Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
ID-COV2-IFU-USA V3
Genedrive® 96 SARS-CoV-2 KitID-CoV2-01-LPU (Low Profile Version) ID-CoV2-01-FPU (Fast Plate Version)ID-CoV2-01-BPU (Bio Plate Version)
Store at temperature 2˚C to 30˚C
For the most up to date version of this document, please visit www.genedrive.com
Validation of this test has not been reviewed by FDA. Review under the EUA program is pending.For In Vitro Diagnostic (IVD) UseRx Only
IMPORTANT: Please read these instructions prior to using this kit.
Table of contents 2 Intended use
2 Summary and explanation
3 Principles of the procedure
3 Compatible RNA extraction methodologies
5 Kit components
5 Materials required but not supplied
6 Specimen collection, handling and storage
6 Storage and shelf life
6 Operating conditions
6 Warnings and precautions
7 Limitations of use
8 Test procedure
16 Result interpretation
19 Genedrive® 96 Exporter – result interpretation
22 Quality controls
23 Troubleshooting
24 Performance characteristics
24 Analytical sensitivity
25 Analytical specificity
26 Precision
26 Inclusivity
27 Diagnostic accuracy
29 Quick reference guide – manual set up
32 Manufacturer details
Terms & Abbreviations cDNA Complementary DNA
COVID-19 Coronavirus disease 2019
DNA Deoxyribonucleic acid
dNTPs Deoxyribonucleotide triphosphates
LoD Limit of Detection
PCR Polymerase Chain Reaction
Real-Time PCR Real-Time Polymerase Chain Reaction
RNA Ribonucleic acid
RNase P Ribonuclease P
RT-PCR Reverse Transcription Polymerase Chain Reaction
SARS-CoV-2 Severe acute respiratory syndrome coronavirus 2
WHO World Health Organization
Summary and explanationUntil recently there were six known members of the
Orthocoronavirinae sub-family of viruses (more commonly
referred to as Coronaviruses) that were capable of
infecting humans. Of these, four are known to cause
common cold symptoms (HKU1, NL63, 229E and OC43).
Whilst two are zoonotic and are associated with more
severe respiratory symptoms (SARS-CoV and MERS-CoV)1.
SARS-CoV-2 emerged from China in late 2019. WHO
declared a global pandemic on 11th March 2020.
Since emergence, it has been demonstrated that the
virus can transfer from person to person and through
environmental contamination. Due to the high infectivity
of the virus it is essential that positive individuals are
identified and quarantined to limit onwards transmission
of the disease.
Intended useGenedrive® 96 SARS-CoV-2 Kit is an end point melt
analysis RT-PCR test intended for the qualitative
detection of RNA from the SARS-CoV-2 in upper
respiratory specimens from individuals suspected of
COVID-19 by their healthcare provider. Testing is limited
to laboratories - certified under the Clinical Laboratory
Improvement Amendments of 1988 (CLIA), 42 U.S.C.
§263a, to perform high complexity tests, or by similarly
qualified non-U.S. laboratories.
Results are for the identification of SARS-CoV-2 RNA.
The SARS-CoV-2 RNA is generally detectable in upper
respiratory specimens during the acute phase of
infection. Positive results are indicative of the presence
of SARS-CoV-2 RNA; clinical correlation with patient
history and other diagnostic information is necessary to
determine patient infection status. Positive results do
not rule out bacterial infection or co-infection with other
viruses. The agent detected may not be the definite
cause of disease. Laboratories within the United States
and its territories are required to report all positive
results to the appropriate public health authorities.
SARS-CoV-2 is a novel Coronavirus currently thought
to have been of zoonotic origin that originally infected
humans from close contact with animals, at open markets
in China. Once infected, the virus causes disease
symptoms known as COVID-19. The symptoms may
include a dry cough, fever, shortness of breath, fatigue,
respiratory failure and death.
The Genedrive 96 SARS-CoV-2 Kit is an in vitro
diagnostic (IVD) molecular assay for the detection of
species-specific genetic sequences within the E and N
genes of the SARS-CoV-2 genome2. The kit is indicated for
use by a laboratory professional for the identification of
SARS-CoV-2 in RNA specimens that have been derived
from upper respiratory specimens.
Negative results do not preclude SARS-CoV-2 infection
and should not be used as the sole basis for patient
management decisions. Negative results must be
combined with clinical observations, patient history, and
epidemiological information.
The Genedrive 96 SARS-CoV-2 Kit is intended for use
by qualified and trained clinical laboratory personnel
specifically instructed and trained in the techniques
of Real-Time PCR and in vitro diagnostic procedures.
Validation of this test has not been reviewed by FDA.
Review under the EUA program is pending.
The kit is intended to be run on the following real-time
thermocycler platform:
• Roche LightCycler® 480 II
• Applied Biosystems® 7500 Fast Real-Time PCR instrument
• Bio-Rad CFX96 Real-Time PCR Detection System
References
1. Chu, D., Pan, Y., Cheng, S., Hui, K., Krishnan, P., Liu, Y., Ng, D., Wan, C., Yang, P., Wang, Q., Peiris, M. and Poon, L., 2020. Molecular Diagnosis of a Novel Coronavirus (2019-nCoV) Causing an Outbreak of Pneumonia. Clinical Chemistry, 66(4),
pp.549-555
2. WHO.int. 2020. Laboratory Testing For 2019 Novel Coronavirus (2019-Ncov) In Suspected Human Cases. [online] Available at: https://www.who.int/publications-detail/laboratory-testing-for-2019-novel-coronavirus-in-suspected-human-cases-20200117
ID-COV2-IFU-USA V3 2
Compatible RNA extraction methodologiesThe specimen input into the Genedrive 96 SARS-CoV-2 Kit
reaction wells is RNA that has been extracted from upper
respiratory specimens using an extraction instrument and kit.
The following extraction methodologies have been validated
Principles of the procedureGenedrive 96 SARS-CoV-2 Kit is a simplified one-step
test designed for the diagnosis of SARS-CoV-2 from RNA
extracts derived from upper respiratory swabs. All the
reagents required for the test are supplied as a lyophilized
bead, to which the specimen is added prior to running
the test.
The genetic targets are species-specific sequences which
are located in the E and N genes of the virus. The target
is amplified via a one-step RT-PCR reaction, whereby
RNA conversion to cDNA and subsequent asymmetric
amplification of cDNA occurs within the same well without
the need of user intervention. The detection is conducted
via light emission of fluorophores during end-point melt
curve analysis.
The fluorescent probe for each target is designed to melt
from the template within discrete temperature ranges. The
instrument measures the resulting change in fluorescence
and displays the changes as peaks on a temperature
profile, or melt curve, for analysis.
The Genedrive 96 SARS-CoV-2 Kit targets and amplifies
endogenous human RNase P, intended to ensure that
human sample is present in the amplification reaction.
However, levels of endogenous human RNA present
may be donor or extraction kit dependent. Users have
the option to “spike” the patient specimen with the
recommended human RNA control material to normalize
RNase P levels prior to the RNA extraction process.
The chemistry also contains primers and probes for a
synthetic RNA amplification control (IPC). The presence
of the associated peak indicates that amplification has
occurred as expected, therefore the test has been set up
correctly and there are no inhibitory substances within the
added specimen. This provides increased confidence that
a negative result is due to absence of CoV-2 target RNA,
rather than failure of amplification.
Each individual Genedrive 96 SARS-CoV-2 Kit contains
all the components necessary to conduct 96 reactions.
The reagents are lyophilized in reaction wells ready for
PCR and end-point melt analysis (dNTPs, buffers, reverse
transcriptase, DNA polymerase, specific primers/probes).
The lyophilized reagents are rehydrated within the wells,
using the specimen or control material. The plate is then
briefly centrifuged and placed into the appropriate
Real-Time PCR instrument.
The test consists of a target DNA amplification step
followed by an end-point melt curve analysis. The data
requires interpretation after the run is completed.
