Calibration and Measurement Capabilities
Electricity and Magnetism, Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrix
NMI Service
IdentifierComments
DC voltage sources:
single values
Solid state voltage
standard
Direct comparison
with a bank of solid
state standards
10 10 V 1 µV/V 2 95% Yes 1Approved on
11 May 2017
DC voltage sources:
low values
DC voltage source,
multifunction
calibrator
Comparison with
reference digital
multimeter
0.01 10 V 4 to 17 µV/V 2 95% Yes DCV_M1.1.2 2Approved on
11 May 2017
DC voltage sources:
intermediate values
DC voltage source,
multifunction
calibrator
Comparison with
reference digital
multimeter
10 1000 V 4 to 8.5 µV/V 2 95% Yes DCV_M1.1.3 3Approved on
11 May 2017
DC voltage meters:
intermediate values
DC voltmeter,
multimeter
Comparison with
reference digital
multimeter
0.01 1000 V 4 to 17 µV/V 2 95% Yes DCV_M1.2.2-1 4Approved on
11 May 2017
DC voltage meters:
intermediate values
DC voltmeter,
multimeter
Direct comparison
with multifunction
calibrator
0.01 1000 V 5 to 45.5 µV/V 2 95% Yes DCV_M1.2.2-2 4Approved on
11 May 2017
DC resistance
standards and
sources:
intermediate values
Fixed resistorDC current
comparator0.1 1 Ω 0.65 µΩ/Ω 2 95% Yes DCR_M2.1.1
Approved on
11 May 2017
DC resistance
standards and
sources:
intermediate values
Fixed resistorDC current
comparator1 1.00E+05 Ω 0.65 µΩ/Ω 2 95% Yes DCR_M2.1.2
Approved on
11 May 2017
DC resistance
standards and
sources: Multiple
ranges
Multifunction
calibrator
Comparison with
reference digital
multimeter
1 1.00E+08 Ω 8 to 135 µΩ/Ω 2 95% Yes DCR_M2.1.5Approved on
11 May 2017
DC resistance
meters: intermediate
values
Ohmeter, multimeterComparison to
standard resistor1 1.00E+09 Ω 1 to 19 µΩ/Ω 2 95% Yes DCR_M2.2.2
Approved on
11 May 2017
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
The BIPM key comparison database, June 2017 1/31
Calibration and Measurement Capabilities
Electricity and Magnetism, Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrix
NMI Service
IdentifierComments
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
DC resistance
meters: intermediate
values
Ohmeter, multimeter
Direct comparison
with reference
digital multimeter
1 1 MΩ 5.2 µΩ/Ω 2 95% YesApproved on
11 May 2017
DC current sources:
low values
DC current generator,
multifunction
calibrator
Drop voltage
across the
standard resistor
1 100 μA 6.5 to 20 μA/A 2 95% Yes DCI_M3.1.1-1Approved on
11 May 2017
DC current sources:
low values
DC current generator,
multifunction
calibrator
Direct comparison
with reference
digital multimeter
10 100 μA 13.5 to 50 μA/A 2 95% Yes DCI_M3.1.1-2Approved on
11 May 2017
DC current sources:
intermediate values
DC current generator,
multifunction
calibrator
Drop voltage
across the
standard resistor
0.0001 0.2 A 6.5 μA/A 2 95% Yes DCI_M3.1.2-1Approved on
11 May 2017
DC current sources:
intermediate values
DC current generator,
multifunction
calibrator
Direct comparison
with reference
digital multimeter
0.0001 1.E+01 A 11 to 620 μA/A 2 95% Yes DCI_M3.1.2-2Approved on
11 May 2017
DC current meters:
low values
DC current meter,
multimeter
Drop voltage
across the
standard resistor
1 100 μA 6.5 to 25 μA/A 2 95% Yes DCI_M3.2.1-1Approved on
11 May 2017
DC current meters:
low values
DC current meter,
multimeter
Direct comparison
with multifunction
calibrator
