European Standards Laboratory CALIBRATION-CERTIFICATE
Yokogawa Europe Solutions B.V.
Amersfoort, The NetherlandsCertificate number 58052P10722
Page 1 of 43
Applicant Demo
Instrument WT1800 Precision Power Analyzer
Manufacturer Yokogawa
Type WT1806-06-F-HE/EX6/G6/FQ/MTR/EC
Serial number
Inventory number CCE_1_843
ID numbers used standards CSE820 CSE976 CSE981 CSE965 CSE1029
Procedure CALWT1800_50A_50A_50A_50A_50A_50A Version: 1 of 28-03-2017
Calibration method The Powermeter was compared to the Yokogawa Power Calibration System
via a phantom technique. During the calibration the distortion of the
applied voltage and current were below 0.1 %. Before calibration the
device was powered on for at least 12 hours.
Environmental conditions Temperature (23.7 ± 1) ºC
Relative Humidity (52 ± 4) %rh
Date of Calibration 26 October 2017
Result The results of the calibration are shown on the next pages.
The reported expanded uncertainty of measurement is stated as the standard
uncertainty of measurement multiplied by the coverage factor 2 such that the coverage
probability corresponds to approximately 95%.
The standard uncertainty of measurement has been determined in accordance with the EA-4/02 M:2013.
The long term stability of the calibrated object is not included in
the reported expanded uncertainty measurement.
This certificate of calibration is issued in compliance with ISO/IEC 17025:2005
Traceability The measurements have been executed using standards for which the traceability
to (inter)national standards has been demonstrated towards the RvA.
Date 26 October 2017
Name N. el Morabit
Function Calibration Engineer
Yokogawa Europe Solutions B.V. The Raad voor Accreditatie is member of the This certificate is issued provided neither the Yokogawa Standards
European Standards Laboratory European Co-operation for Accreditation (EA) Laboratory nor the Raad voor Accreditatie assumes any liability.
P.O. Box 163, 3800 AD Amersfoort (NL) and is one of the signatories to the ILAC Reproduction of the complete certificate is permitted.
Phone +31 (0)88 46410000 Mutual Recognition Arrangements (MRA) for the Parts of the certificate may only be reproduced after written
website http/tmi.yokogawa.com mutual recognition of calibration Certificates. permission of Yokogawa.
Certificate number 58052P10722
Page: 2 of 43
E L E M E N T 1
Voltage Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.98 0.02 V DC -0.016
100V -100.00 -100.00 0.02 V DC -0.001
Voltage Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V AC -0.012
Voltage Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1.5V 1.5000 1.4997 0.0006 V AC -0.021
3V 3.0000 2.9995 0.0004 V AC -0.015
6V 6.0000 5.9995 0.0009 V AC -0.009
10V 10.000 9.999 0.002 V AC -0.015
15V 15.000 14.999 0.002 V AC -0.009
30V 30.000 29.997 0.002 V AC -0.009
60V 60.000 59.994 0.002 V AC -0.010
100V 100.00 99.99 0.02 V AC -0.013
150V 150.00 149.99 0.02 V AC -0.009
300V 300.00 299.97 0.02 V AC -0.010
600V 600.00 599.98 0.02 V AC -0.004
1000V 1.0000 1.0001 0.0002 kV AC 0.006
Voltage Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0011 0.0010 V AC 0.019
100V 100.00 100.00 0.06 V AC 0.004
150V 100.00 99.97 0.06 V AC -0.028
Voltage Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0035 0.0023 V AC 0.058
100V 100.00 100.04 0.14 V AC 0.041
150V 100.00 100.06 0.14 V AC 0.059
Voltage Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.000 6.003 0.010 V AC 0.044
100V 100.00 100.03 0.16 V AC 0.026
150V 100.00 100.16 0.16 V AC 0.164
Certificate number 58052P10722
Page: 3 of 43
Voltage Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V DC -0.010
100V -1.00 -1.00 0.02 V DC 0.020
100V 50.00 50.00 0.02 V DC -0.005
100V -50.00 -50.00 0.02 V DC -0.001
100V 110.00 109.98 0.02 V DC -0.018
100V -110.00 -110.00 0.02 V DC -0.003
Voltage Calibration Linearity at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V AC -0.022
100V 50.00 50.00 0.02 V AC -0.006
100V 110.00 109.99 0.02 V AC -0.012
Current Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
5A 5.000 5.000 0.002 A DC -0.006
5A -5.000 -5.000 0.002 A DC -0.010
Current Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0004 A AC -0.006
Current Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0002 A AC -0.010
2A 2.0000 1.9998 0.0002 A AC -0.011
5A 5.0000 4.9995 0.0002 A AC -0.010
10A 10.000 9.999 0.002 A AC -0.009
20A 18.000 18.000 0.006 A AC 0.001
50A 18.000 18.000 0.006 A AC 0.001
Current Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0005 0.0005 A AC 0.049
5A 5.000 5.002 0.008 A AC 0.046
10A 10.000 10.050 0.012 A AC 0.502
20A 10.000 10.052 0.012 A AC 0.517
Current Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9938 0.0004 A AC -0.618
Certificate number 58052P10722
Page: 4 of 43
Current Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9874 0.0007 A AC -1.255
5A 5.000 4.956 0.019 A AC -0.888
20A 5.000 5.012 0.019 A AC 0.243
Current Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0101 0.0002 A DC 1.340
1A -0.0100 -0.0099 0.0002 A DC -1.004
1A 0.5000 0.5004 0.0004 A DC 0.088
1A -0.5000 -0.4997 0.0005 A DC -0.067
1A 1.1000 1.1001 0.0004 A DC 0.010
1A -1.1000 -1.0995 0.0002 A DC -0.047
Current Calibration Linearity AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A AC -0.012
1A 0.5000 0.5000 0.0002 A AC -0.004
1A 1.1000 1.0999 0.0002 A AC -0.012
External Current Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9999 0.0002 A DC -0.007
1V -1.0000 -1.0000 0.0002 A DC -0.003
External Current Calibration 10 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A AC -0.001
External Current Calibration 60 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50.00 50.00 0.06 mA AC 0.003
100mV 100.00 100.00 0.07 mA AC -0.001
200mV 200.00 199.99 0.13 mA AC -0.004
500mV 500.00 500.00 0.09 mA AC 0.000
1V 1.0000 1.0000 0.0002 A AC -0.003
2V 2.0000 2.0000 0.0006 A AC -0.002
5V 5.0000 5.0000 0.0009 A AC 0.000
External Current Calibration 10 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A AC -0.003
External Current Calibration 50 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9993 0.0005 A AC -0.073
Certificate number 58052P10722
Page: 5 of 43
External Current Calibration 100 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50 50 2 mA AC 0.057
1V 1.0000 1.0005 0.0013 A AC 0.053
Power Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.87 0.15 W -0.027
100V 5A -500.00 -499.80 0.17 W -0.040
