Upload
ridzim
View
216
Download
0
Embed Size (px)
Citation preview
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
1/66
1FIXTURLASERSHAFT100
FIXTURLASER
Shaft100
USERS MANUAL
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
2/66
2FIXTURLASERSHAFT100
Publication No. P-0040-GB
2nd revised edition, January 1998
Copyright 1997 Fixturlaser AB, Mlndal, SwedenAll rights reserved. No part of this manual may be copied or reproduced in any form or by any
means without prior written permission of Fixturlaser AB.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
3/66
3FIXTURLASERSHAFT100
CONTENTS
INTRODUCTION ....................................................................................... 5THE METHOD ........................................................................................ 6THETOUCH SCREEN INTERFACE .................................................................. 8THE SYSTEM CONTENTS.......................................................................... 10SAFETYAND CARE ................................................................................. 12
FIXTUREATTACHMENT & CABLE CONNECTION ............................................. 16THE PROGRAMS.................................................................................... 18THE SYSTEM SET-UP ............................................................................... 20APPLICATION SET-UP .............................................................................. 21PRE-ALIGNMENT FUNCTIONS ................................................................... 23THERMAL OFFSET.................................................................................. 24
SOFTCHECK ...................................................................................... 26SHAFT ALIGNMENT ................................................................................ 28SHAFT ALIGNMENT HORIZONTAL MACHINES ................................................ 29STATICFEETSELECTION ............................................................................ 42VERTICAL SHAFT ALIGNMENT ................................................................... 44TD-VALUES ......................................................................................... 50
MEMORYMANAGER ............................................................................... 52TECHNICAL DRAWINGS ........................................................................... 56ICON INDEX ......................................................................................... 58TECHNICAL SPECIFICATION........................................................................ 64
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
4/66
4FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
5/66
5FIXTURLASERSHAFT100
INTRODUCTION
Congratulations on your choice of Fixturlaser Shaft 100. We hope that this system
will be your favourite tool from now on. This manual will take you through the
different procedures. Since the system is built for easiest possible use, you will not
need to read the entire manual but please take a look at the pages concerning
handling and care before you get started with your first alignment. The Fixturlaser
Shaft 100 user interface is entirely built upon signs and icons, no text. In most cases
the icons are clear and do not cause any misunderstanding, but to your convenienceyou will find an icon index in the end of this manual.
This manual describes the system and its functions. It is not an alignment handbook
sorting out the different problems you might run into when you perform an align-
ment. But it will help you with the basic procedures for each of the programs pro-
vided with the system.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
6/66
6FIXTURLASERSHAFT100
THE METHOD
FIXTURLASER SHAFT 100, HOWITWORKS
The Fixturlaser Shaft 100 is based upon the principle of a reverse indicator alignment
system. Instead of steel bars with dials it uses two laser beams. Unlike steel bars the
laser beams do not have the disadvantage of sagging and this contributes to the
systems high degree of accuracy.
The Fixturlaser Shaft 100 also has two measurement systems but in this case these
are electronic targets rather than mechanical dial indicators. These are built in
together with the laser transmitter in a pair of units, the Transmitter/Detector units
(TD-units).
With standard dial measurement techniques the results need to be graphed out
manually and calculations are needed to establish the corrections required. With theFixturlaser Shaft 100 all this is done automatically. Real time alignment values are
displayed as the machine is moved. You can see the effect of tightening a bolt as
soon as it happens.
The Fixturlaser Shaft 100 is equipped with a permanent memory where you can store
measurement results for later printout or transfer to computers for further processing.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
7/66
7FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
8/66
8FIXTURLASERSHAFT100
THE TOUCH SCREEN INTERFACE
A touch screen is, as it says, touch sensible. A light touch of your fingertip on the
displayed icon activates the icons function. We have found that a clean display
with just the necessary information, and with symbols in stead of text, makes it
much easier to understand, and reduces the input errors from the user to a minimum.
