Click here to load reader
Upload
others
View
10
Download
0
Embed Size (px)
Citation preview
ER 8623
COMPONENTS IRRADIATION TEST NO. 19
GAMMA IRRADIATION
OF
2N914, 2N918, S2N930, 2N2192
AND 2N2369 TRANSISTORS
October 1966
Prepared For:
GEORGE C. MARSHALL SPACE FLIGHT CENTER
Prepared By:
LOCKHEED GEORGIA NUCLEAR LABORATORY
L O C K H E E D G E O R G I A N U C L E A R L A B O R A T O R Y Lockheed-Georgia Company - A Division of Lockheed Aircraft Corporation
https://ntrs.nasa.gov/search.jsp?R=19670009216 2020-03-20T16:51:48+00:00Z
I f this document i s supplied under the requirements of a United States Government contract, the following legend shall apply unless the letter U appears in the coding box.
This data i s furnished under a United States Government contract and only those portions hereof which are marked (for example, by circling, underscoring or otherwise) and indicated as being subject to this legend shall not be released outside the Government (except to foreign gov- ernments, subject to these same limitations), nor be disclosed, used, or duplicated, for procurement or man- ufacturing purposes, except as otherwise authorized by contract, without the permission of Lockheed-Georgia Company, A Division of Lockheed Aircraft Corporation, Marietta, Georgia. This legend shall be marked on any reproduction hereon in whole or in part.
The "otherwise marking" and "indicated portions" as used above shall mean this statement and include al l details or manufacture contained herein respectively.
D I I
I I I I I I I I I I I I I I
m
I
1
I I I E 1
FOREWORD
This report i s submitted to the Astrionics Laboratory of the George C.
Marshal I Space Flight Center, National Aeronautics and Space Admin-
istration, Huntsville, Alabama, in accordance with the requirements
of Task Order No. ASTR-LGC-330f Contract NAS 8-5332. The re-
port i s one of a series describing radiation effects on various electronic
components. This particular report concerns gamma irradiation and an-
nealing of types 2N914, 2N918, S2N930, 2N2192 and 2N2369 tran-
sistors.
The test was performed by the Lockheed Georgia Nuclear Laboratory,
Lockheed-Georgia Company.
I TABLE OF C O N T E N T S
Page
FOREWORD
TABLE OF CONTENTS
LIST OF TABLES AND FIGURES
1 .o
2.0
3.0
3.1
3.2
3.3
3.3.1
3.3.2
3.3.3
3.4
3.4.1
3.4.2
3.4.3
4.0
5.0
5.1
5.2
SUMMARY
INTRODUCTION
TEST PROCEDURE
TEST SPECIMENS
TEST MEASUREMENTS
INSTRUMENTATION
h Measurement Circuit FE Measurement Circuit 'EBO
T Measurement Circuit
TEST ENVIRONMENT
Pressure
X
Tempe ra tu re
Gamma
METHOD OF DATA ANALYSIS
TEST DATA AND DISCUSSION OF RESULTS
ANNEALING AS A FUNCTION OF TIME AFTER
SMALL DOSES OF GAMMA RADIATION
RECOVERY OF hFE BY ANNEALING AT HIGH
TEMPERATURE
i
I l l ...
V
1
3
5
7
8
8
8
9
9
9
9
9
10
1 1
13
14
14
... I l l
I
5.3
5.4
5.5
5.6
5.7
5.8
5.9
TABLE OF C O N T E N T S (Continued)
A COMPARISON OF RADIATION TOLERANCES
DURING FIRST AND SECOND IRRADIATIONS
CORRELATIO N BETWEEN PRE- I RRADl ATlO N
PARAMETERS AND RADIATION TOLERANCE
RELATIVE DEGRADATION OF hFE AT LOW GAMMA
DOSES
CONCLUSIONS
RECOMMENDED METHOD FOR PRE-SELECT1 NG NPN
SILICON TRANSISTORS FOR USE IN A GAMMA
RAD1 AT10 N E NVI RO NME NT
RECOMMENDED METHOD FOR ESTIMATING
MINIMUM GAMMA RADIATION LIFETIME OF
NPN SILICON TRANSISTORS
RECOMMENDED FOLLOW-ON INVESTIGATIONS
REFERENCES
Page
15
16
18
19
20
21
23
25
I V
1 1 I 1 3 I I 1 I I I I
I 1 1 1 I
a
I 'I i I r I I
Tables
TABLE 1
TABLE 2
TABLE 3
TABLE 4
TABLE 5
TABLE 6
TABLE 7
TABLE 8
TABLE 9
L I S T OF TABLES A N D F I G U R E S
TEST SPECIMENS AND TEST CONDITIONS
MANUFACTURERS' SPECIFICATIONS FOR
TEST SPECIMENS
SPEARMAN'S RANK CORRELATION COEFFICIENT
FOR ORDER OF "PARAMETER" VERSUS ORDER OF
FA1 LURE
PERCENT CHANGE IN hFE WITH TIME AT ROOM
TEMPERATURE (25 * 3OC) AFTER GAMMA DOSE OF 3 . 1 5 ~ 10 r
PERCENT CHANGE IN hFE WITH TIME AT ROOM
TEMPERATURE (25 * 3OC) AFTER GAMMA DOSE OF 3 . 1 5 ~ 10 r
PERCENT CHANGE IN hFE WITH TIME AT ROOM
TEMPERATURE (25 3OC) AFTER GAMMA DOSE
OF 9 . 6 7 ~ 10 r
5
5
5
2N914, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),
hFE DEGRADATION, ANNEALING AND FAILURE
DATA
2N914, FAIRCHILD, MOTOROLA, TEXAS INSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),
hFE DEGRADATION, ANNEALING AND FAILURE
DATA
2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),
hFE DEGRADATION, ANNEAL1 NG AND FA1 LURE
DATA
Page
27
28
29
30
31
32
33
34
35
V
Tables
TABLE 10
TABLE 1 1
TABLE 12
TABLE 13
TABLE 14
TABLE 15
TABLE 16
L I S T OF TABLES A N D F I G U R E S (Continued)
2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),
hFE DEGRADATION, ANNEAL1 NG AND FA1 LURE
DATA
2N2369, FAIRCHILD, MOTOROLA, TEXAS I NSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),
hFE DEGRADATION, ANNEALING A N D FAILURE
DATA
2N2369, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),
hFE DEGRADATION, ANNEALING AND FAILURE
DATA
2N918, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 5 mA),
hFE DEGRADATION, ANNEALING AND FAILURE
DATA
2N918, FA1 RCHl LD, MOTOROLA, TEXAS I NSTRU-
MENTS AND GENERAL ELECTRIC, (Ic = 30 mA),
hFE DEGRADATION, ANNEALING AND FAILURE
DATA
2N914 AND 2N2192, FAIRCHILD, MOTOROLA,
TEXAS INSTRUMENTS AND GENERAL ELECTRIC,
hFE DEGRADATION, ANNEALING AND FAILURE
DATA
2N2369 AND 2N918, FA1 RCHl LD, MOTOROLA,
TEXAS INSTRUMENTS AND GENERAL ELECTRIC,
hFE DEGRADATION, ANNEALING AND FAILURE
DATA
Page
36
37
38
39
40
41
42
v i
Tables
TABLE 17
TABLE 18
TABLE 19
TABLE 20
TABLE 21
TABLE 22
TABLE 23
TABLE 24
TABLE 25
TABLE 26
TABLE 27
L IST OF TABLES A N D F I G U R E S (Continued)
S2N930, FAIRCHILD (Ic = 2 mA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
S2N930, FAIRCHILD (Ic = 2 mA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
S2N930, TEXAS INSTRUMENTS (Ic = 2 mA), hFE DEGRA-
DATION, ANNEALING AND FAILURE DATA
S2N930, TEXAS INSTRUMENTS (Ic = 10 mA), hFE DEG-
RADATION, ANNEAL1 NG AND FAILURE DATA
2N914, FAIRCHILD, MOTOROLA, TEXAS I NSTRU-
MENTS AND GENERAL ELECTRIC, Te , lEBO AND h DATA 0 0
FEO
h~~
2N2192 FAIRCHILD, MOTOROLA, TEXAS INSTRU-
MENTS AND GENERAL ELECTRIC, Te , lEBO AND
2N3369, FA1 RCHILD, MOTOROLA, TEXAS I NSTRU-
MENTS AND GENERAL ELECTRIC, Te , lEBO AND
2N818, FAIRCHILD, MOTOROLA, TEXAS INSTRU-
MENTS AND GENERAL ELECTRIC, T , IEBo_ AND
0 0 DATA
0 0 DATA h~~
e-
Page
43
44
45
46
47
48
49
50
51
h DATA F E,
U U 52
2NF192, COMPUTED VALUES OF RADIATION DAM-
AGE CONSTANT (K) AND COEFFICIENTS OF
VARIATION (S/E)
v i i
53
Tables
TABLE 28
TABLE 29
TABLE 30
TABLE 31
TABLE 32
TABLE 33
Figures
FIGURE 1
FIGURE 2
GURE 3
GURE 4
GURE 5
L I S T OF TABLES A N D F I G U R E S (Continued)
Page
I I
I 1 1 I II I I I 1 1 1 I 1 1 I
m
v i i i
SPEARMAN'S RANK CORRELATION COEFFICIENT
FOR ORDER OF h T VERSUS ORDER OF
FAILURE AT IC = 5 rnA
G A I N REAL1 ZED BY SELECT1 NG OUT SPECIMENS
WITH PERCENT DEGRADATION GREATER THAN
MEDIAN VALUE AT GAMMA DOSE EQUAL TO 5%
OF MEDIAN FAILURE DOSE
G A I N REALIZED BY SELECT1 NG OUT SPECIMENS
WITH PERCENT DEGRADATION GREATER THAN
MEDIAN VALUE AT GAMMA DOSE EQUAL TO 5%
OF MEDIAN FAILURE DOSE
RATIO OF MINIMUM RADIATION LIFETIME OF
FEo eo
PRE-SELECTED SPECIMENS TO SCREEN1 NG DOSE
MEDIAN PERCENT DEGRADATION VALUES AT
SCREENING DOSES (5% OF MEDIAN FAILURE
DOSES)
2N914, COMPARISON OF ESTIMATED AND ACTUAL
FAILURE DOSES FOR IC = 5 mA
SCHEMATIC DIAGRAM OF GAMMA IRRADIATION
TEST RIG
hFE MEASUREMENT CIRCUIT
MEASUREMENT CIRCUIT 'EBO
TYPICAL CURVES OF 1o5/hFE VERSUS GAMMA
TRANSIT TIME MEASUREMENT CIRCUIT
DOSE FOR TYPE 2N914
54
55
56
57
58
59
61
62
63
64
65
Figures
FIGURE 6
FIGURE 7
GURE 8
GURE 9
FIGURE 10
FIGURE 11
GURE 12
GURE 13
FIGURE 14
FIGURE 15
FIGURE 16
FIGURE 17
FIGURE 18
L I S T OF TABLES A N D F I G U R E S (Con ti nued)
2N914, FA1 RCHILD, 27OC, PERCENT FAILED
VERSUS GAMMA DOSE
2N914, MOTOROLA, 27OC, PERCENT FA1 LED
VERSUS GAMMA DOSE
2N914, TEXAS INSTRUMENTS, 27OC, PERCENT
FAILED VERSUS GAMMA DOSE
2N914, GENERAL ELECTRIC, 27OC, PERCENT
FAILED VERSUS GAMMA DOSE
2N914, 27OC, PERCENT FAILED VERSUS GAMMA
DOSE, COMPARISON OF MANUFACTURERS
2N2192, FAIRCHILD, 27OC, PERCENT FAILED
VERSUS GAMMA DOSE
2N2192, MOTOROLA, 27OC, PERCENT FAILED
VERSUS GAMMA DOSE
2N2192, TEXAS INSTRUMENTS, 27OC, PERCENT
FAILED VERSUS GAMMA DOSE
2N2192, GENERAL ELECTRIC, 27OC, PERCENT
FAILED VERSUS GAMMA DOSE
2N2192, 27OC, PERCENT FAILED VERSUS GAMMA
DOSE, COMPARISON OF MANUFACTURERS
2N2369, FAIRCHILD, 27OC, PERCENT FAILED
VERSUS GAMMA DOSE
2N2369, MOTOROLA, 27OC, PERCENT FAILED
VERSUS GAMMA DOSE
2N2369, TEXAS INSTRUMENTS, 27OC, PERCENT
FAILED VERSUS GAMMA DOSE
I X
Page
66
67
68
69
70
71
72
73
74
75
76
77
78
Figures
FIGURE 9
FIGURE 20
FIGURE 21
FIGURE 22
