FIGURE 5. Plot of peptide charge state ratios. Quality Control Concept Figure 6 shows a concept for the implementation of quality control as system suitability test as a system check before samples are analyzed and a concept for including a quality control sample in the sample queue. After the data acquisition of the quality control sample is finished, a quality control check is done automatically and the system is stopped if it does not fulfill minimum quality requirements. FIGURE 6. Outlook on a quality control system within Xcalibur / Proteome Discoverer software. 1. System suitability test QC sample and QC analysis, QC sample re- injection after corrective actions if system shows low performance 2. Quality control sample in sample sequence QC sample in sequence (and sequence pause if QC shows low system performance) Conclusion New Rawputator node allows Extraction of additional dynamic precursor ion information Diagnosis of instrument performance (LC + MS) and sample integrity Identification of outliers Method development and optimization References 1.Rudnick PA et al. Performance metrics for liquid chromatography-tandem mass spectrometry systems in proteomics analyses. Mol Cell Proteomics. 2010 Feb;9(2):225-41. 2.Köcher T, Pichler P, Swart R, Mechtler K. Quality control in LC-MS/MS. Proteomics. 2011 Mar;11(6):1026-30. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Overview Purpose: Implementation of proteomics data evaluation for quality control and method optimization. Methods: Hybrid linear ion trap-Orbitrap TM mass spectrometer with specialized proteomics software. Results: Rawputator tool delivers quantitative system performance assessment, method optimization and result verification. Introduction One of the biggest recurring questions in proteomics in a high-throughput environment is whether or not all parameters in the acquisition method are set optimally and the acquired data meets your quality criteria. Good practice therefore is to include quality controls in between the samples to assess the system performance. Figure 1 shows the system in a proteomic workflow which comprises sample preparation as well as LC, MS and data analysis. Why quality control? Optimum performance of the analytical system for the reproducible analysis of complex samples Identification of the source of the sub-optimal performance and variability Evaluation of the impact of the changes or improvements in the analytical system Documentation of the system performance metrics to support the assessment of the proteomic differences between biologically interesting samples Methods Sample Preparation Different enzymatic degradation preparations of complex proteomes were prepared using standard protocols. Liquid Chromatography All samples were separated by a Thermo Scientific EASY-nLC II liquid chromatograph using a peptide trap (C18 BioBasic, 100 μm inner diameter, 2 cm length) and a C18 analytical column (C18 BioBasic, 75 μm inner diameter, 10 cm length, both NanoSeparations, NL), at a flow rate of 300 nL/min using standard data-dependent acquisition methods. Mass Spectrometry Proteomic data sets were acquired using a Thermo Scientific LTQ Orbitrap Velos hybrid mass spectrometer. Data Analysis Data analysis was done using Thermo Scientific Proteome Discoverer software evaluation version 1.3. FIGURE 1. Proteomic workflow. Results Rawputator Node in Proteome Discoverer Software The Rawputator node can be easily added to any database search workflow within Proteome Discoverer TM software and extracts from the raw file additional dynamic precursor ion parameter per MS/MS spectrum as shown in Figure 2. The extracted values are stored in the .msf search output format. The additional time for the extraction is almost neglectable as well as the file size increase of the .msf output. The big advantage of having all that information in the search output is the additional connection to the search score. This can be used to evaluate the system performance. FIGURE 2. Rawputator node in Proteome Discoverer software. Lockmass Correction One example is to plot the lockmass correction over time (see Figure 3). Low (internal) lockmass correction values indicate on the one hand a valid external mass calibration and on the other hand indicate appropriate external mass calibration intervals. FIGURE 3. Lockmass correction over time. Charge state ratios One example for a digestion quality check is the plot of the charge state ratios of all ions triggered for MS/MS. Figure 4 shows a plot of all precursor ions triggered for MS/MS per charge state and identified peptides per charge state for a single run. If the same enzyme is used for digestion and other instrument parameters are not changed then the ratio of the triggered precursor ions (2+/3+, 2+/4+, 3+/4+, ...) should be on the same level (see Figure 5). This plot also allows the identification of possible outliers. In this example the same sample is analyzed in triplicates with different gradient lengths. The number of identified proteins is for the 90-min and 45-min runs at the same level but for the 75-min and 60-min runs one outlier per triplicate can be spotted. For 75-min run 2 and for 60-min run 3 shows deviation from the normal distribution for the precursor ion charge states as well as in the number of identified proteins. FIGURE 4. Number of precursor ions triggered for MS/MS and identified peptides per charge state. Rawputator – A New Tool to Combine Proteomic Data Mining PP443 with Method Development, Result Validation and Quality Control Martin Zeller, Bernard Delanghe, Christoph Henrich, Torsten Ueckert Thermo Fisher Scientific, Bremen, Germany Sample preparation Mass spectrometry Data analysis Proteome Digest Ionization (ESI) Full MS MS/MS Quantitation Identificatio n Extraction of the dynamic parameter: - Precursor S/N - TIC - Ion inject time - Elapsed scan time - Effective collision energy [eV] - Lockmass correction Additional „costs“: - ~ 6% .msf file size increase - < 3 min per search job (complex cell lysate sample, 13665 MS/MS) date dataset proteins 11/6/100:22 90m in 1 676 11/6/102:33 90m in 2 672 11/6/104:43 90m in 3 675 11/6/106:44 75m in 1 634 11/6/108:44 75m in 2 599 11/6/1010:45 75m in 3 644 11/6/1012:30 60m in 1 601 11/6/1014:16 60m in 2 601 11/6/1016:02 60m in 3 638 11/6/1017:32 45m in 1 553 11/6/1019:03 45m in 2 552 11/6/1020:34 45m in 3 550 0 2000 4000 6000 8000 10000 12000 2+ 3+ 4+ 5+ 6+ charge state P recursor ions Identified peptides date dataset proteins 11/6/100:22 90m in 1 676 11/6/102:33 90m in 2 672 11/6/104:43 90m in 3 675 11/6/106:44 75m in 1 634 11/6/108:44 75m in 2 599 11/6/1010:45 75m in 3 644 11/6/1012:30 60m in 1 601 11/6/1014:16 60m in 2 601 11/6/1016:02 60m in 3 638 11/6/1017:32 45m in 1 553 11/6/1019:03 45m in 2 552 11/6/1020:34 45m in 3 550 NO : R estore system performance Q C sam ple real sam ples Q C analysis Q C ok? YES Q C sam ple real sam ples Q C analysis Q C ok? NO : P ause sequence, restore system perform ance and reinject sam ples YES : C ontinue w ith sam ple injections

