18
INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 51 (2006) 5487–5504 doi:10.1088/0031-9155/51/21/007 Generation and modelling of megavoltage photon beams for contrast-enhanced radiation therapy J L Robar Department of Radiation Oncology, Dalhousie University, 5820 University Street, Halifax, Nova Scotia B3H 1V7, Canada and Department of Physics and Atmospheric Science, Dalhousie University, 5820 University Street, Halifax, Nova Scotia B3H 1V7, Canada E-mail: [email protected] Received 1 June 2006, in final form 7 August 2006 Published 9 October 2006 Online at stacks.iop.org/PMB/51/5487 Abstract Contrast-enhanced radiation therapy (CERT) is a treatment approach involving the irradiation of tumours containing high atomic number (Z) contrast media, using low-quality x-ray beams. This work describes the experimental generation of x-ray beams using a linear accelerator with low-Z target materials (beryllium and aluminium), in order to produce photon energy spectra appropriate for CERT. Measurements were made to compare the experimental beams to conventional linear accelerator photon beams in terms of per cent depth dose. Monte Carlo simulation was used to model the generation of each beam, and models were validated against experimental measurement. Validated models were used to demonstrate changes in photon spectra as well as to quantify the variation of tumour dose enhancement with iodinated contrast medium concentration in a simulated tumour volume. Finally, the ratio of the linear attenuation coefficient for iodinated contrast medium relative to water was determined experimentally as a function of iodine concentration. Beams created with low-Z targets show significant changes in energy spectra compared to conventional beams. For the 4 MeV/Be beam, for example, 33% of photons have energies below 60 keV. Measurements and calculation show that both the linear attenuation coefficient ratio and dose enhancement factor (DEF) increase most rapidly at concentrations below 46 mg I ml -1 . There is a significant dependence of DEF on electron energy and a lesser dependence on target material. The 4 MeV/Be beam is the most promising in terms of magnitude of DEF—for example, DEF values of 1.16 and 1.29 are obtained for concentrations of 20 mg I ml -1 and 50 mg I ml -1 , respectively. DEF will increase or decrease, respectively, for shallower or deeper tumours at a rate of approximately 1.1% cm -1 . In summary, we show that significant dose 0031-9155/06/215487+18$30.00 © 2006 IOP Publishing Ltd Printed in the UK 5487

Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY

Phys. Med. Biol. 51 (2006) 5487–5504 doi:10.1088/0031-9155/51/21/007

Generation and modelling of megavoltage photonbeams for contrast-enhanced radiation therapy

J L Robar

Department of Radiation Oncology, Dalhousie University, 5820 University Street, Halifax,Nova Scotia B3H 1V7, CanadaandDepartment of Physics and Atmospheric Science, Dalhousie University, 5820 University Street,Halifax, Nova Scotia B3H 1V7, Canada

E-mail: [email protected]

Received 1 June 2006, in final form 7 August 2006Published 9 October 2006Online at stacks.iop.org/PMB/51/5487

AbstractContrast-enhanced radiation therapy (CERT) is a treatment approach involvingthe irradiation of tumours containing high atomic number (Z) contrastmedia, using low-quality x-ray beams. This work describes the experimentalgeneration of x-ray beams using a linear accelerator with low-Z targetmaterials (beryllium and aluminium), in order to produce photon energy spectraappropriate for CERT. Measurements were made to compare the experimentalbeams to conventional linear accelerator photon beams in terms of per centdepth dose. Monte Carlo simulation was used to model the generation ofeach beam, and models were validated against experimental measurement.Validated models were used to demonstrate changes in photon spectra as wellas to quantify the variation of tumour dose enhancement with iodinated contrastmedium concentration in a simulated tumour volume. Finally, the ratio of thelinear attenuation coefficient for iodinated contrast medium relative to waterwas determined experimentally as a function of iodine concentration. Beamscreated with low-Z targets show significant changes in energy spectra comparedto conventional beams. For the 4 MeV/Be beam, for example, 33% of photonshave energies below 60 keV. Measurements and calculation show that boththe linear attenuation coefficient ratio and dose enhancement factor (DEF)increase most rapidly at concentrations below 46 mg I ml!1. There is asignificant dependence of DEF on electron energy and a lesser dependenceon target material. The 4 MeV/Be beam is the most promising in terms ofmagnitude of DEF—for example, DEF values of 1.16 and 1.29 are obtainedfor concentrations of 20 mg I ml!1 and 50 mg I ml!1, respectively. DEFwill increase or decrease, respectively, for shallower or deeper tumours at arate of approximately 1.1% cm!1. In summary, we show that significant dose

0031-9155/06/215487+18$30.00 © 2006 IOP Publishing Ltd Printed in the UK 5487

Page 2: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5488 J L Robar

enhancement is possible by altering the linear accelerator target and filtration,but the magnitude is highly dependent on contrast medium concentration.