FIGURE 23
FIGURE 24
FIGURE 25
FIGURE 26
FIGURE 27
FIGURE 28
FIGURE 29
L IST OF TABLES A N D F I G U R E S (Con t i nu ed)
2N2369, GENERAL ELECTRIC, 270Cf PERCENT
FAILED VERSUS GAMMA DOSE
2N2369, 270Cf PERCENT FAILED VERSUS GAMMA
DOSE, COMPARISON OF MANUFACTURERS
2N918, FAIRCHILD, 27OC, PERCENT FAILED
VERSUS GAMMA DOSE
2N918, MOTOROLAf 27OC, PERCENT FA1 LED
VERSUS GAMMA DOSE
2N918, TEXAS INSTRUMENTS, 27OC, PERCENT
FAILED VERSUS GAMMA DOSE
2N918, GENERAL ELECTRIC, 27OC, PERCENT
FAILED VERSUS GAMMA DOSE
2N918, 270Cf PERCENT FA1 LED VERSUS GAMMA
DOSE, COMPARISON OF MANUFACTURERS
2N914, 27OC, PERCENT DEGRADATION OF hFE
AT 1.80 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 rnA
2N914, 270Cf PERCENT DEGRADATION OF hFE
AT 3.22 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 rnA
2N2192, 270Cf PERCENT DEGRADATION OF hFE
AT 3.05 x lo5 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR I C = 5 mA 2N2192, 270Cf PERCENT DEGRADATION OF hFE
AT 4.52 x lo5 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA
X
Page
79
80
81
82
83
84
85
86
87
88
89
I
Figures
FIGURE 30
FIGURE 31
FIGURE 32
FIGURE 33
FIGURE 34
FIGURE 35
FIGURE 36
L I S T OF TABLES A N D F I G U R E S (Con t i nu ed)
Page
2N2369, 27OC, PERCENT DEGRADATION OF hFE
AT 2. i o io6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 mA
2N2369, 27OC, PERCENT DEGRADATION OF hFE
AT 3.98 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA
2N918, 27OC, PERCENT DEGRADATION OF hFE
AT 4.48 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 mA
2N918, 27OC, PERCENT DEGRADATION OF hFE AT
4.45 x io6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA
NPN SILICON PLANAR EPITAXIAL TRANSISTORS
(16 TYPE 2N914, 16TYPE 2N2192, 16 TYPE 2N918,
16 TYPE 2N2369), 27OC, PERCENT DEGRADATION OF
hFE AT 1.88 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 5 mA 94
NPN SILICON PLANAR EPITAXIAL TRANSISTORS
( 1 6 TYPE 2N914, 16 TYPE 2N2192, 12 TYPE 2N918,
16 TYPE 2N2369), 27OC, PERCENT DEGRADATION OF
hFE AT 3.05 x lo6 r (5% OF MEDIAN FAILURE DOSE) VERSUS GAMMA DOSE AT FAILURE FOR IC = 30 mA 95
FIRST FAILURE POINT FOR UNSCREENED SPECIMENS
VERSUS FIRST FA1 LURE POINT FOR PRE-SELECTED
SPECIMENS 96
xi
90
91
92
93
8 1 . O S U M M A R Y
NPN silicon planar epitaxial transistors of the types 2N914, 2N2192, 2N2369 and
2N918 were subjected to gamma irradiation and failure points (gamma dose when
50% degradation of h
perature annealed and gamma irradiated a second time to failure along with type
S2N930 specimens from a previous test.
occurred) were determined. The specimens were then tem- FE
investigations were made on possible correlations between pre-irradiation values
h ~ ~ ’ e EBO covery of radiation induced degradation o f h
of gamma irradiation was investigated, and comparison was made between failure
points shown during the first irradiation and those shown during the second irradia-
tion. The ultimate objective of the test was the development of a method to pre-
select transistors for use in a radiation environment.
T and I and radiation tolerance exhibited during the first irradiation. Re-
at mom temperature after small doses FE
Test results indicated that NPN silicon transistors can be pre-selected for use i n a
gamma radiation environment by subjecting the devices to comparatively smal I doses
o f gamma radiation and computing percentage losses of h
vices w i l l have a minimum radiation lifetime about one order of magnitude, or more,
greater than the screening dose.
The pre-selected de- FE *
1
The experiment described in this report i s the nineteenth irradiation of electronic
components and i s the twenty-fourth in a series of radiation effects tests on electron-
i c equipment, circuits and components contemplated for use on a nuclear space ve-
h ic le . Since the use of equipment on this vehic le i s contingent upon its abi l i ty to
withstand the nuclear environment, the Astrionics Laboratory o f the Marshall Space
Flight Center has undertaken to assure that Government furnished or specified equip-
ment w i l l survive this environment. The equipment i s to be subjected to the expected
nuclear environment as simulated at the Lockheed Georgia Nuclear Laboratory. Mea-
surements made on the specimens during the irradiation w i l l describe their radiation
tolerance.
8 I I 1 I
I
1 I
The subjects of this test are the 2N914, 2N918, S2N930, 2N2192 and 2N2369 tran-
sistors. This experiment i s an extension o f the work performed during Components
Irradiation Test No. 18. The purpose of the extended work was to verify correlations
established by the Test No. 18 data, i .e ., the correlation between percentage loss of hFE at comparatively low gamma doses and gamma dose at failure, and the corre-
lation between h x T
o f these correlations would enable pre-selection of transistors for use in radiation en-
vironments. During the experiment, recovery of lost h (annealing) as a function
o f time after small doses of irradiation was also investigated, and comparison of ra-
diation tolerances during in i t ia l irradiation was made with radiation tolerances ex-
hibited during a second irradiation conducted after the specimens had been annealed
at high temperatures.
and dose at failure. Verification of one or both 'IEBO, FEO eo
FE
3
I 3 . 0 T E S T PROCEDURE
The test specimens, except the S2N930 type, were procured "off-the-shelf" from an
electronics supply vendor. The S2N930 specimens had been supplied by the Astrionics
Laboratory of the Marshall Space Flight Center for Components Irradiation Test No. 18.
from each of four manufacturers: Fairchild, Motorola, Texas Instruments, and Gen-
eral Electric. The S2N930 specimens consisted of forty-five specimens from Fairchild
and twenty specimens from Texas instruments. These S2N930 specimens, having been
irradiated during the Test No. 18 experiment, were not used for the first irradiation
of this test.
Seven specimens of each of the 2N914, 2N918, 2N2192 and 2N2369 types were
and T were made on a l l specimens of the FE' 'EBO e Pre-irradiation measurements of h
2N914, 2N918, 2N2192 and 2N2369 type - manufacturer groups. After these mea- surements were completed one specimen of each type - manufacturer group was selected as a control specimen. These control specimens, sixteen in all, were used as checks
on the stability and repeatability of the measurement instrumentation. The remaining
devices were irradiated i n a fixture as shown in Figure 1. Cobalt-60 was used as the
source of the gamma radiation.
4 The irradiation was conducted at a constant gamma dose rate of 1.10 x 10 r/hr unti l
a dose of 5.37 x 10 r had been reached. From that point on, a dose rate of 2.10 x
10 r/hr was used for the remainder of the experiment.
5
5
FE The specimens were removed from the radiation environment for measurement of h
at the following dose accumulation points:
4
4
5
1 . 0 6 ~ 10 r
5 . 3 0 ~ 10 r
1 . 0 2 ~ 10 r
4.40 x
0.27 x
2.70 x
5
5
5
3 . 1 5 ~ 10 r
5 . 3 7 ~ 10 r
9 . 6 7 ~ 1 0 r
o6 r 6
O r 7 0 r (end of first irradiation).
5 At the 5.37 x 10 r dose point the hFE of two specimens of each type - manufacturer group was measured after elapsed periods of three hours, seven hours and sixty-five
hours to detect any annealing at room temperature.
5 At the 9.67 x 10 r dose point two specimens of each type - manufacturer group were removed from the test rig and were not given any additional dose during the first i r -
radiation. This was done to check for annealing at room temperature over a period
of several days, and to investigate the effect of total dose on the degree of anneal-
ing that might be accomplished by subjecting the specimens to high temperatures.
Some of the specimens had not failed (h
terminated at 2.70 x 10 r. The failure points for these specimens were determined
/h FE FEo 5: 0.5) when the first irradiation was 7
by extrapolation of the Q-factor (l/h ) curves. I t i s be l ieved that no errors were FE introduced by this procedure as the Q-factor curves were linear with gamma dose i n
the region of extrapolation.
After post-irradiation measurements of h were made al l specimens except the con-
trol specimens were annealed in an oven. The specimens were left in the oven for a
period of about three hours at the temperatures indicated:
FE
T Y Pe 2N914
2N918
Annealing Temperature
295 i s"C
295 f s"C
TY Pe S2N930
2N2192
2N2369
Ann e a I i ng Tempe rat ure
295 i s"C
295 i s"C
195+ !fC
The annealing temperatures were selected on the basis of the manufacturers' recom-
mended maximum storage temperatures.
