Upload
erick-morris
View
215
Download
1
Tags:
Embed Size (px)
Citation preview
2 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
Introduction
This method is useful, if you want to:
• Dispense solvent in a rack.• Dissolve solid samples that are supplied in a rack.• Dilute samples that are supplied in a rack.
This document describes
How a complete rack or several vials in a rack can be filled with a user defined amount of solvent (Solvent from dilutor or small amounts from tray).
Two procedures will be used:
• The first fills the syringe with the total solvent amount. • The second quickly dispenses equal solvent parts into the selected
destinations.
Tip:
Prep-Gilson always requires source and destination positions. In this case only the second procedure needs the selection of destination positions.
3 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
2. ”Manual destination selection“ is checked.
3. ”Clean ... “ is unchecked.
Step 1: Check Settings
4 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
Open procedure editor dialog
Two main procedures will be defined:
1. Fill syringe with the total solvent amount 1200µl for 12 tubes.Name: ”Load Syringe DMSO 1200.gsp“
2. Dispense partial solvent amount 100µl into one destination.Name: ”Dispense Volume 100.gsp “(with subprocedure ”Sub Disp Volume 100.gsp “)
Step 1: Check Settings
5 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
2. Select the source type ”Solvent in Tray“.
3. Select your own values for:- Solvent name- Volume (Speed ...)
Tip: Never forget ”Update“ and ”Save“ after any modifications.
1. Check the file name ”Dispense Volume 100.gsp“
Step 2: Edit Procedure ”Load Syringe DMSO 1200.gsp“
6 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
Press the ”Edit“ button to edit or create a procedure of type ”Dispense“.
Step 3a: Edit Dispense Subprocedure ”Sub Disp Volume 100.gsp“
7 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
Select your own values for:- Volume - Z offset height (Speed ...).
Check the file name ”Sub Disp Volume 100.gsp“.
Step 3a: Edit Dispense Subprocedure ”Sub Disp Volume 100.gsp“
8 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
Select the source type”Gas Gap“.
Select your dispense procedure.
Check the file name ”Dispense Volume 100.gsp“.
Volume = 0
Step 3b: Edit Dispense Main Procedure ”Dispense Volume 100.gsp“
9 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
Start Automation.
1. Select procedure ”Load Syringe DMSO 1200.gsp“.
2. Select one dummy source and destination position (ones not used in the procedure!).
Tip:Solvent not used in procedure!
3. Create Order and do not start with ”OK“.
Step 4: Start Automation to Fill Sample Loop
10 19.04.23 05:49
Filling a Rack Matrix with Solventin Two Steps - Fast
1. Select procedure “Dispense Volume 100.gsp“.
2. Select 12 source and destination positions (destinations will be used).
Tip: Solvent not used in procedure!
3. Create orders and close dialog with “OK“.
Step 5: Create Orders to Fill Rack Matrix