    Application Specifics Electrode Order no.

    Aqueous acid/base titrations

    General Ecotrode Plus 6.0262.100

    Alkaline samples, Bayer liquorsTitrations performed at high temperatures

    Unitrode 6.0259.100

    Acidity of alcoholic beverages Unitrode (with Pt 1000) 6.0258.010

    Titr ations with small sample volumes Flat membrane pH electr ode 6. 0256.100

    Electroplating and etching baths containing fluorideor hydrofluoric acid

    Combined Sb electrode 6.0421.100

    Carbonate hardness, acid capacity of water,p- & m-value

    Aquatrode Plus (with Pt 1000) 6.0257.000

    Electroplating baths Profitrode 6.0255.100

    Titrations in water having a low conductivity Aquatrode Plus 6.0253.100

    Protein-containing samples Porotrode 6.0235.200

    Non-aqueousacid/basetitrations

    Determination of the acid number (TAN) of mineraloils, titrations with alcoholic KOH , NaOH and TBAOH,potassiummethylate, free acids; hydroxyl number ofoils and fats

    Solvotrode (c(TEABr) =0.4 mol/L in ethylene glycol)

    6.0229.100(6.2320.000)

    Determination of the base number (TBN) of mineraloils, titrations with perchloric acid, cyclohexylamine,alcoholic HCl

    Solvotrode (LiClsat

    in ethanol)6.0229.100(6.2312.000)

    RedoxtitrationsArsenite

    Cerium sulfateIron(III)IodinePotassium bromateSodium nitriteOxalic acidPermanganateThiosulfateTitanium(III)Hg(NO

    3)

    2

    Titrations in Ipol

    mode Double Pt-wire electrode 6.0341.100

    Redox titrations at varying pH value Combined LL Pt-ring elec trode 6. 0451.100

    Chemical oxigen demand in water (ferrometry) Combined LL Au-ring electrode 6.0452.100Penicillin, Ampicillin Combined LL Au-ring electrode 6.0452.100

    Titrations at constant pH value Pt Titrode 6.0431.100

    Bromatometry, iodometry, cerimetry according toPharm. Europe & U.S.P.

    Pt Titrode 6.0431.100

    Karl Fischertitrations

    Water content according to Karl FischerDouble Pt-wire electrode(HF-resistant version on request)

    6.0338.100

    Application Specifics Electrode Order no.

    PrecipitationtitrationsSilver nitrate

    Chloride in general, sodium chloride in food Ag Titrode 6.0430.100

    Chloride in dialysi s and in fusion solut ions Ag T it rode wi th Ag2S coating 6.0430.100S

    Titrations according to Pharm. Europe & U.S.P. Ag Titrode with Ag2S coating 6.0430.100S

    Determination of hydrogen sulfide, mercaptans,carbonyl sulfide, sulfides

    Ag Titrode with Ag2S coa ti ng 6 .0 43 0.10 0S

    Chloride, bromide, iodide and cyanide inelectroplating baths

    Ag Titrode with Ag2S coa ti ng 6 .0 43 0.10 0S

    Fluoride/hydrofluoric acid in etching baths Ion-selective electrode F 6.0502.150

    Complexo-metrictitrations withEDTA, ComplexonIII and IV

    Back-titration of the Ba2+e xces s w it h E DTA Io n- se le ct iv e el ec trod e C a2+ 6.0508.110

    Determination of Ca2+, Mg2+ (Application Bulletin 125) Ion-selective electrode Ca2+ 6.0508.110

    Determination of Al, Ba, Bi, Ca, Cd, Co, Fe, Mg, Ni,Pb, Zn (Application Bulletin 101) Ion-selective electrode Cu2+ 6.0502.140

    Photometrictitrations

    With indicators such as Xylenol Orange, N,N-Diethyl-1,4-phenylenediamine, Phenolphthalein, Thorin,Dichlorophenol indophenol

    Spectrosense 523 nm

    6.5501.100 (Titrino)6.5501.200(Titrando, Titrinoplus)

    With indicators such as Dimethylsulfonazo III,Hydroxy Naphthol Blue, Eriochrome Black T, HHSNN,Diphenylaminosulfonate, Murexide

    Spectrosense 610 nm

    6.5501.110 (Titrino)6.5501.210(Titrando, Titrinoplus)

    Surfactants innon-aqueous

    mediaAromatic/aliphatichydrocarbons,ketones,chloroethanes,methylisobutylketone

    Titration of anionic and cationic surfactants, titrationsin chloroform, surfactant formulations containing oils,

    pH 10

    Surfactrode Refill 6.0507.140

    Surfactantsin aqueousmedia

    Titration of cationic surfactants Cationic Surfactant 6.0507.150

    Titration of anionic surfactants Ionic Surfactant 6.0507.120

    Titration of non-ionic surfactants, titrationof pharmaceutical ingredients with sodiumtetraphenylborate

    NIO Electrode 6.0507.010

    Electrodes for titration

    Which electrode for which application?