After annealing pre-irradiation measurements of h
A l l specimens except the control specimens were then placed i n the test rig and ir- 5 7
radiated at 2.1 x 10 r/hr to a total dose of 5.20 x 10
were made on al I specimens. F E
r.
Measurementsof h
paints:
were made on all devices at the following accumulated dose FE
5
5 1 . 0 5 ~ 10 r
5 . 2 5 ~ 10 r 6 1 . 0 0 ~ 10 r 6
6
7
7
4.38 x 10 r
9 . 3 4 ~ 10 r
2 . 3 0 ~ 10 r
5.20 x 10 r (end o f second irradiation).
The purpose of this second irradiation was to compare failure points experienced prior
to annealing with those experienced after annealing.
3.1 TEST SPECIMENS
The test specimens are l isted i n Table 1 . Manufacturers' specifications are shown in
Table 2.
7
The specimens were irradiated passively and were held for the irradiation by inser-
tion of the device leads into the Styrofoam block shown i n Figure 1 . This arrange-
ment insured that al I specimens were irradiated at the same rate.
3.2 TEST MEASUREMENTS
A complete set of measurements was taken prior to the first irradiation on the 2N914,
FE 2N918, 2N2192 and 2N2369 specimens. These consisted of h for I = 5 mA; h
for I C X E 2N2192 and 2N2369 types; T for IE valuesof 0.33, 0.25, 0.20, 0.167 and 0.125
mA for the 2N918 type; I = 4.0 V for the 2N914, 2N2192 and 2N2369
types; and I
FE C = 30 mA; T for I values of 1 .O, 0.5, 0.33, 0.2 and 0.125 mA for the 2N914,
X
for V E 6 0 EB
for VEB = 2.7 V for the 2N918 type. EBO
Measurementsof h = 30 mA were made at frequent intervals
during the first irradiation. The specimens were removed from the irradiation envi-
ronment to a screened room for these measurements.
for I = 5 mA and I FE C C
Prior to the second irradiation, but after annealing, measurements of h
ues of 5 mA and 30 mA were made on the 2N914, 2N918, 2N2192 and 2N2369 types.
for I val- FE C
At the same time measurements of h
the S2N930 specimens. These same h
for I FE C
FE
values of 2 mA and 10 mA were made on
measurements were taken at frequent intervals
during the second irradiation and after irradiation ceased. The procedure for making
these measurements was the same as that used during the first irradiation.
3.3 INSTRUMENTATION
3.3.1 h Measurement Circuit FE
A pulse measurement system, shown schematically in Figure 2, was used to measure
This circuit l im i ted measurement currents to a time interval of about one millisecond I B *
8
thus minimizing any annealing of radiation damage by the measurement currents.
h~~ The control specimens were used to check on the repeatability of the circuit.
values were calculated from the I measurements. B
3.3.2 lEBO Measurement Circuit
Pre-irradiation measurements of I
shown i n Figure 3.
were made using the circuit scllematica EBO ' Y
3.3.3 T Measurement Circuit X
The circuit shown i n Figure 4 was used to measure transit times for five different val-
ues of I
justing C and R unti l a null was obtained. The use of these values to obtain T i s
Each measurement consisted o f adjusting I to the desired value, then ad- E' E
X X X
explained in Section 4.0.
3.4 TEST ENVl RO NMENT
3.4.1 Pressure
During the test a l l specimens were at atmospheric pressure.
3.4.2 Temperature
0 Al l measurements and the irradiations were conducted at 27 f 2 C.
Annealing Temperatures
Between irradiations the types 2N914, S2N930, 2N2192 and 2N2369 were annealed
for three hours at 295 i 5°C. The 2N918 type was annealed for three hours at 195 * !fC.
9
8 4 I
3.4.3 Gamma
The first irradiation was conducted at a dose rate of 1.10 x 10 r/hr to a dose point 5 5
of 5.37 x 10 r. From that point a dose rate of 2.10 x 10 r/hr was used until 2.70 x 7 5
10 r had been accumulated. The second irradiation was conducted at 2.10 x 10 r/hr
to a total accumulated dose of 5.2 x 10 r. 7
10
4 . 0 M E T H O D OF D A T A A N A L Y S I S
A l l data were recorded manually.
The hFE data were calculated manually from measurements of I
I C ' FE these plots the value of h
FE ure dose for each specimen. (Examples o f typical l/h
in Figure 5.) The data obtained from these plots were used to construct the graphs
discussed i n Section 5.0.
taken at constant B
These data were used to construct plots of l / h versus accumulated dose. From
at any desired dose could be obtained, as could the fail-
versus dose plots are shown FE
Transit time values were calculated from measurements on the bridge shown i n Figure
4. The formula used for calculation of transit t ime was: 1
T = (C R r/(r + R 1). X x x X
where T i s i n nanoseconds, X
C i s i n picofarads,
R i s i n ohms, and
r i s i n ohms.
X
X
The T values were then paired with their corresponding 1/1 values (1 in milliamps)
giving five pairs of values for each specimen. A straight line was then fitted to the
f i v e pairs of values by the least squares method, and the value of T
was calculated from the formula for this line. The value of T for 1/1 = 0 i s T or
base transit t ime.
X E E
for 1/1 = 0
X E e'
X E
Al l of the transit time computations were performed on an IBM 7094 computer.
values were measured directly on a picoammeter and required no calculation. I EBO
11
In order to provide a basis for comparing the correlations obtained between order of
failure and various parameters the parameters were ranked and paired with the order 2 2
= 1 - 6 r D /N(N - l ) , 2 o f failure of the specimens. Spearman's formula , r rank was then used to calculate a rank correlation coefficient. Results are shown in Ta-
ble 3.
Copies of the reduced data are on f i l e in the Lockheed Georgia Nuclear Laboratory,
Lockheed-Georgia Company, Dawsonvil le, Georgia.
12
I 5 . 0 T E S T D A T A A N D D I S C U S S I O N OF R E S U L T S
The test data are presented herein in graphical and tabular form.
The primary objectives of this test were to verify correlations established by the Com-
ponents Irradiation Test No. 18 data, i.e., the correlation between percentage loss
of h at comparatively low gamma doses and gamma dose at failure, and the corre-
lation between h
both of these correlations would enable the development o f a method to pre-se
FE and dose at failure. The establishment o f one or T e i I E B O o FEO
transistors for use i n a gamma radiation environment.
To assist i n the development o f such a method two other objectives were establ
for this test:
ec t
shed
(a) An investigation o f the recovery of lost h
time after small doses of irradiation, and
A comparison of radiation tolerances during in i t ia l irradiation with tol-
erances during a second irradiation conducted after the specimens had
been annealed at high temperatures.
(annealing) as CI function of FE
(b)
These were undertaken because, i f lost h could be regained by annealing without FE adversely affecting subsequent radiation tolerance, that part of a transistor's radia-
tion lifetime lost by subjecting the devices to reasonable doses of gamma radiation
for pre-selection could be recovered.
For purposes of this test failure of a specimen was defined as a 50% decrease in the
value o f h Thus, for the first irradiation failure occurred when h j h first
equaled 0.5, where h
ation; for the second irradiation failure occurred when h /h
FE ' FE FE,, i s defined as the h of a device prior to the first irradi- FE
first equaled 0.5, FEO 1
FE FE-9 i s defined as the h of a device after the first irravdation and after FE where h
FEo2
13
annealing at high temperature but prior to the second irradiation. Since h
not equal to h
ations did not occur at the same absolute values of h
first irradiation are shown in Figures 6 through 25.
was
except in rare cases, failures during the first and second rrradi- FE.2
FEo 1 The failure patterns for the FE'
5.1 ANNEALING AS A FUNCTION OF TIME AFTER SMALL DOSES OF GAMMA RAD1 AT1 0 N
During the first irradiation the radiation was suspended for a period of about seventy-
two hours during which the h of each o f two specimens o f each type - manufacturer group was monitored. Results are shown in Tables 4 and 5. Also during the first i r -
radiation two specimens o f each type -manufacturer group were removed from the
radiation environment after a dose of 9.67 x 10 r had been accumulated. The hFE
o f each of these specimens was measured immediately after removal and again eight
days later.
FE
5
Results are shown in Table 6.
The data i n Tables 4 and 5 show that i n the first three hours after cessation o f irradi-
ation the type 2N914 generally recovers a small percentage of its lost h the type
2N2192 shows essentially no change, and the types 2N2369 and 2N918 generally
continue to lose small percentages of h FE' the 2N914 generally showed smal I percentage gains. It appears the magni tude of the
changes i s sufficiently small and the change pattern i s such that the time interval be-
tween cessation of radiation and measurement of h
are maintained at room temperature.
F E'
After eight days (Table 6) a l l types except
i s not crit ical i f the specimens FE
5.2 RECOVERY OF hFE BY ANNEALING AT HIGH TEhrlPERATURE
Tables 7 through 21
FE tion (h ) to h
the first irradiation FEp 1
contain columns showing the ratio o f h after the first irradia- FE prior to the first irradiation (h
and after annealing at high temperature (h
), and the ratio o f h after
) to hFE prior to FEO I FE
FE- 3 U L
14
the first irradiation (h
through 14 received a total dose of 2.70 x 10
listed i n Tables 15 and 16 received a dose of 9.67 x 10 r, and those l isted i n Tables 6 7 17 through 22 received varying amounts ranging from 3.26 x 10 r to 1.24 x 10 r
(See Tables 6 and 7 i n Components Irradiation Test No. 18 Report).
) for each specimen. The specimens listed i n Tables 7
r during the first irradiation, those FEo 1 7
5
A comparison of corresponding values of h /h and h /h reveals that
i n every case but one (Table 21, specimen T 439) annealing at high temperature causes
an increase in the h
creased to a value greater than the original h The amount of the increase appears
to be more a characteristic o f the type - manufacturer category than a function of the amount of degradation or a function o f the total dose received.
FEY1 FEol FE02 FEO 1
of gamma irradiated transistors. In some cases the h i s in- FE FE
FE'
5.3 A COMPARISON OF RADIATION TOLERANCES DURING FIRST AND SECOND I RRADl AT1 0 N S
Tables 7 through 14 and 17 through 22 contain a column showing the ratio of the dose
at failure during the second irradiation to the dose at failure during the first irradia-
tion for each specimen. The failure points are defined as:
First failure point i s the gamma dose:
/2 ' - 01
h~~ - h~~ when
Second failure point i s the gamma dose:
when hFE = h /2. FEo2
The ratios of second fai l points to first fa i l points are i n most cases greater than one.
