Upload
techne-phobos
View
235
Download
0
Embed Size (px)
Citation preview
7/29/2019 EMR Configuration Guide_v.7.pdf
1/18
WNE CLIENT V.4.2EMR CONFIGURATION GUIDESEPTEMBER 30, 2010
WNE v.4.2
7/29/2019 EMR Configuration Guide_v.7.pdf
2/18
This page left intentionally blank
7/29/2019 EMR Configuration Guide_v.7.pdf
3/18
EMR Configuration Guide WNE v.4.2 i
Table of Contents
Introduction...........................................................................................................................................1Glossary of Abbreviations ...............................................................................................................1
The WNE PC................................................................................................................................2
Supplied Equipment.....................................................................................................................2
Physical Equipment Installation and Connections .....................................................................3
Global satellite coverage ......................................................................................................................4Satellite Parameters ......................................................................................................................4
Accessing the EMR via the web interface ..........................................................................................4Locking to the WNE signal..................................................................................................................5
Saving a configuration file .................................................................................................................10
Troubleshooting..................................................................................................................................11
Unable to lock to the WNE signal.................................................................................................11Connection lost after EMR reboot.............................................................................................12
FAQ .....................................................................................................................................................12EMR Technical Specifications ..........................................................................................................13
Platform...........................................................................................................................................13
Satellite Receiver RF Tuner (single and dual tuner models).......................................................14
Demodulation and Error Correction..............................................................................................14
LNB Power and Switching ............................................................................................................14
Safety...............................................................................................................................................14
7/29/2019 EMR Configuration Guide_v.7.pdf
4/18
EMR Configuration Guide WNE v.4.2 ii
This page left intentionally blank
7/29/2019 EMR Configuration Guide_v.7.pdf
5/18
EMR Configuration Guide WNE v.4..2 1
IntroductionReuters World News Express (WNE) is a file based television delivery system that transmits Reuters Video News(RVN) material, principally over satellite, as MPEG 2 video and XML scripts. The SkyStream Edge Media Receiver(EMR) is a DVB receiver that extracts IP content from the WNE satellite feed. This IP content is then reassembledinto video and script files by the zBand software and stored on your WNE PC.
The EMR that you will have received from Reuters with your WNE installation is shown below:
The Web-based interface for the EMR is shown below:
Although, your EMR comes preconfigured for use, there are several procedures that you must complete yourself in
order to begin receiving data from Reuters WNE satellite. Each of these procedures are described within thisdocument:
Connect the EMR to the WNE PC (described in the Physical Equipment Installation and Connections section of
this document) Tune the receiver to the Reuters Satellite signal and verify that IP data is being received and stored locally on
your WNE PC (the process by which this is accomplished is described within several sections of this document).
Glossary of Abbreviations
Abbrev iation /Term Descr iption
C-BandC Band is the original frequency allocation for communications satellites. C-Band uses3.7-4.2GHz for downlink and 5.925-6.425Ghz for uplink.
DNS Domain Name Server
DVB Digital Video Broadcasting
EMR Edge Media Receiver - the SkyStream satellite receiver used for WNE reception
FECForward Error Correction is a system of error control for data transmission, wherein thesender adds redundant data (also known as an error correction code) to its messages.
GUI Graphic User Interface
7/29/2019 EMR Configuration Guide_v.7.pdf
6/18
EMR Configuration Guide WNE v.4..2 2
Abbrev iation /Term Descr iption
IFL
Generically, IFL (Intra-Facility Link) refers to a cable which runs between two facilities.
In satellite usage, the term IFL refers to the cable which connects the ODU (Out-Door-Unit) with the IDU (In-Door-Unit).
IP/DVBThis is the phrase used to describe the delivery of IP content over a DVB satellitenetwork.
Ku Band
The Ku band (pronounced "kay-yoo") is a portion of the electromagnetic spectrum withinthe microwave range of frequencies. Ku band is primarily used for satellitecommunications particularly from remote locations back to a television network's studiofor editing and broadcasting.
LAN Local Area Network
L-Band
L band is a frequency range between 390MHz and 1.55GHz which is used for satellitecommunications and for terrestrial communications between satellite equipment. An LNBis used to convert Ku and C Band frequency bands to L band, which can be transmitted
over the IFL and processed by the IDU.
LNBA low-noise block converter (LNB, for low-noise block, or sometimes LNC, for low-noiseconverter) is the (receiving, or downlink) antenna of a satellite dish commonly used forsatellite TV reception.
