General Specifications InfraSpec TM NR800 Fourier Transform Near-Infrared Analyzers Yokogawa Electric Corporation 2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan GS 12Y03A03-01E GS 12Y03A03-01E ©Copyright Aug. 2001 8th Edition Oct. 01, 2019 n Overview The NR800 Fourier Transform Near-Infrared Analyzers offers high-resolution, high S/N (signal-to- noise) ratio, and wide wavelength scanning range measurement with its newly developed interferometer and detector. The NR800 also offers high stability, vibration resistance, and durability, inheriting features from NR500 earlier successful models. The NR800 allows online, real-time, continuous, multiple, and simultaneous measurement for properties and component concentration of various processes. n Features (including options) Newly developed interferometer and detector Used for a variety of applications, ranging from over to combination tone. • High resolution: Up to 4 cm –1 , user selectable setting • High S/N ratio: 2250:1 (RMS, 4 cm -1 resolution, 4100 to 4200 cm -1 , 1 sec.) • Wide wavelengths scanning range 900 to 2500 nm (11,000 to 4000 cm -1 ) • Wavelength reproducibility: 0.007 cm -1 • Wavelength accuracy: 0.04 cm -1 Enhanced environmental resistance, durability, and reliability • Provides high vibration resistance by a unique design free of sliding parts • Features a multi-channel measurement optical system free of moving parts • Eliminates the need for a PC for continuous operation. You need an engineering computer only for Calibration Modeling and data loading. Outstanding wavelength accuracy allows calibration model transport between NR800s Outlier detection and self-diagnostic features come as standard function Various standard features and available options for optimal system configuration • Dust-proof and drip-proof: IP53 equivalent or NEMA TYPE 3R Optional Explosion-proof Enclosure: FM, TIIS (*1), KOSHA • Non-moving multi-channel measurement: Expandable to 4 channels • Up to 12 items can be measured per stream: limited to 64 items / unit • The optional I/O unit offers a variety of inputs/outputs: Analogue output (up to 40 points), analogue input, and contact input/output • Communication output: RS-422 • Fast Ethernet communication between Engineering PC and the analyzer (*1) TIIS: The Technology Institution of Industrial Safety Explosion proof Model InfraSpec and SPECTLAND are trademarks or registered trademarks of Yokogawa Electric Corporation. All other company and product names mentioned in this document are trademarks or registered trademarks of their respective companies. Please select appropriate equipment in accordance with the laws and regulations of the relevant country/region, when it is used in a location where explosive atmospheres may be present.

General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

GeneralSpecifications

InfraSpecTM

NR800 Fourier TransformNear-Infrared Analyzers

Yokogawa Electric Corporation2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan

GS 12Y03A03-01E

GS 12Y03A03-01E©Copyright Aug. 2001

8th Edition Oct. 01, 2019

n OverviewThe NR800 Fourier Transform Near-Infrared Analyzers offers high-resolution, high S/N (signal-to-noise) ratio, and wide wavelength scanning range measurement with its newly developed interferometer and detector. The NR800 also offers high stability, vibration resistance, and durability, inheriting features from NR500 earlier successful models. The NR800 allows online, real-time, continuous, multiple, and simultaneous measurement for properties and component concentration of various processes.

n Features (including options) Newlydevelopedinterferometeranddetector

Usedforavarietyofapplications,rangingfromovertocombinationtone.• High resolution: Up to 4 cm–1, user selectable

setting• High S/N ratio: 2250:1 (RMS, 4 cm-1 resolution,

4100 to 4200 cm-1 , 1 sec.)• Wide wavelengths scanning range 900 to 2500

nm (11,000 to 4000 cm-1)• Wavelength reproducibility: 0.007 cm-1

• Wavelength accuracy: 0.04 cm-1

Enhancedenvironmentalresistance,durability,andreliability• Provides high vibration resistance by a unique

design free of sliding parts• Features a multi-channel measurement optical

system free of moving parts• Eliminates the need for a PC for continuous

operation. You need an engineering computer only for

Calibration Modeling and data loading.

OutstandingwavelengthaccuracyallowscalibrationmodeltransportbetweenNR800s

Outlierdetectionandself-diagnosticfeaturescomeasstandardfunction

Variousstandardfeaturesandavailableoptionsforoptimalsystemconfiguration• Dust-proof and drip-proof: IP53 equivalent or

NEMA TYPE 3R• Optional Explosion-proof Enclosure: FM, TIIS (*1),

KOSHA• Non-moving multi-channel measurement:

Expandable to 4 channels• Up to 12 items can be measured per stream:

limited to 64 items / unit• TheoptionalI/Ounitoffersavarietyof

inputs/outputs: Analogue output (up to 40 points), analogue

input, and contact input/output• Communication output: RS-422• Fast Ethernet communication between

Engineering PC and the analyzer(*1) TIIS: The Technology Institution of Industrial Safety

ExplosionproofModel

InfraSpec and SPECTLAND are trademarks or registered trademarks of Yokogawa Electric Corporation.All other company and product names mentioned in this document are trademarks or registered trademarks of their respective companies.

Please select appropriate equipment in accordance with the laws and regulations of the relevant country/region, when it is used in a location where explosive atmospheres may be present.

Page 2: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

2

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E Oct. 01. 2019-00

1.Configuration1.1 Configurationexamples

• Modbusoutputwithsamplingunit(explosion-proof,1measuringchannel)

RS-422/RS-232Cconverter *2

RS-232C(MODBUS)

(15m max.)(1km max.)

Sampleejection

Process

Fiber-optic cable *2

Clean, dry air

Hazardous area Non-hazardous area

Alarm contact output

Optical/electricalsignal converter

Monitor/ControlSystem

Engineering PC

Samplingunit

Ethernet(Fiber-optic cable *1)

RS-422(MODBUS)

Ethernet(10 Base-T)

InfraSpecNR800

explosion-proofmodel

Measurement

cell

• Analogueoutputwithsamplingunit(explosion-proof,2measuringchannels)

I/O unit*2

RS-422Process 2

Process 1

Sampleejection

Sampleejection

Clean, dry air

Fiber-optic cable *2

Hazardous area Non-hazardous area

Optical/electricalsignal

Alarm contact output

Engineering PC

Ethernet(10 Base-T)

Analog signal

Contact signal

Contact signal

Monitor/ControlSystem

Sampling cell 1

Sampling cell 2

Ethernet(Fiber-optic cable *1)

InfraSpecNR800

explosion-proofmodel

Measurementcell 1

Measurementcell 2 (1km max.)

*1: Alwaysuseaspecialfiber-opticcable(seesection2.6)foranexplosion-proofmodel.Electriccablescannotbeused. Forageneralpurposemodel,usespecifictypeofcablesasbelow,accordingtothelengthrequiredfortheirinstallation. •Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Electriccablescannotbeused.*2: UseRS-422/RS-232Cconverter(seesection2.5)orI/Ounit(seesection2.2,selectSuffixcodeforexplosion-proofmodel) when you use RS-422 output of explosion-proof model analyzer. These units convert RS-422 output signals from analyzer

to dedicated ones. These units also block the communication signal upon receiving a failure signal from the analyzer Forsafetyandexplosion-proofintegrity,followtheinstructiontoselectspecificfiber-opticcables.Seesection2.4for

measurement and section 2.6 for Ethernet communication.

Page 3: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

3

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

1.2 Component,softwareandcalibrationmodeling

Item Require-ment(*1) Model Description Reference

1 InfraSpec NR800 analyzer A

NR801AG General purpose model, CSA standard

1.32.1

NR805AG Explosion-proofmodel,FMcertifiedNR801JG General purpose model, Japanese version

NR805JG Explosion-proofmodel,TIIScertified(excluding NR805JG/K), KOSHAcertified(onlyforNR805JG/K)

2 I/O unit BNR893AG To be used with NR801,805AG, CSA standard 1.3

2.1NR893JG To be used with NR801,805JG

3 Measurement cell A

NR510 Flow through cell 2.3.1NR512 Flow through cell with constant temperature water tube 2.3.2NR511 Special measurement cell 2.3.3FIR200 In-situ probe 2.3.4

4 Fiber-optic cable for measurement A

NR821 Applicable wavelength range: 900 to 2100 nm, Silica, single2.4.1

NR822 Applicable wavelength range: 900 to 2100 nm, Silica, dualNR823 Applicable wavelength range: 900 to 2500 nm, Fluoride, single

2.4.2NR824 Applicable wavelength range: 900 to 2500 nm, Fluoride, dual

5 RS-422 / RS-232C converter B K9404LA or

K9404LD Converts the RS-422 signal from the analyzer into RS-232C. 2.5

6 Ethernet cable (*2) BNR895 Dedicatedfiber-opticcable(*3) 2.6

- Electric cable, provided by user 6.2

7 SoftwareA NR831 SPECTLAND2 management and maintenance software 2.7B NR530 Chemometrics software 2.8

8 Sampling unit B J439 Yokogawa will propose an optimum unit based on sample pressure, temperature, properties, and measurement items 2.9

9 On-site guidance on calibration modeling B J964 On-site hands-on practice and guidance 5.1

10 Calibration Modeling B J965 Calibration Modeling by Yokogawa based on user-provided sample with laboratory analysis results 5.2

11 Engineering PC A - Providedbyuser.Seerecommendedspecifications. 6.1

12Optical / electrical signal converter for Ethernet

B -Converts optical signals for an Ethernet output into electrical signals for engineering PC Interface. Provided by user. Seerecommendedspecifications

6.2

13 Customer inspectionB J963 For Instrument except following J443 -B J443 For J439 Sampling unit and FIR200 In-situ probe -

14 Equipment start-up B - Start-up work for analyzers and sampling units -

Notes *1: A: Required, B: optional *2: AnEthernetcableisrequired.Chooseeitheraspecialfiber-opticcableorelectriccabledependingonthefollowingconditions: Foranexplosion-proofmodel,alwaysuseaspecialfiber-opticcable(seesection2.6).Electriccablescannotbeused. Forageneralpurposemodel,usespecifictypeofcablesasbelow,accordingtothelengthrequiredfortheirinstallation.

•Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Anelectriccablecannotbeused. *3: Ifatotalcabledistanceislongerthan20m,anadditionalfiber-opticcable(seesection2.6)shallbeprovidedbytheuser.

1.3Safetystandard,EMCstandardModel SafetyStandard EMCstandard

FT-NIRanalyzer I/O unitNR801AG, NR805AG*2

NR893AG*2

CSA C22.2 No.61010-1 (exclude NR805AG) *1FM Class 3810ANSI/ISA 61010-1ANSI/ISA-82.02.02 (IEC 61010-2-031)

-

NR801JG, NR805JG(excluding NR805JG/K)

NR893JG - -

NR805JG/K - - Korea Electromagnetic Conformity Standard

*1: • Installation category (Overvoltage category) II This number represents the transient overvoltage condition. This also complies with the regulation on impulse withstand voltage. “II”appliestoelectricalequipmentwhosepowerissuppliedfromfixedinstallationssuchasdistributionboards. • Pollution Degree 2 Describes how much the dielectric strength or surface resistivity is being deteriorated by adherence of solid, liquid or gas substances. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.*2: Install 5 A breaker in the power supply circuit

Oct. 01. 2019-00

Page 4: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

4

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.Componentspecifications(including options)

2.1 NR800FourierTransformNear-InfraredAnalyzers

2.1.1 HardwarespecificationsPrinciple: Fourier-transform remote measurement

viafiber-opticcableMeasurement method: Optical transmission absorptionMeasured sample: LiquidClassificationoflaserproduct: Class 1 Laser Product (IEC 60825-1:2007/2014)Beam source: Halogen lamp (recommended

replacement interval for continuous operation: 5000 hours)

Detector: InGaAs (indium gallium arsenide) photodiode,effectivewavelengthrange:900 to 2500 nm

Number of measuring channels: 1 to 4 (non-moving)Housing structures: Field suitable with full

hinged front coverFiber-optic cable entry: Cable glandFiber-optic cable connectors:

For measurement; FCFor Ethernet; ST

Display: LEDKeyboard: Covered with water-proof sheetingOperating location requirements: See chapter 4.Grounding type: Independent, 100ΩorlowerInsulationresistance: 10MΩorgreater(500VDC)Withstanding voltage:

PowerSupply WithstandingVoltage100VAC,115VAC 1000VACfor1min.200VAC,230VAC 1500VACfor1min.

Coating: Epoxy resin coatingPaint colour:

Model NR801AG,NR801JG

NR805AG,NR805JG

Flame proof enclosure - Lamp black

(equivalent to Munsell 0.8Y 2.5/0.4)Air regulator cover -

Case Frosty white ( equivalent to Munsell 2.5Y8.4/1.2)

Display panel Lamp black (equivalent to Munsell 0.8Y 2.5/0.4)

Wall mount bracket, Free standing rack

Deep sea moss green (equivalent to Munsell 0.6GY3.1/2.0)

Oct. 01. 2019-00

Model NR801AG NR801JG NR805AG NR805JG(excluding NR805JG/K) NR805JG/K

Explosion proof General purpose type(non explosion-proof) FM *1, *2 TIIS *3 KOSHA *5

Housing structure IP53, NEMA 3R or CSA Type 3RAir (purge gas) connection - 1/4NPT (F) Rc1/4 or 1/4NPT (F)Cable connection Packing gland 3/4NPT(F) Flame-proofpackingglandorsealingfitting

Weight (kg) *4

Without bracket Approx. 50 Approx. 65With wall mount bracket Approx. 55 Approx. 70Free stand rack Approx. 85 Approx. 100

*1: Type X purging and explosion proof for Class I, Division 1, Groups B, C and D. Temperature Class T5. All wiring shall comply with the manufacture’s instructions, National Electrical Code ANSI/NFPA 70 and Local Electrical Codes.*2: All wiring shall comply with the manufacture’s instructions and Local Electrical Codes.*3: Ex pd II B+H2 T5 X All wiring shall comply with the manufacture’s instructions and Local Electrical Codes.*4: Not including weight of the packaging materials.*5: KOSHA, Ex dp IIB+H2 T5 All wiring shall comply with the manufacture’s instructions and Local Electrical Codes.

2.1.2 PerformanceWavelength scanning range: 900 to 2500 nm

(11000 to 4000 cm –1)Wavelength resolution: 4, 8, 16, 32, and 64 cm -1

(user selectable)Wavelength reproducibility: 0.007 cm -1Wavelength accuracy: 0.04 cm -1S/N ratio 2250:1 (RMS, resolution 4 cm -1, 11000 to

4000 cm -1, 1 sec.)

2.1.3 Inputs/Outputs(seealsosection2.2I/OUnit)

(1)CommunicationInterface:Engineering PC: 1 channel (Ethernet)DCS / I/O unit: 2 channels (RS-422) 1 channel for DCS (Modbus) and

another for I/O unit; or 2 channels for I/O unit

(2)Contactoutputs:Specification: NC/NOselectable,2channelsRating: 0.5A,30VDCorlessAction:

General purpose model:NR800analyzerpower”ON/OFF”;Powerfailsignal

StatusDO1TerminalNumber

1 to 2 2 to 3Power“off” Short OpenPower “on” Open Short

NR800Systemalarm

Status DO2TerminalNumber1 to 2 2 to 3

Power“off” Short OpenAlarm occurred Short OpenAlarm not occurred Open Short

Page 5: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

5

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

Explosion proof model:NR800Systemalarm

Status TerminalNumberD1 to D2 D2 to D3

Power“off” Open ShortAlarm occurred Open ShortAlarm not occurred Short Open

Annunciator signal output

Status TerminalNumberA1 to A2 A2 to A3

Power“off” Open ShortDuring purging Open ShortOverride switch pressed * Open ShortAfter purging Short Open*: Overriding

NR800 has Overriding function. Normally power supplyisautomaticallyshutoffwhenyouopenthedoor of the pressurized explosion-proof enclosure( on the analyzer). However, Overriding function allows you to open the door of explosion-proof sectionwithoutautomaticpoweroff

A switch of the overriding function is located insidetheanalyzer(flameproof).Youcanusetheoverriding to open the door of the explosion-proof section.

When opening the door, always make sure the ambient atmosphere is not hazardous. Use gas sensors to detect and eliminate all hazards.

2.1.4Operatingmodes*1Basicoperatingmodeand

descriptionChanneloperatingmodeNo. Auto/Manual.*3

Maintenance *2

Spectrum analysis of a reference sample, equipment maintenance

- -

RunOn-line measurement (allows spectrum analysis on selected channels)

1 AUTOMANUAL

2 AUTOMANUAL

3 AUTOMANUAL

4 AUTOMANUAL

Notes *1: When the power is turned on, the analyzer starts accordingtoapredefinedoperatingmode.

*2: Can perform spectrum analysis (not continuous measurement).

*3: Auto: Performs continuous measurement Manual: Can perform spectrum analysis (not

continuous measurement).2.1.5 ChangingandSettingOperatingModeSome operations are prohibited depending on the user level.

User levelDescription Changing/

settingmodeAbbreviationUser A UA For operator Unauthorized

User B UB For equipment supervisor Authorized

User C UC For maintenance Authorized

• A user level can be switched on the front console panel of the analyzer or from SPECTLAND2 screen of an engineering PC.

• A password is required to switch levels from UA to UB and switch from UA or UB to UC.

2.1.6 SampleMeasurement(1)Continuousmeasurement

Number of measuring channels: 1to4(Specificationisrequired.)Data updating period: 4 seconds or longer (depending on the

number of averaging and measuring channels, as well as measurement items)

Number of measurement items: Up to 12 per channel (48 max. per analyzer)Number of outlier detection items: Up to 12 per channel (48 max. per analyzer)

Separate maintenance is available for each channel (except for common hardware).

