Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
PRODUCT INFORMATION
FECS40-1000 - for the Detection of Carbon Monoxide
Features:
* High sensitivity/selectivity to CO* Quick response to CO* Linear output* Long life * Stable baseline* Unique leak-proof structure
Applications:
* Portable and fixed installation CO monitors* CO detectors* Ventilation control for indoor parking garages
Figaro’s Carbon Monoxide Sensor FECS40-1000 is a unique electrochemical-type carbon monoxide sensor. Its most notable feature is its unique leak-proof structure, making it ideal for CO monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS40-1000 (20˚C) are shown below.
Typical characteristics (temperature dependency) of FECS40-1000 are shown below.
Temperature Dependency:
0
10
20
30
40
50
60
70
80
0 200 400 600 800 1000
Outp
ut
sign
al (μ
A)
Gas Concentration (ppm)
0
20
40
60
80
100
120
140
-20 -10 0 10 20 30 40 50
% S
ensi
tivi
ty (re
fere
nce 2
0
)
Temperature ()
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20.4+0- 0.3
ø13.
5ø18
C
Counter
Top view
Side view
Bottom view
W
Wor
king
R
Reference
All tolerance ± 0.1mm unless otherwise stated.All dimensions in mm.
Detection Gas Carbon Monoxide
Detection Range 0 ~ 1000 ppm
Maximum Overload 1500 ppm
Output Signal 70 ± 15 nA/ppm (*)
Repeatability ±2% (*)
Resolution 1 ppm (*)
Typical Baseline Range (Pure air) -2 ppm to +3 ppm(*)
Typical Response Time (t90) < 30 sec (*)
Baseline Shift (-20 ~ 50˚C) < 10 ppm(*)
Long Term Output Drift < 5% /year (*)
Expected Life Time 3 years (*)
Operating Temperature -20 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 10 Ω
Bias Voltage Not required
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life 6 months
Cap Color Red
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Gas Concentration(ppm)
Typical Carbon MonoxideConcentration(ppm)Equivalent
Carbon Monoxide 100 100
Hydrogen 100 < 40
Carbon Dioxide 5,000 0
Sulphur Dioxide 30 0
Hydrogen Sulfide 30 0
Nitric Oxide 30 < 3
Nitrogen Dioxide 30 0
Ammonia 100 0
Ethanol 200 0
Table1 shows the typical response of FECS40-1000 to interference gases.
Table1 Cross Sensitivity of FECS40-1000 (20˚C)
PRODUCT INFORMATION
FECS50-100 - for the Detection of Hydrogen Sulfide
Features:
* High sensitivity/selectivity to H2S* Quick response to H2S* Small influence by CO* Linear output* Long life * Stable baseline* Unique leak-proof structure
Applications:
* Portable and fixed installation H2S monitors* H2S detectors
Figaro’s Hydrogen Sulfide Sensor FECS50-100 is a unique electrochemical-type hydrogen sulfide sensor. Its most notable features are small influence by CO and a unique leak-proof structure. These features make the sensor ideal for H2S monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS50-100 (20˚C) are shown below.
Typical characteristics (temperature dependency) of FECS50-100 are shown below.
Temperature Dependency:
0
10
20
30
40
50
60
70
80
0 20 40 60 80 100
Out
put s
igna
l (μA
)
Gas Concentration (ppm)
0
20
40
60
80
100
120
140
-40 -30 -20 -10 0 10 20 30 40 50
% S
ensi
tivi
ty (re
fere
nce 2
0
)
Temperature ()
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20.4+0- 0.3
ø13.
5ø18
C
Counter
Top view
Side view
Bottom view
W
Wor
king
R
Reference
All tolerance ± 0.1mm unless otherwise stated.All dimensions in mm.
Detection Gas Hydrogen Sulfide
Detection Range 0 ~ 100 ppm
Maximum Overload 500 ppm
Output Signal 700 ± 150 nA/ppm (*)
Repeatability ±2% (*)
Resolution 0.1 ppm (*)
Typical Baseline Range (Pure air) -0.1ppm to +0.4 ppm (*)
Typical Response Time (t90) < 30 sec (*)
Baseline Shift (-20 ~ 40˚C) < 0.5 ppm(*)
Long Term Output Drift < 2% /month (*)
Expected Life Time > 2 years (*)
Operating Temperature -40 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 10Ω
Bias Voltage Not required
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life Less than 6 months
Cap Color Dark Blue
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20 ˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Gas Concentration(ppm)
Typical Hydrogen SulfideConcentration(ppm)Equivalent
Hydrogen Sulfide 10 10
Carbon Monoxide 300 < 2
Carbon Dioxide 5,000 0
Hydrogen 1,000 < 3
Sulphur Dioxide 5 0.5
Nitric Oxide 30 0.3
Nitrogen Dioxide 5 -1
Ammonia 100 0
Ethanol 1,000 < 2
Table1 shows the typical response of FECS50-100 to interference gases.
