LSC Meeting - Hanford WA, Aug ust 16th 1 LIGO-G050426-00-R First Results from the Mesa Beam Profile Cavity Prototype Marco Tarallo (Universita` di Pisa) oration with J.Agresti, E.D’Ambrosio, R. DeSalvo, D.Forest(*), B.Lagrange(*) owsky(*), C. Michel(*), J.L. Montorio(*), N.Morgado(*), L.Pinard(*), A.Remil P.Willems Laboratory Collaborators

First Results from the Mesa Beam Profile Cavity Prototype

  • Upload
    prue

  • View
    27

  • Download
    3

Embed Size (px)

DESCRIPTION

First Results from the Mesa Beam Profile Cavity Prototype. Marco Tarallo (Universita` di Pisa). In collaboration with J.Agresti, E.D’Ambrosio, R. DeSalvo, D.Forest(*), B.Lagrange(*), J.M. Mackowsky(*), C. Michel(*), J.L. Montorio(*), N.Morgado(*), L.Pinard(*), A.Remilleux(*), - PowerPoint PPT Presentation

Citation preview

Page 1: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

1LIGO-G050426-00-R

First Results from the Mesa Beam Profile Cavity Prototype

Marco Tarallo(Universita` di Pisa)

In collaboration with J.Agresti, E.D’Ambrosio, R. DeSalvo, D.Forest(*), B.Lagrange(*), J.M. Mackowsky(*), C. Michel(*), J.L. Montorio(*), N.Morgado(*), L.Pinard(*), A.Remilleux(*),B.Simoni, P.Willems

(*) = LMA Laboratory Collaborators

Page 2: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

2LIGO-G050426-00-R

Contents

Environment setup: description and first tests with spherical optics

MH mirrors: their shape and expected resonant beams

Sample C05008: profiles analysis and simulations Systematic and next steps

Page 3: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

3LIGO-G050426-00-R

Environment setup

Input/output optics bench:- Nd:YAG Mephisto laser- Mode match telescope- Fast photodiode for transmitted power readout- CCD camera to control the locked TEM Suspended FP cavity Profile readout bench (CCD camera, high resolution) Feedback control electronics & cavity mirrors DC

driving

Page 4: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

4LIGO-G050426-00-R

PD

Beam ProfilerDAQ

Control electronics

Image analysis and processing

FP cavity

Mode Match Telescope

Page 5: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

5LIGO-G050426-00-R

INVAR rod

MH mirror

Flat input mirror

Flat folding mirror

Vacuum pipe

Thermal shieldSpacer plate

2x 3.5 m

Environment setup

Fabry-Perot cavity structure in detail

Page 6: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

6LIGO-G050426-00-R

Cavity Lock Acquisition

Tested with a R=800cm roc spherical mirror Two techniques:- Side locking: control on the injection current -> easier- Dither locking: modulation of the cavity length ->

possibility to measure coupling with input beam but more sensitive to noise

Results:- TEM patterns characterization- Environment capability to keep a lock

Page 7: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

7LIGO-G050426-00-R

TEMs with spherical end mirrorsHermite-Gauss TEM set

Laguerre-Gauss TEM set

TEM00 TEM10 TEM20 TEM30

TEM10 TEM20

Resonant beams: experimental data

Page 8: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

8LIGO-G050426-00-R

TEMs with spherical end mirrors

Qualitative analysis:- Cylindrical symmetry

gradually lost- Difference between

theoretical Hermite-Gauss and actual TEMs beam profiles (structure in the residual map)

- Marked unbalance between the two TEM10 peaks: not avoided with fine PZTs adjustments

Page 9: First Results from the Mesa Beam Profile Cavity Prototype

LIGO-G050426-00-R

“Mexican hat” mirrors

Numerical eigenmodes for a ideal MH Fabry-Perot interferometer:

The fundamental mode is the so-called “Mesa Beam”, wider and flatter than a gaussian power distribution

Cylindrical symmetry yields TEMs close to the Laguerre-Gauss eigenmodes set for spherical cavities

Page 10: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

10LIGO-G050426-00-R

“Mexican hat” mirrors LMA laboratories provided three mirror samples C05004 (test run):- Thin substrate (20 mm)- large offset on the central bump C05008 & C05009:- Thick substrate (30 mm)- Both affected with a not negligible slope on the central

bump

We can characterize how mirrors imperfections affects the resonant beam in such a interferometer

Page 11: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

11LIGO-G050426-00-R

FFT simulations

Using paraxial approximation, FFT codes can simulate the propagation of actual TEM patterns on optical cavities

A Mathematica FFT routine has been dedicated to simulate our cavity beam behavior: it gave us the best tool to choose the best MH: C05008

Page 12: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

12LIGO-G050426-00-R

The slope on the central bump can be corrected applying the right mirror tilt

FFT simulations

5 nm error central area

Page 13: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

13LIGO-G050426-00-R

MH Cavity Alignment

Spherical optics: tilt is translated in a change of the optical axis

MH mirrors: only cylindrical symmetry

-> resonant beam phase front change with the alignment

Folded cavity: no preferential plane for mirrors alignment

-> very difficult align within rad precision

Page 14: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

14LIGO-G050426-00-R

Experimental Results

No stable Mesa beam profile has been acquired yet Higher order modes were found very easily

Page 15: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

15LIGO-G050426-00-R

FP spectrum analysis:- TEMs identification and

coupling analysis- Non-symmetric spacing:

as expected- More peaks than we

should see?

Experimental Results

Page 16: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

16LIGO-G050426-00-R

Other resonant TEMs:

Experimental Results

2-dimensional nonlinear regression:

Definitively not gaussian

Page 17: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

17LIGO-G050426-00-R

Misalignments and mismatching effects has been modeled to recognize “strange” resonant modes

No way to distinguish between them

Experimental Results

Page 18: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

18LIGO-G050426-00-R

TEM00 tilt simulation TEM00 data

Experimental Results

Page 19: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

19LIGO-G050426-00-R

Experimental Results

Page 20: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

20LIGO-G050426-00-R

Systematic and next steps

Any attempt to “drive” the beam in a centered configuration failed

FFT: even cylindrical symmetry is definitely lost FP spectrum analysis: peaks are separated enough

-> we are observing the actual cavity modes

Page 21: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

21LIGO-G050426-00-R

Coupling efficiency measurements:- Since TEM10 seemed very stable, we investigated

about the actual coupling coefficients and modes finesse

Strange evidence: every time we tried to align the cavity, mode shapes became worse and worse (as with spherical end mirror) -> coupling measurements are not concluded yet

Central part of the cavity seems “unstable”: maybe the problem is not the MH but the other two mirrors

Systematic and next steps

Page 22: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

22LIGO-G050426-00-R

Mechanical clumping, PZTs and screws stress yields deformations on the folder and input mirrors

~ 60 nm deformation -> three times the height of the MH central bump

Marked astigmatism is induced

FFT simulation with actual IM profile in progress

Systematic and next steps

Page 23: First Results from the Mesa Beam Profile Cavity Prototype

LSC Meeting - Hanford WA, August 16th

23LIGO-G050426-00-R

Next steps:- Change mirrors mounts (done!) and test new cavity

behavior- Model folder mirror effects on the resonant modes- Automatic alignment, vacuum operations…- Noise characterization: dithering possible only at low

frequencies (~10 kHz) -> maybe error signal too noisy (work in progress)

Systematic and next steps