Upload
others
View
3
Download
1
Embed Size (px)
Citation preview
UNIVERSITÀ DEGLI STUDI DI MILANO
Facoltà di Scienze Matematiche, Fisiche e Naturali
Corso di Laurea in Fisica
Experimental validation of Monte Carlo simulation of Leksell Gamma
Knife Perfexion stereotactic radiosurgery system
Relatore interno: Prof.ssa Flavia Groppi
Relatore esterno: Dott. Giuseppe Battistoni
Correlatore: Dott. Maria Grazia Brambilla
Tesi di Laurea di:
Nicola Bertolino
Matricola 571204
Codice P.A.C.S. 87.53.-j
Anno Accademico 2008-2009
Riassunto
L'obiettivo del lavoro di questa tesi è la validazione sperimentale della simu-
lazione Monte Carlo con il codice FLUKA della testata per la radiochirurgia stere-
tassica Gamma Knife Perfexion della Elekta.
La radiochirurgia stereotassica si basa su interventi radioterapici di precisione
in cui viene utilizzato un sistema di riferimanto cartesiano per irraggiare la zona
interessata, minimizzanto la dose impartita alle zone cicostanti.
Il sistema Gamma Knife Perfexion è costituito da un'unità radiante dotata di
192 sorgenti di cobalto-60 opportunamente collimate verso un punto focale. Sono
presenti collimatori di tre dimensioni di�erenti (16, 8, 4 mm di raggio). Il sistema
Gamma Knife è utilizzato per radioterapia nella zona cerebrale. Per �ssare la testa
del paziente al letto del GK viene avvitato un casco al cranio del paziente e tramite
una CT (Computer Tomography) inserita in un opportuno software di Elekta viene
eseguita la piani�cazione del trattamento.
Il sistema informatico di Elekta per la piani�cazione dei trattamenti con Gamma
Knife, LGP PFX (Leksell Gamma Plan Perfexion), non tiene conto di cavità, diso-
mogeneità e zone di interfaccia. Infatti, l'algoritmo utilizzato da LGP PFX è basato
su un modello sempli�cato di interazione delle radiazioni gamma con la materia, che
considera omogeneo il mezzo utilizzato. Il metodo Monte Carlo potrebbe risultare
uno strumento utile per a�rontare questa problematica. Da una tesi collegata è
stato sviluppato il primo tentativo di simulazione completa del citato Gamma Knife
con il codice FLUKA.
La parte sperimentale di questo studio è stata eseguita al centro Gamma Knife
dell'ospedale Niguarda Ca' Granda'.
La prima prova è stata una misurazione della dose impartita all'isocentro ef-
fettuata tramite l'utilizzo di una camera a ionizzazione ad aria libera, 31016-PTW
Freiburg, con un piccolo volume sensibile inserita in un fantoccio composto da Gam-
mex 457 SolidWater (Leksell Gamma Knife Dosimetry Phantom) avente un diametro
i
di 160 mm, che modellizza una testa umana standard. La camera a ionizzazione è
stata introdotta dentro il fantoccio, che è stato posizionato in modo tale che la stessa
risultasse nel punto focale dei raggi gamma del Cobalto-60 dell'unità radiante. E'
stata quindi e�ettuata una misurazione con tutti i collimatori da 16 mm aperti,
i cui risultati sono stati successivamente comparati con quelli ottenuti tramite la
simulazione FLUKA eseguita riproducendo le medesime condizioni. L'errore nella
simulazione Monte Carlo (5 %) è dominato dall'incertezza sull'attività delle sorgenti,
mentre l'incertezza sperimentale calcolata è del 3 %. L'accordo fra i due valori risulta
all'interno dei margini di errore. Questa misura sperimentale è particolarmente im-
portante poichè è l'unico valore richiesto come informazione esterna dal software
LGP PFX.
Successivamente è stato e�ettuato un confronto della dose impartita all'isocentro
con le simulazioni per diverse aperture dei collimatori con i valori sperimentali.
Questa è anche una prima veri�ca del sistema computer-assistito per la piani�cazione
dei trattamenti LGP, infatti la dose sperimentale è ottenuta moltiplicando il valore
di dose sperimentale precedentemente ottenuto con camera a ionizzazione per gli
Output Factor (OF) forniti da Elekta. Gli OF sono il rapporto fra la dose impartita
da un set di collimatori (8, 4 mm) e la dose impartita da i collimatori da 16 mm.
Anche in questo caso risulta un buon accordo.
L'ultima veri�ca sperimentale realizzata in questo lavoro è un confronto fra pro�li
di dose relativa. I pro�li di dose sono misurazioni della dose trasversale nei piani
perpendicolari ai fasci di radiadione. Per queste misurazioni sono state utilizzate
pellicole radiocromiche EBT-ISP. Queste pellicole sono state preparate e posizionate
all'interno del fantoccio con l'ausilio di appositi supporti ed esposte a dosi crescenti
da 0 a 10 Gy in modo da ottenere una curva di calibrazione, che corregge la risposta
nonlineare alla radiazione delle pellicole radiocromiche. E�ettuata la calibrazione,
sono state eseguite tre serie di misure per ogni con�gurazione di collimatori (16, 8, 4
mm), impartendo un dose pari a 5 Gy. Le pellicole sono state quindi analizzate per
mezzo di un procedimento digitalizzato mediante uno scanner (Epson Expression
10000XL) per un confronto gra�co con i valori ottenuti dalla simulazione Monte
Carlo, ottenendo un buon accordo.
Dopo queste valutazioni sperimentali, le simulazioni basate sul codice Monte
Carlo FLUKA risultano convalidate. Un ra�namento delle simulazioni, con una
maggiore precisione della geometria e una statistica più grande, e delle misure sper-
imentali, riducendo l'errore statistico e aumentando la risoluzione della scanner, è
ii
prevista nello sviluppo di questo studio.
Dopo questa dovuta convalida del codice FLUKA è previsto uno studio appro-
fondito della sottistima e della sovrastima della dose impartita dovuta all'algoritmo
del software LGP PFX. E' previsto l'utilizzo di un fantoccio modi�cato con inserti
di materiali più densi o di cavità. Sarà poi possibile eseguire TC (Tomogra�e com-
puterizzate) del fantoccio in modo da piani�care trattamenti tramite il software di
Elekta. Con nuove misure sperimentali eseguite utilizzando la nuova con�gurazione
del fantoccio e con simulazioni che riproducono la situazione sperimentali si può
testare il piano di trattamneto del LGP.
iii
Contents
Abstract 2
1 Radiation therapy 5
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Stereotactic radiosurgery . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Gamma Knife 10
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2 Physics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3 Gamma Knife Perfexion model . . . . . . . . . . . . . . . . . . . . . 13
2.4 Leksell Gamma Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3 Isocentre dose measure with ionization chamber 19
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Gamma Knife Perfexion Dosimetry Phantom . . . . . . . . . . . . . . 20
3.2.1 Dosimetry Phantom With Ionization Chamber . . . . . . . . . 21
3.3 Ionization Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.1 Determination of the absorbed dose . . . . . . . . . . . . . . . 24
3.4 The Bragg-Gray Cavity Theory . . . . . . . . . . . . . . . . . . . . . 25
3.4.1 Bragg-Grey Theory and Ionization Chamber . . . . . . . . . . 27
3.5 Experimental procedure . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.6 Data discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.1 Output factor . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4 Relative dose pro�les with radiochromic EBT 31
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.2 Radiochromic EBT �lms . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2.1 Composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
1
CONTENTS CONTENTS
4.2.2 Physical and chemical behavior . . . . . . . . . . . . . . . . . 34
4.3 Leksell Gamma Knife Dosimetry Phantom �lm stack . . . . . . . . . 36
4.4 Experimental Procedure . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.4.1 Cutting marking and positioning . . . . . . . . . . . . . . . . 37
4.4.2 Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.4.3 Scan and analysis . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4.4 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.5 Data discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.5.1 Error in the Gafchromic EBT evaluated dose . . . . . . . . . . 40
4.5.2 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Conclusion 56
Bibliography 57
2
Abstract
The object of this work is to verify experimentally the Monte Carlo simulation
of the Gamma Knife Perfexion stereotactic radiosurgery system performed using the
Fluka code. The LGP PFX (Leksell Gamma Plan PERFEXION), an Elekta infor-
matic system for the planning of the Gamma Knife treatments, does not consider
interface zones and inhomogeneities. The algorithm improved in the LGP PFX is
based in a simplify γ ray interaction model with matter, considering the crossing
media like homogeneous (1, 17, 18, 19). Monte Carlo method (and FLUKA in
particular) could be a valid instrument to face this problem.
The experimental part of the study has been carried out in the Niguarda Ca'
Granda hospital Gamma Knife center.
The �rst test was a isocentre dose measurement performed by means of a free-
air ion chamber, 31016-PTW Freiburg with a small sensible volume, inserted in a
phantom made of Gammex 457 SolidWater (Elekta Dosimetry Phantom) with a
diameter of 160 mm, simulating a simplify standard human head. The ion chamber
was placed in the center of the phantom placed at the 60Co rays focus point of the GK
unit. A dose measurement with all the 16 mm collimators open was performed and
compared with the results of the FLUKA simulation in the same conditions. The
error in the FLUKA simulation (5 %) is determined by uncertainty in the source
activity, while the estimated experimental error is 3 %. The agreement between
the two values is inside the margin of error. This experimental measurement is very
important, because it is the only value requested as input by the LGP PFX software.
