21
Copyright © 2011 Agilent Technologies Copyright © 2011 Agilent Technologies Inc. Optimizing Performance and Best Practices for Millimeter Measurements Ben Zarlingo, Agilent Technologies EuMW 2011 Agilent Workshop Series Copyright © 2011 Agilent Technologies © Agilent Technologies 2011 Agenda Millimeter Measurements Discussion – Information – Best Practices – Optimizing Performance Special Hazards, Problems References, Resources

EuMW 2011 Agilent Workshop Series - keysight.com€¦ · Alignment: Frequent Self Adjustment – Automatic adjustment compensates for temperature, drift, aging – Alignment can affect

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

  • Copyright © 2011 Agilent TechnologiesCopyright © 2011 Agilent Technologies Inc.

    Optimizing Performance and Best

    Practices for Millimeter Measurements

    Ben Zarlingo, Agilent Technologies

    EuMW 2011 Agilent Workshop Series

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Agenda

    Millimeter Measurements Discussion

    – Information

    – Best Practices

    – Optimizing Performance

    Special Hazards, Problems

    References, Resources

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Commercial/Consumer: 60 GHz Wireless HD$600 Transmitter + Receiver

    $300

    What is Millimeter-Wave?

    Wavelengths ~≤ 10 mm, f ~≥ 30 GHz

    Where Everything Gets

    – Difficult

    – Small

    – Expensive

    – Lossy

    – Delicate

    – Inflexible

    – Banded (if not narrowband)

    Graduate Work for the Electrical Engineer

    $180

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Millimeter Performance Expectations

    Performance Generally Less Ideal

    – Harmonic architecture (phase noise, conversion loss, etc.)

    – Path/insertion losses, impedance mismatches

    – Preselector filtering

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Importance of Cables, Connectors

    Influence of Materials, Structures, Geometries

    Amplitude, 30-50 GHz

    – Dropouts due to moding*– Amplitude, phase errors

    – Errors with 3.5mm also, though perhaps more repeatable

    – Error < 2dB in this example• Good enough?

    • Repeatable enough?

    Performance Gain fromcables may be Less Expensive

    2.4 mm

    SMA

    *Moding: Excitation of the first circular waveguide propagation mode in the coaxial structure

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Plastic vs. Air Dielectric

    Plastic (PTFE) vs. Air, Beads

    – Air for metrology grade & many instrumentation grade

    Many Other Tolerances & Construction Differences UsuallyAccompany Dielectric Change

    SMA

    3.5 mm

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Use A Torque Wrench for Consistency

    “Wrench Lift Stress” is a special concern with severaldevices connected or a long device or adapter(s)

    Lift

    Angle between wrenches should be less than 90 deg

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Torque Values & Wrench Sizes

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Torque Problem Example

    Loose Cable AmplitudeError

    – Significant for you?What if it is variable andinconsistent?

    Phase Variability

    Sufficient Torque

    – Enough torque to properly matesurfaces and to produce a strainthat will prevent loosening dueto vibration and/or thermalcycling

    – Not enough to damage or distortconnectors, changing geometry

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Connector Grades

    Metrology Grade– Highest precision, performance, repeatability

    – Often slotless female center conductors (repeatable, lower inductance)

    – Often used in calibration/verification kits

    – Expensive, long life

    PSC = Precision Slotless Connector – Typically metrology grade; PSC versions available for 3.5mm, 2.4 mm

    Instrument Grade– Excellent repeatability, long life

    – Often slotted female center conductors

    General Purpose or Production Grade– Typical on components, cables, microstrip

    – Inspect before connecting to better grades (analyzers), use adapter

    – Limited connection cycles

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Connection Consistency

    Consistent Connections Improve Repeatability– Same Cables (cable loss may be more than 3 dB/meter at 50 GHz)

    – Same connectors, adapters

    – Same cable routing

    Every Connection & Transition is Imperfect– Insertion loss (one adapter = up to 1 dB loss at 50 GHz)

    – Impedance change

    – Bending can change cable characteristics

    – Consider semi-rigid cable

    Analyzer Accuracy is Improving →→→→ Smaller Errors Matter

    Repeatability may be More Important than Absolute Accuracy

    Standardize on Highest Frequency Connectors?– Consistency, but slightly higher loss at lower frequencies

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    mm-Frequency Connectors and Intermating

    24 pt level 1

    – 20 pt level 2

    • 18 pt level 3

    http://na.tm.agilent.com/pna/connectorcare/What_mates_with_what.htm

    Uses a plastic (PTFE) dielectric

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    NMD Connectors

    Precision, Ruggedized

    Provides Support

    Stabilizes Connection

    – The NMD connector is a ruggedized test-port connector used on Agilent test sets and network analyzers. Most often, the connectors used are 3.5mm and 2.4mm male connectors. These connectors include a large threaded body that is specially designed to stabilize the test port cable when attached to the front of the analyzer or test set using Agilent test port cables. The design of these connectors allows the use of standard female 3.5mm and 2.4mm cables and adapters.

