FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware 1. General p.1 2. Installation p.2 3. Connection p.3 4. Operation p.4 5. Maintenance p.5 6. Attention p.5 7. Key-point of measurement p. 6 Chapter Software (FsmⅡ & FsmV) 1. General p.1 2. Operating Environment and Connection Method p.1 3. Display Configuration and Button Function p.2 3.1 Display Configuration p.2 3.2 Button Function p.3 3.3 Mouse Operation (Common) p.4 4. Measurement Procedure p.5 5. Operating Instruction (Details) p.6 5.1 File Input/Output p.6 5.2 History p.6 5.3 Adjusting Measurement Area p.7 5.4 Setting Measurement Condition p.8 5.5 System Setup p.10 5.6 Calibration p.13 5.7 Measurement p.14 6. Specification 7. Guaranty 8. Certificate of inspection FSM-6000LE Manual p.Contents

FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE ManualVer.2015.7

Table of Contents

Chapter Hardware

1. General p.1

2. Installation                           p.2

3. Connection                           p.3

4. Operation                           p.4

5. Maintenance                          p.5

6. Attention p.5

7. Key-point of measurement p. 6

Chapter Software (FsmⅡ & FsmV)

1. General p.1

2. Operating Environment and Connection Method p.1

3. Display Configuration and Button Function p.23.1 Display Configuration p.23.2 Button Function p.33.3 Mouse Operation (Common) p.4

4. Measurement Procedure p.5

5. Operating Instruction (Details) p.65.1 File Input/Output p.65.2 History p.65.3 Adjusting Measurement Area p.75.4 Setting Measurement Condition p.85.5 System Setup p.105.6 Calibration p.135.7 Measurement p.14

6. Specification 7. Guaranty

8. Certificate of inspection

FSM-6000LE Manual p.Contents

Page 2: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

[1]General

Surface Stress Meter FSM-6000LE is a device for measuring the surface compression stress , the depth of layer and the central tension stress of a tempered glasses by using optical waveguide effect in the surface layer of the glass.The meter, however, can not be applied to chemically tempered glass produced by (Li+ in glass) and (Na+ in salt bath) ion exchange and chemically tempered photochromic glass.

[2]Installation

Sets it up in a little dark place and in the place where the vibration is a little.

FSM-6000LEMeasurement head

[3]Connection

FSM-6000LE Harsware Manual p.1

Connect the cable of CCD camera to the USB socket of PC.

Connect the cable of One-touch switch (exclusive use with FsmⅡ Premium) to the USB socket of PC. And plug-in the AC adapter to outlet.

Page 3: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Hardware Manual p.2

(4) Start the PC up and start the FsmⅡ software. (refer the software manual)

It is not use the software without the dongle. Insert the dongle in the USB outlet.

(5) Drop a small amount of immersion liquids onto the prism. Put the sample glass on the prism with the immersion liquid between the sample and the prism. Use a small amount of immersion liquid as much as possible.It can not measure correctly by the dirt of liquid on the side of prism.

(6) Turn the lamp-height adjustment knob so that screen become brighter.

lamp-height is not good lamp-height is good

(7) Adjust the angle of telescope that the fringes appear in the center of the screen. Adjust contrast of the fringes with the brightness adjustment knob.

Adjustment knob for telescope angle

Brightness adjustment knob

[4]Operation

(1) Insert the plug into a line source socket.(2) Turn the power switch on.(3) The lamp becomes enough bright.

The measurement prism

Page 4: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Hardware Manual p.3

(8) It should be preset rotation angle of CCD camera, when the fringes inclines on the screen. Range of inclination for measurement is within ±0.3 deg.

Inclined fringes Accurate vertical fringes

① Crick the “Start Live” button - The fringes appear on the screen② It indicates the inclination degree of fringe on the screen continuously③ Loosen the two screws of the CCD camera mount④ Rotate the CCD camera and set the fringes vertically on the screen⑤ tighten the two screws of the CCD camera mountPlease try to measure your sample by FSM-6000LE

M4 set screws

Camera

Rotatable camera mount

too dark too bright good

Indication of inclination degree

Page 5: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Hardware Manual p.4

(9) The position of each device in which the interference fringes overlap might move up and down a little . At that time, adjust the area as follows (refer Software manual 5.3 Adjusting Measurement Area)

Overlapped position is out of the boundary area and the base area is on the right side fringes

Extend the boundary area and move the base area

(10) Measurment (refer the Software manual 5.7 Measurement Method)

(11) Wipe the surface of prism and turn the lamp off after measuring

Page 6: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

[5]maintenance

(1) Wipe off the immersion liquid or dirt of the surface of the prism with a soft paper wiper. The interference fringe becomes indistinct when there are a lot of scratches in the surface of the prism. At that time, replace it with the new prism with the holder. ( it is not included )

Prism and prism holderRefractive index of the prism is 1.72

(3) Immersion liquid as the accessory is safe. Please use the liquid of safety and non-causticity when you use other liquids. (Transparent liquid of refractive index nd=1.64 or more)

[6]Attention

The object lens fixed to the top of the telescope. Do not move the lens. It is important for the magnification of the device.

Object lens position

FSM-6000LE Hardware manual p.5

High Power LED lamp

(2) The lamp life is about 10,000 hours. However, there is an individual difference in the lamp life. After the lamp is dead, remove the back panel from the main body and replace it with the new lamp. ( it is not included )

Light source wavelength value is printed on the label.

Page 7: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

The fringes reverses to the bright fringes.The fringes always should be the dark fringes.Turn the lamp height knob to the left more.

The lamp height knob

The reversing fringes

The normal fringes

(1) Lamp height condition

The PC detects the dark fringes and calculates the surface stress and the layer depth.