LOA device used to output the frequency of an incoming modulated radio-frequency carriersignal, to produce the required intermediate frequency.
MPEG Motion Pictures Experts Group. Video file format.
NIC Network Interface Card
PID (Process IDentifier) A temporary number assigned to a process or service.
URL(Uniform Resource Locator, previously Universal Resource Locator) is the uniqueaddress for a file that is accessible on the Internet.
WNE World News Express - IP file-based delivery system for Reuters news content.
Reuters Live Linear-based delivery system for Reuters news content.
The WNE PC
The WNE PC is equipped with a Stradis MPEG2 decoder card capable of playing out video content in both compositevideo and serial digital formats. The WNE PC also comes pre-installed with a second network card (NIC) which
allows video content and scripts to be transferred from it to your server infrastructure.Video stories received from the Reuters satellite can be previewed on the WNE PC via the WNE website (providedthat WNE PC has internet access). In addition, the WNE website can be used to control which video stories areplayed out via the Stradis card on the WNE PC (refer to the Online Help system within the WNE website forinstructions on how to do this).
Supplied Equipment
The standard equipment supplied by Reuters with your WNE installation includes:
EMR satellite receiver
WNE PC
able to store 10 days of video content (depending on the size of the files and the amount of content being
stored)
7/29/2019 EMR Configuration Guide_v.7.pdf
7/18
EMR Configuration Guide WNE v.4..2 3
Integrated Stradis MPEG decoder card with composite and SDI Output.
Audio cable with 2 XLR connectors for balanced audio outputs and 1 XLR connector for digital audio output.
Keyboard and mouse
Physical Equipment Installation and ConnectionsThe EMR can be installed at your client site along with the WNE PC. The instructions below assume that:
both the zBand client software and other necessary Reuters applications are already installed on the WNE PC
a monitor (not supplied with the WNE installation), keyboard and mouse are connected to the WNE PC.
Steps
1. Mount the EMR in a 19-inch standard rack (or other appropriate location) using the screws and bracketsprovided.
2. Connect an RF coaxial cable receiving the Reuters Live/WNE signal to the RF input on the rear of the EMR.
3. Connect WNE PC (supplied as part of your WNE installation) to the EMR using an Ethernet cable.
4. Connect a network cable (e.g., a cross over cable) from the LAN 2 port on the EMR to the on-board networkinterface card (NIC) on the rear of the WNE PC.
7/29/2019 EMR Configuration Guide_v.7.pdf
8/18
EMR Configuration Guide WNE v.4..2 4
Global satell ite coverage
Satellite Parameters
Global satellite coverage for the WNE (and Reuters Live) services are transmitted via the regional satellite providerslisted below. You must ensure that your satellite dish is actually positioned to point to the relevant location.
North America Central America South America
Satellite Satmex 6 Satmex 6 Satmex 6
Position 113 West 113 West 113 West
Transponder 15K 15K 15K
D/L Frequency11992.65 MHz(Ku-Band)
11992.65 MHz(Ku-Band)
11992.65 MHz(Ku-Band)
Symbol Rate 10.8 MS/s 10.8 MS/s 10.8 MS/s
FEC 1/2 1/2 1/2
Polarization Horizontal Horizontal Horizontal
Figure 1 Satellite coverage for North, Central and South America
Europe MEA Asia
Satellite NSS-7 Arabsat 5A Asiasat 5
Position 22 West 30.5 East 100.5 East
Transponder 1 5 C9H
D/L Frequency10.9857 MHz(Ku-Band)
3906.75 MHz(C-Band)
3960 MHz(C-Band)
Symbol Rate 7.2 MS/s 7.2 MS/s 27.5 MS/s
FEC 3/4 3/4 3/4
Polarization Horizontal RH Circular Horizontal
Figure 2 Satellite coverage fo r Europe, MEA and Asia
Accessing the EMR via the web interfaceThe most efficient method by which to configure your EMR is via the web-based interface.
1. Turn on your EMR and the WNE PC. Accessing the EMR through the WEB GUI requires that both devicesshare the same subnet address. All EMR LAN 2 ports default to an IP address of 172.16.2.2. (LAN 1 is suppliedwith a default of 172.16.1.2).
2. Change the IP address of the onboard NIC to match the subnet address on the EMRs LAN2 port (e.g.172.16.2.100), with a Subnet Mask of 255.255.255.0.