Please see analysis instructions and Table 7 for a result
interpretation matrix. The time from specimen input to
result is 100-119 minutes dependent on the platform
the test is run on.
for use, with verification of analytical LoD across all Real-
Time PCR instruments listed (See Table 1) and validated with
contrived specimens (Qiagen Viral mini, Table 1) and clinical
RNA specimens (Qiagen Viral Mini, Table 1).
Manufacturer RNA Extraction Kit Product code
Real-Time PCR instrument used for Diagnostic Accuracy
Real-Time PCR instrument used for Analytical Sensivity
(LoD)
Qiagen
Virus/Pathogen Kit with QIAamp® MinElute
Virus Spin Kit57704 Roche LightCycler 480 II
QIAamp Viral RNA Mini Kit
52904/52906 Roche LightCycler 480 II Roche LightCycler 480 II
Beckman CoulterRNAdvance Viral RNA
purification KitC63510
ABI 7500 Fast
Bio-Rad CFX96
Roche LightCycler 480 II
CytivaSera-Xtracta Virus/
Pathogen Kit29514201/ 29506009
Bio-Rad CFX96
ID-COV2-IFU-USA V3 3
Table 1
Product name Manufacturer Product code
LightCycler 480 II Roche Diagnostics05 015 278
001
ABI 7500 FastThermo Fisher
Scientific4351106/7
CFX96 Bio-Rad 1845097-IVD
For all RNA extraction methodologies, ensure optional or
additional steps suggested in the extraction kit instructions
for use are taken to reduce ethanol or wash buffer
carryover as much as possible to reduce the chance of
reaction inhibition, which could impact test sensitivity.
Compatible Real-Time PCR instrumentation
The Genedrive 96 SARS-CoV-2 Kit is available as low
profile, bio or fast 96-well plates and is validated for use
with the following Real-Time PCR instrument.
All instruments used must be installed, calibrated,
checked and maintained according to the
manufacturer’s instructions for best results.
ID-COV2-IFU-USA V3 4
Materials required but not supplied:Reagents
• RNA extraction kit• Negative control• Positive control• Molecular grade Nuclease Free Water (e.g. Sigma W4502)• Human Lymphocyte RNA (AMSBIO, ATR1254148-10)
(Extraction control) This accessory can be obtained from Genedrive
Kit components
Catalogue Number Description
ID-CoV2-01-LPU Genedrive 96 SARS-CoV-2 Kit – Low Profile Plate (96 reactions per kit)
ID-CoV2-01-FPU Genedrive 96 SARS-CoV-2 Kit - Fast Plate (96 reactions per kit)
ID-CoV2-01-BPU Genedrive 96 SARS-CoV-2 Kit - Bio Plate (96 reactions per kit)
Equipment
• RNA extraction
instrument (optional)
• Real-Time PCR instrument
• Powder free
disposable gloves
• Calibrated pipettes (p20,
p200)
• Plate centrifuge
• Class 2 safety cabinet
• Plate vortex
Part
Description
96 well low profile, Bio or fast bar-coded plate containing lyophilized reagent mix in each well. Each plate is supplied foil-sealed. Lyophilized contents (each bead):• DNA polymerase enzyme• Reverse transcriptase enzyme• Oligonucleotide primer pairs;
FAM labelled Oligonucleotide probes• dNTPS (dATP, dCTP, dATP, dTTP)• Stabilisers, including Bovine serum
albumin (<1% (w/w))• Synthetic RNA
Optical film seal Desiccant pouch
Quantity 1 per kit 1 per kit 1 per kit
Information Non-hazardous Non-hazardous Non-hazardous
Low profile block Require GDL Product Code: ID-CoV2-01-LPU
Company Model
Roche LightCycler 480 II Real-Time PCR System
Bio plate block Require GDL Product Code: ID-CoV2-01-BPU
Company Model
Bio-Rad CFX96 Real-Time PCR Detection System
Fast plate block Require GDL Product Code: ID-CoV2-01-FPU
Company Model
Applied Biosystems 7500 Fast Real-Time PCR System
Ordering details: Ensure the correct plate type is used for the instrument
• Single-use RNase, DNase-
free aerosol resistant
pipette tips (p20, p200)
• Any PPE required by local
safety guidelines
• 96 well cool block
• Genedrive 96 Exporter
software (product code:
GE-01) (optional)
ID-COV2-IFU-USA V3 5
For suggested controls please see the
Quality Controls section of this document
*WHO Laboratory testing for 2019 novel coronavirus (2019-nCoV) in suspected human cases Guidance March 2020
ID-COV2-IFU-USA V3 6
Specimen collection, handling and storageThe Genedrive 96 SARS-CoV-2 Kit is intended for use on
RNA extracts stored in viral transport media and derived
from upper respiratory specimens.
Due to the sensitivity of RNA toward shearing by freeze
thaw, do not use specimens that have undergone more
than one freeze-thaw cycle either pre- or post-extraction.
Table 2 indicates the World Health Organization (WHO)
recommendations for handling of specimen types.*
Note: Local regulations for specimen handling must take
priority over these recommendations.
Operating conditionsFor the operating conditions of third party instrumentation
please refer to the associated manufacturer’s user manual.
Nasopharyngeal swabs
Oropharyngeal swabs
Collection
Dacron or
polyester flocked
swabs in viral
transport medium
Dacron or
polyester flocked
swabs in viral
transport medium
Transport temperature
4˚C 4˚C
Initial storage 4˚C for 3≤ days 4˚C for 3≤ days
Long term storage
-70˚C for one
month
-70˚C for one
month
Table 2 WHO recommendations for handling of swab types
Storage and shelf life• Genedrive 96 SARS-CoV-2 Kit should be stored
between 2˚C – 30˚C
• The expiry date for each kit is identified on the kit label
• Expired kits must not be used
Warnings and precautions• RNA is extremely susceptible to shearing by freeze-
thaw cycles. Ensure freeze and thaw of all fresh
specimens and extracted RNA from specimens is
kept to an absolute minimum to prevent failed tests.
• RNA samples are susceptible to degradation by
contaminating RNases. Ensure all specimens are
handled with appropriate laboratory RNA handling
procedures.
• For in vitro diagnostic (IVD) use only
• For prescription use only
• Do not use Genedrive 96 SARS-CoV-2 Kits that have
exceeded the expiry date shown on the outer kit label.
• Each reaction well is designed to perform one test.
Do not reuse any reaction wells.
• The test should be performed in a clean environment,
on a flat surface.
• Each kit is single use, regardless of how many test
wells have been used.
• Immediately discard the kit if the protective foil pouch
has been pierced/broken in transit.
• Visually check for any missing beads in the 96 well
plate. Do not use any wells that do not contain a
lyophilized reagent bead.
• Protect the kits against humidity, prolonged exposure
to humidity may affect the performance of the product.
• Ensure the correct volume of specimen is used
(a change in volume may result in test failure).
• Do not disturb the Real-Time PCR instrument whilst
running a test.
• Do not attempt to remove the optical film seal from
the plate following the completion of the test. This
will result in high risk of local contamination.
• Dispose of plates immediately following run
completion. Prolonged storage following run
completion may increase risk of loss of seal integrity.
• Used plates should be treated as potentially infectious
materials and should be disposed of accordingly as
clinical waste, in accordance with local guidelines.
• Specimens must be treated as potentially infectious.
Take the necessary precautions during collection,
storage, treatment and disposal of specimens.
• PCR must always be carried out using good
laboratory practices.
ID-COV2-IFU-USA V3 7
Limitations of useThe Genedrive 96 SARS-CoV-2 Kit has been validated for
use with upper respiratory specimens run on the Roche
LightCycler 480 II, Applied Biosystems 7500 Fast
Real-Time PCR systems and Bio-Rad CFX96.
Any deviations from the procedures detailed within this
IFU may result in erroneous results or failure.
Samples should be handled and treated as if they are
infectious and all biosafety precautions should be followed.
All results should be interpreted by a trained
healthcare professional.