As second step a comparison between the simulated 4 and 8 mm collimators'
incendiary dose and the experimental value was made with a good agreement.
The �nal veri�cation included in this study is a comparison of the relative dose
pro�les. For this measure radiochromic EBT-ISP �lms were used. Gafchromic �lm
was prepared and placed inside the phantom by apposite holder and exposed to a
rising dose from 0 to 10 Gy in order to obtain a calibration curve to correct the non
3
CONTENTS CONTENTS
linear response of the �lm to radiation dose. After the calibration run, three set
of measurement in the coronal and axial plane were performed for all collimators'
con�guration (16, 8, 4 mm), for a 5 Gy dose. The �lms were scanned with a scanner
(Epson Expression 10000XL) for a graphical comparison with the MC simulation
values. Also in this case a good agreement between experiment and simulation was
achieved.
After these experimental evaluations, the simulation based on the FLUKAMonte
Carlo Code appear to be a valid instrument to test the LGP software and for future
studies about non homogeneous targets.
4
Chapter 1
Radiation therapy
Contents
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Stereotactic radiosurgery . . . . . . . . . . . . . . . . . . . 7
1.1 Introduction
Radiation therapy (also radiotherapy or radiation oncology, sometimes abbrevi-
ated to XRT) is the medical use of ionizing radiation as part of neoplastic pathologies
treatment and some not neoplastic pathologies. Radiation therapy works because
the radiation destroys the cancer cells ability to reproduce and the body naturally
gets rid of these cells.
Radiotherapy may be used for curative or adjuvant cancer treatment. It is used
as palliative treatment (where cure is not possible and the aim is for local disease
control or symptomatic relief) or as therapeutic treatment (where the therapy has
survival bene�t and it can be curative). Radiotherapy has several applications in
non-malignant conditions, such as the treatment of trigeminal neuralgia, severe thy-
roid eye disease, pterygium, pigmented villonodular synovitis, prevention of keloid
scar growth, and prevention of heterotopic ossi�cation. The use of radiotherapy in
non-malignant conditions is limited partly by worries about the risk of radiation-
induced cancers (9).
Radiotherapy is used for the treatment of malignant tumors may be used as
the primary therapy. It is also common to combine radiotherapy with surgery,
5
Chapter1. Introduction
chemotherapy, hormone therapy or some mixture of the three. Most common can-
cer types can be treated with radiotherapy in some way. The precise treatment
intent (curative, adjuvant, neoadjuvant, therapeutic, or palliative) will depend on
the tumour type, location, and stage, as well as the general health of the patient.
The radiation �elds may also include the draining lymph nodes if they are clinically
or radiologically involved with tumour, or if there is thought to be a risk of subclin-
ical malignant spread. It is necessary to include a margin of normal tissue around
the tumour to allow for uncertainties in daily set-up and internal tumor motion.
These uncertainties can be caused by internal movement (for example, respiration
and bladder �lling) and accuracy in patient set-up relative to the tumour position.
To spare normal tissues (such as skin or organs which radiation must pass through
in order to treat the tumour), shaped radiation beams are aimed from several angles
of exposure to intersect at the tumour, providing a much larger absorbed dose there
than in the surrounding, healthy tissue.
Shortly after the discovery of the x-ray by German physicist Wilhelm Conrad
Roentgen in 1895, the "powerful rays" were being used to e�ectively treat cancer.
Today, an increasing number of patients have their cancers treated successfully,
with fewer side e�ects and preservation of normal tissue function, using radiation
therapy.
A cancer patient may be treated with radiation alone. Prostate cancer and
larynx cancer are often treated in this manner.
Sometimes radiation therapy is part of a patient's treatment. Patients can be
treated with radiation therapy and chemotherapy before surgery. This may allow a
patient to have less radical surgery than would otherwise be required. Chemotherapy
may be used simultaneously with radiotherapy without surgery to improve the local
response and reduce metastatic disease (9).
There are three fundamental kind of radiation therapy:
� External beam therapy
� Brachytherapy
� Internal therapy, a form of treatment where a source of radiation is put inside
your body.
The treatment planning for external beam therapy consist of:
6
Chapter1. Stereotactic radiosurgery
� de�nition of position and volume of the target and of patience's anatomic
parameter, based on clinical history or imaging technique
� choice of treatment method
� calculation of dose given to the target, and to the surrounding tissues.
This procedure is now computer-assisted by means of appropriate software (8).
1.2 Stereotactic radiosurgery
Stereotactic is a term that means: the involving, being, utilizing, or used in
a surgical technique for precisely directing the tip of a delicate instrument (as a
needle) or beam of radiation in three planes using coordinates provided by medical
imaging in order to reach a speci�c locus in the body.
Stereotactic surgery is a minimally-invasive form of surgical intervention which
makes use of a three-dimensional coordinates system to locate small targets inside
the body and to perform on them some action such as ablation (removal), biopsy,
lesion, injection, stimulation, implantation, radiosurgery (SRS) etc. Stereotactic in
Greek means "solid ordering".
The open stereotactic method provides the basis for radiosurgery and the �rst
stereotactic instrument was designed for use with probes and electrodes. Despite
its name, stereotactic radiosurgery is a non-surgical procedure that delivers a single
high-dose of precisely-targeted radiation using highly focused gamma-ray or x-ray
beams that converge on the speci�c area or areas of the brain where the tumor or
other abnormality resides, minimizing the amount of radiation to health brain tissue.
Although stereotactic radiosurgery is often completed in a one-day session, physi-
cians sometimes recommend multiple treatments, especially for tumors larger than
2,5 cm in diameter. The procedure is usually referred to as fractionated stereotactic
radiosurgery when two to �ve treatments are given and as stereotactic radiotherapy
when more than �ve treatments are given. (11)
During treatment, the patient will lie on a table and the machine may rotate
around the target while it works.
Sometimes, a head frame may be attached to patient's scalp to keep it very still
during therapy. There are di�erent machines used to perform stereotactic radio-
surgery. Some require the use of a frame, and others do not.
7
Chapter1. Stereotactic radiosurgery
The entire procedure, including the planning stage, takes about half a day or
less. The time period when patient is receiving the radiation is usually limited.
Some patients receive therapy more than once.
In theory, any organ system inside the body can be subjected to stereotactic
surgery. Di�culties in setting up a reliable frame of reference (such as bone land-
marks which bear a constant spatial relation to soft tissues), however, mean that its
applications have been limited to brain surgery. Plain X-ray images (radiographic
mammography) and computed tomography can be used to guide the procedure.
Conventional external beam radiation therapy � the most common form of radia-
tion therapy � delivers full dose radiation to the tumor and some of the surrounding
brain tissue. For several reasons, the target area for conventional radiation deliber-
ately includes a border (called a �margin�) of normal brain around the tumor. These
reasons include uneven tumor borders, the risk of invisible spread of the tumor into
the surrounding tissue, a larger tumor size, or the presence of multiple tumors. Since
normal brain tissue is included in the full-dose region, conventional radiation is bro-
ken down into small daily doses so the normal brain tissue can tolerate it. As a
result, reaching the desired dose of radiation takes several weeks of daily treatment.
Radiosurgery focuses radiation beams more closely to the tumor than conventional
external beam radiation. This is possible through the use of highly sophisticated
computer-assisted equipment. A head frame or facemask used for this treatment
allows very precise set up, localization and treatment of the tumor. Using advanced
computer planning, radiosurgery minimizes the amount of radiation received by nor-
mal brain tissue and focuses radiation in the area to be treated. Since conventional
radiation therapy covers more normal tissue, it can often be given only once. Ra-
diosurgery, however,may be considered for re-irradiation due to its precision and the
possibility of avoiding previously treated areas.
Radiosurgery is used to treat arteriovenous malformations (AVMs), a tangle of
expanded blood vessels that disrupts normal blood �ow in the brain and sometimes
bleeds. AVMs are the leading cause of stroke in young people. When treated with
radiosurgery, arteriovenous malformations (AVMs) begin to thicken and close o�
slowly, typically over several years.
Stereotactic radiosurgery is an important alternative to invasive surgery, espe-
cially for tumors and blood vessel abnormalities located deep within or close to vital
areas of the brain. Radiosurgery is used to treat many types of brain tumors, either
benign or malignant and primary or metastatic and single or multiple. Sometimes
8
Chapter1. Stereotactic radiosurgery
radiosurgery is performed after surgery to treat any residual tumor cells.
Radiosurgery is also a treatment option for other neurological conditions.
Stereotactic radiosurgery works in the same way as other forms of radiation
treatment. It does not actually remove the tumor; rather, it damages the DNA of
tumor cells. As a result, these cells lose their ability to reproduce. Following the
treatment, benign tumors usually shrink over a period of 18 months to two years.
Malignant and metastatic tumors may shrink more rapidly, even within a couple of
months.
There are three basic kinds of stereotactic radiosurgery equipment, each of which
uses di�erent instruments and sources of radiation:
� The Gamma Knife, which uses 192 or 201 beams of highly focused gamma rays
all aiming at the target region. The Gamma Knife is ideal for treating small
to medium size lesions. See the Gamma Knife page for more information.
� Linear accelerator (LINAC) machines, prevalent throughout the world, deliver
high-energy x-rays, also known as photons. The linear accelerator can perform
radiosurgery on larger tumors in a single session or during multiple sessions,
which is called fractionated stereotactic radiotherapy. Multiple manufactur-
ers make this type of machine, which have brand names such as Novalis Tx,
XKnife, and CyberKnife. See the Linear Accelerator page for more informa-
tion.
