Upload
jubin-jain
View
3
Download
0
Embed Size (px)
DESCRIPTION
jj
Citation preview
1
Biomedical InstrumentationWebster Chapter -1
Introduction
Medical InstrumentInvasive, noninvasive; external, implantedDiagnostic, therapeuticdetect biochemical, bioelectrical, or biophysical parametersreproduce the physiologic time response of these parametersprovide a safe interface with biological materials
ExamplesEndocardial catheter; ECG electrodes and monitorExternal & implanted pacemakers, defibrillatorsEEG based neurological monitor; deep brain stimulator for Parkinson’s disease
2
General Medical Instrumentation System Control
Andfeedback
SensorPowersource
PerceptibleoutputOutput
displaySignalprocessing
Datatransmission
Datastorage
VariableConversionelement
Sensor
PrimarySensingelement
Measurand
Calibrationsignal
source
Figure 1.1 Generalized instrumentation system The sensor converts energy or information from the measurand to another form (usually electric). This signal is the processed and displayed so that humans can perceive the information. Elements and connections shown by dashed lines are optional for some applications.
Radiation,electric current,or other appliedenergy
Characteristics of Physiological Signals
Parameter Measurement Range
Frequency Range
Measurement Method
Electrocardiogram (ECG)
0.5 – 4mV 0.01 – 250Hz Skin electrodes
Electroencephalo-gram (EEG)
5 – 300μV DC – 150Hz Scalp electrodes
Electromyogram (EMG)
0.1 – 5mV DC – 10KHz Needle/skin electrodes
Electrooculogram (EOG)
50 – 3500μV DC – 50Hz Contact electrodes( )Blood flow 1 – 300mL/sec DC – 20Hz Ultrasonic flowmeter
Respiratory Rate 2 – 50 breaths/min 0.1 – 10Hz Strain guage
Body Temperature 32 – 40 oC DC – 0.1Hz Thermistor/thermocouple
3
Examples of Medical InstrumentsImplanted pacemaker
External ECG monitor
nerve stimulatornerve stimulator
Laser for coronary angioplasty
X ray or imaging devices
Electrosurgical instrument
Pulse oximeter
Defibrillation electrode
Infrared thermometer
Automatic blood pressure
System Block DiagramInterference“Induced”
Transfer Function
Signal source
InputOutput
Internal interference
(added)
Any measurement includes signal+noiseSignal sources: ECG. EEG, blood
InstrumentSystem
(added) Signal sources: ECG. EEG, blood pressure, temperature…Noise sources
External: 60 Hz, radio frequency (RF), magnetic…Internal: muscle noise, motion artifact, eye blink artifact
4
General instrument static characteristics of• Accuracy (True (from NIST)-measured)
• Precision (No of significant digits)
Static System Properties
digits)
•Resolution (smallest measurable qty)
•Reproducibility (give same output)
•Statistical control (variation of meas red q antities in tolerablemeasured quantities in tolerable limits)
•Static sensitivity – slope, zero-drift (ambient surroundings), sensitivity drift (ex power supply fluctuations)
More next slide
Static system properties cont.Linearity (not all instruments have a perfect linear response; deviation from linear fit line is necessary nonlinearityInput Dynamic Range ratio Linearx1 y1
LINEARITY
Input Dynamic Range - ratio between the maximum undistorted signal (i.e., maximum input signal satisfying the linearity specification for the sensor) and the minimum detectable signal for a given set of operating conditions expressed in dBInput Impedance the
System
LinearSystem
x2 y2
LinearSystem
x1 + x2 y1 + y2
Input Impedance the instantaneous rate at which energy is transferred by a system
System
LinearSystem
Kx1 Ky1
5
Generalized Dynamic CharacteristicsMost biomedical instruments must process signals that change with time. The dynamics of the measurement system, therefore, must be chosen to properly reproduce the dynamics of the physiologic variables the system is sensing. Mostly, we will deal with consider linear, time invariant systems unless otherwise explicitly noted. For such systems the dynamics can be fully described by simple differentialFor such systems, the dynamics can be fully described by simple differential equations of the form:
where x(t) is the input signal (usually the physiologic parameter of interest), y(t) is the output signal (usually the electronic signal), and the a’s and b’s are constants determined by the physical characteristics of the sensor system.
Most practical sensor front-ends are described by differential equations of zero, first or second order (i e n=0 1 2) and derivatives of the input are usually absent so
)(...)(... 0101 txbdtdxb
dtxdbtya
dtdya
dtyda m
m
mn
n
n +++=+++
or second order (i.e., n 0,1,2), and derivatives of the input are usually absent, so m=0.Linear, time invariant systems are simply characterized by their response to sinusoidal inputs of the form x(t) = A sin(wt), where the output is a sinusoid at precisely the same frequency of the form y(t) = B(w) sin(wt + f(w)).
01
01
)(...)()(...)(
)()((
ajajabjbjb
jXjYjH n
n
mm
++++++
==)ωωωω
ωωω
Zero-Order SystemExpression of the input-output relationship
Time-domain R l ti hiRelationship
Transfer Function
Example
)()( 00 txbtya =
0
0)(abjH =ω
Linear potentiometer
6
First-Order SystemSystem contains a single energy-storage elementTime-domain relationship
)()( txbtyadya =+
Transfer Function
)()( 001 txbtyadt
a =+
01
0
)()(
ajabjH+
=ω
ωor
ExampleRC Low-pass or High-pass Filters
RC circuit and response
C
+
−
+
−
y(t)
Output y(t)
Input x(t)
Slope = K = 1x(t)
R
Figure 1.6 (a) A low-pass RC filter, an example of a first-order instrument. (b) Static sensitivity for constant inputs. (c) Step response for larger time constants (τL) and small time constants (τS). (d) Sinusoidal frequency response for large and small time constants.
t
1
(c)
(a)p ( )
(b)
Y (jω)X (jω)
Logscale
1.00.707
Log scale ω(d)
ωSωL
φ
τL
τS
x(t)
y(t)
0°
− 45°
−90°
Log scale ω
t
1
0.63
τLτS
τL
τS
φy(t)
7
Second-Order SystemSecond-order system can approximate higher-order systemsTime-domain Relationship
)()(2
txbtyadyayda =++
Transfer Function
)()( 00122 txbtyadt
adt
a =++
12 20
1 ==aa
aξ12 20
1 >=aa
aξ 12 20
1 <=aa
aξ
Overdamped Underdamped Critically Damped
8
Food and Drug Administration (FDA)
Government body entrusted with the responsibility to l t d di l d i d tregulated medical devices, drugs, etc.
Primary task: certify safety and efficacy
FDA regulates through FDA Instrumentation Categories
Design Control Class I
Process Control Class II
Good Manufacturing Practices Class III
FDA Device Regulations
Class I – General ControlsRequired to perform registration labeling and goodRequired to perform registration, labeling, and good manufacturing practices and to report adverse effects
Class II – Performance StandardsRequired to prove “substantial equivalence” via the 510(k) process
Class III – Pre-market Approval (PMA)Requires extensive testing and expert scrutinyPMA is necessary for devices used in supporting or sustaining human life