Technical Specification FLUXUS® G70x TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 1 FLUXUS FLUXUS Measurement with transducers mounted by PermaRail G704 G709 Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features Precise bi-directional and highly dynamic flow measure- ment with the non-intrusive clamp-on technology High precision at fast and slow flow rates, high temper- ature and zero point stability Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results User-friendly design Transducers available for a wide range of inner pipe di- ameters ( ) and fluid temperatures ( ), approved transducers for hazardous areas available Measurement is unaffected by gas density, viscosity, composition, dust, humidity, temperature or pressure Applications Designed for industrial use in harsh environments, in gas processing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications: Measurement on natural gas pipelines and in natural gas storage installations Measurement of synthesized gas and injection gas Measurement for the gas supply industry 0.3 to 63 in -40 to +392 °F FM Class 1 Div. 2

Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

Technical Specification

FLUXUS® G70x

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 1

FLUXUS

FLUXUS

Measurement with transducers mounted by PermaRail

G704

G709

Gas Ultrasonic Flowmeter for Permanent Installation

Designed for wall mounting or installation in 19'' rack systems

Features

• Precise bi-directional and highly dynamic flow measure-ment with the non-intrusive clamp-on technology

• High precision at fast and slow flow rates, high temper-ature and zero point stability

• Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results

• User-friendly design

• Transducers available for a wide range of inner pipe di-ameters ( ) and fluid temperatures ( ),

• approved transducers for hazardous areas available

• Measurement is unaffected by gas density, viscosity, composition, dust, humidity, temperature or pressure

Applications

Designed for industrial use in harsh environments, in gas processing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications:

• Measurement on natural gas pipelines and in natural gas storage installations

• Measurement of synthesized gas and injection gas

• Measurement for the gas supply industry

0.3 to 63 in-40 to +392 °F

FM Class 1 Div. 2

Page 2: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

FLUXUS® G70x Technical Specification

Table of Contents

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-212

Function ........................................................................................................................................................... 3Measurement Principle ..................................................................................................................................... 3Calculation of Volumetric Flow Rate ................................................................................................................. 3Number of Sound Paths.................................................................................................................................... 4Typical Measurement Setup ............................................................................................................................. 5Standard Volumetric Flow Rate ........................................................................................................................ 5

Flow Transmitter ............................................................................................................................................. 6Technical Data .................................................................................................................................................. 6Dimensions ....................................................................................................................................................... 82 " Pipe Mounting Kit (optional)....................................................................................................................... 10Terminal Assignment ...................................................................................................................................... 11

Transducers................................................................................................................................................... 13Transducer Selection ...................................................................................................................................... 13Transducer Order Code .................................................................................................................................. 16Technical Data ................................................................................................................................................ 17

Transducer Mounting Fixture ...................................................................................................................... 24

Coupling Materials for Transducers............................................................................................................ 26

Damping Mats (optional) .............................................................................................................................. 27

Connection Systems..................................................................................................................................... 29Transducer Cable............................................................................................................................................ 29

Junction Box ................................................................................................................................................. 30Technical Data ................................................................................................................................................ 30Dimensions ..................................................................................................................................................... 302 " Pipe Mounting Kit (optional)....................................................................................................................... 30Terminal Assignment ...................................................................................................................................... 31

Clamp-on Temperature Probe (optional) .................................................................................................... 32

Wetted Temperature Probe (optional) ......................................................................................................... 34

Page 3: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 3

Technical Specification FLUXUS® G70x

Function

Measurement Principle

In order to measure the flow of a medium in a pipe, ultrasonic signals are used, employing the transit time dif-ference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a sec-ond transducer. These signals are emitted alternately in the flow direction and against it.

As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flowdirection is shorter than against the flow direction.

The transit time difference, ∆t, is measured and allows the flowmeter to determine the average flow velocityalong the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to ob-tain the area averaged flow velocity, which is proportional to the volumetric flow rate.

Two integrated microprocessors control the entire measuring process. This allows the flowmeter to removedisturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.

Calculation of Volumetric Flow Rate

= kRe . A . ka . ∆t/(2 . tfl)

where

Path of the ultrasonic signal Transit time difference ∆t

= volumetric flow ratekRe = fluid mechanics calibration factorA = cross-sectional pipe areaka = acoustical calibration factor∆t = transit time differencetfl = transit time in the medium

Page 4: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

4 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Number of Sound Paths

The number of sound paths is the number of transits of the ultrasonic signal through the medium in the pipe.Depending on the number of sound paths, the following methods of installation exist:

• reflect arrangementThe number of sound paths is even. Both of the transducers are mounted on the same side of the pipe. Correct positioning of the transducers is easier.

• diagonal arrangementThe number of sound paths is odd. Both of the transducers are mounted on opposite sides of the pipe.

• direct modeDiagonal mode with 1 sound path. This should be used in the case of a high signal attenuation by the medi-um, pipe or coatings.

The preferred method of installation depends on the application. While increasing the number of sound pathsincreases the accuracy of the measurement, signal attenuation increases as well. The optimum number ofsound paths for the parameters of the application will be determined automatically by the transmitter.

As the transducers can be mounted with the transducer mounting fixture in reflect arrangement or diagonal ar-rangement, the number of sound paths can be adjusted optimally for the application..

a = transducer distance

Reflect arrangement, number of sound paths: 2

Diagonal arrangement, number of sound paths: 3

Direct mode, number of sound paths: 1 Direct mode, number of sound paths: 1,negative transducer distance

a

a

a > 0 a < 0

Page 5: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 5

Technical Specification FLUXUS® G70x

Typical Measurement Setup

Standard Volumetric Flow Rate

The standard volumetric flow rate can be selected as physical quantity to be measured. It will be calculated in-ternally by:

N = . p/pN . TN/T . 1/K

where

The operational pressure p and the operational temperature T of the medium will be entered directly as fixedvalues into the transmitter.

or:

If inputs are installed (optional), pressure and temperature can be measured by the customer and fed in thetransmitter.