(Some figures in this article are in colour only in the electronic version)

1. Introduction

The primary goal of radiation therapy is to achieve tumour control while minimizing toxicityto surrounding tissues and structures. For external beam radiation therapy, improvements inconforming the absorbed dose distribution to the target volume have been achieved mainlythrough refinement of treatment planning techniques, conformal beam shaping and increasedflexibility in beam delivery using, for example, intensity modulated radiation therapy (IMRT).However, a separate and to date experimental approach is the delivery of radiation therapyafter modifying the radiological properties of the tumour itself relative to proximal tissues,through the administration of contrast media (CM). CM are used ubiquitously for contrastenhancement during CT or MRI imaging. CM molecules containing high atomic number (Z)

atoms such as iodine (Z = 53) or gadolinium (Z = 64), for example, localize within braintumours due to local damage of the blood–brain barrier. The presence of the high-Z materialincreases the effective atomic number of the tumour itself. Selective enhancement of absorbeddose can be achieved in the CM-containing tumour with use of x-ray photon spectra for whichthe interaction cross section exhibits a significant Z-dependence. Recently, this method hasbeen termed contrast-enhanced radiation therapy (CERT) (Verhaegen et al 2005).

Eliciting a significant dose-enhancement effect requires the use of spectra containingphotons having a large relative cross section for photoelectric absorption, to take advantageof the Z4 to Z4.8 dependence per atom. Accordingly, most investigations into CERT have usedx-ray beams in the superficial or orthovoltage energy range. Monte Carlo calculation has beenused to estimate achievable magnitudes of dose enhancement using photon beams below200 kVp for simplified (Solberg et al 1992) and realistic (Mesa et al 1999) treatmentgeometries. Norman et al (1991) demonstrated that the presence of iodinated CM has anenhancing effect on cell kill using 140 kVp and 250 kVp beams. Iwamoto et al (1987)demonstrated increased survival times of animals when treated with 120 kVp followingadministration of iodine. This groundwork led to treatment of metastases in humans usinga 140 kVp beam produced by a modified CT scanner (Rose et al 1999). Most recently, thedosimetric and microdosimetric effects of using kilovoltage x-rays in combination with iodinehave been examined (Verhaegen et al 2005). Gold has been shown to have a dose enhancingeffect also, as demonstrated by cell-kill experiments using microspheres (Herold et al 2000),murine models using nanoparticles (Hainfeld et al 2004) and through Monte Carlo modellingof nanoparticles within tumour volumes (Cho 2005).

It is clear that the use of kilovoltage x-ray beams will maximize tumour dose enhancement;however, this approach involves several technical barriers. First, orthovoltage and superficialteletherapy units are not designed to deliver highly conformal, isocentric treatment, and thusCERT would be implemented as an alternative, rather than an adjunct to well-established,linear accelerator-based techniques such as conformal radiosurgery and IMRT. Moreover, theuse of kilovoltage x-rays produces poor dose uniformity within the tumour volume (Verhaegenet al 2005, Mesa et al 1999) and inadvertent dose enhancement within the skull (Mesa et al1999). Previously, we reported on the feasibility of using unflattened beams produced by linearaccelerators for CERT (Robar et al 2002). Removal of the x-ray flattening filter recovers a

Page 3: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5489

population of photons that will undergo photoelectric absorption in the CM-loaded tumour;however, for a 4 MV beam, Monte Carlo simulation demonstrated that the dose enhancementwas limited in magnitude to 9.9% and 13.7% for a tumour containing 30 mg ml!1 iodine andgadolinium, respectively. Achieving greater dose enhancement was shown to be possible byusing lower quality beams, e.g., 2 MV—a mode not readily facilitated by most waveguidesin clinical linacs. The limited magnitude of dose enhancement is partly due to the fact thatthe low-energy photon population is self-absorbed upstream from the flattening filter by thehigh-Z linac target.

In the present work, we extend the previous modelling study (Robar et al 2002) withphysical measurement, and, in addition to removing the flattening filtration, we create linacbeams by using alternative, low-Z target materials. As demonstrated, for example, by Galbraith(1989), compared to standard target materials (e.g. tungsten or copper), low-Z elementsare attractive for this application because (i) the fractional yield of low-energy photonswill be larger (Tseng and Pratt 1979, Motz 1955, Pratt et al 1977), (ii) electron–electronbremsstrahlung is more significant, which has a lower energy peak compared to electron–nuclear bremsstrahlung (Motz 1955), and (iii) self-absorption of low-energy bremsstrahlungphotons will be reduced due to the lower photoelectric cross section (Motz 1955).

The specific goals of this work are the following:

(1) To generate experimental x-ray beams (henceforth called CERT beams) optimized fordose enhancement using beryllium and aluminium targets.

(2) To model the generation of the CERT beams using Monte Carlo simulation and to validatethrough comparison to characteristics of beams produced in step (1).

(3) To measure, for the CERT beams, the ratio of attenuation coefficient for iodine CM towater and the variation of this ratio with iodine concentration.

(4) To determine, using the models validated in step (2), the variation of dose enhancementwith the concentration of iodinated CM in the tumour.

2. Materials and methods

2.1. Generation of CERT x-ray beams

Our installation of low-Z targets in the linear accelerator is similar to the approach used byOstapiak et al (1998), Tsechanski et al (1998) and Flampouri et al (2002) for the purposes ofimproved contrast in portal imaging. Four different low-Z targets, made of either beryllium oraluminium, were installed in previously unoccupied ports of a Varian 2100C (Varian MedicalSystems, Palo Alto, CA, USA) linac carousel. For each material, two target thicknesses wereused—one for each of the 4 MeV and 6 MeV electron energies (table 1). In each of the previouslow-Z portal imaging studies, targets had thicknesses less than the continuously slowing downapproximation (CSDA) range of the incident electrons. For our study, we used a Be targetthickness of 1.5 cm, which is close to that used previously by Ostapiak et al (1.57 cm), for the6 MeV beam. This corresponds to approximately 60% of the CSDA range for the material. Thesame design rule was used for the other three targets. Each target was 5.0 cm in diameter andsecured within an aluminium casing similar to those used to house scattering foils. Transmittedelectrons were absorbed within a 9.0 mm polystyrene filter inserted into the upper wedge tray.This polystyrene filter was used for all measurements except for the validation of the MonteCarlo models of the linac head, described in section 2.2. For experimental measurements, thecarousel-mode/BMAG board was accessed to manually select any one of the four targets. Allbeams were produced with a nominal rep-rate of 400 MU min!1.

