Upload
hichem-tagougui
View
30
Download
0
Embed Size (px)
DESCRIPTION
Manual for FID Chromatograph
Citation preview
Operation & User Manual | DHI Technical Support Team
DHI SERVICES FID CHROMATOGRAPH PANEL
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
1
Contents
1. Introduction 02
2. Operation 09
3. Calibrations 25
4. Maintenance 35
5. Troubleshooting 37
6. Technical Specifications 41
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
2
1. Introduction
1.1 The FID Chromatograph is a compact analyzer that always delivers the information
you need. It provides optimal chromatography including peak symmetry, retention time
repeatability, and retention index accuracy, while allowing for fast GC (Cycle time for C1-C5
of 90sec). When the composition of gas mixtures is critical, it generates more data in less
time for faster and better business decisions.
Fig. 1-1 The FID Chromatograph Panel is a 19” rack mounting at 2U height
1.2 The FID Chromatograph uses the Flame Ionization Detection (FID) system. FIDs are
sensitive primarily to hydrocarbons, and are more sensitive to them than Thermal
Conductivity Detector (TCD); Fig. 1-2.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
3
Fig. 1-2 Flame Ionization Detection (FID)
The detection of organic compounds is most effectively done with flame ionization. FIDs are
best for detecting hydrocarbons, and other easily flammable components. They are very
sensitive to these components, and response tends to be linear across a wide range of
concentrations.
1.3 The FID Chromatograph is able to increase the monitoring frequency by using the
extremely valuable backflushing technique. Backflushing can improve the quality of the
analysis and saves both time and money on every run. By reversing column flow
immediately after the last compound of interest has eluded, we can eliminate long bake-
out times for highly retained sample components. Instead, these materials are swept
backwards through the column and out the split vent, preventing carryover, contamination,
retention time shifts and baseline drift.
1.4 DHI Services uses Clarity Lite chromatography station for the acquisition, processing
and evaluation of data from any gas with analog output. It is effective and user-friendly as
minimal data entry is required to launch methods. Immediate results can be viewed from a
laptop or desktop computer once the Chromatograph is run. Screens control is very simple
and standard or customized reports are easily managed by the operators.
Air or Oxygen for
combustion
Thermo coupler &
Filament Coil Combined Hydrogen &
Capillary Column
Thermo Coupler
Filament Coil
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
4
Fig. 1-3 U-PAD 2 A/D Converter
The U-PAD 2 is the hardware of the Clarity Lite chromatography station. This device has the
following features:
1.4.1 2 data acquisition channels jointly isolated from USB ground.
1.4.2 2 digital inputs.
1.4.3 2 digital outputs also designed as relay contacts.
1.4.4 USB interface.
1.4.5 Excellent reliability and temperature stability.
1.5 The FID Chromatograph Panel is powered by 110 Vac, 3 amps, 50 – 60 Hz. It requires
a power supply module that supplies +/-12 Vdc and +5 Vdc to the Electrometer Board and
the Function Board respectively.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
5
Fig. 1-4 Internal Layout of FID Chromatograph Panel
1.6 Parts descriptions of the FID Chromatograph Panel (Fig. 1-4) are as shown:
1.6.1 Temp Display: Temperature of the oven shown in degree C.
1.6.2 Temp A/B Selector: Switch Selector to ‘A’ to read the actual temperature of the oven; Switch Selector to ‘B’ to read the set temperature of the oven.
Analog Signal
Output
INT Signal
Output
110 Vac Power
Solid State Relays
Power Supply
Module
Function Board
FID Ignition
Switch & LED
3-Way
Solenoid
Valve
SP Inlet
Vent
Sample Inlet
H2 Inlet
Air Inlet
FID Chimney
Out
Oven with R1 – R5
Flow control
FID
Electrometer Board
Temp Display
Temp A/B
Selector
On/Off Switch
& LED
Syringe Inject
Port
Sampling
Switch
Zero Knob
Range Selector
Switch
Signal Display
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
6
1.6.3 Syringe Injection Port: The syringe port is where gas is alternatively
inputted to the chromatograph. This is normally used during the testing
phase of the well.
1.6.4 Sampling Switch: Switch to ‘Sample’ to get either a single
chromatography sample or switch to ‘Auto’ to set the chromatograph so
that it automatically takes in sample as set by the cycle time.
