Upload
lydan
View
214
Download
1
Embed Size (px)
Citation preview
EasySampler
Tips, Tricks and Case Studies
David Place
What is EasySampler?
• All-in-one sample collector AND
diluter/sample preparer
– Enables unattended pull sampling for offline
analysis (great for stability studies)
– Enables Quench-in-place operations
– Enables robust and precise sampling
from reaction Slurries
• Technology – Mechanical Actuation
– Sliding pocket inside a tube
– Carefully machined teflon tip/seal
Pfizer Confidential │ 2
20 mL
Case Studies Guide
Pfizer Confidential │ 3
Sampling Difficulty Scale Considerations:
• Reaction conditions (i.e. Temp, Pressure)
• Reaction texture
• Reaction mixture accessibility
Case Study Stats
Rxn Conc = 33 mg/mL
Temperature = 102 oC
EasySampler settings:
Commercial System
Q,D = 1:1 Acetonitrile/water
R = t-amyl alcohol
df = 160
Reactor
Reaction
SolventDiluentQuench Waste 10 mL
Sample Vial
ID = 0.8 mm
L = 0.15 m
V = 0.075 mL
Piston
Pump
200 uL Stroke
ID = 0.8 mm
L = 0.15+1.8 m
V = 1.0 mL
Needle
ID = 0.8 mm
L =0.1 m
V = 0.05 mL
ID = 0.8 mm
L = 0.8+1.8 m
V = 1.3 mL
Q D R
Dilution factor
Rule of Thumb for new chemistry:
For most Aromatic containing Materials monitored at 210-254 nm,
The following equation will give HPLC/UPLC results on scale
df = [Limiting reagent (mg)/ Total Volume (mL)]/ 0.3 mg/mL
Pfizer Confidential │ 4
Simple Dilution:
“I put one drop of the reaction mixture in 2 mL of diluent
then inject that solution”
Differentiation: Manual vs. EasySampler
Challenge: Try to sample a
solution manually 10 x. Can
you get the same result each
time?
Double Challenge: Now try to
do the same from a slurry.
What do you need to do to
ensure precise sampling?
Now which method do you feel
most confident in using if you
only get to sample once?
Pfizer Confidential │ 5
TRACER Material Sampling
Instrument
Sampling
Reproducibility
Naphthalenefull tip VWR
pippettor4%
Naphthalene EasySampler 5%
Naphthalene plastic pipette 6%
Naphthalene glass pipette 10%
Naphthalene syringe 46%
Acetaminophenfull tip VWR
pippettor22%
Acetaminophencut tip VWR
pippettor19%
Acetaminophen EasySampler 4%
Acetaminophen plastic pipette 52%
Acetaminophen glass pipette 61%
Acetaminophen syringe 37%
Common Sampling (Homogeneous)
Slurry Sampling (heterogeneous)
What does your reaction look like?
