Typical results for dogExigo has been evaluated by a major veterinary teaching hospital, against their reference analyzer, ADVIA 120 and manual differential counts.
Prod
uct s
pec
ifica
tion
Exi
go
art
.nr.
1504
215
- v 3
Product Specification
Veterinary Hematology System
WBC Dogr = 0.99
05
101520253035404550
0 10 20 30 40 50
Advia 120 Hematology System
Exig
o Sy
stem
GRAN Dogr = 0.99
0
5
10
15
20
25
30
35
40
45
50
0 10 20 30 40 50
Advia 120 Hematology SystemEx
igo
Syst
em
HGB Dogr = 0.97
5
7
9
11
13
15
17
19
21
23
5 10 15 20 25
Advia 120 Hematology System
Exig
o Sy
stem
HCT Dogr = 0.98
1520253035404550556065
15 25 35 45 55 65
Advia 120 Hematology System
Exig
o Sy
stem
RBC Dogr = 0.98
2
3
4
5
6
7
8
9
10
2 4 6 8 10
Advia 120 Hematology System
Exig
o Sy
stem
PLT Dogr = 0.97
0
100
200
300
400
500
600
700
800
900
1000
0 100 200 300 400 500 600 700 800 900 1000
Advia 120 Hematology System
Exig
o Sy
stem
®
Key features-Full automatic system with 19 or 17 parameters.
-Multiple species with programmable analysis profi les. (*)
-Built-in Mixer for 6 tubes of 3 diff erent sizes.
-Open tube inlet or direct start with a 20ul microcapillary.
-7” Color touch screen with self intuitive operation.
-Advanced Veterinary WBC diff . (*)
-Hematological Guidelines for pathological samples. (*)
-On board clot prevention system.(*)
-Control Assay Values scanned into system through barcode entry. (*)
-Capacity up to 60 samples per hour.
-USB, Centronics and RS232 output.
-Connector for external PC keyboard.
(*) see description below
Analyzer description& featuresArt no Article name
1400061 Exigo, 17 parameters
1400076 Exigo eos, 19 parameters
Analysis profi lesExigo comes with pre-installed animal analysis profi les and the possiblity to develop your own species profi les. Pre-installed: Dog, Cat, Horse, Cow, Ferret, Rabbit, Sheep, Rat, Mouse, Pig, New World Camel and Goat. New profi les for other species are available as product updates. Up to 20 animal profi les can be used in the instrument at the same time.
Available parameters Using impedance method: -WBC, LYMa, LYM%, MONOa, MONO%, GRANa, GRAN%-NEUTa*, NEUT%*, EOSa*, EOS%*-RBC, MCV, -PLT, MPV
Using photometric method (cyanide free):-HGB Calculated parameters-HCT, MCH, MCHC, RDWa, RDW%
* Available on Exigo eos only
Operation modesOpen tubes with EDTA blood.EDTA coated 20μl micro capillaries.
Advanced WBC diff methodA WBC diff erential from animal blood is one of the more diffi cult parameters to measure. By using an advanced analysis algo-rithm, Exigo takes into consideration both the actual measured cell compositions as well as what is expected form the ana-lyzed species, when performing the analysis. Every profi le uses a dynamic approach on how to analyze the sample for the most accurate result.
Hematological guidelines by ExigoSamples considered as pathological will be fl agged by the system and will give a clear recommended action for further analysis of the sample.
Recommended cleanersExigo should be cleaned periodically to avoid build up of proteins or bacterial growth inside the system. Use only original cleaners. Other cleaners may damage the interior of the system.
Art No Article name
1504126 Boule Hypochlorite CleanerUsed for Manual clot removal procedure.
1504111 Boule Cleaning KitUsed for periodic maintenance.
On-board clot prevention systemAnimal blood specimens have a tendency to cause clots and blockages in cell-counter analysis systems. Therefore Exigo is equipped with on-board cleaner, automatically keeping the system clean.
ReagentsExigo should only be used with its original reagents. All rea-gents are non-toxic and cyanide-free.
Art No Article name
1504201 Exigo Diluent 1,9 L
1504202 Exigo Lyse 1,9 L .
1504203 Exigo System Cleaner 1,9 L
Additional Reagents for Exigo eosArt No Article name
1504278 Exigo EOS Reagent 1,9 L
1504328 Exigo Diluent 10 L
System Verifi cationEvery veterinarian should be able to trust and depend upon their analyzer, which is why a full QC system has been designed into the Exigo software. And, by using barcodes for handling and registering the controls and their assay values, it is simple to use, eff ortless and 100% secure.
Control samplesControl samples should be used on a frequent (preferably daily) basis to verify the systems operation. Using the system Control frequently gives clear and quality proof of the system’s perfor-mance, which can easily be monitored through the built-in QC program.
Art No Article name Description
1504026 Boule Vet Con 1*4.5ml
1504027 Boule Vet Con 6*4.5ml
CalibratorCalibrators should be used for the calibration procedure which is typically made upon installation, after service or after periodic maintenance.
Art No Article name Description
1504028 Boule Vet Calibrator 1*3ml
Clinical dataCorrelation for veterinary samples is very dependent on the hematological composition of the blood from the species being analyzed, as well as sampling methodology and pre-analytic handling. The below correlation data is based on Boule control blood.
Precision and Measuring RangeRBC MCV HGB WBC PLT
Range 0-14.0 x 106 /µl 15-250 fl 0-99.9 g/dl 0-99.9 x 103 /µl 0-1999 x 103 /µl
CV% ≤1.1% ≤0.3% ≤1.0% ≤1.9% ≤3.5%
*Typical values over 10 consecutive measured Boule Control samples by open tube mode and under
good laboratory conditions.
Carry Over RBC HGB WBC PLT
Open Tube ≤2 % ≤1 % ≤1 % ≤2 %
MPA ≤2 % ≤2 % ≤2 % ≤3.5 %
Automatic Hematology Analyzer for veterinary use
-Multiple species with programmable analysis profi les. (*)
-Built-in Mixer for 6 tubes of 3 diff erent sizes.
-Open tube inlet or direct start with a 20ul microcapillary.
-7” Color touch screen with self intuitive operation.
-Advanced Veterinary WBC diff . (*)
-Hematological Guidelines for pathological samples. (*)
Automatic Hematology Analyzer for veterinary use
-Hematological Guidelines for pathological samples. (*)
Open tubes with EDTA blood.EDTA coated 20μl micro capillaries.
Advanced WBC diff methodA WBC diff erential from animal blood is one of the more diffi cult parameters to measure. By using an advanced analysis algo-rithm, Exigo takes into consideration both the actual measured cell compositions as well as what is expected form the ana-lyzed species, when performing the analysis. Every profi le uses a dynamic approach on how to analyze the sample for the most
System Verifi cationEvery veterinarian should be able to trust and depend upon their analyzer, which is why a full QC system has been designed into the Exigo software. And, by using barcodes for handling and registering the controls and their assay values, it is simple to use, eff ortless and 100% secure.
Control samplesControl samples should be used on a frequent (preferably daily) basis to verify the systems operation. Using the system Control frequently gives clear and quality proof of the system’s perfor-
Automatic Hematology Analyzer for veterinary use