34
VPFlowMate User Manual UK / Europe Office Tel: +44 (0)845 9000 601 Fax: +44 (0)845 9000 602 [email protected] www.omniinstruments.co.uk Australia / Asia Pacific Office Tel +61 (0)282 442 363 Fax +61 (0)294 751 278 [email protected] www.omniinstruments.com.au USA / Canada Office Tel +1-866-849-3441 Fax +1-866-628-8055 [email protected] www.omniinstruments.net

Flowmeter Manual

  • Upload
    tepeji

  • View
    49

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Flowmeter Manual

VPFlowMate User Manual

UK / Europe Office Tel: +44 (0)845 9000 601 Fax: +44 (0)845 9000 602 [email protected] www.omniinstruments.co.uk

Australia / Asia Pacific Office Tel +61 (0)282 442 363 Fax +61 (0)294 751 278 [email protected] www.omniinstruments.com.au

USA / Canada Office Tel +1-866-849-3441 Fax +1-866-628-8055 [email protected] www.omniinstruments.net

Page 2: Flowmeter Manual

Contents

1 Getting Started 31.1 Principle of operation . . . . . . . . . . . . . . . . . . . . . . 3

2 Mechanical Installation 52.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.2 Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.3 Installation guidelines . . . . . . . . . . . . . . . . . . . . . . 62.4 Leak testing . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.5 Installation of In-line type flow meters . . . . . . . . . . . . . . 6

2.5.1 In-line flow conditioner . . . . . . . . . . . . . . . . . . 62.5.2 Quick couplings . . . . . . . . . . . . . . . . . . . . . 72.5.3 Compression fittings . . . . . . . . . . . . . . . . . . . 72.5.4 Flanges . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2.6 Installing the VPP insertion probe . . . . . . . . . . . . . . . . 82.6.1 Selecting the installation point . . . . . . . . . . . . . 92.6.2 Piping guidelines . . . . . . . . . . . . . . . . . . . . . 92.6.3 Other sensors and transducers . . . . . . . . . . . . . . 92.6.4 Determination of the insertion depth . . . . . . . . . . 112.6.5 Prepare the installation . . . . . . . . . . . . . . . . . 112.6.6 Sensor fixation . . . . . . . . . . . . . . . . . . . . . . 112.6.7 Sensor alignment . . . . . . . . . . . . . . . . . . . . . 112.6.8 Insertion depth . . . . . . . . . . . . . . . . . . . . . . 112.6.9 Software configuration . . . . . . . . . . . . . . . . . . 11

3 Electrical installation 133.1 Electrical connection guidelines- UL 508 Listing for USA & Canada 133.2 Electrical connection guidelines: general remarks . . . . . . . . 143.3 IO description . . . . . . . . . . . . . . . . . . . . . . . . . . 143.4 Connecting the VPFlowMate to your PC . . . . . . . . . . . . 16

3.4.1 Connecting the power supply . . . . . . . . . . . . . . 163.4.2 Connecting the RS232 . . . . . . . . . . . . . . . . . . 16

iii

Page 3: Flowmeter Manual

iv CONTENTS

4 Measurement & Data readout 174.1 Power and Warm up time . . . . . . . . . . . . . . . . . . . . 174.2 Zero check . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174.3 Reading the flow value (analog and digital signals) . . . . . . . 174.4 Reading the flow value (display) . . . . . . . . . . . . . . . . . 174.5 Time constants . . . . . . . . . . . . . . . . . . . . . . . . . . 18

5 Software 195.1 System requirements . . . . . . . . . . . . . . . . . . . . . . . 195.2 Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

6 Maintenance 216.1 General maintenance . . . . . . . . . . . . . . . . . . . . . . . 216.2 Cleaning the VPFlowMate . . . . . . . . . . . . . . . . . . . . 216.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216.4 Symptoms & causes & solutions . . . . . . . . . . . . . . . . . 21

A Nomenclature 23

B Electrical connections 25

Page 4: Flowmeter Manual

Warnings

In this manual, we have used the following signs:

Do not continue without reading and under-standing of this message. Not following theinstructions may lead to damage of equipment,personal injury of death.