However, there are many devices of the types 2N914, 2N2192, and 2N918 whose
15
ratios are less than one. Thus a technique of irradiating to find the failure point and
then annealing would not guarantee that the failure point during a second irradiation
would be as great as, or greater than, the first.
Rank
1
c n
Such a technique may be applicable to certain types o f transistors. The data for the
2N2369 and the S2N930 types suggest that one might be developed.
Relative Value* Of
0 0 0 h~~ 0 0 Te "EBO 0 h~~ Te e 'EBO T
0 0 hFE
! I
1 Large st I Largest ; Smallest Largest Largest
c Smallest
1 1 I ' 1 ' 4 Smallest 1 Largest Smallest + Smallest
5.4 CORRELATION BETWEEN PRE-I RRADl ATlO N PARAMETERS AND RAD1 ATlON TOLERANCE
and combinations (h T FE e I h~~ Te 'IEBO ) 0 0 0 0 0
FE' Te' 'EBO Pre-irradiation values of h
thereof were ranked as follows:
A Spearman's rank correlation coefficient was computed for each category as men-
tioned above in Section 4.0. These coefficients are shown i n Table 3. Only in the
h category for the type 2N914 were the coefficients .60 greater for both val-
ues of I This leads to the conclusion that these electrical parameters cannot be
used to pre-select transistors for use i n a gamma radiation environment when h
measured at constant I
T FEO eo
C' i s
FE Reference 3 suggests the following relationship:
C'
16
8
where K i s a radiation damage constant for silicon transistors which varies
with emitter current density, and
A 5, i s radiation exposure.
This expression has been shown to accurately depict the degradation of transistors i n
the neutron radiation environment; the similarity of degradation characteris i c s be-
yond the in i t ia l non-linearity (surface effects, reference 3) leads to the con ecture
that the relationship may also be valid for gamma radiation.
Using hFE/'hFE = 0.5 as a definition of failure, the above equation becomes: 0
b dr (Failure Dose) = '/hFE Te K 0 0
Hence, the first failure should correlate with the largest h
wi th the next largest h T etc.
T , the second failure FE e 0 0
FE e 0 0
I n the data obtained during this test only the type 2N914 showed any consistent cor-
relation (Table 3) between h T and order o f failure. An attempt to explain why
this may be so was made by computing the values of K for the 2N2192 specimens for FEo e
0.
= 5 mA. The computed values and the coefficients of variation for each type - manufacturer group are shown in Table 27. This table shows that while K i s fairly
consistent within the type - manufacturer groups i t varies widely within the type. Noting this fact, Spearman's rank coefficients o f correlation were computed for the
type - manufacturer groups and for the type - groups. Results are shown i n Table 28. Note that correlation i s excellent for some groups and wholly lacking or even nega-
tive for others. These data point to the conclusion that K varies so widely among
specimens in some type -manufacturer groups that the h
IC
T parameter cannot be FE e 0 0
17
used as a tool for preselecting transistors for use i n a gamma radiation environment.
5.5 RELATIVE DEGRADATION OF hFE AT LOW GAMMA DOSES
Figures 26 through 33 show percent degradation of h
Dose for each of the types 2N914, 2N2192, 2N2369 and 2N918 at each of two val-
ues of I Based on the results o f Components Irradiation Test No.
18, 5% of the median failure dose has been chosen as the best compromised screen-
ing dose between higher doses which give better correlation and lower doses which
give smaller decreases i n the radiation lifetimes of the devices.
at 5% of the Median Failure FE
5 mA and 30 mA. C'
Note in Figure 26 that by selecting out those specimens with percent degradations
o f h
fai l are eliminated.
dose at failure i s not perfect, one "good" specimen i s also eliminated by the select-
ing out process. However, the important point i s that a l l the "worst" specimens are
eliminated.
larger than the median percent degradation value the first seven specimens to FE Since the correlation between percent degradation of h and FE
Application of the "selecting out'' process to the data i n Figures 27 through
similar results,
Figures 34 a d 35 show percent degradation o f h at 5% of median failure
a l l types (2N914, 2N2192, 2N2369 and 2N918) combined for IC values of FE
33 yields
doses for
5 mA and
30 mA respectively. Application o f the "selecting out'' process to the data i n these
figures eliminates the first twenty-one failures i n Figure 34 (I = 5 mA) and the first
nineteen failures i n Figure 35 (I C
= 30 mA). C
Tables 29 and 30 summarize the gains i n increased median failure dose and dose at
first failure for the pre-selected specimens as compared to median failure dose and
dose at first failure for the whole group. These tables show that the gains are quite
18
substantial.
Table 31 shows that a conservative estimate o f the minimum radiation lifetime (gam-
ma dose at first failure) of the pre-selected specimens would be about eleven times
the screening dose.
Median percent degradation values at screening doses are tabulated for each o f the
types i n Table 32. These values may be used as a guide in determining when to end
the irradiation when screening a group of devices whose median failure dose i s not
known.
5.6 CONCLUSIONS
( 1 ) The correlations between the pre-irradiation electrical parameters h , F E,
U , or combinations thereof, and radiation tolerance are not EBO, T and I sufficiently good for use i n the pre-selection o f transistors for use in a
gamma radiation environment,
(2) The recovery of h at room temperature (annealing) after small doses of
gamma radiation between cessation of irradiation and measurement i s not
a problem i n the proposed pre-selection method provided the measure-
ments are made within two or three days.
FE
(3) Annealing o f transistors at high temperatures 200 - 3OO0C does change (increase or decrease) the i r radiation to I e rance characteristics .
(4) The pre-selection o f silicon transistors for use i n a gamma radiation en-
vironment can be accomplished by subjecting them to comparatively small
doses o f gamma radiation. The proposed method increases the minimum
radiation lifetime o f the pre-selected devices as compared to the minimum
19
8 radiation I ifetime of unscreened devices by factors ranging from about
1 .5 to 50, depending on the failure distribution o f the unscreened de-
vices. See Figure 36.
5.7 RECOMMENDED METHOD FOR PRE-SELECTING NPN SILICON TRANSISTORS FOR USE IN A GAMMA RADIATION ENVIRONMENT
Gamma irradiate to failure a sample of the type transistor under consid-
eration sufficiently large to represent the population. (A sample of thir-
ty or more i s recommended.) Measure h
value of I
that the l/hFE curve can be constructed for each dev ice from these mea-
surements.
at the proposed application FE prior to irradiation and at intervals during the irradiation x,
C
From the data obtained from the l/h
transistor type l i k e those shown in Figures 26 through 33. This curve
w i l l be the basis for determining the cut-off point for percent degrada-
tion of h
curves construct a curve for the FE
in the pre-selection technique. FE
The pre-selection technique i s as follows.
Measure h
at I C variation with I
of a l l devices to be screened. Measurement should be made FE
value equal to planned application value since radiation tolerance
i s not the same for a l l specimens. C
Gamma irradiate a l l devices to 5% of median failure dose as determined
from sample irradiated above. Measure h and compute percent degra-
dation of h FE
for each device. FE
Refer to curve constructed i n paragraph (2) above to determine percent
20
degradation of h corresponding to desired minimum failure dose. E l i -
minate al I devices having percentage degradations of h
this value. The remaining specimens w i l l be the pre-selected specimens.
FE larger than FE
5.8 RECOMMENDED METHOD FOR ESTIMATING MINIMUM GAMMA RADIATION LIFETIME OF NPN SILICON TRANSISTORS
Because of the characteristic shape of the l / h
(See Figure 5) i t i s possible to estimate the minimum dose necessary to give any value
of l/hFE once the early portion of the curve is known. Of course, the greater the
known portion of the curve the more accurate the estimate w i l l be.
curve in a radiation environment FE
The estimate can be made by computing the slope o f the right hand end of the known
portion of the curve and extrapolating this slope to the selected value of l/h FE'
The technique can be best illustrated by example. Let us suppose that we need some
2N914 transistors to be used i n an application requiring a minimum radiation lifetime
of 8.0 x 10 r. We have on hand several 2N914 devices. To estimate their radia-
tion lifetimes we proceed as follows:
7
Gamma irradiate a l l devices to 5% of the required minimum dose (5% x
8 x 10 r = 4 x 10 r). 7 6
Measure h
point as h . of each device and compute l /h FE FE' Designate hFE at this
FE1
Continue gamma irradiation of a l l devices to a total o f 10% of the re-
quired minimum dose (10% x 8 x 10 7 6
r = 8 x 10 r).
Measure h of each device and compute l /h Designate h at this FE FE' FE
21
point as h . FE2
(All hFE measurements should be made at planned application I values.) C
(e) Select failure criterion. Let us suppose that i t has been selected for us
as 50% degradation o f h on the Q-fac-
tor curve.
which corresponds to 2/h E2 E2
The slope of the right hand end of the known portion of the l /h
i s now found as follows.
curve FE
6 (f) Slope = (l/hFE - I/h )/(8 x l o6 - 4 x 10 ).
2 FEl
Next we use this slope to extend the curve to the failure point and esti-
mate the failure dose
6 ) (8 x l o6 - 4 x 10 ) (*IhFE 2 - 1’hFE2
1 /h - l /h (9) Estimated failure dose = FE2 FE1
- - l/h - l /h
E2 FEl
Once we have found the estimated failure dose for each device we can
select those which meet our requirements, and relegate those remaining
to applications less rigorous.
Table 33 shows estimated failure doses for the 2N914 devices irradiated
during this test along with corresponding actual failure doses found from
the experimental data. The ratios of estimated to actual failure doses
22
~
confirm that the estimated dose w i l l always be a conservative estimate.
5.9 RECOMMENDED FOLLOW-ON INVESTIGATIONS
I t i s recommended that experiments be performed to determine whether or not the
method i s applicable to failures caused by neutron irradiation, and that the data al-
ready obtained, as well as any future data obtained, be analyzed to determine the
pre-selection method's applicability to other definitions of failure - such as h degradation to an absolute value.
F E
23
I t 8 1 I 8
REFERENCES
1. Ashar, K. G., "Transit Time in High-speed Switching Transistors, I' Solid State
Design, Vol. 5, No. 2, February 1964.
2. Croxton and Cowden, Applied General Statistics, Second Edition, Prentice-
Hall, Inc., Page 478.
3 . Frank, Max, "Exploratory Development of the Q-Factor Technique, I' The Ben-
d i x Co rpo rat ion, Re se a rc h La bo ra to r i e s Division , So ut h f i e I d, Mi c h i ga n, Tech - nica l Report AFWL-TR-65-166.