Power Calibration 10Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
30V 1A 30.00 29.99 0.02 W -0.019
Power Calibration 60Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 2A 200.00 199.96 0.04 W -0.018
100V 5A 500.00 499.91 0.06 W -0.018
100V 10A 1.0000 0.9999 0.0002 kW -0.015
100V 20A 1.8000 1.7999 0.0008 kW -0.007
100V 50A 1.8000 1.7999 0.0008 kW -0.004
1.5V 1A 1.5000 1.4999 0.0023 W -0.008
3V 1A 3.000 3.000 0.005 W -0.010
6V 1A 6.000 5.999 0.009 W -0.015
10V 1A 10.000 10.002 0.002 W 0.025
15V 1A 15.000 14.997 0.002 W -0.022
30V 1A 30.000 29.993 0.005 W -0.023
60V 1A 60.000 59.988 0.009 W -0.020
100V 1A 100.00 99.98 0.02 W -0.018
150V 1A 150.00 149.97 0.02 W -0.018
300V 1A 300.00 299.94 0.05 W -0.021
600V 1A 600.00 599.92 0.10 W -0.013
1000V 1A 1.0000 0.9999 0.0002 kW -0.005
Certificate number 58052P10722
Page: 6 of 43
Power Calibration 10 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.31 0.17 W 0.061
100V 20A 1.0000 1.0058 0.0020 kW 0.581
6V 1A 6.0000 6.0021 0.0029 W 0.035
100V 1A 100.00 100.03 0.05 W 0.033
150V 1A 100.00 100.00 0.05 W -0.005
Power Calibration 50 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.22 0.17 W 0.043
100V 20A 1.0000 1.0057 0.0020 kW 0.570
6V 1A 6.000 5.953 0.011 W -0.789
100V 1A 100.00 99.38 0.10 W -0.622
150V 1A 100.00 99.37 0.10 W -0.627
Power Calibration 100 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.22 0.17 W 0.043
100V 20A 1.0000 1.0056 0.0020 kW 0.559
100V 1A 100.0 98.8 0.3 W -1.217
150V 1A 100.0 98.9 0.3 W -1.114
Power Calibration Linearity DC PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.035
100V 1A -1.00 -1.00 0.02 W 0.015
100V 1A 50.00 50.00 0.02 W 0.008
100V 1A -50.00 -50.00 0.02 W -0.008
100V 1A 110.00 109.99 0.04 W -0.005
100V 1A -110.00 -110.00 0.04 W 0.001
Certificate number 58052P10722
Page: 7 of 43
Power Calibration Linearity DC PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.01 0.02 W 0.662
100V 1A -1.00 -0.99 0.02 W -0.998
100V 1A 50.00 49.99 0.02 W -0.010
100V 1A -50.00 -49.99 0.02 W -0.028
100V 1A 110.00 109.97 0.02 W -0.031
100V 1A -110.00 -109.95 0.02 W -0.041
Power Calibration Linearity 60 Hz PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.027
100V 1A 50.00 49.99 0.02 W -0.020
100V 1A 110.00 109.98 0.02 W -0.020
Power Calibration Linearity 60 Hz PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.012
100V 1A 50.00 50.00 0.02 W -0.008
100V 1A 110.00 109.97 0.02 W -0.024
External Current Power Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.98 0.02 W -0.029
60V 1V -60.00 -60.00 0.02 W 0.006
External Current Power Calibration 10 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.017
External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 500mV 30.00 30.00 0.05 W -0.011
60V 1V 60.00 59.99 0.02 W -0.017
60V 5V 300.00 299.97 0.12 W -0.011
Power Calibration 60Hz PF=0 Leg
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.01 0.02 W -
Power Calibration 60Hz PF=0 Lead
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Power Calibration 60Hz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.01 0.02 W -
Power Calibration 60Hz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.01 0.02 W -
Certificate number 58052P10722
Page: 8 of 43
Power Calibration 1 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.06 0.12 W -
Power Calibration 10 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.13 0.24 W -
Power Calibration 50 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.8 0.7 W -
Power Calibration 100 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.1 1.2 W -
Power Calibration 1 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.07 0.12 W -
Power Calibration 10 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.26 0.24 W -
Power Calibration 50 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.8 0.6 W -
Power Calibration 100 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.0 1.2 W -
Frequency Calibration
Range Applied Measured ±Uncertainty Unit Deviation %
15V 60.0 60.0 1.2 Hz 0.000
15V 1.0000 1.0000 0.0013 kHz 0.000
15V 10.000 10.000 0.002 kHz 0.000
15V 50.000 50.000 0.002 kHz 0.000
Certificate number 58052P10722
Page: 9 of 43
E L E M E N T 2
Voltage Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V DC -0.010
100V -100.00 -99.99 0.02 V DC -0.010
Voltage Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V AC -0.012
Voltage Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1.5V 1.5000 1.4997 0.0006 V AC -0.018
3V 3.0000 2.9997 0.0005 V AC -0.009
6V 6.0000 5.9994 0.0009 V AC -0.010
10V 10.000 9.999 0.002 V AC -0.013
15V 15.000 14.999 0.002 V AC -0.009
30V 30.000 29.997 0.002 V AC -0.010
60V 60.000 59.996 0.002 V AC -0.007
100V 100.00 99.99 0.02 V AC -0.010
150V 150.00 149.99 0.02 V AC -0.010
300V 300.00 299.97 0.02 V AC -0.011
600V 600.00 599.97 0.04 V AC -0.005
1000V 1.0000 1.0001 0.0002 kV AC 0.006
Voltage Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0024 0.0010 V AC 0.039
100V 100.00 100.02 0.06 V AC 0.024
150V 100.00 99.99 0.06 V AC -0.009
Voltage Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0032 0.0023 V AC 0.054
100V 100.00 100.02 0.14 V AC 0.021
150V 100.00 100.04 0.14 V AC 0.040
Voltage Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.000 6.002 0.010 V AC 0.031
100V 100.00 100.01 0.16 V AC 0.010
150V 100.00 100.07 0.16 V AC 0.067
Certificate number 58052P10722
Page: 10 of 43
Voltage Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V DC 0.323
100V -1.00 -0.99 0.02 V DC -0.647
100V 50.00 50.00 0.02 V DC 0.001
100V -50.00 -49.99 0.02 V DC -0.014
100V 110.00 109.99 0.02 V DC -0.009
100V -110.00 -109.99 0.02 V DC -0.009
Voltage Calibration Linearity at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V AC -0.022
100V 50.00 50.00 0.02 V AC -0.004
100V 110.00 109.99 0.02 V AC -0.006
Current Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
5A 5.000 5.000 0.002 A DC -0.004
5A -5.000 -4.999 0.002 A DC -0.013
Current Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0004 A AC -0.010
Current Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0002 A AC -0.012
2A 2.0000 1.9998 0.0002 A AC -0.009
5A 5.0000 4.9992 0.0002 A AC -0.015
10A 10.000 9.999 0.002 A AC -0.013
20A 18.000 17.999 0.006 A AC -0.003
50A 18.000 17.999 0.006 A AC -0.005
Current Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9998 0.0005 A AC -0.023
5A 5.000 4.999 0.008 A AC -0.026
10A 10.000 10.038 0.012 A AC 0.382
