Click here to load reader

For - NASAspearman's rank correlation coefficient for order of "parameter" versus order of fa1 lure percent change in hfe with time at room temperature (25 * 3oc) after gamma dose

  • Upload
    others

  • View
    10

  • Download
    0

Embed Size (px)

Citation preview

  • ER 8623

    COMPONENTS IRRADIATION TEST NO. 19

    GAMMA IRRADIATION

    OF

    2N914, 2N918, S2N930, 2N2192

    AND 2N2369 TRANSISTORS

    October 1966

    Prepared For:

    GEORGE C. MARSHALL SPACE FLIGHT CENTER

    Prepared By:

    LOCKHEED GEORGIA NUCLEAR LABORATORY

    L O C K H E E D G E O R G I A N U C L E A R L A B O R A T O R Y Lockheed-Georgia Company - A Division of Lockheed Aircraft Corporation

    https://ntrs.nasa.gov/search.jsp?R=19670009216 2020-03-20T16:51:48+00:00Z

  • I f this document i s supplied under the requirements of a United States Government contract, the following legend shall apply unless the letter U appears in the coding box.

    This data i s furnished under a United States Government contract and only those portions hereof which are marked (for example, by circling, underscoring or otherwise) and indicated as being subject to this legend shall not be released outside the Government (except to foreign gov- ernments, subject to these same limitations), nor be disclosed, used, or duplicated, for procurement or man- ufacturing purposes, except as otherwise authorized by contract, without the permission of Lockheed-Georgia Company, A Division of Lockheed Aircraft Corporation, Marietta, Georgia. This legend shall be marked on any reproduction hereon in whole or in part.

    The "otherwise marking" and "indicated portions" as used above shall mean this statement and include al l details or manufacture contained herein respectively.

    D I I

    I I I I I I I I I I I I I I

    m

  • I

    1

    I I I E 1

    FOREWORD

    This report i s submitted to the Astrionics Laboratory of the George C.

    Marshal I Space Flight Center, National Aeronautics and Space Admin-

    istration, Huntsville, Alabama, in accordance with the requirements

    of Task Order No. ASTR-LGC-330f Contract NAS 8-5332. The re-

    port i s one of a series describing radiation effects on various electronic

    components. This particular report concerns gamma irradiation and an-

    nealing of types 2N914, 2N918, S2N930, 2N2192 and 2N2369 tran-

    sistors.

    The test was performed by the Lockheed Georgia Nuclear Laboratory,

    Lockheed-Georgia Company.

  • I TABLE OF C O N T E N T S

    Page

    FOREWORD

    TABLE OF CONTENTS

    LIST OF TABLES AND FIGURES

    1 .o

    2.0

    3.0

    3.1

    3.2

    3.3

    3.3.1

    3.3.2

    3.3.3

    3.4

    3.4.1

    3.4.2

    3.4.3

    4.0

    5.0

    5.1

    5.2

    SUMMARY

    INTRODUCTION

    TEST PROCEDURE

    TEST SPECIMENS

    TEST MEASUREMENTS

    INSTRUMENTATION

    h Measurement Circuit FE Measurement Circuit 'EBO

    T Measurement Circuit

    TEST ENVIRONMENT

    Pressure

    X

    Tempe ra tu re

    Gamma

    METHOD OF DATA ANALYSIS

    TEST DATA AND DISCUSSION OF RESULTS

    ANNEALING AS A FUNCTION OF TIME AFTER

    SMALL DOSES OF GAMMA RADIATION

    RECOVERY OF hFE BY ANNEALING AT HIGH

    TEMPERATURE

    i

    I l l ...

    V

    1

    3

    5

    7

    8

    8

    8

    9

    9

    9

    9

    9

    10

    1 1

    13

    14

    14

    ... I l l

  • I

    5.3

    5.4

    5.5

    5.6

    5.7

    5.8

    5.9

    TABLE OF C O N T E N T S (Continued)

    A COMPARISON OF RADIATION TOLERANCES

    DURING FIRST AND SECOND IRRADIATIONS

    CORRELATIO N BETWEEN PRE- I RRADl ATlO N

    PARAMETERS AND RADIATION TOLERANCE

    RELATIVE DEGRADATION OF hFE AT LOW GAMMA

    DOSES

    CONCLUSIONS

    RECOMMENDED METHOD FOR PRE-SELECT1 NG NPN

    SILICON TRANSISTORS FOR USE IN A GAMMA

    RAD1 AT10 N E NVI RO NME NT

    RECOMMENDED METHOD FOR ESTIMATING

    MINIMUM GAMMA RADIATION LIFETIME OF

    NPN SILICON TRANSISTORS

    RECOMMENDED FOLLOW-ON INVESTIGATIONS

    REFERENCES

    Page

    15

    16

    18

    19

    20

    21

    23

    25

    I V

    1 1 I 1 3 I I 1 I I I I

    I 1 1 1 I

    a

  • I 'I i I r I I

    Tables

    TABLE 1

    TABLE 2

    TABLE 3

    TABLE 4

    TABLE 5

    TABLE 6

    TABLE 7

    TABLE 8

    TABLE 9

    L I S T OF TABLES A N D F I G U R E S

    TEST SPECIMENS AND TEST CONDITIONS

    MANUFACTURERS' SPECIFICATIONS FOR

    TEST SPECIMENS

    SPEARMAN'S RANK CORRELATION COEFFICIENT

    FOR ORDER OF "PARAMETER" VERSUS ORDER OF

    FA1 LURE

    PERCENT CHANGE IN hFE WITH TIME AT ROOM

    TEMPERATURE (25 * 3OC) AFTER GAMMA DOSE OF 3 . 1 5 ~ 10 r

    PERCENT CHANGE IN hFE WITH TIME AT ROOM

    TEMPERATURE (25 * 3OC) AFTER GAMMA DOSE OF 3 . 1 5 ~ 10 r

    PERCENT CHANGE IN hFE WITH TIME AT ROOM

    TEMPERATURE (25 3OC) AFTER GAMMA DOSE

    OF 9 . 6 7 ~ 10 r

    5

    5

    5

    2N914, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),

    hFE DEGRADATION, ANNEALING AND FAILURE

    DATA

    2N914, FAIRCHILD, MOTOROLA, TEXAS INSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),

    hFE DEGRADATION, ANNEALING AND FAILURE

    DATA

    2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),

    hFE DEGRADATION, ANNEAL1 NG AND FA1 LURE

    DATA

    Page

    27

    28

    29

    30

    31

    32

    33

    34

    35

    V

  • Tables

    TABLE 10

    TABLE 1 1

    TABLE 12

    TABLE 13

    TABLE 14

    TABLE 15

    TABLE 16

    L I S T OF TABLES A N D F I G U R E S (Continued)

    2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),

    hFE DEGRADATION, ANNEAL1 NG AND FA1 LURE

    DATA

    2N2369, FAIRCHILD, MOTOROLA, TEXAS I NSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),

    hFE DEGRADATION, ANNEALING A N D FAILURE

    DATA

    2N2369, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),

    hFE DEGRADATION, ANNEALING AND FAILURE

    DATA

    2N918, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),

    hFE DEGRADATION, ANNEALING AND FAILURE

    DATA

    2N918, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-

    MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),

    hFE DEGRADATION, ANNEALING AND FAILURE

    DATA

    2N914 AND 2N2192, FAIRCHILD, MOTOROLA,

    TEXAS INSTRUMENTS AND GENERAL ELECTRIC,

    hFE DEGRADATION, ANNEALING AND FAILURE

    DATA

    2N2369 AND 2N918, FA1 RCHl LD, MOTOROLA,

    TEXAS INSTRUMENTS AND GENERAL ELECTRIC,

    hFE DEGRADATION, ANNEALING AND FAILURE

    DATA

    Page

    36

    37

    38

    39

    40

    41

    42

    v i

  • Tables

    TABLE 17

    TABLE 18

    TABLE 19

    TABLE 20

    TABLE 21

    TABLE 22

    TABLE 23

    TABLE 24

    TABLE 25

    TABLE 26

    TABLE 27

    L IST OF TABLES A N D F I G U R E S (Continued)

    S2N930, FAIRCHILD (Ic = 2 mA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    S2N930, FAIRCHILD (Ic = 2 mA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    S2N930, TEXAS INSTRUMENTS (Ic = 2 mA), hFE DEGRA-

    DATION, ANNEALING AND FAILURE DATA

    S2N930, TEXAS INSTRUMENTS (Ic = 10 mA), hFE DEG-

    RADATION, ANNEAL1 NG AND FAILURE DATA

    2N914, FAIRCHILD, MOTOROLA, TEXAS I NSTRU-

    MENTS AND GENERAL ELECTRIC, Te , lEBO AND h DATA 0 0

    FEO

    h~~

    2N2192 FAIRCHILD, MOTOROLA, TEXAS INSTRU-

    MENTS AND GENERAL ELECTRIC, Te , lEBO AND

    2N3369, FA1 RCHILD, MOTOROLA, TEXAS I NSTRU-

    MENTS AND GENERAL ELECTRIC, Te , lEBO AND

    2N818, FAIRCHILD, MOTOROLA, TEXAS INSTRU-

    MENTS AND GENERAL ELECTRIC, T , IEBo_ AND

    0 0 DATA

    0 0 DATA h~~

    e-

    Page

    43

    44

    45

    46

    47

    48

    49

    50

    51

    h DATA F E,

    U U 52

    2NF192, COMPUTED VALUES OF RADIATION DAM-

    AGE CONSTANT (K) AND COEFFICIENTS OF

    VARIATION (S/E)

    v i i

    53

  • Tables

    TABLE 28

    TABLE 29

    TABLE 30

    TABLE 31

    TABLE 32

    TABLE 33

    Figures

    FIGURE 1

    FIGURE 2

    GURE 3

    GURE 4

    GURE 5

    L I S T OF TABLES A N D F I G U R E S (Continued)

    Page

    I I

    I 1 1 I II I I I 1 1 1 I 1 1 I

    m

    v i i i

    SPEARMAN'S RANK CORRELATION COEFFICIENT

    FOR ORDER OF h T VERSUS ORDER OF

    FAILURE AT IC = 5 rnA

    G A I N REAL1 ZED BY SELECT1 NG OUT SPECIMENS

    WITH PERCENT DEGRADATION GREATER THAN

    MEDIAN VALUE AT GAMMA DOSE EQUAL TO 5%

    OF MEDIAN FAILURE DOSE

    G A I N REALIZED BY SELECT1 NG OUT SPECIMENS

    WITH PERCENT DEGRADATION GREATER THAN

    MEDIAN VALUE AT GAMMA DOSE EQUAL TO 5%

    OF MEDIAN FAILURE DOSE

    RATIO OF MINIMUM RADIATION LIFETIME OF

    FEo eo

    PRE-SELECTED SPECIMENS TO SCREEN1 NG DOSE

    MEDIAN PERCENT DEGRADATION VALUES AT

    SCREENING DOSES (5% OF MEDIAN FAILURE

    DOSES)

    2N914, COMPARISON OF ESTIMATED AND ACTUAL

    FAILURE DOSES FOR IC = 5 mA

    SCHEMATIC DIAGRAM OF GAMMA IRRADIATION

    TEST RIG

    hFE MEASUREMENT CIRCUIT

    MEASUREMENT CIRCUIT 'EBO

    TYPICAL CURVES OF 1o5/hFE VERSUS GAMMA

    TRANSIT TIME MEASUREMENT CIRCUIT

    DOSE FOR TYPE 2N914

    54

    55

    56

    57

    58

    59

    61

    62

    63

    64

    65

  • Figures

    FIGURE 6

    FIGURE 7

    GURE 8

    GURE 9

    FIGURE 10

    FIGURE 11

    GURE 12

    GURE 13

    FIGURE 14

    FIGURE 15

    FIGURE 16

    FIGURE 17

    FIGURE 18

    L I S T OF TABLES A N D F I G U R E S (Con ti nued)

