Upload
sole-gutierrez
View
216
Download
0
Embed Size (px)
Citation preview
8/19/2019 Experimental Setup and Controls for QPCR
1/1
Experimental setup and controls for qPCR
For accurate analysis of qPCR results, each experiment needs to be set up with multiple replicates
and controls (Figure 2)
Replicates: For each experimental and control sample to be compared, at least three technical
replicates are necessary to minimi!e errors in measured gene expression due to pipetting
Figure 2: qPCR assay setup. "utline
of the test and controls needed for an
experiment with two different samples
examined at se#eral time points post
treatment $nclude %&' biologicalreplicates for each time point studied
For each biological replicate studied,
perform 2 re#erse transcription
reactions (R) and * with no R (&
R control) For each c+- sample
generated, set up % technical replicates
for qPCR analysis $nclude a o
emplate Control for each gene
analy!ed to identify any signal due to
contamination
Controls
No RT control: For e#ery re#erse
transcription reaction, it is important to
incorporate a .no R control/ to identify erroneous signal due to genomic +- contamination
No template control: +uring qPCR setup, incorporate a .no template control/ for e#ery different
gene analy!ed to identify possible cross0contamination generated during preparation of samples
Reference genes: 1ultiple reference genes should be analy!ed for stable gene expression elect a
reference gene with an expression pattern that is unaltered by treatment or the different time points to
be tested
Cycling conditions: -lways follow the cycling conditions according to the master mix used For example, do not use a standard master mix with fast cycling conditions