104

Fracture Resistance of Non-Metallic Molar Crowns ... · ractureF Resistance of Non-Metallic Molar Crowns Manufactured with CEREC 3D Dalia A. Madani Master of Science Graduate Department

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Fracture Resistance of Non-Metallic Molar

Crowns Manufactured with CEREC 3D

Dalia A. Madani, BSc, DDS

A thesis submitted in conformity with the requirements

for the degree of Master in Science

Graduate Department of Biomaterials

Faculty of Dentistry

University of Toronto

© Copyright by Dalia Madani (2010)

Fracture Resistance of Non-Metallic Molar Crowns Manufactured withCEREC 3D

Dalia A. MadaniMaster of Science

Graduate Department of Biomaterials, Faculty of DentistryUniversity of Toronto

2010

Abstract

Objectives: To compare fracture strength and fatigue resistance of ce-

ramic (ProCAD, Ivoclar-Vivadent)(C) and resin composite (ParadigmMZ100,

3M/ ESPE)(R) crowns made with CEREC-3D.Methods: A prepared ivorine

molar tooth was duplicated to produce 40 identical prepared specimens made

of epoxy resin (Viade). Twenty (C) crowns and 20 (R) were cemented to their

dies using resin cement. Ten of each group were subjected to compressive

loading to fracture. The remaining 10 of each group were subjected to me-

chanical cyclic loading for 500,000 cycles. The survivors were subjected to

compressive loading to fracture. Results: No signi�cant di�erence in mean

fracture load was found between the two materials. However, only 30% of

the (C) crowns vs. 100% of the (R) crowns survived the cyclic loading test.

Conclusions: (R) crowns demonstrated higher fatigue Resistance than (C)

crowns in-vitro and might better resist cracking in-vivo.

ii

Contents

1 Introduction 1

1.1 Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1.2 Porcelain-Fused-to-Metal Restorations . . . . . . . . . . . . . 3

1.2.1 Advantages of Porcelain-Fused-to-Metal Restorations . 3

1.2.2 Limitations of Porcelain-Fused-to-Metal Restorations . 4

1.3 History of the CEREC technology . . . . . . . . . . . . . . . 4

1.3.1 CEREC I . . . . . . . . . . . . . . . . . . . . . . . . . 5

1.3.2 CEREC II . . . . . . . . . . . . . . . . . . . . . . . . . 6

1.3.3 CEREC III . . . . . . . . . . . . . . . . . . . . . . . . 7

1.4 Advantages and Limitations of the CEREC technology . . . . 9

1.4.1 Advantages of the CEREC technology . . . . . . . . . 9

1.4.2 Limitations of the CEREC technology . . . . . . . . . 12

1.5 Materials Used for Chair-side CEREC . . . . . . . . . . . . . 13

1.5.1 Ceramics . . . . . . . . . . . . . . . . . . . . . . . . . . 14

1.5.2 Composite Resin . . . . . . . . . . . . . . . . . . . . . 18

1.6 Concerns of the current CEREC materials . . . . . . . . . . . 20

iii

1.6.1 Strength . . . . . . . . . . . . . . . . . . . . . . . . . . 20

1.6.2 Esthetics . . . . . . . . . . . . . . . . . . . . . . . . . . 21

1.6.3 Post-operative Sensitivity . . . . . . . . . . . . . . . . 22

1.6.4 Margin Adaptation . . . . . . . . . . . . . . . . . . . . 24

1.6.5 Enamel Wear . . . . . . . . . . . . . . . . . . . . . . . 26

1.6.6 Longevity . . . . . . . . . . . . . . . . . . . . . . . . . 27

1.7 E�ects of porcelain surface pre-treatment and luting cement

on bond strength . . . . . . . . . . . . . . . . . . . . . . . . . 31

2 The Project 33

2.1 Rationale, Objectives and Hypothesis . . . . . . . . . . . . . . 33

2.1.1 Rationale . . . . . . . . . . . . . . . . . . . . . . . . . 33

2.1.2 Objectives . . . . . . . . . . . . . . . . . . . . . . . . . 34

2.1.3 Hypothesis . . . . . . . . . . . . . . . . . . . . . . . . . 34

2.2 Materials and Methods . . . . . . . . . . . . . . . . . . . . . . 35

2.2.1 Sample Size Calculation . . . . . . . . . . . . . . . . . 35

2.2.2 Preparation . . . . . . . . . . . . . . . . . . . . . . . . 36

2.2.3 Optical Impression and Restoration Designing . . . . . 37

2.2.4 Crown Fabrication . . . . . . . . . . . . . . . . . . . . 38

2.2.5 Cementation . . . . . . . . . . . . . . . . . . . . . . . . 41

2.2.6 Cyclic Fatigue and Fracture Test . . . . . . . . . . . . 44

2.2.7 Statistical Analysis . . . . . . . . . . . . . . . . . . . . 46

2.3 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

iv

2.3.1 Direct Loading Fracture Test Results . . . . . . . . . . 49

2.3.2 Compressive Cyclic Loading Test Results . . . . . . . . 55

2.4 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

2.4.1 Signi�cance of this research . . . . . . . . . . . . . . . 57

2.4.2 Fracture strength and fatigue resistance of Paradigm

MZ100 . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

2.4.3 Fracture strength and fatigue resistance of ProCAD . . 60

2.4.4 Fracture strength and fatigue resistance comparison . . 64

2.4.5 Clinical relevance of study design . . . . . . . . . . . . 68

2.4.6 Limitations of study design . . . . . . . . . . . . . . . 71

2.4.7 Future projects . . . . . . . . . . . . . . . . . . . . . . 73

2.5 Conclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

Bibliography 75

v

List of Tables

2.1 Epoxy resin die replicas dimensions . . . . . . . . . . . . . . . 37

2.2 Physical Properties of some of the materials used in this study

(Manufacturer's Data) . . . . . . . . . . . . . . . . . . . . . . 39

2.3 The composition of some of the materials used in this study

(Manufacturer's Data) . . . . . . . . . . . . . . . . . . . . . . 41

2.4 Materials used in this study . . . . . . . . . . . . . . . . . . . 43

2.5 Changes in the fatigued crowns . . . . . . . . . . . . . . . . . 55

2.6 Fracture loads (in N) of the four test groups . . . . . . . . . . 56

vi

List of Figures

1.1 CEREC-3 System, with split acquisition/design (a) and ma-

chining (b) units . . . . . . . . . . . . . . . . . . . . . . . . . 7

1.2 CEREC III: cylindrical diamond and tapered burs . . . . . . 8

1.3 CEREC III: in 2006 a �step bur� replaced the cylinder . . . . 9

2.1 The Epoxy resin replicas. . . . . . . . . . . . . . . . . . . . . . 37

2.2 Paradigm MZ100 bloc in the milling chamber prior milling . . 39

2.3 Paradigm MZ100 crown during milling. . . . . . . . . . . . . . 40

2.4 Di�erent views of the resulting Composite Resin and Ceramic

crowns. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

2.5 Ultrasonic cleaning system that was used to clean the crowns. 42

2.6 Panavia F 2.0 Dual Cure Dental Adhesive Cement (Kuraray

Medical Inc.). . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

2.7 Crowns inserted in their corresponding replicas under static

pressure of 2.2 kg. . . . . . . . . . . . . . . . . . . . . . . . . . 45

2.8 Crown complete seating was veri�ed using a caliper. . . . . . . 46

2.9 Composite Resin (Paradigm MZ100) specimens. . . . . . . . . 47

vii

2.10 Ceramic (ProCAD) specimens. . . . . . . . . . . . . . . . . . . 48

2.11 Comparison between the resulting Composite Resin and the

ceramic crowns. . . . . . . . . . . . . . . . . . . . . . . . . . . 49

2.12 The Instron machine was used to apply load on the crowns. . 50

2.13 The load was applied along the long axis of the replicas with

a hardened steel bar centered at the central groove. . . . . . . 51

2.14 Mechanical Cyclic Loading Machine. . . . . . . . . . . . . . . 52

2.15 Fracture patterns following the direct fracture test. . . . . . . 53

2.16 Fracture Pattern of one of the Ceramic Crowns where the

crown remained cemented and the tooth replica was split in

half. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

2.17 Box-plot of Fracture load test before and after cyclic loading

(fatigue). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

2.18 Kaplan Meier survival curves of failure during cyclic loading

(fatigue) test of Paradigm MZ100 and ProCAD. . . . . . . . . 56

2.19 Number of cycles required for restoration failure graph. . . . . 63

2.20 SEM (500 X) image of milled surface of the Paradigm MZ100

crown showing �ne grooves which is indicative of a plastic

deformation mechanism of material removal. . . . . . . . . . . 66

2.21 SEM (500 X) image of the milled surface of the ceramic crown

showing a fracture mechanism of material removal. . . . . . . 67

2.22 SEM (60 X) comparison between the margin edge of the com-

posite resin vs. the ceramic crown. . . . . . . . . . . . . . . . 67

viii

Chapter 1

Introduction

1.1 Background

The need for esthetic restorations has increased tremendously over the past

decade. Ceramics have gained large popularity as esthetic restorative mate-

rials due to their high esthetic quality [95, 103], wear resistance, durability,

color stability [106], and biocompatibility [112, 4]. However, ceramics have

several disadvantages including low fracture resistance [28], excessive wear

of the opposing teeth [124], the need of more aggressive preparation design

[95, 16], technique sensitivity [16], less than ideal marginal adaptation [9],

repair di�culty [124], and high cost of fabrication. In ceramics, cracks ini-

tiate from defects in the �tted surface of the restoration. Sub-critical crack

growth is facilitated in the aqueous environment of the mouth. [31, 32, 123]

The biocompatibility of all-ceramic restorations [112] has elevated the in-

1

terest in metal-free restorations, especially that metal hypersensitivity has

been a concern to some people. Gold allergy, although rare, had been docu-

mented in the literature [65]. Population prevalence of metal contact allergy

was reported to be 0.78-9.5% for gold [107, 69], 1.6% for silver [63], 9% for

cobalt [69], 8.2% for tin [41], 8% for palladium [63], 8% for chromium [102],

and 29% for nickel [102]. Namikoshi reported a case that developed a hyper-

sensitivity reaction resulting in lichen planus due to the release of palladium

from a dental metal alloy [84]. Contact allergy to metal dental restorations

may even be a risk factor for development of intra-oral squamous cell car-

cinoma [51]. The highest percentage of metal allergic reactions occurred in

patients with lichenoid tissue reactions, lichen planus, burning mouth syn-

drome, stomatitis, gingivitis, and peri-oral dermatitis. [118]

The combination of advancements in dental material, computer technol-

ogy, and equipments has made it possible to fabricate an indirect esthetic

restoration in one appointment. The CAD/CAM CEREC system is used

for electronical designing and milling of restorations. Using this system, the

dentist can manufacture a restoration without the need for lab assistance and

without impressions nor temporary restorations [66]. The restoration can be

designed in less than �ve minutes and then milled in 10 to 12 minutes [119].

Hence, only one appointment is needed to prepare and insert the restoration.

After preparing the tooth, the dentist sprays tin oxide powder on the prepa-

ration. Then, he/she aligns the intra-oral camera's angle of vision for the

scan along the insertion axis of the preparation while checking the image on

2

the monitor. After the camera is stabilized, the preparation is scanned [78]

and the optical impression is completed. The restoration is designed, milled

and polished to prepare it for insertion.

1.2 Porcelain-Fused-to-Metal Restorations

1.2.1 Advantages of Porcelain-Fused-to-Metal Restora-

tions

� Clinically-proven longevity and fracture resistance: fracture rate of

porcelain-fused-to-metal crowns and bridges were reported to be as

low as 2.3% after 7.5 years. [5]

� Combines the esthetics of porcelain with the strength, marginal adap-

tation, and accuracy of cast metal restorations. [103]

� More resistant to fracture than conventional all-ceramic crowns. The

slight di�erence in the coe�cient of thermal expansion between the

metal and ceramic causes the ceramic to undergo residual compressive

stresses following the cooling process during crown fabrication . This

constant residual compressive stresses resists the tensile stresses the

crown is subjected to rendering the crown stronger and more fracture

resistant. [6]

� Amore conservative preparation provides adequate thickness for porcelain-

3

fused-to-metal crowns when compared to all-ceramic crowns. [5]

� No wear of ceramic by abrasion or attrition. [5]

� No change of color due to no microleakage between the veneer and the

metal. [5]

1.2.2 Limitations of Porcelain-Fused-to-Metal Restora-

tions

� Non-ideal esthetics since:

� the metal at the gingival margin may cause gingival discoloration

[131]

� the dark metal core may show through in areas with minimal

porcelain thickness especially at the gingival third

� the metal margin may become visible if gingival recession occurs

� Metal hypersensitivity

1.3 History of the CEREC technology

Advances in computer technology have made it possible to develop the CAD/CAM

technology (Computer-assisted Design/ Computer-aided Manufacturing). The

CEREC system (computer-assisted CERamic REConstruction) was the �rst

operational CAD/CAM system to be used in the dental o�ce [49]. The

4

design was produced by Dr. Brandestini and the software versions were de-

veloped by the CEREC teams at Siemens and Sirona (Bensheim, Germany)

[78]. In 1980, the basic concept was developed by Mormann (University of

Zurich) and Brandestini (Brandestini Instruments, Zurich). Dr. Alain Ferru,

a young French software engineer, was approached by Dr. Mormann to design

the CEREC software. In 1985, the �rst chair-side inlay was fabricated using

the CEREC I by Mormann and Brandestini (Brains, Zurich). In 1988, on-

lays and veneers capabilities were added to the unit. Full and partial crowns

and copings were made possible by introducing the CEREC 2 in 1994 by

Siemens (Munich, Germany). In 2000, the Cerec 3 was introduced and the

three dimensional capability was added in 2003. In 2005, the new software

enabled the automatic virtual occlusal adjustment. [78]

Several studies criticized the marginal �t of the CEREC restorations [121,

106]. However, the improvement in the CEREC unit and software made

it possible to produce more clinically acceptable marginal �t of the milled

restorations. [52, 81, 12, 53, 54, 44, 26]

1.3.1 CEREC I

The CEREC I prototype was available in 1983 [78]. The �rst unit had a

water turbine drive. The modi�ed version with E-drive and CEREC Oper-

ating System 2.0 was launched in 1991. In this system a ceramic block is fed

against a grinding wheel along the mesio-distal axis of the restoration with a

di�erent distance from the inlay axis at each feed step [78]. This system was

5

limited only to inlays, onlays, and veneers [49]. The two dimensional soft-

ware capability did not provide the restoration with any occlusal anatomy or

contact. Hence, the dentist had to carve and �nish the restoration's occlusal

anatomy and contacts him/herself after the unit milled the adapted surface

of the restoration which made this unit impractical to use. [78]

1.3.2 CEREC II

CEREC II was developed in 1994 to try to overcome the limitations of the

CEREC I unit. The team at Siemens (Munich, Germany) added a cylinder

diamond to the system which enabled the CEREC II unit to grind full and

partial crowns and copings [78]. The display was two-dimensional and the

design of the occlusion was introduced in three modes: extrapolation, corre-

lation and function. Bindl and Mormann [15] compared margin adaptation of

Vitablocs Mark II partial crowns fabricated with CEREC-I and CEREC-II

and found signi�cantly improved margin adaptation for CEREC-II manu-

factured partial crowns (207 ± 63µm) compared to those manufactured by

CEREC-I (308± 95µm). Mormann and Schug [82] also compared both sys-

tems in regards to precision of �t of the milled restorations. They found a 30

% improvement in the accuracy �t of CEREC-II manufactured restorations,

with marginal gap of 56±27µm, compared to that of CEREC I, which equals

84± 38µm.