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Torque for Intermating

    Proper Torque for Intermating Can Depend on Gender

    – Example 1: A 56 N-cm. torque wrench should be used to connect male SMA connectors to either 3.5-mm or 2.92-mm (K) connectors

    – Example 2: A 90 N-cm torque wrench should be used to connect male 3.5-mm or male 2.92-mm connectors to mating female connectors including SMA

    Precision Connectors use Higher Torque

    – Use lower torque with SMA ↔↔↔↔ SMA

    – Use lower torque with male SMA male ↔↔↔↔ precision

    Do not connect an SMA to a 3.5 mm calibration standard, VNA port cable or precision 3.5 mm device

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Important Lower Frequency Connectionsin a High Frequency System

    Connectors Make a Difference in Some Low Frequency Measurements!

    3.5 mm or SMA to BNC

    Precision BNC connections on Oscilloscope

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Precision BNC - 3.5 mm vs. BNC - SMA

    Air dielectric vs. PTFE

    Split or slotted or hybrid center and outer connectors

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Understand Your Measurement Plane

    Goal is Consistent Measurement Plane

    Make Connector/Adapter/Cable Choices to Reduce or Eliminate Removable Connection as Part of DUT

    Consistent Torque, Cable Routing

    Goal is Simplest, Shortest Path

    More Critical for Network Analysis

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Connection Corrections-Amplitude:Cables, Adapters, Probes

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Phase-Matched ConnectionsDoes Phase Matter?

    Spectrum, Power, Demodulation

    – Spectrum, Power (channel/band, ACP, etc.) insensitive to phase

    – Demodulation is insensitive, or corrects for phase for single channel

    – Demodulation can be sensitive to phase for 2-channel (i+jQ) measurements (quadrature error varying with frequency)

    – Flatter (over the measured span) is better for demod

    – Wider spans/bandwidths are more demanding

    Phase Stability

    – Cable, connector movement a problem for close-in phase noise

    – See also microphonics

    Phase Repeatability

    – Not a problem for most measurements

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Coax & Waveguide Comparison

    Flexible

    Wide frequency range

    Easy to change routing

    Light weight

    Can carry DC bias

    Can be less expensive

    Good match for component meas.

    Disadvantages– Lossy, increasing with frequency

    – Can be delicate

    – Limited power handling

    Low loss

    High power handling ability

    Mechanically durable

    Disadvantages– Inflexible

    – May need custom construction (expensive)

    – Time delays for construction and reconfiguration

    – Frequency range ≤ 1 octave

    – Transitions needed to connect to analyzers, other coaxial elements

    Coax Waveguide

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Agenda

    Millimeter Measurements Discussion

    – Information

    – Best Practices

    – Optimizing Performance

    Special Hazards, Problems

    References, Resources

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Beware of DC at Analyzer Input

    Millimeter Agilent PXA, PSA Have No DC Block

    Why? Internal DC Block Would Compromise Performance

    Add External DC Block if Needed

    Front Panel Male ConnectorProtects Analyzer

    – Discourages direct mating withtypical coaxial male connector

    – Encourages use of connector saver

    – Male connector more durable,less likely to be damaged fromimproper mating

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Adapters as Connector Savers

    Connectors are “Consumables” with a Limited Lifetime

    Non-Millimeter Instruments Typically use Female Front Panel Connector– Use to mate male cables with

    male analyzer connector

    – Consider other types as neededin your application

    Note “Flats” for Wrenches to Prevent Connector Rotation

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Contact Integrity, Concentricity

    Visual Inspection, Suggest 10x Magnification

    – Bending/displacement

    – Non-concentric male or female

    – Inspect mating planes for deep scratches, dents, debris

    Damage to Female Slotted Connectors Most Common

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Damage Hazard: Non-Captive SMA Nut

    Danger from SMA Male to 2.4 mm and 1.85 mm Female

    3.5 mm to 2.4 mm adapter

    3.5 mm end 2.4 mm end

    SMA with non-captive nutand extended center pin

    SMA with non-captive nut can be inserted into 2.4 mm

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Electrical Damage

    High Frequencies →→→→ Small Size →→→→ Electrically Delicate

    Burnout from Excess Power

    Burnout from Source Transient

    Static Zap

    Detecting Damage (other than outright failure)

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    ESD Precautions

    Wrist Strap

    Table Mat

    Heel Strap

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Microphonics, Other Movement

    Phase Noise, Close-In Modulation

    Can Affect Analyzer, Connections, DUT

    Mitigating Microphonics

    – Separate table, mounting

    – Use isolating feed/pads/mounts

    – Stabilize or support cables, waveguide

    – Use phase-stable cables

    Vibration, Thermal Cycling may Loosen Connections

    – Torque check as part of measurement routine

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Alignment, Calibration, Correction

    Calibration: Initial & Periodic Adjustment, Verification– Factory, service center, calibration lab

    Alignment: Frequent Self Adjustment– Automatic adjustment compensates for temperature, drift, aging

    – Alignment can affect dynamic range

    Correction/Compensation– Often included in demodulation operations

    – Substitution (power meters, etc.)

