Upload
doannguyet
View
215
Download
0
Embed Size (px)
Citation preview
New High Resolution Dynamic Detectors and Flow Modifying Stents
for Neuro-Endovascular Image Guided Interventions (EIGI)
S Rudin, CN Ionita, A Kuhls-Gilcrist, C Keleshis, W Wang, DR Bednarek
(Supported by NIH grants R01 EB002873, NS43924, and EB00842501;
UB Fnd. IRDF and a Toshiba Med. Sys. Corp. equipment grant)
Endovascular vs. InvasiveNeuro-vascular Interventions
1. Arterial puncture vs. larger dissection of body cavity or skull
2. Use of catheter vs. scalpel, drills, clamps, etc.
3. X-ray image guidance vs. visual viewing
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, baloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, polyurethane film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, polyurethane film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Catheters
guiding catheter introducer
microcatheters
flexible
Balloons
Stent on balloon
Stents (on Balloon)
Stenotic vessel with plaque.Stents: drug eluting or bare; covered or not;
material: stainless steel, nitinol, dissolvable(?)expand: balloon or self
Various lengths and diameters
Tetra (Guidant)
Velocity (Cordis)
Stents
1. Integra, nitinol self-expanding (J and J)
2. Multilink, stainless steel, balloon expanding (Guidant)
3. Needle, 26 Ga4. Wallstent, stainless
steel, self-expanding (Boston Scientific)
1
2
3
4
ECOSMicroLysUS
Neurojet
LaTis laser
G. M. Nesbit et al: JVIR 2004; 15:S103–S110
Clot Busting
InTime Retriever and Ensnare
Merci Retriever
Neuronet
Clot Removal Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Flow Modification: Treatment of intracranial aneurysms using asymmetric stents
Perforators (50-200 µm)
Asymmetric Stent in 2-3 mm vessel
C. Ionita
Flow Modifying Stent for Aneurysm Treatment
C. Ionita, et al
Pre-stent Post-stent
Use of 3D forComputational Fluid Dynamics (CFD)
Treatment Planning
Obtain 3D model (vessel or aneurysm)
Create a mathematical grid
Solve the differential equations(Navier-Stokes equations)at each grid pointat each time point
Asymmetric Stent Altering Aneurysm Flow
Untreated StentedStreamlines
Untreated StentedWall Shear Stress
WSS(dyne/cm2)
Untreated Stented
Velocities
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Flow Modifying Asymmetric Vascular Stents (AVS)
C. Ionita, et al
Balloon expandableStainless steel stentStainless steel mesh flow modifierOpen cell
Balloon expandableStainless steel stentPolyurethane film flow modifierOpen cell
Self expandingNitinol stentsPTFE film flow modifier
Open Cell
Closed Cell
Asymmetric Vascular Stents (AVS)
Ionita CN, Chang C, Sinelnikov A, Bednarek DR, Rudin S: Angiographic analysis of aneurysms treated witha novel self-expanding asymmetric vascular stents (SAVS) (abstract). Medical Physics, June 2009, WE-C-304A-5.
