Gas Eye - Medical Gas Alarm System
e eG y
The Geye Medical Gas Alarm System is a
microprocessor based alarm, intended to
show current pressures levels and digital
inputs status, and provide a visual and vocal
alarm in case of deviation of ±20% of the
nominal medical gas pressure from defined
values or when vacuum reaches below 350
mBar.
The purpose of the Geye Medical Gas Alarm
System is to notify both the technical and
the clinical staff that one or more sources of
supply within a supply system are no longer
available for use and it is essential that action
be taken.
The Geye Medical Gas Alarm System uses
information signals to indicate normal and
abnormal status. To facilitate the monitoring
status by hospital personnel, all gases and
vacuum are displayed either with green
7-seg display (Local Modules) or with three
color led lamps (Area Modules), for clear
visibility during day and night. The leds
indicate the direction of the gas / vacuum
pressure. Under normal conditions, the gas
trend indicator is Green, indicating Normal
mode. If the gas pressure approaches alarm
condition, the trend indicator is displayed in
Orange, indicating Caution mode. If an alarm
condition occurs, a Red alarm indicator is
displayed and an audible alarm is heard at
the same time.
The Geye Medical Gas Alarm System
monitors and displays normal and alarm
conditions from up to 4 Local medical gases
or up to 20 Area medical gas conditions
signal in various combinations.
All Local medical gas modules are equipped
with two types of sensors, one for Upstream
pressure (HIGH) and one for Downstream
pressure (LOW) of the same monitored gas
(only one pressure sensor for downstream in
vacuum modules).
The Geye Medical Gas Alarm System uses
gas specific transducers with universal 4-20
mA output, which enables using a wide
range of possible shelf products.
Digital LED readout reflecting a maximum
range of either 0-20 Bar for downstream and/
or 0-250 Bar for upstream service, and 0-999
mBar for vacuum.
Pressing the “Mute” button cancels the
audible alarm. Nevertheless, the unit
remains in alarm condition until the problem
G e ye M e d i c a l G a s A l a r m Sys t e m
e eG yMedical G as Alarm System
is rectified. If the problem is rectified
automatically, the alarm is cancelled, yet a
technician is able to view the causes of the
problems. The system is adjusted to repeat
the alarm every 15 minutes, according to
standards requirements.
The Geye Medical Gas Alarm System can be
used in a Master-Slave configuration, with a
maximum of two Slaves or 10 Report Alarms
per Master.
Each Area alarm module monitors up to 5
alarm points, either Normally Opened or
Normally Closed. Each point represents an
alarm condition which a source equipment
may have. When an alarm condition occurs,
a Red light flashes and the audible alarm
sounds. If several alarm conditions occur
simultaneously, the most recent alarm light
flashes in the specific module as well as in
the master module, while the other alarm
lights remains lit.
The Geye Medical Gas Alarm System is
field-programmable, using several internal
press-buttons or using a Human Machine
Interface, which is supplied along with the
system, for adjustment of High and Low
set-points on the display for each individual
gas service.
The Geye Medical Gas Alarm Systems
comply with the requirements for
Electromagnetic Compatibility (EMC) and
Safety. The alarm systems are not affected
by electromagnetic energy which may be
emitted by other equipment in the area.
The Geye Medical Gas Alarm Systems use
RF energy only for its internal function.
Therefore its RF Emissions are very low and
are not likely to cause any interference in
nearby Electromagnetic equipment.
The Geye Medical Gas Alarm Systems are
easy to use and include build-in interfaces
to facilitate communications from all panels
on-site to PC control in the Gas Centers.
External USB connection for local control
and setting and Internal RS 232 and RS 485
are available.
Alarm history data can be collected and
downloaded into a computer equipped with
proprietary PC Software and can be printed,
for administration analyzing.
e eG yMedical G as Alarm System
Combo Module Configuration
High Pressure (by a colored status led)
Normal Status ledMain Visual Alarm
Gas Label
Normal Open/ Normal Close Status
7-segments display
Main Label
Buzzer
Low Pressure (by a colored status led)
Measurement unit
Power On led
Mute
Test
USB
n Gas specific transducers are used in the system with universal 4-20 mA output, which enables using a wide range of possible shelf products.
n True digital led display and trend indicators for gas pressure direction.
n Lighted led display which is visible from different angles or in dim lighting conditions.
n Self diagnostic circuitry with error display for error recognition.
n High accuracy of the pressure value measurement.
n Dry contacts for remote monitoring of High and Low alarms' status.
n Built-In relay for remote alarm.
n External USB connection for local PC control.
n History data can be logged and printed for administration analyzing.
