GeneralSpecifications
InfraSpecTM
NR800 Fourier TransformNear-Infrared Analyzers
Yokogawa Electric Corporation2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
GS 12Y03A03-01E
GS 12Y03A03-01E©Copyright Aug. 2001
8th Edition Oct. 01, 2019
n OverviewThe NR800 Fourier Transform Near-Infrared Analyzers offers high-resolution, high S/N (signal-to-noise) ratio, and wide wavelength scanning range measurement with its newly developed interferometer and detector. The NR800 also offers high stability, vibration resistance, and durability, inheriting features from NR500 earlier successful models. The NR800 allows online, real-time, continuous, multiple, and simultaneous measurement for properties and component concentration of various processes.
n Features (including options) Newlydevelopedinterferometeranddetector
Usedforavarietyofapplications,rangingfromovertocombinationtone.• High resolution: Up to 4 cm–1, user selectable
setting• High S/N ratio: 2250:1 (RMS, 4 cm-1 resolution,
4100 to 4200 cm-1 , 1 sec.)• Wide wavelengths scanning range 900 to 2500
nm (11,000 to 4000 cm-1)• Wavelength reproducibility: 0.007 cm-1
• Wavelength accuracy: 0.04 cm-1
Enhancedenvironmentalresistance,durability,andreliability• Provides high vibration resistance by a unique
design free of sliding parts• Features a multi-channel measurement optical
system free of moving parts• Eliminates the need for a PC for continuous
operation. You need an engineering computer only for
Calibration Modeling and data loading.
OutstandingwavelengthaccuracyallowscalibrationmodeltransportbetweenNR800s
Outlierdetectionandself-diagnosticfeaturescomeasstandardfunction
Variousstandardfeaturesandavailableoptionsforoptimalsystemconfiguration• Dust-proof and drip-proof: IP53 equivalent or
NEMA TYPE 3R• Optional Explosion-proof Enclosure: FM, TIIS (*1),
KOSHA• Non-moving multi-channel measurement:
Expandable to 4 channels• Up to 12 items can be measured per stream:
limited to 64 items / unit• TheoptionalI/Ounitoffersavarietyof
inputs/outputs: Analogue output (up to 40 points), analogue
input, and contact input/output• Communication output: RS-422• Fast Ethernet communication between
Engineering PC and the analyzer(*1) TIIS: The Technology Institution of Industrial Safety
ExplosionproofModel
InfraSpec and SPECTLAND are trademarks or registered trademarks of Yokogawa Electric Corporation.All other company and product names mentioned in this document are trademarks or registered trademarks of their respective companies.
Please select appropriate equipment in accordance with the laws and regulations of the relevant country/region, when it is used in a location where explosive atmospheres may be present.
2
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E Oct. 01. 2019-00
1.Configuration1.1 Configurationexamples
• Modbusoutputwithsamplingunit(explosion-proof,1measuringchannel)
RS-422/RS-232Cconverter *2
RS-232C(MODBUS)
(15m max.)(1km max.)
Sampleejection
Process
Fiber-optic cable *2
Clean, dry air
Hazardous area Non-hazardous area
Alarm contact output
Optical/electricalsignal converter
Monitor/ControlSystem
Engineering PC
Samplingunit
Ethernet(Fiber-optic cable *1)
RS-422(MODBUS)
Ethernet(10 Base-T)
InfraSpecNR800
explosion-proofmodel
Measurement
cell
• Analogueoutputwithsamplingunit(explosion-proof,2measuringchannels)
I/O unit*2
RS-422Process 2
Process 1
Sampleejection
Sampleejection
Clean, dry air
Fiber-optic cable *2
Hazardous area Non-hazardous area
Optical/electricalsignal
Alarm contact output
Engineering PC
Ethernet(10 Base-T)
Analog signal
Contact signal
Contact signal
Monitor/ControlSystem
Sampling cell 1
Sampling cell 2
Ethernet(Fiber-optic cable *1)
InfraSpecNR800
explosion-proofmodel
Measurementcell 1
Measurementcell 2 (1km max.)
*1: Alwaysuseaspecialfiber-opticcable(seesection2.6)foranexplosion-proofmodel.Electriccablescannotbeused. Forageneralpurposemodel,usespecifictypeofcablesasbelow,accordingtothelengthrequiredfortheirinstallation. •Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Electriccablescannotbeused.*2: UseRS-422/RS-232Cconverter(seesection2.5)orI/Ounit(seesection2.2,selectSuffixcodeforexplosion-proofmodel) when you use RS-422 output of explosion-proof model analyzer. These units convert RS-422 output signals from analyzer
to dedicated ones. These units also block the communication signal upon receiving a failure signal from the analyzer Forsafetyandexplosion-proofintegrity,followtheinstructiontoselectspecificfiber-opticcables.Seesection2.4for
measurement and section 2.6 for Ethernet communication.
3
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
1.2 Component,softwareandcalibrationmodeling
Item Require-ment(*1) Model Description Reference
1 InfraSpec NR800 analyzer A
NR801AG General purpose model, CSA standard
1.32.1
NR805AG Explosion-proofmodel,FMcertifiedNR801JG General purpose model, Japanese version
NR805JG Explosion-proofmodel,TIIScertified(excluding NR805JG/K), KOSHAcertified(onlyforNR805JG/K)
2 I/O unit BNR893AG To be used with NR801,805AG, CSA standard 1.3
2.1NR893JG To be used with NR801,805JG
3 Measurement cell A
NR510 Flow through cell 2.3.1NR512 Flow through cell with constant temperature water tube 2.3.2NR511 Special measurement cell 2.3.3FIR200 In-situ probe 2.3.4
4 Fiber-optic cable for measurement A
NR821 Applicable wavelength range: 900 to 2100 nm, Silica, single2.4.1
NR822 Applicable wavelength range: 900 to 2100 nm, Silica, dualNR823 Applicable wavelength range: 900 to 2500 nm, Fluoride, single
2.4.2NR824 Applicable wavelength range: 900 to 2500 nm, Fluoride, dual
5 RS-422 / RS-232C converter B K9404LA or
K9404LD Converts the RS-422 signal from the analyzer into RS-232C. 2.5
6 Ethernet cable (*2) BNR895 Dedicatedfiber-opticcable(*3) 2.6
- Electric cable, provided by user 6.2
7 SoftwareA NR831 SPECTLAND2 management and maintenance software 2.7B NR530 Chemometrics software 2.8
8 Sampling unit B J439 Yokogawa will propose an optimum unit based on sample pressure, temperature, properties, and measurement items 2.9
9 On-site guidance on calibration modeling B J964 On-site hands-on practice and guidance 5.1
10 Calibration Modeling B J965 Calibration Modeling by Yokogawa based on user-provided sample with laboratory analysis results 5.2
11 Engineering PC A - Providedbyuser.Seerecommendedspecifications. 6.1
12Optical / electrical signal converter for Ethernet
B -Converts optical signals for an Ethernet output into electrical signals for engineering PC Interface. Provided by user. Seerecommendedspecifications
6.2
13 Customer inspectionB J963 For Instrument except following J443 -B J443 For J439 Sampling unit and FIR200 In-situ probe -
14 Equipment start-up B - Start-up work for analyzers and sampling units -
Notes *1: A: Required, B: optional *2: AnEthernetcableisrequired.Chooseeitheraspecialfiber-opticcableorelectriccabledependingonthefollowingconditions: Foranexplosion-proofmodel,alwaysuseaspecialfiber-opticcable(seesection2.6).Electriccablescannotbeused. Forageneralpurposemodel,usespecifictypeofcablesasbelow,accordingtothelengthrequiredfortheirinstallation.
•Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Anelectriccablecannotbeused. *3: Ifatotalcabledistanceislongerthan20m,anadditionalfiber-opticcable(seesection2.6)shallbeprovidedbytheuser.
