GHTF SG2 - Guidance on Reporting of Medical Device Adverse Events N54 - SG2 progress report Dr. Philippe Auclair Abbott EUCOMED Secretary GHTF SG2 "Segundo Foro Internacional y Nacional de Tecnovigilancia" 17 y 18 de junio de 2010, Bogotá, D.C. Colombia

GHTF SG2 Slides for APEC - Invima SG2 - Guidance on Reporting of Medical Device Adverse Events N54 - SG2 progress report Dr. Philippe Auclair –Abbott –EUCOMED Secretary GHTF SG2

  • Upload
    lyminh

  • View
    237

  • Download
    2

Embed Size (px)

Citation preview

GHTF SG2 - Guidance on Reporting of Medical Device Adverse

Events N54 - SG2 progress report

Dr. Philippe Auclair – Abbott – EUCOMED

Secretary GHTF SG2

"Segundo Foro Internacional y Nacional de Tecnovigilancia"17 y 18 de junio de 2010, Bogotá, D.C. Colombia

Post-Market Surveillance Post-Market Surveillance is the

collection of information on the quality, safety or performance of Medical Devices after they have been placed in the market.

A balanced Post-Market Surveillance system will contain an appropriate mix of proactive and reactive activities.

Post-Market Vigilance

(Post-Market) Vigilance is the reporting and investigation of medical device adverse events and incidents. Both the manufacturer and the Regulatory Authority play major roles.

By its very nature, Vigilance is a REACTIVE activity (the manufacturer or authority receive reports and REACT to them) - this is not intended to be a derogatory statement

A Pictorial view of PMS

Post Market

Surveillance

Vigilance(adverse event

report

investigation)

Post-Market Surveillance

Information is used for:

Injury prevention

Development of standards

Regulatory refinement

Product improvement

SG2 Vision for Harmonization

Manufacturer or Authorised Representative

National Regulator (NCA)

Map of SG2 Guidance on AE ReportingGHTF SG2 N54R8

SG2 N54 Sect 8

SG2 N54 Sect 7

SG2 N54 Sect 6

SG2 N54

Sects 3, 4 & 5

SG2 N54

Annex C

To which

NCA(s)?

Within what

timeframe?

What

Information?

Y

?Report?

N

Trend?Document

rationale

for not reporting

N

Y(Report must

be submitted)

National

Regulator

NCA

Report?

Other Participating

National Regulators

Y

SG2 N8 SG2 N79

SG2 N38

Manufacturer’s

Investigation

Adverse

Event

GHTF N54 Section 3.0Three Basic Reporting Criteria

An EVENT must have occurred

AND

The manufacturers device was ASSOCIATEDwith the event

AND

The event led to the death or SERIOUS INJURY of a patient user or other person, ORmight lead to death or serious injury if the event re-occurs

EVENT

Malfunction or deterioration

Inadequate design or manufacture

Inaccuracy in labeling

Significant public health concern

Other information from testing or literature

A change in trend

ASSOCIATION (WITH THE DEVICE)

When the association with the device is difficult to establish, the manufacturer must rely on:

Opinion from healthcare professional

Previous similar events

Other information available to the manufacturer

If there is any doubt, assume that the device was associated with the event.

SERIOUS INJURY

Life threatening illness or injury

Permanent (irreversible) impairment of a body function or permanent damage to a body structure

A condition requiring medical or surgical intervention to prevent permanent impairment of a body function or permanent damage to a body structure

Exemption RulesGHTF N54 Section 4.1- 4.8

1) Deficiency of a new device found by the user prior to its use

2) Adverse event caused by patientconditions

3) Service life or shelf life of the medicaldevice

4) Malfunction protection operated correctly5) Negligible likelihood of occurrence of death or

serious injury6) Expected and foreseeable side effects7) Adverse events described in an advisory notice8) Reporting exemptions granted by NCA

Use Errors reporting N54 Section 5 & Annex D

Use Error: Act, or omission of an act, that has a different result to that intended by the manufacturer or expected by the operator

Use errors related to medical devices, which didresult in death or serious injury or serious public health threat should be reported by the manufacturer to the National Competent Authority