Page 4: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5490 J L Robar

Figure 1. View of the primary collimator and inserted target for the 6 MeV/Al CERT BEAMmodel.

Table 1. CERT beam target design.

Thickness Nominal electronMaterial (cm) energy (MeV)

Beryllium 1.0 4Beryllium 1.5 6Aluminium 0.67 4Aluminium 1.0 6

2.2. Monte Carlo modelling of CERT beams and experimental validation

In order to examine changes in photon energy spectra and to quantify dose enhancement iniodinated CM, generation of the CERT beams was modelled with Monte Carlo simulationusing the BEAMnrc package (Rogers et al 1995). All calculations were performed using acluster of six 3.4 GHz PCs, with distributed processing facilitated by the Sun Grid Engine (SunMicrosystems, Santa Clara, CA, USA). In accordance with the physical set-up, the downstreamsurfaces of the custom targets were located in the same location as electron scattering foils inthe carousel, approximately 88 cm from mechanical isocentre. Separate models were run forfield sizes of 5 " 5 cm2 and 1 " 1 cm2 (specified at isocentre), with 1 " 108 and 1 " 1010

electron histories incident on the targets, respectively. For the 5 " 5 cm2 field size, modelswere run with as well as without the 9.0 mm polystyrene filtration. A conventional 4 MVlinear accelerator head with flattening filtration was also modelled for comparison to the CERTbeams. For all BEAMnrc simulations, the global electron and photon cut-off energies (ECUTand PCUT) were 0.7 MeV and 0.01 MeV, respectively. For both nominal electron energiesexamined, the actual mean energy of the electron beam was determined with reference to thebending magnet shunt current read directly from the linac. Figure 1 shows, for example, theBEAMnrc-modelled generation of photons in the 1.0 cm Al target. This figure emphasizesthat the target location is by no means optimized; after exiting the vacuum section of theprimary collimator, electrons scatter over several cm of air prior to impinging on the target,resulting in a larger focal spot size than is necessary. Eventual refinement of this set-up would

Page 5: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5491

involve placement of the low-Z target at the usual location above the primary collimation—thiswould also introduce the possibility for a retractable arm housing multiple targets (similar tothat for dual-energy linear accelerators) to facilitate a ‘dose-enhancement mode’ as well asconventional photon beams. Interlock systems would be similar to those used presently (i.e.a combination of hardware feedback and software control to ensure correspondence of theselected mode and target arm position).

To validate the Monte Carlo models, particularly in terms of beam quality, measuredand simulated depth doses in water were compared. This comparison was made initially forbeams without the 9.0 mm polystyrene filter, i.e. including all electrons transmitted throughthe target. This approach provides a sensitive comparison because the relative height anddepth of the electron depth dose peak are highly variable with the electron energy as set inthe BEAMnrc simulation, as well as the geometry of the model; for example, the effect oftuning the electron energy by ±0.1 MeV could be detected in the comparison, whereas theeffect would be less marked in comparing depth doses for the filtered photon beams. Depthdose measurements were performed in water using a parallel-plate chamber (Capintec PS-033,Ramsey, NJ, USA) to achieve high resolution in the narrow electron peak region and witha cylindrical mini-ionization chamber (PTW Semiflex 0.125cc, PTW, Freiburg, Germany) atgreater depths. The EGSnrc/DOSXYZnrc code (Nelson et al 1985) was used to calculatedepth dose with the 5 " 5 cm2 beams incident normally on a homogeneous 30 " 30 " 30 cm3

water phantom with a source-to-surface distance of 100 cm. Dose was scored down the centralaxis in depth increments of 0.1 cm from the surface to 1.0 cm depth, in 0.2 cm incrementsfrom 1.0 cm to 2.0 cm depth and in 1.0 cm increments at greater depths. For these simulations,ECUT and PCUT were set to 0.7 and 0.01 MeV, respectively.

2.3. Monte Carlo modelling of iodine CM and attenuation of CERT beams

A commonly used non-ionic, iodinated radiographic contrast agent is iopamidol(C17H22I3N3O), which is commercially available diluted in water to various concentrations.For the purposes of Monte Carlo modelling, cross sections for this material were generatedusing the PEGS utility (Nelson et al 1985) as a ‘compound’ by specifying the relative numberof iodine atoms to water molecules. The highest concentration considered is representative ofundiluted iopamidol containing 370 mg ml!1 organically bound iodine (i.e. Isovue 370, BraccoDiagnostics, Mississauga, Canada), with 18 water molecules per iodine atom. The physicaldensity for this concentration at 20 #C as given by manufacturer and confirmed throughmeasurement is 1.4 g cm!3. Cross-sectional data files were also created for concentrationscorresponding to reducing the undiluted concentration by factors of 2, 4, 8, 16, 32, 64 and 128,i.e. the lowest concentration considered was 2.9 mg I ml!1. The density of each CM mediumwas adjusted according to the dilution ratio. Electron and photon lower cut-off energies usedin creating PEGS data were 0.521 MeV and 0.01 MeV, respectively. Rayleigh scatter crosssections were not generated.