3. Enter the LAN 2 EMR IP address into an internet browser window.
7/29/2019 EMR Configuration Guide_v.7.pdf
9/18
EMR Configuration Guide WNE v.4..2 5
Note:A shortcut link may already exist in your browsers Links section called Skystream Networks Edge MediaRouter(also accessible via 172.16.2.2).
4. The following EMR page will then be displayed. If you select one of the menu options in the left hand navigationbar, you will be prompted to login (as shown below).
Figure 3 EMR web interface with login page
5. To login, enter the default useremradmin. No password is required.
Note: If you are accessing the site for the first time or have just rebooted the EMR, you must supply the default username. Otherwise, you will generally not be prompted for a login.
The initial connectivity between your WNE PC and EMR has now been established. Next, you will need toensure that you have signal lock to the satellite.
Locking to the WNE signal1. Select Satellite Receiverfrom beneath the Statistics menu in the left hand navigation bar. The Statistics-
Satellite Receiverpage will display the NO LOCK status for the key values shown below. This indicates thatsatellite parameters have not yet been entered into the EMR.
7/29/2019 EMR Configuration Guide_v.7.pdf
10/18
EMR Configuration Guide WNE v.4..2 6
Figure 4 Statistics page with no signal lock
2. Select Satellite Receiverfrom the Configuration menu in the left hand navigation bar. The Configuration Tuner 1 page will be displayed.
Figure 5 Configuration of EMR tuner settings
3. Click Change Tuner Settings . The Configuration-Satellite Receiver 1 page will be displayed.
7/29/2019 EMR Configuration Guide_v.7.pdf
11/18
EMR Configuration Guide WNE v.4..2 7
Figure 6 Tuner conf iguration page
4. Enter the parameters for the satellite that is applicable to your location. Refer to the tables in Figures 1 and 2 ofthe Global Satellite Coverage section of this document for the relevant regional downlink parameters. Thesettings you must reference are the D/L Frequency (L-Band Frequency), Symbol Rate and Polarization.
L-Band Frequency is o ften noted as:
KU Band Downlink Frequency LNB LO Frequency.
C Band LNB LO Frequency Downlink Frequency
A high proportion of LNBs used by subscribing Reuters Live clients tend to have a LO (Local Oscillator)frequency of 10GHz. In the example in Figure 6 above, the L-Band Frequency value is 1200MHz.
In the example below, for an EMR unit in Europe, the frequency value is obtained by subtracting the D/L SatelliteFrequency value from NSS-7 for the Europe region (see Figure 1) from a local oscillator (LO) frequency of10GHz or 9.75GHz.
A few sample calcu lations are show below:
10GHz LO i.e. 109857 MHz 10000MHz = 985.7MHz.
9.75GHz LO i.e. 109857 MHz - 9750MHz = 1235.7MHz.
5. Once the correct parameters have been entered, clickApply to accept the changes. If the entered parametersare correct, the Configuration - Tuner 1 page will be displayed. The Data Lock (FEC lock) field should nowdisplay a green lock (as shown in Figure 9). This indicates that the EMR has locked to the WNE satellite signal.
Note: If a red light initially appears in the Data Lock (FEC lock) field, press F5 to refresh. The green lock lightshould then appear in that field.
7/29/2019 EMR Configuration Guide_v.7.pdf
12/18
EMR Configuration Guide WNE v.4..2 8
Figure 7 Successful data lock
Note: Refer to the Troubleshooting section at the end of this document if the signal has not locked.
6. Once you have successfully locked onto the Reuters signal, you must configure the EMR to extract the IP datafrom the satellite signal. This will include setting the relevant PID for the signal. Packet IDs (PIDs) are identifiersused to tag different MPEG component packets within a satellite signal (e.g. video, audio or in the case of WNE,IP data). WNE IP data is carried within the satellite signal on PID 3550.
Select PID Selection from the Configuration menu in the left hand navigation bar. The Configuration PID
Selection page will be displayed.7. Click PID Selection. The PID Selection Receiver 1 page will then be displayed.
Figure 8 New PID addition
7/29/2019 EMR Configuration Guide_v.7.pdf
13/18
EMR Configuration Guide WNE v.4..2 9
8. Enter the PID value of3550 in the New PID field and select the decimal option. ClickAdd to add this PID to theAc tive PIDs list for the EMR.
9. The EMR is now successfully locked to the WNE signal and is extracting IP data from the appropriate PID value.All IP data is carried on a multicast IP address and must be routed to the relevant EMR Ethernet port (e.g.,Ethernet 2/LAN2). Standard WNE installations send received content through Ethernet 2 port on the EMR.