A negative result does not conclusively rule out the
possibility of SARS-CoV-2 infection. Patient medical history
and clinical symptoms must also be taken into account.
Whilst the test is fully exclusive to non-CoV-2 human
pathogens described in Table 11, a negative SARS-CoV-2
test does not rule out disease caused by other pathogens.
Laboratories should include a statement such as
‘the test has been validated but FDA’s independent
review of this validation is pending’ in test reports to
healthcare providers.
False positive results may be caused by:
• Cross-contamination of wells
• Local contamination caused by unsuitable handling of
positive SARS-CoV-2 samples or other positive materials
• Local contamination caused by unsuitable handling of
PCR amplicons.
False negative results may be caused by:
• Unsuitable collection, handling and/or storage of
samples (review RNase P control result)
• Sample taken from patient whilst outside of viraemic
phase of SARS-CoV-2
• IFU has not been followed
• Poor laboratory practice for handling RNA samples.
Test failed results may be caused by:
• Carryover of ethanol from RNA extraction wash buffers
• RNA extraction failure
• IFU has not been followed
• Incorrect automated RNA extraction system
input and eluate volumes
• Not adding specimen or run control to reaction well
• Not including the RNA process extraction control
Use of the Extraction Control - Human Lymphocyte RNA
Human Lymphocyte RNA may be used as an optional
extraction control to assess RNA extraction performance.
When included in the RNA extraction procedure,
successful RNA extraction and test performance will be
indicated by positive detection of RNase P. The Extraction
Control is supplied in an anhydrous 10 µg format. Each vial
may be used to perform between sixty-six and 100 kits
depending on RNA isolation methodology – equivalent to
between 6,336 to 9,600 reactions.
The Extraction Control must first be reconstituted using
Nuclease-Free Water, then dispensed into single use
aliquots, each sufficient to set up one full Genedrive 96
SARS-CoV-2 Kit (1 x 96 well plate). Single use aliquots
must be stored frozen at -80°C. When performing the RNA
extraction procedure for specimens to be analysed using
the Genedrive 96 SARS-CoV-2 Kit, a single aliquot of the
RNA extraction control should be thawed on ice and the
appropriate amount added directly to the RNA extraction
kit lysis/binding buffer before dispensation of the lysis/
binding buffer master mix to each of the specimens.
For specific instructions for each validated RNA extraction
kit see below.
If you require support using the Extraction control
with an extraction kit not listed, please contact
[email protected] for guidance.
1. The Human Lymphocyte RNA is shipped at ambient
temperature. On receipt, store at ambient temperature
(15-30°C) until required.
2. On first use, reconstitute RNA by addition of
200 μL room-temperature NFW, and incubate
at room temperature for 10 minutes. Do not use
ice-cold water for reconstitution. Mix by brief vortex
and centrifugation.
3. Transfer 200 μL reconstituted RNA to a nuclease-free
2 mL microfuge tube and add 1800 μL nuclease-free
water to a final concentration of 5 ng/mL. Mix by
pipetting several times, followed by brief vortex and
centrifugation.
• RNA at 5 ng/μL is stable on ice for up to 8 hours.
• Check Table 3 and the specific instructions below for
the required aliquot volume for the RNA extraction kit
being utilised.
• Prepare aliquots in nuclease-free microfuge tubes and
store as single-use aliquots at -80°C.
• Thaw on ice when required for use.
• Aliquots are intended for single use only, do not
re-freeze thawed aliquots. Repeated freeze-thaw will
impact RNA integrity and risk Extraction Control failure.
• As a guide, each single-use aliquot is sufficient for
1 X 96-well kit with 15% overage.
Calculation of the quantity of extraction control
required per sample (A) is performed according to the
following formula:
Where:
A = Quantity of extraction control to spike per sample
extraction (ng)
B = Quantity of extraction control required per 20 μL
PCR assay (ng)
C = The elution volume of the RNA isolation kit (μL)
20 = Genedrive 96 SARS-CoV-2 Kit volume (20 μL)
ID-COV2-IFU-USA V3 8
Test procedure
1. RNA Extraction The Genedrive 96 SARS-CoV-2 Kit is intended for use with commercially available RNA extraction kits and Universal
Human Lymphocyte RNA (AMSBIO, ATR1254148-10) and is validated for use with the RNA extraction kits listed in Table 3.
Extraction Process Control
RNA Extraction Kit Manufacturer Product
CodeSpecimen Input (µL)
Lysis Buffer Volume/
Specimen (µL)
(A) Quantity of
Extraction Control required per
extraction (ng)
(B) Quantity of Extraction
Control required per 20 µL assay
(ng)
(C) RNA Kit Elution
Volume (µL)
RNAdvance Viral RNA
purification Kit
Beckman Coulter
C63510 200 150 1.25 0.25 100
Sera-Xtracta Virus/Pathogen
KitCytiva
29514201/ 29506009
200 570 1.25 0.25 100
QIAamp Viral RNA Mini Kit
Qiagen52904 or
52906140 560 0.3 0.1 60
Virus/Pathogen Kit with QIAamp MiniElute Virus
Spin Kit
Qiagen 57704 200 200 0.4 0.1 80
Table 3 RNA extraction kits and extraction control requirements
Reagent volume produced will allow for at least 96 extractions (with 15% overage)
Reagent volume produced will allow for at least 96 extractions
1. Remove a single use aliquot (30 μL of 5 ng/μL) of the
RNA extraction control from the freezer (-80°C) and
allow to thaw on ice or a chilled cool block (2-8°C).
2. Mix by laboratory vortex followed by brief centrifugation.
3. Immediately prior to commencing RNA isolation, add
27.5 μL of the RNA extraction control (5 ng/μL) to 16.5
mL RNAdvance Viral Lysis Buffer (LBF) and mix by
laboratory vortex.
4. Use 150 μL of extraction control/LBF mix per
extraction (1.25 ng/extraction – final assay
concentration of 0.25 ng).
5. Perform the extraction according to the RNAdvance
Viral RNA Purification Kit IFU.
1. Remove a single use aliquot (20 μL of 5 ng/μL) of the
RNA extraction control from the freezer (-80°C) and
allow to thaw on ice or a chilled cool block (2-8°C).
2. Mix by laboratory vortex followed by brief
centrifugation.
3. Add carrier RNA provided to AVL buffer (if not
already added) according to the QIAamp Viral
RNA Mini Kit IFU.
4. Add 15 μL of extraction control to 1235 μL Nuclease
Free Water and mix by laboratory vortex, briefly
centrifuge (0.06 ng/μL).
5. Immediately prior to commencing RNA isolation, add
550 μL of diluted RNA extraction control (0.06 ng/μL)
to 61.6 mL AVL buffer and mix by laboratory vortex.
6. Use 560 μL of extraction control/AVL mix per
extraction (0.3 ng per extraction – final assay
concentration of 0.1 ng).
7. Perform the extraction according to the QIAamp Viral
RNA Mini Kit IFU.
RNAdvance Viral RNA Purification Kit (Beckman Coulter):
QIAamp Viral RNA Mini Kit (Qiagen):
ID-COV2-IFU-USA V3 9
ID-COV2-IFU-USA V3 10
The use of carrier RNA is crucial for extraction efficiency and stability of the extracted nucleic acid. Follow RNA extraction kit instructions for processing.CAUTION
The procedure must be conducted in accordance
with the manufacturer’s instructions for use. Ensure
that specimen swabs have been handled according
to the conditions set out in the Specimen collection,
handling and storage section of this IFU.
If using an automated extraction system, perform the
extraction using specimen input to eluate volume
ratio of 5:1 - for example, use a 200 μL specimen
input volume with a 40 μL elution volume.
The quality of the extracted RNA has a profound impact on the performance of the entire test system. It is recommended to ensure that the system used for nucleic acid extraction is compatible with Real-Time PCR technology.