Since the CM was created as a new material for Monte Carlo simulation, it was of interestto validate this material, and the combination of this material with modelled CERT beams, withcomparison to physical measurement. This was done by comparing simulated and measuredtransmissions of the CERT beams through the CM. The arrangement for this measurement isshown in figure 2. A Lucite vial 14 cm in length with 1.0 cm inner diameter was machined tofit into a block tray (i.e. several cm below the polystyrene electron filter), centred on the beamaxis. The end window of this vial was thin (approximately 0.5 mm) to include a minimumof non-CM material in the beam path. Transmission measurements through 370 mg I ml!1

iopamidol were made, minimizing the field size and maximizing distance from the attenuating

Page 6: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5492 J L Robar

Figure 2. Physical set-up for measurements of transmission through iodinated CM.

material to the detector, to avoid detection of attenuator-scattered photons. Accordingly, a1 cm " 1 cm field size was used for all measurements with a mini-ionization chamber (PTWSemiflex 0.125cc, PTW, Freiburg, Germany) located 200 cm from the source. The chamberwas located in an 8 cm thick Solid WaterTM phantom (Gammex RMI, Middleton, WI, USA),with the centre of the chamber volume 1.5 cm below the surface.

If the form of the transmission curve is known, the measured transmission T may beparameterized with as few as two measurements with different attenuation thicknesses (x);however, our intention was to confirm the fit of our data to the form proposed by Bjarngardand Shackford (1994) for polyenergetic spectra:

T (x) = e!µx(1!!x), (1)

where µ and ! are the linear attenuation and beam hardening coefficients, respectively.Measurements were made with thicknesses ranging from 1.0 to 13.0 cm. This was repeatedfor the 4 MeV/Be and 6 MeV/Be CERT beams.

The experimental set-up was replicated as closely as possible using the DOSXYZnrcMonte Carlo code by creating a phantom consisting of a 1 cm " 1 cm column of iodinatedCM within air and a water phantom 200 cm from the source. The dose was scored on thecentral axis within a 0.5 cm " 0.5 cm " 0.5 cm voxel centred 1.5 cm below the surface of thewater phantom. The thicknesses of modelled 370 mg I ml!1 CM were the same as those usedexperimentally. The phase-space files for the 4 MeV/Be and 6 MeV/Be beams with 1 " 1 cm2

field size (at isocentre) were used as input to these calculations.

2.4. Measurement of attenuation coefficient ratio

The goal of this experiment was to quantify, through physical measurement alone, radiologicalchange of the iodinated CM relative to water for the CERT spectra. Accurate measurement ofabsorbed dose in a non-water medium is difficult without Monte Carlo generated parameters,however; for example, if an ionization chamber is used, conversion of ionization to doserequires ratios of averaged restricted stopping powers for the medium to air, for the particularincident spectrum (el-Khatib and Connors 1992). These were not known a priori forour combination of spectra and media and thus direct measurement of dose would not be

Page 7: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5493

possible without relying on further Monte Carlo calculation. In addition, since the doseenhancement arises from low-energy photoelectrons, the fluence of these electrons would beperturbed significantly by the chamber wall, and the measurement would therefore dependon the accuracy of wall correction factors (el-Khatib and Connors 1992). Dose measurementusing alternate detectors such as diodes or thermoluminescent dosimeters (TLDs) would beconfounded by over-response to low-energy photons and spectral change with depth in themedium.

However, absorbed dose can be expressed (Johns and Cunningham 1983) as

Dmedium = "

#

"

mediumEab,medium,

i.e. the product of the fluence ", mass attenuation coefficient (µ/#)medium and averageenergy absorbed per interaction in the medium Eab,medium. Therefore, using transmissionmeasurements parameterized by equation (1), it is possible to determine µCM/µwater, i.e. oneterm in the ratio DCM/Dwater depending on both the medium and incident spectrum. Thequantification of DCM/Dwater itself is left to Monte Carlo simulation (section 2.5).

To determine the variation of µCM/µwater with CM concentration, transmissionmeasurements were made for two thicknesses (3.0 cm and 13.0 cm), for iopamidolconcentrations ranging from 11.6 mg I ml!1 to 370 mg I ml!1. This was done with a serial-dilution method, i.e. after making measurements with the 370 mg I ml!1 CM, this solution wasdiluted by a factor of 2, remixed and reinserted into the linear accelerator tray. Transmissionmeasurements were repeated for distilled water. Complete data sets were acquired in thisfashion for the 4 MeV/Be and 4 MeV/Al CERT beams, and also for a flattened 4 MVbeam for comparison. The 4 MeV CERT beams were chosen for this experiment followingconsideration of results of dose enhancement modelling (section 3.5).

