E Usadi, NPL Low Temperature Blackbody for CAVIAR: Characterisation & Calibration Eric Usadi Thanks to Chris Price Theo Theocharous Title E Usadi 02Jun08

E Usadi, NPL Low Temperature Blackbody for CAVIAR: Characterisation & Calibration Eric Usadi Thanks to Chris Price Theo Theocharous Title E Usadi 02Jun08

  • View
    216

  • Download
    0

Embed Size (px)

Citation preview

E Usadi, NPL

Low TemperatureBlackbody for CAVIAR:Characterisation & Calibration

Eric Usadi

Thanks to

Chris PriceTheo Theocharous

Title

E Usadi 02Jun08

E Usadi, NPL

• blackbody description

• traceability chain

• calibration & validation

plan

E Usadi 02Jun08

E Usadi, NPL

• blackbody description

• traceability chain

• calibration & validation

Blackbody short description

E Usadi 02Jun08

E Usadi, NPL

UniquenessUNIQUE CALIBRATION TOOLS

CAVIAR blackbody• in-air (i.e. non-vacuum)• temperature range• time response• aperture size• emissivity• calibrated & characterised

IR filter radiometer• world leading infrared FR• enables traceability

E Usadi 02Jun08

E Usadi, NPLCAVIAR blackbody

Blackbody schematic (not to scale)

ethanol in

LN2

insulation

PRT s

fluid heater/chiller

& circulator

(sealed system)

-60o C to +30o C

heat

exchanger

E Usadi 02Jun08

E Usadi, NPL

Copper parts

E Usadi 02Jun08

E Usadi, NPL

Copper parts

E Usadi 02Jun08

E Usadi, NPL

• blackbody description

• traceability chain

• calibration & validation

Traceability chain

E Usadi 02Jun08

E Usadi, NPL

Traceability schematic

Cryogenic radiometer

High emissivity BB (@ Ga fixed point)

CBB

TAFTS Aries

IR filter radiometer

IR filter

Transmittance scale

E Usadi 02Jun08

E Usadi, NPL

Traceability schematic 2

Cryogenic radiometer

High emissivity BB (@ Ga fixed point)

CBB

TAFTS Aries

IR filter radiometer

IR filter

Transmittance scale

E Usadi 02Jun08

E Usadi, NPL

Calibration schematicHigh emissivity blackbody@ Gallium Melting Point (29.76 deg C)

IR filter radiometer(4.7 m, et al)

Low temperature blackbody

E Usadi 02Jun08

E Usadi, NPLR

adia

nce

@ 4

.7

m

Temperature (K)

Planck function

Res

pons

ivity

Unc

erta

inty

(%

)

(microns)

Planck function Filter RadiometerResponsivity

E Usadi 02Jun08

Wavelength (m)

Ra

dia

nce

10 20 30 40

2 . ´ 10 6

4 . ´ 10 6

6 . ´ 10 6

8 . ´ 10 6

1 . ´ 10 7

303 K

258 K

213 K

Wavelength (m)

Ra

dia

nce

10 20 30 40

2 . ´ 10 6

4 . ´ 10 6

6 . ´ 10 6

8 . ´ 10 6

1 . ´ 10 7

303 K

258 K

213 K

10 20 30 40

2 . ´ 10 6

4 . ´ 10 6

6 . ´ 10 6

8 . ´ 10 6

1 . ´ 10 7

303 K

258 K

213 K2 .5 3 .0 3 .5 4 .0 4 .5 5 .0 5 .5

1 00

1 00 0

1 04

1 05

1 06

2 .0 2 .5 3 .0 3 .5 4 .0 4 .5 5 .0 5 .5

1 .0

0 .5

2 .0

0 .2

5 .0

1 0 .0

240 260 280 300

500 000

1 . ´ 10 6

1 .5 ´ 10 6

2 . ´ 10 6

E Usadi, NPL

Spectral Radiance slope

E Usadi 02Jun08

0

1

2

3

4

5

6

7

-70 -60 -50 -40 -30 -20 -10 0 10 20 30 40

T (deg C)

% s

lope

of S

pect

ral R

adia

nce 4.7 microns

10 microns

20 microns

30 microns

40 microns

50 microns

% s

lope

of S

pect

ral R

adi

ance

0

1

2

3

4

5

6

7

-70 -60 -50 -40 -30 -20 -10 0 10 20 30 40

T (deg C)

% s

lope

of S

pect

ral R

adia

nce 4.7 microns

10 microns

20 microns

30 microns

40 microns

50 microns

% s

lope

of S

pect

ral R

adi

ance

E Usadi, NPL

• blackbody description

• traceability chain

• calibration & validation

Calibration & validation

E Usadi 02Jun08

E Usadi, NPL

E Usadi 02Jun08

measure BB temperature

measure BB radiance @ 4.7 microns

calculate BB radiance

uncertainty of calculation

uncertainty of measurement

difference between measurement & calculation

difference < uncertainty ?

measure BB temperature

measure BB radiance @ 4.7 microns

calculate BB radiance

uncertainty of calculation

uncertainty of measurement

difference between measurement & calculation

difference < uncertainty ?