    2N914, FA1 RCHILD, 27OC, PERCENT FAILED

    VERSUS GAMMA DOSE

    2N914, MOTOROLA, 27OC, PERCENT FA1 LED

    VERSUS GAMMA DOSE

    2N914, TEXAS INSTRUMENTS, 27OC, PERCENT

    FAILED VERSUS GAMMA DOSE

    2N914, GENERAL ELECTRIC, 27OC, PERCENT

    FAILED VERSUS GAMMA DOSE

    2N914, 27OC, PERCENT FAILED VERSUS GAMMA

    DOSE, COMPARISON OF MANUFACTURERS

    2N2192, FAIRCHILD, 27OC, PERCENT FAILED

    VERSUS GAMMA DOSE

    2N2192, MOTOROLA, 27OC, PERCENT FAILED

    VERSUS GAMMA DOSE

    2N2192, TEXAS INSTRUMENTS, 27OC, PERCENT

    FAILED VERSUS GAMMA DOSE

    2N2192, GENERAL ELECTRIC, 27OC, PERCENT

    FAILED VERSUS GAMMA DOSE

    2N2192, 27OC, PERCENT FAILED VERSUS GAMMA

    DOSE, COMPARISON OF MANUFACTURERS

    2N2369, FAIRCHILD, 27OC, PERCENT FAILED

    VERSUS GAMMA DOSE

    2N2369, MOTOROLA, 27OC, PERCENT FAILED

    VERSUS GAMMA DOSE

    2N2369, TEXAS INSTRUMENTS, 27OC, PERCENT

    FAILED VERSUS GAMMA DOSE

    I X

    Page

    66

    67

    68

    69

    70

    71

    72

    73

    74

    75

    76

    77

    78

  • Figures

    FIGURE 9

    FIGURE 20

    FIGURE 21

    FIGURE 22

    FIGURE 23

    FIGURE 24

    FIGURE 25

    FIGURE 26

    FIGURE 27

    FIGURE 28

    FIGURE 29

    L IST OF TABLES A N D F I G U R E S (Con t i nu ed)

    2N2369, GENERAL ELECTRIC, 270Cf PERCENT

    FAILED VERSUS GAMMA DOSE

    2N2369, 270Cf PERCENT FAILED VERSUS GAMMA

    DOSE, COMPARISON OF MANUFACTURERS

    2N918, FAIRCHILD, 27OC, PERCENT FAILED

    VERSUS GAMMA DOSE

    2N918, MOTOROLAf 27OC, PERCENT FA1 LED

    VERSUS GAMMA DOSE

    2N918, TEXAS INSTRUMENTS, 27OC, PERCENT

    FAILED VERSUS GAMMA DOSE

    2N918, GENERAL ELECTRIC, 27OC, PERCENT

    FAILED VERSUS GAMMA DOSE

    2N918, 270Cf PERCENT FA1 LED VERSUS GAMMA

    DOSE, COMPARISON OF MANUFACTURERS

    2N914, 27OC, PERCENT DEGRADATION OF hFE

    AT 1.80 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 rnA

    2N914, 270Cf PERCENT DEGRADATION OF hFE

    AT 3.22 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 rnA

    2N2192, 270Cf PERCENT DEGRADATION OF hFE

    AT 3.05 x lo5 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR I C = 5 mA 2N2192, 270Cf PERCENT DEGRADATION OF hFE

    AT 4.52 x lo5 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA

    X

    Page

    79

    80

    81

    82

    83

    84

    85

    86

    87

    88

    89

  • I

    Figures

    FIGURE 30

    FIGURE 31

    FIGURE 32

    FIGURE 33

    FIGURE 34

    FIGURE 35

    FIGURE 36

    L I S T OF TABLES A N D F I G U R E S (Con t i nu ed)

    Page

    2N2369, 27OC, PERCENT DEGRADATION OF hFE

    AT 2. i o io6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 mA

    2N2369, 27OC, PERCENT DEGRADATION OF hFE

    AT 3.98 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA

    2N918, 27OC, PERCENT DEGRADATION OF hFE

    AT 4.48 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 mA

    2N918, 27OC, PERCENT DEGRADATION OF hFE AT

    4.45 x io6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA

    NPN SILICON PLANAR EPITAXIAL TRANSISTORS

    (16 TYPE 2N914, 16TYPE 2N2192, 16 TYPE 2N918,

    16 TYPE 2N2369), 27OC, PERCENT DEGRADATION OF

    hFE AT 1.88 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 mA 94

    NPN SILICON PLANAR EPITAXIAL TRANSISTORS

    ( 1 6 TYPE 2N914, 16 TYPE 2N2192, 12 TYPE 2N918,

    16 TYPE 2N2369), 27OC, PERCENT DEGRADATION OF

    hFE AT 3.05 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA 95

    FIRST FAILURE POINT FOR UNSCREENED SPECIMENS

    VERSUS FIRST FA1 LURE POINT FOR PRE-SELECTED

    SPECIMENS 96

    xi

    90

    91

    92

    93

  • 8 1 . O S U M M A R Y

    NPN silicon planar epitaxial transistors of the types 2N914, 2N2192, 2N2369 and

    2N918 were subjected to gamma irradiation and failure points (gamma dose when

    50% degradation of h

    perature annealed and gamma irradiated a second time to failure along with type

    S2N930 specimens from a previous test.

    occurred) were determined. The specimens were then tem- FE

    investigations were made on possible correlations between pre-irradiation values

    h ~ ~ ’ e EBO covery of radiation induced degradation o f h

    of gamma irradiation was investigated, and comparison was made between failure

    points shown during the first irradiation and those shown during the second irradia-

    tion. The ultimate objective of the test was the development of a method to pre-

    select transistors for use in a radiation environment.

    T and I and radiation tolerance exhibited during the first irradiation. Re-

    at mom temperature after small doses FE

    Test results indicated that NPN silicon transistors can be pre-selected for use i n a

    gamma radiation environment by subjecting the devices to comparatively smal I doses

    o f gamma radiation and computing percentage losses of h

    vices w i l l have a minimum radiation lifetime about one order of magnitude, or more,

    greater than the screening dose.

    The pre-selected de- FE *

    1

  • The experiment described in this report i s the nineteenth irradiation of electronic

    components and i s the twenty-fourth in a series of radiation effects tests on electron-

    i c equipment, circuits and components contemplated for use on a nuclear space ve-

    h ic le . Since the use of equipment on this vehic le i s contingent upon its abi l i ty to

    withstand the nuclear environment, the Astrionics Laboratory o f the Marshall Space

    Flight Center has undertaken to assure that Government furnished or specified equip-

    ment w i l l survive this environment. The equipment i s to be subjected to the expected

    nuclear environment as simulated at the Lockheed Georgia Nuclear Laboratory. Mea-

    surements made on the specimens during the irradiation w i l l describe their radiation

    tolerance.

    8 I I 1 I

    I

    1 I

    The subjects of this test are the 2N914, 2N918, S2N930, 2N2192 and 2N2369 tran-

    sistors. This experiment i s an extension o f the work performed during Components

    Irradiation Test No. 18. The purpose of the extended work was to verify correlations

    established by the Test No. 18 data, i .e ., the correlation between percentage loss of hFE at comparatively low gamma doses and gamma dose at failure, and the corre-

    lation between h x T

    o f these correlations would enable pre-selection of transistors for use in radiation en-

    vironments. During the experiment, recovery of lost h (annealing) as a function

    o f time after small doses of irradiation was also investigated, and comparison of ra-

    diation tolerances during in i t ia l irradiation was made with radiation tolerances ex-

    hibited during a second irradiation conducted after the specimens had been annealed

    at high temperatures.

    and dose at failure. Verification of one or both 'IEBO, FEO eo

    FE

    3

  • I 3 . 0 T E S T PROCEDURE

    The test specimens, except the S2N930 type, were procured "off-the-shelf" from an

    electronics supply vendor. The S2N930 specimens had been supplied by the Astrionics

    Laboratory of the Marshall Space Flight Center for Components Irradiation Test No. 18.

    from each of four manufacturers: Fairchild, Motorola, Texas Instruments, and Gen-

    eral Electric. The S2N930 specimens consisted of forty-five specimens from Fairchild

    and twenty specimens from Texas instruments. These S2N930 specimens, having been

    irradiated during the Test No. 18 experiment, were not used for the first irradiation

    of this test.

    Seven specimens of each of the 2N914, 2N918, 2N2192 and 2N2369 types were

    and T were made on a l l specimens of the FE' 'EBO e Pre-irradiation measurements of h

    2N914, 2N918, 2N2192 and 2N2369 type - manufacturer groups. After these mea- surements were completed one specimen of each type - manufacturer group was selected as a control specimen. These control specimens, sixteen in all, were used as checks

    on the stability and repeatability of the measurement instrumentation. The remaining

    devices were irradiated i n a fixture as shown in Figure 1. Cobalt-60 was used as the

    source of the gamma radiation.

    4 The irradiation was conducted at a constant gamma dose rate of 1.10 x 10 r/hr unti l

    a dose of 5.37 x 10 r had been reached. From that point on, a dose rate of 2.10 x

    10 r/hr was used for the remainder of the experiment.

    5

    5

    FE The specimens were removed from the radiation environment for measurement of h

    at the following dose accumulation points:

    4

    4

    5

    1 . 0 6 ~ 10 r

    5 . 3 0 ~ 10 r

    1 . 0 2 ~ 10 r

  • 4.40 x

    0.27 x

    2.70 x

    5

    5

    5

    3 . 1 5 ~ 10 r

    5 . 3 7 ~ 10 r

    9 . 6 7 ~ 1 0 r

    o6 r 6

    O r 7 0 r (end of first irradiation).

    5 At the 5.37 x 10 r dose point the hFE of two specimens of each type - manufacturer group was measured after elapsed periods of three hours, seven hours and sixty-five

    hours to detect any annealing at room temperature.

    5 At the 9.67 x 10 r dose point two specimens of each type - manufacturer group were removed from the test rig and were not given any additional dose during the first i r -

    radiation. This was done to check for annealing at room temperature over a period

    of several days, and to investigate the effect of total dose on the degree of anneal-

    ing that might be accomplished by subjecting the specimens to high temperatures.

    Some of the specimens had not failed (h

    terminated at 2.70 x 10 r. The failure points for these specimens were determined

    /h FE FEo 5: 0.5) when the first irradiation was 7

    by extrapolation of the Q-factor (l/h ) curves. I t i s be l ieved that no errors were FE introduced by this procedure as the Q-factor curves were linear with gamma dose i n

    the region of extrapolation.

    After post-irradiation measurements of h were made al l specimens except the con-

    trol specimens were annealed in an oven. The specimens were left in the oven for a

    period of about three hours at the temperatures indicated:

    FE

    T Y Pe 2N914

    2N918

    Annealing Temperature

    295 i s"C

    295 f s"C

  • TY Pe S2N930

    2N2192

    2N2369

    Ann e a I i ng Tempe rat ure

    295 i s"C

    295 i s"C

    195+ !fC

    The annealing temperatures were selected on the basis of the manufacturers' recom-

    mended maximum storage temperatures.

    After annealing pre-irradiation measurements of h

    A l l specimens except the control specimens were then placed i n the test rig and ir- 5 7

    radiated at 2.1 x 10 r/hr to a total dose of 5.20 x 10

    were made on al I specimens. F E

    r.