2.5. Quantification of tumour dose enhancement

Having validated the Monte Carlo models of the CERT beams as well as simulated attenuationof these beams by modelled CM, we used DOSXYZnrc simulation to calculate doseenhancement within a tumour volume. This was done using a phantom identical to thatdescribed recently by Cho (2005), consisting of a simulated tumour volume inside a cubictissue phantom. The tumour region is 2.4 " 2.4 " 3.5 cm3 with the top surface located4.75 cm below the surface of a 30 " 30 " 30 cm3 tissue slab. Dose was scored in 0.5 cmdepth increments. Separate simulations were run for each of the iodinated CM concentrationsand each of the four CERT beams (4 MeV/Be, 4 MeV/Al, 6 MeV/Be and 6 MeV/Al), aswell as for the 4 MV flattened beam, for comparison. Dose enhancement was quantified bycalculating the dose enhancement factor (DEF) defined (Robar et al 2002) as the mean dosewithin the tumour with CM to the mean dose within a tumour composed of tissue alone.

3. Results and discussion

3.1. CERT x-ray beam characteristics

Figure 3 shows measured per cent depth dose (PDD) in water for the four CERT beams,compared to measured PDD for standard 4 MV and 6 MV beams. All curves correspond toa 5 " 5 cm2 field size. The energy of the incident electron beam is the predominant factorin determining the penetration of the beams in water, but the trend with atomic number ofthe target is as expected. Thus, the 4 MeV/Be target combination gives the lowest beamquality, with a PDD(10 cm) of 48.5%. Note that beyond approximately 5 cm, both 6 MeV

Page 8: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5494 J L Robar

Figure 3. Measured per cent depth dose for the four CERT beams compared to flattened 4 MVand 6 MV beams.

CERT beams have depth dose curves similar (within 3%) to that for a 4 MV flattened beam.The relatively high surface dose for the 6 MeV/Be target beam is caused by the fact thatthe electrons transmitted through the target have a range of almost 1.5 cm and thus the9.0 mm polystyrene filter was not sufficient to eliminate this contamination completely. Thisis less pronounced for the 6 MeV/Al target case because the ratio of photon to electron fluencebelow the target is greater (see section 3.2). Although the magnitude of dose enhancement forlinac-generated CERT beams cannot be expected to reach that for kilovoltage x-ray beams,figure 3 demonstrates an important advantage of the current approach: while the CERT beamsare of significantly lower quality compared to standard megavoltage beams, skin sparing ispreserved.

Figure 4 shows radial profiles for the four CERT beams, measured in the cross-planedimension (i.e. perpendicular to the waveguide axis) at 5 cm depth in water, with the linacjaws set to define the maximum field size (40 cm " 40 cm at isocentre). This demonstrates thatthe use of a higher electron energy and/or a lower-Z target produces a more forward-peakeddose profile. Several investigators have shown that unflattened beam profiles do not presentinsurmountable challenges for treatment planning and delivery. O’Brien et al (1991) havedemonstrated that unflattened beams can be used effectively for stereotactic radiosurgery dueto the small field sizes required. Recently, Titt et al (2006) have shown that for IMRT flattenedbeams are not required and that unflattened beams offer the advantages of increased dose rate,reduced scatter and reduced leakage from the linear accelerator head.

3.2. Validation of Monte Carlo models

Figure 5 compares measured and calculated per cent depth dose curves in water for the CERTbeams, without polystyrene filtration. All four curves show rapid dose fall-off with depth dueto the transmitted electrons, followed by a region corresponding to attenuation of the photoncomponent. Simulated and measured curves were normalized at a shallow depth in the photonregion (1.5 cm for the 4 MeV beams and 2.0 cm for the 6 MeV beams) and then rescaled by the

Page 9: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5495

Figure 4. Radial profiles of the four CERT beams.

Figure 5. Comparison of measured and calculated depth dose curves for the four CERT beamswithout electron filtration.

Page 10: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5496 J L Robar

Figure 6. Comparison of energy spectra for the four CERT beams below 400 keV at a planeimmediately below the polystyrene filtration. The spectrum for a 6 MV flattened beam is shownfor reference.

same factor to set the measured electron peak to 100%. The ranges of the electrons in watersuggest that the mean energies of transmitted electrons are approximately 2 MeV and 3 MeVfor the 4 MeV and 6 MeV CERT beams, respectively. Good agreement between measurementand calculation is observed for all four cases—within 2% throughout the majority of the depthrange. A discrepancy of 5% is seen at the maximum of the 4 MeV/Al target PDD, but this islikely caused by sampling differences between the measurement and simulation in the regionof the very sharp peak. This level of agreement between simulation and measurement wasobserved following only minor tuning (of ±0.1 MeV) of the electron energy in the BEAMnrcmodel from the value derived from the actual bending magnet shunt current. The final electronenergies were 4.5 MeV and 7.0 MeV for the nominal 4 MeV and 6 MeV beams, respectively.

3.3. Energy spectra for CERT beams

Figure 6 shows the spectra below 400 keV for all CERT beams at a plane immediately belowthe 9.0 mm polystyrene filter. These are compared to the spectrum for a flattened 6 MV beamfrom our previous study (Robar et al 2002). All spectra have been normalized to unit area andconsider all photons reaching the phase-space plane within a 40 cm " 40 cm area. Clearly,the use of thin, low-Z targets and absence of flattening filtration produce a marked change inthe photon spectra over the low-energy range. Note that normalization to unit area was usedto compare spectra in a relative sense; however, if the spectra are compared in terms of fluenceper MeV per incident electron, the spectra have peak values of 7.2 " 10!6 (Be 4 MeV), 5.3 "10!6 (Al 4 MeV), 1.7 " 10!5 (Be 6 MeV) and 1.1 " 10!6 (Al 6 MeV) MeV!1 cm!2. Thus,although the normalized spectra shown in figure 6 indicate that both lower Z and lower energyresult in a higher proportion of low-energy photons, for these targets, use of 4 MeV insteadof 6 MeV will reduce the efficiency of x-ray production (near low-energy peak) by a factor of0.42 and 0.48 for Be and Al, respectively.