The user interface is based upon icons and fill-in boxes. The icons represent a func-
tion and activate the specific functions when touched. The fill-in boxes activate a
numerical or alphanumerical keyboard depending on the program you are in. Active
or touchable icons and fill-in boxes have a thick outline or/and a grey background.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
9/66
9FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
10/66
10FIXTURLASERSHAFT100
THE SYSTEM CONTENTS
PARTNO DESCRIPTION QTY
1-0203 Case 1
1-0200 Display Unit, DU10 1
1-0201 TD-M 100 1
1-0202 TD-S 100 1
KA-300569-66 Cable 3 meters 22-0299 V-bracket 2
2-0136 Chain, 500 mm with screw 2
2-0135 Chain, 900 mm 2
OM-PLASTBOX Plastic box 2
3-0777 Nut 2
3-0772 Tool 2
3-0773 Dolly 2
3-0769 Rod, length 60 mm 4
3-0770 Rod, length 150 mm 4
3-0343 Rod holder 2
5-0017 Allen key, 5 mm (optional) 1
1-0083 Magnet base (optional) 2
2-0314 Extension fixture TD 100 (optional) 22-0138 Cable for data transfer to computer (optional) 1
SK-MLC6S8X20 Screw 8x20 mm (optional) 2
2-0320 Measuring tape w. spirit level 1
OE-LR 20 Battery, Duracell Alkaline 1.5 Volts LR 20 4
1-0215 External power unit, universal (optional) 1
5-000* EUR:-3, US:-4, UK:-5, AUS:-7 (optional) 1 of 4
(Power cord to 1-0215)
1-0205 System printer (optional) 1
P-0040 Users manual 1
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
11/66
11FIXTURLASERSHAFT100
1-0203
1-0200
1-0201 1-02022-0299
2-01362-0135OM-PLASTBOX
3-0777
3-0772
3-0773
3-0769 3-07703-0343
5-0017
SK-MLC6S8X20
OM-PLASTBOX
KA-300569-66
P-0040
1-0083
2-0314
KA-30056
9-66
2-03
20
OE-LR 20
1-0215
5-000*
3-0343
1-0205
2-013
8
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
12/66
12FIXTURLASERSHAFT100
SAFETYAND CARE
POWERSUPPLY
The Fixturlaser Shaft 100 is powered by four alkaline batteries, size LR 20, in the
display unit or by the optional external power unit. The lifetime of the batteries is
approximately 24 hours when the system is used for a typical alignment job. The
power indicator in the main menu displays the power from the batteries. When the
power is low a battery replacement warning appears on the screen.
The batteries are replaced by opening the caps revealing the battery tubes. There are
two batteries in each compartment. The batteries should be replaced with the plus
pole pointing inwards the display unit and the minus pole pointing towards the caps.
You can use rechargeable NiCd-batteries but you should expect shorter operating
time than with alkaline batteries. If the system turns off due to low power, the system
returns, after battery replacement or connection of external power, to the applicationwhere it was when it turned off. No information is lost!
If the system is to be stored for a longer period of time the batteries should be
removed to prevent damage.
The external power unit is connected to the terminal next to the serial port on the
display unit and to a wall socket with 110 - 240 Volts.
SAFETY
The Fixturlaser Shaft 100 use two laser diodes with a power output of < 0,5 mW. The
laser classification is Class 2 which is considered safe for its intended use with only
minor precautions required. These are:
Never stare directly into the laser transmitter.Never shine the laser directly into anyone elses eyes.
Your system complies with the requirements in:
SS-EN-60825-1-1994
British Standard BS 4803 Parts 1 to 3
Deutsche Industrie Norm DIN JEC 76 (CO) 6
USA FDA Standard 21 CFR, Ch 1, Part 1040.10 and 1040.11
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
13/66
13FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
14/66
14FIXTURLASERSHAFT100
TREATMENTANDCARE
The Fixturlaser Shaft 100 has been developed for industrial use and is sealed against
water and dust in accordance with IP65. The system should be cleaned with a cotton
cloth moistened with a mild soap solution with exception for the detector surface,
which should be cleaned with alcohol.
Do not use paper tissue, which can scratch the detector surface. For the best
possible function, the laser diode apertures, detector surfaces and connector
terminals should be kept free of grease or dirt. The display unit should be
kept clean and the screen surface protected from scratches.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
15/66
15FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
16/66
16FIXTURLASERSHAFT100
FIXTURE ATTACHMENT& CABLE CONNECTION
Attach the V-block fixtures on the shafts of the measurement object, one on each
side of the coupling. Tighten the tension screws firmly, always using the supplied
tool. Do not overtighten.
Mount the rods to the fixture and tighten firmly. Attach the TD-units on the fixtures.