1.3Safetystandard,EMCstandardModel SafetyStandard EMCstandard
FT-NIRanalyzer I/O unitNR801AG, NR805AG*2
NR893AG*2
CSA C22.2 No.61010-1 (exclude NR805AG) *1FM Class 3810ANSI/ISA 61010-1ANSI/ISA-82.02.02 (IEC 61010-2-031)
-
NR801JG, NR805JG(excluding NR805JG/K)
NR893JG - -
NR805JG/K - - Korea Electromagnetic Conformity Standard
*1: • Installation category (Overvoltage category) II This number represents the transient overvoltage condition. This also complies with the regulation on impulse withstand voltage. “II”appliestoelectricalequipmentwhosepowerissuppliedfromfixedinstallationssuchasdistributionboards. • Pollution Degree 2 Describes how much the dielectric strength or surface resistivity is being deteriorated by adherence of solid, liquid or gas substances. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.*2: Install 5 A breaker in the power supply circuit
Oct. 01. 2019-00
4
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.Componentspecifications(including options)
2.1 NR800FourierTransformNear-InfraredAnalyzers
2.1.1 HardwarespecificationsPrinciple: Fourier-transform remote measurement
viafiber-opticcableMeasurement method: Optical transmission absorptionMeasured sample: LiquidClassificationoflaserproduct: Class 1 Laser Product (IEC 60825-1:2007/2014)Beam source: Halogen lamp (recommended
replacement interval for continuous operation: 5000 hours)
Detector: InGaAs (indium gallium arsenide) photodiode,effectivewavelengthrange:900 to 2500 nm
Number of measuring channels: 1 to 4 (non-moving)Housing structures: Field suitable with full
hinged front coverFiber-optic cable entry: Cable glandFiber-optic cable connectors:
For measurement; FCFor Ethernet; ST
Display: LEDKeyboard: Covered with water-proof sheetingOperating location requirements: See chapter 4.Grounding type: Independent, 100ΩorlowerInsulationresistance: 10MΩorgreater(500VDC)Withstanding voltage:
PowerSupply WithstandingVoltage100VAC,115VAC 1000VACfor1min.200VAC,230VAC 1500VACfor1min.
Coating: Epoxy resin coatingPaint colour:
Model NR801AG,NR801JG
NR805AG,NR805JG
Flame proof enclosure - Lamp black
(equivalent to Munsell 0.8Y 2.5/0.4)Air regulator cover -
Case Frosty white ( equivalent to Munsell 2.5Y8.4/1.2)
Display panel Lamp black (equivalent to Munsell 0.8Y 2.5/0.4)
Wall mount bracket, Free standing rack
Deep sea moss green (equivalent to Munsell 0.6GY3.1/2.0)
Oct. 01. 2019-00
Model NR801AG NR801JG NR805AG NR805JG(excluding NR805JG/K) NR805JG/K
Explosion proof General purpose type(non explosion-proof) FM *1, *2 TIIS *3 KOSHA *5
Housing structure IP53, NEMA 3R or CSA Type 3RAir (purge gas) connection - 1/4NPT (F) Rc1/4 or 1/4NPT (F)Cable connection Packing gland 3/4NPT(F) Flame-proofpackingglandorsealingfitting
Weight (kg) *4
Without bracket Approx. 50 Approx. 65With wall mount bracket Approx. 55 Approx. 70Free stand rack Approx. 85 Approx. 100
*1: Type X purging and explosion proof for Class I, Division 1, Groups B, C and D. Temperature Class T5. All wiring shall comply with the manufacture’s instructions, National Electrical Code ANSI/NFPA 70 and Local Electrical Codes.*2: All wiring shall comply with the manufacture’s instructions and Local Electrical Codes.*3: Ex pd II B+H2 T5 X All wiring shall comply with the manufacture’s instructions and Local Electrical Codes.*4: Not including weight of the packaging materials.*5: KOSHA, Ex dp IIB+H2 T5 All wiring shall comply with the manufacture’s instructions and Local Electrical Codes.
2.1.2 PerformanceWavelength scanning range: 900 to 2500 nm
(11000 to 4000 cm –1)Wavelength resolution: 4, 8, 16, 32, and 64 cm -1
(user selectable)Wavelength reproducibility: 0.007 cm -1Wavelength accuracy: 0.04 cm -1S/N ratio 2250:1 (RMS, resolution 4 cm -1, 11000 to
4000 cm -1, 1 sec.)
2.1.3 Inputs/Outputs(seealsosection2.2I/OUnit)
(1)CommunicationInterface:Engineering PC: 1 channel (Ethernet)DCS / I/O unit: 2 channels (RS-422) 1 channel for DCS (Modbus) and
another for I/O unit; or 2 channels for I/O unit
(2)Contactoutputs:Specification: NC/NOselectable,2channelsRating: 0.5A,30VDCorlessAction:
General purpose model:NR800analyzerpower”ON/OFF”;Powerfailsignal
StatusDO1TerminalNumber
1 to 2 2 to 3Power“off” Short OpenPower “on” Open Short
NR800Systemalarm
Status DO2TerminalNumber1 to 2 2 to 3
Power“off” Short OpenAlarm occurred Short OpenAlarm not occurred Open Short
5
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
Explosion proof model:NR800Systemalarm
Status TerminalNumberD1 to D2 D2 to D3
Power“off” Open ShortAlarm occurred Open ShortAlarm not occurred Short Open
Annunciator signal output
Status TerminalNumberA1 to A2 A2 to A3
Power“off” Open ShortDuring purging Open ShortOverride switch pressed * Open ShortAfter purging Short Open*: Overriding
NR800 has Overriding function. Normally power supplyisautomaticallyshutoffwhenyouopenthedoor of the pressurized explosion-proof enclosure( on the analyzer). However, Overriding function allows you to open the door of explosion-proof sectionwithoutautomaticpoweroff
A switch of the overriding function is located insidetheanalyzer(flameproof).Youcanusetheoverriding to open the door of the explosion-proof section.
When opening the door, always make sure the ambient atmosphere is not hazardous. Use gas sensors to detect and eliminate all hazards.
2.1.4Operatingmodes*1Basicoperatingmodeand
descriptionChanneloperatingmodeNo. Auto/Manual.*3
Maintenance *2
Spectrum analysis of a reference sample, equipment maintenance
- -
RunOn-line measurement (allows spectrum analysis on selected channels)
1 AUTOMANUAL
2 AUTOMANUAL
3 AUTOMANUAL
4 AUTOMANUAL
Notes *1: When the power is turned on, the analyzer starts accordingtoapredefinedoperatingmode.
*2: Can perform spectrum analysis (not continuous measurement).
*3: Auto: Performs continuous measurement Manual: Can perform spectrum analysis (not
continuous measurement).2.1.5 ChangingandSettingOperatingModeSome operations are prohibited depending on the user level.
User levelDescription Changing/
settingmodeAbbreviationUser A UA For operator Unauthorized
User B UB For equipment supervisor Authorized
User C UC For maintenance Authorized
• A user level can be switched on the front console panel of the analyzer or from SPECTLAND2 screen of an engineering PC.
• A password is required to switch levels from UA to UB and switch from UA or UB to UC.
2.1.6 SampleMeasurement(1)Continuousmeasurement
Number of measuring channels: 1to4(Specificationisrequired.)Data updating period: 4 seconds or longer (depending on the
number of averaging and measuring channels, as well as measurement items)
Number of measurement items: Up to 12 per channel (48 max. per analyzer)Number of outlier detection items: Up to 12 per channel (48 max. per analyzer)
Separate maintenance is available for each channel (except for common hardware).
(2)Streamswitchingbyinputsignal(intermittentmeasurement)
Calibration model set for the each channel can be changed by external input signal. This function is used for multi- stream application or multi sample application.
Stream switchingNumber of streams to switch: Up to 16Switching patterns: See table 1. “Stream numbers
assignable to measuring channels corresponding to switching patterns”
Contact inputs: See section “2.2 I/O Unit”RS-422 communication: See section “2.1.8 MODBUS”
Data updating period: 8 seconds or longer (depending on the
switching pattern, numbers of averaging and measuring channels, and stream configuration)
Number of measurement items: Up to 12 per stream (64 max. per analyzer)Number of outlier detection items: Up to 12 per stream (64 max. per analyzer)Separate maintenance is available for each channel
(except for common hardware).Input signals and measurement computation sequenceStreamidentification: Used to identify the selected sample.
The analyzer will choose a calibration model set to suite for the relevant stream based on this signal.
Measurement enabled (valid sample): When closed, the analyzer performs
measurement using the calibration modelspecifiedbythesampleidentificationsignalabove.
Schematictimingchartofmeasurementandcomputationsequence(e.g.2streams)
Stream 1 Stream 2 Stream 1
Computationperformed
Computationperformed
Computationperformed
OpenClosed
Closed
Open
Stream identification*
Measurement enabled
Notes(*) Astreamnumberisdefinedbyacombinationofopen/closedstatesofanidentificationcontactsignal.
Oct. 01. 2019-00
6
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
Table1 Streamnumbersassignabletomeasuringchannelscorrespondingtoswitchingpatterns
Streamswitch MeasuringchannelNo.*3Totalstreams Applicable
channelNo.Case PatternNo. 1 2 3 4None 0 1 2 3 4 4 max 1 to 4
Stream switching per channel *2
1 *1 1 to (17-N) 18-N 19-N 20-N 16 max 1 to 42 1 to 8 9 to 16 None None 16 max 1 to 23 1 to 4 5 to 8 9 to 12 13 to 16 16 max 1 to 4
Notes*1: N: Maximum number of measuring channels included within the analyzer.*2: The stream number for a measuring channel that does not switch paths must be the smallest number in the relevant
column.*3: Measuring channel numbers that equal the number of measuring channels included within the analyzer or smaller are valid.