Figure 7 gives the fraction of photons below given energies for each spectrum. Recently,Verhaegen et al (2005) illustrated that the maximum dose enhancement should be expectedat photon energies exceeding the contrast medium K-edge (i.e., 33.2 keV for I, 50.2 keV for

Page 11: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5497

Figure 7. Fractional fluence below given energies for the four CERT beams, compared to a 6 MVflattened beam.

Gd) by up to 30 keV, where the separation between water and CM absorption coefficients isgreatest. For iodine, this group estimated that this occurs at approximately 60 keV. For the4 MeV/Be, 6 MeV/Be, 4 MeV/Al and 6 MeV/Al spectra, 33%, 29%, 18% and 13% ofphotons have energies less than 60 keV.

Among previous reports in the literature, the portal imaging study by Tsechanski et al(1998) is most similar to our own in terms of energy of incident electrons, target materialsand target thickness used. Using the EGS4 Monte Carlo code (Nelson et al 1985), this groupmodelled 4.0 MeV electrons (actual energy) incident upon 0.727 cm Be and 0.498 cm Altargets (i.e. as compared to our 4.5 MeV electrons incident upon 1.0 cm Be and 0.67 cm Al).Figure 8 shows a comparison of the energy spectra from this study to our own calculatedresults.

3.4. Validation of attenuation of CERT beams by modelled contrast medium

Measured and calculated transmission values agree within associated uncertainties (±1.5%for measurement and approximately ±2% for simulation), as shown in figure 9. The solidand dashed curves result from fitting to the measured and calculated data, respectively, usingequation (1). Fitting yields values of µ agreeing to within 1.0% (µmeasured = 0.1095 cm!1 andµcalculated = 0.1105 cm!1 for the 4 MeV/Be beam and µmeasured = 0.951 cm!1 and µcalculated =0.961 cm!1 for the 6 MeV/Be beam).

3.5. Measurement of linear attenuation coefficient for CM relative to water

Figure 10 gives the measured variation of µCM/µwater with iodine concentration, for the4 MeV/Be, 4 MeV/Al and 4 MV beams. Both CERT beams show a rapid increase ofµCM/µwater between 0 and 46 mg ml!1 to 1.16 and 1.22 for the 4 MeV/Al and 4 MeV/Bebeams, respectively. This initial rise is not observed for the 4 MV flattened beam,which exhibits an approximately linear variation of µCM/µwater with iodine concentration.Considering that the vast majority of photons in this flattened beam will undergo Compton

Page 12: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5498 J L Robar

Figure 8. Comparison of energy spectra to those from Tsechanski et al (1998).

Figure 9. Measured versus calculated transmission through 370 mg I ml!1 iopamidol for the4 MeV/Be and 6 MeV/Be CERT beams.

scatter, the increase can be attributed simply to the increase in density of the CM (1.4 g cm!3

for the 370 mg I ml!1 concentration). We emphasize that figure 10 provides directly measureddata showing the effect of using beams from thin, low-Z targets on attenuation in contrastmedium but does not give dose ratio itself. With increasing iodine concentration, theratio DCM/Dwater will in fact vary according to the product of the ratios µCM/µwater andEab,CM/Eab,water. The latter term is also expected to increase with iodine concentration due toan increase in the photoelectric cross section relative to the total cross section; thus, the resultsin figure 10 likely represent an underestimation of the dose ratio.

Page 13: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5499

(mg ml!1)

Figure 10. Variation of µCM/µwater with concentration of iodinated CM, for the 4 MeV/Be and4 MeV/Al CERT beams, as compared to a standard 4 MV beam. Error bars result from propagationof uncertainty in ionization chamber measurements through equation (1).

Figure 11. Depth dose for the 4 MeV/Be beam within the modelled tumour region. Curves, fromlow to high relative dose, correspond to CM concentrations of 0, 2.9, 5.8, 11.6, 23.1, 46.2, 92.5,185 and 370 mg I ml!1.

3.6. Quantification of tumour dose enhancement

Figure 11 shows the depth dose distribution in the tumour region of the DOSXYZnrc phantomdescribed in section 2.5 for concentrations of 0, 2.9, 5.8, 11.6, 23.1, 46.2, 92.5, 185 and370 mg I ml!1, for the 4 MeV/Be CERT beam. For purposes of comparison, figure 12 showsthe corresponding results for the flattened 4 MV beam. A summary of dose enhancementresults for all CERT beams is given in figure 13, showing the variation of DEF with iodine

Page 14: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5500 J L Robar

Figure 12. Depth dose curves corresponding to those shown in figure 11, for the 4 MV flattenedbeam.

(mg ml!1)

Figure 13. Variation of dose enhancement factor with iodine concentration for the four CERTbeams and the 4 MV flattened beam.

concentration. DEF is seen to increase rapidly at low concentrations, followed by a saturationof the effect. There is a strong dependence on electron energy and a secondary dependenceon target material. This energy dependence, combined with the observation that skin sparingis preserved (figure 3) for the 4 MeV/Be beam, suggests that investigation of energies lowerthan 4 MeV is warranted.