    Measurementsof h

    paints:

    were made on all devices at the following accumulated dose FE

    5

    5 1 . 0 5 ~ 10 r

    5 . 2 5 ~ 10 r 6 1 . 0 0 ~ 10 r 6

    6

    7

    7

    4.38 x 10 r

    9 . 3 4 ~ 10 r

    2 . 3 0 ~ 10 r

    5.20 x 10 r (end o f second irradiation).

    The purpose of this second irradiation was to compare failure points experienced prior

    to annealing with those experienced after annealing.

    3.1 TEST SPECIMENS

    The test specimens are l isted i n Table 1 . Manufacturers' specifications are shown in

    Table 2.

    7

  • The specimens were irradiated passively and were held for the irradiation by inser-

    tion of the device leads into the Styrofoam block shown i n Figure 1 . This arrange-

    ment insured that al I specimens were irradiated at the same rate.

    3.2 TEST MEASUREMENTS

    A complete set of measurements was taken prior to the first irradiation on the 2N914,

    FE 2N918, 2N2192 and 2N2369 specimens. These consisted of h for I = 5 mA; h

    for I C X E 2N2192 and 2N2369 types; T for IE valuesof 0.33, 0.25, 0.20, 0.167 and 0.125

    mA for the 2N918 type; I = 4.0 V for the 2N914, 2N2192 and 2N2369

    types; and I

    FE C = 30 mA; T for I values of 1 .O, 0.5, 0.33, 0.2 and 0.125 mA for the 2N914,

    X

    for V E 6 0 EB

    for VEB = 2.7 V for the 2N918 type. EBO

    Measurementsof h = 30 mA were made at frequent intervals

    during the first irradiation. The specimens were removed from the irradiation envi-

    ronment to a screened room for these measurements.

    for I = 5 mA and I FE C C

    Prior to the second irradiation, but after annealing, measurements of h

    ues of 5 mA and 30 mA were made on the 2N914, 2N918, 2N2192 and 2N2369 types.

    for I val- FE C

    At the same time measurements of h

    the S2N930 specimens. These same h

    for I FE C

    FE

    values of 2 mA and 10 mA were made on

    measurements were taken at frequent intervals

    during the second irradiation and after irradiation ceased. The procedure for making

    these measurements was the same as that used during the first irradiation.

    3.3 INSTRUMENTATION

    3.3.1 h Measurement Circuit FE

    A pulse measurement system, shown schematically in Figure 2, was used to measure

    This circuit l im i ted measurement currents to a time interval of about one millisecond I B *

    8

  • thus minimizing any annealing of radiation damage by the measurement currents.

    h~~ The control specimens were used to check on the repeatability of the circuit.

    values were calculated from the I measurements. B

    3.3.2 lEBO Measurement Circuit

    Pre-irradiation measurements of I

    shown i n Figure 3.

    were made using the circuit scllematica EBO ' Y

    3.3.3 T Measurement Circuit X

    The circuit shown i n Figure 4 was used to measure transit times for five different val-

    ues of I

    justing C and R unti l a null was obtained. The use of these values to obtain T i s

    Each measurement consisted o f adjusting I to the desired value, then ad- E' E

    X X X

    explained in Section 4.0.

    3.4 TEST ENVl RO NMENT

    3.4.1 Pressure

    During the test a l l specimens were at atmospheric pressure.

    3.4.2 Temperature

    0 Al l measurements and the irradiations were conducted at 27 f 2 C.

    Annealing Temperatures

    Between irradiations the types 2N914, S2N930, 2N2192 and 2N2369 were annealed

    for three hours at 295 i 5°C. The 2N918 type was annealed for three hours at 195 * !fC.

    9

  • 8 4 I

    3.4.3 Gamma

    The first irradiation was conducted at a dose rate of 1.10 x 10 r/hr to a dose point 5 5

    of 5.37 x 10 r. From that point a dose rate of 2.10 x 10 r/hr was used until 2.70 x 7 5

    10 r had been accumulated. The second irradiation was conducted at 2.10 x 10 r/hr

    to a total accumulated dose of 5.2 x 10 r. 7

    10

  • 4 . 0 M E T H O D OF D A T A A N A L Y S I S

    A l l data were recorded manually.

    The hFE data were calculated manually from measurements of I

    I C ' FE these plots the value of h

    FE ure dose for each specimen. (Examples o f typical l/h

    in Figure 5.) The data obtained from these plots were used to construct the graphs

    discussed i n Section 5.0.

    taken at constant B

    These data were used to construct plots of l / h versus accumulated dose. From

    at any desired dose could be obtained, as could the fail-

    versus dose plots are shown FE

    Transit time values were calculated from measurements on the bridge shown i n Figure

    4. The formula used for calculation of transit t ime was: 1

    T = (C R r/(r + R 1). X x x X

    where T i s i n nanoseconds, X

    C i s i n picofarads,

    R i s i n ohms, and

    r i s i n ohms.

    X

    X

    The T values were then paired with their corresponding 1/1 values (1 in milliamps)

    giving five pairs of values for each specimen. A straight line was then fitted to the

    f i v e pairs of values by the least squares method, and the value of T

    was calculated from the formula for this line. The value of T for 1/1 = 0 i s T or

    base transit t ime.

    X E E

    for 1/1 = 0

    X E e'

    X E

    Al l of the transit time computations were performed on an IBM 7094 computer.

    values were measured directly on a picoammeter and required no calculation. I EBO

    11

  • In order to provide a basis for comparing the correlations obtained between order of

    failure and various parameters the parameters were ranked and paired with the order 2 2

    = 1 - 6 r D /N(N - l ) , 2 o f failure of the specimens. Spearman's formula , r rank was then used to calculate a rank correlation coefficient. Results are shown in Ta-

    ble 3.

    Copies of the reduced data are on f i l e in the Lockheed Georgia Nuclear Laboratory,

    Lockheed-Georgia Company, Dawsonvil le, Georgia.

    12

  • I 5 . 0 T E S T D A T A A N D D I S C U S S I O N OF R E S U L T S

    The test data are presented herein in graphical and tabular form.

    The primary objectives of this test were to verify correlations established by the Com-

    ponents Irradiation Test No. 18 data, i.e., the correlation between percentage loss

    of h at comparatively low gamma doses and gamma dose at failure, and the corre-

    lation between h

    both of these correlations would enable the development o f a method to pre-se

    FE and dose at failure. The establishment o f one or T e i I E B O o FEO

    transistors for use i n a gamma radiation environment.

    To assist i n the development o f such a method two other objectives were establ

    for this test:

    ec t

    shed

    (a) An investigation o f the recovery of lost h

    time after small doses of irradiation, and

    A comparison of radiation tolerances during in i t ia l irradiation with tol-

    erances during a second irradiation conducted after the specimens had

    been annealed at high temperatures.

    (annealing) as CI function of FE

    (b)

    These were undertaken because, i f lost h could be regained by annealing without FE adversely affecting subsequent radiation tolerance, that part of a transistor's radia-

    tion lifetime lost by subjecting the devices to reasonable doses of gamma radiation

    for pre-selection could be recovered.

    For purposes of this test failure of a specimen was defined as a 50% decrease in the

    value o f h Thus, for the first irradiation failure occurred when h j h first

    equaled 0.5, where h

    ation; for the second irradiation failure occurred when h /h

    FE ' FE FE,, i s defined as the h of a device prior to the first irradi- FE

    first equaled 0.5, FEO 1

    FE FE-9 i s defined as the h of a device after the first irravdation and after FE where h

    FEo2

    13

  • annealing at high temperature but prior to the second irradiation. Since h

    not equal to h

    ations did not occur at the same absolute values of h

    first irradiation are shown in Figures 6 through 25.

    was

    except in rare cases, failures during the first and second rrradi- FE.2

    FEo 1 The failure patterns for the FE'

    5.1 ANNEALING AS A FUNCTION OF TIME AFTER SMALL DOSES OF GAMMA RAD1 AT1 0 N

    During the first irradiation the radiation was suspended for a period of about seventy-

    two hours during which the h of each o f two specimens o f each type - manufacturer group was monitored. Results are shown in Tables 4 and 5. Also during the first i r -

    radiation two specimens o f each type -manufacturer group were removed from the

    radiation environment after a dose of 9.67 x 10 r had been accumulated. The hFE

    o f each of these specimens was measured immediately after removal and again eight

    days later.

    FE

    5

    Results are shown in Table 6.

    The data i n Tables 4 and 5 show that i n the first three hours after cessation o f irradi-

    ation the type 2N914 generally recovers a small percentage of its lost h the type

    2N2192 shows essentially no change, and the types 2N2369 and 2N918 generally

    continue to lose small percentages of h FE' the 2N914 generally showed smal I percentage gains. It appears the magni tude of the

    changes i s sufficiently small and the change pattern i s such that the time interval be-

    tween cessation of radiation and measurement of h

    are maintained at room temperature.

    F E'

    After eight days (Table 6) a l l types except

    i s not crit ical i f the specimens FE

    5.2 RECOVERY OF hFE BY ANNEALING AT HIGH TEhrlPERATURE

    Tables 7 through 21

    FE tion (h ) to h

    the first irradiation FEp 1

    contain columns showing the ratio o f h after the first irradia- FE prior to the first irradiation (h

    and after annealing at high temperature (h

    ), and the ratio o f h after

    ) to hFE prior to FEO I FE

    FE- 3 U L

    14

  • the first irradiation (h

    through 14 received a total dose of 2.70 x 10

    listed i n Tables 15 and 16 received a dose of 9.67 x 10 r, and those l isted i n Tables 6 7 17 through 22 received varying amounts ranging from 3.26 x 10 r to 1.24 x 10 r

    (See Tables 6 and 7 i n Components Irradiation Test No. 18 Report).

    ) for each specimen. The specimens listed i n Tables 7

    r during the first irradiation, those FEo 1 7

    5

    A comparison of corresponding values of h /h and h /h reveals that

    i n every case but one (Table 21, specimen T 439) annealing at high temperature causes

    an increase in the h

    creased to a value greater than the original h The amount of the increase appears

    to be more a characteristic o f the type - manufacturer category than a function of the amount of degradation or a function o f the total dose received.

    FEY1 FEol FE02 FEO 1

    of gamma irradiated transistors. In some cases the h i s in- FE FE

    FE'

    5.3 A COMPARISON OF RADIATION TOLERANCES DURING FIRST AND SECOND I RRADl AT1 0 N S

    Tables 7 through 14 and 17 through 22 contain a column showing the ratio of the dose

    at failure during the second irradiation to the dose at failure during the first irradia-

    tion for each specimen. The failure points are defined as:

    First failure point i s the gamma dose:

    /2 ' - 01

    h~~ - h~~ when

    Second failure point i s the gamma dose:

    when hFE = h /2. FEo2

    The ratios of second fai l points to first fa i l points are i n most cases greater than one.

    However, there are many devices of the types 2N914, 2N2192, and 2N918 whose

    15

  • ratios are less than one. Thus a technique of irradiating to find the failure point and

    then annealing would not guarantee that the failure point during a second irradiation

    would be as great as, or greater than, the first.

    Rank

    1

    c n

    Such a technique may be applicable to certain types o f transistors. The data for the

    2N2369 and the S2N930 types suggest that one might be developed.