FIGURE 5. Plot of peptide charge state ratios. Quality Control Concept Figure 6 shows a concept for the implementation of quality control as system suitability

Embed Size (px)

Citation preview

Page 1: FIGURE 5. Plot of peptide charge state ratios. Quality Control Concept Figure 6 shows a concept for the implementation of quality control as system suitability

FIGURE 5. Plot of peptide charge state ratios.

Quality Control Concept

Figure 6 shows a concept for the implementation of quality control as system suitability test as a system check before samples are analyzed and a concept for including a quality control sample in the sample queue. After the data acquisition of the quality control sample is finished, a quality control check is done automatically and the system is stopped if it does not fulfill minimum quality requirements.

FIGURE 6. Outlook on a quality control system within Xcalibur / Proteome Discoverer software.

1. System suitability test

QC sample and QC analysis, QC sample re-injection after corrective actions if system shows low performance

2. Quality control sample in sample sequence

QC sample in sequence (and sequence pause if QC shows low system performance)

ConclusionNew Rawputator node allows

Extraction of additional dynamic precursor ion information

Diagnosis of instrument performance (LC + MS) and sample integrity

Identification of outliers

Method development and optimization

References1. Rudnick PA et al. Performance metrics for liquid

chromatography-tandem mass spectrometry systems in proteomics analyses. Mol Cell Proteomics. 2010 Feb;9(2):225-41.

2. Köcher T, Pichler P, Swart R, Mechtler K. Quality control in LC-MS/MS. Proteomics. 2011 Mar;11(6):1026-30.

All trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

OverviewPurpose:

Implementation of proteomics data evaluation for quality control and method optimization.

Methods:

Hybrid linear ion trap-OrbitrapTM mass spectrometer with specialized proteomics software.

Results:

Rawputator tool delivers quantitative system performance assessment, method optimization and result verification.

IntroductionOne of the biggest recurring questions in proteomics in a high-throughput environment is whether or not all parameters in the acquisition method are set optimally and the acquired data meets your quality criteria. Good practice therefore is to include quality controls in between the samples to assess the system performance. Figure 1 shows the system in a proteomic workflow which comprises sample preparation as well as LC, MS and data analysis.

Why quality control?

• Optimum performance of the analytical system for the reproducible analysis of complex samples

• Identification of the source of the sub-optimal performance and variability

• Evaluation of the impact of the changes or improvements in the analytical system

• Documentation of the system performance metrics to support the assessment of the proteomic differences between biologically interesting samples

Methods Sample Preparation

Different enzymatic degradation preparations of complex proteomes were prepared using standard protocols.