[7] Key-point of measurement

FSM-6000LE Hardware manual p.6

Page 8: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

(2) A lamp brightness condition

Good condition: left picture has many fringes, right picture has a few fringes.

It is too bright.Turn the light brightness knob to the right.

The fringes of right side are not seen.All fringes have to be seen on the PC .The PC calculates the surface stress and the layer depth according to the position and the number of the fringes.

It is too dark.Turn the light brightness knob to the left.

The fringes of left side are not seen.All fringes have to be seen on the PC .The PC calculates the surface stress and the layer depth according to the position and the number of the fringes.

The light brightness knob

FSM-6000LE Hardware manual p.7

Page 9: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

(3) A telescope angle condition

Good condition: left picture has many fringes, right picture has a few fringes.All of fringes are seen on the center of the PC.

The fringes of the end of left side are not seen.Turn the adjustment knob for telescope angle to the left.

All fringes have to be seen on the PC .The PC calculates the surface stress and the layer depth according to the position and the number of the fringes.

Adjustment knob for telescope angle

The fringes of the end of right side are not seen.Turn the adjustment knob for telescope angle to the right.

All fringes have to be seen on the PC .The PC calculates the surface stress and the layer depth according to the position and the number of the fringes.

FSM-6000LE Hardware manual p.8

Page 10: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.1

1.General

2.Operating Environment and Connection Method

This manual describes how to use Image Analyzing Software for Surface Stress Meter (FSM-6000LE) .For the hardware and software environment, e.g. OS, of this system, refer to the attached manual.

All the malfunctions and damages caused by this software are not responsible.

Some model printer could not print out correctly.

OS: Microsoft Windows 7 Ultimate 32 bit English versionComputer : CPU compatible with Pentium 4.2GHz or higherMemory : 512MB or greaterHDD : 20GB or greaterDisplay : XGA(1280x1024 pixels), 24-bit color or higherDevelopment Microsoft Visual Studio .Net Proffesional 2010 , VisualC++2010Environment:Connection: Connect as below. Before connecting the cables, the equipments power-off

then unplug the power cords from the outlet

Note: Do not use another application program while this application is running.Installation or/and uninstallation of application program(s) or/and driver(s) in the PCmay cause malfunction in this application.

Surface Stress MeterFSM-6000LE

PC

Connect to USB socket of PC

Fig2. Connection Method

Attention:

o This software uses the dongle as copy guard. o Please insert the dongle in the USB outlet when you use this software.o A part of function doesn't operate when the dongle is not used.o Please do not lose the dongle.o It is necessary to purchase a new software with a dongle when losing it.

One-touch switch

Thickness meterPower-supply

box

Connect to Power-supply box

Connect to RS232C port of PC

(Option)

(Option)

Connect to USBSocket of PC

Page 11: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.2

3.Display Configuration and Button Function

Fig3.1 Display Configuration

① Image

⑥ Measurement Result List

⑦ Status

⑧ Location of selected portion

3.1 Display Configuration

Display the four measurement areas. Their locations and sizes are changeable for details of the measurement area and the adjusting method (see item 5.3).

① Image

⑥ Meas. Result List

⑦ Status

⑧ Location of selected portion

② Meas (Upper)

④ Meas (Lower)

⑤ B

ase

③ Boundary

②~⑤ Measure- ment Area

Display 1272x 840 pixels BW image. For image by camera, adjust 636x420 pixels in the center portion of the image twice. The following displays are available (see to item 3.2).

- Display live image and memory image- Display Grid Ruler- Display Auxiliary Figure- Display cross-section brightness and brightness projection etc.

Display measurement results (Refer to item 5.7).

Display the measurement conditions, the names of image filesand the glass thickness, etc.

Display the coordinates and the brightness of the location selected with mouse.

⑨ Operating button

⑨ Operating Button Refer to item 3.2.

Page 12: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.3

3.2 Button Function

Button Function Restriction

Table3.2.1 Button Function (1)

(1) Read Data

(2) Save Data

(4) Print

(3) History

(8) Analysis View

(9) Start Live

(10) Capture

(15) Manual Meas

Read the file of image data and measurement result.

Display the dialog of “Measurement History”.

Print the original image data and the measurement result.

(5) Adjust Area Adjust the measurement areas.Display detail info of the measurement areas in (11) below.

(6) Graph Display stress distribution.

(7) Auxiliary View Display/Hide the information after auto/man measurement.

Display the brightness analysis figure.Select the type of the analysis figure in (24) below.

Display/Hide Live image.

Capture image in the camera to the original image memory.

(11) Meas. Cnd.

(12) System Cnd.

(13) Calib Meas Start the calibration

Start the auto measurement.

(16) Exit Exit the application program.

Note:

*1: “Auxiliary View” is available only after the auto/man measurement finished.*2: The “Analysis View” button is effective only when the permission for usage set up on

shipping. *3, 4: Please make sure that the camera is connected correctly.*5: The program does not support “High Resolution Camera”. *6: The “System Condition” button is effective only when the permission for usage set up on

shipping.*7: “Combo Box” is available only when the “Analysis View” button is selected.

Button

Table3.2.2 Button Function (2)

Item

5.1

5.2

5.35.4

5.7

Select the type of the data analysis figure in (8) above from the followings.- Cross-Section Brightness (Horizontal/Vertical/3D Left/3D Right)- Brightness Projection (Horizontal/Vertical)

-

-

-

-

-

-

-

*1

- *2

- *3

- *4

Display the dialog of “Measurement Condition” to set up data. 5.4 *5

Display the dialog of “System Condition” to set up data. *65.5

5.6 -

(14) Auto Meas 5.7

Start the man measurement. 5.7

-

-

- -

Function Item

(17) Combo Box - *7

Save the file of image data and measurement result.