Table1 Cross Sensitivity of FECS50-100 (20˚C)
PRODUCT INFORMATION
FECS41-250 - for the Detection of Nitric Oxide
Features:
* High sensitivity/selectivity to NO* Quick response to NO* Linear output* Long life * Stable baseline* Unique leak-proof structure
Applications:
* Portable and fixed installation NO monitors* NO detectors* Ventilation control for indoor parking garages
F iga ro ’s N i t r i c Ox ide Sensor FECS41-250 i s a un ique electrochemical-type nitric oxide sensor. Its most notable feature is its unique leak-proof structure, making it ideal for NO monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS41-250 (20˚C) are shown below.
Typical characteristics (temperature dependency) of FECS41-250 are shown below.
Temperature Dependency:
0
20
40
60
80
100
120
140
0 50 100 150 200 250 300
Outp
ut
sign
al (μ
A)
Gas Concentration (ppm)
0
20
40
60
80
100
120
140
-20 -10 0 10 20 30 40 50
% S
ensi
tivi
ty (re
fere
nce 2
0
)
Temperature ()
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20.4+0- 0.3
ø13.
5ø18
C
Counter
Top view
Side view
Bottom view
W
Wor
king
R
Reference
All tolerance ± 0.1mm unless otherwise stated.All dimensions in mm.
Detection Gas Nitric Oxide
Detection Range 0 ~ 300 ppm
Maximum Overload 1000 ppm
Output Signal 400 ± 80 nA/ppm (*)
Repeatability ±2% (*)
Resolution 0.5 ppm (*)
Typical Baseline Range (Pure air) < 3 ppm (*)
Typical Response Time (t90) < 40 seconds (*)
Typical Baseline Shift (-20 ~ 40˚C) < 4 ppm(*)
Long Term Output Drift < 2% /month (*)
Expected Life Time > 2 years (*)
Operating Temperature -20 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 10Ω
Bias Voltage +300 mV
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life Less than 6 months
Cap Color Orange
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Gas Concentration(ppm)
Typical Nitric OxideConcentration(ppm)Equivalent
Nitric Oxide 100 100
Carbon Monoxide 300 0
Carbon Dioxide 5,000 0
Hydrogen 1,000 0
Nitrogen Dioxide 5 < 1
Chlorine 10 0
Hydrogen Sulfide 15 < 4
Sulphur Dioxide 20 0
Ammonia 20 0
Table1 shows the typical response of FECS41-250 to interference gases.
Table1 Cross Sensitivity of FECS41-250 (20˚C)
PRODUCT INFORMATION
FECS42-20 - for the Detection of Nitrogen Dioxide
Features:
* High sensitivity/selectivity to NO2* Quick response to NO2* Linear output* Long life * Stable baseline* Unique leak-proof structure
Figaro’s Nitrogen Dioxide Sensor FECS42-20 is a unique electrochemical-type nitrogen dioxide sensor. Its most notable feature is its unique leak-proof structure, making it ideal for NO2 monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS42-20 (20˚C) are shown below.
Typical characteristics (temperature dependency) of FECS42-20 are shown below.
Temperature Dependency:
0
20
40
60
80
100
120
140
-30 -20 -10 0 10 20 30 40 50
% S
ensi
tivity
(re
fere
nce
20
)
Temperature ()
-15
-10
-5
00 5 10 15 20
Out
put
Sig
nal (
μA
)
Gas Concentration (ppm)
Applications:
* Portable and fixed installation NO2 monitors* NO2 detectors* Ventilation control for indoor parking garages
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20.4+0- 0.3
ø13.
5ø18
C
Counter
Top view
Side view
Bottom view
W
Wor
king
R
Reference
All tolerance ± 0.1mm unless otherwise stated.All dimensions in mm.