1.6.5 Zeroing Knob: The zero knob is used to set the baseline before sampling,
by passing through a zero-hydrocarbon-content medium such as the
compressor air.
1.6.6 Range Selector Switch: The range selector is for selecting the
amplification of the detector signal. The range is to be set in respect to
the Total Gas Reading.
At range X10: Ideal for THA reading at 0 to 2%
At range X100: Ideal for THA reading at 2% to 20%
At range X1K: Ideal for THA reading at 20% to 100%
1.6.7 Signal Display: The signal display indicates the 0-1 volt representation of
the peak detection of each type of gas being detected. It does not give
the exact value of the gas. It is useful though in letting the user identify
that a peak has been reached and detected, and that it gives a clue of
the resulting saturation for a particular range selection. This is tackled in
more detail in calibration of signal.
1.6.8 FID Ignition Switch: The ignite switch is a toggle switch. When it is
flicked and held up for approximately 2 to 4 seconds, 5 Vdc is applied to
the ignition coil on the detector, igniting the flame. The red led above
the switch indicates status of the flame.
1.6.9 FID Flame Indicator: The LED will light up when the FID is ignited.
1.6.10 FID Chimney Out: The chimney out is the exhaust for the burnt out gases.
1.6.11 Air Inlet: Serves as carrier of gas sample, and as combustion air. Setting:
30psi.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
7
1.6.12 H2 Inlet: H2 is serves as a “fuel” to enable the detector to sense
hydrocarbon ions. Setting: 80psi.
1.6.13 Sample Inlet: Sample is the line where gas sample is passed through for
chromatography. Setting: 15psi.
1.6.14 Vent: Vent is where unused sample, especially during back flush state, is
passed through. This should eventually be plumbed to outside the unit.
1.6.15 SP Inlet: SP (Solenoid Pressure) Similar to air, it is also derived from the
unit compressor except that it is maintained at a different pressure. It is
used to switch the commutation valve between its two positions. Setting:
60psi.
1.6.16 Analog Signal Out: This is a 0-1 Vdc signal, which serves as input to the
integrator. 1 volt denotes the maximum gas peak, depending on the
range selected. This is a two-pin terminal (+) and (-). This signal line is
terminating to the integrator (Clarity Lite).
1.6.17 INT Signal Out: This is a trigger to start the integrator whenever the
ample-auto is flicked to sample, or whenever a new cycle starts when
the chromatograph is in auto mode. Note that this two pin terminal is
bipolar – there are no polarities (+ or -) that need to be followed.
1.6.18 Function Board: Aside from controlling the oven temperature, it
also controls timing like Sample Injection State Interval and Cycle Time.
1.6.19 Power Supply: This is a PCB and a transformer mounted on an
aluminium plate. The power supply module supplies +/- 12vdc and + 5
Vdc. The +/-12vdc is required for offset setting of the integrated circuit
chips on the electrometer and temperature control board. The +5V
supply is required for ignition of fid and power for front panel displays.
(Power supply voltages check) lets you identify the power supply board
and adjust the voltages to correct levels.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
8
1.6.20 Solid State Relay (SSR): The SSR is to trigger the Heater in the Oven and
the Solenoid Valve.
1.6.21 Solenoid Valve: The Solenoid Valve is trigger when the Sampling Switch
is toggled. This will activate the 10 port valve.
1.6.22 Oven: The Oven is where the gas is passed through columns to separate
it into its components c1 (methane) up to nc5. To maintain the
consistency of the operation at a certain temperature of the oven, the
insulation material must be present and the cover closed tightly.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
9
2. Operation
2.1 In the mud-logging unit, the FID Chromatograph Panel is installed on the equipment
rack. Connections to the FID Chromatograph Panel are as follows:
2.1.1 110Vac Regulated power supply.
2.1.2 +ve, -ve and Shield Signal Output from the FID Chromatograph
connected to DET1 of the INT7 connector shown in Fig. 2-1.
2.1.3 +ve, -ve INT Output from the FID Chromatograph connected to INT1 of
the INT7 connector shown in Fig. 2-1.
Fig. 2-1 INT7 Connector
2.1.4 AIR Inlet (Green 1/8” tubing labelled ‘AIR CHR1’ or ‘AIR CHR2’ depending
on which FID Chromatograph Panel you are connecting to, CHR1=FID
Chromatograph Online, CHR2=FID Chromatograph Offline.) connected to
Manifold Panel via the Pneumatic Panel.