6

Figure 1.1: CEREC-3 System, with split acquisition/design (a) and machin-ing (b) units

(a) (b)

1.3.3 CEREC III

The CEREC III was launched in 2000 (Figure 1.1 ). The system consists of

an acquisition unit containing a portable computer, design software and an

optical imaging system, and a milling chamber with two diamonds for milling

the �nal restoration from a prefabricated ceramic or composite resin block

[38]. The software was upgraded to its three-dimensional capability in 2003.

The CEREC-3D software is much easier to handle compared to the previous

versions. Automatic virtual occlusal and proximal contact adjustment of a

7

Figure 1.2: CEREC III: cylindrical diamond and tapered burs

selected digital full-crown anatomy was added in 2005. This enabled the

dentist to control the vertical dimension of the restoration prior to milling

[37]. In the CEREC III unit, a two-bur system , a cylindrical diamond and

a tapered bur, was introduced instead of the grinding wheel (Figure 1.2 ).

In 2006, the �step bur�, which has a smaller diameter at the top one-third

of the cylindrical bur, made it possible for the unit to fabricate restorations

with higher precision (Figure 1.3 ). [78]

The marginal �t of CEREC III milled inlays exceeded that of CEREC II,

although both were within the clinically acceptable range of 50-100 microns.

[30]

8

Figure 1.3: CEREC III: in 2006 a �step bur� replaced the cylinder

1.4 Advantages and Limitations of the CEREC

technology

1.4.1 Advantages of the CEREC technology

� CEREC restorations have demonstrated documented clinical success

and longevity [90]. Restoration size, tooth vitality, and tooth location

did not signi�cantly a�ect the prognosis. Posselt and Kerschbaum [90]

reported a survival probability of 95.5% over 9 years.

� The CAD/CAM blocks are fabricated under optimum controlled con-

ditions. This provides a restoration with higher intrinsic strength elim-

inating the material variation found in lab-fabricated restorations. [79]

� Utilizing the CAD/CAM technology, the computer controlled fabrica-

9

tion diminishes the potential inaccuracies resulting from human error

and is able to generate a restoration within a clinically acceptable �t

of about 50 micrometers as established by the American Dental Asso-

ciation. [30]

� Only one appointment is needed to complete the restoration and hence,

� the patient is subjected to only one administration of local anes-

thetic.

� no need to fabricate a temporary restoration [24], this is an ad-

vantage because:

* the temporary restoration is subject to loss, breakage or leak-

age which may lead to contamination of the dentinal tubules

and/or sensitivity. [119]

* requires time to fabricate. [119]

* may be di�cult to clean during the temporization period

which may lead to gingival irritation. [119]

* when removing the temporary restoration, pulpal stress may

occur as a result of excessive cleaning, drying or trauma. [119]

* veneer temporaries are time-consuming, di�cult to fabricate,

and can easily be displaced. [119]

� Decreased second-appointment set-up costs which is inevitable

with conventional laboratory-fabricated restoration, as the pa-

tient needs to come back for a separate appointment to insert

10

the restoration. This leads to fewer instruments needing steril-

ization, less need for chair time set-up/breakdown, and improved

o�ce e�ciency. [119]

� Elimination of the cost of some disposable supplies that are needed

for conventional restorations such as impression materials, wax, stone,

temporary resin material, and temporary cement. [119]

� The clinician has complete control over the �nal result and esthetics of

the restoration since the software delivers a restoration that may only

need some characterization by staining or glazing. [119]

� Minimal reported post-operative sensitivity due to: [38]

� use of rubber dam isolation in the clinical trials which ensures a

clean, isolated tooth surface for adhesion bonding.

� insertion of the restoration at the same appointment of the prepa-

ration which prevents possible tooth contamination during the

temporization phase.

� the use of factory manufactured composite and ceramic blocks

which minimizes the polymerization shrinkage since it is only lim-

ited to the resin cement interface.

� No laboratory fee involved. [119]

11

1.4.2 Limitations of the CEREC technology

� Cost of the equipment , which is approximately $100,000 plus tax,

is a factor especially for dentists in solo practices. To overcome this

problem, several o�ces may form a team and use one centrally located

CEREC machine. Thus, the cost can be shared among several dentists.

[24]

� Color of the �nished restoration is not ideal since the restoration is

milled from a monochromatic block. However, multicolor blocks have

been developed to overcome this limitation. Also, super�cial stains can

be placed by the dentist to mimic any shade variability of the patient's

tooth, but this requires the acquisition of a glazing oven. [24]

� Signi�cant time is required for the dentist to become pro�cient enough

in use of the system to be �nancially successful [24]. However, in

one study, CEREC 3D was introduced to dental students in their last

semester and they were able to produce clinically acceptable inlays with

high short-term success rate (2 years). [127]

� A well-trained, dedicated sta� is essential for a successful cost-e�cient

restoration. Sta� members could make the restoration after the dentist

has prepared the tooth. This saves time which adds to the �nancial

viability of the CEREC technology. [24]

� In severely broken down teeth, it is di�cult to digitally capture sub-

12

gingivally placed margins. Gingival retraction methods are needed in

such cases [24]. However, gingival retraction methods are also needed

in conventional restorations with sub-gingival margins.

� CAD/CAM technology in the dental o�ce is limited to single units

only. A system that can mill several unit restorations is needed to

overcome this problem. [24]

� The time needed to �t, polish and insert the CEREC restoration is

longer than that needed for a lab manufactured restoration. However,

with more experience, the dentist may become faster and more e�cient

in performing these tasks. [24]

1.5 Materials Used for Chair-side CEREC

For a material to be considered appropriate to be used by the CEREC system

it must exhibit the following properties:

� can be milled rapidly. [42]

� can withstand the harshness of the milling process. [34]

� can be easily �nished (polished, stained, or glazed). [42]

� provide a clinically acceptable restoration once cemented. [34]

13

All CEREC blocks have �ne particle-sized micro-structure which leads to

improved machining damage resistance, faster polishability, improved me-

chanical properties, and decreased wear of the opposing dentition. [42]

1.5.1 Ceramics

Feldspathic porcelain-based ceramics:

� Vitablocs Mark II (Vita Zahnfabrik, Bad Sackingen, Germany):

� Introduced in 1991. [34]

� Is a �ne-grained feldspathic ceramic [34] manufactured using �ne-

grained powders producing �ne crystal, pore-free ceramic which

leads to better polishability, lower enamel wear and higher strength

[42]. The strength of this ceramic material when polished is

approximately 130 MPa and it could reach 160 MPa or higher

if glazed. This is about twice as strong as conventional felds-

pathic ceramics and slightly higher than many pressable ceramics.

[101, 43]

� The average particle size is approximately 4µm. [34]

� It can be etched using hydro�uoric acid or abraded with Alu-

minum Oxide particles creating a retentive surface that provides

micro-mechanical retention to the adhesive resin cement. [34]

� CEREC Blocs (Vita Zahnfabrik, Bad Sackingen, Germany).

14

Leucite-reinforced Ceramics:

� ProCAD (Ivoclar Vivadent, Schaan, Lichtenstein) blocks:

� Introduced in 1998. [34]

� They can be etched. [34]

� A leucite-reinforced ceramic [34] with leucite crystals ranging from

5 to 10 micrometers in size. Strength properties are similar to

those of Vitablocs Mark II. [42]

� Empress CAD (Ivoclar Vivadent, Schaan, Lichtestein) blocks:

� The new generation of ProCAD

Advantages of Ceramics

� It is possible to produce a ceramic restoration with natural appearance

[9] in terms of color, translucency, and surface texture [95, 104] due to

the wide range of shades and translucency - opacity combinations that

can be customized to each case. The surface can also be characterized

by add-on stains for improved shade match to the adjacent tooth. [42]

� Ceramics are inert and highly biocompatible [109, 112] due to their

glass-like properties. Dental plaque adherence is minimized when the

ceramic restoration is polished and glazed [59]. A supra-gingival or

at-the-gingiva margin can produce an esthetically pleasing result with

15

all-ceramic restorations. This promotes a healthy periodontium since

,with these margins, the gingival involvement during tooth preparation,

impression, and function is minimized. [59]

� Machined ceramics demonstrated less wear rate when compared to type

I gold alloys, base-metal alloys, cobalt chromium alloys and composite

resin. [128]

� Ceramics have also shown better color stability compared to composite

resin restorations. [106]

Disadvantages of Ceramics

� Low tensile strength. [75]

� Higher yield stress response than enamel. [47]

� Crack propagation that may lead to catastrophic fracture [9]. Cracks

may initiate from two types of defects: fabrication defects (which are

introduced during processing or that exist as micro-structural features)

or surface micro-cracks (which are introduced during the machining

stage) [96]. The interaction of applied loads with those defects may

lead to microscopic damage which may cause restoration failure [28].

Failure can also occur as a result of sub-critical crack growth [31, 32]

which is facilitated by the surrounding aqueous environment.

� Excessive brittleness. [75, 98]

16

� Di�cult to repair if fracture occurs. [23, 98]

� Di�cult to adjust chair-side. [98]

� Technique sensitivity.

� Require �ring for glazing and stain characterization [98]. Glazing is

required for ceramics to attain optimum strength. [98]

� Firing shrinkage. [73]

� Wear of opposing natural dentition. [124, 98]

� Inadequate marginal �t which may lead to microleakage [9]. New ce-

ramic formulations and processing techniques were developed to over-

come such limitation. Also, the improvement of bonding systems and

the use of computed technology have resulted in all-ceramic restorations

with improved marginal adaptation and decreased microleakage. The

use of resin cement with ceramic restorations demonstrated improved

marginal �t and decreased microleakage. [97]

However, dental ceramics have been improved over the past two decades to

overcome the fore-mentioned limitations. Among these improved ceramics

is the injection molded Empress glass ceramic (Vivadent, Schaan, Liechten-

stein), a leucite-reinforced material that has improved fracture resistance.

17

1.5.2 Composite Resin

� Paradigm MZ100 (3M ESPE, St. Paul, Minn.)

� First introduced to the market in 2000. [34]

� ParadigmMZ100 is a bisphenol A-diglycidyldimethacrylate/triethylene

glycol dimethacrylate resin-based composite with �ller composed

of nanocrystalline zirconia in an amorphous silica matrix [98]. The

inorganic �ller is radio-opaque and its loading is 85% by weight

with mean particle size of 0.6 micrometers. Paradigm MZ100 is a

factory processed version of Z100 Restorative (3M ESPE) resin. It

has superior physical properties comparing to conventional Z100

resin due to controlled manufacturing conditions which lead to

a dense, pore-free material that is completely cured throughout.

This processing technique maximizes the degree of cross-linking

[34]. The degree of conversion of the methacrylate groups in

the Paradigm MZ100 is approximately 84% which is higher than

that for Z100 direct restorative composite resin material (74%).

This provides both dependable physical properties and su�cient

amount of unreacted monomers for bonding. [98]

Advantages of Composite Resin

� According to the manufacturers, a more conservative chamfer or bevel

preparation is acceptable with resin compared to ceramics which require

18

a shoulder or a broad chamfer. [121]

� Easier �nishing and polishing than ceramics. [42]

� Can be easily adjusted and/or polished intra-orally. [34]

� Kinder to opposing teeth in regards to wear relative to ceramics. [42,

98]

� Easier to add on adjustments if necessary which is di�cult in ceramic

restorations [132]. The surface can be air-abraded using 50µm silicon

dioxide particles then bonded to hybrid resin. [34]

� Color characterization can be easily accomplished by internal or exter-

nal resin tints and color modi�ers, which are light-cured rather than

oven-�red. [34]

� Using Paradigm MZ100 blocks can elongate the life of the milling di-

amond and increase the number of composite resin restorations milled

by each milling diamond before requiring replacement [34]. Rusin et al

[99] observed that ceramic materials required milling diamond replace-

ment at signi�cantly earlier stage than the Paradigm MZ100 material.