Included Features:
Installation RequirementsPipeline pressure switches (for Area modules only):90 VAC to 240 VAC, 50 to 60 Hz
Item Parameter Value
Specifications
e eG yMedical G as Alarm System
Control system metal box Width 242mm
Depth 41mm
Height 250.6mm
Plastic cover Width 254mm
Depth 35mm
Height 260mm
Electronic circuit for reading 4-20 mA sensor version Reduced- 3
Full- 7
7-SEG Width 8mm
Height 14.2mm
buzzer Sound level 70 dB
80 dB
90 dB
Off
Communication USB
RS232/RS485
Main Switch Toggle (on board)
Available ConfigurationsLocal Module Alarm
GL-104-XXXX
GL-103-XXX
GL-102-XX
Geye Local-104 (including 4 Sensors)
Geye Local-103 (including 3 Sensors)
Geye Local-102 (including 2 Sensors)
Available ConfigurationsArea Module Alarm
GA-104-20L
GA-103-15L
GA-102-10L
Geye Area-104
Geye Area-103
Geye Area-102
Available ConfigurationsCombo Module Alarm
GC-05L-XXX
GC-10L-XX
GC-05L-XX
GC-05L-X
GC-10L-X
GC-15L-X
Geye Combo-3 Gases 05 Leds
Geye Combo-2 Gases 10 Leds
Geye Combo-2 Gases 05 Leds
Geye Combo-1 Gas 05 Leds
Geye Combo-1 Gas 10 Leds
Geye Combo-1 Gas 15 Leds
DescriptionReport Alarm
10 Report Alarms can be connected
per Master.
Report Alarm
DescriptionSoftware
The Geye Medical Gas Alarm System
is field-programmable, using several
internal press-buttons or using a
Human Machine Interface, which
is supplied along with the system,
for adjustment of High and Low
set-points on the display for each
individual gas service.
Geye HMI Software
Ordering Information: X = Gas
L = Leds
GL = Geye Local
GA = Geye Area
GC = Geye Combo
1 = Oxygen
2 = Vacuum
3 = Air
4 = Nitrogen
5 = Nitrous Oxide
6 = Cabron Dioxide
7 = Helium
8 = Evacuation
Manufactured by:
Cat Technologies Ltd.Communication Center,G.G Israel Studios Jerusalem Ltd. Neve Ilan 90850 IsraelTel: +972 25704005, Fax: +972 25704006Website: www.cat-tc.com, Email: [email protected]
Our Authorized EC-Representative:
Celitron Medical Technologies Kft.Avar utca 5, 2600 Vác, HungaryTel: +36 27 512 267 Fax: +36 27 512 268 Website: www.celitron.com, Email: [email protected]
Company Profile:
Distributed by:
Applicable Directives and Standards:Technical Standards:Medical Device Directive (MDD) 93/42/EEC; EN 61326: 1997 + A1:98 + A2:01+ A3:03 Electrical equipment for measurement, control and laboratory use- EMC- requirements, minimum immunity test requirements, standard, harmonized under EMC Directive 2004/108/EC Article 6 (2), complies with class A limits for emission; EN 60601-1-2:01 section 36 (not life-supporting equipment) standard, harmonized under Medical Device Directive 93/42/EEC, CISPR 11:2004+ A2 (06) group 1 Class A limits; EN 60601-1:1988+A1:93+A2:95- Medical Electrical Equipment -Part 1: General Requirements for Safety; ISO 7396-1- Medical gas pipeline systems- Part 1: Pipeline system for compressed medical gases and vacuum- sections 6 (Monitoring and Alarm systems), section 12.6.9 (Tests of monitoring and alarm systems), section 13 (Information to be supplied by the manufacturer) And Annex F (Risk management); EN ISO 14971 – Risk Management for Medical device.
0123
Cat Technologies was founded in 1991 for the purpose of development and production of medical and laboratory equipment which requires unique knowledge and follows strict standardization. Since its foundation, the company has developed Electronic Control Systems for:Medical Gas Alarm Systems, Steam Sterilizers, Gas Sterilizers, Incubators, Oxygen Generators, Ultra-sonic Washers, Heat Sterilizers, Simulators, Testing instruments, Monitors, Validators and so forth. Nowadays the company mostly focuses on the sterilization field.CAT Technologies is owned by a public company whose shares are traded on the London Stock Exchange and held mainly by public institutions and investors. The company’s engineers use advanced development tools, which include Emulators and Simulators, testing equipment, integration tools and software.CAT Technologies accompanies the customer from the Idea stage, through the product’s characterization, writes the Requirements documents, and together with the customer, consolidates the design, planning and implementation of the product, as well as the product’s integration, testing and after-sales services. The company has various customers from the United States, France, Germany, Hungary, Israel and Mexico.
The Quality Management system in Cat Technologies is reviewed and approved by TUV Germany since 1997, in compliance with International standards ISO 9001:2008, ISO 13485:2003 (Quality Management System for Medical Device) and the Medical Device Directive MDD
93/42/EEC.