20A 10.000 10.039 0.012 A AC 0.395
Current Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9873 0.0004 A AC -1.269
Certificate number 58052P10722
Page: 11 of 43
Current Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9869 0.0007 A AC -1.312
5A 5.000 4.960 0.019 A AC -0.802
20A 5.000 5.010 0.019 A AC 0.205
Current Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A DC 0.329
1A -0.0100 -0.0100 0.0002 A DC 0.003
1A 0.5000 0.5003 0.0004 A DC 0.060
1A -0.5000 -0.4998 0.0005 A DC -0.049
1A 1.1000 1.1001 0.0004 A DC 0.009
1A -1.1000 -1.0996 0.0002 A DC -0.033
Current Calibration Linearity AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A AC -0.015
1A 0.5000 0.5000 0.0002 A AC -0.008
1A 1.1000 1.0999 0.0002 A AC -0.008
External Current Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9999 0.0002 A DC -0.007
1V -1.0000 -1.0000 0.0002 A DC -0.003
External Current Calibration 10 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9999 0.0002 A AC -0.009
External Current Calibration 60 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50.00 50.00 0.06 mA AC -0.008
100mV 100.00 100.00 0.07 mA AC 0.000
200mV 200.00 199.99 0.13 mA AC -0.005
500mV 500.00 499.99 0.09 mA AC -0.001
1V 1.0000 0.9999 0.0002 A AC -0.008
2V 2.0000 1.9999 0.0006 A AC -0.003
5V 5.0000 5.0000 0.0009 A AC -0.001
External Current Calibration 10 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9994 0.0002 A AC -0.057
External Current Calibration 50 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9986 0.0005 A AC -0.140
Certificate number 58052P10722
Page: 12 of 43
External Current Calibration 100 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50 50 2 mA AC 0.000
1V 1.0000 0.9999 0.0013 A AC -0.012
Power Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.87 0.15 W -0.026
100V 5A -500.00 -499.79 0.17 W -0.042
Power Calibration 10Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
30V 1A 30.00 29.99 0.02 W -0.019
Power Calibration 60Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 2A 200.00 199.96 0.04 W -0.019
100V 5A 500.00 499.90 0.06 W -0.020
100V 10A 1.0000 0.9998 0.0002 kW -0.016
100V 20A 1.8000 1.7999 0.0008 kW -0.007
100V 50A 1.8000 1.7999 0.0008 kW -0.008
1.5V 1A 1.5000 1.5000 0.0023 W -0.003
3V 1A 3.000 3.000 0.005 W -0.012
6V 1A 6.000 6.000 0.009 W -0.007
10V 1A 10.000 10.002 0.002 W 0.024
15V 1A 15.000 14.997 0.002 W -0.018
30V 1A 30.000 29.994 0.005 W -0.018
60V 1A 60.000 59.988 0.009 W -0.020
100V 1A 100.00 99.98 0.02 W -0.017
150V 1A 150.00 149.97 0.02 W -0.018
300V 1A 300.00 299.94 0.05 W -0.019
600V 1A 600.00 599.92 0.10 W -0.013
1000V 1A 1.0000 0.9999 0.0002 kW -0.005
Certificate number 58052P10722
Page: 13 of 43
Power Calibration 10 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.02 0.17 W 0.004
100V 20A 1.0000 1.0046 0.0020 kW 0.462
6V 1A 6.0000 5.9990 0.0029 W -0.017
100V 1A 100.00 99.98 0.05 W -0.018
150V 1A 100.00 99.94 0.05 W -0.057
Power Calibration 50 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.93 0.17 W -0.013
100V 20A 1.0000 1.0045 0.0020 kW 0.453
6V 1A 6.000 5.921 0.011 W -1.325
100V 1A 100.00 98.76 0.10 W -1.245
150V 1A 100.00 98.75 0.10 W -1.246
Power Calibration 100 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.93 0.17 W -0.014
100V 20A 1.0000 1.0045 0.0020 kW 0.449
100V 1A 100.0 98.7 0.3 W -1.281
150V 1A 100.0 98.8 0.3 W -1.209
Power Calibration Linearity DC PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.025
100V 1A -1.00 -1.00 0.02 W -0.015
100V 1A 50.00 50.00 0.02 W -0.007
100V 1A -50.00 -50.00 0.02 W -0.009
100V 1A 110.00 109.98 0.04 W -0.016
100V 1A -110.00 -109.99 0.04 W -0.010
Power Calibration Linearity DC PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 0.99 0.02 W -0.672
100V 1A -1.00 -1.01 0.02 W 0.673
100V 1A 50.00 49.99 0.02 W -0.025
100V 1A -50.00 -49.99 0.02 W -0.014
100V 1A 110.00 109.96 0.02 W -0.037
100V 1A -110.00 -109.95 0.02 W -0.045
Power Calibration Linearity 60 Hz PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.023
100V 1A 50.00 49.99 0.02 W -0.020
100V 1A 110.00 109.98 0.02 W -0.020
Certificate number 58052P10722
Page: 14 of 43
Power Calibration Linearity 60 Hz PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.011
100V 1A 50.00 50.00 0.02 W -0.006
100V 1A 110.00 109.97 0.02 W -0.025
External Current Power Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.017
60V 1V -60.00 -60.01 0.02 W 0.012
External Current Power Calibration 10 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.012
External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 500mV 30.00 30.00 0.05 W -0.010
60V 1V 60.00 59.99 0.02 W -0.015
60V 5V 300.00 299.97 0.12 W -0.010
Power Calibration 60Hz PF=0 Leg
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.02 0.02 W -
Power Calibration 60Hz PF=0 Lead
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Power Calibration 60Hz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.01 0.02 W -
Power Calibration 60Hz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Certificate number 58052P10722
Page: 15 of 43
Power Calibration 1 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.07 0.12 W -
Power Calibration 10 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.23 0.24 W -
Power Calibration 50 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.2 0.7 W -
Power Calibration 100 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 2.2 1.2 W -
Power Calibration 1 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.07 0.12 W -
Power Calibration 10 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.2 0.6 W -
Power Calibration 50 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.2 0.6 W -
Power Calibration 100 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -2.2 1.2 W -
Certificate number 58052P10722
Page: 16 of 43
E L E M E N T 3
Voltage Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V DC -0.009
100V -100.00 -99.99 0.02 V DC -0.010
Voltage Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V AC -0.012
Voltage Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1.5V 1.5000 1.4997 0.0006 V AC -0.021
3V 3.0000 2.9997 0.0004 V AC -0.010
6V 6.0000 5.9997 0.0009 V AC -0.005
10V 10.000 9.999 0.002 V AC -0.010
15V 15.000 14.999 0.002 V AC -0.006
30V 30.000 29.998 0.002 V AC -0.008
60V 60.000 59.996 0.002 V AC -0.006
100V 100.00 99.99 0.02 V AC -0.009
150V 150.00 149.99 0.02 V AC -0.009
300V 300.00 299.98 0.02 V AC -0.008
