Upload
lykhue
View
218
Download
3
Embed Size (px)
Citation preview
Detection efficiency
Marie-Christine LépyLaboratoire National Henri Becquerel
CEA Saclay, F-91191 Gif-sur-Yvette cedex, FranceE-mail: [email protected]
IAEA - ALMERA Technical Visit – July 2010
Detector efficiency
• Full energy peak efficiency
– Definition
– Experimental calibration
– Curve fitting
• Total efficiency
– Definition
– Experimental calibration
• Monte Carlo simulation
FULL ENERGY PEAK EFFICIENCY
Ratio of the number of counts in full-energy peak corresponding
to energy E (NP(E)), by the number of photons with energy E
emited by the source (F(E))
εP(E) depends on the source-detector geometry
and on the energy
Full-energy Peak Efficiency (FEPE): εP(E)
)(
)()(
EF
ENE
P
P =ε
NP(E)
Energy
Counts per channel
εp(E) depends on the geometrical conditions and on the energy
Measurement
conditionsDetector
characteristics
εP(E) = εG. εI(E)
Geometrical
efficiency
Intrinsic
efficiency
Sample
Distance
Shielding
Full-energy Peak Efficiency (FEPE): εP(E)
Geometrical efficiency
πε
4G
Ω=
d
r
Point source
Detector
Ω
Ω = solid angle between source and detector (sr)
For a point source :
)1(222 rd
d
+−=Ω π
εG depends only on the source-detector geometry
Ratio of the number of photons emitted towards the detector
by the number of photons emited by the source
εI(E) depends on the energy of the
incident photons:
transmission
absorption
full-energy deposition
εI(E) = Ratio of the number of counts in
full-energy peak by the number of impiging photons
Difficulty: exact composition badly known
Intrinsic efficiency
Ge dead layer
Be window
Capvacuum
vacuum
HPGe
Electrical contact(Au, Ni, ... thickness?)
Incoming photons
Interacting photons
Transmission probability through material i
with thickness xi : (Beer-Lambert law) :
Thus
Interaction probability in the same material:
PT (E, xi) = exp (-µi(E) . xi)
PI (E, xi) = 1 - PT(E, xi)
Calculation of the detector FEP efficiency
To result in a count in the FEP peak:
The photon must :
- be emitted in the Ω solid angle,
- cross the screens (air,n window, dead layer,…) without being absorbed,
- and be totally absorbed in the detector active volume.
GeHP
Ω
Source
Transmission
through screens
Interaction in the
detectorr volume
( )( ) ( ) Ω⋅⋅⋅−−⋅⋅−= ∫∏Ω
dEPxExEER Pdd
i
ii
P )()(exp(1)(exp)( µµ
For the low-energy range:
Probability of total absorption in the detector
For higher energies:
Total absorption is due to succesive effects : multiple scattering
Thus the calcul is not possible
Pp(E) ≈ τd(E)/µd(E)
Calculation of the detector efficiency
Many difficulties for an accurate calculation
Radiography of a HPGe detector:
Rounded crystal, axially shifted, tilted …
- Exact knowledge of the detector parameters:
materials (composition)
geometry (thickness, position …)
- Data used in the calculation :
Attenuation coefficients
Material density
- Semi-conductor effects: parameters and physical
interaction :
Electrical field
Electrodes
Calculation of the detector efficiency
Experimental FEP efficiency calibration
This is performed using standard radionuclides with standardized activity A (Bq) with
photon emission intensities, Iγ well known
NP(E) : peak net area
Energy (keV)
Counts
per
channel
100 200 300 400 500 600 700
1
2
10
100
1000
10000
20000
50000
NP(E)Spectrum 137 Cs
εP(E) Full-energy peak (FEP) efficiency depends on the energy and
on the source-detector geometrical arrrangement
)E(F
)E(N)E(
P
P =ε)E(IA
)E(N)E(
P
P
γ
ε⋅
=
Associated standard uncertainty
)E(IA
)E(N)E(
γ
ε⋅
=Efficiency calibration
2222
+
+
=
)E(I
)E(I
A
A
N
)E(N)E( ∆∆∆
ε
ε∆
33 101)(
)(105
)(
1
)(
)(
)(
)( −− ⋅=∆
⋅=∆
==∆
EI
EI
A
A
ENEN
EN
EN
EN
γ
γ
Influence of the peak area :
if N = 104 ∆N/N = 10-2 -> ∆ε/ε = 1,1 10-2
if N = 105 ∆N/N = 3,1 10-3 -> ∆ε/ε = 6 10-3
FEP efficiency calibrationTo get an efficiency values at any energy : energy calibration over the whole
energy range
1. Use different radionuclides to get energies regularly spaced over the range of interest
Single gamma-ray emitters : 51Cr (320 keV), 137Cs (662 keV) 54Mn (834 keV) : one efficiency value per one measurement
Multigamma emitters : 60Co, 133Ba, 152Eu, 56Co : several efficiencies values per one
measurement , but coincidence summing effects !