25
I I I I 1 I I I I I I I 1 8 I I I 1 1
24
TABLE 1 TEST SPECIMENS AND TEST CONDITIONS
- -~ VCE = 5V, IC = 30 rnA
VEB = 4 v
I = 1, .5, .33, .2 & .125mA E
Description
T ransi s to r
2N914
NPN, S i
Planar Epitaxial
*
Transistor
2N918
NPN, S i
Planar Epitaxial
*
Transistor
2N2192
NPN, S i
Planar Epitaxial
*
Transistor
2N2369
NPN, S i
Planar Epitaxial
*
T ran s i s to r
S2N930
NPN, Si, Planar
Fairchild (40)
Tex. lnst. (17)
No.1 Test Conditions
24
VCE = 5V, I C = 5 rnA VCE = 5V, IC = 30 rnA
IE= 1, .5, .33, .2 & .125rnA
I Signal at 600 kc
24 1 IE= .33, .25, .2G, .I67 & .125 rnA
Signal at 600 kc
VCE = 5V, I C = 5 rnA
I Signal at 600 kc VCE = 5V, IC = 5 rnA
VCE = 5V, IC = 30 rnA
24
I = 1, .5, .33, .2 & . I25 rnA E I Signal at 600 kc = 5V, 1 = 2 rnA
=5V, I = 10rnA 'CE C
'CE C
Signal at 600 kc
~~
Pararnete r
hFE
h~~
'EBO
X T
h~~
h~~ I
'EBO
X T
h~ E
h~~
'EBO
X T
h~~
h~ E
I EBO
X T
h~~
FE h
I
'EBO
X T
*Six from each of Fairchild, Motorola, Texas Instruments and General Electric
27
TABLE 2 MANUFACTURERS' SPECIFICATIONS FOR TEST SPECIMENS
Description I 2N914
NPN, Silicon
Double-Diffused
Planar Epitaxial
2N918
NPN, Silicon
Planar Epitaxial
2 N930
NPN, Sil icon
Planar
2N2192
NPN, Silicon
Planar Epitaxial
2N2369
NPN, Silicon
Planar Epitaxial
Conditions
VCE = lV, IC = 10 mA
IC =o, v E B = 4 v
fob
VCE = lV, I C = 3 mA
fob
VCE = 5V, IC = 10 mA
IC =o, v E B = 5 v f Gain-Bandw idth Product T
-
VCE = lOV, IC= 10 mA
IC VEB = 3v
f Gai n-Bandw idth Product T
VCE = lV, I C = 10 mA
fab
Specification
hFE= 30 to 120
'EBO = 0.1 MA max
480 M c
hFE = 20 min
960 M c
hFE = 600 max
= 10 nA max I EBO 30 Mc min
hFE = 75 min
= 50 nA max 'EBO 130 Mc
hFE = 60 to 120
800 M c
1 1 1 1
I I 1 I I I 1 I 1 1 I I I 1
a
W CY
J 3
z LL
0" 2 ce W
0 m v, & W
3
> - - & W I-
? 6 Q e- 0
n
&
LL
CY W
5 e LL I-
W
0
Z V - - LL LL W
0 V Z 0 6 - I- 4 W & &
0 V 2 Z 6
Z
CY m -
2 2 W L m m W -I
7 I-
(u u)
03 9
9 m
3 2
3 %
-t m
u) 0
0. *
0 W L
II
I -
I\ u)
I
(u (u
I
9 0 I
I
I
N * I
2
9 u)
0 W
I-
f N 6
u) II V -
9 0 - t I
- 0
m * 8 I
0. d I
0
0 -
a3 (u
I
N -t I
oo rn - UJ
L
c W W
L 5 0 a
- c 0
I
(u 0 h m
- u)
2 %
m a3 0 0
I
m 7 (u (u
I I
u) 0. h m
u) u) 9 rr)
O0
W T . I- 0 u o
m o w
LL I: 0
W LL L
29
I
9 0.
0 a3
a3 0.
(u 0.
I
c 9
x
Q E 0 m II v -
6 E u) II V -
L
E .- V al Q v,
d 0.
hl
c
Z
30
I 1 1 I I I I I I I I I 1 I I I I I
I
L
E .- V al Q v,
0 0 00
c m w w LL LL I: I:
31
IT 9)
E .- V al Q v,
V al a v,
0. 9 m N
c\l Z
d 0.
(u
- Z
co 0.
(u
- Z
TABLE 7 2N914, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS
AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-
TION, ANNEALING AND FAILURE DATA
1 . 1 1
.88 1.22
.97
.94
Specimen
77 .87
1.8 7.78 2.0 1 .15 2.8 1.75 2.6 2.73
F2 F3 F4 F5
M2 M3 M4 M5
T2 T3 T4 T5
G 2 G 3 G4 G 5
01 h~~
57.5 64.1 80.6 64.1
69.4 63.3 79.4 74.6
~
50. C 50.5 64.1 55.6
62.5 96.2 82.9 59.5
PI h~~
.59
.56
.67
.63
.33
.34
.36
.37
.61
.59
.67
.66
.46
.37
.57 -48
Point
1.21 .52 1.08 I 1: 1 1.61 1.11 .34
- Beta prior to first i r rad ia t ion . - Beta a f t e r f i rs t i r radiat ion to 2.70 x 10 7
01
P I
F E h
r . h~ E
h - Beta a f t e r a n n e a l i n g a n d pr ior to s e c o n d i r rad ia t ion . FEo2
1st Fail Point - Dose w h e n hFE- -hFE /2. 01
2 n d Fail Point - Dose w h e n hFE = h /2. FEo2
33
TABLE 8 2N914, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS
AND GENERAL ELECTRIC, (Ic = 3C mA), hFE DEGRADA-
TION, ANNEALING AND FAILURE DATA
2.9 8.0 9.9
10
-7-c Specimen
1 1 .,-: .76
1.82 1.80
F2 55.8 F3 60.0 F4 72.1 F5 61.5
I M2 69.9 M3 60.5 M4 77.7 M5 72.8
T2 T3 T4 T5
G2 G3 G4 G 5
60.5 60.4 84.2 70.3
57.6 83.1 75.6 54.0
.65 1.09
.61 1 .c3
.70 1.09
.67 1.02
---I--- .40 .88 .45 1.17 .40 .91 .41 .89
1 .71 .86 .72 .94 .70 .91 .68 .84
1 s t Fai l Point
2nd Fai l Poini 1 s t Fai l Point
61 48 72 75
.89 1.54
.50 1.07
i
100 146 97
104
1.59 .87
1.21 .85
I 61 1.69
27 I 2.04 82 1 .oo 1.43
68 I - Beta prior to first irradiation. - Beta after first irradiation to 2.70 x 10 - Beta after annealing and prior to second irradiation.
7 01 h~ E
h~ E
FEo2
r.
P l h
1s t Fail Point - Dose when hFE = hFE /2. 01
FE FEo2 2nd Fail Point - Dose when h = h /2;
34
~
1 1 : . 'I I I I I I I I I I I I I I I I I I
TABLE 9 2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS
AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-
TION, ANNEALING AND FAILURE DATA
1 s t Fail Point
(x lo6 r) 1.8 0.4 1.7 2.7
1.2 0.9 0.2 1.7
30 48 21 28
9.5 32 27 1 1
- Beta prior to first irradiation.
- Beta after first irradiation to 2.70 x 10
- Beta after annealing and prior to second irradiation.
- Dose when h
7 01 h~~
h~~
FEo2
r.
P l h
1 s t Fail Point
2nd Fail Point - Dose when hFE= h /2
01 FE = h~~
/2. FE02
35
2nd Fail Point 1 s t Fai l Point
1.44 4.50 1.82 1.19
.58
.78 1.50
.82
1.30 1.17 1.86 1.46
2.84 1 .oo 1.48 2.91
TABLE 10 2N2192, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS
AND GENERAL ELECTRIC, (Ic = 30 mA), hFE DEGRADA-
TION, ANNEALING AND FAILURE DATA
105.3 169.5 193.4 123.5
163.9 172.4 227.3 104.2
Specimen
.34
.23
.31
.36
.26
.25
.21
.34
F2 F3 F4 F5 4.0
1.7 2.2
.4 7.1
M2 M3 M4 M5
1.20
1.18 .95
2.00 .44
T2 T3 T4 T5
25 35 14 16
G2 G3 G4 G5
1.40 1.43 2.36 2.13
h~~
142.9 137.6 185.2 153.1
140.8 141.4 133.3 142.9
.48
.55
.38
.41
.40
.54
.47
.41
I
1 1 34 24 14
3.00 1 .oo 1.63 2.71
h FE
.91
.95 1.03 1.12
01
1.01 1.04 1.04 1.15
.74
.84
.61
.65
.68 1.01
.79
.67
1 s t Fai l Point
x 10 r
2nd Fai l Point 1 s t Fai l Point
1.04 E:; I 5.20 3.2 1.28
- Beta prior to first irradiation. 01 h~ E 7 - Beta after first irradiation to 2.70 x 10 r. P I
h~ E
h
1st Fail Point - Dose when h = h / 2 .
- Beta after annealing and prior to second irradiation. FE02
FE FEol
2nd Fail Point - Dose when h = h /2. FE FEo2
36
1m I I I I 0 I I I I I I I I I I I I I
TABLE 1 1 2N2369, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS
AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-
TION, ANNEALING AND FAILURE DATA
.50
.65
.49
.84
.40
.37
.45
.32
.57
.53
.55
.58
.58
.70
.53
.67
I hFE 01 Specimen .74 .77 .71
1.03
.51
.53
.54
.49
.77
.77
.76
.73
.89
.88
.80
.92
F2 I 90.9 F3 F4 F5
94.3 92.6 51 .O
M2 M3 M 4 M5
T2 T3 T4 T5
102.0 74.6
1 1 1 . 1 83.3
68.5 102.0 76.9 53.2
G 2 G3 G 4 G 5
63.3 50.5 64.1 64.9
1 s t Fail Point
6 (x 10 r)
27 80 25
250
5.3 2.6
1.8 14
62 34 51 69
49 110 35
100
!nd Fai l Point 1s t Fai l Point
1.89 1.49 4.00 1 .oo
17.36 33.46 7.86
30.56
1.66 2.21 2.31 3.26
1.49 1.55 2.20 1.51
- Beta prior to first irradiation. 01 h~ E
P l h~ E
7 - Beta after first irradiation to 2.70 x 10 r. h
1 s t Fa i l Point
- Beta after annealing and prior io second irradiation.
- Dose when hFE = FEo2
/2 01 h~ E
2nd Fai l Point - Dose when hFE = h /2. FE02
37
TABLE 12 2N2369, FAIRCHI LD, MOT0 WLA, TEXAS INSTRUMENTS
AND GENERAL ELECTRIC, (Ic = 30 mA), hFE DEGRADA-
TION, ANNEALING AND FAILURE DATA
1 s t Fail Point 2nd Fai l Point
.73
.75
.73
.86
.56
.58
.56
.55
.81
.75
.78
.82
.84
.85
.79
.84
42 4.10 97 1.57 55 2.60
180 1.45
21 4.67 16 9.69 29 3.90 13 1.85
140 2.79 73 2.63
120 2.74 120 2.13
86 1.66 150 1.73 50 2.22
170 1.46
- Beta prior to first irradiation. - Beta after first irradiation to 2.70 x 10 - Beta after annealing and prior to second irradiation.
