Agilent Technologies Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory For the Agilent Intuvo 9000 Gas Chromatograph Installation Instructions Parts Supplied Table 1 Parts supplied with an Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory G3969A Description Quantity Part number Transfer line interface accessory 1 G4589-68500 Transfer line interface assembly G4599-60500 Headspace transfer line GC cover G4599-60340 Headspace transfer line inlet cover G4599-20461 9000 HS TL Interface Accessory ship kit: 1 G3969-68000 Capillary tubing cutter 4/pk 1 5181-8836 XLSI transfer line nut for CFT fitting 2 G3520-20210 Column storage fitting 2 G2855-20590 Magnifier, 3x, 6x, paddle, plastic 1 G2855-40001 Plug for microfluidic manifold or unions 2 G2855-60570 Ferrule pre-swage tool, capillary flow 1 G2855-60200 O-ring, 2-010, fluoroelastomer 5/pk 1 5188-6405 11 mm LOWBLD septa 5/pk 1 5182-3413 Ferrule flexi inert 0.53 mm col 5/pk NFS 1 G3188-27553 Flexible metal ferrule swaging guide 1 G3188-90001 UltiMetal flexible metal ferrule guide 1 G3188-90010 Liner, ultra inert, straight 2 mm ID 1 5190-6168

G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Embed Size (px)

Citation preview

Page 1: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

For the Agilent Intuvo 9000 Gas Chromatograph

Installation Instructions

Parts Supplied

Table 1 Parts supplied with an Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory G3969A

Description Quantity Part number

Transfer line interface accessory 1 G4589-68500

• Transfer line interface assembly G4599-60500

• Headspace transfer line GC cover G4599-60340

• Headspace transfer line inlet cover G4599-20461

• 9000 HS TL Interface Accessory ship kit: 1 G3969-68000

• Capillary tubing cutter 4/pk 1 5181-8836

• XLSI transfer line nut for CFT fitting 2 G3520-20210

• Column storage fitting 2 G2855-20590

• Magnifier, 3x, 6x, paddle, plastic 1 G2855-40001

• Plug for microfluidic manifold or unions 2 G2855-60570

• Ferrule pre-swage tool, capillary flow 1 G2855-60200

• O-ring, 2-010, fluoroelastomer 5/pk 1 5188-6405

• 11 mm LOWBLD septa 5/pk 1 5182-3413

• Ferrule flexi inert 0.53 mm col 5/pk NFS 1 G3188-27553

• Flexible metal ferrule swaging guide 1 G3188-90001

• UltiMetal flexible metal ferrule guide 1 G3188-90010

• Liner, ultra inert, straight 2 mm ID 1 5190-6168

Agilent Technologies

Page 2: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Parts Identification

2

Figure 1 Parts identification - Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory (transfer line interface accessory)

Figure 2 Parts identification - Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory (GC cover and inlet cover)

EPC module weldment

Turn top insert assemblyCarrier gas line

Heating block

Spacer block

GC cover

Inlet cover

Installat

ion Instructions
Page 3: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Installation Instructio

Figure 3 Parts identification - Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory (ship kit)

Ferrules

Capillary tubing cutter

Column storage fitting

XLSI transfer line nut for CFT fitting

Ferrule pre-swage tool

Ferrules

O-rings

Plug

Magnifier

Septa

Inlet liners

ns

3
Page 4: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Tools Required

4

• T-20 Torx driver

• T-10 Torx driver

• ESD wrist strap

Compatibility

The G3969A transfer line interface accessory is used to connect an Agilent 7697A Headspace Sampler (HS) to an Agilent Intuvo 9000 GC through an installed split/splitless (S/SL) or multimode inlet (MMI).

The G3969A transfer line interface accessory cannot be used with a G4514A sample tray on the Agilent Intuvo 9000 GC.

For information about changing the inlet liner or changing the Intuvo guard chip, see “Changing the Inlet Liner and Intuvo Guard Chip" on page 17.

Installation Instructions

Page 5: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Installation Procedure

Installation Instructio

This procedure explains how to install the transfer line interface accessory on the Agilent 9000 Gas Chromatograph (GC).

Read the Agilent Intuvo 9000 Gas Chromatograph Safety Manual before using the GC. The Safety Manual describes precautions to follow when installing and using the GC, as well as important regulatory information.

Before proceeding, turn off the column heater and other heated zones, and

WARNINGlet them cool down. When they have cooled, turn off any inlet gases at their supply, then power down the GC as directed, and unplug the power cord.

