John E Shulze, CTO
Biosensors International Inc
Drug Application in
Stent Technology:How to Diversify?
My Conflicts of interest are:
Chief Technology Officer
and a shareholder of
Biosensors International Inc.1
(1Biosensors’ development pipeline products to be discussed
in this Presentation)
BIOLIMUS A9™ DRUG
RAPAMYCIN DERIVATIVE
Developed specifically for stent application
by Biosensors
Potent immunosuppressive and anti-
inflammatory properties
LIPOPHILICITY COMPARISON
Highest lipophilic and hydrophobic
properties of commercially available limus
drugs
Mainly localized effects, minimal drug
release into bloodstream
BIOLIMUS A9™ DRUGBiosensors’ proprietary rapamycin derivative
Highest lypophilicity of the common limus drugs1
BIODEGRADABLE PLASimultaneous PLA biodegradation and BA9™ elution
No PLA /BA9™ coating on the stent after 6 to 9months2
The Abluminal Biodegradable Polymer DES
ABLUMINAL BIODEGRADABLE COATINGEarly BMS-like endothelial coverage1
More targeted tissue release
Less systemic exposure
1 Data on file at Biosensors Intl2 In vivo testing in porcine model demonstrates abluminal coating is absorbed after 6 to 9 months - Data on file at Biosensors Intl
10810-0
00-E
N –
Rev.0
1
Trial Design (LEADERS)
Assessor-blind
1:1 Randomisation
1o endpoint: CV death, MI, clinically-indicated TVR (9 months)
2o endpoints: Death, CV death, MI, TLR, TVR
Stent thrombosis according to ARC
Angiographic study: In-stent % diameter stenosis
Late loss, binary restenosis
DAPT recommended for 12 month
N=1700
Patients
SES
Cypher Select N=850
BES
BioMatrix Flex N=850
Clinical F/U
N=640
Angio F/U
N=210
Clinical F/U
N=640
Angio F/U
N=210
1:3 Randomisation
Stable and ACS Patients Undergoing PCI
3-year follow-up
(N=1,621; 95.0%)
0
5
10
15
0 1 2 3 4 5 6 7 8 9
Months of Follow-up
Cypher® Select™ 10.5%
BioMatrix Flex™ 9.2%
pnon-inferiority = 0.003
Primary EndpointMACE (Cardiac Death, MI and TVR) @ 9 Months
BioMatrix Flex™ reached its primary endpoint
Windecker S. et al., The Lancet 2008; 372 No. 9644: 1163-1173
MA
CE
%
MACE
%
Number at risk
BES 857 851 761 743 729 712 668
SES 850 846 749 732 713 686 639
Months
13.0%
15.4%
2-year HR0.84 [0.65 to 1.08]
P = 0.18*
12.1%
10.7%
1-year HR0.88 [0.66 to 1.17]
P = 0.37*
0
5
10
15
20
0 6 12 18 24
BES
SES
3630
15.7%
19.0%
3-year HR0.82 [0.65 to 1.03]
P = 0.09*
*P values for superiority
MACE = Cardiac Death, MI, or Clinically-Indicated TVR
Serruys, P. W., oral presentation ,TCT 2010
Definite ST through 3 years
1.0
2.0
3.0
% 2.2%
2.9%
BES
SES
0 6 12 18 24 3630Months
2.2%
2.5%
2-year HR
0.90 [0.48 to 1.67
P = 0.73*
2.0%
2.0%
1-year HR
0.99 [0.51 to 1.94]
P = 0.98*
3-year HR
0.78 [0.43 to 1.43]
P = 0.43*
0
4.0
Number at risk
BES 857 846 808 797 787 774 732
SES 850 841 801 792 779 758 715
Δ 0.0%Δ 0.3%
Δ 0.7%
*P values for superiority
Serruys P.W., oral presentation, TCT 2010
PROVISIONAL STENTING USE
Micro-CT Biomatrix results
MV STENTING AFTER KISSING
MACE in Complex Patients
%
Months
8.4%
19.7%
15.9%
6.8%
0
5
10
15
20
0 6 12 18 24 3630
10.0%
21.3%
3-year HR0.43 [0.23 to 0.84]
P = 0.01*
25 STEMI
%
Months
15.3%
21.8%
18.6%
12.7%
0
5
10
15
20
0 6 12 18 24 3630
17.0%
23.8%
3-year HR
0.69 [0.46 to 1.04]
P = 0.08*
25High Syntax Score (>16)
Bifurcations
%15.6%
18.9%16.4%
12.9%
0
5
10
15
20
0 6 12 18 24 3630
18.5%
23.3%
3-year HR
0.79 [0.53 to 1.16]
P = 0.22*
25
16.0%
24.1%
3-year HR
0.65 [0.41 to
1.03]
P = 0.06*
%
15.0%
23.5%
17.1%
13.0%
0
5
10
15
20
0 6 12 18 24 3630
25
Multi-vessel
*P values for superiority
Windecker, S., oral presentation ,TCT 2010
BES
SES
BioFreedom™Selectively micro-structured surface holds drug in
abluminal surface structures
Proprietary Highly Lipophilic Limus drug
Hypothesis: Polymer-free drug
release via porous-eluting stents
may reduce late events caused by
polymer stent coatings.
