guide is an aid in identifying problems within your liquid chromatographic system andhelping to return your system to a proper level of performance. Your system may be acombination of Waters and non-Waters instrumentation.

    The primary intent of this guide is to:

    Provide a logical approach to troubleshooting an LC system

    Outline good chromatography/operating practices

    Maximize system operation time

    This guide includes a separate quick reference chart that synopsizes the systemtroubleshooting information.

    AudienceThis guide is intended for use by a wide variety of LC system operators, whose experiencemay range from novice to expert.

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    10/214

    10 How to Use This Guide

    Structure of This Document

    This guide is divided into our chapters and a series of Appendixes. Each page includes a

    footer, providing easy access to information within the chapter.The table below describes the material covered in each chapter and Appendix.

    Chapter 1, Introduction toSystem Troubleshooting

    Includes a general discussion on thesystem troubleshooting process.

    Chapter 2, Troubleshooting anLC System

    Divides the system troubleshooting processinto primary LC system problems:

    System pressure (high, no or low,erratic)

    Baseline noise Changes in chromatographic resolution

    or results, including:

    Incorrect retention time

    Abnormal peak shape

    Incorrect qualitative/quantitative results

    Loss of resolution

    Possible symptoms, causes, and correctiveactions included for each topic.

    Chapter 3, TroubleshootingSystem Components

    Troubleshooting tables for individual hardwarecomponents in an LC system:

    Pump Manual injector Autoinjector Column Detector Data-handling device

    Possible symptoms, causes, and correctiveactions included for each topic.

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    11/214

    How To Use This Guide 11

    Related Documents

    Refer to the appropriate instrumentation and column operator's manuals and servicemanuals for specific troubleshooting information.

    Related Adobe Acrobat Reader DocumentationFor detailed information about using the Adobe Acrobat Reader, refer to the AdobeAcrobat Reader Online Guide . This Online Guide covers procedures such as viewing,navigating and printing electronic documentation from Adobe Acrobat Reader.

    Printing From This Electronic Document

    Adobe Acrobat Reader lets you easily print pages, pages ranges, or the entire electronicdocument by selecting Print from the File menu. For optimum print quantity, Watersrecommends that you specify a Postscript printer driver for your printer. Ideally, use aprinter that supports 600 dpi print resolution.

    Conventions Used in This Guide

    This guide uses the following conventions to make text easier to understand.

    Purple Text indicates user action. For example:Press 0, then press Enter for the remaining fields.

    Italic text denotes new or important words, and is also used for emphasis. Forexample:

    An instrument method tells the software how to acquire data.

    Chapter 4, GoodChromatography/ OperatingPractices

    Proper care and use of a system, intended asa set of operator preventative actions:

    Solvent preparation and use System plumbing Chromatographic performance tests

    (resolution, k, , and N) Measuring System bandspreading

    Appendix A, ReferenceInformation

    Reference tables

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    12/214

    12 How to Use This Guide

    Underlined, Blue Color text indicates hypertext cross-references to a specificchapter, section, subsection, or sidehead. Clicking this topic using the hand symbolautomatically brings you to this topic within the electronic document. Right-clicking

    and selecting Go Back from the popup context menu brings you back to theoriginating topic. For example:

    To begin troubleshooting, refer to Chapter 1.2, Check Simple Things First .

    Notes, Attentions, and Cautions

    Notes call out information that is important to the operator. For example:

    Note: Record your results before you proceed to the next step.

    Attentions provide information about preventing possible damage to the system orequipment. For example:

    Attention: To avoid damaging the detector flow cell, do not touch the flow cellwindow.

    Cautions provide information essential to the safety of the operator. For example:

    Caution: To avoid chemical or electrical hazards, always observe safe laboratorypractices when operating the system.

    Caution: To avoid the possibility of electrical shock and possible injury, always turnoff the detector and unplug the power cord before performing maintenanceprocedures.

    Caution: To avoid the possibility of burns, turn off the lamp at least 30 minutesbefore removing it for replacement or adjustment.

    STOP

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    13/214

    1

    13

    1Introduction to SystemTroubleshooting

    Troubleshooting a LC system can be a frustrating and sometimes mysterious process. Has

    one of these happened your lab? You left your autoinjector running samples over the weekend. When you return on

    Monday, you realize that your results gradually progressed from highly reproducibleresults to no peaks in the chromatogram.

    The retention times of sample peaks have suddenly shifted within the past hour.

    What do you do now? Where do you start?

    This chapter outlines a common-sense process for troubleshooting an LC system. Thetroubleshooting information in this chapter is presented with the following assumptions:

    Your analysis method has already been developed

    Proper system benchmarks have been recorded in the normal operation log

    System or instrument performance has degraded as compared with previouslyestablished system benchmarks (see Section 1.3, Compare System Performance toEstablished Benchmarks )

    Troubleshooting Stragety

    A troubleshooting strategy includes five primary processes:

    Identifying the symptom(s).

    Understanding possible causes of the symptom. Is it hardware, software, operation,or chemistry related?

    Isolating the exact possible causes of the problem. Determining if you have the expertise and items to correct the problem?

    Resolving the problem and resuming operation or contacting your local CustomerSupport Representative.

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    14/214

    14 Introduction to System Troubleshooting

    1

    Troubleshooting process

    To implement this strategy, you need a systematic troubleshooting process that works withall types of problems. This process is to:

    1. Get the facts.

    2. Check simple things first.

    3. Compare system performance to established benchmarks.

    4. Indentify possible causes.

    5. Use a systematic troubleshooting approach.

    6. Getting help.

    The flowchart in Figure 1-1 illustrates the troubleshooting process. The remainder of thischapter discusses each phase of the strategy.

    1.1 Get the Facts

    Before troubleshooting, be sure to get all the facts. The fact gathering stage of systemtroubleshooting is an important first step in determining how the system is malfunctioning.

    Gather general facts

    When you initially believe there is a problem with the performance of your LC system, stepback and consider:

    What makes you think something is malfunctioning? What is the evidence (shiftedpeaks, no peaks, baseline drift)?

    Is anything different now with the system as compared with establishedperformance (as documented in a log)?

    Gather important facts

    After this initial gathering process, write down answers to the following questions in your

    log: Could someone have changed something in the system (such as detector

    sensitivity)?

    Are you asking the system to do something that may be beyond its capabilities?

    Has the problem ever occurred before?

    Is the problem reproducible?

    Does this problem occur at any particular time of the day or when anotherinstrument is turned on and off?

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    15/214

    15 Introduction to System Troubleshooting

    1

    Figure 1-1 The System Troubleshooting Process

    Your LC SystemExhibits a Problem

    NoDid you gather all ofthe facts about the

    problem?

    Did you perform a visualinspection of the

    system?

    Did you compare present

    system performance withestablished

    performance?

    Did you indentify allproblem(s) per the exhibited

    symptom(s)?

    Did you perform asystematic troubleshooting

    process?

    Can you resolve theproblem?

    Problem resolved

    Review(Section 1.1 )

    No Review(Section 1.2 )

    No Review(Section 1.4 )

    No Review(Section 1.5 )

    No Review(Section 1.6 )

    Yes

    Yes

    Yes

    Yes

    Yes

    Yes

    No Review(Section 1.3 )

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    16/214

    16 Introduction to System Troubleshooting

    1

    1.2 Check Simple Things First

    When you are certain that something is affecting system performance, always check thesimple things. Finding the easy solution saves time and frustration.

    Attention: Always observe safe laboratory practices when troubleshooting. Know thechemical and physical properties of the solvents. Refer to the Material Safety Data Sheetfor the solvents in use.

    Attention: To avoid electric shock, power down the instrument and disconnect thepower cord before removing the cover and examining the instrument.

    Look for obvious "clues"

    Make a visual inspection of the system, investigating:

    Alarms triggered on any components?

    Fluid leaks?

    Pump pressure reading normal? Power cord inserted in both the outlet and the instrument rear panel?

    Power switch on?

    Fuses blown?

    Sufficient solvent in the reservoir?