The TD-M should be mounted on the moveable machine and the TD-S on the station-
ary machine. If the shaft diameter is too large the chains can be extended with the
optional extension chains.Connect the cables between the display unit and the TD-units. There are two ways
you can do this:
1. Connect one cable between the display unit and one of the TD units. Then
connect the second cable between the TD-units.
2. Connect one cable from one of the terminals on the display unit to one of
the terminals on TD-M. Then connect the second cable from the other
terminal on the display unit to one of the terminals on TD-S.
You can use any of two terminals on each unit.
If, for any reason, you detach a cable during measurement you will need to return to
the main menu and then restart the measurement program.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
17/66
17FIXTURLASERSHAFT100
The optional extension fixtures are used together with either the V-block fixtures or
the magnet bases. If the space between the machine casing and the coupling/shaft
end is too short for a standard set-up this fixture solves the problem.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
18/66
18FIXTURLASERSHAFT100
THE PROGRAMS
Fixturlaser Shaft 100 is provided with different programs for specific purposes. When
you have pressed the red button the main screen appears. You have to return to this
screen to turn the system off.
Program 1: Horizontal Shaft Alignment
Program 2: Vertical Shaft Alignment
Program 3: TD-value displayProgram 4: File Manager
Program 5: The System Set-up
The other functions such as softfoot, static feet selection and thermal growth are
found as sub-functions to the programs.
The screen displays also abattery indicatorshowing the voltage level of the
batteries or anexternal power icon, and a contrast adjuster for the screen.
The bulb icon activates the screen background light. The background light
will be on for 5 minutes after the last use.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
19/66
19FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
20/66
20FIXTURLASERSHAFT100
THE SYSTEM SET-UP
The system Set-up menu includes settings that are common for all applications.
Displayed measurement units Select between metric and imperial units.
Auto off time Select between 1 and 98 minutes, When choosing
99 minutes the auto-off function is disabled. Auto-
off function is also disabled when using external
power.
Current time Sets the system clock.
Current date Sets the date. Display format will change accord-
ing to chosen measurement entity.
Printer selection Currently no options
Factory settings Resets all values to factory settings, also in the
application set-ups.
Factory settings are: Unit: mm
Printer: System printer
Auto-off: 5 min.
Resolution: 0,01
Filter: 3 sec.
Method: Clock method
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
21/66
21FIXTURLASERSHAFT100
APPLICATIONSET-UP
In each program there is different settings available valid for the specific program
only. The settings differ depending on the program. Below you will find a summary of
the different settings and what they do.
Displayed resolution Sets the displayed resolution of the measurement values.
Available settings are 0.1, 0.01 and 0.001 mm/10, 1 and
0.1 mils.Measurement method Sets whether you want to use the Clock-method or the
Tripoint method. Using the clock method you should
take readings at the 9, 3 and 12 oclock positions. With
the Tripoint method you can take three readings wher-
ever you want as long as it is at least 30 degrees between
each reading.
Filtered values Sets the sample rate for the internal filter calculations from
1 - 99 seconds.
Static feet selection Starts the static feet selection program.
Softcheck Starts the Softcheck program.
Thermal offset Sets the system in the mode for input of compensation
values.Tolerance table Displays a table with most often used tolerances. Imperial
or metric display depending on settings for displayed
entity
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
22/66
22FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
23/66
23FIXTURLASERSHAFT100
PRE-ALIGNMENTFUNCTIONS
In the efforts to obtain the best possible conditions for a shaft alignment it is
necessary to check whether the shafts are bent, the machine base is warped or
if there is a soft foot condition. Alignment of machines exposed to pipe strain
or thermal growth is easily performed with the Compensation value function.
The Thermal offset program provides the possibility to pre-set known compensa-
tion values for thermal growth. Values required are normally to find in themachine specifications. The Fixturlaser Shaft 100 accepts foot values, dial
values or angular-/parallel error values.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
24/66
24FIXTURLASERSHAFT100
THERMAL OFFSET
Most machines develop a certain amount of heat while running. In the best case
both the driving and the driven machine are affected equally requiring no input of
compensation values. But in several applications the driven machine is either
hotter, i.e. a pump for hot liquid, or cooler than the driving machine. The machine
manufacturers define the thermal growth of the machines differently, but in most
cases you will find it as a factor of deliberately misalignment expressed in parallel
offset and angular error.In the Fixturlaser Shaft 100 you can pre-set these values from the shaft alignment
program before starting your alignment work. Accepted values are feet values, clock
values and parallel offset/angular error.
