Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
Evaluation and Verification of Aerosol Diluters:
Accuracy and Particle Loss
by
Terry Hoon Suk Jung
A thesis submitted in conformity with the requirements
for the degree of Master of Applied Science and Engineering
Graduate Department of Mechanical and Industrial Engineering
University of Toronto
© Copyright by Terry Hoon Suk Jung 2014
ii
Evaluation and Verification of Aerosol Diluters: Accuracy and Particle Loss
Terry Hoon Suk Jung
Master of Applied Science and Engineering 2014
Department of Mechanical and Industrial Engineering
University of Toronto
Abstract
The aerosol diluter characteristics of three different systems, the single-stage and the two-
stage TSI 379020A rotary disk thermodiluters and Dekati FPS-4000 ejector diluter, were tested
using gases and particles over a range of dilution ratios. The upstream and downstream gas and
particle concentrations of the diluters were measured in real-time to compute the actual dilution
ratio achieved by the three systems. Dilution ratios from approximately 15 to 100 were found to
fall within the expected operating error margin of ± 10% for CO2 and CH4. Dilution ratios covering
a similar range were also achieved to within ± 10% for particles with diameters from 9.3 to 200
nm. However, when engine exhaust was sampled, significant loss of particles smaller than 29.4
nm occurred during the dilution process. As the dilution ratio increased, the deviation from the
expected value increased due to an increase in measurement uncertainty.
iii
Acknowledgements
First and foremost I would like to thank Professor Wallace and Professor Evans for their
insightful guidance and support through this process. I would also like to extend my thanks to all
of the laboratory mates and staff in the Engine Research and Development Laboratory and
Southern Ontario Center for Atmospheric Aerosol Research, with a special thanks to The
Exhaust Measurement and Inhalation Toxicology Testing of Emerging Diesel Fuel study group
members – Naomi Zimmerman, Krystal Godri-Pollitt, Cheol-Heon Jeong and Josephine Cooper.
Finally, I would like to give my profound thanks to my parents, grandparents and brother
for their never ending support of all my pursuit in life. Without their support and encouragement,
this work may have never been completed.
iv
Table of Contents
Abstract ………………………………………………………………………………………….. ii
Acknowledgements ……………………………………………………………………………... iii
Table of Contents ………………………………………………………………………………... iv
List of Figures ………………………………………………………………………………….. viii
List of Tables …………………………………………………………………………………….. xi
List of Acronyms ……………………………………………………………………………….. xii
List of Symbols ………………………………………………………………………………… xiv
Chapter 1 Introduction ………………………………………………………………………... 1
1.1 Introduction …………………………………………………………………………... 1
1.2 What is an Aerosol? …………………………………………………………………... 2
1.3 Properties of Aerosol …………………………………………………………………. 3
1.3.1 Dynamics of Single Aerosol Particle …………………………………………… 4
1.3.2 Dynamics of Aerosol Populations ………………………………………………. 6
1.4 Why Study Automotive Aerosol? …………………………………………………….. 7
1.5 Other Related Concern: Visibility Degradation ………………………………………. 7
1.6 What is Aerosol Dilution? …………………………………………………………….. 8
1.7 Literature Review …………………………………………………………………… 12
1.7.1 Nucleation of Nanoparticles …………………………………………………... 12
1.7.2 Influence of Dilution Condition ……………………………………………….. 14
1.7.3 Ultrafine Diesel Exhaust Particles …………………………………………….. 16
1.7.4 Emission Regulations …………………………………………………………. 18
1.7.5 Diesel Engine Improvements ………………………………………………….. 18
v
1.7.6 Emission Control Technologies Improvements ……………………………….. 20
1.7.7 Summary ….…………………………………………………………………… 21
1.8 Objectives …………………………………………………………………………… 21
Chapter 2 Aerosol Diluter Review …………………………………………………………... 23
2.1 Introduction …………………………………………………………………………. 23
2.2 Description of Diluters ………………………………………………………………. 23
2.2.1 TSI 379020A Rotary Disk Thermodiluter ……………………………………... 23
2.1.2 Dekati FPS-4000 Ejector Diluter ……………………………………………… 25
2.3 Aerosol Diluter Systematic Comparison …………………………………………….. 27
2.4 Known Problems of Aerosol Diluters ……………………………………………….. 29
Chapter 3 Description of Test Instrumentation ……………………………………………… 32
3.1 Introduction …………………………………………………………………………. 32
3.2 Gas Concentration Measurement ……………………………………………………. 32
3.2.1 Heated Flame Ionization Detector …………………………………………….. 32
3.2.2 Non-dispersive Infrared Gas Analyzer ………………………………………… 34
3.3 Particle Number Measurement ……………………………………………………… 36
3.3.1 EEPS/FMPS …………………………………………………………………... 36
3.4 Soot Generator ………………………………………………………………………. 40
Chapter 4 Evaluation of Dilution in the Diluter Systems using the Gas Phase Species …….. 42
4.1 Introduction …………………………………………………………………………. 42
4.2 Dilution Evaluation Apparatus ……………………………………………………… 42
4.3 Dilution Evaluation Methods ………………………………………………………... 45
4.4 Sampling Time Parameters ………………………………………………………….. 48
4.5 Leakage Check ……………………………………………………………………… 48
4.6 Calculation Methods ………………………………………………………………… 49
vi
4.7 Results ………………………………………………………………………………. 50
4.7.1 Single-Stage TSI 379020A Rotary Disk Thermodiluter ……………………….. 50
4.7.2 Two-Stage TSI 379020A Rotary Disk Thermodiluter …………………………. 53
4.7.3 Dekati FPS-4000 Ejector Diluter ……………………………………………… 55
4.8 Error Analysis and Discussion of Results …………………………………………… 57
Chapter 5 Particle Loss in Aerosol Diluters …………………………………………………. 64
5.1 Introduction …………………………………………………………………………. 64
5.2 Particle Loss Measurement Apparatus ………………………………………………. 64
5.2.1 Diesel Emission Test Apparatus ……………………………………………….. 65
5.2.2 Soot Generator Test Apparatus ………………………………………………... 68
5.3 Particle Loss Measurement Methods ………………………………………………... 70
5.4 Sampling Time Parameters ………………………………………………………….. 72
5.5 Leakage Check ……………………………………………………………………… 72
5.6 Calculation Methods ………………………………………………………………… 73
5.7 Results ………………………………………………………………………………. 74
5.7.1 Single-Stage TSI 379020A Rotary Disk Thermodiluter ……………………….. 74
5.7.2 Two-Stage TSI 379020A Rotary Disk Thermodiluter …………………………. 84
5.7.3 Dekati FPS-4000 Ejector Diluter ……………………………………………… 92
5.8 Investigation of Heating and Thermal Conditioning Elements ……………………… 98
5.8.1 TSI 379020A Rotary Disk Thermodiluter ……………………………………... 99
5.8.2 Dekati FPS-4000 Ejector Diluter …………………………………………….. 103
5.9 Error Analysis and Discussion of Results ………………………………………….. 104
Chapter 6 Conclusions and Recommendations …………………………………………….. 110
6.1 Conclusions ………………………………………………………………………... 110
6.2 Recommendations …………………………………………………………………. 111
Appendices ……………………………………………………………………………………. 114
vii
Appendix A Preliminary Experiment Result for Single-Stage TSI 379020A
using Methane Gas ……………………………………………………….. 114
Appendix B Derivation of Uncertainty Analysis Equations …………………………… 116
References …………………………………………………………………………………….. 118
viii
List of Figures
Figure 1-1 Photomicrograph of Diesel Particulates: Cluster (Upper Left), Chain
(Upper Right), and Collection from Filter (Heywood, 1988) ………………….. 3
Figure 1-2 General Aerosol Dilution Process ………………………………………………… 9
Figure 2-1 Principle of Dilution Method for TSI 379020A (TSI, 2009) …………………….. 24
Figure 2-2 TSI 379020A Single and Two-Stage Layout (TSI, 2009) ………………………... 25
Figure 2-3 Principle of Dilution Method for Dekati FPS-4000 (Dekati, 2010) ……………... 27
Figure 3-1 Flame Ionization Detection Technology
(Rosemount Analytical NGA 2000 Manual) ………………………………….. 33
Figure 3-2 Non-dispersive Infrared Detector Technology (LI-COR, 2009) …………………. 36
Figure 3-3 Differential Mobility Analyzer Technology (TSI, 2004) ………………………… 37
Figure 3-4 EEPS Concentration Range (TSI, 2012) …………………………………………. 39
Figure 3-5 FMPS Concentration Range (TSI, 2004) ………………………………………… 39
Figure 3-6 Soot Generator Technology (Jing, 2009) ………………………………………… 40
Figure 3-7 Quenching Gas Flow Rate Dependent Soot Particle Size Distribution ………….. 41
Figure 4-1 Dilution Evaluation Apparatus for the TSI 379020A (Left)
and the Dekati FPS-4000 (Right) ……………………………………………... 43
Figure 4-2 Diluted CO2 Concentration Measurement at Various Dilution Ratios
for the Single-Stage TSI Diluter ………………………………………………. 51
Figure 4-3 Experimental Dilution Ratios for the Single-Stage TSI Diluter …………………. 52
Figure 4-4 Diluted CO2 Concentration Measurement at Various Dilution Ratios
for the Two-Stage TSI Diluter ………………………………………………… 53
Figure 4-5 Experimental Dilution Ratios for the Two-Stage TSI Diluter …………………… 54
Figure 4-6 Diluted CH4 Concentration Measurement at Various Dilution Ratios
for the Dekati Diluter ………………………………………………………….. 55
Figure 4-7 Experimental Dilution Ratios for the Dekati Diluter …………………………….. 57
Figure 5-1 Diesel Exhaust Particle Loss Test Apparatus for Diluters ……………………….. 66
Figure 5-2 Soot Generator Particle Loss Test Apparatus for Diluters ……………………….. 69
ix
Figure 5-3 EEPS/FMPS Equivalency Correction Factor (Zimmermann et al, 2013) ……….. 71
Figure 5-4 Engine Exhaust Particle Distribution at Various Dilution Ratio
for the Single-Stage TSI Diluter ………………………………………………. 75
Figure 5-5 Engine Exhaust Experiment Percent Particle Penetration
for the Single-Stage TSI Diluter ………………………………………………. 76
Figure 5-6 Engine Exhaust Experimental Dilution Ratios
for the Single-Stage TSI Diluter ………………………………………………. 80
Figure 5-7 Soot Generator Particle Distribution at Various Dilution Ratio
for the Single-Stage TSI Diluter ………………………………………………. 81
Figure 5-8 Soot Generator Experiment Average Percent Particle Penetration
for the Single-Stage TSI Diluter ………………………………………………. 82
Figure 5-9 Soot Generator Experimental Dilution Ratios
for the Single-Stage TSI Diluter ………………………………………………. 83
Figure 5-10 Engine Exhaust Particle Distribution at Various Dilution Ratio
for the Two-Stage TSI Diluter ………………………………………………… 85
Figure 5-11 Engine Exhaust Experiment Percent Particle Penetration
for the Two-Stage TSI Diluter ………………………………………………… 86
Figure 5-12 Engine Exhaust Experimental Dilution Ratios
for the Two-Stage TSI Diluter ………………………………………………… 87
Figure 5-13 Soot Generator Particle Distribution at Various Dilution Ratio
for the Two-Stage TSI Diluter ……………………………………………….... 88
Figure 5-14 Soot Generator Experiment Average Percent Particle Penetration
for the Two-Stage TSI Diluter ………………………………………………… 89
Figure 5-15 Soot Generator Experimental Dilution Ratios
for the Two-Stage TSI Diluter ………………………………………………… 90
Figure 5-16 Engine Exhaust Particle Distribution at Various Dilution Ratio
for the Dekati Diluter ………………………………………………………….. 92
Figure 5-17 Engine Exhaust Experiment Percent Particle Penetration
for the Dekati Diluter ………………………………………………………….. 93
x
Figure 5-18 Engine Exhaust Experimental Dilution Ratios
for the Dekati Diluter ………………………………………………………….. 94
Figure 5-19 Soot Generator Particle Distribution at Various Dilution Ratio
for the Dekati Diluter ………………………………………………………….. 95
Figure 5-20 Soot Generator Experiment Average Percent Particle Penetration
for the Dekati Diluter ………………………………………………………….. 96
Figure 5-21 Soot Generator Experimental Dilution Ratios
for the Dekati Diluter ………………………………………………………….. 97
Figure 5-22 Engine Exhaust Dilution Ratios Corresponding to
the Primary Dilution Air Temperature
for the Single and Two-stage TSI Diluters …………………………………….. 99
Figure 5-23 Engine Exhaust Percent Particle Penetration Corresponding to
the Evaporation Tube Heater Temperature
for the Two-Stage TSI Diluter ……………………………………………….. 100
Figure 5-24 Engine Exhaust Dilution Ratios Corresponding to
the Evaporation Tube Heater Temperature
for the Two-Stage TSI Diluter ……………………………………………….. 101
Figure 5-25 Engine Exhaust Percent Particle Penetration Corresponding to
the Evaporation Tube Heater Temperature
for the Two-Stage TSI Diluter ……………………………………………….. 102
Figure 5-26 Engine Exhaust Dilution Ratios Corresponding to
the Primary Dilution Air Temperature
for the Dekati Diluter ………………………………………………………… 103
Figure 5-27 Engine Exhaust Percent Particle Penetration Corresponding to
the Primary Dilution Air Temperature
for the Dekati Diluter ………………………………………………………… 104
Figure A-1 Diluted CH4 Concentration Measurement at Various Dilution Ratios
For the Single-stage TSI Diluter ……………………………………………... 115
xi
List of Tables
Table 2-1 Aerosol Diluters Specification Summary Table (TSI, 2009; Dekati, 2010) ………. 28
Table 2-2 Percent Particle Penetration Values for Varying Particle Sizes for the Diluters …... 30
Table 4-1 Table of Uncertainty Analysis for TSI Diluter Experimental Apparatus …………. 61
Table 4-2 Table of Uncertainty Analysis for Dekati Diluter Experimental Apparatus ………. 62
Table 5-1 Summary of Average Percent Dilution Ratio Difference
and Highest Deviation Point for the Single-Stage TSI Diluter ………………... 84
Table 5-2 Summary of Average Percent Dilution Ratio Difference
and Highest Deviation Point for the Two-Stage TSI Diluter ………………….. 91
Table 5-3 Summary of Average Percent Dilution Ratio Difference
and Highest Deviation Point for the Dekati Diluter …………………………… 98
Table 5-4 Table of Uncertainty Analysis for Experimental Apparatus ……………………... 107
xii
List of Acronyms
BTE Brake Thermal Efficiency
DHT Dilution Air Heating Temperature
DMA Differential Mobility Analyzer
DOC Diesel Oxidation Catalyst
DPF Diesel Particulate Filter
DPM Diesel Particulate Matter
DR Dilution Ratio
EEPS Engine Exhaust Particle Sizer
EGR Exhaust Gas Recirculation
EHT Evaporation Heater Temperature
EMITTED Exhaust Measurement and Inhalation Toxicology Testing of
Emerging Diesel Fuels
FMPS Fast Mobility Particle Sizer
GSA Gaseous Sulfuric Acid
HC Hydrocarbon
HDD Heavy-Duty Diesel
HEI Health Effects Institute
HFID Heated Flame Ionization Detector
LDD Light-Duty Diesel
LNT Lean NOx Trap
NDIR Non-Dispersive Infrared
NMP Nucleation Mode Particle
NTE Not-to-Exceed
OXICAT Oxidation Catalyst
PDR Particle Dilution Ratio
PDT Primary Dilution Temperature
PM Particulate Matter
xiii
PMP Particulate Measurement Program
PRDR Primary Dilution Ratio
PRH Primary Dilution Relative Humidity
RR Relative Risk
RT Residence Time
SCR Selective Catalytic Redcution
SI Spark-Ignition
SVI Sulfur VI
VDR Volumetric Dilution Ratio
xiv
List of Symbols
∝ Related to
ρ Density
≈ Approximately equal to
~ Approximately
��� Potentiometer dial number of the primary and secondary flow settings
for TSI 379020A diluter
���� Diameter of particles
��� Dilution ratio of sample ������� Average dilution ratio of sample ��� � Rate of carbon atom input
� Electrical current
�� � Ionization rate of hydrocarbon
∑�� Sum of mass of particles �� Concentration of sample ��� Average concentration of sample ∑�� Sum of concentration of particles
� �� � �� Loss/Conversion rate of particle number
� Particle penetration value
�� Particle number
�� Volumetric flow of sample �� � Volumetric flow rate of sample �� Variance of the average measured value of sample � ! Standard deviation for any function R
[$%&] Total hydrocarbon concentration
( Volume
1
Chapter 1
Introduction
1.1 Introduction
Diesel and gasoline fuel combustion is a major source of atmospheric air contaminants.
Further, a range of adverse human health effects are associated with particulate matter (PM) from
engine emissions. Numerous studies report that long term human exposure to PM results in
increased risk of cardiac ischemia and arrhythmias, increased blood pressure, decreased heart rate
variability, and increased circulating markers of thrombosis and inflammation (Froines, 2005).
Ultrafine particles are categorized as the particles smaller than 100 nm in diameter. Inhalation and
exposure to ultrafine particles, especially to the volatile fraction of such particles, is hypothesized
to induce acute cardiopulmonary responses and compromise cardiorespiratory status. Existing
regulations cover particles > 2.5 µm in diameter (PM2.5). Compared to the PM2.5 standard, the
mass of these ultrafine particles is relatively small and yet they are more effective in transporting
hazardous substances absorbed on their surface due to their high surface area to mass ratio
compared to larger particles (Stoeger, 2006; Oberdörster, 2000).
Studies of air pollution exposure and mortality suggest that the cardiopulmonary mortality
is associated with living near a high traffic density area. People living within 100 m of a freeway
or 50 m of a major urban road had their cardiopulmonary mortality relative risk (RR) value
increased to 1.95 (Hoek et al, 2002). The RR value is a ratio of the probability of the event
occurring in the exposed people compared to non-exposed. Hence the RR value of 1.95 indicates
that people living near a high traffic density area have 1.95 times higher risk of cardiopulmonary
2
mortality. In addition, wheezing in school children living within 150 m of a main road increases
RR to 1.08 (Venn et al, 2001).
Awareness of the potential risk of ultrafine particles leading to adverse effects is increasing.
Studies pertaining to ultrafine particles are valuable in order to characterize the effectiveness of
current automotive combustion and emission control technologies, and to address adverse health
related concerns.
1.2 What is an Aerosol?
An aerosol is a gaseous suspension of fine and liquid particles. It can be a typical cloud in
the sky or undesired air pollutions such as automotive exhaust and smog. Under typical
atmospheric conditions, the sizes of aerosol particles measured vary with different contributing
factors for specific size ranges. For instance, human activities, including combustion processes,
contribute primarily to the submicron size range. The aerosol from primary natural sources, such
as sea salt and soil dust, is concentrated in the size range larger than 1.0 µm diameter. Also, the
mode in the volume distribution that occurs in the 0.1 to 1.0 µm diameter range results from the
growth of particles by gas-phase chemical reactions with condensable products (gas-to-particle
conversion). Regardless of where the aerosol comes from, exposure to these particles is
unavoidable. Therefore the surrounding atmospheric particles are studied to understand and define
what is breathed in.
The particles in diesel engine exhaust are referred to as the organic and inorganic species
that can exist in both liquid, such as hydrocarbons, water, and sulfuric acid, and solid, such as
elemental carbon and ash. Also some droplets of liquid content can exist, which can provide
coating for the solid particles. Representative images of diesel particulates are shown in Figure 1-
3
1, which includes photomicrographs of diesel particulates in a cluster, in a chain, and a selection
collected from a filter.
Figure 1-1
Photomicrographs of Diesel Particulates: Cluster (Upper Left), Chain (Upper Right), and
Collection from Filter (Heywood, 1988)
1.3 Properties of Aerosol
The study of aerosol physico-chemical properties is crucial in explaining why certain
particles behave in a particular way and how they are formed. The particles can undergo various
chemical or physical processes as they move from one place to another. Such knowledge aids in
4
determining the fate of the pre-existing and emitted particles in the atmosphere. The overview of
aerosol properties in this section is based on the textbook by Senifeld and Pandis (2006).
1.3.1 Dynamics of Single Aerosol Particle
The mean free path of a gas molecule can be defined as the average distance travelled
between collisions with other gas molecules, The mean free path of an aerosol particle is difficult
to define because the aerosol particles collide only very infrequently with other particles. In a case
where the collision actually occurs, it is usually assumed that the two particles adhere to each other.
The aerosol particles have large size and mass relative to gas molecules. Thus, they experience a
large number of collisions per unit time with the surrounding gas molecules and are not influenced
significantly by any one collision. Consequently, the motion of an aerosol particle can be seen as
continuous in nature.
In the elementary kinetic theory of gases, transport properties such as viscosity, thermal
conductivity, and molecular diffusivity are related to the mean free path by the flux of gas
molecules across planes separated by a distance. Therefore, the diffusional mean free path is
defined in terms of the molecular diffusivity of the vapor and its mean speed.
