FLS HF6 LEVEL & PRESSURE TRANSMITTER
SAFETY INSTRUCTIONS General Statements• Do not install and service the product without following the Instruction Manual.• This item is designed to be connected to other instruments which can be hazardous if used improperly. Read and follow all associated instrument manuals before using with it.• Product installation and wiring connections should only be performed by qualified staff.• Do not modify product construction.
Installation and Commissioning Statements • Remove power to the sensor before wiring any connection.• Depressurize and vent the system before installing or removing the sensor.• Check and confirm the chemical compatibility of the materials in contact with the liquid.• Do not exceed maximum temperature/pressure data.• To clean the sensor, use only chemical compatible products.
2
PACKING LIST
Please verify that the product is complete and without any damage. The following items must be included:• HF6 Level and Pressure transmitter• Instruction Manual for HF6 Level and Pressure transmitter
DESCRIPTION
FLS HF6 is a new family of transmitters suitable for hydrostatic level/depth measurement and for pressure monitoring. PVDF body and ceramic diaphragm with FPM sealing system offer a smart combination of valuable wetted materials. The semi-flush diaphragm can work properly and for long time in pasty media and with liquids tending to crystallize. HF6 family can be also the finest choice for clean water applications in case customer would maximize operating time. The compact design can fit all main applications related to aggressive and corrosive liquids granting different solutions of installation: screwin (combined with a FIP valve or fitting), submersed with PUR cable running into pipe and submersed with PUR/FEP cable in contact with liquid. Other operating ranges and selectable length of cable in addition to sealing material options offer a customized solution to match the customer needs in the best possible way.
Specific Statements• The ceramic diaphragm is subject to breakage by over-pressure conditions or mechanical contact. Handle with care. Do not attempt to test the sensor by pressing on the diaphragm.• Use a suitable thread sealant to provide a watertight connection.• Hand tighten the nut. Do not use tools.• Do not allow the cable to be bent or compressed.• Do not allow damp spreads down the vent capillary to the rear of the diaphragm, the sensor may be damaged.• Blocked or damaged vent capillary cause an improper operating.• Short duration pressure pulses or shocks, called ‘‘water hammer,’’ are generated by a rapid change in flow rate caused by the operation of system components such as compressors, pumps, pistons and valves. The water hammer can reach pressure levels much higher than the nominal pressure risking to damage the ceramic membrane.• It’s suggested to install a pressure snubber (properly sized) at or near the input of a pressure sensor in order to protect it from water hammer damage. Snubber slows the rate of change of system flow.• When threading the sensor into a piping system, any backpressure can damage the ceramic diaphragm.• Open nearby valves to relieve any backpressure while threading the sensor into the pipe.• If the nearby valves cannot be open while threading the sensor into the pipe, a bleed valve can be installed to avoid overpressure.• Make sure there are no air bubbles in the pipe, particularly if a long and flexible hose is used to tap into the process pipe. Air bubbles could cause reading fluctuation and errors.
3
General • Operating range: see the table “Specific for HF6”• Accuracy*: ≤ ± 0.5 % FSO *Accuracy according to IEC 60770 – limit point adjustment (non-linearity, hysteresis, repeatability) • Wetted Materials: - Pressure port / housing: PVDF - Cable: 8 m PUR (FEP option)- Max cable length: 700 m - Seals: FKM (EPDM option) - Diaphragm ceramics: Al2O3 96% - Electromagnetic compatibility emission and immunity according to EN 61326• Thermal effects (Offset and Span) / Permissible Temperatures: - Thermal error: ≤ ± 0.2 % FSO / 10 K - Compensated range: -25 °C to 85 °C - Permissible temperatures: * Medium: -30 to 125 °C * Electronics/environment: -30 to 85 °C * Storage: -30 to 100 °C • Mechanical stability: - Vibration (25 to 2000 Hz) according to DIN EN 60068-2-6 - Shock 500 g / 1 msec according to DIN EN 60068-2-27 • Enclosure: IP68
Electrical• Supply voltage (VS): 8 to 32 VDC• Current consumption: max. 25 mA
• 1 X Current output: 4-20mA• Max loop impedance: Rmax = [(VS – VS min) / 0.02 A] Ω • Influence effects: - supply: 0.05 % FSO / 10 V - load: 0.05 % FSO / kΩ • Response time: ≤ 10 msec • Long term stability: ≤ ± 0,3 % FSO / year at reference conditions • Thermal error (Offset and Span): ≤ ± 0.2 % FSO/10 K• Permissible Temperature Range: -25 °C to 85 °C • Vibration: 10 g RMS (25 ... 2000 Hz) according to DIN EN 60068-2-6• Shock 500 g / 1 msec according to DIN EN 60068-2-27
Enviromental• Operating temperature: - medium: -30 to 125 °C (-22 to 257 °F) (referred to in line installation)- electronics / environment: -30 to 85 °C (-22 to 185 °F)• Storage temperature: -30 to 100 °C (-22 to 212 °F)
Standard & Approvals• Manufactured under ISO9001• CE
TECHNICAL DATA
Specific for HF6
ModelNominal pressure
gauge [bar]
Level[mH2O]
Overpressure [bar]
Burst pressure
[bar]
Vacuum resistance [bar]
HF6.004 0.4 4 1 2PN ≥ 1 bar: unlimited vacuum resistance
PN < 1 bar: on request
HF6.010 1 10 2 4
HF6.100 10 100 20 30
HF6.160 16 160 40 50
4
Maximum Operating Pressure/Temperature (using thread for coupling)
For water and non-hazardous fluids for which the material is classified as chemically resistant(25 years lifetime).
