Upload
faizan-akram
View
334
Download
0
Embed Size (px)
Citation preview
P a g e | 1
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Gas Chromatography
REAL PHARMACISTS
GROUP (RPG)
P a g e | 2
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Table Of Contents
History & Introduction
Advantages Of Gas Chromatography
Components Of Gas Chromatography
Detectors
Gas Chromatography-Mass
Spectrometry (GC-MS)
P a g e | 4
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
History:-
The father of modern gas chromatography is
Nobel Prize winner John Porter Martin, who also developed
the first liquid-gas chromatograph. (1950).
Introduction:-
Gas chromatography – “It is a process of separating
component(s) from the given crude drug by using a
gaseous mobile phase”‟
It involves a sample being vaporized and injected onto the
head of the chromatographic column. The sample is
transported through the column by the flow of inert,
gaseous mobile phase. The column itself contains a liquid
stationary phase which is adsorbed onto the surface of an
inert solid
Two major types
• Gas-solid chromatography
(stationary phase: solid)
• Gas-liquid chromatography
(stationary phase: immobilized liquid)
Gas - Solid Chromatography (GSC)
o The stationary phase, in this case, is a solid like silica or
alumina.
o It is the affinity of solutes towards adsorption onto the
stationary phase which determines, in part, the retention
time.
o The mobile phase is, of course, a suitable carrier gas.
o Most useful for the separation and analysis of gases like
CH4, CO2 CO, ... etc.
o The use of GSC in practice is considered marginal when
compared to gas liquid chromatography.
o
o
P a g e | 5
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Gas - Liquid Chromatography (GLC)
The stationary phase is a liquid with very low volatility
while the mobile phase is a suitable carrier gas.
GLC is the most widely used technique for separation of
volatile species.
The presence of a wide variety of stationary phases with
contrasting selectivity‟s and easy column preparation add
to the assets of GLC or simply GC.
o Advantages of Gas Chromatography
o The technique has strong separation power and even
complex mixture can be resolved into constituents.
o The sensitivity of the method is quite high.
o It gives good precision and accuracy.
o The analysis is completed in a short time.
o The cost of instrument is relatively low and its life is
generally long.
o The technique is relatively suitable for routine analysis.
P a g e | 6
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Types of Column
a):Open tubular (capillary) column:-
o When the stationary phase (liquid or solid) is uniformly
distributed on the interior surface of column it is called an
open tubular column.
o When the stationary phase is uniform film of few μm thick
liquid that is uniformly coat the inner capillary tubing.
The type of column is capillary column.
o They are further classified into three
types:-
i. Wall-coated open tubular column (WCOT)
ii. Support-coated open tubular column (SCOT)
iii. Porous-layer open tubular column (PLOT)
b).Packed Columns:-
o These columns are fabricated from glass, stainless steel,
copper, or other suitable tubes.
o Stainless steel is the most common tubing used with
internal diameters from 1-4 mm.
o The column is packed with finely divided particles (<100-
300 mm diameter), which is coated with stationary phase.
However, glass tubes are also used for large-scale
separations.
P a g e | 7
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Components Of Gas Chromatography
o Carrier gas
o Flow regulators & Flow meters
o Injection devices
o Columns
o Temperature control devices
o Detectors
P a g e | 8
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
schematic diagram of a gas chromatograph:-
P a g e | 9
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Carrier Gas
Hydrogen ( H2
)
o better thermal conductivity
Disadvantage: it reacts with unsaturated compounds &
inflammable
Helium ( He)
o excellent thermal conductivity
o it is expensive
Nitrogen ( N2
)o reduced sensitivity o it is inexpensive
Requirements Of A Carrier Gas:- o Inertness
o Suitable for the detector
o High purity
o Easily available
o Cheap
o Should not cause the risk of fire
o Should give best column performance
o Impurities in the carrier gas:-
o Impurities in the carrier gas degrade the stationary phase.
High-quality gases should be used, and even they should
be passed through purifiers to remove O2, H2O, and traces
of organic compounds prior to entering the column.
o Steel or copper tubing, rather than plastic or rubber
tubing, should be used for gas lines because metals are
less permeable to air and do not release volatile
contaminants into the gas stream.
P a g e | 10
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Flow regulators & Flow meters
deliver the gas with uniform pressure/flow rate
o Rota meter:-
placed before column inlet it has a glass
tube with a float held on to a spring the level of the float
is determined by the flow rate of carrier gas
o Soap Bubble Meter:-
Similar to Rota meter & instead of
a float, soap bubble formed indicates the flow rate
Injection devices
1)-Split injection: o routine method o % sample to column
o remainder to waste
o The injector temperatureis kept high
(for example,350 ) to promote
fast evaporation
o Split Ratio:-
The proportion of sample that does
not reach the column is called the
split ratio and typically ranges from
50:1 to 600:1.