    Electrical cables connected between devices? Are connections correct?

    Instrument settings changed or incorrectly set?

    Solvent flowing out of detector waste tubing?

    Correct column?

    If the visual inspection does not uncover anything obvious, compare current systemperformance to established system operation (see Section 1.3, Compare SystemPerformance to Established Benchmarks ).

    STOP

    STOP

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    17/214

    1

    Compare System Performance to Established Benchmarks 17

    1.3 Compare System Performance to EstablishedBenchmarks

    To help you identify normal operation conditions:

    Record a map of your LC system

    Keep a log

    Run test chromatograms regularly

    These three practices allow you to compare present system operation with establishedsystem performance.

    Note: A log is especially important in a lab where the system is shared by multipleoperators.

    Recording a System Map

    When your LC system is initially installed, sketch a general map of the fluidic and electricalconfiguration. Label the individual connections. Use this map to review these fluidic andelectrical connections and reconfigure when necessary. Figure 1-2 is an example of asystem map.

    Figure 1-2 System Map Example

    Pump

    Pump/ system

    controller

    Auto-injector

    ComputerA/D

    interface

    Detector

    In-Linefilter

    Column

    INJ GND + -1 V

    OUT

    IEEE-488 CONNECTIONRS-232 CONNECTION

    Chan 11 V analog out signal

    (from detector)

    Inject startsignal

    RS-232connection

    (225 l pump heads) (250 l sample loop) Waste

    IEEE-488 connectionIEEE-488

    connection

    Controlcable

    Start int.

    Electrical cable

    Res A Res B

    Fluid line- 1/16 inch (1.6 mm) O.D.,0.020 inch (0.5 mm) I.D.

    stainless steel tubing (3 ft)

    Fluid line- 1/16 inch ( 1.6 mm) O.D.,0.060 inch (1.5 mm) I.D. teflon

    tubing

    Fluid line- 1/16 inch (1.6 mm) O.D.,0.009 inch (0.23 mm) I.D.

    stainless steel tubing (kept to a minimum)

    Fluid line- 1/16 inch ( 1.6 mm) O.D.,0.009 inch (0.23 mm) I.D. stainlesssteel tubing (1 ft) (30 cm)

    (C18 3.9 mm X 150 mm)

    Red line

    CONTROL CABLE

    FLUID CONNECTION

    ELECTRICAL SIGNALS

    --

    Blue line

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    18/214

    18 Introduction to System Troubleshooting

    1

    Keeping a log

    Record operating conditions (pressure, flow rate, and so on) in a logbook. When you haveproblems with your system, use the logbook to compare system information, such as:

    Instrument settings

    Parts and/or components recently replaced (with manufacturer serial number anddate that the change was made)

    Maintenance procedures (what and when)

    Number of samples run (system throughput) in your method

    Sample, standard, and mobile phase information for each method used

    Test chromatogram, including specific operating conditions (column, flow rate,mobile phase, backpressure, and so on) for each column used

    Method table(s) used for your application(s) (such as gradient conditions, integrationparameters)

    Running a Test Chromatogram

    Always run a test chromatogram when: A new system is installed

    An instrument is replaced or added

    A column is replaced or added

    New mobile phase is prepared

    When running the standard, establish a set of standard conditions. Collect the data andrecord:

    Pressure

    Resolution (Rs) of critical peak pairs

    Capacity factor (k') for each peak

    Selectivity ( ) Column efficiency (N)

    Refer to Chapter 4, Good Chromatography/ Operating Practices for information oncalculating and troubleshooting resolution, capacity factor, selectivity, and columnefficiency.

    Keep the test chromatogram and associated data in your logbook. When you believe thereis a problem, rerun the standard. Compare the results to establish:

    If the test results are not significantly different from those previously recorded, theproblem is method-specific.

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    19/214

    1

    Identify Possible Causes 19

    If the test results are different, either the column or an instrument have changed.Repeat the test with a new (or known good) column. If the results of the secondcolumn are satisfactory, the problem is with the first column. If both columns fail, the

    problem is most likely with an instrument.

    1.4 Identify Possible Causes

    Use the exhibited symptoms to narrow down the possible causes within the system. Toidentify a problem(s):

    Identify all of the symptoms Match the symptoms with the potential possible causes

    Identifying the Symptoms

    Perform a survey of your system to determine where the exhibited symptom(s) mayoriginate.

    Some symptoms are:

    Variable or unusual system pressure (high, no or low, erratic)

    Noisy or drifting baseline

    Incorrect or changing retention time(s)

    Abnormal peak shapes (such as broad, tailing, or fronting peaks)

    Incorrect qualitative/quantitative results (too many peaks, too few peaks, wrong

    answer)

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    20/214

    20 Introduction to System Troubleshooting

    1

    Identifying the Possible Causes

    From the isolated symptom(s), write down a list of suspected possible causes in yourlogbook.

    For example:

    The troubleshooting tables in Chapter 2, Troubleshooting an LC System , and Chapter 3,Troubleshooting System Components are examples of possible causes for a particularsymptom.

    Isolating the Possible Cause

    From your list of possible causes, isolate the problem areas within your LC system. Thisnarrows down whether the problem is hardware, software, operation, or chemistry related.You can then follow a systematic troubleshooting approach to resolve the problem (seeSection 1.5 ).

    Symptom Possible Causes

    Sudden high systembackpressure

    Blocked column frit Sample precipitation on column

    Changed or incorrect flow rate Blocked tubing

    Defective pump pressuretransducer

    Drop in operating temperature

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    21/214

    1

    Use a Systematic Troubleshooting Approach 21

    1.5 Use a Systematic Troubleshooting Approach

    Once all of the possible causes are identified, resolve the problem using a systematicapproach:

    Follow a logical sequence to correct the problem

    Make only one change at a time to the system, starting with the easiest changes

    Document the results of each change

    Following a Logical Sequence

    After listing the possible causes of the problem, follow a logical troubleshooting sequenceto narrow down the exact source of the problem.

    For example, if the symptom is unusually high backpressure, loosen each fitting, startingat the detector waste outlet, and observe if pressure drops. Continue to work toward thepump until the pressure drops, identifying the possible cause of the problem.

    Note: For detailed information on system troubleshooting, refer to Chapter 2,Troubleshooting an LC System .

    Making Only One Change at a Time

    During troubleshooting, make only one change to the system at a time to the system untilthe problem is resolved. Parts of the system you may make changes to include:

    Instrument settings

    Column Sample

    Operating conditions

    Mobile phase

    Replacement of suspected malfunctioning parts

    Replacement of suspected malfunctioning instruments

    Note: If the replacement part does not correct the problem, be sure to put back theoriginal part. Conversely, rebuild or discard a malfunctioning part/assembly.

    Making only one change at a time allows you to understand the affects on systemperformance.

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    22/214

    22 Introduction to System Troubleshooting

    1

    Documenting Changes

    Always take notes as you make changes to the system, such as changing settings,swapping component parts, or swapping an instrument. Record-keeping is a valuable partof the systematic approach because it provides you with:

    A trail back to configuring the system in its original form if your troubleshooting pathis incorrect

    A trail back if new problems or symptoms arise

    If you proceed through the system troubleshooting process and the identical symptom(s)persist, reevaluate your diagnosis and investigate another area in the system.

    1.6 Getting Help

    It is important that you feel confident with your troubleshooting and maintenance abilities.If you feel that the problem is beyond your area of expertise, consult someone within yourlab with more experience. If that does not help resolve the problem, call the CustomerSupport Department (or service specialist) for help.

    Follow the troubleshooting process outlined in this guide to narrow down the symptom tothe most likely problem source. With that knowledge, you are able to concretely discussyour diagnosis with the Customer Support representative, saving valuable time andconfusion.