Use errors related to medical devices which did notresult in death or serious injury or serious public health concerns, need not be reported by the manufacturer to the national competent authorities

Abnormal Use reporting N54 Section 5 & Annex D

Abnormal Use:Act, or omission of an act by the operator or user of a medical device as a result of conduct that is beyond any reasonable means of risk control by the manufacturer

Abnormal use need not to be reported by the manufacturer to the national competent authority under adverse event reporting procedure. Abnormal use should be handled by the healthcare facility and appropriate regulatory authorities If manufacturers become aware of instances of abnormal use,

they may bring this to the attention or other appropriate organizations and healthcare facility personnel

The Universe of Device Associated Adverse Events

= Serious Injury

= Death

= Malfunction

/potential Serious injury

RR R

RR NR*

NR*NR*

NR*

R=ReportNR*= No Report*

Manufacturer or Authorised Representative

National Regulator (NCA)

Trends

SG2 N54 Sect 8

SG2 N54 Sect 7

SG2 N54 Sect 6

SG2 N54

Sects 3, 4 & 5

SG2 N54

Annex C

To which

NCA(s)?

Within what

timeframe?

What

Information?

Y

?Report?

N

Trend?Document

rationale

for not reporting

N

Y(Report must

be submitted)

National

Regulator

NCA

Report?

Other Participating

National Regulators

Y

SG2 N8 SG2 N79

SG2 N38

Manufacturer’s

Investigation

Adverse

Event

AE Trend Reporting

Adverse events specifically exempted from reporting become reportable, if there is a change in trend (usually an increase in frequency) or pattern is identified.

The SG2 document on trend reporting describes the criteria for identifying a significant increase in the rate of adverse events.

AE Trend Reporting

Example of an upward shift in trend

time

IB baseline

Incidence of events (i)

IT threshold

RoV* new baseline RoV*

* normal Range of Variance

Report

new threshold

Manufacturer or Authorised Representative

National Regulator (NCA)

To Which NCA’s to Report?

SG2 N54 Sect 8

SG2 N54 Sect 7

SG2 N54 Sect 6

SG2 N54

Sects 3, 4 & 5

SG2 N54

Annex C

To which

NCA(s)?

Within what

timeframe?

What

Information?

Y

?Report?

N

Trend?Document

rationale

for not reporting

N

Y(Report must

be submitted)

National

Regulator

NCA

Report?

Other Participating

National Regulators

Y

SG2 N8 SG2 N79

SG2 N38

Manufacturer’s

Investigation

Adverse

Event

To Whom to ReportN54 Section 6

Adverse Events must be reported to a National Competent Authority (NCA) according to applicable requirements in each jurisdiction. NCA’s should provide a contact point to manufacturer from reporting.

SG2 considered several options that might resolve this situation, including the establishment of a global database for submission of adverse event reports.

Manufacturer or Authorised Representative

National Regulator (NCA)

Within What Timeframe?

SG2 N54 Sect 8

SG2 N54 Sect 7

SG2 N54 Sect 6

SG2 N54

Sects 3, 4 & 5

SG2 N54

Annex C

To which

NCA(s)?

Within what

timeframe?

What

Information?

Y

?Report?

N

Trend?Document

rationale

for not reporting

N

Y(Report must

be submitted)

National

Regulator

NCA

Report?

Other Participating

National Regulators

Y

SG2 N8 SG2 N79

SG2 N38

Manufacturer’s

Investigation

Adverse

Event

Reporting TimeframesN54 Section 7 & Annex B

Adverse events that result in unanticipated death or unanticipated serious injury or represent a serious public health threat must be reported immediately by the manufacturer.

All other reportable events must be reported as soon as possible by the manufacturer, but not later than 30-elapsed calendar days following the date of awareness of the event.

Reporting Timeframes

Immediately: For purposes of adverse event reporting, immediately means as soon as possible, but not later than 10 elapsed calendar days following the date of awareness of the event.

Serious public heath threat: Any event type, which results in imminent risk of death, serious injury, or serious illness that may require prompt remedial action.

Manufacturer or Authorised Representative

National Regulator (NCA)

What Information (Dataset)?

SG2 N54 Sect 8

SG2 N54 Sect 7

SG2 N54 Sect 6

SG2 N54

Sects 3, 4 & 5

SG2 N54

Annex C

To which

NCA(s)?