2.1.5 H2 Inlet (Red 1/8” tubing labelled ‘H2 CHR1’ or ‘H2 CHR2’ depending on
which FID Chromatograph Panel you are connecting to, CHR1=FID
Chromatograph Online, CHR2=FID Chromatograph Offline.) connected to
Manifold Panel via the Pneumatic Panel.
2.1.6 SMPL Inlet (Blue 1/8” tubing labelled ‘Sample CHR1’ or ‘Sample CHR2’
depending on which FID Chromatograph Panel you are connecting to,
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
10
CHR1=FID Chromatograph Online, CHR2=FID Chromatograph Offline.)
connected to Manifold Panel via the Pneumatic Panel.
2.1.7 VENT Outlet (Black 1/8” tubing labelled ‘Vent CHR1’ or ‘Vent CHR2’
depending on which FID Chromatograph Panel you are connecting to,
CHR1=FID Chromatograph Online, CHR2=FID Chromatograph Offline.)
connected to Manifold Panel via the Pneumatic Panel.
2.1.8 SP Inlet (Clear 1/8” tubing labelled ‘SP CHR1’ or ‘SP CHR2’ depending on
which FID Chromatograph Panel you are connecting to, CHR1=FID
Chromatograph Online, CHR2=FID Chromatograph Offline.) connected to
Manifold Panel via the Pneumatic Panel.
Fig. 2-2 FID Chromatograph Panel - Back view
2.2 Switching on the FID Chromatograph Panel will lit-up the Red LED above the power
switch, TEMP and SIGNAL Display simultaneously. As all equipments in the mud-logging unit
are tested, therefore it is unlikely that one should find difficulty in powering up the FID
Chromatograph Panel.
2.3 Upon switching on the FID Chromatograph Panel, switch the TEMP switch on the
front of the FID Chromatograph to “B” and adjust VR2 of the Function PCB (Refer to Fig. 2-3)
until the front display is showing 62.2 immediately. (This will make the oven temperature
AIR Inlet Vent Outlet Signal Output 110Vac
Regulated Power
Signal Shield
Sample Inlet H2 Inlet SP Inlet INT Output
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
11
stabilized at 60.0 degree Celsius.) This will also prevent the oven from burning out because
of the unknown VR2 of the Function PCB setting.
Fig. 2-3 FID Chromatograph Panel Function PCB
2.4 Allow the FID Chromatograph Panel approximately 2 hours to warm-up. After which,
refer to the Chapter 3. Calibration and carry out the FID Chromatograph Panel Electrometer
PCB Zeroing Calibration accordingly.
2.5 Next we have to set the Jumpers Settings of the FID Chromatograph Panel Function PCB. There are 5 Banks of 10 jumpers’ settings at the top part of the Function board shown in Fig 2-4.
Function PCB VR2
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
12
Fig. 2-4 FID Chromatograph Panel Function PCB
2.5.1 The First bank of 10 jumpers sets the integrator start pulse delay from 0
to 9 tenths of a second. Set this to 3. (Put the jumper 4th pin from the left).
2.5.2 The Second and Third banks of 10 jumpers each set the Forward Flow
Time. Second bank to 5 (Put the jumper 6th pin from the left), which is about 5
seconds.
Set the Third bank to 10 (Put the jumper 2nd pin from the left), which is
about 10 seconds.
That will make it a total of 15 seconds. But with the columns we are using, this usually gets adjusted down to about 12 seconds but has to be determined for each new column.
2.5.3 The Fourth and Fifth banks of 10 jumpers each set the Auto Recycle Time.
Set the Fourth bank to 40 (Put the jumper 5th pin from the left), which is
about 40 seconds.
Set the Fifth bank to 200 (Put the jumper 3rd pin from the left), which is
about 200 seconds.
That will result in an Auto Recycle Time of 240 seconds in total.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
13
2.6 Set the regulator after the Air Compressor Output to 60PSI, this is also your SP for the
FID Chromatograph. And ensure that you have the following settings on your Pneumatic
Panel. Air=30 PSI (Regulated by the Air Regulator Knob), Zero=15 PSI (Regulated by the
Zero Regulator Knob) and Carrier=80 PSI (Regulated from the Hydrogen Generator).