    Relative Value* Of

    0 0 0 h~~ 0 0 Te "EBO 0 h~~ Te e 'EBO T

    0 0 hFE

    ! I

    1 Large st I Largest ; Smallest Largest Largest

    c Smallest

    1 1 I ' 1 ' 4 Smallest 1 Largest Smallest + Smallest

    5.4 CORRELATION BETWEEN PRE-I RRADl ATlO N PARAMETERS AND RAD1 ATlON TOLERANCE

    and combinations (h T FE e I h~~ Te 'IEBO ) 0 0 0 0 0

    FE' Te' 'EBO Pre-irradiation values of h

    thereof were ranked as follows:

    A Spearman's rank correlation coefficient was computed for each category as men-

    tioned above in Section 4.0. These coefficients are shown i n Table 3. Only in the

    h category for the type 2N914 were the coefficients .60 greater for both val-

    ues of I This leads to the conclusion that these electrical parameters cannot be

    used to pre-select transistors for use i n a gamma radiation environment when h

    measured at constant I

    T FEO eo

    C' i s

    FE Reference 3 suggests the following relationship:

    C'

    16

  • 8

    where K i s a radiation damage constant for silicon transistors which varies

    with emitter current density, and

    A 5, i s radiation exposure.

    This expression has been shown to accurately depict the degradation of transistors i n

    the neutron radiation environment; the similarity of degradation characteris i c s be-

    yond the in i t ia l non-linearity (surface effects, reference 3) leads to the con ecture

    that the relationship may also be valid for gamma radiation.

    Using hFE/'hFE = 0.5 as a definition of failure, the above equation becomes: 0

    b dr (Failure Dose) = '/hFE Te K 0 0

    Hence, the first failure should correlate with the largest h

    wi th the next largest h T etc.

    T , the second failure FE e 0 0

    FE e 0 0

    I n the data obtained during this test only the type 2N914 showed any consistent cor-

    relation (Table 3) between h T and order o f failure. An attempt to explain why

    this may be so was made by computing the values of K for the 2N2192 specimens for FEo e

    0.

    = 5 mA. The computed values and the coefficients of variation for each type - manufacturer group are shown in Table 27. This table shows that while K i s fairly

    consistent within the type - manufacturer groups i t varies widely within the type. Noting this fact, Spearman's rank coefficients o f correlation were computed for the

    type - manufacturer groups and for the type - groups. Results are shown i n Table 28. Note that correlation i s excellent for some groups and wholly lacking or even nega-

    tive for others. These data point to the conclusion that K varies so widely among

    specimens in some type -manufacturer groups that the h

    IC

    T parameter cannot be FE e 0 0

    17

  • used as a tool for preselecting transistors for use i n a gamma radiation environment.

    5.5 RELATIVE DEGRADATION OF hFE AT LOW GAMMA DOSES

    Figures 26 through 33 show percent degradation of h

    Dose for each of the types 2N914, 2N2192, 2N2369 and 2N918 at each of two val-

    ues of I Based on the results o f Components Irradiation Test No.

    18, 5% of the median failure dose has been chosen as the best compromised screen-

    ing dose between higher doses which give better correlation and lower doses which

    give smaller decreases i n the radiation lifetimes of the devices.

    at 5% of the Median Failure FE

    5 mA and 30 mA. C'

    Note in Figure 26 that by selecting out those specimens with percent degradations

    o f h

    fai l are eliminated.

    dose at failure i s not perfect, one "good" specimen i s also eliminated by the select-

    ing out process. However, the important point i s that a l l the "worst" specimens are

    eliminated.

    larger than the median percent degradation value the first seven specimens to FE Since the correlation between percent degradation of h and FE

    Application of the "selecting out'' process to the data i n Figures 27 through

    similar results,

    Figures 34 a d 35 show percent degradation o f h at 5% of median failure

    a l l types (2N914, 2N2192, 2N2369 and 2N918) combined for IC values of FE

    33 yields

    doses for

    5 mA and

    30 mA respectively. Application o f the "selecting out'' process to the data i n these

    figures eliminates the first twenty-one failures i n Figure 34 (I = 5 mA) and the first

    nineteen failures i n Figure 35 (I C

    = 30 mA). C

    Tables 29 and 30 summarize the gains i n increased median failure dose and dose at

    first failure for the pre-selected specimens as compared to median failure dose and

    dose at first failure for the whole group. These tables show that the gains are quite

    18

  • substantial.

    Table 31 shows that a conservative estimate o f the minimum radiation lifetime (gam-

    ma dose at first failure) of the pre-selected specimens would be about eleven times

    the screening dose.

    Median percent degradation values at screening doses are tabulated for each o f the

    types i n Table 32. These values may be used as a guide in determining when to end

    the irradiation when screening a group of devices whose median failure dose i s not

    known.

    5.6 CONCLUSIONS

    ( 1 ) The correlations between the pre-irradiation electrical parameters h , F E,

    U , or combinations thereof, and radiation tolerance are not EBO, T and I sufficiently good for use i n the pre-selection o f transistors for use in a

    gamma radiation environment,

    (2) The recovery of h at room temperature (annealing) after small doses of

    gamma radiation between cessation of irradiation and measurement i s not

    a problem i n the proposed pre-selection method provided the measure-

    ments are made within two or three days.

    FE

    (3) Annealing o f transistors at high temperatures 200 - 3OO0C does change (increase or decrease) the i r radiation to I e rance characteristics .

    (4) The pre-selection o f silicon transistors for use i n a gamma radiation en-

    vironment can be accomplished by subjecting them to comparatively small

    doses o f gamma radiation. The proposed method increases the minimum

    radiation lifetime o f the pre-selected devices as compared to the minimum

    19

  • 8 radiation I ifetime of unscreened devices by factors ranging from about

    1 .5 to 50, depending on the failure distribution o f the unscreened de-

    vices. See Figure 36.

    5.7 RECOMMENDED METHOD FOR PRE-SELECTING NPN SILICON TRANSISTORS FOR USE IN A GAMMA RADIATION ENVIRONMENT

    Gamma irradiate to failure a sample of the type transistor under consid-

    eration sufficiently large to represent the population. (A sample of thir-

    ty or more i s recommended.) Measure h

    value of I

    that the l/hFE curve can be constructed for each dev ice from these mea-

    surements.

    at the proposed application FE prior to irradiation and at intervals during the irradiation x,

    C

    From the data obtained from the l/h

    transistor type l i k e those shown in Figures 26 through 33. This curve

    w i l l be the basis for determining the cut-off point for percent degrada-

    tion of h

    curves construct a curve for the FE

    in the pre-selection technique. FE

    The pre-selection technique i s as follows.

    Measure h

    at I C variation with I

    of a l l devices to be screened. Measurement should be made FE

    value equal to planned application value since radiation tolerance

    i s not the same for a l l specimens. C

    Gamma irradiate a l l devices to 5% of median failure dose as determined

    from sample irradiated above. Measure h and compute percent degra-

    dation of h FE

    for each device. FE

    Refer to curve constructed i n paragraph (2) above to determine percent

    20

  • degradation of h corresponding to desired minimum failure dose. E l i -

    minate al I devices having percentage degradations of h

    this value. The remaining specimens w i l l be the pre-selected specimens.

    FE larger than FE

    5.8 RECOMMENDED METHOD FOR ESTIMATING MINIMUM GAMMA RADIATION LIFETIME OF NPN SILICON TRANSISTORS

    Because of the characteristic shape of the l / h

    (See Figure 5) i t i s possible to estimate the minimum dose necessary to give any value

    of l/hFE once the early portion of the curve is known. Of course, the greater the

    known portion of the curve the more accurate the estimate w i l l be.

    curve in a radiation environment FE

    The estimate can be made by computing the slope o f the right hand end of the known

    portion of the curve and extrapolating this slope to the selected value of l/h FE'

    The technique can be best illustrated by example. Let us suppose that we need some

    2N914 transistors to be used i n an application requiring a minimum radiation lifetime

    of 8.0 x 10 r. We have on hand several 2N914 devices. To estimate their radia-

    tion lifetimes we proceed as follows:

    7

    Gamma irradiate a l l devices to 5% of the required minimum dose (5% x

    8 x 10 r = 4 x 10 r). 7 6

    Measure h

    point as h . of each device and compute l /h FE FE' Designate hFE at this

    FE1

    Continue gamma irradiation of a l l devices to a total o f 10% of the re-

    quired minimum dose (10% x 8 x 10 7 6

    r = 8 x 10 r).

    Measure h of each device and compute l /h Designate h at this FE FE' FE

    21

  • point as h . FE2

    (All hFE measurements should be made at planned application I values.) C

    (e) Select failure criterion. Let us suppose that i t has been selected for us

    as 50% degradation o f h on the Q-fac-

    tor curve.

    which corresponds to 2/h E2 E2

    The slope of the right hand end of the known portion of the l /h

    i s now found as follows.

    curve FE

    6 (f) Slope = (l/hFE - I/h )/(8 x l o6 - 4 x 10 ).

    2 FEl

    Next we use this slope to extend the curve to the failure point and esti-

    mate the failure dose

    6 ) (8 x l o6 - 4 x 10 ) (*IhFE 2 - 1’hFE2

    1 /h - l /h (9) Estimated failure dose = FE2 FE1

    - - l/h - l /h

    E2 FEl

    Once we have found the estimated failure dose for each device we can

    select those which meet our requirements, and relegate those remaining

    to applications less rigorous.

    Table 33 shows estimated failure doses for the 2N914 devices irradiated

    during this test along with corresponding actual failure doses found from

    the experimental data. The ratios of estimated to actual failure doses

    22

    ~

  • confirm that the estimated dose w i l l always be a conservative estimate.

    5.9 RECOMMENDED FOLLOW-ON INVESTIGATIONS

    I t i s recommended that experiments be performed to determine whether or not the

    method i s applicable to failures caused by neutron irradiation, and that the data al-

    ready obtained, as well as any future data obtained, be analyzed to determine the

    pre-selection method's applicability to other definitions of failure - such as h degradation to an absolute value.

    F E

    23

  • I t 8 1 I 8

    REFERENCES

    1. Ashar, K. G., "Transit Time in High-speed Switching Transistors, I' Solid State

    Design, Vol. 5, No. 2, February 1964.

    2. Croxton and Cowden, Applied General Statistics, Second Edition, Prentice-

    Hall, Inc., Page 478.

    3 . Frank, Max, "Exploratory Development of the Q-Factor Technique, I' The Ben-

    d i x Co rpo rat ion, Re se a rc h La bo ra to r i e s Division , So ut h f i e I d, Mi c h i ga n, Tech - nica l Report AFWL-TR-65-166.