• Successfully sampled for kinetics in all of these cases w/ EasySampler to
provide modeling quality data sets
Pfizer Confidential │ 6
Thick Slurry Viscous w/ solids Undissolved
Inorganics
Tri-phasic
mixtures Just plain Ugly
(Cuprate salts)
O2
Oxygen sensitive
EasySampler = Representative Sampling
Representative sampling
• Sample containing all species,
solid/liquid/gas, contained
within a mixture which
represents the distributions of
those species in the bulk
media
• Examples:
– In Process Controls (IPC)
– Quality Control
– Kinetic Analysis
Selective Sampling
• Sample containing only a
portion of analyzable species
in a mixture due to physical
partitioning of the bulk media
• Examples:
– Mother Liquor analysis
– Solubility testing
– Phase splits
Slurries: Probe orientation is important
Test System in isopropanol:
• Dissolved tracer = Diphenyl ether
• Undissolved tracer = Caffeine
Best Tradeoff is pocket facing 180o
AWAY from agitator
Pfizer Confidential │ 8
Agi
90o
180o
270o
Reaction Mass
0o
Case Ullman: Dealing with clogs (Kristin Wiglesworth)
Pfizer Confidential │ 9
Easy Hard *
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
80.00
90.00
100.00
0.0 5.0 10.0 15.0 20.0 25.0 30.0
UP
LC A
% a
t 2
20
nm
Time, h
Conversion: A% relative to Toluene (Int Std)
ArBr
ArOH
Product
Rxn Conc = ~125 mg/mL
Temperature = 85 oC
EasySampler settings:
Alpha Prototype
Q,D = 1 v/v% water in DMSO
R = Toluene
df = 500
Reaction Matrix thickens
Preventing Clogs: Dissolution Rate important
• Lesson Learned: Choose the quench (Q) and diluent (D) solution wisely
• Use Quench operation in your EasySampler Method
– Quench oscillates the flow back and forth through the pocket allowing better mixing
and more time for solids to dissolve
Pfizer Confidential │ 10
0
2
4
6
8
10
12
14
16
0 2 4 6 8
Con
cent
rati
on
, mg/
mL
Time, minutes
Conc, mg/mL (50 % IPA)
Conc, mg/mL (75 % IPA)
Conc, mg/mL (90 % IPA)
Thin
ing
of S
lurr
y O
bser
ved
Dis
solu
tion
Ach
ieve
d
kLA = 2.2 ± 1.9% sec-1
Test system: Caffeine in IPA/water Mixtures
Empirical observation:
If at least 90% of the solid
Does not dissolve by 4 min
Clogging or inconsistent
Sampling will result.
Not recommended to use
Dilution factors less than 160
In commercial systems
Case Imidazolide : Quench-In-Place (Shu Yu)
Pfizer Confidential │ 11
Easy Hard *
Rxn Conc = 100 mg/mL
Temperature = 22 – 50 oC
EasySampler settings:
Beta Prototype
Q,D = 1:1 DMSO/MeOH
R = THF
df = 330
Dete
rmin
e A
ctivation E
nerg
ies
Slurries – Things to Watch for
Sampling from Slurry depends
on Agitation rate and
Probe Position
Stratification
Solid
crust
EasySampler Facilitates Workflow
• EasySampler is an enabling technology that facilitates integration
between hardware and software assets at Pfizer. Integration leads to
stronger fundamental and dynamic models of chemical processes.
Pfizer Confidential │ 13
Without a technology like this
All other steps become
More tedious and difficult
Case C-H Activation: Trending and IPC specification
(Brian Vanderplas)
Rxn Conc = 33 mg/mL
Temperature = 102 oC
EasySampler settings:
Commercial System
Q,D = 1:1 Acetonitrile/water
R = t-amyl alcohol
df = 160
Pfizer Confidential │ 14
7 h
IPC Target 5 A% Substrate
Easy Hard *
Challenge: Calling reaction endpoint
Benefit: Further mechanistic insights
Case De-Boc: Dynochem Modeling (Marshall Greenwood, Brian Vanderplas)
Pfizer Confidential │ 15
Easy Hard *
Rxn Conc = 250190 mg/mL
Temperature = 60 oC
EasySampler settings:
Commercial System
Q,D = 3:1 Acetonitrile/water
R = Methanol
df = 350
Base-lining kinetics sets stage for future
mechanistic understanding
Challenge: catalytic reaction showing
Dependence on catalyst why?
k1= 14 x 10-4 1/s, C.I. 11%
k2 = 9.6 x 10-4 1/s, C.I. 11%
Case Amination: Virtual Design Space (Jeff Sperry)
Best Trade-off:
T = 27 oC, NH4OH = 5.9 mol equiv.