P a g e | 11
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
2)-Split less injection:
o all sample to column
o best for quantitative analysis
o only for trace analysis, low [sample]
o
o Injector temperature for splitless
injection is lower (220 ) than that for split
injection, because the sample spends more
time in the port and we do not want it to
decompose.
3)-On-Column Injection:
o On-column injection is used for samples
that decompose above their boiling points
and is preferred for quantitative analysis.
o Solution is injected directly into the
column,Without going through a hot injector.
Gas Chromatography – Columns
o Important part of GC
o Made up of glass or stainless steel
o Glass column- inert , highly fragile
COLUMNS can be classified
o Depending on its use
1. Analytical column
1-1.5 meters length & 3-6 mm d.m
2. Preparative column
3-6 meters length, 6-9mm d.m
P a g e | 12
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
o Depending on its nature:
a)-Packed columns
o contain a finely divided, inert, solid support material
( diatomaceous earth) coated with liquid stationary phase.
o Most packed columns are 1.5 - 10m in length and have an
internal diameter of 2 - 4mm
b).Open tubular or Capillary column
or Golay column
o Long capillary tubing 30-90 M in length
o Uniform & narrow d.m of 0.025 - 0.075 cm
o Made up of stainless steel & form of a coil
Disadvantage: more sample cannot loaded
o They are further classified into three types
a. Wall-coated Open Tubular Column (WCOT)
b. Support-coated Open Tubular Column (SCOT)
c. Porous-layer Open Tubular Column (PLOT)
P a g e | 13
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
a. Wall-coated Open Tubular Column (WCOT):
o Inner coated with the liquid stationary phase
o The most common type of wall coated open tubular
column used is fused-silica, because it is stronger, inert,
reliable, easy to use, and flexible
b. Support-coated Open Tubular Column (SCOT):
o Inner layer is of Solid support coated with liquid stationary
phase
o Open tubular column has a greater amount of stationary
phase than the wall coated column, so it can handle a
larger quantity of sample.
c. Porous-layer Open Tubular Column (PLOT):
o Inner layer is coated with stationary solid phase particles.
o The only difference b/w SCOT & PLOT is that a PLOT does
not have a liquid stationary phase.
P a g e | 14
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Gas Chromatography-Detectors
Definition:-
“Detector, sensor, sensing element any device
that receives a signal or stimulus (as heat or pressure or light or
motion etc.) and responds to it in a distinctive manner”
Detectors can be grouped into:-
a)-concentration dependant detectors
b)-mass flow dependant detectors
The signal from a concentration dependant detector is related
to the concentration of solute in the detector, and does not
usually destroy the sample Dilution of with make-up gas will
lower the detectors response.
Mass flow dependant detectors usually destroy the sample,
and the signal is related to the rate at which solute molecules
enter the detector. The response of a mass flow dependant
detector is unaffected by make-up gas.
G C – IDEAL DETECTORS
o Sensitive (10-8-10-15 g solute/s)
o Operate at high
o Tempreture (0-400 °C)
o Stable and reproducible
o Linear response
o Wide dynamic range
o Fast response
o Simple (reliable)
o Nondestructive
o Uniform response to all analytes
P a g e | 15
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Types Of Detectors
1) Thermal conductivity (TCD)
2) Flame Ionization Detector (FID)
3) Electron capture Detector (ECD)
4) Nitrogen-phosphorus Detector (NPD)
5) Flame photometric Detector (FPD)
6) Photo-ionization Detector (PID)
7) sulfur chemiluminescence detector
1-Thermal conductivity (TCD)
(Katharometer, Hot Wire Detector)
In the past, thermal conductivity detectors were most common
in gas chromatography because they are simple and universal.
Principle:-
“TCD is based upon changes in the thermal
conductivity of the gas stream brought about by the presence of
analyte molecules”
The sensing element of TCD is an electrically heated element
whose temperature at constant electrical power depends upon
the thermal conductivity of the surrounding gas.
o Helium is the carrier gas commonly used with a thermal
conductivity detector.
o Helium has the second highest thermal conductivity (after
H2), so any analyte mixed with helium lowers the
conductivity of the gas stream.
o The heated element may be a fine platinum, gold, or
tungsten wire or a semiconducting thermistor.
