FDR MS-3500 Service Manual 026-201-05(1)E Spec-11 2. Dimensions and Weight of the Machine External Dimensions and Weight Width Depth Height Weight Exposure stand Approx. 624 mm (24.57 in.) Approx. 1270 mm (50 in.) (Base plate approx. 787 mm (30.98 in.)) Approx. 1974 mm 2285 mm (77.72 in. to 89.96) Approx. 370 kg (815.7 lb.) Control cabinet Approx. 205 mm (8.07 in.) Approx. 503 mm (19.8 in.) Approx. 530 mm (20.87 in.) Approx. 20 kg (44.09 lb) Generator (X-ray high voltage unit) Approx. 445 mm (17.52 in.) Approx. 315 mm (12.4 in.) Approx. 825 mm (32.48 in.) Approx. 70 kg (154.3 lb) Positioner control cabinet Approx. 160 mm (6.3 in.) Approx. 350 mm (13.78 in.) Approx. 473 mm (18.62 in.) Approx. 17 kg (37.48 lb) Hardware specification, dimensions and weight are subject to change without notice. SPEC

FDR MS-3500 05(1)E - Fujifilm Healthcare

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-11

2. Dimensions and Weight of the Machine

External Dimensions and Weight

Width Depth Height Weight

Exposure stand Approx. 624 mm

(24.57 in.)

Approx. 1270 mm (50 in.)

(Base plate approx. 787 mm (30.98 in.))

Approx. 1974 mm 2285 mm

(77.72 in. to 89.96)

Approx. 370 kg

(815.7 lb.)

Control cabinet Approx. 205 mm (8.07 in.)

Approx. 503 mm (19.8 in.)

Approx. 530 mm (20.87 in.)

Approx. 20 kg

(44.09 lb)

Generator (X-ray high voltage unit)

Approx. 445 mm

(17.52 in.)

Approx. 315 mm (12.4 in.)

Approx. 825 mm (32.48 in.)

Approx. 70 kg

(154.3 lb)

Positioner control cabinet Approx. 160 mm (6.3 in.)

Approx. 350 mm (13.78 in.)

Approx. 473 mm(18.62 in.)

Approx. 17 kg

(37.48 lb)

Hardware specification, dimensions and weight are subject to change without notice.

SPEC

Page 2: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-12

2.1 External View

Controller unitPositioner control cabinet

(Accessory sold separately)

Generator (X-ray high voltage unit)

Install the controller unit and generator within 5 m (196.9 in.) of the exposure stand. Install the positioner control cabinet within 3 m (118.1 in.) of the exposure stand.

Exposure stand

DRMS1001_0_E.ai

SPEC

Page 3: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-13

2.2 Mammographic Stand

DRMS1001_01_1_E.ai

Units: mm (in.)

SPEC

Page 4: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-14

2.3 Control Unit

DRMS1001_221_01_E ai

Units: mm (in.)

2.4 Control Pad

DRMS1001_222_E.ai

Units: mm (in.)

SPEC

Page 5: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-15

2.5 Foot Switch

DRMS1001_223_E.ai

Units: mm (in.)

2.6 Generator (X-ray High Voltage Device)

825(

32.4

8)

DRMS1001_224_E.ai

Units: mm (in.)

SPEC

Page 6: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-16

2.7 AWS: PC and Keyboard

DRMS1001_225_E ai

Units: mm (in.)

SPEC

Page 7: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-17

2.8 HUB Unit

DRMS1001_226_E.ai

Units: mm (in.)

The external view and dimensions are just examples, as this is general-purpose electric equipment/component.

SPEC

Page 8: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-18

2.9 24x30 Compression Plate (High)

183(

7.20

)

341(13.43)

DRMS1001_227_H_E.ai

Units: mm (in.)Permissible range: ±10%

SPEC

Page 9: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-19

2.10 Operation Table

Characteristics - Can store the AWS PC.

DRMS1001_228_E.ai

Units: mm (in.)

SPEC

Page 10: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-20

Protective Shield (Optional)

DRMS1001_229_E.ai

Units: mm (in.)

Characteristics Comes with protective glass (equivalent to 0.5 mm Pb), and can be used in the study room. Can store the AWS PC.

600(23.6)

SPEC

Page 11: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-21

Protective Plate (Optional), Storage Rack (Accessory Sold Separately)

Characteristics Comes with protective glass (equivalent to 0.5 mm Pb), and can be used in the study room. Can store the AWS PC. Can store each compression plate or accessories during exposure.