600V 600.00 599.99 0.04 V AC -0.002
1000V 1.0000 1.0001 0.0002 kV AC 0.007
Voltage Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0032 0.0010 V AC 0.054
100V 100.00 100.03 0.06 V AC 0.034
150V 100.00 100.00 0.06 V AC 0.002
Voltage Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0024 0.0023 V AC 0.039
100V 100.00 100.03 0.14 V AC 0.031
150V 100.00 100.03 0.14 V AC 0.030
Voltage Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.000 6.003 0.010 V AC 0.056
100V 100.00 100.06 0.16 V AC 0.060
150V 100.00 100.11 0.16 V AC 0.108
Certificate number 58052P10722
Page: 17 of 43
Voltage Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V DC 0.337
100V -1.00 -0.99 0.02 V DC -0.667
100V 50.00 50.00 0.02 V DC 0.001
100V -50.00 -49.99 0.02 V DC -0.014
100V 110.00 109.99 0.02 V DC -0.006
100V -110.00 -109.99 0.02 V DC -0.009
Voltage Calibration Linearity at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V AC -0.022
100V 50.00 50.00 0.02 V AC -0.005
100V 110.00 109.99 0.02 V AC -0.009
Current Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
5A 5.000 5.000 0.002 A DC -0.008
5A -5.000 -5.000 0.002 A DC -0.008
Current Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0004 A AC -0.011
Current Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0002 A AC -0.012
2A 2.0000 1.9998 0.0002 A AC -0.010
5A 5.0000 4.9994 0.0004 A AC -0.013
10A 10.000 9.999 0.002 A AC -0.011
20A 18.000 17.999 0.006 A AC -0.003
50A 18.000 18.000 0.006 A AC 0.000
Current Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0002 0.0005 A AC 0.019
5A 5.000 5.001 0.008 A AC 0.018
10A 10.000 10.040 0.012 A AC 0.401
20A 10.000 10.042 0.012 A AC 0.423
Current Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9902 0.0004 A AC -0.982
Certificate number 58052P10722
Page: 18 of 43
Current Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9938 0.0007 A AC -0.620
5A 5.000 4.998 0.019 A AC -0.036
20A 5.000 5.048 0.019 A AC 0.952
Current Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A DC 0.339
1A -0.0100 -0.0100 0.0002 A DC -0.006
1A 0.5000 0.5003 0.0004 A DC 0.055
1A -0.5000 -0.4997 0.0005 A DC -0.060
1A 1.1000 1.1000 0.0004 A DC 0.003
1A -1.1000 -1.0996 0.0002 A DC -0.040
Current Calibration Linearity AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A AC -0.015
1A 0.5000 0.5000 0.0002 A AC -0.008
1A 1.1000 1.0998 0.0002 A AC -0.015
External Current Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A DC 0.003
1V -1.0000 -0.9999 0.0002 A DC -0.013
External Current Calibration 10 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A AC -0.005
External Current Calibration 60 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50.00 50.00 0.06 mA AC 0.004
100mV 100.00 100.00 0.07 mA AC -0.001
200mV 200.00 200.00 0.13 mA AC 0.000
500mV 500.00 500.00 0.09 mA AC 0.000
1V 1.0000 0.9999 0.0002 A AC -0.007
2V 2.0000 2.0000 0.0006 A AC -0.001
5V 5.0000 5.0002 0.0008 A AC 0.004
External Current Calibration 10 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9997 0.0002 A AC -0.035
External Current Calibration 50 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9990 0.0005 A AC -0.096
Certificate number 58052P10722
Page: 19 of 43
External Current Calibration 100 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50 50 2 mA AC -0.092
1V 1.0000 1.0002 0.0013 A AC 0.023
Power Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.90 0.15 W -0.021
100V 5A -500.00 -499.85 0.17 W -0.030
Power Calibration 10Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
30V 1A 30.00 30.00 0.02 W -0.013
Power Calibration 60Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 2A 200.00 199.98 0.04 W -0.011
100V 5A 500.00 499.91 0.06 W -0.017
100V 10A 1.0000 0.9999 0.0002 kW -0.012
100V 20A 1.8000 1.8000 0.0008 kW 0.000
100V 50A 1.8000 1.8001 0.0008 kW 0.005
1.5V 1A 1.5000 1.5000 0.0023 W -0.001
3V 1A 3.000 3.000 0.005 W -0.004
6V 1A 6.000 6.000 0.009 W -0.007
10V 1A 10.000 10.002 0.002 W 0.021
15V 1A 15.000 14.998 0.002 W -0.012
30V 1A 30.000 29.995 0.005 W -0.015
60V 1A 60.000 59.991 0.009 W -0.016
100V 1A 100.00 99.98 0.02 W -0.017
150V 1A 150.00 149.98 0.02 W -0.015
300V 1A 300.00 299.95 0.05 W -0.015
600V 1A 600.00 599.94 0.10 W -0.010
1000V 1A 1.0000 1.0000 0.0002 kW -0.004
Certificate number 58052P10722
Page: 20 of 43
Power Calibration 10 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.40 0.17 W 0.080
100V 20A 1.0000 1.0051 0.0020 kW 0.511
6V 1A 6.0000 6.0033 0.0029 W 0.055
100V 1A 100.00 100.05 0.05 W 0.050
150V 1A 100.00 100.01 0.05 W 0.012
Power Calibration 50 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.31 0.17 W 0.062
100V 20A 1.0000 1.0051 0.0020 kW 0.507
6V 1A 6.000 5.942 0.011 W -0.973
100V 1A 100.00 99.06 0.10 W -0.942
150V 1A 100.00 99.05 0.10 W -0.950
Power Calibration 100 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.30 0.17 W 0.060
100V 20A 1.0000 1.0050 0.0020 kW 0.499
100V 1A 100.0 99.5 0.3 W -0.536
150V 1A 100.0 99.5 0.3 W -0.493
Power Calibration Linearity DC PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.061
100V 1A -1.00 -1.00 0.02 W 0.062
100V 1A 50.00 50.00 0.02 W 0.004
100V 1A -50.00 -50.00 0.02 W 0.004
100V 1A 110.00 109.98 0.04 W -0.015
100V 1A -110.00 -110.00 0.04 W 0.000
Power Calibration Linearity DC PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.01 0.02 W 1.003
100V 1A -1.00 -0.99 0.02 W -1.003
100V 1A 50.00 50.00 0.02 W 0.006
100V 1A -50.00 -49.99 0.02 W -0.022
100V 1A 110.00 109.97 0.02 W -0.028
100V 1A -110.00 -109.96 0.02 W -0.039
Power Calibration Linearity 60 Hz PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.027
100V 1A 50.00 49.99 0.02 W -0.018
100V 1A 110.00 109.98 0.02 W -0.015
Certificate number 58052P10722
Page: 21 of 43
Power Calibration Linearity 60 Hz PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.013
100V 1A 50.00 50.00 0.02 W -0.006
100V 1A 110.00 109.98 0.02 W -0.022
External Current Power Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.010
60V 1V -60.00 -60.01 0.02 W 0.023
External Current Power Calibration 10 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 60.00 0.02 W -0.007
External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 500mV 30.00 30.00 0.05 W -0.007
60V 1V 60.00 60.00 0.02 W -0.008
60V 5V 300.00 299.99 0.12 W -0.002
Power Calibration 60Hz PF=0 Leg
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Power Calibration 60Hz PF=0 Lead