Fig. 2-5 Pneumatic Control Panel – Front view
2.7 Set the Oven flow as per the instructions on the top oven plate (Fig. 2-6)
ZERO @
15 PSI
AIR @
30 PSI
Carrier @
80 PSI
ZERO Regulator
Knob
AIR Regulator
Knob
Select ZERO
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
14
Fig. 2-6 Flow Setting Instructions on the Oven top plate
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
15
2.8 After you had ignited (Hearing the faint ‘pop’ sound) the FID Chromatograph Panel,
the Ignite LED should light up. Now you need to warm up the FID Chromatograph for 24
hours before it is ready for testing.
2.9 Ensure that you have the Clarity Lite Software installed in the DrillSense Master
Computer. You must also ensure that the Clarity Lite USB Dongle and U-PAD2 Device are
already connected to the USB ports of the DrillSense Master Computer. Double-click on the
Clarity Lite Shortcut to launch the program.
2.10 Enter ‘Administrator’ and click “OK”.
2.11 Go to “File” -> “Project...”.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
16
2.12 Click on “New”…
2.13 Create a new project “DrillSense” and click “OK”.
2.14 Put the unit number of the respective unit under “Description” and click “OK”.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
17
2.15 Go to “Analysis” -> “Single”.
2.16 Key in the details as shown.
You’ll need to point the filename to the directory you want to use, and make sure you have
the %R parameter set to use the date format. Then click “OK”.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
18
2.17 Go to “Method” -> “Measurement”.
2.18 Fill in these parameters as shown. Any parameters you fill from here may need to be
tweaked for your chromatograph. Next, click on the “Acquisition” tab below.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
19
2.19 Select the Range and Sample Rate as shown. Next, click on the “Integration” tab.
2.20 Fill in the Global Peak Width=0.020 min and Global Threshold=1.000mV as shown.
Next, click on the “Calculation” tab.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
20
2.21 Fill in the details as shown and click “ok”. This is where you will create your first
calibration file. You won’t calibrate anything yet, you’re just setting the reference
calibration file in the directory.
2.22 At the “Clarity Lite – Method” window -> Setting -> Export Data...
CHR1x10CAL
1
PPM
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
21
2.23 Only Select the details as shown and click “OK”. Do not put in any additional
information as it may cause the DrillSense unable to process the created text file.
2.24 At the “Clarity Lite – Method” window -> Setting -> Postrun...
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
22
2.25 Only Select “Open Chromatograph Window” and “Export Data”. Then click “OK”. Do
not put in any additional information.
2.26 To complete the Installation and Setup of the Clarity Lite Program, perform the
following instructions.
At the “Clarity Lite – Method” window -> File -> Save Method As... and save these settings
you had set up to a Method name of your choice (i.e. CHR1x10MET). Take note of where
the file was saved to, so that you will be able to retrieve it again when required.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
23
2.27 Flow C1 – C5 0.1% Mixture to SPAN, On the Pneumatic Panel, switch from ZERO to
SPAN and wait for 1 minute. Trigger the Sample Switch up on the FID Chromatograph Panel.
After it completed its run, you should get a Clarity Lite – Chromatograph pop up window
shown. Click on the Clarity Lite – Calibration Window. (Indicated by the black arrow)
2.28 Go to “Open Standard”, search under “Data” folder for the results of the sample you
had just ran and click “OK”. It should be the latest file created.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
24
2.29 After opening the data file, click on “Add All” indicated by the black arrow. This will add all the 7 peaks of C1 to C5 to the calibration table. Rename the “Compound Name” and key in Analytical values accordingly. The names of the peaks MUST be key in as follows else DrillSense will not be able to “recognize” it. C1 (Methane): methane or C1 or c1 C2 (Ethane): ethane or C2 or c2 C3 (Propane): propane or C3 or c3 iC4 (iso-Butane): i-butane or IC4 or ic4 nC4 (n-Butane): butane or n-butane, NC4 or nc4 (Note: n-butane only works in DrillSense version November 30, 2009 and after) iC5 (iso-Pentane): i-pentane or IC5 or ic5 nC5 (n-Pentane): pentane or n-pentane or NC5 or nc5 (Note: n-pentane only works in DrillSense version November 30, 2009 and after)
Go to File -> Save Calibration As.. Save the calibration as the one you put in Step 2.21. (I.e.