� Proton beam or heavy-charged-particle radiosurgery. The number of centers
o�ering proton therapy has increased dramatically in the last several years.
(29)
9
Chapter 2
Gamma Knife
Contents
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2 Physics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3 Gamma Knife Perfexion model . . . . . . . . . . . . . . . 13
2.4 Leksell Gamma Plan . . . . . . . . . . . . . . . . . . . . . 16
2.1 Introduction
In the 1950s, Swedish professors Borje Larsson of the Gustaf Werner Institute,
University of Uppsala, and Lars Leksell at the Karolinska Institute in Stockholm,
Sweden, began to investigate combining proton beams with stereotactic (guiding)
devices capable of pinpointing targets within the brain. This approach was eventu-
ally abandoned because it was complex and costly. Instead, in 1967, the researchers
arranged for construction of the �rst Gamma Knife device using cobalt 60 as the
energy source. The prototype unit, used for 12 years in Sweden, was speci�cally
designed for functional neurological surgery, that is, for treatment of patients with
pain, movement disorders, and even certain behavioral disorders that were not re-
sponsive to conventional psychiatric treatment.
Realizing the potential of stereotactic radiosurgery for treating brain tumors,
professor Leksell and his colleagues built a second Gamma Knife in 1975. It was
installed at the Karolinska Institute and became an integral part of the neurosurgical
service there. The third and fourth units, built in the early 1980s, were installed in
10
Chapter2. Physics
Buenos Aires, Argentina, and She�eld, England.
The initial gamma unit design was intended for lesion generation in functional
neurosurgery. Through the development of stereotactic angiography, AVMs became
suitable targets for stereotactic irradiation as did cranial base tumors imaged with
pneumoencephalography or cisternography. In 1980's, an increasing number of pa-
tients had radiosurgery for AVMs, selected benign tumors, and small-volume ma-
lignant tumors. Today over 300000 patients have undergone Gamma Knife surgery
and over 35000 per year receive the treatment.
Currently, the Gamma Knife is used primarily to treat benign brain tumors,
AVMs, acoustic neuromas, pituitary adenomas, craniopharyngiomas, brain metas-
tases, other tumors of the skull base, and pineal region tumors. Most users of
Gamma Knife technology have restricted lesion size to a mean spherical diameter of
35 mm (and usually less). Although larger lesions were treated, to maintain safety
a signi�cant increase in lesion volume must be paralleled by a decrease in delivered
dose. A large decrease in dose for larger lesions leads to a relatively ine�ective total
dose from a radiobiological standpoint, and probably does not improve upon what
might be obtained by standard fractionated techniques. Failure to decrease the dose
for larger volumes can lead to a higher complication rate.
For certain patients with various deep�seated tumors or arterio�venous malfor-
mations, the Gamma Knife may be preferable to conventional surgery. As the unit
can carefully model the radiosurgical doses to lesions of suitable size (preferably less
than three centimeters in diameter), it is used as an alternative approach to stan-
dard microsurgical tools in these patients. Because of the fact the radiation fallo�
is very steep outside the target area, the surrounding brain tissue is spared harmful
after e�ects. Gamma Knife radiosurgery also is safer than many existing procedures
because patients need not undergo risky, open�skull procedures, and adult patients
do not require general anesthesia. Thus, the Gamma Knife is especially useful when
conventional surgical techniques would pose high risk, such as in the presence of
other illnesses or when a patient's age prohibits standard surgery. (28)
2.2 Physics
The basic physics of the Gamma Knife has remained substantially the same since
its conception. The device uses cobalt-60 as a radiation source that decays through
emission to a stable isotope of nickel; 60Ni, with a half life of 5.26 years. As a part
11
Chapter2. Physics
of the decay process, one electron with an energy of up to 315 keV and two gamma
rays with energies of 1.17 MeV and 1.33 MeV are emitted. It is the gamma radiation
that is used to clinical e�ect in the gamma knife and contributes to the naming of
the device.
Figure 2.1: Cobalt-60 decay scheme
The details of the internal design of the gamma knife changes slightly among the
four models currently in use around the world (the U, B, and C models, and the new
Perfexion model). Inside the Gamma Knife unit are an array of 60Co sources (201
sources in the U, B, and C models, 192 in the Perfexion) which are aligned with a
collimation system. The collimation system focuses the individual beams of gamma
radiation to a very precise focus point. While an individual beam has a relatively
low dose rate and causes minimal biological e�ect, the superposition of all beams
at the focus point have a much higher dose rate. The Gamma Knife can therefore
target very precise areas of tissue without causing signi�cant collateral damage to
areas outside of the targeted area.
Modi�cation of the isodose distribution is achieved by using combinations of
isocentres using di�erent collimators, di�erent stereotactic locations, and di�ering
dwell times.
12
Chapter2. Gamma Knife Perfexion model
2.3 Gamma Knife Perfexion model
Figure 2.2: Gamma Knife Perfexion
The Perfexion is the last Gamma Knfe model. The couch serves as the patient
positioning system (PPS), moving the patient to the preselected stereotactic coordi-
nates. The PPS can move between coordinates at speeds of up to 10 mm/s, which is
an order of magnitude faster than the previous Automat Positioning System (APS)
implemented on model C (the �rst model with automatized positioning), which has
a speed of 0.8 mm/s. Docking of the patient to the PPS is achieved by means of an
adaptor that attaches to the standard stereotactic Leksell G-frame with three clips.
The adaptor is then mounted directly to the PPS with a simple locking mechanism.
Angulation of the head in the sagittal plane, the so-called gamma angle, can be at
70, 90, or 110 degrees.
The shielding doors now move horizontally to the left and right, rather than
vertically, before the couch/PPS moves in.
The beam delivery system (radiation unit) has been redesigned with a di�erent
beam geometry. An array of 192 Cobalt-60 sources is arranged in a cone section con-
�guration. This di�ers substantially from the previous hemispherical arrangement.
As a result, the sources have a di�erent source to focus distance, varying from 374
mm to 433 mm. The majority of the sources are closer to the focus/isocenter com-
pared with previous models, yielding a slightly higher dose rate for a given source
13
Chapter2. Gamma Knife Perfexion model
activity.
Figure 2.3: Gamma Knife Perfexion's collimators
By the inverse square law, beams originating from sources farther from the focal
point contribute a lower dose to the isocentre; thus, more complex modeling is
required to reproduce dose distributions for treatment planning.
The primary and secondary collimators have been replaced by a single, larger, 12
cm thick tungsten collimator array, in which collimators are arranged in a series of
�ve concentric rings, similar to Models B and C. The collimator array is subdivided
into eight independently variable sectors, each containing 24 sources and 72 collima-
tors (24 collimators for each of the three collimator sizes) around its circumference.
The increase in the volume of the radiation cavity to more than three times the
previous volume allows for a greater treatment range in the x and y dimensions of 160
mm and 180 mm, respectively. The z coordinate is limited by the physical distance
from the focal point to the inner surface of the collimator assembly. These ranges
are much greater than the previous models for the automatic positioning system
(APS). This increased distance from focal point to the inner surface, combined
14
Chapter2. Gamma Knife Perfexion model
with the nearcylindrical shape of the radiation cavity compared with the previously
hemispherical shape of previous Models makes mechanical collisions between the
frame or patient's head and the interior surface of the collimator assembly far less
likely.
The treatment cavity is lined with a 1 mm thick insert made of aluminum,
denoted the collimator cap. This acts as a collision touchguard, causing immediate
retraction of the sources to the home position in response to any pressure being
placed on it, allowing the safe manual removal of the patient. In addition, it protects
the collimator assembly from the ingress of dirt, and it can be easily removed and
cleaned if necessary.
The range of beam diameters has changed from previous models. There are
collimators of three di�erent size 4 mm, 8 mm and 16 mm.
Beam diameters are automatically changed by moving the sources over the se-
lected collimator set, an action performed by servo�controlled motors located in the
sector drive unit at the rear of the radiation unit. Sources have �ve possible posi-
tions: 4 mm, 8 mm, 16 mm, sector o�, and home. Although sector o� and home are
both positions in which the beams are blocked, in sector o�, the sources are closer to
the collimators, taking less than 1 second to reach any of the beam on positions. In
the sector o� position, the sources are located between the 4 and 8 mm collimators,
allowing minimal source travel time to block the beams.
This is used when an individual sector is blocked, when the patient is being
moved between stereotactic coordinates, or when the patient is being transported
into or out of the radiation unit. As a result, the couch no longer needs to move to
the defocus position, as is the case for APS treatments with Model C. This greatly
lowers transit doses to the patient, as beams are only turned on when the patient is
in the correct treatment position. Home is the position occupied by the sources when
the machine is switched o�, idle, or when an emergency stop has been activated.
In this position, the sources are withdrawn further into the radiation unit, several
centimeters away from any of the collimators, giving a lower dose rate outside the
unit.
The sources are licensed and meet the ANSI standard N�542 for medical radio-
therapy sources.
Each of the 192 radioactive sources located in the radiation unit is composed of60Co pellets which are encapsulated in double stainless steel capsules with welded
closures.
