25
AquFlow ® ECCA TM Electric Capacity Control Series 1000 & 2000 INSTALLATION, OPERATION AND MAINTENANCE MANUAL AquFlow Pumps.

ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

AquFlow®

ECCATM

Electric Capacity ControlSeries 1000 & 2000

INSTALLATION, OPERATION ANDMAINTENANCE MANUAL

Aq

uF

low

Pu

mp

s.

Page 2: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

i

Table of Contents

Warranty ........................................................................................................................................... iiGeneral Description ......................................................................................................................... 1

Description of Electonic Circuitry ..................................................................................................... 1Power Supply/Driver Card .......................................................................................................... 1

Installation Instructions ..................................................................................................................... 1Electrical Connections ............................................................................................................... 3Signal Inputs ............................................................................................................................... 3

Specifications ................................................................................................................................... 2

Startup Instructions .......................................................................................................................... 3

Calibration of the Circuit Board - Series 1000Equipment Required................................................................................................................... 4Input Amplifier ............................................................................................................................. 4Checking Motor Limits ................................................................................................................ 4Setting Ratio Control .................................................................................................................. 4Returning to Service ................................................................................................................... 4

DisassemblyRemoval of P.C. Card................................................................................................................. 4Removal of Synchronous Motor ................................................................................................. 5Removal of Actuator Housing from Pump Series 1000 .............................................................. 5Removal of Actuator Housing from Pump Series 2000 .............................................................. 6

ReassemblyInstalling Actuator Housing on Pump Series 1000...................................................................... 7Installing Actuator Housing on Pump Series 2000...................................................................... 7Installation of the Synchronous Motor ......................................................................................... 7Re-Synchronozation Equipment Needed ................................................................................... 8

Returning to Service ......................................................................................................................... 9

Outdoor Installations ........................................................................................................................ 9

DrawingsHookup Diagram - Electronic Capacity Control .......................................................................... 10Board Layout - Power Supply Driver .......................................................................................... 11Board Layout - Input Buffer/Motor Logic ..................................................................................... 12Electrical Schematic - Input Buffer/Motor Logic ......................................................................... 13Electrical Schematic - Power Supply & Drive Card ................................................................... 14Test Cable - Synchronization - Local Control ............................................................................. 15Test Cable Hookup Schematic ................................................................................................... 15Series 1000/2000 Motor Coupling Installation ............................................................................. 16ECCATM Cross Section & Parts List - Series 1000 All Plungers ................................................. 17ECCATM Cross Section & Parts List - Series 2000 All Plungers ................................................. 18

Troubleshooting Sequence ............................................................................................................... 19Replacement Part Numbers ............................................................................................................ 20

Parts Ordering Instructions.................................................................................................................21

Page 3: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

ii

WARRANTY

AquFlow warrants its products against defects in workmanship or material under specified use andservice for a period of two years from date of shipment. The obligation and liability of AquFlow islimited to repairing or replacing (at its option) such allegedly defective units as are returned freight prepaid to NPA’s plant.No liability is assumed for warranties of merchantability or fitness for a particular purpose except as specified, or forreinstallation costs. Liability for consequential damages (including personal injury) is hereby excluded; liability in anyevent being limited to the original purchase price of the goods involved.

Defects shall not include failure by errosion, abrasion, chemical action, material compatibility, or operation of productsoutside of the recommended performance envelope. The component materials set forth in the proposal and/or thespecifications of an order are recommended by AquFlow for the particular application. But suchrecommendations shall not be construed as a warranty or guarantee against wear and/or corrosion, and suchrecommendations are subject in all cases to verification and acceptance by the purchaser. When repairs or replacementsare made under our warranty, incoming freight costs must be paid by the purchaser.

NIKKISO Pumps America, Inc. will not be responsible for work performed by others. AquFlowreserves the right to void applicable warranties if work performed by others is determined by AquFlow to be improper.

Read this instruction completely before installing,operating or servicing AquFlow® ECCATM Electric Capacity Controllers!

Always keep safety considerations foremost! Use proper tools, protective clothing andeye protection when working, and install the equipment with a view toward assuring safeoperation. Be extremely careful when working in the presence of hazardous substances!

Page 4: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

NIKKISO HYDROFLO® ECCATM

page 1

n this instruction manual, the Note, CAUTIONand WARNING notations have the followingmeanings:

Note: This heading is used if a special featureor operating requirement is being pointed out.

CAUTION: This heading is used to denote whenfailure to follow the operating instructions asprescribed herein can lead to pump or systemdamage.

WARNING: This heading is used to denote whenfailure to follow the operating insructions asprescribed herein could result in PERSONALINJURY!

GENERAL DESCRIPTIONThe electric capacity control actuator can befurnished in place of the standard manualmicrometer adjustment. The actuator adjustspump capacity from an electronic signal ormanually by means of an override handwheel.Local capacity indication is read directly from anindicating scale on the actuator, or remotely byuse of the electronic retransmission signal (4-20MADC) feature. Control is achieved through theuse of down sized state of the art electronics andsynchronous motor technology in a singlecompact enclosure.

Synchronous Motor

The use of this type of motor permits rapidresponse and precision positioning in a smallpackage. Since this kind of motor is magneticallydetented, overshoot is eliminated, a probleminherent in shaded pole or gear motors.

Electronics

The operation of the electric capacity controlactuator is controlled by two electronic circuitcards. The top one serving as the input and logicsection and the bottom serving as the powersupply and motor driven section. These circuitsaccept an input signal from a process instrumentand commands the synchronous motor to driveto a corresponding position. Closed looppositioning is accomplished through use of afeedback slide wire.

Description of Electronic Circuitry

Input Buffer/Motor Logic 170071. Input Buffer Amplifier: The input voltage signal

is converted from 1-5 VDC to a 0-10 VDC signalusing a high impedance amplifier, U4A. Thisvoltage is applied to the ratio control which isused to scale the signal to the appropriate value.The resultant voltage is the positive input to theerror amplifier, U4B, which compares it to theactual capacity control position as representedby the negative feedback potentiometer whichis gear driven by the actuator motor.

2. Error Amplifier and Motor Logic: The errordeveloped at U4B is applied to twocomparators, U5A & U5D, which have slightlydifferent set points. Any error in excess of thesepre-selected values will cause the motor logic,U6A or U6B, to respond with either an up (+) ordown (-) signal. The signal then enters the motordrive Section which provides currentamplification through transistors Q3 or Q4.

3. Feedback Buffer: The previously mentionedfeedback potentiometer picks off a voltage from+5 VDC to -1.2 VDC, depending upon itsposition as the motor rotates. This voltage isapplied to the feedback buffer amplifier forcomparison with the input signal. At the sametime, this signal is applied to the electroniclimiting circuitry U5B and U5C. Thesecomparators are set (via high and lowpotentiometer) to limit actuator travel in theevent of an overdriven or underdriven input.

