Upload
amberlynn-george
View
217
Download
0
Tags:
Embed Size (px)
Citation preview
EET286 – Fall 2015 follow-up 2015-12-10
• Four types of clinical equipment• Power Connectors• NFPA99 Chassis Leakage (touch) Current• Human Heart & Cardiac Cycle• Wiggers diagram• Pressure Conversions• Common Electronic Component Abbreviations• Quiz #2 recap• Quiz #3 recap
Four types of clinical equipment
Monitoring:Detects and displays physiologic data for the purposes of clinical observation.
Diagnostic:Detects, displays, and records physiologic data in order to confirm or rule out
a disorder or disease.
Therapeutic:Applies treatment in the form of energy or medication in order to treat a
disorder or disease.
Assistive:Used as an accessory to facilitate a medical or surgical procedure.
Power Connectors
plug: male connector, furthest away from power source
outlet or receptacle: female connector providing power
NEMA: (U.S.) National Electrical Manufacturers Association
IEC: International Electrotechnical Commission
NEMA wire colors:Hot – Black Neutral – White Ground – Green
IEC wire colors:Hot – Brown Neutral – Blue Ground Green/Yellow
Screw colors on some connectors:Hot – Gold Neutral – Silver Ground – Green
HOT (L1) Neutral (L2) Ground (earth)
NEMA colors: Black White Green
IEC colors: Brown Blue Green/Yellow
connectors: Gold Silver Green
NFPA99 Chassis Leakage (touch) Current• 1999 edition • 2005 edition
• 2012 edition
• 2015 edition 10.2.6* Touch Current – Portable Equipment. The touch current for cord- connected equipment shall not exceed 500 uA with normal polarity and
the ground wire disconnected (if a ground wire is provided).
PSI KiloPascal cm of H2O mm of Hg atmosphere millibar
1 PSI = 1 6.89 70.3 51.7 0.068 68.9
1 KiloPascal = 0.145 1 10.19 7.5 0.0098 10
1 cm of H2O = 0.014 0.1 1 0.735 0.00097 1
1 mm of Hg = 0.019 0.133 1.36 1 0.0013 1.33
1 atmosphere = 14.7 101 1033 760 1 1013
1 millibar = 0.0147 0.1 1.02 0.75 0.0009 1
Pressure Conversions
Most useful conversions in bold
Manometers, two styles;Pressure measured as a column of mercury:
• closed end (contains vacuum)• measures atmospheric pressure• 760 mmHg ‘normal’• higher atmospheric pressure = higher number• (pressure in outer space vacuum would be 0 mmHg)
• open end (contains ambient air pressure)• measures applied pressure• 0 mmHg is ‘unpressurized’ state• higher applied pressure = higher number
ECG trace and display
Calibration (y axis):10 mm = 1 mV
Speed (x axis):25mm = 1 Second
|---- 25 mm (1 Second) ----------------------------------------------------------------------------------|
|---
- 10
mm
(1 m
illiV
olt)
----
----
---|
Pacemaker Modes:
Asynchronous: provides pulses at a fixed rate
Demand (or ‘Synchronous’): provides pulses when the heart fails to beat
Rate-Modulated: provides pulses based on the heart’s current activity
Defibs for Cardioversion
•The use of a small energy pulse from a defibrillator to convert an non-lethal arrhythmia (eg atrial flutter) to a normal sinus rhythm
•Delivers a pulse about 30ms after the peak of the R-wave, synchronized by the defibrillator
•Modern units can discriminate the R-wave from the T-wave or noise spikes
Timing of cardioversion pulse
Do NOT cardiovert across the T-wave !
1) Line in (not shown here)2) Transform (only the output is shown here)3) Rectify (usually a 4-diode bridge)4) Filter (usually a big cap & a small cap)5) Regulate (often with 78xx & 79xx ICs on heat sinks)
Linear power supply(2) (3) (4) (5)
from Electrical Safety presentation (class 6) CORRECTED:(per NFPA-99, 2012 edition)
Resistance of Ground wire of power cord: 500 milliOhms (500mΩ or .5Ω)
Leakage current at the chassis: NC (normal condition) 100 microAmps (100 μA)SFC (Single Fault Condition, e.g. open ground) 500 microAmps (500 μA)
Leakage current for any one patient contact lead to ground: 10 μA with ground closed, 50 μA with ground open.