The most important concept that explains the dynamics of aerosol particles is Brownian
motion. Brownian motion refers to the particles suspended in a fluid undergoing irregular random
motion due to bombardment by surrounding fluid molecules. It can be described as a diffusion
process. A particle of sufficiently large size, approximately 100 µm, would experience only the
drag force and its motion would be unaffected by the molecular bombardment. However, if the
diameter of the particle is continually reduced, the fluctuations in its motion due to molecular
bombardment become increasingly noticeable and Brownian motion becomes dominant. As the
5
particle gets smaller, the force associated with Brownian motion must be considered in addition to
gravitational and drag forces. To further explain, Brownian diffusion is comparable to gravitational
settling in terms of the distance that a particle travels in both cases (Seinfeld and Pandis, 2006):
1. Over a period of 1 s, a 1 µm radius particle diffuses a distance of 4 µm, while it falls
about 200 µm under gravity.
2. Under the same time period, a 0.1 µm radius particle diffuses a distance of 20 µm
compared to a distance of 4 µm due to gravity.
The temperature of a Brownian particle suspended in a fluid is the same as that of the fluid,
but the kinetic energy of its motion must be determined from the kinetic energy of the molecules
of the fluid. The transfer of kinetic energy to the Brownian particle must be accompanied by a
local cooling of the fluid. Thus, small random fluctuations in temperature about the equilibrium
temperature always exist.
Phoretic effects produce a directional preference in the Brownian diffusion of aerosol
particles due to a difference in momentum imparted to a particle by molecules coming from
different directions. The directional preference depends on the local gradient of molecular
momentum caused by a difference in energy, or velocity, or by differences in mass. Phoretic effects
in the Brownian diffusion can be categorized into three types. Thermopheresis is the particle
motion caused by the higher energy molecules on one side of the particle due to a macroscopic
temperature gradient. This result causes the aerosol particles to diffuse away from warmer regions
toward cooler regions. Photophoresis results when incident radiation heats one side of the particle
more than the other, leading to differences in the energies of gas molecules adjacent to the surface
of the particle. Diffusiophoresis occurs in the presence of a gradient of vapor molecules that are
6
either lighter or heavier than air molecules due to a balance in the directional fluxes between the
two molecules.
1.3.2 Dynamics of Aerosol Populations
The aerosol particles suspended in a fluid may come into contact because of their Brownian
motion. They evolve in size by coagulation and gas-to-particle conversion or condensation. It has
been found that 0.01 µm to 1.0 µm particles grow principally by gas-to-particle conversion, the
process by which vapor molecules diffuse to the surface of a particle and subsequently are
incorporated into the particle. The rate-controlling step in condensation may be a result of one or
a combination of these three mechanisms (Seinfeld and Pandis, 2006):
1. The rate of diffusion of the vapor molecule to the surface of the particle (diffusion-
controlled growth).
2. The rate of a surface reaction involving the absorbed vapor molecule and the particle
surface (surface reaction-controlled growth).
3. The rate of a reaction involving the dissolved species occurring uniformly throughout
the volume of the particle (volume reaction-controlled growth).
The understanding of Brownian motion and condensation mechanisms allow us to describe
how the aerosol particles will behave when subjected to varying conditions. The aerosol particles
found in the atmosphere can be very small, which makes them difficult to trace and analyze. As
such, when studying these particles, it is important to recognize the governing principles to further
identify why a certain phenomenon has occurred.
7
1.4 Why Study Automotive Aerosol?
The exhaust particles emitted from diesel and spark-ignition (SI) engines are of concern to
engine builders for their influence on engine performance and wear. Not only that, but the focus
has been shifted towards the health and environmental impacts that these engine-emitted particles
induce, more so for diesel than SI engines.
The inhalation of air is a natural process to provide oxygen throughout the body. Thus,
breathing of the surrounding atmospheric air is inevitable. Studies have shown that PM can induce
health related problems, not to mention that pollutants can cause serious environmental issues.
Thus it is important to characterize aerosol in automotive emissions to help avert adverse effects,
whether on health or the environment.
1.5 Other Related Concern: Visibility Degradation
One concern arising from automotive aerosol emissions is visibility reduction. The
prevailing visibility is defined as the greatest distance in a given direction for which it is just
possible to see and identify an object. Absorption and light scattering are the two effects that
aerosol particles have on visible radiation which causes the visibility reduction. Light scattering is
usually the more important phenomenon responsible for visibility impairment. Visibility is reduced
due to significant scattering of light into the line of sight, which decreases the contrast between the
object and the background sky. The scattering by PM of sizes comparable to the wavelength of
visible light is responsible for visibility reduction. Particles in the range of 0.1 to 1 µm in radius
are the most effective per unit mass. The scattering coefficient is directly dependent on the
atmospheric aerosol concentration in this range. Scattering by particles causes 60 to 95 percent of
visibility reduction. The chemical species sulfates, nitrates, and organics are the main contributors
8
to the light-scattering coefficient. Sulfate, SO42- is often the most important scattering material,
followed by organic carbon.
In many cases, light absorption by black carbon particles is a significant contributor to
visibility reduction. Black carbon is more effective than non-absorbing aerosol particles in
attenuating light. Absorption by soot particles causes 5 to 40 percent of visibility reduction. Soot
is about three times more efficient than SO42-, NO3
-, or organics in terms of visibility reduction per
unit mass of airborne particles (Seinfeld and Pandis, 2006).
1.6 What is Aerosol Dilution?
Before the composition and concentration of the raw PM from an engine can be studied,
the exhaust must be diluted by a known factor in order to cool the exhaust and reduce the
concentrations to values within the ranges of the instruments available. The PM concentration
released from engine exhaust is orders of magnitude higher than the mixture (of air and exhaust)
found in the atmosphere. The process of dilution is a simple act to reduce this raw amount by a
known ratio in order to obtain a condition more representative of the air we breathe. In certain
cases, dilution is performed to reduce the concentration below the maximum detection capacities
of the particle analyzers. In automotive studies, a dilution process is essential in all parts of
emissions experiments. Therefore, it is crucial to identify the operating behavior of the dilution
system prior to carrying out an extensive experimental matrix. In a general dilution process, the
number of particles per unit volume in the given raw sample stream is reduced by introducing a
particle-free air flow.
The generalized premise of a dilution process is described in Figure 1-2. The volumetric
flowrate from the upstream, Qupstream (volume/unit time), with concentration Nupstream (particle
9
number/unit volume) is mixed with the dilution flowrate, Qdilution, with concentration Ndilution at a
certain ratio. During the dilution phase, associated size dependent particle loss,
������ ���,�� ����� (particle number/unit time) occurs at some rate where PN denotes the particle
number. The outflow from the process, Qdownstream, with concentration Ndownstream is the diluted
sample with a defined dilution ratio (DR).
Figure 1-2
General Aerosol Dilution Process
Assuming constant pressure and temperature and density, the conservation of particle
number and the volumetric flow rate equations of the flow in Figure 1-2 are described as (Collins,
2010)
����� ��,�� ���������� �� � ���������,�� �������������� � ������ �����, ���
������� ���,�� ����� � ������� ��,�� ������������ �� Equation 1-1
����� �� � ��������� � ������� �� Equation 1-2
10
Using the assumption that the dilution flow is particle-free, ���������,�� ����� � 0, and
neglecting the effect of coagulation, evaporation and associated losses such as wall deposition and
diffusion, ������ �����, ��� � ������ ���,�� ����� � 0, the appropriate DR can be described as
!��� � �"#$%&'()�*+,-$&'() �
."#$%&'()/.*01"%0+-."#%&'() Equation 1-3
This equation is defined as the volumetric DR of the process since it is dependent only on the
volumetric flow with zero particle loss – other terms are assumed negligible.
Similarly, the particle size dependent DR, defined as the particle DR, can be generalized
as
!������� ,�� ����� � �"#$%&'(),$02'3*#-*�*+,-$&'(),$02'3*#-* Equation 1-4
However in case where a loss is present such that ������ ���,�� ����� ≠ 0 but all other
assumptions remain valid, ������ �����, ��� � ���������,�� ����� � 0, then Equation 1-1 can be
rewritten as
1 ������� ���,�� ���������� ��,�� ���������� �� � ������� ��,�� ���������� ��,�� ����� 6����� �� � �������������� �� 7
Equation 1-5
Using the previous Equations 1-3 and 1-4, the above expression can be simplified to
11
1 ������� ���,�� ���������� ��,�� ���������� �� � !���,�� ����� !������� ,�� ����� � 8�� �����
Equation 1-6
where 8�� ����� is defined as the (diluter) penetration number (Collins, 2010). The
particle penetration number quantifies the fraction of particles lost during the dilution process. In
this expression, unity is defined as the condition where there are no particle losses present in the
system. This is referred to as an “ideal dilution”, meaning that a particle DR equal to that of the
volumetric DR has occurred in all particle size distributions. It should be noted that such a
condition is most unlikely to occur in a typical operating condition. Frequently, the value of P will
be lower or greater than unity. The case where the value of P falls below unity describes a system
in which particle loss is present, where the DRparticle is greater than the DRvol. In contrast, when the
value of P rise above unity, the system is generating particles when DRparticle is less than the DRvol.
Apparent particle generation in the system might occur due to condensation on particles in a certain
particle size range during the dilution process. The resulting particle growth would cause a shift of
size distribution causing a reduction of particle number in smaller size bins and an increase in
larger size bins.
It is important to note that such particle generation or loss present within the dilution
system can result in a skewed outcome when measuring the concentration of the raw exhaust. In
emission testing and aftertreatment technology development, it is crucial to discern the mode
particle size. Consequently, a thorough investigation of the operating diluter is necessary to avoid
any misrepresentation of the data.
12
1.7 Literature Review
The chemical and physical properties of diesel particulate matter (DPM) can vary for
different fuels and emission control technologies. Many studies have characterized the influences
of biodiesel fuel and the aftertreatment systems, but further investigation is still required. Such
technologies include the diesel particulate filter (DPF) for removal of PM or soot from the diesel
emissions and selective catalytic reduction (SCR) systems for converting NOx into diatomic
nitrogen and water. In all of these research areas, it should be noted that an appropriate aerosol
dilution technology is necessary to make accurate measurements of system performance. Thus, in
order to provide context for PM measurement requirements, the underlying relevant research
papers from the literature are reviewed to provide an overview of the current state of these
technologies.
This review provides an overview of the basic characteristics of nucleation of nanoparticles,
the influence of dilution condition on particle number measurement, the investigation of ultrafine
diesel exhaust particles, diesel emission regulations, and the improvements in diesel emission
control strategies. Unless stated otherwise, the literature reviewed in this section relates to diesel
engines (Note that gasoline direct injection engines also produce particles).
1.7.1 Nucleation of Nanoparticles
Kittelson (1998) describes diesel exhaust particles as consisting mainly of highly
agglomerated solid carbonaceous material and volatile organic and sulfur compounds. The
particles are classified into different modes depending on the particle diameter: accumulation and
nuclei mode. Most particle mass exists in the accumulation mode range from 50 -500 nm in
diameter. In contrast only 5-20% of the particle mass is contributed by the nuclei mode, D < 50
13
nm. However, more than 90% of the total particles by number, exist in this mode.
Studies by many institutes and research centers have raised concerns about the increase of
nanoparticle emissions from new diesel engine technologies (Kittelson 2001, Heikkilä et al. 2009,
Park et al. 2009). Although the newly developed engine reduce particle mass emissions, the
number emissions of these nanoparticles sharply increases. Kittelson (1998) postulated that
particle formation by future engines would rely on the relative amount of condensable species and
solid surface area on which the vapor species can condense or absorb. Furthermore, the primary
composition of these nanoparticles must be characterized, because if these particles are volatile,
they may have different health impacts than solid particles.
The nanoparticle number emissions from diesel engines are mostly found in the nucleation
mode. Exhaust gas cooling and dilution in the atmosphere induces nucleation, which leads to the
formation of these nanoparticles. Concerns about nanoparticles have been raised because of their
adverse human health effects. There have been numerous studies done to explain the nanoparticle
nucleation mechanism.
Some authors suspect ion-induced nucleation as the mechanism describing the formation
of nucleation mode particles (NMPs) (Yu et al., 2001, 2002, 2004; Ma et al., 2008). However,
Collings et al. (1988) and Moon (1984) found that particles in the nucleation mode had little or no
charge. Jung et al. (2005) confirmed that diesel NMPs carry little or no electrical charge and
concluded that ion-induced nucleation is not the primary mechanism for the nucleation. Ma et al.
(2008) also concluded that an ion trap prior to dilution did not significantly influence nucleation
mode formation due to the low ion concentration. Conversely, Yu (2001) proposed that the
charging occurred from the attachment of chemi-ions, produced from the combustion process
14
under high temperature conditions. Yu (2001) claimed that the ion induced nucleation theory
agrees well with measurements in terms of total nanoparticle concentration. This claim was further
supported by Yu et al. (2004) through measurement of ion concentrations in diesel exhaust.
Conclusions arising from studies pertaining to particle nucleation, as suggested, are still
unclear. Many studies seem to contradict or support the other findings but there are no definite
conclusions as to how these particles nucleate. Thus, particles and their physicochemical
mechanisms are a subject of increased interest for many researchers.
1.7.2 Influence of Dilution Conditions
Many studies have focused on the measurement of engine-out emissions and the influence
of the various schemes for reducing emissions. In many cases, the sampling and dilution systems
are not clearly defined. Consequently, extreme difficulties are encountered in interpretation of
measurement results because the concentration of ultrafine and nanoparticles are strongly
influenced by dilution. Thus, variations in the measured concentration of such particles are evident
between the different dilution systems.
The typical diesel-emitted emissions are conducted in either two standard test setups:
engine dynamometer and chassis dynamometer setups. The dynamometer is a device for
measuring force, torque and power of an engine. In an engine dynamometer layout, the engine is
removed from the vehicle and direct emissions from the engine can be studied. A chassis
dynamometer is used to simulate road loading conditions and the tailpipe emissions of a fully
functional vehicle can be studied. In most cases reviewed here, light-duty diesel emissions were
studied using an engine dynamometer setup and heavy-duty diesel emissions were studied using a
chassis dynamometer. This is in contrast to the regulatory emissions tests, where light duty vehicles
15
are tested on a chassis dynamometer and heavy duty engines are tested on an engine dynamometer.
Numerous studies have been conducted measuring the NMPs under real-world dilution
conditions in the vehicle exhaust plume (Kittelson 2002; Vogt et al. 2003). Mohr et al (2003)
proposed that the repeatable measurements could only be achieved through a dilution scheme that
does not favor formation of NMPs and is not sensitive to small changes of dilution parameters.
However, Mathis et al. (2004) proposed that the accumulation-mode particles and NMPs should
not be disregarded for a comprehensive characterization of particulate emission. Hence, strict
requirements and conditions should be suggested for the exhaust dilution process. The main
influential dilution parameters, which have been investigated in several studies (Kittelson 1998;
Abdul-Khalek et al., 1999; Shi et al., 1999; Khalek et al., 2000; Mathis 2002; Ntziachristos et al.,
2004a), are primary dilution temperature (PDT), primary dilution ratio (PRDR), residence time
(RT), and primary dilution relative humidity (PRH).
Abdul-Khalek et al. (1999) concluded that the accumulation mode was stable and was not
influenced by the dilution conditions. On the other hand, the nuclei mode was highly sensitive to
PDT and PRH (Mathis et al., 2004). They concluded that a decrease in PDT or an increase in PRH
initiated the formation of NMPs and consequently increased the number concentration of NMPs.
Lowering of the PDT favored the formation and growth of NMPs due to possible higher partial
vapor pressures and the reaction of volatile compounds during nucleation. Abdul-Khalek et al.
(1999) found that at low PRDR and low PDT, the influence of the RT is the strongest and the
nucleation mode concentration was at highest; vice versa at the opposite conditions. Additionally,
Collings et al. (2000) proposed that the total number of NMPs produced is extremely non-linear
and highly sensitive to PDT. However, when the concentration of accumulation mode is
16
sufficiently large, the formation of NMPs can be suppressed, and the total number of NMP
concentration cannot be associated to the influence of dilution condition (Kittelson 1998; Mohr et
al., 2001).
From the result of Mathis et al. (2004), it was determined that a PRDR between 20 and 30
produced the highest NMPs number concentration. A number reduction and decreasing NMP size
was observed at a PRDR above 30 due to the reduction in partial vapor pressures. A number
reduction was observed at a PRDR below 20 due to coagulation. Size growth was induced from
condensation of organic compounds on NMPs and acid catalyzed particle-phase heterogeneous
reactions. In addition, the highest NMP volume concentration was found at a low PRDR because
of the size growth.
Abdul-Khalek et al. (1999) indicated that the increase in PRH at constant DT led to an
increase in number concentrations by 30%. Mathis et al. (2004) however, found that the increase
in PRH resulted in a shift of the particle number size distribution toward larger diameter and a
decrease in the particle number concentration, due to coagulation of NMPs and possible interaction
between particles.
Aside from the sampling parameters indicated, a significant effect of engine cooling from
the local airstream speed around the vehicle was observed. With increased airstream speed,
nucleation mode was decreased (Mathis et al., 2004). Lastly, the utilization of ultra low sulfur fuel
resulted in a 70% decrease in number emissions (Abdul-Khalek et al., 1999).
1.7.3 Ultrafine Diesel Exhaust Particles
Commonly, an oxidation catalyst (OXICAT) is used to treat engine exhaust (aftertreatment)
17
to control DPM emissions of the diesel engine. Although the OXICAT reduces the concentration
of accumulation mode particles to undetectable levels, it increases emissions of ultrafine particles
in the nucleation mode (Kittelson et al., 2006). Gaseous sulfuric acid (GSA) is the most important
nucleating gas present in modern diesel vehicle exhaust. The presence of GSA triggers the
formation of new aerosol particles and growth by condensation and coagulation (Arnold et al.,
2006).
The sulfur VI (SVI) emission increases with increasing fuel sulfur mass fraction and
fraction F (fraction of fuel-sulfur conversion to SVI) of SO2 converted to SO3. Subsequently, F
increases in an OXICAT with high exhaust temperatures, through conversion of SO2 to SO3 and
H2SO4 (Arnold et al., 2006; Grose et al., 2006). Additionally, the absence of soot induces the build
up of a larger GSA super-saturation, which tends to ultimately increase GSA nucleation leading to
the formation of fresh nanoparticles. The atmospheric residence time of these nanoparticles
increases with increasing particle diameter and decreases with increasing larger particle
concentration (Arnold et al., 2006).
The results of Grose et al. (2006) showed that the diesel particles behaved in a manner
similar to ammonium bisulfate. But their observation indicated that a fraction of pure sulfuric acid
particles became neutralized by presumed ammonia contamination of the apparatus. Nonetheless,
12 nm particles all behaved like the neutralized sulfate and no additive volatile particles were found
in comparison to 12 nm sulfuric acid. Thus the observed results confirmed that the diesel exhaust
particles were neutralized to some extent, and the main composition of the particles was sulfuric
acid.
Lastly, the review of Sakurai et al. (2003) and Cooper et al. (1989) indicated that the
18
OXICAT effectively oxidized the high molecular weight semi-volatile organics (e.g. hydrocarbons;
HCs) and induced the conversion of SO2 to sulfate. Herner et al. (2007) confirmed that the removal
of gas phase HCs by the OXICAT reduced the formation of HC NMPs, or prevented the growth of
sulfate 1 nm particles to detectable size. It was predicted that the unexpected nucleation events
could have occurred due to storage or release of materials in the constant volume sampling system.
The conclusion was drawn that even with the use of ultra low sulfur diesel fuel to enhance the
formation of a nucleation mode, considerable amount of sulfates were in existent. The growth of
nucleated 1 nm sulfate particles to a detectable size was indicated as a function of availability of
additional condensates, notably HCs. Therefore, the existence of the ultrafine particles in the
nucleation mode from the diesel exhaust should be recognized and regulated for further reference.
1.7.4 Emission Regulations
In 2007, the finalization of the light-duty diesel (LDD) Euro 5 and 6 emissions standards
occurred in Europe. Euro 5 regulates NOx emission control measures to 180 mg/km and PM to 5
mg/km; and Euro 6 regulates NOx to 80 mg/km and PM to 4.5 mg/km with a particle number
standard of 6 × 1011/km that are determined using the UN/ECE Particulate Measurement Program
(PMP). The European Commission proposed Euro VI heavy-duty diesel (HDD) standards of 400
mg/kW-hr NOx and PM standards of 10 mg/kW-hr as measured on the European steady state and
transient cycles. In addition to criteria pollutant standards, the first standards on CO2 emission
limits at 130 g/km were proposed.
1.7.5 Diesel Engine Improvements
In general, the diesel engine developers are responding to the change in regulations by
using advanced fuel injection technologies, better exhaust gas recirculation (EGR) control as well
19
as EGR cooling, advanced and two-stage turbocharging, variable valve actuation, close loop
combustion control, and advanced model-based control for LDD (Johnson, 2007; 2008). It was
reported that the Euro 5 regulations could be satisfied with DPF without NOx aftertreatment. In
contrast, NOx aftertreatment is needed to meet Euro 6. However, Akmadza (2007) reported that
going from a Euro 5 calibration to a Euro 6 calibration results in a nominal 7% fuel penalty by
using NOx aftertreatment. It was pointed out that the choice between additional engine
technologies and NOx aftertreatment would need to balance costs and fuel economy benefits
(Johnson, 2008).