600(23.6)

DRMS1001_230_E.ai

Units: mm (in.)

SPEC

Page 12: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-22

Protective Glass Cover for Additional Height (0.5 mm Pb Equivalent) (Optional) (Outside Japan)

609(23.98)

310(

12.2

)

23(0.91)

1900

(74.

8) 2140

(84.

25)

DRMS1001_231_E.ai

Units: mm (in.)

SPEC

Page 13: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-23

2.11 FDR MS-3500 Options

(1)-1 24x30 compression plate (Low) (1) Compression plate

(1)-2 24x30 compression plate (Shift)

DRMS1001_232_E.ai

Units: mm (in.)

341(13.4)

183(7.20)

341(13.4)

183(7.20)

SPEC

Page 14: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-24

(1)-3 24x30 compression plate (SS)

(1)-4 24x30 compression plate (Flex)

314(13.4)

183(7.20)

341(13.4) 313(12.3)

DRMS1001_233_E.ai

Units: mm (in.)

SPEC

Page 15: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-25

DRMS1001_234_E.ai

Units: mm (in.)

313(12.3)341(13.4)

215(8.46)

(1)-5 24x30H compression plate (FS)

(1)-6 18x24 compression plate (Low)

SPEC

Page 16: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-26

(1)-7 18x24 compression plate (High)

(1)-8 18x24 compression plate (Small)

276(10.9)312(12.3)

183(7.20)

276(10.9)

183(7.20)

311(12.2)

DRMS1001_236_E ai

Units: mm (in.)

SPEC

Page 17: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-27

(1)-9 18x24 compression plate (Flex)

(1)-10 18x24L compression plate (FS)

215(8.46)

310(12.2)276(10.9)

276(10.9)

DRMS1001_236_E ai

Units: mm (in.)

SPEC

Page 18: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-28

123(4.84)

183(7.20)

313 (12.3) 264 (10.4)

183 (7.20)

(1)-11 Rectangular spot compression plate

(1)-12 Compression plate for axilla radiography

DRMS1001_237_E.ai

Units: mm (in.)

SPEC

Page 19: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-29

(1)-13 Compression plate for magnifications

(1)-14 Rectangle spot magnification compression plate

251(9.88)

126(4.96)

311(12.2)

123(4.84)

311(12.2)

DRMS1001_238_E.ai

Units: mm (in.)

SPEC

Page 20: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-30

(1)-15 2D biopsy compression plate

183(7.20)

DRMS1001_239_E.ai

Units: mm (in.)

SPEC

Page 21: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-31

(2) Magnification table

Magnification table (when stored)

DRMS1001_240_E.ai

Units: mm (in.)

SPEC

Page 22: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-32

(3) Remote exposure hand switch

(4) Remote exposure foot switch

(5) Shielding plate

Units: mm (in.) DRMS1001_241_E.ai

SPEC

Page 23: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-33

(6) Under-armrest

Installed

DRMS1001_242_E.ai

Units: mm (in.)

SPEC

Page 24: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-34

(7) Compression plate storage holder

* Can be attached to the wall surface in either horizontal or vertical orientation.

57(2

.25)

DRMS1001_243_E.ai

Units: mm (in.)

SPEC

Page 25: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-35

2.12 Accessories Sold Separately(1) Biopsy Positioner

(1)-1 Positioner

(1)-2 Positioner Control Cabinet

DRMS1001_246_E.ai

Units: mm (in.)

SPEC

Page 26: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-36

(1) -3 Control panel

(1)-4 Biopsy compression plate

DRMS1001_247_E.ai

Units: mm (in.)

SPEC

Page 27: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-37

(1)-5 Needle holder C

(1)-6 Needle holder M

DRMS1001_248_E.ai

Units: mm (in.)

SPEC

Page 28: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-38

(1)-7 Needle holder Y

(1)-8 Needle holder M8

Units: mm (in.) DRMS1001_249_E ai

SPEC

Page 29: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-39

(1)-9 Needle holder E

(1)-10 Biopsy spacer

Units: mm (in.) DRMS1001_250_E ai

SPEC

Page 30: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-40

(1)-11 Needle guide cover

Units: mm (in.) DRMS1001_251_E ai

SPEC

Page 31: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-41

(2) Lateral approach kit (2)-1 Lateral adapter

(2)-2 Lateral biopsy compression plate

Units: mm (in.) DRMS1001_252_E ai

SPEC

Page 32: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-42

(3) Decoration label

594 23.39

1670(65.75)

Decoration label attachment surface

DRMS1001_254_E ai

Units: mm (in.)