10 100 μA 41 to 190 μA/A 2 95% Yes DCI_M3.2.1-2Approved on
11 May 2017
DC current meters:
low values
DC current meter,
multimeter
Direct comparison
with reference
digital multimeter
10 100 μA 13.5 to 50 μA/A 2 95% Yes DCI_M3.2.1-3Approved on
11 May 2017
The BIPM key comparison database, June 2017 2/31
Calibration and Measurement Capabilities
Electricity and Magnetism, Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrix
NMI Service
IdentifierComments
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
DC current meters:
intermediate values
DC current meter,
multimeter
Drop voltage
across the
standard resistor
0.1 200 mA 6.5 μA/A 2 95% Yes DCI_M3.2.2-1Approved on
11 May 2017
DC current meters:
intermediate values
DC current meter,
multimeter
Direct comparison
with multifunction
calibrator
0.0001 10 A 27 to 625 μA/A 2 95% Yes DCI_M3.2.2-2Approved on
11 May 2017
DC current meters:
intermediate values
DC current meter,
multimeter
Direct comparison
with reference
digital multimeter
0.0001 10 A 11 to 620 μA/A 2 95% Yes DCI_M3.2.2-3Approved on
11 May 2017
AC voltage up to
1100 V: sources
Multifunction
calibrator
Direct comparison
with reference
multimeter
0.01 1000 V Frequency 50 Hz to 1 MHz 0.08 to 31 mV/V 2 95% Yes ACV_M5.2.1Approved on
11 May 2017
AC voltage up to
1100 V: meters
AC voltmeter,
multimeter
Direct comparison
with reference
multimeter
0.01 1000 V Frequency 50 Hz to 1 MHz 0.08 to 31 mV/V 2 95% Yes ACV_M5.2.2-1Approved on
11 May 2017
AC voltage up to
1100 V: meters
AC voltmeter,
multimeter
Direct comparison
with multifunction
calibrator
0.01 1000 V Frequency 50 Hz to 1 MHz 0.15 to 31 mV/V 2 95% Yes ACV_M5.2.2-2Approved on
11 May 2017
AC current up to
100 A: sources
AC current generator,
multifunction
calibrator
Direct comparison
with reference
multimeter
1.00E-04 10 A Frequency 50 Hz to 10 kHz 0.36 to 3.5 mA/A 2 95% Yes ACI_M6.2.1Approved on
11 May 2017
AC current up to
100 A: meters
AC ammeter,
multimeter
Drop voltage
across the
standard current
shunt
10 10 mA Frequency 300 Hz, 1 kHz 90 μA/A 2 95% YesApproved on
11 May 2017
AC current up to
100 A: meters
AC ammeter,
multimeter
Direct comparison
with a reference
standard
1.00E-04 10 A Frequency 50 Hz to 10 kHz 0.36 to 3.5 mA/A 2 95% Yes ACI_M6.2.2 15Approved on
11 May 2017
The BIPM key comparison database, June 2017 3/31
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCV_M1.1.2
Expanded uncertainty (µV/V)
10 mV 17
20 mV 10
50 mV 6.5
100 mV 4.5
190 mV 4.5
0.2 V 4.5
0.5 V 4.5
1 V 4
1.9 V 4
5 V 4
10 V 4
The BIPM key comparison database, June 2017 4/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCV_DCV_M1.1.3
Expanded uncertainty (µV/V)
15 V 4
19 V 4
20 V 6.5
50 V 5.5
100 V 5.5
190 V 6.5
200 V 8.5
500 V 6.5
900 V 5.5
1000 V 5.5
The BIPM key comparison database, June 2017 5/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCV_M1.2.2-1
Expanded uncertainty (µV/V)
10 mV 17
20 mV 10
50 mV 6.5
100 mV 4.5
190 mV 4.5
0.2 V 4.5
0.5 V 4.5
1 V 4
1.9 V 4
5 V 4
10 V 4
15 V 4
19 V 4
20 V 6.5
50 V 5.5
100 V 5.5
190 V 6.5
200 V 8.5
500 V 6.5
900 V 5.5
1000 V 5.5
The BIPM key comparison database, June 2017 6/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCV_M1.2.2-2
Expanded uncertainty (µV/V)
10 mV 45.5
20 mV 25
50 mV 12
100 mV 8
190 mV 6.5
0.2 V 6.5
0.5 V 6
1 V 5
1.9 V 5
5 V 5
10 V 5
15 V 5