PROCEDURE
1. When in the Horizontal Shaft
Alignment program touch theTool
icon.
2. Choose one of three ways to express
the offset values: feet values, TD
values or parallel/angular error.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
25/66
25FIXTURLASERSHAFT100
Feet values
3A Touch the feeticon where you want to
compensate.
4A Touch the foot valueicons. Enter the
distances required and the foot values
in mm or mils according to the pre-set
unit. You can change the enteredvalues by touching the foot value
icons.
TD values (Clock values)
3B Touch theclockicon.
4B Touch theTD-S value boxand enter the
value in mm or mils according to thepre-set unit. Do the same thing for the
TD-M value. Enter the distance
between shaft ends (DBSE).
Parallel/angular error
3C Touch themisalignment icon.4C Touch theValue boxesand enter the
misalignment values in mm/100 mm
and mm, or mils/inch and mils,
according to the pre-set unit.
5. If required you can now change to the
horizontal view for entering of com-pensation values. Touch OK. The
entered values are now pre-set and the
shaft alignment can be performed with
adjustments towards zero-values.
Note: The use of compensation values is indicated in the alignment program as asmall icon in the upper right corner (the lower left corner in static feet).
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
26/66
26FIXTURLASERSHAFT100
SOFTCHECK
A soft foot condition needs to be corrected before any alignment takes place. If not,
the measurement result will be of no value. It is more or less impossible to establish
if there is a soft foot condition without using some kind of measurement tool. The
Fixturlaser Shaft 100 built-in Softcheck program checks each foot and displays the
result in mm or mils.
The Softcheck program is entered from the Horizontal Shaft Alignment program set-
up. The measurements A, B, C and D have to be set before checking soft foot.
PROCEDURE
1. Start the Softcheck by touching its icon from the application set-up.
2. Check that all feet bolts are firmly tightened.
3. Select a bolt of your choice by touching its icon.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
27/66
27FIXTURLASERSHAFT100
4. Loosen the bolt fully and then tighten it firmly. Preferable with a dynamo-
metric wrench. The measurement value is registered by touching OK.
5. Continue with the rest of the bolts. Re-measurements can be done at any
time.
6. Do the necessary corrections and then check each foot again.
7. Exit by touching theExit icon.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
28/66
28FIXTURLASERSHAFT100
SHAFTALIGNMENT
INTRODUCTION
Shaft alignment - correction of the relative position of two machines that are con-
nected, such as a motor and a pump, so that the centre lines of the shafts form a
straight line when the machines are working at normal operating temperature. Shaft
alignment means moving the front and the rear pair of feet of one machine, verti-
cally and horizontally, until the shafts are aligned to within given tolerances.Alignment tolerances depend to a large extent on the speed of rotation of the shafts.
Machine alignment should be carried out to within the manufacturers tolerances.
The table below can be helpful if no limits are specified. The suggested tolerances
can be used as a starting point for developing in-house tolerances when the machin-
ery manufacturers recommended tolerances are not available. The tolerances are the
maximum allowable deviation from desired values, whether zero or targeted offset
for thermal growth.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
29/66
29FIXTURLASERSHAFT100
SHAFTALIGNMENTHORIZONTAL MACHINES
APPLICATION SET-UP
Here you can do the following settings valid for this specific program:
Displayed measurement value resolution: 0.1, 0.01 and 0.001 mm / 10, 1, 0.1
mils.
Filtered values: Sample time from 1-99 seconds
Thermal offset: Input mode for correction valuesStatic feet selection: Recalculates the measurement values
according to static feet choice
Softcheck: Sub-function for checking of soft foot
conditions
Choice of measurement method: The Clock method or the Tripoint
method.
Tolerance table: Displays a table with most often usedtolerances. Imperial or metric display
depending on settings for displayed
units.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
30/66
30FIXTURLASERSHAFT100
THE CLOCK METHOD
PROCEDURE
1. From the main menu touch theHorizontal Shaft Alignment icon (program 1).
2. The screen shows the movable machine. The grey areas are data entry fields.
Measure the distance A. Touch the A field and enter the value. Confirm with
OK. Continue with the B and C dimension. The D dimension is set by default
to half the A measure, but can be changed by touching the D field and
entering the correct value.