LED display: The display depends on the operating
status.Run mode
Basic operating mode LEDs: Only RUN lights up.Channel operating mode LEDs: Lit in accordance with each setting.Self-diagnosis LEDs: Lit in accordance with the results of
self-diagnosis.LED display: The display depends on the operating
status.
2.1.8 MODBUS
(1)Communicationspecifications
Item SpecificationCommunication standard*
RS-422 or RS-232C (a converter is required)
Start bit 1Number of data bits 7 (ASCII mode) or 8 (binary mode)Parity bit 1Stop bit 1Communication speed 4800, 9600 or 19200 (selectable)
Error detection Odd number parity, Even number Parity or none (selectable)
*: Only RS-232C can be used in NR805*G explosion-proof type.
See section 1.2 and 2.5
(2)Serialcommunication
Item ASCIImode RTUmodeNumber of data bits 7 (ASCII) 8 (binary)
Message starting character Colon “:” None
Message ending character
Carriage return/Line feed “<CR><LF>” None
Error detection LRC (Logical Redundancy Check)
CRC-16 (Cyclic Redundancy Check)
Inter-character time out None 100msec
Oct. 01. 2019-00
2.1.7 ConsolePanel
(1)Displayitems
Operating mode LEDs (MODE) The following LEDs indicate the current
basic operating mode:RUN: Lit when in the Run mode.MAINT: Lit when in the Maintenance mode. The operating modes of each channel
are displayed by LEDs when in the Run mode.
Two LEDs are provided for each channel, amounting to a total of 8 LEDs.
AUTO: Lit when in the Auto mode.MAN: Lit when in the Manual mode.
Self-diagnosis LEDs (DIAGNOSIS)ALARM: Lit when an alarm occurs.LAMP: Lit when a lamp has burned out or after
atimeperioddefinedbytheservicelifesetting elapses.
LASER: Lit when the laser has burned out orafteratimeperioddefinedbytheservice life setting elapses.
LED display (16 digits, STREAM/CONSTITUENT/VALUE)
The content depends on the operating status or operation.
Operation keys The following six keys are provided:
ITEM: Used to change items.ENT: Usedtoconfirmtheentry.Arrow keys: Used to move the cursor or change
display.
(2)Analyzerbehaviorforeachbasicoperatingmode
Maintenance modeBasic operating mode LEDs: Only MAINT lights up.Channel operating mode LEDs: Lit in accordance with each setting.Self-diagnosis LEDs: Lit in accordance with the results of
self-diagnosis.
7
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(3)SupportfunctionintheMODBUSprotocol
Function No. Description
01 Reads the ON/OFF status of coils.02 Reads the ON/OFF status of input relays.03 Reads the current value of holding registers.04 Reads the current value of input registers.05 Forcibly changes the status of a coil.06 Write a value to a holding register.
08Loop-back testSends back the same message as the command message.
(4)Recommendedvalueofmasterside
Item RecommendedvalueTransmission speed (bps) 4800/9600/19200Time monitor of Inter-character gap 1000msNo reply time 2 sec.Number of transmission retries 5 timesCommunication Recovery Period 30 sec.
Oct. 01. 2019-00
2.1.9 OtherFunctions
(1)Baselinecompensation:Upto10points
(2)Outputoftheresultcompensatedusinganalogue inputs
The following operation result can be output.Compensation operation processing:
P2 = C1 * P1 + C0 + D D = An2 * D2 + An * D1 + D0
An: Analogue input (n: 4 or less)P1: Measurement value before the
compensationP2: Measurement value after the
compensationD0,D1,D2,C0,C1: Coefficient
(3)OutputsmoothingSimple moving average (Up to 10 points)Exponential moving average
(4)On-linemeasurementspectrasaving
2.1.10ModelandSuffixCodes
(1)NR801AG,Generalpurposemodel
[Style:S2]Model SuffixCode OptionCode Description
NR801AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, General Purpose modelLanguage -E ••••••••••••••••• EnglishPower supply 1
346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelength scanning range W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDual cable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1
2
•••••••••••••••••
•••••••••••••••••
Electric cable, only for general purpose model and less than 40 mFiber-optic cable
Mounting ABC
•••••••••••••••••••••••••••••••••••••••••••••••••••
Without bracketsWith wall-mounting bracketsWith free standing rack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS With stream switch input function
8
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(2)NR805AG,TypeXpurgedandFMapprovedexplosion-proofmodel
[Style:S2]Model SuffixCode OptionCode Description
NR805AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion proof model, FM approved version
Language -E ••••••••••••••••• EnglishPower supply 1
346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N"Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelength scanning range W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDual cable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A
BC
•••••••••••••••••••••••••••••••••••••••••••••••••••
Without bracketsWith wall-mounting bracketsWith free standing rack
- -N ••••••••••••••••• Always “-N"- -N ••••••••••••••••• Always “-N"Cable entrance -2 ••••••••••••••••• Female 3/4NPTPurge air connection 2 ••••••••••••••••• Female 1/4NPTOption /SS With stream switch input function
Oct. 01. 2019-00
9
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(3)NR801JG,Generalpurposemodel
[Style:S2]Model SuffixCode OptionCode Description
NR801JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, General Purpose modelLanguage -E
-J••••••••••••••••••••••••••••••••••
EnglishJapanese
Power supply 1346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelength scanning range W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDual cable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1
2
•••••••••••••••••
•••••••••••••••••
Electric cable, only for general purpose model and less than 40 mFiber-optic cable
Mounting ABC
•••••••••••••••••••••••••••••••••••••••••••••••••••
Without bracketsWith wall-mounting bracketsWith free standing rack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS With stream switch input function
Oct. 01. 2019-00
10
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(4)NR805JG,TIISapprovedexplosion-proofmodel*1
[Style:S2]Model SuffixCode OptionCode Description
NR805JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion-proof model, TIIS approved *1
Language -E-J
••••••••••••••••••••••••••••••••••
EnglishJapanese *1
Power supply 1346
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels
-S1-M1-M2-M3-M4
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels
Wavelength scanning range W1W2
••••••••••••••••••••••••••••••••••
900 to 2100 nm900 to 2500 nm
Fiber-optic cable -1-2
••••••••••••••••••••••••••••••••••
Single cableDual cable
- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A
BC
•••••••••••••••••••••••••••••••••••••••••••••••••••
Without bracketsWith wall-mounting bracketsWith free standing rack
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Cable entrance -1
-2-3-4
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Metal conduit (G3/4)Metal conduit (3/4NPT)Flameproof packing (G3/4)Flameproof packing (3/4NPT)
Purge air connection 12
••••••••••••••••••••••••••••••••••
Female Rc1/4Female 1/4 NPT
Option /SS/K
With stream switch input functionKOSHA approved explosion proof model *1
*1: “TIIS” approval is not applicable when option code “/K” is selected. “Japanese” can not be selected with option code “/K.”
Oct. 01. 2019-00
11
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.1.11Externaldimensions
(1)Generalpurposemodel(NR801*Gtype)•Withoutbrackets(Suffixcode:A)
6282 60 60 60 60 62 62
68±5
30±5
480
92
60430
277
665
355550
Mounting screws4-M8
2-cable outlets (For NR801AG)(Packing glands, acceptable cables: ø10 to 13)
Earth terminal(M4 screw)
Power supply cable connection inlet (For NR801AG)(Packing gland, acceptable cable: ø10 to 15.9)
3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)
Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)
4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)
Unit: mm
For NR801AG
Part A
Part A
Note: The wall shall be designed to withstand four times the weight of NR800.
Oct. 01. 2019-00
12
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
• Wall-mounting model(Suffixcode:B)30
±566
5550
60 430
50 50
603086
0
355 20
920
6282 60 60 60 62 6260
277
68±5
Mounting holes4-ø10.5 holes
Earth terminal(M4 screw)
Power supply cable connection inlet (For NR801AG)(Packing gland, acceptable cable: ø10 to 15.9)
2-cable outlets (For NR801AG)(Packing glands, acceptable cables: ø10 to 13)
3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)
Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)
4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)
Wall mounting bracket
Unit: mm
For NR801AG
Part A
Part A
Note: The wall shall be designed to withstand four times the weight of the NR800.