If using a spin column based sample preparation procedure including washing buffers containing ethanol, it is highly recommended to perform an additional centrifugation step for 1 min at approximately 17000 x g (~ 13000 rpm), using a new collection tube, prior to the elution of the nucleic acid.
CAUTION
ID-COV2-IFU-USA V3 11
2.9 Being careful to contact only the end of the optical
film seal and not the adhesive surface, place the
seal on the plate evenly, and ensure a tight seal is
obtained for each well. A microtitre sealing brayer
(or similar) may be used to ensure an even seal.
Carefully place seal on the plate and then rub the
FLAT edge of the applicator back and forth along the
long edge of the plate, sealing from the middle of
the plate outwards in all directions. Next rub the flat
edge of the applicator back and forth along the short
edge (width) of the plate. Finally, rub the end of the
applicator horizontally and vertically between all wells
and around all the outside edges of the plate, using
small back and forth motions to form a complete seal
around all the outside wells.
2.10 Carefully drop from a 1cm height 3 times onto the
work surface, to promote liquid movement to the
bottom of the well.
2.11 Using a tube vortex with the flat attachment, or plate
vortex, pulse mix cool block/plate for 3 x 15 seconds
at full power.
CAUTION
It is crucial to ensure sufficient mixing at this stage
of the procedure.
2.12 Drop from 1cm height 3 times onto the work
surface, to promote liquid movement to the
bottom of the well.
2.13 Incubate on cool block for 5 minutes.
2.14 Using a tube vortex with the flat attachment, or
plate vortex, pulse mix cool block/plate for 3 x 15
seconds at full power (total 45 seconds).
(Proceed to step 2.18)
CAUTION
It is crucial to ensure sufficient mixing at this stage
of the procedure.
Alternatively, pipette based reagent mixing can be
pursued to prepare the reactions following set up steps
on the next page:
2. Reaction setup and run protocol – amplification and detection on RT-PCR instrument
To ensure high test sensitivity and specificity, keep
the 96 well plate cool during specimen addition,
mixing and incubation steps. We recommend using
a 96 well plate cool block that has been cooled
down to 2-8 °C.
• Follow good laboratory practice for RNA
handling and PCR.
• It is important to ensure the set up area is free
of RNase and sources of PCR contamination.
• Perform all steps in a clean PCR hood.
2.1 Open the kit pouch using the tear seal.
2.2 Remove the 96-well plate and the optical film seal
provided. Once removed from the pouch, the sealed
assay kit is stable for 5 days.
2.3 Place the Genedrive 96 SARS-CoV-2 Kit 96 well plate
in a 96 well cold block.
2.4 Carefully peel back the plate foil seal. It is
recommended to peel back in sections so as to
provide further protection against contamination.
Once foil seal is removed, the assay kit is stable for
4 hours.
2.5 Inspect each well for the presence of a reagent bead.
2.6 Add 20 μL of specimen (or control) to each well,
using a fresh pipette tip. Ensure you do not touch
the lyophilised reagent bead with the pipette tip. The
specimen will begin to rehydrate the reagent bead.
2.7 Once reconstituted, the assay kit is stable for 1
hour (from reconstitution to initiation of cycling
programme).
If reactions are insufficiently mixed performance of
the kit may be variable.
It is critical that no well is left non-reconstituted. Empty wells containing a reagent bead will cause substantial interference that results in incorrect readings on LightCycler and CFX96 systems, and with ABI 7500 Fast may result in overall plate failure.
CAUTION
Manual reaction set up
If setting up the reactions manually perform the
following actions:
2.8 Once plate dispensation is complete, prepare to seal
the plate with the optical film seal provided. Separate
the optical film seal from the paper backing by
separation of seal/paper at the ends.
ID-COV2-IFU-USA V3 12
Pipette based reagent mixing
2.15 Once plate dispensation is complete leave to
reconstitute for 5 mins.
2.16 Set multi-channel pipette step volume to 15 μL,
and mix by pipetting in and out 10 times.
2.17 Remove the plate from the automated laboratory
system and apply the optical film seal – being
careful to contact only the end of the optical film
seal and not the adhesive surface. Place the
seal on the plate evenly, and ensure a tight seal
is obtained for each well. A microtitre sealing
brayer (or similar) may be used to ensure an
even seal. Carefully place seal on the plate
and then rub the FLAT edge of the applicator
back and forth along the long edge of the plate,
sealing from the middle of the plate outwards
in all directions. Next rub the flat edge of the
applicator back and forth along the short edge
(width) of the plate. Finally, rub the end of the
applicator horizontally and vertically between
all wells and around all the outside edges of the
plate, using small back and forth motions to form
a complete seal around all the outside wells.
2.18 Centrifuge in a plate centrifuge for 2 minutes
at 2000 rpm.
Ensure that the time between reconstitution of reagent
beads and initiation of cycling programme is no longer
than a maximum of 1 hour.
2.19 If using the Roche LightCycler 480 II, ensure the plate
barcode is on the correct side. The instrument will not
load the plate if the plate is labelled incorrectly. Ensure
the barcode is on the back of the plate. If not, remove
the barcode and re-attach it.
2.20 Load the plate into the Real-Time PCR instrument and
start the PCR cycling immediately.
Following completion of run, remove plate and dispose
of immediately.
ID-COV2-IFU-USA V3 13
Manual set up instructions for Roche LightCycler 480 II
• Input the following cycling parameters into the
instrument according to Table 4.
StageTemperature
(°C)Time
(hh:mm:ss)Ramp rate (°C/s)
Number of cycles
Reverse Transcription 42 00:10:00 4.4 1
Amplification
Denaturation 95 00:00:20 4.4
50Annealing 58 00:00:20 2.2
Extension 72 00:00:20 4.4
Melt
curve analysis
95 00:01:00 4.4
140 00:02:00 2.2
95 Continuous 0.11
Cooling 40 00:00:30 2.2 1
Table 4 Reverse Transcription-PCR cycling and melt analysis parameters
3. Real-Time PCR instrument setup
How to Setup a Genedrive 96 SARS-CoV-2 assay on the
Roche LightCycler 480 II System
To set up instrument parameters using a template file follow
the instructions below:
1. Download the Roche LightCycler 480 II template file
from the genedrive website at:
https://genedrive.com/assays/genedrive-96-exporter.php
2. On the instrument software home screen, click on the
‘Navigator’ icon and click on ‘Import’
3. Locate the Run Template File Filepath and click ‘Open’
4. The Run Template File will now be open
5. Click on the ‘Save’ icon and save in the Template Folder
6. Click on the ‘Close’ icon
7. To Start a run, Open New Experiment from Template
8. Locate the Saved template and open
9. Use the “Sample Editor” tab to enter specimen names
10. Click start run to open the save experiment dialog box,
save the run in a suitable location and click “OK” to start
the run
* Note – The Genedrive 96 SARS-CoV-2 test does not use SYBR green chemistry. This information is only used for the Roche LightCycler 480 II instrument wavelength filter settings.
The set up information contained in the template file is
outlined below in case you need to check or troubleshoot
the set up.
Instrument settings
Reaction volume 20 μL
Detection format SYBR green*/HRM dye
Target Reporter Quencher
SARS-CoV-2 E/N gene FAM* None
Internal RNA assay control
FAM None
Human RNase P target control
FAM None
• Input the following settings: • Fluorescent dyes
*FAM - Fluorescein labelled thymine residues
Dispose of plate after cycling complete. DO NOT OPEN as this will result in local PCR contamination.
ID-COV2-IFU-USA V3 14
How to Setup a Genedrive 96 SARS-CoV-2 assay
on the ABI 7500 Fast Real-Time PCR System
These instructions and associated run template file are
for use with the 7500 Software only.