Restriction

Page 13: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.4

3.3 Mouse Operation (Common)

L ButtonMovement and Function

Table3.3 Mouse Operation

Note:*1: Changing the image view page is effective only when the permission for usage set up on shipping.*2: Some mouse or driver does not respond to mouse wheel movement.*3: “Analysis View” is effective only when the permission for usage set up on shipping.

Mouse Operation

- Change the image view page.

Common mouse operation is shown in Table 3.3. For mouse operation in “Adjust Area” and “Man Measurement”, refer to item 5.3 and item 5.7 respectively.

D-Click

R Button Wheel

Click

D-Click

(1) Normal Function

- Move “Cursor”. The location is shown in Fig 3.1 ⑧.

- Change the image view page to No.1 (Original).

Restriction

*1

Click - (NA)

Rotate *1,2

Rotate - Move “Cursor” (just after left clicked). *2

- -

(2) Function in “Analysis View” mode

Click -- Move “Cursor”. The location is shown in Fig 3.1 ⑧.- Display the Cross-Section Brightness or the Brightness Projection of the portion selected.

D-Click - - - Same as (1)

Click - Move “Cursor” only when “Brightness (Horizontal)/(Vertical)” specified.- -

- Same as (1)D-Click Rotate-

Rotate- - - Same as (1)

*1

*1,2

*1,2

*3

Page 14: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.5

Fig.4 Measurement Procedure

4.Measurement Procedure

Standard measurement procedure is shown below.

First Measurement

Read Image “Capture”

Setting “Measurement Condition”(item 5.4)

Normal Measurement

Optical Adjusting

Setting “System Condition” (item 5.5)

Fine “Adjusting Area” (item5.3)

Setting “Measurement Condition” (item 5.4)

Fine “Adjusting Area” (item 5.3)

“Manual Meas.” (item 5.7) “Auto Meas.” (item 5.7)

“Save Data” (if necessary: item 5.1)

“Print”(if necessary: item 3.2)

“Calibration” (item 5.6)

Premium measurement procedure is shown below.

First Measurement

Push “Aute Meas.” for more than 2 seconds

Setting “Measurement Condition”(item 5.4)

Normal Measurement

Optical Adjusting

Setting “System Condition” (item 5.5)

Fine “Adjusting Area” (item5.3)

Setting “Measurement Condition” (item 5.4)

Fine “Adjusting Area” (item 5.3)

“Meas.”

Page 15: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.6

5. Operating Instruction (Details)

5.1 File Input/Output

5.2 History

(1) Save File

Click the “Save Data” button to save the measurement image (original image) data and the measurement result. Display only the saved data in the “History View” (see to item 5.2).

(2) Read File

Click the “Read Data” button to read the measurement image (original image) data and the measurement result. It is also able to read the data from the “History View” (see item5.2).

Click the “History View” button to open the “History View” dialog. This dialog shows the list of the measurement result saved in “Save File”(see item 5.1 (2)). The functions of the “History View” dialog are: - Click the “Delete History” button to delete all histories and the related files(see item 5.1). - Select one row from the history list, then click the “Read Data” button to read and show the measurement image data and the measurement result specified from the file(same as item 5.1 (2 - Select files you want from the history list, then click the “Save Data” button and save the files as CSV file . (shit key+selection:block selection, alt key+selection:individual selection)

Fig.5.2 History View“” dialog

(3) Type and Operation of File

Save File: specify extension (.bmp)

Save File: specify extension (.fsm)

Read File: specify extension (.bmp)

Read File: specify extension (.fsm)

File Name and Type

***.bmp : Image File, BMP format, 1272x840 pixels, 8bit B/W

***.bmp : Image File, BMP format, 1272x840 pixels, 24bit full color

***(B).bmp : Image File, BMP format, 1272x840 pixels, 24bit full color

***(R).bmp : Image File, BMP format, 1272x840 pixels, 24bit full color

***(G).bmp : Image File, BMP format, 1272x840 pixels, 24bit full color

***.bmp : Image File, BMP format, 640x480 pixels, 8bit B/W

***.fsm : Data for the application (measurement result etc.)

***.csv : For user, can be read into EXCEL (measurement result list)

c:\Fsm\Fsm_SystemData\Fsm_HistoryData.hst: History data for the application

Table5.1 File Type and Operation

×

×

×

×

×

×

×

×

×

×

×

×

×

◎*1

◎*1

◎*1

◎*1

◎*2

×

×

×

×

×

×

×

×

◎: shows the file that can be specified directly in saving and reading. ○: shows the file that cannot be specified directly but is saved and read simultaneously in saving and reading.×: shows the file that is not covered in saving and reading.*1: Read the color component (one of R/G/B; R if not specified) specified in the file name () as B/W image.*2: Enlarge 636x420 pixels in the center portion of the image twice to read.

Page 16: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.7

5.3 Adjusting Measurement Area

There are four measurement areas as shown in Table 5.3.

(1) Define the Measurement Area

Table5.3 Adjusting Measurement Area

Area

Measure “(Upper)”

Measure“ (Lower)”

Non-Measure“Boundary”

“Base”

Calib/Meas

Auto ManHints for setting up

- Set up the sample so that the whole horizontal portion of the fringe and the high contrast portion in vertical direction stay within the area.

(non-measurement)-

-Req

Click the “Adjust Area” button to adjust the location and range of the individual area. Setup them so that the image stays within the area necessary for measurement (see Table 5.3). The location (coordinates) of the area is shown in the “Measurement Condition” dialog (see item 5.4).