19 V 5
20 V 7
50 V 6.5
100 V 6.5
190 V 7
200 V 9
500 V 7.5
900 V 6.5
1000 V 6.5
The BIPM key comparison database, June 2017 7/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCR_M2.1.1
Expanded uncertainty (µΩ/Ω)
0.1 Ω to 1 Ω 0.65
The BIPM key comparison database, June 2017 8/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCR_M2.1.2
Value Expanded uncertainty (µΩ/Ω)
1 Ω to 10 Ω 0.65
10 Ω to 10 kΩ 0.65
10 kΩ to 100 kΩ 0.65
The BIPM key comparison database, June 2017 9/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCR_M2.1.5
Expanded uncertainty (µΩ/Ω)
1 Ω 16.0
10 Ω 9.0
100 Ω 8.0
1 kΩ 8.0
10 kΩ 8.0
100 kΩ 8.0
1 MΩ 8.5
10 MΩ 20
100 MΩ 135
The BIPM key comparison database, June 2017 10/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCR_Μ2.2.2
Expanded uncertainty (µΩ/Ω)
1 Ω 1
1.9 Ω 1
10 Ω 1
100 Ω 1
1 kΩ 1
10 kΩ 1
19 kΩ 1
100 kΩ 1
1 MΩ 4
10 MΩ 4
100 MΩ 6.50
1 GΩ 19
The BIPM key comparison database, June 2017 11/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.1.1-1
Expanded uncertainty (µA/A)
1 μA 20
2 μA 15
5 μA 9
10 μA 7
20 μA 6.5
50 μA 6.5
100 μA 6.5
The BIPM key comparison database, June 2017 12/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.1.1-2
Expanded uncertainty (µA/A)
10 μA 50
20 μA 29
50 μA 18.5
100 μA 13.5
The BIPM key comparison database, June 2017 13/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.1.2-1
Expanded uncertainty (µA/A)
200 μA 6.5
500 μA 6.5
1 mA 6.5
2 mA 6.5
5 mA 6.5
10 mA 6.5
20 mA 6.5
50 mA 6.5
100 mA 6.5
200 mA 6.5
The BIPM key comparison database, June 2017 14/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.1.2-2
Value Expanded uncertainty (µA/A)
190 μA 12
0.2 mA 27.5
0.5 mA 16
1 mA 12.5
1.9 mA 11
2 mA 29
5 mA 17
10 mA 13.5
19 mA 12
20 mA 75
50 mA 50
100 mA 45
190 mA 40
0.2 A 330
0.5 A 275
1 A 210
1.9 A 200
2 A 620
5 A 500
10 A 430
The BIPM key comparison database, June 2017 15/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.2.1-1
Value Expanded uncertainty (µA/A)
1 μA 25
2 μA 15
5 μA 9
10 μA 7
20 μA 6.5
50 μA 6.5
100 μA 6.5
The BIPM key comparison database, June 2017 16/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.2.1-2
Value Eexpanded uncertainty (µA/A)
10 μA 190
20 μA 110
50 μA 58
100 μA 41
The BIPM key comparison database, June 2017 17/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.2.1-3
Value Expanded uncertainty (µA/A)
10 μA 50
20 μA 29.5
50 μA 18.5
100 μA 13.5
The BIPM key comparison database, June 2017 18/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.2.2-1
Value Expanded uncertainty (µA/A)
200 μA 6.5
500 μA 6.5
1 mA 6.5
2 mA 6.5
5 mA 6.5
10 mA 6.5
20 mA 6.5
50 mA 6.5
100 mA 6.5
200 mA 6.5
The BIPM key comparison database, June 2017 19/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.2.2-2
Value Expanded uncertainty (µA/A)
190 μA 33
0.2 mA 41
0.5 mA 35
1 mA 29
1.9 mA 27
2 mA 38
5 mA 35
10 mA 29
19 mA 27
20 mA 79
50 mA 60
100 mA 53
190 mA 48
0.2 A 335
0.5 A 280
1 A 220
1.9 A 205
2 A 625
5 A 520
10 A 450
The BIPM key comparison database, June 2017 20/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix DCI_M3.2.2-3
Value Expanded uncertainty (µA/A)
190 μA 12
0.2 mA 27.5
0.5 mA 16
1 mA 12.5
1.9 mA 11
2 mA 29
5 mA 17
10 mA 13.5
19 mA 12
20 mA 75
50 mA 50
100 mA 45
190 mA 40
0.2 A 330
0.5 A 275
1 A 210
1.9 A 200
2 A 620
5 A 500
10 A 430
