Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
Copyright © 2011 Agilent TechnologiesCopyright © 2011 Agilent Technologies Inc.
Optimizing Performance and Best
Practices for Millimeter Measurements
Ben Zarlingo, Agilent Technologies
EuMW 2011 Agilent Workshop Series
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Agenda
Millimeter Measurements Discussion
– Information
– Best Practices
– Optimizing Performance
Special Hazards, Problems
References, Resources
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Commercial/Consumer: 60 GHz Wireless HD$600 Transmitter + Receiver
$300
What is Millimeter-Wave?
Wavelengths ~≤ 10 mm, f ~≥ 30 GHz
Where Everything Gets
– Difficult
– Small
– Expensive
– Lossy
– Delicate
– Inflexible
– Banded (if not narrowband)
Graduate Work for the Electrical Engineer
$180
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Millimeter Performance Expectations
Performance Generally Less Ideal
– Harmonic architecture (phase noise, conversion loss, etc.)
– Path/insertion losses, impedance mismatches
– Preselector filtering
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Importance of Cables, Connectors
Influence of Materials, Structures, Geometries
Amplitude, 30-50 GHz
– Dropouts due to moding*– Amplitude, phase errors
– Errors with 3.5mm also, though perhaps more repeatable
– Error < 2dB in this example• Good enough?
• Repeatable enough?
Performance Gain fromcables may be Less Expensive
2.4 mm
SMA
*Moding: Excitation of the first circular waveguide propagation mode in the coaxial structure
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Plastic vs. Air Dielectric
Plastic (PTFE) vs. Air, Beads
– Air for metrology grade & many instrumentation grade
Many Other Tolerances & Construction Differences UsuallyAccompany Dielectric Change
SMA
3.5 mm
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Use A Torque Wrench for Consistency
“Wrench Lift Stress” is a special concern with severaldevices connected or a long device or adapter(s)
Lift
Angle between wrenches should be less than 90 deg
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Torque Values & Wrench Sizes
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Torque Problem Example
Loose Cable AmplitudeError
– Significant for you?What if it is variable andinconsistent?
Phase Variability
Sufficient Torque
– Enough torque to properly matesurfaces and to produce a strainthat will prevent loosening dueto vibration and/or thermalcycling
– Not enough to damage or distortconnectors, changing geometry
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Connector Grades
Metrology Grade– Highest precision, performance, repeatability
– Often slotless female center conductors (repeatable, lower inductance)
– Often used in calibration/verification kits
– Expensive, long life
PSC = Precision Slotless Connector – Typically metrology grade; PSC versions available for 3.5mm, 2.4 mm
Instrument Grade– Excellent repeatability, long life
– Often slotted female center conductors
General Purpose or Production Grade– Typical on components, cables, microstrip
– Inspect before connecting to better grades (analyzers), use adapter
– Limited connection cycles
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Connection Consistency
Consistent Connections Improve Repeatability– Same Cables (cable loss may be more than 3 dB/meter at 50 GHz)
– Same connectors, adapters
– Same cable routing
Every Connection & Transition is Imperfect– Insertion loss (one adapter = up to 1 dB loss at 50 GHz)
– Impedance change
– Bending can change cable characteristics
– Consider semi-rigid cable
Analyzer Accuracy is Improving →→→→ Smaller Errors Matter
Repeatability may be More Important than Absolute Accuracy
Standardize on Highest Frequency Connectors?– Consistency, but slightly higher loss at lower frequencies
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
mm-Frequency Connectors and Intermating
24 pt level 1
– 20 pt level 2
• 18 pt level 3
http://na.tm.agilent.com/pna/connectorcare/What_mates_with_what.htm
Uses a plastic (PTFE) dielectric
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
NMD Connectors
Precision, Ruggedized
Provides Support
Stabilizes Connection
– The NMD connector is a ruggedized test-port connector used on Agilent test sets and network analyzers. Most often, the connectors used are 3.5mm and 2.4mm male connectors. These connectors include a large threaded body that is specially designed to stabilize the test port cable when attached to the front of the analyzer or test set using Agilent test port cables. The design of these connectors allows the use of standard female 3.5mm and 2.4mm cables and adapters.