P a g e | 16
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Advantages of Katharometer:
o Linearity is good
o Applicable to most compounds
o Non destructive
o Simple & inexpensive
Disadvantages:
o Low sensitivity
o Affected by fluctuations in temperature and flow rate
o Biological samples cannot be analyzed
2-Flame Ionization Detector
o It operates by the principle that by change in conductivity
of the flame as the compound is burnt.
o The change in conductivity of the flame does not arise by
simple ionization of the compound , it is partial or
complete stripping of the compound to give charged
hydrogen-deficient polymers or aggregates of carbon of
low ionization potential.
o Carbon atoms (except carbonyl and carboxyl carbons)
produce CH radicals, which are thought to produce CHO
ions and electrons in the flame.
P a g e | 17
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
CH + O CHO + e
o Nitrogen gives best detection limit
o signal proportional to number of susceptible
o carbon atoms
o 100-fold better detection than thermal conductivity
o 107 linear response range
o FIDs are mass sensitive rather than conc. Sensitive
ADVANTAGES:
o µg quantities of the solute can be detected
o Stable
o Responds to most of the organic compounds
o Linearity is excellent
DISADVANTAGE:
o destroy the sample
P a g e | 18
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
3-Electron-Capture Detectors(ECD)
Principle:-
“The electron-capture detector is selective in its
response being highly sensitive to molecules containing
electronegative functional groups such as halogens, peroxides,
quinones, and nitro groups‟‟
Mechanism:-
The detector consists of a cavity that contains two electrodes
and a radiation source that emi - radiation (e.g.63Ni, 3H)
The collision between electrons and the carrier gas
(methane plus an inert gas) produces a plasma containing
electrons and positive ions.
Remember the following facts about ECD:
1. Electrons from a b-source ionize the carrier gas (nitrogen)
2. Organic molecules containing electronegative atoms
capture electrons and decrease current
3. Simple and reliable
4. Sensitive (10-15 g/s) to electronegative groups (halogens)
5. Largely non-destructive
6. Insensitive to amines, alcohols and hydrocarbons
7. Limited dynamic range (102)
8. Mass sensitive detector
APPLICATION:-
This detector is frequently used in the
analysis of chlorinated compound (Pesticides).
P a g e | 19
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
4-Nitrogen-phosphorus
4-Nitrogen-phosphorus Detector
(alkali flame detector)
o The nitrogen-phosphorus detector, also called an alkali
flame detector, is a modified flame ionization detector that
is especially sensitive to compounds containing N and P. Its
response to N and P is times greater than the response to
carbon.
o “It is particularly important for drug, pesticide, and
herbicide analyses”.
o Ions such as NO2, CN and PO2 , produced by these
elements when they contact a Rb2SO4 containing glass
bead at the burner tip, create the current that is
measured. N2 from air is inert to this detector and does
not interfere.
o The bead must be replaced periodically because RbSO4 is
consumed shows a chromatogram from a nitrogen-
phosphorus detector.
P a g e | 20
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Nitrogen-phosphorus
Detector
5-flame photometricdetector
(FPD)
o A flame photometric detector measures optical emission
from phosphorus, sulfur, lead, tin, or other selected
elements.
o When eluate passes through a H2-air flame, as in the
flame ionization detector, excited atoms emit
characteristic light.
o Phosphorus emission at 536 nm or sulfur emission at
394 nm can be isolated by a narrow-band interference
filter and detected with a photomultiplier tube.
P a g e | 21
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
6-Photo-ionization Detector (PID)
o A photoionization detector uses a vacuum
ultraviolet source to ionize aromatic and
unsaturated compounds, with little response
to saturated hydrocarbons or halocarbons.
o Electrons produced by the ionization
o are collected and measured.
7-sulfur chemiluminescence detector
o A sulfur chemiluminescence detector takes exhaust from
a flame ionization detector, in which sulfur has been
oxidized to SO, and mixes it with ozone to form an excited
state of that emits blue light and ultraviolet radiation.
o A nitrogen chemiluminescence detector is analogous.
Combustion of eluate at 1800°C converts nitrogen into NO,
which reacts with Ozone to create chemiluminescent
product. The response to N is 107 times greater than the
response of Cabon (C).
P a g e | 23
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
GC-MS Contents
A. GC-Mass Spectrometry Introduction
B. Application Of GC-MS
C. Sample Preparation
D. Method Development in Gas Chromatography
A:-Mass Spectrometry (MS)
Mass spectrometry (MS) is an analytical chemistry technique
that helps identify the amount and type of chemicals present in
a sample by measuring the mass-to-charge ratio and
abundance of gas-phaseions
Principle of GC-MS
o The sample solution is injected into the GC inlet where it is
vaporized and swept onto a chromatographic column by
the carrier gas (usually helium).
o The sample flows through the column and the compounds
comprising the mixture of interest are separated by virtue
of their relative interaction with the coating of the column
(stationary phase) and the carrier gas (mobile phase).
o The latter part of the column passes through a heated
transfer line and ends at the entrance to ion source where
compounds eluting from the column are converted to ions
and detected according to their mass to charge m/z ratio.