    1.7 Troubleshooting ReferencesFor additional information on liquid chromatography system troubleshooting or liquidchromatography in general, refer to the following reference material:

    The operator's manuals and service manuals for your LC instrumentation andcolumns

    Dolan, John, W., and Snyder, Lloyd, R. (1989), Troubleshooting LC Systems , TheHumana Press Inc., P.O. Box 2148, Clifton, NJ, 07015

    Snyder, L.R. and Kirkland, J.J., Introduction to Modern Liquid Chromatography ,Second Edition, Wiley-Interscience, New York, 1979

    The "LC Troubleshooting" column in LC/GC Magazine

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    23/214

    23

    2

    2Troubleshooting an LCSystem

    This chapter outlines LC system troubleshooting. It divides system troubleshooting into:

    Initial survey of system problems Isolating possible causes of system problems, including:

    System backpressure (high, no or low, erratic)

    Baseline noise

    Changes in chromatographic resolution or results, including:

    Incorrect/changing retention times

    No peaks or abnormal peak shapes

    Incorrect qualitative/quantitative results

    Loss of resolution

    Troubleshooting information in this chapter is presented with the assumption:

    Your analysis method has already been developed

    Proper system benchmarks have been recorded in the normal operation log System or instrument performance has degraded as compared with previously

    established system benchmarks (see Section 1.3 )

    For background information on the system troubleshooting process, refer to Chapter 1,Introduction to System Troubleshooting .

    For specific instrument and column troubleshooting, refer to Chapter 3, Troubleshooting

    System Components . When troubleshooting, keep the following safety considerations inmind:

    Attention: Always observe safe laboratory practices when troubleshooting. Know thechemical and physical properties of the solvents. Refer to the Material Safety Data Sheetfor the solvents in use.

    Attention: To avoid electric shock, power down the instrument and disconnect the powercord before removing the cover and examining the instrument.

    STOP

    STOP

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    24/214

    24 Troubleshooting an LC System

    2

    Attention: Ultraviolet light is emitted during UV and fluorescence detector operation. Toprevent eye damage, eye protection must be worn while troubleshooting the detector withits covers removed.

    If handling integrated circuit boards, use an anti-static mat and wear an anti-staticwrist strap to remove excess static charge and prevent damage to the board.

    Do not touch any of the integrated circuit chips or other components which do notspecifically require manual adjustment.

    When You Call

    Many problems with your LC system Customer Support can be easily corrected by theprocess outlined in this chapter. If you cannot correct a condition for a Waters product,contact one of the following:

    Your Waters service specialist

    Waters Customer Support Department

    Waters Service offices (listed on the rear cover of this guide)

    To expedite service, have the following information available when you call:

    1. Completed Normal Operation Log and test sample chromatogram for method

    2. Nature of symptom(s)

    3. Type and model number of pump (single or multiple solvent)

    4. Flow rate

    5. Operating pressure

    6. Mobile phase(s)

    7. Type and model number of injector (manual or autoinjector)

    8. Type and model number of detector (UV, RI, fluorescence, conductivity,electrochemical)

    9. Detector settings (wavelength, sensitivity, and so on)

    10. Type and serial number of column

    11. Sample matrix and components

    12. Data system

    13. Software version (if applicable)

    STOP

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    25/214

    Initial Survey of System Problems 25

    2

    2.1 Initial Survey of System Problems

    When a problem occurs, first perform a visual check of the system. Look for leaks,disconnected tubing, disconnected cables, changed instrument settings, and so on.Review your system map System map (as described in Section 1.3, Compare SystemPerformance to Established Benchmarks ) to verify all fluidic and electrical connections.

    Table 2-1 is a summary of the most common problems affecting system operation.

    If you determine one (or more) of these are the problem, refer to Chapter 3,Troubleshooting System Components , or the instrument/column operator's manual for thecorrective action.

    Table 2-1 System Problems

    Component Areas to Check

    General System Instrument not plugged in/not turned on No power at wall outlet Blown power fuse Incorrect instrument settings Cooling fans in instrument not running Leaks

    Incorrect air or gas supplies Disconnected or improper electrical cabling Too many instruments on same circuit Incorrect grounding between instruments

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    26/214

    26 Troubleshooting an LC System

    2

    Pump/SolventFlow

    Incorrect flow rate setting Solvent reservoir(s) empty

    Solvent not degassed or sparged

    Obstruction or crimp in tubing

    Improper tubing ID

    Improperly cut tubing Incorrect or worn ferrules/compression screws

    Plugged in-line filter

    Plugged solvent inlet filter

    Air in pump inlet lines or pump heads

    Leaking plunger seal

    Leaking fittings Malfunctioning inlet/outlet check valve

    Pressure transducer out of calibration

    Solvent reservoir positioned lower than pumpinlet

    Manual Injector Obstruction in sample loop

    Insufficient injection volume Inconsistent injection volume

    Incorrect or damaged syringe

    Leaking vent tube or valve

    Wrong sample loop

    Inconsistent injection sequence

    Leaking or worn seals

    Contaminated syringe

    Injector out of adjustment

    Table 2-1 System Problems (Continued)

    Component Areas to Check

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    27/214

    Initial Survey of System Problems 27

    2

    Autoinjector Air in syringe Insufficient sample in vial Improper sample vial position Incorrect or improper settings Leaking or worn seals Valve failure No or low air pressure Wrong syringe size

    Detector Defective source lamp Incorrect or improper settings (sensitivity,

    attenuation, wavelength, time constant)

    Insufficient time for lamp to stabilize Auto zero left enabled Incorrect optical filter and/or lamp Dirty flow cell Air bubble in flow cell Dirty reference electrode Dirty working electrode Incorrect reference and sample solvent

    balance (RI detector) Leaking flow cell

    Computer Wrong A/D sampling time Improper output voltage signal from detector Improper attenuation, peak width, area reject,

    noise rejection parameters Poor peak integration Incorrect calibration or sample table

    Table 2-1 System Problems (Continued)

    Component Areas to Check

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    28/214

    28 Troubleshooting an LC System

    2

    Integrator Improper attenuation Improper input voltage range Auto zero left enabled by detector Improper output voltage signal from detector Improperly inserted paper

    Chart recorder Improper chart speed Improper input voltage range Improper output signal from detector Gain improperly adjusted Reversed polarity Auto zero left enabled on detector

    Auto zero left enabled on chart recorderChromatography Operating environment

    Column not equilibrated Mobile phase problem (improperly degassed,

    prepared, or filtered; incorrect for application;contaminated)

    Sample problem (improperly prepared orfiltered; insoluble with mobile phase;degraded)

    Contaminated column Obstruction in guard column or column inlet

    frit Improper column or precolumn

    Table 2-1 System Problems (Continued)

    Component Areas to Check

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    29/214

    Isolating System Problems 29

    2

    2.2 Isolating System Problems

    This section covers the most common symptoms exhibited during operation, including:

    Note: If a specific instrument or the column in your system exhibits a symptom notaddressed in this section, refer to Chapter 3, Troubleshooting System Components , forinstrumentation and column troubleshooting.

    2.2.1 System PressureTroubleshooting overview

    This section covers troubleshooting:

    High system pressure

    Low system pressure Erratic (fluctuating) system pressure

    System pressure troubleshooting is presented in a flow diagram format. Use Figure 2-2 through Figure 2-4 to investigate pressure problem sources.

    Exhibited symptom Refer to

    System pressure(high, no/low, or erratic

    Section 2.2.1

    Baseline noise Section 2.2.2

    Changes to chromatographic resolutionand results

    Incorrect retention time

    No peaks or abnormal peaks

    Incorrect qualitative/quantitative

    results Loss of resolution

    Section 2.2.3

    Section 2.2.3.1Section 2.2.3.2Section 2.2.3.3

    Section 4.3

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    30/214

    30 Troubleshooting an LC System

    2

    System pressure reference point

    To identify a pressure change from normal operation, it is critical that you have a pressurereference point. System pressure is affected by the column, flow rate, mobile phase, andtemperature, and can vary greatly with different methods.When running a gradient,fluctuations in system pressure may be due to viscosity changes between solvents.

    Each time you install a new column or start a new method, equilibrate the system andrecord the system pressure (both with and without the column in line) to use as acomparison. Use the normal operation log to record system pressure.