- Dose when h
7 01 h~~
h~~
FEo2
r.
P l h
1 s t Fai l Point
2nd Fai l Point - Dose when hFE = h
/2 01 FE = h~~
/2. FEo2
38
~ ~~ ~~ ~
TABLE 13 2N918, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS
2nd Fai l Point 1 s t Fail Point
I
.41 1.54 .57 .55
.86 1.30
.65 1.29
1.22 1.26
.60
.99
A
1.22 4.13
.73 1.45
AND GENERAL ELECTRIC, (Ic = 5 mA), hFE DEGRADA-
TION, ANNEALING AND FAILURE DATA
Specimen
F2 F3 F4 F5
M2 M3 M4 M5
T2 T3 T4 T5
01 h~~
G2 G3 G 4 G 5
- Beta prior to first irradiation. 7 - Beta after first irradiation to 2.70 x 10 r.
P l h~~
h - Beta after annealing and prior to second irradiation. FE02
1 s t Fail Point - Dose when h - /2 2nd Fai l Point - Dose when h = h /2.
01
FE FEo2
FE - h~~
39
TABLE 14 2N918, FAIRCHILD, MOTOROLA, TEXAS INSTRUMENTS
AND GENERAL ELECTRIC, (Ic = 30 mA), hFE DEGRADA-
TION, ANNEALING, AND FAILURE DATA
62.1 48.4 57.7 50.4
Specimen
.77
.72
.75
.68
F2 F3 F4 F5
71.9 100.0 103.1 100.3
M2 M3 M4 M5
.65
.64 0 59 .59
T2 T3 T4 T5
48.2 48.2 79.8 61.5
68.5 G 2 G3 G 4 G 5
.74
.76
.63
.69
01
hFEo2
hFE 01 1.01 1.02
.99 1.02
.99
.93
.98
.95
.99 1 .oo
.99 1.07
1st Fail Point
(x lo6 r)
300 82
1 50 91
82 85 87 75
150 140 64 93
No Useable Data
2nd Fail Point 1 s t Fail Point
.72 1.96
.83 1.60
1.54 1.79
.78 1.85
1.71 1.71 1.31 1.32
(All Specimens, 9 Including Control Specimens,
58.8 Very Unstable)
1 I I I - Beta prior to first irradiation.
- Beta after first irradiation to 2.70 x 10 7 01 h~ E
h~ E r. P l
h - Beta after annealing and prior to second irradiation. FEo2
1s t Fai l Point - Dose when hFE - - h~~ /2 9 2nd Fai l Point - Dose when h = h /2.
01
FE FEo2
40
41
7-
ambo O O N O
. . . a I
c L ' m u ) c o m - - 0
. . . a
a N h h 0 - m o . . . a
- - I - -
9 b L L L L
N 0.
N
N
- Z
0 0 . .
h a 0.h . .
h 0 ma . .
0 0
a * . . - -
o c a - w n
* ,
- h a 0 . ,
t ",I"" o m a c 90 c o o . . -
NC: h a
L
u) 0
X h a a
0
-
Lc
; - 0 -I- e 0
S 0
O -0 0
-I-
.- -I- .- L L .- c v) L
v- L
c v-
.- al
0 w w L L L L S L II I I
C 0
O -0 0
.- -I- .- L L .-
-0 c ou 2 e h
L 0 .- -0 C 0 0 C
0 W C C 0
W
0
.- -
L
c Lc
W LL L
I I - - N
Q O wo w w L L L L L
S L L
Z cv
a)
0.
cv
c
Z
42
C 0
0 -a 0
, - c
,-
L L 8 -
.I- v) L
.I- ,-
e L 0
C i ,- L
L
ul 0
X h 9 0.
0 W 0 U
0
0
-
Y
In
- + c
e S 0
0
-0
.- c .- 2 L
.I- v) L
Y L
.I- u-
.-
.- W
0 w w L L C L I1 II
C 0
0 U
.-
.I- .- 2
8 x
L .- U S
0
0
c L .- & U S 0 m S
0 W S S 0
W
0
.- -
L
c u-
w u,
5
I1 - - c v
Q O wo w w L L L L L L
L L 5
TABLE 17 S2N930, FAIRCHILD (Ic = 2 mA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
Specimen
AF046 AF 508 AF985 AF 1083 AF1213
BF210 BF373 BF1316 BF 1875 BF2173
F047 F052 F055 F066 F114 F160 F197 F217 F229 F263
01 h~~
488 526 400 435 435
385 57 1 426 377 377
435 392 38 5 526 488 385 444 51 3 408 408
P l h~ E
hFE 01 .32 .34 .29 .39 .37
.39
.31
.36
.33
.31 ~
.38
.24
.26
.31
.31
.31
.33
.37
.39
.36
hFE 01 .46 .42 .50 .46 .51
.52
.39
.47
.48
.48 ~
.51
.46
.47
.38
.41
.47
.45
.43
.49
.45
~~
1 s t Fai I Point
6 (x 10 r)
.31
.54
.76
.80
.70
.74
.44
.05 2.40
.33 ~ ~~
.54
.58 1.37 1.50
.51
.45
.22
.80
.86
.61
!nd Fail Point 1 s t Fa i l Point
39.0 27.4 18.3 23.0 21.1
23.6 28.9
324.0 8.8
42.1 ~
23.5 31.7 14.8 19.2 24.3 20.7 63.2 24.5 21.4 31.3
- Beta prior to first irradiation.
- Beta after first irradiation. 01 h~~
h~ E P I
h
1 s t Fail Point - Dose when hFE = h /2.
2nd Fail Point - Dose when h = h /2.
- Beta after annealing and prior to second irradiation. FE02
FEo 1
FE FEO2
43
TABLE 18 S2N930, FAIRCHILD (Ic = 2 rnA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
S pe c i rn en
F 296 F 299 F345 F 399 F446 F475 F 644 F 648 F687 F1174 F1232 F1317 F 1370 F1389 F1511 F 1 605 F 1 665 F 1898 F 1987 F2195
01 h~ E
-
408 455 444 465 370 435 385 465 408 455 51 3 377 408 476 476 3 70 455 476 385 392
P l h~~
h ~ ~ o l
.27
.23
.32
.35
.36
.32
.31
.31
.34
.38
.32
.31
.36
.34
.33
.40
.25
.35
.40
.28
h FEo2
.45
.44
.41
.43
.49
.46
.47
.39
.45
.44
.39
.48
.45
.38
.42
.54
.40
.47
.58
.57
1 s t Fai I Point
6 (x 10 r)
.89
.83
.67
.58 3.25
.40
.89
.89
.64
.59
.44
.33
.61
.38
.31
.86
.94
.59
.76
.24
2nd Fa i l Point 1 s t Fa i l Point
22.8 24.5 13.9 35.0 8.2
34.8 25.8 45.4 25.3 47.1 52.3 53.0 37.7 86.0 65.5 20.4 36.8 27.5 19.5 38.8
- Beta prior to first irradiation, - Beta after first irradiation.
01
P I
h~~
h~ E
h - Beta after annealing and prior to second irradiation. FEo2
1 s t Fa i l Point - Dose when h - /2 01 FE - h~~
2nd Fail Point - Dose when h = h /2. FE FEo2
44
TABLE 19 S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
Specimen
AF046 AF508 AF985 AF 1083 AF1213
BF210 BF373 BF1316 BF 1875 BF2173
F047 F052 F055 F066 F114 F160 F197 F217 F229 F263
469 538 405 426 422
43 1 568 459 361 389
~
433 364 392 529 493 3 58 452 472 406 41 3
h~~ P I
hFE 01 .42 .42 .38 .49 .47
.45
.39
.46
.43
.39
.46
.35
.35
.43
.41
.42
.41
.48
.47
.46
hFEo2
01
.76
.77
.73
.78
.79
h~ E
.77
.73
.66
.73
.73
.83
.74
.77
.65
.65
.73
.76
.79
.82
.73
1st Fai I Point
6 (x 10 r)
1.46 1.60 4.20 3.00 2.50
2.00 1 .oo 2.00 5.58
.94
2.30 3.90 4.20 4.90 1.70 1.68 1.38 2.84 2.63 2.40
- Beta prior to first irradiation.
- Beta after first irradiation. 01 h~ E
h~~ P l
h
1 s t Fail Point
2nd Fail Point - Dose when h = h /2.
- Beta after annealing and prior to second irradiation.
- Dose when h FE02
/2 01
FE FEo2
FE = h~~
45
2nd Fail Poini 1 s t Fa i l Point
2.73 1.83 2.30 2.38 2.32
2.81 3.17 5.72 2.42 4.47
1.71 2.55 1.96 2.83 2.72 2.74 2.74 2.04 1.87 3.07
TABLE 20 S2N930, FAIRCHILD (Ic = 10 mA), hFE DEGRADATION,
ANNEALING AND FAILURE DATA
Specimen
F 296 F 299 F 345 F 399 F446 F475 F 644 F 648 F687 F1174 F 1 232 F1317 F 1 370 F 1 389 F1511 F1605 F 1 665 F 1898 F 1 987 F2195
01 h~ E
3 72 437 406 450 368 405 379 44 1 377 472 51 3 364 397 488 465 366 429 444 389 424
P l h~~
hFE,I
.37
.32
.46
.48
.45
.41
.39
.40
.48
.46
.39
.41
.45
.44
.43
.49
.35
.45
.49
.34
h FEo2
.79
.76
.79
.72
.80
.82
.83
.78
.83
.82
.75
.86
.84
.76
.83
.88
.80
.87
.92
.84
1 s t Fai l Point
6 (x 10 r)
3.70 3.28 2.30 2.70 7.00 1.30 3.40 3.88 2.73 2.25
.98 1.40 2.15 1.90 1.50 3.02 3.25 2.24 2.95 3.65
2nd Fail Point 1 s t Fa i l Point
46
1.67 1.98 1.50 2.34 1.42 2.41 1.29 1.86 1.53 1.76 3.60 2.32 1.78 2.45 2.21 1.39 1.69 1.55 1.33 1.20
- Beta prior to first irradiation.
- Beta after first irradiation. 01 h~~
h~ E P l
h - Beta after annealing and prior to second irradiation. FEo2
1 s t Fai l Point - Dose when h - /2 01
FE FEo2
FE - h~~ 2nd Fai l Point - Dose when h = h /2.