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Torque for Intermating
Proper Torque for Intermating Can Depend on Gender
– Example 1: A 56 N-cm. torque wrench should be used to connect male SMA connectors to either 3.5-mm or 2.92-mm (K) connectors
– Example 2: A 90 N-cm torque wrench should be used to connect male 3.5-mm or male 2.92-mm connectors to mating female connectors including SMA
Precision Connectors use Higher Torque
– Use lower torque with SMA ↔↔↔↔ SMA
– Use lower torque with male SMA male ↔↔↔↔ precision
Do not connect an SMA to a 3.5 mm calibration standard, VNA port cable or precision 3.5 mm device
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Important Lower Frequency Connectionsin a High Frequency System
Connectors Make a Difference in Some Low Frequency Measurements!
3.5 mm or SMA to BNC
Precision BNC connections on Oscilloscope
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Precision BNC - 3.5 mm vs. BNC - SMA
Air dielectric vs. PTFE
Split or slotted or hybrid center and outer connectors
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Understand Your Measurement Plane
Goal is Consistent Measurement Plane
Make Connector/Adapter/Cable Choices to Reduce or Eliminate Removable Connection as Part of DUT
Consistent Torque, Cable Routing
Goal is Simplest, Shortest Path
More Critical for Network Analysis
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Connection Corrections-Amplitude:Cables, Adapters, Probes
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Phase-Matched ConnectionsDoes Phase Matter?
Spectrum, Power, Demodulation
– Spectrum, Power (channel/band, ACP, etc.) insensitive to phase
– Demodulation is insensitive, or corrects for phase for single channel
– Demodulation can be sensitive to phase for 2-channel (i+jQ) measurements (quadrature error varying with frequency)
– Flatter (over the measured span) is better for demod
– Wider spans/bandwidths are more demanding
Phase Stability
– Cable, connector movement a problem for close-in phase noise
– See also microphonics
Phase Repeatability
– Not a problem for most measurements
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Coax & Waveguide Comparison
Flexible
Wide frequency range
Easy to change routing
Light weight
Can carry DC bias
Can be less expensive
Good match for component meas.
Disadvantages– Lossy, increasing with frequency
– Can be delicate
– Limited power handling
Low loss
High power handling ability
Mechanically durable
Disadvantages– Inflexible
– May need custom construction (expensive)
– Time delays for construction and reconfiguration
– Frequency range ≤ 1 octave
– Transitions needed to connect to analyzers, other coaxial elements
Coax Waveguide
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Agenda
Millimeter Measurements Discussion
– Information
– Best Practices
– Optimizing Performance
Special Hazards, Problems
References, Resources
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Beware of DC at Analyzer Input
Millimeter Agilent PXA, PSA Have No DC Block
Why? Internal DC Block Would Compromise Performance
Add External DC Block if Needed
Front Panel Male ConnectorProtects Analyzer
– Discourages direct mating withtypical coaxial male connector
– Encourages use of connector saver
– Male connector more durable,less likely to be damaged fromimproper mating
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Adapters as Connector Savers
Connectors are “Consumables” with a Limited Lifetime
Non-Millimeter Instruments Typically use Female Front Panel Connector– Use to mate male cables with
male analyzer connector
– Consider other types as neededin your application
Note “Flats” for Wrenches to Prevent Connector Rotation
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Contact Integrity, Concentricity
Visual Inspection, Suggest 10x Magnification
– Bending/displacement
– Non-concentric male or female
– Inspect mating planes for deep scratches, dents, debris
Damage to Female Slotted Connectors Most Common
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Damage Hazard: Non-Captive SMA Nut
Danger from SMA Male to 2.4 mm and 1.85 mm Female
3.5 mm to 2.4 mm adapter
3.5 mm end 2.4 mm end
SMA with non-captive nutand extended center pin
SMA with non-captive nut can be inserted into 2.4 mm
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Electrical Damage
High Frequencies →→→→ Small Size →→→→ Electrically Delicate
Burnout from Excess Power
Burnout from Source Transient
Static Zap
Detecting Damage (other than outright failure)
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
ESD Precautions
Wrist Strap
Table Mat
Heel Strap
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Microphonics, Other Movement
Phase Noise, Close-In Modulation
Can Affect Analyzer, Connections, DUT
Mitigating Microphonics
– Separate table, mounting
– Use isolating feed/pads/mounts
– Stabilize or support cables, waveguide
– Use phase-stable cables
Vibration, Thermal Cycling may Loosen Connections
– Torque check as part of measurement routine
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Alignment, Calibration, Correction
Calibration: Initial & Periodic Adjustment, Verification– Factory, service center, calibration lab
Alignment: Frequent Self Adjustment– Automatic adjustment compensates for temperature, drift, aging
– Alignment can affect dynamic range
Correction/Compensation– Often included in demodulation operations
– Substitution (power meters, etc.)