(2)Streamswitchingbyinputsignal(intermittentmeasurement)

Calibration model set for the each channel can be changed by external input signal. This function is used for multi- stream application or multi sample application.

Stream switchingNumber of streams to switch: Up to 16Switching patterns: See table 1. “Stream numbers

assignable to measuring channels corresponding to switching patterns”

Contact inputs: See section “2.2 I/O Unit”RS-422 communication: See section “2.1.8 MODBUS”

Data updating period: 8 seconds or longer (depending on the

switching pattern, numbers of averaging and measuring channels, and stream configuration)

Number of measurement items: Up to 12 per stream (64 max. per analyzer)Number of outlier detection items: Up to 12 per stream (64 max. per analyzer)Separate maintenance is available for each channel

(except for common hardware).Input signals and measurement computation sequenceStreamidentification: Used to identify the selected sample.

The analyzer will choose a calibration model set to suite for the relevant stream based on this signal.

Measurement enabled (valid sample): When closed, the analyzer performs

measurement using the calibration modelspecifiedbythesampleidentificationsignalabove.

Schematictimingchartofmeasurementandcomputationsequence(e.g.2streams)

Stream 1 Stream 2 Stream 1

Computationperformed

Computationperformed

Computationperformed

OpenClosed

Closed

Open

Stream identification*

Measurement enabled

Notes(*) Astreamnumberisdefinedbyacombinationofopen/closedstatesofanidentificationcontactsignal.

Oct. 01. 2019-00

Page 6: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

6

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

Table1 Streamnumbersassignabletomeasuringchannelscorrespondingtoswitchingpatterns

Streamswitch MeasuringchannelNo.*3Totalstreams Applicable

channelNo.Case PatternNo. 1 2 3 4None 0 1 2 3 4 4 max 1 to 4

Stream switching per channel *2

1 *1 1 to (17-N) 18-N 19-N 20-N 16 max 1 to 42 1 to 8 9 to 16 None None 16 max 1 to 23 1 to 4 5 to 8 9 to 12 13 to 16 16 max 1 to 4

Notes*1: N: Maximum number of measuring channels included within the analyzer.*2: The stream number for a measuring channel that does not switch paths must be the smallest number in the relevant

column.*3: Measuring channel numbers that equal the number of measuring channels included within the analyzer or smaller are valid.

LED display: The display depends on the operating

status.Run mode

Basic operating mode LEDs: Only RUN lights up.Channel operating mode LEDs: Lit in accordance with each setting.Self-diagnosis LEDs: Lit in accordance with the results of

self-diagnosis.LED display: The display depends on the operating

status.

2.1.8 MODBUS

(1)Communicationspecifications

Item SpecificationCommunication standard*

RS-422 or RS-232C (a converter is required)

Start bit 1Number of data bits 7 (ASCII mode) or 8 (binary mode)Parity bit 1Stop bit 1Communication speed 4800, 9600 or 19200 (selectable)

Error detection Odd number parity, Even number Parity or none (selectable)

*: Only RS-232C can be used in NR805*G explosion-proof type.

See section 1.2 and 2.5

(2)Serialcommunication

Item ASCIImode RTUmodeNumber of data bits 7 (ASCII) 8 (binary)

Message starting character Colon “:” None

Message ending character

Carriage return/Line feed “<CR><LF>” None

Error detection LRC (Logical Redundancy Check)

CRC-16 (Cyclic Redundancy Check)

Inter-character time out None 100msec

Oct. 01. 2019-00

2.1.7 ConsolePanel

(1)Displayitems

Operating mode LEDs (MODE) The following LEDs indicate the current

basic operating mode:RUN: Lit when in the Run mode.MAINT: Lit when in the Maintenance mode. The operating modes of each channel

are displayed by LEDs when in the Run mode.

Two LEDs are provided for each channel, amounting to a total of 8 LEDs.

AUTO: Lit when in the Auto mode.MAN: Lit when in the Manual mode.

Self-diagnosis LEDs (DIAGNOSIS)ALARM: Lit when an alarm occurs.LAMP: Lit when a lamp has burned out or after

atimeperioddefinedbytheservicelifesetting elapses.

LASER: Lit when the laser has burned out orafteratimeperioddefinedbytheservice life setting elapses.

LED display (16 digits, STREAM/CONSTITUENT/VALUE)

The content depends on the operating status or operation.

Operation keys The following six keys are provided:

ITEM: Used to change items.ENT: Usedtoconfirmtheentry.Arrow keys: Used to move the cursor or change

display.

(2)Analyzerbehaviorforeachbasicoperatingmode

Maintenance modeBasic operating mode LEDs: Only MAINT lights up.Channel operating mode LEDs: Lit in accordance with each setting.Self-diagnosis LEDs: Lit in accordance with the results of

self-diagnosis.

Page 7: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

7

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(3)SupportfunctionintheMODBUSprotocol

Function No. Description

01 Reads the ON/OFF status of coils.02 Reads the ON/OFF status of input relays.03 Reads the current value of holding registers.04 Reads the current value of input registers.05 Forcibly changes the status of a coil.06 Write a value to a holding register.

08Loop-back testSends back the same message as the command message.

(4)Recommendedvalueofmasterside

Item RecommendedvalueTransmission speed (bps) 4800/9600/19200Time monitor of Inter-character gap 1000msNo reply time 2 sec.Number of transmission retries 5 timesCommunication Recovery Period 30 sec.

Oct. 01. 2019-00

2.1.9 OtherFunctions

(1)Baselinecompensation:Upto10points

(2)Outputoftheresultcompensatedusinganalogue inputs

The following operation result can be output.Compensation operation processing:

P2 = C1 * P1 + C0 + D D = An2 * D2 + An * D1 + D0

An: Analogue input (n: 4 or less)P1: Measurement value before the

compensationP2: Measurement value after the

compensationD0,D1,D2,C0,C1: Coefficient

(3)OutputsmoothingSimple moving average (Up to 10 points)Exponential moving average

(4)On-linemeasurementspectrasaving

2.1.10ModelandSuffixCodes

(1)NR801AG,Generalpurposemodel

[Style:S2]Model SuffixCode OptionCode Description

NR801AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, General Purpose modelLanguage -E ••••••••••••••••• EnglishPower supply 1

346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelength scanning range W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDual cable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1

2

•••••••••••••••••

•••••••••••••••••

Electric cable, only for general purpose model and less than 40 mFiber-optic cable

Mounting ABC

•••••••••••••••••••••••••••••••••••••••••••••••••••

Without bracketsWith wall-mounting bracketsWith free standing rack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS With stream switch input function

Page 8: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

8

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(2)NR805AG,TypeXpurgedandFMapprovedexplosion-proofmodel

[Style:S2]Model SuffixCode OptionCode Description

NR805AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion proof model, FM approved version

Language -E ••••••••••••••••• EnglishPower supply 1

346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N"Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelength scanning range W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDual cable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A

BC

•••••••••••••••••••••••••••••••••••••••••••••••••••

Without bracketsWith wall-mounting bracketsWith free standing rack

- -N ••••••••••••••••• Always “-N"- -N ••••••••••••••••• Always “-N"Cable entrance -2 ••••••••••••••••• Female 3/4NPTPurge air connection 2 ••••••••••••••••• Female 1/4NPTOption /SS With stream switch input function

Oct. 01. 2019-00

Page 9: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

9

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(3)NR801JG,Generalpurposemodel

[Style:S2]Model SuffixCode OptionCode Description

NR801JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, General Purpose modelLanguage -E

-J••••••••••••••••••••••••••••••••••

EnglishJapanese

Power supply 1346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelength scanning range W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDual cable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1

2

•••••••••••••••••

•••••••••••••••••

Electric cable, only for general purpose model and less than 40 mFiber-optic cable

Mounting ABC

•••••••••••••••••••••••••••••••••••••••••••••••••••

Without bracketsWith wall-mounting bracketsWith free standing rack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS With stream switch input function

Oct. 01. 2019-00

Page 10: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

10

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(4)NR805JG,TIISapprovedexplosion-proofmodel*1

[Style:S2]Model SuffixCode OptionCode Description

NR805JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion-proof model, TIIS approved *1

Language -E-J

••••••••••••••••••••••••••••••••••

EnglishJapanese *1

Power supply 1346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelength scanning range W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDual cable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A

BC

•••••••••••••••••••••••••••••••••••••••••••••••••••

Without bracketsWith wall-mounting bracketsWith free standing rack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Cable entrance -1

-2-3-4

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Metal conduit (G3/4)Metal conduit (3/4NPT)Flameproof packing (G3/4)Flameproof packing (3/4NPT)

Purge air connection 12

••••••••••••••••••••••••••••••••••

Female Rc1/4Female 1/4 NPT

Option /SS/K

With stream switch input functionKOSHA approved explosion proof model *1

*1: “TIIS” approval is not applicable when option code “/K” is selected. “Japanese” can not be selected with option code “/K.”