    Gradual versus sudden pressure increase

    When high system pressure occurs, it is important to note whether the pressure increasewas gradual or sudden.This can help you isolate the problem source.

    If the pressure has risen gradually (over a series of injections), it may be due to:

    Particulates in the sample or the mobile phase which have accumulated in thein-line filter or column frits

    Debris from failed fluid seals

    If the pressure has risen suddenly , it may be due to:

    Particulates in one sample

    A system hardware problem (such as blocked tubing)

    Collapse of the column packed bed

    High pressure locations within an LC system

    To isolate the origin of high system pressure, loosen inlet or outlet fittings as instructed inFigure 2-2 and observe if pressure stays the same or reduces.

    Attention: Always observe safe laboratory practices when troubleshooting. Always wearsafety glasses and gloves. Know the chemical and physical properties of the solvents.Refer to the Material Safety Data Sheet for the solvents in use.

    Figure 2-1 presents pressure release points in a typical LC system. Use Figure 2-1 as areference when reviewing the high pressure troubleshooting flow diagram in Figure 2-2 .

    To relieve pressure, slowly loosen the fitting. Use a tissue to prevent solvent spray and alsoto collect any spilled solvent.

    STOP

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    31/214

    Isolating System Problems 31

    2

    Attention: f the source of high pressure is within the detector flow cell, use extremecaution when relieving high pressure buildup. Many detectors (especially RI andfluorescence) have fragile flow cells. Before backflushing the detector to remove the

    blockage, review the flow rate specifications and backpressure limits for that flow cell in thedetector operator's manual.

    Figure 2-1 Pressure Release Points within a Typical LC System

    STOP

    Single pumpgradient system

    Solvent reservoirsA, B, C, D

    Gradientmixer

    (optional)

    n ec or(manual or

    autoinjector)Outlet

    OutletInlet

    OutletInlet

    Column

    Detector

    Wastereservoir

    In-linefilter

    or guardcolumn

    (optional)

    Outlet Inlet

    Outlet Inlet

    Multiple pump

    Inlet

    Outlet

    OutletInlet

    Gradientmixer

    (optional)

    Pump A Outlet

    Solvent reservoirA

    Solvent reservoirB

    Pump B

    OutletPumpOption 2

    Option 1

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    32/214

    32 Troubleshooting an LC System

    2

    Figure 2-2 High System Pressure Troubleshooting

    System pressureproblem?

    Are highly viscoussolvents being used?

    YesHas ambienttemperature changed?

    Is pump flow rate set

    too high?

    Figure 2-2 continuedon next page.

    Check solvent viscosity (seeAppendix A, Reference

    Information ). The observedpressure may be normal for this

    column and solvent blend. Ifnecessary, change to a less

    viscous solvent.

    Set to correct flow rate.

    Check the operating temperature.If temperature is too low, stabilize

    operating environment.

    High, no/low, orerratic?

    Go toFigure 2-3

    Go toFigure 2-4

    Yes

    Yes

    No

    No

    Erratic

    High

    No/Low

    No

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    33/214

    Isolating System Problems 33

    2

    Figure 2-2 High System Pressure Troubleshooting (Continued)

    Yes

    No

    Is pressuretransducer operating

    correctly?

    Loosen fitting atdetector waste outlet.

    Pressure reduced?

    Loosen fitting at

    detector inlet.Pressure reduced?

    Figure 2-2 continuedfrom previous page.

    Loosen fitting atcolumn inlet.

    Pressure reduced

    Loosen fitting at guard

    column or in-line filter outlet.Pressure reduced?

    Loosen fitting atcolumn outlet.

    Pressure reduced?

    Figure 2-2 continuedon next page.

    Loosen fitting at

    transduceroutlet. Set flow

    rate to zero.Does pressurefall to zero? (Ifyes, retighten

    fitting).

    Adjusttransducer. Can

    pressure bezeroed?

    Defectivetransducer.Replace per

    pump operatorsmanual.

    Blocked outlet connectingtubing. Clean or replace tubing

    per Chapter 4 and detectoroperators manual.

    Blocked fluid path (flow cell).

    Clean blockages per detectoroperators manual.

    Blocked outlet connecting tubing.Clean or replace tubing per

    Chapter 4 and columnoperators manual.

    Blocked or voided column.Clean column per column

    operators manual. If problempersists, replace the column.

    Blocked outlet connecting tubing.Clean or replace tubing perChapter 4 and column/filter

    operators manual.

    Yes

    Yes

    Yes

    Yes

    No

    No

    No

    Yes

    No No No

    No

    Yes Yes

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    34/214

    34 Troubleshooting an LC System

    2

    Figure 2-2 High System Pressure Troubleshooting (Continued)

    Yes

    No

    Loosen fitting at guardcolumn or in-line filter

    inlet. Pressure reduced?

    Loosen fitting atinjector outlet.

    Pressure reduced?

    Loosen fitting at mixerinlet. Pressure

    reduced?

    Loosen fitting at pumpoutlet. Pressure

    reduced?

    Loosen fitting atinjector inlet. Pressure

    reduced?

    Pump problem, callCustomer Support.

    Blocked in-line filter or guardcolumn. Replace components.

    Blocked outlet connectingtubing. Clean or replace tubing

    per Chapter 4 and injectoroperators manual.

    Blockage within injector orconnecting tubing. Clean

    blockages per injector operatorsmanual. Verify sample solubility.

    Blockage within connecting tubingor mixer. Clean or replace tubing

    per Chapter 4 . If mixer problem,see operators manual.

    Blocked outlet connecting tubing.Clean or replace tubing per

    Chapter 4 and pump operatorsmanual. Verify solvent miscibility.

    Yes

    Yes

    Yes

    Yes

    No

    No

    No

    Figure 2-2 continuedfrom previous page.

    No

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    35/214

    Isolating System Problems 35

    2

    Figure 2-3 Low System Pressure Troubleshooting

    No or low systempressure

    Figure 2-3 continuedon next page

    Check fuse, replace ifnecessary.

    Yes

    NoIs pump fuse blown?

    Turn pump on.

    Yes

    NoIs pump on (are fansrunning)?

    Is there solvent in thereservoir?

    Yes

    NoIs there solvent flowingfrom the detector

    outlet?Refill reservoir.

    Is the low pressure limitset to a value higher

    than operating pressure?

    Can solvent be drawneasily from draw-off

    valve?

    Has the pump beenproperly primed?

    Set correctpressure limit.

    Plugged in-line filter.Clean filter. Replace if

    necessary.

    Prime the pump perpump operators

    manual.

    Yes

    Yes

    No

    Yes

    No

    Yes

    No

    No

    Figure 2-3 continuedon next page

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    36/214

    36 Troubleshooting an LC System

    2

    Figure 2-3 Low System Pressure Troubleshooting (Continued)

    Figure 2-3 continued

    from previous page

    Figure 2-3 continuedon next page

    Is there air in thesolvent inlet lines?

    Are both pump headsworking? Place finger onindicator rod (if present)

    to check.

    Set correct flow rate.

    No

    YesHas flow rate been set

    too low?

    Yes

    Figure 2-3 continued

    from previous page

    Remove air from eachinlet line.

    Pump problem. Referto Section 3.1, Pump

    Troubleshooting .

    Check operatingtemperature. If too high,stabilize environment.

    No

    YesHas ambienttemperature changed?

    Tighten any loosefittings (not overtight).

    No

    YesAre there any visibleleaks?

    Verify solvent used. Ifnecessary, change

    solvent.

    Yes

    NoIs correct solvent

    being used?

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    37/214

    Isolating System Problems 37

    2

    Figure 2-3 Low System Pressure Troubleshooting (Continued)

    Figure 2-3 continuedfrom previous page

    Call CustomerSupport

    Autoinjector problem.Refer to autoinjectoroperators manual.

    No

    YesIs autoinjector stuck inpurge?

    Measure flow rate atdetector outlet with a

    graduated cylinder anda stop watch. Is solventflow rate much differentthan the pump setting?Yes

    NoIs the correct flow rateactually being deliveredthrough fluid system?