The gas compressibility coefficient K of the gas is entered in the transmitter:

• as fixed value or

• as approximation according to e.g. AGA8 or GERG

Example of a measurement setup in reflect arrangement with connection of the inputsto an external process pressure and process temperature measurement for standard volumetric flow rate calculation

N = standard volumetric flow rate= operating volumetric flow rate

pN = standard pressure (absolute value)p = operating pressure (absolute value)TN = standard temperature in KT = operating temperature in KK = compressibility coefficient of the gas: ratio of the compressibility factors of the gas at operating

conditions and at standard conditions Z/ZN

P T

transducers damping mat

transmitter

RS232

analog/binary outputs analog inputs

power supply

temperature probee.g. external pressure sensor

RS485(optional)

temperature probe

V· V·

Page 6: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

6 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Flow Transmitter

Technical Data

FLUXUS G704 G704.316SE G709design standard field device field device

with stainless steel housing19 " module

measurementmeasurement principle transit time difference correlation principleflow velocity 0.03 to 115 ft/s, depending on pipe diameterrepeatability 0.15 % of reading ±0.03 ft/smedium all acoustically conductive gases,

e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propanetemperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011

accuracyvolumetric flow rate ± 1 to 3 % of reading ±0.03 ft/s depending on application

± 0.5 % of reading ±0.03 ft/s with field calibrationflow transmitterpower supply 100 to 230 V/50 to 60 Hz or

20 to 32 V DCpower consumption < 15 Wnumber of flow measuring channels

1, optional: 2

signal attenuation 0 to 100 s, adjustablemeasuring cycle (1 channel) 100 to 1000 Hzresponse time 1 s (1 channel), option: 70 mshousing material aluminum, powder coated stainless steel 316L aluminumdegree of protection NEMA 4 NEMA 4X NEMA 1dimensions see dimensional drawing 42HP x 3U

(without back panel)

see dimensional drawingweight 6.2 lb 10.5 lb 3.8 lbfixation wall mounting, optional: 2 " pipe mounting 19 " rack mountingambient temperature -4 to +140 °Fdisplay 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanishexplosion protection (optional)

FM

marking NI/Cl. I,II,III/Div. 2/ GP. A,B,C,D,E,F,G/

T5 Ta = 60 °C

-

measuring functionsphysical quantities operating volumetric flow rate, standard volumetric flow rate, mass flow rate, flow velocitytotalizer volume, masscalculation functions average, difference, sum

(2 measuring channels necessary)diagnostic functions sound speed, signal amplitude, SNR, SCNR,

standard deviation of amplitudes and transit timesdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuescommunicationinterface - process integration (optional): RS485 (emitter) or Modbus RTU or HART or BACnet MS/TP

- diagnosis: RS232serial data kit (optional)

software (all Windows™ versions)

-FluxData: download of measurement data, graphical presentation,conversion to other formats (e.g. for Excel™)

-FluxKoef: creating medium data sets

-FluxSubstanceLoader: upload of medium data setsсable RS232

adapter RS232 - USB

Page 7: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 7

Technical Specification FLUXUS® G70x

outputs (optional)The outputs are galvanically isolated from the transmitter.

number on requestcurrent output

current output- range 0/4 to 20 mA- accuracy 0.1 % of reading ±15 μA- active output Rext < 500 Ω- passive output Uext = 4 to 24 V, depending on Rext, Rext < 1 kΩcurrent output I1 in HART mode- range 4 to 20 mA- passive output Uext = 10 to 24 V

voltage outputrange 0 to 1 V or 0 to 10 Vaccuracy 0 to 1 V: 0.1 % of reading ±1 mV

0 to 10 V: 0.1 % of reading ±10 mVinternal resistance Ri = 500 Ω

frequency outputrange 0 to 5 kHzopen collector 24 V/4 mA

binary outputReed relay 48 V/100 mA 48 V/100 mAopen collector 24 V/4 mA 24 V/4 mAoptorelay 26 V/100 mA -binary output as alarm output- functions limit, change of flow direction or error limit, change of flow direction or

errorbinary output as pulse output- pulse value 0.01 to 1000 units 0.01 to 1000 units- pulse width optorelay: 1 to 1000 ms

Reed relay, open collector: 80 to 1000 ms80 to 1000 ms

inputs (optional)The inputs are galvanically isolated from the transmitter.

number max. 4, on requesttemperature input

type Pt100/Pt1000connection 4-wirerange -238 to +1040 °Fresolution 0.01 Kaccuracy ±0.01 % of reading ±0.03 K

current inputaccuracy 0.1 % of reading ±10 μA 0.1 % of reading ±10 μA 0.1 % of reading ±10 μAactive input Ui = 24 V, Ri = 50 Ω, Pi < 0.5 W,

not short-circuit proofUi = 24 V, Ri = 50 Ω, Pi < 0.5 W,

not short-circuit proofUi = 15 V, Ri = 50 Ω, Pi < 0.5 W,

not short-circuit proof- range 0 to 20 mA 0 to 20 mA 0 to 20 mApassive input Ri = 50 Ω, Pi < 0.3 W Ri = 50 Ω, Pi < 0.3 W Ri = 50 Ω, Pi < 0.3 W- range -20 to +20 mA -20 to +20 mA -20 to +20 mA

voltage inputrange 0 to 1 Vaccuracy 0.1 % of reading ±1 mVinternal resistance Ri = 1 MΩ

FLUXUS G704 G704.316SE G709

Page 8: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

8 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Dimensions

FLUXUS G704

2.7

60.