Stent designs such as Enterprise and Wingspan
Clo
sed
Cel
l
Ope
n C
ell
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Region of Interest (ROI) Micro-angiography and Fluoroscopy for Image-guided
Interventions (IGI)
Detector requirements:� Angiography, ROI-CT, and new device imaging:
• High resolution (> 4 Lp/mm)� Real-time fluoroscopy for image guidance:
• High sensitivity - low instrumentation noise (quantum noise limited)• High speed (30 fps)• No lag (high temporal resolution)
� ROI imaging: reduce integral radiation dose
Detectors:� High Sensitivity Micro-Angiographic Fluoroscope, MAF� Solid State X-ray Image Intensifier, SSXII
New High Resolution Detectors• Micro-angiographic detector, MA (not fluoro capable )• High Sensitivity Micro-angiographic Fluoroscope,
MAF (with light amplifier for gain)• Solid State X-ray Image Intensifier, SSXII
(built-in EMCCD gain, fluoro capable)
FOT
FOP
FOP
CCD Camera
CsI
MA detector (no gain) MAF detector
FOP
FOP
FOPs
Gen2 LA withMCPs
CCD Camera
CsI
Photocathode
FOTFOT
FOP
FOP
EMCCD Camera
CsI
SSXII detector
New High Resolution Detectors• Micro-angiographic detector, MA (not fluoro capable )• High Sensitivity Micro-angiographic Fluoroscope,
MAF (with light amplifier for gain)• Solid State X-ray Image Intensifier, SSXII
(built-in EMCCD gain, fluoro capable)
FOT
FOP
FOP
CCD Camera
CsI
MA detector (no gain) MAF detector
FOP
FOP
FOPs
Gen2 LA withMCPs
CCD Camera
CsI
Photocathode
FOTFOT
FOP
FOP
EMCCD Camera
CsI
SSXII detector
The High-Sensitivity Microangiographic Fluoroscopic (MAF) Detector
� High sensitivity for fluoroscopic applications
• Direct fiber-optic coupling (no lenses): Fiber-optic plate (FOP) windows, Fiber-optic taper (FOT)
• LII (LA) with MCPs for variable high system gain
� High resolution (~ 35 µm pixel in equal-area-framing)
MCP LII
CsI(Tl)
CCD Camera
FOT
FOP
FOP
FOP
MCP LII
CsI(Tl)
CCD Camera
FOT
FOP
FOP
FOP
CCD CameraCCD CameraCCD Camera
FOT
FOP
FOP
FOP LIICCD Camera
FOT
Power Supply for the LII
Ionita CN, Keleshis C, Jain A, Bednarek DR, Rudin S: Testing of the high-resolution ROI micro-angio fluoroscope (MAF) detector using a modified NEMA XR-21 phantom (abstract). Medical Physics, June 2009, MO-FF-A4-3.Jain A, Bednarek DR, Rudin S: Performance evaluation of a custom-made anti-scatter grid used for the high-resolution Micro-Angiographic Fluoroscope (MAF) (abstract). Medical Physics, June 2009, WE-C-304A-3.
Line-Pair Phantom
MAF and FPD Detectors
Wang W, Keleshis C, Kuhls-Gilcrist A, Ionita CN, Jain A, Bednarek DR, RudinS: New High-Resolution-Detector Changer for a Clinical Fluoroscopic C-Arm Unit (abstract). Medical Physics, June 2009, SU-FF-I-171
High Resolution ROI Detector Implementation
DeployedRetracted
Roadmaps of AVS treated-AneurysmsMAF XII
Un-
depl
oyed
Dep
loye
d
New High Resolution Detectors• Micro-angiographic detector, MA (not fluoro capable )• High Sensitivity Micro-angiographic Fluoroscope,
MAF (with light amplifier for gain)• Solid State X-ray Image Intensifier, SSXII
(built-in EMCCD gain, fluoro capable)
FOT
FOP
FOP
CCD Camera
CsI
MA detector (no gain) MAF detector
FOP
FOP
FOPs
Gen2 LA withMCPs
CCD Camera
CsI
Photocathode
FOTFOT
FOP
FOP
EMCCD Camera
CsI
SSXII detector
Imaging Imaging AreaArea
FrameFrame
Serial Serial RegisterRegister
MultiplicationMultiplicationRegisterRegister
Transfer AreaTransfer AreaOutputOutput
AmplifierAmplifier
Electron Multiplying CCD
( ) xGain 2000019.1 400 ≈=
( ) xGain 10006.1 400 ≈=22 V
15 V
Current State-of-the-ArtSSXIIAcquisition Parameters: 70kVp; 160mA; 45ms; 2” PMMA; 0.3mm Focal Spot; Identical Geometry
High-Resolution Imaging
Kuhls-Gilcrist A, Bednarek DR, Rudin S: Component analysis of a new solid state x-ray image intensifier (SSXII) using photon transfer. SPIE vol. 7258, 2009. In: Proc. from Med. Imaging 2009: Physics of Med. Imaging, Orlando, FL, # 7258-42, 725817:1-10.
Initial Images: Asymmetric Stent
Kuhls AT, Yadava G, Rudin S: Progress in electron-multiplying CCD (EMCCD) based, high-resolution, high-sensitivity x-raydetector for fluoroscopy and radiography. SPIE 6510-47, 2007. In: Proc. Med. Imag. 2007: Phys. of Med. Imag., San Diego, CA.