Potential advantage
• Avoid long term late adverse
effects that might be attributable
to the polymer
• Improved surface integrity since
there is no polymer to be sheared
or peeled away from the stent
struts
• Possible shorter need of dual
antiplatelet therapy
Grube E., oral presentation, TCT 2010
BioFreedom FIM Design
BioFreedom FIM182 patients
12 Month Angio FU107 patients
BioFreedom
standard
dose
(BFD SD)
N=35
BioFreedom
low
dose
(BFD LD)
N=36
TAXUS®
Liberté ®
N=36
Second Cohort
Enrollment Period
Jan 2009 – Jun 2009
BioFreedom
standard
dose
(BFD SD)
N=25
BioFreedom
low
dose
(BFD LD)
N=26
TAXUS®
Liberté ®
N=24
4 Month Angio FU75 patients
First Cohort
Enrollment Period
Sept 2008 – Jan 2009
Angio FU 92%
12 Month Clinical FU 99%
Angio FU 92%
12 Month Angiographic FUIn-Stent Late Lumen Loss: 2nd Cohort
0.17[0.09, 0.39]
0.22[0.17, 0.66]
0.35[0.22, 0.57]
0.0
0.1
0.2
0.3
0.4
BFD SD BFD LD Taxus
P = 0.001
P = 0.21
Non – inferiority
P values
(mm
)
N = 31 N = 31N = 35
Primary Endpoint
Grube E., oral presentation, TCT 2010
All values are presented as median [IQR]
12 Month MACEAll Patients (1st + 2nd Cohorts)
BFD SD
N = 60
BFD LD
N = 62
Taxus
N = 60
MACE*(All Death, MI, Emergent Bypass or TLR)
3 (6.1%) 7 (11.6%) 3 (5.5%)
All Death 1 (1.8%) 0 (0.0%) 0 (0.0%)
MI 1 (1.8%) 1 (1.6%) 0 (0.0%)
Q Wave MI 0 (0.0%) 0 (0.0%) 0 (0.0%)
Non-Q Wave MI 1 (1.8%) 1 (1.6%)** 0 (0.0%)
Emergent Bypass 0 (0.0%) 0 (0.0%) 0 (0.0%)
TLR 1 (1.8%) 6 (10.0%) 3 (5.5%)
All p values are non significant.
Tests were performed for BFD SD vs. Taxus and BFD LD vs. Taxus.
*Time to first event
**In-hospital MI
No ST event out to 12 months: ARC defined Definite, Probable or Possible!
AXXESS
The AXXESS stent
The AXXESS stent is a dedicated bifurcation stent
• Nitinol self explandable stent
• Using PLA/BA9™ abluminal coating technology
• 4 references: 3.0 and 3.5 mm in diameter; 11
and 14 mm in lengths
• 2 models: bifurcation and LM (difference in
sizes of the conical shape)
The AXXESS stent is associated to a clinical program
• AXXESS Plus and DIVERGE for de novo
bifurcation
• AXXENT for LM lesion
THE AXXESS stent received CE mark in August 2010
• Indication: bifurcation
Restenosis in DIVERGE
Any in-Bifurcation restenosis:
6.4% (9/140 at 9 months)
Lowest restenosis rates ever
reported in a bifurcation
study of any kind
Parent
Vessel RS
3 pts
Side
Branch RS
4pts
2 pts
Both
What is next?
• The smallest profile drug-eluting stent system in the world
• 60% smaller than the nearest available coronary stent system.
• limus drug with a biodegradable polymer matrix
• Self-expanding ultra-thin Nitinol stent
• 0.014-inch guidewire delivery system
• unique “stent-on-the-wire” construction
• BMS System CE Marking 3/2010
Sparrow™ Microstent:
designed for small vessels
Non-Sterile,
reusable, hand-held
Power Supply
Sterile,
Disposable
Adapter
Delivery
System
Stent
Lead Connection/ Wire Extension
CARE II RCT4:4:3 (80:80:60)
Total N=220 + roll in
3 Arm Sub-Set
1:1:1 randomization
Sparrow DES: Sparrow BM:
Micro-Driver/Driver BM
14 Centers
IVUS sub-study (n = 120 @ 7
centers)
2 Arm Sub-Set
1:1 randomization
Sparrow DES: Sparrow BM
4 Centers
Sparrow - BM(N=20)
Sparrow – DES(~6 ug/mm Sirolimus)
(N=20)
Sparrow DES= 80 Sparrow BM = 80 Micro-Driver/Driver = 60
Sparrow - BM(N=60)
Micro-Driver/
Driver(N=60)
Sparrow – DES(~6ug/mm Sirolimus)
(N=60)
CARE II DESIGN
Primary Endpoint:In Stent Late lumen loss @ 8 mo
CARE II 8 month QCA
SES Sparrow Stent demonstrated superior
in-stent LLL -
CARE II Primary Endpoint met
% CMI-SES CMI-BMS Driver
Mean
(mm)0.29±0.45 0.86±0.54 0.94±0.39
Range
(mm)-0.59-1.28 -0.20-1.89 0.44-2.11
p 0.0001 < 0.0001
n 30 31 25-10
0
10
20
30
40
CMI-SES
(n=30)
(%)
CMI-BMS
(n=31)
Driver
(n=25)
P=0.0012
Binary Restenosis
(in Lesion)
6.7 4445.2
P=.00006
Abiziaid A.C. et al., TCT 2010
BA9 shows sustained Safety and Efficacy –Independent from stent platform or method of drug delivery!
Take-home message
BioMatrixTM
BioMatrixTM
FlexBioFreedom
TMAXXESS
TMPlus Sparrow-DES Nobori
(Terumo)
Custom
NXTM
(XTENT)
Stainless SteelBalloonexpandable
Abluminal
biodegradable
polymer
Stainless SteelBalloonexpandable
Polymer Free
Nitinol
Self-expandable
Abluminal
biodegradable
polymer
Designed for
Bifurcations
Nitinol
Self-
expandable
Sirolimus*/PLA
Designed for
small vessels
SS
BA9/PLA
Cobalt-Chrome
BA9/PLA
*BA9 application under development
BioMatrix Flex™ Drug Eluting Coronary Stent System is CE approvedNot available for sale in the United States and certain other countries