39

Ø 0.18

6.42

7.87

11.310.43 fixing holes for wall mounting

thread: 6x M20 x 1.5cable gland: max. 6x 1/2 NPT

11.5

Ø 0

.17

7.3

brackets for wall mounting (optional)

Page 9: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 9

Technical Specification FLUXUS® G70x

FLUXUS G704.316SE

FLUXUS G709

in inch

6.42

12.6

cable gland: max. 6x 1/2 NPS with counter nut

14.57

10.0

4

3.43

3.86

8.19

5.08

8.74

0.59

4.84

4.33

M2.

51.

73

8.39

5.0

8

6.69

7.8

Page 10: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

10 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

2 " Pipe Mounting Kit (optional)

FLUXUS G704

FLUXUS G704.316SE

Page 11: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 11

Technical Specification FLUXUS® G70x

Terminal Assignment

2 The number, type and terminal assignment of the outputs and inputs will be customized.

FLUXUS G704, G704.316SE

power supplyterminal strip KL3

terminal connection (AC) connection (DC)PE earth earthN(-) neutral -L(+) phase +

transducersterminal strip KL1

extension cable (transducers ****LI*, *****52)transducer cable (transducers ****LI*)

transducer cable (transducers *****52)

measuring channel A measuring channel B measuring channel A measuring channel Bterminal connection terminal connection terminal connection

AV signal BV signal X_AV X_BV SMB connectorAVS shield BVS shield X_AR X_BR SMB connectorARS shield BRS shieldAR signal BR signal

outputs2 RS485, Modbus, BACnet (optional)

terminal strip KL4 terminal strip KL4terminal connection terminal connection

P1+ to P4+, P1- to P4- current output, voltage output, frequency output or binary output (Reed relay, open collector)

A+ signal +B- signal -

P5a to P7a, P5b to P7b

binary output 101 shield

inputs2

terminal strip KL2temperature probe passive current

sourceactive current

sourceterminal with connector without connector connection of

an active inputconnection of a passive inputdirect

connectionconnection with extension cable

direct connection

connection with extension cable

T1a to T4a red red red white not connected not connectedT1A to T4A red/blue gray red black - +T1b to T4b white/blue blue white red + not connectedT1B to T4B white white white green not connected -S1 to S4 shield shield - - not connected not connected

! " # ! " $ $ $

% % %

$ % $ % $ % $ %

#

#

#

& & ' & & '

#

( &

(

( &

(

Page 12: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

12 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

2 The number, type and terminal assignment of the outputs and inputs will be customized.

FLUXUS G709

power supplyterminal strip KL2

terminal connection (AC) terminal connection (DC)PE earth PE earthN neutral L- -L1 phase L+ +

transducersterminal strip KL6, KL8

extension cable (transducers ****LI*, *****52)transducer cable (transducers ****LI*)

measuring channel A measuring channel Bterminal connection terminal connection

AV signal BV signal

AVS shield BVS shield

ARS shield BRS shieldAR signal BR signal

outputs2 RS485, Modbus, BACnet (optional)

terminal strip KL1 terminal strip KL4terminal connection terminal connection

P1+ to P7+, P1- to P7- current output, voltage output, frequency output or binary output 4A+ signal +4B- signal -

P5a to P7a, P5b to P7b

binary output 43 shield

inputs2

terminal strip KL2temperature probe passive current

sourceactive current

sourceterminal with connector without connector connection of

an active inputconnection of a passive inputdirect

connectionconnection with extension cable

direct connection

connection with extension cable

T1a to T4a red red red white not connected not connectedT1A to T4A red/blue gray red black - +T1b to T4b white/blue blue white red + not connectedT1B to T4B white white white green not connected -S1 to S4 shield shield - - not connected not connected

# # # #

# #

& &

# #

) #

&

&

'

&

&

'

#

#

#

%

%

%

$

$

$

%

$

$

%

$

$

%

%

Page 13: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 13

Technical Specification FLUXUS® G70x

Transducers

Transducer Selection

Step 1a

Select a Lamb wave transducer:

Step 1b

If the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:

Step 2

inner pipe diameter d dependent on the flow velocity v of the medium in the pipe

The transducers are selected from the characteristics (see next page). Lamb wave transducers are selectedfrom the left column, shear wave transducers from the right column.

Lamb wave transducers: If the values d and v are not in the range, the diagonal arrangement with 1 soundpath may be used, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the val-ues are still not in the range, shear waves transducers regarding the pipe wall thickness have to be selectedin step 1b.

transducer order code

GLG 0.43 0.91

GLH 0.28 0.59

GLK 0.16 0.35

GLM 0.08 0.2

GLP 0.04 to 0.12

GLQ 0.02 to 0.04

0.2 0.4 0.6 0.8 1pipe wall thickness [in]

transducer order code

GSG 0.55 0.91

GSK 0.2 0.35

GSM 0.1 0.2

GSP 0.06 0.12

0.2 0.4 0.6 0.8 1pipe wall thickness [in]

recommended possible

Page 14: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

14 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Lamb wave transducer1 shear wave transducer1

GLG GSG

GLH

GLK GSK

GLM GSM

GLP

GSP

GLQ

1 inner pipe diameter and max. flow velocity for a typical application with natural gas, nitrogen, oxygen in reflect arrangement with2 sound paths (Lamb wave transducers)/1 sound path (shear wave transducers)

0 20 40 60 80 100 120vfts

10

20

30

40

DmaxinGRG

0 20 40 60 80 100 120vmaxfts

10

20

30

40

Dmaxin

0 20 40 60 80 100 120vmaxfts

10

20

30

40

DmaxinGRH

0 20 40 60 80 100 120vmaxfts

5

10

15

20Dmaxin

0 20 40 60 80 100 120vmaxfts

5

10

15

20Dmaxin

0 20 40 60 80 100 120vmaxfts

1

2

3

4

DmaxinGRM

0 20 40 60 80 100 120vmaxfts

1

2

3

4

DmaxinGDM

0 20 40 60 80 100 120vmaxfts

0.5

1.0

1.5

2.0

2.5Dmaxin

GRP

0 20 40 60 80 100 120vmaxfts

0.5

1.0

1.5

2.0

2.5Dmaxin

GDP

0 20 40 60 80 100 120vmaxfts

0.5

1.0

1.5

2.0

2.5Dmaxin

GRQ

Page 15: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 15

Technical Specification FLUXUS® G70x

Step 3

min. medium pressure

Example

Step 4

for the characters 4 to 11 of the transducer order code (ambient temperature, explosion protection, connectionsystem, extension cable) see page 16

Step 5

for the technical data of the selected transducer see page 17 et seqq.