5
DIMENSIONS
INSTALLATION
FLS HF6 Level & Pressure transmitter is a gauge pressure sensor. Gauge pressure sensor measures the difference in pressure between the process on one side of a diaphragm and the atmospheric pressure on the opposite side of the diaphragm. A tiny capillary tube inside the sensor body is used to ensure that the back of the diaphragm remains at atmospheric pressure. If moisture is allowed to propagate down this tube to the rear of the diaphragm, the sensor may be damaged. Hydrostatic pressure measurement offers a level measurement can be a simple and practical solution to those applications where the fluid temperature is stable and constant. Be careful to apply this technology if the fluid is subject to temperature variations or any other variable that causes the density of the fluid to change.
Cable PUR 8mThread (R ½”)Wrench flatThread (G ¾”)FPM O-ringCeramic diaphragm Body: PVDF
1234567
6
ReccomendationPlease refer to SAFETY INSTRUCTIONS/Specific Statements
WIRING DIAGRAM
Electrical connections
Supply + wh (white)Supply - bn (brown)Shield gnye (green-yellow)
2-wire-system (current)
Supply +
Supply -
Relation between Pressure and Level is:h = p / (ρ * g)p = hydrostatic pressure [barg]ρ = fluid specific weight [kg/m³]g = force or gravitational acceleration [m/s²] = 9,80665 m/s² h = height of water column [m]
FLS transmitter can be installed in-line mode using a FIP saddle, FIP valves or a FIP T fitting through ¾” G thread male on the bottom of body, as well as, in submersion mode using a FIP double socket fitting ½” R thread female or, in according with the chemical compatibilities of cable, with cable in contact with fluid directly.
7
ORDERING DATA
HF6.XXX Level and Pressure Transmitter
Part No.
Description/Name
Applications/Operative
Range
Process connection
Main Wetted
Materials
Measuring Mode
Diaphragm CableWeight
(g)
HF6.004
Level/Pressure
Hydrostatic Transmitter
0-400mbarg
STD: ¾” G (class A)
SUBMERSED: ½” R (class A)
PVDF, PUR, FPM,
ceramic
Relative Pressure
Ceramic/ Semi flush
PUR 8m 550
HF6.010
Level/Pressure
Hydrostatic Transmitter
0-1000mbarg
STD: ¾” G (class A)
SUBMERSED: ½” R (class A)
PVDF, PUR, FPM,
ceramic
Relative Pressure
Ceramic/ Semi flush
PUR 8m 550
HF6.100
Level/Pressure
Hydrostatic Transmitter
0-10barg
STD: ¾” G (class A)
SUBMERSED: ½” R (class A)
PVDF, PUR, FPM,
ceramic
Relative Pressure
Ceramic/ Semi flush
PUR 8m 550
HF6.160
Level/Pressure
Hydrostatic Transmitter
0-16barg
STD: ¾” G (class A)
SUBMERSED: ½” R (class A)
PVDF, PUR, FPM,
ceramic
Relative Pressure
Ceramic/ Semi flush
PUR 8m 550
Other versions available:• SS AISI630 body equipped with SS AISI630 flush diaphragm and ranges from 0-1bar up to 0-100bar (gauge mode).• SS AISI316L body equipped with ceramic diaphragm and ranges from 0-0.1bar up to 0-25bar (gauge or absolute mode).
8
NOTE
9
NOTE
10
NOTE
11
NOTE