CHR1x10CAL)
2.30 Now go back to the Clarity Lite – Method Window -> File -> Save the Current Method
(i.e. CHR1x10MET).
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
25
3. Calibrations
3.1 After you had obtained your first Clarity Lite-Calibration file, the FID Chromatograph
is ready for calibration to its unique settings. (Note: Each and every FID Chromatograph is
unique equipment as it depends on the columns and FID which may have different
properties even if they are manufactured with the same specifications.)
3.2 FID Chromatograph Panel Electrometer PCB Zeroing Calibration will help you to
calibrate the FID Chromatograph Panel Electrometer PCB. You may have done this under
the operation procedure, but there will be some variation after the FID Chromatograph had
been fully warmed up. Thus you will need to do the FID Chromatograph Panel Electrometer
PCB Zeroing Calibration again.
3.3 FID Chromatograph Panel Electrometer PCB Zeroing Calibration procedure as
follows:-
3.3.1 Unplug the FID Detector connector from the Electrometer Board. (See Fig. 3-1)
3.3.2 Unplug the Zero Potentiometer connector from the Electrometer Board.
(See Fig. 3-1)
Fig. 3-1 FID Chromatograph Electrometer PCB – Top view
FID
Detector
connect
or
Zero Potentiometer connector
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
26
3.3.3 At the front of the FID Chromatograph Panel, place the Range Switch knob at 10.
Fig. 3-2 FID Chromatograph Panel – Front view
3.3.4 Adjust VR2 on the FID Chromatograph Panel Electrometer PCB until the Signal
display on the front of the FID Chromatograph Panel reads 0.0.
Fig. 3-3 FID Chromatograph Panel Electrometer PCB – Top view
3.3.5 Plug the FID Detector and Zero Potentiometer connector back onto the FID
Chromatograph Panel Electrometer PCB. 3.3.6 Adjust the Zero knob at the front of the FID Chromatograph until the Signal
display reads 0.0. If you are unable to adjust till the Signal display reads 0.0, secure back the top cover of the FID Chromatograph Panel, and let it warm up overnight and try again.
VR2 on FID Chromatograph Panel Electrometer PCB
Zero knob at the front of the Chromatograph Panel
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
27
3.4 By now the FID Chromatograph Panel should be ignited and warmed up for at least
24 hours. Now we need to do the Calibration of the FID Chromatograph Panel with
Integrator (Clarity Lite)
3.5 Set the Range = 10, re-adjust the Zero Knob on the front panel to show 0.0 on the
display. It should be 0.0 when the Pneumatic selector switch is at Zero position.
3.6 Flow C1 to C5 0.1% to SPAN. Switch to SPAN on the Pneumatic Panel, wait for 1
minute, and trigger the Sample / Auto Switch upwards to take a sample. Switch SPAN to
ZERO once the FID Chromatograph 10-port valve had switched back. Basically, when the 10-
port valve switched back, the FID Chromatograph had completed its sampling, and we
switch back to ZERO to save the Cal Gas.Having the setting up done in Chapter 2, you
should be able to see the C1 to C5 7 peaks on the Clarity Lite-Chromatograph Window
when the Clarity Lite program completed its run.
3.7 After the Clarity Lite Program completed its run, you should get a 7-peak graph with
reference to the Calibration File (i.e. CHRx10CAL) created in Chapter 2. Ensure that the 7
peaks’ Retention Time (RT) is captured correctly on the Clarity Lite Program, else recalibrate
the present run as your calibration file and run again. You must ensure that the Retention
Time must coincide with the ones you had calibrated.
3.8 Ensure that the 7 peaks appear in the range of the 1.5 minutes (90 Seconds) run time
set. However this Run Time can be determined by how fast or how slow they appear. The
Default Run Time setting was 1.4 minutes.
3.9 Apply the C1 to C5 0.1% mixture a few more times to determine the Run Time.
Determine the Run Time. The run time should be round to the nearest 0.1 minutes after the
last peak (nC5) had fully appeared. At the same time, observe the signals picked up by the
CHR by referring the current run Area amount with the previous run Area amount. The Area
amount should be around the same.