Lamb wave transducer shear wave transducertransducer order code

medium pressure1 [psi] transducer order code

medium pressure1 [psi]metal pipe plastic pipe metal pipe plastic pipe

min. min. extended min. min. min. extended min.GLG 218 145 15 GSG 435 290 15

GLH 218 145 15 GSK 435 290 15

GLK 218 (d > 4.7 in)145 (d < 4.7 in)

145 (d > 4.7 in)73 (d < 4.7 in)

15 GSM 435 290 15

GLM 145 (d > 2.4 in)73 (d < 2.4 in)

- 15 GSP 435 290 15

GLP 145 (d > 1.4 in)73 (d < 1.4 in)

- 15

GLQ 145 (d > 0.59 in)73 (d < 0.59 in)

- 15

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air

d = inner pipe diameter

step1 pipe wall thickness in 0.47 0.47 0.47 1.2

selected transducer GLG or GLH GLG or GLH GLG or GLH GS2 inner pipe diameter in 31.5 23.6 31.5 11.8

max. flow velocity ft/s 49 49 98 49selected transducer GLG GLG or GLH values not in the range

of the characteristics, but by using direct mode, the inner pipe diameter in the cha-racteristics is doubled:GLG

GSK

3 min. medium pressure psi 247 247 247 508selected transducer GLG GLG or GLH

influence of acoustic noise is reduced with increased transducer frequency, thus recommended:GLH

GLG GSK

Page 16: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

16 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Transducer Order Code

1, 2 3 4 5, 6 7, 8 9 to 11 12, 13 no. of character

tran

sduc

er

tran

sduc

er fr

e-q

uenc

y

- am

bien

t te

mpe

ratu

re

exp

losi

on p

rote

c-tio

n

conn

ect

ion

sys-

tem

- ext

ensi

on c

able

/ opt

ion

description

GL set of ultrasonic flow transducers for gas measurement, Lamb wave

GS set of ultrasonic flow transducers for gas measurement, shear wave

G 0.2 MHz

H 0.3 MHz (Lamb wave only)

K 0.5 MHz

M 1 MHz

P 2 MHz

Q 4 MHz (Lamb wave only)

N normal temperature range

E extended temperature range (shear wave transducers with trans-ducer frequency M, P)

F2 FM Class I Div. 2

NN not explosion proof

TS direct connection or connection via junction box

XXX cable length in m, for max. length of extension cable see page 29

0 m: without junction box

> 0 m: with junction box JB03 or JBP3 (transducers NEMA 6P)

IP68 degree of protection NEMA6P

OS housing with stainless steel 316

example

GL K - N F2 TS - 030 Lamb wave transducer 0.5 MHz, normal temperature range, FM Class I Div. 2, connection system TS with junction box JB03 and extension cable 30 m (98 ft)

- - /

Page 17: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 17

Technical Specification FLUXUS® G70x

Technical Data

Shear Wave Transducers (FM or not explosion proof)

technical type GDG1N52 GDK1N52order code GSG-NF2TS

GSG-NF2TS/OSGSG-NNNTSGSG-NNNTS/OS

GSK-NF2TSGSK-NF2TS/OSGSK-NNNTSGSK-NNNTS/OS

transducer frequency MHz 0.2 0.5medium pressure1

min. extended psi metal pipe: 290 metal pipe: 290min. psi metal pipe: 435

plastic pipe: 15metal pipe: 435plastic pipe: 15

inner pipe diameter d2

min. extended in 9.8 2.8min. recommended in 15 3.1max. recommended in 31.9 19.7max. extended in 43.3 28.3pipe wall thicknessmin. in 0.55 0.2max. in - -materialhousing PEEK with stainless steel

cap 304, option OS: 316LPEEK with stainless steel cap 304, option OS: 316L

contact surface PEEK PEEKdegree of protection NEMA 6 NEMA 6transducer cabletype 1699 1699length ft 16 16dimensionslength l in 5.1 4.98width b in 2.01 2.01height h in 2.64 2.66dimensional drawing

ambient temperaturemin. °F -40 -40max. °F +266 +266temperature compen-sation

x x

explosion protection

FM

order code GSG-NF2TSGSG-NF2TS/OS

GSK-NF2TSGSK-NF2TS/OS

explosion protection temperaturemin. °F -40 -40max. °F +257 +257marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 49 ft/s

l

hb

l

hb

Page 18: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

18 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Shear Wave Transducers (FM or not explosion proof)

technical type GDM2N52 GDP2N52 GDQ2N52order code GSM-NF2TS

GSM-NF2TS/OSGSM-NNNTSGSM-NNNTS/OS

GSP-NF2TSGSP-NF2TS/OSGSP-NNNTSGSP-NNNTS/OS

GSQ-NF2TSGSQ-NF2TS/OSGSQ-NNNTSGSQ-NNNTS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended psi metal pipe: 290 metal pipe: 290 metal pipe: 290min. psi metal pipe: 435

