Eagle III — Micro-EDXRF System

Page 2: Eagle III — Micro-EDXRF System. Eagle System Schematic

Eagle System SchematicEagle System Schematic

Page 3: Eagle III — Micro-EDXRF System. Eagle System Schematic

XRF AdvantagesXRF Advantages

Non-destructive: No beam damage or coating of sample

Minimal Sample Preparation:• conductivity not required• sample shape can be irregular

Low Vacuum (~ 100 mTorr) or No Vacuum (Air)

Navigation by Optical Microscope

Detection limits improve: 10x or better (vs. SEM-EDS)

X-rays are penetrating (microns to millimeters)

Page 4: Eagle III — Micro-EDXRF System. Eagle System Schematic

Advantages to EDS (Matt’s addition)Advantages to EDS (Matt’s addition)

Cheaper to add EDS to a microscope than to buy an XRF system

Orders of magnitude better image quality• CCD camera in XRF has magnification of 150 – 200X• Resolution comparable to XRF: about 10 nm• SEM image quality can be orders of magnitude better

Smaller analytical volume• One order of magnitude always• Another order of magnitude if you can live with lower voltage

Page 5: Eagle III — Micro-EDXRF System. Eagle System Schematic

Non-Destructive TestingNon-Destructive Testing

10x Eagle Video (B/W)

100x Eagle Video (color)Small Fracture in Diamond Table

? Glass-Filled ?

Small Fracture in Diamond Table? Glass-Filled ?

* Rh Tube* Aperture/Rh filter

Conclusion: Yes!Conclusion: Yes!

Page 6: Eagle III — Micro-EDXRF System. Eagle System Schematic

“As Delivered” Sample Analysis“As Delivered” Sample Analysis

Chemical Residues from suspected drug lab

X-ray Excitation minimizes sample preparation

Qualitative answer in < 2 minutes

Page 7: Eagle III — Micro-EDXRF System. Eagle System Schematic

High SensitivityHigh Sensitivity

Reduced background

Page 8: Eagle III — Micro-EDXRF System. Eagle System Schematic

Eagle System SchematicEagle System Schematic

Page 9: Eagle III — Micro-EDXRF System. Eagle System Schematic

Standard features• Rh or Mo tube (40kV, 40W)• 300µm monocapillary• Video: 10× colour; 100× colour (plus 2× digital zoom)• Sapphire™ 80 mm2 Si(Li) detector• Genesis 2000 (Windows XP)• Vision32 version 4 software (patented FP and

Comb32)

Configuration — Standard Eagle IIIConfiguration — Standard Eagle III

Page 10: Eagle III — Micro-EDXRF System. Eagle System Schematic

Configuration — Eagle III - OPTIONSConfiguration — Eagle III - OPTIONS

Options• 100µm monocapillary in lieu of the 300µm• collimators (1 & 2mm)• manually interchangeable filters (for collimator only)• 40kV, 20W Cr-anode X-ray tube• 50 kV, 50W X-ray tube (Mo, Rh or W anodes)• 30 mm2 Si(Li) detector• rotation table OR sample backlighting• LineScan, Mapping & Image processing software

Page 11: Eagle III — Micro-EDXRF System. Eagle System Schematic

Sample Illumination: White LEDsSample Illumination: White LEDs

Directionally adjustable LED arrays

Individual arrays for both Low- & High-mag image views

Individual light output adjustment to both arrays at both magnification views

Low-mag

High-mag

Page 12: Eagle III — Micro-EDXRF System. Eagle System Schematic

$20 banknote (US)

Color Low-Magnification Image (single)Color Low-Magnification Image (single)

Page 13: Eagle III — Micro-EDXRF System. Eagle System Schematic

$20 banknote (US)

Color Low-Magnification Image (montage)Color Low-Magnification Image (montage)