Pfizer Confidential │ 16
Easy Hard *
Case Pressure Reactors (Teresa Makowski, Jason Mustakis)
• Mock System using JM
Carbon support (~1 wt% vs.
reaction mass)
• Slow leaks observed when
probe is actuated
• Consistent Sampling
observed up to 75 psig
• Heterogeneous catalyst
actually benefit stopping
leaks
Pfizer Confidential │ 17
Pressure, psig Carbon TRACER %RSD Sample Weight %RSD Accuracy, % deviation
0.10 no 0.68% 0.62% +5.4%
75 no 0.70% 0.76% +181%
75 yes 2.05% 0.99% +6.1%
Disclaimer: This is Out-of-spec use of the EasySampler but…shows promise
• Homogeneous Catalyst system
• Seeing solvent dependent
degradation of chiral center
• EasySampler provided a
method to sample without
breaking pressure
Case Pd Catalysed Carbonylation (Teresa Makowski)
Pfizer Confidential │ 18
Easy Hard *
Rxn Conc ~ 116 mg/mL
Pressure, Temperature = 50 psi, 80 oC
EasySampler settings:
Commercial System
Q,D = Acetonitrile; R = Methanol
df = 160
Case Heterogenous Catalysis (Teresa Makowski)
Pfizer Confidential │ 19
Easy Hard *
Rxn Conc ~ 50 mg/mL
Pressure, Temperature = 70 psi, 70 oC
EasySampler settings:
Commercial System
Q,D = Methanol; R = Methanol
df = 160
Rxn Conc ~ 50 mg/mL
Pressure, Temperature = 25-50 psi, 70 oC
EasySampler settings:
Commercial System
Q,D = Methanol; R = Methanol
df = 160
Seeing evidence of pump/clog issues
PUMP ERROR: Pressure testing Reactors with
EasySampler installed
• Most probe technologies allow pressure testing with an empty reactor
• Not recommended for EasySampler, fill with solvent to avoid EasySampler
Pump errors Gas leaks cause complete evacuation of lines over time
Pfizer Confidential │ 20
Take-Away #1
• Two key features of the reaction mixture should be
present in order to acquire Representative samples:
1.Reaction mixture is flowable/suspenable
2.Reaction mixture is dissolvable
• When Mixture doesn’t have these properties, selective
sampling is achieved and results obtained will be skewed
to favor observation of certain reaction components over
others
Pfizer Confidential │ 21
Take Away #2
THINGS TO CONSIDER BEFORE SAMPLING for UPLC/HPLC
These concepts are important regardless of the equipment used:
1.Can I dissolve the mixture (and with what)?
2.What dilution factor should I use?
3.Where should I sample the mixture?
4.Can I repeat and get the same result?
5.Easysampler: is the probe pocket oriented 180o away from the agitator
For most Aromatic containing Materials monitored at 210-254 nm,
The following equation will give HPLC/UPLC results on scale
df = [Limiting reagent (mg)/ Total Volume (mL)]/ 0.3 mg/mL
Pfizer Confidential │ 22
Acknowledgements
• Pfizer
Joel Hawkins
Stephanie Dolph
George Reid
Hahdi Perfect
Barb Sitter
Daniel Price
• Mettler Toledo
Peter Blacklin
Vaso Vlachos
Reni Mueller
Cis Vandereycken
Roel Ferwerda
Erich Schellinger
Nilesh Shah
• Development Partners
Luc Moens (JNJ)
Tom Maes (JNJ)
Chris Nichols (JNJ)
Antonio Ramirez (BMS)
Selin Aytar (BMS)
Jon Jurica (Merck)
Charles Orella (Merck)
Mats Ridemark (AZ)
Christian Bernlind (AZ)
