1 EDGES MEMO #082 MASSACHUSETTS INSTITUTE OF TECHNOLOGY HAYSTACK OBSERVATORY WESTFORD, MASSACHUSETTS 01886 December 21, 2011 Telephone: 781-981-5407 Fax: 781-981-0590 To: EDGES Group From: Alan E.E. Rogers Subject: A laboratory test of the absolute calibration using an antenna simulator 1] Introduction: EDGES Memo #80 described a method of absolute calibration (more details are in a paper “Absolute calibration of a wideband antenna and spectrometer for sky noise spectral index measurements by Rogers and Bowman recently submitted to Radio Science). Memo 81 outlines some initial tests of the method. A more complete test has now been made using a “hot filament” source which includes a balun. 2] Antenna simulator Figure 1 shows the circuit diagram of the antenna simulator and Figure 2 is a photograph of the unit with the cover removed. The basic concept behind the “antenna simulator” is that the hot filament emulates the sky noise temperature presented to the EGES 3-position switched receiver via various stray reactance to which is added the loss and ambient thermal noise from the balun. The incandescent lamp provides a poorly matched source of noise at a temperature over 1000 K and should be spectrally flat once the mismatch and balun are taken into account although these remain an issue concerning the effects of the skin effect producing a frequency dependent weighting of the variations of temperature along the filament. The use of the tungsten filament as a hot noise source can also be used to check or set the temperature scale since its temperature can be measured via the change in resistance. For the measured 8.26±0.1 fold increase in resistance the temperature is 1670±30 K from tables of tungsten resistance, assuming 294 K for the ambient room temperature. 3] Initial results Figure 3 shows the antenna simulator reflection coefficient measurements and polynomial fits. While measurements could be made at more closely spaced frequencies a least squares fit using a smooth curve is needed to remove noise and allow interpolation to the precise frequencies of the spectrometer channels. These show that in general a polynomial can be fit with fewer free parameters than a Fourier series. Figure 4 is the measured LNA reflection coefficient rms and Figure 5 shows the spectrum of the noise wave calibration cable. Figure 6 shows the partially calibrated spectrum prior correction for the balun. Figure 7 shows the observed spectrum prior to correction for the mismatch and balun along with the corrected spectrum and the best fit constant. The rms deviation from a flat spectrum is 1 K and the constant is 1666±0.5 K.

EDGES M #082 MASSACHUSETTS INSTITUTE OF …

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

1

EDGES MEMO #082 MASSACHUSETTS INSTITUTE OF TECHNOLOGY

HAYSTACK OBSERVATORY WESTFORD, MASSACHUSETTS 01886

December 21, 2011 Telephone: 781-981-5407 Fax: 781-981-0590 To: EDGES Group From: Alan E.E. Rogers Subject: A laboratory test of the absolute calibration using an antenna simulator 1] Introduction:

EDGES Memo #80 described a method of absolute calibration (more details are in a paper “Absolute calibration of a wideband antenna and spectrometer for sky noise spectral index measurements by Rogers and Bowman recently submitted to Radio Science). Memo 81 outlines some initial tests of the method. A more complete test has now been made using a “hot filament” source which includes a balun.

2] Antenna simulator Figure 1 shows the circuit diagram of the antenna simulator and Figure 2 is a photograph of the unit with the cover removed. The basic concept behind the “antenna simulator” is that the hot filament emulates the sky noise temperature presented to the EGES 3-position switched receiver via various stray reactance to which is added the loss and ambient thermal noise from the balun. The incandescent lamp provides a poorly matched source of noise at a temperature over 1000 K and should be spectrally flat once the mismatch and balun are taken into account although these remain an issue concerning the effects of the skin effect producing a frequency dependent weighting of the variations of temperature along the filament. The use of the tungsten filament as a hot noise source can also be used to check or set the temperature scale since its temperature can be measured via the change in resistance. For the measured 8.26±0.1 fold increase in resistance the temperature is 1670±30 K from tables of tungsten resistance, assuming 294 K for the ambient room temperature.

3] Initial results Figure 3 shows the antenna simulator reflection coefficient measurements and polynomial fits. While measurements could be made at more closely spaced frequencies a least squares fit using a smooth curve is needed to remove noise and allow interpolation to the precise frequencies of the spectrometer channels. These show that in general a polynomial can be fit with fewer free parameters than a Fourier series. Figure 4 is the measured LNA reflection coefficient rms and Figure 5 shows the spectrum of the noise wave calibration cable. Figure 6 shows the partially calibrated spectrum prior correction for the balun. Figure 7 shows the observed spectrum prior to correction for the mismatch and balun along with the corrected spectrum and the best fit constant. The rms deviation from a flat spectrum is 1 K and the constant is 1666±0.5 K.

Page 2: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

2

4] The Measurements needed are listed in the table below

Item Measurement Comments Purpose Antenna Impedance In field Calibration Antenna Spectrum In field Observation Balun Impedance In lab open Calibration Balun Impedance In lab shorted Calibration Hot load Spectrum In lab Noise diode

calibration Open cable Impedance In lab Noise wave

calibration Open cable Spectrum In lab Noise wave

calibration Hot load Impedance In lab Calibration check Antenna simulator Impedance In lab Calibration check Antenna simulator Spectrum In lab Calibration check Note: If connector of the device under test doesn’t match the SOLT calibration connector corrections for adapters need to be made. Also it is assumed that the SOLT offset delay correction has been entered in the VNA. 5] Sources of error

While initial results provide some confidence that the method is correct the rms deviation is more than an order of magnitude larger than the goal of under 100 milli K. The problem is largely lack of accuracy and attention to detail. For example, an error of about 0.1 dB or 5 degrees in phase in antenna reflection coefficient produces an error of about 20 K. the high sensitivity is due to the relatively poor match of the antenna simulator. The following are some “lessons learned:”

1. Use sufficient averaging time when calibration and making measurements with VNA. 2. Use the appropriate sex of SOL (short open load) when calibrating the or correctly

account for the use of a male to male adapter when measuring the LNA impedance at the female input. In each case the reference plane needs to be precisely at the interface of the LNA to the antenna. For example a 1 cm long adapter has a 2-way delay of 100 ps which corresponds to a phase shift of 7 degrees at 200 MHz.