Page 14: Eagle III — Micro-EDXRF System. Eagle System Schematic

Hi-Magnification Image - Montage 5×5Hi-Magnification Image - Montage 5×5

Page 15: Eagle III — Micro-EDXRF System. Eagle System Schematic

Hi-Magnification Image - Montage 3×3Hi-Magnification Image - Montage 3×3

Page 16: Eagle III — Micro-EDXRF System. Eagle System Schematic

Hi-Magnification Image (Single) + Digital ZoomHi-Magnification Image (Single) + Digital Zoom

Normal (100×) Digital Zoom (2× “normal”)

Blue security-fibre in banknote

Page 17: Eagle III — Micro-EDXRF System. Eagle System Schematic

Transmission Sample BacklightingTransmission Sample Backlighting

Reflective lighting Transmission lighting

Fine “Hi-Purity” Silica particles

Page 18: Eagle III — Micro-EDXRF System. Eagle System Schematic

Transmission Sample BacklightingTransmission Sample Backlighting

Transmission lighting (Low Mag View)

Transmission lighting (High Mag View)

Page 19: Eagle III — Micro-EDXRF System. Eagle System Schematic

Si(Li) Detector propertiesSi(Li) Detector properties

Active area(mm2)

Be(coated)

window

Processing TC(µsec)

Countrate(cps)

Resolution @MnKa (eV)

30nominal

8µm

35 5000 ≤145

10 15000 ≤165

80nominal

12µm

35 5000 ≤155

10 15000 ≤185

100,000cps processing capabilityAbsolute intensities: I30 ≈ I80× 55%

Page 20: Eagle III — Micro-EDXRF System. Eagle System Schematic

30mm2

80mm2

NaK MgK AlK SiK

500 700 900 1100 1300 1500 1700 1900 (eV)

Glass sample (srm620)Spectra normalised to CaK (3690eV)

Detector’s relative low energy performancesDetector’s relative low energy performances

Page 21: Eagle III — Micro-EDXRF System. Eagle System Schematic

Si(Li) CoolingSi(Li) Cooling

Standard: Liquid Nitrogen• 30 mm2 or 80 mm2

• 5 L dewar• ≥ 3 day hold time• Detector can be allowed to warm when not in

use. Detector High Voltage bias is switched off when detector warms.

Page 22: Eagle III — Micro-EDXRF System. Eagle System Schematic

Capillary X-ray OpticsCapillary X-ray Optics

“Total” Reflection of X-rays inside glass capillary

c = f(1/E)

Page 23: Eagle III — Micro-EDXRF System. Eagle System Schematic

Incident X-Ray Spectral Distribution(Modified Excitation Spectrum)Incident X-Ray Spectral Distribution(Modified Excitation Spectrum)

Page 24: Eagle III — Micro-EDXRF System. Eagle System Schematic

Multilevel Sample AnalysisMultilevel Sample Analysis

Page 25: Eagle III — Micro-EDXRF System. Eagle System Schematic

Filter BenefitsFilter Benefits

Improve Limits of Detection

Make analysis possible

Remove Tube Characteristic Lines

Reduce Bremsstrahlung in limited region

Eliminate Bragg Diffraction Peaks in limited region

This is accomplished by …

Page 26: Eagle III — Micro-EDXRF System. Eagle System Schematic

Example: Ni FilterExample: Ni Filter

High Sensitivity Region

Useful Region

Ni Absorption Edge

Filter Band Pass

Page 27: Eagle III — Micro-EDXRF System. Eagle System Schematic

Example: Ni Filter – Improve Limits of Detection Example: Ni Filter – Improve Limits of Detection

Page 28: Eagle III — Micro-EDXRF System. Eagle System Schematic

“Vision” Software: Modes of Operation “Vision” Software: Modes of Operation

Manual point to point

Automated multiple point, lines or matrices

Analyze within an area and add spectra together

Line Scan (generates a plot)

Elemental Imaging and Spectral Mapping

Page 29: Eagle III — Micro-EDXRF System. Eagle System Schematic

“Vision” Software: Applications“Vision” Software: Applications

Qualitative Analysis (what elements and where)