Power Supply/Driver Card 17006Power Supply

The transformer output is rectified by a full wavebridge, filtered and routed to a pre-regulator stage.This darlington pair is controlled by zener CR3referenced to the output of the primary regulator,U1. The 5 volt reference U2 is tied to U1 through adropping diode CR2 and held 1.2 VDC above thesupply return in order to permit live zero voltagefeedback potentiometer operation without offset.

Motor Driven

The current amplified up and down signal entersthe motor driver selection at pin two of the opticallyisolated triac drivers U7 and U8. these optoisolators

Page 5: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 2

SPECIFICATIONS

Input voltage: 120 VAC 50/60 hz single phasePower consumption: 44 va maximumLine protection: Fuse type, 3 amp Bussman PCC3Line voltage variations: ±10% of nominalLine frequency variations: ±2% 50/60 hzAmbient temperatures: 0-40°C (32-104°F)Relative humidity: 5-96% (witout condensation)Altitude: 3300 feet (1000 meters)Motor type: AC synchronous (impedance protected)Response time: Series 1000 8 seconds 0-100% Series 2000 16 seconds

200 8 secondsSeries977/997 8 seconds

Turndown ration: better than 20:1Repeatability: ±1%Drive connection: Direct on Series 900, 1000, 2000 and HydromyteInput signal voltage: 1-5 VDC, 3 meg ohm impedance current: 4-20 MADC, 250 ohm impedanceOutput signal voltage: 1-5 VDC across a 250 ohm resistorCurrent: 4-20 MADC into 500 ohms maximum

Adjustments: High limit factory setLow limit factory setSpan factory setZero factory setGain field adjustable, 4-20 retransmissionOffset field adjustableRatio field adjustable 100%

Field connections: AC powerProcess signal in-captive wire typePosition signal out-captive wire typeMotor disable-MTA . 100” 2 Pin

Enclosure type: NEMA 4 standardNEMA 7 optional (class 1, div. 1, group C and D)

Indicator type: Local - mechanicalRemote - 4-20 MADC into 500 ohm or1-5 VDC across 250 ohm

Actuator control: Electronic comparator with synchronous motor andmechanical manual override

Actuator limits: Electronic - no limit switches

ACTUATOR MODEL SELECTIONACTUATOR MODEL OF PUMP SERIES OF PUMPMODEL NO. WHERE USED WHERE USED

17100-1100 100017100-2100 200017100-3100 300017100-4100 AqFlow 400017100-5100 977/99717100-5100 200

Page 6: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 3

trigger triacs CR4 or CR5 which are directly coupledto the directional windings of the capacity adjustingmotor.

INSTALLATION INSTRUCTIONSThe electric capacity control has been factory testedand calibrated on the pump prior to shipment. Nofurther field adjustments or calibrations to the factorysettings should be necessary prior to placing unitinto service.

Note: To maintain correct alignment andsupport, it is recommended that the pump/actuator combination remain bolted to its factorysupplied base plate for Series 200, 977, 997,1000 and 2000 pumps.

CAUTION: Do not adjust capacity handwheelbeyond the limits of the indicating scale ordamage to the pump and actuator may result.

Electrical Connections1. Unscrew end cover to expose with connection

points. Refer to drawing on page 12 for hookupdiagram.

2. Power and signal wires SHOULD BE RUNSEPARATELY. Use shielded cable for signal line.The electric capacity control requires a powersource of 120 VAC 50/60 hz single phase. Thebranch circuit should be protected by a fuse orcircuit breaker and contain a disconnect switch.The circuit boards are protected by a selfcontained fuse.

3. The black wire should be the phase side of theline and connect to terminal “L”; the neutral orwhite wire to terminal “N”. Equipment groundingis effected by connecting a green wire to one ofthe screws in the housing, the other end of thiswire is returned to protective ground at thesource of supply.

Note: Appropriate conduit hubs must be used,and all covers must be in place and tight toprevent liquid entry and maintain NEMA integrity.

Signal Inputs1. The electric capacity control accepts a voltage

signal of 1-5 VDC. If the controlling signal is acurrent signal, the 249 ohm resistor (R133) willtransform it into the voltage required. It may beleft in place when a voltage signal is used, but

the transmitter must be low impedance output.Better practice is to remove the resistorcompletely.

2. The 249 ohm resistor is the correct value for 4-20 MADC input. This resistor (R133) is locatedadjacent to J1 and in back of TB1 on the inputbuffer/motor logic card (17007). If a highimpedance voltage source is used, the resistorcan be removed by carefully clipping the leadat the resistor body (R133).

3. Connections have been provided that allows theactuator to be disabled. A N.O. dry contactconnected to J1 (when closed) will disable themotor drive circuit while allowing the remainingcircuitry to be fully activated. This permits theactuator to be set and then disabled, while stilltransmitting a positioner signal (4-20 MADC). Italso permits manual override while the unit ispowered up. Connections to J1 are made viaan MTA two pin connector on .100” Centers.

START UP INSTRUCTIONS1. Apply power to the actuator and inject a 4 MADC

signal at B1 to bring the actuator to zero percentcapacity. Place a mark on the slotted coupling(3409 drawing 17017, 17018). Inject a 20 MADCsignal and count the number of turns. For Series1000, 977, 997 and 200, there should be tenturns for 100% output. For Series 2000, thereshould be twenty turns for 100% output. Anydifferences can be corrected by adjusting theratio potentiometer R130 on input buffer board17007.

The electronic limiting feature built into theactuator circuitry permits only a few degrees ofpositioner rotation past 100%, likewise, a fewdegrees below zero percent. This is to preventmechanical jamming of the actuator. Whenadjusting the actuator by means of the overridehandrail, it is important to maintain indicatortravel within the limits of the indicating scale, ordamage to the actuator can occur. The manualoverride should be used in conjunction with theactuator disable feature or with power removedfrom the actuator. See section 3, under “SIGNALINPUTS” for the actuator disable feature.

2. The ratio control potentiometer is used to adjustthe full scale output of the pump for a full scaleinput signal. If it is desirable to have less than100% pump output for a 100% input signal, theratio control pot should be turned in a

Page 7: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 4

counterclockwise direction from its normallymidscale position. If 100% output is required forless than 100% input signal, the pot should beadjusted in a clockwise direction. Smalladjustments are in order because of the highgain system employed. (Any change in the ratiocontrol from the factory setting will require re-calibration of the pump).

3. Remote indication or closed loop operation canbe accomplished using the 4-20 MADCretransmission feature located at TB3. “COM”is the low or negative side and “OT” is the highor positive side. Apply 4 MADC to the input,(allow actuator to position itself) and adjust thegain pot to 20 MADC output. Since the offsetand gain pots interact with each other, severaladjustments from 4 to 20 MADC may be requiredto complete the calibration. The input and outputsignals are isolated from phase and ground butnot from each other. The maximum load thatcan be connected across TB3 is 500 ohms.