TABLE 21 S2N930, TEXAS INSTRUMENTS (Ic = 2 mA), hFE DEGRA-
DATION, ANNEALING AND FAILURE DATA
Specimen
AT 1 203 AT 1 262 AT 1 263
BT986 BT 1 209 BT1401
T1136 T1144 T1154 T1188 T1216 T1311 T1370 T1397 T 1 406 T1439 T1484
01 h~~
278 3 28 227
244 222 31 2
235 41 7 303 377 263 225 230 263 235 31 7 247
P I h~ E
.57
.43
.54
.52
.36
.41
.39
.32
.54
.37
.37
.61
.67
.44
.53
.67
.43
h FE02
hFE,I .60 .55 .63
.63
.69
.58
.65
.48
.66
.53
.69
.81
.73
.63
.71
.63
.68
1 s t Fai I Point
(x l o6 r)
17.60 8.95
14.20
13.40 2.80
.50
1.90 3.80
15.70 2.00 6.10
20.50 25.50 8.10
13.10 24.10 7.20
2nd Fai l Point 1 s t Fai l Point
~~~ ~~
2.72 2.38 2.75
2.80 6.25
35.80
10.89 5.34 2.39
11.50 3.03 2.36 1.88 3.10 2.16 2.46 3.64
- Beta prior to first irradiation.
- Beta after first irradiation. 01 h~~
h~ E P I
h
1 s t Fail Point
2nd Fail Point - Dose when hFE = h
- Beta after annealing and prior to second irradiation.
- Dose when h FEo2
/2 01 FE = h~~
/2. FE02
47
TABLE 22 S2N930, TEXAS INSTRUMENTS (Ic = 10 mA), hFE DEGRA-
DATION, ANNEALING AND FAILURE DATA
Spe c i me n
AT 1 203 AT 1 262 AT 1 263
BT986 BT 1209 BT 1401
T1136 T1144 T1154 T1188 T1216 T1311 T 1370 T1397 T 1 406 T 1 439 T 1 484
01 FE h
287 332 238
25 1 239 3 22
2 50 433 31 3 380 282 233 242 267 243 344 258
P I h~ E
5 .63 .56 .61
.58
.49
.5t:
.56
.46
.55 0 52 .55 .65 .71 .56 .62 .69 .55
h FEo2
hFE,I .83 .81 .81
.83
.80
.82
.85
.70
.80
.75
.82
.81
.83
.80
.92
.83
.81
1 st Fai I Polnt
(x lo6 r)
22.40 14.90 19.50
17.70 11.20 8.50
17.20 10.10 22.80 13.50 13.70 24.00 29.90 16.40 19.80 26.40 15.40
2nd Fail Point 1 s t Fail Point
1.52 1.34 1.38
1.41 1.82 2.19
2.12 1.90 1.58 1.84 1.79 1.48 1.23 1.53 1.26 1.34 1.68
- Beta prior to first irradiction. - Beta after first irradiation.
01 h~~
h~~ P l
h - Beta after annealing and prior to second irradiation. FEo2
1 s t Fail Point - Dose when h = h /2. FE FEol
2nd Fail Point - Dose when h = h /2 FE FEo2
0.-o TtMhO. I - m . - a3--0.03 o* c o o o m w k - - 9 m m - . . . . . . . .
0 W L I:
m * m m 9-90
9903k m o o k m . . . . . . . 0°*0 L m m *
0 3 0 - m 0 l u ) k C o m o m - 0 . O k C v 109h9 9 9 k k
. . . . . . . .
0 0
P
x I - &
oo m - 0 9 9 u ) m
0 W LL C
. . . . I . . . . . . . . . a m . - - m m h l
- L 3 I W 4 - 0
m - I - v
w 0 : L -
* o c o O . m 9 m 0 . 0-99 m - 0 . - m h l m h l m m m m . . . . . . . .
0 W L
-r
* O h 0 I 0 - m - . . . . 0 3 0 9 m A d d d I " i?.q? *-09 . . . .
V ale z
I - &
m 0 . m o * o - - 0.--hlm **o- 9 k 9 9 9 9 k k . . . . . . . .
> 0 * II m
C :
49
O h 0 0 O o h l h 9 9 h - h hl - c u
o o o m m o m h l
0 7 - m m 9
0 3 0 9 m - m m - W 0 4
u . 5 L
0 w U I
m o o o 0 3 9 0 0 . - - o h m m
m o o o 25%%
O h 0 0 m m o n - m * -
-009 m - o o m h l m h l
0 0 9 h
C 4 c h - h 0 3 - m m
. . . . - -
o m * m . . . . 009hl . . . . 0 w LL I
* O m d h l
0 0 0 0 . . . . hl-l-hl
0 0 0 0 . . . . - m h l -
0 0 0 0 . . . . - - - - . . . .
0 0 0 0
V P)O 4
h . 5 . . . .
hl--hl . . . . c y - -
> 0
* II m w
2
C
E .- u e, a v,
50
1 T - R 1 I 1
I lil
( u m * o 00- - a s s * a 9 -0 r\ d o 0
- L a - . . . . m a *
h m h m - - * N
. . . . . . . . . c
co - ( u 9 m c o d 0 7 m s m -
O h - ( u N N N N
. . . . . . . . c o * s m m m m -
-
o x L L . 5 W
4 < I- n
0 W u- I n Z Q
O0 m W
c1 . 0
aJ I-
u' z U I-
W -4 w -I
oi W
W
< Z (3 0 Z Q
- m m 9 o o * o m o m k . . . . o * m o -ma)-
O h O h I\ocos . . . . L Z Z G 1 ( u m - 0 m a c o m . . . . 0 W u-
I o o m m a m s s - I - - I - N m m m - 9 s m 0 0 - ( u . . . .
0 W - 0 u
LL - -
n
. . . . . . sod h d
co h
0 W n
I - v w 0 : L L - I
m m * n . . . . m m m h - - o m ( u - N -
. . . . s m * m m h d o . . . . - - h - 0 Nh("
m m 9 0 O d N - m 0 . o . m . . . . o s - m m o m m m m - m 0 w
LL x
. . . . ( u * - m O h - a 3
. . . . c o m s m a o h m & 6 4 4 s m s s
cv - h * W 0 0 0 - . . . . (u-l- . . .
m o o 0 (u
O O O ( u o * m m h * -
c o o 0 0 90 h c o -
c9
n V
aJo x I - 5
m - h m m h m o - 0 0 0 . . . . . . . .
C
E
51
0 h aJ 0-
' 0 w o m ' < w v L L - aJ
v) v
I 0 3 0 3 0 3 . . . a m . . h m * o m 0.h hl m
0 W
- * h * 0 . 0 - m h l m c o m m m h o - o m 0 03030.- 9 d m m h 0 0 0 * * h a . . . . . . . . . . . .
LL -r
d 9 I I a h -
0 ( U h
I-- v aJ 0 - w c
L L - x
m o m 0 . . . .
m m m o . . . . 0 w
LL 2
~ ~~~
- - m m * * a m m * m m ohloo 0000 0000
. . . . . . . . . . . . - 0 m - o m 0 0 . . . .
.- V 0 P m
52
v u a cu'
OI
W
03
c7 09
m 9 m ch
0 9 0
0 d -
I-
53
m m * m
m hl hl
0
03 d 9 c
h l r n d l n ln c S
E" 3
S
0 X a I-
L c ln
CI
ln
hl m hl
0
hl
c703om c7mcoco hl-hlhl . . . .
Cd W
2 0
7 II
z z
v, 7 a, - 0
I--
3 wo C d L L < I : W % E
2
co hl W J
I-
co 0.
hl
- Z
0. 9 m hl
N Z
hl 0.
hl
(u
c
Z
d 0.
hl
c
Z
I
W 0 I
0 c
co 0
- 03
0
9 0 I
54
09 0
d 0
- 0 c
- hl
0
hl
0 I
co 0
hl h
0
rn
0
c
m hl
I
o?: W I-
9 cd
(3 Z 0 F: Q n
(3 a
2
5 2
2 2
W
W
W a r I-
W z e(
U W a v,
I- 3 0 (3 Z i= U W A W v,
>- m
N i Q
Z rz (3
a W
W cd
0. hl W -I
2 +
c .- h 08 (3-
L
c (I) w L 5 0 a
w Q x I-
- m
h l m 0 3 9 c
- \ h 00
x x hhl d - 9 m
c c
. .
- \ n 0 0
x x m03 m - m
c c
. . -
.. .. $ 2 a = 0 3
U
> e - v ) .- L O K
U S
LL
64
s a .-
- u7 -
d - 0. Z hl
h( 0.
hl
hl
c
Z
0 m
55
03 - Cr 7 L hl
In c a, E .- V e, 0 V .- L c V 0)
W
0 e, c
- - L
3 In In
w -I
*
% 0 U C 0
V e,
.- c
- 0 2
P
I
a L
-0 e, V c
E L
Ou C w 0)
A
0 II In
E 0, 3 - In .- II I-
* i t
ni W z W
0 m W
I-
h h 0 0
x x 9 - 9 0 .
h h 0 0
x x l n h l h 0
m
c c
.. % B 9)
3
0 LL
C 0 U
L - .-
.- 9
.. L 3
0 LL
- .- c v) L .- LL
2 9) B
u)
h h 0 0
x x 0 . 0 .
0 . m
c c
c -
h h 0 0
x x 2 8 9
.. + 9) B E 3
0 LL
- .- C 0
73 .- 9
1. + 0)
3
0 LL
L
- .- c v) L .-
LL
2 9) B
0 c3
n
56
Q
0 -0
ln
C 0
V Q
.- c
+ c +
57
Z hl
9 0 hl
- L
9 0
X co m m
-
- L
n 0
X 0 hl
03
-
- 0 0
-
L hl v)
TABLE 32 MEDIAN PERCENT DEGRADATION VALUES AT SCREENING
DOSES (5% OF MEDIAN FAILURE DOSES)
d
I = 5 m A IC = 30 mA Type C
I
21 % 2N914 29.5% I
2N2192 21 % 18.5%
2 N2369 28.5% 22 % I
2N918 22 Yo 19.5%
A l l Combined 26 % 22 Yo
* I C = 2 m A IC = 10 mA
S2N930 12.5% 19 Yo I
58
- * - * m m h ( m c c - 9 . . . . m m o m 90-9 h - c v c v om.-*
* d c v d . . . . . . . .
c n l c v c
hhl*hl h h m k c
h h m o . . . . c v d c v 0 0 . . . . m 9 9 m - c v 0 . 0 . * c v d O . . . .
9 Q d O * d d 9
0000 o m - 0 m m m o m m n l m
0000 o n l m o h ( m Q 0 0 mmcvcv
0000 O m - Q m 0 . h c o m e - c v
c v m v m c c z z c v m * m 0000
59
Y V 0 - m E
9 0
2 x c v,
61
t
L e, u, (5)
t- .- L
4
62
C al
V Q, Q m
E .-
63
> > > > 0 0 0. ha . .