This leads to reduced operating cost and fewer delays from diamond

bur exchange. [99]

19

1.6 Concerns of the current CEREC materials

1.6.1 Strength

The blocks used for CEREC are fabricated under ideal manufacturing con-

ditions in a reproducible constant manner eliminating human error. This

results in a dense, defect-free, high-quality material. On the other hand,

conventionally fabricated restorations are made by hand which may intro-

duce human error that could a�ect its mechanical and esthetic properties.

[42]

In one study [117], it was concluded that industrially prepared ceramics

are more structurally reliable than conventional lab-fabricated ceramics al-

though CAD-CAM procedures may introduce surface and subsurface �aws

that may adversely a�ect this property. However, the strength can be gained

back by polishing using rubber wheels and diamond paste. Further enhance-

ment of strength of about 160 MPa can be acquired using a combination of

polishing and glazing. [42]

Attia concluded that the fracture resistance of teeth restored with CEREC

manufactured crowns was equivalent to that of unprepared natural teeth, but

was signi�cantly higher than that of teeth restored by conventional low-fusing

ceramic crowns. [9]

20

1.6.2 Esthetics

Esthetics have been a concern in CEREC materials due to the monochro-

matic nature of the blocks. However, the porcelain restoration can be stained

and glazed after milling. Color modi�ers and internal and external tints can

be easily added to composite restorations then cured [34]. Also, the blocks

come in a large variety of shades to match the adjacent natural dentition.

Vitablocs TriLuxe (Vita Zahnfabrik) contain three shades in a graded vari-

ation in which the middle layer has a regular chroma, the top layer has a

translucent, low intense chroma, and the lower layer has low translucency

and an intense chroma. This makes it possible to imitate the optical charac-

teristics of the natural tooth and hence make the restoration more blendable

with the adjacent teeth. [42]

Herrguth et al. conducted a clinical study to compare the esthetics of

individually stained CEREC-manufactured crowns to Cergogold crowns (De-

gussa Dental GmbH, Hanau, Germany) which is fabricated by the layering

technique [48]. They concluded that regardless of the type of crown used, the

restorations were esthetically acceptable to all patients with no statistically

signi�cant di�erence in the esthetic ratings between the two crowns. In a

clinical study of 109 inlays in 46 patients over 7 years, Cerutti et al. [21]

reported an initial good shade match of 88 % Alfa which decreased to 62.4 %

Alfa and 33 % Bravo after 7 years. Several studies reported that the degree

of color mismatch increased over time [76, 39]. In a series of studies by Sjo-

gren et al. [108, 110, 111] of 66 Vitablocs Mark II inlays, the color mismatch

21

increased from 16% to 38% after 10 years. The observed increase in color

mismatch was due to a color change of the tooth rather than the color shift

of the milled restoration [39]. In a di�erent clinical study, Fasbinder et al.

reported that the color match of Paradigm composite inlays was signi�cantly

better than Vitablocs Mark II ceramic inlays at three years. This was due to

the fact that composite resin appeared to re�ect the surrounding tooth color

to a better degree than did the ceramic inlays. [38]

In conclusion, CEREC-manufactured restorations can provide an esthet-

ically acceptable restoration when polished and an esthetically optimum one

when stained and glazed. The decrease in color match over time can be at-

tributed to a change in tooth shade and translucency, rather than a change

in the color of the CEREC-manufactured restoration. [36]

1.6.3 Post-operative Sensitivity

Signi�cant level of post-operative sensitivity was reported in early clinical

studies addressing CEREC-manufactured restorations. Magnuson et al. [67]

reported 9% immediate post-operative sensitivity in a study of 301 CEREC-

manufactured inlays. Most of these cases involving post-operative sensitivity

resolved within one month. However, three cases remained problematic for

six months and which necessitated endodontic therapy. In a di�erent study

by Sjogren et al. [108], 13.8 % of patients with Vitablocs Mark I or II inlays

complained of post-operative hypersensitivity. In a three-year clinical study

by Fasbinder et al. [40], 13 % of 92 Vitablocs Mark II onlays were slightly

22

sensitive after one week, and 4 % were slightly sensitive in two weeks. All

post-operative sensitivity resolved after one month, and no post-operative

sensitivity was reported in the three-year period of observation. Otto and

De Nisco [87] observed 13 % immediate post-operative sensitivity in 200

CEREC-manufactured inlays that were caused by premature occlusal con-

tact. Out of the 17 cases reported, 12 resolved within three weeks and the

rest resolved in seven months. The majority of post-operative sensitivity

can be attributed to occlusal interference since the CEREC-manufactured

restorations are inserted in a single appointment. Hence, it is advisable to

equilibrate the occlusal contacts after the e�ects of the local anesthetic have

worn away. [36]

More recent studies have reported less post operative sensitivity which

could be attributed to the signi�cant improvement of the adhesive materials.

In a study of 20 CEREC- generated Vitablocs Mark I inlays, Molin and

Karlsson reported no post-operative sensitivity throughout the observation

period of 5 years [76]. The same results were observed by Heymann et al. in

their four-year clinical trial of �fty CEREC-generated inlays [49]. Fasbinder

et al. reported only one sensitive restoration at one week in their randomized

clinical trial of 80 Vitablocs Mark II and Paradigm inlays. However, it was

resolved in two weeks and no sensitivity was reported in the remainder of the

observation period of three years. [38]

The lack of signi�cant post-operative sensitivity in CEREC-manufactured

restorations can be attributed to several factors. The optical imaging of the

23

preparation requires careful isolation. This ensures that optimum �uid con-

trol is possible which maximizes the adhesive cementation predictability [36].

The fact that CEREC-manufactured restorations eliminate the need of tem-

porization contributes to the lack of post-operative sensitivity, as it prevents

the possible contamination of the dentin tubules during the temporization

period if the temporary restoration is lost, fractured, or leaked. [38]

1.6.4 Margin Adaptation

Nakamura et al. [83] studied the e�ect of abutment occlusal conversion angle

and the computer luting space setting on the internal �t and marginal adap-

tation of the ceramic CEREC III milled restoration. They found that when

the computer luting space was set to 30 to 50 micrometers, the marginal gap

ranged from 53 to 67 micrometers and was not in�uenced by the abutment

angle of occlusal conversion. Other researchers measured margins of about

50 micrometers [30, 91] which suggested that the marginal �t of CAD/CAM

generated restorations is adequate for clinical use. Denissen et al. [26] re-

ported a margin gap of 85µm for CEREC III manufactured onlays, this is not

signi�cantly di�erent from that of the laboratory-fabricated onlays. These

recorded gaps are well within the reported maximum clinically accepted gap

of 120 µm. [74]

In a study of CEREC-manufactured restorations, Fasbinder reported less

detectable margins in resin-based composite inlays (Paradigm) compared to

ceramic inlays (Vitablocs Mark II) after 1 year [38]. However, in 3 years, no

24

signi�cant di�erence in margin adaptation between the ceramic and resin-

based composite inlays was detected. The nature of the tooth antagonist to

the restoration did not a�ect the amount of margin ditching [11]. Posselt

and Kerschbaum [90] reported 47.4 % of under�lled margins of 44 CEREC I

and CEREC II-manufactured restorations after 9 years.

A well-�tting margin is expected to maximize the longevity of a restora-

tion. In almost all the clinical studies of CEREC-manufactured restorations,

ditching due to wear of the composite resin cement at the margin was re-

ported. However, this ditching was not associated with margin discoloration

or recurrent decay. In a clinical study of 121 Vitablocs Mark II and Dicor

inlays cemented with micro�ll or hybrid resin-based composite luting agent,

a linear wear rate was reported in the �rst year but then it decreased by

approximately 50% [54]. They found that the vertical loss of cement at the

margin stopped when it reached 50% of the margin width. No microleak-

age or secondary decay was reported at the margin. Heymann et al. [49]

reported an increase in the wear of the adhesive luting agent at the occlusal

margin of inlays over the �rst 3 years, and then a decrease of that wear after

3 years. No enamel or porcelain inlay margin chipping or margin staining

was identi�ed as the adhesive cement started to wear. In a study compar-

ing Vitablocs Mark I and Dicor MGC porcelain inlays and P-50 resin-based

composite inlays, no signi�cant di�erence in margin adaptation was reported

[44]. No marginal staining or recurrent decay was noted either. In Otto and

De Nisco's study [87], a 74% occurrence of under�lled margins after 10 years

25

was not associated with clinical failure.

These studies agree to the fact that the margin wear is just a surface

phenomenon that is not associated with a deterioration in the adhesive bond

to the tooth. Also, the wear tends to localize at the margin of the occlusal

surface of the restoration rather than the margins of the proximal surfaces.

Micro�ll resin cements demonstrated better wear pattern compared to hybrid

resin cements. [57, 133]

1.6.5 Enamel Wear

Enamel wear is always a concern when ceramics are used as a restorative ma-

terial. Several factors may in�uence the harshness of ceramics against enamel

tooth structure. It is possible to minimize enamel wear by using �ne-grained

ceramics and by polishing or glazing the ceramic surface [42]. Several studies

have demonstrated that the wear of enamel against polished or glazed ceramic

restorations was essentially the same of that against enamel [61, 72, 71, 1].

Researchers measured the amount of tooth structure lost when in function

with di�erent dental materials and normalized the data relative to enamel

versus enamel. They found that Vitablocs Mark II and ProCAD blocks be-

haved much like enamel, whereas Paradigm MZ100 exhibited slightly higher

resin material loss, i,e. gentler on enamel than porcelains. [42]

26

1.6.6 Longevity

CEREC manufactured restorations have proven their success and longevity

through several studies. The �rst clinical trial was conducted by Mormann

et al. [80] who evaluated 94 CEREC I -manufactured Vitablocs Mark I in-

lays between September 1985 and August 1987 and reported two fractured

inlays. This low level of failure was repeatedly reproduced by a large number

of clinical studies. Isenberg et al. [54] observed 121 CEREC I -generated

inlays and reported a 94.2% success rate over three years. All the fractures

of the failed restorations (3 out of 121 inlays) occurred through the occlusal

isthmus and the thickness of ceramic at fracture site was less than 2 millime-

ters. In a clinical study over a 5-year period, Berg and Derand [11] reported

3 fractures in 115 Vitablocs Mark I inlays. A systemic review by Martin

and Jedynakiewicz undertook a comprehensive literature search from 1986

to 1997 to identify the survival rate of such restorations and their causes of

failure [70]. It determined that the mean survival rate of CEREC manufac-

tured restorations was 97.4% over a period of 4.2 years. Causes of failure

included fracture of ceramic, fracture of supporting tooth, post-operative

sensitivity, and wear of the adhesive. Vitablocs Mark II was used in most

of these clinical studies but ProCAD was used occasionally. Mormann and

Schug studied only Vitablocs Mark II and concluded that the success rate

was 95% in 5 years [82]. Brauner and Bieniek [18] reported a survival rate

of 88% for 238 Vitablocs Mark II inlays after 5.5 years of clinical service of

inlays. In a di�erent study, only one Vitabloc Mark II inlay fractured out

27

of 16 pairs of CAD/CAM inlays during an eight-year period [88]. During

the recall period, no tooth fractures were observed. A 90.4% survival rate

for 200 Vitabloc Mark II inlays was observed by Otto and De Nisco after

10 years of clinical service of inlays [87]. The failure was caused by ceramic

fractures in 8 cases and by tooth fracture in 3 cases. In an in-vivo study,

18 anterior Vitablocs Mark II crowns were compared to 18 anterior Ceramic

core (In-Ceram Spinell) crowns. Survival rate was determined to be 91.7%

for In-Ceram Spinell and 94.4% for Vitablocs Mark II over 2-5 years without

a statistically signi�cant di�erence [14]. This raised the interest in evalu-

ating monoceramic Vitablocs Mark II crowns clinically as an alternative to

ceramic-core crowns. Bindl et al. compared Vitablocs Mark II crowns in

regards to type of preparation and tooth type. They classi�ed type of prepa-

ration design to classic crown preparation (preparation wall height of at least

3 mm, 6-8 degree taper, and 1-1.2 mm shoulder), reduced crown preparation

(preparation wall height less than 3 mm), and endodontic crown preparation

(no clinical crown remaining which used only the pulp chamber for reten-

tion). In molars, the survival probability was 94.6% in classic crowns, 92.1

% in reduced crown, and 87.1 % in endodontic crowns. In premolars, the

reported survival probability was 97 % in classic crowns, 92.9 % in reduced

crown, and 68.8 % in endodontic crowns. A signi�cant di�erence was found

between the premolar classic crown and endodontic crown. The success of

the reduced crown was attributed to the use of the adhesive resin. They

concluded that CEREC-manufactured restorations had good prognosis for

28

classic and reduced crowns on both molars and premolars, and endodontic

crowns only for molars.

Another study evaluated 2,328 CEREC fabricated ceramic inlays and on-

lays in 794 patients and the survival rate was found to be 95.5% at 9 years

[90]. The majority of failure was caused by inlay fracture, tooth fracture,

tooth extraction, and replacement for occlusal reconstruction. In this study,

a successful prognosis was not a�ected by restoration size, tooth vitality,

treatment of caries profunda (CP), type of tooth treated, or whether the

restoration was placed in the maxilla or mandible. Reiss and Walther pub-

lished a series of articles on 1,011 CEREC-manufactured inlays and onlays

placed between 1987 and 1990 in 299 patients and evaluated them for up

to 18 years. They determined a survival probability of 95% after 5 years,

91.6% after 7 years, and 90.0% at 10 years [93, 94, 92]. Prognosis for inlays

in premolars (90%) was more favorable than those in molars (80 %). The lo-

cation of the restoration (maxillary or mandibular) and the number of tooth

surfaces restored did not signi�cantly in�uence the survivability of the inlay.

Sjogren et al. [108, 110, 111] published 3 reports on 66 Vitablocs Mark II

inlays from 1995 to 2004 and determined a survival probability of 89% after

10 years.