Oct. 01. 2019-00
13
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
• Free standing rack model(Suffixcode:C)66
5
1500
70
350 8080
30±5
68±5
550 355 50
630
550
490
56827
7
6282 60 60 60 62 6260
Mounting holes 4-ø17 (for anchor bolts)
4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)
Earth terminal(M4 screw)
Unit: mm
Power supply cable connection inlet (For NR801AG)(Packing gland, acceptable cable: ø10 to 15.9)
2-cable outlets (For NR801AG)(Packing glands, acceptable cables: ø10 to 13)
3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)
Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)
For NR801AG
Part A
Part A
Oct. 01. 2019-00
14
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(2)Explosion-proofmodel(NR805*Gtype)• Without brackets(Suffixcode:A)
62
82180
93
36
6262332
665
355550
App
rox.
300
200
430 60
9248
0
4-M8Mounting screws
95
Internal-pressure protection unit(flameproof enclosure)Cable connectors(with sealing fitting orflameproof packing gland *1)
Air (purge gas) connection: Rc1/4 or 1/4NPT(only 1/4NPT for NR805AG)
Connector *3
Connection:3/4NPT
For NR805AG
Earth terminal(M4 screw)
86
277
216
65
4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)
2-cable outlets *2 *3(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)
Special Ethernet fiber-optic cable outlet
Unit: mm
For NR805AGPart B
Part A
Part A
Part B
*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: InthecaseofNR805AG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*3: The connector of NR805AG is 3/4NPT.Note: The wall shall be designed to withstand four times the weight of the NR800.
Oct. 01. 2019-00
15
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
• With wall-mounting brackets(Suffixcode:B)
62
App
rox.
300
36
82180
6093
6262332
665
550
60 430
50 50
3086
0
200
For NR805AG
Connection: 3/4NPT
Mounting holes 4-ø10.5 holes
Internal-pressure protection unit(flameproof enclosure)
Cable connectors(with sealing fitting or flameproof packing gland *1)
Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)
4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)
95
Earth terminal(M4 screw)
86
277
21665
Special Ethernet fiber-optic cable outlet
2-cable outlets *2 *3(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)
920
355 20
Wall mounting bracket *4
Connector *3
Unit: mm
For NR805AG
Part A
Part A
Part B
Part B
*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: InthecaseofNR805AG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*3: The connector of NR805AG is 3/4NPT.*4: The wall shall be designed to withstand four times the weight of the NR800.
Oct. 01. 2019-00
16
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
• With free standing rack(Suffixcode:C)
568
490
550
630
70
1500
665
App
rox.
300
80
200
550
Mounting holes 4-ø17 (for anchor bolts)
36
93
Connection: 3/4NPT
For NR805AG
For NR805AGInternal-pressure protection unit(flameproof enclosure)
Cable connectors(with sealing fitting or flameproof packing gland *1)
Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)
82180
350 80
95
Connector *3
50355
Bottom of analyzer
277
216
8665
Earth terminal (M4 screw)
62332 62 62
4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)
2-cable outlets *2 *3(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)
Special Ethernet fiber-optic cable outlet
Unit: mm
Part B
Part A
Part A
Part B
*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: InthecaseofNR805AG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*3: The connector of NR805AG is 3/4NPT.
Oct. 01. 2019-00
17
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.2 I/OUnit2.2.1 OverviewThe I/O unit is an input/output interface between the analyzer and external monitor/control systems such as DCS. The I/O unit connects with NR800 analyzer via serial communications (RS-422). Up to two I/O units can be connected to the analyzer, and one of which can incorporate the contact input for stream switching and analogue input.
2.2.2 SpecificationsPowersupply: Seesection“3.UtilitySpecifications”.Insulationresistance: 5MΩormore,500VDCWithstandvoltage: 1500VACfor1min.Analogue output:
Output data: Measurement results (Properties and concentration)
Number of outputs: 0 to 40Outputspecifications:
Item Description
Output range 4 to 20 mA DC (3.0 to 21.0 mA DC, Floating-common type)
Isolation methodBetween output terminals and internal circuit: Photo coupler isolationBetween output terminals: non-isolated, Common negative
Withstand voltage 500VDCfor1min.Allowable load resistance 600Ωorless12-bit D/A converter resolution 5.7 µA
System accuracy ±0.5%offullscale
Separately,24VDCpowersupplyisrequired (see section “2.2.3. Model and Suffixcodes”and“3.UtilitySpecifications”).
Connect a short-cable (*) to unused analog output terminal between [OUT+] and [OUT-]. If not, the ALM indicator blinks of an analog output module.
(*)Cablespeccifications: Size: More than 0.2 mm2 (AWG24) Voltagerating: Morethan300V Temperature rating: More than 60°C Flamerating: EquivalenttoVW-1 Screw size of terminal on the side of
the analog output module: M3.5Contact output:
Alarm Quantityofoutputs Alarm Quantity
ofoutputsGeneral 1 Outlier 4Communication failure 1 I/O unit failure 1
Operating mode 5 - -
Oct. 01. 2019-00
Contactoutputspecifications:Item Description
I/O unit failure
Rating 24VDC.0.3A
Action and number1 normally open and 1 normally closed (shared common)
Alarm, operating mode, and outlier detection
Insulation method Mechanical isolationRated load voltage
DC 24VAC 100to240V
Maximum load current 2 Amps/point, 8 Amps/common
Servicing life
Mechanical At least 20 million actions
Electrical At least 100 thousand actions
Surge killer NoneNumber per common 8 points/commonExternal power supply Not required.
Contact input:Description:Streamidentificationformulti-streamsamplingunit: 8 points, status signals to identify
sample streams that pass through measurement cells.
Measurement enabled (stream valid): 4points,statussignalstoconfirmthat
samples inside the measurement cells are ready for measurement.
Inputunitspecifications:Item Description
Input type VoltagefreecontactCommon terminal Common to 8 pointsIsolation method Transformer isolation
Withstand voltageBetween external connectors collectively and internal circuit:500VDCfor1min.
Off-stateopen-circuit voltage 5to7VOn-state load current 1 to 3 mAOn-state load resistance 200ΩorsmallerOff-stateloadresistance 100Ωorlarger
Provide external contact-input signals meeting the above requirements.
Analogue input:Input data and number: Analogue output compensation signal, 4 pointsInputspecifications:
Item DescriptionInput range (actual) 1to5VDC(-0.25to5.25VDC)
Isolation methodBetween input terminals and internal circuit: photo coupler isolationBetween input terminals: non-isolated, negative common
Withstand voltage 500VDCfor1min.Input resistance 1MΩ12-bit A/D converter resolution 1to5VDC:1.4mV
System accuracy ±0.5%offullscaleat0to40°C
Installation requirements: See chapter 4.Mounting: Wall mountingExternal connection terminal: M3.5 screwDrip-proof and dust-proof construction: IP4X (with enclosure)
18
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.2.3 ModelandSuffixcodes
(1)NR893AG:forNR801AGandNR805AGanalyzer
[Style:S2]Model SuffixCode OptionCode Description
NR893AG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IO unit (non explosion-proof)Language -E ••••••••••••••••• EnglishAnalyzer 1
2••••••••••••••••••••••••••••••••••
NR805AG (Explosion-proof model) (*1)NR801AG (General Purpose model)
Contact input for stream switching
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Analogue input for compensation
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Number of analogue outputs -00-04-08-12-16-20-24-28-32-36-40
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
None481216202428323640
Enclosure -A-B-C
•••••••••••••••••••••••••••••••••••••••••••••••••••
Number of analog outputs: 0 to 12Number of analog outputs: 0 to 28Number of analog outputs: 0 to 40
- -N ••••••••••••••••• Always “-N”
*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.
(2)NR893JG:forNR801JGandNR805JGanalyzer
[Style:S2]Model SuffixCode OptionCode Description
NR893JG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IO unit (non explosion-proof)Language -E
-J••••••••••••••••••••••••••••••••••
EnglishJapanese
Analyzer 12
••••••••••••••••••••••••••••••••••
NR805JG (Explosion-proof model) (*1)NR801JG (General Purpose model)
Contact input for stream switching
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Analogue input for compensation
AB
••••••••••••••••••••••••••••••••••
AvailableNone
Number of analogue outputs (*2)
-00-04-08-12-16-20-24-28-32-36-40
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
None481216202428323640
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Option /C Withenclosureincluded24VDCpowersupplyunit
for analog output. (*2)
*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.
*2: Separate24VDCpowersupplyrequiredwhenusingananalogueoutputandoptioncode“/C”arenotspecified.(seechapter 3).