Important warning message.

Application hint.

v

Page 5: Flowmeter Manual

vi CONTENTS

GENERAL WARNINGS

BY OPERATING THE VPFLOWMATE YOU CONFIRM THAT YOU HAVE READ

AND FULLY UNDERSTOOD THIS USER MANUAL AND YOU HAVE AGREED

WITH OUR WARRANTY STATEMENT.

FOLLOW ALL SAFETY PRECAUTIONS AND WARNINGS! NOT FOLLOWING

THE GUIDELINES COULD RESULT IN PERSONAL INJURY, DEATH AND

/ OR DAMAGE TO THE EQUIPMENT AND IT’S SURROUNDINGS.

NEVER USE THE VPFLOWMATE OUTSIDE THE SPECIFICATIONS AS

IT MIGHT DAMAGE THE INSTRUMENT OR GIVE MISREADINGS. ALWAYS

REFER TO THE SPECIFICATIONS OF YOUR FLOWMETER BEFORE USE!

TOLERANCES: UNLESS SPECIFIED OTHERWISE, ALL INDICATED

DIMENSIONS ARE NOMINAL.

VPINSTRUMENTS TAKES NO LIABILITY FOR ANY DAMAGES OR

MALFUNCTIONING OF THE FLOWMETER CAUSED BY MISUSE. IN ANY

CASE, WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF THE

DEFECTIVE PRODUCT.

THE VPFLOWMATE IS NOT APPROVED FOR ANY FISCAL METERING

APPLICATION. CONSULT YOUR NATIONAL OFFICE FOR APPLICABLE

LAWS AND DIRECTIVES.

THE INFORMATION IN THIS MANUAL HAS BEEN REVIEWED AND

IS BELIEVED TO BE CORRECT. NO RESPONSIBILITY IS ASSUMED

FOR ANY INACCURACIES. THE MATERIAL IN THIS MANUAL IS FOR

INFORMATION PURPOSES ONLY AND SUBJECT TO CHANGE WITHOUT

NOTICE

Page 6: Flowmeter Manual

viii CONTENTS

CE Declaration

DECLARATION OF CONFORMITY

Product type : Thermal mass flow

meter

Brand name(s) : VPInstruments,

VPFlowMate

Model numbers : VPF-R..-M..-D..-S...-E..

VPP-R..-M..-D..-S...-E..

Manufacturer : Van Putten

Instruments B.V.

Buitenwatersloot

335

2614 GS Delft

The Netherlands

Year of first

making

: 2001

Date of issue : November 21, 2001

The complete product family has been tested and found

to be compliant with the requirements of the following

European directives:

EN 61326-1EN 50082-1

Van Putten Instruments BV

Pascal F.A.M. van Putten

Director⋆ No signature has been printed for security reasons; contact us if a

hardcopy is required

Page 7: Flowmeter Manual

UL/ C-UL Listing mark

The VPFlowMate is UL/cUL Listed in accordance with UL 508 (Industrial Con-trol Equipment) and CAN/CSA 14-95 (Industrial Control Equipment), UL FileE231429. Details can be found on the UL website

www.ul.com

Please refer to chapter 3, section 3.2 for UL compliant use of the product.

1

Page 8: Flowmeter Manual

Chapter 1

Getting Started

Thank you for choosing the VPFlowMate digital mass flow meter. The VPFlow-Mate and the VPFlowMate insertion probe are easy to use mass flow meters forCompressed Air.

Check the packaging box for any inconsistencies. Should there be any ship-ping damage, notify the local carrier. At the same time a report should besubmitted to Van Putten Instruments BV, PO BOX 151, 2600 AD DELFT,The Netherlands.

1.1 Principle of operation

The VPFlowMate uses the Thermal Mass Flow principle. The sensor is madeof silicon, just like a computer chip.

Flow induces a temperature gradient in the silicon sensor. This gradient canbe measured with on-chip sensing elements. A heating element keeps the sensorat a constant temperature. Our silicon flow sensors have the following uniquefeatures:

• High dynamic range;

• High sensitivity and resolution;

• High reproducibility and stability over time.