plastic pipe: 15metal pipe: 435plastic pipe: 15

metal pipe: 435plastic pipe: 15

inner pipe diameter d2

min. extended in 1.2 0.59 0.24min. recommended in 1.6 0.79 0.39max. recommended in 3.1 1.6 0.79max. extended in 4.7 2.4 1.2pipe wall thicknessmin. in 0.1 0.06 0.04max. in - - -materialhousing PEEK with stainless steel

cap 304, option OS: 316LPEEK with stainless steel cap 304, option OS: 316L

PEEK with stainless steel cap 304, option OS: 316L

contact surface PEEK PEEK PEEKdegree of protection NEMA 6 NEMA 6 NEMA 6transducer cabletype 1699 1699 1699length ft 13 13 9dimensionslength l in 2.52 2.52 1.57width b in 1.26 1.26 0.87height h in 1.59 1.59 1dimensional drawing

ambient temperaturemin. °F -40 -40 -40max. °F +266 +266 +266temperature compen-sation

x x x

explosion protection

FM

order code GSM-NF2TSGSM-NF2TS/OS

GSP-NF2TSGSP-NF2TS/OS

GSQ-NF2TSGSQ-NF2TS/OS

explosion protection temperaturemin. °F -67 -67 -67max. °F +374 +374 +374marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 49 ft/s

l

hb

l

hb

hb

l

Page 19: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 19

Technical Specification FLUXUS® G70x

Shear Wave Transducers (not explosion proof, NEMA 6P)

technical type GDG1LI8 GDK1LI8 GDM2LI8 GDP2LI8order code GSG-NNNTS/IP68 GSK-NNNTS/IP68 GSM-NNNTS/IP68 GSP-NNNTS/IP68transducer frequency MHz 0.2 0.5 1 2

medium pressure1

min. extended psi metal pipe: 290 metal pipe: 290 metal pipe: 290 metal pipe: 290min. psi metal pipe: 435

plastic pipe: 15metal pipe: 435plastic pipe: 15

metal pipe: 435plastic pipe: 15

metal pipe: 435plastic pipe: 15

inner pipe diameter d2

min. extended in 9.8 2.8 1.2 0.59min. recommended in 15 3.1 1.6 0.79max. recommended in 31.9 19.7 3.1 1.6max. extended in 43.3 28.3 4.7 2.4pipe wall thicknessmin. in 0.55 0.2 0.1 0.06max. in - - - -materialhousing PEEK with stainless steel

cap 316TiPEEK with stainless steel cap 316Ti

PEEK with stainless steel cap 316Ti

PEEK with stainless steel cap 316Ti

contact surface PEEK PEEK PEEK PEEKdegree of protection NEMA 6P NEMA 6P NEMA 6P NEMA 6Ptransducer cabletype 2550 2550 2550 2550length ft 39 39 39 39dimensionslength l in 5.12 5.12 2.76 2.76width b in 2.13 2.13 1.26 1.26height h in 3.29 3.29 1.81 1.81dimensional drawing

ambient temperaturemin. °F -40 -40 -40 -40max. °F +212 +212 +212 +212temperature compen-sation

x x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 49 ft/s

l

hb

l

hb

l

hb

l

hb

Page 20: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

20 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Shear Wave Transducers (extended temperature range, FM or not explosion proof)

technical type GDM2E52 GDP2E52 GDQ2E52order code GSM-EF2TS

GSM-EF2TS/OSGSM-ENNTSGSM-ENNTS/OS

GSP-EF2TSGSP-EF2TS/OSGSP-ENNTSGSP-ENNTS/OS

GSQ-EF2TSGSQ-EF2TS/OSGSQ-ENNTSGSQ-ENNTS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended psi metal pipe: 290 metal pipe: 290 metal pipe: 290min. psi metal pipe: 435

plastic pipe: 15metal pipe: 435plastic pipe: 15

metal pipe: 435plastic pipe: 15

inner pipe diameter d2

min. extended in 1.2 0.59 0.24min. recommended in 1.6 0.79 0.39max. recommended in 3.1 1.6 0.79max. extended in 4.7 2.4 1.2pipe wall thicknessmin. in 0.1 0.06 0.04max. in - - -materialhousing PI with stainless steel

cap 304, option OS: 316LPI with stainless steel cap 304, option OS: 316L

PI with stainless steel cap 304, option OS: 316L

contact surface PI PI PIdegree of protection NEMA 4 NEMA 4 NEMA 4transducer cabletype 6111 6111 6111length ft 13 13 9dimensionslength l in 2.52 2.52 1.57width b in 1.26 1.26 0.87height h in 1.59 1.59 1dimensional drawing

ambient temperaturemin. °F -22 -22 -22max. °F +392 +392 +392temperature compen-sation

x x x

explosion protection

FM

order code GSM-EF2TSGSM-EF2TS/OS

GSP-EF2TSGSP-EF2TS/OS

GSQ-EF2TSGSQ-EF2TS/OS

explosion protection temperaturemin. °F -49 -49 -49max. °F +455 +455 +455marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 49 ft/s

l

hb

l

hb

hb

l

Page 21: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 21

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (FM or not explosion proof)

technical type GRG1N52 GRH1N52 GRK1N52order code GLG-NF2TS

GLG-NF2TS/OSGLG-NNNTSGLG-NNNTS/OS

GLH-NF2TSGLH-NF2TS/OSGLH-NNNTSGLH-NNNTS/OS

GLK-NF2TSGLK-NF2TS/OSGLK-NNNTSGLK-NNNTS/OS

transducer frequency MHz 0.2 0.3 0.5medium pressure1

min. extended psi metal pipe: 145 metal pipe: 145 metal pipe:145 (d > 4.7 in)73 (d < 4.7 in)