    – May be only practical way to dramatically improve accuracy

    – Enter parameters in analyzer for front-panel operation

    – May be manually cumbersome or need switching

    Perform Alignments Frequently and Whenever Measurement conditions Change– Temperature sensitivity generally goes up with frequency

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Preselector Centering

    YTF Preselector (Filter)

    – YTF operates open-loop

    – Preselector tuning is characterized automatically by analyzer for full frequency range

    – Centering performed by analyzer, very brief operation

    – Perform centering for best amplitude accuracy at a single frequency

    – Centering is initiated by user; not part of normal alignment

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Preselector Bypass

    “Microwave Preselector Bypass”

    Typically a Hardware Option

    Removes YIG-Tuned Preselector Filter from Measurement Path

    – Essential for wide-bandwidth measurements

    – Removes preselector insertion loss but passeswideband noise; effect on analyzer DANLis variable

    – May allow other signals into measurement

    – Can tune to signal with preselector in placeand then bypass

    Preselector

    Bypass Path

    Norm

    Bypass

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Vector Calibrations for Modulated, Wideband Signal Generation & Meas.

    Scalar Amplitude Corrections

    – Scalar measurements: amplitude, power, phase noise, noise figure

    – Narrowband

    – CW, narrowband signals

    Vector (Complex) Corrections

    – I/Q or amplitude/phase

    – Modulated signals

    – Wideband measurements

    32

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Example: RADAR Signal with IQ Errors

    2 GHz Chirp: Gated sweep shows chirp spectrum and unwanted image from modulator IQ errors

    33

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    RADAR Signal with IQ Corrections

    800 MHz Chirp, 900 MHz Span: IQ-Correcting Modulated Signal Removes Almost All Image Energy

    34

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Combining Preamp and Attenuation

    Preamp Improves Sensitivity

    Attenuation Improves Match (reduces mismatch error)

    Combine to Optimize Match & Sensitivity

    35

    Std PathLNP

    Preamp

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Hardware Features & Signal Processing Improve Performance

    Low NoisePath

    Noise FloorExtension

    36

    • Standard

    • With LNP

    • With NFE

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    New Calc. Method for Mismatch Uncertainty

    Mismatch Uncertainty typically dominates

    Better understanding, verification of probability distributions

    Rayleigh model is an excellent fit, provides more accurate but still conservative estimates of uncertainty due to mismatch

    Result: Lower Uncertainties due to mismatchby factor of 3 to 6

    AN 1449-3 Power Measurement Uncertainty Per International Guidelines (updated April 2011)

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Better Calculations Indicate a Rayleigh Distribution

    Probability density of the magnitude of the reflection coefficient is Rayleigh distributed if the probability density of both complex parts is Gaussian distributed

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Validating Assumptions with Real World Examples

    Power

    Sensor

    Signal

    Generator

    Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Validating Assumptions with Real World Examples

    Signal Analyzer

  • Copyright © 2011 Agilent Technologies

    © Agilent Technologies 2011

    Additional InformationMicrowave and Millimeter Signal Measurements: Tools and Best Practices http://cp.literature.agilent.com/litweb/pdf/5990-8892EN.pdfFundamentals of RF and Microwave Power Measurements (part 3)Power Measurement Uncertainty Per International Guidelines AN-1449-3, April 2011 http://cp.literature.agilent.com/litweb/pdf/5988-9215EN.pdfAverage Power Sensor Uncertainty Calculatorhttp://www.home.agilent.com/upload/cmc_upload/All/Average_Power_Sensor_Uncertainty_calculator_Rev6.xlsx

    Coaxial Connector Overviewwww.home.agilent.com/upload/cmc_upload/All/CoaxialConnectorOverview.pdf

    What Mates With Whathttp://na.tm.agilent.com/pna/connectorcare/What_mates_with_what.htm

    Connector Gradeshttp://na.tm.agilent.com/pna/connectorcare/Connector_Grades.htm

    INTERMATEABILITY OF SMA, 3.5 MM AND 2.92 MM CONNECTORSMicrowave Journal (Cables & Connectors Supplement) 3/2007, available at www.gore.com

    Connector Care Quick Reference Cardhttp://cp.literature.agilent.com/litweb/pdf/08510-90360.pdf

    Principles of Connector Carehttp://cp.literature.agilent.com/litweb/pdf/5954-1566.pdf

    External Waveguide Mixing and Millimeter Wave Measurements with Agilent PSA Spectrum Analyzers (app-note 1485)http://cp.literature.agilent.com/litweb/pdf/5988-9414EN.pdf