This survival rate is comparable to that of conventional IPS-Empress ce-

ramic restorations which was reported in a literature review to range from

96% at 4.5 years to 91% at 7 years for inlays and onlay and to range from

92% to 99% at 3-3.5 years for crowns [29]. In a separate literature review by

29

Hickel and Manhart, annual failure rates in posterior stress-bearing restora-

tions were reported to be: 0-7% for amalgam restorations, 0-9% for direct

composites, 1.4-14.4% for glass ionomers and derivatives, 0-11.8% for com-

posite inlays, 0-7.5% for ceramic restorations, 0-5.9% for cast gold inlays and

onlays, and 0-4.4% for CAD/CAM ceramic restorations [50]. While recurrent

decay was the primary cause of failure in direct restorations, fracture of the

restoration and tooth fracture were the the most frequent causes of failure in

indirect restorations.

In a clinical study, resin-based composite CAD/CAM inlays performed as

good as porcelain CAD/CAM inlays after three years of clinical observation

[38]. Fasbinder et al. also concluded that porcelain fracture was the primary

cause of failure for Vitablocs Mark II inlays, as two of the 40 Vitablocs Mark

II inlays and none of the Paradigm inlays fractured over the three year period

of study.

A low failure rate of CEREC-manufactured restorations was consistently

reported in the literature. This documents the reliability and the clinical

predictability of such restorations. Similar to conventional ceramic restora-

tions, the primary cause of failure of CEREC-manufactured restorations is

ceramic and tooth fractures. [50]

30

1.7 E�ects of porcelain surface pre-treatment

and luting cement on bond strength

Di�erent treatments of the adapted surface of the ceramic have been devel-

oped to improve the bonding strength to the resin luting cement [55, 77].

This includes mechanical roughening of the ceramic adapted surface with

a coarse diamond bur [2], air abrasion using Alumina particles [126], and

etching with hydro�uoric acid [114]. Each technique has some limitations.

Mechanical roughening could cause ceramic to chip which may a�ect the �t

of the restoration. Hydro�uoric acid is toxic, caustic, and extremely irritable

to the skin and lungs. [55]

Following the micro-roughening of the ceramic surface, a silane coupling

agent is applied which has bi-functional molecules capable of reacting with

both porcelain surface at one end and the resin luting cement at the other

end forming a chemical bond [33]. The use of this method has yielded a high

fracture resistance comparable to that of porcelain-fused to metal restora-

tions [27, 116, 55, 129, 56]. The fracture resistance is also a�ected by the

fabrication technique, the restoration's �nal surface �nish, and the type of

luting cement used [22, 58]. Chen et al. [22] concluded that oven-glazing

of ProCAD crowns resulted in signi�cantly higher strength and higher resis-

tance to cyclic loading than surface polishing.

Resin-based composites are the material of choice for adhesive luting for

all-porcelain restorations. Physical properties and wear behavior of �ne par-

31

ticle hybrid-type resin-based composites exceed those of other materials [62].

Enamel and dentin bonding was proven clinically acceptable when utilize

multi-step systems with separate primer and bonding agents, as it provides

a perfect internal seal with almost no hypersensitivity. [62]

32

Chapter 2

The Project

2.1 Rationale, Objectives and Hypothesis

2.1.1 Rationale

The demand for esthetic restorations has increased dramatically over the past

2 decades. The development of CAD/CAM technology has made it possible

to fabricate an indirect restoration at one appointment while the patient is

waiting. Several machinable materials are currently available to be used for

the CEREC chair-side restoration manufacturing. The recent introduction of

the composite resin blocks to be used for crowns has raised the question for

their strength and durability compared to ceramic blocks. The majority of

published CAD/CAM-manufactured restorations long-term clinical studies

were based on the CEREC I system. Hence, most of them addressed ceramic

inlays and onlays only and very few more recent studies focused on crowns

33

[36]. As a result, CEREC manufactured ceramic restorations were heavily

studied and their long term failure rate was reported to be low [68, 92, 111,

87, 88]. However, studies addressing CEREC manufactured composite resin

restorations are very limited. Hence, it is crucial to test the mechanical

properties such as fracture strength and fatigue resistance of these materials

in-vitro prior to conducting costly in-vivo studies.

Although laboratory studies are valuable to test the physical and mechan-

ical properties of the dental material, clinical studies are still mandatory to

test the clinical performance of the restoration in its actual setting in the

oral environment under normal function.

2.1.2 Objectives

The objectives of this study were:

1. To measure and compare fracture strength of crowns made with two dif-

ferent materials (ProCAD and Paradigm MZ100) utilizing CAD/CAM

CEREC-3D technology.

2. To investigate the e�ects of mechanical cyclic loading on the fatigue

resistance and fracture strength of CAD/CAM manufactured ceramic

(ProCAD) and composite resin (Paradigm MZ100) crowns.

2.1.3 Hypothesis

The null hypotheses are:

34

1. There is no di�erence between the fracture strength of CEREC manu-

factured ceramic and composite resin crowns.

2. Cyclic loading has no e�ect on fracture strength of crowns made with

the two di�erent materials (ceramic and composite resin).

3. There is no di�erence in fatigue resistance between CEREC manufac-

tured ceramic and composite resin crowns.

2.2 Materials and Methods

2.2.1 Sample Size Calculation

Using the following formula [86], the sample size was calculated at (α = 0.05)

and (β = 0.1):

n = 2[(Zα + Zβ)δ/∆]2

where:

n =sample size per group

Zα =standard score (Z value) corresponding to alpha error

Zβ =standard score (Z value) corresponding to beta error

δ =common standard deviation for both populations

∆ =di�erence in the mean that is considered to be clinically signi�cant

The values used for sample size calculation were based on a similar previ-

ous study [130] where the mean fracture strength of Vitabloc Mark II crowns

35

was 1272.65±109.06. A di�erence of 15% was set arbitrary for the di�erence

in mean between the samples to be considered clinically signi�cant.

n = 2[(1.96 + 1.28)[(109.06/1.15(1272.65) − 1272.65]2 = 6.85 (7 samples

per group)

Several studies comparing fracture strength of di�erent materials used a

sample size of 8-10 per group [130, 10, 9]. Hence, we selected n = 10 for the

sample size of our study.

2.2.2 Preparation

An arti�cial ivorine mandibular molar replica was prepared to receive an all-

ceramic crown following the preparation guidelines suggested for the CEREC

3D system [35]. The tooth received a reduction of 2 mm of the functional

cusp and 1.5 mm of the non-functional cusp. Minimal axial reduction was

1.2 mm and the gingival margin was a circumferential shoulder of at least 1

mm width. The lingual and facial surfaces were prepared in two planes and

all line angles were rounded to decrease stress concentration. The angle of

convergence was 12o. The prepared tooth replica was duplicated to form 40

replicas fabricated from highly �lled epoxy resin material (Viade Products

Inc., Camarillo, California, USA) (Figure 2.1 ). This material has a modulus

of elasticity equivalent to that of human dentin (12.9 GPa) and responds to

34% phosphoric acid etching by forming micro-roughness for bonding [85].

Using a caliper, the replicas' dimensions were measured occluso-gingivally,

facio-lingually, and mesio-distally to verify accuracy of reproduction (Table

36

Figure 2.1: The Epoxy resin replicas.

Table 2.1: Epoxy resin die replicas dimensionsDimension Mesio-distal Dimensions (mm) Bucco-lingual Dimensions (mm)

Cervical Occlusal Cervical Occlusal

Mean 6.88 6.34 7.51 6.89Standard deviation 0.02 0.01 .02 0.01

2.1 ). The CEREC 3D system (Sirona Dental Systems GmbH, Benshein,

Germany) was utilized to fabricate the crowns from two di�erent materials:

ProCAD (Ivoclar/Vivadent) and Paradigm MZ100 blocks (3M/ESPE). A

complete CEREC 3D system is available on site of the research laboratory.

2.2.3 Optical Impression and Restoration Designing

The Prepared teeth replicas were inserted into a dentoform with mesial and

distal adjacent ivorine teeth in place. Both the prepared and adjacent teeth

were covered by CEREC Imaging Liquid (Ploysorbate liquid; Vita Zahn-

fabrik, Bad Sackingen, Germany) to create a sticky surface. Then, the teeth

surface was covered with a uniform thin layer of optical re�ective powder (Ti-

37

tanium dioxide; Vita Zahnfabrik, Bad Sackingen, Germany). The CEREC

3D intra-oral camera was used to capture optical impressions of the pre-

pared teeth replicas. An arti�cial ivorine unprepared mandibular molar was

inserted in a dentoform with adjacent teeth and a preoperative occlusal im-

pression was captured. CEREC 3D (Serial number 04996) and CEREC 3D

software V3.10 (Sirona Dental Systems GmbH, Bensheim, Germany) were

used to design the crown utilizing the correlation mode and the crown cop-

ings.

2.2.4 Crown Fabrication

In this experiment, 20 ProCAD blocks and 20 Paradigm MZ100, shade A

3.5- Vitapan, 14 mm long were used to fabricate monolithic crowns (Figure

2.2 and Figure 2.3 ). According to manufacturer's instructions, burs were

replaced after milling each 20 crowns. The Paradigm MZ100 crowns were

polished using polishing discs and points. The ProCad crowns were glazed

using IPS Empress Universal Glaze Paste (Ivoclar/Vivadent) (Figure 2.4 )

(Table 2.2 and Table 2.3 ). Manufacturer's instructions were followed for

the �ring cycle. We �rst pre-dried the crown at 600oC for 6 minutes, then

increased the temperature at a rate of 58oC/min to 950oC and held it for 1

minute.

38

Figure 2.2: Paradigm MZ100 bloc in the milling chamber prior milling

Table 2.2: Physical Properties of some of the materials used in this study(Manufacturer's Data)

Material Basic Chemical Structure Modulus of Flexural FractureElasticity Strength Toughness(GPa) (MPa) (MPa)

Paradigm MZ100 Resin Composite with 85 wt% 15-20 (more 140-160 1.35ultra�ne zirconia-silica ceramic �exible moreparticles (0.6 µm), cured resilient)

ProCAD Leucite reinforced glass ceramic, 62-70 127-140 1.3crystallized [13]

Panavia F 2.0 Adhesive resin cement 9.6 79 -Epoxy Resin dies 12.9 - -

39

Figure 2.3: Paradigm MZ100 crown during milling.

Figure 2.4: Di�erent views of the resulting Composite Resin and Ceramiccrowns.

40

Table 2.3: The composition of some of the materials used in this study(Manufacturer's Data)

Material Composition

Paradigm MZ100 Polymer matrix of bisGMA (Bisphenol A diglycidyl ether dimethacrylate) and

TEGDMA (tri[ethylene glycol] dimethacrylate), Ultra�ne zirconia-silica ceramic

particles (0.6 µm)(85%)

ProCAD Silicon dioxide (SiO2) (59-63%), Aluminum oxide (Al2O3) (16-20%),

Potassium oxide (K2O) (10-14%), Sodium oxide (Na2O) (3.5-6.5%),

Boron oxide (B2O3) (0-1.0%), Barium oxide (BaO) (0-1.5%), Calcium

oxide (CaO) (0.5-2.5%), Cesium oxide (CeO2) (0-1.0%), Titanium oxide

(TiO2) (0-0.5%), Pigments (0.2-1.0%)

2.2.5 Cementation

Prior to cementation, each crown was placed in its respective tooth replica

to check for �t accuracy. A dental probe with 50 µm tip size was used to

check the marginal �t and any crown with a margin gap exceeding 50 µm

would have been excluded and remade [62]. No crowns were excluded in

this study. The internal surfaces of the ceramic crowns were treated with

9.6% hydro�uoric acid etching gel (Pulpdent Corporation, Watertown, MA,

USA) for 2 minutes [17]. The internal surfaces of the composite resin crowns

were sandblasted with 50 µm Aluminum oxide particles at 0.20 MPa pressure

for 5 seconds [8]. The etched/ sandblasted internal surface was rinsed with

water spray followed by 60 seconds of ultrasonic cleaning in distilled water

(Figure 2.5 ). Finally, it was dried using oil-free compressed air [8]. A silane

coupling agent (Silane, Pulpdent, Watertown, MA, USA) and resin cement

containing adhesive phosphate monomer (Clear�l SE bond, Kuraray Medical

Inc., Okayama, Japan) were applied to the cleaned internal surface of each

41

Figure 2.5: Ultrasonic cleaning system that was used to clean the crowns.

crown and air dried (according to manufacturer's recommendation).

The surfaces of teeth replicas were etched using 40% phosphoric acid

(3M/ESPE) for 1 minute, rinsed, and dried using oil-free compressed air

[130]. All crowns were cemented with dual-cured resin cement (Panavia F

2.0; Kuraray) (Figure 2.6 ) according to manufacturer's recommendations.

Equal amounts of ED Primer A&B were mixed and evenly applied to the

tooth surfaces of the replicas for 30 seconds then gently dried with compressed

air. This initiates the set of the cement. Equal amounts of paste A and B

were expressed from the syringe, mixed for 20 seconds, and then applied to

the internal surface of the crowns (Table 2.4 ). The crowns were inserted

in their corresponding replicas under static pressure of 2.2 kg for 5 minutes

42

Figure 2.6: Panavia F 2.0 Dual Cure Dental Adhesive Cement (KurarayMedical Inc.).