Kuhls-Gilcrist AT, Yadava GK, Patel V, Bednarek DR, Rudin S: The Solid-State X-Ray Image Intensifier (SSXII): AnEMCCD-Based X-Ray Detector. SPIE 6913-19, 2008. In: Proc. Med. Imag. 2008: Phys. of Med. Imag., San Diego, CA.
SSXII XII
3.3 mR
90 µR
100 µm Au
50 µm Pt
100 µm I
AVS w, polyUPatch marked
Detector Mosaic Array(Extended FOV)
Rudin S, Yadava G, Josan G, Kuhls A, Rangwala H, Wu Y, Ionita C, Bednarek DR: New light-amplifier-baseddetector designs for high spatial resolution and high sensitivity CBCT mammography. SPIE vol. 6142,pp. 6142R1-11, 2006. In: Proc. Med. Imag. 2006: Phys. of Med. Imag., San Diego, CA, paper #63.
Wang W, Keleshis C, Kuhls-Gilcrist A, Bednarek DR, Hoffmann KR, Rudin S: Control, Acquisition, Processing, and Image Display System (CAPIDS) for the Solid-State X-Ray Image Intensifier (SSXII) (abstract). Medical Physics, June 2009, SU-FF-I-170.
Control, Acquisition, Processing, and Image Display System
(CAPIDS)
Control, Acquisition, Processing, and Image Display System(CAPIDS)
Control, Acquisition, Processing, and Image Display System(CAPIDS)
Control, Acquisition, Processing, and Image Display System(CAPIDS)
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Errors in Conventional Slit Method
(Simulation)
Slit Width
Slit Angle
3 % Added Noise
Jain A, Patel V, Kuhls-Gilcrist A, Bednarek DR, Hoffmann KR, Rudin S:Effect of point spread function, x-ray quantum noise, and additiveinstrumentation noise on the accuracy of the angulated slit method fordetermination of pre-sampled detector MTF. (abstract).Medical Physics, June 2009, SU-FF-I-108.
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
• GOAL: Provide a Simple and Accurate Presampled MTF Measurement Technique for Digital Radiography Systems
• Use only the Detector Noise Response
• Inherently 2-D
Line Spread Function
Noise Response Method
Kuhls-Gilcrist A, Jain A, Bednarek DR, Rudin S: A method for measuring the MTF of digital radiography systems using noise response (abstract). Medical Physics, June 2009, WE-C-304A-7.
Cascaded Linear Systems: Results
Output NPS:
( ) ( )[ ] ( )vuNPSygxvuTvuAgyxvuNPS ADDSYSS , ,,~),( 4421 +Φ∆∆+∆∆= −
(((( ))))(((( ))))(((( ))))
signaloutput
factor conversionnumber digital-to-electron
noise electronic additive,
factorSwank dependent -frequency,
MTF presampled,
ray)- xabsorbedper (DNgain system ~ widthpixel ,
4
4
============================
Φ
∆∆
g
vuNPS
vuA
vuT
g
yx
ADD
S
SYS
Cascaded Linear Systems: Results
Output NPS:
( ) ( )[ ] ( )vuNPSygxvuTvuAgyxvuNPS ADDSYSS , ,,~),( 4421 +Φ∆∆+∆∆= −
Primary Quantum + Poisson Excess Noise
Secondary Quantum Noise
Additive Noise Primary Quantum + Poisson Excess Noise
Cascaded Linear Systems: Results
Output NPS:
( ) ( )[ ] ( )vuNPSygxvuTvuAgyxvuNPS ADDSYSS , ,,~),( 4421 +Φ∆∆+∆∆= −
Secondary Quantum Noise
Additive Noise
Presampled MTF is Inherently in the Noise
Response!
Image Simulations: Simple Detector Model
• 1000 x 1000, 32 µm Pixels
• 150 µm CsI Phosphor
• “True” MTF Known Exactly
NR Method: Procedure• Acquire 30 Flat-Field Images at Several mAs
Values [IEC Guidelines; Standardized Spectrum (RQA)]
• Measure NPS(u, v)
NR Method: Procedure• Acquire 30 Flat-Field Images at Several mAs
Values [IEC Guidelines; Standardized Spectrum (RQA)]
• Measure NPS(u, v)
• Plot NPS(u, v) versus Signal and Fit with 2nd Order Poly.