Note: The A dimension is measured from centre to centre of the rods. The B
dimension is measured from rod centre to first bolt centre and the C dimen-sion is measured between the bolt centres.
3. Stand by the machine to be adjusted facing the stationary machine.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
31/66
31FIXTURLASERSHAFT100
4. Rotate the shafts to the 12 oclock position, using the inclinometer displaywith the graphic spirit level. The led on the TD-M unit turns from flashing
green to alternate red and green when within 3 of correct position. The
targets should be slid over the detectors. Adjust the lasers to the centre of
both targets using the blue adjustment screws.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
32/66
32FIXTURLASERSHAFT100
5. Rotate the TD-units to the 9 oclock position. Slide open the targets, waituntil the TD-values appear, and press the9 oclockicon.
6. Using the inclinometer display, rotate the shafts to the 3 oclock position
and press the3 oclockicon. The screen now displays the current horizontalposition of the machine.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
33/66
33FIXTURLASERSHAFT100
8. Adjust the machine vertically until the values for both parallel and angular
alignment are within tolerances required.
9. Rotate the shafts to the 3 oclock position and adjust the machine horizon-tally until the required alignment is achieved.
7. Rotate the shafts to the 12 oclock position and press the12 oclockicon. Thescreen now displays the current vertical position.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
34/66
34FIXTURLASERSHAFT100
10. Rotate the shafts back to the 12 oclock position, press the12 oclockiconand check that the machine lies within the required tolerances.
11. Measurement and adjustment are now completed. To confirm the result,
redo the measurement. The result can be printed out by touching the Print
icon, or saved for later print-out or transfer to PCs by touching the File
Drawer icon. The measurement will be stored in the permanent memorylabelled with the current date and time. For more details see The Memory
Manager.
Note: It is the measurement results that is stored in memory and not the
displayed measurement after performed adjustments. To store this, redo the
measurement and then touch theFile drawer icon.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
35/66
35FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
36/66
36FIXTURLASERSHAFT100
TRIPOINTTMMETHOD
When shafts have a limited rotation or can only be rotated in one direction.
1. From the Main menu, touch theHorizontal Shaft Alignment icon (program 1
2. Touch theApplication Set-up icon to enter the set-up.
3. Touch theTripointTMicon to select the measurement method to use.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
37/66
37FIXTURLASERSHAFT100
5. Touch theRegister icon. This registers the first reading.
4. The screen displays the movable machine. Set dimension as for the Clockmethod. Set the TD units so that they are approximately parallel. Adjust the
lasers to the centre of the targets using the blue adjustment screws.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
38/66
38FIXTURLASERSHAFT100
7. Rotate the shafts to the third position and touch theRegister icon to register
the reading.
6. Rotate the shafts to the next desired position. The shafts have to be rotatedover a minimum of 30 and are indicated when theRegister icon is dis-
played. Touch theRegister icon to register the reading.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
39/66
39FIXTURLASERSHAFT100
8. To display the current machine position values touch the horizontal orvertical icon. If the shafts are rotated to a position where the TD-units is
NOT positioned in the 12/6 oclock or 9/3 oclock position, the values are
displayed with a black mark in the top right corner of the value box and are
not real time values. You can toggle between horizontal and vertical posi-
tion values by using the different icons. See Icon index
To adjust the machine in TripointTM mode
1. Real time adjustment is only available at 12 and 6 oclock position for
vertical adjustment and 3 and 9 oclock position for horizontal adjustment.
The inclinometer senses which position it is at and automatically updates
the readings as the machine is moved. The shaft must be within 3 from12/6/9/3 oclock positions before real time values are available. This is
indicated by the led on the TD-M changing from flashing green to alternate
red and green light.
2. If shaft rotation is not possible the vertical and horizontal views, without
real time values, can be used to obtain corrective moves but after a move-
ment is carried out the three point measuring sequence must be repeated.
This can be done at any time using there-measureicon.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
40/66
40FIXTURLASERSHAFT100
3. The result can be printed out by touching the Print icon, or saved for laterprint-out or transfer to PCs by touching theFile Drawer icon. The measure-
ment will be stored in the permanent memory labelled with the current date
and time. For more details see The Memory Manager.
Note: It is the measurement results that is stored in memory and not the
displayed measurement after performed adjustments. To store this, redo the
measurement and then touch theFile drawer icon.