15
Chapter2. Leksell Gamma Plan
The total activity at the time of loading Leksell Gamma Knife Perfexion is
approximatively 6000 Ci ≡ 222 TBq (1 Ci = 3,7x1010 Bq). The cobalt sources are
delivered to the site in a specially designed and approved protective container, called
cask. Once loaded into the radiation unit, the sources are maintenance free. They
are handled again only when it is necessary to reload the unit with new sources.
The sources are loaded into the positions within the radiation unit by use of a
loading machine. The loading machine is speci�cally designed to transfer sources
from the shipping cask to the radiation unit with full radiation safety. It is built
almost entirely of lead and weights approximately 12000 kg. Due to the heavy
weight of the equipment, radiation unit, loading machine and cask, the room �oor
must withstand a total weight of 38000 kg. A certain working area is also required.
(24) (30)
2.4 Leksell Gamma Plan
Figure 2.4: Leksell Gamma Plan Perfexion's screenshot
The Leksell GammaPlan PFX (LGP PFX) is a new version (vs 8.2) of the LGP
that runs on a PC platform with a Linux operating system as an SQL client server
16
Chapter2. Leksell Gamma Plan
database. This con�guration accommodates remote planning and unlimited near-
instant access to previous patient treatments. Leksell Gamma Plan PERFEXION
supports DICOM image studies obtained from CT, MRI and PET scans as well
as DICOM angiograms. Using the positioning information provided by the Leksell
Coordinate Frame, these image studies are given precise Leksell location coordinates.
New features in the treatment shot dialogue window include composite shots
and dynamic shaping as well as mouse and keyboard shortcuts, enabling faster,
more user friendly treatment planning.
The most outstanding original feature of the collimator design is the ability to
generate a single isocenter composed of di�erent beam diameters. This composite
shot feature allows each individual isocenter of a treatment plan to have its own
optimized shape, thereby increasing the conformity of the overall treatment plan.
In the treatment of multiple metastases in which a single isocenter may be used
for each tumor, this individual isocenter can now be shaped to the target, without the
time penalty of changing collimator helmets or increasing the number of isocenters
to increase conformity.
Intermediate collimator sizes can be mimicked by alternating sectors with di�er-
ent collimator sizes. For example, a 6 mm collimator can be created by placing 4
and 8 mm collimators in opposing sectors. The ability to change collimators in less
than 1 second removes the previous time penalty of approximately 8 minutes every
time a collimator helmet needed to be changed.
Extreme elongation of isodoses from a single isocenter can be achieved by se-
lective blocking of sectors. This feature allows blocking of more sources than that
previously achievable on Models B and C by beam plugging, allowing greater control
of isodoses and potentially greater dose gradients. Doses to critical structures can
be limited by a process called dynamic shaping. This feature has the same func-
tion as the automatic generation of plug patterns using \textit{shields} in previous
versions of LGP as used for Models U, B, and C. Instead of placing a shield on a
volume at risk, the critical structure is outlined and de�ned as a risk volume. Sec-
tors that contain beams that pass through the risk volume are blocked. The severity
of blocking can be varied between four levels. The sectors are then automatically
blocked during treatment, resulting in no extra set up time, as would be required
with Models U, B, and C. However, beam on times will increase accordingly in order
to compensate for the lower dose rate from the partially blocked collimator array,
just as with traditional plugging. A potential disadvantage with the new system is
17
Chapter2. Leksell Gamma Plan
the inability to block individual beams to protect distant structures, e.g., the lens.
However, the manufacturer's measurements on anthropomorphic phantoms show
lower doses to the eyes as a result of increased shielding and lower transport doses.
For targets closer to the eyes, where larger numbers of beams need to be blocked,
dynamic shaping adapts for protection of the lens. (25)
18
Chapter 3
Isocentre dose measure with
ionization chamber
Contents
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Gamma Knife Perfexion Dosimetry Phantom . . . . . . 20
3.2.1 Dosimetry Phantom With Ionization Chamber . . . . . . 21
3.3 Ionization Chamber . . . . . . . . . . . . . . . . . . . . . . 23
3.3.1 Determination of the absorbed dose . . . . . . . . . . . . 24
3.4 The Bragg-Gray Cavity Theory . . . . . . . . . . . . . . . 25
3.4.1 Bragg-Grey Theory and Ionization Chamber . . . . . . . . 27
3.5 Experimental procedure . . . . . . . . . . . . . . . . . . . 28
3.6 Data discussion . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.1 Output factor . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.1 Introduction
The purpose of this chapter is to illustrate the materials, the experimental pro-
cedure and the theory behind the �rst step of this study, the isocentre dose measure
made with a ion chamber.
The experimental part of the study was done in the Niguarda Ca' Granda hos-
pital's Gamma Knife center.
19
Chapter3. Gamma Knife Perfexion Dosimetry Phantom
The �rst test was a punctual dose measure made with a free-air ion chamber,
31016-PTW Freiburg with a small sensible volume, inserted in a phantom made of
Gammex 457 SolidWater (Elekta Dosimetry Phantom) with a diameter of 160 mm,
simulating a simplify standard human head. The ion chamber was at the center of
the phantom placed at the 60Co rays focus point of the GK unit. A dose measure with
all the 16 mm collimators open was made and compared with the Fluka simulation
in the same condition. The error in the �uka simulation was a�ected by the 5 %
error in sources activity and the experimental estimated was 3 %. The agreement
between the two value is inside the margin of error. This experimental measure is
very important, because is the only value we can put in the LGP PFX.
As second step a comparison between the simulated 4 and 8 mm collimators'
punctual dose and the experimental value multiply by the output factor (ratio be-
tween a collimators' calculated dose and the reference 16 mm collimators' dose)
given by Elekta (the constructor company) was made with a good agreement.
3.2 Gamma Knife Perfexion Dosimetry Phantom
Figure 3.1: Leksell Gamma Knife Dosimetry Phantom(courtesy of Elekta)
For the experimental measurements has been used a Gamma Knife Perfexion
dosimetry phantom.
The shape and size of Gamma Knife Perfexion Dosimetry Phantom simulate the
head of an adult human, in fact it is a spherical phantom with a 8 cm ray. Gam-
20
Chapter3. Gamma Knife Perfexion Dosimetry Phantom
mex 457 Solid Water responds to 60Co radiation �eld as water would do regarding
absorbed dose, and, exactly for this reason this material has been selected to build
the Gamma Knife Perfexion Dosimetry Phantom.
Elekta Dosimetry Phantom allow several con�guration with di�erent components
depending on the application and the measures type. Generally speaking, the main
con�guration of the phantom are two:
� Dosimetry Phantom with ionization chamber, for measuring absorbed dose
rate.
� Dosimetry Phantom �lm stack, for checking precision in dose delivery.
Figure 3.2: Ionization Chamber position inside Gamma Knife Perfexion DosimetryPhantom
3.2.1 Dosimetry Phantom With Ionization Chamber
The con�guration with ionization chamber consists in several components as
show in the following �gure:
21
Chapter3. Gamma Knife Perfexion Dosimetry Phantom
Figure 3.3: Exploded view of Leksell Gamma Knifer Dosimetry Phantom with ion-ization chamber, mounted on a frame.
The phantom base À is mounted on the frame Á, there are three cylindrical
holes in the ion chamber clamp  the ion chamber housing à can be �tted into the
largest hole. The ion chamber housing is cylindrical and divided into two section by
a mechanical stop. The �lm holder align rods Ä are inserted into the two smaller
holes in the ion chamber clamp. The ion chamber clamp and the upper �lm holder
Å are �xated to the phantom by the front clamp Æ using two screws Ç. (4)
Measures are correctly performed when a small ionization chamber is inserted
in the appropriate phantom set-up, the phantom is attached into the frame adapter
22
Chapter3. Ionization Chamber
and it is moved to the treatment position, corresponding to Leksell coordinates
X = Y = Z = 100mm. In this con�guration the radiological focus point will
coincide with sensitive measuring point of the chamber, �nally the measures of
absorbed dose, according to the dose rate protocol, will be performed with all 192
sources on at 16mm collimators open. (5)
3.3 Ionization Chamber
Experimental measures has been performed with ionization chamber model Pin-
Point PTW 31016 waterproof, this chamber type is simple to use for measurements
in a phantom. The chamber cavity volume is about 0.016 cm3, this size is a com-
promise between the need for su�cient sensitivity and the ability to measure dose
at a point.
Figure 3.4: Ionization chamber PTW31016
PTW-31016 is a cylindrical chamber-type with a 2.9 mm diameter and length.
This chamber shows an identical spatial resolution in all three dimensions due to
the special design, this provides additional precision when scanning beams in the x
and y direction without changing the chamber orientation.
In use, the chamber, must be aligned in such a way that the radiation �uence
is approximately uniform over the cross-section of the cavity-chamber. The cavity
length therefore sets a lower limit on the size of the �eld in which measurements
may be made.
23
Chapter3. Ionization Chamber
The construction of the chamber should be as homogeneous as possible but for
technical reasons the central electrode is likely to be of a material di�erent from
that of the walls. Indeed the choice of materials may play an important role in
ensuring that the energy response of the chamber does not vary considerably. It is
also necessary for the air cavity not to be sealed; it should be designed so that it
will equilibrate rapidly with the ambient temperature and air pressure.
In PTW-31016 the wall material is compound by 0.57mm of PMMA and 0.09mm
of graphite, wall density is about 1.19 g/cm3 for PMMA and 1.84 g/cm3 for graphite,
the central electrode is made of Aluminum 99.98 R.