3. Be careful to avoid flexing the cable to the VNA between calibration and measurement.

4. VNA measurements of the open cable used to measure the LNA noise waves, need to be made rather than assuming a delay and loss from cable datasheet. The VNA measurements are fit to model parameters of delay, impedance, and attenuation which depends on the square root of frequency. Alternately a polynomial fit could be made to the reflection magnitude and phase but the rapid variations due to a deviation from 50 ohms require a high order polynomial.

Page 3: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

3

5. The hot load used to calibrate the internal noise diode needs to have a near perfect

match of 50 ohms with 0.1 ohms at the operating temperature. 6. The ferrite impedance in the balun needs to be measured using VNA measurements

with open and shorted balanced ports. It is not yet clear if the model of the balun as a perfect transformer with twice the ferrite impedance in parallel with the balanced port is sufficiently accurate. It is noted that while the ferrite chokes are largely resistive their stray reactance makes a transition from inductive to capacitive at about 100 MHz.

7. The EDGES electronics is not temperature controlled so that temperature coefficients of the change in LNA impedance, noise waves etc. will be needed so corrections can be made based on a measurement of the temperature. However these coefficients are quite small, for example, the temperature coefficient of LNA impedance is of the order of 0.01 Ω/K.

Summary In summary a great attention to detail and care is needed to obtain accurate calibration. The table below summarizes the level of error measurement in current tests.

Measurement Number of terms in fit Rms Antenna S11 9 5×10-4

LNA S11 13 1×10-4 Balun impedance 5 2Ω Noise wave calibration spectrum

4 0.3 K

Noise wave calibration cable delay

1 50 ps

Noise wave calibration cable loss

1 0.1 dB

Noise wave calibration cable impedance

1 0.1Ω

Page 4: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

4

Figure 1

SMA male

Ch ok e Balu n

l uH

2 x Zf ~ 800

Tu n gsten 273 K 472 K 835 K 1485K 1794K

rsistance 1.0 2 .0 4.0 8 .0

10.0

280 0 . 02

~0 .02

28v la m p 22 ohms cold 220 ohms at 1794K

l u H

0.1 uH

+28 v

EDGES an t e n na s im u la t o r

ant si m. d wg aee r 8decll

Page 5: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

5

Figure 2. Antenna simulator with balun and hot filament noise source

Page 6: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

6

Figure 3. Antenna simulator reflection coefficient. Thick line is the phase. Cures are a polynomial fit to the measurements.

co' ~

(1) "O

-~ C: bJ)

"' E

Cl)

0 180

-1 150

120 -2

90

-3 60

-4 30

-5 0

-6 -30

-60 -7

-90

-8 -120

-9 -150

-10 .____,_ _ _.__....___.....___,_ _ _.__....___.....___,_ _ _._ _ _.__.._____. _ _, -180

50.0 60.0 70.0 80.0 90.0 100.0110.0120.0130.0140.0150.0 160.0 170.0 180.0 190.0

Frequency (MHz)

~ V, (1) (1)

'"' bJ) (1)

~ (1)

"' "' ..c: 0.

..... Cl)

Page 7: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

7

Figure 4. LNA reflection coefficient. Thick line is the phase. Curves polynomial fits to the measurements.

0 180

-2 150

120 -4

90

-6 60

-8 30

,,...._ <J> Q)

0 Q)

6h Q)

~ Q)

-30 <J)

"" ..c:: 0.. --C/J C/J

-60 -14

-90

-16 -120

-18 -150

-20 .____.__ _ _.___.____.__ _ _.___.____.__ _ _.___.....___._ _ _.___..____.____, -180

50.0 60.0 70.0 80.0 90.0 100.0 110.0 120.0 130.0140.0 150.0160.0 170.0180 .0 190.0

Frequency (MHz)

Page 8: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

8

Figure 5. Spectrum of the noise wave calibration cable and fit using 4 term polynomial for each of the 3 noise wave parameters.

144 K

128 K

112 K

96 K

80 K

64K

48 K

32K

16 K

OK

50.0 60.0 70.0 80.0 90.0 100.0110.0120.0130.0140.0150.0160.0170.0180.0190.0 Frequency (MHz)

Page 9: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

9

Figure 6. Partially calibrated spectrum prior to correction for the balun. The lower curve is the observed spectrum before calibration. The upper curves are the partially calibrated spectrum and best fit constant of 1377 K.

1800 K

1600K

1200 K

1000 K

800 K

600 K

400 K

200 K

OK 50.0 60.0 70.0 80.0 90.0 100.0110.0 120.0130.0140.0 150.0 160.0170.0 180.0 190.0

Frequency (MHz)

Page 10: EDGES M #082 MASSACHUSETTS INSTITUTE OF …

10

Figure 7. Calibrated spectrum. The average is 1666 K and the rms deviation from a constant is 1 K .

1800 K

1600 K

1400 K

1200 K

1000 K

800 K

600 K

400 K

200 K

OK

50.0 60.0 70.0 80.0 90.0 100.0 llO.O 120.0130.0140.0150.0160.0170.0180.0 190.0 Frequency (MHz)