3

Page 9: Flowmeter Manual

Chapter 2

Mechanical Installation

WARNING

VPInstruments TAKES NO LIABILITY FOR ANY DAMAGES OR

MALFUNCTIONING OF THE FLOWMETER CAUSED BY POLLUTION,

DIRT, DUST, WATER, OIL OR ANY OTHER CONTAMINATION OF

THE MEDIUM IN WHICH YOU ARE USING THE FLOWMETER.

DIRT, RUST PARTICLES, SAND, OIL DROPLETS OR DUST MIGHT

DAMAGE THE SENSOR, SO MAKE SURE THAT ALL PIPING AND

THE AIR IS CLEAN!

2.1 Introduction

The VPFlowMate is designed in such a way that it should be very easy for youto install. However, it is important to read all installation instructions, especiallywhen the product is new to you. The VPFlowMate intended for use with cleanair in light industrial environments (non hazardous). Experience with processcontrol equipment, sensors and other transducers is required.

5

Page 10: Flowmeter Manual

6 CHAPTER 2. MECHANICAL INSTALLATION

2.2 Applications

WARNING

These devices are only for use with air, nitrogen and

other non hazardous fluids or non combustible gases.

The maximum working pressure is 16 bar g.

Main applications of the VPFlowMate are: Compressed air flow measure-ment, air flow monitoring, metering, energy monitoring, testing of pneumaticsystems. Take a look at our Website www.vpinstruments.com for applicationexamples.

2.3 Installation guidelines

• Refer to the installation drawing or technical drawing for up- and down-stream length guidelines.

• Refer to the user manual and instructions of the Swagelok process con-nectors for instructions.

• Check the PIPING TABLE before installation of the flowmeter!

• Make sure that no Teflon tape is trapped within the tubing!

2.4 Leak testing

Perform a leak test at all times! Leakage can lead to dangerous situations orbad measurements.

Check the system on leaks using leak testing liquid or leak testing equipment.

2.5 Installation of In-line type flow meters

Depending on the in-line model you have, the VPFlowMate requires at least20D straight piping at the inlet and 5D at the outlet.

2.5.1 In-line flow conditioner

Your VPFlowMate can be equipped with built-in flow conditioners. These con-ditioners will reduce the required upstream and downstream length.

Page 11: Flowmeter Manual

2.5. INSTALLATION OF IN-LINE TYPE FLOW METERS 7

2.5.2 Quick couplings

Your VPFlowMate can be equipped with quick couplings for compressed airapplications. Follow the instructions of the quick coupling manufacturer.

2.5.3 Compression fittings

Follow the instructions of the fitting manufacturer.

2.5.4 Flanges

Flanges may be screwed into the threaded openings of the in-line VPFlowMate.Use Teflon tape or use O-ring sealed flanges. NEVER use liquid sealant, becausethey might permanently damage the screw thread!!!!

Page 12: Flowmeter Manual

8 CHAPTER 2. MECHANICAL INSTALLATION

2.6 Installing the VPP insertion probe

The VPFlowMate insertion mass flow meter is designed to operate under a widevariety of conditions. To ensure its longevity of operation, precautions shouldbe taken before and during its installation.

WARNING

INSERTION SENSORS CAN BECOME DANGEROUS

PROJECTILES!

THE IMPROPER USE OF INSERTION PROBES UNDER HIGH

PRESSURES (ANYTHING ABOVE AMBIENT PRESSURE)

IS VERY DANGEROUS. USE A SAFETY CHAIN UNDER

OPERATING CONDITIONS.

PLEASE KEEP CLEAR OF THE METER WHEN THE LINE

IS BEING PRESSURIZED. THIS IS A GOOD SAFETY

PROCEDURE FOR THE INSTALLATION OF ANY EQUIPMENT

IN A PRESSURIZED LINE.

