GE Healthcare
Giraffe WarmersFor Neonatal Intensive Care
Helping make sick babies well…The Giraffe® Warmer is designed to meet the needs of the NICU patient and clinician along with the Giraffe Incubator and Giraffe OmniBed.® The revolutionary heater design eliminates a large overhead heater making it easier to observe the patient and perform procedures such as X-ray examinations and even surgery. Silence alarms using the new Hands Free Alarm Silence. A brilliant full color control panel displays patient parameters and trending. The Giraffe Warmer eliminates alarms during Warmup Mode so you can warm the bed before the baby arrives without disturbing other patients, family or clinicians. Integrated resusicitation and SpO2 options support the latest resuscitation guidelines. Weighing your baby can be a difficult procedure, but with our integrated in-bed scale option you can weigh the baby at the bedside and easily chart daily weight trends.
Technical SpecificationsGiraffe Warmer• Height:(adjustable)193-218cm (76-86in)
• Depth:119.4cm(47in)
• Width:63.5cm(25in)
• Weight:Approx.100kg(220lbs)
• Standard2Drawer:6.8kg(15lbs)
• StandardDeepDrawer:6.8kg(15lbs)
• Footprint:119.4x63.5cm(47x25in)
• FloorSpace:7,582sqcm(1,175sqin)
• MattressSize:65x48x4cm (25.5x19x1.5in)
• BedtoFloor(fixedbase):83.8-107.9cm (33-42.5in)with4cm(1.6in) fixed increments
• BedtoFloor(elevating):83.8-107.9cm (33-42.5in)
• BedCapacity:Patientweightupto 13.6kg(30lbs)
• BedTilting:± 12°continuoustilt; Bubble level indicates level bed
Operational EnvironmentGiraffe Warmer
• OperatingTemp.Range:18°Cto30°C
• OperatingHumidityRange:5%to75%
• StorageTemp.Range:-25°Cto+60°C
• StorageHumidityRange:0%to85%
Resuscitation Unit
• OperatingTemp.Range:18°Cto41°C
• OperatingHumidityRange:0%to90%
• StorageTemp.Range:-25°Cto+60°C
• StorageHumidityRange:0%to95%
• StoragePressureRange:50kPato106kPa
Electrical Power Requirements• 115VAC,50/60HzModels: 115V± 10%;6.6Amps
• 220-240VAC,50/60HzModels: 230V±10%;3.3Amps*
©2012GeneralElectricCompany–Allrightsreserved.
General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without noticeorobligation.ContactyourGERepresentativeforthe most current information.
GiraffeisatrademarkofDatex-Ohmeda,Inc.
GE and GE Monogram are trademarks of General Electric Company.
All other company and product names mentioned may be trademarks of the companies with which they are associated.
GE Medical Systems, a General Electric company doing business as GE Healthcare.
• NominalPowerConsumption:520W at max. heater power, max. lighting
• ComplieswithIEC60601-1for electrical safety
Integrated Resuscitation Characteristics• Inputpressure:40-75psi(275-517kPa)
• MinimumInputFlow:70lpm
Venturi Vacuum Subsystem
• Range:0-150mmHg
• Accuracy:+/-5%offullscale
Adjustable PIP (For T-Piece Resuscitation System only)
• MaximumPIP:45±5cmH2O
• Override:>30±4cmH2O
• FlowCapacity:15lpm
Flowmeters
• Range:0-15lpm
• Accuracy: 0-5lpm± ½ lpm 5-10lpm±1.5lpm 10-15lpm± 2 lpm
Air/O2 Blender
• Range:21-100%O2
• Accuracy:+/-5%O2
Airway Pressure Manometer
• DisplayRange:-10to80cmH2O
• Accuracy:+/-5%offullscale
Integrated SpO2 Characteristics
Parameter Masimo SET®
Nellcor Oximax™
MeasurementRange
SpO2 30-100% 1-100%
PulseRate 25-240bpm 20-300bpm
SpO2 Accuracy
From70-100%±3digitsfor
neonates±3digitsfor
neonates
Below69% Unspecified Unspecified
PulseRateAccuracy
No Motion ±3bpm ±3bpm
Motion ±5bpm N/A
Integrated Scale Characteristics• Accuracy:+/-10g(0.35oz)
• Range:300gto8kg(0.66lbsto17.6lbs)
System Control Characteristics• FullColorDisplay:16.5cm(6.5in)
• ObservationLights:2dimmablehalogen lights(35Weach),0-1500luxaverage
•ProcedureLight:50W,2000luxaverage; estimatedlife3000hrsbasedon manufacturer’s rating
Warmup Mode
• Automaticallystartedatpowerup
• Heateroperatesat100%powerfor 7minutes,thenrevertsto50%power
• Checkpatientalarmseliminatedduring Warmup mode
Microprocessor Based Control System
• Self-testfunctionperformedatpoweron and continuously during operation
• HeaterWarmupTimeat100%Power: lessthan3minutes
• HeaterElement:360watts
• HeaterOutput:adjustablefrom0-100% in5%increments
• PatientControl(Baby)Mode:34°Cto37.5°C inincrementsof0.1°C(95°Fto99.5°F)
• ManualMode:Indicatesmanualmode heatselectionrangefrom0to100% in5%increments
• TemperatureMeasurementAccuracy:± 0.3°C Resolution:± 0.1°C
• ProbeAccuracy:± 0.1°C Range:30-42°C(71.6-107.6°F)
* International specification, not available in the USA
GE HealthcareP.O.Box900,FIN-00031GE,FinlandGEDirectUnitedKingdom:+44(0)8000329201 www.gehealthcare.com