  • 5/20/2018 Electrodes_for_Titration.pdf

    2/2

    Practical tips, care and maintenanceEcotrode Plus Unitrode Aquatrode Plus Profitrode Solvotrode Porotrode Ag/Pt/Au Titrodes

    Combined Ag/Pt/Auelectrodes

    SurfactrodesSurfactantelectrodes

    Spectrosense

    Fixed ground-jointdiaphragm

    Insensitive tocontamination

    Attractive price/

    performance ratioOrdering info

    6.0262.100

    Fixed ground-jointdiaphragm

    Very low alkali error

    Insensitive tocontamination

    High temperatureresistance

    Ordering info

    6.0259.100without cable

    6.0258.010 (with Pt1000) with fixed cable,plug F+2 x 2 mm

    Fixed ground-jointdiaphragm

    Rapid response in ion-deficient or weaklybuffered solutions

    Insensitive tocontamination

    Ordering info

    6.0253.100without cable

    6.0257.000 (with Pt1000) with fixed cable,plug F+2 x 4 mm

    Ground-jointdiaphragm

    For difficult matrices

    Double-junctionconstruction

    Ordering info

    6.0255.100 (Fittinglength = 113 mm)

    6.0255.110 (Fittinglength = 170 mm)

    6.0255.120 (Fittinglength = 310 mm)

    Ground-jointdiaphragm

    For titrations in non-aqueous matrices

    Rapid response in

    organic solventsElectrically shielded

    Ordering info

    6.0229.100

    Ceramic capillarydiaphragm

    For protein-containingor viscous samples

    Reference electrolyte:

    Porolyte (6.2318.000)Ordering info

    6.0235.200

    Maintenance-free pHglass reference system

    Also available asmicro-electrodes with6.4 mm diameter

    Ordering infoAg Titrode 6.0430.100

    Pt Titrode 6.0431.100

    Micro-Au Titrode6.0435.110

    Ceramic-pindiaphragm

    For precipitation orredox titrations withchange of pH value

    Ordering infoAg: 6.0450.100

    Pt: 6.0451.100

    Au: 6.0452.100

    For surfactanttitrations in non-aqueous media

    SurfactrodeResistant

    (6.0507.130)Resistant tochloroform and manyother solvents

    Surfactrode Refill(6.0507.140)

    Refillable surfactantelectrode, thuspractically unlimitedworking life

    Not resistant tochloroform

    For surfactanttitrations in aqueousmedia

    NIO Electrode(6.0507.010)

    For titration of non-ionic surfactants

    Ionic Surfactant(6.0507.120)

    Optimized for anionicsurfactants

    CationicSurfactant(6.0507.150)

    Optimized for cationicsurfactants

    Two wavelengthsavailable:523 and 610 nm

    Power supplied bytitrator, no separate

    power adapterneeded.

    Ordering info

    Titrino:

    6.5501.100 (523 nm)6.5501.200 (610 nm)

    Titrando, Titrino plus:

    6.5501.110 (523 nm)6.5501.210 (610 nm)

    all cables included

    Store in 6.2323.000storage solution only!

    Do not wipe electrode!

    Use 6.2308.040Idrolyte for titrationsat temperatures80...100 C or in

    samples containingorganic solvents.

    Rinse with water orethanol to removecontamination.

    Do not wipe electrode!

    Unitrodes filled withc(KCl) = 3 mol/Lshould be stored in6.2323.000 storagesolution.

    Store in 6.2323.000storage solution only!

    Do not wipe electrode!

    Store in bridgeelectrolyte.

    If the sleeve ring isblocked, place theelectrode in hot waterand try again.

    Spare ground-jointdiaphragm:

    Order no. 6.1243.020

    Store in referenceelectrolyte

    Condition in dist.water before nexttitration (glassmembrane only!).

    If the sleeve ringis blocked, placeelectrode in hot waterand try again.

    Alternative electrolytes:c(LiCl) = 2 mol/L inethanol (6.2312.010)

    c(TEABr) = 0.4 mol/Lin ethylene glycol(6.2320.000)

    Store in 6.2323.000storage solution.

    Rinse with water orethanol to removecontamination.

    Store in distilled wateronly.

    Testing of theelectrodes accordingto AB48.

    Also available withAg

    2S or Ag-halide

    coating.

    Store in referenceelectrolyte.

    Testing of theelectrodes accordingto AB48.

    Also available withAg

    2S or Ag-halide

    coating.

    Allow a few titrationsfor conditioning of theSurfactrodes.

    Reactivate SurfactrodeResistant with fine-grained sandpaper ifresponse is poor.

    Testing of theelectrodes accordingto AB305.

    Rinse with dist. wateror 20% methanol inwater.

    Wipe carefully witha methanol-soakedtissue to removeany adheringcontamination.

    Do not use in organicmatrices or at tem-peratures >40 C!

    Testing of theelectrodes accordingto AB305.

    Allow three minutesfor warming up theLED.

    Clean mirror withsoft cloth only; do nottouch with fingers.

    8.0

    00.6

    003EN