Detection Gas Nitrogen Dioxide
Detection Range 0 ~ 20 ppm
Maximum Overload 150 ppm
Output Signal -600 ± 150 nA/ppm (*)
Repeatability ±2% (*)
Resolution 0.1 ppm (*)
Typical Baseline Range (Pure air) < ±0.2 ppm (*)
Typical Response Time (t90) < 25 sec (*)
Baseline Shift (-20 ~ 40˚C) < 0.2 ppm(*)
Long Term Output Drift < 2% /month (*)
Expected Life Time > 3 years (*)
Operating Temperature -20 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 33Ω
Bias Voltage Not required
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life Less than 6 months
Cap Color Black
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Gas Concentration(ppm)
Typical Nitrogen DioxideConcentration(ppm)Equivalent
Nitrogen Dioxide 20 20
Hydrogen 1000 0
Carbon Monoxide 300 0
Carbon Dioxide 5,000 0
Sulphur Dioxide 20 0
Hydrogen Sulfide 10 -0.7 ~ 0.3
Nitric Oxide 30 0
Ammonia 100 0
Ethanol 100 0
Chlorine 1 1
Table1 shows the typical response of FECS42-20 to interference gases.
Table1 Cross Sensitivity of FECS42-20 (20˚C)
PRODUCT INFORMATION
FECS45-10 - for the Detection of Chlorine
Features:
* High sensitivity/selectivity to Cl2* Linear output* Long life * Stable baseline* Unique leak-proof structure
Applications:
* Portable and fixed installation Cl2 monitors* Cl2 detectors
Figaro’s Chlorine Sensor FECS45-10 is a unique electrochemical-type chlorine sensor. Its most notable feature is its unique leak-proof structure, making it ideal for Cl2 monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS45-10 (20˚C) are shown below.
Typical characteristics (temperature dependency) of FECS45-10 are shown below.
Temperature Dependency:
0
20
40
60
80
100
120
140
-20 -10 0 10 20 30 40 50
% S
ensi
tivi
ty (re
fere
nce 2
0
)
Temperature ()-7
-6
-5
-4
-3
-2
-1
00 2 4 6 8 10
Outp
ut
sign
al (μ
A)
Gas Concentration (ppm)
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20.4+0- 0.3
ø13.
5ø18
C
Counter
Top view
Side view
Bottom view
W
Wor
king
R
Reference
All tolerance ± 0.1mm unless otherwise stated.All dimensions in mm.
Detection Gas Chlorine
Detection Range 0 ~ 10 ppm
Maximum Overload 50 ppm
Output Signal -600 ± 150 nA/ppm (*)
Repeatability ±2% (*)
Resolution 0.1 ppm (*)
Typical Baseline Range (Pure air) < ±0.2 ppm (*)
Typical Response Time (t90) < 60 seconds (*)
Baseline Shift (-20 ~ 50˚C) < ±0.5 ppm (*)
Long Term Output Drift < 2% /month (*)
Expected Life Time > 2 years (*)
Operating Temperature -20 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 33Ω
Bias Voltage Not required
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life Less than 6 months
Cap Color Brown
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Gas Concentration(ppm)
Typical ChorineConcentration(ppm)Equivalent
Chlorine 10 10
Carbon Monoxide 300 0
Carbon Dioxide 5,000 0
Hydrogen 1,000 0
Nitrogen Dioxide 10 10
Nitric Oxide 35 < -0.3
Hydrogen Sulfide 15 < -7.5
Sulphur Dioxide 20 0
Ethanol 100 0
Table1 shows the typical response of FECS45-10 to interference gases.
Table1 Cross Sensitivity of FECS45-10 (20˚C)
PRODUCT INFORMATION
FECS43-20 - for the Detection of Sulfur Dioxide
Features:
* High sensitivity/selectivity to SO2* Quick response to SO2* Linear output* Long life * Stable output signal at high-temperature* Unique leak-proof structure
Applications:
* Portable and fixed installation SO2 monitors* SO2 detectors
Figaro’s Sulfur Dioxide Sensor FECS43-20 is a unique electrochemical-type sulfur dioxide sensor. Its most notable features are stable output signal at high-temperature and a unique leak-proof structure. These features make the sensor ideal for SO2 monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS43-20 (25˚C) are shown below.
Typical characteristics (temperature dependency) of FECS43-20 are shown below.