HDD engine developments are primarily aimed at improved fuel economy, reliability cost,
and durability. Hence, advancements are conservative and incremental. HDD engine researches
are more focused on traditional diesel combustion but some advanced combustion strategies have
emerged to meet the US 2010 Not-to-Exceed (NTE) in-use emissions limits (Johnson, 2007; 2008).
The incremental technologies of the research engines showed 10% fuel saving with significant
reductions in NOx (Dellmeyer et al., 2007). These engines showed brake thermal efficiency (BTE)
of 47%, and the potential to hit 53% BTE by 2013 (Shanton, 2007). Where BTE refers to the
efficiency of an engine in converting the heat from a fuel to the mechanical energy.
Dreisback (2007) examined the non-road engine technologies to attain the interim Tier 4
and final Tier 4 emission levels. The EGR and DPF were capable of hitting the interim levels, but
to meet the final level further development was needed. These developments include 2-stage
turbocharging; increasing EGR, cooling, and control; high-pressure flexible fuel injection; and
premixed or low temperature combustion strategies. Nishimura (2007) and Signer (2007) however,
proposed only DPFs as the non-road DPM reduction technology solution.
20
1.7.6 Emission Control Technologies Improvements
For the purpose of PM control, DPF technology was reported to be in a state of
optimization and cost reduction (Johnson 2007; 2008). Although SiC has been the main filter
material for LDD, the advancement in technology enabled utilization of alternative materials, such
as cordierite (Pidria et al., 2006; Craig et al., 2005; Maramatsu et al., 2006). Fischer et al. (2006)
showed that the fuel penalties due to DPF regeneration are minimized due to much lower thermal
conductivity for the cordierite. Moreover, the new DPF regeneration strategies pertaining to
control, in-cylinder injection, the fundamentals of how soot interacts with the catalyst, and the
impact of DPF pore structure were examined (Ootalke et al., 2007; Parks et al., 2007). Lastly, a
new DPF substrate material, aluminum titanate, was also studied (Ingram-Ogunwami et al., 2007).
Improvements in NOx control were focusing on SCR for diverse applications. SCR
efficiencies reached up to 90% with better mixing and control (Johnson, 2007). Low temperature
deNOx efficiency was addressed with better understanding on limitations. Solid urea and gaseous
ammonia storage in magnesium dichloride were suggested as substitutes for liquid urea (Mueller,
2007; Johanssen, 2007). Moreover, the suggested solution to meet LDD requirements of nominally
65% reduction was to use a lean NOx trap (LNT). A LNT refers to a NOx absorber which is
designed to reduce the oxides in the lean burn (combustion with excess in air) engine. Newly
developed SCR catalysts are emerging with improved durability and better ammonia slip
conversions, such that the aged LNTs are effective up to 60-70 % deNOx efficiency (Hu et al.,
2006). Also, the combination of LNT and SCR systems have been reported to convert NOx on the
LNT to ammonia during the rich regeneration, which in turn is stored on the SCR catalysis for
additional NOx removal during the lean treatment (Lambert et al., 2005). The LNT durability and
21
performance was enhanced with a sulfur trap (Yoshida et al., 2007). In addition, lean NOx catalysts
were significantly improved with a double layer concept utilizing ammonia generated in a NOx
absorber material (Satoh et al., 2006; Wada et al., 2007; Morta et al., 2007).
Finally, studies on diesel oxidation catalysts (DOCs) showed that the hydrocarbon
emissions from a low temperature combustion engine are difficult to treat potentially due to the
class of HC generated (Knafl et al., 2007). Noack et al. (2007) and Katare et al. (2007) also
proposed that NO2 was not formed at temperatures below the light-off temperature of HC and the
CO. NO2 was consumed as long as HC and CO were present in the exhaust. In addition, Punke et
al. (2006) reported a method of incorporating the DOC function onto the DPF.
1.7.7 Summary
Building upon these reviews, the important parameters that should be noted for the
measurement and collection of PM emissions data can be determined. Such parameters include:
the particle measurements relating the effects of dilution conditions; the particle size distribution
of diesel particulates; and the expected particle concentrations and the particulate composition.
Knowing the basics along with these parameters will further help to understand the behavior of the
diesel emissions. Better understanding will lead to improvements in diesel emission control
strategies. Finally, the literature review documented above will support generalizing observations
made through the study and will be used to derive conclusions or validate the collected data.
1.8 Objectives
At this point, the importance of a dilution system in emission testing and aftertreatment
technology development is clear. Investigating the dilution system behavior in terms of loss is then
22
necessary to prevent misrepresentation of the sample measurement. Thus, the primary goal of this
thesis was to verify the validity of the dilution systems – mainly to identify the portion of
concentration and particle loss during the dilution process. In order to quantify such loss during
the dilution process, experiments were conducted with base gases and particles. As indicated by
the Equation 1-6, the loss within the dilution system (or penetration number) can be described in
terms of the volumetric and particle dilution ratios. Therefore the body of work done to investigate
the behavior of diluters for both gases and particles was crucial. The secondary objective was to
validate the functionality of the dilution system features, such as heating and thermal conditioning.
From this portion of work, conclusions could be derived regarding the optimal configuration for
operating under typical exhaust testing conditions.
23
Chapter 2
Aerosol Diluter Review
2.1 Introduction
As discussed in Chapter 1, an aerosol diluter is used to lower the concentration of the
engine exhaust to simulate near-atmospheric conditions and to reduce the concentration below the
maximum detection capacities of the particle analyzers. In a general dilution process, a particle-
free air flow is introduced into a given raw sample stream to reduce the number of particles per
unit volume. Validation and analysis of the dilution system is essential in all parts of engine
emission studies. Particle loss from dilution may result in under estimation of the emitted particle
number in certain size bins. Thus, any conclusions drawn from tests where unknown loss occur
can be very misleading.
2.2 Description of Diluters
Two types of diluters will be considered in this thesis: a rotary disk diluter and an ejector
diluter. The systems reviewed and compared are single-stage and two-stage TSI 379020A rotary
disk thermodiluters and a Dekati FPS-4000 ejector diluter.
2.2.1 TSI 379020A Rotary Disk Thermodiluter
The TSI 379020A is a rotating disk type diluter. A schematic is shown in Figure 2-1. The
diluter construction consists of a rotating disk having 10 small cavities placed in a dilution block
with heating elements. A portion of the raw sample is pumped in at approximately 1.0 lpm and is
captured by the cavities of the rotating disk and mixed with HEPA-filtered, particle-free dilution
24
air in the measurement channel. The dilution air flow is controlled by an internal pump over a
dilution factor range of 0 – 10. This dilution air can be heated up to 150 °C along with the diluter
block, which allows the removal of condensed volatile materials. The dilution ratio is a linear
function of the disk rotation frequency and the dilution air flow rate as shown in Equation 2-1.
DR � <�������=���>������?�����@<�AB�������?� C� ��D<�������E��?���B�� Equation 2-1
The flow rate of diluted sample available for measurement ranges from 0.5 – 5 lpm.
Figure 2-1
Principle of Dilution Method for TSI 379020A (TSI, 2009)
Additional features of the two-stage TSI 379020A include a thermal conditioner and an
internal air supply to the built-in single stage as shown in Figure 2-2. The supply air to the primary
dilution is controlled to 1.5 lpm. The thermal conditioner consists of an evaporation tube, which
allows the elimination of the condensed nanodroplets and prevents volatile particle formation. The
temperature of this tube can be adjusted up to 400 °C but is normally operated at 300°C, as
25
recommended by the TSI manual. Secondary dilution is achieved in a mixing assembly through
adjustment of a calibrated dilution airflow over a range from 0 – 15 lpm, which corresponds to a
dilution factor range of 1 – 11. The flow rate of the diluted sample exiting the system and available
for measurement, ranges up to 16.5 lpm.
Figure 2-2
TSI 379020A Single and Two-Stage Layout (TSI, 2009)
2.2.2 Dekati FPS-4000 Ejector Diluter
The Dekati FPS-4000 is an ejector type diluter. There are two stages to the operation of
the Dekati FPS-4000 as shown in Figure 2-3. In the primary stage, the filtered, particle-free dilution
air flows into the inner tube of the probe through perforated walls. Introduction of the dilution air
through entire length of the tube minimizes particle loss. In addition, primary dilution air and the
probe itself can be heated for the purpose of evaporating the condensed volatile material. In the
second stage, the primary diluted sample is drawn through the nozzle. The nozzle causes a high
26
velocity flow and the resulting pressure drop creates suction inside the tube. The DRs for both
stages are governed by the diluter dimensions, the dilution airflow, the sample temperature, and
the pressure drop in the sample input and the ejector nozzle. These parameters are pre-calibrated
by the manufacturer, controlled, and monitored by the control unit in real-time to calculate the
spontaneous DR within the system. With proper venting, the diluted sample exiting the second
stage is always at ambient pressure regardless of the initial condition.
Dilution air is provided to the Dekati as compressed air at a minimum pressure of 6 bar
and a maximum pressure of 9 bar. The compressed air must have a particle concentration of less
than 100 particle/cc and a very low relative humidity such that the air is non-condensing at -40 oC.
An internal pressure regulator maintains a constant pressure of 4.5 bar to the FPS valve unit.
Dilution air flow to both stages is controlled by the FPS valve unit. Pressurized filtered air is
supplied to the valve units, which has eight magnetic valves, each of which is connected to a
critical flow orifice. Valves 1 and 2 are used to control the ejector diluter (secondary dilution) air
flow, while valves 3, 4, 5, 6, 7, and 8 control the primary dilution air flow. Each critical orifice has
a different size. Opening and closing various combinations of the 8 magnetic valves produces
various dilution ratios. If the dilution air into the primary stage is greater than what is drawn with
the ejector, the dilution ratio becomes infinite – indicated by the sign NaN on the display. When
this occurs, the primary dilution air flow rate should be lowered.
27
Figure 2-3
Principle of Dilution Method for Dekati FPS-4000 (Dekati, 2010)
2.3 Aerosol Diluter Systematic Comparison
The specifications and capabilities of the two aerosol diluter types, shown in Table 2-1
below, are significantly different from one another. As such, each diluter has distinct technical
advantages that makes it suitable for different experimental conditions. Note that these
specifications are directly sourced from their manufacturers, TSI and Dekati.
The TSI 379020A rotary disk thermodiluters are designed with more focus on simplicity
and robustness. For instance, the adjustment of DR can be done without the use of tools or
recalibration. The rotating cavity disk can be easily accessed and maintained. The engineered
coatings on the disks allow great reduction in wear and improve lifetime. In addition, replaceable
28
disks are readily available for added convenience and to reduce maintenance downtime. As such,
these advantages make the TSI 379020A diluters to be more suitable for uses in simple engine
exhaust emission studies where the accurate measurement of concentration with numerous
repeated sessions is critical.
TSI Model 379020A Dekati FPS-4000
Raw sample temp. 0 – 200 ºC 0 – 600 ºC
Raw sample press. (abs) 900 – 1100 mbar 750 – 2000 mbar
Sample flow Approx. 1 lpm 0 – 10 lpm1
Primary dilution air flow 0.5 – 5 lpm 2 – 40 lpm
Secondary dilution air flow 0 – 15 lpm2 40 – 140 lpm
Diluted sample flow Up to 16.5 lpm 60 – 160 lpm
Dilution ratio 1:15 – 1:30003 1:20 – 1:2004
Primary dilution ratio 10-cavity disk: 1:15 – 1:300
8-cavity disk: 1:150 – 1:3000
1:3 – 1:20
Secondary dilution ratio 1:1 – 1:11 1:7 – 1:15
Dilution temp. OFF, 80, 120, or 150 ºC 0 – 350 ºC
Evaporation tube temp.5 Ambient to 400 ºC
1 Measurement at 1.013 bar and 20 ºC
2 With accuracy of 3% of set value +0.1 lpm
3 Dilution accuracy within ±10% range using the calibrated factors supplied with each disk
4 Dilution ratios displayed within ±10% of reading
5 Temp. measurement within ±2 ºC and temp. control within ±3 ºC accuracy
Table 2-1
Aerosol Diluters Specification Summary Table (TSI, 2009; Dekati, 2010)
29
The Dekati FPS-4000 ejector diluter is designed with an emphasis on the ability to create
a wide variety of dilution conditions. Using the Dekati diluter, conducting a study of the particle
dynamics is more manageable than with the TSI diluters. For instance, volatile vapors can be
handled in three ways: preventing condensation with heated dilution; preventing condensation and
possibly capturing some of the vapor in absorbing material with a Dekati thermodenuder; and
facilitating nucleation of vapors with cooled dilution and a residence time chamber. As such, both
heated and cooled dilution can be performed. The residence time within the system is nominally
less than 0.5 seconds, but can be altered by simply installing a longer residence time chamber. In
addition, the particles sampled through the Dekati diluter travel on a straight path with no bends
or moving parts, making it more traceable. These advantages make the Dekati diluter to be more
suitable for uses in nucleation and particle dynamic studies where dilution condition variability
and controlling mechanisms are vital.
2.4 Known Problems of Aerosol Diluters
A previous study of loss in diluters was conducted as part of the European PMP (PMP
report GRPE-PMP-25-5, 2010). The report presented percent particle penetration (Refer to
Equation 1-6) values for various particle sizes, and these are reproduced and shown in Table 2-2.
As shown, the Dekati diluter has a higher percent penetration over the different size bins,
meaning that particle losses within the Dekati diluter are less than those within the TSI diluters.
One thing to note is that the 68% particle penetration value of the TSI 379020A at the 30 nm size
bin is low, indicating that particle numbers in the 30 nm range will be understated in particle
measurements made using the TSI 379020A.
30
P (30 nm) P (50 nm) P (100 nm)
TSI 379020A 68% 88% 95%
Dekati FPS-4000 (heated) 96% 98% 100%
Dekati FPS-4000 (not heated) 100% 99% 100%
Table 2-2
Percent Particle Penetration Values for Varying Particle Sizes for the Diluters
In addition to the percent particle penetration values at different particle sizes, the PMP
report indicated some problems with regards to the concentration measurements made using all of
these diluters.
The PMP report indicated that there has been overestimation of particle number
concentrations measured using the TSI 379020A diluters. The measurement error was caused by
the production of wear particles from the diamond-like carbon rotating disk of the primary diluter.
In response, the manufacturer has developed an alternative coating to prevent this. However, to
the extent that the disk coating deteriorates, it could contribute to the overestimation of the particle
number emissions values.
For the Dekati diluter, the report states that there has been underestimation of the measured
particle number concentrations caused by particle loss inside the thermodenuder evaporating tube
due to evaporation of condensed species.
Any particle loss within the aerosol diluters contribute to over and underestimation of the
measured particle number concentrations, so that an analysis based on the inaccurate measurement
will not be correct and could be misleading. Therefore, verification of diluter characteristics and
31
performance is very important.
In addition, the manufacturers specify that the dilution ratio presented by these diluters will
be within ±10% of the reading. Therefore, to use these diluters with confidence, further
investigation of dilution accuracy is necessary.
32
Chapter 3
Description of Test Instrumentation
3.1 Introduction
Investigations of particle loss in aerosol diluters were done on both vapor and particle
phases. In order to collect quantifiable data to signify the dilution efficiency of the system, effective
methods of measuring concentration of designated gases and particle numbers as a function of
particle size are necessary. However, working with particles can be problematic under various
circumstances, especially for smaller particles (< 30 nm). Smaller particles are subject to
coagulation, surface growth and agglomeration into bigger particle; and bigger particles can
evaporate into smaller ones. Therefore, a means to produce stable solid particles of all sizes is
required to evaluate particle loss through the diluters.
3.2 Gas Concentration Measurement
Verification of actual DRs was done by measuring the dilution of a known tracer gas
passing through the diluter. Two different gases, each at a specified concentration and balanced
with nitrogen were used as tracer gases: methane at 6500 ppm (nominal) and carbon dioxide at
2.15% (nominal). The concentration of methane was measured using a heated flame ionization
detector (HFID) and that of carbon dioxide with a non-dispersive infrared (NDIR) gas analyzer.
3.2.1 Heated Flame Ionization Detector
Measurement of methane concentration was done using a California Analytical Model 600
HFID. The HFID is a very sensitive gas detector for hydrocarbons. In principle, a regulated flow
33
of sample gas (1.5 – 3.0 lpm) is introduced into a hydrogen flame sustained by a flow of hydrogen
fuel gas (40% H2/60% He at 120 cc/min) and zero air (< 1 ppm C at 220 cc/min) in the burner as
shown in Figure 3-1.
Figure 3-1
Flame Ionization Detection Technology (Rosemount Analytical NGA 2000 Manual)
As the sample passes through the burning flame, the hydrocarbon components in the sample
undergo a complex ionization as they are burnt. The burning hydrocarbons produce electrons and
positive and negative ions. The negative ions are collected by the polarized electrodes, which are
biased with a high DC voltage (+ 90 V). Collection of the negative ions causes a current to flow
through an electronic measuring circuit. The ionization current in the measuring circuit is
proportional to the rate of the ionization, which in turn is proportional to the rate at which the
carbon atoms enter the burner. Therefore, the number of carbon atom entering corresponds to the
concentration of hydrocarbon within the sample gas, as shown in Equation 3-1:
34
F ∝ �HI�� � �=0-
�� � [KLM] Equation 3-1
where I is the current through an electronic measuring circuit; IH is the ionization of hydrocarbon;
and Cin is the carbon atom entering the burner.
The CH4 or THC (total hydrocarbon) measurement capacity that can be detected by the
Model 600 HFID ranges from 0 – 3 ppm to 3% carbon (30,000 ppm). Therefore, the 6500 ppm
(nominal) concentration of methane in the tracer gas, falls within the measurement range of the
HFID. The repeatability of the instrument is indicated to be better than 0.5% of full scale – meaning
the measurement error is expected to be always less than 150 ppm (0.5% of 3% carbon). Similarly,
the linearity between the ranges is expected to be better than 0.5% of full scale. The amount of
deviation throughout the instrument’s range is to be less than 150 ppm. The response time of the
Model 600 HFID can be adjusted from 1.0 second to 60 seconds. By default, the response time
was always set to 1.0 second in order to simulate real-time measurement of concentration for better
accuracy.
3.2.2 Non-dispersive Infrared Gas Analyzer
The measure of carbon dioxide concentration was done using the LI-COR LI-820 CO2 gas
analyzer. The NDIR gas analyzer is a simple spectroscopic instrument used to measure CO/CO2
contents in a sample gas. As shown in Figure 3-2, the infrared (IR) source is directed through a 14
cm gold plated optical path, also known as the sample chamber, towards an optical detector. When
a sample flows through the sample chamber, exposure to the IR beam induces absorption at a
specific frequency for each constituent gas. Each element of gas introduced to the IR light absorbs
different wavelengths of light due to differences in their chemical properties. For instance, CO2
35
molecules will absorb IR light at a wavelength of approximately 4.26 µm. The term “non-
dispersive” is used because the wavelength passing through the sampling chamber is not pre-
filtered until a short way before the detector, meaning that the wavelength absorbed by the gas is
allowed to propagate without disturbance or change in shape. This non-dispersive exposed sample
is then passed through 3.95 and 4.26 µm optical filters to eliminate any light other than the
wavelengths that can be absorbed by the CO2 molecules. The 3.95 µm optical filter is used to
measure the reference wavelength for higher measurement accuracy. The electro-optical detector
then measures the amount of IR absorbed by the sample and computes the concentration.
As shown below, the LI-820 is constructed with a gold plated reflector and optical path to
maximize energy transmission. The sample passing through the sample chamber is kept close to
constant conditions by the controlling mechanisms; thermally by a heating element; barometrically
by a pressure transducer; and mechanical shock and vibration protected by the foam enclosure.
The CO2 measurement capacity of this system ranges from 0 – 2% carbon (20,000 ppm), where
the maximum threshold was around 2.5% carbon. Therefore sampling a carbon dioxide gas
concentration below 2.5% is accepted for measurement, in particular the mix containing 2.15%
(nominal) of carbon dioxide that was used as a tracer gas and calibration. The repeatability of the
instrument is indicated to be better than 3% of reading – meaning the measurement error is
expected to be within ± 3% of the current measurement. For instance, when a measurement
displays 2,000 ppm, the expected margin of error is ± 60 ppm. The response time of the LI-820 is
at 1.0 second to simulate real-time measurement of concentration for improved accuracy.
36
Figure 3-2
Non-dispersive Infrared Detector Technology (LI-COR, 2009)
3.3 Particle Number Measurement
The complexity of the particle phase condition arises from the unexpected behavior of
particles in different size bins. When investigating dilution efficiency under the particle condition,
loss due to coagulation, evaporation, condensation, and other possible variables must be
considered. As such, a dependable means to accurately measure particles with a widespread size
distribution is required. The engine exhaust particle sizer (EEPS) and fast mobility particle sizer
(FMPS) each provide such a reliable method for proper measurement.
3.3.1 EEPS/FMPS
The EEPS and FMPS are spectrometers that are used to measure the size distribution of
particles in the range from 5.6 to 560 nm. The core operating principle is identical in the two
instruments. However, the naming of the instruments reflects characteristics that differentiate them.
37
The EEPS provides faster time resolution that is more beneficial in engine exhaust studies.