10. In order to distinguish between the different transmission standards, PAL (625) and NTSC (525), each formatexists on a separate multicast address. These are provided below:
Transmission paths multicas t addresses
525 Satellite Transmission Path 224.14.14.14
625 Satellite Transmission Path 224.14.14.15
11. Select SkyCast from the Configuration menu in the left hand navigation menu. The Configuration - SkyCastpage will then be displayed.
12. Click Edit on themulticast IP address (224.14.14.14 or 224.14.14.15).The Edit SkyCast Rule page will bedisplayed. Select the Forward and Ethernet2 options in theAc tion field for thismulticast address.
Figure 9 SkyCast Rule for the EMR
13. ClickApp ly to designate that IP data should be sent to Ethernet 2 and confirm your SkyCast Rule settings.
14. Your IP data should now be being received and processed correctly.
You can verify data reception via the Network Interface Traffic and Statistics - Skycast pages. You canaccess these pages by selecting the Network Interface orSkyCast options from the Statistics menu in the lefthand navigation bar.
7/29/2019 EMR Configuration Guide_v.7.pdf
14/18
EMR Configuration Guide WNE v.4..2 10
Figures 13 & 14 below demonstrate the processing and destination of IP packets via the EMR.
Figure 10 Network Interface Statistics
Figure 11 SkyCast Statistics
15. Be sure to save all the configuration values you have just set to a conf iguration file! The process by whichto do this is described in the Saving a Configuration File section of this document.
Saving a configuration fi leOnce the EMR is correctly connected to the WNE signal, extracting IP data from the appropriate multicast addressand forwarding this IP data to Ethernet 2, this configuration should be saved as the default setting.
7/29/2019 EMR Configuration Guide_v.7.pdf
15/18
EMR Configuration Guide WNE v.4..2 11
1. Select Configuration Filefrom the Maintenance menu in the left hand navigation bar. The Maintenance-Configuration File page will be displayed.
Figure 12 Default configuration f ile for an EMR
Note:A default configuration file may already exist for this EMR. However, it is recommended that a new default filebe created with the settings you have just configured. This default profile will then be loaded whenever the EMR isrebooted. In addition, if during the routine operation of your EMR, you or a Reuters support representative makeany changes to the configuration settings of the EMR, you should save these changes to the configuration file toensure that they are preserved whenever the EMR is rebooted.
2. Set the following fields on the Maintenance-Configuration File page:
Enter a suitable name for this configuration in the Save current configuration as field.
It is recommended that you enter the date that the profile was created In the Configuration descriptionfield.
Ensure that the Default option has been selected.
3. Once these details have been entered, click Save to save this configuration.
4. It is recommended that you now reboot the EMR to ensure that the configuration is loaded correctly upon startup.
5. To reboot, select System Reboot from the Maintenance menu in the left hand navigation bar. A Reboot Nowbutton will be displayed. Click it. The EMR requires approximately 60 seconds to fully reboot and lock onto theWNE signal.
You EMR configuration is now complete.
Note: If the configuration file is not correctly loaded when your EMR is rebooted, ensure that you have properly setall your configuration settings and saved them to the configuration file or contact a Reuters support representative
for assistance.
Troubleshooting
Unable to lock to the WNE signalItems to verify:
Verify the LNB values that have been entered (refer to Figures 7 & 8). Several WNE installations haveuncovered a d ifferent LNB value when examined by an engineer. For example, if it is believed that a 10GHzLNB is installed, try altering the L-Band frequency value by 500MHz.
Check the quality of the RF coaxial cable connecting to the EMR. The F connector is a type of coaxial RF
connector commonly used for:
7/29/2019 EMR Configuration Guide_v.7.pdf
16/18
EMR Configuration Guide WNE v.4..2 12
"over the air" terrestrial television
cable television and
universally for satellite television and cable modems, usually with RG-6/U cable or, in older installations, withRG-59/U cable.
Figure 13 RG59 Flexible Coaxial Cable
Further information can be found online regardingtroubleshooting cable issues at:
http://en.wikipedia.org/wiki/Coaxial_cable#Issues
Details Illustrated
A: Outer plastic sheath
B: Copper screen
C: Inner dielectric insulator
D: Copper core
Connection lost after EMR reboot
It is possible that the satellite parameters were not saved to the EMR as a default configuration. Re-enter the L-Bandfrequency details for the EMR, acquire the WNE signal and be sure to save these values as the default configuration,as outlined in Figure 12.