    25

  • I I I I 1 I I I I I I I 1 8 I I I 1 1

    24

    TABLE 1 TEST SPECIMENS AND TEST CONDITIONS

    - -~ VCE = 5V, IC = 30 rnA

    VEB = 4 v

    I = 1, .5, .33, .2 & .125mA E

    Description

    T ransi s to r

    2N914

    NPN, S i

    Planar Epitaxial

    *

    Transistor

    2N918

    NPN, S i

    Planar Epitaxial

    *

    Transistor

    2N2192

    NPN, S i

    Planar Epitaxial

    *

    Transistor

    2N2369

    NPN, S i

    Planar Epitaxial

    *

    T ran s i s to r

    S2N930

    NPN, Si, Planar

    Fairchild (40)

    Tex. lnst. (17)

    No.1 Test Conditions

    24

    VCE = 5V, I C = 5 rnA VCE = 5V, IC = 30 rnA

    IE= 1, .5, .33, .2 & .125rnA

    I Signal at 600 kc

    24 1 IE= .33, .25, .2G, .I67 & .125 rnA

    Signal at 600 kc

    VCE = 5V, I C = 5 rnA

    I Signal at 600 kc VCE = 5V, IC = 5 rnA

    VCE = 5V, IC = 30 rnA

    24

    I = 1, .5, .33, .2 & . I25 rnA E I Signal at 600 kc = 5V, 1 = 2 rnA

    =5V, I = 10rnA 'CE C

    'CE C

    Signal at 600 kc

    ~~

    Pararnete r

    hFE

    h~~

    'EBO

    X T

    h~~

    h~~ I

    'EBO

    X T

    h~ E

    h~~

    'EBO

    X T

    h~~

    h~ E

    I EBO

    X T

    h~~

    FE h

    I

    'EBO

    X T

    *Six from each of Fairchild, Motorola, Texas Instruments and General Electric

    27

  • TABLE 2 MANUFACTURERS' SPECIFICATIONS FOR TEST SPECIMENS

    Description I 2N914

    NPN, Silicon

    Double-Diffused

    Planar Epitaxial

    2N918

    NPN, Silicon

    Planar Epitaxial

    2 N930

    NPN, Sil icon

    Planar

    2N2192

    NPN, Silicon

    Planar Epitaxial

    2N2369

    NPN, Silicon

    Planar Epitaxial

    Conditions

    VCE = lV, IC = 10 mA

    IC =o, v E B = 4 v

    fob

    VCE = lV, I C = 3 mA

    fob

    VCE = 5V, IC = 10 mA

    IC =o, v E B = 5 v f Gain-Bandw idth Product T

    -

    VCE = lOV, IC= 10 mA

    IC VEB = 3v

    f Gai n-Bandw idth Product T

    VCE = lV, I C = 10 mA

    fab

    Specification

    hFE= 30 to 120

    'EBO = 0.1 MA max

    480 M c

    hFE = 20 min

    960 M c

    hFE = 600 max

    = 10 nA max I EBO 30 Mc min

    hFE = 75 min

    = 50 nA max 'EBO 130 Mc

    hFE = 60 to 120

    800 M c

    1 1 1 1

    I I 1 I I I 1 I 1 1 I I I 1

    a

  • W CY

    J 3

    z LL

    0" 2 ce W

    0 m v, & W

    3

    > - - & W I-

    ? 6 Q e- 0

    n

    &

    LL

    CY W

    5 e LL I-

    W

    0

    Z V - - LL LL W

    0 V Z 0 6 - I- 4 W & &

    0 V 2 Z 6

    Z

    CY m -

    2 2 W L m m W -I

    7 I-

    (u u)

    03 9

    9 m

    3 2

    3 %

    -t m

    u) 0

    0. *

    0 W L

    II

    I -

    I\ u)

    I

    (u (u

    I

    9 0 I

    I

    I

    N * I

    2

    9 u)

    0 W

    I-

    f N 6

    u) II V -

    9 0 - t I

    - 0

    m * 8 I

    0. d I

    0

    0 -

    a3 (u

    I

    N -t I

    oo rn - UJ

    L

    c W W

    L 5 0 a

    - c 0

    I

    (u 0 h m

    - u)

    2 %

    m a3 0 0

    I

    m 7 (u (u

    I I

    u) 0. h m

    u) u) 9 rr)

    O0

    W T . I- 0 u o

    m o w

    LL I: 0

    W LL L

    29

    I

    9 0.

    0 a3

    a3 0.

    (u 0.

    I

    c 9

    x

    Q E 0 m II v -

    6 E u) II V -

  • L

    E .- V al Q v,

    d 0.

    hl

    c

    Z

    30

    I 1 1 I I I I I I I I I 1 I I I I I

  • I

    L

    E .- V al Q v,

    0 0 00

    c m w w LL LL I: I:

    31

  • IT 9)

    E .- V al Q v,

    V al a v,

    0. 9 m N

    c\l Z

    d 0.

    (u

    - Z

    co 0.

    (u

    - Z

  • TABLE 7 2N914, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS

    AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-

    TION, ANNEALING AND FAILURE DATA

    1 . 1 1

    .88 1.22

    .97

    .94

    Specimen

    77 .87

    1.8 7.78 2.0 1 .15 2.8 1.75 2.6 2.73

    F2 F3 F4 F5

    M2 M3 M4 M5

    T2 T3 T4 T5

    G 2 G 3 G4 G 5

    01 h~~

    57.5 64.1 80.6 64.1

    69.4 63.3 79.4 74.6

    ~

    50. C 50.5 64.1 55.6

    62.5 96.2 82.9 59.5

    PI h~~

    .59

    .56

    .67

    .63

    .33

    .34

    .36

    .37

    .61

    .59

    .67

    .66

    .46

    .37

    .57 -48

    Point

    1.21 .52 1.08 I 1: 1 1.61 1.11 .34

    - Beta prior to first i r rad ia t ion . - Beta a f t e r f i rs t i r radiat ion to 2.70 x 10 7

    01

    P I

    F E h

    r . h~ E

    h - Beta a f t e r a n n e a l i n g a n d pr ior to s e c o n d i r rad ia t ion . FEo2

    1st Fail Point - Dose w h e n hFE- -hFE /2. 01

    2 n d Fail Point - Dose w h e n hFE = h /2. FEo2

    33

  • TABLE 8 2N914, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS

    AND GENERAL ELECTRIC, (Ic = 3C mA), hFE DEGRADA-

    TION, ANNEALING AND FAILURE DATA

    2.9 8.0 9.9

    10

    -7-c Specimen

    1 1 .,-: .76

    1.82 1.80

    F2 55.8 F3 60.0 F4 72.1 F5 61.5

    I M2 69.9 M3 60.5 M4 77.7 M5 72.8

    T2 T3 T4 T5

    G2 G3 G4 G 5

    60.5 60.4 84.2 70.3

    57.6 83.1 75.6 54.0

    .65 1.09

    .61 1 .c3

    .70 1.09

    .67 1.02

    ---I--- .40 .88 .45 1.17 .40 .91 .41 .89

    1 .71 .86 .72 .94 .70 .91 .68 .84

    1 s t Fai l Point

    2nd Fai l Poini 1 s t Fai l Point

    61 48 72 75

    .89 1.54

    .50 1.07

    i

    100 146 97

    104

    1.59 .87

    1.21 .85

    I 61 1.69

    27 I 2.04 82 1 .oo 1.43

    68 I - Beta prior to first irradiation. - Beta after first irradiation to 2.70 x 10 - Beta after annealing and prior to second irradiation.

    7 01 h~ E

    h~ E

    FEo2

    r.

    P l h

    1s t Fail Point - Dose when hFE = hFE /2. 01

    FE FEo2 2nd Fail Point - Dose when h = h /2;

    34

    ~

  • 1 1 : . 'I I I I I I I I I I I I I I I I I I

    TABLE 9 2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS

    AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-

    TION, ANNEALING AND FAILURE DATA

    1 s t Fail Point

    (x lo6 r) 1.8 0.4 1.7 2.7

    1.2 0.9 0.2 1.7

    30 48 21 28

    9.5 32 27 1 1

    - Beta prior to first irradiation.

    - Beta after first irradiation to 2.70 x 10

    - Beta after annealing and prior to second irradiation.

    - Dose when h

    7 01 h~~

    h~~

    FEo2

    r.

    P l h

    1 s t Fail Point

    2nd Fail Point - Dose when hFE= h /2

    01 FE = h~~

    /2. FE02

    35

    2nd Fail Point 1 s t Fai l Point

    1.44 4.50 1.82 1.19

    .58

    .78 1.50

    .82

    1.30 1.17 1.86 1.46

    2.84 1 .oo 1.48 2.91

  • TABLE 10 2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS

    AND GENERAL ELECTRIC, (Ic = 30 mA), hFE DEGRADA-

    TION, ANNEALING AND FAILURE DATA

    105.3 169.5 193.4 123.5

    163.9 172.4 227.3 104.2

    Specimen

    .34

    .23

    .31

    .36

    .26

    .25

    .21

    .34

    F2 F3 F4 F5 4.0

    1.7 2.2

    .4 7.1

    M2 M3 M4 M5

    1.20

    1.18 .95

    2.00 .44

    T2 T3 T4 T5

    25 35 14 16

    G2 G3 G4 G5

    1.40 1.43 2.36 2.13

    h~~

    142.9 137.6 185.2 153.1

    140.8 141.4 133.3 142.9

    .48

    .55

    .38

    .41

    .40

    .54

    .47

    .41

    I

    1 1 34 24 14

    3.00 1 .oo 1.63 2.71

    h FE

    .91

    .95 1.03 1.12

    01

    1.01 1.04 1.04 1.15

    .74

    .84

    .61

    .65

    .68 1.01

    .79

    .67

    1 s t Fai l Point

    x 10 r

    2nd Fai l Point 1 s t Fai l Point

    1.04 E:; I 5.20 3.2 1.28

    - Beta prior to first irradiation. 01 h~ E 7 - Beta after first irradiation to 2.70 x 10 r. P I

    h~ E

    h

    1st Fail Point - Dose when h = h / 2 .

    - Beta after annealing and prior to second irradiation. FE02

    FE FEol

    2nd Fail Point - Dose when h = h /2. FE FEo2

    36

    1m I I I I 0 I I I I I I I I I I I I I

  • TABLE 1 1 2N2369, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS

    AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-

    TION, ANNEALING AND FAILURE DATA

    .50

    .65

    .49

    .84

    .40

    .37

    .45

    .32

    .57

    .53

    .55

    .58

    .58

    .70

    .53

    .67

    I hFE 01 Specimen .74 .77 .71

    1.03

    .51

    .53

    .54

    .49

    .77

    .77

    .76

    .73

    .89

    .88

    .80

    .92

    F2 I 90.9 F3 F4 F5

    94.3 92.6 51 .O

    M2 M3 M 4 M5

    T2 T3 T4 T5

    102.0 74.6

    1 1 1 . 1 83.3

    68.5 102.0 76.9 53.2

    G 2 G3 G 4 G 5

    63.3 50.5 64.1 64.9

    1 s t Fail Point

    6 (x 10 r)

    27 80 25

    250

    5.3 2.6

    1.8 14

    62 34 51 69

    49 110 35

    100

    !nd Fai l Point 1s t Fai l Point

    1.89 1.49 4.00 1 .oo

    17.36 33.46 7.86

    30.56

    1.66 2.21 2.31 3.26

    1.49 1.55 2.20 1.51

    - Beta prior to first irradiation. 01 h~ E

    P l h~ E

    7 - Beta after first irradiation to 2.70 x 10 r. h

    1 s t Fa i l Point

    - Beta after annealing and prior io second irradiation.

    - Dose when hFE = FEo2

    /2 01 h~ E

    2nd Fai l Point - Dose when hFE = h /2. FE02

    37

  • TABLE 12 2N2369, FAIRCHI LD, MOT0 WLA, TEXAS INSTRUMENTS

    AND GENERAL ELECTRIC, (Ic = 30 mA), hFE DEGRADA-

    TION, ANNEALING AND FAILURE DATA

    1 s t Fail Point 2nd Fai l Point

    .73

    .75

    .73

    .86

    .56

    .58

    .56

    .55

    .81

    .75

    .78

    .82

    .84

    .85

    .79

    .84

    42 4.10 97 1.57 55 2.60

    180 1.45

    21 4.67 16 9.69 29 3.90 13 1.85

    140 2.79 73 2.63

    120 2.74 120 2.13

    86 1.66 150 1.73 50 2.22

    170 1.46

    - Beta prior to first irradiation. - Beta after first irradiation to 2.70 x 10 - Beta after annealing and prior to second irradiation.

    - Dose when h

    7 01 h~~

    h~~

    FEo2

    r.

    P l h

    1 s t Fai l Point

    2nd Fai l Point - Dose when hFE = h

    /2 01 FE = h~~

    /2. FEo2

    38

  • ~ ~~ ~~ ~

    TABLE 13 2N918, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS

    2nd Fai l Point 1 s t Fail Point

    I

    .41 1.54 .57 .55

    .86 1.30

    .65 1.29

    1.22 1.26

    .60

    .99

    A

    1.22 4.13

    .73 1.45

    AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-

    TION, ANNEALING AND FAILURE DATA

    Specimen

    F2 F3 F4 F5

    M2 M3 M4 M5

    T2 T3 T4 T5

    01 h~~

    G2 G3 G 4 G 5

    - Beta prior to first irradiation. 7 - Beta after first irradiation to 2.70 x 10 r.