Table 2.4: Materials used in this studyMaterial Manufacturer LOT number

Paradigm MZ100 3M ESPE, St. Paul, MN, USA 20060718

ProCAD Ivoclar Vivadent, Schaan, Liechtenstein J01294

Panavia F 2.0 Kuraray Medical Inc., Okayama, Japan 51249

IPS Empress (Glaze and stain liquid) Ivoclar Vivadent, Schaan, Liechtenstein J06718

Clear�l SE Bond Kuraray Medical Inc., Okayama, Japan 00719A

Silane (Porcelain repair kit) Pulpdent Corporation, Watertown, MA, USA 070516

9.6% HF Etch Gel (Porcelain repair kit) Pulpdent Corporation, Watertown, MA, USA 070516

43

[89] (Figure 2.7 ). The extruded excess cement was removed and each surface

(Buccal, Lingual, and occlusal) was light-cured (Optilux 501, Kerr Demetron,

Danbury, CT, USA) for 40 seconds. A layer of Oxyguard II was applied to the

margins for 3 minutes according to the manufacturer's instructions. Using

a caliper, the occluso-apical distance of each tooth was recorded before and

after cementation to verify the degree of seating (Figure 2.8 ). Any tooth

with more than 50 µm increase in that dimension as a result of cementation

would have been excluded and remade [130]. No crowns were excluded in

this study due to seating error. One hour following cementation, the crowns

were stored in distilled water at a temperature of 37oC for 1 week (Figure

2.9 , Figure 2.10 , and Figure 2.11 ). [10]

2.2.6 Cyclic Fatigue and Fracture Test

Each group of 20 crowns was randomly divided into two equal groups. One

half was loaded in the Instron machine (Instron, Canton, Mass) , a hydraulic

driven universal testing machine, using a cross head speed of 1 mm/min

(Figure 2.12 ). The load was applied along the long axis of the replicas

with hardened steel bar centered in the central groove (Figure 2.13 ). The

other half was subjected to mechanical cyclic loading in distilled water at

room temperature. The cyclic load ranged from 50 N to 600 N for 500,000

cycles at a frequency of 20 Hz (Figure 2.14 ). The Instron machine was

adjusted to stop if the deformation increased to more than 0.15 mm during

the mechanical cyclic loading. Those crowns that did not fracture during the

44

Figure 2.7: Crowns inserted in their corresponding replicas under static pres-sure of 2.2 kg.

45

Figure 2.8: Crown complete seating was veri�ed using a caliper.

cyclic loading were fractured according to the load-to-failure parameters.

2.2.7 Statistical Analysis

Mean direct fracture load was analyzed using Wilcoxon rank sum test (less

sensitive to non-normal data compared to Student's T test). Failure during

cyclic loading was tested with the log-rank test for comparing Kaplan-Meier

survival curves. All statistical tests were two-tailed and the level of signif-

icance was set at alpha=0.05. The estimate power of this study was 96%

based on a 15% e�ect size between materials. All tests were preformed using

the R statistical package (www.r-project.org).

46

Figure 2.9: Composite Resin (Paradigm MZ100) specimens.

47

Figure 2.10: Ceramic (ProCAD) specimens.

48

Figure 2.11: Comparison between the resulting Composite Resin and theceramic crowns.

2.3 Results

2.3.1 Direct Loading Fracture Test Results

During the direct loading fracture test, all crowns exhibited catastrophic

fracture in the buccal-lingual plane direction (Figure 2.15 ). The proximal

half of each crown was retained by the adhesive cement while the other

proximal half was completely dislodged. In one ceramic specimen, the crown

fractured in half while both parts were still retained by the adhesive cement

which caused the tooth replica to split down the middle (Figure 2.16 ).

The mean fracture loads for the composite resin and the ceramic crowns

in N were 1, 078.64 ± 283.36 and 972.28 ± 256.33, respectively. No statisti-

cally signi�cant di�erence in mean fracture load was found between the two

materials by Wilcoxon rank sum test (p = 0.42). Figure 2.17 summarizes the

49

Figure 2.12: The Instron machine was used to apply load on the crowns.

50

Figure 2.13: The load was applied along the long axis of the replicas with ahardened steel bar centered at the central groove.

51

Figure 2.14: Mechanical Cyclic Loading Machine.

52

Figure 2.15: Fracture patterns following the direct fracture test.

Figure 2.16: Fracture Pattern of one of the Ceramic Crowns where the crownremained cemented and the tooth replica was split in half.

53

Figure 2.17: Box-plot of Fracture load test before and after cyclic loading(fatigue).

Paradigm MZ100 ProCAD

500

1000

1500

2000

Non−fatigued

n=10 n=10

p=0.42

Paradigm MZ100 ProCAD

500

1000

1500

2000

Fatigued

n=9 n=3

54

Table 2.5: Changes in the fatigued crownsGroup N No evidence of fracture Fractured Cracked

and/or cracks

Paradigm MZ100 (fatigued) 10 10 (100%) 0 0ProCAD (fatigued) 10 0 7 3

results in graphical format. Since no statistically signi�cant di�erence was

found between the non-fatigued fracture strength of the two materials, the

�rst null hypothesis cannot be rejected.

2.3.2 Compressive Cyclic Loading Test Results

All the composite resin crowns survived the compressive cyclic loading test.

However, only 30% of the ceramic crowns survived that test (Table 2.5 ). The

ceramic crowns fractured at a mean of 61,103 cycles which is equivalent to 1.2

years of function (Figure 2.18 ). The di�erence in survival of the two materials

was statistically signi�cant, p=0.0012. Visible cracks were evident on those

3 surviving ceramic crowns. The cracks started to spontaneously propagate

following the compressive cyclic loading test and during the following loading

fracture test.

Following the compressive cyclic loading test, the mean fracture load for

the composite resin crowns was 1,517.56 N ±231.21 N. The mean fracture

load for the remaining 3 ceramic crowns was found to be 1,023.33 N ±271.56

N (Table 2.6 ).

The fracture strength of the composite resin crown specimens has signif-

55

Figure 2.18: Kaplan Meier survival curves of failure during cyclic loading(fatigue) test of Paradigm MZ100 and ProCAD.

Loading cycles X 1000

Pro

port

ion

inta

ct

0 60 120 180 240 300 360 420 480

0.0

0.2

0.4

0.6

0.8

1.0

10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 Paradigm MZ100

10 6 4 4 4 4 4 4 4 4 4 3 3 3 3 3 3 ProCAD

Paradigm MZ100

ProCAD

Log−rank test p−value=0.0012

Table 2.6: Fracture loads (in N) of the four test groupsGroup N Median Mean SD Maximum Minimum

Paradigm MZ100 (Not fatigued) 10 990.6 1078.6 283.36 1581 702.4ProCAD (Not fatigued) 10 955.7 972.3 256.33 1414 595.4

Paradigm MZ100 (fatigued) 9 1508 1517.56 231.21 1840 1174ProCAD (fatigued) * 3 1017 1023.33 271.56 1298 755

* Crowns that fractured during the cyclic loading (fatigue) test were omitted.

56

icantly increased following the compressive cyclic loading while the fracture

strength of the three surviving ceramic crown specimens demonstrated no sta-

tistically signi�cant di�erence to the non-fatigued specimens. Theoretically,

this leads to the rejection of the second null hypothesis for the composite

resin crown specimens. The second null hypothesis cannot be rejected for

the ceramic crown specimens. However, these results should be interpreted

cautiously since the conditions of the specimens were not standardized be-

tween pre- and post-fatiguing. While the specimens used for direct loading

fracture test were tested one week following cementation, those specimens

used for the compressive cyclic loading test were tested 3 to 4 months fol-

lowing cementation. In addition, the three surviving fatigued-ceramic crown

specimens demonstrated visible cracks prior the fracture test. This may have

resulted in inaccuracy of the second fracture test values. Since the di�erence

in survival of the ceramic and composite resin materials was statistically

signi�cant, the third null hypothesis can be rejected.

2.4 Discussion

2.4.1 Signi�cance of this research

Improvement in adhesive dentistry has made it possible for all-ceramic restora-

tions to be clinically successful. This encouraged the development of new

restorative materials to be used for anterior and posterior crowns, such as

ceramic and composite resin blocks that are used with CAD/CAM technol-

57

ogy. Although these blocks have been improving throughout the past several

years, a major disadvantage in this system is the adaptation of the restoration

and the surface smoothness following the milling procedure which may lead

to failure due to crack propagation and/or poor marginal adaptation. How-

ever, we believe that the new CEREC 3D software will prevent or minimize

such problems.

This study provides essential information regarding the in-vitro survival

of all-ceramic and composite crowns manufactured by CEREC 3D. The most

promising material will be selected for subsequent clinical trial supported by

industry and/or external funding agencies.

2.4.2 Fracture strength and fatigue resistance of Paradigm

MZ100

Since the introduction of composite resin blocks to be used for CEREC-3D

full-coverage crowns was fairly recent, studies addressing fracture strength

and fatigue resistance of composite resin crowns are very limited. For com-

posite resin crowns, Attia et al. [8] reported that the fracture load in N

for Paradigm MZ100 crowns was of 827.1 (86.3), 914.7 (131.7), and 955.9

(130.6) when used zinc phosphate, Fuji CEM, and RelyX ARC, respectively.

After subjecting his crowns to 3500 thermal cycles (58oC/4oC) and 600,000

masticatory cycles at 1.2 Hz ranging from 0 N to 49 N, the resulting frac-

ture loads in N of those fatigued composite resin crowns were reported to

58

be 772.3 (134.7), 923.6 (153.5), and 929.1 (148.5) when used zinc phosphate,

Fuji CEM, and RelyX ARC, respectively.

Tsitrou et al. [120] reported a fracture load of 1682 N (315 N) for

Paradigm MZ100 composite resin crowns and a fracture load of 1751 N (338

N) for minimally prepared composite resin crowns. Tyan et al. [122] found

that the fracture load for composite resin crowns was 2251N (714N) and

the fatigued fracture load of those crowns was 1464 N (266 N). In a di�er-

ent study [3], the mean fracture load for Paradigm MZ100 composite resin

crowns without fatigue and with 50,000 load cycles was stated to be 1680

N and 1330 N, respectively. In this study, the mean fracture load for of

Paradigm MZ100 crowns was 1,078.64 N (283.36 N) which falls within the

reported range of the mean fracture strengths (827.1 N to 2251 N).

All of the Paradigm MZ100 crowns subjected to cyclic loading survived

the test without any evidence of cracking. None of the previous limited stud-

ies reported any failure with Paradigm MZ100 crowns during cyclic loading

testing [8, 120, 122, 3]. However, the testing criteria in this study have

subjected the specimens to the most extreme testing conditions among the

previously reported studies.

The mean fracture load of the fatigued composite resin crowns was found

to be 1517.56 N (231.21 N). Most of the literature has reported a decrease in

fracture loads following the cyclic loading test [3, 8]. However, these studies

used di�erent cements than the one used in this study under di�erent cyclic

loading conditions. Attia and Kern [10] found that cyclic loading did not

59

decrease the median fracture load of crowns cemented with Panavia F. In

our study, the mean fracture load increased signi�cantly (p = 0.003). This

can be attributed to one or more of the following:

� The cyclic loading test high frequency (20 Hz) and load generated high

temperature that caused a higher degree of conversion in the resin ce-

ment and hence a more complete resin cure. However, the test was

conducted under water which would have acted as coolant and pre-

vented temperature rise.

� The specimens that received cyclic compressive loading were stored for

3 months in water prior testing due to unavailability of the Instron

machine for that period. This may have provided the resin cement

with more time to fully mature.

� The operator who conducted the direct fracture test is di�erent than

that who performed the cyclic loading and the following fracture test.

Inter-operator variability may have contributed to such results.

2.4.3 Fracture strength and fatigue resistance of Pro-

CAD

Attia and Kern [9] measured fracture load of ProCAD crowns to be 715.9

N (105.2 N) when used hydro�uoric acid etching and silane coupling agent

(Mirage ABC silane); and 708.4 N (108.7 N) when used phosphoric acid clean-

ing and Porcelain Liner-M primer. In a di�erent study, Attia and Kern [10]

60

recorded the fracture resistance of ProCAD crowns cemented with Panavia F

to be 960.2 N (211.8 N) and 809.2 N (96.4 N) without and with cyclic load-

ing, respectively. No statistically signi�cant di�erence was found between

the non-fatigued and fatigued ProCAD crowns when Panavia F was used for

cementation. Chen et al. [22] reported a fracture load of 2120 N (231 N) and

2254 N (186 N) for polished and oven glazed ProCAD crowns, respectively.

For the fatigued ProCAD crowns, Chen at al. reported a fracture load of 1613

(296) N and 2033 (413) N for polished and oven glazed crowns, respectively.

They found that prior cyclic loading decreased the fracture resistance of all

tested crowns signi�cantly. They also concluded that oven-glazing ProCAD

crowns improved their strength and fatigue properties signi�cantly. Stappert

et al. [115] studied partial coverage ProCAD crowns and recorded a mean

fracture strength of 2134 N (333 N) after subjecting all of the crowns to a

load of 49 N for 1.2 million masticatory cycles at 1.2 Hz. All of their Pro-

CAD specimens survived mastication simulation. Tyan et al. [122] reported

an initial load of 1250 N (255 N) and a fatigue load of 798 N (178 N) for

ProCAD crowns. Those crowns were subjected to 10,000 fatigue cycles with

a load that varied between the selected maximum value (40% of minimum

fracture load = 500 N) and a minimum value of about 1/3 of the maximum

load (166 N).

The recorded fracture resistance in this study for the ProCAD crowns

was 972.28 (256.33) N which falls within the literature reported means range

(715.9 N to 2254 N). The variability in the reported fracture loads can be

61

attributed to the use of di�erent cements, bonding techniques, abutment

types, and testing modalities.

As opposed to the previous studies that reported no ProCAD crown fail-

ure during the cyclic loading test, we experienced catastrophic fracture of

70 % and cracking of the remaining 30 % of our ProCAD specimens. While

most of the previous studies [9, 10, 115] used a load of 0-49 N during the

compressive cyclic loading test, our study used a range of 50-600 N when

fatiguing the specimens. This load falls into the highest range of forces that

crowns might be subjected to in the molar area [7]. For a given material,

the failure load vs. the no. of cycles required to fracture that material is

represented in the following graph (Figure 2.19 ). If the load applied during

the cyclic loading test is too low, the crown may never undergo failure as the

number of cycles required for the crown to break reaches in�nity. In those

fore-mentioned studies, a maximum load of 49 N might have been too low to

cause any catastrophic fracture. Tyan [122] used a cyclic load that is 13.3-40

% of the minimal load to fracture which corresponds to 166-500 N. However,

the number of cycles used was only 10,000 cycle which is 2% of the number of

cycles used in this study and believed to be too low to cause any catastrophic

fracture of the restorations.