To set up instrument parameters using a template file
follow the instructions below:
1. Download the ABI 7500 Fast template file from the
Genedrive website at:
https://genedrive.com/assays/genedrive-96-exporter.php
2. On the instrument software home screen, click on ‘New
Experiment’ drop down tab (top left of screen) and
select ‘From Template’
3. Locate the Template File Filepath and click ‘Open’
4. The ‘Experiment Menu’ will now be open
5. Input an ‘Experiment Name’ within the ‘Experiment
Properties’ tab
6. Click “Start run”. An alert message will appear saying
“Applied Biosystems recommends the SYBR Green
reagents to perform a melt curve experiment”, click
“OK” to start run
7. Save the run file in a suitable location, the reaction will
now start
8. Once the run has started select the “Plate setup” tab.
Insert specimen names into the “Define Samples” box,
then assign to wells using the “Assign Targets and
Samples” tab
Experimental Properties
Instrument Type: 7500 Fast (96 Wells)
Experiment Type: Quantitation - Standard Curve
Reagent Type: Other – Include Melt Curve
Ramp Speed: Fast
Run Method
Reaction volume: 20 μL
Expert mode: Tick, select filters 1, 2 and 3
Stage Step Temp (°C) Time (mm:ss)
Ramp Rate (%)
Holding Stage Step 1 42 10:00 100
Cycling Stage
(50 Cycles)
Step 1 95 00:20 100
Step 2 (Data Collection On) 58 00:20 100
Step 3 72 00:20 100
Melt Curve Stage
Step 1 95 01:00 100
Step 2 40 02:00 100
Step 3 (Data Collection On - Continuous) 95 00:05 3
Step 4 40 00:30 100
Target data
Reporter: JOE
Quencher: None
Passive reference dye: None
Manual set up instructions for ABI 7500 Fast Real-Time
PCR System
The set up information contained in the template file is
outlined below in case you need to check or troubleshoot
the set up
Table 5 Reverse Transcription-PCR cycling and melt analysis parameters
ID-COV2-IFU-USA V3 15
How to Setup a Genedrive 96 SARS-CoV-2 assay on the
Bio-Rad CFX96 Real-Time PCR Detection System
To set up instrument parameters using a template file
follow the instructions below:
1. Download the Bio-Rad CFX96 template zip file
containing the protocol template and the plate template
from the Genedrive website at: https://genedrive.com/
assays/genedrive-96-exporter.php
2. Open the Bio-Rad CFX Maestro software.
3. On the Startup Wizard pop-up, select ‘CFX96’ as the
instrument then select ‘User-defined’ run type.
4. On the Protocol tab of the Run Setup window, click
‘Select Existing’ and locate the Protocol Template
Filepath and click ‘Open’. Click ‘Next’ to move onto
the Plate tab.
5. On the Plate tab, click ‘Select Existing’ and locate the
Plate Template Filepath and click ‘Open’. Click ‘Next’ to
move on to the ‘Start Run’ tab.
6. If running multiple units, ensure the correct block is
selected.
7. Click ‘Start Run’ and save the run when prompted.
8. Sample names and biological groups for each well
may be input whilst the reaction is running. Click the
Real-Time Status tab then select “Plate Setup” then
“View/Edit Plate”. Use the plate editor window to detail
biological groups (i.e. wells assigned to controls and
clinical specimens), and also to enter specimen names
for individual wells.
Note: two templates are required to set up the test, a
protocol template and a plate template. Ensure that
you have used both templates.
Stage Step Temp (°C) Time (hh:mm:ss)
Holding Stage Step 1 42 00:10:00
Cycling Stage
Step 2 95 00:00:20
Step 3 (Plate read) 58 00:00:20
Step 4 72 00:00:20
Step 5 - Go to Step 2 44 times n/a n/a
Melt Curve Stage
Step 6 95 00:01:00
Step 7 40 00:02:00
Step 8 (0.5°C Increment, Plate read) 40-85 00:00:05
Step 9 40 00:00:30
Within ‘Plate Editor’ select the following parameters:
Fluorophore – FAM
Scan Mode – SYBR/FAM only
Plate Type – BR Clear
Manual set up instructions for Bio-Rad CFX96
• Within User-Defined run type and Protocol Editor,
input the following cycling parameters, accoring to
Table 5 using the ‘Protocol’ tab.
The set-up information contained in the template file is
outlined below in case you need to check or troubleshoot
the set up
Table 6 Cycling and melt analysis parameters
Sample volume - 20 μL
ID-COV2-IFU-USA V3 16
4. Result interpretation
Graph showing 6 Tm windows
4.1 Data analysis instructions for use with the Roche LightCycler 480 II• Select “Tm Calling”
• Set Max peaks to “6 or less”, followed by
“Calculate”
• Check the “Manual Tm Method” box, followed by
clicking the corner of the plate map to activate.
• Check all 6 Tm boxes in “Manual Tm”
• Enter the following:
• TM 1 56.1°C
• TM 2 61.5°C
• TM 3 65.0°C
• TM 4 72.7°C
• TM 5 73.5°C
• TM 6 80.1°C
RNase P SARS-CoV-2 IPC
56.1°C – 61.5°C 65.0°C – 72.7°C 73.5°C – 80.1°C
Human RNA control (RNase P) peak
SARS-CoV-2 peak Internal Process Control (IPC) peak
TM1 TM2 TM3 TM4 TM5 TM6
40
40.472
35.972
31.472
26.972
22.472
17.972
13.472
-0.028
4.472
-4.528
8.972
42 44 46 48 50 52 54 56 58 60 62 64 66
Temperature (˚C)
(d/d
T)
Flu
ore
sce
nce
(4
65
-510
)
68 70 72 74 76 78 80 82 84 86 88 90 92 94
Inspect First Order Derivative melt data visually
for presence of peaks in the respective target
Tm ranges. Control wells should be interpreted
first to validate the performance of the run. If
controls have passed, move on to analysing
specimen wells.
SARS-COV-2 Positive melt curve graph
Target Tm ranges
RNase P SARS-CoV-2 IPC
56.1°C – 61.5°C 65.0°C – 72.7°C 73.5°C – 80.1°C
40
33.614
30.114
26.614
23.114
19.614
16.114
12.614
9.114
-1.836
2.114
-4.886
5.614
42 44 46 48 50 52 54 56 58 60 62 64 66
Temperature (˚C)
(d/d
T)
Flu
ore
sce
nce
(4
65
-510
)
68 70 72 74 76 78 80 82 84 86 88 90 92 94
RNase P SARS-CoV-2 IPC
ID-COV2-IFU-USA V3 17
4.2 Manual data analysis instructions for use with the ABI 7500 Fast Real-Time PCR System• Select “Analyse Experiment” if opening a saved
file and/or proceed to select “Melt Curve” in the
Analysis module.
• Select “Derivative Reporter” in the drop down
menu plot settings.
• FODs for each well are displayed.
• Enter the following Tm ranges.
Target Tm ranges
Graph to show SARS-CoV-2 positive melt curve on ABI 7500 Fast Real-Time PCR System
SARS-CoV-2RNAse P IPC
RNase P SARS-CoV-2 IPC
48.8 - 54.1°C 54.8 - 64.3°C 65.8 - 75.0°C
4.3 Manual data analysis instructions for use with the Bio-Rad CFX96 Real-Time PCR Detection System• Once the run has finished the ‘Data Analysis’
window will open.
• Enter the following Tm ranges.
RNase P SARS-CoV-2 IPC
46.0°C - 56.0°C 59.0 - 66.0°C 67.0 to 75.0°C
Target Tm ranges
Graph to show example of SARS-Cov-2 melt curve on Bio-Rad CFX96 Real-Time PCR Detection System
RNase P IPCSARS-CoV-2
ID-COV2-IFU-USA V3 18
Follow instructions in Troubleshooting section for interpretation of failed results.
Table 7 Interpretation matrix
RNase PSARS-
CoV-2 E/NIPC Status Result Action
NEG NEG NEG INVALID TEST FAIL
Ensure RNA extraction
procedure is set up correctly.
Ensure the test procedure has
been followed correctly. Repeat
test using 50% dilution of
extracted specimen. If repeated
result remains invalid, consider
collecting a new specimen
NEG NEG POS INVALID
TEST FAIL
(No extraction
control detected)
Ensure RNA extraction
procedure is set up correctly.