(2) Setup the Measurement Area

図5.3 Adjusting Measurement Area

-

Req Req

Req Req

It detects the brightness of dark area in “Measurement Condition”,”Depth Measurement”,”Adopt boundary opaque”

Page 17: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.8

5.4 Setting Measurement Condition

Click the “Meas . C n d . ” button to open the “Measurement Condition” dialog. Apply the current measurement condition to the calibration measurement and the auto measurement.

Table5.4 Setting Items of Measurement Condition (1)

Items Contents

Type of Temper

Refraction ( Sample)

Refraction(Surface of Sample)

Photo erastic constant

Specify the type of temper (Thermal/Chemical ). (*2)

Enter the inside/surface refractive index of a sample.It generally uses “inside” refractive index.

Enter the photo erastic constant of a sample

Remarks For comment. Do not affect action of the application.

Setting Name Name is “ConfigNo.*” (*=1 to16). Can save max 16 settings. (*1)

DOL Measurement (*3)

This function is only for measuring Chemical Tempered samples. Specify the method of detecting location where the fringe used for depth measurement disappears.◇Adopt Boundary (Transparent) : Use the left end location in the low brightness portion on the right end image for transparent glass by VR light source. It is suitable for clear boundary.◇Adopt the right end Fringe : Use the location of the fringe detected in the right end.※ Usually select “Adopt Fringe Boundary”.◇Adopt Boundary (Opaque) : Use the left end location in the low brightness portion on the right end image as the fringe boundary for Opaque glass by IR light source.◇Adopt Boundary (BP) : Use the left end location in the low brightness portion on the right end image for transparent glass by VR light source. It is suitable for obscure boundary.

(*1)  Measurement condition is saved to file "c;\Fsm\Fsm_ConfigData\Fsm_ConfigData*.cfg". (*=1~16)    Default value is saved to file "c;\Fsm\Fsm_ConfigData\Fsm_ConfigData.def".

(*2) In FsmV, select only “Chemical”

(*3) Unnecessary setting on the Man Measurement.

(*3) Only for Thickness meter

Measure the right end Fringe (*3)

This function is only for Thermally Tempered samples.Specify whether to use the fringe detected in the right end for measuring surface stress.※ Do not select this in case the location of the fringe detected in the right end is unstable.

<Sample Characteristic>

Fig.5.4(A) “Measurement Condition” dialog

Glass thickness Enter the thickness of a sample for indicating the internal tension stress as a reference value.

・・(Master) (*4)

・・(Calib.:) (*4)

Enter thickness of Master glass.

Displayed calibration value

Calibrate Thickness meter based on Master glass.Calibration (*4)

Measure (*4) Displayed measurement result by Thickness meter.

Page 18: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.9

Fringe Distance Definition

Meas. Range(*3)

Save as Default Save to the Default File at the same time on saving the settings.Click the “Read Default” button to read and display the default saved.

Table5.4 Setting Items of Measurement Condition (2)

Items Contents

(*3) Do not need setting up in the Man Measurement.

Fringe Color Specify the color of fringe to be measured. It generally selects “black ”.

Measurement (Upper) / Non Measurement / Measurement (Lower) / Base

Display the measurement area. Cannot adjust the area in this dialog.Adjust the area in “Adjusting Measurement Area” (see item 5.3).

<Measurement Method>

Specify the method of measuring the distance between a pair of upper/lower fringes.

Definition

“Parallel”

“Non-Parallel”

“V-Parallel”

Measurement Range

Table Definition of Fringe Distance and Measurement Range

No Change

Specify

No Change

Measurement Method

See Fig.5.4(B1)

See Fig.5.4(B2)

See Fig.5.4(B3)

<Measurement Area>

Read Default Read and display the default.

Save

Return

Save the settings, then close the dialog.

Close the dialog without saving the settings.

Fig.5.4(B1) “Parallel”

Fig.5.4(B2) “Non-Parallel”

Fig.5.4(B3) “V-Parallel”

N1

0

Nc

100N2

The Fringe Distance is a distance between two parallel lines⑫/⑭ that go through the center points⑪/⑬ of the upper/lower fringes.In case the lines are not parallel, the gradient of the lines⑫/⑭ is a average gradient of the upper/lower fringes.

Fig.5.4(B1~3) Measurement Method of the Fringe Distance①“Measure (Upper)” area ②“Non Measure” area ③“Measure (Lower)” area ④“Base” area

When the “Measurement Range” is specified as N1 to N2[%], the Fringe Distance is a distance between two lines that go through the center points between the point N1 and N2 in the upper/lower area and are parallel in vertical direction.

The Fringe Distance is a distance between two lines that go through the center points ⑪/⑬ of the upper/lower fringes and are parallel in vertical direction.

Image compositing processing

Use two setting of camera. Set camera condition. See Fig.5.5.In case of no checking, use 1st camera condition.

Page 19: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.10

5.5 System Setup ※ Each item is effective only when the permission for usage set up on shipping.

Click the “System Cnd.” button to open the “System Condition” dialog.

Table5.5 Setting Items of System Condition (1)

Items Contents

Light Source Wavelength

Refractive (Prism)

Optical Path Length

Master Piece Stress

Enter the light source wavelength of the equipment. It is indicated on the label of lamp

Enter the refractive index of prism of the equipment.

Enter the optical path length of the equipment.

Enter the stress value of the Standard glass.

※ Save the system setup data to File "c;\Fsm\Fsm_SystemData\Fsm_SystemData.ini".   Save the default data to File "c;\Fsm\Fsm_SystemData\Fsm_SystemData.def".

< Setting Equipment >

<Setting Display>

<Setting Calibration>

Pixel Resolution Specify the side length of one pixel.