Oct. 01. 2019-00

Page 11: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

11

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.1.11Externaldimensions

(1)Generalpurposemodel(NR801*Gtype)•Withoutbrackets(Suffixcode:A)

6282 60 60 60 60 62 62

68±5

30±5

480

92

60430

277

665

355550

Mounting screws4-M8

2-cable outlets (For NR801AG)(Packing glands, acceptable cables: ø10 to 13)

Earth terminal(M4 screw)

Power supply cable connection inlet (For NR801AG)(Packing gland, acceptable cable: ø10 to 15.9)

3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)

Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)

4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)

Unit: mm

For NR801AG

Part A

Part A

Note: The wall shall be designed to withstand four times the weight of NR800.

Oct. 01. 2019-00

Page 12: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

12

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

• Wall-mounting model(Suffixcode:B)30

±566

5550

60 430

50 50

603086

0

355 20

920

6282 60 60 60 62 6260

277

68±5

Mounting holes4-ø10.5 holes

Earth terminal(M4 screw)

Power supply cable connection inlet (For NR801AG)(Packing gland, acceptable cable: ø10 to 15.9)

2-cable outlets (For NR801AG)(Packing glands, acceptable cables: ø10 to 13)

3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)

Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)

4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)

Wall mounting bracket

Unit: mm

For NR801AG

Part A

Part A

Note: The wall shall be designed to withstand four times the weight of the NR800.

Oct. 01. 2019-00

Page 13: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

13

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

• Free standing rack model(Suffixcode:C)66

5

1500

70

350 8080

30±5

68±5

550 355 50

630

550

490

56827

7

6282 60 60 60 62 6260

Mounting holes 4-ø17 (for anchor bolts)

4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)

Earth terminal(M4 screw)

Unit: mm

Power supply cable connection inlet (For NR801AG)(Packing gland, acceptable cable: ø10 to 15.9)

2-cable outlets (For NR801AG)(Packing glands, acceptable cables: ø10 to 13)

3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)

Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)

For NR801AG

Part A

Part A

Oct. 01. 2019-00

Page 14: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

14

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(2)Explosion-proofmodel(NR805*Gtype)• Without brackets(Suffixcode:A)

62

82180

93

36

6262332

665

355550

App

rox.

300

200

430 60

9248

0

4-M8Mounting screws

95

Internal-pressure protection unit(flameproof enclosure)Cable connectors(with sealing fitting orflameproof packing gland *1)

Air (purge gas) connection: Rc1/4 or 1/4NPT(only 1/4NPT for NR805AG)

Connector *3

Connection:3/4NPT

For NR805AG

Earth terminal(M4 screw)

86

277

216

65

4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)

2-cable outlets *2 *3(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)

Special Ethernet fiber-optic cable outlet

Unit: mm

For NR805AGPart B

Part A

Part A

Part B

*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: InthecaseofNR805AG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*3: The connector of NR805AG is 3/4NPT.Note: The wall shall be designed to withstand four times the weight of the NR800.

Oct. 01. 2019-00

Page 15: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

15

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

• With wall-mounting brackets(Suffixcode:B)

62

App

rox.

300

36

82180

6093

6262332

665

550

60 430

50 50

3086

0

200

For NR805AG

Connection: 3/4NPT

Mounting holes 4-ø10.5 holes

Internal-pressure protection unit(flameproof enclosure)

Cable connectors(with sealing fitting or flameproof packing gland *1)

Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)

4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)

95

Earth terminal(M4 screw)

86

277

21665

Special Ethernet fiber-optic cable outlet

2-cable outlets *2 *3(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)

920

355 20

Wall mounting bracket *4

Connector *3

Unit: mm

For NR805AG

Part A

Part A

Part B

Part B

*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: InthecaseofNR805AG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*3: The connector of NR805AG is 3/4NPT.*4: The wall shall be designed to withstand four times the weight of the NR800.

Oct. 01. 2019-00

Page 16: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

16

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

• With free standing rack(Suffixcode:C)

568

490

550

630

70

1500

665

App

rox.

300

80

200

550

Mounting holes 4-ø17 (for anchor bolts)

36

93

Connection: 3/4NPT

For NR805AG

For NR805AGInternal-pressure protection unit(flameproof enclosure)

Cable connectors(with sealing fitting or flameproof packing gland *1)

Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)

82180

350 80

95

Connector *3

50355

Bottom of analyzer

277

216

8665

Earth terminal (M4 screw)

62332 62 62

4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)

2-cable outlets *2 *3(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)

Special Ethernet fiber-optic cable outlet

Unit: mm

Part B

Part A

Part A

Part B

*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: InthecaseofNR805AG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*3: The connector of NR805AG is 3/4NPT.

Oct. 01. 2019-00

Page 17: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

17

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.2 I/OUnit2.2.1 OverviewThe I/O unit is an input/output interface between the analyzer and external monitor/control systems such as DCS. The I/O unit connects with NR800 analyzer via serial communications (RS-422). Up to two I/O units can be connected to the analyzer, and one of which can incorporate the contact input for stream switching and analogue input.

2.2.2 SpecificationsPowersupply: Seesection“3.UtilitySpecifications”.Insulationresistance: 5MΩormore,500VDCWithstandvoltage: 1500VACfor1min.Analogue output:

Output data: Measurement results (Properties and concentration)

Number of outputs: 0 to 40Outputspecifications:

Item Description

Output range 4 to 20 mA DC (3.0 to 21.0 mA DC, Floating-common type)

Isolation methodBetween output terminals and internal circuit: Photo coupler isolationBetween output terminals: non-isolated, Common negative

Withstand voltage 500VDCfor1min.Allowable load resistance 600Ωorless12-bit D/A converter resolution 5.7 µA

System accuracy ±0.5%offullscale

Separately,24VDCpowersupplyisrequired (see section “2.2.3. Model and Suffixcodes”and“3.UtilitySpecifications”).

Connect a short-cable (*) to unused analog output terminal between [OUT+] and [OUT-]. If not, the ALM indicator blinks of an analog output module.

(*)Cablespeccifications: Size: More than 0.2 mm2 (AWG24) Voltagerating: Morethan300V Temperature rating: More than 60°C Flamerating: EquivalenttoVW-1 Screw size of terminal on the side of

the analog output module: M3.5Contact output:

Alarm Quantityofoutputs Alarm Quantity

ofoutputsGeneral 1 Outlier 4Communication failure 1 I/O unit failure 1

Operating mode 5 - -

Oct. 01. 2019-00

Contactoutputspecifications:Item Description

I/O unit failure

Rating 24VDC.0.3A

Action and number1 normally open and 1 normally closed (shared common)

Alarm, operating mode, and outlier detection

Insulation method Mechanical isolationRated load voltage

DC 24VAC 100to240V

Maximum load current 2 Amps/point, 8 Amps/common

Servicing life

Mechanical At least 20 million actions

Electrical At least 100 thousand actions

Surge killer NoneNumber per common 8 points/commonExternal power supply Not required.

Contact input:Description:Streamidentificationformulti-streamsamplingunit: 8 points, status signals to identify

sample streams that pass through measurement cells.

Measurement enabled (stream valid): 4points,statussignalstoconfirmthat

samples inside the measurement cells are ready for measurement.

Inputunitspecifications:Item Description

Input type VoltagefreecontactCommon terminal Common to 8 pointsIsolation method Transformer isolation

Withstand voltageBetween external connectors collectively and internal circuit:500VDCfor1min.

Off-stateopen-circuit voltage 5to7VOn-state load current 1 to 3 mAOn-state load resistance 200ΩorsmallerOff-stateloadresistance 100Ωorlarger

Provide external contact-input signals meeting the above requirements.

Analogue input:Input data and number: Analogue output compensation signal, 4 pointsInputspecifications:

Item DescriptionInput range (actual) 1to5VDC(-0.25to5.25VDC)

Isolation methodBetween input terminals and internal circuit: photo coupler isolationBetween input terminals: non-isolated, negative common

Withstand voltage 500VDCfor1min.Input resistance 1MΩ12-bit A/D converter resolution 1to5VDC:1.4mV

System accuracy ±0.5%offullscaleat0to40°C

Installation requirements: See chapter 4.Mounting: Wall mountingExternal connection terminal: M3.5 screwDrip-proof and dust-proof construction: IP4X (with enclosure)

Page 18: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

18

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.2.3 ModelandSuffixcodes

(1)NR893AG:forNR801AGandNR805AGanalyzer

[Style:S2]Model SuffixCode OptionCode Description

NR893AG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IO unit (non explosion-proof)Language -E ••••••••••••••••• EnglishAnalyzer 1

2••••••••••••••••••••••••••••••••••

NR805AG (Explosion-proof model) (*1)NR801AG (General Purpose model)

Contact input for stream switching

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Analogue input for compensation

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Number of analogue outputs -00-04-08-12-16-20-24-28-32-36-40

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

None481216202428323640

Enclosure -A-B-C

•••••••••••••••••••••••••••••••••••••••••••••••••••

Number of analog outputs: 0 to 12Number of analog outputs: 0 to 28Number of analog outputs: 0 to 40

- -N ••••••••••••••••• Always “-N”

*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.