The BIPM key comparison database, June 2017 21/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix ACV_M5.2.1
50 Hz 1 kHz 20 kHz 50 kHz 100 kHz 300 kHz 1 MHz
10 mV 0.72 0.50 1.20 3.00 3.00 - -
20 mV 0.40 0.30 0.75 1.80 1.80 - -
100 mV 0.16 0.15 0.40 1.00 1.00 - -
190 mV 0.15 0.15 0.35 0.85 0.85 - -
1 V 0.11 0.09 0.25 0.72 0.72 - -
1.9 V 0.10 0.08 0.25 0.62 0.62 - -
10 V 0.11 0.09 0.25 0.72 0.72 5.10 31.00
19 V 0.10 0.08 0.25 0.62 0.62 4.10 21.00
100 V 0.11 0.10 0.25 0.72 0.72 - -
190 V 0.10 0.08 0.25 0.62 0.62 - -
500 V 0.13 0.13 0.30 - - - -
700 V 0.13 0.13 0.35 - - - -
1000 V 0.11 0.11 0.29 - - - -
The expanded uncertainties are expressed in mV/V
The BIPM key comparison database, June 2017 22/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix ACV_M5.2.2-1
50 Hz 1 kHz 20 kHz 50 kHz 100 kHz 300 kHz 1 MHz
10 mV 0.72 0.51 1.20 2.85 3.00 - -
20 mV 0.40 0.30 0.75 1.80 1.80 - -
100 mV 0.16 0.15 0.40 1.00 1.00 - -
190 mV 0.15 0.15 0.35 0.85 0.85 - -
1 V 0.11 0.10 0.25 0.72 0.72 - -
1.9 V 0.10 0.08 0.25 0.62 0.62 - -
10 V 0.11 0.09 0.25 0.72 0.72 5.10 31.00
19 V 0.10 0.08 0.25 0.62 0.62 4.10 21.00
100 V 0.11 0.10 0.25 0.72 0.72 - -
190 V 0.10 0.08 0.25 0.62 0.62 - -
500 V 0.13 0.13 0.29 - - - -
700 V 0.13 0.13 0.35 - - - -
1000 V 0.11 0.11 0.29 - - - -
The expanded uncertainties are expressed in mV/V
The BIPM key comparison database, June 2017 23/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix ACV_M5.2.2-2
50 Hz 1 kHz 20 kHz 50 kHz 100 kHz 300 kHz 1 MHz
10 mV 0.90 0.70 1.30 3.00 3.20 - -
20 mV 0.50 0.50 0.85 2.00 2.00 - -
100 mV 0.25 0.25 0.50 1.10 1.20 - -
190 mV 0.20 0.20 0.40 0.90 1.10 - -
1 V 0.15 0.15 0.30 0.80 0.80 - -
1.9 V 0.15 0.15 0.30 0.70 0.70 - -
10 V 0.15 0.15 0.30 0.80 0.80 5.50 31.00
19 V 0.15 0.15 0.30 0.70 0.70 4.50 21.00
100 V 0.15 0.15 0.30 0.75 0.75 - -
190 V 0.15 0.15 0.30 0.70 0.70 - -
500 V 0.20 0.20 0.40 - - - -
700 V 0.20 0.20 0.40 - - - -
1000 V 0.15 0.15 0.40 - - - -
The expanded uncertainties are expressed in mV/V
The BIPM key comparison database, June 2017 24/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix ACI_M6.2.1
Value 50 Hz 1 kHz 5 kHz 10 kHz
100 μA 0.50 0.50 0.50 0.55
190 μA 0.45 0.45 0.45 0.45
0.2 mA 1.30 1.30 1.30 1.30
1 mA 0.45 0.45 0.45 0.45
1.9 mA 0.36 0.36 0.36 0.36
2 mA 1.30 1.30 1.30 1.30
10 mA 0.45 0.45 0.45 0.45
19 mA 0.36 0.36 0.36 0.36
20 mA 1.30 1.30 1.30 1.30
100 mA 0.45 0.45 0.45 0.45
190 mA 0.36 0.36 0.36 0.36
0.2 A 1.60 1.60 1.70 1.70
1 A 0.80 0.80 0.90 0.90
1.9 A 0.71 0.71 0.82 0.82
2 A 1.80 1.8 3.50 3.50
10 A 1.00 1.00 2.70 2.70
The expanded uncertain The expanded uncertainties in the table are expressed in mA/A
The BIPM key comparison database, June 2017 25/26
Calibration and Measurement Capabilities
Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)
Matrix ACI_M6.2.2
50 Hz 1 kHz 5 kHz 10 kHz
100 μA 0.50 0.50 0.50 0.55
190 μA 0.45 0.45 0.45 0.45
0.2 mA 1.30 1.30 1.30 1.30
1 mA 0.45 0.45 0.45 0.45
1.9 mA 0.36 0.36 0.36 0.36
2 mA 1.30 1.30 1.30 1.30
10 mA 0.45 0.45 0.45 0.45
19 mA 0.36 0.36 0.36 0.36
20 mA 1.30 1.30 1.30 1.30
100 mA 0.45 0.45 0.45 0.45
190 mA 0.36 0.36 0.36 0.36
0.2 A 1.60 1.60 1.70 1.70
1 A 0.80 0.80 0.90 0.90
1.9 A 0.71 0.71 0.82 0.82
2 A 1.80 1.80 3.50 3.50
10 A 1.00 1.00 2.70 2.70
The expanded uncertain The expanded uncertainties in the table are expressed in mA/A