Oct. 01. 2019-00
19
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.2.4 Externaldimensionsandweight:
(1)NR893AGPaint color: Light beige (Munsell 5Y 7/ 1 or equivalent)
B D F
CA
E G
H
75 17
16118°
1743
5028
0
400
68.5
200
6100 505050505050
L + 50
L50 50
4-ø14 holesfor mounting *1
ø20 holefor protective ground wiring
Number ‘N’ of wiring holesSuitable cable OD *2 A: ø10 to 15.9 (to the power supply) B to H: ø10 to 18.5
Unit: mm
Enclosure code
Numberofanalog outputs L(mm) N Approximate
Weight-A 0 to 12 400 6 (A to F) 15.5 kg-B 0 to 28 500 7 (A to G) 17.5 kg-C 0 to 40 600 8 (A to H) 20.0 kg
*1: The wall shall be designed to withstand four times the equipment's weight.*2: Sinceattachedferritecoresareused,thepracticalcableODareuptoø13mm.
Oct. 01. 2019-00
20
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(2)I/OunitWithoutenclosure(onlymodelNR893JG)• For Explosion proof analyzer unit•ThesizeofIOunitforgeneralpurposeanalyzerunitisthesameasabovefigurebutRS-422/232Cconverterpower
supply case removed.Paint color: Baseplate, RS-422/232C converter power supply case;
Module colors: Light cobalt blue (Munsell 6.2PB4.6/ 8.8 or equivalent), lamp black (Munsell 0.8Y2.5/0.4 or equivalent)
215
199
L
303 151510
(101)Approx. 120
N-ø7.2 holesfor mounting *1
Safety barrier power supply caseused only with NR805JG explosion-proof analyzer
Unit: mm
Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 333 4 6.5 kg-16, -20, -24, -28 16, 20, 24, 28 450 6 7.0 kg
-32, -36, -40 32, 36, 40 537 6 7.5 kg
*1: The wall construction should be withstood a force of four times the weight.
(3)Withenclosure(Optioncode:/C)Paint color: Light beige (Munsell 5Y 7/ 1 or equivalent)
C E
D
A
F H
G
5050
B
5050 50505075 17
10118°
1743
5028
0
400
App
rox.
30
200
6
L + 50
L50 50
4-ø14 holesfor mounting *1
ø20 holefor protective ground wiring
Number ‘N’ of wiring holesSuitable cable OD: ø10 to 18.5
Unit: mm
I/O Unit (inside)
Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 400 6 (A to F) 15.5 kg-16, -20, -24, -28 16, 20, 24, 28 500 7 (A to G) 17.5 kg
-32, -36, -40 32, 36, 40 600 8 (A to H) 20.0 kg
*1: The wall shall be designed to withstand four times the weight of the NR800
Oct. 01. 2019-00
21
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.3MeasurementCells
2.3.1 FlowthroughCell
(1)SpecificationsOptical path length: 1, 2, 5, 10, or 20 mmFiber-optic connector: FC typeWetted part material: Fluoro rubber, Kalrez 4079, 316SS, PTFE, Borosilicate crown glass, or SapphireSample pressure range: Atmospheric pressure up to 1.9 MPa GSample temperature range: –5°C to +80°C (It varies by the materials of O-ring. Contact Yokogawa for further
information.)Sample connection: Female Rc1/4Installationangle: VerticalInstallation location requirements: See section 4.Weight: Approx. 3 kg
(2)ModelandSuffixCodes
[Style:S1]Model SuffixCode OptionCode Description
NR510 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cellWindow material and optical path length
-B00
-B01-B02-B05-B10-B20-S00
-S01-S02-S05-S10-S20
•••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Borosilicate crown glass, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm
Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ring material -B
-K••••••••••••••••••••••••••••••••••
Fluoro rubberKalrez 4079
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”
(3)ExternalDimensions
Unit: mm(96)55 29
102
92
L4-ø5.2 holes
2-Rc1/4
Optical path length (mm)
1251020
L (mm)
64±2
65±2
68±2
73±2
83±2
Fiber-optic connection(in the other side)
Oct. 01. 2019-00
22
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.3.2 FlowthroughCellwithConstantTemperatureWaterTube
(1)SpecificationsOptical path length: 1, 2, 5, 10, or 20 mmFiber-optic connector: FC typeWetted part material: Fluoro rubber, Kalrez 4079, 316SS, PTFE, Borosilicate crown glass, or SapphireSample pressure range: Atmospheric pressure up to 1.9 MPaSample temperature range: +5°C to +80°C (It varies by the materials of O-ring. Contact Yokogawa for further
information.)Constant water temperature range: +5°C to +80°C (It varies by the materials of O-ring. Contact Yokogawa for
further information.)Sample connection: Female Rc1/4Connection for water with constant temperature: 6-mm outside diameter, copper tubeInstallationangle: VerticalInstallation location requirements: See section 4.Weight: Approx. 3 kg
(2)ModelandSuffixCodes
[Style:S1]Model SuffixCode OptionCode Description
NR512 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cell with constant temperature water tube
Window material and optical path length
-B00
-B01-B02-B05-B10-B20-S00
-S01-S02-S05-S10-S20
•••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Borosilicate crown glass, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire, 10 mm, with variable optical path adapter (1, 2, 5, or 20 mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm
Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ring material -B
-K••••••••••••••••••••••••••••••••••
Fluoro rubberKalrez 4079
- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”
(3)ExternalDimensions
Unit: mm
4-ø5.2 holes
2-Rc1/4
L29
Approx. 160
App
rox.
160
102
92
B-SIDEB-SIDE A-SIDE A-SIDE
Thermal insulation jacket
Approx. 145
70111
Optical path length (mm)
1251020
L (mm)
64±2
65±2
68±2
73±2
83±2Fiber-optic connection (in the other side)
Oct. 01. 2019-00
23
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.3.3 SpecialmeasurementcellThisclausedescribesPFAflow-throughtypespecialcell.
(1)SpecificationsOptical path length: 1, 2, 5, or 10 mmFiber-optic connector: FC typeWetted part material: PFA,Kalrez4079,Moriseiperflioro(MP4275B),orSapphire (Stainless steel is partly used as non-wetted parts.)Sample pressure range: Atmospheric pressure up to 0.3 MPa GSample temperature range: 5°C to 80°CSample connection: 1/4 inchs (Ø6.35 x Ø3.95 mm) PFA tube, or Ø6 x Ø4 mm PFA tube (length: 3000 mm)Installationangle: VerticalInstallation location requirements: See section 4.Weight: Approx. 1.5 kg
(2)ModelandSuffixCodes
[Style:S1]Model SuffixCode OptionCode Description
NR511 ••••••••••••••••••••••• ••••••••••••••••• Special measurement cellWindow material and optical path length
-S01-S02-S05-S10
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mm
Body material P ••••••••••••••••• PFAO-ring material -K
-P••••••••••••••••••••••••••••••••••
Kalrez 4079Moriseiperflioro(MP4275B)
Tube -A-B
••••••••••••••••••••••••••••••••••
1/4 inchs (Ø6.35 x Ø3.95 mm), 3000 mm, PFAØ6 x Ø4 mm, 3000 mm, PFA
Option /Z Always add “/Z”
(3)ExternalDimensions
Unit: mm(80)(50)
90
(300
0)(3
000)
74
77(94)
4-ø5.2 holes
Fiber-optic connection(in the other side)
Oct. 01. 2019-00
24
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.3.4 In-situprobe
(1)SpecificationsFiber-optic connector: FC typeWet part material:
Body; Hastelloy C-276 or UNS S31603 (Stainless steel 316L)Others; Sapphire, Hastelloy C-276, Gold
Sample pressure range: Atmospheric pressure to 0.98 MPa *Sample temperature range: Normal temperature to 80°C *Purge gas connection: Female 1/8 NPT x 2Fiber-optic entrance: Female 1 NPTInstallationangle: Vertical(Insertportionisdownward)orHorizontalInstallation location requirements: See section 4Weight: Approx.5kg(excludingflange)*:*: ThespecificationsaboveapplytothecasewhentheflangeisJIS10K. If you use sample pressure or sample temperature
outsidetherangeshownabove,pleasecontactoursalesrepresentativesregardingthespecificconditionsofyourdemand.Wewillprovidesomeadequateproposalswhichfityourparticularneeds.(Installationexample10MPa,300°C)
(2)ModelandSuffixCodesFIR200/Z
(3)ExternalDimensions
L *3
( 560 ) *4 ( 80 )
Approx. 35 A *1
Flange *2
Fiber-optic connection box
Fiber-optic entrance(Exclusive packing)
Purge-gas connections2-1/8NPT
Approx. ø27
Unit: mm
*1: Optical path length “A”: 2/5/10 mm, selectable (Dependent on sample conditions, measurement items, measurement range, etc. Consult Yokogawa.)