WARNING If an MS is installed on the GC, it must be cooled and vented prior to installation of the inlet. Refer to the Operation Manual for the MS for cooldown and venting procedures.

WARNING If an MMI inlet is installed on the GC, high pressure or extremely cold gas may be present. Before replacing a cryo valve or connected inlet, run the GC maintenance method, and turn off the cryo line at the source.

Always allow time for the column heater and other heated zones to cool

CAUTIONbefore turning off the gas supplies. Failure to allow cooling may damage the flow path.

CAUTION Ensure you are properly grounded with an ESD wrist strap before continuing.

ns 5

Page 6: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Remove GC top cover and inlet components

6

1 If an MS is installed, cooldown and vent the MS. Refer to the Agilent 5977B Series MSD Operating Manual for details.

2 On the touch screen, navigate to Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Start Maintenance.

3 When the GC becomes ready, turn it off as follows:

a On the touch screen, navigate to Settings > Power. The Power Options dialog box appears.

b Press Shut Down. The GC shuts down.

c Switch off the power on/off switch on the rear panel.

4 Remove the GC power cord from the wall outlet.

5 Turn off all connected gases at the source.

6 If present, remove the injector from the GC.

7 Put on an ESD wrist strap, and attach the ground to the GC sheet metal frame for electrostatic protection.

8 Using a T20 Torx driver, completely loosen the five captive screws securing the top cover to the GC, and lift the cover off of the GC. See Figure 4.

Figure 4 GC top cover removal

GC cover retaining screws

Detector cover (typical)

Inlet cover

Installation Instructions

Page 7: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Installation Instructio

9 Remove the detector cover components from the GC top cover for use on the top cover supplied with the accessory. Typically, this requires removal of the flat portion of the cover by using a T-20 Torx driver to remove four screws.

10 Store the GC top cover, including the inlet cover, in a safe, dry location for future use.

11 Use a T-10 Torx driver to completely loosen the two captive screws securing the EPC bracket to the GC, and remove the bracket. See Figure 5.

12 Use a T-10 Torx driver to completely loosen the captive screw securing the EPC module weldment to the EPC module. See Figure 6.

13 Remove the weldment from the EPC module.

Figure 5 EPC bracket and retaining screws

Figure 6 EPC module weldment and screw

EPC bracket retaining screws

Inlet EPC module

EPC module weldment

EPC module weldment screw

ns

7
Page 8: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

8

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

14 Free the turn top insert assembly from the inlet weldment by the using turn top handle to rotate the insert counter-clockwise until the insert is loosened from the inlet weldment. See Figure 7.

15 Lift the turn top insert assembly out of the inlet weldment.

16 Remove the turn top insert assembly, septum purge tubing assembly and EPC module weldment out of the GC. See Figure 8.

Figure 7 Inlet weldment and turn top insert assembly

Figure 8 Removing the turn top insert assembly, septum purge tubing assembly and EPC module weldment

Inlet weldment

Turn top handle

Turn top insert assembly

EPC module weldment

Septum purge tubing

Turn top insert assembly

Installatio

n Instructions
Page 9: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Install the accessory on the GC

Installation Instructio

Avoid sharp bends when handling the septum purge tubing assembly.

CAUTIONFailure to do so may result in damage to the hardware.

1 The accessory is shipped configured for installation on an S/SL inlet. If the accessory is being installed on an MMI inlet, do the following:

a Use a T-20 Torx driver to remove the three screws securing the two spacer blocks to the accessory. See Figure 9.

b Flip the spacer blocks over, and reinstall them by securing with the three screws. See Figure 10.

Figure 9 Transfer line interface accessory spacer blocks

Figure 10 Transfer line interface accessory spacer blocks (oriented for MMI installation)

Spacer blocks

Screws

Spacer blocks

ns

9
Page 10: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

10

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

2 Carefully unfurl the septum purge tubing to place the transfer line interface accessory into the GC, leaving the looped portion of the tubing coiled. Avoid making any sharp bends. See Figure 11.

Figure 11 Positioning the transfer line interface accessory in the GC

EPC module weldment

Turn top insert assembly

Looped tubing

Carrier gas line

Installatio

n Instructions
Page 11: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Installation Instructio

3 Place the turn top insert portion of the transfer line interface accessory on top of the inlet, taking care to keep it level. When properly positioned, the arm on the transfer line connector is roughly parallel to the front of the GC. A positioning pin on the GC inserts into the accessory heating block to align the accessory. See Figure 12 and Figure 13.