– May be only practical way to dramatically improve accuracy
– Enter parameters in analyzer for front-panel operation
– May be manually cumbersome or need switching
Perform Alignments Frequently and Whenever Measurement conditions Change– Temperature sensitivity generally goes up with frequency
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Preselector Centering
YTF Preselector (Filter)
– YTF operates open-loop
– Preselector tuning is characterized automatically by analyzer for full frequency range
– Centering performed by analyzer, very brief operation
– Perform centering for best amplitude accuracy at a single frequency
– Centering is initiated by user; not part of normal alignment
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Preselector Bypass
“Microwave Preselector Bypass”
Typically a Hardware Option
Removes YIG-Tuned Preselector Filter from Measurement Path
– Essential for wide-bandwidth measurements
– Removes preselector insertion loss but passeswideband noise; effect on analyzer DANLis variable
– May allow other signals into measurement
– Can tune to signal with preselector in placeand then bypass
Preselector
Bypass Path
Norm
Bypass
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Vector Calibrations for Modulated, Wideband Signal Generation & Meas.
Scalar Amplitude Corrections
– Scalar measurements: amplitude, power, phase noise, noise figure
– Narrowband
– CW, narrowband signals
Vector (Complex) Corrections
– I/Q or amplitude/phase
– Modulated signals
– Wideband measurements
32
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Example: RADAR Signal with IQ Errors
2 GHz Chirp: Gated sweep shows chirp spectrum and unwanted image from modulator IQ errors
33
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
RADAR Signal with IQ Corrections
800 MHz Chirp, 900 MHz Span: IQ-Correcting Modulated Signal Removes Almost All Image Energy
34
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Combining Preamp and Attenuation
Preamp Improves Sensitivity
Attenuation Improves Match (reduces mismatch error)
Combine to Optimize Match & Sensitivity
35
Std PathLNP
Preamp
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Hardware Features & Signal Processing Improve Performance
Low NoisePath
Noise FloorExtension
36
• Standard
• With LNP
• With NFE
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
New Calc. Method for Mismatch Uncertainty
Mismatch Uncertainty typically dominates
Better understanding, verification of probability distributions
Rayleigh model is an excellent fit, provides more accurate but still conservative estimates of uncertainty due to mismatch
Result: Lower Uncertainties due to mismatchby factor of 3 to 6
AN 1449-3 Power Measurement Uncertainty Per International Guidelines (updated April 2011)
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Better Calculations Indicate a Rayleigh Distribution
Probability density of the magnitude of the reflection coefficient is Rayleigh distributed if the probability density of both complex parts is Gaussian distributed
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Validating Assumptions with Real World Examples
Power
Sensor
Signal
Generator
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Validating Assumptions with Real World Examples
Signal Analyzer
Copyright © 2011 Agilent Technologies
© Agilent Technologies 2011
Additional InformationMicrowave and Millimeter Signal Measurements: Tools and Best Practices http://cp.literature.agilent.com/litweb/pdf/5990-8892EN.pdfFundamentals of RF and Microwave Power Measurements (part 3)Power Measurement Uncertainty Per International Guidelines AN-1449-3, April 2011 http://cp.literature.agilent.com/litweb/pdf/5988-9215EN.pdfAverage Power Sensor Uncertainty Calculatorhttp://www.home.agilent.com/upload/cmc_upload/All/Average_Power_Sensor_Uncertainty_calculator_Rev6.xlsx
Coaxial Connector Overviewwww.home.agilent.com/upload/cmc_upload/All/CoaxialConnectorOverview.pdf
What Mates With Whathttp://na.tm.agilent.com/pna/connectorcare/What_mates_with_what.htm
Connector Gradeshttp://na.tm.agilent.com/pna/connectorcare/Connector_Grades.htm
INTERMATEABILITY OF SMA, 3.5 MM AND 2.92 MM CONNECTORSMicrowave Journal (Cables & Connectors Supplement) 3/2007, available at www.gore.com
Connector Care Quick Reference Cardhttp://cp.literature.agilent.com/litweb/pdf/08510-90360.pdf
Principles of Connector Carehttp://cp.literature.agilent.com/litweb/pdf/5954-1566.pdf
External Waveguide Mixing and Millimeter Wave Measurements with Agilent PSA Spectrum Analyzers (app-note 1485)http://cp.literature.agilent.com/litweb/pdf/5988-9414EN.pdf