Repeat extraction step
and re-test
POS or NEG POS POS or NEG VALIDSARS-CoV-2
POSITIVENone
POS NEG POS VALIDSARS-CoV-2
NEGATIVENone
POS NEG NEG VALIDSARS-CoV-2
NEGATIVENone
The absence and/or presence of the RNase P, SARS-CoV-2 and IPC peaks should be noted for each specimen and
the interpretation matrix used to perform result interpretation, see Table 7.
ID-COV2-IFU-USA V3 19
*Genedrive 96 Exporter software has been validated for use on a 64 bit computer running Windows 10.
Roche LightCycler 480 II1. Within the LightCycler 480 software:
2. Use the ‘Melt Curve Genotyping’ analysis settings,
highlight the wells to be analysed and click
‘calculate’
3. Right click on the ‘Melting curve’ graph (NOT the
melt peak first order derivative (FOD) data) and
click ‘export chart’
4. Change the tab to ‘data’ and ensure text file
is selected
5. Name the file and save in a suitable location
ABI 7500 FAST1. Within the 7500 Fast software:
2. Ensure all wells on the plate are selected in the
software
3. Select “Export”
4. Ensure “Results” and “Multicomponent data” check
boxes are selected
5. Choose file name and a suitable location – use .xls
format for export
6. Select “Start Export”
4.4 Genedrive 96 Exporter – result interpretation
Bio-Rad CFX961. Within the CFX Maestro software:
2. Once the run has finished the ‘Data Analysis’
window will open
3. On the toolbar click ‘Export’ then ‘Export all data
sheets’ as ‘CSV’. Choose a file name and save in a
suitable location
4. The “Filename - Melt Curve RFU Results_FAM”
can then be imported into the Genedrive 96
Exporter software
Genedrive 96 Exporter steps1. Open the software
2. Click ‘Input file’ and select export file type in the box
on the bottom right of the input screen. Navigate to
where the text file was saved and import the file
3. The analysis will complete automatically
Inspect run controls
1. Inspect run controls to ensure the validity of the run
before reporting the specimen results
2. When using an extraction control, ensure the
“Options” “Extraction control used” box has been
checked
3. If the run control has been directly added to the
plate without being taken through RNA extraction
– i.e. without the use of the extraction control –
ensure the “Options” “Extraction control used” box
has been not been checked
Result interpretation may be automated using the
optional Genedrive 96 Exporter software*
1. Go to the website address below and download the
Genedrive 96 Exporter software installer
https://genedrive.com/assays/genedrive-96-exporter.php
2. Unzip the file
3. Double-click the ‘Genedrive 96 Exporter’ Icon to
install the software
4. The Genedrive 96 Exporter will be installed in
C:\Genedrive96Exporter
5. Double-click the ‘Genedrive 96 Exporter’ icon to open
the software
ID-COV2-IFU-USA V3 20
Borderline specimen result interpretationA “Check Result” flag indicates that whilst Genedrive 96
Exporter classified the specimen as SARS-CoV-2-Positive
or Negative, the result is borderline. The graph for the
specimen should be visually inspected to determine the
final result
1. To perform a manual check - whilst holding down the
‘ctrl’ key on the keyboard - select a well containing
the positive run control, a well containing a negative
run control and the well with either the ‘Check result’
or ‘Test failed’ result
2. Selecting the positive and negative control wells
will allow any potential peaks to be compared to the
baseline (negative run control well) and a known
positive reaction (positive run control well)
NOTE: If a well contains a reagent bead that has not
been reconstituted it will result in interference and
produce an erroneous result. Ensure each plate is
inspected for any such wells and the results from
these wells are discounted.
3. Individual or multiple wells can be selected to display
the raw melt peak data (Tm, peak height, peak width)
and a graphical visualisation of FOD melt data
Colour Symbol ResultRed + SARS-CoV-2 POSITIVE
Hashed Red + CHECK RESULT
Dark Green - SARS-CoV-2 NEGATIVE
Hashed Green - CHECK RESULT
Blue X TEST FAIL
Result reporting
1. To interpret results from specimens ensure the
“Options” “Extraction control used” box has been
checked
2. Final sample result calls will be displayed
automatically on the image of the plate. Each well will
be colour coded and contain a symbol based upon
the result. Coding is as follows:
Export results
Results may be exported as .csv or .xls format
1. Select wells required for export
2. Data may be exported for each well individually, or by
row, column or entire plate
3. Select the wells of interest, then Select ‘Export
Results’ and save to desired location as either a .csv
or .xls format
ID-COV2-IFU-USA V3 22
Quality controlsEach test in the Genedrive 96 SARS-CoV-2 Kit contains
primers and probes for an extraction control (RNaseP).
The presence of the associated peak indicates a RNA
extraction was performed successfully and gives
increased confidence that negative results were not due
to poor RNA extraction.
Each test also contains primers, probes and target RNA
for an internal PCR control (IPC). This control amplifies
even within negative controls and the presence of the
peak indicates that PCR has functioned as expected and
has not been impacted by inhibition.
It is recommended that a positive and negative run control
are used when running the Genedrive 96 SARS-CoV-2 Kit.
Nuclease Free Water may be taken through the RNA
extraction method alongside clinical specimens to
function as a negative run control that can help monitor
the full process, including RNA extraction.
The Genedrive 96 SARS-CoV-2 Kit has been validated
using the following controls:
* Do not use for LoD verification studies
** If taken through RNA extraction process when using the RNA extraction control.
Positive run control materials
Product name ManufacturerProduct
Code
RNA extraction required?
Instruction Expected result
SARS-CoV-2
Standard
Exact
DiagnosticsCOV019 No
Dilute to 30 copies/
reaction or 1500
copies/mL. Follow test
procedure steps
RNase P peak
SARS-CoV-2 peak
IPC Peak
2019-nCoV_N
Positive Control
Integrated DNA
Technologies
(IDT)
10006625 No
Dilute to 1:2000 and
add 20 μl/well. Follow
test procedure steps*
SARS-CoV-2 peak
IPC peak
AccuPlex™
SARS-CoV-2
Reference
Material Kit
(positive control)
Seracare 0505-0126 Yes
Perform RNA
extraction. Follow test
procedure steps
SARS-CoV-2 Peak
IPC peak
AccuPlex™
SARS-CoV-2
verification panel
Seracare 0505-0168 Yes
Perform RNA
extraction. Follow test
procedure steps
SARS-CoV-2 Peak
IPC peak
Table 8 Positive control material
For the running of instrumentation control tests please refer to the respective instrument manufacturer user manuals.
Negative run control materials
Product name Manufacturer Product codeRNA
extraction required?
InstructionExpected
result
Nuclease Free
WaterVarious Various No
Follow test
procedure steps
IPC peak
RNase P peak**
Table 9 Negative control material
ID-COV2-IFU-USA V3 23
Troubleshooting
Sample type
Issue/Error Probable Cause Solution
Positive
control
SARS-CoV-2 peak
absent
Positive control
material not
added
Repeat test with new kit
PCR inhibitors
present
Clean area and repeat test with new kit. Ensure
appropriate PPE worn to reduce contamination risk
Human RNase P peak
present*
Contamination
with Human RNA
Clean area and repeat test with new kit. Ensure
appropriate PPE worn to reduce contamination risk
No peaks present PCR fail
Clean area and repeat test with new kit. Ensure
appropriate PPE worn to reduce contamination risk.
Ensure instrument parameters are correct
Negative
control
SARS-CoV-2 peak or
Human RNase P peak
present
Local
contamination
Clean area and repeat test with new kit. Ensure
appropriate PPE worn to reduce contamination risk.
Negative control material may be contaminated.
Consider using a fresh tube
Internal RNA control
peak absentPCR fail
Clean area and repeat test with new kit. Ensure
appropriate PPE worn to reduce contamination risk.