Figure 3-3
Differential Mobility Analyzer Technology (TSI, 2004)
Measurement of particle size distribution was performed using one or both of the TSI
EEPS spectrometer 3090 and TSI FMPS spectrometer 3091, depending on the specific experiment
being conducted. Collectively, these two instruments are known as the differential mobility
analyzers (DMA). The DMA utilizes an electric field to classify and analyze charged aerosol
particles of different sizes in a gas phase. When aerosol particles of a specified sample are drawn
continuously into the system, they are positively charged to a predictable level by a corona charger,
as illustrated in Figure 3-3.
These charged particles are introduced to and transported down a high voltage electrode
column by the HEPA-filtered sheath air. The electrode column is also charged with a positive
38
voltage, which causes the charged particles to be repelled outwards according to their electrical
mobility. It is important to note that the electrical mobility of a particle depends on its size. Larger
particles with larger surface area will have lower electrical mobility and smaller particles will have
higher electrical mobility. Therefore, smaller particles are repelled further outwards and larger ones
less so. These charged particles then strike the appropriate highly sensitive electrometers and
transfer their charges. Figure 3-3 shows that the smaller diameter particles will strike an
electrometer near the top (E1) whereas the larger diameter particles strike one at the bottom (EN).
The counts of transferred charges from the particle to the respective electrometers are then
computed to result in a particle number at appropriate size bins simultaneously.
As previously mentioned, both the EEPS and FMPS are capable of measuring particles in
the range from 5.6 to 560 nm. The resolution for particle size distributions is 16 channels per
decade, meaning there are 32 size bins measured between 5.6 and 560 nm. The size bins indicated
by the analyzers are the average of a local range of similar particle sizes. For example, the size bin
of 6.98 nm include particle sizes from 6.51 to 7.52 nm. The maximum and minimum concentration
measurements for the two instruments are not identical however. As shown in Figures 3-4 and 3-
5, the maximum and minimum concentration measurement at various averaging times (or response
times) for the EEPS is an order of magnitude higher than for the FMPS. However, for better
comparison and relation to one another, the averaging time was set to 1.0 seconds for both
instruments. When direct comparison between the concentrations measured by both instruments
was necessary, the appropriate correction factor was applied Details of the correction procedure
are provided in Chapter 5.
39
Figure 3-4
EEPS Concentration Range (TSI, 2012)
Figure 3-5
FMPS Concentration Range (TSI, 2004)
40
3.4 Soot Generator
In the particle phase dilution study, having a particle-generating source capable of
producing a constant size distribution over time is crucial. A diesel engine is a great source for
producing particles having a widespread size range. However, the ultrafine particles produced are
significantly subjected to the particle coagulation, evaporation, or condensation (gas-to-particle
conversion) mechanisms that can change the distribution Thus, a means to produce stable small
solid particles is necessary to avoid such particle conversion mechanisms. The soot generator
provides the capacity of producing suitable particles in sufficient concentration.
Figure 3-6
Soot Generator Technology (Jing, 2009)
The desired particles were produced using the Jing miniCAST MOD6203 soot generator.
It consists of an isolated flame chamber where a stable flame enables the production of a particle
stream. As shown in Figure 3-6, a co-flow diffusion flame generates soot particles due to
hydrocarbon pyrolysis in the diffusion flame core. Pyrolysis is a process where thermochemical
decomposition of a vapor phase organic compound occurs at elevated temperatures in the absence
41
of oxygen. It is a process where the solid residue is produced from the result of burning organic
material.
The miniCAST uses propane gas supplied at 43.5 psi as the fuel source to produce soot
particles. The fuel flow rate is constant and the generated particle size distribution is adjusted by
varying the quenching gas flow rate. Nitrogen gas, supplied at 43.5 psi, is used as a quench gas
The flow rate of nitrogen is adjusted to prevent further combustion processes from occurring at
various stages of burning and to stabilize the soot particles. The quenching process inhibits particle
condensation at ambient air conditions. As shown in Figure 3-7, increasing the flow rate of the
quenching gas shifts the size distribution towards the ultrafine region. With a higher quenching
flow rate, the mode size of generated particles becomes smaller. For instance, the particles
concentrated around 20 nm diameter can be produced at 80 sccm (Standard Cubic Centimeter per
Minute, cm3/min) of nitrogen flow.
Figure 3-7
Quenching Gas Flow Rate Dependent Soot Particle Size Distribution
dN
/dlo
gD
p (
#/c
m3)
42
Chapter 4
Evaluation of Dilution Accuracy Using Gas Phase Species
4.1 Introduction
The performance of an aerosol diluter is significantly affected by any loss of sample within
the dilution process. The actual dilution ratio delivered by the diluter must be known accurately.
This portion of the work will focus on evaluating the dilution accuracy in the aerosol diluters using
gas phase species as tracers. This investigation will quantify the difference between the actual
dilution ratio delivered and the user selected (theoretical) dilution ratio performed by the diluter
system, neglecting the effect of particle loss such as coagulation, evaporation and condensation.
Particle loss will be investigated in Chapter 5. The diluters tested were the single-stage and two-
stage TSI 379020A rotary disk thermodiluters, and the Dekati FPS-4000 ejector diluter.
4.2 Dilution Evaluation Apparatus
Figure 4-1 shows the apparatus used for the dilution evaluation measurement of the
diluters investigated. A specified gas mixture with a known concentration was used as the input
source for the diluter. The downstream concentration level was monitored continuously with an
appropriate analyzer (i.e a HFID for a hydrocarbon containing gas mixture or an NDIR instrument
for a CO2 containing mixture).
Varied experimental layouts were used to verify the volumetric dilution condition for the
different types of diluters. Substantial loss of gas phase organic compounds can occur in rotary
disk thermodiluters according to the TSI 379020A manual (TSI, 2009) due to the presence of an
43
evaporation tube in the flow path. This loss was verified in a series of preliminary experiments
(described in Appendix A). Consequently, a hydrocarbon mixture, such as methane, is unsuitable
as the input source to test diluter performance. Thus, to evaluate these TSI diluter measurements,
a mixture of 2.15% (nominal) carbon dioxide balanced in nitrogen was used as the tracer gas. The
ejector diluter, on the other hand, did not scavenge volatile compounds. As such, a mixture of 6500
ppm (nominal) of methane balanced in nitrogen was used as the upstream sample.
Figure 4-1
Dilution Evaluation Apparatus for the TSI 379020A (Left) and the Dekati FPS-4000 (Right)
As shown in Figure 4-1, an excess flow airway was placed before the diluters to ensure
near atmospheric pressure at the sample inlet point. Since the volumetric DRs are subjected to
fluctuation due to pressure change, a “positive flow” to the ambient is required to relieve any
44
pressure above atmosphere.
The pressure has a significant effect on the behavior of the dilution process in the diluters.
As shown in figure 2-3, the Dekati diluter consists of an ejector pump. The pressure drop across
the inlet and the outlet points determines the flow rate through the ejector pump. Hence, the
accumulation of pressure at the inlet of the ejector pump causes an undesirably high flow rate and
affects the DR. Although the effect of such a pressure build up is less in the TSI diluters, inaccurate
dilution can still occur. The higher flow rate caused by the pressure can increase the pump
controlled flow rate. Therefore a mitigating strategy for relieving such pressure build up was
necessary.
The flow rate at the sampling point for the Dekati diluter varied from 0-10 lpm depending
on the DR, so constant monitoring for the presence of excess flow was crucial. The excess flow
was vented to the atmosphere through the 1/2” diameter Teflon tubing to accommodate the large
flows. For the TSI diluters, the undiluted sample is pulled in at a rate of approximately 1.0 lpm,
thus there was a sufficiently greater flow of inlet gas (> 1.0 lpm) and 1/4” diameter Teflon tubing
was adequate.
A bypass line was installed for the Dekati test apparatus to provide an option for
monitoring the upstream concentration level. As mentioned before, the Dekati diluter sampling
flow rate varies significantly depending on the set DR. For the case in which the sample flow rate
becomes greater than the source, an undesired dilution occurs at the excess flow tee – by drawing
in atmospheric air from the ventilation. Observing the upstream concentration of the Dekati
provided secondary insurance to avoid such “negative flow” from happening. Sudden changes in
the input sample concentration could be used to indicate that such “negative flow” had occurred.
45
A modification was made to the commercial single-stage TSI 379020A for this test
apparatus. The original single-stage TSI 379020A drew dilution air from an open rear port, covered
inside the aluminum rear panel with small ventilation openings. Evidently, the system was
circulating a mix of atmospheric and dusty air from the internal component as the dilution air. Thus
inconsistent concentrations of unfiltered air from these sources were injected in the dilution
process. To compensate for this, the rear port was extended outwards through the aluminum panel
using quick-disconnecting tube couplings and 1/4’’ flexible polyvinyl chloride tubing. This new
inlet was supplied with a “positive flow” of 5.0 zero grade air (CO2 ≤ 1 ppm) at the excess airway.
Again, the “positive flow” was important to avoid any pressure induced variations. The two-stage
TSI 379020A has the single-stage unit imbedded within system in addition to the secondary
dilution unit. The secondary dilution component generates the HEPA-filtered, particle-free primary
dilution air for both stages with a calibrated and controlled flow of 1.5 lpm (TSI, 2009).
A TSI flowmeter was installed downstream of the TSI 379020A diluters to monitor the
diluted output flow rate. The dilution in the TSI diluters was governed by the output flow of the
diluted sample. Variance in the dilution ratio occurred when the flow rate of the diluted output was
changed. The NDIR analyzer, downstream of the flowmeter, had an internal pump, which allowed
a relatively constant flow rate to enter the instrument. This pump was used to provide a flow rate
of approximately 1.4 lpm and monitored with an external flowmeter to avoid any fluctuation in
DR during the tests.
4.3 Dilution Evaluation Methods
In order to properly characterize dilution, it is crucial to calibrate the analyzers using gas
concentrations covering the desired range prior to the installation of diluters. The calibration gas
46
concentrations must be within 80% of the measurement range capacity of the analyzers. This
precursor step ensures higher accuracy at both the high and low concentrations being measured.
Thus, the calibration gas cylinders were also used as the sample gas in the appropriate range of the
analyzers.
Once the analyzers were calibrated with the sample gases, the downstream concentration
measurements were done on varying DRs. The DRs were adjusted according to the system being
tested, ranging from the lowest possible DR to a DR of approximately 100 in increments of 10 ±
5. This range covered the dilutions most often used in engine emission testing. A DR higher than
100 may dilute the concentration of minor exhaust constituent to a level below the detection limit
of analyzers. As such, the investigation focused on the DRs up to approximately 100. However,
the diluters are capable of performing dilution at higher DRs.
For the two-stage TSI 379020A, the primary dilution potentiometer was kept constant at
100% while the secondary dilution potentiometer was adjusted. This isolated the effect of the
primary from the secondary dilution for analysis. Behavior of the primary dilution was verified by
testing the single-stage TSI 379020A. It should be noted that a higher DR is possible if the primary
dilution potentiometer is adjusted. Furthermore, any heating features associated with the diluters
were disabled during the experiment as the sample gases were balanced in nitrogen with minimal
or no volatile materials and heating was therefore not needed.
When operating the single-stage TSI 379020A, it is important to note that the supplied
zero dilution air was always at “positive excess flow” as mentioned before. It was observed that
the DR range coverage was greatly altered when the supplied air quantity was insufficient or
pressure buildup occurred at the inlet. This had no impact for the two-stage diluter as the air was
47
supplied through an internal air generator at a controlled flow rate of 1.5 lpm.
Determining the relative theoretical DR when operating with the two-stage TSI 379020A
was found to be problematic. The DR given by the conventional TSI method (calculation
spreadsheet provided by TSI) deviated more than 20% from the measured value. To correct this
issue, it was necessary to measure both the sample outlet flow and the excess flow from the diluter.
The summation of these flows was used to back-calculate for the corrected DR, which gave better
results with ±10% error margin at 99.7% confidence level. More on this is covered in section 4.6.
In addition, the particle-free air generated internally by the two-stage diluter also
contained undesired supplementary CO2 from the room which resulted in a misrepresentation of
the concentration measured downstream. To compensate for this, quantification of the zero
condition at various DRs was necessary prior to each experiment. It was established that the
internal generator was producing approximately 471.4 ppm of CO2 consistently regardless of the
DR variations. Therefore 471.4 ppm was subtracted from the successive measured mean
downstream concentration for the two-stage diluter. It should be noted that the approximate
concentration level of CO2 found in the atmosphere is approximately 398.58 ppm (Mauna Loa
Observatory: NOAA-ESRL, 2013). However, the concentration measured in the two-stage TSI
diluter was higher than this average level. This phenomenon occurred because the experiment was
conducted in a closed environment filled with potential CO2 sources. These sources include various
instruments running and the number of people present. Therefore, it is important to conduct a zero
condition testing prior to the experiment in order to specify the CO2 concentration level specific
to the current environment.
48
4.4 Sampling Time Parameters
The sample time is defined as the allocated length of time the data is logged for a specified
experimental condition. The sample time used in the diluter evaluation was 180 seconds. During
the sampling period, the analyzers were recording data at an averaging time of 1 second. The
averaging time of the analyzers corresponds to the time interval between the two consecutive
recorded data points. Hence a sample time of 180 seconds with an averaging time of 1 second
resulted in 180 data for each condition. This sample size ensured a large enough population for a
more precise evaluation. In between the varying sample conditions, sufficient time was allowed
for the measured concentration to settle down so as to adjust for the change in the dilution air flows.
It was experimentally determined that approximately 60 seconds of stabilization time was
necessary to measure a comparatively consistent value. In response to this finding, the inter-sample
time was set to 120 seconds during which the DR was effectively changed and stabilized.
4.5 Leakage Check
Another important preparatory step before the actual measurements was inspecting for
any source of leakage in the experimental setup. When measuring the concentration of carbon
dioxide and methane, it was absolutely necessary to verify that there was no source of leakage
occurring in the sampling lines. A small leak allowing atmospheric carbon dioxide or methane to
enter the system under negative pressure could considerably skew the results. Similarly, when the
system was under positive pressure, the sample gas could leak out to the atmosphere causing the
concentration level to decrease. Thus the system was checked thoroughly using the soapy water
leak test. Soapy water was applied to all junctions and lengthy lines while the gas flowed through
the apparatus. Any bubble formation on the applied surface indicated that there was a leak and the
49
appropriate part was replaced prior to initializing the test matrix. To guarantee that there was no
leak in the system, the zero air was flowed into the apparatus and measured by the analyzers. The
system was considered leak tight if the concentration determined by the appropriate analyzer was
less than 1 ppm.
4.6 Calculation Methods
Equation 1-3 described the general calculation method for the volumetric DR. All of the
data collected during the sampling period was averaged over the sample population of 180. Thus,
the average concentration was used to calculate the mean experimental volumetric DR as follows:
!OOOO��� � �P"#$%&'()�*+,-$%&'() Equation 4-1
where �P is the average gas concentration measured by the analyzer in ppm.
As mentioned in section 4.3, back-calculation was necessary to correct for the inherent
deviation of the theoretical DR value given by the TSI spreadsheet. The conventional method
calculates the appropriate DR as follows:
!������� ,QRH � =*[email protected]@<�$'X+-*<�#&0)(&Y Equation 4-2
�Z ��� � 1.5 @ � ���� � 0.3(^_`) Equation 4-3
�Z ��� � �Z���� ��Z b� (^_`) Equation 4-4
where M��A = 1543 is the constant coefficient specific to the ten cavities disk supplied with the
diluter; �� is the potentiometer dial number of the primary and secondary flow settings; and �Z � is the average flow to the specific outlet of the diluter. Therefore, �Z���� and �Z b� were
50
measured for each specified DR and recorded. Using Equations 4-3 and 4-4, the corrected
� ���� was computed. With �������D being kept constant at 100%, the corrected
!������� ,QRH was calculated using Equation 4-2.
4.7 Results
In each case, the diluter performance was studied using the concentration measured
upstream and downstream to calculate the experimental DRs. The recorded values were then
compared with the theoretical values given by the diluter setting. A mentioned previously, the
heating elements were disabled for the duration of the experiments.
4.7.1 Single-Stage TSI 379020A Rotary Disk Thermodiluter
As shown in Figure 4-2, the downstream CO2 concentration was measured by the NDIR
analyzer. The resulting concentration showed that the measured values were within 4.0% of the
theoretical values with exception of an 11.1% deviation at a DR of 45. The uncertainty graphed
for each point was at the 99.7% confidence interval (3 sigma). These y-axis error bars correspond
to the precision of the measured values over 180 points. The results show that greater fluctuation
occurs at DRs below 30 and the fluctuations became visibly smaller as the DR value increased.
Overall, the concentration values were very consistent at a stabilized condition with 99.7%
confidence.
51
Figure 4-2
Diluted CO2 Concentration Measurement at Various Dilution Ratios for the Single-Stage
TSI Diluter
The measured concentration trend line was graphed and compared with the theoretical line.
A power function best describes the experimental results since Equation 4-2 shows that the
downstream concentration is inversely proportional to the DR for a fixed upstream concentration.
The coefficient of the determination, R2, indicates that the equation y = 21352x-1 matched the data
well (R2 = 0.997). There is a discrepancy between this empirical model and the theoretical model.
The constant term (i.e. 21352), if the dilution processes were performed ideally, should correspond
to the initial tracer gas concentration (21500 ppm). The difference in this constant term suggests
that there may have been a leak in the system or imprecision in measured values. These issues will
be discussed later in section 4.8.
The prediction of measurement outcome was made based on this function over the
y = 21352x-1
R² = 0.997
0
200
400
600
800
1000
1200
1400
1600
1800
2000
0 50 100 150 200
CO
2C
once
ntr
ation [ppm
]
Dilution Ratio
Measured Conc. Theoretical Conc. Power (Measured Conc.)
52
measured DRs up to DR = 106. Comparison of this projection to the hypothesized line illustrates
that the probability of the system operating within 4.0% of the theoretical DR is very likely.
Figure 4-3
Experimental Dilution Ratios for the Single-Stage TSI Diluter
Based on the concentration measurement results, the corresponding DR was calculated as
shown in Figure 4-3. Direct comparison of the measured outcome to the theoretical trend indicated
that the values were generally within 3.9%. The highest deviation of 10.2% occurred again at a
DR of 45. The standard deviation of the result showed that, unlike the concentration result, greater
fluctuation generally occurred at a higher DR. The reason for this was because at higher DRs the
fluctuation in the measured downstream concentration was relatively larger than at lower DRs,
regardless of consistency (i.e. a change of 5 ppm at 2000 ppm is minuscule compared to the same
change at 200 ppm). However, the overall regression of the measured DRs fit the theoretical DR
well. Thus the actual DR can be expected to deviate by less than 3.9% from theoretical DR over
0
20
40
60
80
100
120
0 20 40 60 80 100 120
Mea
sure
d D
ilution R
atio
Theoretical Dilution Ratio
Measured/Theoretical Theoretical/Theoretical
53
the DR range of 12 to 106.
4.7.2 Two-Stage TSI 379020A Rotary Disk Thermodiluter
The CO2 concentrations downstream of the diluter, measured with the NDIR analyzer are
shown in Figure 4-4. The measured concentration values are within 7.9% of the theoretical values,
with exception a difference of 21.3% at a DR of 15. The standard deviations calculated over the
range were small. This suggested that the measured values remained very consistent at the
stabilized condition with 99.7% confidence.
Figure 4-4
Diluted CO2 Concentration Measurement at Various Dilution Ratios for the Two-Stage TSI
Diluter
A trend line was fitted to the measured concentration data and compared with the
theoretical line. As before, a power function is used to best describe the experimental result. The
0
200
400
600
800
1000
1200
1400
1600
1800
0 50 100 150 200
CO
2C
once
ntr
ation [ppm
]
Dilution Ratio
Measured Conc. Theoretical Conc. Power (Measured Conc.)
54
coefficient of the determination indicated that the equation y = 22587x-1 described the data well
(R2 = 0.988). A prediction of measurement outcome was made based on this function over the
measured DRs up to DR = 110. Comparing this projection to the theoretical line showed that the
likelihood of the system operating within 7.9% of the theoretical DR was very high. Again, there
is some discrepancy in the constant term (22587), where the theoretical model corresponds to y =
21500 x-1. This issue will be discussed in section 4.8.
Figure 4-5
Experimental Dilution Ratios for the Two-Stage TSI Diluter
The DR calculated from the concentration measurement results are shown in Figure 4-5.
Direct comparison of the measured outcome to the theoretical trend indicated that the measured
values were generally lower by 7.1%. The highest deviation, 17.5%, occurred again at a DR of 15.
The standard deviation of the result showed that the fluctuations became more significant with the
0
20
40
60
80
100
120
140
0 20 40 60 80 100 120
Mea
sure
d D
ilution R
atio
Theoretical Dilution Ratio
Measured/Theoretical Theoretical/Theoretical
55
higher DR due to the difficulty of measuring the lower concentration at the outlet with high DRs.
Regardless of these variations, the regression of the measured DRs matched the theoretical DR
within an acceptable range (≤ 10%). Therefore the two-stage TSI 379020A is expected to deviate
by less than 3.9% from the theoretical DR over the DR range of 22 to 110.
4.7.3 Dekati FPS-4000 Ejector Diluter
Figure 4-6
Diluted CH4 Concentration Measurement at Various Dilution Ratios for the Dekati Diluter
The methane concentrations measured downstream of the diluter by the HFID analyzer
are shown in Figure 4-6. The measured values were within 6.2% of the theoretical values, with
exception of a difference of 14.9% at a DR of 125. The results indicate that the increase in deviation
was directly proportional to the increase in DR. The graph showed a divergence of measured
concentration from the ideal values as the DR values got higher. The y-axis error bars computed
0
100
200
300
400
500
600
0 50 100 150 200
CH
4C
once
ntr
ation [ppm
]
Dilution Ratio
Measured Conc. Theoretical Conc. Power (Measured Conc.)