Within what

timeframe?

What

Information?

Y

?Report?

N

Trend?Document

rationale

for not reporting

N

Y(Report must

be submitted)

National

Regulator

NCA

Report?

Other Participating

National Regulators

Y

SG2 N8 SG2 N79

SG2 N38

Manufacturer’s

Investigation

Adverse

Event

Report Data Set

Event information: Dates, Reporter details, Healthcare facility details, Patient details, Event type and description, Notified CA’s, Resolution description

Device Information: Manufacturer, Generic device group, Disposition, Results of analysis, Corrective action taken.

Other: Comments, Notified Body details, CA’s notified of Corrective action

Courtesy of Isabelle Demade, Principal Administrator, European Commission, DG Health and

Consumers, Unit B2, Cosmetics and Medical Devices ,

Chair GHTF SG2

SG2 STATUS REPORT- May 2010

SG2 Main Tasks

Updating/developing guidance on A/E reporting and other PM surveillance activities, when appropriate.

Improving/developing/maintaining the NCAR Exchange Programme

Providing training on SG2 Guidance and NCAR Exchange Program upon request

SG2 2009 Main Achievements

Updating and improving guidance on NCAR Exchange Program: N38:2009;N79:2009

Establishing closer links with AHWP WG2

SG2 guidance in progress (1)

Definition and Classification of FSCA, incl. Recalls

Reason: term ‘recall’ not uniformally applied. Has impeded communication between NCA’s+ led to anxieties among general public. Recall classification equally disparate.

Expected completion: Q4 2010/ Q1 2011

SG2 guidance in progress (2)

Severe Adverse Events REPORTINGDURING CLINICAL INVESTIGATION

Reason: C.I. conducted globally

Joint SG2-SG5 TF

Draft: along lines GHTF SG2 N54 R8

Expected completion : Q4 - 2010

Training SG2 Activities

Ongoing:

- Centralizing training requests

- Developing/updating training slide presentations

Request: development of SG2 Training

Documents

Scope: trainers qualified, type of training sessions, feedback on training.

GHTF/SC/N2R12:2010 :GHTF Roles and Responsibilities

GHTF Membership Update

Rules on participation to GHTF (SG2)GHTF/SC/N2R12:2010 :GHTF Roles

and Responsibilities

GHTF - voluntary international group of medical device regulatory authorities, and medical device trade associations grouped into three geographical areas, namely Europe, North America and Asia-Pacific

Different categories of membership – Founding Members, Regional Members, Participating Members and Liaison Bodies.

Founding Members : Regulatory authorities or industry representatives from Europe, the United States of America (USA), Canada, Japan and Australia.

Regional members: Members represent the regulatory authorities and industry representatives in a formal regional grouping of economies, not otherwise represented by a Founding Member.

Participating members : representatives of medical device regulatory authorities or medical device trade associations of countries other than the Founding Members

Liaison bodies: public health organizations, international standard-setting bodies or other groups who can contribute to or benefit from participation in GHTF.

GHTF Member Categories

Steering

Committee

Closed Session

Steering Committee

Open Session

Study Groups Other meetings,

such as

trainings....

Founding

Members

Yes Yes Yes Yes

Regional

Members

Yes Yes Yes Yes

Participating

Members

No Yes, on invitation of

Chair

Yes, can propose members Yes

Liaison Bodies No Yes, on invitation of

Chair

Can propose observers Yes

Observers

from Founding

Members

Yes, on meeting-

by-meeting

basis and

agreement

by Chair

Yes, on meeting-by-

meeting basis and

agreement by

Chair

Yes, on meeting-by-

meeting basis and

agreement by Study

Group Chair

Yes

Observers

Regional

Members

Yes, on meeting-

by-meeting

basis and

agreement

by Chair

Yes, on meeting-by-

meeting basis and

agreement by

Chair

Yes, on meeting-by-

meeting basis and

agreement by Chair

Yes

Observers

Participating

Members

No No Yes, on meeting-by-

meeting basis and

agreement by Study

Group Chair

-

SG2 Vision for Harmonization

Thanks for your attention

Questions and suggestions