*2:Mountingflange(HastelloyC-276/UNSS31603comparable)shouldbeprovidedbycustomer.*3: Specify length “L” within the range 150 to 250 mm.*4: The shape and size of connection box may change.
Note1:Keepthetemperatureattheconnectionofthefiber-opticcableat80°Corbelow.Ifthetemperatureattheconnectionexceeds 80°C, the connection box should be purged with clean, dry air, or other appropriate measures should be taken. Inthiscase,theconnectionboxmaybedifferentinshapeandsizefromtheoneshownabove.Ifafiber-opticcableformeasurement is exposed to radiant heat, it should be shielded.
Note 2: Dimensions of each part may change without prior notice.
2.4 Fiber-opticCablesformeasurementFiber-optic cables connect an analyzer to measurement cell. Select appropriate cables to meet your application requirement, e.g., wavelengths scanning range.
2.4.1 SilicaFiber-opticCable
(1)SpecificationsApplicable wavelength range: 900 to 2100 nmConnector: Double-end FC or FC-SMA typeStructure: Dual (for sample and reference) or single (sample), two-core, protected by resin-
coatedstainlessflexibletubeMinimum bending radius: 100 mm. To reduce optical attenuation, make the radiuses along the cable as
large as possible when laying cables.Installation location requirements: See section 4.Cabling: Conduit protected cabling is recommended.
Oct. 01. 2019-00
25
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(2)ModelandSuffixCodes
a.Singlefiber-opticcable[Style:S1]
Model SuffixCode OptionCode DescriptionNR821 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2100nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-end FC typeFC type on analyzer side and SMA type on measurement cell side
Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
5 m10 m20 m30 m50 m100 m150 m200 m300 m
- -000 ••••••••••••••••• Always “-000”
b.Dualfiber-opticcable[Style:S1]
Model SuffixCode OptionCode DescriptionNR822 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2100nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-end FC typeFC type on analyzer side and SMA type on measurement cell side
Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
5 m10 m20 m30 m50 m100 m150 m200 m300 m
- -000 ••••••••••••••••• Always “-000”
(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.
• Double-endFCconnector
BA
A B
Unit: mmFC connector FC connector
(Cable length)600±50
(57) 3010
1510
• FC(analyzer)-SMA(measurementcell)connector
BA
A B
Unit: mmFC connector (analyzer) SMA connector
(measurement cell)
(Cable length)600±50
(57) 3010
600±50
(53) 8
1510
Oct. 01. 2019-00
26
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.4.2 FluorideFiber-opticCable(1)Specifications
Applicable wavelength range: 900 to 2500 nmConnector: Double-end FC type or FC-SMA typeStructure: Dual (for sample and reference) or single (sample), two-core, protected by resin-
coatedstainlessflexibletubeMinimum bending radius: Analyzer side; 100 mm, Measurement side;120 mm To reduce optical attenuation, make the radiuses along the cable as large as
possible when laying cables.Installation location requirements: See section 4.Note: Fluoride,amaterialusedintheNR823andNR824fiber-opticcables,hasthedeliquescenceproperties.Whenstoring
these cables, take appropriate measures to prevent moisture, e.g., a cable should be placed with desiccant in a sealed moisture-barrier plastic bag. Replace the desiccant regularly.
(2)ModelandSuffixCodes(ModelcodetoincludefibersforAnalyzer–Cell(probe)–Analyzer)a.Singlefiber-opticcable
[Style:S1]Model SuffixCode OptionCode Description
NR823 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2500nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-end FC typeFC type on analyzer side and SMA type on measurement cell side
Cable length -L003-L004-L005
•••••••••••••••••••••••••••••••••••••••••••••••••••
3 m4 m5 m
- -000 ••••••••••••••••• Always “-000”
b.Dualfiber-opticcable[Style:S1]
Model SuffixCode OptionCode DescriptionNR824 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2500nmConnector -FF
-FS••••••••••••••••••••••••••••••••••
Double-end FC typeFC type on analyzer side and SMA type on measurement cell side
Cable length -L003-L004-L005
•••••••••••••••••••••••••••••••••••••••••••••••••••
3 m4 m5 m
- -000 ••••••••••••••••• Always “-000”
(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.(Connectthelongerendtothe measuremrnt cell side.)• Double-endFCconnector
BA
A B
Unit: mmFC connector (analyzer) FC connector
(measurement cell)
(Cable length)600±50 800±50
(57) 3010
1510
• FC(analyzer)-SMA(measurementcell)connector
BA
A B
Unit: mmFC connector (analyzer) SMA connector
(measurement cell)
(Cable length)600±50
(57) 3010
800±50
(53) 8
1510
Oct. 01. 2019-00
27
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.5 RS-422-to-RS-232CConverter(PartNumber:K9404LAorK9404LD)This unit receives data from the analyzer via RS-422 and converts them to RS-232C serial interface, therefore it helps the data transmission between the analyzer and I/O unit or monitor/control system. This unit also blocks communicationsignalswhenitreceivesapower-offsignalfromtheanalyzer,thusensuringtheexplosion-proofintegrity.
2.5.1 SpecificationsPower supply: See section 3.Signal terminals: Analyzer main unit side (RS-422): M4, output side (RS-232C): Female Dsub 25-pinGroundingtype: 100ΩorlowerInstallation location requirements: See section 4.Housing structure: Desktop (K9404LA), Screw cramp (K9404LD)Weight: Approx. 2 kgNote: The converter does not carry the CE Marking. An appropriate housing should be provided by customer, if necessary.
2.5.2 TerminalDiagram
CN3 CN2
LN-+RDBRDASDBSDA
RS-422
From analyzer
D/I 12V AC power supply
Female D-SUB 25-pin connectorCN1
RS-232C
2.5.3 RS-232CCommunicationStraightCable
GND
123456789
CDRDSDERSGDRRSCSCI
FGCDRDSDERSGDRRSCSCI
GND183220764522
Internal connection
To converter To host
2.5.4 ExternalDimensions
4-ø4.5 holes203
126
126
39.5
15 15
9611
.5 218
203
37
K9404LA Desktop type K9404LD Screw cramp type
Unit: mm
Oct. 01. 2019-00
28
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.6 EthernetFiber-opticCableCable for Ethernet communication between the analyzer and the engineering PC.
2.6.1 SpecificationsLength: Up to 20 mConnector: ST typeStructure: Two-core,protectedbystainlessflexibletubeInstallation location requirements: See section 4.Minimum bending radius: 50 mm. To reduce optical attenuation, make the radiuses along the cable as
large as possible when laying cables.
2.6.2 ModelandSuffixCodes[Style:S1]
Model SuffixCode OptionCode DescriptionNR895 ••••••••••••••••• ••••••••••••••••• Ethernetfiber-opticcableLength *1 -L003
-L005-L010-L020
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
3 m5 m10 m20 m
- -000 ••••••••••••••••• Always “-000”Option /JB With junction box *2
*1: Foratotalcabledistanceof20morlonger,anadditionalfiber-opticcable(fittedwithSTconnectors)mustbeprovidedbythe user.
Seechapter6.2forthecablespecificationofuserprovisions.*2: TheJunctionboxisforconnectionbetweenanNR895fiber-opticcableandtheadditionalfiberonesprovidedbytheuser.
2.6.3 ExternalDimensions(1)Fiber-opticCable
15
9.310
Blue
Yellow
Unit: mm
1000 0+100
43 30
(Cable length)
Oct. 01. 2019-00
29
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
(2)JunctionBoxforEthernetfiber-opticcables(Optioncode:/JB)
400 430
305
300
142
24120
280
60
20260
4-ø10
Handle to open
100 30Hole
2-Cable gland
Unit : mm
NR895_JB.ai
Housing structure: IP44Coating: Epoxy resin coatingPaint color: Light beige (equivalent to Munsell 5Y7/1)Material: IronWeight: Approx. 10 kg
Oct. 01. 2019-00
30
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.7 SPECTLAND2MeasurementandMaintenanceSoftware
2.7.1 OverviewSPECTLAND2 is an application software that controls NR800 operation and monitoring on a PC. It displays the analyzer status, instructs measurement and sets parameters for the analyzer. To use SPECTLAND2, firstinstallitintheengineeringPC.The continuous operation of SPECTLAND2 for extended periods of time is not guaranteed.
MainFeatures• Instructs the analyzer to measure spectra, stores
data and displays. The measured spectra can be used for calibration modeling.
• Shows trend graphs of the measurements and saves themtofilesduringacontinuousmeasurement.
• Sets parameters for continuous measurement.• Displays various data of the analyzer, such as
operating modes, alarm status, and maintenance data.