CALIBRATION OF THE CIRCUIT BOARD -Series 1000

Note: The P.C. Cards are tested, calibrated andmatched to each actuator motor and pump atthe factory, and are ready to be placed intoservice as received. The following instructionsare included as an aid in troubleshooting in theunlikely event of a malfunction.

Equipment Required1. 120 VAC supply.

2. Adjustable DC voltage source capable ofsimulating 1-5 VDC or a current source capableof 25 MADC.

3. A digital voltmeter reading .01 volts minimum.

4. Miscellaneous jumpers.

Extent of Calibration: This procedure covers thesetting of the zero and the span potentiometer,checking the high and low limit potentiometer, andsetting the ratio control. This procedure assumesthat the input signal range is 1 to 5 VDC, or 4-20MADC.

Preliminary: Connect (but do not turn on) the ACpower to terminal L&N (TB2) on power supply board17006. Safety ground can be connected at thehousing or under the screw head of a standoff.Connect adjustable DC voltage supply to terminalC-and V1 + (TB1) on input board 17007.

Input Amplifier1. Apply line power to board. Read voltage from

TP2 & to C-.

2. Connect a jumper (across the zero pot) fromR102 (inside leg) to C. Apply 5 VDC to input andadjust span pot for 12.5 VDC at TP2.

3. Remove jumper across zero pot. Apply 1 volt toinput and adjust zero pot for 0.0 V at TP2.

4. Apply 5 VDC to input and adjust span pot for 10VDC. All voltages are ±0.1 VDC.

5. Repeat steps 3 and 4 as required.

Motor Limits: The high and low potentiometers onthe input board are factory set for each actuator/pump combination and require no furtheradjustment.

Checking Motor Limits1. Input 0.98 VDC. Read voltage from test point h

(+) to common.

2. Input 0.94 VDC. Observe that voltage goes onhigh.

3. Input 2 VDC then 0.98 VDC. Observe thatvoltage goes on high.

4. Input 5.08 VDC. Read voltage from test point g(+) to common. Observe that voltage is high.

5. Input 5.13 VDC. Observe that voltage goes low.

6. Input 4 VDC then 5.08 VDC. Observe thatvoltage goes high. This procedure is for checkingproper operation of motor limits only.

Note: The above voltages are typical and willvary with pump/actuator combination.

Setting Ratio Control1. Input desired voltage for full scale operation.

2. Adjust ratio potentiometer for desired pumpoutput, turn clockwise to increase output,counterclockwise to decrease output.

Returning to Service1. Remove power and disconnect all test

equipment and jumpers.

2. Reconnect power and input/output signal cablesif other sources were used for calibration check.

3. Replace all covers securely to maintain NEMAintegrity.

Page 8: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 5

DISASSEMBLYRemoval of P.C. Cards1. Remove and lock out power and signal to

actuator and pump.

2. If possible set pump to 0% capacity usingmanual override hand wheel and capacity scale.

3. Remove cover from actuator.

4. Disconnect incoming power and signal lines andoutgoing signal line if used.

5. Remove feedback potentiometer cable from P2on card 17007.

6. Remove motor control cable from P1 on bottomcard 17006.

7. Remove the three screws (3417) retaining themotor logic card to standoffs (3548). Drawings17018 & 17017 (depending on Series of pump).

8. Remove the three standoffs (3548) retainingpower supply to standoffs (3547).

9. Unplug phase shift network leads from the wireson the underside of the power supply card.

10. Remove P.C. Cards as a pair and place them ina safe location. Care must be taken to preventbreakage of interconnect cables.

Removal of Synchronous Motor1. Remove the three standoffs (3547) attached to

the motor plate (3402). It is recommended toidentify these standoffs soles on the motor platefor reassembly later on. These standoffs are twoinches long and should not be confused withstandoffs (3458) which are one and one halfinches long. Clip off and discard any tie wrapsretaining interconnect wiring to these standoffs(3547).

2. Remove the Zerust anticorrosion pad thenremove the four flathead screws (3416) retainingthe motor plate (3407) to the four motor standoffs(3415). Carefully remove the motor plateassembly and inspect all components forbreakage or wear. Replace worn or brokenpotentiometer and drive gears. It may benecessary to loosen (not remove) the potbracket screws on the lower side of the motorplate (3407) to allow removal of motor plateassembly from actuator housing.

3. Remove the four motor standoffs (3415)retaining the motor in place. The motor is placedinside the actuator housing with the connector

at the three o’clock position relative to the conduitentry holes. During reassembly the motor mustbe installed in this same position.

4. Carefully remove the motor and inspect for wearor breakage of the drive components. The motorshaft must turn freely, but will exhibit a detentedor slightly “rough feeling” when turned by hand.The motor must be replaced if the shaft is hardto turn or “frozen”. Check the motor coupling(3431) for breakage or wear. If the coupling isusable be sure the set screws (3432) are tight,securing the coupling to the motor shaft.

The motor shaft is flatted and one of the two setscrews must engage this flat. It is also advisableto use “retaining” grade Loctite to aid in securingthe motor coupling (3431) to the motor shaft. Toavoid damage to the motor apply Loctite to theI.D. of the coupling (3431) at the large end beforesliding it over the motor shaft. See drawing17124-1 for coupling replacement procedure.

5. Check the motor gear (3437) for wear orbreakage. If the gear must be replaced be sureit is oriented with the teeth facing outward (awayfrom the motor). Care should be taken to assurethe spacer under the motor gear is in placebefore installing the new gear. Reassemble theoverride coupling (3409) to the motor shaftabove the motor gear (3437). If any drivecomponents appear worn (beyond use) orbroken, contact NIKKISO Pumps America, inc.or your local factory representative forreplacement parts. See “replacement partnumbers”.

Removal of Actuator Housing from PumpSeries 1000Most serviceable parts are contained within theactuator housing, and are accessible withoutremoving the housing from the pump. Theseinstructions are provided for your convenience inthe unlikely event the housing must be removed.

1. Remove support stud (3425). In mostapplications this will be a length of “all thread”with accompanying lock washers and nuts.

2. Remove the four screws (3418) retaining thecapacity plate (3414) to the actuator housing(3402). This plate has been factory sealed andmay require some gentle persuasion to break itfree. Set the plate in a safe area.

3. Reach inside the housing, using a flat object(three quarter inch wide machinist scale) and

Page 9: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 6

turn the actuator (3404) until the slotted end ofthe indicator pin (3413) is accessible throughthe capacity plate opening. This pin has beenfactory installed using low shear grade Loctite,use the correct size screwdriver and removethe indicator pin and set it in a safe place.

4. Loosen, but do not remove, the four clamp ringscrews (3442). This releases clamp ringpressure, permitting the actuators to beunscrewed from the mounting adapter (3401).