3.3.1 Determination of the absorbed dose
The absorbed dose to water at the reference point in water-like phantom for a
reference beam of quality Q0 and in the absence of the chamber is given by:
DwQ0 = MQ0ND,w,Q0 (3.1)
whereMQ0 is the reading of the dosimeter under the reference conditions used in the
standards laboratory and ND,w,Q0 is the calibration factor in terms of absorbed dose
to water of the dosimeter obtained from a standards laboratory. In most clinical
situations the measurement conditions do not match the reference conditions used
in the standards laboratory. This may a�ect the response of the dosimeter and
it is then necessary to di�erentiate between the reference conditions used in the
standards laboratory and the clinical measurement conditions.
When a dosimeter is used in a beam of quality Q di�erent from that used in its
calibration, Q0, the absorbed dose to water is given by:
DwQ0 = MQND,w,Q0kQQ0 (3.2)
where the factor kQQ0 corrects for the e�ects of the di�erence between the reference
beam quality Q0 and the actual user quality Q, and the dosimeter reading MQ
has been corrected to the reference values of in�uence quantities, other than beam
quality, for which the calibration factor is valid. The beam quality correction factor
kQQ0 is de�ned as the ratio, at the qualities Q and Q0, of the calibration factors in
24
Chapter3. The Bragg-Gray Cavity Theory
terms of absorbed dose to water of the ionization chamber.
kQQ0 =ND,w,Q
ND,w,Q0
=Dw,Q/MQ
Dw,Q0/MQ0
(3.3)
The most common reference quality Q0 used for the calibration of ionization cham-
bers is 60Co gamma radiation, in which case the symbol kQ is used for the beam
quality correction factor. Ideally, the beam quality correction factor should be mea-
sured directly for each chamber at the same quality as the user beam. However,
this is not achievable in most standards laboratories. Such measurements can be
performed only in laboratories having access to the appropriate beam qualities. For
this reason the technique is at present restricted to a few primary standard dosime-
try laboratories (PSDLs) in the world. The procedure requires the availability of
an energy-independent dosimetry system, such as a calorimeter, operating at these
qualities. A related problem is the di�culty in reproducing in a standards labora-
tory beam qualities identical to those produced by clinical accelerators. When no
experimental data are available, or it is di�cult to measure kQQ0 directly for realistic
clinical beams, in many cases the correction factors can be calculated theoretically.
Where Bragg-Gray theory can be applied, an expression for kQQ0 is:
kQQ0 =(Sw,air)Q
(Sw,air)Q0
(Wair)Q
(Wair)Q0
PQ
PQ0
(3.4)
which depends only on quotients of water to air stopping-power ratios and per-
turbation factors at the beam qualities Q and Q0. The only chamber speci�c factors
involved are the perturbation correction factors PQ and PQ0 . (6)
3.4 The Bragg-Gray Cavity Theory
Fundamental theory of radiation dosimetry and �rst theory developed, Bragg-
Gray cavity theory relates the absorbed dose to the material in a small cavity in
a uniformly irradiated piece of matter, to the absorbed dose to the surrounding
material. Cavity sizes are referred to as small, intermediate or large in comparison
with the ranges of secondary charged particles produced by photons in the cavity
medium. The case where the range of charged particles (electrons and positrons) is
much larger than the cavity dimensions (i.e., the cavity is regarded as small) is of
special interest. Brag-Grey theory is founded on two conditions:
25
Chapter3. The Bragg-Gray Cavity Theory
1. The cavity must be small when compared with the range of charged particles
incident on it, so that its presence does not perturb the �uence of charged
particles in the medium.
2. The absorbed dose in the cavity is deposited solely by those electrons crossing
the cavity.
If a �uence Φ of identical particles of kinetic energy T passes through an interface
between two di�erent media, g and w, then the absorbed dose on the g-side of
boundaries can be written as:
Dg = Φ
[(dT
ρdx
)c,g
]T
(3.5)
in a similar way, on the w-side:
Dw = Φ
[(dT
ρdx
)c,w
]T
(3.6)
where [(dT/ρdx)c,g]T and [(dT/ρdx)c,w]T are the mass collision stopping-powers of
the two media, evaluated at energy T. The result of �rst condition is that the electron
�uences are almost the same and equal to the equilibrium �uence established in the
surrounding medium, but this condition can only be valid in regions of charged
particle equilibrium or transient charged particle equilibrium.
Second condition implies that:
� Photon interactions in the cavity are negligible and thus ignored.
� All electrons depositing the dose inside the cavity are produced outside the
cavity and completely cross the cavity.
� No secondary electrons are produced inside the cavity and no electrons stop
within the cavity.
Therefore, assuming that the value of Φ is continuous across the interface (i.e. ig-
noring backscattering) the ratio of absorbed doses in the two media adjacent to their
boundary can be written as the ratio of the mass collision stopping-powers of the
two media:
Dw
Dg
=(dT/ρdx)c,w
(dT/ρdx)c,g
. (3.7)
26
Chapter3. The Bragg-Gray Cavity Theory
Mathematical statement often referred as Bragg-Grey relation.
Where [(dT/ρdx)c,w]/[(dT/ρdx)c,g] is the ratio of the average unrestricted mass
collision stopping powers of the medium and the cavity.
The validity of Bragg-Grey relation, under the term of two Bragg-Grey con-
ditions, holds only for each monoenergetic component of the spectrum for charge
particles crossing the cavity. For a di�erential energy distribution Φt (particles
per cm2MeV ) the appropriate average mass collision stopping-power in the cavity
medium g is:
mS̄g =
∫ Tmax
0
ΦT
( dTρdx
)c,gdT∫ Tmax
0
ΦT dT
=1
Φ
∫ Tmax
0
ΦT
( dTρdx
)c,gdT =
Dg
Φ. (3.8)
And likewise, for a thin layer of wall material w ;
mS̄w =
∫ Tmax
0
ΦT
( dTρdx
)c,wdT∫ Tmax
0
ΦT dT
=1
Φ
∫ Tmax
0
ΦT
( dTρdx
)c,wdT =
Dw
Φ. (3.9)
The Bragg-Gray relation in terms of absorbed dose in the cavity, is then:
Dw
Dg
=mS̄w
mS̄g
≡ mS̄wg . (3.10)
3.4.1 Bragg-Grey Theory and Ionization Chamber
One of the most applications of ion chamber is in the measurement of gamma-ray
exposure. An air-�lled ion chamber is particularly well-studied for this application
because exposure is de�ned in terms of amount of ionization charge create in air.
Under the proper conditions a determination of ionization charge in a air-�lled
ionization chamber can give an accurate measure of exposure, and a measurement
of the ionization current will indicate the exposure rate.
27
Chapter3. Experimental procedure
Dose and exposure are related by the following relation:
Dg =∆Q
∆m
(W̄
e
)g
(3.11)
where, ∆Q is is the charge of the ions of one sign, expressed in Coulomb, collected
from the active volume of the gas during irradiation, ∆m is the mass, expressed in
Kg, of the active volume of the gas, and (W̄/e)g is the average energy deposited in
the gas divided by the charge of one of the ion pairs, per ion pair created.
The concept of active volume just means that volume of the gas for which electric
�eld lines can sweep ions toward the electrodes. There may be guard volumes or in-
active volumes that still logically comprise the cavity volume, but do not participate
in sweeping the ions toward the electrodes.
Therefore, assuming an exposure measurement, we may write for the absorbed
dose in the medium:
Dg =∆Q
∆m
(W̄
e
)g
mS̄wg (3.12)
This equation allows one to calculate absorbed dose in the medium immediately
surrounding a Bragg-Grey cavity, on the basis of the charge produced in the cavity-
gas, provided hat the appropriate values of ∆m, (W̄/e)g, and mS̄wg are known. (7)
3.5 Experimental procedure
In this work the measurements of exposition were performed under Bragg-Gray
conditions, and introducing all corrective factors relative to ionization chamber cal-
culation about pression and room temperature. The Elekta Dosimetry Phantom has
been used with the PinPoint ionization chamber inserted in it and positioned at the
isocentre of the 192 sources of 60Co. The measurement was acquired with all the 16
mm collimators open and the time of exposition was 100.8 s in order to obtain an
absorbed dose of 5 Gy. This experimental measurement is very important, because
it is the only value requested as input by the LGP PFX, when a new radiating device
is installed this measure is asked by the software. The percentage error, for 60Co
photons beam quality Q0, is of 3 % and has been provided by National Laboratories
of Standard Measures.
Because of the di�erences of ambient condition ( temperature and humidity),
28
Chapter3. Data discussion
compared to Standard measures, has been necessary to correct respect work-conditions.
In this study this �rst measure has been compared with the Monte Carlo Fluka
value obtained by the simulation in the same condition. The error in the MC simu-
lation is a�ected by the error in the sources activities, which is the 5 %, compared
to that the statistical error can be neglected.
3.6 Data discussion
Exposition Time(s) Experimental Dose(Gy) MC Dose(Gy) Di�erence
100,8 5±3% 4,95±5% 1%
Table 3.1: Experiment-Simulation comparison
In the table above it is showed the absorbed dose obtained with the ionization
chamber (Experimental Dose) and with the FLUKA simulation (MC Dose). The
last term is the percentage di�erence between the two value:
Difference =Experimental Dose−MC Dose
Experimental Dose· 100 (3.13)
There is a good agreement between the two measure.