Ensure instrument parameters are correct
Test
sample
Human RNase P
peak absent
Inefficient / failed
RNA extraction
or no extraction
control added
Repeat extraction step and re-test
No peaks present PCR fail
Clean area and repeat test with new kit. Ensure
appropriate PPE worn to reduce contamination risk.
Ensure instrument parameters are correct. Check
specimen was added to well. Ensure automated RNA
extraction parameters are correct. - e.g. Ensure input
and eluate volumes are correct (See Table 2). Specimen
input volume: eluate volume ratio is 5:1
*In cases where an RNase P peak is not expected - i.e. when an
extraction control has not been used or the control used does not
contain RNase P RNA.
Table 10 Troubleshooting
ID-COV2-IFU-USA V3 24
Analytical sensitivityThe analytical sensitivity, or Limit of detection (LoD), is
defined as the lowest concentration which >95% of the
tested samples generate a positive result.
LoD of the Genedrive 96 SARS-CoV-2 kit was determined
by spiking of either Seracare Accuplex CoV-2 reference
material or intact heat inactivated SARS-CoV-2 virus into
CoV-2 negative oropharyngeal and nasopharyngeal
samples separately collected in viral transport media.
Isolated RNA was utilized for probit analysis at 24
replicates per dilution level, with the the LoD of the
Genedrive 96 SARS-CoV-2 assay being established as
10 copies/reaction (CI 8.2 – 12.77) or 0.5 copies/μl, and
verified across all platforms listed.
Performance characteristicsAll performance characteristics data were determined using manual result interpretation.
RNA Extraction
Kit
Real-Time PCR
Platform
SARS CoV-2 Standard
Target Titration
pre- or post-extraction
Copies / ml
Sample
Copies / Reaction
Detected Positive
LoD (copies per
reaction (95% CI))
LoD (copies/µL (95% CI))
Qiagen QIAamp
Viral RNA Mini Kit
Roche LightCycler
480 II
*Accuplex (V2)
Post-extraction
NA
20 10 5
2.5 1.25
48/48 45/48 35/48 23/48 14/48
9.82 (8.19 - 12.77)
0.49 (0.41-0.64)
Beckman Coulter
RNAdvance Viral RNA
purification Kit
ABI 7500 Fast
*Accuplex (V2)
Post-extraction
NA
20 10 5
2.5 1.25
24/24 24/24 23/24 20/24 15/24
4.53 (3.33 - 10.59)
0.23 (0.17-0.53)
ABI 7500 Fast
**Intact Heat
Inactivated SARS CoV-2
Pre-extraction
500 250 125 62.5 31.25
20 10 5
2.5 1.25
24/24 21/24 24/24 9/24 6/24
9.55 (7.58 - 13.97)
0.48 (0.38-0.70)
Roche LightCycler
480 II
**Intact Heat
Inactivated SARS CoV-2
Pre-extraction
500 250 125 62.5 31.25
20 10 5
2.5 1.25
24/24 21/24 22/24 16/24 7/24
10.25 (7.76 - 17.15)
0.51 (0.39-0.86)
Bio-Rad CFX96
**Intact Heat
Inactivated SARS CoV-2
Pre-extraction
500 250 125 62.5 31.25
20 10 5
2.5 1.25
24/24 24/24 22/24 14/24 10/24
5.7 (4.44 - 9.42)
0.29 (0.22-0.47)
Cytiva Sera-Xtracta
Virus/Pathogen
Kit
Bio-Rad CFX96
*Accuplex (V2)
Post-extraction
NA
20 10 5
2.5 1.25
24/24 22/24 16/24 9/24 4/24
10.27 (8.28 - 14.49)
0.51 (0.41-0.72)
*AccuPlex™ SARS-CoV-2 Verification Panel; V2, partial genome (SeraCare, 0505-0129)
**Heat inactivated SARS CoV-2 virus (ATCC® VR-1986HK™, strain 2019-oV/USA-WA1/2020)
Table 11 LoD across Real-Time PCR platforms
ID-COV2-IFU-USA V3 25
The Genedrive 96 SARS-CoV-2 Kit is exclusive for non-
CoV-2 human pathogens listed in the table below, assessed
firstly by in silico analysis (<80% match of primers and probe,
Analytical specificity
Table 12 Exclusivity
with none resulting in viable amplification products detected
in the detection windows), and also by experimental
assessment at the copies per reaction detailed.
Pathogen In silico
Experimental (Copies per
reaction)
Human coronavirus 229E Yes 4.0 X 105
Human coronavirus OC43 Yes 3.4 X 105
Human coronavirus HKU1 Yes Not tested
Human coronavirus NL63 Yes 1.0 X 105
SARS-Coronavirus Yes 2.8 X 105
MERS coronavirus Yes 3.2 X 105
Parainfluenza virus 1 Yes 3.0 X 105
Parainfluenza virus 2 Yes 2.5 X 105
Parainfluenza virus 3 Yes 3.6 X 105
Parainfluenza virus 4 Yes 3.6 X 105
Rhinovirus Yes 1.0 X 105
Influenza A (H1N1) Yes 2.8 X 105
Influenza A /(H3N2) Yes 3.6 X 105
Influenza B Yes 2.5 X 105
Respiratory syncytial virus
(subtype A)Yes 3.6 X 105
Respiratory syncytial virus
(subtype B)Yes 3.0 X 105
Adenovirus Yes 3.8 X 105
Adenovirus (Ad.41) Yes 2.4 X 105
Haemophilus ducreyi Not tested 3.6 X 105
Haemophilus influenzae Yes 4.0 X 105
Human Metapneumovirus (hMPV) Yes 1.0 X 105
Streptococcus pneumoniae Yes 3.4 X 105
Mycoplasma pneumoniae Yes 3.2 X 105
Streptococcus pyogenes Yes 1.0 X 105
Bordetella pertussis Yes 3.8 X 105
Bordetella holmesii Yes 3.2 X 105
Bortedella Parapertussis Yes 3.4 X 105
Pathogen In silico
Experimental (Copies per
reaction)
Chlamydia trachomatis Yes 3.4 X 105
Candida albicans Yes 2.5 X 105
Candida auris Yes 2.6 X 105
Legionella pneumophila Yes 2.8 X 105
Staphylococcus salivarius Yes 1.0 X 106
Staphylococcus epidermis Yes 1.0 X 106
Pseudomonas aeruginosa Yes 1.0 X 106
Mycobacterium tuberculosis Yes 3.4 X 105
Chlamydia psittaci Yes Not tested
Chlamydia pneumonia Yes Not tested
Influenza C Yes Not tested
Parechovirus Yes Not tested
Corynebacterium diptheriae Yes Not tested
Legionella feeleei Yes Not tested
Legionella micdadei Yes Not tested
Legionella longbeachae Yes Not tested
Bacillus anthracosis (Anthrax) Yes Not tested
Moraxella cararrhalis Yes Not tested
Neisseria elongate and
miningitidisYes Not tested
Leptospirosis Yes Not tested
Coxiella burneti (Q-Fever) Yes Not tested
Staphylococcus aureus Yes Not tested
Enterovirus (type 71) Yes Not tested
Pneumocystis jirovecii (PJP) Yes Not tested
Bacterioides oralis Yes Not tested
Chlamydophila pneumoniae Yes Not tested
Nasal wash N/ANo assay cross-
reactivity (N/A)
ID-COV2-IFU-USA V3 26
InclusivityIn silico analysis concluded that the Genedrive 96 SARS
CoV-2 assay will detect all analyzed SARS-CoV-2 sequences
in the GISAID database by using a dual target design.
As of 22nd October 2020, 155,276 CoV-2 sequences
were available in the GISAID database for detailed
analysis of Genedrive 96 SARS CoV-2 Assay E & N gene
primer and probe homology. Of these, 137,246 genomes
(88.4%) covered the assay E & N gene target regions fully
to permit sequence alignment analysis of both target
assays together, and demonstrated 100% homology to
all Genedrive 96 SARS CoV-2 assay target primers and
probes. Summary of individual N and E gene assay target
alignments are shown in Table 14.