5. Turning counterclockwise, unscrew the actuatorhousing from the mounting adapter (3401).COUNT THE NUMBER OF TURNS AS YOUGO. This information will be necessary forreassembly of the actuator to the pump.

6. Inspect the indicator ring (3412) and pin (3413)for wear or damage. Replace as necessary,

7. Inspect the actuator coupling (3404) for wear ordamage. The actuator must turn freely insidethe housing. End movement between housingand actuator is a normal condition though radialmovement should be negligible to nonexistent.If excess radial movement is detected, housing(3402) and/or actuator (3404) should bereplaced.

8. To replace a worn or damaged actuator (3404),set the housing (3402) on end and reach insidewith a pair of snap ring pliers and remove snapring (3433). This will allow the actuator to beremoved from the housing. When replacing theactuator (3404), apply a thin film of moly greaseto its minor diameter before inserting it into thehousing, and retain it in place using a new snapring (3433). Electric capacity controldisassembly instructions stop at this point.Further disassembly of the pump is coveredunder separate print. Contact NIKKISO PumpsAmerica, Inc. for this information.

Removal of Actuator Housing from PumpSeries 2000Most serviceable parts are contained within theactuator housing, and are accessible withoutremoving the housing from the pump. Theseinstructions are provided for your convenience inthe unlikely event the housing must be removed.

1. Remove the support stud (3425). In mostapplications this will be a length of “all thread”with accompanying lockwashers and nuts.

2. Remove the four screws (3418) retaining thecapacity plate (3414) to the actuator housing

(3402). This plate has been factory sealed andmay require some gentle persuasion to break itfree. Set the plate in a safe area.

3. Loosen, but do not remove, the two adaptorhousing set screws (3429) on adaptor housing(3428). Grasp the actuator housing (3402) firmlyand in a clockwise direction, turn the adapterhousing (3428) six to eight turns to completelyseparate actuator housing (3402) from adaptorhousing (3428).

4. Inspect capacity indicator and indicator pin forwear or damage. Replace as necessary.

5. To remove the capacity indicator and pin, turnthe indicator until the pin is in the vertical position.Support the threaded adjuster (4633) andcarefully drive the pin out using an appropriatesized drift (punch). To install a new indicator andpin, support the adjuster and align one of theholes in the indicator over the hole in the adjustershaft with the raised section if the indicator facingaway from the pump housing. Place the pinthrough the hole in the indicator and hold it inplace over the hole in the adjuster (4633).

Carefully drive the pin through the hole in theadjuster until there is an equal amount ofexposed pin on either side of the adjuster shaft.Care should be taken not to deform the indicatorduring installation or binding may occur.

6. Inspect the actuator coupling (3404) for wearand damage. Actuator must turn freely insidethe housing. End movement between thehousing and the actuator is a normal conditionthough radial movement should be negligible tononexistent. If excess movement is detected,housing (3402) and actuator (3404) should bereplaced.

7. To replace a worn or damaged actuator (3404)set the housing (3402) on end and reach insidewith a pair of snap ring pliers and remove snapring (3433). This will allow the actuator to beremoved from the housing. When replacing theactuator (3404), apply a thin film of moly greaseto the minor diameter before inserting it into thehousing, and retain it in place using a new snapring (3433). Electric capacity controldisassembly instructions stop at this point.Further disassembly of the pump is coveredunder separate print. Contact NIKKISO PumpsAmerica, Inc. for this information.

Page 10: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 7

REASSEMBLY

Installing Actuator Housing on Pump Series10001. Side the capacity indicating ring (3412) over the

actuator (3404) with the raised end toward thehousing (3402).

2. Align the housing (3402) with the pump adaptor(3401). Turning clockwise, start and then screwthe housing onto the adaptor using the samenumber of turns required for removal bringingthe flatted surface by the conduit entry holesparallel to the base plate (plate the pump ismounted to).

3. Tighten the four clamp ring screws (3442).Check for any radial movement of the housing.If movement is detected, loosen the clampscrew and turn the housing (3402) one turnclockwise and retighten the four clamp ringscrews.

4. Turn actuator (3404) to align a slot with thethreaded hole in the control spool assembly(3439). Slide the indicating ring (3412) in place,and insert the indicator pin (3413) through oneof the two holes in the ring, down through theslot, and into the threaded hole of the controlspool (3439). Turning clockwise, screw theindicator pin (3413) through the control spooland down through the hole in the opposite sideof the indicating ring (3412). Ends of the pin maynot protrude past the raised section of theindicator ring.

5. Remove old sealant from the capacity plate(3414) and its mating surface on the housing.Apply a film of clear RTV (silicone) to the matingsurface on the housing and secure the capacityplate in place over the RTV using the four screws(3418).

6. Install the stud (3425) with its hardware into thebottom side of the housing (3402). It may benecessary to loosen the clamp ring screws(3442) to allow slight housing repositioning wheninstalling this stud. Retighten the clamp ringscrews first then secure the stud (3425). Thisstud is intended for support only. Care shouldbe taken when tightening the jam nuts to preventthe introduction of lateral forces that might causemisalignment between the pump and theactuator housing.

Installing Actuator Housing on Pump Series20001. Apply a thin film of moly grease to the threaded

end of adaptor (3428). Align housing (3402),actuator (3404), capacity indicator (3412) andhousing adaptor (3428). In a counterclockwisedirection, screw the housing adaptor (3402) untilintimate contact is made between these twoparts. Bring the flat surface of the conduit entryholes in the housing (3402) parallel to the baseplate (plate the pump is mounted on). Hold theactuator housing in this position and install thestud (3425) with its hardware into the bottomside of the housing (3402). Tighten the two setscrews (3429) first, then secure the stud. It isimportant to seat the two set screws (3429) inthe groove (undercut) portion of the adjusterhousing (3436). Stud (3425) is intended forsupport only, and care should be taken whentightening the jam nuts to prevent the introductionof lateral forces that might cause misalignmentbetween the pump and the housing (3402).

2. Remove old sealant from the capacity plate(3414) and its mating surface on the housing.Apply a film of clear RTV (silicone) to the matingsurface on the housing and secure the capacityplate in place over the RTV using the four screws(3418).

Installation of the Synchronous MotorThe motor assembly must be complete with coupling(3431), spacer, gear (3437), and override coupling(3409). Connector with red, black and white wiresis also considered part of the motor assembly andmust be in place.

1. Insert the motor into the housing engaging thecoupling (3431) with actuator (3404). The motormust sit squarely in its mount with no movementother than rotational.

2. Orient the motor so the connector is at the threeo’clock position relative to the conduit entry hole(conduit holes at the twelve o’clock position).Secure the motor in place using the fourstandoffs (3415). If new standoffs are beingused, it is recommended to apply a thin film ofgrease to the male threads before installation.