3.6.1 Output factor
For all LGK models, the absolute output is calibrated in reference to the dose
rate of the largest open collimator, i.e. 16 mm for the PFX system and 18 mm for
the other LGK models. Such a change largely re�ects the new design of the PFX
system (7).
The output factors (OF) for LGP (version 8.2) are the ratio between the dose
given by a set of collimators and the dose given by the 16 mm set of collimators.
OF =Di
D16
(3.14)
29
Chapter3. Data discussion
where i = 4, 8.
With the experimental value obtained from the ionization chamber, it is possi-
ble to make a �rst test of the LGP . We can multiply the output factors by the
experimental measure made with the ionization chamber (Experimental Dose) and
compare the results with the Fluka MC Dose obtained simulating the 4 and 8 mm
collimators set up.
Collimator(mm)
Outputfactor
Experimentaldose (Gy)
MC Dose(Gy)
Di�erence
8 0,924 4,62 4,54 1,7 %
4 0,805 4,02 3,93 2,2 %
Table 3.2: Output factor comparison
Where Di�erence is the same as in 3.13.
Based in this study, the MC simulation validate the output factor of the LGP.
30
Chapter 4
Relative dose pro�les with
radiochromic EBT
Contents
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.2 Radiochromic EBT �lms . . . . . . . . . . . . . . . . . . . 32
4.2.1 Composition . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.2.2 Physical and chemical behavior . . . . . . . . . . . . . . . 34
4.3 Leksell Gamma Knife Dosimetry Phantom �lm stack . . 36
4.4 Experimental Procedure . . . . . . . . . . . . . . . . . . . 37
4.4.1 Cutting marking and positioning . . . . . . . . . . . . . . 37
4.4.2 Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.4.3 Scan and analysis . . . . . . . . . . . . . . . . . . . . . . . 38
4.4.4 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.5 Data discussion . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.5.1 Error in the Gafchromic EBT evaluated dose . . . . . . . 40
4.5.2 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.1 Introduction
The purpose of this chapter is to illustrate the materials, the experimental pro-
cedure and the theory behind the measures of the relative dose pro�les and their
31
Chapter4. Radiochromic EBT �lms
comparison with the MC simulation's data.
For this measure radiochromic EBT �lms, made by ISP (International Specialty
Products), were used. Gafchromic EBT �lm o�ers many advantages making it
an appropriate instrument for this kind of dosimetry measures. It is near tissue
equivalence, easy to handle in room arti�cial ligth, self-developing and it has high
spatial resolution.
GafCrhomic �lms were cut, marked, placed inside the Elekta Dosimetry Phantom
in �lm stack con�guration by apposite holder and exposed to a rising dose from 0
to 10 Gy in order to obtain a calibration curve to correct the non linear response
to radiation's dose of the �lm. After calibration, three set of measure for coronal
and axial plane were made for all collimators con�guration (16, 8, 4 mm), giving a
5 Gy dose. The �lms were scanned with a scanner (Epson Expression 10000XL) for
a graphical comparison with the MC simulation values. Even in this case a good
agreement between experiment and simulation was achieved.
4.2 Radiochromic EBT �lms
In radiation dosimetry there are several problems associated to the measurement
of isodose curves (lines delimiting body areas receiving the same quantity of radia-
tion in radiotherapy) and depth-dose distributions in high-gradient regions of beams
using conventional measuring system such as ion chambers, semiconductors, ther-
moluminescent detectors and radiographic �lms. These di�culties have resulted in
a search for radiation dosimeter with high spatial resolution which does not require
a special developmental procedure and give permanent absolute values of absorbed
dose with an acceptable accuracy and precision and ease of handling and data anal-
ysis (3).
Some of this feature have been achieved with the introduction of radiochromic
dosimeters. These dosimeters, with very high spatial resolution and relatively low
spectral sensitivity variation are insensitive to visible light, thus o�ering ease of
handling and preparation in room light (4).
Radiochromic EBT �lms has been used for several years as a dosimetry media. It
is near tissue equivalence, and its high spatial resolution makes it ideal for measuring
in a steep dose gradient region. Gafchromic EBT (International Specialty Products,
NJ, USA), radiochromic �lm is one of the newest radiation-induced auto-developing
x-ray analysis �lms available for therapeutic radiation dosimetry in radiotherapy
32
Chapter4. Radiochromic EBT �lms
applications.
This �lm does not require any physical, chemical or thermal processing and it's
optical density (OD) is relatively stable within 24 hours, have a dimension of 20,3 x
25,4 cm and they can be use for a dose interval from 0,1 to 8 Gy. GafChromic EBT
tolerate a temperature up to 70oC. Anyway ISP suggest to keep gafchromic �lms at
room temperature (20-25 oC).
Gafchromic ISP-EBT has been developed speci�cally to address the needs of the
medical physicist and dosimetrist working in the radiotherapy environment.
4.2.1 Composition
Figure 4.1: Con�guration of Gafchromic EBT dosimetry �lm
Most radiochromic �lm products have active components, which include a radiation-
sensitive monomer. Upon irradiation, this active monomer polymerizes to form a
polymer colored dye; thus, the automated darkening process upon irradiation with
x-rays takes place. The �nal color, or more correctly the absorption spectrum, is
dependent on the proprietary chemical within the �lm itself. Radiation-induced re-
actions can have an incubation period of at least 1ms and polymerization can proceed
after irradiation has ceased causing a post-exposure density growth which manifests
itself as a signi�cant increase in the optical absorption. This fact is well known
within the medical physics community through the use of various radiochromic �lm
33
Chapter4. Radiochromic EBT �lms
products and the general rule of leaving �lms for a period of 24 h before readout is
performed (16).
Gafchromic EBT consist on two active layers 17 µm thick separated by a surface
layer 6 µm thick protected by two polyester layer each one 97 µm thick.
The active layers are water-equivalent material with a Zeff equal to 6,98. The
atomic composition of the GafChromic active layer is H (37,9%), C (42,3%), O
(16,2%), N (1,1%), Li (0,3%) and Cl (0,3%). (27)
4.2.2 Physical and chemical behavior
Gafchromic EBT �lm, has been commercially released which produces a di�erent
absorption spectrum from older radiochromic �lm products. The �lm also produces a
higher dose sensitivity which more closely matches the dose criteria for radiotherapy
applications.
When the active component is exposed to radiation, it reacts to form a blue
colored polymer with absorption maxima at about 636 nm, as seen in Figure 4.2 .
Figure 4.2: Absorption spectra of GAFCHROMIC EBT
Consequently the response of this dosimetry �lm will be enhanced by measure-
ment with red light.
34
Chapter4. Radiochromic EBT �lms
The post-exposure density growth of GAFCHROMIC EBT dosimetry �lm has
been investigated. In common with the currently available GAFCHROMIC dosime-
try �lms, the density of the developmental �lm increases with time following expo-
sure. For all GAFCHROMIC dosimetry �lms the density growth is approximately
proportional to log(time after exposure). However, for the EBT �lm the post-
exposure growth is substantially less than for previous GAFCHROMIC �lms. Also
the time period over which signi�cant post-exposure occurs is much less for EBT
�lm. Finally the magnitude and rate of post exposure changes in EBT �lm are
dependent upon the mode of readout (14, 26).
Figure 4.3: Percentage increase in optical density for EBT-�lm post irradiation atvarious dose level
In Cheung et al. (14) work as shown in Figure 4.3 an approximate 7%�9% increase
in coloration occurs up to 24 h post-irradiation. In absolute OD terms the increase
in postirradiation coloration is dependent on the magnitude of the delivered dose.
Approximately 1% of this coloration is from 6 h after irradiation. (3)
35
Chapter4. Leksell Gamma Knife Dosimetry Phantom �lm stack
4.3 Leksell Gamma Knife Dosimetry Phantom �lm
stack
The phantom with �lm stack consist of several components which is shown in
�gure.
Figure 4.4: Exploded view of Elekta Gamma Knife Dosimetry Phantom with vertical�lm stack, mounted on a frame.
The phantom base À is mounted in a frame Á. The ion chamber plug  is placed
in the bottom hole of the phantom base. The bottom �lm holder Ã, the thin �lm
36
Chapter4. Experimental Procedure
Holder Ä and upper �lm holder Å have engravements on the arched side and are
positioned together by two �lm holder aligne rods Æ. When positioned together the
engravements from a diagonal track across the arched surface. The front clamp Ç
is �xated to the phantom by two screws È. (4)
4.4 Experimental Procedure
This experimental measuring took place at the Niguarda Ca' Granda hospital's
Gamma Knife center. It is a dosimetric measurement of the relative dose pro�le,
transverse dose measurements performed in the coronal plane or axial plane at a
given depth in the phantom, using radiochromic EBT-ISP �lms in the focus point
of the Gamma Knife unit. The position was 100,100,100 in the Gamma Knife
reference system.
4.4.1 Cutting marking and positioning
EBT �lm can be safely handled in the normal indoor lighting before and after
exposure. However is not totally insensitive to light. When the �lm is not being set
up in the phantom, or being exposed to the radiation �elds, or being digitized or
measured, it is recommended to minimize the exposure to light, so the �lms were
kept in their box.
GafChromic EBT were cut with a paper cutter in pieces 6,3 ± 0,3 cm x 4,0 ±0,2 cm. It is essential to scan all pieces in the same direction, for this reason a track
of the orientation relative to the original sheet was kept. The �lms were marked
with a number and Cartesian axis orientation, so it is possible to track the set up
used for the measure. Since the outside surfaces are polyester GafChromic �lm can
be marked using a pen.