3.10 Change the Run Time in Clarity Lite Program in the Clarity Lite – Method window
under Method / Measurement; so that it will take effect next time you Sample Trigger the
FID Chromatograph Panel. Note down the RUN TIME.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
28
3.11 In the Clarity Lite – Chromatograph window where you can see the Results of that
particular run. You can try to vary the Global Threshold here under the Integration tab to
determine the value of the Global Threshold. Set the Global Threshold at a level where all
the 7 peaks are detected with minimum or no noise peaks picked up. (Default: 1.000 mV)
You can change the Global Threshold in Clarity Lite – Method window under Method /
Integration so that it will take effect next time you Sample Trigger the FID Chromatograph
Panel. Note down the Global Threshold:
3.12 Now you got to set the Forward Flow Time (sampling time). This is to prevent
unwanted heavy hydrocarbons from being introduced into the main separation column.
The default Forward Flow Time set is 15 sec. (Set by the Second and Third Bank of Jumpers
settings.)
Bring the Forward Flow Time down to 14 sec by setting the Second bank of 10 jumpers to 4
(5th pin from the left) and sample C1 to C5 0.1%.
Slowly reduce the Forward Flow Time until you see a significant drop (10% or more) in C5.
When the amount of C5 is significantly reduced, add 1 or 2 seconds back to the current
forward flow time setting. (As an example, if the last forward flow time is 11 seconds and
the C5 amount is reduced by more than 50%, set the configuration jumpers to 13 seconds.
If the last forward flow time is 11 seconds and the C5 amount is reduced by less than 50%,
set the configuration jumpers to 12 seconds.) Note down the Final Forward Flow Time.
3.13 Save the current Method again as per the file name you had saved earlier. (Note: The
General Calibration for the FID Chromatograph Panel had been completed.)
With the Clarity Lite Program calibrated at RANGE 10, apply C1 to C5 0.1% again and check
the accuracy of the resulting readings. Take note of the proximity of the readings to the
nominal values. Average reading accuracy from nominal values (e.g., 50 ppm):
If the accuracy is greater than 50 ppm, the Clarity Lite Calibration file has to be
recalibrated again else this calibration file reference is ok.
3.14 Now proceed to the Calibration of the High Level Gas. With the Current Method still
open, save it as High Level Gas file (eg. CHR1x1000MET). This is to prevent unnecessary
overwriting of the data previously saved.
(Disregard the Calibration File tentatively.)
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
29
3.15 This is to calibrate the High Gas at Range 1k. At RANGE 1k, apply C1 to C5 0.1% and
Sample Trigger the FID Chromatograph Panel. This is to create the C1 to C5 readings at
RANGE 1k for the High Gas Calibration File in Clarity Lite.
After the run completed, at the Clarity Lite – Chromatograph window, you will notice that
some of the C1 to C5 Compounds were not shown. It is very obvious that there is a peak
there, but Clarity Lite is not recognising them as peaks. At the Clarity Lite – Chromatograph
window, click on the “Integration” tab below and reduce the Global Threshold Voltage by
steps of 0.1mV. Reduce until you are able to see all the 7 peaks. Take note of the mV
reading (i.e. 0.5mV) all the 7 peaks appear.
3.16 Go back to the Clarity Lite – Chromatograph window, Results Tab, and click to save
the Chromatograph. After which, go and change the Global Threshold to the value you had
obtained in step 3.15, by going to the Clarity Lite – Method window under Method /
Integration and saved to this High Gas Method (i.e.CHRx1000MET), so that it will take effect
next time you Sample Trigger the FID Chromatograph Panel.
3.17 Now go to the Clarity Lite – Calibration window, select a “New” calibration, then click
on “Open Standard”, and select the C1 to C5 0.1% at RANGE 1K chromatograph you had