min. psi metal pipe: 218plastic pipe: 15

metal pipe: 218plastic pipe: 15

metal pipe:218 (d > 4.7 in)145 (d < 4.7 in)plastic pipe: 15

inner pipe diameter d2

min. extended in 7.5 4.7 2.4min. recommended in 8.7 5.5 3.1max. recommended in 35.4 23.6 11.8max. extended in 63 39.4 19.7pipe wall thicknessmin. in 0.43 0.28 0.16max. in 0.91 0.59 0.35materialhousing PPSU with stainless steel

cap 304, option OS: 316LPPSU with stainless steel cap 304, option OS: 316L

PPSU with stainless steel cap 304, option OS: 316L

contact surface PPSU PPSU PPSUdegree of protection NEMA 6 NEMA 6 NEMA 6transducer cabletype 1699 1699 1699length ft 16 16 16dimensionslength l in 5.06 5.06 5.06width b in 2.01 2.01 2.01height h in 2.66 2.66 2.66dimensional drawing

ambient temperaturemin. °F -40 -40 -40max. °F +338 +338 +338temperature compen-sation

x x x

explosion protection

FM

order code GLG-NF2TSGLG-NF2TS/OS

GLH-NF2TSGLH-NF2TS/OS

GLK-NF2TSGLK-NF2TS/OS

explosion protection temperaturemin. °F -40 -40 -40max. °F +329 +329 +329marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflect arrangement and for a flow velocity of 49 ft/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 82 ft/s

l

hb

l

hb

l

hb

Page 22: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

22 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Lamb Wave Transducers (FM or not explosion proof)

technical type GRM1N52 GRP1N52 GRQ1N52order code GLM-NF2TS

GLM-NF2TS/OSGLM-NNNTSGLM-NNNTS/OS

GLP-NF2TSGLP-NF2TS/OSGLP-NNNTSGLP-NNNTS/OS

GLQ-NF2TSGLQ-NF2TS/OSGLQ-NNNTSGLQ-NNNTS/OS

transducer frequency MHz 1 2 4medium pressure1

min. extended psi - - -min. psi metal pipe:

145 (d > 2.4 in)73 (d < 2.4 in)plastic pipe: 15

metal pipe:145 (d > 1.4 in)73 (d < 1.4 in)plastic pipe: 15

metal pipe:145 (d > 0.59 in)73 (d < 0.59 in)plastic pipe: 15

inner pipe diameter d2

min. extended in 1.2 0.59 0.28min. recommended in 1.6 0.79 0.39max. recommended in 3.5 2 0.87max. extended in 5.9 2.8 1.4pipe wall thicknessmin. in 0.08 0.04 0.02max. in 0.2 0.12 0.04materialhousing PPSU with stainless steel

cap 304, option OS: 316LPPSU with stainless steel cap 304, option OS: 316L

PPSU with stainless steel cap 304, option OS: 316L

contact surface PPSU PPSU PPSUdegree of protection NEMA 4 NEMA 4 NEMA 4transducer cabletype 1699 1699 1699length ft 13 13 9dimensionslength l in 2.91 2.91 1.65width b in 1.26 1.26 0.87height h in 1.59 1.59 1dimensional drawing

ambient temperaturemin. °F -40 -40 -40max. °F +338 +338 +338temperature compen-sation

x x x

explosion protection

FM

order code GLM-NF2TSGLM-NF2TS/OS

GLP-NF2TSGLP-NF2TS/OS

GLQ-NF2TSGLQ-NF2TS/OS

explosion protection temperaturemin. °F -67 -67 -67max. °F +329 +329 +329marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflect arrangement and for a flow velocity of 49 ft/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 82 ft/s

l

hb

l

hb

lh

b

Page 23: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 23

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (not explosion proof, NEMA 6P)

technical type GRG1LI8 GRH1LI8 GRK1LI8order code GLG-NNNTS/IP68 GLH-NNNTS/IP68 GLK-NNNTS/IP68transducer frequency MHz 0.2 0.3 0.5medium pressure1

min. extended psi metal pipe: 145 metal pipe: 145 metal pipe:145 (d > 4.7 in),73 (d < 4.7 in)

min. psi metal pipe: 218plastic pipe: 15

metal pipe: 218plastic pipe: 15

metal pipe:218 (d > 4.7 in),145 (d < 4.7 in)plastic pipe: 15

inner pipe diameter d2

min. extended in 7.5 4.7 2.4min. recommended in 8.7 5.5 3.1max. recommended in 35.4 23.6 11.8max. extended in 63 39.4 19.7pipe wall thicknessmin. in 0.43 0.28 0.16max. in 0.91 0.59 0.35materialhousing PPSU with stainless steel

cap 316TiPPSU with stainless steel cap 316Ti

PPSU with stainless steel cap 316Ti

contact surface PPSU PPSU PPSUdegree of protection NEMA 6P NEMA 6P NEMA 6Ptransducer cabletype 2550 2550 2550length ft 39 39 39dimensionslength l in 5.65 5.65 5.65width b in 2.13 2.13 2.13height h in 3.29 3.29 3.29dimensional drawing

ambient temperaturemin. °F -40 -40 -40max. °F +212 +212 +212temperature compen-sation

x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflect arrangement and for a flow velocity of 49 ft/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 82 ft/s

l

hb

Page 24: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

24 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Transducer Mounting Fixture

Order Code

1, 2 3 4 5 6 7 to 9 10, 11 no. of character

tran

sduc

er

mo

untin

g fi

xtur

e

tran

sduc

er

- me

asur

ing

mod

e

size

- fixat

ion

out

er p

ipe

dia

met

er

/ opt

ion

description

PL PermaLok

VL PermaRail

K transducers with transducer frequency G, H, K

M transducers with transducer frequency M, P

Q transducers with transducer frequency Q

D reflect arrangement or diagonal arrangement/direct mode

R reflect arrangement

S small

M medium

L large

S tension straps

W welding

N without fixation

SK1 0.5 to 2.5 in

SK2 3 to 6 in

SK3 8 to 10 in

SK4 12 to 18 in

SK5 20 to 36 in

SK6 42 to 100 in

IP68 degree of protection NEMA6P

OS housing with stainless steel 316

Z special design

example

VL K - D S - S 200 PermaRail and tension straps for transducers with transducer frequency G, H, K