NR Method: Procedure• Acquire 30 Flat-Field Images at Several mAs
Values [IEC Guidelines; Standardized Spectrum (RQA)]
• Measure NPS(u, v)
• Plot NPS(u, v) versus Signal and Fit with 2nd Order Poly.
• Fit Quantum NPS (Slope Data)( ) ( )( )
5
2
4
23
21
MixtureGuassian expF hh
hfh
fA
hf
S
+
−−=
Presampled MTF
NR Method: ResultsAveraged
Deviation = 0.3%
NR Method: ResultsER Averaged
Deviation > 30%
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Detector Noise Evaluation: Instrumentation Noise Equivalent
Exposure (INEE) Model*
Nexp2 = k ( E + INEE )
IF E>INEE, then quantum noise limited.IF E<INEE, then instrumentation noise limited.
*Kuhls-Gilcrist A, Bednarek DR, Rudin S: The Instrumentation Noise Equivalent Exposure (INEE): Including conversion, secondary quantum, structure and electronic noise (abstract). Medical Physics, June 2009, WE-C-304A-8.
Noise Versus Exposure
INEE = 2.0 µR
INEE
InstrumentationNoise Limited
QuantumNoise
Limited
Instrumentation Noise Limited?
Quantum Noise Limited
Instrumentation Noise Limited
Advances in EIGIs Devices and Detectors*• Devices
– Catheters and stents– Clot busters– Flow modifying asymmetric vascular stents (AVS) for aneurysms
• Stainless, balloon expandable, open cell, steel mesh flow modifier• Stainless, balloon expandable, open cell, polyurethane film modifier• Nitinol, self-expanding, open cell, PTFE porous film flow modifier• Nitinol, self-expanding, closed cell, PTFE porous film flow modifier
• High Resolution Detectors– Micro-angiographic (MA) detector– Micro-angiographic Fluoroscope (MAF)– Solid State X-ray Image Intensifier (SSXII)
• New Imaging System Evaluation Concepts– MTF only from noise response measurements– Instrumentation Noise Equivalent Exposure and Quantum
Limited Performance– GMTF and GDQE
*Rudin S, Bednarek DR, Hoffmann KR: Endovascular Image Guided Interventions (EIGI).Vision 20/20 paper. Medical Physics 35(1): 301-309, Jan 2008.
Phantom: Scattering fraction, ρ
MAF Detector, D
X-ray tubeFocal spot, F
Patient Table
Factors affecting GMTF and GDQE for given object plane:detectorscattering objectfocal spot
Image mag, m
)()()1
()1(),,(m
fMTF
m
fMTFf
m
mMTFmfGMTF DSF
+−−= ρρρ
� Generalized Modulation Transfer Function (GMTF)
� Generalized Noise Power Spectrum (GNPS)
222
),(
)(
),(),,(
m
Xm
fNNPS
Xdm
Xm
fNPS
mXfGNNPSDD
==
� Generalized Noise Equivalent Quanta (GNEQ)
),,,(
),,(),,,(
2
mXfGNNPS
mfGMTFmXfGNEQ
ρρρ =
� Generalized Detective Quantum Efficiency (GDQE)
),(
),,,(),,,(
2 mXm
mXfGNEQmXfGDQE
inΦ= ρρ
� The generalized parameters are defined with reference to the object plane
Generalized System Evaluation Metrics
Kyprianou I, Rudin S, Bednarek DR, Hoffmann KR: Generalizing the MTF and DQE to include x-ray scatter and focalspot unsharpness: Application to a new micro-angiographic system for clinical use. Medical Physics, 32(2): 613-626, 2005.
Impact of varying the x-ray tube focal-spot size and air gaps
GMTFs
� Yadava et al., AAPM 2005, Medical Physics 32 (6), 2080 (2005)
0.3 mm focal-spot 0.6 mm focal-spot
DQE/GDQE
Summary (Educational Objectives):
1. Appreciate the progress being made in improved EIGI devices and in particular flow modifiers such as the asymmetric vascular stent (AVS) for aneurysm treatment.
2. Understand the operation of new high-resolution micro-angiographic systems including the MAF and SSXII.
3. Understand new objective image detector evaluations including INEE, GMTF, GDQE, and determination of MTF from noise response measurements alone.