SPEC

Page 33: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-43

3. Installation Space and Servicing Space

■ Mammographic Stand/Control Unit

● Front/rear direction

R700(27.6)R200(7.87)

DRMS1002_1_01_E.ai

Units: mm (in.)

SPEC

Page 34: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-44

● Height direction

26

5( 1

0.43

)

C arm variable range

DRMS1003_1_01_E.aiUnits: mm (in.)

SPEC

Page 35: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-45

4. Range of Movement■ Mammographic Stand

REFERENCE

Although the machine is set in the factory so that the C arm stops at 2,350 mm when it is moved up to its uppermost position, the uppermost position of the C arm can be adjusted within a range of 2,050 to 2,350 mm by repositioning the C arm elevation upper-limit switch.The corresponding adjustment range of the uppermost position of the FPD upper surface is 1,200 to 1,500 mm.

DRMS1060_1 ai

Unit:mm

1935

~228

5(fa

ctor

y de

faul

t)

+120°

Arm uppermost adjustable range

350

460~

810

-120°

224

C arm adjustable range

+90° -90°

690

1150

~150

0

Lowermost position of FPD upper surface

Uppermost position of FPD upper surface

974

1503

+180° -180°

Lowermost position at which c arm can swivel 180°

SPEC

Page 36: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-46SPEC

Page 37: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-47

5. Anchoring Locations

5.1 Mammographic Stand

DRMS1003_21_01_E.aiUnits: mm

Through-hole

SPEC

Page 38: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-48

5.2 Generator (X-ray High Voltage Unit)300(11.81) recommended

(Min. 200(7.87))

DRMS1003_22_E.ai

Units: mm (in.)

SPEC

Page 39: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-49

5.3 Control Unit

DRMS1003_21__01_E.ai

Units: mm (in.)

SPEC

Page 40: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-50

5.4 Operation Desk and Storage Rack

DRMS1003_23_E.ai

Units: mm (in.)

SPEC

Page 41: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-51

6. System Configuration Example CAUTION

Do not use a multi-socket/extension cord for the quality and safety of the product.

NOTES - The FDR-3500DRLH can be installed in an exposure chamber (patient

environment).- Appropriately furnish the exposure chamber and the operation chamber to

visually and acoustically check the patient conditions.

6.1 Installing only the Mammographic Stand in the Exposure Chamber

DRMS1005_1_E.ai

Do not place obstacles on the flow line of a patient.

Do not place peripheral devices in the safety guaranteed area.

Control Pad Cable: approx. 9m

Power Supply, I/F cable: approx. 4m

2503 or more

SPEC

Page 42: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-52

6.2 Installing Everything Except the Operation Desk in the Exposure Chamber

Length of the cable for the control pad : Approx. 10m

Power and I/F cable length : Approx. 5m

Power and I/F cable length : Approx. 5m

Power and I/F cable length : Approx. 11m

3090 or more

No peripheral devices are allowed to be installed in the safety guaranteed area.

No obstacle is allowed to be installed on the flow line of a patient.

DRMS1005_2_en.ai

SPEC

Page 43: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-53

6.3 Installing the Operation Desk in the Exposure Chamber

WARNINGWhen the operation desk is to be installed in the exposure chamber, put the optional protective board for radiation protection. Place the operation desk at a distance of 1,500 mm or more from the X-ray focus.

DRMS1008_1_E ai

Unit:mmFocus has to be on the same vertical line of center of the protective board

Do not place peripheral devices in the safety guaranteed area

Do not place obstacles on the flow line of a patient.

SPEC

Page 44: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-54SPEC

Page 45: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-55

7. Machine SpecificationsSpecifications of the FDR MS-3500 including processing capabilities are described below.They are subject to change without prior notice.

7.1 Exposure Unit Specifications

DRMS1008_2_E ai

Power RequirementsInput voltage Input current (standby) Input current (normal) Input current (peak) Frequency

AC200V (Single phase) 2.4A

16.5A 90A 50 Hz/60 HzAC208V (Single phase) 2.3AAC220V (Single phase) 2.18AAC230V (Single phase) 2.09AAC240V (Single phase) 2.0A

Requirements for Maximum Allowable Impedance between Substation Equipment and Circuit Breaker

Input voltage Maximum allowable impedance Input voltage Maximum allowable

impedanceAC200V (Single phase) 0.143Ω AC230V (Single phase) 0.165ΩAC208V (Single phase) 0.150Ω AC240V (Single phase) 0.172ΩAC220V (Single phase) 0.158Ω

Requirements for Circuit Breaker Connected to Equipment

Number of electrodes 2Rated insulation voltage Ui (V) 600 (V) or moreRated current (A) 50 (A) (AC240V)

* Install a breaker upstream of the device that does not trip when

or 130A flows for 10 seconds, and that trips when 1500A flows continuously for 0.1 seconds.