Note:The shafts should be coupled during the measurement using the Tripoin method inorder to achieve as reliable and accurate results as possible.TIPS:The larger the angle over which the three points are measured, the fewer the movesand repeat measurements that will have to be made. Minimum angle betweenreadings is 30.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
41/66
41FIXTURLASERSHAFT100
STATIC FEET SELECTION
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
42/66
42FIXTURLASERSHAFT100
STATICFEETSELECTION
In some cases the machine that is displayed as the moveable machine is not move-able, or just some of the feet of the moveable machine are not adjustable. In order to
perform a proper alignment in these cases the Static feet selection program will be
helpful. In this program you can freely select which feet to be adjustable or locked.
The horizontal shaft alignment has to be accomplished before starting the program.
Procedure
1. From within the Horizontal Shaft Alignment program, touch theApplication
set-up iconand then theStatic feet selection icon.
2. Enter the E and F dimensions.3. Switch to horizontal/vertical view if necessary by touching theSwitch view
icon.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
43/66
43FIXTURLASERSHAFT100
4. Locking symbols for each foot are now displayed. Touch the two feet youchoose to lock. The feet values are displayed at the unlocked feet as live
values.
5. View the horizontal values by touching theSwitch view icon.
6. Different feet to lock can be selected by first unlocking the locked feet.
7. Go back to the horizontal shaft alignment program by touching theExit icon.
Note:
If the measurement is going to be saved after adjustment is made in the Static FeetSelection-mode, the measurement has to be re-made before saving.
VERTICAL SHAFTALIGNMENT
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
44/66
44FIXTURLASERSHAFT100
VERTICAL SHAFTALIGNMENT
APPLICATION SET-UP
Here you can do the following settings valid for this specific program:
Displayed measurement value resolution: 0,1 - 0,01 - 0,001mm / 10, 1, 0.1 mils
Filtered values: Sample time from 1-99 seconds
PROCEDURE
The vertical shaft program calculates the shims required under each bolt to correct
angular error and the live display shows the corrections required for concentricity.
1. From the Main menu touch theVertical Shaft Alignment
icon (program 2).
2. The screen shows the movable machine. The grey areas
are data entry fields. Measure the distance
between the TD units. Touch the A field and
enter the value. Confirm with OK. Continue
with the B value (Diameter of bolt pattern) and the
number of bolts. The D measure is by default set to
half the A measure, but can be changed by touching
the D field and entering the correct value. Any
values can be corrected if necessary. Note. The A
dimension is measured from centre to centre of the
rods. Set the TD units so that they are approximately
parallel at the 12 oclock position.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
45/66
45FIXTURLASERSHAFT100
3. Position yourself where it is easiest to turn the shafts through 180. Thisposition ideally should correspond to a 6 oclock position. The first bolt is at
the position 9 oclock. Turn the shafts to where the TD units are positioned at
9 oclock and touch the corresponding icon.
Tip: Mark up the different positions before you start measuring.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
46/66
46FIXTURLASERSHAFT100
4. Turn the shafts 180 to the 3 oclock position. Touch the corresponding iconto register the reading. The displayed values show the current position of the
machine in the 9 to 3 oclock axis.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
47/66
47FIXTURLASERSHAFT100
5. Rotate the shaft to the 12 oclock position and touch the12 oclock icon. Thevalues displayed are the alignment in the 12 to 6 oclock axis. The list of
values displayed is the position values for each bolt. These correspond to the
shim thickness that should be placed under each bolt to eliminate the
angular error. The first bolt value corresponds to the bolt at the 9 oclock
position. The parallelism error is corrected using the live display.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
48/66
48FIXTURLASERSHAFT100
6. Start with correction of the angular error by adding shims where required.
7. To adjust the machine in the 6 to 12 oclock direction, rotate the shafts to
the 12 oclock position and touch the corresponding icon. Use the parallel-
ism display to correct the error.
8. To adjust the machine in the 3 to 9 oclock direction, rotate the shaft to the
3 oclock position and touch the corresponding icon. Use the parallelism
display to correct the error.
9. The result can be printed out by touching the Print icon, or saved for later
print-out or transfer to PCs by touching theFile Drawer icon. The measure-
ment will be stored in the permanent memory labelled with the current date
and time. For more details see The Memory Manager.Note: It is the measurement results that is stored in memory and not the
displayed measurement after performed adjustments. To store this, redo the
measurement and then touch theFile drawer icon.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
49/66
49FIXTURLASERSHAFT100
TD-VALUES
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
50/66
50FIXTURLASERSHAFT100
The TD-value program is quite a simple one, but also very useful. The screen displaysthe TD values straight off.