    P l h~~

    h - Beta after annealing and prior to second irradiation. FE02

    1 s t Fail Point - Dose when h - /2 2nd Fai l Point - Dose when h = h /2.

    01

    FE FEo2

    FE - h~~

    39

  • TABLE 14 2N918, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS

    AND GENERAL ELECTRIC, (Ic = 30 mA), hFE DEGRADA-

    TION, ANNEALING, AND FAILURE DATA

    62.1 48.4 57.7 50.4

    Specimen

    .77

    .72

    .75

    .68

    F2 F3 F4 F5

    71.9 100.0 103.1 100.3

    M2 M3 M4 M5

    .65

    .64 0 59 .59

    T2 T3 T4 T5

    48.2 48.2 79.8 61.5

    68.5 G 2 G3 G 4 G 5

    .74

    .76

    .63

    .69

    01

    hFEo2

    hFE 01 1.01 1.02

    .99 1.02

    .99

    .93

    .98

    .95

    .99 1 .oo

    .99 1.07

    1st Fail Point

    (x lo6 r)

    300 82

    1 50 91

    82 85 87 75

    150 140 64 93

    No Useable Data

    2nd Fail Point 1 s t Fail Point

    .72 1.96

    .83 1.60

    1.54 1.79

    .78 1.85

    1.71 1.71 1.31 1.32

    (All Specimens, 9 Including Control Specimens,

    58.8 Very Unstable)

    1 I I I - Beta prior to first irradiation.

    - Beta after first irradiation to 2.70 x 10 7 01 h~ E

    h~ E r. P l

    h - Beta after annealing and prior to second irradiation. FEo2

    1s t Fai l Point - Dose when hFE - - h~~ /2 9 2nd Fai l Point - Dose when h = h /2.

    01

    FE FEo2

    40

  • 41

    7-

    ambo O O N O

    . . . a I

    c L ' m u ) c o m - - 0

    . . . a

    a N h h 0 - m o . . . a

    - - I - -

    9 b L L L L

    N 0.

    N

    N

    - Z

    0 0 . .

    h a 0.h . .

    h 0 ma . .

    0 0

    a * . . - -

    o c a - w n

    * ,

    - h a 0 . ,

    t ",I"" o m a c 90 c o o . . -

    NC: h a

    L

    u) 0

    X h a a

    0

    -

    Lc

    ; - 0 -I- e 0

    S 0

    O -0 0

    -I-

    .- -I- .- L L .- c v) L

    v- L

    c v-

    .- al

    0 w w L L L L S L II I I

    C 0

    O -0 0

    .- -I- .- L L .-

    -0 c ou 2 e h

    L 0 .- -0 C 0 0 C

    0 W C C 0

    W

    0

    .- -

    L

    c Lc

    W LL L

    I I - - N

    Q O wo w w L L L L L

    S L L

  • Z cv

    a)

    0.

    cv

    c

    Z

    42

    C 0

    0 -a 0

    , - c

    ,-

    L L 8 -

    .I- v) L

    .I- ,-

    e L 0

    C i ,- L

    L

    ul 0

    X h 9 0.

    0 W 0 U

    0

    0

    -

    Y

    In

    - + c

    e S 0

    0

    -0

    .- c .- 2 L

    .I- v) L

    Y L

    .I- u-

    .-

    .- W

    0 w w L L C L I1 II

    C 0

    0 U

    .-

    .I- .- 2

    8 x

    L .- U S

    0

    0

    c L .- & U S 0 m S

    0 W S S 0

    W

    0

    .- -

    L

    c u-

    w u,

    5

    I1 - - c v

    Q O wo w w L L L L L L

    L L 5

  • TABLE 17 S2N930, FAIRCHILD (Ic = 2 mA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    Specimen

    AF046 AF 508 AF985 AF 1083 AF1213

    BF210 BF373 BF1316 BF 1875 BF2173

    F047 F052 F055 F066 F114 F160 F197 F217 F229 F263

    01 h~~

    488 526 400 435 435

    385 57 1 426 377 377

    435 392 38 5 526 488 385 444 51 3 408 408

    P l h~ E

    hFE 01 .32 .34 .29 .39 .37

    .39

    .31

    .36

    .33

    .31 ~

    .38

    .24

    .26

    .31

    .31

    .31

    .33

    .37

    .39

    .36

    hFE 01 .46 .42 .50 .46 .51

    .52

    .39

    .47

    .48

    .48 ~

    .51

    .46

    .47

    .38

    .41

    .47

    .45

    .43

    .49

    .45

    ~~

    1 s t Fai I Point

    6 (x 10 r)

    .31

    .54

    .76

    .80

    .70

    .74

    .44

    .05 2.40

    .33 ~ ~~

    .54

    .58 1.37 1.50

    .51

    .45

    .22

    .80

    .86

    .61

    !nd Fail Point 1 s t Fa i l Point

    39.0 27.4 18.3 23.0 21.1

    23.6 28.9

    324.0 8.8

    42.1 ~

    23.5 31.7 14.8 19.2 24.3 20.7 63.2 24.5 21.4 31.3

    - Beta prior to first irradiation.

    - Beta after first irradiation. 01 h~~

    h~ E P I

    h

    1 s t Fail Point - Dose when hFE = h /2.

    2nd Fail Point - Dose when h = h /2.

    - Beta after annealing and prior to second irradiation. FE02

    FEo 1

    FE FEO2

    43

  • TABLE 18 S2N930, FAIRCHILD (Ic = 2 rnA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    S pe c i rn en

    F 296 F 299 F345 F 399 F446 F475 F 644 F 648 F687 F1174 F1232 F1317 F 1370 F1389 F1511 F 1 605 F 1 665 F 1898 F 1987 F2195

    01 h~ E

    -

    408 455 444 465 370 435 385 465 408 455 51 3 377 408 476 476 3 70 455 476 385 392

    P l h~~

    h ~ ~ o l

    .27

    .23

    .32

    .35

    .36

    .32

    .31

    .31

    .34

    .38

    .32

    .31

    .36

    .34

    .33

    .40

    .25

    .35

    .40

    .28

    h FEo2

    .45

    .44

    .41

    .43

    .49

    .46

    .47

    .39

    .45

    .44

    .39

    .48

    .45

    .38

    .42

    .54

    .40

    .47

    .58

    .57

    1 s t Fai I Point

    6 (x 10 r)

    .89

    .83

    .67

    .58 3.25

    .40

    .89

    .89

    .64

    .59

    .44

    .33

    .61

    .38

    .31

    .86

    .94

    .59

    .76

    .24

    2nd Fa i l Point 1 s t Fa i l Point

    22.8 24.5 13.9 35.0 8.2

    34.8 25.8 45.4 25.3 47.1 52.3 53.0 37.7 86.0 65.5 20.4 36.8 27.5 19.5 38.8

    - Beta prior to first irradiation, - Beta after first irradiation.

    01

    P I

    h~~

    h~ E

    h - Beta after annealing and prior to second irradiation. FEo2

    1 s t Fa i l Point - Dose when h - /2 01 FE - h~~

    2nd Fail Point - Dose when h = h /2. FE FEo2

    44

  • TABLE 19 S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    Specimen

    AF046 AF508 AF985 AF 1083 AF1213

    BF210 BF373 BF1316 BF 1875 BF2173

    F047 F052 F055 F066 F114 F160 F197 F217 F229 F263

    469 538 405 426 422

    43 1 568 459 361 389

    ~

    433 364 392 529 493 3 58 452 472 406 41 3

    h~~ P I

    hFE 01 .42 .42 .38 .49 .47

    .45

    .39

    .46

    .43

    .39

    .46

    .35

    .35

    .43

    .41

    .42

    .41

    .48

    .47

    .46

    hFEo2

    01

    .76

    .77

    .73

    .78

    .79

    h~ E

    .77

    .73

    .66

    .73

    .73

    .83

    .74

    .77

    .65

    .65

    .73

    .76

    .79

    .82

    .73

    1st Fai I Point

    6 (x 10 r)

    1.46 1.60 4.20 3.00 2.50

    2.00 1 .oo 2.00 5.58

    .94

    2.30 3.90 4.20 4.90 1.70 1.68 1.38 2.84 2.63 2.40

    - Beta prior to first irradiation.

    - Beta after first irradiation. 01 h~ E

    h~~ P l

    h

    1 s t Fail Point

    2nd Fail Point - Dose when h = h /2.

    - Beta after annealing and prior to second irradiation.

    - Dose when h FE02

    /2 01

    FE FEo2

    FE = h~~

    45

    2nd Fail Poini 1 s t Fa i l Point

    2.73 1.83 2.30 2.38 2.32

    2.81 3.17 5.72 2.42 4.47

    1.71 2.55 1.96 2.83 2.72 2.74 2.74 2.04 1.87 3.07

  • TABLE 20 S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,

    ANNEALING AND FAILURE DATA

    Specimen

    F 296 F 299 F 345 F 399 F446 F475 F 644 F 648 F687 F1174 F 1 232 F1317 F 1 370 F 1 389 F1511 F1605 F 1 665 F 1898 F 1 987 F2195

    01 h~ E

    3 72 437 406 450 368 405 379 44 1 377 472 51 3 364 397 488 465 366 429 444 389 424

    P l h~~

    hFE,I

    .37

    .32

    .46

    .48

    .45

    .41

    .39

    .40

    .48

    .46

    .39

    .41

    .45

    .44

    .43

    .49

    .35

    .45

    .49

    .34

    h FEo2

    .79

    .76

    .79

    .72

    .80

    .82

    .83

    .78

    .83

    .82

    .75

    .86

    .84

    .76

    .83

    .88

    .80

    .87

    .92

    .84

    1 s t Fai l Point

    6 (x 10 r)

    3.70 3.28 2.30 2.70 7.00 1.30 3.40 3.88 2.73 2.25

    .98 1.40 2.15 1.90 1.50 3.02 3.25 2.24 2.95 3.65

    2nd Fail Point 1 s t Fa i l Point

    46

    1.67 1.98 1.50 2.34 1.42 2.41 1.29 1.86 1.53 1.76 3.60 2.32 1.78 2.45 2.21 1.39 1.69 1.55 1.33 1.20

    - Beta prior to first irradiation.

    - Beta after first irradiation. 01 h~~

    h~ E P l

    h - Beta after annealing and prior to second irradiation. FEo2

    1 s t Fai l Point - Dose when h - /2 01

    FE FEo2

    FE - h~~ 2nd Fai l Point - Dose when h = h /2.

  • TABLE 21 S2N930, TEXAS INSTRUMENTS (Ic = 2 mA), hFE DEGRA-

    DATION, ANNEALING AND FAILURE DATA

    Specimen

    AT 1 203 AT 1 262 AT 1 263

    BT986 BT 1 209 BT1401

    T1136 T1144 T1154 T1188 T1216 T1311 T1370 T1397 T 1 406 T1439 T1484

    01 h~~

    278 3 28 227

    244 222 31 2

    235 41 7 303 377 263 225 230 263 235 31 7 247

    P I h~ E

    .57

    .43

    .54

    .52

    .36

    .41

    .39

    .32

    .54

    .37

    .37

    .61

    .67

    .44

    .53

    .67

    .43

    h FE02

    hFE,I .60 .55 .63

    .63

    .69

    .58

    .65

    .48

    .66

    .53

    .69

    .81

    .73

    .63

    .71

    .63

    .68

    1 s t Fai I Point

    (x l o6 r)

    17.60 8.95

    14.20

    13.40 2.80

    .50

    1.90 3.80

    15.70 2.00 6.10

    20.50 25.50 8.10

    13.10 24.10 7.20

    2nd Fai l Point 1 s t Fai l Point

    ~~~ ~~

    2.72 2.38 2.75

    2.80 6.25

    35.80

    10.89 5.34 2.39

    11.50 3.03 2.36 1.88 3.10 2.16 2.46 3.64

    - Beta prior to first irradiation.