56
over the range were narrow, suggesting that the measured values were very consistent at a
stabilized condition with 99.7% confidence.
As before, a power function trend line was fitted to the measured concentration data and
compared with the theoretical line. The function fit describing the experimental outcome was
closely in line with the theoretical line. The coefficient of the determination indicated that the
equation y = 6608x-1 fit the data well (R2 = 0.993). Once again, the power law constant would be
expected to be the 6500 ppm concentration of the methane test mixture. The discrepancies between
the fitted power function constant (6608) and the theoretical value 6500 is believed to be due to
measurement imprecision (see section 4.8).
The measurement result was predicted based on the power function over the measured
DRs up to DR = 125. Comparison of this projection to the hypothesized line illustrates that it is
very likely that the system will operate within 6.2% of the theoretical DR.
Relative DRs based on the concentration measurements are shown in Figure 4-7. The
direct comparison of the computed result to the theoretical trend indicated that the values were
generally higher by 6.3%. The highest deviation of 17.3% occurred again at a DR of 125. At this
point, a DR = 125, the system was not operating within the manufacturer’s stated range of ≤ 10%
due to a sudden change in test conditions. Such changes were caused by the unstable (fluctuation)
pressure at the inlet point. The standard deviation of the result showed that the divergence became
more significant as DR increased. With the exception of DR = 125, the regression of the measured
DRs fit the theoretical values within the margin of tolerance. Therefore, the performance of the
Dekati FPS-4000 was found to adequate in that it deviated by less than 6.3% from the theoretical
DR over the DR range of 13 to 107.
57
Figure 4-7
Experimental Dilution Ratios for the Dekati Diluter
4.8 Error Analysis and Discussion of Results
All three diluters were expected to perform the dilution process with an actual DR ±10%
of the theoretical dilution ratio. As the result indicated, the diluters tested generally performed
within this tolerance, with minimal outliers. The volumetric dilution ratios of the single-stage TSI
diluter indicated a 3.9% deviation over the dilution ratio range of 12 to 106. The two-stage TSI
diluter operated within 7.1% deviation over the DR range of 22 to 110. The Dekati diluter had less
than 6.3% of deviation over the DR range of 13 to 107.
As was described by Equation 1-3, the volumetric dilution ratios are dependent only on
the volumetric flow. Thus it can be speculated that the expected DR variance within the tested
diluters are contributed from the measurement precision, the consistency of the operating
0
20
40
60
80
100
120
140
160
180
0 20 40 60 80 100 120 140
Mea
sure
d D
ilution R
atio
Theoretical Dilution Ratio
Measured/Theoretical Theoretical/Theoretical
58
conditions, and the dilution accuracy.
Through leakage testing and numerous secondary procedures, such as monitoring of the
upstream concentration for the Dekati, the consistency of the testing conditions were ensured.
From the literature review, it was stated that the primary dilution temperature, residence time and
relative humidity of the primary dilution air could have influence on the dilution conditions.
However, the experiments were performed under a closed environment where the effects of such
factors are minimal. Thus, any variance contributed from such factors was presumably negligible.
Therefore, the measurement precision and the dilution accuracy were considered as the major
contributing factors to the variance from the expected DR values.
The precision and accuracy of the dilution and measurement instruments were very
important in investigating performance, since small inaccuracies in the data could significantly
affect the result.
To investigate the source of inaccuracies from the test apparatus, an error analysis was
performed on the instruments. The purpose of this analysis was to determine to what extent
measurement error contributed to the discrepancies between the measured DRs and the theoretical
DRs computed.
The uncertainties associated from the gas analyzers are solely dependent on the
measurement precision and accuracies. Therefore the equation for a volumetric DR, Equation 4-2,
is used to determine the uncertainties in the analyzers/particle sizers. This expression can be related
to the general form found in Appendix B as shown below, where k and d terms are simply unity:
!OOOO��� � eP"#$%&'()eP*+,-$%&'() ≈ R � k ghi ≈ R � g
i Equation 4-5
59
Then the expression for the percent uncertainty for the analyzers can be re-expressed as
�&B�j � �kg �
j � �li �j Equation 4-6
where mg and mi are the variances of the average measured downstream and upstream gas
concentrations n and o.
For example, at DR = 12 for the single-stage TSI diluter: mg � 171, n � 20,877 ppm,
mi � 29.3 , and o � 1872 ppm. These values are experimentally measured using the NDIR
analyzer. Computing these values into the Equation 4-6, the percent uncertainties value for DR
=12 can be calculated.
�m�!�j � t 171
20877uj � t29.31872u
j → m�! � 0.018
Using the data collected for the other DRs, the appropriate percent uncertainties values
can be computed for the concentration measurement analyzers, the NDIR and the HFID.
Similarly, the uncertainties in the DRs contributed from the flowmeter used to set the TSI
DRs can be derived. The uncertainties associated with the flowmeter are dependent on the two
measured parameters: the sample flow rate and the excess flow rate, as shown in Appendix B.
Rearranging the expression, the resulting function can be related to the general form found in
Equation 4-5 as shown below, where M��A and �������D are constants:
!������� ,QRH � M��A @ 0.83 @ (�Z���� � �Z b� � 0.3)1.5 @ �������D ≈ R′ � k(Xy � Yy)
Equation 4-7
60
Re-expressing this equation in terms of &{By gives
�&{By�j � � |/hygy/hy mgy�
j � � |/gygy/hy mhy�j Equation 4-8
where mgy and mhy are the manufacturer’s specified accuracy (±2% of the current reading) of the
sample and the excess flow rate readings n′ and d′. For example, at DR = 15 for the two-stage TSI diluter: mgy � 0.019, n′ � 0.93 lpm,
mhy � 0.015, and d′ � 0.76 lpm. These values are experimentally measured using the flowmeter.
Substituting these values into Equation 4-8, the percent uncertainties value for the DR =15 case
can be calculated.
�m�y!′�j � t 1 � 0.76
0.93 � 0.76 ∙ 0.019uj � t 1 � 0.93
0.93 � 0.76 ∙ 0.015uj → m�! � 0.026
Using the data collected for the other DRs, the appropriate percent uncertainties values
can be computed for the flowmeter measurements.
Table 4-1 shows the uncertainties for the instruments used in the TSI diluter experiments
at various range of DRs measured. The analysis indicates that the NDIR has relatively low
uncertainties at lower DRs of 1.8%/1.3% (single /two-stage TSI diluters). The uncertainty values
of the diluters, however, increased as the DRs became higher. The uncertainties associated with
the NDIR for the single-stage TSI rose to 9.2% at DR of 106, and to 5.3% at DR of 110 for the
two-stage TSI. This phenomenon was observed in the previous experimental results, where the
higher DRs deviated more from the expected than the lower DRs.
61
Instrument Percent Uncertainties
Diluters
DR
NDIR
Flowmeter
Single-stage
TSI
12 1.8 2.0
53 3.8 2.0
106 9.2 2.0
Two-stage TSI
15 1.3 2.6
55 3.0 5.9
110 5.3 10.3
Table 4-1
Table of Uncertainty Analysis for TSI Diluter Experimental Apparatus
Similarly, the uncertainties from the flowmeters also increased along with the DR
increment. At lower DRs, the percent uncertainties for the single and two-stage TSI were 2% and
2.6% respectively. One thing to note is that the flowmeter measurement for the single-stage TSI
diluter remained relatively constant throughout the change in DRs (~1.4 lpm). Thus the
uncertainties contributing to the DR value of the single-stage TSI was minimal regardless of the
DR increase. The excess flow rate measured for the two-stage TSI diluter, however, increased
along with DR values, resulting in significantly higher uncertainties at DR = 110 (10.3%).
Direct comparison of the uncertainties associated from the NDIR and the flowmeter
suggests that as the DR increases, the imprecision of the flowmeter becomes the major contributing
factor. At DR = 110, the uncertainties from the flowmeter are twice the uncertainties from the
NDIR. Therefore, accurate measurement of flow is crucial at higher DR to set an accurate DR
value.
62
Likewise, the uncertainties for the instruments used in the Dekati diluter experiment at
various range of DRs measured as shown in Table 4-2. Again, the uncertainty values of the HFID
increased along with the DR increment. At a low DR of 13, the uncertainty value was 1.7%, where
at a higher DR of 125, it rose to 4.7%.
These analyses suggest that the major factor contributing to discrepancies in the
experimental and theoretical DRs measurements are indeed from the precision of the instruments.
It should also be noted that the uncertainties contributed from the flowmeter become more
significant with higher flow rates at higher DRs.
Instrument Percent
Uncertainties
Diluters
DR
HFID
Dekati
13 1.7
56 3.7
125 4.7
Table 4-2
Table of Uncertainty Analysis for Dekati Diluter Experimental Apparatus
As previously indicated in the result sections, there were some discrepancies in power law
constants between the empirical and the theoretical models. The difference in these constant terms
for the TSI single-stage, TSI two-stage, and Dekati diluters were found to be 0.7%, 5.1%, and 1.7%
respectively. These percent differences are well within the expected uncertainties due to the
imprecision in measurement apparatuses. Thus, the discrepancies in power law constants can be
believed to be due to the measurement imprecision quantified in the calculations above.
63
In addition, the result indicated that there were some cases where the dilution ratio
deviated more than the expected error margin. The uncertainty analysis on such a case is
represented by the values of the two-stage TSI. Compared to the single-stage TSI test setup, the
uncertainty on the diluter increased and the others remained constant. This indicates that such
outliers are caused solely by the effect of dilution conditions changing within the diluter. The
primary reason for the cause of change in dilution condition is believed to be a sudden fluctuation
in input pressure. A stabilized pressure at the sample inlet is very difficult to maintain. However,
an extended inter-sample time to allow for a longer adjustment period for a dilution ratio change
can alleviate the effect of the pressure change. Therefore, for future experiments it is suggested to
implement a longer inter-sample time of 5 min for stabilization purposes.
This chapter has examined the accuracy of the dilution ratio actually achieved by the
various diluters. With this knowledge in hand, it is now possible to quantify particle loss in the
diluters, which is the topic of Chapter 5.
64
Chapter 5
Particle Loss in Aerosol Diluters
5.1 Introduction
The results of Chapter 4 show that the actual dilution ratio achieved by all of the diluters
can be expected to be within ±10% of the user set dilution ratio. With dilution ratio accuracy
established, this chapter will focus on the particle phase loss within the aerosol diluters. The effect
of particle loss, neglected in the vapor phase such as coagulation, evaporation, condensation, and
other diffusion loss, will be discussed. The influence of these effects on different particle sizes is
investigated to further characterize the systematic behavior of the diluters. In addition, the
operating features such as heating and thermal conditioning are explored. Again, the diluters tested
were the single-stage and two-stage TSI 379020A rotary disk thermodiluters, and the Dekati FPS-
4000 ejector diluter.
5.2 Particle Loss Measurement Apparatus
Due to the complexity of particle dynamics of the engine-emitted emission particles, the
particle loss measurement experiments were conducted in two phases with varying particle sources:
the exhaust emitted from the diesel engine and the soot particles generated from the Jing
miniCAST MOD6203 soot generator. The ultrafine particles produced from the diesel engine may
undergo coagulation, evaporation, or condensation mechanisms due to their significant organic
carbon content. Thus, particles below the 29.4 nm size range were significantly lost during the
dilution process (As indicated by the results in section 5.7). Hence accurate measurements of
particle distribution were extremely difficult. Therefore, the soot generator was used as a means to
65
produce stable small solid particles to minimize such particle conversion mechanisms.
5.2.1 Diesel Emission Test Apparatus
Figure 5-1 shows the apparatus used for the particle loss measurement of the diluters
investigated utilizing the diesel exhaust. The sample particles were generated from a Tier 1 direct
injection, turbocharged, water-cooled, Cummins diesel engine operating in stabilized condition at
mode 9 of the ISO 8178 engine test procedure (1400 rpm, 25% load). The upstream and
downstream concentration levels were monitored continuously with the EEPS and the FMPS.
As discussed in Chapter 1, the direct emission from the engine was sampled using an
engine dynamometer setup, where the engine is isolated from the vehicle body. The engine out-
emission is directed through the 3” thermally insulated stainless steel exhaust pipe to a sampling
canister with an 11.5” diameter. This canister is used as a placement holder for diesel aftertreatment
systems such as a DPF or DOC. The engine exhaust then can be sampled at various points along
the canister through a 1/2” outlet port – i.e. pre-treated and post-treated emissions can be examined.
After the diesel exhaust is treated by the aftertreatment system, the number of particles in all size
ranges is greatly reduced. The particle numbers in certain size ranges may fall below the detection
limit of the particle sizer. Thus for the purpose of verifying the particle loss within the diluter
system, the exhaust upstream of the aftertreatment system was used due to the higher particle
number available in all size bins.
66
Figure 5-1
Diesel Exhaust Particle Loss Test Apparatus for Diluters
As illustrated in Figure 5-1, the raw exhaust from the diesel engine was pre-diluted with
HEPA-filtered, particle-free, dry air to dilute the raw exhaust down to the appropriate concentration
level. This step was necessary since the upper particle number limit of the EEPS is lower than the
untreated concentration from the engine exhaust. The flow rate of the dilution air was gradually
increased until the monitored mode particle concentration fell below the detection capacity of the
particle sizer (approximately 108 #/cm3). It should be noted that the dilution ratio of this pre-
dilution process does not have to be known since the concentration upstream and downstream of
the diluters are collected simultaneously. Therefore, regardless of the dilution ratio used in the pre-
dilution phase, the sample upstream concentration can be related to the sample downstream
concentration.
67
A pre-dilution also helps to decrease the inlet sample temperature. This is essential for the
proper functioning of the EEPS. The temperature of the raw exhaust from the operating engine at
1400 rpm is approximately 140 °C. This temperature is well beyond the range of accepted inlet
sample temperature of the particle sizer: 10 to 52 °C. Thus, by introducing the dilution air at room
temperature (approximately 24 °C), the sample temperature can be reduced to the acceptable range.
Dilution with ambient temperature air also forces condensation of volatiles, ensuring a large
concentration of very small particles.
A vent line to the atmosphere was installed before the diluters to avoid any pressure
accumulation at the sampling point. The presence of any pressure buildup has a direct effect on
the dilution ratio as discussed in the previous chapter.
A post-dilution was also performed before the downstream measurement. The reason for
this additional dilution process was due to the difference between the flow rates of the diluters and
the FMPS. The FMPS samples at 9.15 lpm, whereas the TSI diluters provide ~ 1.425 lpm and the
Dekati diluter operates at various flow rates depending on the dilution ratio chosen. Thus the flow
rates from the outlet of the diluters are well below the inlet flow required by the FMPS. Directly
connecting the FMPS to the outlet of the diluters will cause pressure fluctuations and as discussed
before, will cause significant change in the dilution conditions. Therefore, a post-dilution was done
with HEPA-filtered, particle-free air to provide make-up flow for the FMPS.
A flowmeter was installed before the connection between the FMPS and the diluter to
measure the post-dilution ratio achieved. It was necessary to quantify this dilution ratio as opposed
to the pre-dilution process in order to identify the diluted concentration at the diluter output –
which was the true downstream concentration of the diluter.
68
5.2.2 Soot Generator Test Apparatus
Figure 5-2 shows the apparatus used for the particle loss measurement of the diluters
investigated utilizing stable ultrafine particles produced by the soot generator. As discussed in
Chapter 3, the particle size distribution generated from the soot generator can be varied using
different flow rates of the quenching gas, nitrogen. To produce particles below 29.4 nm, the
quenching flow rate was gradually increased until the mode particle size fell within the range of
8.06 and 22.1 nm. A quenching flow rate of approximately 60 sccm produced such a particle
distribution and this could be varied through the fuel gas condition (i.e. proportions of fuel and N2
and pressure). Fuel pressure can also have an effect in producing a different particle size
distribution.
The diffusion flame in the soot generator is very sensitive to the pressure at the outlet. The
flame will extinguish in the presence of a pressure build-up at this point. Therefore, to produce a
stable flame, a 1/2” vent line to the atmosphere was used to relieve any buildup pressure. There
was a flow pipe reduction from 1/2" to 1/4" between the soot generator and the diluter. Hence, a
second vent line to atmosphere was necessary to relieve the pressure at the diluter inlet. As
discussed in Chapter 4, a “positive flow” to the atmosphere was always ensured during the
experiment to ensure a precise dilution process.
69
Figure 5-2
Soot Generator Particle Loss Test Apparatus for Diluters
Unlike the engine exhaust particle loss test apparatus, producing constant particle
concentration over the various size bins was difficult for the soot generator particle loss test
apparatus due to unstable flame conditions. The placement of the EEPS upstream of the diluter
extinguished the flame by creating too much pressure from the sampling flow rate. Hence, the
measurements of the particle concentrations upstream of the diluter were made prior to the
experiment. Once the flame was stabilized, the particle size distribution produced from the soot
generator was very consistent. Hence the distribution change over time was minimal and it was
safe to assume that the upstream measurement made prior to the experiment was valid. These
measurements will be presented in section 5.7.
For the reasons discussed in the previous section, the diluter output was further diluted
and a flowmeter was installed to allow appropriate measurement of the true downstream flow rate.
70
5.3 Particle Loss Measurement Methods
For the measurement of diluter particle loss, it was crucial to stabilize the engine speed
and the soot generator flame prior to the measurement. At an unstable condition these sources will
produce particle distributions with fluctuations in the mode particle size. The engine or the soot
generator should be allowed to pre-run for 30 minutes before the experiment. This preliminary
step ensured higher precision with less fluctuation in the size distribution.
Once the particle producing sources are stabilized, simultaneous measurement was done
on the upstream and downstream concentration for varying DRs, except that when the soot
generator was used, the upstream concentration was measured prior to that downstream. For the
soot generator, the DRs were adjusted ranging from the lowest possible DR to a DR of
approximately 100 in increments of 10 ± 5. When testing with the engine exhaust, three specific
DRs (8, 15, 26 for the single-stage TSI; 50, 72, 100 for the two-stage TSI; 9, 12, 14 for the Dekati)
were verified. These DRs were chosen in accordance to the range of DR used in the Exhaust
Measurement and Inhalation Toxicology Testing of Emerging Diesel Fuels (EMITTED) study. The
EMITTED study characterizes the diesel engine exhaust emission at various points along the
aftertreatment systems and evaluates the effect of these control technologies. The single-stage TSI
diluter is used to dilute the post-treated emissions at lower DRs.; the two-stage TSI diluter is used
to dilute the pre-treated emission at higher DRs; and the Dekati diluter for the filter collection of
exhaust at lower DRs. A wider range was tested through the soot generator experiment, from
lowest to the highest DR possible without affecting the state of the flame.
It should be noted that the same operating procedures and conditions were followed as
discussed in the dilution evaluation measurement methods section (Chapter 4) when operating the
71
diluters. Any heating features associated with the diluters were disabled during the experiment
(primary air and evaporation tube temperature at room temperature ~ 24 °C). The heating and
thermal conditioning features were investigated separately in section 5.8.
When measuring the same source simultaneously there were discrepancies between the
measurements by the EEPS and the FMPS. The particle concentrations measured at varying size
bins were not identical to one another. Since the two particle sizers were reading different number
concentrations under the same condition due to the systematic behavior difference, equivalency
testing was performed. Such testing was essential for the DR analysis, as the disagreement between
the two instruments will result in a misrepresentation of the particle loss.
Figure 5-3
EEPS/FMPS Equivalency Correction Factor (Zimmerman et al, 2013)
As shown in Figure 5-3, appropriate correction factors were derived for the individual
0.80
0.90
1.00
1.10
1.20
1.30
1.40
1.50
5 50
EEPS/F
MPS -
Cle
aned
Colu
mn
Diameter (nm)
40 10 15 20 25 30 35 45 100 150 200 250
72
particle size bins from 9.31 to 220.7 nm. The difference between the two instruments at the varying
particle size bins was up to 30%, thus indicating that if the DR analysis is performed with raw data,
the result of particle loss within the system will be exaggerated by a factor of 1.3. Therefore, to
appropriately equate the readings from the two instruments, these correction factors were applied
to the individual size bins of the FMPS.
5.4 Sampling Time Parameters
The sample time used in the particle loss measurements was 10 minutes. During the
sampling period, the analyzers were recording the data at the averaging time of 1 second. The
averaging time of the analyzers corresponded to the time interval between the two consecutive
recorded data. Hence a sample time of 10 minutes with an averaging time of 1 second resulted in
600 data points for each condition. This sample size ensured a large enough population for a better
averaging value.
After changing the test conditions (i.e. DR), adequate time was required for the measured
concentration to settle down to adjust for the change in the dilution air flows. It was observed
experimentally that approximately 180 seconds of stabilization time was necessary to measure a
comparatively consistent value. In response to this finding, a period of up to 180 seconds was
allowed before taking data under a new test condition.