DO NOT LOOSEN THE NUT UNLESS THE LINE IS

DEPRESSURIZED OR A HOT-TAP MECHANISM HAS BEEN

INSTALLED PROPERLY ON THE METER SHAFT.

USE HOT TAPPING EQUIPMENT WHICH IS RATED FOR THE

OPERATING PRESSURE OF YOUR COMPRESSED AIR LINE.

REFER TO THE MANUAL OF THE HOT TAPPING EQUIPMENT

BEFORE USE.

REFER TO YOUR PLANT SAFETY MANUAL FOR ADDITIONAL

SAFETY MEASURES.

Installation is a four step process. This four step approach can be regardedas a checklist for mobile measurements and air audits.

1. Select the proper installation site; follow all piping guidelines;

2. Prepare the installation (mount an insertion point for the Swagelock cou-pling, prepare electrical connections and power supply);

3. Insert and fix VPFlowMate;

Page 13: Flowmeter Manual

2.6. INSTALLING THE VPP INSERTION PROBE 9

4. Configure the VPFlowMate (this step may be done before step 2 if theinstallation site is completely known).

2.6.1 Selecting the installation point

• A) The VPFlowMate insertion flow meter requires minimum upstream anddownstream piping. Refer to the piping table for details!

• B) Choose a site with minimal mechanical vibrations.

• C) Avoid areas of high humidity or corrosive atmosphere where possible.

• D) When installing the meter choose a site which is accessible, and allowsease of wiring and maintenance.

2.6.2 Piping guidelines

To ensure the high accuracy of flow measurement specified on the identificationsheet, piping and installation instructions must be followed carefully.

• A) The pipe immediately up and down stream of the flow meter must beof sufficient length, straight and free of obstructions. Refer to the formerparagraph and to the piping table (figure 2.1);

• B) Weld beads on the internal wall of the pipe before or after the flowmeter should be ground flush with the pipe wall before the meter is in-stalled;

• C) Proper style and proper size gaskets should always be used when in-stalling the flow meter;

• D) The meter can be placed horizontally. When placed in a vertical linewith low air velocities, thermal convection which is caused by the flowsensor may give wrong readings.

Refer also to the ISO 14511:2001 (”Measurement of fluid flow inclosed conduits- Thermal mass flow meters”) International standard.

2.6.3 Other sensors and transducers

Any other transducer should be installed 5 times D downstream the VPFlowMateinsertion flow meter, because they might cause distortion of the flow profile.

Page 14: Flowmeter Manual

10 CHAPTER 2. MECHANICAL INSTALLATION

Fig. 2.1. In the piping table, you can find the requirements for the minimum pipinglength up- and downstream of the flow meter. Note that regulator valves shouldalways be installed downstream the flow sensor!

Page 15: Flowmeter Manual

2.6. INSTALLING THE VPP INSERTION PROBE 11

2.6.4 Determination of the insertion depth

Refer to the technical drawing provided with your VPFlowMate to determinethe insertion depth. The bottom of the measurement head should be in themiddle of the tube.

In de-pressurized lines, the internal diameter depth can be measured usinga caliper. A piece of tubing or a pencil can be used to mark the insertion depthon the probe.

2.6.5 Prepare the installation

Make sure the installation point is made according the technical drawings pro-vided on the VPInstruments Web site! If possible, place a 3

4inch ball valve,

so that the VPFlowMate insertion probe can be retracted and inserted underpressurized conditions.

2.6.6 Sensor fixation

SAFETY FIRST: START WITH MOUNTING THE SAFETY LINE!

The VPFlowMate inserts through a 1/2 inch compression fitting. Watchout with steel compression fittings as they might damage the sensor tube. Ifyou are planning to use the flowmeter at different places, use a Teflon ferruleinstead of the stainless steel ferrule. NEVER over-tight the fitting, because itmay damage the sensor tube.

2.6.7 Sensor alignment

After installing the meter, check for alignment. The flowmeter is shaped in away that alignment is very easy. Alignment ”by the eye” is sufficient.

2.6.8 Insertion depth

Check the insertion depth and make sure the sensor is inserted properly. Thebottom of the probe should be aligned with the middle of the tube. NOTE thatin small tubes, wrong insertion depth can result in large measurement errors,because of the blockage effect.