Serial No. Display the serial no. of the equipment.

Correction Value Show the correction value acquired in the calibration.

Photoelastic Constant

Refraction (Sample)

Photoelastic constant of the Standard glass.

Refractive index of the Standard glass.

Calibration Type Select the standard. (Others are engineer items)

“Display”“Calib Meas.””Button”

Calibration Measure button becomes available.

Display fringe of non-calculating

Also display the fringes not used for measuring surface stress out of fringes detected as well as normal fringes.※ Usually set Off.

Display Meas value & line on Error

Display the measurement value and line as well as in normal measuring, even if error occurred in measuring.※ Usually set Off.

Display Pseudo-Color

Specify the display color.◇ “Check” ON Display pseudo-color equivalent to visual image.◇ “Check” OFF Display B/W grayscale image. (Default)※ Usually set Off.

Grid Distance Specify the division of Grid Ruler displayed when “Grid Ruler View” is selected.

Display the fringe no. & scale in man meas.

Display the fringe no. and the scale in the man measurement.In case numbers of fringes is counted to under the decimal place, select this.

Noise Margin N(sigma)

Min vertical length of fringe in area

Allowable Grad of Fringe in Area

Specify the sensitivity in distinguishing fringe and noise.For detecting a fringe in a small brightness deviation, enter smaller figure.

Specify the allowable fringe gradient. Specify the gradient against the screen vertical line.

Specify the min vertical length of the fringe to be detected in the individual area.

For SmoothingSpecify the area size of smoothing processing for image noise rejection.e.g. Enter H=4, V=5 to conduct the processing in the area [H:9=4x2+1, V:11=5x2+1].

<Setting Data Processing>

Allowable Distance of Fringes (Horizontal)

Specify the threshold for detecting abnormal distance of fringes and issuing warning. The processing depends on “Type of Temper” of “Measurement Condition” (see Table 5.4) as follows:◇ Select “Chemical” Specify allowable change rate of the distance between adjacent fringes.◇ Select “Thermal” In addition to the above, specify allowable change rate of the distance between adjacent pairs of upper/lower fringes.

Page 20: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.11

Table5.5 Setting Items of System Setup (2)

Items Contents

Cross-Section Brightness(3D Shift)

Specify the rotation value of 3D Cross-Section figure displayed when selecting “Cross-Section (3D Right) or (3D left) ” in “Analysis View”. With entering 3 (pix), it will rotate 45 degree.

Compare Adjacent Fringe Brightness

Reject the noise similar to a fringe shape by comparing brightness. ◇ “One Side” Compare the brightness of two adjacent fringes, then the fringe(i) is determined as noise in the following.

Brightness of Fringe(i)<Brightness of Fringe(i+1) x Setting[%]◇ “Both Sides” Compare the brightness of a fringe with fringes on both sides, then the fringe(i) is determined as noise in the following.

Brightness of Fringe(i)<Brightness of Fringe(i-1) x Setting[%] andBrightness of Fringe(i)<Brightness of Fringe(i+1) x Setting[%]

< Setting Data Processing > (continued from the previous page)

DOL (1)Change Rate of Brightness change

For the condition of detecting a fringe boundary, specify the threshold for change rate of the brightness at the boundary for transparent glass. (by selected “Adopt Boundary (Transparent)” of “Measurement Condition “dialog.)

AllowableDeviation between Upper/Lower Fringe Boundaries.

For the condition for detecting abnormal and issuing warning, specify the allowable deviation between upper/lower fringe boundaries.

Apply Brightness Correction (1), (2)

Correct the brightness distribution in the right-left direction on the image to improve the measurement accuracy of fringe location.※ Usually select only “Apply Brightness Correction (2). These are engineer items.

Save as DefaultSave to the Default File at the same time on saving the settings.Click the “Read Default” button to read and display the default saved.

Save

Cancel

Save the settings, then close the dialog.

Close the dialog without saving the settings.

Lower Limit of Fringe Brightness Grad

Select the lower limit of fringe brightness gradient (tangent value).

DOL (3) Change Rate of Brightness change

For the condition of detecting a fringe boundary, specify the threshold for change rate of the brightness at the boundary for apaque glass. (by selected “Adopt Boundary (Opaque)” of “Measurement Condition “dialog.).

<I/O Condition>

No output No output to any printer.

BL2-58 For Sanei Electric (BL2-58). Set the serial port.

FSM-STD For general printer. Set the serial port.

Every print items Print header items one by one.

RAM Disk

PIO

Thickness meter ・・ Use Thickness meter (required optional Thickness meter)

Use One-touch switch. (only for FsmXP Premium)

Use RAM disk (required commercially available RAM disk)

Page 21: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

Fig.5.5  “System Setup” dialog

Camera Condition

FSM-6000LE Software Manual p.12

Page 22: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.13

5.6 Calibration

(*1) Verify the update of “Calibration Value” in the “System Condition” dialog.(*2) Optical path length of the device is displayed in the Certificate of inspection. (*3) The surface stress value of the standard glass is displayed in the inspection report. (*4) The photoelastic constant of the standard glass is displayed in the inspection report. (*5) The refractive index value of the master piece is displayed in the inspection report.

Preparation Specify the followings in the “System Condition” dialog. (see item 5.5)(*1)

① Select the “Standard” in “Calibration Type” ② “Master Piece Stress” in “Calibration”(*4) ③ “Photoerastic constant” in “Calibration”(*5)       ④ “Refraction (Sample)” in “Calibration”(*6)       ⑤ Check “ Display Calib. Mesas. Button ” ⑥ Crick “Save” button

Click the “Start Live” button to display the image of fringes correctly on the screen.Adjust measurement aRea. (see item 5.3)Click the “Capture” button to read the image data.