- - /

Page 25: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 25

Technical Specification FLUXUS® G70x

PermaRail (VLK, VLM, VLQ) material: stainless steel 304, 301, 410, 410option OS: 316, 316L, 17-7PH

inner length:VLK: 13.7 in,option IP68: 14.5 inVLM: 9.2 inVLQ: 6.9 in

dimensions:VLK: 16.65 x 3.54 x 3.66 in,option IP68: 17.44 x 3.7 x 4.13 inVLM: 12.17 x 2.24 x 2.48 inVLQ: 9.72 x 1.69 x 1.85 in

PermaLok PL

Page 26: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

26 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Coupling Materials for Transducers

Technical Data

normal temperature range(4th character of transducer order code = N)

extended temperature range(4th character of transducer order code = E)

< 212 °F < 338 °F < 302 °F < 392 °F< 24 h coupling compound

type N orcoupling foil type VT

coupling compoundtype E orcoupling foil type VT

coupling compoundtype E orcoupling foil type VT

coupling compoundtype E or H orcoupling foil type VT

long time measurement

coupling foiltype VT1

coupling foiltype VT2

coupling foiltype VT1

coupling foiltype VT2

1 < 5 years2 < 6 months

type order code ambient temperature material remark°F

coupling compound type N

990739-1 -22 to +266 mineral grease paste

coupling compound type E

990739-2 -22 to +392 silicone paste

coupling compound type H

990739-3 -22 to +482 fluoropolymer paste

coupling foil type VT 990739-0 14 to +392 fluoroelastomer for transducers with transducer frequency G, H, K

990739-6 for shear wave transducers with transducer frequency M, P

990739-14 for shear wave transducers IP68 and Lambwave transducers with transducer frequency M, P

990739-5 for transducers with transducer frequency Q

Page 27: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 27

Technical Specification FLUXUS® G70x

Damping Mats (optional)Damping mats will be used for the gas measurement to reduce acoustic noise influences on the measure-ment.

transducer damping matTransducer damping mats will be installed below the transducers.

pipe damping matPipe damping mats will be installed if the sound propagation is disturbed at reflection points (e.g. flange,weld). Depending on the noise, the pipe damping mats will be installed at one or both sides of the transducerdamping mat. If the local conditions

Technical Data

type E30R4 E30R3width in 8.9 2thickness in 0.03length (per roll) ft 32weight lb/ft² 2.2ambient temperature °F -22 to +1760properties self-adhesive

outer pipe diameter ≥ 35.4 in outer pipe diameter < 35.4 in

transducer damping matpipe damping mat (optional)

installation length

diagonal arrangement/direct mode

reflect arrangement

Page 28: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

28 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Dimensioning

transducer damping mattransducer

mounting fixtureorder code

type number of layers

transducer damping mat transducer damping mat+

2x pipe damping matmax. instal-lation length number of rolls1

max. instal-lation length number of rolls1

[in] standard2 extended 2 [in] standard extendedPermaRailVLK GLG E30R4 3 35 4 4 72 9 13

GSG 3 4 4 9 11GLH 2 2 3 4 7GLK 1 1 1 1 2GSK 1 1 1 2 3

VLK-**-****/IP68 GLG E30R4 3 36 5 5 75.2 10 14GSG 3 5 5 10 11GLH 2 2 3 5 7GLK 1 1 1 2 2GSK 1 1 2 2 3

VLM GLM E30R3 1 26 1 1 53.5 1 2GSM 1 1 1 1 2GLP 1 1 1 1 1GSP 1 1 1 1 1

VLQ GLQ E30R3 1 21.3 1 1 44.1 1 11 calculation on the base of:

- max. installation length (installation of one transducer mounting fixture per transducer in reflect arrangement) and- max. recommended pipe diameter (standard) or max. extended pipe diameter (extended) (for inner pipe diameter max. recommended and max. extended see Technical Data of the Transducers from page 17)

2 calculation for the number of rolls when both transducers are mounted in one transducer mounting fixture (reflect arrangement) or in diagonal arrangement/direct mode: number of rolls/2 and round up to the nearest integer

Page 29: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 29

Technical Specification FLUXUS® G70x

Connection Systems

Transducer Cable

Technical Data

connection system TSconnection with extension cable direct connection transducers

technical type

****LI*

*****52

transducer frequency(3d character of transducer

order code)

G, H, K M, P Q S

TS

x l x l x l x lcable length ft 16 ≤ 984 13 ≤ 984 9 ≤ 295 6 ≤ 131

cable length (option LC) ft 29 ≤ 984 - - - - - -cable length (option IP68) ft 39 ≤ 984 39 ≤ 984 - - - -

x = transducer cable lengthl = max. length of extension cable

transducer cable extension cabletype 1699 2550 (option IP68) 6111 2615standard length ft see table above -max. length ft - see table aboveambient temperature °F -67 to +392 -40 to +212 -148 to +437 -40 to +158properties longitudinal water

tighthalogen free

fire propagation test according to IEC 60332-1

combustion test according to IEC 60754-2

sheathmaterial stainless steel 304 - stainless steel 304 -

option OS: 316L option OS: 316Louter diameter in 0.31 - 0.31 -cable jacketmaterial PTFE PUR PFA PURouter diameter in 0.11 0.2 ±0.01 0.11 0.47thickness in 0.01 0.04 0.02 0.08color brown gray white blackshield x x x x