    Pump problem. Referto Section 3.1, Pump

    Troubleshooting .

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    38/214

    38 Troubleshooting an LC System

    2

    Figure 2-4 Erratic System Pressure Troubleshooting

    Yes

    Is pressure transduceroperating properly?

    Has the pump beenproperly primed?

    Erratic systempressure.

    Call CustomerSupport.

    Set flow rate tozero and checktransducer. Is

    pressure stable atzero flow?

    Adjust transducer.Can pressure be

    zeroed?

    Defectivetransducer.Replace per

    pump operatorsmanual.

    Prime the pump per thepump operators manual. If

    pressure erratic, refer toSection 3.1, PumpTroubleshooting .

    No

    Yes

    No No No

    Yes

    Do pressure changesreproducibly follow the

    gradient change?

    No

    Are you using volatilesolvents (such as

    ethers)?

    Pump problem. Refer toSection 3.1, PumpTroubleshooting .

    Check solvent miscibility.Change solvents (see

    Appendix A, ReferenceInformation ).

    Yes

    No

    Yes

    No

    Yes

    No

    No

    Degas/helium sparge

    solvent(s). See Chapter 4 .

    Degas/sparge solvents (seeChapter 4 ).

    Are miscible solventsbeing used?

    Is the mobile phaseproperly sparged and

    degassed?

    Are both pump headsworking? Place finger on

    indicator rod (ifappropriate) to verify.

    Are you running agradient?

    Pressure fluctuations arenormal due to viscosity

    changes between solvents.

    No

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    39/214

    Isolating System Problems 39

    2

    2.2.2 Baseline NoiseTroubleshooting overview

    This section covers troubleshooting baseline noise. It assumes baseline noise is eitherfluid path-related (mobile phase, pump, column) or detector electronics- related.

    Baseline noise is characterized as:

    Non-cyclic (erratic) baseline noise

    Cycling (short or long-term) baseline noise

    Baseline drift

    Noise spikes on the baseline

    Table 2-3 covers fluid path-related noise and Table 2-4 covers detector electronics-relatednoise.

    Note: Data-handling device electronics noise is covered in Section 3.6 .

    Table 2-2 summarizes the baseline noise symptoms used in Table 2-3 andTable 2-4 . From Table 2-2 :

    1. Review the baseline noise symptoms and select the one that best typifies yoursystem problem.

    2. Proceed to the page number of the baseline noise symptom (in Table 2-3 orTable 2-4 ).

    3. Review the list of possible causes and follow the corrective actions.

    Isolating the source of baseline noise

    To isolate the source of the baseline noise:

    1. Turn off your pump to stop solvent flow to the system.

    2. Monitor the baseline for a few minutes. Note the following:

    If there is significant improvement in the baseline, the problem is within the fluidpath (pump/mobile phase/flow path/column). Refer to Table 2-3 .

    Note: Some flow sensitive detectors, such as RI and electrochemical, mayrequire a significant time to stabilize once flow stops.

    If the noise continues, the problem is within the detector or its electricalconnections. Proceed to step 3.

    3. Disconnect the detector electrical cables from the data-handling device (A/Dinterface to the computer, computer, integrator, or chart recorder).

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    40/214

    40 Troubleshooting an LC System

    2

    4. Attach a jumper source to the input terminals on the data-handling device (suchas a wire or paperclip).

    If the noise continues, the problem is within the data-handling device. Refer to

    Section 3.6, Data-Handling Device Troubleshooting . If the noise stops, the problem is within the detector or its electrical

    connections. Refer to Table 2-4 .

    Baseline noise summary

    Table 2-2 is a summary of the baseline noise symptoms listed in Table 2-3 and Table 2-4 . These baseline noise examples were generated using the following operating conditions:

    Note: The amplitude of the noise shown in the examples in this section have beengenerated in a lab environment. The signal amplitude can vary depending upon theseverity of the problem.

    Parameter Setting

    Pressure 1000 psi

    Detector absorbanceunits

    1 AU

    Recorder sensitivity 10 mV full-scale

    Chart speed 1 cm/min

    Output signal fromdetector

    10 mV

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    41/214

    Isolating System Problems 41

    2

    Table 2-2 Baseline Noise Summary

    Baseline Noise Symptom Refer to

    Non-cyclic (erratic) noiseFluid path problem: Table 2-3

    Detector electronics problem: Table 2-4

    Short-term cycling noise

    Fluid path problem Table 2-3

    Air bubble in flow cell

    Unstabilized system

    Time

    A U o r m

    V

    Time

    A U o r m

    V

    Leak in detector flow cellTime

    A U o r m

    V

    Defective lamp Operating at high sensitivityTime

    A U o r m

    V

    Time

    A U o r m

    V

    Inadequate solvent blending

    Air bubbles indetector flow cell

    Time

    A U

    o r m

    V

    Time

    A U o r m

    V

    Time

    A U o r m

    V

    Inadequate solvent blending

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    42/214

    42 Troubleshooting an LC System

    2

    Short-term cycling noise(Continued)

    Detector electronics problem:

    Table 2-4

    Long-term cycling noise

    Fluid path problem: Table 2-3

    Detector electronics problem: Table 2-4

    Table 2-2 Baseline Noise Summary (Continued)

    Baseline Noise Symptom Refer to

    Cycling equipment

    Time

    A U o r m

    V

    Temperature fluctuations

    A U o r m

    V

    Time

    Temperature fluctuations

    A U o r m

    V

    Time

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    43/214

    Isolating System Problems 43

    2

    Baseline driftFluid path problem: Table 2-3

    Detector electronics problem Table 2-4

    Noise spikesFluid path problem: Table 2-3

    Table 2-2 Baseline Noise Summary (Continued)

    Baseline Noise Symptom Refer to

    Contaminated column

    A U o r m

    V

    Time

    Insufficient stabilization

    Time

    A U o r m

    V

    Air bubbles in fluid path

    Time

    A U o r m

    V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    44/214

    44 Troubleshooting an LC System

    2

    Detector electronics problem: Table 2-4

    Table 2-2 Baseline Noise Summary (Continued)

    Baseline Noise Symptom Refer to

    Defective lamp

    Time

    A U o r m

    V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    45/214

    Isolating System Problems 45

    2

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting

    Symptom Possible Cause Corrective Action

    Non-cyclic (erratic) noise(originating from fluid path)

    Large air bubble trappedin detector flow cell

    To remove the air bubble,purge the detector flow cell orapply slight pressure on thedetector waster outlet (seedetector operators manual).Note: To prevent air bubblesfrom forming in the flow cell,add a 1 to 3 foot (30 to 90 cm)length of 0.009 -inch (0.23mm) ID, 1/16 -inch (1.58 mm)ID, 1/16 -inch (1.58 mm) ODtubing to the detector wasteoutlet. This tubing functionsas a flow restrictor to increasebackpressure. A 3 foot (90cm) piece of tubing provides30 to 50 psi (2 to 3 atm) ofbackpressure at 1 mL/min inwater.

    Keep in mind thebackpressure limits of the flowcell (with RI, fluorescence,

    conductivity, andelectrochemical detectors)before attaching this tubing.

    Air bubble trapped inreference electrode(electrochemical detectoronly)

    Remove the referenceelectrode and gently shake itto remove the air bubble.Replace the referenceelectrode.

    Normal baseline

    Air bubble in flow cell

    Unstabilized system

    Time

    A U o r m V

    Time

    A U o r m V

    Time

    A U o r m V

    Leak in detector flow cellTime

    A U o r m V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    46/214

    46 Troubleshooting an LC System

    2

    Non-cyclic (erratic) noise(originating from fluid path)(Continued)

    Small air bubblestraveling through the flowpath

    Prime the pump to remove air(see pump operators manual).To prevent additional airbubbles, ensure the mobilephase is properly degassed orhelium sparged (seeChapter 4 for information).