2.6.9 Software configuration

The VPSuite software can be used to set the pipe diameter and the outputunits. Refer to the VPSuite manual.

Page 16: Flowmeter Manual

Chapter 3

Electrical installation

The VPFlowMate complies with the CE requirements as stated in the ce dec-laration. CE compliance can only be achieved when grounding and shieldingdirections are followed and proper cables and connector assemblies are used.

3.1 Electrical connection guidelines- UL 508 Listing for

USA & Canada

The VPFlowMate is intended to be used with a Class 2 power source or Class2 transformer in accordance with UL1310 or UL1585.

As an alternative a LVLC (Low Voltage Limited Current) power source, withthe following properties shall be used:

• The device shall be used with an suitable isolating source such that themaximum open circuit voltage potential available to the product is notmore than 24 Vdc and the current is limited to a value not exceeding 8amperes measured after 1 minute of operation;

• A fuse in accordance with the UL248 series and rated max 4A shall beInstalled in the 24Vdc power supply to the device in order to limit theavailable current.

WARNING Wrong connections will damage or DESTROY the

instrument. Check ALL wiring diagrams!

Page 17: Flowmeter Manual

14 CHAPTER 3. ELECTRICAL INSTALLATION

3.2 Electrical connection guidelines: general remarks

Make sure that the following conditions are met:

• USE GALVANIC INSULATED POWER SUPPLIES: Switched mode powersupplies might generate a high voltage, high-frequency noise. They alsomight cause ground loops, which could affect the current loop output ofthe flowmeter.

• For portable, non-critical applications, a switched mode 12 V DC, 1Apower adapter may be used.Low cost or poor quality switched mode powersupplies might affect the accuracy!

• PAY EXTRA ATTENTION TOP RS232 PORTS. These are often sourcesof noise and other pollution, since some PC manufacturers do not takeEMC requirements seriously.

• ALWAYS GROUND THE INSTRUMENT HOUSING. GROUND the in-strument, as switched mode power supplies sometimes inject a high volt-age on the metal parts of the flowmeter.

• BEWARE OF GROUND LOOPS. User either one of the proposed groundpoints. Using both ground points might cause ground loop effects.

3.3 IO description

The VPflowMate features the following IO. Please note that the some outputsare optional; refer to the product type number to determine which IO yourVPFlowMate offers.

Page 18: Flowmeter Manual

3.3. IO DESCRIPTION 15

Power supply : 12 to 24 VDC (24V ±10%) Class 2. Seealso appendix B.1.

Serial interface : This is a standard RS232 interface, usingthree wires. Refer to figure ?? for wiringdetails

4..20 mA output : This is a passive, optical isolated linearoutput 4 mA corresponds with zero flow,20 mA corresponds with the maximumflow. See the flowmeter calibration reportfor the maximum flow value. See alsoappendix ?? and ??.

Pulse output : This is a passive, optical isolated low fre-quency pulse output. 0 mA = low and20 mA is high. It can be directly con-nected to pulse based energy counterswith active (12..24 volt powered) inputswhich work according to the DIN43864(S0) standard. Depending on the con-figuration of the flowmeter (see vpfcon-fig software) one pulse is equal to a fixedvolume, for example 1 normal cubic me-ter.

Depending on the model, the following electrical outputs are available. Pleasenote that the connections are based upon colors of the individual leads. Thenumbering of connectors may differ.

Page 19: Flowmeter Manual

16 CHAPTER 3. ELECTRICAL INSTALLATION

A separate leaflet indicating different or special wiring may be issued withthe product!

3.4 Connecting the VPFlowMate to your PC

WARNING: Wrong connection of the power supply may destroy theRS232 input of your laptop! Never apply +12..24 Volt to any of thepins of the SUB D connector!

In order to connect your VPFlowMate to a personal computer, you will need theVPK02 interface kit, or you can use the RS232 connection on the VPFlowTer-minal. From here, a standard RS232 connection cable is used to make theconnection.