For the E gene, greater than 99% of sequences had 100%
homology to amplification primers and detection probe.
The majority of those with less than 100% homology contain
non-significant single mismatches distal to 3’ ends, being
PrecisionIntra-assay (within run), Inter-assay (between run),
Inter-instrument and Inter-lot (between reagent lot)
variability end-point melt analysis signal intensity for
SARS-CoV-2 RNase P
%CV SD %CV SD
Intra assay 11.68 0.21 14.64 0.14
Inter assay 5.11 0.09 9.10 0.09
Inter lot 3.00 0.05 6.61 0.06
Inter instrument 11.98 0.11 9.82 0.10
N gene E gene
Amplification primers
Detection Probe
Amplification primers
Detection Probe
Number of sequences 153,587 (100%) 154,234 (100%)
Number of sequences with 100% homology 141,219 (91.94%) 152,868 (99.53%) 152,873 (99.12%) 154,078 (99.9%)
Number of sequences with <100% homology 12,368 (8.05%) 719 (0.47%) 1,361 (0.88%) 156 (0.1%)
Number of sequences with predicted
potential to impact detection
386 (0.25%) 2 (0.001%) 137 (0.09%) 0 (0%)
Number detected by alternative dual target 385 / 386 719 / 719 136 / 137 n/a
Number of sequences remaining potentially
not detected
1 (0.0007%)* 0 (0%) 1 (0.0006%)* n/a
Table 13 Precision data
Table 14 Summary of Individual N and E gene target alignments
predicted not to affect amplification or detection. The
remaining 137 sequences with 2 or more mismatches
(0.09%) were 100% homologous to the alternative N
gene assay (with the exception of 1 sequence*; Egypt/
MASRI-025/2020; 3 consecutive mismatches in the E
gene forward primer at positions -6 to -8 relative to 3’).
Conversely, for the N gene, greater than 92% of
sequences are 100% homologous for amplification
primers (>99% for detection probe). The majority of
the remaining 12,368 sequences (11,982 sequences)
contain non-significant single mismatches distal to
3’ ends, being predicted not to affect amplification.
Those 386 sequences with 2 or more mismatches
(0.25%) were 100% homologous to the alternative
E-gene assay (with the exception of 1 sequence*; Egypt/
MASRI-025/2020; single mismatch in the N gene
reverse primer at position -3 relative to 3’ (estimated to
still result in 80% efficiency).
Genedrive 96 SARS-CoV-2 Kit probes when expressed as
a ratio of the internal PCR control signal intensity. All results
displayed to 2 decimal points.
ID-COV2-IFU-USA V3 27
Diagnostic accuracyContrived Upper Respiratory Tract Specimens
A clinical performance study was performed to evaluate
the performance of the Genedrive 96 SARS-CoV-2 Kit
using nasopharyngeal (NP) and oropharyngeal (OP) swabs.
Sixty contrived upper respiratory tract specimens were
tested:
• Thirty contrived positive NP/OP specimens
• Of which twenty were 2x the Limit of Detection (429
copies/mL, projected assay concentration of 20
copies/reaction)
• Of which ten were 5x the Limit of Detection (1071
copies/mL, projected assay concentration of 50
copies/reaction)
• Thirty contrived negative NP/OP specimens
Specimens were contrived by spiking known
concentrations of viral protein-coated RNA (Seracare
Accuplex Verification Panel), relative to the product
LoD, into pooled NP/OP specimen matrix which was
determined to be negative by the collection site prior
to spiking in the RNA. Specimens were randomized
and blinded. RNA Extraction was performed using the
Qiagen QIAmp Viral RNA mini Kit. RT-PCR analysis was
performed using the Roche LightCycler 480 II across
three reagent lots.
100.00% (30/30, 95% CI 88.65 to 100.00) of positive
specimens returned a positive result, and 100.00% (30/30,
95% CI 88.65 to 100.00) of negative specimens returned a
negative result, see Table 15.
Table 15 Clinical performance of the Genedrive 96 SARS-CoV-2 Kit on contrived upper respiratory
tract specimens performed using the Roche LightCycler 480 II Real-Time PCR Instrument.
Expected result
Positive Negative Total
Genedrive 96
SARS-CoV-2 Kit
Positive 30 0 30
Negative 0 30 30
Total 30 30 60
Positive Specimens Detected 100.00% 95% CI 88.65% to 100.00%
Negative Specimens Not Detected 100.00% 95% CI 88.65% to 100.00%
ID-COV2-IFU-USA V3 28
EUA SARS-CoV-2 Assay
Positive Negative Total
Genedrive 96
SARS-CoV-2 Kit
Positive 100 0 100
Negative 0 50 50
Total 100 50 150
Positive Percent Agreement 100.00% 95% CI 96.30% to 100.00%
Negative Percent Agreement 100.00% 95% CI 92.87% to 100.00%
Table 16: Clinical performance of the Genedrive 96 SARS-CoV-2 Kit on upper respiratory tract
specimens performed using the Roche LightCycler 480 II Real-Time PCR Instrument.
Upper Respiratory Specimens – EUA SARS-CoV-2 Assay
comparator
A clinical performance study was performed to evaluate the
performance of the Genedrive 96 SARS-CoV-2 Kit using
nasopharyngeal (NP) and oropharyngeal (OP) swabs.
150 natural NP/OP specimens were tested using the
Genedrive 96 SARS-CoV-2 Kit on the Roche LightCycler
480 II Real Time PCR Instrument. Specimens were
collected from one site, and each specimen was tested
using an EUA SARS-CoV-2 Assay.
Extraction of specimens was performed using three
different RNA extraction methods: Maxwell® RSC Buccal
Swab DNA Kit (N=120); Roche cobas extraction (N=21); and
Qiagen Virus/Pathogen Kit with QIAamp MinElute Virus
Spin Kit (N=9).
The positive percent agreement was 100.00% (95%
CI 96.30% to 100.00%), whilst the negative percent
agreement was 100.00% (95% CI 92.87% to 100.00%),
see Table 16.
ID-COV2-IFU-USA V3 29
Quick reference guide – manual set up
Step 1Place 96 well plate in a cool block
for the duration of reaction setup.
Step 2Carefully peel back the foil seal. It is
recommended to do this in sections
to minimise open wells during
pipetting. Check there is a reagent
bead in each well.
Step 5Carefully drop cool block 3 times from
1 cm height. Vortex the cool block/
plate for 3 x 15 seconds at full power
(total 45 seconds).
Step 8Centrifuge the plate for 2 minutes
at 2000 RPM.
Step 4Carefully peel optical film seal from its
backing and seal the plate. Use of a
microtitre sealing brayer or similar is
recommended.
Step 7Vortex the cool block/plate for
3 x 15 seconds at full power
(total 45 seconds).
Step 3Pipette 20 μL of extracted specimen
or control into the side of each well.
Add 20 μL of Nuclease Free Water
to any wells that do not contain a
specimen. Ensure all beads have
been rehydrated.
Step 6Carefully drop cool block 3 times from
1 cm height. Incubate on cool block
for 5 minutes.
Step 9Load the plate on the Real-Time PCR
instrument and start the program
immediately. Ensure all the run
parameters are correct before starting.
ID-COV2-IFU-USA V3 32
Genedrive® is a registered trademark of Genedrive Diagnostics Ltd.
Copyright © 2021 genedrive plc.
All rights reserved
Use By Single useConsult Instructions
for Use
Temperature limitation Contains sufficient for <n> tests
Product code
ManufacturerIn Vitro Diagnostic
Medical DeviceIVD Batch codeLOT
REF
Date issued: February 2021
IVD
Manufacturer Details
Genedrive Diagnostics Ltd. 48 Grafton Street,
Manchester M13 9XX United Kingdom
Tel +44 (0)161 989 0245
For general and technical enquires: [email protected]
For complaints: [email protected]
For distributor details please visit the website at:
www.genedrive.com
Prescription
Use only