Leakage current for each patient contact lead to any other lead: 10 μA with ground closed, 50 μA with ground open.
Leakage current for patient leads exposed to line voltage: 50 μA
or “distributed capacitance”
Leakage current:
Leakage current is both capacitive, caused by intrinsiccapacitance between conductors; and resistive, causedby imperfect insulation.
Where does the leakage current go?
Open ground !
Where does the leakage current go now?
Test device measuringchassis leakage current:
“Applied Part” of medical equipmentas defined by the IEC standard 60601-1:
Type B: Applied parts that are generally not conductive and can be immediately released from the patient. May be grounded.
eg: Non-invasive BP monitors [think: Body]
Type BF: Devices that have direct contact with the patient, or parts that have long term contact with the patient.
eg: ECG monitors [think: Body, Floating (ground)]
Type CF: Applied parts that have direct contact with the heart.eg: Invasive pressure monitors, defib paddles
[think: Cardiac Floating (ground)]
ONE of these ratings should be present on a medical device
If ‘Defib Proof’, ONE of these
Almostideal
Line is partially occluded
Air bubblesin the line
Troubleshooting invasive BP lines:
Damping in any oscillating system:
Damping is an effect on any oscillating system that reduces or restricts its oscillations. In a mechanical system, damping is produced by drag in the system. In a fluid system, damping is caused by viscosity. In an electronic system, damping is caused by impedance.
Undamped: The system oscillates at its natural resonant frequency.
Critically damped: When disturbed, the system returns to equilibrium quickly, without oscillation.
Overdamped: When disturbed, the system returns to equilibrium without oscillating.
Underdamped: When disturbed, the system oscillates, but the amplitude decreases to zero.
Swan-Ganz catheter (in ‘right’ heart)Starts here Ends up ‘wedged’ here or here
Swan-Ganz Catheter – From the Right Atrium through the Right Ventricle, into the Pulmonary Artery
Pressure waveforms during insertion ofthe Swan-Ganz catheter:
Most used:
Manual determination of NIBP blood pressure :
NIBP Blood pressure determination by oscillometryi.e. the automated electronic method:
Pressure pulsations during oscillometry:
non-invasive blood pressure monitor
Quiz #3 recap
LR#1: Intra-aortic balloon pump (IABP) is inserted into descending AORTA (output of LEFT heart)
#3: IABP is typically used for 1-2 days
#6: Capnography is measurement of CO2.
Measured by how much IR (infrared) light is absorbed by the gas sample.
(to lung)
(from lung) (from lung)
(from body)
(to lung)
(from body)(TO body)
Question #7:
Deep breath(or sigh)
normal breathing
exhaling as much as you can
this much air always stays in your lungs
note: TLC is often shown as 6 liters, which is normal for a 75 kg (165 #) person.80 ml per kg of body weight (as shown here) is more accurate.
Question #14: Flow-Volume Loop (time is shown as a counter-clockwise loop):
PEFR: Peak Expiratory Flow Rate:measures how fast a person can exhale (breathe out).PIFR: Peak Inspiratory Flow Rate FVC: Functional Vital Capacity (same as VC)
Pressure-Volume-FlowDiagrams
(X axis for these is time)
inhale
exhale
note
: Res
idua
l Vol
ume
is s
how
n on
the
right
sid
e(s
till p
art o
f Tot
al L
ung
Capa
city
– 6
lite
rs)
Respiratory Gasses • Inspired (inhaled) air: N2 79% O2 21%
• Expired (exhaled) air: N2 79% O2 16%CO2 5%
Blood Components
• Hemoglobin (hb without attached O2)
• Oxyhemoglobin (hb with attached O2)• Carboxyhemoglobin (hb with attached CO, Carbon Monoxide) gives inaccurate
SpO2
• Methemoglobin (pron: MET-hemoglobin) cannot bind to O2 and gives inaccurate SpO2
this ratio gives SpO2 in Pulse Oximetry