(2)NR893JG:forNR801JGandNR805JGanalyzer

[Style:S2]Model SuffixCode OptionCode Description

NR893JG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IO unit (non explosion-proof)Language -E

-J••••••••••••••••••••••••••••••••••

EnglishJapanese

Analyzer 12

••••••••••••••••••••••••••••••••••

NR805JG (Explosion-proof model) (*1)NR801JG (General Purpose model)

Contact input for stream switching

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Analogue input for compensation

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Number of analogue outputs (*2)

-00-04-08-12-16-20-24-28-32-36-40

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

None481216202428323640

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Option /C Withenclosureincluded24VDCpowersupplyunit

for analog output. (*2)

*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.

*2: Separate24VDCpowersupplyrequiredwhenusingananalogueoutputandoptioncode“/C”arenotspecified.(seechapter 3).

Oct. 01. 2019-00

Page 19: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

19

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.2.4 Externaldimensionsandweight:

(1)NR893AGPaint color: Light beige (Munsell 5Y 7/ 1 or equivalent)

B D F

CA

E G

H

75 17

16118°

1743

5028

0

400

68.5

200

6100 505050505050

L + 50

L50 50

4-ø14 holesfor mounting *1

ø20 holefor protective ground wiring

Number ‘N’ of wiring holesSuitable cable OD *2 A: ø10 to 15.9 (to the power supply) B to H: ø10 to 18.5

Unit: mm

Enclosure code

Numberofanalog outputs L(mm) N Approximate

Weight-A 0 to 12 400 6 (A to F) 15.5 kg-B 0 to 28 500 7 (A to G) 17.5 kg-C 0 to 40 600 8 (A to H) 20.0 kg

*1: The wall shall be designed to withstand four times the equipment's weight.*2: Sinceattachedferritecoresareused,thepracticalcableODareuptoø13mm.

Oct. 01. 2019-00

Page 20: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

20

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(2)I/OunitWithoutenclosure(onlymodelNR893JG)• For Explosion proof analyzer unit•ThesizeofIOunitforgeneralpurposeanalyzerunitisthesameasabovefigurebutRS-422/232Cconverterpower

supply case removed.Paint color: Baseplate, RS-422/232C converter power supply case;

Module colors: Light cobalt blue (Munsell 6.2PB4.6/ 8.8 or equivalent), lamp black (Munsell 0.8Y2.5/0.4 or equivalent)

215

199

L

303 151510

(101)Approx. 120

N-ø7.2 holesfor mounting *1

Safety barrier power supply caseused only with NR805JG explosion-proof analyzer

Unit: mm

Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 333 4 6.5 kg-16, -20, -24, -28 16, 20, 24, 28 450 6 7.0 kg

-32, -36, -40 32, 36, 40 537 6 7.5 kg

*1: The wall construction should be withstood a force of four times the weight.

(3)Withenclosure(Optioncode:/C)Paint color: Light beige (Munsell 5Y 7/ 1 or equivalent)

C E

D

A

F H

G

5050

B

5050 50505075 17

10118°

1743

5028

0

400

App

rox.

30

200

6

L + 50

L50 50

4-ø14 holesfor mounting *1

ø20 holefor protective ground wiring

Number ‘N’ of wiring holesSuitable cable OD: ø10 to 18.5

Unit: mm

I/O Unit (inside)

Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 400 6 (A to F) 15.5 kg-16, -20, -24, -28 16, 20, 24, 28 500 7 (A to G) 17.5 kg

-32, -36, -40 32, 36, 40 600 8 (A to H) 20.0 kg

*1: The wall shall be designed to withstand four times the weight of the NR800

Oct. 01. 2019-00

Page 21: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

21

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.3MeasurementCells

2.3.1 FlowthroughCell

(1)SpecificationsOptical path length: 1, 2, 5, 10, or 20 mmFiber-optic connector: FC typeWetted part material: Fluoro rubber, Kalrez 4079, 316SS, PTFE, Borosilicate crown glass, or SapphireSample pressure range: Atmospheric pressure up to 1.9 MPa GSample temperature range: –5°C to +80°C (It varies by the materials of O-ring. Contact Yokogawa for further

information.)Sample connection: Female Rc1/4Installationangle: VerticalInstallation location requirements: See section 4.Weight: Approx. 3 kg

(2)ModelandSuffixCodes

[Style:S1]Model SuffixCode OptionCode Description

NR510 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cellWindow material and optical path length

-B00

-B01-B02-B05-B10-B20-S00

-S01-S02-S05-S10-S20

•••••••••••••••••

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Borosilicate crown glass, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm

Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ring material -B

-K••••••••••••••••••••••••••••••••••

Fluoro rubberKalrez 4079

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”

(3)ExternalDimensions

Unit: mm(96)55 29

102

92

L4-ø5.2 holes

2-Rc1/4

Optical path length (mm)

1251020

L (mm)

64±2

65±2

68±2

73±2

83±2

Fiber-optic connection(in the other side)

Oct. 01. 2019-00

Page 22: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

22

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.3.2 FlowthroughCellwithConstantTemperatureWaterTube

(1)SpecificationsOptical path length: 1, 2, 5, 10, or 20 mmFiber-optic connector: FC typeWetted part material: Fluoro rubber, Kalrez 4079, 316SS, PTFE, Borosilicate crown glass, or SapphireSample pressure range: Atmospheric pressure up to 1.9 MPaSample temperature range: +5°C to +80°C (It varies by the materials of O-ring. Contact Yokogawa for further

information.)Constant water temperature range: +5°C to +80°C (It varies by the materials of O-ring. Contact Yokogawa for

further information.)Sample connection: Female Rc1/4Connection for water with constant temperature: 6-mm outside diameter, copper tubeInstallationangle: VerticalInstallation location requirements: See section 4.Weight: Approx. 3 kg

(2)ModelandSuffixCodes

[Style:S1]Model SuffixCode OptionCode Description

NR512 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cell with constant temperature water tube

Window material and optical path length

-B00

-B01-B02-B05-B10-B20-S00

-S01-S02-S05-S10-S20

•••••••••••••••••

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Borosilicate crown glass, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm

Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ring material -B

-K••••••••••••••••••••••••••••••••••

Fluoro rubberKalrez 4079

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”

(3)ExternalDimensions

Unit: mm

4-ø5.2 holes

2-Rc1/4

L29

Approx. 160

App

rox.

160

102

92

B-SIDEB-SIDE A-SIDE A-SIDE

Thermal insulation jacket

Approx. 145

70111

Optical path length (mm)

1251020

L (mm)

64±2

65±2

68±2

73±2

83±2Fiber-optic connection (in the other side)

Oct. 01. 2019-00

Page 23: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

23

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.3.3 SpecialmeasurementcellThisclausedescribesPFAflow-throughtypespecialcell.

(1)SpecificationsOptical path length: 1, 2, 5, or 10 mmFiber-optic connector: FC typeWetted part material: PFA,Kalrez4079,Moriseiperflioro(MP4275B),orSapphire (Stainless steel is partly used as non-wetted parts.)Sample pressure range: Atmospheric pressure up to 0.3 MPa GSample temperature range: 5°C to 80°CSample connection: 1/4 inchs (Ø6.35 x Ø3.95 mm) PFA tube, or Ø6 x Ø4 mm PFA tube (length: 3000 mm)Installationangle: VerticalInstallation location requirements: See section 4.Weight: Approx. 1.5 kg

(2)ModelandSuffixCodes

[Style:S1]Model SuffixCode OptionCode Description

NR511 ••••••••••••••••••••••• ••••••••••••••••• Special measurement cellWindow material and optical path length

-S01-S02-S05-S10

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mm

Body material P ••••••••••••••••• PFAO-ring material -K

-P••••••••••••••••••••••••••••••••••

Kalrez 4079Moriseiperflioro(MP4275B)

Tube -A-B

••••••••••••••••••••••••••••••••••

1/4 inchs (Ø6.35 x Ø3.95 mm), 3000 mm, PFAØ6 x Ø4 mm, 3000 mm, PFA

Option /Z Always add “/Z”

(3)ExternalDimensions

Unit: mm(80)(50)

90

(300

0)(3

000)

74

77(94)

4-ø5.2 holes

Fiber-optic connection(in the other side)

Oct. 01. 2019-00

Page 24: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

24

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.3.4 In-situprobe

(1)SpecificationsFiber-optic connector: FC typeWet part material:

Body; Hastelloy C-276 or UNS S31603 (Stainless steel 316L)Others; Sapphire, Hastelloy C-276, Gold

Sample pressure range: Atmospheric pressure to 0.98 MPa *Sample temperature range: Normal temperature to 80°C *Purge gas connection: Female 1/8 NPT x 2Fiber-optic entrance: Female 1 NPTInstallationangle: Vertical(Insertportionisdownward)orHorizontalInstallation location requirements: See section 4Weight: Approx.5kg(excludingflange)*:*: ThespecificationsaboveapplytothecasewhentheflangeisJIS10K. If you use sample pressure or sample temperature

outsidetherangeshownabove,pleasecontactoursalesrepresentativesregardingthespecificconditionsofyourdemand.Wewillprovidesomeadequateproposalswhichfityourparticularneeds.(Installationexample10MPa,300°C)

(2)ModelandSuffixCodesFIR200/Z

(3)ExternalDimensions

L *3

( 560 ) *4 ( 80 )

Approx. 35 A *1

Flange *2

Fiber-optic connection box

Fiber-optic entrance(Exclusive packing)

Purge-gas connections2-1/8NPT

Approx. ø27

Unit: mm

*1: Optical path length “A”: 2/5/10 mm, selectable (Dependent on sample conditions, measurement items, measurement range, etc. Consult Yokogawa.)