Quantification:• Fundamental Parameter Modeling Quantification

without standards and with type standard(s) {Patented}• Semi-empirical quantification with type standards

Page 30: Eagle III — Micro-EDXRF System. Eagle System Schematic

“Vision” Software: Applications (cont’d)“Vision” Software: Applications (cont’d)

Coating thickness• FP modeling• FP modeling with standards correction

Spectral Match (Known alloys - ID unknown)

Line Scan

Elemental Imaging and Spectral Mapping

Image Manipulation and Overlay

Page 31: Eagle III — Micro-EDXRF System. Eagle System Schematic

Manual Control and AnalysisManual Control and Analysis

Page 32: Eagle III — Micro-EDXRF System. Eagle System Schematic

Automated Multiple Point AnalysesAutomated Multiple Point Analyses

Navigate to Feature

Save Coordinates in Stage Table

Page 33: Eagle III — Micro-EDXRF System. Eagle System Schematic

Automated Multi-Point Analysis: Automated Multi-Point Analysis:

Example: Foreign Particulates

Page 34: Eagle III — Micro-EDXRF System. Eagle System Schematic

Foreign Particulates in SilicaForeign Particulates in Silica

Transmission lighting (Low Mag View)

Transmission lighting (High Mag View)

Page 35: Eagle III — Micro-EDXRF System. Eagle System Schematic

FP “Standardless” Analysis: Particle 1FP “Standardless” Analysis: Particle 1

  Particle 1

Element: Wt%

Cr (K) 18.88

Mn (K) 0.44

Fe (K) 69.47

Ni (K) 11.21

Particle 1 = Stainless Steel

Page 36: Eagle III — Micro-EDXRF System. Eagle System Schematic

FP “Standardless” Analysis: AccuracyFP “Standardless” Analysis: Accuracy

Bulk Compositional Standard: Stainless 310 

Element:   Measured Wt%   Given Wt%   % Error

Si(K)   0.53   0.51   3.9

Cr(K)   24.97   24.88   0.4

Mn(K)   1.44   1.39   3.6

Fe(K)   53.03   52.8   0.4

Ni(K)   19.7   19.6   0.5

Mo(K)   0.32   0.23   39.1

Total   100   99.41    

Note: Measured with Poly-capillary lens

Page 37: Eagle III — Micro-EDXRF System. Eagle System Schematic

Foreign Particulates in SilicaForeign Particulates in Silica

Particle “2” Particles “3” and “4”

Page 38: Eagle III — Micro-EDXRF System. Eagle System Schematic

Foreign Particulates in SilicaForeign Particulates in Silica

Particle 2

Particle 1

“Stainless” Steels Same Alloy

Page 39: Eagle III — Micro-EDXRF System. Eagle System Schematic

Foreign Particulates in SilicaForeign Particulates in Silica

Particle 3

Particle 4

Silica particles with impurities

Page 40: Eagle III — Micro-EDXRF System. Eagle System Schematic

Multi-Point Analysis: Chemical DistributionMulti-Point Analysis: Chemical Distribution73

.24

72.7

72.1

7

71.6

3

71.0

9

70.5

6

70.0

2

45.82

45.28

44.75

44.21

43.67

43.14

42.6

Y Position (mm)X Position (mm)

86.1

85.69

85.28

84.86

84.45

84.04

83.63

83.22

82.81

46

.77

46

.36

45

.95

45

.53

45

.12

44

.71 44

.3

43

.89

43

.48

Y Position (mm)

X Position (mm)

• Automated Matrix Point CollectionAutomated Matrix Point Collection• Data ported into ExcelData ported into Excel

Page 41: Eagle III — Micro-EDXRF System. Eagle System Schematic

Spectral Mapping DefinitionSpectral Mapping Definition

Collect and save XRF spectrum at each map pixel

Database correlating each spectrum to position

(X, Y)