APPLICATION SET-UP
Here you can do the following settings valid for this specific program:
Displayed measurement resolution: 0,1 - 0,01 - 0,001mm/10 - 1 - 0,1 mils
Filtered values: Sample time from 1 - 99 seconds
PROCEDURE
1. Touch theTD-value icon inthe Main menu.
2. The screen displays the physical readings from TD-M and TD-S.
3. By pressing the Zero icon, the values will be zeroed and the additional
functions will be available, such as zero the values, half the values, reset
values to physical, apply filter to the values. The angle sensor displays the
TD-M position.
4. By touching thePrint icon the TD-values and the angle value are printed
out, not a complete screen dump.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
51/66
51FIXTURLASERSHAFT100
MEMORYMANAGER
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
52/66
52FIXTURLASERSHAFT100
Touching the Memory Manager Icon from the Main Menu starts the memory man-ager. The Memory Manager screen displays a list of all stored measurements in the
memory. The measurements are stored in a chronological order with the latest
measurement at the first position. If the memory is full, the oldest measurement will
be erased automatically when a new measurement is stored.
ENTERINGAMEASUREMENT
The measurement is stored from the application program. The measurement will
automatically be stored labelled with the current date and time if no additional
labelling is made. It is the measurement results that is stored in memory and not the
displayed measurement after performed adjustments.
REVIEWINGAMEASUREMENT
To review a stored measurement, just touch it in the list. You can scroll the list by
touching theupand down icons. Open the selected measurement by touching the
File drawer icon. The appearing screen displays all information about the measure-
ment including time and date and eventually any made compensations, i.e. thermal
offset.
PRINTINGAMEASUREMENT
First of all you will need to connect the Fixturlaser system printer to the printer port
on the display unit. Select the measurement to be printed and then touch thePrint
icon. It is also possible to print out a screen dump of a reviewed measurement. Just
touch thePrint icon and the displayed screen will be printed out.
ADDORCHANGEAMEASUREMENTLABEL
Select the measurement to add or change label on. Touch theLabel icon.The alpha-
numerical keypad is displayed. Enter the new label and touch OK. Confirm with OK.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
53/66
53FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
54/66
54FIXTURLASERSHAFT100
DELETINGASELECTEDMEASUREMENT
Select the measurement by touching it in the list. Touch theDustbin icon. Confirm
the deletion by touchingOK.
DELETINGALLSTOREDMEASUREMENTS
Touch theFile drawer to dustbin icon. Confirm the operation by touchingOK.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
55/66
55FIXTURLASERSHAFT100
TECHNICAL DRAWINGS
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
56/66
56FIXTURLASERSHAFT100
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
57/66
57FIXTURLASERSHAFT100
ICON INDEX
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
58/66
58FIXTURLASERSHAFT200
THE MAIN MENU
Starts the Horizontal shaft
alignment program.
Starts the Vertical shaft alignment
program.
Starts the TD display program.
Starts the Memory manager
program.
Starts the Tool program.
Turns on/off the background
light.
Screen contrast adjustment.
Voltmeter indicating current
battery status.
Indicates that external power is
connected.
Turns off the system
THE HORIZONTAL SHAFT ALIGNMENTPROGRAM
Starts the Application set-up
program.
Registers measurement value
when using Tripoint method.
Registers measurement value at
the 9 oclock position when
using the Clock method
Registers measurement value at
the 3 oclock position when
using the Clock method.
Registers measurement value atthe 12 oclock position when
using the Clock method.
Prints out a screen dump.
Saves the measurement in the
permanent memory.
Alters between horizontal and
vertical result view.
Restarts the measurements with
entered values.
Indicates that settings in Thermal
growth program are activated.
Exits the program and returns to
the main menu.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
59/66
59FIXTURLASERSHAFT200
THE VERTICALSHAFTALIGNMENT
PROGRAM
Starts the Application set-up
program.
Opens the entry field for number
of bolts.
Registers measurement value at
the 9 oclock position when
using the Clock method
Registers measurement value at
the 3 oclock position when
using the Clock method.
Registers measurement value atthe 12 oclock position when
using the Clock method.
Prints out a screen dump.
Restarts the measurements with
entered values.