62

Figure 2.19: Number of cycles required for restoration failure graph.

63

2.4.4 Fracture strength and fatigue resistance compar-

ison

In our study, Paradigm MZ100 crowns performed as well as ProCAD crowns

when subjected to direct fracture loads. There was no signi�cant di�erence

between both crowns in terms of direct fracture strength. All of the pre-

vious studies comparing Paradigm MZ100 crowns to CEREC manufactured

ceramic crowns have reported no signi�cant di�erence between the mean frac-

ture loads of both crown systems without cyclic loading [8, 122]. Similarly,

the previous studies have reported no failure of composite resin crowns when

subjected to compressive cyclic loading. This is consistent with the �ndings

of this study.

Although most clinicians assume that all-ceramic crowns have better me-

chanical properties when compared to composite resin crowns, we reported no

signi�cant di�erence in the mean direct fracture loads of Paradigm MZ100

and ProCAD crowns. This may be attributed to the optimized manufac-

turing conditions that CAD-CAM materials undergo eliminating the risk of

voids and volume defects [10, 9]. It may also be due to the fact that com-

posite resin crowns have improved elastic properties when compared to all

ceramic crowns [64]. As a result, during the process of loading, the composite

resin crowns may exhibit higher resiliency with more absorption of load, and

thus, the fracture load is increased [8]. Most published literature compar-

ing composite resin to ceramic various restorations reported similar �ndings

64

[25, 20, 19, 105]. Gracis et al. [45] concluded that composite resin reduced

the impact forces by 57% more than porcelain.

All of our composite resin crowns survived the rigorous compressive cyclic

loading test while only 30 % of the ceramic crowns survived the test. The

surviving 30 % ceramic crowns demonstrated visible cracks. When the milled

surfaces of the crowns were examined under 500 X magni�cation with the

Scanning Electron Microscope (SEM), a di�erence in the pattern of material

removal was observed. In the composite resin specimen, �ne grooves can

be seen on the milled surface indicating a plastic deformation mechanism

of material removal (Figure 2.20 ). However, ceramic specimens demon-

strated a surface typical of fracture mechanism material removal (Figure

2.21 ). The 3M Paradigm MZ100 technical product pro�le reported similar

�ndings. When the margins of milled crowns examined under 60 X SEM

magni�cation, the same broad contours and peaks can be seen in both crown

types which is re�ective of the margins shape. However, the margin edge

appears more sharply de�ned in the Paradigm MZ100 crown compared to

the ProCAD crown which also show chipping (Figure 2.22 ). While large

milling grooves matching the outer shape of the cutting bur can be seen in

both, smaller grooves corresponding to individual diamond grains can only

be seen on the Paradigm MZ100 material.

The fracture mechanism of material removal in the ProCAD crowns may

have introduced micro-cracks in the material during milling. Cyclic loading,

especially under wet conditions, results in propagation of micro-cracks that

65

Figure 2.20: SEM (500 X) image of milled surface of the Paradigm MZ100crown showing �ne grooves which is indicative of a plastic deformation mech-anism of material removal.

66

Figure 2.21: SEM (500 X) image of the milled surface of the ceramic crownshowing a fracture mechanism of material removal.

Figure 2.22: SEM (60 X) comparison between the margin edge of the com-posite resin vs. the ceramic crown.

67

may have initiated from the milling process. These cracks propagate to form

an advancing �ssure that weakens the crown [22, 113] and hence leads to

catastrophic fracture during cyclic loading.

In addition, Zohairy et al. [132] stated that �CAD/CAM restorations

fabricated from resin blocks may have advantage over ceramic blocks with

regards to the higher bond strength with resin cements�. This stronger bond

strength of composite resin crowns may have contributed to the survival of

those crowns when underwent the compressive cyclic loading test.

2.4.5 Clinical relevance of study design

In an in-vitro study, it is crucial to mimic the actual oral conditions as

closely as possible since this will a�ect the performance and the outcome of

the tested material. The tested in-vitro crowns should ideally be subjected

to the same forces and force vectors along with thermal changes, the aqueous

environment and pH variations that a crown may undergo during its clini-

cal function. Hence, the crown should be tested in an environment that is

clinically relevant. [8]

In our study, the master die was prepared according to the CEREC 3D

recommendation to receive an all-ceramic crown. This preparation was du-

plicated using epoxy resin material which has a modulus of elasticity similar

to that of dentin (12.9 MPa) and responds to 34% phosphoric acid etching

by forming micro-roughness for bonding. An arti�cial material was used

instead of natural tooth to standardize the preparation dimensions since dif-

68

ferent preparation criteria, such as position of �nish line and the amount

of tooth structure removed, may in�uence the fracture load of these non-

metallic crowns [8]. This enabled us to compare the di�erent crown materials

fracture loads and fatigue resistance while keeping the shape/size variability

associated with natural teeth constant. When the dimensions of the epoxy

resin dies were measured, a low standard deviation of 10-20 µm was recorded

(Table 2.1). A pilot study conducted by Zahran et al. [130] concluded that

these epoxy resin dies were susceptible to phosphoric acid-etching and that

they bonded well to resin cements. Hence, the risk of these epoxy resin die

replicas not bonding to the adhesive resin cement is minimized.

In this study, the crowns were fabricated according to manufacturer's

instructions as in the clinical situation to produce optimum results. The

Paradigm MZ100 and the ProCAD crowns' dimensions were standardized by

using the same preoperative optical impression to mill them. The Paradigm

MZ100 crowns were polished and the ProCAD crowns were glazed to produce

maximum strength according to manufacturer's recommendation.

It was crucial in this study to obtain an optimum bond between the

crown material and the adhesive resin cement to ensure that any di�erences

in fracture strength or fatigue resistance reported in this study were not

due to the cementation technique used. The �tted surfaces of the Paradigm

MZ100 crowns were sandblasted using Al2O3 particles and the �tted surfaces

of the ProCAD crowns were etched. Both crown materials were treated with

a silane coupling agent (Silane, Pulpdent, Watertown, MA, USA) and resin

69

cement containing adhesive phosphate monomer (Clear�l SE bond, Kuraray

Medical Inc., Okayama, Japan). All crowns were cemented with the same

dual-curing phosphate-modi�ed resin cement Panavia F 2.0. No failure was

noted to occur between the cement and the crown material.

In the oral environment, non-metallic crowns fail as a result of slow crack

propagation that occurs due to fatigue under masticatory stress in a wet envi-

ronment. The crown specimens in this study were fatigued using a maximum

load of 600 N which corresponds to a high level of force a crown might be sub-

jected to in the posterior region. It was reported that the mean masticatory

forces during mastication and swallowing in humans to be approximately 40

N. However, the mean maximum posterior masticatory forces may vary from

150 N to 665 N [129, 8]. In our study, the crowns were loaded for 500,000

cycles which represents 10 years of clinical normal mastication function [125],

or about half a year of continuous bruxism as suggested by Kelly. [58]

The crack propagation in the oral cavity is accelerated by the presence of

saliva. All the compressive cyclic loading in this study was performed in wet

environment to simulate the hydrolytic e�ect of saliva on the crown mate-

rial �static fatigue: a stress-dependent chemical reaction between water and

surface �aws that causes the �aws to grow to a critical dimension, allowing

spontaneous crack propagation�. [10, 58]

Kelly [58] recommended several criteria for in-vitro failure testing for all-

ceramic crowns to be clinically relevant. Some of these recommendations

were followed in this study which includes using die material with elastic

70

modulus equivalent to that of dentin, preparing the dies according to clin-

ical recommendations, using crowns with dimensions similar to the clinical

situation, and using a reliable, commonly used luting cement.

2.4.6 Limitations of study design

1. Teeth replicas were used instead of natural teeth to allow for stan-

dardizing teeth sizes, preparation design, and di�erent factors a�ecting

the dentin modulus of elasticity. The dissimilarity in the coe�cient

of thermal expansion between the teeth replicas and the monolithic

crowns prevented thermocycling of the non-metallic crowns. In addi-

tion, epoxy resin is not identical to dentine in terms of cellular struc-

ture, bonding mechanism to the adhesive resin, and its physical and

mechanical properties.

2. In this study, all the forces during the cyclic loading were applied along

the vertical axes of the teeth replicas while ignoring all the other force

directions due to limitations in the loading machine. Crowns are sub-

jected to non-axial forces that is caused by the elliptical jaw movement

and the teeth contact on the cusps' inclines in the clinical environment

[46]. However, the majority of masticatory forces are vertical along the

long axis of the teeth in the molar area [46]. Due to the large damage

in�ected by the lateral forces [100], the clinical implication of the re-

sults of this in-vitro study should be limited only to the vertical loading

71

situation.

3. During the cyclic loading test, the use of high frequency of (20 Hz) com-

pared to a more clinically relevant frequency of (1-2 Hz) was inevitable

due to budget limitations. This high frequency may have lead to more

heat generation and may have not provided enough time for stress re-

laxation. In a study by Kelly et al. [60], a frequency of 20 Hz was

used for cyclic loading of leucite-reinforced all-ceramic crowns using a

staircase approach for loads between 200-800 N with a 100 N step size

for 1,000,000 cycles in water. They were able to record fracture loads

within reasonable clinical relevance. The high frequency used in this

study should not be a concern when comparing the two tested material

since both materials were fatigued under identical testing conditions.

4. The use of a stainless steel rounded tip rod would have increased the

contact pressure in the crown specimen compared to the clinical contact

pressure. This contact pressure is a�ected by the ratio of the elastic

modulus of the ceramic to that of the loading steel rod, and by the

radius of the tip of that steel rod. To overcome such limitation, we could

have used a loading ball with a modulus of elasticity lower than that

of the stainless steel rod to get a clinically relevant contact pressure.

However, the increase in the contact pressure due to using the stainless

steel rod with high modulus of elasticity and small diameter was applied

to both crown material specimens and should not alter the comparison

72

between them.

5. It is expected that crack initiation occurs at lower forces compared to

those needed for complete fracture. If a crack initiates in a crown, there

is a high possibility that the crack will propagate leading to catastrophic

fracture. In this study, the load needed to initiate cracking was not

recorded because the crack initiation was not accompanied by any drop

in the load since the loading piston was completely supported by the

intact crown. In this study, fracture loads were recorded when a sharp

drop in the load-displacement curve occurred. However, the load at

catastrophic fracture in not an adequate indicator of the load that the

material would withstand prior to crack initiation.

2.4.7 Future projects

According to the results of this study, the following need further investiga-

tions:

� In-vitro fracture strength and fatigue resistance of Paradigm MZ100

crowns compared to the gold-standard porcelain-fused-to-metal crowns.

This will provide signi�cant information regarding the strength and

durability of Paradigm MZ100 crowns compared to the clinically proven

strength and durability of porcelain-fused-to-metal crowns prior to con-

ducting expensive and time consuming clinical trials.

� If the Paradigm MZ100 crowns showed encouraging results in the pre-

73

vious in-vitro study, a clinical investigation of those resin composite

crowns should be conducted to test their durability and longevity un-

der the conditions of the oral environment and function.

2.5 Conclusions

Within the limitations of this in-vitro study, the following can be concluded:

1. There was no statistically-signi�cant di�erence between the in-vitro

mean compressive fracture loads of the resin composite (Paradigm

MZ100) and ceramic (ProCAD) molar crowns. However, this �nd-

ing should be cautiously interpreted since the two materials behaved

di�erently under conditions of cyclic loading.

2. Resin composite crowns sustained compressive cycling load without

any fractures while 70% of the ceramic crown specimens underwent

catastrophic fracture.

3. These in-vitro results suggest that resin composite molar crowns (Paradigm

MZ100) milled with CEREC-3D would be expected to resist fracture

well under forces of mastication.

74

Bibliography

[1] B. Abozenada, R. Pober, and R. Giordano. In-vitro wear of restorative

dental materials. In IADR Abstract no.1693, 2002.

[2] M. Aida, T. Hayakawa, and K. Mizukawa. Adhesion of composite to

porcelain with various surface conditions. The Journal of Prosthetic

Dentistry, 73(5):464�470, May 1995.

[3] K. AL-Gergawi, R. Giordano, and R. Pober. Cyclic fatigue of cad-cam

milled full thickness and layered crowns. In IADR Abstract no. 1367,

2004.

[4] K. J. Anusavice et al. Phillips' science of dental materials. W. B.

Saunders, Philadelphia, PA, 10th edition, 1996. pg. 585.

[5] K. J. Anusavice et al. Phillips' science of dental materials. W. B.

Saunders, Philadelphia, PA, 10th edition, 1996. pg. 607-608.

[6] K. J. Anusavice et al. Phillips' science of dental materials. W. B.

Saunders, Philadelphia, PA, 10th edition, 1996. pg. 593.

75

[7] K. J. Anusavice et al. Phillips' science of dental materials. W. B.

Saunders, Philadelphia, PA, 10th edition, 1996. pg. 66.

[8] A. Attia, K. M. Abdelaziz, S. Freitag, and M. Kern. Fracture load

of composite resin and feldspathic all-ceramic cad/cam crowns. The

Journal of Prosthetic Dentistry, 95(2):117�123, Feb 2006.

[9] A. Attia and M. Kern. Fracture strength of all-ceramic crowns luted

using two bonding methods. The Journal of Prosthetic Dentistry,

91(3):247�252, Mar 2004.

[10] A. Attia and M. Kern. In�uence of cyclic loading and luting agents

on the fracture load of two all-ceramic crown systems. The Journal of

Prosthetic Dentistry, 92(6):551�556, Dec 2004.

[11] N. G. Berg and T. Derand. A 5-year evaluation of ceramic inlays

(cerec). Swedish Dental Journal, 21(4):121�127, 1997.