    System not stabilized orchemically equilibrated

    Allow all system components(such as the column anddetector) sufficient time tostabilize and chemicallyequilibrate. Note the operatingconditions of your application(such as mobile phase,detector settings, detectortype). Refer to the instrumentor column operators manualfor recommended equilibrationtimes.If running an automatedgradient method, ensure

    sufficient and reproducibleequilibration times are usedbetween injections.Note: If using ion-pairingreagents, ensure that the firsttime you use the column youprovide a sufficient time andvolume of solvent to

    adequately equilibrate thecolumn (for example, runninga total volume of 100 mL of a5 mM solution at 1 ml/min).

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    47/214

    Isolating System Problems 47

    2

    Non-cyclic (erratic) noise(originating from fluid path)(Continued)

    Mobile phasecontaminated

    Discard the contaminatedmobile phase and:

    Clean the solventreservoir. Clean orreplace the solvent inletfilter. To clean the fil ter,

    remove and sonicateusing 6N nitric acid,water (repeat 3 times),followed by methanol.

    Prepare and filter freshsolvent using only highquality reagents andHPLC-grade solvents(see Section 4.1 forsolvent preparation anduse considerations)

    Flush and re-equilibratethe system.

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    48/214

    48 Troubleshooting an LC System

    2

    Non-cyclic (erratic) noise(originating from fluid path)(Continued)

    Detector flow cell leaking Remove the detector coverand check for leaks. If leaksare not visible, perform thefollowing:1. Flush the detector with a

    non-buffered misciblesolvent, followed bymethanol.

    2. Attach a nitrogen or heliumsource to the detector inlet.Slowly blow the flow celldry.

    3. Monitor the baseline for

    noise. If noise disappears,there is a leak within thedetector flow cell.

    Repair/replace the detectorflow cell (see detectoroperators manual)

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    49/214

    Isolating System Problems 49

    2

    Non-cyclic (erratic) noise(originating from fluid path)(Continued)

    Column contaminated To verify this problem as apossible cause, replace allcolumns in the system with aunion (or with a known goodcolumn of the same type).Return the mobile phase, andmonitor the baseline.Note: If you require additionalsystem backpressure duringthis verification, use a piece of0.009 -inch (0.23 mm) IDtubing (or suitable restrictor)instead of a union.

    If the problem stops, clean or

    replace the contaminatedcolumn as outlined in thecolumn operators manual.If the problem continues, thiscould be due to:

    Solvent properties suchas miscibility. Refer to

    Section 4.1 andAppendix A, ReferenceInformation forinformation.

    Contaminated mobilephase (refer to theMobile phasecontaminated PossibleCause listed above forthe corrective action

    Contaminated guardcolumn in-line filter(clean or replace asoutlined in the operatorsmanual).

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    50/214

    50 Troubleshooting an LC System

    2

    Short-term cycling noise(seconds to minutes)(originating from fluid path)

    Erratic pump pressure/pump pulsations

    Refer to Figure 2-4 (inSection 2.2.1 ) to verify andcorrect the source of erraticpressure. If erratic pressurecontinues, refer to Section 3.1,Pump Troubleshooting .

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Normal baseline

    Time

    A U o r m

    V

    Inadequate solvent blending

    Air bubbles in detectorflow cell

    Time

    A U o r m

    V

    Time

    A U

    o r m

    V

    Time

    A U o r m

    V

    Inadequate solvent blending

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    51/214

    Isolating System Problems 51

    2

    Short-term cycling noise(seconds to minutes)(originating from fluid path)

    Inadequate solventbending

    To confirm a mixing problem:1. With the column in-line,

    pump 5-10 columnvolumes of 100% A andmonitor the baseline. Thisprovides both a constant

    composition and sufficientvolume to equilibrate thecolumn and reach thedetector flow cell.

    2. Pump several pre-mixedsolvents (for example,50/50, 95/5, 70/30, or the

    mixture you are running)and monitor the baseline.

    If the baseline is acceptablewith 100% A, but becomesnoisier when running mixtures,there is a mixing problem.Verify the mixing problem andcorrect as outlined below:1. Use of immiscible solvents.

    Verify miscibility of solvents(see Appendix A) andchange to more misciblesolvents.

    2. Malfunction in the pump,the pump solventproportioning valve, or thehigh-pressure mixer.Troubleshoot perSection 3.1, PumpTroubleshooting

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    52/214

    52 Troubleshooting an LC System

    2

    Short-term cycling noise(seconds to minutes)(originating from fluid path)(Continued)

    Inadequate solventbending (continued)

    3. Inadequate blending afterthe pump. The solution isto add additional mixing.However, the mixingrequired depends on theseverity of the problem.Verify and correct asfollows:

    Increase laminar flowmixing and residencetime (the time to passthrough the mixer). Adda 6 to 12 inch (150 to300 mm) length of0.040-inch (1.0 mm) IDtubing between thepump outlet and theinjector. This tubinglength adds only a smallsystem delay. volumeand does not distortgradient shape.

    If the tubing does notresolve the problem, adda larger mixer volume fora more vigorous mixing.Add one or more WatersGradient Flow Mixers

    between the pump outletand injector. The mixingchamber providesconsistent solventblending with minimalvolume and gradientdistortion.

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    53/214

    Isolating System Problems 53

    2

    Short-term cycling noise(seconds to minutes)(originating from fluid path)(Continued)

    Inadequate solventbending (continued)

    Note: Ensure that you use amixing chamber with thesmallest possible volume.This prevents introducingadditional delay volume ordistorting gradient shape.

    4. Use premixed solvent(s).

    Pump Inlet tubing loose,bent, or blocked

    Check the tubing. If loose,tighten. If bent, straighten. Ifblocked, replace (seeChapter 4, GoodChromatography/ OperatingPractices ).

    Large air bubble indetector flow cell

    Refer to the Corrective Actionunder the Non-cyclic (erratic)noise symptom listed above.

    Dirty or malfunctioningpump inlet check valve

    Refer to Section 3.1, PumpTroubleshooting

    Worn pump plunger seal Refer to Section 3.1, PumpTroubleshooting

    Solvent from detectoroutlet dripping into wastecontainer (primarilyaffecting flow sensitivedetectors such as RI andelectrochemical)

    Position detector outletagainst the side next to thewaste container.

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    54/214

    54 Troubleshooting an LC System

    2

    Long-term cycling noise(minutes to hours)(originating from fluid path)

    Ambient temperaturefluctuations

    Stabilize operatingenvironment temperature. Ifthe problem continues:

    Use a column heater(run 5 Celsius aboveambient).

    Relocate the system orcolumn to a thermallystable environment.

    Avoid placing thesystem in directsunlight.

    Solvent is being recycledfrom detector waste outletback through LC system

    Unless absolutely necessary,do not recycle solvent throughyour LC system. Only usefresh and filtered solvent foryour application.

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Normal baselineTime

    A U o r m

    V

    Temperature fluctuations

    A U o r m

    V

    Time

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    55/214

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    56/214

    56 Troubleshooting an LC System

    2

    Baseline drift(originating from fluid path)(Continued)

    Detector flow cell leaking Refer to Corrective Actionunder the Non-cyclic (erratic)noise symptom listed above.

    Contaminated column Refer to the Corrective Actionunder the Non-cyclic (erratic)noise symptom listed above.

    Solvent is being recycledfrom detector waste outletback through LC system

    Unless absolutely necessary,do not recycle solvent throughyour LC system. Only usefresh and filtered solvent foryour application.

    Leak(s) in system Check all fitting for leaks. If

    there is a leaky fitting, tightenit (do not overtighten).If the leak continues, replacethe fitting and ferrule (seeSection 4.2 forconsiderations).

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    57/214

    Isolating System Problems 57

    2

    Baseline drift(originating from fluid path)(Continued)

    Stationary phase bleed To verify this problem, replaceall columns in the system witha union. Rerun the mobilephase and monitor thebaseline.Note: If you require additionalsystem backpressure duringthis verification, use a piece of0.009 -inch (0.23 mm) IDtubing instead of a union.