Page 42: Eagle III — Micro-EDXRF System. Eagle System Schematic

Spectral Mapping: Search and Use of DataSpectral Mapping: Search and Use of Data

Spectral Display:• Point by point• Summation of selected region or total map• Display of Linear Distributions

Return to Sample using Map for collection of spectrum with improved statistical significance

Quantitative mapping

Page 43: Eagle III — Micro-EDXRF System. Eagle System Schematic

Spectral Mapping: Spectral Mapping:

Mapping Examples

Page 44: Eagle III — Micro-EDXRF System. Eagle System Schematic

Elemental Spatial Distribution Maps: PaperElemental Spatial Distribution Maps: Paper

Mg Map Al Map Fe Map

• Generation of BMP Elemental Maps

Fe X-rays penetrate paper

Page 45: Eagle III — Micro-EDXRF System. Eagle System Schematic

Spatial Distribution Maps: Facial TissueSpatial Distribution Maps: Facial Tissue

• Tissue masked with carbon tape for Si-free zone• Mapping region 15.6 mm x 11.3 mm

Page 46: Eagle III — Micro-EDXRF System. Eagle System Schematic

Spatial Distribution Maps: Facial TissueSpatial Distribution Maps: Facial Tissue

• Recall spectra from mapped pixels

• Hot Si spots hide low-level Silicone coverage

Page 47: Eagle III — Micro-EDXRF System. Eagle System Schematic

Spatial Distribution Maps: Facial TissueSpatial Distribution Maps: Facial Tissue

• 3 individual color logarithmic scales (NIST)

• Low level Silicone distribution exposed in Green

Page 48: Eagle III — Micro-EDXRF System. Eagle System Schematic

Quantitative Mapping: Geological SampleQuantitative Mapping: Geological Sample

• Sedimentary rock

• Epoxy-embedded “puck” used to make thin sections

• Map area defined by 5x5 Hi-Mag montage

Map Image: Total XRF counts in each map pixel

Page 49: Eagle III — Micro-EDXRF System. Eagle System Schematic

Quantitative Mapping: Geological SampleQuantitative Mapping: Geological Sample

FeK Intensity Fe2O3 Wt%

Page 50: Eagle III — Micro-EDXRF System. Eagle System Schematic

Quantitative Mapping: Geological SampleQuantitative Mapping: Geological Sample

SiK Intensity SiO2 Wt%

Page 51: Eagle III — Micro-EDXRF System. Eagle System Schematic

Multi-Field Mapping: Geological SampleMulti-Field Mapping: Geological Sample

• 7 adjacent High Mag Camera FOV

• Map more layers in shorter time

• Maps are stitched together in SW utility while adjusting map intensities

Page 52: Eagle III — Micro-EDXRF System. Eagle System Schematic

Spectral Mapping - Bone Fossilization Spectral Mapping - Bone Fossilization

Fe

Na

K

Si

P

Page 53: Eagle III — Micro-EDXRF System. Eagle System Schematic

Map Image Overlays: Bone Fossilization Map Image Overlays: Bone Fossilization

Fe – Red

K – Blue

Si – Yellow

P – Gray

Na - Green

Page 54: Eagle III — Micro-EDXRF System. Eagle System Schematic

Metal Analysis: Coins (Non-Destructive)Metal Analysis: Coins (Non-Destructive)

* Rare Coin (2 Reichsmark - 1927?)

* Pixels: 64 x 50 Map* Dwell time: 0.3 s/pixel* Total time ~ 20 minutes

Conclusion:

Counterfeit Coin

Conclusion:

Counterfeit Coin

Page 55: Eagle III — Micro-EDXRF System. Eagle System Schematic

Eagle ApplicationsEagle Applications

Glass, Ceramics (inhomogeneity, inclusions, particles)

Metal alloys (inhomogeneity, particles, wire filament)

Inorganic contaminants, residues, deposits (ex. Corrosion)

Inorganic additives polymers, paints, inks

Inclusions in plastics, “light element” materials

Coating thickness and distribution of coating thickness