Figure 12 Turn top inlet and transfer line heater block installation and alignment

Figure 13 Turn top inlet and transfer line heater block installation and alignment

Transfer line connector arm

Turn top insert assembly

Turn top handle

Accessory heating block

Inlet interface PCA

Accessory heater cable connector

Positioning pin

Uninstalled Installed

ns

11
Page 12: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

12

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

4 Rotate the yellow turn top handle on the turn top insert assembly to lock it in place.

5 Insert the accessory heater cable into the connector on the inlet interface PCA. The connector is keyed to prevent improper insertion.

6 Use a T-10 Torx driver to tighten the captive screw securing the EPC module weldment to EPC module. See Figure 6 on page 7.

7 Route the carrier gas line out of the rear of the GC. See Figure 11 on page 10.

8 Ensure that the vent tubing within the GC is properly oriented. The looped portion of tubing should reside near the split vent trap.

9 Install the EPC bracket to the GC and use a T-10 Torx driver to secure it with two captive screws. See Figure 5 on page 7.

10 Place a Swagelok fitting on the end of the carrier gas line, and connect it to the Carrier fitting on the back of the Agilent 7697A Headspace Sampler. See Figure 14.

Figure 14 Carrier gas connection at 7967A Headspace Sampler

Carrier gas tubing

In

stallation Instructions
Page 13: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Prepare the transfer line

Installation Instructio

Materials are provided in the ship kit to make connections to 0.53-mm id columns and tubing. If you are using a different transfer line, use the appropriate ferrules (see Table 2). Ensure the flexible metal ferrule dimension matches the transfer line. Cut fused silica, and install the flexible metal ferrules and the transfer line.

Before performing this procedure, ensure the transfer line is installed in the 6-port valve. Refer to the procedure “To Install a Fused Silica Column into the Transfer Line” in the Agilent 7697A Headspace Sampler Maintenance Manual.

To prepare the GC interface end of the transfer line:

1 Pass the transfer line fused silica tubing through the XLSI transfer line nut, flexible metal ferrule, and ferrule pre-swage tool. See Figure 15.

2 With the XLSI transfer line nut tight against the transfer line, hold the nut with a 1/4-inch wrench, and thread the ferrule pre-swage tool onto the nut.

Table 2 Flexible metal ferrules

Description Part number

.25 mm flexible metal ferrules, inert G3188-27501

.32 mm flexible metal ferrules, inert G3188-27502

.53 mm flexible metal ferrules, inert G3188-27503

Figure 15 Threading the XLSI transfer line nut, ferrule, and ferrule pre-swage tool

XLSI transfer line nut

Flexible metal ferrule

Transfer line fused silica tubing

Ferrule pre-swage tool

ns

13
Page 14: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

14

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

The flexible metal ferrules are delicate. They must not be overtightened.

CAUTIONFailure to carefully follow the instructions in the following steps may cause the fused silica tubing to break.

3 Using the 1/4-inch wrench and ferrule pre-swage tool, tighten the nut a little at a time, occasionally checking to see if the ferrule is gripping the tube.

4 When the ferrule just starts to grip, notice the position of the nut and then tighten by turning 45 to 60 degrees of rotation, but no more than 60 degrees (one flat). If you can pull the column free, it is not tight enough.

5 Remove the ferrule pre-swage tool.

6 Using a ceramic column cutter, trim the tubing at the small end of the ferrule, leaving approximately 0.3 mm of tubing extending beyond the ferrule.

The column tubing end cannot extend beyond 0.5 mm from the end of the

NOTEferrule.

It is important that the ceramic column cutter has one side dedicated to only making contact with the column and the other side dedicated to riding on the edge of the flexible metal ferrule. This preserves the ceramic edge sharpness, which is necessary for making good column cuts. See Figure 16.

7 Check the end of the tube with a magnifier. The end of the tube need not be perfectly square, but should not have cracks extending under the ferrule.

Figure 16 Ceramic column cutter and ferrule

Installation Instructions

Page 15: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Attach the transfer line to the accessory

Installation Instructio

1 Install the accessory GC top cover, without the accessory inlet cover installed, and secure it by using a T-20 Torx driver to tighten the five captive screws. See Figure 4 on page 6.

2 Unlatch and open the clamp on the transfer line interface accessory. See Figure 17.

3 Carefully place the transfer line end-cap into the clamp on the transfer line interface accessory. The end-cap should lie squarely against the bottom and back surface.