    Ensure your operatingconditions are suitable for thecolumn (for example, solventcompatibility, pH range, and so

    on). If the operating conditionsaffect the column: Select a different mobile

    phase Select a different

    columnRefer to the column operators

    manual for information.Incorrect wavelength forsolvent

    Verify the backgroundabsorbance of the mobilephase using a spectrometer. Ifthe background is high (youare unable to zero the detectorbaseline), the mobile phase

    contains a UV-absorbingcompound, which causesbaseline drift.Operate at a wavelength thatis above the UV cutoff for themobile phase (see Table A-3 )or change the solvent(s).

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    58/214

    58 Troubleshooting an LC System

    2

    Baseline drift(originating from fluid path)(Continued)

    Mobile phase contains astabilizer or there is achange in the stabilizer

    Use a preservative-freesolvent(s).Note: Separations mayrequire adjustment ifchanging to apreservative-free solvent.

    Late-eluting samplecomponent

    Wash the column with anappropriate strong solvent.

    Unbalanced solvents(gradient operation)

    Attempt to balance the UVabsorbance with the mobilephase. If constrained bychemistry, run a gradient blankand subtract the baseline.

    Noise spikes(originating from fluid path)

    Small air bubblestraveling through the fluidpath

    Refer to the Corrective Actionunder the Non-cyclic (erratic)noise symptom listed above.

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Normal baselineTime

    A U o r m

    V

    Air bubbles influid path

    Time

    A U o r m

    V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    59/214

    Isolating System Problems 59

    2

    Noise spikes(originating from fluid path)(Continued)

    Pump head cavitation Refer to Section 3.1, PumpTroubleshooting .

    Particles in detector flowcell

    Clean or backflush thedetector flow cell (asdescribed in the detectoroperators manual).

    Improper grounding ofpump or autoinjectorelectrical connections

    Use a shielded signal cableand attach the shield to onedevice only .If shielding is not the problem,plug the pump or autoinjector

    into another outlet on adifferent electrical circuit, If aseparate outlet is unavailable,use a line power conditioner.

    Table 2-3 Fluid Path-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    60/214

    60 Troubleshooting an LC System

    2

    Table 2-4 Detector Electronics-Related Baseline Noise Troubleshooting

    Symptom Possible Cause Corrective Action

    Non-cyclic (erratic) noise(originating from detector)

    Detector not stabilized Allow detector lampsufficient time to stabilize(until baseline is stable).Detector equilibration timevaries based on the type ofdetector used and operatingparameters (such as

    wavelength, sensitivity,background, potential,and/or current). Refer to thedetector operators manualfor recommendedequilibration times.

    Normal baselineTime

    A U o r m

    V

    Defective lamp

    Operating at highsensitivity

    Time

    A U o r m

    V

    Time

    A U o r m

    V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    61/214

    Isolating System Problems 61

    2

    Non-cyclic (erratic) noise(originating from detector)(Continued)

    Defective detector lamp Verify lamp energy usingdetector diagnostics. Ifenergy is belowspecification (as comparedwhen the lamp was new),replace the lamp.Note: Some detectorsallow you to adjust lampenergy to compensate fordecreased energy. Refer toyour detector operatorsmanual for information onadjusting lamp energy.

    Contaminated detector flowcell

    Refer to Section 3.5,Detector Troubleshooting .

    Detector electronicsproblem

    Detector malfunction.Contact Customer Support.

    Cable loose or improperlyconnected betweendetector and data-handling

    system (computer,integrator, or chart recorder)

    Verify that the correctdetector output signal isproperly connected to the

    data-handling device.Ensure that any relatedoutput signal switch settingsare in the proper position.Refer to the detectorsoperators manual anddata-handling deviceoperators manual.

    Table 2-4 Detector Electronics-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    62/214

    62 Troubleshooting an LC System

    2

    Non-cyclic (erratic) noise(originating from detector)(Continued)

    Detector improperlygrounded

    Use a shielded signal cableand attach the shield to onedevice only .If shielding is not theproblem, plug the detectorinto another outlet on adifferent electrical circuit. Ifa separate outlet isunavailable, use a linepower conditioner.

    Cycling equipment or radiofrequency (RF) interference

    Isolate the detector fromother equipment in the lab,especially devices withlarge electric motors. Then:

    Check circuitgrounding and linevoltage quality (referto the Detectorimproperly groundedPossible cause abovefor the corrective

    action). Ensure that the

    detector and thedata-handling deviceare on the samecommon ground.Relocate if necessary.

    If necessary, relocatethe detector to anarea where RF is nota problem or use aFaraday cage aroundthe detector.

    Table 2-4 Detector Electronics-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    63/214

    Isolating System Problems 63

    2

    Non-cyclic (erratic) noise(originating from detector)(Continued)

    Data-handling devicegain/sensitivity setting toohigh

    Change to a lowergain/sensitivity setting (seedata-handling deviceoperators manual).

    Leak from referenceelectrode (electrochemical

    detector only)

    Repair or replace thereference electrode (see

    detector operators manual).

    Foulded reference electrode(electrochemical detectoronly)

    Renew reference electrodefilling solution and/orreplace the frit (see detectoroperators manual).

    Contaminated or scratched

    working electrode(electrochemical detectoronly)

    Clean or polish working

    electrode (see detectoroperators manual). Ifproblem persists, replacethe working electrode.

    Short-term cycling noise(originating from detector)(seconds to minutes)

    Cycling equipment or radiofrequency (RF) interference

    Refer to the CorrectiveAction under the Non-cyclic(erratic) noise symptom

    listed above.

    Table 2-4 Detector Electronics-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Normal baselineTime

    A U o r m

    V

    Cycling equipment

    Time

    A U o r m

    V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    64/214

    64 Troubleshooting an LC System

    2

    Short-term cycling noise(originating from detector)(seconds to minutes)(Continued)

    Internal detectortemperature improperly set(heater cycles on and offtoo frequently)

    Correctly set the detectorinternal heater (seedetector operators manual).

    Long-term cycling noise(originating from detector)

    (minutes to hours)

    Ambient temperaturefluctuations

    Stabilize operatingenvironment temperature to

    allow full equilibration. Ifproblem continues:

    Relocate the detectorto a thermally stableenvironment or closeany open air vents.

    Avoid placing the

    system in directsunlight.

    Cycling equipment or radiofrequency (RF) interference

    Refer to the CorrectiveAction under the Non-cyclic(erratic) noise symptomlisted above.

    Table 2-4 Detector Electronics-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Normal baseline

    Time

    A U o r m

    V

    Temperature fluctuations

    A U o r m

    V

    Time

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    65/214

    Isolating System Problems 65

    2

    Baseline drift(originating from detector)

    Detector not stabilized Refer to the CorrectiveAction under the Non-cyclic(erratic) noise symptomlisted above.

    Change in ambient

    temperature

    Refer to the Corrective

    Action under the Long-termcycling noise symptomlisted above.

    Contaminated detector flowcell

    Refer to Section 3.5,Detector Troubleshooting .

    Fouled reference electrode(electrochemical detectoronly)

    Renew reference electrodefilling solution (see detectoroperators manual). Ifproblem persists, replacethe frit.

    Table 2-4 Detector Electronics-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Normal baselineTime

    A U o r m

    V

    Insufficient stabilization

    Time

    A U o r m

    V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    66/214

    66 Troubleshooting an LC System

    2

    Contaminated or scratchedworking electrode(electrochemical detectoronly)

    Clean or polish workingelectrode (see detectoroperators manual). Ifproblem persists, replacethe working electrode.

    Noise spikes

    (originating from detector)

    Defective detector lamp Refer to the Corrective

    Action under the Non-cyclic(erratic) noise symptomlisted above.

    Cycling equipment or radiofrequency (RF) interference Refer to the CorrectiveAction under the Non-cyclic(erratic) noise symptomlisted above.

    Detector electronicsproblem

    Detector malfunction.Contact your CustomerSupport Department.