3. The motor plate should be a built up assemblyof (3407), phase shift network consisting of2.5UF capacitor and a 500 ohm resistor, and afeedback pot assembly. Locate this motor plateassembly over the four motor standoffs (3415)

Page 11: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 8

with the feedback pot at the bottom (sideopposite conduit entry holes). The two feedbackpot bracket screws (6-32 x 1/2) should be looseat this time to allow full disengagement of thetwo gears (3437) and (3438).

4. Install the three standoffs (3547) in the holesidentified during disassembly.

5. This step is for Series 1000 pumps only. Turnthe motor shaft counterclockwise (via slot onthe override coupling 3409) until it reaches itsmechanical stop. Then turn the shaft 1/4 turnclockwise. This is 0% capacity of the pump.

6. This step is for Series 2000 pumps only. Turnthe motor shaft clockwise (via slot on overridecoupling 3409) until it reaches mechanical stop.Place a mark on the coupling at the twelveo’clock position. Now, turn the couplingcounterclockwise one and one half turns. Thisis the 100% capacity. Now turn the couplingcounterclockwise twenty turns. This is 0%capacity.

7. The pump and the actuator are now ready to besynchronized.

Re-Synchronization Equipment NeededA. 1-5 VDC or 4-20 MADC source.B. 115 VAC power cord set.C. Test cable with plug, Hydroflo part number

17120-1.

1. Connect feedback pot cable to P2 on inputbuffer/motor logic card 17007.

2. Connect (do not apply power) 115 VAC cord setto TB2 on power supply/drive card 17006.

3. Connect voltage or current source TB1 on inputbuffer/motor logic card 17007.

4. Connect test cable plug (NIKKISO Hydroflo® partnumber 17120-1) to P1 on power supply/drivercard 17006.

5. Place these two cards on an insulated surfacesuch as cardboard, plastic, etc. Do not allowthe two cards to touch each other or any metallicsurface.

6. Connect 115 VAC light bulbs with pig tails (anywattage) to the test cable wires. The white wirefrom P1 is common or neutral to both lights, theblack is hot to one light, and the red is hot to theother light. See test cable schematic drawing17121-1. Do not connect the phase shift networkwires at this time.

7. Apply a 4 MADC or 1 VDC input signal to TB1.Make sure that negative, common or low isconnected to “C” on TB1.

8. Apply 115 VAC power to the card set. One ofthe two light should come on. If none of the lightscome on, check the following:

A. Power is present at TB1 of 7006.

B. Power supply fuse open or broken lead.

C. That J1 is NOT jumpered.

D. Light bulbs are operational.

E. Feedback pot is already at 0% position. (Thiscan be checked by rotating the pot gear oneturn clockwise or counterclockwise).

F. Interconnect wiring for breaks ormisconnections.

9. Turn the feedback pot gear clockwise until bothlights go out.

10. Remove 115 VAC power first, then remove theinput signal.

11. Carefully engage the feedback pot gear to themotor gear and tighten both 6-32 x 1/2” screws.On Series 2000 pumps, care must be taken tomaintain equal distance from standoffs to potgear. (Gear centered between both standoffs.)Actuator and pump are now synchronized.Remove power, feedback pot and test cableconnections from both cards.

12. Replace the Zerust vapor capsule. If a newcapsule is being used, do not forget to date it.This capsule should be replaced on a yearlybasis.

13. Position the power supply card over the threestandoffs (3547) oriented with TB2 atapproximately the five o’clock position. Be sureall three black fiber washers are in place on theunderside of the power supply card and overthe mounting holes securing it in place with thethree shorter standoffs (3548).

14. Connect the motor leads red, black and whiteto P1 on the bottom card, feedback pot leadsgray, brown and yellow to P2 on the top card,and the phase shift network leads to the redwires on the underside of the bottom card.These wires should be tied back and clear of allmoving parts, override shaft area and the conduitentry holes.

15. Reconnect 115 VAC power and signal lines toP.C. Cards. Apply a 4 MADC signal and then

Page 12: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 9

apply 115 VAC power. Place a mark on theoverride coupling (3409). Inject a 20 MADCsignal and count the turns from 4 MADC to 20MADC. There should be ten turns for Series 900,1000 and Hydromyte, and twenty turns for Series2000. Any differences can be adjusted out usingthe ratio pot R130. Before making the finaladjustment to the ratio pot, high limit must bechecked. Apply 20 MADC and very carefullyadjust the ratio pot clockwise (extremely smalladjustments are in order) and note the amountof rotation past one hundred percent. Therecommended allowable is one-eight turn. If thecoupling rotates more tan one-eight (1/8) turnpast one hundred percent, adjust the ratio potcounterclockwise (decreasing capacity) enoughto bring the coupling (3409) one turncounterclockwise from one hundred percent.Adjust the high limit pot (R128) counterclockwiseone turn. Readjust the ratio pot (R130) clockwiseand check maximum rotation past one hundredpercent. Repeat this procedure using smallincremental adjustments until the actuator fallswithin stated operating parameters. After initialadjustment, increments of one half turn or lessshould be used on “HI” pot (R128). Adjust theratio pot (R130) for the one hundred percentposition. Inject a 4 MADC signal to bring theactuator to zero percent capacity. Remove theinput signal altogether and note the amount ofrotation below zero. There should be no morethan one-eight (1/8) turn. If more than one-eight(1/8) turn or jamming occurs, inject 6 MADC andthen adjust the low pot (R129) clockwiseapproximately one half turn. Inject 4 MADC tobring the actuator to zero capacity and thenremove the signal altogether and again observecounterclockwise rotation below zero. Repeatthis procedure using small incrementaladjustments until actuator falls within statedoperating parameters. This concludes theresynchronization procedure.

RETURNING TO SERVICE1. Reconnect wiring to actuator and pump motor

making sure all conduit and hubs are liquid tight.Do not apply power at this time.

2. Connect suction and discharge piping.

3. Adjust oil level as required and replace oil cover.

4. Open any suction and discharge valves andcheck supply vessel for minimum fluid level.

5. Inject a 4 MADC signal and apply 115 VAC powerto the actuator and start the pump motor.

6. Gradually increase input signal to 20 MADCwhile checking for leaks and correct pump/actuator performance (noisy orientation,overcurrent, jamming, etc.). If pump is installedoutdoors in temperatures below 40°F,intermittent actuator stalling may occur. This willnot cause any damage or an out of synccondition. To obtain maximum pump/actuatorperformance, it is recommended that ambienttemperatures above 40°F be maintained.

7. Inject a 4 MADC signal to bring the actuator to0% capacity. Remove power from the pumpmotor, correct any leaks as necessary andrecheck the oil level in the pump (refer to thepump manual for the correct oil grade and level).Be sure all covers are on tight. This is importantto maintain NEMA integrity and prevent liquidentry, which may cause damage to the pumpand the actuator.