[12] K. W. Bieniek and A. Peters. Clinical evaluation of ceramic restora-

tions machined by cad/cam technique. In W. H. Mormann, editor,

International symposium on computer restorations: state of the art of

the CEREC-method, pages 327�335. Quintessence, Chicago, 1991.

[13] A. Bindl, H. Luthy, andW. H. Mormann. Strength and fracture pattern

of monolithic cad/cam-generated posterior crowns. Dental Materials,

22(1):29�36, Jan 2006.

76

[14] A. Bindl and W. H. Morman. Survival rate of mono-ceramic and

ceramic-core cad/cam-generated anterior crowns over 2-5 years. Eu-

ropean Journal of Oral Sciences, 112(2):197�204, Apr 2004.

[15] A. Bindl and W. H. Mormann. Clinical and sem evaluation of all-

ceramic chair-side cad/cam-generated partial crowns. European Jour-

nal of Oral Sciences, 111(2):163�169, Apr 2003.

[16] M. B. Blatz. Long-term clinical success of all-ceramic posterior restora-

tions. Quintessence International, 33(6):415�426, Jun 2002.

[17] M. B. Blatz, A. Sadan, and M. Kern. Resin-ceramic bonding: a review

of the literature. The Journal of Prosthetic Dentistry, 89(3):268�274,

Mar 2003.

[18] A. W. Brauner and K. W. Bieniek. Seven years of clinical experience

with the cerec inlay system. In W. H. Mormann, editor, CAD/CIM

in aesthetic dentistry: CEREC 10 year anniversary symposium, pages

217�228. Quintessence, Chicago, 1996.

[19] B. D. Bremer andW. Geurtsen. Molar fracture resistance after adhesive

restoration with ceramic inlays or resin-based composites. American

Journal of Dentistry, 14(4):216�220, Aug 2001.

[20] P. A. Brunton, P. Cattell, F. J. Burke, and N. H. Wilson. Fracture

resistance of teeth restored with onlays of three contemporary tooth-

77

colored resin-bonded restorative materials. The Journal of Prosthetic

Dentistry, 82(2):167�171, Aug 1999.

[21] A. Cerutti, G. Venturi, and P. L. Sapelli. Seven year clinical evaluation

of 109 cad/cam inlays (abstract 2249). Journal of Dental Research, 77

(special issue B):913, 1998.

[22] H. Y. Chen, R. Hickel, J. C. Setcos, and K. H. Kunzelmann. E�ects

of surface �nish and fatigue testing on the fracture strength of cad-

cam and pressed-ceramic crowns. The Journal of Prosthetic Dentistry,

82(4):468�475, Oct 1999.

[23] L. Cho, J. Choi, Y. J. Yi, and C. J. Park. E�ect of �nish line vari-

ants on marginal accuracy and fracture strength of ceramic optimized

polymer/�ber-reinforced composite crowns. The Journal of Prosthetic

Dentistry, 91(6):554�560, Jun 2004.

[24] G. J. Christensen. Computerized restorative dentistry. state of the art.

Journal of the American Dental Association (1939), 132(9):1301�1303,

Sep 2001.

[25] P. H. P. Dalpino, C. E. Francischone, A. Ishikiriama, and E. B. Franco.

Fracture resistance of teeth directly and indirectly restored with com-

posite resin and indirectly restored with ceramic materials. American

Journal of Dentistry, 15(6):389�394, Dec 2002.

78

[26] H. Denissen, A. Dozic, J. van der Zel, and M. van Waas. Marginal �t

and short-term clinical performance of porcelain-veneered cicero, cerec,

and procera onlays. The Journal of Prosthetic Dentistry, 84(5):506�

513, Nov 2000.

[27] A. M. Diaz-Arnold, M. A. Vargas, and D. R. Haselton. Current status

of luting agents for �xed prosthodontics. The Journal of Prosthetic

Dentistry, 81(2):135�141, Feb 1999.

[28] J. L. Drummond. Ceramic behavior under di�erent environmental con-

ditions. dental materials in vivo: Aging and related phenomena. In

G. Eliades, T. Eliades, W. A. Brantley, and D. C. Watts, editors, Den-

tal Materials In Vivo: Aging and Related Phenomena, pages 35�46.

Quintessence, 2003.

[29] O. El-Mowafy and J. Brochu. Longevity and clinical performance of

ips-empress ceramic restorations�a literature review. Journal of the

Canadian Dental Association, 68(4):233�237, Apr 2002.

[30] D. Estafan, F. Dussetschleger, C. Agosta, and S. Reich. Scanning

electron microscope evaluation of cerec ii and cerec iii inlays. General

Dentistry, 51(5):450�454, Sep-Oct 2003.

[31] A. G. Evans. Slow crack growth in brittle materials under dynamic

loading conditions. International Journal of Fracture, 10(2):251�259,

1974.

79

[32] A. G. Evans and H. Johnson. The fracture stress and its dependence on

slow crack growth. Journal of Materials Science, 10(2):214�222, 1975.

[33] P. L. Fan. Porcelain repair materials. council on dental materials, in-

struments and equipment. Journal of the American Dental Association,

122(8):124, 126, 128�124, 126, 130, Aug 1991.

[34] D. J. Fasbinder. Restorative material options for cad/cam restorations.

Compendium of Continuing Education in Dentistry, 23(10):911�924,

Oct 2002.

[35] D. J. Fasbinder. CEREC 3D Operator's manual, volume 2. Cereconline

(Cited 2005 Apr 3), 2003.

[36] D. J. Fasbinder. Clinical performance of chairside cad/cam restora-

tions. Journal of the American Dental Association, 137 Suppl:22S�31S,

Sep 2006.

[37] D. J. Fasbinder. Predictable cerec occlusal relationships. In W. H.

Mormann, editor, State of the Art of CAD/CAM Restorations: 20

Years of CEREC, pages 93�100. Quintessence, 2006.

[38] D. J. Fasbinder, J. B. Dennison, D. R. Heys, and K. Lampe. The

clinical performance of cad/cam-generated composite inlays. Journal

of the American Dental Association, 136(12):1714�1723, Dec 2005.

[39] D. J. Fasbinder, K. Lampe, J. B. Dennison, M. C. Peters, and Ne-

matollahi K. Three year clinical performance of cad/cim generated

80

ceramic onlays (abstract 1883). Journal of Dental Research, 80 (spe-

cial issue):271, 2001.

[40] D. J. Fasbinder, K. Lampe, J. B. Dennison, M. C. Peters, and K. Ne-

matollahi. Clinical performance of CAD/CIM generated ceramic onlays

(abstract 2711). Journal of Dental Research, 78, 1999.

[41] J. F. Jr. Fowler. Allergic contact dermatitis to metals. American

Journal of Contact Dermatitis, 1:212�223, 1990.

[42] R. Giordano. Materials for chairside cad/cam-produced restorations.

Journal of the American Dental Association (1939), 137 Suppl:14S�

21S, Sep 2006.

[43] R. Giordano, K. Kanchanatawewat, P. Asvanund, and D. Nathanson.

Flexural strength of ceramics for celay restorations (abstract 860).

Journal of Dental Research, 75:125, 1996.

[44] S. Gladys, B. Van Meerbeek, S. Inokoshi, G. Willems, M. Braem,

P. Lambrechts, and G. Vanherle. Clinical and semiquantitative

marginal analysis of four tooth-coloured inlay systems at 3 years. Jour-

nal of Dentistry, 23(6):329�338, Dec 1995.

[45] S. E. Gracis, J. I. Nicholls, J. D. Chalupnik, and R. A. Yuodelis. Shock-

absorbing behavior of �ve restorative materials used on implants. The

International Journal of Prosthodontics, 4(3):282�291, 1991.

81

[46] H. Graf. Occlusal forces during function. In N. H. Rowe, editor, Pro-

ceedings of symposium on Occlusion: Research in Form and Function,

pages 90�111. Ann Arbor, Michigan, University of Michigan Press,

September 1975.

[47] L. H. He and M. V. Swain. Nanoindentation derived stress-strain prop-

erties of dental materials. Dental Materials, 23(7):814�821, Jul 2007.

[48] M. Herrguth, M. Wichmann, and S. Reich. The aesthetics of all-ceramic

veneered and monolithic cad/cam crowns. Journal of Oral Rehabilita-

tion, 32(10):747�752, Oct 2005.

[49] H. O. Heymann, S. C. Bayne, J. R. Sturdevant, A. D. Wilder Jr, and

T. M. Roberson. The clinical performance of cad-cam-generated ce-

ramic inlays: a four-year study. Journal of the American Dental Asso-

ciation (1939), 127(8):1171�1181, Aug 1996.

[50] R. Hickel and J. Manhart. Longevity of restorations in posterior teeth

and reasons for failure. Journal of Adhesive Dentistry, 3(1):45�64, 2001.

[51] F. G. Hougeir, J. A. Yiannias, M. L. Hinni, J. G. Hentz, and R. A.

el Azhary. Oral metal contact allergy: a pilot study on the cause of

oral squamous cell carcinoma. International Journal of Dermatology,

45(3):265�271, Mar 2006.

[52] M. B. Huerzeler, H. Fett, and W. H. Morman. Clinical evaluation

of ceramic restorations machined by cad/cam technique. In W. H.

82

Mormann, editor, International symposium on computer restorations:

state of the art of the CEREC-method, pages 327�335. Quintessence,

Chicago, 1991.

[53] S. Inokoshi, B. Van Meerbeek, G. Willems, P. Lambrechts, M. Braem,

and G. Vanherle. Marginal accuracy of cad/cam inlays made with the

original and the updated software. Journal of Dentistry, 20(3):171�177,

Jun 1992.

[54] B. P. Isenberg, M. E. Essig, and K. F. Leinfelder. Three-year clini-

cal evaluation of cad/cam restorations. Journal of Esthetic Dentistry,

4(5):173�176, Sep-Oct 1992.

[55] K. Kamada, K. Yoshida, and M. Atsuta. E�ect of ceramic surface

treatments on the bond of four resin luting agents to a ceramic material.

The Journal of Prosthetic Dentistry, 79(5):508�513, May 1998.

[56] E. C. Kao and W. M. Johnston. Fracture incidence on debonding of

orthodontic brackets from porcelain veneer laminates. The Journal of

Prosthetic Dentistry, 66(5):631�637, Nov 1991.

[57] K. Kawai, B. P. Isenberg, and K. F. Leinfelder. E�ect of gap dimension

on composite resin cement wear. Quintessence International, 25(1):53�

58, Jan 1994.

83

[58] J. R. Kelly. Clinically relevant approach to failure testing of all-ceramic

restorations. The Journal of Prosthetic Dentistry, 81(6):652�661, Jun

1999.

[59] J. R. Kelly. Dental ceramics: current thinking and trends. Dental

Clinics of North America, 48(2):513�530, Apr 2004.

[60] J. R. Kelly, B. D. Hunter, and M. R. Brenyo. Simulating clinical failure

during in vitro testing of all-ceramic crowns (abstract 1175). Journal

of Dental Research, 77 (special issue B):778, 1998.

[61] I. Krejci, F. Lutz, and M. Reimer. Wear of cad/cam ceramic inlays:

restorations, opposing cusps, and luting cements. Quintessence Inter-

national, 25(3):199�207, Mar 1994.

[62] N. Krämer, U. Lohbauer, and R. Frankenberger. Adhesive luting of in-

direct restorations. American Journal of Dentistry, 13(Spec No):60D�

76D, Nov 2000.

[63] B. Kränke, M. Binder, J. Derhaschnig, P. Komericki, D. Pirkhammer,

V. Ziegler, and W. Aberer. Patch testing with the "austrian standard

series"�epidemiologic test values and results. Wien Klin Wochenschr,

107(11):323�330, 1995.

[64] K. H. Kunzelmann, B. Jelen, A. Mehl, and R. Hickel. Wear evalua-

tion of mz100 compared to ceramic cad/cam materials. International

Journal of Computerized Dentistry, 4(3):171�184, Jul 2001.

84

[65] R. Laeijendecker and T. van Joost. Oral manifestations of gold allergy.

Journal of the American Academy of Dermatology, 30(2 Pt 1):205�209,

Feb 1994.

[66] K. F. Leinfelder, B. P. Isenberg, and M. E. Essig. A new method

for generating ceramic restorations: a cad-cam system. Journal of the

American Dental Association (1939), 118(6):703�707, Jun 1989.

[67] D. Magnuson, A. Oden, and I. Krystek. Clinical evaluation of cerec

restorations. In W. H. Mormann, editor, International symposium on

computer restorations: state of the art of the CEREC-method, pages

339�346. Quintessence, chicago, 1991.

[68] J. Manhart, H. Chen, G. Hamm, and R. Hickel. Buonocore memorial

lecture: Review of the clinical survival of direct and indirect restora-

tions in posterior teeth of the permanent dentition. Operative Dentistry,

29(5):481�508, 2004.

[69] J. G. Marks, D. V. Belsito, V. A. DeLeo, J. F. Fowler, A. F. Fran-

sway, H. I. Maibach, C. G. Mathias, M. D. Pratt, R. L. Rietschel, E. F.

Sherertz, F. J. Storrs, and J. S. Taylor. North american contact der-

matitis group patch-test results, 1996-1998. Archives of Dermatology,

136(2):272�273, Feb 2000.

85

[70] N. Martin and N. M. Jedynakiewicz. Clinical performance of cerec

ceramic inlays: a systematic review. Dental Materials, 15(1):54�61,

Jan 1999.

[71] E. A. McLaren and R. A. Giordano. Zirconia-based ceramics: ma-

terial properties, esthetics, and layering techniques of a new veneer-

ing porcelain, vm9, high-alumina frameworks. Quintessence of Dental

Technology, 28:1�12, 2005.