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.04 0.02 W -
Power Calibration 60Hz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.04 0.02 W -
Power Calibration 60Hz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.06 0.02 W -
Certificate number 58052P10722
Page: 22 of 43
Power Calibration 1 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.09 0.12 W -
Power Calibration 10 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.04 0.24 W -
Power Calibration 50 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.2 0.7 W -
Power Calibration 100 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.1 1.2 W -
Power Calibration 1 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.09 0.12 W -
Power Calibration 10 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.17 0.24 W -
Power Calibration 50 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.3 0.6 W -
Power Calibration 100 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.1 1.2 W -
Certificate number 58052P10722
Page: 23 of 43
E L E M E N T 4
Voltage Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V DC -0.009
100V -100.00 -99.99 0.02 V DC -0.010
Voltage Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V AC -0.012
Voltage Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1.5V 1.5000 1.4996 0.0006 V AC -0.025
3V 3.0000 2.9995 0.0005 V AC -0.015
6V 6.0000 5.9995 0.0009 V AC -0.008
10V 10.000 9.999 0.002 V AC -0.014
15V 15.000 14.999 0.002 V AC -0.008
30V 30.000 29.996 0.002 V AC -0.012
60V 60.000 59.994 0.002 V AC -0.011
100V 100.00 99.99 0.02 V AC -0.013
150V 150.00 149.98 0.02 V AC -0.015
300V 300.00 299.96 0.02 V AC -0.012
600V 600.00 599.97 0.02 V AC -0.005
1000V 1.0000 1.0001 0.0002 kV AC 0.006
Voltage Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0019 0.0010 V AC 0.031
100V 100.00 100.01 0.06 V AC 0.015
150V 100.00 99.98 0.06 V AC -0.018
Voltage Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0035 0.0023 V AC 0.059
100V 100.00 100.03 0.14 V AC 0.031
150V 100.00 100.04 0.14 V AC 0.040
Voltage Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.000 6.002 0.010 V AC 0.040
100V 100.00 100.02 0.16 V AC 0.021
150V 100.00 100.12 0.16 V AC 0.121
Certificate number 58052P10722
Page: 24 of 43
Voltage Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V DC 0.003
100V -1.00 -0.99 0.02 V DC -0.667
100V 50.00 49.99 0.02 V DC -0.012
100V -50.00 -49.99 0.02 V DC -0.014
100V 110.00 109.98 0.02 V DC -0.015
100V -110.00 -109.99 0.02 V DC -0.009
Voltage Calibration Linearity at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V AC -0.022
100V 50.00 50.00 0.02 V AC -0.004
100V 110.00 109.99 0.02 V AC -0.012
Current Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
5A 5.000 5.000 0.002 A DC 0.002
5A -5.000 -5.000 0.002 A DC -0.007
Current Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0000 0.0004 A AC 0.000
Current Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0000 0.0002 A AC -0.001
2A 2.0000 1.9999 0.0002 A AC -0.005
5A 5.0000 4.9996 0.0004 A AC -0.008
10A 10.000 10.000 0.002 A AC 0.000
20A 18.000 17.999 0.006 A AC -0.003
50A 18.000 18.001 0.006 A AC 0.006
Current Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0009 0.0005 A AC 0.088
5A 5.000 5.004 0.008 A AC 0.086
10A 10.000 10.047 0.012 A AC 0.473
20A 10.000 10.049 0.012 A AC 0.487
Current Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9910 0.0004 A AC -0.903
Certificate number 58052P10722
Page: 25 of 43
Current Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9929 0.0008 A AC -0.708
5A 5.000 4.991 0.019 A AC -0.183
20A 5.000 5.043 0.019 A AC 0.857
Current Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A DC -0.011
1A -0.0100 -0.0100 0.0002 A DC 0.007
1A 0.5000 0.5004 0.0004 A DC 0.079
1A -0.5000 -0.4998 0.0005 A DC -0.043
1A 1.1000 1.1001 0.0004 A DC 0.005
1A -1.1000 -1.0996 0.0002 A DC -0.033
Current Calibration Linearity AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A AC -0.014
1A 0.5000 0.5000 0.0002 A AC -0.008
1A 1.1000 1.0999 0.0002 A AC -0.007
External Current Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9999 0.0002 A DC -0.006
1V -1.0000 -1.0001 0.0002 A DC 0.007
External Current Calibration 10 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A AC 0.003
External Current Calibration 60 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50.00 50.00 0.06 mA AC 0.001
100mV 100.00 100.00 0.07 mA AC -0.001
200mV 200.00 200.01 0.13 mA AC 0.004
500mV 500.00 500.01 0.09 mA AC 0.003
1V 1.0000 1.0000 0.0002 A AC -0.005
2V 2.0000 2.0001 0.0007 A AC 0.004
5V 5.0000 5.0003 0.0009 A AC 0.005
External Current Calibration 10 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0001 0.0002 A AC 0.007
External Current Calibration 50 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9995 0.0005 A AC -0.045
Certificate number 58052P10722
Page: 26 of 43
External Current Calibration 100 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50 50 2 mA AC -0.096
1V 1.0000 1.0009 0.0013 A AC 0.086
Power Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.87 0.15 W -0.026
100V 5A -500.00 -499.79 0.17 W -0.043
Power Calibration 10Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
30V 1A 30.00 29.99 0.02 W -0.020
Power Calibration 60Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 2A 200.00 199.96 0.04 W -0.018
100V 5A 500.00 499.92 0.06 W -0.017
100V 10A 1.0000 0.9998 0.0002 kW -0.015
100V 20A 1.8000 1.7999 0.0008 kW -0.007
100V 50A 1.8000 1.8000 0.0008 kW -0.001
1.5V 1A 1.5000 1.5000 0.0023 W -0.001
3V 1A 3.000 3.000 0.005 W -0.005
6V 1A 6.000 6.000 0.009 W -0.006
10V 1A 10.000 10.002 0.002 W 0.025
15V 1A 15.000 14.997 0.002 W -0.018
30V 1A 30.000 29.995 0.005 W -0.017
60V 1A 60.000 59.990 0.009 W -0.017
100V 1A 100.00 99.99 0.02 W -0.014
150V 1A 150.00 149.98 0.02 W -0.015
300V 1A 300.00 299.95 0.05 W -0.016
600V 1A 600.00 599.93 0.10 W -0.011
1000V 1A 1.0000 0.9999 0.0002 kW -0.006
Certificate number 58052P10722
Page: 27 of 43
Power Calibration 10 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.59 0.17 W 0.118
100V 20A 1.0000 1.0056 0.0020 kW 0.559
6V 1A 6.0000 6.0059 0.0029 W 0.099
100V 1A 100.00 100.10 0.05 W 0.097
150V 1A 100.00 100.04 0.05 W 0.043
Power Calibration 50 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.51 0.17 W 0.102
100V 20A 1.0000 1.0055 0.0020 kW 0.550
6V 1A 6.000 5.944 0.011 W -0.925
100V 1A 100.00 99.14 0.10 W -0.856
150V 1A 100.00 99.12 0.10 W -0.876