• Instructs the analyzer to perform operation mode change, spectra measurement, property value setting, calibration models, and setting measurement conditions.
2.7.2 MainWindows(1)ManualSpectrumWindow
Enables the analyzer to measure spectra, which are to be processed by Chemometrics (Calibration Modeling software). Also allows data to be saved to filesanddisplayed.
(2)AutoSpectrumWindowAllows the user to upload spectra data to the Engineering PC during continuous measurement upon receiving a signal at periodic intervals, outlier detection, or a property value variation failure. This data is displayed for each measuring channel.
(3)PowerSpectrumWindowIt displays power spectra. This window is available for UB level (equipment supervisor) users.
(4)InterferogramWindowIt displays interferogram data collected. This window is available for UB level users.
(5)Real-timeTrendWindowsDisplay measurement values of Nos. 1 to 6 and Nos. 7 to 12 components in two separate trend graph windows for each stream. Up to 10 windows can be open at the same time.
(6)HistoricalTrendWindowsDisplay historical trend data saved. Trend data of 24hoursforeachstreamissavedtoafile.Upto4windows can be open at the same time.
(7)ParameterWindowDisplays the current parameter settings for the analyzer. In addition, UB level users can change the settings.
(8)AlarmStatus/HistoryWindowsThe Alarm Status window displays the active alarms for the analyzer, while the Alarm History window displays all the past alarms. The alarm history can be deleted with commands.
(9)Tab-controlledMaintenanceWindowDisplays the A/D reference value and servo-related data of the analyzer. This window is available for UC level users.
(10)Tab-controlledCommunicationStatusWindowDisplays the communication status between the engineering computer and the analyzer. This window is available for UC level (maintenance) users.
2.7.3 ModelandSuffixCodes
Model SuffixCode
Option Code Description
NR831 ••••••••••• ••••••••••• SPECTLAND2 measurement and maintenance software
Language -E-J
••••••••••••••••••••••
EnglishJapanese
- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”Option /UP Versionup
Package contents: One CD-ROM One user’s manual
2.8 ChemometricsSoftware
2.8.1 Specifications• Calibration Modeling technique: Partial least square (PLS) and others
2.8.2 ModelandSuffixCodes
Model SuffixCode
Option Code Description
NR530 ••••••••••• ••••••••••• Chemometrics softwareLanguage -E ••••••••••• English- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”
Package contents: Two CD-ROMs (Install disc/User's manual)
One set of user registration document
Oct. 01. 2019-00
31
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.9 SamplingUnitSome applications permit direct link between the process and the measurement cell or you don’t need the sampling unit.Pleasefillouttheinquireform(lasttwopages)toshowandcontactourrepresentatives.Wewillselecttheoptimum system to suit the application of your demand.
lExampleofSamplingSystemThefollowingintroducesaconceptualsamplingunitformeasuringtheproperties(RON,RVP,Distillationetc.)ofgasolineblendinginapetroleumrefinery.Notethateachsamplingunitshouldbedesignedforeachapplication;designs will vary.
MeasurementSampleConditionsItem Condition
Fluid to be measured Gasoline
Sample inlet/outlet condition
Inlet pressure 0.3 to 0.9 MPa GOutlet pressure 0.3 to 0.9 MPa GInlet temperature 0°C to +40°C
Notes1: When samples do not contain free water content, a coalescer (CF1) is not required.2: When samples do not have bubbles, a degasser (DG1) is not required.3: Ifthepressuredifferencebetweenthesampleinletandoutletisgreaterthan0.3MPa,asamplingpump(PU1)isnot
required.
com comcom com
PU1CHV1, 2, 3BV1, 2, 3, 4BV5, 6, 7, 8, 9V1, 2, 3, 4, 5, 6NV1, 2, 3, 4, 5FM1, 2, 3, 4PR1, 2CF1DG1P1, 2DT1TK1, 2, 3, 4CON1RN1
PUMPCHECK VALVE3-WAY BALL VALVE2-WAY BALL VALVESTOP VALVENEEDLE VALVEFLOW METERPRESSURE REGULATORCOALESCER FILTERDEGASSERPRESSURE GAUGEDETECTORTANKInfraSpec NR800RESISTANCE TEMP. SENSOR
: JIS 10K-15A-RF: Rc1/2: Rc1/4
<connection>
: SUS316TP 15A sch40: ø12/ø10 SUS316: ø8/ø6 SUS316: ø4/ø3 Nylon tube: ø6/ø4 Nylon tube: ø22/ø6 Braid tube
<piping>
outlet
Cooling waterinlet
*2PR2
Air inlet
*3
*4
*1PR1
PU1Sample inlet
P1
V1
NV1
FM1NV3NV4
BV1
DG1 BV2
CF1
BV9
Sample outlet
P2
V2
FM4FM3
Cooling water
*3 *4
*1
TK1
V3
V4
CH
V2
DT1 BV3
CH
V1
*2NV5 RN1
CON1
TK2 V5V6
a'
NV2FM2
BV4
Drain
TK4
BV
7
TK3
CH
V3
BV
8a'
BV
5
BV
6
SAMPLING UNIT
N2 gas inlet
N2 gas
Oct. 01. 2019-00
32
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
2.10 AuxiliaryParts*Therecommendedreplacementintervalsforauxiliaryparts,listedinSection2.10.1,ModelandSuffixCodes,areset for preventive maintenance and do not guarantee the service life of parts. Actual replacement intervals for auxiliary parts vary depending on operating conditions, sample conditions, and other factors. Be sure to perform preventive maintenance based on operational status of instrument.Theauxiliarypartsareclassifiedintotwotypes:thosethatcanbereplacedbycustomerandthosethatmustbereplaced by a Yokogawa service engineer.Prior to replacement, refer to the user’s manual and follow the instruction therein.
* Auxiliary parts mean limited life components that are excepted to be in their wear-out failure period within 5 years under normal operating or storage conditions.
Model SuffixCode
Option Code Description
RecommendedReplacement
IntervalNR8SP01 •••••••• ••••••••• Auxiliary parts set for NR800
-01-02-03-04-05-06
-07-08-10-11-12-13-14-15-16-17
••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Sealing packing for analyzer door (for NR801AG/NR805AG/NR801JG/NR805JG)Analyzer cooling fan (5 pcs)Spectrometer support insulator (4 pcs)Lamp (adjusted)Laser headDesiccant for spectrometer (9 pcs)
Desiccantforflow-throughcell(10pcs)*1Sealing packing for analyzer door (for NR805JG/K)O-ringforflow-throughcell(JISP38,FKM(fluororubber),2pcs)O-ringforflow-throughcell(JISP21,FKM(fluororubber),2pcs)O-ringforflow-throughcell(JISP16,FKM(fluororubber),2pcs)O-ringforflow-throughcell(JISP38,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP21,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP16,Kalrez4079,2pcs)Windowpaneofflow-throughcell(borosilicatecrownglass,2pcs)*3Windowpaneofflow-throughcell(sapphire,2pcs)*3
1 year3 years3 years5,000 hours12,000 hours1 year (Shelf life: 1 year after shipment)3 months1 year*1, *2*1, *2*1, *2*1, *2*1, *2*1, *2
*1: When a measurement cell is disassembled for service/cleaning, it is recommended that O-rings should be replaced.*2: Replacement intervals vary depending on sample composition, humidity, pressure, etc. Determine replacement intervals
based on actual operational status.*3: If a window is scratched or fogged, replace the whole set of windows (2 pieces/set).
Please contact Yokogawa, when you need the O-ring for Special measurement cell.
3.UtilitySpecifications3.1 PowerSupplies
(1)NR800analyzerPower supply: 100,115,200,230VAC,singlephase,
50/60 Hz( *1)Voltagefluctuation: Rating±10%,50/60±2HzPower consumption: Approx.250VA(*1): Tobespecified.Seethecorrespondingmodel
andsuffixcodesfordetails.
(2)RS-422/RS-232CconverterPower supply: 100to120VACor200to240VAC,
single phase, 50/60 HzVoltagefluctuation: Rating±10%,50/60±2Hz(Don’texceed
250VAC)Power consumption: Approx.15VA
(3) I/O unitAC Power supply: 100to240VAC±10%,singlephase,
50/60Hz±5%Power consumption:
NR893AG, NR893JG with enclosure (AO points 0 to 12):Approx.150VA (AO points 16 to 28):Approx.190VA (AO points 32 to 40):Approx.220VANR893JG without enclosure (AO points 0 to 40):Approx.100VA
DC Power supply: 24VDC±10%
Power consumption: Approx. 200 mA / AO 4 points*: Required when in the NR893AG, NR893JG I/O
unit,optioncode“/C”isnotspecified.