Using the Elekta dosimetry phantom in the �lm stack con�guration, radiochromic
EBT �lm was placed in the phantom in the radiative unit for the experimental mea-
sures centered in the focus point. The Gamma Knife PERFEXION reference system
coordinates of the focus point were 100,100,100.
4.4.2 Exposure
For the experimental measurement the EBT �lms were placed in the phantom
inside the GK units at the focus point of the gamma rays of the 60Co sources. Film's
37
Chapter4. Experimental Procedure
exposures in six di�erent con�guration were performed:
1. All 16 mm collimators open and �lm in coronal plane
2. All 16 mm collimators open and �lm in axial plane
3. All 8 mm collimators open and �lm in coronal plane
4. All 8 mm collimators open and �lm in axial plane
5. All 4 mm collimators open and �lm in coronal plane
6. All 4 mm collimators open and �lm in axial plane
A set of three measurement for each con�guration was performed.
4.4.3 Scan and analysis
Figure 4.5: Imagej screenshot
Radiochromic EBT �lms were scanned, after about 24 hours from exposure, by
an Epson Expression 10000XL high quality �at bed scanner. The resolution is 72 dpi
38
Chapter4. Experimental Procedure
and the GafChromics were digitalized with a 16 bit pixel depth as not compressed
TIFF images. The software used for acquisition was �Image Acquisition� by Epson.
In order to optimize the process it is necessary to perform 5 blank scanning to warm
up the Xenon �uorescent lamp. The scanner's borders were not used, because in
the central zone it is possible to obtain a �at response (0,3 % from mean value) and
a mask around the scanning zone was �xed to minimize the di�used light. The gap
between the values read by the scanner is about 1 %.
A scanner's limitation is the saturation level. In fact it is possible to separate
grey scales equivalent to an optical density equal to 2.9-3. For this reason it is
necessary to pay attention to the dose interval during the exposure.
The �rst step is to scan the unexposed piece of �lm, then the exposed pieces
were scanned all with the same orientation. The scanning procedure thus results in
two sets of images: unexposed �lm and irradiated �lms.
After the scanning process the images were analyzed by means of the �ImageJ�
program version 1.40a.
4.4.4 Calibration
Even if this study concerns relative dose pro�les, it is necessary to introduce
a dose calibration for the EBT dosimetry �lm because of the non-linear response
to gamma radiations. Radiochromic �lms were staked in the Elekta Dosimetry
Phantom and irradiated with rising radiation dose from 0 to 10 Gy in order to
obtain a calibration curve necessary to correct the non linear response of the �lms.
The calibration curve is obtained by a �t with an eighth degree polynomial:
y = a+ bx+ cx2 + dx3 + ex4 + fx5 + gx6 + hx7 + ix8 (4.1)
where a = -8,3742x10−6, b = 4,8567x10−5, c = 1,2399x10−8, d = -3,5511x10−12,
e = 4,6299x10−16, f = -3,0954x10−20, g = 1,1261x10−24, h = -2,114x10−29,
i = 1,6127x10−34; r2 = 0.9996.
39
Chapter4. Data discussion
Figure 4.6: Calibration curve
4.5 Data discussion
4.5.1 Error in the Gafchromic EBT evaluated dose
The response curve of a radiochromic �lm is usually expressed by the optical
density (OD) in relation with the given dose, in this case represented by the gray
scale of the scanner.
The optical density is de�ned as:
OD = −log10(I
I0)
where I is the intensity of light at a speci�ed wavelength l that has passed
through a sample (transmitted light intensity) and I0 is the intensity of the light
before it enters the sample (incident light intensity).
The experimental error is a�ected by to many factors like the di�erence of re-
sponse from a �lm to another, the error in the determination of net optical density,
the time for development after exposure.
40
Chapter4. Data discussion
Therefore the experimental error was determined using the Devic et al. method
(14) that was applied by Dr Parini in his work (19) at the Niguarda Ca' Granda
hospital.
The �rst step of this work is to calculate the net optical density and it's standard
deviation in this way:
netOD = log10Iunexp − 65535
Iexp − 65535(4.2)
σnetOD =1
ln10
√σ2
unexp
Iunexp
+σ2
exp
Iexp
(4.3)
where Iunexpand Iexp are respectively the intensity value of the exposed and the
unexposed �lm, σunexpand σexptheir standard deviation.
In order to �nd the most suitable function for the given system, the following
criteria are used:
1. the �t function has to be monotonically increasing;
2. the �t function has to pass through zero;
3. we choose the function that gives the minimum relative uncertainty for the
�tting parameters.
Based on these criteria, the chosen �t function is of the form:
Dfit = b · netOD + c · netODn (4.4)
The term n is introduced to account for the nonlinear dose dependence while
approaching the high dose region close to saturation level for radiochromic EBT-
ISP. The best curve was found with a, b and as �tting parameters and n was varied
from 0,5 to 5 with a step of 0,5.
b σb c σc n r2
558 34 3043 545 3 0,998
Table 4.1: Fitting parameters
41
Chapter4. Data discussion
Finally the error is dependent from the �t parameters and the scanner (1%),
applying the error's propagation theory:
σtot(%) =
√netOD2σ2
b + netOD2nσ2c + (b+ ncnetODn−1)2σ2
netOD + σ2scanner
D· 100
(4.5)
The percent total error depend on dose and it is represented by the curve of �g.
4.7.
Figure 4.7: Uncertainty estimates fof GafChromic EBT �lm (19)
This procedure shows that for a given dose over 0,3 Gy the uncertainty is esti-
mated to be smaller than 8 %.
4.5.2 Results
In the following �gures the graphical comparison between the FLUKA MC sim-
ulation and the radiated radiochromic EBT-ISP �lms data is shown.
The EBT �lms errobars represent the statistical error of the radiochromic �lm,
which in the dose interval of the measurement of this study is under the 8 %, while
the FLUKA errorbars is the statistical error of the simulation.
42
Chapter4. Data discussion
16 mm collimators
The agreement for this collimators set up is inside the margin of error and the
most accurate. In this case the simulation was ran with the largest number of
primaries particles.
The 16mm diagrams are split in two (positive and negative axis), allowing an
easier reading of data.
Coronal plane
20
30
40
50
60
70
80
90
100
110
0 2 4 6 8 10 12 14
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.8: Film 22
43
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-14 -12 -10 -8 -6 -4 -2 0
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.9: Film 22
20
30
40
50
60
70
80
90
100
110
0 2 4 6 8 10 12 14
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.10: Film 23
44
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-14 -12 -10 -8 -6 -4 -2 0
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.11: Film 23
20
30
40
50
60
70
80
90
100
110
0 2 4 6 8 10 12 14
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.12: Film 24
45
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-14 -12 -10 -8 -6 -4 -2 0
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.13: Film 24
Axial plane
20
30
40
50
60
70
80
90
100
110
0 2 4 6 8 10 12 14
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.14: Gaf 31
46
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-14 -12 -10 -8 -6 -4 -2 0
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.15: Gaf 31
20
30
40
50
60
70
80
90
100
110
0 2 4 6 8 10 12 14
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.16: Gaf 32
47
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-14 -12 -10 -8 -6 -4 -2 0
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.17: Gaf 32
20
30
40
50
60
70
80
90
100
110
0 2 4 6 8 10 12 14
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.18: Gaf 33
48
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-14 -12 -10 -8 -6 -4 -2 0
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.19: Gaf 33
8 mm collimators
As shown in the diagrams of �g. 4.20 to 4.25, the statistic for the 8 mm collimator
con�guration is limited. In fact the statistical error is greater than 10 % at the sides
of the diagrams. In this case, it would be usefull to run a simulation with a larger
number of primaries to achieve a more accurate comparison.
49
Chapter4. Data discussion
Coronal plane
20
30
40
50
60
70
80
90
100
110
-6 -4 -2 0 2 4 6
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.20: Film 19
20
30
40
50
60
70
80
90
100
110
-6 -4 -2 0 2 4 6
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.21: Film 20
50
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-6 -4 -2 0 2 4 6
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.22: Film 21
Axial plane
20
30
40
50
60
70
80
90
100
110
-6 -4 -2 0 2 4 6
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.23: Film 28
51
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-6 -4 -2 0 2 4 6
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.24: Film 29
20
30
40
50
60
70
80
90
100
110
-6 -4 -2 0 2 4 6
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.25: Film 30
52
Chapter4. Data discussion
4 mm collimators
In the 4 mm collimators con�guration the percentage error is small even at the
sides of the curve, because the statistic is lower than in the centre. The agreement,
especially in the axial plane pro�les, is not so good probably, but the resolution of
radiochromic �lm does not allow to draw de�nitive conclusions.
It is possible to improve the accuracy of the experimental measurements, in
order to reduce the statistical error, repeating the procedure described by Devic et
al. (14). Particularly it is necessary to take �ve times over the same scanning region
as the acquired images with the �lm pieces. This allows one to correct for �defective�
pixels, de�ned as pixels that di�er in intensity from the blank unattenuated signal.
It would be also helpfull to use an higher resolution scanner to improve the resolution
of the images.