Pfizer Confidential │ 23
Back-up Slides
Pfizer Confidential │ 24
Slurry vs. Temp
Control
@ 25 °C
60 °C to
reflux
10 °C +
1 day
Slurry
thins Recryst
Solid
crust
5/11/2015
Site Install Performance Metrics
Pfizer Confidential │ 26
Leak Detection Metrics Sampling Performance Metrics
SN# Pump SN# Probe Back Pressure, barg Tube ID, mm DYE TEST Pump performance Sampling %RSD Pocket Volume, uL Real Stroke, mL
Criteria For Success --> 1.7 +/- 0.3 0.80 +/- 0.16 < 1% < 5% 20 +/- 5 0.2000 +/- 0.0060
B432897873 B438042793 1.64 0.81 PASS 0.13% 5.3% 29.95 0.1983
B443190018 B443190028 1.63 0.81 PASS 0.39% 3.3% 20.60 0.1977
B432897872 B438042794 1.52 0.83 PASS 0.50% 3.2% 20.2 0.1983
B435975955 B443190040 1.64 0.81 PASS 0.49% 5.0% 22.2 0.2000
B435975954 B441110195 1.63 0.81 PASS 0.31% 2.4% 22.0 0.1974
B435975956 B441110190 1.50 0.83 PASS 0.52% 1.5% 21.8 0.1973
B435975962 B441110192 1.73 0.80 PASS 0.14% 4.2% 21.3 0.1992
B443190019 B443190029 1.76 0.80 PASS 0.09% 4.3% 20.93 0.1992
Ismatec MODEL
78017
Isopropanol
Solvent Resrevoir
P
Alpha Pfastrack
Pressure Tranducer
1/8" PTFE
1/8" PTFE
PTFE
ID=0.8 mm
L= 1.8 m
PTFE
ID=0.8 mm
L= 1.8 m
Displayed Text Description Manufacturer Material ModelIsmatec MODEL
78017 Peristaltic pump Ismatec
Isopropanol
Solvent ResrevoirVented Schott bottle
Equipment List Toluene tracer in IPA
Leak Detection Metrics
Test # Notes SN# Probe Back Pressure, barg Tube ID, mm DYE TEST
1 Install B438042794 2.99 0.70 PASS
2 Changed Tubing B438042794 3.20 0.69 PASS
3 Changed Tip B438042794 2.15 0.76 PASS
4 Changed Head B438042794 1.52 0.83 PASS
Hagen-Poisieulle used to Calculate Tube ID
What to expect when the Teflon Tip dies
Checklist:
1. Was the sample tip completely
immersed?
2. For slurries, Probe optimally
oriented in the flow?
3. Does diluent dissolve solids?
4. UPLC/HLC integrator correctly
identifying peaks in each
sample?
Pfizer Confidential │ 27
Prim
ary
Convers
ion, A
%
Impurities, A
%
Change Teflon Tip out with new one
Time, h
Observed Teflon Tip lifetimes (so far)
• Running at 102 oC and/or 50 psi + for 12-24 h = 400-450 actuations ( ~ 14 expts )
• HOWEVER, one system up to 770 actuations and still functioning optimally
Cross Contamination
Scenario: Q = Water, D= MeCN, R = THF; 1 g scale
• Assume 100 ppm water contaminates reaction solvent ( R ) in the 20
uL pocket.
• How many actuations are required to add 1% of the water required to
completely quench reaction mixture?
Pfizer Confidential │ 28
610
Tracer Study (Zero Series)
Tracer = Diphenyl ether
(DPE)
• Sample Reaction
mixture 10 x
• Collect Samples at
max df = 450
• Sample reaction pot
manually
Pfizer Confidential │ 29
Reactor
Reaction
SolventDiluentQuench Waste 10 mL
Sample Vial
ID = 0.8 mm
L = 0.15 m
V = 0.075 mL
Piston
Pump
200 uL Stroke
ID = 0.8 mm
L = 0.15+1.8 m
V = 1.0 mL
Needle
ID = 0.8 mm
L =0.1 m
V = 0.05 mL
ID = 0.8 mm
L = 0.8+1.8 m
V = 1.3 mL
IPA D IPA
25.4 wt% DPE
In IPA 8.9 mg/g Toluene
In IPA
(4 mL)
Result:
DPE in Reaction pot = 71 ppm
Diluent Contamination = 279 ppm