    Table 2-4 Detector Electronics-Related Baseline Noise Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Normal baselineTime

    A U o r m

    V

    Defective lampTime

    A U o r m

    V

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    67/214

    Isolating System Problems 67

    2

    2.2.3 Changes in Chromatographic Resolution or ResultsChromatographic problems typically exhibit themselves as:

    Incorrect or changing retention time(s)

    No peaks or abnormal peak shape(s)

    Loss of resolution

    Incorrect qualitative or quantitative results

    When your chromatographic results are unacceptable (as compared with establishedperformance), reinject your application standard(s) and translate your observations into

    specific problems with your system.For example, problems that appear to be due to abnormal peak shape can be due to otherissues with chromatographic performance. Before investigating solutions to abnormalpeak shape, verify that incorrect retention time is not affecting your chromatography.

    Troubleshooting unacceptable chromatographic results involves two stages:

    Sequentially evaluate your chromatographic results (retention time, peak shape,

    resolution, and qualitative/quantitative results) Isolate the source of the problem

    The flow diagram in Figure 2-5 presents questions to help you narrow down your problem.Once you isolate the problem, refer to the appropriate troubleshooting section.

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    68/214

    68 Troubleshooting an LC System

    2

    Figure 2-5 Isolating the Source of Changing Chromatographic Resolution/Results

    Yes

    Are retection timescorrect?

    Has peak shapechanged?

    Are the peaks properlyidentified and theresults correct?

    Has resolutionchanged?

    Troubleshoot incorrect/changingretention times. Proceed to

    Section 2.2.3.1 .

    Troubleshoot abnormal peakshapes. Proceed to

    Section 2.2.3.2

    Troubleshoot resolution. Proceedto Section 4.3 .

    Troubleshoot qualitative and/orquantitative results. Proceed to

    Section 2.2.3.3

    No

    Yes

    Yes

    No

    No

    No

    Changes inchromatographic

    resolution or results.

    Yes

    Chromatographyis acceptable.

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    69/214

    Isolating System Problems 69

    2

    2.2.3.1 Incorrect/Changing Retention TimesTroubleshooting overview

    Incorrect retention time is characterized as: Erratic (changing back and forth from run to run)

    Steadily increasing

    Steadily decreasing

    Changed to a constant new value (incorrect but reproducible)

    Table 2-5 covers incorrect/changing retention time. From the table:

    1. Review the retention time symptoms and select the symptom that best typifies yourproblem.

    2. Review the possible causes and follow the corrective actions.

    Retention time stability factors

    Retention time stability is affected by:

    System and column equilibration Mobile phase (composition, preparation, improper degassing, stability)

    Column age

    Operating temperature

    Pump performance (flow rate, pressure, loss of prime)

    Isolating retention time changes

    When evaluating retention times changes, it is critical that you have previously establishedsystem performance benchmarks. From those benchmarks, you determine whetherretention times:

    Erratically change in a different direction (either increasing or decreasing) from runto run.

    Change in the same direction (increasing or decreasing) from run to run.

    Suddenly change to a new value (increased or decreased), which is outside theallowable range for the assay, and then hold at that value for a number of runs.

    T bl 2 5 I /Ch i R i Ti T bl h i

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    70/214

    70 Troubleshooting an LC System

    2

    Table 2-5 Incorrect/Changing Retention Time Troubleshooting

    Symptom Possible Cause Corrective Action

    Erratic retention times(changing from run to run)

    System not stabilized orchemically equilibrated

    Allow all systemcomponents (such as thecolumn and detector)sufficient time to stabilizeand chemically equilibrate.Note the operatingconditions of your

    application (such as mobilephase, detector settings,and detector type). Refer tothe instrument or columnoperator's manual forrecommended equilibrationtimes.If running an automatedgradient method, ensuresufficient and reproducibleequilibration times areused between injections.Note: If using Ion-pairingreagents, ensure that thefirst time you use thecolumn you provide asufficient time and volumeof solvent to adequatelyequilibrate the column (forexample, running a totalvolume of 100 mL of a5 mM solution at1 mL/min).

    T bl 2 5 I t/Ch i R t ti Ti T bl h ti (C ti d)

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    71/214

    Isolating System Problems 71

    2

    Erratic retention times(changing from run to run)(Continued)

    Erratic pumppressure/pump pulsations(due to problems such asair bubbles, worndirty/malfunctioning checkvalves, worn plunger seals,worn plunger)

    Refer to Figure 2-4 (inSection 2.2.1 ) to verify thesource of erratic pressure.If erratic pressurecontinues, refer toSection 3.1, PumpTroubleshooting .

    Injection volume/sampleconcentration too high(sample overload),disrupting equilibrium

    Reduce the injectionvolume or dilute thesample with mobile phase.If using a weaker solvent,you can inject up to 10% ofcolumn void volume. Ifusing a stronger solvent,

    you can inject up to 1% ofcolumn void volume.

    Ambient temperaturefluctuations

    Stabilize operatingenvironment temperatureto allow full equilibration. Ifproblem continues:

    Use a column heater

    (run 5 Celsius aboveambient).

    Ensure the solvent isstirred.

    Relocate the systemor column to athermally stableenvironment or closeany open air vents.

    Table 2-5 Incorrect/Changing Retention Time Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Table 2 5 Incorrect/Changing Retention Time Troubleshooting (Continued)

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    72/214

    72 Troubleshooting an LC System

    2

    Erratic retention times(changing from run to run)(Continued)

    Inadequate solventblending

    To confirm a mixingproblem:1. Premix, filter, and degas

    the mobile phase.

    2. With the column in-line,pump a minimum of

    5 to 10 column volumesthrough a single pump(or solvent line) toequilibrate the column.

    3. Inject a standard aminimum of 3 times andcompare the

    reproducibility of theretention times to theprevious injections (witherratic retention times).

    If retention times arereproducible with thepremixed solvent, thisindicates a solventblending problem.Verify the mixing problemand correct as outlinedbelow:1. Use of immiscible

    solvents. Verifymiscibility of solvents

    (see Appendix A,Reference Information )and change to moremiscible solvents.

    Table 2-5 Incorrect/Changing Retention Time Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Table 2 5 Incorrect/Changing Retention Time Troubleshooting (Continued)

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    73/214

    Isolating System Problems 73

    2

    Erratic retention times(changing from run to run)(Continued)

    Inadequate solventblending

    2. Malfunction in thepump, the pump solventproportioning valve, orthe high-pressure mixer.Troubleshoot perSection 3.1, PumpTroubleshooting .

    3. Inadequate blendingafter the pump. Thesolution is to addadditional mixing.However, the mixingrequired depends onthe severity of theproblem. For theCorrective Action, referto Table 2-3 , the"Short-term cyclingnoise" symptom.

    4. Use premixedsolvent(s).

    Table 2-5 Incorrect/Changing Retention Time Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Table 2-5 Incorrect/Changing Retention Time Troubleshooting (Continued)

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    74/214

    74 Troubleshooting an LC System

    2

    Erratic retention times(changing from run to run)(Continued)

    Column contamination To verify this problem as apossible cause, replace thecolumn with a known goodcolumn of the same type.Rerun the analysis andobserve if retention timesstabilize. If the problem

    stops, clean or replace thecolumn as outlined in thecolumn operator's manual.If the retention timecontinues to be erratic, thiscould be due to:

    Solvent propertiessuch as miscibility.Refer to Section 4.1 and Appendix A,Reference Information for information.

    Contaminated mobilephase (refer to the"Mobile phasecontaminated"Possible Cause laterin this table for thecorrective action).

    Contaminated guardcolumn or in-line filter(clean or replace asoutlined in theoperator's manual).

    Increased or decreasedretention times(continuously changing inthe same direction)

    Pump flow rate changed Verify the solvent flow ratesetting. Set to appropriateflow rate for the application.

    Table 2-5 Incorrect/Changing Retention Time Troubleshooting (Continued)

    Symptom Possible Cause Corrective Action

    Table 2-5 Incorrect/Changing Retention Time Troubleshooting (Continued)

  • 8/9/2019 Guide to Successful Operation of Your LC System.pdf

    75/2