For each energy, discrete values of the FEP efficiency ε(E1), ε(E2), … ε (En)
2. Computation of the efficiency for
1. Local interpolation
2. Fitting a mathematical function to the experimental values
!
Efficiency calibration for different source-to-detector distances
0.01
0.10
1.00
10.00
10 100 1 000 10 000Energy(keV)
D = 10 cm
D = 15 cm
D = 5 cm
D = 4 cm
D = 3 cm
D = 2 cm
D = 1 cm
D =7 cm
FEP efficiency calibration
Efficiency calibration - Interpolation
Local interpolation
E1 -> ε1
E2 -> ε2
To get ε(E) for E1 < E < E2
Solution 1 : linear interpolation )()EE(
)EE()E( 12
12
11 εεεε −⋅
−
−+=
Solution 2 : logarithmic interpolation
: valid only for close energies!
)ln(ln)ElnE(ln
)ElnE(ln)ln())E(ln( 12
12
11 εεεε −⋅
−
−+=
Efficiency calibration - InterpolationLocal interpolation : example
Actual value (at 662 keV) : ε(E) = 1.90 10-3
Conclusion : approximation to be used only if high uncertainties (> 10 %) are acceptable
Determine efficiency for E = 662 keV (137Cs) knowing :
1. E1 = 569 keV (134Cs) ε1 = 2.21 10-3
E2 = 766 keV (95Nb) ε2 = 1.67 10-3
Linear interpolation : ε(E) = 1.96 10-3
Logarithmic interpolation : ε(E) = 1.92 10-3
2. E1 = 122 keV (57Co) ε1 = 8.22 10-3
E2 = 1 173 keV (60Co) ε2 = 1.13 10-3
LInear interpolation : ε(E) = 4.57 10-3
Logarithmic interpolation : ε (E) = 1.87 10-3
Determination of the best fitted function to a given set of experimental data (energy, efficiency)
In the logarithmic scale , the shape is smoother than in the linear scale.
Efficiency calibration : mathematical fitting (1)
Experimental efficiency calibration of a
HPGe detector (100 cm3) for a point source
at 12 cm between 10 and 1500 keV (linear scale)
Energy (keV)
0 200 400 600 800 1000 1200 1400 1600 1800
Effic
iency (
%)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
Experimental efficiency calibration of a
HPGe detector (100 cm3) for a point source
at 12 cm between 10 and 1500 keV (logarithmic scale)
Energy (keV)
10 100 1000
Effic
ien
cy (
%)
0.2
0.3
0.4
0.5
0.6
0.70.80.9
0.1
1
Functions frequently used:
Polynomial fitting in the log-log scale: i
n
0ii )E(lna)E(ln ∑
=
⋅=ε
in
0ii Ea)E(ln −
=∑ ⋅=ε
Efficiency calibration : mathematical fitting
Remarks :
-ai coefficients are determined using a least-squares fitting method
- experimental data must be weighted
- the polynomial degree (n) must be adjusted depending on the number of experimental data (p) : n << p
- in some case two different functions can be used with a cross point
- check the resulting fitted curves !
Efficiency calibration : mathematical fitting
Example : 40 experimental values
in the 122-to-1836 keV range
Fitting function :
Adjusted coefficients :
in
ii EaE )(ln)(ln
0∑=
⋅=ε
fitting 122 to 1836 keV
deg0 -34,11961
deg1 48,16797
deg2 -25,89215
deg3 5,80219
deg4 -0,40503
deg5 -0,01632
Efficiency fitting must be
visually checked
Efficiency calibration : mathematical fitting
In the case of cross
points be carefull :- Avoidzones with
important inflexion - Avoid highdegreepolynomials
Spline functions
Simply a curve
Special function defined piecewise by polynomials
A piecewise polynomial f(x) is obtained by dividing of X into contiguous intervals, and representing f(x) by a separate polynomial in each interval
The polynomials are joined together at the interval endpoints (knots) in such a way that a certain degree of smoothness of the resulting function is guaranteed
Slides from Dr. Oliver Ott - PTB
Piecewise defined polynomials
constraints
K1 ≤ xmin < K2 < … < Kk < xmax < Kk+1
Series of knots
xmin: minimum x-value of the data points
xmax: maximum x-value of the data points
Ki ( i = 1, …, k+1)
• 2 consecutive knots establish an interval [Ki, Ki+1) where a polynomial Pi(x) of order n (degree n-1) is defined.
• In total k intervals are fixed by the knots.
• Outside the interval i the polynomial Pi(x) is not defined.