*2:Mountingflange(HastelloyC-276/UNSS31603comparable)shouldbeprovidedbycustomer.*3: Specify length “L” within the range 150 to 250 mm.*4: The shape and size of connection box may change.

Note1:Keepthetemperatureattheconnectionofthefiber-opticcableat80°Corbelow.Ifthetemperatureattheconnectionexceeds 80°C, the connection box should be purged with clean, dry air, or other appropriate measures should be taken. Inthiscase,theconnectionboxmaybedifferentinshapeandsizefromtheoneshownabove.Ifafiber-opticcableformeasurement is exposed to radiant heat, it should be shielded.

Note 2: Dimensions of each part may change without prior notice.

2.4 Fiber-opticCablesformeasurementFiber-optic cables connect an analyzer to measurement cell. Select appropriate cables to meet your application requirement, e.g., wavelengths scanning range.

2.4.1 SilicaFiber-opticCable

(1)SpecificationsApplicable wavelength range: 900 to 2100 nmConnector: Double-end FC or FC-SMA typeStructure: Dual (for sample and reference) or single (sample), two-core, protected by resin-

coatedstainlessflexibletubeMinimum bending radius: 100 mm. To reduce optical attenuation, make the radiuses along the cable as

large as possible when laying cables.Installation location requirements: See section 4.Cabling: Conduit protected cabling is recommended.

Oct. 01. 2019-00

Page 25: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

25

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(2)ModelandSuffixCodes

a.Singlefiber-opticcable[Style:S1]

Model SuffixCode OptionCode DescriptionNR821 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2100nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-end FC typeFC type on analyzer side and SMA type on measurement cell side

Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

5 m10 m20 m30 m50 m100 m150 m200 m300 m

- -000 ••••••••••••••••• Always “-000”

b.Dualfiber-opticcable[Style:S1]

Model SuffixCode OptionCode DescriptionNR822 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2100nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-end FC typeFC type on analyzer side and SMA type on measurement cell side

Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

5 m10 m20 m30 m50 m100 m150 m200 m300 m

- -000 ••••••••••••••••• Always “-000”

(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.

• Double-endFCconnector

BA

A B

Unit: mmFC connector FC connector

(Cable length)600±50

(57) 3010

1510

• FC(analyzer)-SMA(measurementcell)connector

BA

A B

Unit: mmFC connector (analyzer) SMA connector

(measurement cell)

(Cable length)600±50

(57) 3010

600±50

(53) 8

1510

Oct. 01. 2019-00

Page 26: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

26

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.4.2 FluorideFiber-opticCable(1)Specifications

Applicable wavelength range: 900 to 2500 nmConnector: Double-end FC type or FC-SMA typeStructure: Dual (for sample and reference) or single (sample), two-core, protected by resin-

coatedstainlessflexibletubeMinimum bending radius: Analyzer side; 100 mm, Measurement side;120 mm To reduce optical attenuation, make the radiuses along the cable as large as

possible when laying cables.Installation location requirements: See section 4.Note: Fluoride,amaterialusedintheNR823andNR824fiber-opticcables,hasthedeliquescenceproperties.Whenstoring

these cables, take appropriate measures to prevent moisture, e.g., a cable should be placed with desiccant in a sealed moisture-barrier plastic bag. Replace the desiccant regularly.

(2)ModelandSuffixCodes(ModelcodetoincludefibersforAnalyzer–Cell(probe)–Analyzer)a.Singlefiber-opticcable

[Style:S1]Model SuffixCode OptionCode Description

NR823 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2500nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-end FC typeFC type on analyzer side and SMA type on measurement cell side

Cable length -L003-L004-L005

•••••••••••••••••••••••••••••••••••••••••••••••••••

3 m4 m5 m

- -000 ••••••••••••••••• Always “-000”

b.Dualfiber-opticcable[Style:S1]

Model SuffixCode OptionCode DescriptionNR824 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2500nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-end FC typeFC type on analyzer side and SMA type on measurement cell side

Cable length -L003-L004-L005

•••••••••••••••••••••••••••••••••••••••••••••••••••

3 m4 m5 m

- -000 ••••••••••••••••• Always “-000”

(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.(Connectthelongerendtothe measuremrnt cell side.)• Double-endFCconnector

BA

A B

Unit: mmFC connector (analyzer) FC connector

(measurement cell)

(Cable length)600±50 800±50

(57) 3010

1510

• FC(analyzer)-SMA(measurementcell)connector

BA

A B

Unit: mmFC connector (analyzer) SMA connector

(measurement cell)

(Cable length)600±50

(57) 3010

800±50

(53) 8

1510

Oct. 01. 2019-00

Page 27: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

27

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.5 RS-422-to-RS-232CConverter(PartNumber:K9404LAorK9404LD)This unit receives data from the analyzer via RS-422 and converts them to RS-232C serial interface, therefore it helps the data transmission between the analyzer and I/O unit or monitor/control system. This unit also blocks communicationsignalswhenitreceivesapower-offsignalfromtheanalyzer,thusensuringtheexplosion-proofintegrity.

2.5.1 SpecificationsPower supply: See section 3.Signal terminals: Analyzer main unit side (RS-422): M4, output side (RS-232C): Female Dsub 25-pinGroundingtype: 100ΩorlowerInstallation location requirements: See section 4.Housing structure: Desktop (K9404LA), Screw cramp (K9404LD)Weight: Approx. 2 kgNote: The converter does not carry the CE Marking. An appropriate housing should be provided by customer, if necessary.

2.5.2 TerminalDiagram

CN3 CN2

LN-+RDBRDASDBSDA

RS-422

From analyzer

D/I 12V AC power supply

Female D-SUB 25-pin connectorCN1

RS-232C

2.5.3 RS-232CCommunicationStraightCable

GND

123456789

CDRDSDERSGDRRSCSCI

FGCDRDSDERSGDRRSCSCI

GND183220764522

Internal connection

To converter To host

2.5.4 ExternalDimensions

4-ø4.5 holes203

126

126

39.5

15 15

9611

.5 218

203

37

K9404LA Desktop type K9404LD Screw cramp type

Unit: mm

Oct. 01. 2019-00

Page 28: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

28

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.6 EthernetFiber-opticCableCable for Ethernet communication between the analyzer and the engineering PC.

2.6.1 SpecificationsLength: Up to 20 mConnector: ST typeStructure: Two-core,protectedbystainlessflexibletubeInstallation location requirements: See section 4.Minimum bending radius: 50 mm. To reduce optical attenuation, make the radiuses along the cable as

large as possible when laying cables.

2.6.2 ModelandSuffixCodes[Style:S1]

Model SuffixCode OptionCode DescriptionNR895 ••••••••••••••••• ••••••••••••••••• Ethernetfiber-opticcableLength *1 -L003

-L005-L010-L020

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

3 m5 m10 m20 m

- -000 ••••••••••••••••• Always “-000”Option /JB With junction box *2

*1: Foratotalcabledistanceof20morlonger,anadditionalfiber-opticcable(fittedwithSTconnectors)mustbeprovidedbythe user.

Seechapter6.2forthecablespecificationofuserprovisions.*2: TheJunctionboxisforconnectionbetweenanNR895fiber-opticcableandtheadditionalfiberonesprovidedbytheuser.

2.6.3 ExternalDimensions(1)Fiber-opticCable

15

9.310

Blue

Yellow

Unit: mm

1000 0+100

43 30

(Cable length)

Oct. 01. 2019-00

Page 29: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

29

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

(2)JunctionBoxforEthernetfiber-opticcables(Optioncode:/JB)

400 430

305

300

142

24120

280

60

20260

4-ø10

Handle to open

100 30Hole

2-Cable gland

Unit : mm

NR895_JB.ai

Housing structure: IP44Coating: Epoxy resin coatingPaint color: Light beige (equivalent to Munsell 5Y7/1)Material: IronWeight: Approx. 10 kg

Oct. 01. 2019-00

Page 30: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

30

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.7 SPECTLAND2MeasurementandMaintenanceSoftware

2.7.1 OverviewSPECTLAND2 is an application software that controls NR800 operation and monitoring on a PC. It displays the analyzer status, instructs measurement and sets parameters for the analyzer. To use SPECTLAND2, firstinstallitintheengineeringPC.The continuous operation of SPECTLAND2 for extended periods of time is not guaranteed.