Liquid Chromatography

All samples were separated by a Thermo Scientific EASY-nLC II liquid chromatograph using a peptide trap (C18 BioBasic, 100 μm inner diameter, 2 cm length) and a C18 analytical column (C18 BioBasic, 75 μm inner diameter, 10 cm length, both NanoSeparations, NL), at a flow rate of 300 nL/min using standard data-dependent acquisition methods.

Mass Spectrometry

Proteomic data sets were acquired using a Thermo Scientific LTQ Orbitrap Velos hybrid mass spectrometer.

Data Analysis

Data analysis was done using Thermo Scientific Proteome Discoverer software evaluation version 1.3.

FIGURE 1. Proteomic workflow.

Results

Rawputator Node in Proteome Discoverer Software

The Rawputator node can be easily added to any database search workflow within Proteome DiscovererTM software and extracts from the raw file additional dynamic precursor ion parameter per MS/MS spectrum as shown in Figure 2. The extracted values are stored in the .msf search output format. The additional time for the extraction is almost neglectable as well as the file size increase of the .msf output.

The big advantage of having all that information in the search output is the additional connection to the search score. This can be used to evaluate the system performance.

FIGURE 2. Rawputator node in Proteome Discoverer software.

Lockmass Correction

One example is to plot the lockmass correction over time (see Figure 3). Low (internal) lockmass correction values indicate on the one hand a valid external mass calibration and on the other hand indicate appropriate external mass calibration intervals.

FIGURE 3. Lockmass correction over time.

Charge state ratios

One example for a digestion quality check is the plot of the charge state ratios of all ions triggered for MS/MS. Figure 4 shows a plot of all precursor ions triggered for MS/MS per charge state and identified peptides per charge state for a single run. If the same enzyme is used for digestion and other instrument parameters are not changed then the ratio of the triggered precursor ions (2+/3+, 2+/4+, 3+/4+, ...) should be on the same level (see Figure 5). This plot also allows the identification of possible outliers. In this example the same sample is analyzed in triplicates with different gradient lengths. The number of identified proteins is for the 90-min and 45-min runs at the same level but for the 75-min and 60-min runs one outlier per triplicate can be spotted. For 75-min run 2 and for 60-min run 3 shows deviation from the normal distribution for the precursor ion charge states as well as in the number of identified proteins.

FIGURE 4. Number of precursor ions triggered for MS/MS and identified peptides per charge state.

Rawputator – A New Tool to Combine Proteomic Data Mining PP443with Method Development, Result Validation and Quality Control

Martin Zeller, Bernard Delanghe, Christoph Henrich, Torsten Ueckert

Thermo Fisher Scientific, Bremen, Germany

Sample preparation Mass spectrometry Data analysis

Proteome

Digest

Ionization (ESI)

Full MS

MS/MS

QuantitationIdentification

Extraction of the dynamic parameter:

- Precursor S/N

- TIC

- Ion inject time

- Elapsed scan time

- Effective collision energy [eV]

- Lockmass correction

Additional „costs“:

- ~ 6% .msf file size increase

- < 3 min per search job (complex cell

lysate sample, 13665 MS/MS)

date dataset proteins11/6/10 0:22 90 min 1 67611/6/10 2:33 90 min 2 67211/6/10 4:43 90 min 3 67511/6/10 6:44 75 min 1 63411/6/10 8:44 75 min 2 599

11/6/10 10:45 75 min 3 64411/6/10 12:30 60 min 1 60111/6/10 14:16 60 min 2 60111/6/10 16:02 60 min 3 63811/6/10 17:32 45 min 1 55311/6/10 19:03 45 min 2 55211/6/10 20:34 45 min 3 550

0

2000

4000

6000

8000

10000

12000

2+ 3+ 4+ 5+ 6+

charge state

Precursor ions Identified peptides

date dataset proteins11/6/10 0:22 90 min 1 67611/6/10 2:33 90 min 2 67211/6/10 4:43 90 min 3 67511/6/10 6:44 75 min 1 63411/6/10 8:44 75 min 2 599

11/6/10 10:45 75 min 3 64411/6/10 12:30 60 min 1 60111/6/10 14:16 60 min 2 60111/6/10 16:02 60 min 3 63811/6/10 17:32 45 min 1 55311/6/10 19:03 45 min 2 55211/6/10 20:34 45 min 3 550

NO: Restore system performance

QC sample real samplesQC analysis QC ok? YES

QC sample real samples QC analysis QC ok?

NO: Pause sequence, restore system performance and reinject samples

YES: Continue with sample injections