(only G704, G704.316SE)

tran

smitt

er

x

l

JBP3

tran

smitt

er

x

tran

smitt

er

x

l

JB03

tran

smitt

er

x

Page 30: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

30 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Junction Box

Technical Data

Dimensions

2 " Pipe Mounting Kit (optional)

technical type JB03 JBP3dimensions see dimensional drawing see dimensional drawingfixation wall mounting, optional: 2 " pipe mounting wall mounting, optional: 2 " pipe mountingmaterialhousing stainless steel 304

option OS: 316Lstainless steel 316L

gasket silicone siliconedegree of protection NEMA 6 NEMA 6ambient temperaturemin. °F -40 -40max. °F +176 +176

in inch

6.85

4.690.08

wall mounting holder

2.766.44

6.14

Ø 0.35

thread: 3x M20 x 1.5cable gland: max. 2x 1/2 NPT

Page 31: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 31

Technical Specification FLUXUS® G70x

Terminal Assignment

JB03

JBP3

&

# &

&

'

'

transducers

terminal connectionXV transducer , SMB connectorXR transducer , SMB connectorcable gland external shield

extension cable

terminal strip KL2terminal connection

TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield

shield terminal

&

&

'

'

&

&

# #

transducers

terminal strip KL1terminal connection

TV transducer , signalTVS transducer , internal shieldTRS transducer , internal shieldTR transducer , signalcable gland external shield

extension cable

terminal strip KL2terminal connection

TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield

shield terminal

Page 32: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

32 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Clamp-on Temperature Probe (optional)

Technical Datatechnical type PT13N PT13N PT13N PT13N PT13F PT13Forder code 670413-1 670412-1 770413-1 770412-1 670413-2 670412-2design with connector without connector with connector

short response timetype Pt1000 Pt1000 matched

according to EN 1434-1

Pt1000 Pt1000 matched according to EN 1434-1

Pt1000 Pt1000 matched according to EN 1434-1

connection 4-wire 4-wire 4-wiremeasuring range °F -22 to +482 -40 to +752

(see also Technical Data of Cable)-58 to +482

accuracy T ±(0.27 °F + 2 . 10-3 . (T [°F] - 32 °F))class A

±(0.27 °F + 2 . 10-3 . (T [°F] - 32 °F))class A

±(0.27 °F + 2 . 10-3 . (T [°F] - 32 °F))class A

accuracy ∆T - ≤ 0.1 K(3K < ∆T < 6 K), more corre-sponding to EN 1434-1

- ≤ 0.03 °F(at 50 °F)3 point ∆T check, more cor-responding to EN 1434-1

- ≤ 0.1 K(3K < ∆T < 6 K), more corre-sponding to EN 1434-1

response time s 50 8housing aluminum 360 brass alloy PEEK, stainless steel 304, copperdegree of protection NEMA 4 NEMA 4 NEMA 4weight (without connector)

lb 0.6 1.1 0.437 0.875 0.7 1.4

fixation clamp-on clamp-on clamp-onaccessories - - plastic protection plate,

insulation foamdimensionslength l in 0.59 0.59 0.55width b in 0.59 0.49 1.18height h in 0.79 0.79 1.06dimensional drawing A C B

extension cableh

l

junction box

C

B

h

l

A

Cextension cable

h

l

Page 33: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21 33

Technical Specification FLUXUS® G70x

Connection

Temperature Probe

Connector

Cable data

pin cable of temperature probe extension cable1 white/blue blue2 red/blue gray3, 4, 5 not connected6 red red7 white white8 not connected

with connector without connectorcable of temperature

probeextension cable cable of temperature

probeextension cable

type 4 x 0.25 mm² black or white

LIYCY 8 x 0.14 mm² gray 4 x 24 AWG 4 x 18 AWG

standard length ft 9 16/32/82 23 -max. length ft - 656 - 656cable jacket PTFE PVC fiberglass, PTFE LS PVCambient temperature °F max. +752 (fiberglass)

max. +266 (transition, PTFE)

Junction Box

technical type JBT3dimensions see dimensional drawingfixation wall mounting

optional: 2 " pipe mountingmaterialhousing stainless steel 304 gasket siliconedegree of protection NEMA 6cable gland max. 2x 1/2 NPTambient temperaturemin. °F -40max. °F +176

without connector red red white whitewith connector red/blue red white/blue white

fiberglass PTFEtransition

Page 34: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

34 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

Terminal Assignment

Wetted Temperature Probe (optional)

JBT3

# #

temperature probe (with connector)

terminal strip KL1terminal connection1 red2 red/blue3 white4 white/blue

extension cable (with connector)

terminal strip KL2terminal connection1 red2 gray3 white4 blue

temperature probe (without connector)

terminal strip KL1terminal connection1 red2 red3 white4 white

extension cable (without connector)

terminal strip KL2terminal connection1 white2 black3 green4 red

shield terminal

type Pt1000A insertion length 6 " or specified lengthB resistance 1 000 Ω, 00385C insertion length 6 " or specified length

sheath material stainless steel 316D thread 1/2 " NPT HEX CPLG. spring loadedE head aluminum screw cover head

4 terminal blockF thread 3/4 " NPT

Page 35: Gas Ultrasonic Flowmeter for Permanent Installationquotation.flexim.com/sales-docs/Tech-Specifications/TS... · 2014-11-13 · Gas Ultrasonic Flowmeter for Permanent Installation

36 TSFLUXUS_G7V1-5-4US_Lus, 2013-06-21

FLUXUS® G70x Technical Specification

FLEXIM AMERICAS Corporation

Edgewood, NY 11717

USA

Tel.: (631) 492-2300

Fax: (631) 492-2117

internet: www.flexim.com

e-mail: [email protected]

1-888-852-7473

Subject to change without notification. Errors excepted.

FLUXUS® is a registered trademark of FLEXIM GmbH.