8. Adjust capacity decal (3440) to line up withcapacity indicator ring (3412) while the pump isat zero percent capacity position. Replace thescale if it is damaged or hard to read.

9. Pump is ready for service.

OUTDOOR INSTALLATIONSSpecial consideration must be given when the pumpand actuator are used in outdoor service. Avoidlocating the pump and actuator in an area whereflooding or direct water runoff can occur.

Although the pump and actuator are designed foroutdoor service, condensation will form within theactuator housing as temperatures fluctuate,especially in sub-tropical or humid environments.

In all outdoor installations, a space heater shouldbe installed in the actuator enclosure, or the unitshould be heat traced such that a temperature of55-70°F can be maintained.

Properly sized space heaters are available fromNIKKISO Hydroflo®. Consult the factory or your localrepresentative for additional information.

Page 13: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 10

TB1

- 4-2

0maD

C In

put S

igna

lTB

2 - 1

-60/

50-1

20 P

ower

Inpu

tTB

3 - 4

-20m

aDC

/1-5

VD

C R

etra

nsm

issi

on

Not

es:

1.R

ecom

men

ded

wire

siz

e 16

AW

G a

mx.

2.In

put a

nd re

trans

mis

sion

sig

nal m

ust b

ein

sep

erat

e co

ndui

t fro

m p

ower

inpu

t.3.

Ref

eren

ce b

oard

layo

ut d

raw

ings

-To

p ca

rd -

Inpu

t buf

fer/m

otor

logi

cB

-170

07Bo

ttom

car

d - P

ower

sup

ply/

driv

erB

-170

064.

Con

duit

conn

ectio

n si

ze -

1/2”

- 14

NPT

5.R

emov

e Pl

astic

Plu

g an

d re

plac

e w

ithm

etal

lic to

mai

ntai

n N

MEA

enc

losu

rein

tegr

ity.

End

Vie

w

Hook-Up DiagramElectronic Capactiy Control (Local)

Page 14: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 11

Top

of A

ctua

tor

Board LayoutPower Supply/Driver

Des

Des

crip

tion

CA,

C5

Cap

acito

r - .0

1UF

500V

C2C

apac

itor E

lec.

470

UF

80V

C3,

C4,

C6

Cap

acito

r .1U

F 50

VC7

Cap

acito

r .01

UF

100V

CR1

Dio

de B

ridge

Rec

tifie

r 1 A

MP

600

PR

VCR

2D

iode

Zen

er 1

3VCR

3D

iode

Zen

er 4

.7V

CR

4, C

R5

Tria

c 80

0V @

8 A

MP

F1Fu

se 3

Am

pP1

Plug

hea

der 4

pos

ition

Q1

Tran

sist

or 2

N22

22A

Q2

Tran

sist

or 2

N30

54R

100,

R10

3R

esis

tor 3

0 O

hms

1/2

wat

t C.C

.R

101,

R10

2R

esis

tor 5

60 O

hms

1/2

Wat

t C.C

.R

104,

R10

5R

esis

tor 3

.9K

Ohm

s 1/

2 W

att C

.C.

R10

6R

esis

tor 1

6.5K

Ohm

s 1/

4 W

att H

.F.

T1Tr

ansf

orm

er 1

20 x

30

100M

ATB

2Te

rmin

al B

lock

2 p

ositi

on ri

ght a

ngle

V1Va

risto

rU1

Volta

ge R

egul

ator

+24

vol

t an7

824

U2Vo

ltage

Reg

ulat

or +

5 vo

lt EC

G 9

77U3

Volta

ge R

efer

ence

1.2

V LM

385Z

1.2

U7,

U8

Tria

c dr

iver

opt

oiso

lato

r MO

C 3

083

Page 15: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 12

Top

of A

ctua

tor

Ran

geR

ES

Board LayoutInput Buffer/Motor Logic

Des

Des

crip

tion

C10

0C

apac

itor,

.01U

F 10

0V C

eram

icC

101,

103

Cap

acito

r .1

UF

50V

Cer

amic

C10

2C

apac

itor 1

0UF

16V

TA

NT

CR

1D

iode

IN 4

003

P2

Jum

per

Plu

g H

eade

r 4

Pos

ition

Loc

king

Q3,

Q4

Tran

sist

or 2

N22

22A

Q5

Tran

sist

or 2

N24

84Q

6Tr

ansi

stor

2N

2270

R10

0, R

101

Res

isto

r 5.

6K 1

/4W

C.C

. 5%

R10

2R

esis

tor

5.76

K 1

/4W

M.F

. 1%

R10

3R

esis

tor

3.3M

1/4

W C

.C.

5%R

104

Res

isto

r 24

9K 1

/4W

M.F

. 1%

R10

5R

esis

tor

11K

1/4

W M

.F.

1%R

106

Res

isto

r 49

9K 1

/4W

M.F

. 1%

R10

7R

esis

tor

453K

1/4

W M

.F.

1%R

108

Res

isto

r 10

0K 1

/4W

M.F

. 1%

R10

9R

esis

tor

200K

1/4

W M

.F.

1%R

110,

R11

3R

esis

tor

17.4

K 1

/4W

M.F

. 1%

R11

1, R

112

Res

isto

r 20

0 O

HM

S 1

/4W

M.F

. 1%

R11

4R

esis

tor

24.9

K 1

/4W

M.F

. 1%

R11

5, R

116

Res

isto

r 1M

, 1/

4W M

.F.

1%R

117,

R13

3R

esis

tor

249

OH

M 1

/4W

M.F

. 1%

R11

8R

esis

tor

10K

OH

M 1

/4W

M.F

. 1%

R11

9, R

120

Res

isto

r 20

K O

HM

1/4

W M

.F.

1%R

121

Res

isto

r 4.

22K

OH

M 1

/4W

M.F

. 1%

R12

2R

esis

tor

46.4

K O

HM

1/4

W M

.F.

1%R

123

Res

isto

r 3.

01K

OH

M 1

/4W

M.F

. 1%

R12

4R

esis

tor

158K

1/4

W M

.F.

1%R

125

Res

isto

r 49

.9K

1/4

W M

.F.

1%R

126,

R12

7P

oten

tiom

eter

2K

1/2

W M

ulti-

turn

R12

8, R

129

Pot

entio

met

er 1

0K 1

/2W

Mul

ti-tu

rnR

130

Pot

entio

met

er 1

00K

1/2

W S

ingl

e Tu

rnR

131,

R13

2P

oten

tiom

eter

5K

1/2

W M

ulti-

turn

RN

1R

esis

tor

netw

ork

3.9K

OH

MR

N2,

RN

3R

esis

tor

Net

wor

k 10

K O

HM

TB1,

TB

3Te

rmin

al b

lock

2 p

ositi

on r

ight

ang

leTP

2Tu

rret

term

inal

U4

I.C. L

H32

4NU

5I.C

. LH

339N

U6

I.C. C

D40

82B

CN

U7

I.C. L

H35

8NJ1

Jum

per,

shor

ting

Page 16: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 13

Electrical SchematicInput Buffer/Motor Logic

Not

es:

1. R

ange

resi

stor

sol

dere

d to

boa

rd.

f

or 1

- 5V

DC

rem

ove

249

OH

M2.