just created.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
30
3.18 Click on “Add all” and key in the compound names and Analytical Values accordingly.
The names of the peaks MUST be key in as follows else DrillSense will not be able to “recognize” it. C1 (Methane): methane or C1 or c1 C2 (Ethane): ethane or C2 or c2 C3 (Propane): propane or C3 or c3 iC4 (iso-Butane): i-butane or IC4 or ic4 nC4 (n-Butane): butane or n-butane, NC4 or nc4 (Note: n-butane only works in DrillSense version November 30, 2009 and after) iC5 (iso-Pentane): i-pentane or IC5 or ic5 nC5 (n-Pentane): pentane or n-pentane or NC5 or nc5 (Note: n-pentane only works in DrillSense version November 30, 2009 and after)
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
31
3.19 You must take note that 0.1% at RANGE 10 = 1000ppm. Therefore 0.1% at RANGE 1k
= 10ppm. But if you want DrillSense to recognise the small reading at RANGE 1k = 1000ppm,
you may key in the desired value. (i.e. At RANGE 1k, 0.1% = 1000ppm.) But you MUST
remember to change to the appropriate Method (i.e. CHRx10MET for RANGE 10 and
CHRx1000MET for RANGE 1k.) for the RANGE you had selected on the front of the FID
Chromatograph Panel. Save this calibration file as a file name you know which FID
Chromatograph Panel you are referring to. (i.e. CHR1x1000CAL)
3.20 After you had saved the calibration file, go back to the Clarity Lite – Chromatograph
window, click on “Set…” and select the calibration file (i.e. CHR1x1000CAL) in you had
entered in step 3.19. You should see the ppm results changed and correspond with the
Analytical Values you entered.
3.21 After you had set the calibration file, go to File and save the chromatograph with the
new reference calibration file you had just set.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
32
3.22 Then go to the Clarity Lite – Method window. Go to “Method” -> “Calculation”, , click
on “Set…” and select the calibration file (i.e. CHR1x1000CAL) in you had entered in step
3.19, and close. This is to tell the integrator to use this calibration file as reference for
future analysis.
3.23 Under the Clarity Lite – Method window, go to “File” and save this method to
“CHR1x1000MET”.
3.24 Flow the C1 50% to SPAN and switch to SPAN on the Pneumatic Panel. Adjust
SPAN=15PSI. Wait for 1 minute, ensure that your FID Chromatograph Panel is at RANGE 1K,
and sample trigger it.
3.25 After the Clarity Lite program completed its run, you should see a C1 50% peak on
the Clarity Lite – Chromatograph window. Click on the Clarity Lite –Calibration window,
open the calibration file you had saved in 3.19 (i.e. CHR1x1000CAL), disregard if the
calibration file is correct and already opened.
3.26 Click on “Open Standard”, and open the C1 50% chromatograph you had just created.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
33
3.27 Click on “Add All”, you should see that the calibration file created a 2nd level of
Calibration of the C1 50% result. Just enter the Analytical values of the C1 50% in the
Amount column, the rest just leave it as 0.00 and save the calibration file.
3.28 Click back to the Clarity Lite – Chromatograph window, you should see your C1 50%
amount showing the amount you had keyed in the Clarity Lite –Calibration window.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
34
3.29 Calibration of the FID Chromatograph is fully completed; you may try a few runs of
the C1 to C5 0.1% calibration gas or other C1 to C5 gas compound to see if the results are
close to the analytical values. You may also create a CHRx100MET method file and a
CHRx100CAL calibration file for RANGE 100.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
35
4. Maintenance
4.1 The FID Chromatograph Panel is easy to maintain. However due to the environment
of mud-logging activities, it requires continuous team effort to upkeep the cleanliness in the
mud-logging unit.
4.2 Periodic checks on the FID Chromatograph Panel will definitely allow the equipment
to achieve its best performance. It is important to follow the Unit Equipment Maintenance
Schedule to perform checks on the FID Chromatograph Panel.
4.3 These are some simple steps to perform in maintaining the FID Chromatograph Panel:
4.3.1 Switch off the FID Chromatograph Panel and remove the power cord
from the panel.
4.3.2 Remove 4 x securing rack screws of the FID Chromatograph panel from
the equipment rack.
4.3.3 Gently slide the FID Chromatograph Panel from the equipment rack.
Caution if there is any entanglement from the gas tubing or signal cables
at the back of the FID Chromatograph Panel.
4.3.4 There is no requirement to remove the FID Chromatograph Panel
completely from the equipment rack. Once the panel is slide out,
remove the top screws and remove the top panel cover.
4.3.5 Check for any loose wires within the FID Chromatograph Panel. Gently
dust off any dirt that may have settled inside the panel, do not attempt
to use wet-base treatment for cleaning. The use of wet cloth or tissues is
strictly prohibited.
4.3.6 Refrain from contacting the Electrometer and Function boards, as both
have already been calibrated and functioning, it is not advisable to have
any unnecessary adjustment to the PCB board settings.
4.3.7 Simply cover the panel top cover and secure the panel screws when
done. Fasten the rack screws to secure the FID Chromatograph Panel
onto the equipment rack. Insert the power cord and switch on the panel.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
36
4.3.8 Perform a quick check on the panel; ensure the FID Chromatograph
Panel is in operational mode after the maintenance. Verify the
calibration values.