Power Calibration 100 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.51 0.17 W 0.102
100V 20A 1.0000 1.0056 0.0020 kW 0.556
100V 1A 100.0 99.3 0.3 W -0.657
150V 1A 100.0 99.4 0.3 W -0.554
Power Calibration Linearity DC PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.055
100V 1A -1.00 -1.00 0.02 W 0.035
100V 1A 50.00 49.99 0.02 W -0.012
100V 1A -50.00 -50.00 0.02 W 0.005
100V 1A 110.00 109.97 0.04 W -0.025
100V 1A -110.00 -109.98 0.04 W -0.019
Power Calibration Linearity DC PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 0.99 0.02 W -1.001
100V 1A -1.00 -1.01 0.02 W 1.000
100V 1A 50.00 49.98 0.02 W -0.039
100V 1A -50.00 -50.01 0.02 W 0.011
100V 1A 110.00 109.96 0.02 W -0.039
100V 1A -110.00 -109.96 0.02 W -0.034
Power Calibration Linearity 60 Hz PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.027
100V 1A 50.00 49.99 0.02 W -0.021
100V 1A 110.00 109.99 0.02 W -0.013
Certificate number 58052P10722
Page: 28 of 43
Power Calibration Linearity 60 Hz PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.013
100V 1A 50.00 50.00 0.02 W -0.010
100V 1A 110.00 109.97 0.02 W -0.024
External Current Power Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.020
60V 1V -60.00 -60.00 0.02 W 0.003
External Current Power Calibration 10 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.010
External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 500mV 30.00 30.00 0.05 W -0.009
60V 1V 60.00 59.99 0.02 W -0.013
60V 5V 300.00 299.98 0.12 W -0.008
Power Calibration 60Hz PF=0 Leg
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.01 0.02 W -
Power Calibration 60Hz PF=0 Lead
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Power Calibration 60Hz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.01 0.02 W -
Power Calibration 60Hz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Certificate number 58052P10722
Page: 29 of 43
Power Calibration 1 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.11 0.12 W -
Power Calibration 10 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.09 0.24 W -
Power Calibration 50 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.4 0.7 W -
Power Calibration 100 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 2.4 1.2 W -
Power Calibration 1 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.10 0.12 W -
Power Calibration 10 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.22 0.24 W -
Power Calibration 50 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.3 0.6 W -
Power Calibration 100 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -2.4 1.2 W -
Certificate number 58052P10722
Page: 30 of 43
E L E M E N T 5
Voltage Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.98 0.02 V DC -0.020
100V -100.00 -100.00 0.02 V DC -0.003
Voltage Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V AC -0.012
Voltage Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1.5V 1.5000 1.4997 0.0006 V AC -0.018
3V 3.0000 2.9996 0.0005 V AC -0.012
6V 6.0000 5.9997 0.0009 V AC -0.005
10V 10.000 9.999 0.002 V AC -0.014
15V 15.000 14.999 0.002 V AC -0.008
30V 30.000 29.997 0.002 V AC -0.012
60V 60.000 59.994 0.004 V AC -0.009
100V 100.00 99.99 0.02 V AC -0.013
150V 150.00 149.99 0.02 V AC -0.009
300V 300.00 299.97 0.02 V AC -0.010
600V 600.00 599.96 0.02 V AC -0.007
1000V 1.0000 1.0000 0.0002 kV AC 0.003
Voltage Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0027 0.0010 V AC 0.044
100V 100.00 100.03 0.06 V AC 0.035
150V 100.00 100.01 0.06 V AC 0.005
Voltage Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 5.9999 0.0023 V AC -0.002
100V 100.00 100.04 0.14 V AC 0.041
150V 100.00 100.06 0.14 V AC 0.060
Voltage Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.000 5.997 0.010 V AC -0.043
100V 100.00 100.02 0.16 V AC 0.021
150V 100.00 100.10 0.16 V AC 0.099
Certificate number 58052P10722
Page: 31 of 43
Voltage Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V DC -0.013
100V -1.00 -1.00 0.02 V DC 0.020
100V 50.00 49.99 0.02 V DC -0.020
100V -50.00 -50.00 0.02 V DC 0.000
100V 110.00 109.98 0.02 V DC -0.018
100V -110.00 -110.00 0.02 V DC -0.003
Voltage Calibration Linearity at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V AC -0.022
100V 50.00 50.00 0.02 V AC -0.005
100V 110.00 109.99 0.02 V AC -0.011
Current Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
5A 5.000 5.000 0.002 A DC -0.004
5A -5.000 -4.999 0.002 A DC -0.011
Current Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0000 0.0004 A AC -0.004
Current Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0002 A AC -0.011
2A 2.0000 1.9998 0.0002 A AC -0.010
5A 5.0000 4.9995 0.0002 A AC -0.010
10A 10.000 9.999 0.002 A AC -0.009
20A 18.000 17.999 0.006 A AC -0.005
50A 18.000 17.999 0.006 A AC -0.006
Current Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0004 0.0005 A AC 0.038
5A 5.000 5.002 0.008 A AC 0.038
10A 10.000 10.045 0.012 A AC 0.450
20A 10.000 10.046 0.012 A AC 0.464
Current Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9900 0.0004 A AC -1.004
Certificate number 58052P10722
Page: 32 of 43
Current Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9896 0.0007 A AC -1.036
5A 5.000 4.972 0.019 A AC -0.561
20A 5.000 5.022 0.019 A AC 0.442
Current Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0101 0.0002 A DC 1.329
1A -0.0100 -0.0100 0.0002 A DC -0.333
1A 0.5000 0.5004 0.0004 A DC 0.086
1A -0.5000 -0.4998 0.0005 A DC -0.047
1A 1.1000 1.1001 0.0004 A DC 0.011
1A -1.1000 -1.0997 0.0002 A DC -0.026
Current Calibration Linearity AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A AC -0.015
1A 0.5000 0.5000 0.0002 A AC -0.006
1A 1.1000 1.0999 0.0002 A AC -0.007
External Current Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9999 0.0002 A DC -0.007
1V -1.0000 -1.0001 0.0002 A DC 0.006
External Current Calibration 10 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A AC 0.001
External Current Calibration 60 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50.00 50.01 0.06 mA AC 0.010
100mV 100.00 100.00 0.07 mA AC -0.003
200mV 200.00 200.00 0.13 mA AC 0.000
500mV 500.00 500.01 0.09 mA AC 0.002
1V 1.0000 1.0000 0.0002 A AC -0.003
2V 2.0000 2.0000 0.0006 A AC 0.000
5V 5.0000 5.0002 0.0009 A AC 0.003
External Current Calibration 10 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9993 0.0002 A AC -0.066
External Current Calibration 50 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9985 0.0005 A AC -0.149