Temperature Dependency:
0
2
4
6
8
10
12
0 5 10 15 20
Outp
ut
sign
al (μ
A)
Gas Concentration (ppm)
0
20
40
60
80
100
120
140
-40 -30 -20 -10 0 10 20 30 40 50 60
% S
ensi
tivi
ty (re
fere
nce 2
0
)
Temperature ()
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20 +0- 0.3
ø13.5
ø18
CCounter
Top View
Side View
Bottom View
W
Working
R Reference
All tolerance ± 0.1mm Unless otherwise stated.All dimensions in mm.
Detection Gas Sulfur Dioxide
Detection Range 0 ~ 20 ppm
Maximum Overload 150 ppm
Output Signal 500 ± 100 nA/ppm (*)
Repeatability ±2% (*)
Resolution 0.1 ppm (*)
Baseline Range (Pure air) -0.2 ppm to +0.5 ppm equivalent (*)
Typical Response Time (t90) < 25 sec (*)
Baseline Shift (-20 ~ 40˚C) < ±0.2 ppm equivalent(*)
Long Term Output Drift < 2% /month (*)
Expected Life Time > 2 years (*)
Operating Temperature -30 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 10Ω
Bias Voltage Not required
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life 6 months
Cap Color Green
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Gas Concentration(ppm)
Typical Sulfur DioxideConcentration(ppm)Equivalent
Sulfur Dioxide 10 10
Carbon Monoxide 300 < 1
Carbon Dioxide 5,000 0
Hydrogen 400 < 1
Hydrogen Sulfide 15 0
Nitric Oxide 50 1
Nitrogen Dioxide 10 -10
Ammonia 20 0
Chlorine 5 -1
Table1 shows the typical response of FECS43-20 to interference gases.
Table1 Cross Sensitivity of FECS43-20 (25˚C)
PRODUCT INFORMATION
FECS44-100 / -200
Features:
* High selectivity to NH3* Small influence by H2S* Excellent durability to NH3 exposure* Linear output* Long life * Stable baseline* Unique leak-proof structure
Applications:
* Portable and fixed installation NH3 monitors* NH3 detectors* Ammonia leak detection in refrigerators
Figaro’s Ammonia Sensor FECS44 is a unique electrochemical-type ammonia sensor. Its most notable features are small influence by H2S, excellent durability to NH3 exposure, and a unique leak-proof structure. These features make the sensor ideal for NH3 monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS44-100 / -200 (20˚C) are shown below.
Typical characteristics (temperature dependency) of FECS44-100 / -200 are shown below.
Temperature Dependency:
- for the Detection of Low Concentration Ammonia
0
2
4
6
8
10
0 50 100 150 200
Outp
ut
sign
al (μ
A)
Gas Concentration (ppm)
FECS44-100FECS44-200
0
20
40
60
80
100
120
140
-20 -10 0 10 20 30 40 50
% S
ensi
tivi
ty (re
fere
nce 2
0
)
Temperature ()
FECS44-100FECS44-200
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20.4+0- 0.3
ø13.
5ø18
C
Counter
Top view
Side view
Bottom view
W
Wor
king
R
Reference
All tolerance ± 0.1mm unless otherwise stated.All dimensions in mm.
Model Number FECS44-100 FECS44-200
Detection Gas Ammonia
Detection Range 0 ~ 100 ppm 0 ~ 200 ppm
Maximum Overload 200 ppm 500 ppm
Output Signal 100 ± 30 nA/ppm (*) 40 ± 12 nA/ppm (*)
Repeatability ±10% (*) ±10% (*)
Resolution 1 ppm (*) 2 ppm (*)
Baseline Range (Pure air) < ±2 ppm (*) < ±4 ppm (*)
Typical Response Time (t90) 60 sec (*) 90 sec (*)
Baseline Shift (-30 ~ 50˚C) < 3 ppm(*) < 6 ppm(*)
Long Term Output Drift < 2% /month (*)
Expected Life Time > 2 years (*)
Operating Temperature -30 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 33Ω
Bias Voltage Not required
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life Less than 6 months
Cap Color Purple
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Model Number FECS44-100 FECS44-200
Gas Concentration(ppm)
Typical AmmoniaConcentration(ppm)
Equivalent
Concentration(ppm)
Typical AmmoniaConcentration(ppm)
Equivalent
Ammonia 100 100 100 100
Hydrogen Sulfide 10 -1.5 ~ 0 10 -1.5 ~ 0
Sulphur Dioxide 10 -3 10 -2
Carbon Dioxide 5,000 0 5,000 0
Carbon Monoxide 300 0 300 0
Hydrogen 1,000 0 1,000 0
Nitrogen Dioxide 20 -1.5 ~ 0 20 0
Nitric Oxide 30 0 30 0
Ethanol 100 0 100 0
Table1 shows the typical response of FECS44-100 / -200 to interference gases.