I-
w Z
0 m W
C Y 0 hl I v)
? N
V Y
I
I
h
X CY
+ L
5
ux
X CY
I1
I- X
64
u)
0 VI
us 'W
LL -c
0 0 0 0 0 0 9 u) d
0 0 8
0 u)-
X
d
c3
0 0 0 0 h(
0 0 0 0
65
99.99
99.9 99.8
99
98
95
9a
80
-0 70 al
0 u- C
- .- 60 c 50
40
LL 30
20
10
5
2
1
0.5
0.2 0.1
0.05
0.01
I Q,
10" 10' 10" 1
Gamma Dose (r)
FIGURE 6 2N914, FAIRCHILD, 27OC, PERCENT FAILED VERSUS GAMMA DOSE
66
99.99
99.9 99.8
99
98
95
90
80
70 U a,
0 LL
S
- e- 60
+. 50 40
2 30
20
10
5
2
1
0.5
0.2 0.1
0.05
0.01
L
11 7
1 o6 10' Gamma Dose (r)
1
FIGURE 7 2N914, MOTOROLA, 27'C, PERCENT FAILED VERSUS GAMMA DOSE
67
99.99
99.9 99.8
99
98
95
90
80
70
K 60
50 40
2 2 30
20
10
5
2
1
0.5
0.2 0.1 9.95
0.01
-0 e,
0
+
1 0' 10" 10'
Gamma Dose (r)
1 c)
FIGURE 8 2N914, TEXAS INSTRUMENTS, 27OC, PERCENT FAILED VERSUS GAMMA DOSE
6a
99.99
99.9 99.8
-0 a,
U L L
C a, V
a, a
- .- +
L
99
98
95
90
80
70
40 50 40
30
20
10
5
2
1
0.5
0.2 0.1
0.05
0.01
I
I
4-
I
1
f 1 o6 1 0’ 1 o8 10’
Gamma Dose (r)
FIGURE 9 2N914, GENERAL ELECTRIC, 27OC, PERCENT FAILED VERSUS GAMMA DOSE
69
99.99
99.9 99 .s
99
9s
95
90
80
70
._ 40 50 40
2 d 30
20
10
5
2
1
0.5
0.2 0.1 0.05
0.01
-0 a,
0 -
4-
1 o6
FIGURE 18 2N914, 27OC, PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS
70
10’
Gamma Dose (r)
0 13’
~ _ _ ~ ~ ~
99.99
99.9 99.8
99
98
95
YO
8 1
70
.- 69 U al
D LL
c V
- r. t3
a, 43 ., +
L
2 3 :
23
13
5
2
1
0.5
0.2 3.1
3.05
0.01
1 o6 Gamma Dose (r)
7
10' 1 o5
FIGURE 1 1 2N2192, FAIRCHILD, 27OC, PERCENT FAILED VERSUS GAMMA DOSE
71
s'9.99
99.9 99.8
99
98
95
90
33
70
.- 60 50
U
0
+
s 40 2 2 30
20
10
5
2
1
0.5
0 .2 0.1
0.05
0.01 1 o3 1 o6 3
Gamma Dose ( r )
FIGURE 12 2N2192, MOTOROLA, 27OC, PERCENT FAILED VERSUS GPMMP DOSE
72
I 1 ’ I 8 I 8 I E 8 I I 8 I I 8 I 8 I 1
99.99
99.9 99.8
99
98
95
90
80
70
“3 60
50 6 40
CL 30
20
10
5
2
1
0.5
0.2 0.1
0.05
0.01
U
.- 0
+
u
1 o6 1 o8 1 Gamma Dose ( r )
FIGURE 13 2N2192, TEXAS INSTRUMENTS, 27OC, PERCENT FAILED VERSUS GAMMA DOSE
73
99.99
99 .Y 99 .
8 I ’ I 1 I 8 1 I 1 I I( I 1 B 8 I t 8 1
99.99
99.9 99.8
99
98
95
90
80
70
60 50 40
30
20
13
5
2
1
0.5
0.2 0.1
0.05
0.01 1 o6
Gamma Dose (r)
10’ 1 o8
FIGURE 15 2N2192, 27’C, PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS
75
1 o6 1 o8 1 Gamma Dose ( r )
FIGURE 16 2N2369, FAIRCHILD, 27'C, PERCENT FAILED VERSUS GAMMA DOSE
76
8 8 ’ 8 8 8 8 I 1 I 8 8 8 8 8 B I 8 8 1
99.99
99.9 99.8
99
98
95
90
83
70
69 50 40
30
20
10
5
2
1 0.5
0.2 0.1
0.05
0.01
1 o5 1 o6 1 o8 Gamma Dose (r)
FIGURE 17 2N2369, MOTOROLA, 27’C, PERCENT FAILED VERSUS GAMMA DOSE
77
39.9 33.8
99
5;
95
4'c'
3L
IC
- .- 6C 5c
-0 (u
0
+
2 4c a" 3c Y
2c
1c
F - i 1
6.5
G . i G. 1
c.cc
c.01 1 G6 10' 1 c8
Gamma Dose (r)
FIGURE 18 2N2369, TEXAS INSTRUMENTS, 27OC, PERCENT FAILED VERSUS GAMMA DOSE
78
I I I I 8 I 8 8 8 8 I I 8 8 I 8 8 I 8
99.99
99.9 99.8
99
98
95
90
80
70 -0 a,
0 L L
t
- -- 60 + 50
40 2 2 30
20
10
5
2 1
0.5
0.2 0.1
0.05
0.01
Failure Defined As
-
FIGURE 19 2N2369, GENERAL ELECTRIC,
3 1 O'r
Gamma Dose
27'C, PERCENT FAILED VERSUS GAMMA DOSE
79
99.99
99.9 99.8
99
98
95
90
80
70 -0
0)
0 L
t
- .- 60 50
e, 40 2 ti? 30
20
10
5
2
1
0 .5
0.2 0.1
0.05
0.01
+
Gamma Dose
FIGURE 20 2N2369, 27OC, PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS
80
1 1 1 1
1 1 1 1 1 1 1 1 1 1 1 1 1 I
a
I 1 " I 8 8 8 8 8 I I I 1 8 8 I 8 8 I 8
99.99
99.9 99.8
99
98
95
9c
8C
7c
.- 6C c 5c
4c 2
-0 W
0 LI
C
-
2 3c
2c
1c
C w
2
1 0.5
0.2 0.1
0.05
0.01
Failure Defined As
7 10 r 6 10 r
I
9 10 r 8 10 r
Gamma Dose
FIGURE 21 2N918, FAIRCHILD, 27'C, PERCENT FAILED VERSUS GAMMA DOSE
81
99.9s
99.4 99.8
9s
98
9'
9c
8C
7c
6C
c 5c e, 4c 2 2 3c
73 0)
u LL
c
2c
1c
r; i
2
1 0.5
0.2 0.1
0.05
0.01 6 10 r 8 10 r
7 10 r
Gamma Dose
9 10 r
FIGURE 22 2N918, MOTOROLA, 27'C, PERCENT FAVILED VERSUS GAMMA DOSE
02
1
8 8 ' 8 8 I I 8 8 II 8 I I 8 I 8 8 1 8 I
99.99
99.9 99.8
99
98
95
90
80
70
60 50
40
30
20
10'
5
2
1
0.5
0.2 0.1
0.05
0.01 7
10 r 8 10 r 9 10 r
Gamma Dose
FIGURE 23 2N918, TEXAS INSTRUMENTS, 27'C, PERCENT FAILED VERSUS GAMMA DOSE
83
99.99
99.9 99.8
95
9e
92
9c
8C
7(
* - 6(
+ 5( 4(
-73 a,
0 LL
c
-
L
2 3(
2(
1 (
0..
0.: 0.
0.0.
0.0
.
7 10 r 6 10 r
9 10 r 8 10 r
Gamma Dose
FIGURE 24 2N918, GENERAL ELECTRIC, 27OC, PERCENT FAILED VERSUS GAMMA DOSE
84
I 8 1 1 I I I 8 I 8 I 8 8 I 8 8 8 I 1
99.99
99.9 99.8
99
98
95
90
80
70 U W
0 u- c
- -- 60 t 50 a, U 40
2 30
20
10
5
2
1
0.5
0.2 0.1
0.05
0.01
I
1 ( r 1 O'r 1 08r
Gamma Dose
FIGURE 25 2N918, 2 7 O C , PERCENT FAILED VERSUS GAMMA DOSE, COMPARISON OF MANUFACTURERS
85
r
3 *
0 3
0 N
0 -
0
86
1 - 1
I I I 1 I 1 I 1
I 1 I I I I 1 1
a
u
I I X O
v) c C (u
5 L c v)
C - : X 0) I-
I a
87
LL
0
L
9 0
X h( (u
( u *
Q
Q
0 n Q 2 z Q c3
LL
I-
w v,
Ln
v, @i w
3
>
0
h
w v,
n W @i
3 - Q LL
0 0 0 0 0 c3 hl c 0 0 0 9 u) *
0 hl
u) c
0 c
u)
88
1 . I
I I I I I 1 I 1 I 1 I I I I I I I
~~
I I 1 I I I I 8 I I I I I I I I 1 I I 89
> f
7 3
> 3
2 3
0 LL
- .- 0 U
I z 9 o z - 2 X : 6 E
0 0 c? (u 0 0 u) d
I
I I
,I I
I I I
I
I
I
' I I n W
I 0 c
Z 6 n -
- 0 * 7
0 N c
0 0 - 0 00
0 9
0 d
0 (u
0 I
I I I I I I I 1 I I I I i I I I I I B
0 0 0 0 0 0 0 9 m d m CJ -
91
0 03
0 9
0 *
3 N
3
0 0 0 0 0 0 c
0 hl m 9 d
92
1 1 I I I I I I I 1 I I I I I I I I I
7-t- I 4.__ +- I
+-+- 0 0 0 0 0 9 u) d (3 hl
93
3
- t - - - - T ----!----.
I 1
-tt --+* TI- - -- t
0 0 c
0 0 *
0 9 c7
0 hl c7
0 co hl
0 s
8 -0 L
N c
X
u. 0 $ 4 u) W E L O
a - 0 l l
94
0 N ?-,
0 m N
0
3
g 9o L
N - X
0 9 7
0 N -
0 03
0 *
0
0 9 m
0 cu 0
0 co cu
0 d hl
L
8 *o h l -
X
0 9 c
0 N 7
0 03
3 d
LLI 3 3
95
Q Q E € 0 m c ) II II u u - -
0 0
TT
I I
t-t r i I 1
I I I I +!
L 3 0 -
L . 0 c
DL 0
-)L 0
v)
C
E .- V e, Q v)
-0 e, C W e, V
C
L
v)
3
6 L 3
U LL
- .- + v) L .- LL
0 L
L
d E E 0
I I I I
I I