DUAL SOURCE CARDIAC CT ANGIOGRAPHY

Dr Ravi Mathai, MD.

Consultant Radiologist,

Dar Al Shifa Hospital

Page 2: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

2

Role of CTA

risk stratification ; high CAC score + intermediate FRS = reclassified as high risk

acute chest pain - presence of stenoses + determine the necessity of further treatment.

ruling out stenosis in patients with intermediate pretest likelihood of disease

Detect stenotic lesions in symptomatic patients

Follow-Up of Percutaneous Coronary Intervention -in-stent restenosis

Follow-Up After Bypass Surgery - patency of the bypass graft - in course, anastomotic site and native vessels

Page 3: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

3

Role of CTA exact analysis of anomalous coronary arteries.

assess morphology and function eg valvular motion, wall motion, EF, CO.

CT angiography (CTA) has high negative predictive value.

Page 4: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

4

Caveats Use of CT angiography in asymptomatic persons as a screening test for

atherosclerosis (noncalcific plaque) is not yet recommended

Not recommended for acute coronary syndrome

Page 5: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

5

Limitations of CTA overestimate disease severity

limited spatial resolution = +-accurate grading of the severity of stenosis

Pronounced coronary calcifications

motion artefacts

trigger artefacts - cardiac cycle phase

high image noise can prevent reliable evaluation

radiation dose

Page 6: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

6

DSCT advantage High temporal resolution - 0.28 seconds rotation time = temporal resolution

of upto 0.75 seconds - at a pitch of 3.2 (FLASH)

Mayo Clinic study 2009 showed no differences in quantitative measures of image quality between single-source scans at pitch = 1 and dual-source scans at pitch = 3.2

Regular and low heart rates prerequisite for CCTA by 64 slice MDCT

Page 7: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

7

Page 8: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

8

Flash advantage high pitch (3.2)

dual tube quarter rotation data acquisition

0.28s scan time

very low dose

limitations - obese, high HR >75

Page 9: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

9

Page 10: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

10

Goals of CTA primary goal diagnostic IQ

second goal - low dose

protocol aims at the above

Page 11: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

11

Components of a cardiac scan: Patient selection

Breathold

HR

Medications

Ecg and gating

Contrast timing and dose

Scan mode selection - prospective, retrospective, pulse off, flash.

kV techniques

ROI

Reconstruction kernels

Low dose techniques

Image processing

Page 12: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

12

Patient selection CTA rule out in acute chest pain

Ruling out stenosis in patients with intermediate pretest likelihood of disease

For the assessment of obstructive disease in symptomatic patients

For detecting re-stenosis after stent placement

Follow-Up After Bypass Surgery -bypass graft - native vessels-

Exact analysis of anomalous coronary arteries.

Page 13: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

13

Contraindications to CTA Contrast hypersensitivity (absolute)

Renal failure (absolute)

Poor breathhold

AF

Page 14: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

14

Breathold Breathe in -- breathe out – breathe in – hold your breath (13 seconds)

Patient training

Page 15: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

15

Heart rate for S64 - <60

for DSCT <95

look for ectopics, arrythmias

Page 16: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

16

Medications for S64 -Oral ß-Blockers 1 h before scan if heart rate > 60/min e.g., 100 mg

Atenolol

i.v. metoprolol (up to 6 x 5 mg) if heart rate in CT scanner is still > 60/min

for DSCT - no BB required for HR upto 85. 85-105 no BB if dose increased (retropective scan). >105, BB

Page 17: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

17

ECG ECG -must be noise free

Gating - prospective and restrospective

Scan protocols

Page 18: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

18

Noisy

Page 19: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

19

Standard Scan Protocol - DSCT

Collimation 0.6 mm

Rotation: 330 ms

kV: 120 kV

mAs: 400 mAs

Pitch: Enter expected heart rate manually

ECG Pulsing: 70-70% for heart rates < 65/min

40-70% for heart rates > 65/min

Delay: Contrast time + 2 seconds

Page 20: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

20

Contrast Injection

5 ml/s for the duration of the scan

At least 50 ml + Follow by 50 ml saline (or 20% contrast) at 5 ml/s

Page 21: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

21

Contrast injection and Bolus Type of CM: Concentration min. 350mg/ ml, better 370mg/ ml