P a g e | 24
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
B:-Application Of GC-MS
Petrochemical and hydrocarbons analysis
Geochemical research
Forensic (arson, explosives, drugs, unknowns)
Environmental analysis
Pesticide analysis, food safety and quality
Pharmaceutical and drug analysis
C. Sample Preparation
Sample preparation is the process of transforming a sample
into a form that is suitable for analysis. This process might
entail extracting analyte from a complex matrix,
preconcentrating very dilute analytes to get a concentration
high enough to measure, removing or masking interfering
species, or chemically transforming (derivatizing) analyte
into a more convenient or more easily detected form.
a)- Solid-phase micro extraction
b)- Purge and trap
c)- Thermal desorption
a)-Solid-phase micro extraction:
Solid-phase microextraction is a method to extract compounds
from liquids, air, or even sludge without using any solvent.16
The key component is a fused-silica fiber coated with a film of
stationary phase similar to those used in gas chromatography.
b)- Purge and trap:
Purge and trap is a method for removing volatile
analytes from liquids or solids (such as groundwater
or soil), concentrating the analytes, and introducing
them into a gas chromatograph.
P a g e | 25
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
In contrast with solid-phase microextraction, which
removes only a portion of analyte from the sample,
the goal in purge and trap is to remove 100% of
analyte from the sample. Quantitative removal of
polar analytes from polar matrices can be difficult.
c)- Thermal desorption:
Thermal desorption is a method to release volatile
compounds from solid samples. Aweighed sample is
placed in a steel or glass tube and held in place with
glass wool. The sample is purged with carrier gas to
remove O2, which is vented to the air, not into the
chromatography column.
The desorption tube is then connected to the
chromatography column and heated to release
volatile substances that are collected by cold trapping
at the beginning of the column. The column is then
heated rapidly to initiate chromatography.
D. Method Development in Gas
Chromatography
Order Of Decision:-
1. Goal Of Analysis
2. Sample Preparation
3. Choosing the Detector
4. Selecting the Column
5. Choosing the Injection Method
P a g e | 26
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
1. Goal Of Analysis
What is required from the analysis? Is it qualitative
identification of components in a mixture?
Will you require high-resolution separation of everything or
do you just need good resolution in a portion of the
chromatogram?
Can you sacrifice resolution to shorten the analysis time?
Do you need quantitative analysis of one or many
components?
Do you need high precision?
Will analytes be present in adequate concentration or do you
need
preconcentration or a very sensitive detector for ultratrace
analysis? How much can the Analysis cost?
2. Sample Preparation
The key to successful chromatography of a complex sample is to
clean it up before it ever sees the column.
solid-phase microextraction,
purge and trap
& thermal desorption
to isolate volatile components from complex matrices.
Other methods include:-
liquid extraction, supercritical fluid extraction, and solid-phase
extraction,
P a g e | 27
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
3. Choosing the Detector
The next step is to choose a detector. Do you need
information about everything in the sample or do you
want a detector that is specific for a particular element or
a particular class of compounds?
The most general purpose detector for open tubular
chromatography is a mass spectrometer.
1. Thermal conductivity (TCD)
2. Flame Ionization Detector (FID)
3. Electron capture Detector (ECD)
4. Nitrogen-phosphorus Detector (NPD)
5. Flame photometric Detector (FPD)
6. Photo-ionization Detector (PID)
4. Selecting the Column
The basic choices are the stationary phase, column
diameter and length, and the thickness of stationary
phase. A nonpolar stationary phase is most useful.
An intermediate polarity stationary phase will handle most
separations that the nonpolar column cannot.
For highly polar compounds, a strongly polar column
might be necessary.
Optical isomers and closely related geometric isomers
require special stationary phases for separation.
5. Choosing the Injection Method
The last major decision is how to inject the sample.
Split injection:
o concentrated sample
o high resolution
o dirty samples (use packed liner)
o could cause thermal decomposition
P a g e | 28
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Splitless injection:
o dilute sample
o high resolution
o requires solvent trapping or cold trapping
On-column injection:
o best for quantitative analysis
o thermally sensitive compounds
o low resolution
Application Of Gas Chromatography
1)- Qualitative Analysis – by comparing the retention time
or volume of the sample to the standard / by collecting the
individual components as they emerge from the chromatograph
and subsequently identifying these compounds by other method
2)- Quantitative Analysis- area under a single component
elution peak is proportional to the quantity of the detected
component/response factor of the detectors.