With regard to magnitude of dose enhancement, the key finding is that the DEF increasesrapidly over a range of iodine concentrations achievable in brain tumours. Previous studieshave estimated maximal iodine concentrations ranging from 20 mg I ml!1 (Mesa et al 1999,

Page 15: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5501

Figure 14. Comparison of CERT beam spectra within the dose enhancement phantom, 1.25 cmbelow the surface, in tissue.

Rose et al 1999, Adam et al 2005) to 50 mg I ml!1 (Verhaegen et al 2005); for the4 MeV/Be beam, this would yield DEF values of 1.16–1.29, respectively. The sensitivityof DEF to CM concentration in figure 13 underlines the importance of maximizing uptake ofCM concentration within tumours. This challenge is outside the scope of this paper—however,recently, Adam et al (2005) have reported a five-fold increase in iodine tumour uptake withthe use of mannitol to increase blood–brain-barrier permeability.

Note that the relative magnitudes of DEF for the four CERT beams in figure 13 appearto contradict the spectral comparison in figure 6, i.e. the spectra immediately below the exitwindow of the linac would suggest that the 4 MeV/Be and 6 MeV/Be beams should yieldthe highest values of DEF. We investigated this issue by using the FLURZ Monte Carlo code(Nelson et al 1985) to record the photon spectra at various depths in a phantom similar to thatdescribed in section 2.5. Figure 14 shows the spectra in tissue 1.25 cm below the surface ofthe tissue phantom, where each spectrum has been normalized to unit area. While the relativeorder of the spectra in the low-energy range (e.g. below 100 keV) is the same as that at the exitwindow of the linac (figure 6), it can be seen that the difference between the 6 MeV/Be and4 MeV/Al beams is less significant. The spectra are shown at a depth of 6.25 cm, correspondingto the tumour region, in figure 15. In terms of relative population of low-energy photons,the order of the four spectra has changed and now corresponds to the trend of DEF withenergy/target combination shown in figure 13. This study thus demonstrates the effect ofbeam hardening: with increasing depth, both 4 MeV beams contain a higher proportion oflow-energy photons compared to the 6 MeV beams.

The diminution of the low-energy photon populations in figure 15 suggests that beamhardening will also reduce the magnitude of DEF with depth. Note that the results in figure 13correspond to a tumour centred 6.5 cm within the tissue phantom, i.e. with a proximal surface4.75 cm deep. Repeating the simulation with the tumour surface at varying depths, from 0to 10 cm, for the 4 MeV/Be beam and a fixed concentration of 23.1 mg I ml!1, yields therelationship given in figure 16. The dose enhancement decreases approximately linearly withdepth in phantom at a rate of approximately 1.1% cm!1.

Page 16: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5502 J L Robar

Figure 15. Spectra 6.25 cm below the surface, in the tumour region containing 23.1 mg I ml!1.The low-energy peak (0.033 MeV) results from characteristic photons from iodine.

Figure 16. Reduction of DEF with increased depth of the tumour surface, for the 4 MeV/Be beamand a CM concentration of 23.1 mg I ml!1. DEF is diminished at a rate of 1.1% cm!1.

4. Conclusions

This work demonstrates the generation and modelling of megavoltage beams with berylliumand aluminium targets for maximizing dose enhancement in contrast-enhanced radiationtherapy (CERT). The results of our studies show that, compared to flattened photon beamswith standard tungsten/copper targets, low-Z targets produce (i) reduced penetration in water,(ii) a significant increase in the relative population of photons in the vicinity of the K-edge

Page 17: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

Generation and modelling of x-ray beams for contrast-enhanced radiation therapy 5503

of iodine, (iii) a more rapid rise of µCM/µwater with iodine concentration, and (iv) a morerapid increase of DEF with iodine concentration. The 4 MeV/Be beam is the most promisingin terms of magnitude of dose enhancement. This beam has a per cent depth dose at depth10 cm of 48.5%, yet preserves skin sparing. Monte Carlo results demonstrate that thisenergy/target combination yields an energy spectrum with 33% of photons below 60 keV.Physical measurement and Monte Carlo modelling demonstrate, respectively, significantincreases in the attenuation coefficient and absorbed dose, for iodinated CM at concentrationsbelow 50 mg I ml!1. The 4 MeV/Be beam yields the most rapid increase in dose enhancementfactor (DEF) with iodine concentration, with DEF values equal to 1.16 and 1.29 for CMconcentrations of 20 mg I ml!1 and 50 mg I ml!1, respectively, for a tumour centred 6.5 cmbelow the surface of a tissue phantom. The DEF will increase or decrease for shallower ordeeper tumours, at a rate of approximately 1.1% cm!1, due to effects of beam hardening. Iflinac-based CERT is to become a viable technique, maximizing the concentration of contrastmedium in the tumour will be essential.

Acknowledgments

The author is grateful to Varian Medical Incorporated for funding of this project. The authorthanks David Burgess, Scott Purcell and Carl Murphy for invaluable electronic and technicalsupport. Thanks are extended to John Grant at the Cape Breton Cancer Centre for assistancewith initial experiments. Assistance from Dr Michael Hale in establishing a distributedprocessing resource is greatly appreciated. The author thanks Ian Porter for machining ofrequired apparatus. Insightful discussions with Dr Richard Morse at Varian Medical, withDr Dennis Fujii at Bracco Diagnostics Incorporated and with Dr Sang Hyun Cho at MDAnderson Cancer Centre were greatly valued.