U4C

not

use

d

Page 17: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 14

Electrical SchematicPower Supply & Drive Card

Not

es:

1. R

esis

tor -

Ser

ies

1000

, 200

0 - 5

00 O

HM

S, 2

5 W

atts

S

erie

s 30

00 -

200

OH

MS

, 25

Wat

ts2.

Cap

acito

r - S

erie

s 10

00, 2

000

- 2.5

OU

F, 2

80 V

AC

S

erie

s 30

00 -

4.0U

F, 2

80VA

C3.

Pha

se s

hift

netw

ork

is m

ount

ed s

eper

ate

from

P.C

. Car

d4.

º D

irect

ly w

ired

to s

chem

atic

170

01

Page 18: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 15

Item Qty Description1 1 4 Pin Connector2 1 4 Pin Dust Cover3 12” 22 AWG Black4 12” 22 AWG Red5 12” 22 AWG White

Black

Red

White

Test Cable - SychronizationLocal Control

Customer Supplied

P/N 17120-1

Test Cable Hook-up Schematic

Wire nut

Ligh

ts w

ith p

igta

ils10

0 W

atts

@ 1

15 V

AC

RedBlack

White

Page 19: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 16

Series 1000/2000 Motor CouplingInstallation

To a

void

dam

age

to m

otor

appl

y lo

ctite

to I.

D. o

fco

uplin

g ‘O

NLY

.010

± .0

05 g

ap

Page 20: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 17

Not

es:

1. L

ocal

con

trol s

how

n.2.

Rem

ote

cont

rol d

elet

e th

e fo

llow

ing

i

tem

s: 3

417,

342

0, 3

527,

354

7, 3

548

3. P

ump

mod

el &

ser

ial n

umbe

r req

uire

d

to

acc

urat

ely

spec

ify p

arts

whi

ch

va

ry b

y si

ze.

4. A

ll inf

orm

atio

n co

ntai

ned

here

in

su

bjec

t to

chan

ge w

ithou

t not

ice.

Item

Qty

Des

crip

tion

3431

1C

oupl

ing,

mot

or34

322

Set

scr

ew, c

oupl

ing

3433

1S

nap

ring,

act

uato

r34

371

Gea

r, m

otor

3439

1C

ontro

l spo

ol a

ssem

bly

3440

1D

ecal

, cap

acity

3527

1P

C b

oard

ass

y.,

pow

er s

uppl

y/dr

iver

3533

1S

nap

ring

3547

3St

ando

ff, P

c bo

ard

3548

3St

ando

ff, P

C b

oard

3441

1C

lam

p rin

g34

424

Scr

ew, c

lam

p rin

g34

431

Stop

col

lar

Item

Qty

Des

crip

tion

2601

Ref

Hou

sing

Pum

p26

23R

efO

-rin

g, c

ontro

l spo

ol a

ssem

bly

2641

Ref

Scr

ew,

cont

rol s

pool

3401

1A

dapt

er,

hous

ing

3402

1H

ousi

ng34

031

Cov

er,

hous

ing

3404

1Ac

tuat

or34

051

Sha

ft, o

verri

de34

061

Kno

b, o

verri

de34

071

Pla

te, m

otor

3408

1M

otor

3409

1C

oupl

ing

3411

1Sp

ring,

ove

rride

3412

1In

dica

tor,

capa

city

Item

Qty

Des

crip

tion

3413

1P

in,

indi

cato

r34

141

Pla

te, c

apac

ity34

154

Stan

doff,

mot

or p

late

3416

4S

crew

, mot

or p

late

3417

3S

crew

, PC

boa

rd34

184

Scr

ew, c

apac

ity p

late

3420

1P

C b

oard

ass

y. in

put b

uffe

r/m

otor

log

ic.

retra

nsm

issi

on34

22R

efS

et s

crew

, kno

b34

232

Set

scr

ew, c

oupl

ing

3424

1P

in, o

verri

de34

25A/

RB

olt

3426

1O

-rin

g, h

ousi

ng34

271

Sea

l, sh

aft o

verri

de

Electronic Capacity Control ActuatorCross Section & Parts List - Series 1000, all plungers

Page 21: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 18

Electronic Capacity ActuatorCross Section & Parts List - Series 2000 all plungers

Not

es:

1. L

ocal

con

trol s

how

n.2.

Rem

ote

cont

rol d

elet

e th

e fo

llow

ing

i

tem

s: 3

417,

342

0, 3

527,

354

7, 3

548

3. P

ump

mod

el &

ser

ial n

umbe

r req

uire

d

to

acc

urat

ely

spec

ify p

arts

whi

ch

va

ry b

y si

ze.

4. A

ll inf

orm

atio

n co

ntai

ned

here

in

su

bjec

t to

chan

ge w

ithou

t not

ice.

Item

Qty

Des

crip

tion

3431

1C

oupl

ing,

mot

or34

322

Set

scr

ew, c

oupl

ing

3433

1S

nap

ring,

act

uato

r34

341

Dec

al, c

apac

ity34

351

Gea

r, m

otor

3527

1P

C B

oard

ass

y.,

pow

er s

uppl

y/dr

iver

3533

1S

nap

ring

3547

3St

ando

ff, P

C b

oard

3548

3St

ando

ff, P

C b

oard

3436

1A

djus

ter

hous

ing

3443

1A

djus

ter

4636

1O

-rin

g, a

djus

ter

Item

Qty

Des

crip

tion

3402

1H

ousi

ng34

031

Cov

er,

hous

ing

3404

1Ac

tuat

or34

051

Sha

ft, o

verri

de34

061

Kno

b, o

verri

de34

071

Pla

te, m

otor

3408

1M

otor

3409

1C

oupl

ing

3410

1G

er,

pote

ntio

met

er34

111

Sprin

g, o

verri

de34

121

Indi

cato

r, ca

paci

ty34

131

Pin

, in

dica

tor

3414

1P

late

, cap

acity

3415

4St

ando

ff, m

otor

pla

te

Item

Qty

Des

crip

tion

3416

4S

crew

, mot

or p

late

3417

3S

crew

, PC

boa

rd34

184

Scr

ew, c

apac

ity p

late

3420

1P

C b

oard

ass

y. in

put b

uffe

r/m

otor

log

ic.

retra

nsm

issi

on34

22R

efS

et s

crew

, kno

b34

232

Set

scr

ew, c

oupl

ing

3424

1P

in, o

verri

de34

25A/

RB

olt

3426

1O

-rin

g, h

ousi

ng34

271

Sea

l, sh

aft o

verri

de34

281

Sea

l, sh

aft o

verri

de34

292

Set

scr

ew,

adap

ter

hous

ing

Page 22: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 19

TROUBLESHOOTING SEQUENCE

SYMPTOM PROBABLE CAUSES RECOMMENDED ACTION

1. Actuator does not Power not present at control Check for power at TB2 “N” and “L” for 115 VAC.respond to input. cards.