Separate Structure, Quantum, Electronic Noise
INEE – Effect on DQE
Always Instrumentation Noise Limited
Instrumentation Noise Limited @ 2 µR
(((( )))) (((( ))))EvuTKgaEvu CCDADCpixel ,~ˆ,,NPS 22Secondary ηΦ====
(((( )))) (((( ))))EvuTgaEvu SYSpixel ,~ˆ,,NPS 222Primary ηΦ====
(((( )))) (((( )))) Eg
vuTgaEvuCsI
gSYSpixel
CsIε
η ,~ˆ,,NPS 222Excess Φ====
(((( )))) (((( ))))vuSEvu ADD ,,,NPSElectronic ====
(((( )))) (((( )))) 2Structure ,,,NPS EvuSEvu Structure====
Theoretical NPS(((( )))) (((( )))) (((( )))) (((( )))) (((( ))))vuSEvuTK
ggvuTgaEvuSEvu ADDPixelADC
CsI
gSYSpixelStructure
CsI ,,1~,~ˆ,,,NPS 2222Measured ++++
++++
++++++++====
εηΦ
(((( )))) (((( )))) (((( )))) (((( ))))vuCEvuBEvuAEvu ,,,,,NPS 2Measured ++++++++====
ElectronicSecondaryExcessPrimaryStrctureMeasured NPSNPSNPSNPSNPSNPS ++++++++++++++++====
Detector Comparison*
Detector System INEE Pixel (µm) Lag (frames) Frame Rate (fps)
MA 20.4 µR 43 None 4
MAF, SSXII <0.1 µR 35-48 None 30
XII <0.2 µR 120-300+ None 30
FPD** 2.75 µR ~300 5+ 30
*Kuhls-Gilcrist A, Jain A, Bednarek DR, Rudin S: Instrumentation Noise Equivalent Exposure (INEE):an investigation of spatial frequency effects (abstract). Medical Physics 2008, SU-DD-A4-6.
*Szczykutowicz, T, Kuhls-Gilcrist A, Bednarek DR, Rudin S: Instrumentation noise equivalent exposure(INEE) for routine quality assurance: INEE measurements on a clinical flat panel detector (abstract). Medical Physics 2008, MO-E-332-7.
**Roos et al., “Multiple gain ranging readout method to extend the dynamic range of amorphous silicon flat panel imagers”, Physics of Medical Imaging, Proc SPIE 5368, 139-149 (2004).
Dual Detector ROI Cone Beam CT
1. Chityala R, Hoffmann KR, Rudin S, Bednarek DR: Region-of-interest (ROI) Computed Tomography: Combining dual resolution XRII images (abstract). Medical Physics, 32(6): 2056, June 2005, MO-D-I-611-4. (AAPM05)2. Patel V, Ionita CN, Keleshis C, Sherman J, Hoffmann KR, Bednarek DR, Rudin S: First implementation of high-resolution dual-detector Region-of-Interest Cone-Beam Computed Tomography (ROI-CBCT) for a rotating C-arm gantry system (abstract). Medical Physics, TH-C-332-2. (AAPM08)3. Patel V, Hoffmann KR, Bednarek DR, Rudin S: Automatic registration technique for rotating-gantry dual-detector Region-of-Interest Cone-Beam Computed Tomography (ROI-CBCT) (abstract). Medical Physics, TH-D-332-4 (AAPM08)
Dual Detector ROI CBCT
Ionita CN, Patel V, Keleshis C,Hoffmann KR, Bednarek DR, Rudin S: Update on the development of a new dual detector (Micro-Angiographic Fluoroscope/Flat Panel) C-arm mounted system for endovascular image guided interventions (EIGI) (abstract). Medical Physics 2008, MO-D-332-7.
High Resolution ROI Detector Implementation
DeployedRetracted
MAF projection of stent deployed in rabbit carotid artery near constriction.
Dual Detector ROI-CBCTFPD-CBCT
First in –vivo3D ROI-CBCT
of a Stent
DQE of MAF
Asymmetric Vascular Stent(AVS)
Aneurysm treated with AVS
Side View Transverse View
Schematics of the AVS-treated aneurysms