MainFeatures• Instructs the analyzer to measure spectra, stores

data and displays. The measured spectra can be used for calibration modeling.

• Shows trend graphs of the measurements and saves themtofilesduringacontinuousmeasurement.

• Sets parameters for continuous measurement.• Displays various data of the analyzer, such as

operating modes, alarm status, and maintenance data.

• Instructs the analyzer to perform operation mode change, spectra measurement, property value setting, calibration models, and setting measurement conditions.

2.7.2 MainWindows(1)ManualSpectrumWindow

Enables the analyzer to measure spectra, which are to be processed by Chemometrics (Calibration Modeling software). Also allows data to be saved to filesanddisplayed.

(2)AutoSpectrumWindowAllows the user to upload spectra data to the Engineering PC during continuous measurement upon receiving a signal at periodic intervals, outlier detection, or a property value variation failure. This data is displayed for each measuring channel.

(3)PowerSpectrumWindowIt displays power spectra. This window is available for UB level (equipment supervisor) users.

(4)InterferogramWindowIt displays interferogram data collected. This window is available for UB level users.

(5)Real-timeTrendWindowsDisplay measurement values of Nos. 1 to 6 and Nos. 7 to 12 components in two separate trend graph windows for each stream. Up to 10 windows can be open at the same time.

(6)HistoricalTrendWindowsDisplay historical trend data saved. Trend data of 24hoursforeachstreamissavedtoafile.Upto4windows can be open at the same time.

(7)ParameterWindowDisplays the current parameter settings for the analyzer. In addition, UB level users can change the settings.

(8)AlarmStatus/HistoryWindowsThe Alarm Status window displays the active alarms for the analyzer, while the Alarm History window displays all the past alarms. The alarm history can be deleted with commands.

(9)Tab-controlledMaintenanceWindowDisplays the A/D reference value and servo-related data of the analyzer. This window is available for UC level users.

(10)Tab-controlledCommunicationStatusWindowDisplays the communication status between the engineering computer and the analyzer. This window is available for UC level (maintenance) users.

2.7.3 ModelandSuffixCodes

Model SuffixCode

Option Code Description

NR831 ••••••••••• ••••••••••• SPECTLAND2 measurement and maintenance software

Language -E-J

••••••••••••••••••••••

EnglishJapanese

- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”Option /UP Versionup

Package contents: One CD-ROM One user’s manual

2.8 ChemometricsSoftware

2.8.1 Specifications• Calibration Modeling technique: Partial least square (PLS) and others

2.8.2 ModelandSuffixCodes

Model SuffixCode

Option Code Description

NR530 ••••••••••• ••••••••••• Chemometrics softwareLanguage -E ••••••••••• English- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”

Package contents: Two CD-ROMs (Install disc/User's manual)

One set of user registration document

Oct. 01. 2019-00

Page 31: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

31

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.9 SamplingUnitSome applications permit direct link between the process and the measurement cell or you don’t need the sampling unit.Pleasefillouttheinquireform(lasttwopages)toshowandcontactourrepresentatives.Wewillselecttheoptimum system to suit the application of your demand.

lExampleofSamplingSystemThefollowingintroducesaconceptualsamplingunitformeasuringtheproperties(RON,RVP,Distillationetc.)ofgasolineblendinginapetroleumrefinery.Notethateachsamplingunitshouldbedesignedforeachapplication;designs will vary.

MeasurementSampleConditionsItem Condition

Fluid to be measured Gasoline

Sample inlet/outlet condition

Inlet pressure 0.3 to 0.9 MPa GOutlet pressure 0.3 to 0.9 MPa GInlet temperature 0°C to +40°C

Notes1: When samples do not contain free water content, a coalescer (CF1) is not required.2: When samples do not have bubbles, a degasser (DG1) is not required.3: Ifthepressuredifferencebetweenthesampleinletandoutletisgreaterthan0.3MPa,asamplingpump(PU1)isnot

required.

com comcom com

PU1CHV1, 2, 3BV1, 2, 3, 4BV5, 6, 7, 8, 9V1, 2, 3, 4, 5, 6NV1, 2, 3, 4, 5FM1, 2, 3, 4PR1, 2CF1DG1P1, 2DT1TK1, 2, 3, 4CON1RN1

PUMPCHECK VALVE3-WAY BALL VALVE2-WAY BALL VALVESTOP VALVENEEDLE VALVEFLOW METERPRESSURE REGULATORCOALESCER FILTERDEGASSERPRESSURE GAUGEDETECTORTANKInfraSpec NR800RESISTANCE TEMP. SENSOR

: JIS 10K-15A-RF: Rc1/2: Rc1/4

<connection>

: SUS316TP 15A sch40: ø12/ø10 SUS316: ø8/ø6 SUS316: ø4/ø3 Nylon tube: ø6/ø4 Nylon tube: ø22/ø6 Braid tube

<piping>

outlet

Cooling waterinlet

*2PR2

Air inlet

*3

*4

*1PR1

PU1Sample inlet

P1

V1

NV1

FM1NV3NV4

BV1

DG1 BV2

CF1

BV9

Sample outlet

P2

V2

FM4FM3

Cooling water

*3 *4

*1

TK1

V3

V4

CH

V2

DT1 BV3

CH

V1

*2NV5 RN1

CON1

TK2 V5V6

a'

NV2FM2

BV4

Drain

TK4

BV

7

TK3

CH

V3

BV

8a'

BV

5

BV

6

SAMPLING UNIT

N2 gas inlet

N2 gas

Oct. 01. 2019-00

Page 32: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

32

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

2.10 AuxiliaryParts*Therecommendedreplacementintervalsforauxiliaryparts,listedinSection2.10.1,ModelandSuffixCodes,areset for preventive maintenance and do not guarantee the service life of parts. Actual replacement intervals for auxiliary parts vary depending on operating conditions, sample conditions, and other factors. Be sure to perform preventive maintenance based on operational status of instrument.Theauxiliarypartsareclassifiedintotwotypes:thosethatcanbereplacedbycustomerandthosethatmustbereplaced by a Yokogawa service engineer.Prior to replacement, refer to the user’s manual and follow the instruction therein.

* Auxiliary parts mean limited life components that are excepted to be in their wear-out failure period within 5 years under normal operating or storage conditions.

Model SuffixCode

Option Code Description

RecommendedReplacement

IntervalNR8SP01 •••••••• ••••••••• Auxiliary parts set for NR800

-01-02-03-04-05-06

-07-08-10-11-12-13-14-15-16-17

••••••••••••••••••••••••••••••••••••••••••••••••••••••

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Sealing packing for analyzer door (for NR801AG/NR805AG/NR801JG/NR805JG)Analyzer cooling fan (5 pcs)Spectrometer support insulator (4 pcs)Lamp (adjusted)Laser headDesiccant for spectrometer (9 pcs)

Desiccantforflow-throughcell(10pcs)*1Sealing packing for analyzer door (for NR805JG/K)O-ringforflow-throughcell(JISP38,FKM(fluororubber),2pcs)O-ringforflow-throughcell(JISP21,FKM(fluororubber),2pcs)O-ringforflow-throughcell(JISP16,FKM(fluororubber),2pcs)O-ringforflow-throughcell(JISP38,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP21,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP16,Kalrez4079,2pcs)Windowpaneofflow-throughcell(borosilicatecrownglass,2pcs)*3Windowpaneofflow-throughcell(sapphire,2pcs)*3

1 year3 years3 years5,000 hours12,000 hours1 year (Shelf life: 1 year after shipment)3 months1 year*1, *2*1, *2*1, *2*1, *2*1, *2*1, *2

*1: When a measurement cell is disassembled for service/cleaning, it is recommended that O-rings should be replaced.*2: Replacement intervals vary depending on sample composition, humidity, pressure, etc. Determine replacement intervals

based on actual operational status.*3: If a window is scratched or fogged, replace the whole set of windows (2 pieces/set).

Please contact Yokogawa, when you need the O-ring for Special measurement cell.

3.UtilitySpecifications3.1 PowerSupplies

(1)NR800analyzerPower supply: 100,115,200,230VAC,singlephase,

50/60 Hz( *1)Voltagefluctuation: Rating±10%,50/60±2HzPower consumption: Approx.250VA(*1): Tobespecified.Seethecorrespondingmodel

andsuffixcodesfordetails.

(2)RS-422/RS-232CconverterPower supply: 100to120VACor200to240VAC,

single phase, 50/60 HzVoltagefluctuation: Rating±10%,50/60±2Hz(Don’texceed

250VAC)Power consumption: Approx.15VA

(3) I/O unitAC Power supply: 100to240VAC±10%,singlephase,

50/60Hz±5%Power consumption:

NR893AG, NR893JG with enclosure (AO points 0 to 12):Approx.150VA (AO points 16 to 28):Approx.190VA (AO points 32 to 40):Approx.220VANR893JG without enclosure (AO points 0 to 40):Approx.100VA

DC Power supply: 24VDC±10%

Power consumption: Approx. 200 mA / AO 4 points*: Required when in the NR893AG, NR893JG I/O

unit,optioncode“/C”isnotspecified.