5.5 Leakage Check
Another important preparatory step before the actual tests was inspecting for any source
of leakage in the experimental setup. When measuring the concentration or particle size
distribution, it was absolutely necessary that there was no source of leakage occurring in the
73
sampling lines. A small leak that allowed atmospheric particles to enter the system could
considerably skew the results. Similarly, operating the system under positive pressure could allow
particles to leak out to the atmosphere causing the number concentration to decrease. Thus, the
system was checked thoroughly using the soapy water leak test. Soapy water was applied to all
junctions and lengthy lines while particles were flowed into the apparatus. Any bubble formation
on the applied surface indicated that there was a leak and the appropriate part was replaced prior
to initializing the test matrix. To guarantee that there was certainly no leak in the system, zero air
was flowed into the apparatus and measured by the particle sizers. The system was considered leak
tight if the particle number concentration determined by the particle sizers were below the
detection limit.
5.6 Calculation Methods
Equation 1-4 described the general calculation method for the particle DR. But, all of the
data collected during the sampling period was averaged over the sample population of 600. Thus,
the average concentration is used to calculate the mean experimental particle DR as follows:
!OOOO������� ( ����� ) � eP"#$%&'()( �#
�$02')eP*+,-$%&'()( �#
�$02') Equation 5-1
where �̅ is the average particle concentration in individual particle size bin measured by the sizers
in #/cm3.
The same method for correcting theoretical DR for the TSI two-stage diluter described in
Chapter 4 was used. Using Equations 4-4 and 4-5, the corrected � ���� was computed. With
�������D being kept constant at 100%, the corrected !������� ,QRH was calculated using the
Equation 4-3.
74
In the particle loss test apparatus, additional post-dilution was performed in order to
compensate for the detection capacity of the particle sizers. As such, a correction factor had to be
applied to the measured downstream particle size bins to compute the appropriate particle
concentration. For this correction, the post-dilution process was assumed to be a volumetric phase
and any particle loss was assumed to be negligible. With this assumption, Equation 1-3 was used
to calculate the correction factor. Since the upstream concentration of the post-dilution process
was desired, the equation was rewritten as
����� �� � ."#$%&'()/.*01"%0+-."#$%&'() @ ������ �� Equation 5-2
where the term ."#$%&'()/.*01"%0+-
."#$%&'() � M is the determined correction factor. This correction factor
was then applied to the individual particle size bins. The computed concentration from this
correction was used as the downstream concentration for further analysis.
5.7 Results
In each case, the particle loss within the diluter system was studied using the concentration
measured upstream and downstream to calculate the experimental DRs. The recorded value was
then compared with the theoretical values given by the diluter setting.
5.7.1 Single-Stage TSI 379020A Rotary Disk Thermodiluter
As shown in Figure 5-4, the upstream and downstream engine exhaust particle
distributions at various DRs were measured by the appropriate particle sizers (i.e. an EEPS for the
upstream concentration and an FMPS for the downstream concentration). The resulting
concentrations showed that the measured downstream distributions at the three DRs tested, 8, 15,
75
and 26 were following a similar trend to the undiluted (raw exhaust) concentration. The undiluted
concentration was measured simultaneously for each DR. Because these were very similar, only
the undiluted concentration measurement at DR = 8 is presented in Figure 5-4. From the direct
comparison of undiluted and diluted data for DR = 8, it can be seen that the shape of two
distributions is indeed similar. The mode particle size for each distribution was found to be around
93.1 nm for DR = 8 but it was shifted to 60.4 nm for DR = 15 and 26. This, however, did not
suggest that the dilution behavior had changed because the simultaneous measurement of the
upstream concentration indicated a similar trend as the data graphed for DR = 15 and 26.
Figure 5-4
Engine Exhaust Particle Distribution at Various Dilution Ratio for the Single-Stage TSI
Diluter
It should be noted that the standard deviation graphed for each point was at the 99.7%
confidence level. These y-axis error bars correspond to the precision of measured values over 600
1.E-01
1.E+00
1.E+01
1.E+02
1.E+03
1.E+04
1.E+05
1.E+06
5 50
dN
/dlo
g D
p (#/c
m3)
Particle Size (nm)
Undiluted DR = 8 DR = 15 DR = 26
40 10 15 20 25 30 35 45 100 150 200 250
76
points. The result indicated that a greater fluctuation in particle counts occurred for smaller
particles. For particle size bins above 29.4 nm, the error bars were very small. This indicated that
the values were very consistent at a stabilized condition with 99.7% confidence. However below
a particle size of 29.4 nm, particles were being significantly lost after the dilution process. The
increased standard deviation for these particles were a direct result of the undiluted concentration
measurement, where the existence of such variation was evident. Hence, the fluctuation in the
undiluted gas concentrations were reflected on the diluted concentration measurements.
Figure 5-5
Engine Exhaust Experimental Percent Particle Penetration for the Single-Stage TSI Diluter
Figure 5-5 shows the percent particle penetration values corresponding to the measured
concentrations. The resulting values suggested that a significant portion of particles smaller than
29.4 nm were indeed being lost. In fact, the low particle penetration for these small particles may
have been due to a combination of particle loss and particle growth. The percent particle
0
20
40
60
80
100
120
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
PDR = 8 PDR = 15 PDR = 26
40 10 15 20 25 30 35 45 100 150 200 250
77
penetration values above 100% for size bins 29.4 to 52.3 nm suggested that volatile components
in these smaller particles, such as organic carbon, were evaporating and condensing onto the larger
29.4 to 52.3 nm particles. The overall average percent particle penetration for the DRs tested were
98.5%, with lowest penetration at 191.1 nm with 90.8%.
Particle penetration values greater than 100% indicate that mass was gained during the
dilution process. In order to evaluate the extent to which the disappearing small particles contribute
to the increase in numbers (and consequently their mass) of the larger size particles, a mass balance
was performed between the particle size bins below 29.4 nm and the size bins that have particle
penetration values above 100%. If, during the dilution process, the mass of all particles in the
smaller size bins that have disappeared is the only contributor to the increase in mass of the larger
size bins that show penetration values above 100%, the two masses should be equal to one another.
The bins for particle sizes less than 29.4 nm include 9.31, 10.8, 12.4, 14.3, 16.5, 19.1, 22.1, and
25.5 nm diameters while those bins that have particle penetration numbers greater than 100%
include bins with 29.4, 34, 39.2, 45.3, and 52.3 nm diameters. The mass balance relation can be
expressed as:
∑ (`�� � �`� �)j�.��.W| � ∑ (`� � �`�� �)�j.Wj�.� Equation 5-3
where: ∑`� represents the sum of the mass of particles in each bin � , the subscript �ℎ��
represents the theoretical particle mass and the subscript `��m represents the particle mass
calculated from the measured particle numbers in the bin. Here, theoretical means the value
calculated by applying the volumetric dilution ratio to the undiluted particle number concentration
in that bin. There is a mass loss on the left hand side of the equation, where the measured
concentration values are much less than the expected (theoretical) values. There is a gain of mass
78
on the right hand side where the measured concentration exceeds the expected value.
Particle mass can be computed from the number of particles, but the measurement
provides concentration, particles/unit volume of diluted engine exhaust. In order to make use of
the concentration measurements, equation 5-3 can be expressed on a per unit volume basis as:
∑ ������� � � ����� ��j�.��.W| � ∑ ������ � � �
����� ���j.Wj�.� Equation 5-4
Particle mass per unit volume can be calculated from the concentration �� , particle
diameter �� and density �:
���� � �� ∙ ���W ��#j �W�� ∙ � Equation 5-5
Substituting equation 5-5 into equation 5-4 for ���� gives:
∑ �����,�� �� � ���,� ��� ∙ ���W ��#j �W�� ∙ ��j�.��.|
� ∑ �����,� �� � ���,�� ��� ∙ ���W ��#j �W�� ∙ ���j.Wj�.� Equation 5-6
The equations above assume a common density for all particles. The particles less than
29.4 nm in diameter were assumed to be liquid and to have the density (ρ= 0.770 g/cm3) of cetane,
a surrogate for diesel fuel.
Performing the mass balance for DR = 8, the theoretical mass of all particles below 29.4
nm that were lost in dilution was 4 @ 10�|� g/cm3 (mass of cetane/volume of air) and the mass
of all particles gained in the size bins where particle penetration exceeded 100% was 5.7 @ 10�|W
g/cm3. The particles below 29.4 contributed only 7% of the added mass of particles in the bins
79
where greater than above 100% penetration occurred. Hence, it can be hypothesized that the
evaporation and condensation of volatile components in the smaller particles onto the larger 29.4
to 52.3 nm particles are a minor contributor to the increase in particle penetration value. The major
contributor to the particle growth is unclear. However, diesel exhaust contains many organic and
inorganic species that are in the gas phase at normal exhaust system temperature. As the gas
temperature falls during the dilution process these species could condense or adsorb on existing
particles, resulting in growth. Small particles have a larger surface to volume ratio than larger
particles, which means that growth by surface addition would have a much larger effect on particle
mass for small particles.
Based on the concentration measurement results, the corresponding DRs were calculated
for each individual particle size bin as shown in Figure 5-6. Direct comparison of the outcomes
(PDR, particle DR) to the theoretical trend (VDR, volumetric DR) indicated that the values were
generally within 6.6% at DR = 8, 6.6% at DR = 15 and 3.8% at DR = 26 for all size bins above
29.4 nm. The highest deviation of 9.7% occurred at a particle size of 220.7 nm for DR = 8, 11.6%
at a particle size of 191.1 nm for DR = 15, and 9.3% at a particle size of 191.1 nm for DR = 26.
The standard deviation of the result showed that, again, greater fluctuation occurred at lower
particle sizes where a significant portion of the particles was being lost. However, disregarding the
particle concentration measured below 29.4 nm, the overall regression of the measured DRs fit the
theoretical volumetric DR well. Thus, from the engine exhaust studies, the particle loss within the
single-stage TSI 379020A was minimal, and this unit can be expected to dilute to within 6.6% of
the theoretical value for particles larger than 29.4 nm.
The compounds in diesel exhaust are in both the vapor and solid phases, and therefore the
80
particles are subjected to coagulation, condensation and evaporation. Thus, larger particles can
shrink due to hydrocarbon condensation onto smaller particles, which can cause them to grow to
a larger size. Therefore to accurately verify particle loss in the smaller particle ranges, the soot
generator was used as the source instead of the engine exhaust.
Figure 5-6
Engine Exhaust Experimental Dilution Ratios for the Single-Stage TSI Diluter
The appropriate particle sizers measured the upstream and downstream soot generator
particle distributions at various DRs as illustrated in Figure 5-7. The resulting concentrations
showed that the measured downstream distributions at various DRs are very similar in shape. The
mode particle size for each distribution was found to be around 10.8 nm for all DRs tested with
the soot generator. Direct comparison of the undiluted distribution to the diluted distribution
indicates that the general trend of the particle distribution is consistent regardless of the change in
DR. The standard deviations calculated over the ranges were also noticeably small for both
1.E+00
1.E+01
1.E+02
1.E+03
1.E+04
1.E+05
5 50
Dilution R
atio
Particle Size (nm)
PDR = 8 VDR = 8 PDR = 15
VDR = 15 PDR = 26 VDR = 26
40 10 15 20 25 30 35 45 100 150 200 250
81
undiluted and diluted conditions. This indicates that the measured values were very consistent at a
stabilized condition with 99.7% confidence.
Figure 5-7
Soot Generator Particle Distribution at Various Dilution Ratio for the Single-Stage TSI
Diluter
Figure 5-8 shows the average percent particle penetration values corresponding to the
measured concentrations for all the DRs tested with the soot generator. The resulting values
indicate a relatively consistent percent particle penetration over the particle size bins. The percent
particle penetration values above 100% for size bins 16.5 to 29.5 nm were likely due to the
measurement precision of the instruments used in the experiment. As will be discussed in section
5.8, these instruments have certain associated uncertainties that lead to imprecision in
measurement (< 5.9%). However, the average percent particle penetration for the DRs tested were
98.1%, with lowest penetration at 9.31 nm with 92.1%.
1.E+03
1.E+04
1.E+05
1.E+06
1.E+07
1.E+08
5 50
dN
/dlo
g D
p (#/c
m3)
Particle Size (nm)
Undiluted DR = 12 DR = 20 DR = 30
DR = 40 DR = 50 DR = 60 DR = 70
DR = 80 DR = 90 DR = 100
40 10 15 20 25 30 35 45
82
Figure 5-8
Soot Generator Experimental Average Percent Particle Penetration for the Single-Stage
TSI Diluter
The DRs calculated from the particle concentrations measurement results are shown in
Figure 5-9. Comparing individual measured PDR values to the theoretical trend indicated that the
computed PDR values for individual size bins at various DRs were generally within ± 6.0%. The
average percent DR difference (between particle and set volumetric DRs) and the highest
deviations for the various DR conditions are summarized in Table 5-1. From the result, it can be
seen that the highest deviation points occur at various particle size bins with no specific size
preference. These repeated outcomes suggest that the single-stage TSI diluter can suffer particle
loss in any particle size bin. However, the particle loss within the single-stage TSI 379020A is
generally small, less than 12.8% over the DR range of 12 to 100.
0
20
40
60
80
100
120
5 50
Part
icle
Pen
etart
ion (%
)
Particle Size (nm)
Average of Tested PDRs
40 10 15 20 25 30 35 45
83
Figure 5-9
Soot Generator Experimental Dilution Ratios for the Single-Stage TSI Diluter
0
10
20
30
40
50
60
5 50
Dilution R
atio
Particle Size (nm)
PDR = 12 VDR = 12 PDR = 20 VDR = 20
PDR = 30 VDR = 30 PDR = 40 VDR = 40
PDR = 50 VDR = 50
40
50
60
70
80
90
100
110
120
5 50
Dilution R
atio
Particle Size (nm)
PDR = 60 VDR = 60 PDR = 70 VDR = 70
PDR = 80 VDR = 80 PDR = 90 VDR = 90
PDR = 100 VDR = 100
40 10 15 20 25 30 35 45
40 10 15 20 25 30 35 45
84
DR
DR Difference (%)
Highest Deviation Point
Particle Size (nm) DR Difference (%)
12 12.8 39.2 16.1
20 4.6 34 8.3
30 3.5 34 6.3
40 6.5 25.5 9.7
50 6.3 22.1 10.8
60 5.7 22.1 11.5
70 5.9 16.5 11.4
80 4.3 6.04 10.6
90 5.6 12.4 9.8
100 6.0 9.31 11.6
Table 5-1
Summary of Average Percent Dilution Ratio Difference and Highest Deviation Point for the
Single-Stage TSI Diluter
It should be noted that the trends of the PDR values over the range of particle size bins at
various DRs are relatively similar to one another. Such behavior suggests that the systematic
operation of the single-stage TSI diluter is consistent with high repeatability.
5.7.2 Two-Stage TSI 379020A Rotary Disk Thermodiluter
The particle concentrations downstream and upstream of the diluter, measured with the
particle sizers are shown in Figure 5-10. Comparison of the measured downstream distributions to
the upstream distribution at three DRs of 50, 72, and 100 showed similar trends. The shape of the
undiluted size distribution plotted corresponds to that of the upstream measurement at DR = 50.
85
From the direct comparison of these two data sets, it can be seen that the shape of two distributions
are indeed similar. The mode particle size for each distribution was found to be around 69.8 nm
for all DRs tested. This suggested that the engine was producing a consistent concentration of
particles distributed over the entire size range.
Again, the results indicated that the greatest fluctuation in particle counts occurred in the
lower particle size ranges. Particles smaller than 29.4 nm were being lost through the dilution
process and the standard deviation was increased. For ranges above 29.4 the deviations over the
ranges were noticeably smaller, suggesting greater consistency at a stabilized condition with 99.7%
confidence.
Figure 5-10
Engine Exhaust Particle Distribution at Various Dilution Ratio for the Two-Stage TSI
Diluter
Figure 5-11 shows the percent particle penetration values corresponding to the measured
1.E-02
1.E-01
1.E+00
1.E+01
1.E+02
1.E+03
1.E+04
1.E+05
1.E+06
5 50
dN
/dlo
g D
p (#/c
m3)
Particle Size (nm)
Undiluted DR = 50 DR = 72 DR = 100
40 10 15 20 25 30 35 45 100 150 200 250
86
concentrations. Again, the resulting values indicated that a significant portion of particles were
being lost below a particle size of 29.4 nm. The percent particle penetration values were above
100% for size bins 29.4 to 52.3 nm suggesting condensation of hydrocarbons into larger particles
in those particle size bins. However, the average percent particle penetration for the DRs tested
were 98.8%, with lowest penetration at 165.5 nm with 91.1% at DR = 50, 91.2% at 165.5 nm for
DR = 72, and 96.6% at 107.5 nm for DR = 100.
Figure 5-11
Engine Exhaust Experimental Percent Particle Penetration for the Two-Stage TSI Diluter
Once again, a mass balance for DR = 50 was performed. The lost mass per unit volume of
diluted exhaust gas of all particles below 29.4 nm was 3.4 @ 10�|� g/cm3. The mass per unit
volume of diluted exhaust gas gained by all particle sizes where the particle penetration was above
100% was 2.3 @ 10�|� g/cm3. Thus, the particles smaller than 29.4nm that disappeared during
dilution only contributed 14.5% of the mass gained by the particles having a particle penetration
0
20
40
60
80
100
120
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
PDR = 50 PDR = 72 PDR = 100
40 10 15 20 25 30 35 45 100 150 200 250
87
value above 100%. Therefore, similar to the TSI single-stage diluter, the volatile components that
have evaporated and condensed onto the larger 29.4 to 52.3 nm particles are a minor contributor
to the particle growth in these size bins.
Figure 5-12
Engine Exhaust Experimental Dilution Ratios for the Two-Stage TSI Diluter
The DRs corresponding to each individual particle size bin calculated from the
concentration measurements are shown in Figure 5-12. Direct comparison of the PDRs to the
theoretical VDRs indicated that the values were generally within 3.0% at DR = 50, 1.9% at DR =
72 and 4.3% at DR = 100 for all size bins above 29.4 nm. The highest deviation of 7.2% occurred
at a particle size of 34 nm for DR = 50, 6.7% at a particle size of 39.2 nm for DR = 72, and 10.9%
at a particle size of 34 nm for DR = 100. The standard deviation of the result showed that greater
fluctuation occurred at a lower DR where a significant portion of the particles was being lost.
However, disregarding the particle concentrations measured below 29.4 nm, the overall regression
1.E+01
1.E+02
1.E+03
1.E+04
5 50
Dilution R
atio
Particle Size (nm)
PDR = 50 VDR = 50 PDR = 72
VDR = 72 PDR = 100 VDR = 100
40 10 15 20 25 30 35 45 100 150 200 250
88
of the measured PDRs fit the theoretical VDRs well. Thus, from the engine exhaust studies, the
particle loss within the two-stage TSI 379020A was minimal. The expected DR deviation was less
than 4.3% from the theoretical DR over the particle size bins above 29.4 nm.
Figure 5-13
Soot Generator Particle Distribution at Various Dilution Ratio for the Two-Stage TSI
Diluter
The appropriate particle sizers measured the upstream and downstream soot generator
particle distributions at various DRs as illustrated in Figure 5-13. The resulting concentrations
showed that the measured downstream distributions at various DRs are very similar in shape. The
mode particle size for each distribution was found to be around 10.8 nm for all DRs tested with
the soot generator. Direct comparison of the undiluted distribution to the diluted distributions
indicated that the general trend of the particle distribution was consistent regardless of the change
in DR. The standard deviations calculated over the ranges were also noticeably small for both
1.E+04
1.E+05
1.E+06
1.E+07
1.E+08
5 50
dN
/dlo
g D
p (#/c
m3)
Particle Size (nm)
Undiluted DR = 16 DR = 24 DR = 35
DR = 45 DR = 56 DR = 64 DR = 74
DR = 85 DR = 98 DR = 113
40 10 15 20 25 30 35 45
89
undiluted and diluted conditions. This suggests that the measured values were very consistent at a
stabilized condition with 99.7% confidence. For DR = 113, however, the standard deviation
increased. The reason for this increase in error bar was due to fluctuation in pressure causing the
flame to become unstable.
Figure 5-14
Soot Generator Experimental Average Percent Particle Penetration for the Two-Stage TSI
Diluter
Figure 5-14 shows the average percent particle penetration values corresponding to the
measured concentrations for all the DRs tested with the soot generator. The resulting values
indicated a relatively consistent percent particle penetration over all particle size bins. The percent
particle penetration values above 100% for size bins 6.04 to 9.31 nm were likely due to the
measurement precision of the instruments used in the experiment. However, the average percent
particle penetration for the DRs tested were 99.4%, with lowest penetration at 29.4 nm with 94.2%.
0
20
40
60
80
100
120
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
Average of Tested PDRs
40 10 15 20 25 30 35 45
90
Figure 5-15
Soot Generator Experimental Dilution Ratios for the Two-Stage TSI Diluter
The DRs calculated from the particle concentrations measurement results were shown in
Figure 5-15. Comparing individual PDR values to the theoretical volumetric DR trend indicated
10
20
30
40
50
60
70
5 50
Dilution R
atio
Particle Size (nm)
PDR = 16 VDR = 16 PDR = 24 VDR = 24
PDR = 35 VDR = 35 PDR = 45 VDR = 45
PDR = 56 VDR = 56
50
70
90
110
130
5 50
Dilution R
atio
Particle Size (nm)
PDR = 64 VDR = 64 PDR = 74 VDR = 74
PDR = 85 VDR = 85 PDR = 98 VDR = 98
PDR = 113 VDR = 113
40 10 15 20 25 30 35 45
40 10 15 20 25 30 35 45
91
that the computed DR values for individual size bins at various DRs were generally within ± 3.9%.