• At the joining points K2 …Kk they have to fit smoothly – continous up to the (n-2) derivative
Spline functions mathematically
Condition that a spline function solution must satisfy:
For the derivatons of neighboured polynomials at all knots:
( ) ( ) 2-n m 0 ,01
)(
11
)(≤≤=− +++ i
m
ii
m
i KPKP
k polynomials in k intervals: ( )∑=
=k
i
i xPxF1
)(
where
( ) =xP( )∑
=+
−≤≤−⋅
n
j
i
j
iij KxKxa1
1i
1Kfor
0 else
Efficiency calibration at 25 cm
Efficiency calibration at 10 cm
Multigamma emitters -> Coincidence summing effets
• Efficiency calibration for reference geometry
– For point source : relative uncertainty 1-2 %
• Corrective factors needed if differentmeasurement geometry
FEP Efficiency calibration : remarks
TOTAL EFFICIENCY
Ratio of the total number of counts in the spectrum (NT(E)), by the
number of photons with energy E emitted by the source (F(E))
εT(E) depends on the source-detector geometry
and on the energy
Total Efficiency (TE): εT(E)
)(
)()(
EF
ENE T
T =ε
NT(E)
Energy
Counts per channel
( )( ) ( ) Ω⋅⋅−−⋅⋅−= ∫∏Ω
dxExEET dd
i
ii )(exp(1)(exp)( µµε
xi : i shield thickness
xd : detector thickness
Total efficiency depends on:
- Solid angle Ω
- Attenuations in absorbing layers (air, window, dead layer, etc)
- Interaction in the detector active volume (any effect)
And on the environment ! (scattering)
where
Attenuation in screens Detector interaction
Calculation of the total efficiency
Experimental total efficiency calibration εT(E)
Energy (keV)
Co
un
tsp
er
ch
an
nel
100 200 300 400 500 600 700
1
2
3
57
10
20
30
5070
100
200
300
500700
1000
2000
3000
50007000
10000
20000
30000
50000
NT (E)
Spectrum 137Cs
Total spectrum minus background
background
)(.
)()(
EIA
ENTET
γ
ε =
1. Using mono-energetic radionuclides
241Am (60 keV), 109Cd (88 keV), 139Ce (166 keV-T1/2=138 d), 51Cr (320 keV- T1/2= 28d),
85Sr (514 keV- T1/2=65 j), 137Cs (662 keV), 54Mn (834 keV - T1/2=302 d)
2. For the low-energy range, if P-type detector, 57Co can be used (mean energy and sum of emission intensities)
3. For the high energy range, 88Y and 60Co can be used... But complex decay scheme
Experimental total efficiency calibration εT(E)
Approximation of the total efficiency with 2-photons emitters:
NT = A (Iγ1. εT 1 + Iγ2
. εT 2 - Ιγ1. εT 1
. P12. Iγ2
. εT 2)
88Y : E1 =898 keV and E2 = 1836 keV (or 65Zn : 511 and 1115keV)
E1 = 898 keV → εT1 can be extrapolated from previous data
(54Mn – 834 keV)
P12 = 1
( )112
11
21 T
T
TIIA
IANT
ε
εγε
γγ ⋅−⋅⋅
⋅⋅−=
Sr88
γ1
2γ
1β
0
2 734
Y88
1 836γ3
EC3
EC4+,
( ) A
N
A
AN T
T
TTT −−=
−⋅
⋅−= 11
1 ε
εε
Approximation of the total efficiency with 2-photons emitters:
NT = A (Iγ1. εT1 + Iγ2
. εT2 - Ιγ1. εT1
. P12. Iγ2
. εT2)
Co
60
γ1
2γ
1β
2β
2 505
1 332
0
60
Pβ2: 0,001
Pβ1: 0,999
60Co :
E1 = 1 173 keV Iγ1 = 0,99
E2 = 1 332 keV Iγ2 = 0,98
P12 = 1 Iγ1 ≈ Iγ2 = 1
εT1 ≈ εT2 = εT
• Efficiency calibration for reference geometry
– Difficult, even with point sources:
• relative uncertainty 5-10 %
• But limited influence on the coincidence correction factor
• Monte Carlo simulation ?
Total Efficiency calibration : remarks
MONTE CARLO SIMULATION
Monte Carlo simulation
• Difficulties: bad knowledge of the detector internal
parameters
• Accurate description is time-consuming:
– Radiography (external dimensions)
– Collimated beam (hole - dead layer)
– Window to crystal distance (source at different distance)
– Comparison with some experimental values
• Different energies
• Different geometries
X-ray tube ->
Geometrical dimensions
Cristal shape (rounding)
60Co source ->
Hole dimensions
Comparison with experimental data
10 100 1000Energy (keV)
0.001
0.01S
r-85
Co-5
7 Am
-241
Cs-1
37
Ce-1
39
Am
-241
Eu-1
55
Cd-1
09
Eu-1
55
Eu-1
52
Co-5
7
Co-5
7
Ce-1
39
Cr-
51
Eu-1
52
Sn-1
13
Sr-
85
Cs-1
37
Nb-9
4E
u-1
52
Mn-5
4
Nb-9
4Y
-88
Zn-6
5C
o-6
0E
u-1
54
Na-2
2
Co-6
0E
u-1
52
Eu-1
54
Y-8
8
-1
0
1
Z-s
core
Experimental
Calculated
Extended-range coaxial HPGe with carbon-epoxy window (61 x 61 mm)Point sources at 16 cm from the detector window
Picture from V. Peyres - CIEMAT
• Can provide
– FEP efficiency
– Total efficiency (very dependent on the
environment)
– Absolute calculation should be compared with
accurate experimental data
– Strong interest for efficiency transfer calculation
Monte Carlo simulation