Flow rate: average size patients (~70kg / 150lb) 5cc/ sec; larger patients 7cc/ sec

i.v. line: min 18g, better 16g

Test Bolus: 10cc contrast/ 50cc saline (Care bolus ROI AA, threshold 100HU)

ROI : Measure in ascending aorta

Page 22: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

22

Page 23: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

23

Coronary CTA injector options Normal injector:

Volume of contrast = scan time x flow rate + 10cc + 50cc saline; min 45cc, max 100cc

Dual flow option:

Volume of contrast = scan time x flow rate + 10cc contrast

1st phase: total volume of contrast

2nd phase: 50cc of volume (20% contrast + 80%saline) = (10cc contrast + 40cc of saline)

Page 24: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

24

Page 25: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

25

Scan modes Scan mode selection - prospective, retrospective, pulse off, flash.

Page 26: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

26

ECG controlled dose modulation - retrospective

Page 27: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

27

ROI AA for test and care bolus

extend for bypass grafts

Page 28: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

28

Page 29: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

29

Reconstruction Reconstruction Slice thickness: 0.6 mm

Kernel: B26f (B46 f for Stents, Ca++)

Phase: Initially: Best Diast / Best Syst

DSCT: Usually 75% R-R best ; Preset: BD, BS, 70%, 75%, 40%

S64: Usually 70% R-R best; Preset: 65%, 70%, 35%, 40%

Page 30: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

30

Multiphase reconstructions

Page 31: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

31

low dose techniques 1.“CARE Dose4D” – Real-time Anatomic Exposure Control

2. “Adaptive ECG-Pulsing” – ECG-Controlled Dose Modulation for Cardiac Spiral CT

3. “Adaptive Cardio Sequence” – ECG-triggered Sequential CT

4. “Adaptive Dose Shield” – Asymmetric Collimator Control

5. “Flash Spiral” – ECG-Triggered Dual Source Spiral CT Using High Pitch Values

6. “X-CARE” – Organ Based Dose Modulation

7. “IRIS” – Iterative Reconstruction in Image Space

Page 32: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

32

Page 33: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

33

Adaptive ECG-Pulsing -ECG-Controlled Dose Modulation for Cardiac Spiral CT

Page 34: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

34

Page 35: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

35

Page 36: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

36

Page 37: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

37

Page 38: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

38

Adaptive Cardio Sequence -ECG-Triggered Sequential CT

Page 39: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

39

ADAPTIVE

Page 40: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

40

High pitch -FLASH With a single source CT, the spiral pitch is limited to values below 1.5 to

ensure gapless volume coverage along the z-axis. If the pitch is increased, sampling gaps occur

With DSCT systems, data acquired with the second measurement system a quarter rotation later can be used to fill these gaps.In this way, the pitch can be increased up to 3.4

Page 41: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

41

FLASH

Page 42: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

42

FLASH ECG-triggered DSCT scan data acquisition and image reconstruction at very

high pitch.

images reconstructed in this mode with an acquisition time of 250 ms, a temporal resolution of 75 ms, 100 kV and 0.8 mSv.

Page 43: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

43

Modification of protocols

Beta Blockers - preferable above HR 85.

Saline vs. 20% contrast flush. - full functional assessment

6 (7) ml flow for heavy patients

XXL for heavy patients

100 kV for slim patients

Page 44: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

44

Kv modification as per weight

Page 45: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

45

sub mSv CTA

100kv

320mAs

120mm scan range

flash mode

Page 46: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

46

FLASH

Page 47: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

47

pediatric protol

Flash mode

80kv

104mAs

120mm scan range

Dose less than 1mSv

Page 48: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

48

FLASH

Page 49: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

49

Image processing 2d images MIP best for diagnosis

3d complementary

stenosis grading - software Syngovia - automatic calculation.

Page 50: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

50

Graft

Page 51: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

51

Stenosis

Page 52: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

52

HR 85 Prospective

Page 53: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

53

HR 83 Retrospective

Page 54: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

54

Stent

Page 55: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

55

Pediatric

Page 56: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

56

Page 57: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

57

Adaptive with arrythmia

Page 58: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

58

FLASH

Page 59: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

59

FLASH

Page 60: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

60

FLASH

Page 61: DUAL SOURCE CARDIAC CT ANGIOGRAPHY Dr Ravi Mathai, MD. Consultant Radiologist, Dar Al Shifa Hospital 1

61

TEAM CTA