[72] E. A. McLaren, R. A. Giordano, R. Pober, and B. Abozenada. Material

testing and layering techniques of a new two-phase all glass veneering

porcelain. Quintessence of Dental Technology, 26:69�81, 2003.

[73] J. W. McLean, J. R. Hubbard, and M. I. Kedge. The Science and

Art of Dental Ceramics, chapter II: Construction of Porcelain Veneer

Crowns: Technical Considerations, page 45. Quintessence, Chicago,

1979.

[74] J. W. McLean and J. A. von Fraunhofer. The estimation of cement �lm

thickness by an in vivo technique. British Dental Journal, 131(3):107�

111, Aug 1971.

[75] P. F. Messer, V. Piddock, and C. H. Lloyd. The strength of dental

ceramics. Journal of Dentistry, 19(1):51�55, Feb 1991.

86

[76] M. K. Molin and S. L. Karlsson. A randomized 5-year clinical eval-

uation of 3 ceramic inlay systems. The International Journal of

Prosthodontics, 13(3):194�200, May-Jun 2000.

[77] T. Morikawa, H. Matsumura, and M. Atsuta. Bonding of a mica-based

castable ceramic material with a tri-n-butylborane-initiated adhesive

resin. Journal of Oral Rehabilitation, 23(7):450�455, Jul 1996.

[78] W. H. Mormann. The evolution of the cerec system. Journal of the

American Dental Association, 137 Suppl:7S�13S, Sep 2006.

[79] W. H. Mormann and A. Bindl. The new creativity in ceramic restora-

tions: dental cad-cim. Quintessence International, 27(12):821�828, Dec

1996.

[80] W. H. Mormann, T. Gotsch, I. Krejci, F. Lutz, and F. Barbakow.

Clinical status of 94 cerec ceramic inlays after 3 years in situ. In W. H.

Mormann, editor, International symposium on computer restorations:

state of the art of the CEREC-method, pages 355�362. Quintessence,

Chicago, 1991.

[81] W. H. Mormann and I. Krejci. Cinical and sem evaluation of cerec

inlays after 5 years in situ. In W. H. Mormann, editor, International

symposium on computer restorations: state of the art of the CEREC-

method, pages 25�32. Quintessence, Chicago, 1991.

87

[82] W. H. Mormann and J. Schug. Grinding precision and accuracy of �t

of cerec 2 cad-cim inlays. Journal of the American Dental Association,

128(1):47�53, Jan 1997.

[83] T. Nakamura, N. Dei, T. Kojima, and K. Wakabayashi. Marginal and

internal �t of cerec 3 cad/cam all-ceramic crowns. The International

Journal of Prosthodontics, 16(3):244�248, May-Jun 2003.

[84] T. Namikoshi. Case of oral lichen planus due to dental metal allergy.

Nihon Hotetsu Shika Gakkai Zasshi, 50(3):461�463, Jul 2006.

[85] G. Neiva, P. Yaman, J. B. Dennison, M. E. Razzoog, and B. R. Lang.

Resistance to fracture of three all-ceramic systems. Journal of Esthetic

Dentistry, 10(2):60�66, 1998.

[86] G. R. Norman and D. L. Streiner. Biostatistics : the bare essentials.

B.C. Decker, Hamilton, 3rd edition, 2008. Page 324.

[87] T. Otto and S. De Nisco. Computer-aided direct ceramic restorations:

a 10-year prospective clinical study of cerec cad/cam inlays and onlays.

The International Journal of Prosthodontics, 15(2):122�128, Mar-Apr

2002.

[88] U. Pallesen and J. W. van Dijken. An 8-year evaluation of sintered ce-

ramic and glass ceramic inlays processed by the cerec cad/cam system.

European Journal of Oral Sciences, 108(3):239�246, Jun 2000.

88

[89] K. Pallis, J. A. Griggs, R. D. Woody, G. E. Guillen, and A. W. Miller.

Fracture resistance of three all-ceramic restorative systems for posterior

applications. The Journal of Prosthetic Dentistry, 91(6):561�569, Jun

2004.

[90] A. Posselt and T. Kerschbaum. Longevity of 2328 chairside cerec inlays

and onlays. International Journal of Computerized Dentistry, 6(3):231�

248, Jul 2003.

[91] S. Reich, M. Wichmann, E. Nkenke, and P. Proeschel. Clinical �t of all-

ceramic three-unit �xed partial dentures, generated with three di�erent

cad/cam systems. European Journal of Oral Sciences, 113(2):174�179,

Apr 2005.

[92] B. Reiss. Clinical results of cerec inlays in a dental practice over a

period of 18 years. International Journal of Computerized Dentistry,

9(1):11�22, Jan 2006.

[93] B. Reiss and W. Walther. Survival analysis and clinical evaluation of

cerec restorations in a private practice. In W. H. Mormann, editor,

International symposium on computer restorations: state of the art of

the CEREC-method, pages 215�223. Quintessence, Chicago, 1991.

[94] B. Reiss andW.Walther. Clinical long-term results and 10-year kaplan-

meier analysis of cerec restorations. International Journal of Comput-

erized Dentistry, 3(1):9�23, Jan 2000.

89

[95] S.F. Rosenstiel, F. Martin, and J. Fujimoto. Contemporary Fixed

Prosthodontics. Mosby St. Louis, 4th edition, 2001. Page 262-271.

[96] S.F. Rosenstiel, F. Martin, and J. Fujimoto. Contemporary Fixed

Prosthodontics. Mosby St. Louis, 4th edition, 2001. Page 643-672.

[97] M. Rosentritt, M. Behr, R. Lang, and Handel G. In�uence of cement

type on the marginal adaptation of all-ceramic mod inlays. Dental

Materials, 20(5):463�469, Jun 2004.

[98] R. P. Rusin. Properties and applications of a new composite block

for cad/cam. Compendium of Continuing Education in Dentistry, 22(6

Suppl):35�41, Jun 2001.

[99] R. P. Rusin, R. P. Costin, and L. A. Goldsworthy. Tool life-time on the

cerec 2 system (abstract 582). Journal of Dental Research, 80, 2001.

[100] K. Sasaki, A. G. Hannam, and W. W. Wood. Relationships between

the size, position, and angulation of human jaw muscles and unilateral

�rst molar bite force. Journal of Dental Research, 68(3):499�503, Mar

1989.

[101] R. R. Seghi and J. A. Sorensen. Relative �exural strength of six

new ceramic materials. The International Journal of Prosthodontics,

8(3):239�246, 1995.

[102] A. Sertoli, S. Francalanci, MC Acciai, and M. Gola. Epidemiological

survey of contact dermatitis in Italy (1984-1993) by GIRDCA (Gruppo

90

Italiano Ricerca Dermatiti da Contatto e Ambientali). American Jour-

nal of Contact Dermatitis, 10(1):18�30, 1999.

[103] H. T. Shillingburg. Fundamentals of �xed prosthodontics. Chicago:

Quintessence Pub. Co., 3rd edition, 1997. page 455.

[104] H. T. Shillingburg. Fundamentals of �xed prosthodontics. Chicago:

Quintessence Pub. Co., 3rd edition, 1997. page 433-454.

[105] A. Shor, J. I. Nicholls, K. M. Phillips, and W. J. Libman. Fatigue load

of teeth restored with bonded direct composite and indirect ceramic

inlays in mod class ii cavity preparations. The International Journal

of Prosthodontics, 16(1):64�69, 2003.

[106] S. Siervo, A. Pampalone, P. Siervo, and R. Siervo. Where is the gap?

machinable ceramic systems and conventional laboratory restorations

at a glance. Quintessence International, 25(11):773�779, Nov 1994.

[107] R. Silva, F. Pereira, O. Bordalo, E. Silva, A. Barros, M. Gonçalo,

T. Correia, G. Pessoa, A. Baptista, and M. Pecegueiro. Contact allergy

to gold sodium thiosulfate. a comparative study. Contact Dermatitis,

37(2):78�81, Aug 1997.

[108] G. Sjogren, M. Bergman, M. Molin, and C. Bessing. A clinical ex-

amination of ceramic (cerec) inlays. Acta Odontologica Scandinavica,

50(3):171�178, Jun 1992.

91

[109] G. Sjogren, R. Lantto, A. Granberg, B. O. Sundstrom, and A. Till-

berg. Clinical examination of leucite-reinforced glass-ceramic crowns

(empress) in general practice: a retrospective study. The International

Journal of Prosthodontics, 12(2):122�128, Mar-Apr 1999.

[110] G. Sjogren, M. Molin, and J. W. van Dijken. A 5-year clinical evalua-

tion of ceramic inlays (cerec) cemented with a dual-cured or chemically

cured resin composite luting agent. Acta Odontologica Scandinavica,

56(5):263�267, Oct 1998.

[111] G. Sjogren, M. Molin, and J. W. Van Dijken. A 10-year prospective

evaluation of cad/cam-manufactured (cerec) ceramic inlays cemented

with a chemically cured or dual-cured resin composite. The Interna-

tional Journal of Prosthodontics, 17(2):241�246, Mar-Apr 2004.

[112] G. Sjogren, G. Sletten, and J. E. Dahl. Cytotoxicity of dental alloys,

metals, and ceramics assessed by millipore �lter, agar overlay, and mtt

tests. The Journal of Prosthetic Dentistry, 84(2):229�236, Aug 2000.

[113] L. C. Sobrinho, M. J. Cattell, R. H. Glover, and J. C. Knowles. Investi-

gation of the dry and wet fatigue properties of three all-ceramic crown

systems. The International Journal of Prosthodontics, 11(3):255�262,

1998.

92

[114] I. Stangel, D. Nathanson, and C. S. Hsu. Shear strength of the compos-

ite bond to etched porcelain. Journal of Dental Research, 66(9):1460�

1465, Sep 1987.

[115] C. F. Stappert, P. C. Guess, S. Chitmongkolsuk, T. Gerds, and J. R.

Strub. All-ceramic partial coverage restorations on natural molars.

masticatory fatigue loading and fracture resistance. American Journal

of Dentistry, 20(1):21�26, Feb 2007.

[116] J. R. Strub and S. M. Beschnidt. Fracture strength of 5 di�erent all-

ceramic crown systems. The International Journal of Prosthodontics,

11(6):602�609, Nov-Dec 1998.

[117] J. Tinschert, D. Zwez, R. Marx, and K. J. Anusavice. Structural relia-

bility of alumina-, feldspar-, leucite-, mica- and zirconia-based ceram-

ics. Journal of Dentistry, 28(7):529�535, Sep 2000.

[118] R. R. Torgerson, M. D. P. Davis, A. J. Bruce, S. A. Farmer, and

R. S. Rogers. Contact allergy in oral disease. Journal of the American

Academy of Dermatology, 57(2):315�321, Aug 2007.

[119] L. Trost, S. Stines, and L. Burt. Making informed decisions about

incorporating a cad/cam system into dental practice. Journal of the

American Dental Association (1939), 137 Suppl:32S�36S, Sep 2006.

93

[120] E. A. Tsitrou, M. Helvatjoglu-Antoniades, S.E. Northeast, and R. Van

Noort. Fracture resistance of minimally prepared cerec crowns. In

IADR Abstract no. 85781, 2006.

[121] E. A. Tsitrou, S. E. Northeast, and R. van Noort. Evaluation of the

marginal �t of three margin designs of resin composite crowns using

cad/cam. Journal of Dentistry, 35(1):68�73, Jan 2007.

[122] B. Tyan, R. Pober, and R. Giordano. Fatigue of cad/cam milled crowns.

In IADR Abstract no. 3963, 2002.

[123] S. M. Wiederhorn. Moisture assisted crack growth in ceramics. Inter-

national Journal of Fracture, 4(2):171�177, 1968.

[124] M. G. Wiley. E�ects of porcelain on occluding surfaces of restored

teeth. The Journal of Prosthetic Dentistry, 61(2):133�137, Feb 1989.

[125] H. W. Wiskott, J. I. Nicholls, and U. C. Belser. Stress fatigue: basic

principles and prosthodontic implications. The International Journal

of Prosthodontics, 8(2):105�116, 1995.

[126] D. M. Wolf, J. M. Powers, and K. L. O'Keefe. Bond strength of com-

posite to porcelain treated with new porcelain repair agents. Dental

Materials, 8(3):158�161, May 1992.

[127] K. T. Wrbas, N. Hein, J. F. Schirrmeister, M. J. Altenburger, and

E. Hellwig. Two-year clinical evaluation of cerec 3d ceramic inlays

94

inserted by undergraduate dental students. Quintessence International,

38(7):575�581, Jul 2007.

[128] K. H. K Yip, R. J. Smales, and J. A. Kaidonis. Di�erential wear of

teeth and restorative materials: clinical implications. The International

Journal of Prosthodontics, 17(3):350�356, 2004.

[129] M. Yoshinari and T. Derand. Fracture strength of all-ceramic crowns.

The International Journal of Prosthodontics, 7(4):329�338, Jul-Aug

1994.

[130] M. Zahran, O. El-Mowafy, L. Tam, P. A. Watson, and Y. Finer.

Fracture strength and fatigue resistance of all-ceramic molar crowns

manufactured with cad/cam technology. Journal of Prosthodontics,

17(5):370�377, Jul 2008.

[131] Z. T. Zhang, L. Yan, M. Zhong, X. D. Yang, and H. J. Ai. The ultra-

structure and activities of free radical scavenger in discolored gingiva

adjacent to porcelain fused to metal crowns. Shanghai Journal of Stom-

atology, 16(2):176�179, April 2007.

[132] A. A. El Zohairy, A. J. De Gee, M. M. Mohsen, and A. J. Feilzer.

Microtensile bond strength testing of luting cements to prefabricated

cad/cam ceramic and composite blocks. Dental Materials, 19(7):575�

583, Nov 2003.

95

[133] R. Zuellig-Singer and R. W. Bryant. Three-year evaluation

of computer-machined ceramic inlays: in�uence of luting agent.

Quintessence International, 29(9):573�582, Sep 1998.

96