The average percent DR difference and the highest deviations for the various DR conditions are
summarized in Table 5-2. From the result, it can be seen that the highest deviation points occurred
in various particle size bins. This suggested that the two-stage TSI diluter can suffer particle loss
in any particle size bin. However, the particle loss within the two-stage TSI 379020A is expected
to be small, less than 5.7% over the DR range of 16 to 100.
DR
DR Difference (%)
Highest Deviation Point
Particle Size (nm) DR Difference (%)
16 2.7 45.3 7.0
24 2.7 45.3 7.0
35 5.7 34 10.3
45 3.5 34 6.7
56 3.7 34 8.3
64 3.7 6.98 7.8
74 3.2 6.04 7.2
85 3.9 29.4 8.1
98 4.8 45.3 9.9
113 5.0 45.3 10.8
Table 5-2
Summary of Average Percent Dilution Ratio Difference and Highest Deviation Point for the
Two-Stage TSI Diluter
A similar pattern of particle dilutions over the ranges of size bins suggest that the
systematic operation of the two-stage TSI diluter is consistent with high repeatability. The standard
92
deviation of the individual particle dilutions were also less than 10% suggesting that the system is
operating within the accepted error margin.
5.7.3 Dekati FPS-4000 Ejector Diluter
Figure 5-16
Engine Exhaust Particle Distribution at Various Dilution Ratio for the Dekati Diluter
The particle concentrations downstream and upstream of the diluter, measured with the
particle sizers are shown in Figure 5-16. The comparison of the measured downstream distributions
to the upstream distribution at three DRs of 9, 12, and 14 showed a similar trend. The undiluted
distribution plotted corresponds to the upstream measurement at DR = 9. From direct comparison
of these two data sets, it can be seen that the shape of the two distributions are indeed similar. The
mode particle size for each distribution was found to be around 69.8 nm for all DRs tested. This
suggested that the engine was producing a consistent concentration of particles distributed over
1.E+00
1.E+01
1.E+02
1.E+03
1.E+04
1.E+05
1.E+06
5 50
dN
/dlo
g D
p (#/c
m3)
Particle Size (nm)
Undiluted DR = 9 DR = 12 DR = 14
40 10 15 20 25 30 35 45 100 150 200 250
93
the entire particle size range.
The result indicated that the greater fluctuation in particle counts occurred at lower particle
size ranges. For particle size bins below 29.4 nm, the particles were being significantly lost after
the dilution process and the standard deviation is increased. For ranges above 29.4 nm, the
deviations over the ranges were noticeably small, suggesting great consistency at a stabilized
condition with 99.7% confidence.
Figure 5-17
Engine Exhaust Experimental Percent Particle Penetration for the Dekati Diluter
Figure 5-17 shows the percent particle penetration value corresponds to the measured
concentrations. Again, the resulting values indicated that a significant portion of particles were
being lost below a particle size of 29.4 nm. The percent particle penetration values were above
100% for size bins 29.4 to 52.3 nm suggesting condensation of hydrocarbons into larger particles
in those particle size bins. However, the average percent particle penetration for the DR tested was
0
20
40
60
80
100
120
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
PDR = 9 PDR = 12 PDR = 14
40 10 15 20 25 30 35 45 100 150 200 250
94
99.9%, with lowest penetration at 220.7 nm with 96.4%.
The mass balance comparison shows that at DR = 9, the lost mass of all particles smaller
than 29.4 nm was 2.9 @ 10�|� g/cm3 and the mass gain of all particles having greater than 100%
particle penetration was 3.9 @ 10�|W g/cm3. Thus, the particles smaller than 29.4 nm only
contributed 7.6% of the mass gained by the particles having a particle penetration above 100%.
Therefore, similar to both TSI diluters, the volatile components that evaporated and condensed
onto the larger 29.4 to 52.3 nm particles are a minor contributor to the particle growth in these size
bins.
Figure 5-18
Engine Exhaust Experimental Dilution Ratios for the Dekati Diluter
Relative DRs based on the concentration measurements corresponding to each individual
particle size bin are shown in Figure 5-18. Direct comparison of the PDRs to the theoretical VDRs
1
10
100
1000
5 50
Dilution R
atio
Particle Size (nm)
PDR = 9 VDR = 9 PDR = 12
VDR = 12 PDR = 14 VDR = 14
40 10 15 20 25 30 35 45 100 150 200 250
95
indicated that the values were generally within 5.7% at DR = 9, 7.3% at DR = 12 and 9.6% at DR
= 14 for all size bins above 29.4 nm. The highest deviation of 14.1% occurred at a particle size of
34 nm for DR = 9, 18.2% at a particle size of 34 nm for DR = 12, and 14.6% at a particle size of
39.2 nm for DR = 14. The standard deviation of the result showed that greater fluctuation occurred
at a lower DR where a significant portion of the particles was being lost. However, disregarding
the particle concentration measured below 29.4 nm, the overall regression of the measured PDRs
fit the theoretical VDRs well. Thus, from the engine exhaust studies, the particle loss within the
Dekati FPS-400 was relatively small. The measured DR was expected to deviate less than 9.6% of
theoretical DR over the particle size bins above 29.4 nm.
Figure 5-19
Soot Generator Particle Distribution at Various Dilution Ratios for the Dekati Diluter
Figure 5-19 shows the upstream and downstream soot generator particle distributions at
various DRs measured using the appropriate particle sizers. The resulting concentrations showed
1.E+03
1.E+04
1.E+05
1.E+06
1.E+07
1.E+08
1.E+09
5 50
dN
/dlo
g D
p (#/c
m3)
Particle Size (nm)
Undiluted DR = 12 DR = 23 DR = 35
DR = 44 DR = 55 DR = 66
40 10 15 20 25 30 35 45
96
that the measured downstream distributions at various DRs are very similar in shape. The mode
particle size for each distribution was found to be around 10.8 nm for all DRs tested. The direct
comparisons of the undiluted distribution to the diluted distributions indicate that the general trend
of the particle distribution is consistent regardless of the change in DR. The standard deviations
calculated over the ranges were also noticeably small for both undiluted and diluted conditions.
This suggests that the measured values are very consistent at a stabilized condition with 99.7%
confidence. For DR = 44, however, the standard deviation increased. The reason for this increase
in error was due to fluctuations in pressure causing the flame to become unstable.
Figure 5-20
Soot Generator Experimental Average Percent Particle Penetration for the Dekati Diluter
Figure 5-20 shows the average percent particle penetration values corresponding to the
measured concentrations for all the DRs tested with the soot generator. The resulting values
indicate a relatively consistent percent particle penetration over the particle size bins. The percent
0
20
40
60
80
100
120
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
Average of Tested PDRs
40 10 15 20 25 30 35 45
97
particle penetration values above 100% for size bins 8.06 to 16.5 nm were likely due to the
measurement precision of the instruments used in the experiment. However, the average percent
particle penetration for the DRs tested were ~100 % (99.9 %), with lowest penetration at 6.04 nm
with 93.6%.
The DRs calculated from the particle concentrations measurement results are shown in
Figure 5-21. Comparing individual PDR values to the theoretical VDR indicated that the computed
DR values for individual size bins at various DRs were generally within ± 5.1%. The average
percent DR difference and the highest deviations for the various DR conditions are summarized in
Table 5-3. The highest particle loss at various size bins suggest that the Dekati FPS-4000 diluter
can suffer particle loss in any particle size bin. The particle loss within the Dekati FPS-4000 is
expected to be small, less than 6.4% over the DR range of 12 to 66.
Figure 5-21
Soot Generator Experimental Dilution Ratios for the Dekati Diluter
0
10
20
30
40
50
60
70
80
5 50
Dilution R
atio
Particle Size (nm)
PDR = 12 VDR = 12 PDR = 23 VDR = 23
PDR = 35 VDR = 35 PDR = 44 VDR = 44
PDR = 55 VDR = 55 PDR = 66 VDR = 66
40 10 15 20 25 30 35 45
98
DR
DR Difference (%)
Highest Deviation Point
Particle Size (nm) DR Difference (%)
12 4.5 19.1 7.8
23 4.5 25.5 8.4
35 6.2 14.3 8.2
44 6.4 25.5 9.7
55 4.3 6.04 6.6
66 4.8 6.04 9.2
Table 5-3
Summary of Average Percent Dilution Ratio Difference and Highest Deviation Point for the
Dekati Diluter
A relatively similar pattern of the particle dilutions over the ranges of size bins suggest
that the systematic condition of the Dekati diluter is consistent with high repeatability. The
standard deviation of the individual particle dilutions were also less than 10% suggesting that the
system is operating within the accepted error margin.
5.8 Investigation of Heating and Thermal Conditioning Elements
As mentioned before, the diluters investigated have the capacity to heat the dilution air to
a higher temperature for removal of condensed volatile materials from the sample. In addition, the
two-stage TSI 379020A has a thermal conditioner that allows the elimination of condensed
nanodroplets and volatile particle formation. In the previous experiments, any heating features
associated with the diluters were disabled (primary air and evaporation tube temperatures were
room temperature ~ 24 °C). In this section, the effect of these operating features will be quantified
99
in regards to the particle dilution ratio using the engine exhaust – a known source with a volatile
fraction.
5.8.1 TSI 379020A Rotary Disk Thermodiluter
Figure 5-22
Engine Exhaust Dilution Ratios Corresponding to the Primary Dilution Air Temperature
for Single-stage TSI Diluters
Figure 5-22 illustrates the engine exhaust particle DRs corresponding to the primary
dilution air heating temperature (DHT) for the single-stage TSI diluters above 29.4 nm size bin.
The size bins below 29.4 nm were neglected due to significant loss of particles as noted in section
5.7. The result indicated that as the primary dilution air temperature was increased, the particle
DRs along the recorded size bins converged towards the theoretical dilution ratio. At DR = 5
without heating, the highest deviation was 30% from the theoretical value. These data suggest that
3
4
5
6
7
8
9
10
20 200
Dilution R
atio
Particle Size (nm)
DHT = Off VDR = 5 DHT = 80 °C
VDR = 7 DHT = 150 °C VDR = 8
160 40 60 80 100 120 140 180
100
heating is required to get good agreement for particle sizes below 124.1 nm. However, at 150 °C
or DR = 8, the highest deviation was greatly reduced to 4.4%. One thing to note about this result
is that the dilution settings were kept constant except the temperature. Thus, the increase in dilution
temperature also increases the dilution ratio because the density changes. Therefore the dilution
ratio without heating was at 5 but with the increase in temperature up to 150 °C, the dilution ratio
rose to 8.
Figure 5-23
Engine Exhaust Percent Particle Penetration Corresponding to the Primary Dilution Air
Temperature for the Single-stage TSI Diluters
Figure 5-23 shows the percent particle penetration values corresponding to the measured
concentrations for all the primary dilution air temperature tested with the engine exhaust. The
resulting values indicate that, when the heating element is disabled, condensation causes growth
in the particle range 29.4 to 60.4 nm, where the particle penetration values exceed 100% (discussed
0
20
40
60
80
100
120
140
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
DHT = Off DHT = 80 °C DHT = 150 °C
40 10 15 20 25 30 35 45 100 150 200 250
101
in section 5.7). However, at 150 °C, there is an evident improvement in particle penetration below
29.4 nm and a relatively consistent percent particle penetration over all particle size bins above
29.4 nm (98.2 %). The particle condensation was thus minimized with the primary dilution air
heating.
Figure 5-24
Engine Exhaust Dilution Ratios Corresponding to the Evaporation Tube Heater
Temperature for the Two-Stage TSI Diluter
The effect of evaporation heater temperature (EHT) on the dilution ratio above 29.4 nm is
shown in Figure 5-23. The result suggests that there are no significant effects on particle dilution
ratio corresponding to the evaporation heater temperature. The overall deviation percentage at
300 °C (or On) was comparable to the percentage at Off, 6.6% to 6.3%. This suggests that there
was no significant formation of nanodroplets or volatile particles during the dilution process. Thus
the evaporation heater temperature had no influence towards the particle DRs.
60
65
70
75
80
85
90
95
100
20 200
Dilution R
atio
Particle Size (nm)
ETH = On ETH = Off VDR = 72
160 40 60 80 100 120 140 180
102
Figure 5-25
Engine Exhaust Percent Particle Penetration Corresponding to the Evaporation Tube
Heater Temperature for the Two-Stage TSI Diluter
Figure 5-25 shows the percent particle penetration values corresponding to the measured
concentrations for evaporation tube heater temperature tested with the engine exhaust. The overall
average percent particle penetration values above 29.4 nm size bin of heating temperature at 300 °C
(or On) and at room temperature (or Off: ~ 24 °C) were 98.8% and 99% respectively. As discussed,
there was no significant effects on particle dilution ratio corresponding to the evaporation heater
temperature. However, it is suggested to have the evaporation tube heater at 300 °C to actively
prevent formation of undesired particles and droplets
0
20
40
60
80
100
120
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
ETH = On ETH = Off
40 10 15 20 25 30 35 45 100 150 200 250
103
5.8.2 Dekati FPS-4000 Ejector Diluter
Figure 5-26
Engine Exhaust Dilution Ratios Corresponding to the Primary Dilution Air Temperature
for the Dekati Diluter
Figure 5-26 illustrates the particle DRs corresponding to the primary dilution air
temperature for the Dekati diluter. It should be noted that the size bins below 29.4 nm were
excluded to make the effect of dilution air heating more visible. The result indicated that as the
primary dilution air temperature was increased, the particle DRs along the recorded size bins
diverged from the theoretical dilution ratio. At DR = 14 without heating (DHT = Off), the highest
deviation neglecting the first point was 19.6% from the theoretical value. However at 300 °C, the
highest deviation was greatly increased to 45.3%. Similar to the TSI experiment, the dilution
settings were kept constant except the temperature. As the DHT increases, the particles in size bins
39.2 to 93.1 nm seems to deviate more from the theoretical.
10
15
20
25
30
35
30 300
Dilution R
atio
Particle Size (nm)
DHT = Off VDR = 14 DHT = 80 °C
VDR = 15 DHT = 150 °C VDR = 15
DHT = 300 °C VDR = 16
240 60 90 120 150 180 210 270
104
Figure 5-27
Engine Exhaust Percent Particle Penetration Corresponding to the Primary Dilution Air
Temperature for the Dekati Diluter
Figure 5-27 shows the percent particle penetration values corresponding to the measured
concentrations for all the primary dilution air temperature tested with the engine exhaust. Overall
percent particle penetration values for temperatures above 80 °C (83.3%) is comparably lower than
when the temperature was at ~ 24 °C (90.3%). However, the convergences of points at higher
temperatures are visible in the result, suggesting that volatile fractions may have been successfully
removed at temperatures above 80 °C.
5.9 Error Analysis and Discussion of Results
The expected margin of tolerance for the operating diluters was below 10% for particle
loss. All three diluters are expected to perform the dilution process with an actual DR ±10% of the
0
20
40
60
80
100
5 50
Part
icle
Pen
etra
tion (%
)
Particle Size (nm)
DHT = Off DHT = 80 °C DHT = 150 °C DHT = 300 °C
40 10 15 20 25 30 35 45 100 150 200 250
105
theoretical (user set) DR. As the results indicate, the particle loss within the diluters tested were
often below such an error margin, with minimal outliers, for particles larger than 29 nm. The
overall particle DRs of the single-stage TSI diluter indicated 6.0% discrepancies between the
measured and the user set DRs over the dilution ratio range of 12 to 100. The two-stage TSI diluter
operated within 3.9% over the DR range of 16 to 110. The Dekati diluter had less than 5.1% over
the DR range of 12 to 66. The overall averaged particle penetration value also suggested high
percent of particle penetration through all size bins. The single-stage TSI diluter operated at 98.8%
penetration value, the two-stage TSI diluter at 99.4%, and the Dekati diluter at ~100%.
As observed from the engine exhaust experiment, the particles below 29.4 nm are greatly
influenced by evaporation. The presence of such effects is indicated by the constantly higher
overall particle DRs. Under the engine exhaust experiment condition; the overall particle DR
(excluding the size bins below 29.4 nm) deviation was 6.6% for the single-stage TSI diluter, 4.3%
for the two-stage TSI diluter, and 9.6% for the Dekati diluter compared to the values stated above
(also indicated by low particle penetration below 29.4 nm). The divergence of the overall particle
dilution ratio suggests that the particles below 29.4 nm evaporated and condensed onto the larger
particles above that range. Particle growth contributes to the decrease in the downstream particle
concentrations for these smaller particles. Thus referring to the general formulation of the DR,
Equation 1-4, the computed DR values will be higher than the expected values. Such cases are
evident in the previous results where a significant portion of the particles was lost in those size
bins and the resulting DRs were magnitudes higher.
However, such effects are absent in the soot generator experiments. The solid particles
generated from the soot generator are more stable and gas-to-particle conversions can be avoided.
106
The result from this experiment, however, also indicated some particles loss, resulting in deviation
from the theoretical values.
As discussed previously, through leakage testing the consistencies of the testing conditions
were ensured. From the literature review, it was stated that the primary dilution temperature,
residence time and the relative humidity of primary dilution air could have an influence on the
dilution conditions. However, the experiment was performed under a closed environment where
the effects of such factors are minimal. Thus, the loss contributed from such changes in dilution
conditions can be neglected. Therefore, the measurement precision and the dilution accuracy are
considered as the major contributing factors to the DR discrepancies between the measured and
the expected.
The precision and accuracy of the dilution and measurement instruments are very
important in investigating particle loss. As previously mentioned, the misrepresentation of the data
can significantly affect the result.
To investigate the source of loss from the test apparatus, an error analysis was performed
on the instruments at various range of DRs measured for the mode particle size bin (10.8 nm) using
Equations 4-9 and 4-11 as shown in Table 5-4.
As shown in Table 5-4, individual uncertainties for the instruments used in the experiment
at various range of DRs measured were calculated. The analysis indicates that the uncertainties of
the particle sizers increased along with the increase in the DR. At lower DRs, the uncertainties
were 5.4%/3.1% for the single/two-stage TSI and 3.0% for the Dekati diluter. At higher DRs, the
uncertainties associated with the particle sizer for the Single-stage TSI rose to 5.9% at DR of 100,
and to 7.3% at DR of 113 for the two-stage TSI, and to 7.2% at DR of 66 for the Dekati diluter. As
107
mentioned before, a similar phenomenon was observed in the experimental results, where the
higher DRs deviated more from the expected than the lower DRs.
Instrument Percent Uncertainties
Diluters
DR
Particle Sizers
Flowmeter
Single-stage
TSI
12 5.4 2.0
50 5.8 2.0
100 5.9 2.0
Two-stage TSI
16 3.1 2.5
56 3.3 5.8
113 7.3 10.3
Dekati
12 3.0 2.0
35 3.5 2.0
66 7.2 2.0
Table 5-4
Table of Uncertainty Analysis for Experimental Apparatus
Likewise, the uncertainties from the flowmeters also increased along with the DR
increment. At lower DRs, the percent uncertainties for the single-stage TSI and Dekati diluters
were 2% and 2.6% for the two-stage TSI diluter. The flowmeter measurement for the single-stage
TSI and the Dekati diluter remained relatively constant throughout the change in DRs. Thus the
uncertainties contributing to the DR value of the single-stage TSI and the Dekati was minimal
regardless of the DR increase. The flow rate measured for the two-stage TSI diluter, however,
increased along with DR values, resulting in significantly higher uncertainties at DR = 113 (10.3%).
108
This analysis suggests that the major factor contributing to discrepancies in the experimental and
theoretical DRs measurements are, once again, from the precision and accuracy of the instruments.
It should also be noted that the uncertainties contributed from the flowmeter used with the TSI
diluters become more significant with higher flow rates at higher DRs.
The result indicated that there were some cases where the dilution ratio deviated more
than the expected error margin. As discussed previously, such outliers are caused most likely by
the effect of dilution condition change within the diluter. The primary reason for the cause of
change in dilution condition can be speculated to be a sudden fluctuation in input pressure.
Stabilization of the pressure at the sample inlet is very difficult to maintain. However, with
extended inter-sample time allowing for a longer adjustment period for the dilution ratio change
can alleviate the pressure change. Therefore, for future experiments using a longer inter-sample
time of 10 min (600 seconds) is recommended for stabilization purposes. Such a lengthy time is
especially required for the engine exhaust experiment because the compositions and effects of
diesel particles are more unpredictable than the particles generated by the soot generator.
Based on the various findings about the operating features of the diluters, a few
suggestions can be made to improve the sampling condition for the future engine emission studies.
When using the TSI 379020A rotary diluters, the primary diluter air temperature should
be set to 150 °C for better convergence of particle dilutions over the measured size bins. Use of
this higher temperature is found to be effective in reducing the overall DR deviation value by
approximately 25.6%. This suggests that the heated primary diluter air will effectively remove the
condensed volatile material from the engine exhaust. As mentioned previously, the study showed
that the evaporation heater temperature had no significant effect towards the particle loss or
109
dilution ratio. However, the purpose of the thermal conditioner is to prevent formation of
nanodroplets and volatile particles. Therefore it is suggested to keep the thermal conditioner
temperature at 300 °C consistently as a precautionary measure to reduce the formation of particles
from condensate and volatile fractions.