Calibration Click the “Calibration” button to start the calibration. The calibration value is averaged by 5 times measurement of the Standard glass. It must put the Standard glass again on the prism every measurement.

Verification

(1) Procedure of the Calibration

Fig.5.6 Calibration

When the Calibration is done, the magnification of the device is automatically changed.Please inspect the device by measuring the standard glass usually. Please calibrate when the malfunction is admitted by the inspection result.

Attention

Page 23: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.14

5.7 Measurement

(1) Preparation for the Measurement (Common items for auto/man measurement)

Specify the measurement condition in the “Measurement Condition” and “System Condition” dialog. (see item 5.4 & 5.5) (*1)Click the “Start Live” button to display the image of fringes.Adjust measurement area. (see item 5.3)Click the “Capture” button to read the image data.

(*1) The items in “System Condition” are effective only when the permission for usage set up on shipping.

(2) Auto Measurement Method and Understanding of the Results

After the above (1) Preparation finished, click the “Auto Measure” button. The measurement result can be saved after measured. (see item 5.1 & 5.2)

◇ Measurement item is the Surface Stress.◇ Measurement objects are max 5 pairs of fringes (No.1 to No.5).◇ Display the detection line and the fringe no. in green color on the location detected after measured properly.◇ “Measurement Result List” is as follows: - Column: “Select No.” Shows the fringe no. numbered from the left end. - Column:Fringe (Upr) mm“” / “Fringe (Lwr) mm” Shows the location of the fringe detected. - Column: “Gradient Deviation” Shows the deviation of gradients between a pair of the upper fringe and the lower fringe in absolute value. - Column: “CS” Shows the surface stress value calculated from a pair of the upper fringe and the lower fringe.

◇ Measurement items are the Surface Compression Stress and the Depth of Layer and Central Tension.◇ Measurement objects are Fringe No.1 & No.2 and the location of fringe boundary.◇ Display the detection line and the fringe no. in green color on the location detected after measured properly.◇ “Measurement Result List” is as follows: - Column: “Select No.” The first and second row are for Fringe No.1 and No.2 respectively. The 4th row is for the fringe boundary, and the fringe no. is showed down to the first decimal place. - Column: “Fringe (Upr) mm” / “Fringe (Lwr) mm” Shows the locations of the fringe and the fringe boundary detected. - Column: “Gradient Deviation” Shows the deviation of gradients between a pair of the upper fringe and the lower fringe and a pair of the fringe boundaries in absolute value. - Column: “CS” Shows the surface compression stress value calculated from Fringe No.1 and No.2. - Column: “DOL” Shows the depth of layer from the number of fringe and the density of fringe. - Column: “CT” Shows the central tension from CS, DOL and thickness of glass. (CT is reference value)

① Select “Thermal Temper” in “Measurement Condition”

Fig.5.7.2 ① Auto Measurement (Thermal Temper)

② Select “Chemical Temper” in “Measurement Condition”

Fig.5.7.2 ② Auto Measurement (Chemical Temper)

Fig.5.7.2 ③ Status column (Chemical Temper)

<FsmⅡ>

Page 24: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.15

③ Display of measurement error

o The error message is displayed and the measurement is interrupted when the inclination of the interference fringes and the spacing and the number are conditions other than the definition.

o Please measure it again after adjusting the angle of the camera or adjusting the source of light etc.

o The density of the interference fringes might be different according to the condition of strengthening.

o The ghost of the interference fringes might be generated and the array of the interference fringes are not theoretical.

o The image processing can be likely to be done automatically though PC might display the error message for attention.

o The measurement has ended normally when it can be confirmed to have selected the interference fringes correctly.

o Example displayed that fringes spacing is illegal.

o The interference fringes that appears in a bright area is normal.

o However, fringes like the interference fringes appear in a dark area though it is indistinct.

o In this case, it advances previously clicking "OK" button.

o All detected fringes are displayed in the green and majenta color lines.

o The majenta color lines are drawn in the dark area.

o All fringes that PC recognized are displayed.

o Two left green lines correspond two left fringes position.

o One of the right green line corresponds the boundary of a bright area and a dark area.

o This measurement is correctly done.

o The majenta color line is displayed to check "System setting" - "Setting Display" -"Disply fringe of non-

calculating" by “レ” mark.

o It can not measure stress and depth when the number of fringes are less than two.

o The depth of layer is less than five micron in this case, and it can not measure the surface stress.

o It can measure the outline value of the surface stress by manual measurement method.

o It sets the first line on the left end fringe.

o It sets the second line near left of the boundary .

o It sets the fourth line on the boundary.

o It enters two as the number of fringes.

Page 25: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

④ Continuous mesurement (only for FsmⅡ standard)

It measures and saves data automatically at “AUTO MEASURE METHOD”It sets all condition for measurement and push the “AUTO MEASURE BUTTON” for 2 seconds or more. The below dialog is displayed on the screen.

① Check “Save Data Folder”, the result is saved into the specified folder every measurement.② Check “Print Data (Label Printer)”, the result is printed out to the specified printer every measurement.③ Push “Folder” button and specify the folder for saving data.④ Push “Clear” button and clear the File Title and put the file title into the space.⑤ Sample No. is incremented one by a measurement.⑥ Sample No. Is changed by <>button and sample number is changed to zero by Clear button.⑦ Push “Measure” button and measure the stress and depth.⑧ Push “Back” button and come back to before measurement.⑨ Push “Next” button and go ahead to next measurement.⑩ Push “Cancel” button and close the this dialog and come back to standard measurement.