Process signal (4-20 MADC/ Check for correct signal at TB1 on top card1-5 VDC) not present at 17007. Connect dvm across “C” and “V1” atTB1. controller. Actuator responds to 4-20 MADC

signal which is converted to 1-5 VDC at TB1.

Process signal connected to Check input/output connections. Refer tosignal output instead of signal hookup diagram.input.

Phase shift network not Check phase shift network connections onconnected or defective. underside of power supply card.

Check phase shift network for shorted or opencomponents. Set OHM meter to Rx10K scale.Needle deflection should rise to approximately1/2 scale and return to 1000K OHMS. If meterdoes not return to 1000K OHMS or does notmove at all, phase shift network is defective.

2. Actuator power supply Shorted motor, phase shift Remove connector from P1 and checkfuse F-1 clears after network or power supply resistance between RED and BLACK wire.application of power. transformer. Reading should be approximately 450-500

ohms; across black or red to white 225-250ohms, and 1000k ohms from any three to theframe. If motor is shorted, contact NIKKISOHydroflo® for a replacement motor.

Check both phase shift network leads for ashort circuit to ground (Casing).

Check power supply transformer for a shortedor open primary/secondary. Resistance at theprimary should be approximately 300 ohms andat the secondary 50 ohms. If the transformer isshorted or open replace entire card set. CallNIKKISO Hydroflo® for replacement parts.

3. Actuator exhibits poor Plus and minus process CHECK CONNECTIONS!response and controls signal wires are switched. Minus (com, lo-), side of process MUST beonly half the output of connected to terminal “C” of TB1, high sidethe pump. to terminal V1.

4. Manual override feature Power still applied to Remove power to actuator.is hard to turn or “feels actuator. orrough”. Does not stay Disable actuator motor via J1. See page 5,positioned. section3, under “SIGNAL INPUTS”.

5. Pump does not respond Drive coupling loose or Tighten or replace drive coupling. See page __to manual override or the broken. under ‘REMOVAL OFautomatic positioner. SYNCHRONOUS MOTOR”.(Positioner turns butactuator coupling 3404does not). Series 900,1000, 2000 and Hydromyte.

Note: The two red wires onthe underside of the powersupply connect to thephase shift network.Disconnect these leads atthe wires for access to thenetwork connectors.

Page 23: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

Replacement Part NumbersItem Description Part No. Direct Drive Series3401 adapter, housing 17093-1 10003402 actuator housing 17083-1 all3403 cover housing 17086-1 all3404 actuator 17103-1 all3405 shaft, override 17074-1 all3406 knob, override 17065-0001 all3407 plate motor 17071-1 all

capacitor 2.5Mf 17014-0401 allresistor 500 ohm 17021-0103 allpot bracket 17088-1 allpotentiometer 5k with cable 17123-1 allscrew, rh (4-40 x 1/4”) 07430-2104 allscrew, rh (6-32 x 1/2”) 07430-2208 all terminal qd f 02F6-18279 all terminal qd m 02F8-18278 all vapor capsules 17111-1 all

3408 motor, ac syn 90 oz/in 17109-1 allmotor cable assembly 17122 all

3409 coupling 17087-1 all3410 gear potentiometer 17097-1004 20003411 spring, override 17070-1 all3412 capacity indicator 17090-1 all3413 drive pin, indicator 17102-1 1000

drive pin, indicator 07750-41528 20003414 plate capacity 17092-1 all3415 standoff 17042-1 all3416 screw, motor plate (10-32 x 1/2”) 07460-2408 all3417 screw, pcb (6-32 x 3/8”) 07430-2206 all3418 screw, capacity plate (8-32 x 3/8”) 07410-3306 all3420 pcb set, power supply/input buffer 17108-1 all3422 set screw knob (10-32 x 3/8”) 07310-4406 all3423 set screw, coupling (8-32 x 1/4”) 07310-4304 all3424 pin override 12530-0207 all3425 stud a/r all3426 o-ring housing 12020-1125 all3427 seal, shaft override 15036-1005 all3428 adapter, housing 17091-1 20003429 set screw (3/8-16 x 1/2”) 07310-4908 20003431 coupling motor 17104-1 all3432 set crews motor (8-32 x 3/16”) 07310-4303 all3433 snap ring, actuator 07900-4100 all3434 decal capacity scale 12135-0 20003435 gear motor 17091-1001 20003437 gear, motor 17097-1002 10003438 gera, potentiometer 17097-1003 10003440 decal capacity scale 12544-0 10003441 clamp ring 17049-1 10003442 screw, clamp ring (10-32 x 1/2”) 07100-4408 10003533 snap ring 17075-1101 all3547 standoff 2” 17024-31250 all

washer #8 fiber, non-metallic 07800-10200 all3548 standoff 1-1/2” 17024-31203 all

page 20

Page 24: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

page 21

PARTS ORDERING INSTRUCTIONS

Always have the complete NIKKISO Hydroflo®

pump model and serial number available whencontacting NIKKISO Pumps America, Inc. for partsor service. This information can be found on thenameplate mounted on the side of the pump.

Because of the complicated solid-state circuitrycontained in the control unit, we do notrecommend field servicing unless the necessarytools and test equipment are available.

Should it be necessary to send the pump andelectric capacity control to the factory for repair orservice, thoroughly flush the pump liquid end ofall process fluid, drain all hydraulic oil, pack securelyfor shipment, and call factory for a returnauthorization number. A material safety data sheet(MSDS) and a signed Return Goods Authorizationform will be required. Equipment received withoutreturn authorization number will not be accepted.

Direct all inquiries and orders to your localrepresentative or directly to:

AquFlow1642 McGaw Ave

Irvine, CA 92614Telephone: (949) 757-1753

Fax: (949) 757-1687Website: www.aquflow.comE-mail: [email protected]

AquFlow®

CHEMICAL METERING PUMP

MODEL CJ4T 11314-04014

SERIAL 1234567

CAPACITY 25.7 GPH

PRESSURE 425 PSIG

AquFlow Pumpwww.aquflowpumps.com

Page 25: ECCA Instruction Manual - AquFlow Pumps · supply and motor driven section. These circuits accept an input signal from a process instrument and commands the synchronous motor to drive

2001 NPA AquFlow Pumps

1642 McGaw Ave • Irvine, CA 92620Phone (949) 757-1753 • Fax (949) 757-1687

email: [email protected] http://www.aquflowpumps.com

Printed in U.S.A. 11/01

Bulletin 24200.60