    - Beta after first irradiation. 01 h~~

    h~ E P I

    h

    1 s t Fail Point

    2nd Fail Point - Dose when hFE = h

    - Beta after annealing and prior to second irradiation.

    - Dose when h FEo2

    /2 01 FE = h~~

    /2. FE02

    47

  • TABLE 22 S2N930, TEXAS INSTRUMENTS (Ic = 10 mA), hFE DEGRA-

    DATION, ANNEALING AND FAILURE DATA

    Spe c i me n

    AT 1 203 AT 1 262 AT 1 263

    BT986 BT 1209 BT 1401

    T1136 T1144 T1154 T1188 T1216 T1311 T 1370 T1397 T 1 406 T 1 439 T 1 484

    01 FE h

    287 332 238

    25 1 239 3 22

    2 50 433 31 3 380 282 233 242 267 243 344 258

    P I h~ E

    5 .63 .56 .61

    .58

    .49

    .5t:

    .56

    .46

    .55 0 52 .55 .65 .71 .56 .62 .69 .55

    h FEo2

    hFE,I .83 .81 .81

    .83

    .80

    .82

    .85

    .70

    .80

    .75

    .82

    .81

    .83

    .80

    .92

    .83

    .81

    1 st Fai I Polnt

    (x lo6 r)

    22.40 14.90 19.50

    17.70 11.20 8.50

    17.20 10.10 22.80 13.50 13.70 24.00 29.90 16.40 19.80 26.40 15.40

    2nd Fail Point 1 s t Fail Point

    1.52 1.34 1.38

    1.41 1.82 2.19

    2.12 1.90 1.58 1.84 1.79 1.48 1.23 1.53 1.26 1.34 1.68

    - Beta prior to first irradiction. - Beta after first irradiation.

    01 h~~

    h~~ P l

    h - Beta after annealing and prior to second irradiation. FEo2

    1 s t Fail Point - Dose when h = h /2. FE FEol

    2nd Fail Point - Dose when h = h /2 FE FEo2

  • 0.-o TtMhO. I - m . - a3--0.03 o* c o o o m w k - - 9 m m - . . . . . . . .

    0 W L I:

    m * m m 9-90

    9903k m o o k m . . . . . . . 0°*0 L m m *

    0 3 0 - m 0 l u ) k C o m o m - 0 . O k C v 109h9 9 9 k k

    . . . . . . . .

    0 0

    P

    x I - &

    oo m - 0 9 9 u ) m

    0 W LL C

    . . . . I . . . . . . . . . a m . - - m m h l

    - L 3 I W 4 - 0

    m - I - v

    w 0 : L -

    * o c o O . m 9 m 0 . 0-99 m - 0 . - m h l m h l m m m m . . . . . . . .

    0 W L

    -r

    * O h 0 I 0 - m - . . . . 0 3 0 9 m A d d d I " i?.q? *-09 . . . .

    V ale z

    I - &

    m 0 . m o * o - - 0.--hlm **o- 9 k 9 9 9 9 k k . . . . . . . .

    > 0 * II m

    C :

    49

  • O h 0 0 O o h l h 9 9 h - h hl - c u

    o o o m m o m h l

    0 7 - m m 9

    0 3 0 9 m - m m - W 0 4

    u . 5 L

    0 w U I

    m o o o 0 3 9 0 0 . - - o h m m

    m o o o 25%%

    O h 0 0 m m o n - m * -

    -009 m - o o m h l m h l

    0 0 9 h

    C 4 c h - h 0 3 - m m

    . . . . - -

    o m * m . . . . 009hl . . . . 0 w LL I

    * O m d h l

    0 0 0 0 . . . . hl-l-hl

    0 0 0 0 . . . . - m h l -

    0 0 0 0 . . . . - - - - . . . .

    0 0 0 0

    V P)O 4

    h . 5 . . . .

    hl--hl . . . . c y - -

    > 0

    * II m w

    2

    C

    E .- u e, a v,

    50

  • 1 T - R 1 I 1

    I lil

    ( u m * o 00- - a s s * a 9 -0 r\ d o 0

    - L a - . . . . m a *

    h m h m - - * N

    . . . . . . . . . c

    co - ( u 9 m c o d 0 7 m s m -

    O h - ( u N N N N

    . . . . . . . . c o * s m m m m -

    -

    o x L L . 5 W

    4 < I- n

    0 W u- I n Z Q

    O0 m W

    c1 . 0

    aJ I-

    u' z U I-

    W -4 w -I

    oi W

    W

    < Z (3 0 Z Q

    - m m 9 o o * o m o m k . . . . o * m o -ma)-

    O h O h I\ocos . . . . L Z Z G 1 ( u m - 0 m a c o m . . . . 0 W u-

    I o o m m a m s s - I - - I - N m m m - 9 s m 0 0 - ( u . . . .

    0 W - 0 u

    LL - -

    n

    . . . . . . sod h d

    co h

    0 W n

    I - v w 0 : L L - I

    m m * n . . . . m m m h - - o m ( u - N -

    . . . . s m * m m h d o . . . . - - h - 0 Nh("

    m m 9 0 O d N - m 0 . o . m . . . . o s - m m o m m m m - m 0 w

    LL x

    . . . . ( u * - m O h - a 3

    . . . . c o m s m a o h m & 6 4 4 s m s s

    cv - h * W 0 0 0 - . . . . (u-l- . . .

    m o o 0 (u

    O O O ( u o * m m h * -

    c o o 0 0 90 h c o -

    c9

    n V

    aJo x I - 5

    m - h m m h m o - 0 0 0 . . . . . . . .

    C

    E

    51

  • 0 h aJ 0-

    ' 0 w o m ' < w v L L - aJ

    v) v

    I 0 3 0 3 0 3 . . . a m . . h m * o m 0.h hl m

    0 W

    - * h * 0 . 0 - m h l m c o m m m h o - o m 0 03030.- 9 d m m h 0 0 0 * * h a . . . . . . . . . . . .

    LL -r

    d 9 I I a h -

    0 ( U h

    I-- v aJ 0 - w c

    L L - x

    m o m 0 . . . .

    m m m o . . . . 0 w

    LL 2

    ~ ~~~

    - - m m * * a m m * m m ohloo 0000 0000

    . . . . . . . . . . . . - 0 m - o m 0 0 . . . .

    .- V 0 P m

    52

  • v u a cu'

    OI

    W

    03

    c7 09

    m 9 m ch

    0 9 0

    0 d -

    I-

    53

    m m * m

    m hl hl

    0

    03 d 9 c

    h l r n d l n ln c S

    E" 3

    S

    0 X a I-

    L c ln

    CI

    ln

    hl m hl

    0

    hl

    c703om c7mcoco hl-hlhl . . . .

  • Cd W

    2 0

    7 II

    z z

    v, 7 a, - 0

    I--

    3 wo C d L L < I : W % E

    2

    co hl W J

    I-

    co 0.

    hl

    - Z

    0. 9 m hl

    N Z

    hl 0.

    hl

    (u

    c

    Z

    d 0.

    hl

    c

    Z

    I

    W 0 I

    0 c

    co 0

    - 03

    0

    9 0 I

    54

    09 0

    d 0

    - 0 c

    - hl

    0

    hl

    0 I

    co 0

    hl h

    0

    rn

    0

    c

    m hl

    I

  • o?: W I-

    9 cd

    (3 Z 0 F: Q n

    (3 a

    2

    5 2

    2 2

    W

    W

    W a r I-

    W z e(

    U W a v,

    I- 3 0 (3 Z i= U W A W v,

    >- m

    N i Q

    Z rz (3

    a W

    W cd

    0. hl W -I

    2 +

    c .- h 08 (3-

    L

    c (I) w L 5 0 a

    w Q x I-

    - m

    h l m 0 3 9 c

    - \ h 00

    x x hhl d - 9 m

    c c

    . .

    - \ n 0 0

    x x m03 m - m

    c c

    . . -

    .. .. $ 2 a = 0 3

    U

    > e - v ) .- L O K

    U S

    LL

    64

    s a .-

    - u7 -

    d - 0. Z hl

    h( 0.

    hl

    hl

    c

    Z

    0 m

    55

    03 - Cr 7 L hl

    In c a, E .- V e, 0 V .- L c V 0)

    W

    0 e, c

    - - L

    3 In In

    w -I

    *

    % 0 U C 0

    V e,

    .- c

    - 0 2

    P

    I

    a L

    -0 e, V c

    E L

    Ou C w 0)

    A

    0 II In

    E 0, 3 - In .- II I-

    * i t

  • ni W z W

    0 m W

    I-

    h h 0 0

    x x 9 - 9 0 .

    h h 0 0

    x x l n h l h 0

    m

    c c

    .. % B 9)

    3

    0 LL

    C 0 U

    L - .-

    .- 9

    .. L 3

    0 LL

    - .- c v) L .- LL

    2 9) B

    u)

    h h 0 0

    x x 0 . 0 .

    0 . m

    c c

    c -

    h h 0 0

    x x 2 8 9

    .. + 9) B E 3

    0 LL

    - .- C 0

    73 .- 9

    1. + 0)

    3

    0 LL

    L

    - .- c v) L .-

    LL

    2 9) B

    0 c3

    n

    56

    Q

    0 -0

    ln

    C 0

    V Q

    .- c

    + c +

  • 57

    Z hl

    9 0 hl

    - L

    9 0

    X co m m

    -

    - L

    n 0

    X 0 hl

    03

    -

    - 0 0

    -

    L hl v)

  • TABLE 32 MEDIAN PERCENT DEGRADATION VALUES AT SCREENING

    DOSES (5% OF MEDIAN FAILURE DOSES)

    d

    I = 5 m A IC = 30 mA Type C

    I

    21 % 2N914 29.5% I

    2N2192 21 % 18.5%

    2 N2369 28.5% 22 % I

    2N918 22 Yo 19.5%

    A l l Combined 26 % 22 Yo

    * I C = 2 m A IC = 10 mA

    S2N930 12.5% 19 Yo I

    58

  • - * - * m m h ( m c c - 9 . . . . m m o m 90-9 h - c v c v om.-*

    * d c v d . . . . . . . .

    c n l c v c

    hhl*hl h h m k c

    h h m o . . . . c v d c v 0 0 . . . . m 9 9 m - c v 0 . 0 . * c v d O . . . .

    9 Q d O * d d 9

    0000 o m - 0 m m m o m m n l m

    0000 o n l m o h ( m Q 0 0 mmcvcv

    0000 O m - Q m 0 . h c o m e - c v

    c v m v m c c z z c v m * m 0000

    59

  • Y V 0 - m E

    9 0

    2 x c v,

    61

  • t

    L e, u, (5)

    t- .- L

    4

    62

  • C al

    V Q, Q m

    E .-

    63

    > > > > 0 0 0. ha . .