③ ④

⑦ ⑧ ⑨ ⑩

The saved data are original picture and result of measurement and saved by CSV file.

FSM-6000LE Software Manual p.16

Page 26: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.18

⑤ Judgment (only for FsmⅡ Premium)

Judge OK or NG based on the result of measurement and the judgment condition on automatic measurement. It sets all condition for measurement and push the “AUTO MEASURE BUTTON” for 2 seconds or more. The below dialog is displayed on the screen.

① Select among judgment condition No.1 ~ 16.② Set every particular item a ~ f.   a. Enter the name of judgment condition.   b. Enter the value of judgment condition.    CS...compression stress / DOL...depth of layer / CT...central tension / GT...glass thickness   c. Select measurement condition.   d. Save the data by checking a box.   e. Designate a location to save data.   f. Enter the header of data name.③ Displayed the measurement result and judgment.④ Measure.⑤ Test measuring without saving data.⑥ Finish judging.

<A> <B>

<A>① ② b

ced

a ③

f

<B>

④⑥⑤

The measurement data is saved as csv file. The measurement data which have the same header of data name, date and the judgment condition are saved on the same csv file. The file name is shown as follows.

EX) Judgment condition: 01 / Header of data name: ABC / Date: 2013.12.01 “ABC_20131201_No.01.csv”

Page 27: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.16

(3) Man Measurement Method and Understanding of the Results

The measurement procedure is shown in Table 5.7.3. The measurement result can be saved. (see item 5.1 & 5.2).

After item 5.7 (1) Preparation for the Measurement finished, click the “Man Measure” button.In case “Auto Measurement” conducted just before “Man Measurement”, the message in fig. 5.7.3A is shown. When select “Yes”, keep the last measured detection location of fringe.

Fig.5.7.3A Message

Thermal Temper Chemical Temper

② Select the fringe no.(No.1 to 5)

Select the checkbox for Fringe (No.1 to No.5)to be measured in the measurement result sheet.

Specify whether to measure Fringe No.3in the measurement result sheet. (*2)

L-Click “Cursor”, it turns red-purple. (see Fig. 5.7.3.B1)L-Click & drag “Cursor” or click the “<<” / “>>” button to move it on the fringe/fringe boundary.R-Click “Cursor” to confirm its location. (see Fig. 5.7.3 B2)Select all fringes and fringe boundary in the same way as the above.

Fig.5.7.3.B1 Selecting Fig.5.7.3.B2 Confirmed

④ Calculate Surface Stress (and Depth)

Click the “Adopt Value ” button.Display the measurement result in the result list.The format is same as item 5.7.2①

Click the “Adopt Value ” button to display the dialog below. (see Fig. 5.7.3 C)In case of selecting Fringe No.3(/No.4), enterthe numbers of fringes. (*3)

③ Select the fringe (and the fringe boundary)

Fig.5.7.3.C Enter the numbers of Fringes

The measurement result is shown in the result list. The format is same as item 5.7.2②. (*2)

*1 *2 In case of choosing Fringe No.3, the numbers of fringes (*3) at the fringe boundary (Fringe No.4) is estimated from Fringe No.3. *3 The numbers of fringes is calculated from specifying the leftend fringe end as the first fringe.

Table5.7.3 The Man Measurement Procedure for FsmⅡ

① Getting started

Fig.5.7.3.B3 

It selects the fringe so that the upper and lower fringes are pair. It selects the fringe 1,2,… from the left sequentially. (Fig. 5.7.3.B1 B2)

It selects the fringe so that the upper and lower fringes are pair. It selects the fringe 1,2 from the left sequentially and selects the fringe No.4 from boundary. (Fig 5.7.3.B3)

Page 28: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.17

◇ Measurement items are the Surface Compression Stress and the Depth of Layer and Central Tension.◇ Measurement objects are Fringe No.1 & No.2 and the location of fringe boundary.◇ Display the detection line and the fringe no. in green color on the location detected after measured properly.◇ “Measurement Result List” is as follows: - Column: “Select No.” The first and second row are for Fringe No.1 and No.2 respectively. The 4th row is for the fringe boundary, and the fringe no. is showed down to the first decimal place. - Column: “Fringe (Upr) mm” / “Fringe (Lwr) mm” Shows the locations of the fringe and the fringe boundary detected. - Column: “Gradient Deviation” Shows the deviation of gradients between a pair of the upper fringe and the lower fringe and a pair of the fringe boundaries in absolute value. - Refer to P19 about “CS2”, “DOL_Tale”,“CT_CV”, “CS1”, “CS_TP”, “DOL_TP”, “DOL_50”, “DOL_Zero”, “GR”.

③ Select “Chemical Temper” in “Measurement Condition” (FsmV is unavailable for thermally tempered glass.)

<FsmV>

Fig.5.7.4 ② Auto Measurement (Chemical Temper)

Fig.5.7.4 ③ Status column (Chemical Temper)

(1) Preparation for the Measurement (Common items for auto/man measurement)

Specify the measurement condition in the “Measurement Condition ” and “System Condition” dialog. (*1), (*2)Click the “Start Live” button to display the image of fringes.Adjust measurement area. (see item 5.3)Click the “Capture” button to read the image data.

(*2) The items in “System Condition” are effective only when the permission for usage set up on shipping. (refer to item 5.5)

(2) Auto Measurement Method and Understanding of the Results

After the above (1) Preparation finished, click the “Auto Measure” button. The measurement result can be saved after measured. (see item 5.1 & 5.2)

(*1) Double ion exchanged glass sometimes require to check “Image compositing processing” in “Measurement condition”. (refer to item 5.4)

Page 29: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.18

(3) Click “Graph” button

The graph of stress distribution appear.