E Usadi, NPL

-40

-30

-20

-10

0

10

20

30

0 2 4 6 8 10

Time (hours)

Tem

p (d

eg C

) AMBER

Ch 2

Ch 5

Ch 1

22May08 data

Blackbody thermometer readings

AMBER radiometerreading of referenceblackbody (arb units)

AMBER radiometerreading of low-tempblackbody

E Usadi 02Jun08

E Usadi, NPL

E Usadi 02Jun08

-40

-30

-20

-10

0

10

20

30

0 2 4 6 8 10

Time (hours)

Tem

p (d

eg C

) AMBER

Ch 2

Ch 5

Ch 1

22May08 data

-2 mK0.007%

-16 mK0.06%

-25 mK0.1%

-13 mK0.05%

-53 mK0.24%

-78 mK0.38%

-112 mK0.59%

-62 mK0.29%

-31 mK0.13%

-11 mK0.04%

3 mK

0.01%

Difference between end platethermometer readings

Resultant radiance difference

AMBER radiometerreading of referenceblackbody (arb units)

AMBER radiometerreading of low-tempblackbody

-40

-30

-20

-10

0

10

20

30

0 2 4 6 8 10

Time (hours)

Tem

p (d

eg C

) AMBER

Ch 2

Ch 5

Ch 1

22May08 data

-2 mK0.007%

-16 mK0.06%

-25 mK0.1%

-13 mK0.05%

-53 mK0.24%

-78 mK0.38%

-112 mK0.59%

-62 mK0.29%

-31 mK0.13%

-11 mK0.04%

3 mK

0.01%

Difference between end platethermometer readings

Resultant radiance difference

AMBER radiometerreading of referenceblackbody (arb units)

AMBER radiometerreading of low-tempblackbody

E Usadi, NPL

Equation 6

E Usadi 02Jun08

E Usadi, NPL

Equation 6 detailPlanck function

Filter radiometer responsivity

measured at discrete ’s

Signal from

reference blackbody

at Gallium Point

Emissivities known by

design & modelling

Signal from

CAVIAR blackbody

1) MEASURED

2) CALCULATED

E Usadi 02Jun08

E Usadi, NPL

E Usadi 02Jun08

0.25

0.75

1.25

1.75

2.25

2.75

3.25

3.75

2 3 4 5 6 7 8

Time (hours)

Radia

nce

(arb

unit

s)

AMBER

Ch 2

Ch 5

Ch 1

0.05%

-0.10%

0.12%

0.03%

-0.33%

0.08%

0.25%

0.30%

0.31%

0.09%

-0.69%

Difference betweenmeasured & calculatedradiance

22May08 data

-0.13%

Radiometer uncertainty : 0.22% Calculated spectral radiance uncertainty: 0.2% @ T = 303K

0.9% @ T = 233K

0.25

0.75

1.25

1.75

2.25

2.75

3.25

3.75

2 3 4 5 6 7 8

Time (hours)

Radia

nce

(arb

unit

s)

AMBER

Ch 2

Ch 5

Ch 1

0.05%

-0.10%

0.12%

0.03%

-0.33%

0.08%

0.25%

0.30%

0.31%

0.09%

-0.69%

Difference betweenmeasured & calculatedradiance

22May08 data

-0.13%

Radiometer uncertainty : 0.22% Calculated spectral radiance uncertainty: 0.2% @ T = 303K

0.9% @ T = 233K

E Usadi, NPL

Uncertainty budget

E Usadi 02Jun08

CAVIAR BB Uncertainty Budget

PRT calibrationPRT measurement by Agilent DMMGradient across paint due to convective heat tansferNextel reflectanceeffective emissivityuniformitynumerical integration of ratioIntegration: filter radiometer calibration uncertaintiesIntegration: temperature uncertaintiesReflection of ambientTemp gradient along cylinder

E Usadi, NPL

Combined uncertainty

E Usadi 02Jun08

Tbb=303K, Tbgnd=293K

0

2000000

4000000

6000000

8000000

10000000

12000000

2 3 4 5 6 7 8 9 10 12 14 16 18 20 25 30 35 40 45 50

wavelength (microns)

0

0.05

0.1

0.15

0.2

0.25

0.3

Spectral Radiance (arb units)

Combined standard uncertainty (%)

T b b = 2 3 3 K , T b g n d = 2 7 3 K

0

5 0 0 0 0 0

1 0 0 0 0 0 0

1 5 0 0 0 0 0

2 0 0 0 0 0 0

2 5 0 0 0 0 0

3 0 0 0 0 0 0

2 3 4 5 6 7 8 9 1 0 1 2 1 4 1 6 1 8 2 0 2 5 3 0 3 5 4 0 4 5 5 0

w a v e le n g th (m ic ro n s)

0

0 .2

0 .4

0 .6

0 .8

1

1 .2

1 .4

1 .6

1 .8

2

S p e c t ra l R a d ia n c e (a rb u n it s )

C o m b in e d s ta n d a rd u n c e rt a in t y (% )

E Usadi, NPL

Conclusion

E Usadi 02Jun08

Conclusions

• blackbody tested -60 to +30 deg C

• performance CALIBRATED & VALIDATED

• full uncertainty analysis

• unique characterisation of low-temp blackbody

• ready for ARIES & TAFTS……