4.3.9 The recommended maintenance mentioned above is based on weekly
checks. Leak-check, re-calibration for Electrometer and Function boards
will have to be carried out monthly.
4.3.10 The Offline FID Chromatograph Panel maintenance will be as per the
Online FID Chromatograph Panel.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
37
5. Troubleshooting
5.1 Q: Unable to switch on the FID Chromatograph Panel?
A: Ensure there is 110Vac regulated power.
A: Check that the power cord is serviceable.
A: Check that the fuse in the panel power entry is serviceable.
A: Check that the power supply module in the FID Chromatograph Panel is
serviceable; by measuring the power in (110Vac) / out (5Vdc & 12Vdc).
5.2 Q: FID Chromatograph Panel signal display or temperature display blanked off?
A: Check for loose connector on the display chicklet board.
A: Check that there are no loose wires to the connector.
5.3 Q: Signal display produces unstable readings?
A: Check if the range selection is set correctly.
A: Check that the Signal Shield is properly fastened; this is to eliminate signal noise.
Has the signal stabilised after checking the shield contact? Otherwise continue with
troubleshooting.
A: Check that the mud-logging unit is free from any form of gaseous in the air that
could contribute to the unstable reading.
A: Check that the mud-logging unit is operating at a stable temperature (21 – 25
degree C). This will help to stabilise the FID readings.
A: Check that the U-PAD2 device is connected to a stable power supply source (i.e
the DrillSense PC is powered by stable power source such as a regulated UPS power
source). You may remove the Signal Outputs and observe whether the reading
stabilises, if it does stabilises then it shows that the Chromatograph signal is affected
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
38
by unstable power source that is generating noise to the signals. Please switch the
power source to provide regulated power to both the PC and the FID Chromatograph
Panel.
A: It could be that the FID has been contaminated and becomes faulty, switch to a
new FID and check again.
5.4 Q: Why the Temp display, Signal display and the power LED blanked off when the
Ignite switch is activated?
A: It is either the FID pins or filament has contact with the chimney (internal), as such
this will create short-circuit to the FID Chromatograph Panel.
Switch off the panel; simply reposition the FID in the chimney by sliding it slightly
away from any possible contact within the chimney.
Take note that the short-circuitry does not damage any of the components, thus
power up the panel and prepare to ignite the FID.
5.5 Q: Unable to ignite the FID Chromatograph Panel?
A: Ensure that the FID filament glows when the ignite switch is activated. You may go
to the back of the panel and in a safe distance of ½ meter away from the chimney
exhaust, observe that the filament glows (inside the chimney) when a helper is
activating the ignite switch.
A: If the filament is unable to glow, please proceed to check the 5Vdc supply at the
power supply module in the FID Chromatograph Panel. Ensure that the 5Vdc is
supplying to the ignite switch. You may also use the multimeter to measure the
filament pins for 5Vdc as you activate the ignite switch (Fig. 5-1). Check for possible
loose soldering joints.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
39
Fig. 5-1 Measuring for 5Vdc on Filament Pins
A: Measure the Conductivity across the 2 points below, it shown be a very low ohms (Less than 10 ohm). Reading that is greater than 10 ohm is most likely that the igniting coil is broken or no good. A: When you hold up the ignite switch, measure the voltage across the 2 points of the FID Connector as shown in Fig 5-2. It should be around -5V or +5V. If not, the ignite switch or 5V from the power supply may be faulty or no good.
Fig. 5-2 Measuring points for 5 volts
Filament contact
pin
Filament contact
pin
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
40
A: With a glowing filament, inject the sample gas and reduce the Air by 50% to
approximately 15 psi. This method is to obtain higher concentration of the sample
gas for combustion; therefore a louder “pop” is expected to be heard. Please
increase the Air to 30 psi after the ignition is achieved.
A: If none of the above mentioned help to solve the problem, you may need to check
the flow settings again, please refer to Chapter 2. Operation; Fig. 2-6 .
A: It could be that the FID has been contaminated and becomes faulty, switch to a
new FID and check again.
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
41
6. Technical Specifications
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
42
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
43
DHI SERVICES | FID Chromatograph Panel Operation & User Manual (v.01)
44