(4) Click “Setting” button.

The detail data of stress distribution appear.

① … The graph of 1st strengthening. It is estimated from the stress distribution of twice strengthening. ② … The graph of 2nd strengthening. It is estimated from the stress distribution of twice strengthening.③ … Cross point (CP). The point where the two lines cross.④ … Turning point (TP). The point where the nearest point to CP on the stress distribution.⑤ … Refraction graph or birefringence graph.⑥ … Refraction data and depth data.⑦ … Stress data on every depth.

Fig. 5.7.5

⑥ ⑦

Fig. 5.7.6

Page 30: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.19

Items Contents

DOL

CS Surface stress from the fringe no.1, no.2 and boundary.

<Liner approx.> Calculated as the refraction distribution is liner. Same measurement algorism as the FsmⅡ.

Depth of layer from the number of fringe and the density of fringe.

CT Central tension from CS, DOL and thickness of glass.

<Curve approx.> Calculated as the refraction distribution is curve.

CS2

CS_TP

Surface stress from the fringe no.1, no.2 and boundary.

Surface stress on the Turning point.

CS1 Expected Surface stress which is given by 1st strengthening.

CT_CV Central tension according to stress distribution.

GR The gradient of 1st strengthening-to-the gradient of 2nd strengthening ratio.

DOL_TP Depth of layer on the Turning point.

DOL_50

DOL_Zero

DOL_Tale

Depth of layer where the surface stress is 50 MPa.

Depth of layer where the surface stress is 0 MPa.

The deepest Depth of layer which can be measured by FSM system.

CS_CP

DOL_CP

Thickness

Surface stress on the Cross point.

Depth of layer on the Cross point.

Glass thickness.

Fitting

Meas.

TPCP

Show the fitting curve of refraction or birefringence.

Show the stress curve.

Show the Turning point and Cross point.

Add Data

Add Fitting Data

Transfer the stress data to comparison graph.

Transfer the fitting stress data to comparison graph.

Cnd.

Renew

Save Data

Condition to make a graph.

Save the text dada of refraction distribution and stress distribution.

Apply Apply the display condition on the “Stress Distribution” window.

Minimize Minimize a “Stress Distribution” window.

⑧ … Select the kind of graph.

Show the both Upper refraction distribution and Lower refraction distribution.

Show the refraction distribution from Upper fringes.

Show the refraction distribution from Lower fringes.

Show the refraction difference.

Upper:

Lower:

Upper & Lower:

For Caluculation:

⑧ … Select the stress calculation method.

Can change configuration arbitrarily

Calculate from refraction difference.

Calculate from the average of refraction distribution of both

Upper refraction and Lower refraction.

Exact Mode:

Approx. Mode:

Test Mode:

Fig. 5.7.7

Fig. 5.7.8

Table 5.7.9 Stress Distribution window

Renew a graph after change condition.

Page 31: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Software Manual p.20

(5) Man Measurement Method and Understanding of the Results

After item 5.7 (1) Preparation for the Measurement (FsmV) finished, click the “Man Measure” button.In case “Auto Measurement” conducted just before “Man Measurement”, it is inherited from “Auto Measurement”. (recommended)

Upper

Luminance

Select Upper fringe.

Items Contents

Lower Select Lower fringe.

Move Move a fringe in a transverse direction.

Select fringe number. “RE” means boundary position.

Add Add a fringe.

Delete a fringe.Delete

<<< or >>>

<< or >>

< or > Move a fringe in 1 pixel.

Defaultposition

Reconvey a fringe.

Show the brightness pattern.

Measure Measure the stress.

Cancel Cancel the manual measurement.

Move a fringe in 10 pixels.

Move a fringe in 50 pixels.

Fig. 5.7.9

Page 32: FSM-6000LE Manual p.Contents FSM-6000LE Manual …img53.chem17.com/1/20160307/635929627728964496436.pdf2016/03/07  · FSM-6000LE Manual Ver.2015.7 Table of Contents Chapter Hardware

FSM-6000LE Specification

6. Specification

Name: Surface Stress Meter FSM-6000LEMeasuring range (stress): 0-1000mpa (ARTCAM-285-BW-OHS: 0-1500mpa)Measuring range (depth): 10-100 micrometers (Chemical tempered glass only, over two fringes) Measurement decomposition ability (stress): 1mpa Measurement decomposition ability (stress layer): 0.1 micrometers Measurement accuracy : (stress)± 20mpa (at the master piece)Measurement accuracy : (depth)± 5 micrometersMeasurement area: 12*7mm (Regular size) Specimen size: 10*10mm or moreMeasuring principle: Photoelastic analysis by refractometerSource of light: High power LED + band pass filter made by ORIHARASource of light starter: +5VDCMeasurement prism: size=12*7mm RI=1.72/590nmImage reader: CCD XC-ES51 made by SONY , C-MOS ARTCAM-130SS made by ARTRAY CCD ARTCAM-150P3 or ARTCAM-150P5 or ARTCAM-285 made by ARTRAYImmersion liquid: Special liquid for measurement (nD=1.64) with MSDS

PC for analysis: Model with Pentium 4.2Ghz or more Display: Model of XGA 24bit color displaysAnalytical software: Special software for FSM-6000LE FsmⅡ Standard or FsmⅡ Premium or FsmVPower supply: AC100-240V, 5A or less (whole)Size (measuring head): 280*600*220mmWeight (measuring head): About 10kg

Attention:

o This software uses the dongle as copy guard. o Please insert the dongle in the USB outlet when you use this software.o A part of function doesn't operate when the dongle is not used.o Please do not lose the dongle.o It is necessary to purchase a new software with a dongle when losing it.