Coronal plane
20
30
40
50
60
70
80
90
100
110
-4 -3 -2 -1 0 1 2 3 4
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.26: Film 16
53
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-4 -3 -2 -1 0 1 2 3 4
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.27: Gaf 17
20
30
40
50
60
70
80
90
100
110
-4 -3 -2 -1 0 1 2 3 4
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.28: Gaf 18
54
Chapter4. Data discussion
Axial plane
20
30
40
50
60
70
80
90
100
110
-4 -3 -2 -1 0 1 2 3 4
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.29: Film 25
20
30
40
50
60
70
80
90
100
110
-4 -3 -2 -1 0 1 2 3 4
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.30: Film 26
55
Chapter4. Data discussion
20
30
40
50
60
70
80
90
100
110
-4 -3 -2 -1 0 1 2 3 4
Dm
ax/D
(%
)
mm
EBT FILMFLUKA
Figure 4.31: �lm 27
56
Conclusion
The FLUKA Monte Carlo simulation of the Gamma Knife Perfexion stereotactic
radiosurgery system is validated by the experiments described in this thesis. A re-
�nement is probably necessary both in the simulation (statistics, geometrical model)
and in the accuracyof the measurements with radiochromic �lm to improve the cor-
respondence between data and simulation. This is a preliminary work for a future
testing of the Leksell Gamma Plan version 8.2 developed by Elekta. As shown in
other publications (1, 17, 18), the LGP algorithm is based in a simpli�ed model
of radiation interaction with matter. In fact the LGP does not consider air cavity,
inhomogeneities and zone of interface. The future prosecution of this study should
include in the analysis of this topic.
The proposed method for the future work is to acquire a CT (Computer Tomog-
raphy) scan of the phantom in order to perform a treatment plan by means of the
LGP PFX software. It is possible to insert in the Elekta Dosimetry Phantom �lm
holders made by di�erent media (aluminum, delrin), or air cavities, and perform CT
scans and new experimental measurements using the modi�ed phantom. It is also
expected to run new simulations with a target reproducing the phantom with the
insertion of inhomogeneous �lm holders. At last, a comparison between the LGP's
treatment planning and the simulated and experimental results will be performed.
By means of the combined results of simulation (computation) and measure-
ments it should be possible to quantify how much the LGP PFX treatment plan
overestimates or underestimates the dose given to the patients. An information of
this kind will be helpfull for the future treatments planning.
57
Bibliography
[1] W. Xiong, D. Huang, L. Lee, J. Feng, K. Morris, E. Calugaru, C. Burman, J
Li, , C-M Ma, �Implementation of Monte Carlo simulation for Gamma Knife�.
Journal of Phisycs: Conference Series 74 (2007).
[2] Fluka User Guide 2008
[3] Azam Niroomand-Rad, C. R. Blackwell, B. M. Cousey, K. P. Gall, J. M. Galvin,
W. L. McLaughlin, A. S. Meigooni, R. Nath, J. E. Rodgers, C. G. Soares,
�Radiochromic �lmdosimetry: Recommendations of AAPM Radiation Therapy
Committee Task Group 55 �. Med. Phys. 25, pp 2093-2115 (1998).
[4] S. Devic, J. Seuntgens, E. Sham, G. Hegyi, E. B. PodGorsak, G. C. Soares, A.
S. Kirov, I. Ali, �Dosimetric property of improved GafChromic �lm for seven
di�erent digitizers�. Med. Phys. Vol. 31 (9), pp. 2392-2401 (2004).
[5] Elekta Instruments, �Leksell Gamma Knife Dosimetry Phantom, Instruction
for use� (2007).
[6] P. Andreo, D.T. Burns, K. Hohlfeld, M.S. Huq, T. Kanai, F. Laitano, V. Smith,
S. Vynkier, �Absorbed Dose Determination in External Beam Therapy: An In-
ternacional Code of Practice for Dosimetry based on Standard of Absorbed Dose
to Water �. IAEA TRS-398 (2004).
[7] H.F. Attix, �Introduction to Radiological Physic and Radiation Dosimetry�. A
Wiley-Interscience Publication, (1986).
[8] L. Ma, P. Kjäll, J. Novtny Jr., H. Nordstrom, J. Johanssön, L. Verhey, �A
simple and e�ective method for validation and measurement of collimator out-
put factors for Leksell Gamma Knife Perfexion�. stacks.iop.org/PMB/54/3897
(2009).
58
BIBLIOGRAPHY BIBLIOGRAPHY
[9] G. Agati, �Introduzione alla �sica radiologica�. Seconda edizione, edizioni libre-
ria cortina (1988).
[10] G. Cittadini, �Diagnostica per immagini e radioterapia�. ECIG (2002).
[11] L. Leksell, �Stereotactic Radiosurgery�. Journal of Neurology, Neurosurgery,
and Psichiatry 46, 797-803 (1983).
[12] J. Novotny Jr, L. Dade Lunsford, M.S. Huq, �Measurement of relative output
factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by
�lm dosimetry�. Med. Phys. Vol. 36 (5), pp. 1768-1774 (2009).
[13] M. Ruschin, H. Nordström, P. Kjäll, Y.-B. Cho, D. Ja�ray, �Investigation of
intracranial peripheral dose arising from the treatment of large lesions with
Leksell GammaKnife Perfexion�. Med. Phys. Vol. 36 (6), pp. 2069-2073 (2009).
[14] ISP, International Specialty products, �Gafchromic EBT, Self-developing �lm
for radiotherapy dosimetry�. (Agosto 2007).
[15] S. Devic, J. Seuntjens, E. Sham, E.B. Podgorsak, C. Ross Schmidtlein, S. Kirov,
Christopher G. Soares, �Precise radiochromic �lm dosimetry using a �at-bed
document scanner �. Med. Phys. Vol. 32 (7), pp. 2245-2253 (2005).
[16] T. Cheung, M. J. Butson, P.K.N. Yu, �Post-irradiation colouration of
Gafchromic EBT radiochromic �lm�. Phys. Med. Biol. Vol. 50, pp. 281-285
(2005).
[17] V.Moskvin, R. Timmerman, C. DesRosiers, M. Randall, P. DesRosiers, P.
Dittmer, L. Papiez, �Monte Carlo simulation of Leksell Gamma Knife: Source
modelling and calculation in homogeneus media�. Phys. Med. Biol. Vol. 47, pp.
1995-2011 (2002).
[18] V.Moskvin, R. Timmerman, C. DesRosiers, M. Randall, P. DesRosiers, P.
Dittmer, L. Papiez, �Monte Carlo simulation of Leksell Gamma Knife: II. Ef-
fects of heterogeneous versus homogeneous for stereotactic radiosurgery�. Phys.
Med. Biol. vol. 49, pp. 4879-4895 (2004).
[19] F.M. O Al-Dweri, E.L. Rojas, A.M. Lallena, �E�ects of bone- and air-tissue
inhomogeneities on the dose distributions of the Leksell Gamma Knife calculated
with PENELOPE �. Phys. Med. Biol. Vol. 50, pp. 5665-5678 (2005).
59
BIBLIOGRAPHY BIBLIOGRAPHY
[20] V. Parini, �Caratterizzazione di un sistema per trattamenti radioterapici con
modulazione della �uenza del fascio�. Tesi di specializzazione in �sica sanitaria,
Università degli Studi di Milano (2008).
[21] J. Trnka, J. Novotny Jr, J Kluson, �MCNP-based computational model for the
Leksell Gamma Knife�. Med. Phys. 34 (1), pp. 63-75 (2007).
[22] J. Y. C. Cheung, K. N. Yua, C. P. Yu, R. T. K. Ho, �Monte Carlo calculation
of single-beam dose pro�les used in a gamma knife treatment planning system�.
Med. Phys. Vol. 25 (9), pp. 1673-1675 (1998).
[23] Y.C. Cheung, K.N. Yu, R.T.K. Ho, C.P. Yu, �Stereotactic dose planning system
used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic
lms MD-55 �. Applied Radiation and Isotopes Vol. 53, pp. 427-430 (2000).
[24] C. Lindquist, I. Paddick, �The Leksell Gamma Knife Perfexion and comparison
with its predecessors�. ONS-130 Vol. 61 (2007).
[25] Elekta Insruments AB Sweden, �Reference Manual for Leksell GamaPlan�.
[26] M. Yamauchi, T. Tomonaga, O. Nakamura, �Gafchromic �lm dosimetry with a
�atbed color scanner for leksell gamma knife therapy�. Medical Physics Vol. 31
(5), pp. 1243-1248 (2004).
[27] ISP, International Specialty products, www.ispcorp.com.
[28] Elekta instruments, www.elekta.com.
[29] RadiologyInfo, www.radiologyInfo.org.
[30] J. Régis, M. Tamura, C. Guillot, X. Muracciolle, M. Nagaje, D. Porcheron,
�Radiosurgery of the head and neck with the world's �rst fully robotized 192
Cobalt-60 source Leksell Gamma Knife Perfexion in clinical use�. (2006)
60
Aknowledgements
I would like to thank Elekta industries and Dr. Johansson in particular, Prof.
Collice (Head physician of neurosurgery S.C.,Niguarda), Dr. La Camera (Head of
GK centre) and all his sta�, Dr. Torresin (Chief of medical physics C.S.,Niguarda),
Dr. Mainardi (Niguarda), Ing. Agosteo (Politecnico di Milano), Dr. Mairani
(INFN), Dr. Muraro (INFN).
A special thank go to my supervisors Dr. Battistoni and Dr. Brambilla. They
have provided assistance in numerous ways.
I want to express my gratitude to all my family for the support and patience in the
last years.
At last I would like to Fabrizio for sharing worries and di�culties.
61