3.4.1 Connecting the power supply

Always use a proper power supply. Refer to the specifications to make sure yourpower supply is +12..24 Volt Class 2. Connect the black wire to the 0 Volt sideof your DC power supply. Connect the red wire to +12..24 Volt side of yourpower supply.

3.4.2 Connecting the RS232

In order to connect your VPFlowMate to a personal computer, you will needthe VPK02 interface kit, which splits the cable into RS232 and power supply.Alternativelu, you will use the RS232 connection on the VPFlowTerminal. Astandard RS232 connection cable is used to make the connection.

Page 20: Flowmeter Manual

Chapter 4

Measurement & Data readout

4.1 Power and Warm up time

Before switching on the power check if all connections have been made accordingto figures in chapter 3. Check gas connections again and make sure there is noleakage.

Turn on the power. The instrument will start up and performs a self-calibration. Then it initializes. After this sequence, the instrument is partiallywarmed up. Complete warm-up time depends on the temperature of the sur-roundings and the flowmeter temperature. It may take up to one hour forcomplete warm-up.

4.2 Zero check

The VPFlowMate needs no zeroing as long as you keep the instrument clean.

4.3 Reading the flow value (analog and digital signals)

After warming up the instrument is ready to use. Generally, you can immediatelyread out the flow value. In some circumstances you might want to wait for afew seconds until the flow is stabilized. For instance, when the gas temperatureis changing.

4.4 Reading the flow value (display)

On the built-on display, you can see the flow on the first row. Total gas con-sumption is shown on the second row. The small underscore or minus sign onthe right indicates whether the flow is in ln/minute or m3n/hour. If an asterisk

17

Page 21: Flowmeter Manual

18 CHAPTER 4. MEASUREMENT & DATA READOUT

(*) is shown instead, the output is in m/sec, and the total count (second rowon the display) WILL NOT BE UPDATED.

Generally, you can immediately read out the flow value.In some circumstances you might want to wait for a few seconds until the

flow is stabilized. For instance, when the gas temperature is changing.

4.5 Time constants

Your VPFlowMate has a time constant, which depends on the flow rate. Whengoing from zero flow to a certain flow value, the sensor’s response is a fewseconds, since it is cooled down by the flow.

When you abruptly shut off the flow, it generally takes longer to stabilize,since the sensor has to cool down using thermal conduction to the non-movingair, which is a slower process.

Page 22: Flowmeter Manual

Chapter 5

Software

VPSuite is provided with your flowmeter. A SEPARATE MANUAL is issuedwith the software.

5.1 System requirements

System requirements are:

• PC or Laptop with at least Windows 95;

• Pentium or faster model;

• True color VGA screen, at least 1024x768;

• One free COM port;

• No PDA, modem, cellular phone software or other serial software activeduring measurements, as these might interfere with the COM port.

Install the software according to the instruction manual which can be foundon the CD.

5.2 Purpose

The VPFlowMate software package is designed to ease the installation of in-sertion probes in different tube diameters. It also enables you to change theflowmeter configuration. You can change sample frequency, zero cutoff value,span and so on.

The software also contains a data logger feature, for direct on site measure-ments with your laptop.

19

Page 23: Flowmeter Manual

Chapter 6

Maintenance

6.1 General maintenance

The VPFlowMate is a LOW MAINTENANCE product, as long as it is used inthe right way, with clean gases.

6.2 Cleaning the VPFlowMate

In case of extreme pollution you can clean the flowmeter by gently purging withnitrogen or clean air.

6.3 Calibration

Calibration is performed at VPInstruments.

6.4 Symptoms & causes & solutions

Use the table below to check on basic failures. Contact us if you have any otherproblem not stated in this table.

21

Page 24: Flowmeter Manual

22 CHAPTER 6. MAINTENANCE

Table 6.1. Possible Solutions to common Basic Failures.