Saves the measurement in the
non-volatile memory.
THE TD VALUEDISPLAYPROGRAM
Starts the application set-up
program.
Prints out the readings and
additional information.
Restarts the measurements with
entered values.
Zeroes the TD-M value/resets to
raw data.
Halves the TD-M value.
Turns on/off filter function. Filter
time is set in the application set-
up menu.
Exits the program and returns to
the main menu.
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
60/66
60FIXTURLASERSHAFT200
THE
MEMORY
MANAGER
PROGRAM
Changes/adds a tag on selected
measurement.
Displays the selected measure-
ment results.
Deletes selected measurement
from memory.
Deletes all stored measurements
in memory.
Scrolls one page at a time up/
down.
Prints displayed measurement.
Exits the program and returns to
the previous menu.
Alters between viewing of
coupling values and feet values.
THE TOOLSPROGRAM
Selects between metric or
imperial display.
Sets the auto-off time in minutes.
99 minutes = deactivated.
Sets the system date. Display
format depends on measure unit
format.
Sets the system clock.
Sets the type of printer (if options
available)
Resets the system settings to
factory default. Affects applica-
tion set-up as well.
Confirms changes and exits to
the Main menu.
T A S U S
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
61/66
61FIXTURLASERSHAFT200
THE APPLICATION SET-UP
PROGRAMMES
Sets the displayed resolution. In
inch-mode the values are 10, 1
and 0.1 mils.
Sets the filter value.
Selects measurement method
Tripoint or Clock method.
Starts the thermal growth
function.
Starts the static feet function.
Starts the Softcheck function.
Displays the tolerance table.
STATICFEET
Locks the selected foot.
Unlocks all locked feet.
Indicates locked foot
Alters the displayed result
between horizontal and vertical.
Prints out a screen dump.
Exits back to the Application set-
up menu.
M SC O ST G O
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
62/66
62FIXTURLASERSHAFT200
MISCELLANEOUS
Value entry field.
Softfoot value box.
Confirms previous selection or
finishes sequence in program.
Result box with locked value
(not live).
Low power, battery replacement
necessary.
Confirmation of exit command.
Confirmation of deleting selected
measurement.
Confirmation of deleting entire
memory.
Confirmation of restarting
measurement.
THERMALGROWTH
Starts entering mode for pre-
setting of angular and parallel
errors.
Entry fields for pre-setting of
angular and parallel errors.
Starts entering mode for pre-
setting of feet values.
Entry field for pre-setting of feet
values.
Starts entering mode of pre-set
clock values.
Entry field for pre-setting of
clock values.
Prints out a screen dump.
Confirms changes and exits back
to the application program.
Exits back to the Application set-
up menu
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
63/66
63FIXTURLASERSHAFT200
TECHNICAL SPECIFICATION
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
64/66
64FIXTURLASERSHAFT100
TRANSMITTER/DETECTOR UNITS
Housing material Die-cast aluminium
Laser class Class 2
Laser wavelength 675 nm, visible red light
Detector area 20x20 mm (3/4x3/4)
Displayed resolution Selectable 0,1 - 0,01 - 0,001 mm
(10 - 1 - 0,1 mils)Measuring distance 20 meters (66 feet)
Inclinometer accuracy 1 (TD-M unit)
Temperature range 0 - 50 C (32 - 122F)
Filter for difficult measurement environment Steplessly variable
Light stabilization No interference from ambient light
DISPLAYUNIT
Housing material PP covered with DryflexTM
Display type 6 EL monochrome VGA
Power supply 4 x 1,5 Volts alkaline batteries
Measurement accuracy 1%0,003 mm
Temperature range 0 - 40C (32 - 104F)Battery lifetime 24 h in cycles of 50% operation
and 50% sleep mode
PPLICATION PROGRAMS
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
65/66
65FIXTURLASERSHAFT100
APPLICATION
PROGRAMS
Shaft alignment horizontal machines
Shaft alignment vertical machines
Measurement with compensation for offset/angularity
PROGRAMFUNCTIONS
Measurement of thermal growth
Softfoot measurement - SoftcheckTM
Base-bound or bolt-bound feet selection
Filter function for measurement values
Selection of measurement method - Tripoint or Clock method
Memory management for up to 100 measurements
Printer function
7/28/2019 Fixturlaser_Shaft_100_r2.pdf
66/66
66FIXTURLASERSHAFT200