    I-

    w Z

    0 m W

  • C Y 0 hl I v)

    ? N

    V Y

    I

    I

    h

    X CY

    + L

    5

    ux

    X CY

    I1

    I- X

    64

  • u)

    0 VI

    us 'W

    LL -c

    0 0 0 0 0 0 9 u) d

    0 0 8

    0 u)-

    X

    d

    c3

    0 0 0 0 h(

    0 0 0 0

    65

  • 99.99

    99.9 99.8

    99

    98

    95

    9a

    80

    -0 70 al

    0 u- C

    - .- 60 c 50

    40

    LL 30

    20

    10

    5

    2

    1

    0.5

    0.2 0.1

    0.05

    0.01

    I Q,

    10" 10' 10" 1

    Gamma Dose (r)

    FIGURE 6 2N914, FAIRCHILD, 27OC, PERCENT FAILED VERSUS GAMMA DOSE

    66

  • 99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70 U a,

    0 LL

    S

    - e- 60

    +. 50 40

    2 30

    20

    10

    5

    2

    1

    0.5

    0.2 0.1

    0.05

    0.01

    L

    11 7

    1 o6 10' Gamma Dose (r)

    1

    FIGURE 7 2N914, MOTOROLA, 27'C, PERCENT FAILED VERSUS GAMMA DOSE

    67

  • 99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70

    K 60

    50 40

    2 2 30

    20

    10

    5

    2

    1

    0.5

    0.2 0.1 9.95

    0.01

    -0 e,

    0

    +

    1 0' 10" 10'

    Gamma Dose (r)

    1 c)

    FIGURE 8 2N914, TEXAS INSTRUMENTS, 27OC, PERCENT FAILED VERSUS GAMMA DOSE

    6a

  • 99.99

    99.9 99.8

    -0 a,

    U L L

    C a, V

    a, a

    - .- +

    L

    99

    98

    95

    90

    80

    70

    40 50 40

    30

    20

    10

    5

    2

    1

    0.5

    0.2 0.1

    0.05

    0.01

    I

    I

    4-

    I

    1

    f 1 o6 1 0’ 1 o8 10’

    Gamma Dose (r)

    FIGURE 9 2N914, GENERAL ELECTRIC, 27OC, PERCENT FAILED VERSUS GAMMA DOSE

    69

  • 99.99

    99.9 99 .s

    99

    9s

    95

    90

    80

    70

    ._ 40 50 40

    2 d 30

    20

    10

    5

    2

    1

    0.5

    0.2 0.1 0.05

    0.01

    -0 a,

    0 -

    4-

    1 o6

    FIGURE 18 2N914, 27OC, PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS

    70

    10’

    Gamma Dose (r)

    0 13’

  • ~ _ _ ~ ~ ~

    99.99

    99.9 99.8

    99

    98

    95

    YO

    8 1

    70

    .- 69 U al

    D LL

    c V

    - r. t3

    a, 43 ., +

    L

    2 3 :

    23

    13

    5

    2

    1

    0.5

    0.2 3.1

    3.05

    0.01

    1 o6 Gamma Dose (r)

    7

    10' 1 o5

    FIGURE 1 1 2N2192, FAIRCHILD, 27OC, PERCENT FAILED VERSUS GAMMA DOSE

    71

  • s'9.99

    99.9 99.8

    99

    98

    95

    90

    33

    70

    .- 60 50

    U

    0

    +

    s 40 2 2 30

    20

    10

    5

    2

    1

    0.5

    0 .2 0.1

    0.05

    0.01 1 o3 1 o6 3

    Gamma Dose ( r )

    FIGURE 12 2N2192, MOTOROLA, 27OC, PERCENT FAILED VERSUS GPMMP DOSE

    72

  • I 1 ’ I 8 I 8 I E 8 I I 8 I I 8 I 8 I 1

    99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70

    “3 60

    50 6 40

    CL 30

    20

    10

    5

    2

    1

    0.5

    0.2 0.1

    0.05

    0.01

    U

    .- 0

    +

    u

    1 o6 1 o8 1 Gamma Dose ( r )

    FIGURE 13 2N2192, TEXAS INSTRUMENTS, 27OC, PERCENT FAILED VERSUS GAMMA DOSE

    73

  • 99.99

    99 .Y 99 .

  • 8 I ’ I 1 I 8 1 I 1 I I( I 1 B 8 I t 8 1

    99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70

    60 50 40

    30

    20

    13

    5

    2

    1

    0.5

    0.2 0.1

    0.05

    0.01 1 o6

    Gamma Dose (r)

    10’ 1 o8

    FIGURE 15 2N2192, 27’C, PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS

    75

  • 1 o6 1 o8 1 Gamma Dose ( r )

    FIGURE 16 2N2369, FAIRCHILD, 27'C, PERCENT FAILED VERSUS GAMMA DOSE

    76

  • 8 8 ’ 8 8 8 8 I 1 I 8 8 8 8 8 B I 8 8 1

    99.99

    99.9 99.8

    99

    98

    95

    90

    83

    70

    69 50 40

    30

    20

    10

    5

    2

    1 0.5

    0.2 0.1

    0.05

    0.01

    1 o5 1 o6 1 o8 Gamma Dose (r)

    FIGURE 17 2N2369, MOTOROLA, 27’C, PERCENT FAILED VERSUS GAMMA DOSE

    77

  • 39.9 33.8

    99

    5;

    95

    4'c'

    3L

    IC

    - .- 6C 5c

    -0 (u

    0

    +

    2 4c a" 3c Y

    2c

    1c

    F - i 1

    6.5

    G . i G. 1

    c.cc

    c.01 1 G6 10' 1 c8

    Gamma Dose (r)

    FIGURE 18 2N2369, TEXAS INSTRUMENTS, 27OC, PERCENT FAILED VERSUS GAMMA DOSE

    78

  • I I I I 8 I 8 8 8 8 I I 8 8 I 8 8 I 8

    99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70 -0 a,

    0 L L

    t

    - -- 60 + 50

    40 2 2 30

    20

    10

    5

    2 1

    0.5

    0.2 0.1

    0.05

    0.01

    Failure Defined As

    -

    FIGURE 19 2N2369, GENERAL ELECTRIC,

    3 1 O'r

    Gamma Dose

    27'C, PERCENT FAILED VERSUS GAMMA DOSE

    79

  • 99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70 -0

    0)

    0 L

    t

    - .- 60 50

    e, 40 2 ti? 30

    20

    10

    5

    2

    1

    0 .5

    0.2 0.1

    0.05

    0.01

    +

    Gamma Dose

    FIGURE 20 2N2369, 27OC, PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS

    80

    1 1 1 1

    1 1 1 1 1 1 1 1 1 1 1 1 1 I

    a

  • I 1 " I 8 8 8 8 8 I I I 1 8 8 I 8 8 I 8

    99.99

    99.9 99.8

    99

    98

    95

    9c

    8C

    7c

    .- 6C c 5c

    4c 2

    -0 W

    0 LI

    C

    -

    2 3c

    2c

    1c

    C w

    2

    1 0.5

    0.2 0.1

    0.05

    0.01

    Failure Defined As

    7 10 r 6 10 r

    I

    9 10 r 8 10 r

    Gamma Dose

    FIGURE 21 2N918, FAIRCHILD, 27'C, PERCENT FAILED VERSUS GAMMA DOSE

    81

  • 99.9s

    99.4 99.8

    9s

    98

    9'

    9c

    8C

    7c

    6C

    c 5c e, 4c 2 2 3c

    73 0)

    u LL

    c

    2c

    1c

    r; i

    2

    1 0.5

    0.2 0.1

    0.05

    0.01 6 10 r 8 10 r

    7 10 r

    Gamma Dose

    9 10 r

    FIGURE 22 2N918, MOTOROLA, 27'C, PERCENT FAVILED VERSUS GAMMA DOSE

    02

    1

  • 8 8 ' 8 8 I I 8 8 II 8 I I 8 I 8 8 1 8 I

    99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70

    60 50

    40

    30

    20

    10'

    5

    2

    1

    0.5

    0.2 0.1

    0.05

    0.01 7

    10 r 8 10 r 9 10 r

    Gamma Dose

    FIGURE 23 2N918, TEXAS INSTRUMENTS, 27'C, PERCENT FAILED VERSUS GAMMA DOSE

    83

  • 99.99

    99.9 99.8

    95

    9e

    92

    9c

    8C

    7(

    * - 6(

    + 5( 4(

    -73 a,

    0 LL

    c

    -

    L

    2 3(

    2(

    1 (

    0..

    0.: 0.

    0.0.

    0.0

    .

    7 10 r 6 10 r

    9 10 r 8 10 r

    Gamma Dose

    FIGURE 24 2N918, GENERAL ELECTRIC, 27OC, PERCENT FAILED VERSUS GAMMA DOSE

    84

  • I 8 1 1 I I I 8 I 8 I 8 8 I 8 8 8 I 1

    99.99

    99.9 99.8

    99

    98

    95

    90

    80

    70 U W

    0 u- c

    - -- 60 t 50 a, U 40

    2 30

    20

    10

    5

    2

    1

    0.5

    0.2 0.1

    0.05

    0.01

    I

    1 ( r 1 O'r 1 08r

    Gamma Dose

    FIGURE 25 2N918, 2 7 O C , PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS

    85

    r

  • 3 *

    0 3

    0 N

    0 -

    0

    86

    1 - 1

    I I I 1 I 1 I 1

    I 1 I I I I 1 1

    a

  • u

    I I X O

    v) c C (u

    5 L c v)

    C - : X 0) I-

    I a

    87

    LL

    0

    L

    9 0

    X h( (u

    ( u *

    Q

    Q

    0 n Q 2 z Q c3

    LL

    I-

    w v,

    Ln

    v, @i w

    3

    >

    0

    h

    w v,

    n W @i

    3 - Q LL

  • 0 0 0 0 0 c3 hl c 0 0 0 9 u) *

    0 hl

    u) c

    0 c

    u)

    88

    1 . I

    I I I I I 1 I 1 I 1 I I I I I I I

  • ~~

    I I 1 I I I I 8 I I I I I I I I 1 I I 89

    > f

    7 3

    > 3

    2 3

    0 LL

    - .- 0 U

    I z 9 o z - 2 X : 6 E

  • 0 0 c? (u 0 0 u) d

    I

    I I

    ,I I

    I I I

    I

    I

    I

    ' I I n W

    I 0 c

    Z 6 n -

    - 0 * 7

    0 N c

    0 0 - 0 00

    0 9

    0 d

    0 (u

    0 I

  • I I I I I I I 1 I I I I i I I I I I B

    0 0 0 0 0 0 0 9 m d m CJ -

    91

    0 03

    0 9

    0 *

    3 N

    3

  • 0 0 0 0 0 0 c

    0 hl m 9 d

    92

    1 1 I I I I I I I 1 I I I I I I I I I

  • 7-t- I 4.__ +- I

    +-+- 0 0 0 0 0 9 u) d (3 hl

    93

    3

    - t - - - - T ----!----.

    I 1

    -tt --+* TI- - -- t

    0 0 c

    0 0 *

    0 9 c7

    0 hl c7

    0 co hl

    0 s

    8 -0 L

    N c

    X

    u. 0 $ 4 u) W E L O

    a - 0 l l

  • 94

    0 N ?-,

    0 m N

    0

    3

    g 9o L

    N - X

    0 9 7

    0 N -

    0 03

    0 *

    0

  • 0 9 m

    0 cu 0

    0 co cu

    0 d hl

    L

    8 *o h l -

    X

    0 9 c

    0 N 7

    0 03

    3 d

    LLI 3 3

    95

  • Q Q E € 0 m c ) II II u u - -

    0 0

    TT

    I I

    t-t r i I 1

    I I I I +!

    L 3 0 -

    L . 0 c

    DL 0

    -)L 0

    v)

    C

    E .- V e, Q v)

    -0 e, C W e, V

    C

    L

    v)

    3

    6 L 3

    U LL

    - .- + v) L .- LL

    0 L

    L

    d E E 0

    I I I I

    I I