Insulation Distance between L and NCreepage 3mm (0.11 in.) or longerClearance 1.6mm (0.06 in.) or longer

Operating Characteristic Curves

Operating characteristic curves

Maximum

Maximum (60A~63A)

Maximum (40A~50A)

Current (Percentage of rated current)

Minimum

Operation tim

e

Temperature Characteristic Curve

Percentage of operation

time variation (%

)

Reference am

bient tem

perature

Ambient temperature (°C)

Reference ambient temperature ( 0°C)Reference ambient temperature ( 5°C)

* Recommended product : NF63-CV (50A) (Mitsubishi)

Total shutdown time

Delayed trip Instantaneous trip

SPEC

Page 46: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-56

DRMS1008_3_E.ai

Environmental Requirements(1) Operating conditions

Temperature : 20°C to 30°CHumidity : 30% RH to 75% RH (No dew condensation)Air pressure : 750 hPa to 1060 hPa

(2) Non-operating conditionsTemperature : 15°C to 35°CHumidity : 10% RH to 80% RH (No dew condensation)Air pressure : 750 hPa to 1060 hPa

* This equipment cannot be used at an altitude of over 3000 m (9842 ft).

Attenuation Equivalent• Collimator : At least 0.4 mmAl (to the conditions of 49kV radiation)• Exposure table : 0.1 mmAl (to the conditions of 30kV radiation)

: 0.21 mmAl (to the conditions of 30kV radiation)• Biopsy protection cover : 0.18 mmAl (to the conditions of 30kV radiation)

Operating ModeContinuous operation with intermittent loading

Continuous Operation60 shots/hour (28kV, W/Rh, 71mA, 4 consecutive hours)40 shots/hour (28kV, W/Rh, 71mA, 8 consecutive hours)However, maintain the interval shown in the “Minimum exposure interval for energy input to X-ray tube” graph at least.

SID (Focal Spot to Image Receptor Distance)65 cm (25.59 in.)

Characteristics of X-ray ExposureTube voltage range 22kV to 49kV (at 1kV increments)Tube voltage accuracy ±1kVmAs setting range Large focal spot: 2mAs to 600mAs

Small focal spot: 2mAs to 300mAsTomosynthesis: 16mAs to 90mAs(Depending on load conditions such as tube voltage)

mAs setting accuracy ±10%+0.2mAsTube current Large focal spot: 20mA to 200mA

Small focal spot: 20mA to 61mANominal maximum output power 7kW (35kV x 200mA) * Load time: 40ms

Nominal maximum output power (at 30kV, 40ms) 5kW (168mA)

Maximum output power for making exposures at 1-minute intervals

2.8kW (28kV x 100mA) Load time 1s

Total allowable mAs per hour (at 35kV) total 7560mAs

Anode material 10%Re/Tungsten face on molly alloy

X-ray tube focal spot sizeLarge focal spot: 0.3Small focal spot: 0.1* Per IEC 60336, determined with the slit camera method.

0.63mm Be

W/50μmRh, W/700μmAl

SPEC

Page 47: FDR MS-3500 05(1)E - Fujifilm Healthcare

FDR MS-3500 Service Manual026-201-05(1)E Spec-57

DRMS1008_5_E.ai

Cooling curve of the X-ray system (tube)----- Tube [°C]----- Ambient temperature

Time [min]

Temperature [°C

)]

Minimum exposure interval for energy input to X-ray system (tube)

■ 2D imaging● Tomosynthesis imaging

Min. exposure interval [sec]

Input for exposure [kV x mA x sec]

Condition: 25°C, large focal spot

X-ray Conversion MethodDirect conversion

X-ray Conversion Materiala-Se

236.4 mm (9.31 in.) (vertical) x 296.4 mm (11.7 in.) (horizontal)

Pixel Resolution50 x 50 μm/pix

Number of Output Pixels4728 (vertical) x 5928 (horizontal) = 28,027,584 pixels

Output gradation(1) FDR-3000DRL

16bit(2) FDR-3000AWS

SPEC