Evaluation and Implementation of
Machine-assisted Slide Processing
Charlotte GregorySouth-East Scotland Cytogenetics Service
Western General
Edinburgh
Time Consuming
Labour Intensive
Inconsistent
Messy
Slide Processing
Genetix SL-120
Metaphase Finder/Analysis
Transgenomics Hanabi
Automatic Harvester
Aims of the Project
• Integrate new equipment
(Set up/Programming/SOPS/Training)
• Compare quality and consistency to manual method
• Time, Space, financial implications
Thermotron (Fisher)
Environmental Chamber
• Controlled Environment
• Gives Consistency
• Facilitates Spreading
• Optimum values?
TemperatureOC
RelativeHumidity %
%Metaphase ≥ QA5
B F
AverageCrossovers
23.8 Control in Hood
45 23 27 8 7
20 50 50 50 9 6
25 50 49 56 5 6
30 30 50 33 9 8
Variables
•20, 25, 30oC
•20, 30, 40, 50 ,60% RH
for each temperature as above
Varistain Gemini (Shandon)
Automatic Stainer
• Robotic arm/troughs
• Multiple Samples
• Accurate Timing
• Heat/Running Water
Postnatal
Pre-treatment
Hot Plate/Peroxide
Time 1hr/30mins
TrypsinConcentration
1/10
Time 50s
PBS wash 5s + agitation
Leishman’sConcentration
1/4
Time 2.5m
Gurr’sWash
5 secs +agitation
•Postnatal
•Prenatal
•Tissue
•Haematology
•Tumour
Adapted from Manual Protocol
1m 1m 1m5m
Load Trypsin PBS Stain Gurr’s Store
1 rack every 8 minutes (32 minutes for 4)
Previous Version
Updated Version
Load Store Trypsin PBS Stain Gurr’s Store
1m 1m 1m
5m
5m
5m
5m
4 racks at same time (11 minutes for 4)
Clearvue (Shandon)
Automatic Coverslipper
• Daily maintenance
• Preloaded Coverslips
• Humid atmosphere
• No pressure
Methods % Cells ≥ QA5
Spreading Banding
Thermo/Man 67 OK OK/S. Under
Thermo/Gem 62 OK OK
Man/Gem 56 Thick S. Over
Man/Man 53 OK OK pale
•Pattern consistent in all cases
•Pattern consistent for all users
•Thermo/Man best quality due to not batch banding
•Thermo/Gemini good, quicker, more consistent
SIZE M2
Per Unit TotalSIZE M2
Per Unit Total
MonoairHood
0.965x2
1.93 ClearvueH 500D 575W 645
0.37+hood
0.965
SensairHood
0.82x2
1.64 ThermotronH 1425D 715W 1095
0.78X2
1.56
Worktop 0.57x4
2.28 VaristainH 880D 710W 720
0.51+bench
1.12
Total 5.85 3.65
Manual MachineTime Walk away Time Walk away
Mon 3h30 1h 3h30 1h30
Tues 7h20 1h 3h20 1h55
Wed 6h25 1h 6h20 1h35
Thur 6h 1h 2h05 2h
Fri 4h55 1h 4h25 1h45
TIME 29h10 5h 19h40 8h45
TOTAL 24h10 10h55
PBS ml
Man Auto
BloodFluidOncology
299200390
380
Total
Litres
2.5 1.9
Cost £2.08 £1.57
•Gurr's
•Leishman's
•Trypsin
•Coverslip
•Mountant
Automatic method costs ~£80 more, per year
£• Saving: 689 MTO hours per year
• Extra cost: Machines + Extra Reagent + Maintenance
• Overall Saving• £6000 a year
QA 6 6Crossovers 8 5Space Needed 5.85m2 3.65m2
Time taken 4.72mins 2.11minsCost £1.76 £1.52
ConclusionStatistics on an average slide basis