Symptom Possible causes Action

No output signals No power supply Check powerBad cable Check cableWrong connections Check wiring diagramGeneral failure Consult factoryWrong installation The flow is too low to be

detected or the flowme-ter is mounted in a wrongway

Wrong output signal Out of calibration Consult factoryExtreme pollution Consult factoryWrong installation Check alignment of

probeCheck up- and down-stream lengthsCheck for obstructions inflow pipe

Offset signal at zero flow Extreme pollution 1. Wipe sensor clean2. Remove any particles3. Gently dry with cleancompressed air or Nitro-gen, or let product dry toAir.

No display or wrong dis-play

TX and RX wire swapped Check wiring

Display shows

O&@*&-

RX Line open Connect RX line (white)to ground

Page 25: Flowmeter Manual

Appendix A

Nomenclature

Table A.1. Nomenclature; x can be any letter/number

VPF -Rxxx -Pxxx -Dx -Sxxx -ExxxI II III IV V VI

Table A.2. Explanation of Nomenclature

I. Basic Type Designation VPP Insertion probeVPF In-line meter

II. Range Rxxx 0..xxx (mn/sec)III. Probe length: Pxxx 0..xxx (Length in mm)

or in-line meter diameter: Mxxx 0..x.xx inch diameter tubeIV. Display options Dx 0 or 1

0 = without Display1 = with Display

V. Output options Sxxx 0..xxx (manufacturer identification)VI. Connector options Exxx 0..xxx (manufacturer identification)

23

Page 26: Flowmeter Manual

Appendix B

Electrical connections

This appendix describes the electrical connection schemes for the VPFlowMate.

25

Page 27: Flowmeter Manual

26 APPENDIX B. ELECTRICAL CONNECTIONS

Fig. B.1. Power connection only. For models with built-on display

Page 28: Flowmeter Manual

Fig. B.2. Serial connection via RS232

Page 29: Flowmeter Manual

28 APPENDIX B. ELECTRICAL CONNECTIONS

Fig. B.3. Active current loop readout, where the readout unit also generates thepower for the 4..20 mA loop (”loop-powered”).

Page 30: Flowmeter Manual

Fig. B.4. Current loop readout example using a separate power supply and a passivereadout device.

Page 31: Flowmeter Manual

30 APPENDIX B. ELECTRICAL CONNECTIONS

Fig. B.5. Pulse output. The pulse output is a potential free, opto isolated output,and can be used as described in DIN43864.

Page 32: Flowmeter Manual

Fig. B.6. Pulse output, converted to Voltage.

Page 33: Flowmeter Manual

Index

CE-Declaration, viiicleaning, 23compression fittings, 7configuration, 11connection

electrical, 13power supply, 16RS232, 16to PC, 16

connectionselectrical, 27

couplings, 7current loop readout, 28

Data readout, 19display, 19

electrical outputs, 15EMC, 13

Failures, 23fixation

sensor, 11flanges, 7flow conditioner, 6

in-line, 6

gradient, 3

insertion depth, 11insertion probe, 8Installation

in-line meters, 6installation

electrical, 13

guidelines, 6insertion depth, 11mechanical, 5sensor, 11sensor fixation, 11

IO, 14

leak testing, 6

maintenance, 23

operationprinciple, 3

outputselectrical, 15

piping guidelines, 9

readoutcurrent loop, 28

RS232, 16

sensor, 3silicon, 3

sensor alignment, 11silicon sensor, 3software, 11, 21

installation, 21system requirements, 21

time constants, 20

UL-listing, ixwebsite, ix

VPConfig, 21

32

Page 34: Flowmeter Manual

INDEX 33

VPFlowDAQ, 21VPFlowmate

applications, 6software, 21

VPSuite, 11, 21

warm-up time, 19warnings, v

zero check, 19

UK / Europe Office Tel: +44 (0)845 9000 601 Fax: +44 (0)845 9000 602 [email protected] www.omniinstruments.co.uk

Australia / Asia Pacific Office Tel +61 (0)282 442 363 Fax +61 (0)294 751 278 [email protected] www.omniinstruments.com.au

USA / Canada Office Tel +1-866-849-3441 Fax +1-866-628-8055 [email protected] www.omniinstruments.net