Table1 Cross Sensitivity of FECS44-100 / -200 (20˚C)
PRODUCT INFORMATION
FECS44-1000 / -5000 - for the Detection of High Concentration Ammonia
Features:
* High selectivity to NH3* Small influence by H2S* Excellent durability to NH3 exposure* Linear output* Long life * Stable baseline* Unique leak-proof structure
Applications:
* Portable and fixed installation NH3 monitors* NH3 detectors* Ammonia leak detection in refrigerators
Figaro’s Ammonia Sensor FECS44 is a unique electrochemical-type ammonia sensor. Its most notable features are small influence by H2S, excellent durability to NH3 exposure, and a unique leak-proof structure. These features make the sensor ideal for NH3 monitors and detectors in various fields.
Sensitivity Characteristics:Typical characteristics (linearity) of FECS44-1000 / -5000 (20˚C) are shown below.
Typical characteristics (temperature dependency) of FECS44-1000 / -5000 are shown below.
Temperature Dependency:
0
5
10
15
20
0 1000 2000 3000 4000 5000
Outp
ut
sign
al (μ
A)
Gas Concentration (ppm)
FECS44-1000FECS44-5000
0
20
40
60
80
100
120
140
-20 -10 0 10 20 30 40 50
% S
ensi
tivi
ty (re
fere
nce 2
0
)
Temperature ()
FECS44-1000FECS44-5000
REV.01/16
Specifications: Dimensions:
ø15
4.3
16.5
0.6
ø1.5
ø20.4+0- 0.3
ø13.
5ø18
C
Counter
Top view
Side view
Bottom view
W
Wor
king
R
Reference
All tolerance ± 0.1mm unless otherwise stated.All dimensions in mm.
Model Number FECS44-1000 FECS44-5000
Detection Gas Ammonia
Detection Range 0 ~ 1000 ppm 0 ~ 5000 ppm
Maximum Overload 2500 ppm -
Output Signal 8 ± 4 nA/ppm (*) 4 ± 2 nA/ppm (*)
Repeatability ±10% (*)
Resolution 10 ppm (*) 20 ppm (*)
Baseline Range (Pure air) < 6 ppm (*) < 10ppm (*)
Typical Response Time (t90) 120 sec (*) 150 sec (*)
Baseline Shift (-30 ~ 50˚C) < 20 ppm(*) < 40 ppm(*)
Long Term Output Drift < 2% /month (*)
Expected Life Time Over 2 years (*)
Operating Temperature -30 ~ 50˚C
Operating Humidity 15 ~ 90% RH
Operating Pressure Range 1013 hPa ±10%
Recommended Load Resistor 33Ω
Bias Voltage Not required
Position Sensitivity None
Recommended Storage Temp. 0 ~ 20˚C
Storage Life Less than 6 months
Cap Color Purple
Weight 4.5g (approx.)
PRODUCT INFORMATION
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.figaro.co.jpemail: [email protected]
(*) Performance data conditions: 20˚C, 50%RH and 1013 hPa.
Cross Sensitivity Data :
Model Number FECS44-1000 FECS44-5000
Gas Concentration(ppm)
Typical AmmoniaConcentration(ppm)
Equivalent
Concentration(ppm)
Typical AmmoniaConcentration(ppm)
Equivalent
Ammonia 100 100 100 100
Hydrogen Sulfide 20 0 20 0
Sulphur Dioxide 20 -8 20 -12
Carbon Dioxide 5,000 0 5,000 0
Carbon Monoxide 300 0 300 0
Hydrogen 1,000 0 1,000 0
Nitrogen Dioxide 20 0 20 0
Nitric Oxide 30 0 30 0
Ethanol 100 0 100 0
Table1 shows the typical response of FECS44-1000 / -5000 to interference gases.
Table1 Cross Sensitivity of FECS44-1000 / -5000 (20˚C)