Oct. 01. 2019-00

Page 33: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

33

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

3.2 Others

(1)Clean,dryairforAnalyzerpurge(explosion-proofmodel)

Item DescriptionTemperature -10°C to +40°CPressure 0.3 to 0.9 MPa

Dew point-20°C or lower (under pressure)Compressed air should not dew at the installation ambient temperature.

Cleanliness

Must be free from dust, corrosive, and toxic elements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.

Volume; Approx. 75 l/min (stp.)

(2)Clean,dryairforin-situprobeItem Description

Temperature -10°C to +40°C

PressureTo be supplied at atomspheric pressure. A regulator and a needle valve to be prepared by the customer.

Dew point-20°C or lower (at pressure)Compressed air should not dew at the installation ambient temperature.

Cleanliness

Must be free from dust, corrosive, and toxic elements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.

Volume;Approx. 0.3 to 0.5 l/min (stp.).Depend on sample temperature and ambient temperature.

(3)Waterforflowthroughcellwithwatertube;Dependonmeasurementspecifications.Temperature,response speed, etc.,

(4)Utilityforsamplingunit(whenused)Dependonsamplingsystemconfiguration,sampletemperature, response speed, etc.

4.InstallationLocationRequirementsAvoid physical shock as it may cause damage to the equipment.

4.1 NR800analyzer,measurementcell,andfiber-opticcable

Item Description

Location

Hazardous area/non-hazardous area.Outdoor/indoor.Avoid direct exposure to wind, rain, sunlight and radiation heat.Altitude 2,000 m max.

Ambient temperature -10°C to +40°CAmbient humidity 5%to95%RH(nocondensation)

VibrationInstall the equipment in a place with minimum vibration (vibration acceleration of 2 m/s2 or less)

AtmosphereMust not contain corrosive or toxic substances.Must be free from dust.

4.2 RS-422/RS-232CconverterandI/OunitItem Description

LocationNon-hazardous area, indoor.Avoid direct exposure to wind and rain, sunlight, or radiation heat.Altitude 2,000 m max.

Ambient temperature +5° C to +35 °CAmbient humidity 5%to95%RH(nocondensation)

VibrationInstall the equipment in a place with minimum vibration (vibration acceleration of 2 m/s 2 or less)

AtmosphereMust not contain corrosive or toxic substances.Must be free from dust.

4.3MaintenanceSpaceforNR800AnalyzerSecure space as below for the operation and easy maintenance.

561

300 577 200305

1200

1077

(Maintenance space)

50

Unit: mm

*NR800

Analyzer

* JunctionboxforEthernetfiber-opticcables.(When NR895/JB is selected.) See section 2.6.3 for external dimensions.

4.4 StorageConditionsAvoid physical shock as it may cause damage to the equipment.

Item Description

LocationIndoor.Avoid direct exposure to wind and rain, sunlight, or radiation heat

Ambient temperature -10°C to +50°CAmbient humidity 5%to95%RH(nocondensation)

VibrationInstall the equipment in a place with minimum vibration (vibration acceleration of 2 m/s2 or less)

AtmosphereMust not contain corrosive or toxic substances.Must be free from dust.

Note: Fluoride, a material used in the NR823 and NR824 fiber-opticcables,hasthedeliquescenceproperties.When storing these cables, take appropriate measures to prevent moisture, e.g., a cable should be placed with desiccant in a sealed, moisture-barrier plastic bag and replace the desiccant regularly.

Oct. 01. 2019-00

Page 34: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

34

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

5.SupportforCalibrationModeling5.1 On-siteGuidanceonCalibrationModelingUsers prepare a sample with its laboratory analysis results.AYokogawarepresentativewillofferon-sitehands on instruction on Calibration Modeling to a personnel from the user’s facility while using the sample and the installed analyzer.

5.2 CalibrationModelingYokogawa make calibration modeling using the necessary quantity of user-provided samples with laboratoryanalysisresults.ApredefinedSEP(standarderrorofprediction)valueof1σwillbeusedasthemeasurement target value. The target value, sample quantity, and other details are determined separately for each application upon request.

5.3 OthersOther support options for calibration modeling and maintenance include:• Sampling test for potential users• Maintenance contracts• Sampling/model generation/maintenance consulting

service.Contact a Yokogawa sales representative for further information.Wewillprovidethebestsolutiontofityourneeds.

6.RecommendedSpecificationsforEngineeringPCandperipherals(providedbyuser)

6.1 PC

Computer IBM PC/AT compatibleOperating system (OS) Microsoft Windows 10 Professional 64 bit

CPU 1 GHz or higherRAM 2 GB or more

Hard disk space 10 GB (for program) and 25 GB (for data storage) or more

Ethernet adapter 10 BASE-TDisplay 1024 x 768 pixels or moreOther CD-R drive is recommended

Note: Tablet mode is not supported on Windows 10.

l ColorPrinterPrepare if necessary.

l ConnectperipheralsandconsumablesPrepare if necessary.

6.2 EthernetCommunication

l Whenfiber-opticcablesareusedforEthernetcommunication(seesection2.6)

Customer should provide:An optical-to-electrical signal converter (850 nm wavelength,foramultimodefiber-optic

cable that complies with Ethernet version 2.0 and IEEE 802.3 10Base-T and 10Base-FL standards andisfittedwithSTconnectors)

Fiber-optic extension cablesElectric cables for Ethernet communication

l WhenelectriccablesareusedforEthernetcommunication

Electric cables for Ethernet communication should be provided by customer.

l AdditionalEthernetfiber-opticcableMaterialandconnector: SilicaglassfiberwithST

connectorsMode: Multi-mode GINumber of cores: 2Core/Clad diameters: 50/125 µmApplicable wavelength: 850 nmLength: 1000 m or less in total including an

NR895fiber-opticcable

l SpecificationsofelectriccableforEthernetcommunication

Type: 10Base-T, 8-core shielded, straight connection, RJ-45 connectors

NotethatwhenanNR895fiber-opticcable for Ethernet communication is not used, use an Ethernet crossover cable. If you use a hub, consult Yokogawa.

Length: 3 to 40 mFinished outside diameter: 5.0 to 7.4 mm

Jan. 10. 2020-01

Page 35: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

35

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

SpecificationSheetsforinquiryofNR800series(1/2)Thank you for your inquiry. Please provide us with the following information on your request. We will quote in response to your inquiry. Fill in the blank and tick/check the appropriate box ().

1. ScopeofQuotationRequired: Analyzer Measurement cell/probe, Fiber-optic, Software, User’s manual (1 copy)Optional: Document(Specificationforapproval copy Testcertificate copy) Factoryacceptancetest CalibrationModelingbyYokogawa On-sitehands-ontrainingofCalibrationModeling

2. GeneralUser’s name: Purpose : Control Monitor/alarm OthersPlant name: Document: English Japanese

3. SpecificationsNumberofmeasuringchannel: 1ch 2ch 3ch 4ch PossibilityofthechannelexpansionAnalogueoutput: Need( points) NotneedWithstreamswitchinputfunction:Need NotneedMeasurementcell/probe: No.1ch(cellprobe) No.2ch(cellprobe) No.3ch(cellprobe) No.4ch(cellprobe)Length of fiber-optics: No.1 ch (Laying distance: m) No.2 ch (Laying distance: m) No.3 ch (Laying distance: m) No.4 ch (Laying distance: m)EthernetFiber-opticCable: Need(3m,5m,10m,20m) Withjunctionbox

4. InstallationConditionAnalyzer: Indoors Outdoors explosion-proofnon-ex-proof Toxic/Corrosiveambient Restrictedinstallationarea( m x m)Measurement cell / probe: Indoors Outdoors Toxic/Corrosiveambient

5. UtilitiesPower supply: VAC, HzInstrumentair: (Pressure MPa) None

Note: Clear and dry air is required for purging explosion-proof type analyzer. See section 3.2.

6. SampleconditionPlease fill in the next page.

7. SpecialnotePlease make other notes upon device installation if any.

Oct. 01. 2019-00

Page 36: General InfraSpecTM 6SHFL¿FDWLRQV Near-Infrared Analyzers · 2020. 9. 18. · FM *1, *2 TIIS *3 KOSHA *5 Housing structure IP53, NEMA 3R or CSA Type 3R Air (purge gas) connection

36

All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E

SpecificationSheetsforinquiryofNR800series(2/2)Samplecomposition/MeasuringConstituentList

Pleasefilloutallsampleconstituentincludingnon-measuringconstituent.No.1chSamplename:

Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Variation/ measuring range Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

No.2chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Variationmeasuringrange Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

No.3chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Variation/measuringrange Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

No.4chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Variation/measuringrange Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

Oct. 01. 2019-00Subject to change without notice.