Certificate number 58052P10722
Page: 33 of 43
External Current Calibration 100 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50 50 2 mA AC 0.009
1V 1.0000 0.9998 0.0013 A AC -0.022
Power Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.88 0.15 W -0.024
100V 5A -500.00 -499.80 0.17 W -0.041
Power Calibration 10Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
30V 1A 30.00 29.99 0.02 W -0.022
Power Calibration 60Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 2A 200.00 199.96 0.04 W -0.018
100V 5A 500.00 499.89 0.06 W -0.022
100V 10A 1.0000 0.9999 0.0002 kW -0.015
100V 20A 1.8000 1.7999 0.0008 kW -0.007
100V 50A 1.8000 1.7999 0.0008 kW -0.003
1.5V 1A 1.5000 1.5000 0.0023 W -0.003
3V 1A 3.000 3.000 0.005 W -0.007
6V 1A 6.000 6.000 0.009 W -0.007
10V 1A 10.000 10.002 0.002 W 0.025
15V 1A 15.000 14.997 0.002 W -0.019
30V 1A 30.000 29.994 0.005 W -0.021
60V 1A 60.000 59.989 0.009 W -0.019
100V 1A 100.00 99.98 0.02 W -0.017
150V 1A 150.00 149.98 0.02 W -0.015
300V 1A 300.00 299.95 0.05 W -0.017
600V 1A 600.00 599.92 0.10 W -0.013
1000V 1A 1.0000 0.9999 0.0002 kW -0.006
Certificate number 58052P10722
Page: 34 of 43
Power Calibration 10 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.43 0.17 W 0.085
100V 20A 1.0000 1.0055 0.0020 kW 0.550
6V 1A 6.0000 6.0038 0.0029 W 0.063
100V 1A 100.00 100.06 0.05 W 0.063
150V 1A 100.00 100.02 0.05 W 0.022
Power Calibration 50 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.33 0.17 W 0.066
100V 20A 1.0000 1.0054 0.0020 kW 0.542
6V 1A 6.000 5.933 0.011 W -1.125
100V 1A 100.00 99.03 0.10 W -0.966
150V 1A 100.00 99.04 0.10 W -0.963
Power Calibration 100 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.34 0.17 W 0.068
100V 20A 1.0000 1.0054 0.0020 kW 0.538
100V 1A 100.0 99.0 0.3 W -0.966
150V 1A 100.0 99.1 0.3 W -0.918
Power Calibration Linearity DC PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.018
100V 1A -1.00 -1.00 0.02 W 0.002
100V 1A 50.00 50.00 0.02 W 0.005
100V 1A -50.00 -50.00 0.02 W 0.003
100V 1A 110.00 110.00 0.04 W -0.002
100V 1A -110.00 -109.99 0.04 W -0.009
Power Calibration Linearity DC PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.01 0.02 W 0.668
100V 1A -1.00 -1.00 0.02 W 0.333
100V 1A 50.00 50.00 0.02 W 0.002
100V 1A -50.00 -49.99 0.02 W -0.015
100V 1A 110.00 109.96 0.02 W -0.040
100V 1A -110.00 -109.96 0.02 W -0.037
Power Calibration Linearity 60 Hz PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.023
100V 1A 50.00 49.99 0.02 W -0.016
100V 1A 110.00 109.98 0.02 W -0.018
Certificate number 58052P10722
Page: 35 of 43
Power Calibration Linearity 60 Hz PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.011
100V 1A 50.00 49.99 0.02 W -0.013
100V 1A 110.00 109.98 0.02 W -0.023
External Current Power Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.98 0.02 W -0.030
60V 1V -60.00 -60.00 0.02 W 0.002
External Current Power Calibration 10 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.014
External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 500mV 30.00 30.00 0.05 W -0.010
60V 1V 60.00 59.99 0.02 W -0.014
60V 5V 300.00 299.98 0.12 W -0.008
Power Calibration 60Hz PF=0 Leg
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.02 0.02 W -
Power Calibration 60Hz PF=0 Lead
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Power Calibration 60Hz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.02 0.02 W -
Power Calibration 60Hz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.01 0.02 W -
Certificate number 58052P10722
Page: 36 of 43
Power Calibration 1 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.08 0.12 W -
Power Calibration 10 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.18 0.24 W -
Power Calibration 50 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.1 0.7 W -
Power Calibration 100 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 1.7 1.2 W -
Power Calibration 1 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.09 0.13 W -
Power Calibration 10 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.31 0.24 W -
Power Calibration 50 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.0 0.6 W -
Power Calibration 100 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -1.7 1.2 W -
Certificate number 58052P10722
Page: 37 of 43
E L E M E N T 6
Voltage Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V DC -0.013
100V -100.00 -99.99 0.02 V DC -0.010
Voltage Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 100.00 99.99 0.02 V AC -0.012
Voltage Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1.5V 1.5000 1.4997 0.0006 V AC -0.018
3V 3.0000 2.9997 0.0005 V AC -0.011
6V 6.0000 5.9998 0.0009 V AC -0.004
10V 10.000 9.999 0.002 V AC -0.013
15V 15.000 14.999 0.002 V AC -0.009
30V 30.000 29.997 0.002 V AC -0.008
60V 60.000 59.994 0.002 V AC -0.010
100V 100.00 99.99 0.02 V AC -0.006
150V 150.00 149.98 0.02 V AC -0.011
300V 300.00 299.97 0.02 V AC -0.011
600V 600.00 599.98 0.02 V AC -0.004
1000V 1.0000 1.0001 0.0002 kV AC 0.006
Voltage Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0007 0.0010 V AC 0.011
100V 100.00 100.00 0.06 V AC 0.005
150V 100.00 99.97 0.06 V AC -0.028
Voltage Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.0000 6.0017 0.0023 V AC 0.029
100V 100.00 100.05 0.14 V AC 0.051
150V 100.00 100.10 0.14 V AC 0.100
Voltage Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
6V 6.000 6.001 0.010 V AC 0.017
100V 100.00 100.07 0.16 V AC 0.069
150V 100.00 100.19 0.16 V AC 0.188
Certificate number 58052P10722
Page: 38 of 43
Voltage Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V DC 0.373
100V -1.00 -1.00 0.02 V DC -0.363
100V 50.00 49.99 0.02 V DC -0.012
100V -50.00 -49.99 0.02 V DC -0.014
100V 110.00 109.98 0.02 V DC -0.015
100V -110.00 -109.99 0.02 V DC -0.009
Voltage Calibration Linearity at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1.00 1.00 0.02 V AC -0.022
100V 50.00 50.00 0.02 V AC -0.003
100V 110.00 109.99 0.02 V AC -0.012
Current Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
5A 5.000 5.000 0.002 A DC -0.004
5A -5.000 -4.999 0.002 A DC -0.012
Current Calibration AC at 10 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0004 A AC -0.013
Current Calibration AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9999 0.0002 A AC -0.010
2A 2.0000 1.9998 0.0002 A AC -0.010
5A 5.0000 4.9995 0.0002 A AC -0.011
10A 10.000 9.999 0.002 A AC -0.013
20A 18.000 18.000 0.006 A AC -0.001