4 Partially latch the clamp, closing it enough to loosely secure the transfer line and end-cap. This provides strain relief, and aligns the end-cap to the heated block portion of the transfer line interface accessory.

5 Carefully thread the XLSI transfer line nut into the transfer line interface accessory fitting while keeping the transfer line end-cap aligned with the fitting.

6 Finger tighten the XLSI transfer line nut on the end of the transfer line, screwing it into the transfer line interface accessory fitting.

7 Completely latch the clamp.

8 Use a 1/4-inch wrench to tighten the XLSI transfer line nut by 1/2 a flat (30 degrees).

Figure 17 Attaching the transfer line to the GC

XLSI transfer line nut

Clamp

Transfer line end-cap

ns

15
Page 16: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

16

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

9 Install the inlet cover on the GC. See Figure 2 on page 2.

10 If present, install the injector on the GC.

11 Restore gas pressure, and leak check all fittings. Refer to the Agilent 7697A Headspace Sampler Installation and First Startup Manual for more information.

12 Connect the GC power cord to the power outlet.

13 Switch on the GC. The GC will power up in maintenance mode.

14 Exit maintenance mode. Maintenance > Instrument > Perform Maintenance > Maintenance Mode > End Maintenance.

15 Configure inlet setpoints as required. Methods Edit Active > Inlets.

16 Configure transfer line interface setpoints as required. Methods > Edit Active> Thermal Aux 1 settings are used to control the setpoints for the transfer line interface.

17 Purge the entire system prior to turning on any heated zones.

Installation Instructions

Page 17: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Changing the Inlet Liner and Intuvo Guard Chip

Installation Instructio

When the G3969A transfer line interface accessory is installed on the GC, inlet maintenance and guard chip maintenance can be performed by lifting the turn top inlet assembly (still connected to the transfer line accessory) and gently moving it aside to gain access to the inlet liner and guard chip access port.

Before proceeding, turn off the column heater and other heated zones, and

Disconnecting the accessory

WARNINGlet them cool down. When they have cooled, turn off any inlet gases at their supply, then power down the GC as directed, and unplug the power cord.

1 If an MS is installed, cooldown and vent the MS. Refer to the Agilent 5977B Series MSD Operating Manual for details.

2 On the touch screen, navigate to Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Start Maintenance.

3 Wait until the GC becomes ready.

4 If present, remove the injector from the GC.

5 Put on an ESD wrist strap, and attach the ground to the GC sheet metal frame for electrostatic protection.

6 Remove the inlet cover from the GC. See Figure 2 on page 2.

ns 17

Page 18: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

18

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

7 Free the turn top insert assembly from the inlet weldment by using the turn top handle to rotate the insert counter-clockwise until the accessory is loosened from the inlet weldment. See Figure 18.

Figure 18 Turn top inlet assembly and inlet weldment

Inlet weldment

Turn top insert assembly

Turn top handle

Guard chip access port

Septum purge tubing

When moving the accessory, avoid sharp bends in the septum purge tubing.

CAUTIONFailure to do so may result in damage to the hardware.

8 Carefully lift the turn top insert portion of the transfer line interface accessory from the top of the inlet, and move it toward the rear of the GC until the inlet weldment and guard chip access port are visible.

Installation Instructions

Page 19: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Reconnecting the accessory

Installation Instructio

1 Place the turn top insert portion of the transfer line interface accessory on top of the inlet, taking care to keep it level. When properly positioned, the arm on the transfer line connector is roughly parallel to the front of the GC. A positioning pin on the GC inserts into the accessory heating block to align the accessory. See Figure 18 on page 18.

2 Lock the turn top insert assembly on the inlet weldment by using the turn top handle to rotate the insert clockwise until the accessory is firmly attached to the inlet weldment.

3 Install the inlet cover on the GC. See Figure 2 on page 2.

4 If present, install the injector on the GC.

5 Exit maintenance mode by pressing Finished.

ns 19

Page 20: G3969A Transfer Line Interface for an Agilent 7697A HS ... · PDF fileAgilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory ... WARNING Before proceeding, turn off

Agilent Technologies

Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warran-ties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

© Agilent Technologies, Inc. 2017Printed in USA and China

Third edition, July 2017Agilent Technologies, Inc.

2850 Centerville RoadWilmington, DE 19808-1610 USA

G3969-90010

G3969-90010