The qualitative observation of the dilution ratio change in response to an increase in the
primary dilution air temperature for the Dekati diluter suggested that with increase in the primary
dilution air temperature, the DR deviation value increased in size bins 39.2 to 93.1 nm. However,
the result indicated convergence of points at higher temperature for the individual size bins,
suggesting that the sampled exhaust is rich in volatile material and large portion of particles was
removed. Therefore, for the purpose of removing the volatile fraction from the engine emissions,
the heater should be kept at 80 – 150 °C.
110
Chapter 6
Conclusions and Recommendations
6.1 Conclusions
The results confirmed the ability of all three diluters tested to dilute a particle laden flow
without significant particle loss when the particles are solid carbon. Tests with a soot generator
showed average percent particle penetration values of 98.1% and 99.4% for the single and two-
stage TSI 379020A diluters respectively, and nearly 100% for the Dekati diluter.
As previously discussed in Chapter 2, the percent particle penetration value of the TSI
379020A at the 30 nm size bin was significantly lower, 68%, according to the European PMP
report (PMP report GRPE-PMP-25-5, 2010). The European PMP test was conducted using diesel
engine exhaust as the sample. The findings of this work also showed similar percent penetration
to the PMP test for engine exhaust particles below 29.4 nm. However, the result obtained with the
soot generator indicated ~ 100 % penetration of particles smaller than 29.4 nm for the single-stage
and 94.2% for the two-stage diluters.
The different response to particles produced by the engine and by the soot generator
suggests that when the particle laden flow contains significant amount of liquid droplets (e.g.
engine exhaust), there can be a substantial loss of very small particles (< 29.4 nm diameter). This
loss is accompanied by an increase in number of particle in the size ranges 29.4 to 52.3 nm,
suggesting that the liquid particles agglomerate, coagulate, or condense on larger particles.
Sample flow through all of the diluters depends on the pressure difference between the
111
sample inlet and the diluter outlet. Consequently the performance of all three diluters was affected
by the actual pressure difference encountered in use. The two TSI 3709020A diluters were not as
sensitive to pressure difference as the Dekati FPS-4000 diluter because they have a controlled
sample pump. However, the theoretical (user set) DR given by the system was deviating ±20%
from the measured DR. Use of a supplemental flowmeter to measure the TSI diluter sample flow
and excess flow and then using these to compute the corrected DR was found to improve the DR
accuracy from ±10% to ±5%.
The Dekati FPS-4000 diluter uses an air ejector to pump the sample flow and is therefore
extremely sensitive to the pressure difference between the sample point and the diluter outlet
(generally at atmospheric pressure). Great care must be taken in use of the Dekati FPS-4000 diluter
to ensure that the pressure difference encountered in a test (i.e. engine exhaust system pressure and
the diluter outlet pressure) are within the operating range of the Dekati diluter. In this work, the
Dekati FPS-4000 performed well (~100% percent penetration value) because precautions were
taken to ensure minimal pressure difference across the diluter.
The different characteristics exhibited by the two diluter designs (TSI 379020A and Dekati
FPS-4000) suggest that there are specific applications for which each diluter is best suited. The
TSI 379020A diluters are more suitable for uses in simple engine exhaust emission studies where
the accurate measurement of concentration with numerous repeated sessions is critical. The Dekati
FPS-400 diluter is more suitable for use in nucleation and particle dynamic studies where dilution
condition variability and controlling mechanisms are vital.
6.2 Recommendations
Prior to operating the diluters discussed in this thesis (the single and two-stage TSI
112
379020A rotary thermodiluters and the Dekati FPS-4000 ejector diluter), there are a few
preparatory steps that must be followed in order to achieve satisfactory results.
For best results, the actual volumetric dilution ratio obtained in use should be continuously
monitored by the use of a tracer gas. This is especially true for the Dekati diluter due to its extreme
sensitivity to sample inlet (engine exhaust) and diluter outlet pressures. Carbon dioxide, a product
of combustion of all hydrocarbon fuels and therefore present in engine exhaust in large
concentrations, is the obvious choice to use as a tracer gas since it is a highly superheated vapor in
the exhaust as well as non-reactive. It is therefore unchanged by any physical or chemical processes
that might occur in the diluter. The carbon dioxide concentration in engine exhaust is routinely
measured by exhaust emissions analyzers. However, the low concentration of carbon dioxide after
dilution requires the use of a different CO2 analyzer designed for the appropriate concentration
range, since the exhaust emissions CO2 analyzer is not accurate at low concentrations. Quantifying
the actual volumetric dilution ratio through use of a CO2 tracer will ensure accurate operation of
the diluters.
For high dilution ratios, the diluted concentration can approach the concentration of
carbon dioxide in the ambient air. Since ambient air is used for dilution, the local background
CO2 concentration must be measured and taken into account in calculating the measured dilution
ratio. For example, the particle-free aerosol generated internally from room air by the two-stage
diluter also introduced undesired supplementary 471.4 ppm of CO2 from the room. Therefore,
471.4 ppm was subtracted from the successive measured mean downstream concentration for the
two-stage diluter. The reason for the discrepancy between the atmospheric CO2 level (398.58 ppm)
and this measured concentration was because the experiment was conducted in a closed
113
environment filled with potential CO2 sources including people and instruments. Therefore, it is
important to measure the CO2 concentration level specific to the local environment prior to
conducting any experiments.
The primary diluter air temperature of the TSI 379020A diluters should be set to 150 °C
for better convergence of particle dilutions over the measured size bins. This temperature was
found to be effective in reducing the overall DR deviation value by approximately 25.6%. It is also
suggested to keep the thermal conditioner temperature at 300 °C consistently to reduce the
formation of condensate and volatile fraction particles.
Although the deviation of the measured DR increased with a primary dilution air
temperature increase, the heater should be kept in the range 80 – 150 °C. The result indicated
convergence of points at higher temperature for the individual size bins, suggesting that the
sampled exhaust was rich in volatile material and large portion of particles was removed.
The results showed that the diluters tested in this thesis are expected to perform dilution
within ±10% of the theoretical dilution ratio in both volumetric and particle phases as stated by
the manufacturer. However, outliers are present when pressure fluctuations or a change in dilution
conditions occur. To maintain relatively consistent dilution condition for improved dilution
precision and accuracy, it is suggested that a longer inter-sample time of 10 min be implemented
for stabilization purpose for the future experiments.
Due to concern about its pressure sensitivity, further study of the Dekati FPS-4000 diluter
under engine sampling conditions is warranted to identify the limits of its exhaust pressure
capability.
114
Appendices
Appendix A
Preliminary Experiment Result for Single-Stage TSI 379020A using
Methane Gas
According to the TSI 379020A manual (TSI, 2011), substantial loss of gas phase organic
compounds can occur in rotary disk thermodiluters, where the cause of such losses are unclear. In
order to verify that such phenomenon indeed occurs, the preliminary experiment, discussed in
Chapter 4, was conducted using a mixture of 1200 ppm (nominal) of methane balanced in nitrogen
as the upstream sample. The experiment layout used in this test followed the one illustrated in
Figure 4-1 (left for TSI 379020A).
With known concentration of the gas mixture at upstream, the downstream concentration
was measured using the FHID analyzer. The result of the concentration measurements is shown in
Figure B-1.
The measured concentrations were well below the theoretical concentration at various
DRs measured, with the exception of the first two points. It should be noted that even the first two
points fail to fall within the expected error margin of ± 10%. The results indicate that the increase
in deviation was directly proportional to the increase in DR. The graph showed a divergence of
measured concentration from the ideal values as the DR values got higher. At higher DR values,
the TSI 379020A diluter was over-diluting the mixture of methane gas to a point where almost
entire volume of gas was lost (~ 0.4 ppm).
115
Figure A-1
Diluted CH4 Concentration Measurement at Various Dilution Ratios for the single-stage
TSI Diluter
Thus, it was verified that the substantial loss of gas phase organic compounds occur in
rotary disk thermodiluters. Therefore, to evaluate these TSI diluter measurements, a mixture of
2.15% (nominal) carbon dioxide balanced in nitrogen was used as the sampling gas.
0
10
20
30
40
50
60
70
80
90
100
10 20 30 40 50 60 70 80 90 100
CH
4C
once
ntr
ation [ppm
]
Dilution Ratio
Measured DR Theoretical DR
116
Appendix B
Derivation of Uncertainty Analysis Equations
The discussion of uncertainty that follows is based on the method presented in Theories
of Engineering Experimentation (Schenck, 1974). The term s2 is defined as the deviation envelope
that encloses 95 percent of all the readings, also known as the criterion of “twenty-to-one odds”.
This means that this value is the numerical expression for the size of an envelope that has twenty-
to-one odds of containing the true value, or simply as standard deviation. The m�j (also known as
standard deviation) for any function R of measured variables X, Y, Z, etc. can be expressed as
m�j � ��B�g�hj , … mgj � ��B�h�g
j , … mhj � ��B�i�hj , … mij �⋯ Equation B-1
where mg, mh,mi, etc., are the variance for the measured variables X, Y, Z, etc…(Schenck, 1974).
With the assumption that the measured variables are independent of one another and the function
R is of the general form
R � k ghi , Equation B-2
then the Equation B-1 can be re-expressed as
m�j � �� hi�j mgj � �� gi�
j mhj � ��� ghi��j mij Equation B-3
Dividing this equation by R2, it can be simplified to
�&B�j � �kg �
j � � h �j � �li �
j Equation B-4
The terms kg ,
h , and
li represent percentage error of the individual measurements. Unless
117
stated otherwise, this term will be referred to as the “uncertainty” of the measured variable. Then
the term &B represents the percent error of the result.
118
References
Abdul-Khalek IS and Kittleson DB. 1999. The influence of dilution conditions on diesel exhaust
particle size distribution measurements. SAE Technical Paper 1999-01-1142.
Abu-Allaba M, Coulomb W, Gertler AW, Gillies J, Pierson WR, Rogers CF, Sagebiel JC, Tarnay
L. 2002. Exhaust particle size distribution measurements at the Tuscarora mountain tunnel.
Aerosol Science and Technology 36:
Air Supply/Thermal Conditioner ASET15-1. 2005. Matter Engineering AG. SKM 04127-21.
Andersson J, Giechaskiel B, Muñox-Bueno R, Sandbach E, Dilara P. 2007. Particle measurement
propgramme (PMP) light-duty inter-laboratory correlation exercise (ILCE_LD) final
report. EUR 22775 EN.
Andersson J, Mamakos A, Giechaskiel B, Carriero M, Martini G. 2010. Particle measurement
propgramme (PMP) heavy-duty inter-laboratory correlation exercise (ILCE_HD) final
report. GRPE-PMP-25-05.
Arnold F, Pirjola L, Aufmhoff H, Schuck T. Lähde T, Hämeri K. 2006. First gaseous sulfuric acid
measurement in automobile exhaust: Implications for volatile nanoparticle formation.
Atmospheric Environment 40.
Atmospheric CO2 for July 2013. Mauna Loa Observatory: NOAA-ESRL [Internet]. c2013.
Available from: http://co2now.org.
Baron PA and Willeke K. 1993. Aerosol measurement: principles, techniques, and applications.
van Nostrand Reinhold. New York.
119
Burtscher H. 2005. Physcial characterization of particulate emissions from diesel engines: A
review Journal of Aerosol Science 36(7)
Collins AM and Kittelson DB. 2010. Characterization and reduction of particle loss in aerosol
diluters. AFS Abstract. University of Minnesota: Department of Mechanical Engineering.
Collins AM. 2010. Ultrafine particle loss in aerosol diluters. M.A.Sc. Thesis. University of
Minnesota: Department of Mechanical Engineering.
Collings N and Graskow BR. 2000. Particles from internal combustion engeins – what we need to
know. Philos. Trans. R. Soc. Lond. Ser. A-Math. Phys. Eng. Sci. 358:2611-2622
Dekati FPS-4000 fine particle sampler. 2010. Dekati, Ltd.
Dockery DW, Pope CA 3rd, Xu X, Spengler JD, Ware JH, Fay ME, Ferris BG, Speizer FE. 1993.
An association between air pollution and mortality in six U.S. cities. N. Engl. J. Med.
329(24):1753-9.
Engine Exhaust Particle Sizer spectrometer model 3090. 2012. TSI, Inc. P/N 2980244, Rev. B.
Froines JR. 2005. Ultrafine particle health effects. Southern California Particle Center. Available
from: http://www.aqmd.gov/tao/ultrafine_presentations/Pre-Conference_2_Froines.pdf.
FPS-4000 user manual ver. 6.00. 2010. Dekati, Ltd.
Grose M, Sakurai H, Savstrom J, Stolzenburg M, Watts W, Morgan C, Murray I, Twigg M,
Kittleson DB, McMurry P. 2006. Chemical and physical properties of ultrafine diesel
exhaust particles sampled downstream of a catalytic trap. Environmental Science and
Technology 40:5502-5507.
120
Heeb NV, Ulrich A, Emmenegger L, Czerwinski J, Mayer A, Wyser M. 2005. Secondary
emissions risk assessment of diesel particulate traps for heavy duty applications. SAE
International SIAT 2005-ABS-165.
Heikkilä J, Virtanen A, Rönkkö T, Keskinen J, Aakko-Saksa P, Murtonen M. 2009. Nanoparticle
emissions from a heavy-duty engine running on alternative diesel fuels. . Environmental
Science and Technology 43:9501-6.
Herner JD, Robertson WH, Ayala A. 2007. Investigation of ultrafine particle number
measurements from a clean diesel truck using the European PMP protocol. SAE
Technical Paper 2007-01-1114.
Heywood JB. 1988. Internal combustion engine fundamentals. United States of America:
McGraw-Hill, Inc.
Hoek G, Brunekreef B, Goldbohm S, Fischer P, van den Brandt PA. 2002. Association between
mortality and indicators of traffic-related air pollution in the Netherlands: a cohort study.
Lancet. 360:1203-1209.
Johnson JH, Bagley ST, Gratz LD, Leddy DG. 1994. A review of diesel particulate control
technology and emission effects. 1992 Horning Memorial Award Lecture. SAE Paper
940233.
Johnson TV. 2006. Diesel emission control in review. SAE Technical Paper 2006-01-0233.
Johnson TV. 2007. Diesel emission control in review. SAE Technical Paper 2007-01-0233.
Johnson TV. 2008. Diesel emission control in review. SAE Technical Paper 2008-01-0069.
121
Khalek IA. 2006. The particulars of diesel particle emissions: new research looks at particle
numbers and size as well as mass. Technology Today. Available from:
http://www.swri.org/3pubs/ttoday/spring06/Particulars.htm.
Khalek IA, Kittelson DB, Brear F. 2000. Nanoparticle growth during dilution and cooling of diesel
exhaust: experimental investigation and theoretical assessment. SAE Technical Paper
2000-01-0515.
Kittelson DB. 1998. Engines and nanoparticles: A review. J Aerosol Sci. 29:575-588.
Kittleson DB, Arnold M, Watts W. 1999. Review of diesel particulate matter sampling methods.
EPA Report. University of Minnesota: Department of Mechanical Engineering.
Kittleson DB, Watts W, Johnson J. 2002. Diesel aerosol sampling methodology – CRC E-43:
Final report. University of Minnesota: Report for the Coordinating Research Council.
Kittleson DB. 2001. Recent measurements of nanoparticle emissions from engines. University of
Minnesota: Department of Mechanical Engineering.
Kulmala M. 2003. How particles nucleate and grow. Science 302:1000-1001.
Kwon SB, Lee KW, Saito K, Shinozaki O, Seto T. 2003. Size-dependent volatility of diesel
nanoparticles: chassis dynamometer experiments. Environ. Sci. Technol. 37:1794-1802.
Lang MC, Dhaniyala S, Yapa SD. 2008. Characterization of electrospray aerosol generator.
Summer Project Report. Clarkson University: Department of Mechanical Engineering.
122
Lyyränen J, Jokiniemi J, Kauppinen EI, Backman U, Vesala H. 2004. Comparison of different
dilution methods for measuring diesel particle emissions. Aerosol Sci. Thecnol. 3:327-
334.
Ma H, Jung H, Kittleson DB. 2008. Investigation of diesel nanoparticle nucleation mechanism.
Aerosol Science and Technology 42:335-342.
Mathis U, Ristimäki J, Mohr M, Keskinen J, Ntziachristos L, Samaras Z, Mikkanen P. 2004.
Sampling conditions for the measurement of nucleation mode particles in the exhaust of a
diesel vehicle. Aerosol Science and Technology 38:1149-1169.
Mathis U, Mohr M, Zenobi R. 2004. Effect of organic compounds on nanoparticle formation in
diluted diesel exhaust. Atmos. Chem. Phys. 4:609-620.
miniCAST Series 6200. Jing Ltd. [Internet]. c2013. Available from:
http://www.sootgenerator.com/miniCAST_g.htm.
Model 3091 Fast Mobility Particle Sizer spectrometer. 2004. TSI, Inc. P/N 2980290.
Model 3480 electrospray aerosol generator manual. 2003. TSI, Inc.
Model 379020A rotating disk thermodiluter operation and service manual. 2011. TSI, Inc. P/N
6002726, Rev. A.
NGA 2000 heated flame ionization detector analyzer module. 1999. Rosemount Analytical, Inc.
748297-D.
123
Ntziachristos L, Samaras Z, Mohr M, Mathis U, Keskinen J, Ristimäki J, Mikkanen P, Vogt R.
2004b. Performance evaluation of a novel sampling and measurement system for exhaust
particle characterization. SAE Technical Paper 2004-01-1439.
Oberdörster, G. 2000. Pulmonary effects of inhaled ultrafine particles. International Archives of
Occupational and Environmental Health, 74(1), 1-8.
Park S, Kim H, Choi B. 2009. Emission characteristics of exhaust gases and nanoparticles from a
diesel engine with biodiesel-diesel blended fuel (BD20). Journal of Mechanical Science
and Technology, 23:2555-2564.
Rotating disk thermodiluters and thermal conditioners. 2009. TSI, Inc. P/N 2980487, Rev. B.
Sakurai H, Park K, McMurry PH, Zarling DD, Kittleson DB, Ziemann PJ. 2003. Size-dependent
mixing characteristics of volatile and nonvolatile components in diesel exhaust aerosols.
Environmental Science and Technology 37:5487-5495.
Sappok AG and Wong VW. 2007. Detailed chemical and physical characterization of ash species
in diesel exhaust entering aftertreatment systems. SAE Technical Paper 2007-01-0318.
Schenck H. 1979. Theories of engineering experimentation. Hemisphere Publishing Corporation.
New York
Schneider J, Hock N, Weimer S, Borrmann S, Kirchner U, Vogt R, Scheer V. 2005.
Nucleation particle in diesel exhaust: Composition inferred from in situ mass
spectrometric analysis. Environental Science and Technology 39:6159-6161.
Seinfeld JH and Pandis SN. 2000. Atmospheric chemistry and physics: from air pollution to
climate Change. Wiley & Sons. New York.
124
Shi JP and Harrison RM. 1999. Investigation of ultrafine particle formation during diesel exhaust
dilution. Environ. Sci. Technol. 33:3730-3736.
Smith OI. 1981. Fundamentals of soot formation in flames with application to diesel engine
particulate emissions. Prog Energy Combust Sci 7:275-291.
Stoeger T, Reinhard C, Takenaka S, Schroeppel A, Karg E, Ritter B, Heyder J, Schulz H. 2006.
Instillation of six different ultrafine carbon particles indicates a surface area threshold dose
for acute lung inflammation in mice. Environmental Health Perspectives, 114(3), 328.
The LI-820 CO2 gas analyzer. 2009. LI-COR, Inc. DOC#980-07505, Rev. 1.
Tyree C and Allen J. 2004. Diffusional particle loss upstream of isokinetic sampling inlets. Aerosol
Science and Technology 38:1019-1026.
U.S. Environmental Protection Agency. 2002. A comprehensive analysis of biodiesel impacts on
exhaust emissions Report. EPA420-P-02-001.
Venn AJ, Lewis SA, Cooper M, Hubbard R, Britton J. 2001. Living near a main road and the risk
of wheezing illness in children. Am. J. Crit. Care Med. 164:2177-2180.
Yu FQ. 2001. Chemiions and nanoparticle formation in diesel engine exhaust. Geophys. Res. Lett.
28:4191-4194.
Yu FQ. 2002. Chemiion evolution in motor vehicle exhaust: further evidence of its role in
nanoparticle formation. Geophys. Res. Lett. 29:1717.
125
Zhang KM and Wexler AS 2004. Evolution of particle number distribution near roadways – part
I: Analysis of aerosol dynamics and its implications for engine emission measurement.
Atmospheric Environment 38:6643-6653
Zheng Z, Durbin TD, Karavalakis G, Johnson KC, Chaudhary A, Cocker III DR, Herner JD,
Robertson WH, Huai T, Ayala A, Kittelson DB, Jung HS. 2012. Nature of Sub-23-nm
particles downstream of the European Particle Measurement Programme (PMP)-
compliant system: a real-time data perspective. Aerosol Science and Technology 46:886-
896.
Zimmerman N, Godri-Pollitt K, Jeong C, Jung T, Cooper J, Wallace J, Evans G, 2013. Inter-
Comparison of three nanoparticle sizing instruments measuring urban ambient, laboratory
generated, and diesel engine particles. Submitted to Atmospheric Environment.