50A 18.000 18.000 0.006 A AC 0.000
Current Calibration AC at 10 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 1.0007 0.0005 A AC 0.066
5A 5.000 5.003 0.008 A AC 0.066
10A 10.000 10.044 0.012 A AC 0.437
20A 10.000 10.046 0.012 A AC 0.461
Current Calibration AC at 50 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9896 0.0004 A AC -1.044
Certificate number 58052P10722
Page: 39 of 43
Current Calibration AC at 100 kHz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 1.0000 0.9919 0.0009 A AC -0.809
5A 5.000 4.988 0.019 A AC -0.246
20A 5.000 5.035 0.019 A AC 0.702
Current Calibration Linearity DC
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A DC 0.335
1A -0.0100 -0.0100 0.0002 A DC -0.001
1A 0.5000 0.5003 0.0004 A DC 0.062
1A -0.5000 -0.4998 0.0005 A DC -0.037
1A 1.1000 1.1001 0.0004 A DC 0.005
1A -1.1000 -1.0997 0.0002 A DC -0.030
Current Calibration Linearity AC at 60 Hz
Range Applied Measured ±Uncertainty Unit Deviation %
1A 0.0100 0.0100 0.0002 A AC -0.014
1A 0.5000 0.5000 0.0002 A AC -0.008
1A 1.1000 1.0999 0.0002 A AC -0.007
External Current Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A DC 0.004
1V -1.0000 -1.0000 0.0002 A DC -0.005
External Current Calibration 10 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 1.0000 0.0002 A AC -0.001
External Current Calibration 60 Hz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50.00 50.00 0.06 mA AC 0.009
100mV 100.00 100.00 0.07 mA AC 0.004
200mV 200.00 200.00 0.13 mA AC 0.000
500mV 500.00 500.00 0.09 mA AC 0.001
1V 1.0000 1.0000 0.0002 A AC -0.003
2V 2.0000 2.0000 0.0006 A AC 0.001
5V 5.0000 5.0001 0.0009 A AC 0.002
External Current Calibration 10 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9996 0.0002 A AC -0.036
External Current Calibration 50 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
1V 1.0000 0.9988 0.0005 A AC -0.116
Certificate number 58052P10722
Page: 40 of 43
External Current Calibration 100 kHz. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
50mV 50 50 2 mA AC -0.043
1V 1.0000 1.0001 0.0013 A AC 0.009
Power Calibration DC
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 499.84 0.15 W -0.031
100V 5A -500.00 -499.78 0.17 W -0.044
Power Calibration 10Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
30V 1A 30.00 29.99 0.02 W -0.019
Power Calibration 60Hz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 2A 200.00 199.96 0.04 W -0.018
100V 5A 500.00 499.89 0.06 W -0.022
100V 10A 1.0000 0.9998 0.0002 kW -0.016
100V 20A 1.8000 1.7999 0.0008 kW -0.008
100V 50A 1.8000 1.7999 0.0008 kW -0.007
1.5V 1A 1.5000 1.5000 0.0023 W -0.003
3V 1A 3.000 3.000 0.005 W -0.007
6V 1A 6.000 5.999 0.009 W -0.009
10V 1A 10.000 10.002 0.002 W 0.022
15V 1A 15.000 14.997 0.002 W -0.018
30V 1A 30.000 29.994 0.005 W -0.021
60V 1A 60.000 59.990 0.009 W -0.017
100V 1A 100.00 99.98 0.02 W -0.016
150V 1A 150.00 149.97 0.02 W -0.022
300V 1A 300.00 299.94 0.05 W -0.020
600V 1A 600.00 599.91 0.10 W -0.015
1000V 1A 1.0000 1.0000 0.0002 kW -0.004
Certificate number 58052P10722
Page: 41 of 43
Power Calibration 10 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.40 0.17 W 0.080
100V 20A 1.0000 1.0050 0.0020 kW 0.499
6V 1A 6.0000 6.0031 0.0029 W 0.052
100V 1A 100.00 100.06 0.05 W 0.058
150V 1A 100.00 100.02 0.05 W 0.020
Power Calibration 50 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.30 0.17 W 0.060
100V 20A 1.0000 1.0049 0.0020 kW 0.492
6V 1A 6.000 5.935 0.011 W -1.085
100V 1A 100.00 99.02 0.10 W -0.975
150V 1A 100.00 99.06 0.10 W -0.937
Power Calibration 100 kHz PF=1
Range Applied Measured ±Uncertainty Unit Deviation %
100V 5A 500.00 500.31 0.17 W 0.062
100V 20A 1.0000 1.0050 0.0020 kW 0.496
100V 1A 100.0 99.3 0.3 W -0.672
150V 1A 100.0 99.4 0.3 W -0.551
Power Calibration Linearity DC PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.038
100V 1A -1.00 -1.00 0.02 W 0.018
100V 1A 50.00 50.00 0.02 W -0.003
100V 1A -50.00 -50.00 0.02 W 0.003
100V 1A 110.00 109.98 0.04 W -0.017
100V 1A -110.00 -109.99 0.04 W -0.007
Power Calibration Linearity DC PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W 0.335
100V 1A -1.00 -0.99 0.02 W -1.000
100V 1A 50.00 49.99 0.02 W -0.010
100V 1A -50.00 -49.98 0.02 W -0.035
100V 1A 110.00 109.96 0.02 W -0.035
100V 1A -110.00 -109.95 0.02 W -0.041
Power Calibration Linearity 60 Hz PF=1 at constant 1 A
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.023
100V 1A 50.00 49.99 0.02 W -0.018
100V 1A 110.00 109.98 0.02 W -0.019
Certificate number 58052P10722
Page: 42 of 43
Power Calibration Linearity 60 Hz PF=1 at constant 100 V
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 1.00 1.00 0.02 W -0.013
100V 1A 50.00 50.00 0.02 W -0.006
100V 1A 110.00 109.97 0.02 W -0.024
External Current Power Calibration DC. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.009
60V 1V -60.00 -60.01 0.02 W 0.013
External Current Power Calibration 10 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 1V 60.00 59.99 0.02 W -0.012
External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.
Range Applied Measured ±Uncertainty Unit Deviation %
60V 500mV 30.00 30.00 0.05 W -0.008
60V 1V 60.00 59.99 0.02 W -0.011
60V 5V 300.00 299.99 0.12 W -0.005
Power Calibration 60Hz PF=0 Leg
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.01 0.02 W -
Power Calibration 60Hz PF=0 Lead
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.01 0.02 W -
Power Calibration 60Hz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.02 0.02 W -
Power Calibration 60Hz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.02 0.02 W -
Certificate number 58052P10722
Page: 43 of 43
Power Calibration 1 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.09 0.12 W -
Power Calibration 10 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 -0.14 0.24 W -
Power Calibration 50 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 0.3 0.7 W -
Power Calibration 100 kHz PF=0 Lagging
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 2.6 1.2 W -
Power Calibration 1 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.10 0.12 W -
Power Calibration 10 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.00 0.26 0.24 W -
Power Calibration 50 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -0.3 0.6 W -
Power Calibration 100 kHz PF=0 Leading
Range Applied Measured ±Uncertainty Unit Deviation %
100V 1A 0.0 -2.6 1.2 W -
Comments : No comment