References

Adam J F et al 2005 Enhanced delivery of iodine for synchrotron stereotactic radiotherapy by means of intracarotidinjection and blood–brain barrier disruption: quantitative iodine biodistribution studies and associated dosimetryInt. J. Radiat. Oncol. Biol. Phys. 61 1173–82

Bjarngard B E and Shackford H 1994 Attenuation in high-energy x-ray beams Med. Phys. 21 1069–73Cho S H 2005 Estimation of tumour dose enhancement due to gold nanoparticles during typical radiation treatments:

a preliminary Monte Carlo study Phys. Med. Biol. 50 N163–73el-Khatib E and Connors S 1992 Conversion of ionization measurements to radiation absorbed dose in non-water

density material Phys. Med. Biol. 37 2083–94Flampouri S, Evans P M, Verhaegen F, Nahum A E, Spezi E and Partridge M 2002 Optimization of accelerator target

and detector for portal imaging using Monte Carlo simulation and experiment Phys. Med. Biol. 47 3331–49Galbraith D M 1989 Low-energy imaging with high-energy bremsstrahlung beams Med. Phys. 16 734–46Hainfeld J F, Slatkin D N and Smilowitz H M 2004 The use of gold nanoparticles to enhance radiotherapy in mice

Phys. Med. Biol. 49 N309–15Herold D M, Das I J, Stobbe C C, Iyer R V and Chapman J D 2000 Gold microspheres: a selective technique for

producing biologically effective dose enhancement Int. J. Radiat. Biol. 76 1357–64Iwamoto K S, Cochran S T, Winter J, Holburt E, Higashida R T and Norman A 1987 Radiation dose enhancement

therapy with iodine in rabbit VX-2 brain tumors Radiother. Oncol. 8 161–70Johns H E and Cunningham J R 1983 The Physics of Radiology 4th edn (Springfield, IL: Thomas) p 223Mesa A V, Norman A, Solberg T D, Demarco J J and Smathers J B 1999 Dose distributions using kilovoltage x-rays

and dose enhancement from iodine contrast agents Phys. Med. Biol. 44 1955–68Motz J W 1955 Bremsstrahlung differential cross-section measurements for 0.5 and 1.0 MeV electrons Phys. Rev.

100 1560–71Nelson W R, Hirayama H and Rogers D W O 1985 The EGS4 code system SLAC Report 265 (Stanford, CA: Stanford

University)

Page 18: Generation and modelling of megavoltage photon beams for ......2.2. Monte Carlo modelling of CERT beams and experimental validation In order to examine changes in photon energy spectra

5504 J L Robar

Norman A, Iwamoto K S and Cochran S T 1991 Iodinated contrast agents for brain tumor localization and radiationdose enhancement Invest. Radiol. 26 (Suppl. 1) S120–1 (discussion S125–8)

O’Brien P F, Gillies B A, Schwartz M, Young C and Davey P 1991 Radiosurgery with unflattened 6-MV photonbeams Med. Phys. 18 519–21

Ostapiak O Z, O’Brien P F and Faddegon B A 1998 Megavoltage imaging with low Z targets: implementation andcharacterization of an investigational system Med. Phys. 25 1910–8

Pratt R H, Tseng H K, Lee C M, Kissel L, MacCallum C and Riley M 1977 Bremsstrahlung energy spectra fromelectrons of kinetic energy 1 keV ! T ! 2000 keV incident on neutral atoms 2 ! Z ! 92 At. Data Nucl. DataTables 20 175–209

Solberg T D, Iwamoto K S and Norman A 1992 Calculation of radiation dose enhancement factors for doseenhancement therapy of brain tumours Phys. Med. Biol. 37 439–43

Robar J L, Riccio S A and Martin M A 2002 Tumour dose enhancement using modified megavoltage photon beamsand contrast media Phys. Med. Biol. 47 2433–49

Rogers D W, Faddegon B A, Ding G X, Ma C M, We J and Mackie T R 1995 BEAM: a Monte Carlo code to simulateradiotherapy treatment units Med. Phys. 22 503–24

Rose J H, Norman A, Ingram M, Aoki C, Solberg T and Mesa A 1999 First radiotherapy of human metastatic braintumors delivered by a computerized tomography scanner (CTRx) Int. J. Radiat. Oncol. Biol. Phys. 45 1127–32

Titt U, Vassiliev O N, Ponisch F, Dong L, Liu H and Mohan R 2006 A flattening filter free photon treatment conceptevaluation with Monte Carlo Med. Phys. 33 1595–602

Tsechanski A, Bielajew A F, Faermann S and Krutman Y 1998 A thin target approach for portal imaging in medicalaccelerators Phys. Med. Biol. 43 2221–36

Tseng H K and Pratt R H 1979 Electron bremsstrahlung energy spectra above 2 MeV Phys. Rev. A 19 1525–28Verhaegen F, Reniers B, Deblois F, Devic S, Seuntjens J and Hristov D 2005 Dosimetric and microdosimetric study

of contrast-enhanced radiotherapy with kilovolt x-rays Phys. Med. Biol. 50 3555–69