FAQWhat type of LNB should be used on a satellite dish for receiving WNE?
The Reuters recommended specifications for each type of LNB (KU and C Band) are listed below:
KU Band
SMW LNB Single - Swedish Microwave - http://www.smw.se/
The LNB Single has a very Low Phase Noise and is optimized for applications like Broadband, High speed data,VSAT systems and Commercial satellite headends.
Figure 14 A Light Single Ku Band LNB
The LNB has a standard Low Noise figure, F- or N-connector and 2 Year warranty. As option available with Highstability LO, Low gain, Low output VSWR and extended frequency range up to 3000 MHz.
7/29/2019 EMR Configuration Guide_v.7.pdf
17/18
EMR Configuration Guide WNE v.4..2 13
SMW LNB Frequency Range LO Frequency Output Frequency
Single 9.75 10.7 - 11.8 GHz 9.75 GHz 950 - 2050 MHz
Single 10.0 10.95 - 12.1 GHz 10.0 GHz 950 - 2100 MHz
Single 10.05 11.0 - 12.1 GHz 10.05 GHz 950 - 2050 MHz
Single 10.25 11.2 - 11.7 GHz 10.25 GHz 950 - 1450 MHz
Single 10.6 11.7 - 12.75 GHz 10.6 GHz 1100 - 2150 MHz
Single 10.75 11.7 - 12.75 GHz 10.75 GHz 950 - 2000 MHz
Single 11.2 12.2 - 12.75 GHz 11.2 GHz 1000 - 1550 MHz
Single 11.25 12.2 - 12.75 GHz 11.25 GHz 950 - 1500 MHz
Single 11.3 12.25 - 12.75 GHz 11.3 GHz 950 - 1450 MHz
Single 11.475 12.5 - 12.75 GHz 11.475 GHz 1025 - 1275 MHz
Figure 15Specification SMW LNB Single
C Band
C-Band & Ku-Band Digi-Ready LNBs California Amplifier - http://www.calamp.com/
Figure 16 Extended C-Band Professional II LNB
Cost effective PLL local oscillator stability
Part Number: 140194-1
Input Frequency: 3.4 to 4.2 GHz
Output Frequency: 950 to 1750 MHz
Noise Temp. Options: 20 or 25 K
Gain: 60dB Min, 63 dB Typ
Local Oscillator Stability: 125 KHz Max
EMR Technical Specifications
Platform 19-inch rack-mountable chassis
Rack height: 1 RU
Dimensions: 1.75 x 17 x 9.75 inches (4.5 x 43.2 x 24.7 cm)
Weight: 6.5lbs for single tuner (5520S) and 7.5 lbs for dual tuner (5540S) Power Input: 100 to 240 VAC @ 2/1 Amps 50/60 Hz
7/29/2019 EMR Configuration Guide_v.7.pdf
18/18
EMR Configuration Guide WNE v.4..2 14
Two 10/100 Ethernet ports
One RS-232 serial port / console port
Embedded Linux operating system
Satelli te Receiver RF Tuner (single and dual tuner
models) Receiving Frequency: 950 MHz 2150 MHz
Input Signal Level: -25 dBm to -65 dBm
RF Input Impedance: 75 ohm
Optional second tuner card (EMR-5540S)
Demodulation and Error Correction QPSK and BPSK demodulator
1.5 45 Msymbols/sec
Reed-Solomon outer code: (204, 188), T=8
Viberbi inner code: K=7, R=1/2, 2/3, 3/4, 5/6, 7/8
LNB Power and Switching Supply voltage selectable: Off, 13V or 18V
Max LNB Current: 400mA per L-Band Interface
400mA for one tuner in EMR5520
200mA for each tuner in EMR5540
Antenna Control: 22kHz signal
DiSeqC: 1.1 Ready
Over current and short circuit protection
Safety UL60950 3rd Edition
CSA 22.2 No.950
EN60950
IEC-950 CB Report and Certificate
Thermal Operation ANSI T1 304 (7 day Operational at -5 to +50 C
2010 Thomson Reuters. All rights reserved.
Republication or redistribution of Thomson Reuterscontent, including by framing or similar means, isprohibited without the prior written consent of ThomsonReuters. 'Thomson Reuters' and the Thomson Reuterslogo are registered trademarks and trademarks ofThomson Reuters and its aff iliated companies.
For more information visitreuters.com/newsagency