3)- Volatile Oils, official monograph gives chromatography
profile for some drugs. E.g. to aid distinction between anise oil
from star anise and that from Pimpinelle anisum.
4)- Separation of fatty acids derived from fixed oils.
5)- Miscellaneous-analysis of foods like carbohydrates, proteins,
lipids, vitamins, steroids, drug and pesticides residues, trace
elements
6)- Pollutants like formaldehyde, carbon monoxide, benzene,
DDT etc
7)- Dairy product analysis- rancidity
P a g e | 29
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
8)- Separation and identification of volatile materials, plastics,
natural and synthetic polymers, paints, and microbiological
samples.
9)- Inorganic compound analysis.
10)- Manufacturers of Cosmetics also use gas chromatography
to effectively measure how much of each chemical is used for
their products.
11)- Barbiturates & Benzodiazepines
12)- It can also be used for the
volatile samples such as human breathe,
blood, saliva and other secretions which
contains organic volatiles and can be easily
analyzed with Gas Chromatography.
P a g e | 30
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
References
1)- Quantitative Chemical Analysis 7th Edition Daniel C. Harris
Page # 528 ( Accessed at December 22,2014)
2)- Chromatographic methods 5th Edition By A. Braithwaite &
F.J. Smith Page # 166 (Accessed at December 23,2014)
3)- Chemical Analysis Modern Instrumentation Methods and
Techniques 2nd Edition Francis Rouessac and Annick
Rouessac Page # 31 (Accessed at December 24,2014)
4)- UNDERGRADUATE INSTRUMENTAL ANALYSIS 6th Edition By
James W. Robinson Page # 750 (December 26, 2014)
5)- VOGES Textbook Of Quantitative Chemical Analysis By
MENDHAM. Page # 317,745 (December 27, 2014)
6)- Organic Spectroscopy & Chromatography By M.YOUNAS
7)- Mass Spectrometry Principles & Applications By ED
Hoffmann & Stroobant Page # 217 (December 27,2014)
8)- Mass Spectrometry By H-GROSS Page # 482
9)- Principles Of Instrumental Analysis By Holler & Skoog
Page # 788 (Accessed at December 28,2014)
10)- Instrumental Techniques For Analytical Chemistry By
FRAUT SELLER Page # 125 (Accessed at December 29,2014)
11)- Dr.S. Ravi Sankar. Text book of pharmaceutical
analysis:Rx Publication, Tiruneveli: Edition 1997,1999,2001.
12)- A Textbook Of Analytical Chemistry Mahinder Singh
Page # 110 (Accessed at December 29,2014)
P a g e | 31
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
13)- Analytical Chemistry- A Modern Approach To Analytical
Science 2nd Edition By M. Mermet, M. Otto Page # 536,553
14)- Analytical Chemistry By Gary D. Christian 6th Edition
Page # 584,591,593 (December 30, 2014)
15)- Chromatography CONCEPTS & CONTRASTS By James M.
Miller Page # 141,165,278,285 (December 30,2014)
16)- Forensic Applications Of Gas Chromatography By Michelle
Groves Carlin.
17)- Amirav, A. Gordin, A. Poliak, M. Alon, T. and Fialkov, A.
B. (2008), "Gas Chromatography Mass Spectrometry with
Supersonic Molecular Beams". Journal of Mass Spectrometry 43:
141–163.
18)- A. Hites and K. Biemann, Analytical Chemistry, 855
19)- H.F. Waltor, J. Reyes: Modern Chemical analysis and
instrumentation.
20)- Structural Identification Of Organic Compounds With
Spectroscopic Techniques By Richard R. Ernst Page# 252
21)- Introduction To Pharmaceutical Chemical Analysis By Steen
Hansen Page # 191-200
22)- G Chatwal,S anand. Instrumental Methods of Chemical
analysis: Himalaya publishing house, Delhi: Edition June 1995.
23)- www.GasChromatography/Slideshare.com
24)- www.Gas Chromatography/Authorstream.com
25)- www.GasChromatography/Orgchem.com
P a g e | 32
Anf-Pharma.blogspot.com Real Pharmacists Group (RPG)
Presented to:
Respected Mam Saharish & 3rd Prof (Eve)
Presentation Given by Members of:
„„Real Pharmacist Group (RPG)‟‟
Made By:
Faizan Akram (Faizy)
Contributors:-
Ali Hammad Tariq
Zeeshan Saeed
For comments and mistakes correction please
contact.
Anf-Pharma.blogspot.com