Oct. 01. 2019-00
33
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
3.2 Others
(1)Clean,dryairforAnalyzerpurge(explosion-proofmodel)
Item DescriptionTemperature -10°C to +40°CPressure 0.3 to 0.9 MPa
Dew point-20°C or lower (under pressure)Compressed air should not dew at the installation ambient temperature.
Cleanliness
Must be free from dust, corrosive, and toxic elements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.
Volume; Approx. 75 l/min (stp.)
(2)Clean,dryairforin-situprobeItem Description
Temperature -10°C to +40°C
PressureTo be supplied at atomspheric pressure. A regulator and a needle valve to be prepared by the customer.
Dew point-20°C or lower (at pressure)Compressed air should not dew at the installation ambient temperature.
Cleanliness
Must be free from dust, corrosive, and toxic elements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.
Volume;Approx. 0.3 to 0.5 l/min (stp.).Depend on sample temperature and ambient temperature.
(3)Waterforflowthroughcellwithwatertube;Dependonmeasurementspecifications.Temperature,response speed, etc.,
(4)Utilityforsamplingunit(whenused)Dependonsamplingsystemconfiguration,sampletemperature, response speed, etc.
4.InstallationLocationRequirementsAvoid physical shock as it may cause damage to the equipment.
4.1 NR800analyzer,measurementcell,andfiber-opticcable
Item Description
Location
Hazardous area/non-hazardous area.Outdoor/indoor.Avoid direct exposure to wind, rain, sunlight and radiation heat.Altitude 2,000 m max.
Ambient temperature -10°C to +40°CAmbient humidity 5%to95%RH(nocondensation)
VibrationInstall the equipment in a place with minimum vibration (vibration acceleration of 2 m/s2 or less)
AtmosphereMust not contain corrosive or toxic substances.Must be free from dust.
4.2 RS-422/RS-232CconverterandI/OunitItem Description
LocationNon-hazardous area, indoor.Avoid direct exposure to wind and rain, sunlight, or radiation heat.Altitude 2,000 m max.
Ambient temperature +5° C to +35 °CAmbient humidity 5%to95%RH(nocondensation)
VibrationInstall the equipment in a place with minimum vibration (vibration acceleration of 2 m/s 2 or less)
AtmosphereMust not contain corrosive or toxic substances.Must be free from dust.
4.3MaintenanceSpaceforNR800AnalyzerSecure space as below for the operation and easy maintenance.
561
300 577 200305
1200
1077
(Maintenance space)
50
Unit: mm
*NR800
Analyzer
* JunctionboxforEthernetfiber-opticcables.(When NR895/JB is selected.) See section 2.6.3 for external dimensions.
4.4 StorageConditionsAvoid physical shock as it may cause damage to the equipment.
Item Description
LocationIndoor.Avoid direct exposure to wind and rain, sunlight, or radiation heat
Ambient temperature -10°C to +50°CAmbient humidity 5%to95%RH(nocondensation)
VibrationInstall the equipment in a place with minimum vibration (vibration acceleration of 2 m/s2 or less)
AtmosphereMust not contain corrosive or toxic substances.Must be free from dust.
Note: Fluoride, a material used in the NR823 and NR824 fiber-opticcables,hasthedeliquescenceproperties.When storing these cables, take appropriate measures to prevent moisture, e.g., a cable should be placed with desiccant in a sealed, moisture-barrier plastic bag and replace the desiccant regularly.
Oct. 01. 2019-00
34
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
5.SupportforCalibrationModeling5.1 On-siteGuidanceonCalibrationModelingUsers prepare a sample with its laboratory analysis results.AYokogawarepresentativewillofferon-sitehands on instruction on Calibration Modeling to a personnel from the user’s facility while using the sample and the installed analyzer.
5.2 CalibrationModelingYokogawa make calibration modeling using the necessary quantity of user-provided samples with laboratoryanalysisresults.ApredefinedSEP(standarderrorofprediction)valueof1σwillbeusedasthemeasurement target value. The target value, sample quantity, and other details are determined separately for each application upon request.
5.3 OthersOther support options for calibration modeling and maintenance include:• Sampling test for potential users• Maintenance contracts• Sampling/model generation/maintenance consulting
service.Contact a Yokogawa sales representative for further information.Wewillprovidethebestsolutiontofityourneeds.
6.RecommendedSpecificationsforEngineeringPCandperipherals(providedbyuser)
6.1 PC
Computer IBM PC/AT compatibleOperating system (OS) Microsoft Windows 10 Professional 64 bit
CPU 1 GHz or higherRAM 2 GB or more
Hard disk space 10 GB (for program) and 25 GB (for data storage) or more
Ethernet adapter 10 BASE-TDisplay 1024 x 768 pixels or moreOther CD-R drive is recommended
Note: Tablet mode is not supported on Windows 10.
l ColorPrinterPrepare if necessary.
l ConnectperipheralsandconsumablesPrepare if necessary.
6.2 EthernetCommunication
l Whenfiber-opticcablesareusedforEthernetcommunication(seesection2.6)
Customer should provide:An optical-to-electrical signal converter (850 nm wavelength,foramultimodefiber-optic
cable that complies with Ethernet version 2.0 and IEEE 802.3 10Base-T and 10Base-FL standards andisfittedwithSTconnectors)
Fiber-optic extension cablesElectric cables for Ethernet communication
l WhenelectriccablesareusedforEthernetcommunication
Electric cables for Ethernet communication should be provided by customer.
l AdditionalEthernetfiber-opticcableMaterialandconnector: SilicaglassfiberwithST
connectorsMode: Multi-mode GINumber of cores: 2Core/Clad diameters: 50/125 µmApplicable wavelength: 850 nmLength: 1000 m or less in total including an
NR895fiber-opticcable
l SpecificationsofelectriccableforEthernetcommunication
Type: 10Base-T, 8-core shielded, straight connection, RJ-45 connectors
NotethatwhenanNR895fiber-opticcable for Ethernet communication is not used, use an Ethernet crossover cable. If you use a hub, consult Yokogawa.
Length: 3 to 40 mFinished outside diameter: 5.0 to 7.4 mm
Jan. 10. 2020-01
35
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
SpecificationSheetsforinquiryofNR800series(1/2)Thank you for your inquiry. Please provide us with the following information on your request. We will quote in response to your inquiry. Fill in the blank and tick/check the appropriate box ().
1. ScopeofQuotationRequired: Analyzer Measurement cell/probe, Fiber-optic, Software, User’s manual (1 copy)Optional: Document(Specificationforapproval copy Testcertificate copy) Factoryacceptancetest CalibrationModelingbyYokogawa On-sitehands-ontrainingofCalibrationModeling
2. GeneralUser’s name: Purpose : Control Monitor/alarm OthersPlant name: Document: English Japanese
3. SpecificationsNumberofmeasuringchannel: 1ch 2ch 3ch 4ch PossibilityofthechannelexpansionAnalogueoutput: Need( points) NotneedWithstreamswitchinputfunction:Need NotneedMeasurementcell/probe: No.1ch(cellprobe) No.2ch(cellprobe) No.3ch(cellprobe) No.4ch(cellprobe)Length of fiber-optics: No.1 ch (Laying distance: m) No.2 ch (Laying distance: m) No.3 ch (Laying distance: m) No.4 ch (Laying distance: m)EthernetFiber-opticCable: Need(3m,5m,10m,20m) Withjunctionbox
4. InstallationConditionAnalyzer: Indoors Outdoors explosion-proofnon-ex-proof Toxic/Corrosiveambient Restrictedinstallationarea( m x m)Measurement cell / probe: Indoors Outdoors Toxic/Corrosiveambient
5. UtilitiesPower supply: VAC, HzInstrumentair: (Pressure MPa) None
Note: Clear and dry air is required for purging explosion-proof type analyzer. See section 3.2.
6. SampleconditionPlease fill in the next page.
7. SpecialnotePlease make other notes upon device installation if any.
Oct. 01. 2019-00
36
All Rights Reserved. Copyright © 2001, Yokogawa Electric Corporation GS 12Y03A03-01E
SpecificationSheetsforinquiryofNR800series(2/2)Samplecomposition/MeasuringConstituentList
Pleasefilloutallsampleconstituentincludingnon-measuringconstituent.No.1chSamplename:
Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Variation/ measuring range Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to
No.2chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Variationmeasuringrange Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to
No.3chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Variation/measuringrange Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to
No.4chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa, normal MPaBubble/slurry: Bubble Slurry Others( )
Tick for measuring constituent
ConstituentFill out all constituent Variation/measuringrange Normal Unit
1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to
Oct. 01. 2019-00Subject to change without notice.